1. Genie 225
Membrane Separator
The Model 225 protects liquid hydrocarbon
systems from water, caustic, sulphuric acid or
other immiscible liquids
The Supreme 200 Series remove 100% of suspended, immiscible
liquids in liquid hydrocarbon samples, which allow only hydrocarbon
KEYBENEFITS
• Protection against liquids
• On-line and portable analysers
• GC’s, Mass Specs, O2, H2S, Moisture,
and others
• Superior analyser protection
• Economical
• Safe and easy to install and maintain
• Allows for continuous liquid sampling
in any process industry including
natural gas, petrochemical, and oil
liquid sample to flow to an analyser. This action protects analysers
against damage to analysers and sample system components.
Remove Liquid Flow from Samples
Ideal for high-pressure applications, the Genie 225 also removes
absorbed gases, gas bubbles, or volatile organic carbon (VOC)
compounds from water sample at the same pressure rating.
Easily Accessible Membrane Maintenance
Genie Supreme Series Membrane Separators are safe and easy
to install and maintain, especially in heated, densely populated
cabinets. The Genie 225 has the same design as both the original
Genie Model 205 and Model 205HP, except it includes a
standard screw off cover to allow easier membrane maintenance.
refining.
“
“
The original brand known for
sample conditioning and
analyser protection
2. TECHNICALSPECIFICATION
• Upstream of membrane: 0.47 cubic inches
• Downstream of membrane: 0.26 cubic inches
PRESSURE RATINGS
• Maximum pressure rating: 2,000 psig
• Maximum recommended supply pressure:
FLOW RATES
• 150 cc/min in Diesel*
• 200 cc/min in Kerosene*
• 450 cc/min in Gasoline *
2.000 psig
MAXIMUM TEMPERATURE
*maximum flow results in approx. 10 psi membrane differential pressure
• 302 ºF (150 ºC)
WETTED MATERIALS
• Machined parts: 316 stainless steel / NACE compliant
• All other metal parts: stainless steel / NACE compliant
• Sealing material: Fluoroelastomer standard
• Membrane: Inert
DIMENSIONS
• Port sizes: Inlet, Outlet, & Bypass: 1/4” female NPT
INTERNAL VOLUME
• Total: 0.73 cubic inches
PARTBUILDER
MODEL NUMBERING & ADDITIONAL PART NUMBERS
•
•
•
•
Sealing material
0 = fluoroelastomer
Membrane type
3 = Liquid/Liquid membrane (with bonded o-ring)
(other materials available upon request)
Mounting bracket accessory
Part # 225-509-SS (sold separately)
O-ring replacement
8 = Liquid/Liquid Backed membrane
Part # 225-500 (sold separately)
How to build the model number:
How to build the replacement membrane kit:
11 Ark Royal Way, Lairdside Technology Park, Tranmere, Wirral, CH41 9HT
T: +44(0)151 666 8300 F: +44(0)151 666 8329 E: sales@a1-cbiss.com
W: www.a1-cbiss.com Buy Online: shop.a1-cbiss.com
Cert No. 996