ULTRAVIOLET-VISIBLE
SPECTROSCOPY
TOPIC:
Instrumentation, choice of
solvents and solvent
effect.
Submitted by
BINUJA.S.S
First year M.Pharm
Department of Pharmaceutical chemistry
Ezhuthachan college of pharmaceutical
sciences
 INTRODUCTION
 INSTRUMENTATION
 INSTRUMENTS
 Uv- visible Photometers
 Uv – visible Spectrophotometers
 CHOICE OF SOLVENTS AND SOLVENT
EFFECT
 REFERENCE
CONTENTS
3
INTRODUCTION
 Oldest physical method for quantitative
analysis and structural elucidation.
 qualitative and quantitative analysis of
substances.
 The spectrophotometer technique is to
measures light intensity as a function of
wavelength.
4
 It does this by:
 diffracting the light beam into a spectrum of wavelengths
 direct it to an object
 receiving the light reflected or returned from the object
 detecting the intensities with a charge-coupled device
 displaying the results as a graph on the detector.
5
6
Instrumentation
Instruments for measuring the absorption of UV-Visible are
made up of following components.
1. Sources
2. Wavelength selectors
3. Sample containers
4. Detectors
5. Readout device
7
Sources
REQUIREMENTS OF AN IDEAL SOURCE
 Stable and should not allow fluctuations.
 Emit light of continuous radiation.
 Provide sufficient intensity.
8
UV Radiation sources Visible sources
 Deuterium lamp
 Hydrogen lamp
 Tungsten halogen
lamp
 Xenon arc lamp
 Mercury arc lamp
 Tungsten halogen
lamp
 Mercury vapour lamp
 Carbonone lamp
9
Tungsten Halogen lamp
 filament of Tungsten fixed in evacuated
condition and then filled with inert gas.
 The filament can be heated up to 3000 k,
beyond this tungsten starts sublimating.
 It is used when polychromatic light is
required. To prevent this along with inert gas
some amount of halogen is introduced
10
 Sublimated form of tungsten reacts with Iodine to form
Tungsten –Iodine complex.
 Which migrates back to the hot filament where it
decomposes and Tungsten get deposited.
 Wavelength region:350~2500nm
DEMERIT:
 It emits the major portion of its radiant energy in near IR
region of the spectrum.
11
Deuterium and Hydrogen lamp
 By electrical excitation of hydrogen or deuterium at
low pressure.
 A pair of electrodes is enclosed in a glass tube filled
with hydrogen gas.
 Current is passed → discharge of electrons occurs
→ excitation of hydrogen molecules → emission of
UV radiations in near UV region.
 They are stable.
12
 Wavelength range: 160-800nm.
 Quartz windows must be employed.
13
XENON ARC LAMP:
 Two tungsten electrodes separated by some distance.
 enclosed in a glass tube (for visible) with quartz or
fused silica and xenon gas is filled under pressure.
 An intense arc is formed between electrods.
 Good source of continuous plus additional intense
radiation. Its intensity is higher than the hydrogen
discharge lamp.
DEMERIT:
 The lamp since operates at high voltage becomes very
hot during operation and hence needs thermal
insulation.
14
 Wavelength range: 200-1000nm
 Peak intensity at about 500nm.
15
MERCURY ARC LAMP
 Mercury vapour is stored under high pressure
and excitation of mercury atoms is done by
electric discharge.
 Not suitable for continuous spectral studies,
because of the presence of sharp lines or
bands.
16
LIGHT EMITTING DIODE
 p-n junction device.
 When forward biased, produces radiant
energy.
 Diodes are made from,(mixtures of)
 Gallium aluminium arsenide
 Gallium arsenic phosphide
 Gallium phosphide
 Gallium nitride
 Indium gallium nitride.
17
 Wavelength region: 375-1000nm.
MERITS
 Long life time
 smaller environmental impact than tungsten
filament lamp.
18
19
WAVELENGTHSELECTORS
Filters and Monochromators
Filters
 A device that allows light of the required
wavelength to pass but absorbs light of other
wavelength wholly or partially.
 Suitable filter can select a desired wavelength
band.
 A particular filter may be used for a specific
analysis.
Absorption filters
Interference filters
20
Absorption filters - works by selective
absorption of
unwanted radiation and transmits the radiation
which is required.
 Examples- Glass and Gelatin filters.
 Selection of absorption filter is done
according to
the following procedure:
21
 Draw a filter wheel.
 Write the colour VIBGYOR in clockwise or
anticlockwise manner, omitting Indigo.
22
 If solution to be analyzed is BLUE in colour a filter
having a complimentary colour ORANGE is used in the
analysis.
 Similarly, we can select the required filter in colorimeter,
based upon the colour of the solution.
 An Absorption glass filter is made of solid sheet of glass
that has been coloured by pigments which is dissolved
or dispersed in the glass.
 The colour in the glass filters are produced by
incorporating metal oxides like V, Cr, Mn, Fe, Ni, Co, Cu
etc.
23
INTERFERENCE FILTERS
 Works on the interference phenomenon,
causes rejection of unwanted wavelength by
selective reflection.
 It is constructed by using two parallel glass
plates, which are silvered internally and
separated by thin film of dielectric
material(CaF2, SiO, MgF2).
 These filters have a band pass of 10-15nm
with peak transmittance of 40-60%.
24
Merits -
 Provide greater
transmittance and
narrower band pass
(10-15nm) as
compare to
absorption filter.
 Inexpensive
 Additional filters can
be used to cut off
undesired
wavelength.
25
MONOCHROMATORS
 A device used to isolate the radiation of the
desired wavelength from wavelength of the
continuous spectra.
PARTS
1. An entrance slit
2. A collimating lens
3. A dispersing device (usually a prism or
grating)
4. A focusing lens
5. An exit slit
26
The dispersing element in monochromator is 2 types;
• Prism
• Grating
27
PRISM MONOCHROMATOR
 Prism is made from glass, Quartz or fused
silica.
 Quartz or fused silica is the choice of
material of UV spectrum.
 Light is passes → dispersion of
polychromatic light→ rotation of the prism
different wavelengths of the spectrum can be
made to pass through in exit slit on the
28
29
Refraction type
30
Reflection type
31
GRATING MONOCHROMTOR
 Are most effective one in converting a
polychromatic light to monochromatic light. As a
resolution of ±0.1nm could be achieved by using
gratings, they are commonly used in
spectrophotometers.
2 types.
1. Diffraction grating.
2. Transmission gratings.
32
DIFFRACTION GRATING
 More refined dispersion of light is obtained by
means of diffraction gratings.
 consist of large number of parallel lines (
grooves) about 15000-30000/ inch is ruled on
highly polished surface of aluminium.
33
34
TRANSMISSION GRATING
 Similar to diffraction grating
 Refraction takes place instead of reflection.
 Refraction produces reinforcement.
 Occurs when radiation transmitted through
grating reinforces with the partially refracted
radiation.
35
36
SAMPLE CONTAINERS
 The cells or cuvettes are used for handling
liquid samples.
 The cell may either be rectangular or
cylindrical in nature.
UV region- quartz or fused silica
Visible region- glass
37
38
DETECTORS
 Device which converts light energy into
electrical signals.
 The transmitted radiation falls on the detector
which determines the intensity of radiation
absorbed by sample.
3 TYPES.
 Barrier layer cell/Photovoltaic cell
 Phototubes/ Photo emissive tube
 Photomultiplier tube
39
Barrier layer cell/Photovoltaic
cell
Consists of thin film metallic layer coated with
silver or gold and acts as an electrode.
 metal base plate which acts as another electrode.
 These two layers are separated by a
semiconductor layer of selenium.
40
 Light radiation falls on selenium layer →
electrons become mobile → taken up by
transparent metal layer.
 Creates a potential difference between two
electrodes & causes the flow of current.
 A flow of current observed in galvanometer
which is proportional to the intensity and
wavelength of light falling on it.
41
42
PHOTOTUBES/ PHOTOEMISSIVE
TUBES
 Consists of a evacuated glass tube with a
photocathode and a collector anode.
 surface of photocathode is coated with a layer
of elements like caesium, silver oxide or
mixture of them.
 radiant energy falls on photosensitive
cathode, electrons are emitted which are
attracted to anode causing current to flow.
 More sensitive compared to barrier layer cell.
43
44
PHOTOMULTIPLIER TUBE
 Multiplication of photoelectrons by secondary
emission of electrons.
 In a vacuum tube, a primary photo-cathode is
fixed which receives radiation from the sample.
 Eight to ten dynodes are fixed each with
increasing potential of 75-100V higher than
preceding one.
 last dynode is fixed an anode or electron collector
electrode.
 extremely sensitive to light and is best suited
where weaker or low radiation is received.
45
46
47
READOUT DEVICE
 The data from a detector are displayed by a
readout device, such as an analogue meter, a
light beam reflected on a scale, or a digital
display , or LCD .
 The output can also be transmitted to a
computer or printer.
48
INSTRUMENTS
PHOTOMETER
SPECTOPHOTOMETER
49
PHOTOMETER
An instrument for measuring the of
light or the relative intensity of a pair of
lights. Also called an illuminometer. It
utilizes filter to isolate a narrow
wavelength region.
 Single beam
 Double beam
50
SPECTOPHOTOMETER
An instrument measures the ratio, or a
function of the two, of the radiant power of
two EM beams over a large wavelength
region. It utilizes dispersing element
(Prisms/Gratings) instead of filters, to scan
large wavelength region.
Single beam
Double beam
51
SINGLE BEAM VISIBLE
PHOTOMETERS
 A direct reading instrument consisting of
 tungsten filament/LED as a source
 lens to provide a parallel beam
 a filter and a photodiode transducer
 The current produced by the photodiode is
processed with electronics or a computer to give a
readout in absorbance.
 Sample is inserted into the light path.
 The absorbance is calculated as the logarithm of
the ratio of these signals.
52
53
Double beam visible
photometer
 Used to measure the absorbance of sample in a flowing
stream.
 The light beam is split by a 2 branched fibre optic.
 One beam is passes – sample
 Other beam- reference.
 Filters are placed after the cells before the photodiode
with nearly identical response.
 Electrical outputs→ voltages and signals are processed by a
log ratio amplifier or a computer to give a readout
proportional to absorbance.
54
55
UV absorption photometers
 Serve as detectors in HPLC.
 Mercury vapour lamp- source.
 Wavelength range 254nm.
56
Spectrophotometers
Visible spectrophotometers
 Wavelength range:380-800nm.
 Simple and inexpensive.
 Eg; Spectronic20
 Tungsten filament – source
 Diffraction by a simple reflection grating→
radiation passes→ sample/reference
57
58
59
UV spectrophotometers
Single beam computerised spectrophotometers
 Light from the source is carried through lens
and/or through aperture to pass through a
suitable filter.
 The type of filter to be used is governed by the
colour of the solution.
 The sample solution to be analysed is placed in
cuvettes.
60
61
 After passing through the solution, the light
strikes the surface of detector (barrier-layer
cell or phototube) and produces electrical
current.
 The output of current is measured by the
deflection of needle of light-spot galvanometer
or micro ammeter.
62
63
Double beam
spectrophotometers
 two beams are formed in the space by a U
shaped mirror called as beam splitter or beam
chopper.
 Chopper is a device consisting of a circular disc.
One third of the disc is opaque and one third is
transparent, remaining one third is mirrored. It
splits the monochromatic beam of light into two
beams of equal intensities.
64
65
 Eg; Varian Cary100
66
CHOICE OF SOLVENTS
 It should not itself absorb radiations in the
region under investigations.
 It should be less polar so that it has minimum
interaction with the solute molecule.
 Most commonly: 95%ethanol
 Cheap, good dissolving power, does not
absorb radiation above 210nm.
 Typical examples
67
Solvent Wavelength(nm)
Water 205
Methanol 210
Ethanol 210
Ether 210
Chloroform 245
Carbon tetra chloride 265
Cyclohexane 210
68
SOLVENT EFFECT
 The position and intensity of an absorption band
may shift when the spectrum is recorded in
different solvents.
 A dilute sample solution is prefered for analysis.
 Most commonly used solvent is 95% ethanol. It is
best solvent as it is cheap,transparent down to
210nm.
 Position as well as intensity of absorption maxima
get shifted for a particular chromophore by
changing the polarity of solvent.
 By increasing polarity of solvent→dienes,
conjugated hydrocarbons→ no shift
69
 α,β unsaturated carbonyl compound show 2
different shifts.
n →Π* transition.
 Absorption band moves to shorter
wavelength(blue shift) by increasing the
polarity of the solvent.
 Ground state is more polar as compared to the
excited state.
Eg: absorption maximum of acetone is at
279nm in hexane as compared to 264nm in
water.
70
Π→Π* transition.
 Absorption band moves to longer
wavelength(red shift) by increasing the polarity
of the solvent.
 The dipole-dipole interaction with the solvent
molecules lower the energy of excited state
more than that of the ground state.
71
Effect of solvent polarity on the
various types of bands.
K-band
 Due to conjugated enes & enones are affected
differently by changing the polarity of the solvent.
 K bands due to conjugated dienes are not
affected by changing the polarity of the solvent,
while these bands due to enones shows a red
shift by increasing the
 polarity of solvent.
R band
 The absorption shifts to shorter wavelength (blue
shift) with increasing polarity of solvent.
72
B band
 The position as well as the intensity of the
band is not shifted by increasing the polarity of
the solvent.
 But the heterocyclic aromatic compound, a
marked hyperchromic shift is observed by
increasing the polarity of the solvent.
73
REFERENCE
 Principles of instrumental analysis – Doglas A
Skoog, James holler, Timothy a nieman,
5thedition.
 Instrumental methods of analysis by Willards,
7thedition.
 Instrumental methods of chemical analysis by
H.Kaur
 Instrumental methods of chemical analysis by
B.K Sharma.
 Instrumental methods of chemical analysis by
Gurdeep R.Chatwal, Sham K.Anand.
74
75
QUERIES
76

More Related Content

What's hot

PRINCIPLES of FT-NMR & 13C NMR
PRINCIPLES of FT-NMR & 13C NMRPRINCIPLES of FT-NMR & 13C NMR
PRINCIPLES of FT-NMR & 13C NMRAditya Sharma
 
Mass fragmentation & rules
Mass fragmentation & rulesMass fragmentation & rules
Mass fragmentation & rulesMehulJain143
 
IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...
IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...
IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...Dr. UMESH KUMAR SHARMA
 
Fragmentation techniques in mass spectroscopy
Fragmentation techniques in mass spectroscopyFragmentation techniques in mass spectroscopy
Fragmentation techniques in mass spectroscopyMahendra G S
 
uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS
 uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS
uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICSHarevindarsingh
 
Mass spectrometry(Ionization Techniques) by Ashutosh Panke
Mass spectrometry(Ionization Techniques) by Ashutosh PankeMass spectrometry(Ionization Techniques) by Ashutosh Panke
Mass spectrometry(Ionization Techniques) by Ashutosh PankeAshutosh Panke
 
Types of crystals & Application of x ray
 Types of crystals & Application of x ray Types of crystals & Application of x ray
Types of crystals & Application of x raykajal pradhan
 
Factors affecting vibrational frequencies in IR
Factors affecting vibrational frequencies in IRFactors affecting vibrational frequencies in IR
Factors affecting vibrational frequencies in IRRashmi Raveendran
 
Ionizaion Techniques - Mass Spectroscopy
Ionizaion Techniques - Mass SpectroscopyIonizaion Techniques - Mass Spectroscopy
Ionizaion Techniques - Mass SpectroscopySuraj Choudhary
 
Sampling techniques of Infrared Spectroscopy
Sampling techniques of Infrared SpectroscopySampling techniques of Infrared Spectroscopy
Sampling techniques of Infrared SpectroscopyPARITALA JAGADEESH
 
UV-Visible Spectroscopy
UV-Visible SpectroscopyUV-Visible Spectroscopy
UV-Visible SpectroscopySudha Reddy
 
Spectroflourimetry
SpectroflourimetrySpectroflourimetry
SpectroflourimetryXenChisti
 
IR Spectroscopy - Sudheerkumar Kamarapu, M. Pharmacy Lecture pdf
IR Spectroscopy - Sudheerkumar Kamarapu, M. Pharmacy Lecture pdfIR Spectroscopy - Sudheerkumar Kamarapu, M. Pharmacy Lecture pdf
IR Spectroscopy - Sudheerkumar Kamarapu, M. Pharmacy Lecture pdfDr. Sudheer Kumar Kamarapu
 
Quenching of Fluorescence
Quenching of FluorescenceQuenching of Fluorescence
Quenching of FluorescenceFatema Zohora
 
NMR Spectroscopy By Himaja Donthula
NMR Spectroscopy By Himaja DonthulaNMR Spectroscopy By Himaja Donthula
NMR Spectroscopy By Himaja Donthulahimaja donthula
 
Solvents and solvent effect in UV - Vis Spectroscopy, By Dr. Umesh Kumar sh...
Solvents and  solvent effect in UV -  Vis Spectroscopy, By Dr. Umesh Kumar sh...Solvents and  solvent effect in UV -  Vis Spectroscopy, By Dr. Umesh Kumar sh...
Solvents and solvent effect in UV - Vis Spectroscopy, By Dr. Umesh Kumar sh...Dr. UMESH KUMAR SHARMA
 
Quantum numbers and its roles in nmr
Quantum numbers and its roles in nmrQuantum numbers and its roles in nmr
Quantum numbers and its roles in nmrAtul Adhikari
 
Uv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentationUv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentationAnusreeAnu11
 

What's hot (20)

PRINCIPLES of FT-NMR & 13C NMR
PRINCIPLES of FT-NMR & 13C NMRPRINCIPLES of FT-NMR & 13C NMR
PRINCIPLES of FT-NMR & 13C NMR
 
Mass fragmentation & rules
Mass fragmentation & rulesMass fragmentation & rules
Mass fragmentation & rules
 
IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...
IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...
IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...
 
Fragmentation techniques in mass spectroscopy
Fragmentation techniques in mass spectroscopyFragmentation techniques in mass spectroscopy
Fragmentation techniques in mass spectroscopy
 
Flourimetry factors
Flourimetry factorsFlourimetry factors
Flourimetry factors
 
uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS
 uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS
uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS
 
Mass spectrometry(Ionization Techniques) by Ashutosh Panke
Mass spectrometry(Ionization Techniques) by Ashutosh PankeMass spectrometry(Ionization Techniques) by Ashutosh Panke
Mass spectrometry(Ionization Techniques) by Ashutosh Panke
 
Types of crystals & Application of x ray
 Types of crystals & Application of x ray Types of crystals & Application of x ray
Types of crystals & Application of x ray
 
Factors affecting vibrational frequencies in IR
Factors affecting vibrational frequencies in IRFactors affecting vibrational frequencies in IR
Factors affecting vibrational frequencies in IR
 
Ionizaion Techniques - Mass Spectroscopy
Ionizaion Techniques - Mass SpectroscopyIonizaion Techniques - Mass Spectroscopy
Ionizaion Techniques - Mass Spectroscopy
 
Sampling techniques of Infrared Spectroscopy
Sampling techniques of Infrared SpectroscopySampling techniques of Infrared Spectroscopy
Sampling techniques of Infrared Spectroscopy
 
UV-Visible Spectroscopy
UV-Visible SpectroscopyUV-Visible Spectroscopy
UV-Visible Spectroscopy
 
Spectroflourimetry
SpectroflourimetrySpectroflourimetry
Spectroflourimetry
 
IR Spectroscopy - Sudheerkumar Kamarapu, M. Pharmacy Lecture pdf
IR Spectroscopy - Sudheerkumar Kamarapu, M. Pharmacy Lecture pdfIR Spectroscopy - Sudheerkumar Kamarapu, M. Pharmacy Lecture pdf
IR Spectroscopy - Sudheerkumar Kamarapu, M. Pharmacy Lecture pdf
 
Quenching of Fluorescence
Quenching of FluorescenceQuenching of Fluorescence
Quenching of Fluorescence
 
NMR Spectroscopy
NMR SpectroscopyNMR Spectroscopy
NMR Spectroscopy
 
NMR Spectroscopy By Himaja Donthula
NMR Spectroscopy By Himaja DonthulaNMR Spectroscopy By Himaja Donthula
NMR Spectroscopy By Himaja Donthula
 
Solvents and solvent effect in UV - Vis Spectroscopy, By Dr. Umesh Kumar sh...
Solvents and  solvent effect in UV -  Vis Spectroscopy, By Dr. Umesh Kumar sh...Solvents and  solvent effect in UV -  Vis Spectroscopy, By Dr. Umesh Kumar sh...
Solvents and solvent effect in UV - Vis Spectroscopy, By Dr. Umesh Kumar sh...
 
Quantum numbers and its roles in nmr
Quantum numbers and its roles in nmrQuantum numbers and its roles in nmr
Quantum numbers and its roles in nmr
 
Uv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentationUv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentation
 

Similar to UV-Vis Spectroscopy: Instrumentation and Solvent Effects

UV-VISIBLE SPECTROSCOPY INSTRUMENTATION.pptx
UV-VISIBLE SPECTROSCOPY INSTRUMENTATION.pptxUV-VISIBLE SPECTROSCOPY INSTRUMENTATION.pptx
UV-VISIBLE SPECTROSCOPY INSTRUMENTATION.pptxDr. Vijaya Barge
 
Instrumentation of uv spectroscopy
Instrumentation of uv spectroscopyInstrumentation of uv spectroscopy
Instrumentation of uv spectroscopyArabinda Changmai
 
Seminar on Uv Visible spectroscopy by Amogh G V
Seminar on Uv Visible spectroscopy by Amogh G VSeminar on Uv Visible spectroscopy by Amogh G V
Seminar on Uv Visible spectroscopy by Amogh G VAmoghGV
 
UV Spectroscopy 1 (1).ppsx
UV Spectroscopy 1 (1).ppsxUV Spectroscopy 1 (1).ppsx
UV Spectroscopy 1 (1).ppsxAyush Chavhan
 
Ultra violet spectrometer
Ultra violet spectrometerUltra violet spectrometer
Ultra violet spectrometeranshu singh
 
UV SPECTROSCOPY AND INSTRUMENTATION
UV SPECTROSCOPY AND INSTRUMENTATION UV SPECTROSCOPY AND INSTRUMENTATION
UV SPECTROSCOPY AND INSTRUMENTATION Geeta Prasad Kashyap
 
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRYINSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRYAshwini Somayaji
 
uv spectelectronic transition in the roscopy
uv spectelectronic transition in the roscopyuv spectelectronic transition in the roscopy
uv spectelectronic transition in the roscopyRiyaDas765755
 
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...mariomS7
 
Ultraviolet and visible spectroscopy
Ultraviolet and visible  spectroscopyUltraviolet and visible  spectroscopy
Ultraviolet and visible spectroscopyGeetaraniLoushigam
 
Uv/visible spectroscopy
Uv/visible spectroscopyUv/visible spectroscopy
Uv/visible spectroscopySavitaBhatt2
 
INSTRUMENTATIONS UV-VIS.pdf
INSTRUMENTATIONS UV-VIS.pdfINSTRUMENTATIONS UV-VIS.pdf
INSTRUMENTATIONS UV-VIS.pdfDr. Dinesh Mehta
 
Uv visible spectrphotometery
Uv visible spectrphotometeryUv visible spectrphotometery
Uv visible spectrphotometeryWasim Sajjad
 
Spectrophotometry: basic concepts, instrumentation and application
Spectrophotometry: basic concepts, instrumentation and applicationSpectrophotometry: basic concepts, instrumentation and application
Spectrophotometry: basic concepts, instrumentation and applicationBasil "Lexi" Bruno
 

Similar to UV-Vis Spectroscopy: Instrumentation and Solvent Effects (20)

UV-VISIBLE SPECTROSCOPY INSTRUMENTATION.pptx
UV-VISIBLE SPECTROSCOPY INSTRUMENTATION.pptxUV-VISIBLE SPECTROSCOPY INSTRUMENTATION.pptx
UV-VISIBLE SPECTROSCOPY INSTRUMENTATION.pptx
 
U.V Spectroscopy.
U.V Spectroscopy.U.V Spectroscopy.
U.V Spectroscopy.
 
Instrumentation of uv spectroscopy
Instrumentation of uv spectroscopyInstrumentation of uv spectroscopy
Instrumentation of uv spectroscopy
 
Seminar on Uv Visible spectroscopy by Amogh G V
Seminar on Uv Visible spectroscopy by Amogh G VSeminar on Uv Visible spectroscopy by Amogh G V
Seminar on Uv Visible spectroscopy by Amogh G V
 
UV Spectroscopy 1 (1).ppsx
UV Spectroscopy 1 (1).ppsxUV Spectroscopy 1 (1).ppsx
UV Spectroscopy 1 (1).ppsx
 
Ultra violet spectrometer
Ultra violet spectrometerUltra violet spectrometer
Ultra violet spectrometer
 
UV SPECTROSCOPY AND INSTRUMENTATION
UV SPECTROSCOPY AND INSTRUMENTATION UV SPECTROSCOPY AND INSTRUMENTATION
UV SPECTROSCOPY AND INSTRUMENTATION
 
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRYINSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
 
Basic spectrophotometry
Basic spectrophotometryBasic spectrophotometry
Basic spectrophotometry
 
UV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptxUV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptx
 
UV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptxUV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptx
 
uv spectelectronic transition in the roscopy
uv spectelectronic transition in the roscopyuv spectelectronic transition in the roscopy
uv spectelectronic transition in the roscopy
 
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
 
Ultraviolet and visible spectroscopy
Ultraviolet and visible  spectroscopyUltraviolet and visible  spectroscopy
Ultraviolet and visible spectroscopy
 
Uv/visible spectroscopy
Uv/visible spectroscopyUv/visible spectroscopy
Uv/visible spectroscopy
 
Uv
UvUv
Uv
 
INSTRUMENTATIONS UV-VIS.pdf
INSTRUMENTATIONS UV-VIS.pdfINSTRUMENTATIONS UV-VIS.pdf
INSTRUMENTATIONS UV-VIS.pdf
 
Spectrophotometer introducing
Spectrophotometer  introducingSpectrophotometer  introducing
Spectrophotometer introducing
 
Uv visible spectrphotometery
Uv visible spectrphotometeryUv visible spectrphotometery
Uv visible spectrphotometery
 
Spectrophotometry: basic concepts, instrumentation and application
Spectrophotometry: basic concepts, instrumentation and applicationSpectrophotometry: basic concepts, instrumentation and application
Spectrophotometry: basic concepts, instrumentation and application
 

More from Binuja S.S

Transdermal drug delivery system
Transdermal drug delivery systemTransdermal drug delivery system
Transdermal drug delivery systemBinuja S.S
 
RESEARCH METHODOLOGY
RESEARCH METHODOLOGYRESEARCH METHODOLOGY
RESEARCH METHODOLOGYBinuja S.S
 
APPLICATIONS OF QSAR
APPLICATIONS OF QSARAPPLICATIONS OF QSAR
APPLICATIONS OF QSARBinuja S.S
 
PHOTO-OXIDATION
PHOTO-OXIDATIONPHOTO-OXIDATION
PHOTO-OXIDATIONBinuja S.S
 
ION-EXCLUSION CHROMATOGRAPHY
ION-EXCLUSION CHROMATOGRAPHYION-EXCLUSION CHROMATOGRAPHY
ION-EXCLUSION CHROMATOGRAPHYBinuja S.S
 
SYNTHETIC REAGENTS AND APPLICATION
SYNTHETIC REAGENTS AND APPLICATIONSYNTHETIC REAGENTS AND APPLICATION
SYNTHETIC REAGENTS AND APPLICATIONBinuja S.S
 
ENZYME INHIBITION
ENZYME INHIBITIONENZYME INHIBITION
ENZYME INHIBITIONBinuja S.S
 
CASE PRESENTATION ON RHEUMATOID ARTHRITIS
CASE PRESENTATION ON RHEUMATOID ARTHRITISCASE PRESENTATION ON RHEUMATOID ARTHRITIS
CASE PRESENTATION ON RHEUMATOID ARTHRITISBinuja S.S
 
VACUUM ASSISTED WOUND THERAPY
VACUUM ASSISTED WOUND THERAPYVACUUM ASSISTED WOUND THERAPY
VACUUM ASSISTED WOUND THERAPYBinuja S.S
 

More from Binuja S.S (12)

Transdermal drug delivery system
Transdermal drug delivery systemTransdermal drug delivery system
Transdermal drug delivery system
 
Aerosols
AerosolsAerosols
Aerosols
 
RESEARCH METHODOLOGY
RESEARCH METHODOLOGYRESEARCH METHODOLOGY
RESEARCH METHODOLOGY
 
APPLICATIONS OF QSAR
APPLICATIONS OF QSARAPPLICATIONS OF QSAR
APPLICATIONS OF QSAR
 
PHOTO-OXIDATION
PHOTO-OXIDATIONPHOTO-OXIDATION
PHOTO-OXIDATION
 
ION-EXCLUSION CHROMATOGRAPHY
ION-EXCLUSION CHROMATOGRAPHYION-EXCLUSION CHROMATOGRAPHY
ION-EXCLUSION CHROMATOGRAPHY
 
EVAPORATION
EVAPORATIONEVAPORATION
EVAPORATION
 
SYNTHETIC REAGENTS AND APPLICATION
SYNTHETIC REAGENTS AND APPLICATIONSYNTHETIC REAGENTS AND APPLICATION
SYNTHETIC REAGENTS AND APPLICATION
 
ENZYME INHIBITION
ENZYME INHIBITIONENZYME INHIBITION
ENZYME INHIBITION
 
Steroids
SteroidsSteroids
Steroids
 
CASE PRESENTATION ON RHEUMATOID ARTHRITIS
CASE PRESENTATION ON RHEUMATOID ARTHRITISCASE PRESENTATION ON RHEUMATOID ARTHRITIS
CASE PRESENTATION ON RHEUMATOID ARTHRITIS
 
VACUUM ASSISTED WOUND THERAPY
VACUUM ASSISTED WOUND THERAPYVACUUM ASSISTED WOUND THERAPY
VACUUM ASSISTED WOUND THERAPY
 

Recently uploaded

专业一比一美国俄亥俄大学毕业证成绩单pdf电子版制作修改
专业一比一美国俄亥俄大学毕业证成绩单pdf电子版制作修改专业一比一美国俄亥俄大学毕业证成绩单pdf电子版制作修改
专业一比一美国俄亥俄大学毕业证成绩单pdf电子版制作修改yuu sss
 
Industrialised data - the key to AI success.pdf
Industrialised data - the key to AI success.pdfIndustrialised data - the key to AI success.pdf
Industrialised data - the key to AI success.pdfLars Albertsson
 
RadioAdProWritingCinderellabyButleri.pdf
RadioAdProWritingCinderellabyButleri.pdfRadioAdProWritingCinderellabyButleri.pdf
RadioAdProWritingCinderellabyButleri.pdfgstagge
 
dokumen.tips_chapter-4-transient-heat-conduction-mehmet-kanoglu.ppt
dokumen.tips_chapter-4-transient-heat-conduction-mehmet-kanoglu.pptdokumen.tips_chapter-4-transient-heat-conduction-mehmet-kanoglu.ppt
dokumen.tips_chapter-4-transient-heat-conduction-mehmet-kanoglu.pptSonatrach
 
Indian Call Girls in Abu Dhabi O5286O24O8 Call Girls in Abu Dhabi By Independ...
Indian Call Girls in Abu Dhabi O5286O24O8 Call Girls in Abu Dhabi By Independ...Indian Call Girls in Abu Dhabi O5286O24O8 Call Girls in Abu Dhabi By Independ...
Indian Call Girls in Abu Dhabi O5286O24O8 Call Girls in Abu Dhabi By Independ...dajasot375
 
毕业文凭制作#回国入职#diploma#degree澳洲中央昆士兰大学毕业证成绩单pdf电子版制作修改#毕业文凭制作#回国入职#diploma#degree
毕业文凭制作#回国入职#diploma#degree澳洲中央昆士兰大学毕业证成绩单pdf电子版制作修改#毕业文凭制作#回国入职#diploma#degree毕业文凭制作#回国入职#diploma#degree澳洲中央昆士兰大学毕业证成绩单pdf电子版制作修改#毕业文凭制作#回国入职#diploma#degree
毕业文凭制作#回国入职#diploma#degree澳洲中央昆士兰大学毕业证成绩单pdf电子版制作修改#毕业文凭制作#回国入职#diploma#degreeyuu sss
 
vip Sarai Rohilla Call Girls 9999965857 Call or WhatsApp Now Book
vip Sarai Rohilla Call Girls 9999965857 Call or WhatsApp Now Bookvip Sarai Rohilla Call Girls 9999965857 Call or WhatsApp Now Book
vip Sarai Rohilla Call Girls 9999965857 Call or WhatsApp Now Bookmanojkuma9823
 
Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...
Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...
Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...ThinkInnovation
 
Call Girls In Dwarka 9654467111 Escorts Service
Call Girls In Dwarka 9654467111 Escorts ServiceCall Girls In Dwarka 9654467111 Escorts Service
Call Girls In Dwarka 9654467111 Escorts ServiceSapana Sha
 
20240419 - Measurecamp Amsterdam - SAM.pdf
20240419 - Measurecamp Amsterdam - SAM.pdf20240419 - Measurecamp Amsterdam - SAM.pdf
20240419 - Measurecamp Amsterdam - SAM.pdfHuman37
 
办理(Vancouver毕业证书)加拿大温哥华岛大学毕业证成绩单原版一比一
办理(Vancouver毕业证书)加拿大温哥华岛大学毕业证成绩单原版一比一办理(Vancouver毕业证书)加拿大温哥华岛大学毕业证成绩单原版一比一
办理(Vancouver毕业证书)加拿大温哥华岛大学毕业证成绩单原版一比一F La
 
科罗拉多大学波尔得分校毕业证学位证成绩单-可办理
科罗拉多大学波尔得分校毕业证学位证成绩单-可办理科罗拉多大学波尔得分校毕业证学位证成绩单-可办理
科罗拉多大学波尔得分校毕业证学位证成绩单-可办理e4aez8ss
 
INTERNSHIP ON PURBASHA COMPOSITE TEX LTD
INTERNSHIP ON PURBASHA COMPOSITE TEX LTDINTERNSHIP ON PURBASHA COMPOSITE TEX LTD
INTERNSHIP ON PURBASHA COMPOSITE TEX LTDRafezzaman
 
Call Us ➥97111√47426🤳Call Girls in Aerocity (Delhi NCR)
Call Us ➥97111√47426🤳Call Girls in Aerocity (Delhi NCR)Call Us ➥97111√47426🤳Call Girls in Aerocity (Delhi NCR)
Call Us ➥97111√47426🤳Call Girls in Aerocity (Delhi NCR)jennyeacort
 
Saket, (-DELHI )+91-9654467111-(=)CHEAP Call Girls in Escorts Service Saket C...
Saket, (-DELHI )+91-9654467111-(=)CHEAP Call Girls in Escorts Service Saket C...Saket, (-DELHI )+91-9654467111-(=)CHEAP Call Girls in Escorts Service Saket C...
Saket, (-DELHI )+91-9654467111-(=)CHEAP Call Girls in Escorts Service Saket C...Sapana Sha
 
VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130
VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130
VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130Suhani Kapoor
 
RS 9000 Call In girls Dwarka Mor (DELHI)⇛9711147426🔝Delhi
RS 9000 Call In girls Dwarka Mor (DELHI)⇛9711147426🔝DelhiRS 9000 Call In girls Dwarka Mor (DELHI)⇛9711147426🔝Delhi
RS 9000 Call In girls Dwarka Mor (DELHI)⇛9711147426🔝Delhijennyeacort
 
04242024_CCC TUG_Joins and Relationships
04242024_CCC TUG_Joins and Relationships04242024_CCC TUG_Joins and Relationships
04242024_CCC TUG_Joins and Relationshipsccctableauusergroup
 
RA-11058_IRR-COMPRESS Do 198 series of 1998
RA-11058_IRR-COMPRESS Do 198 series of 1998RA-11058_IRR-COMPRESS Do 198 series of 1998
RA-11058_IRR-COMPRESS Do 198 series of 1998YohFuh
 

Recently uploaded (20)

专业一比一美国俄亥俄大学毕业证成绩单pdf电子版制作修改
专业一比一美国俄亥俄大学毕业证成绩单pdf电子版制作修改专业一比一美国俄亥俄大学毕业证成绩单pdf电子版制作修改
专业一比一美国俄亥俄大学毕业证成绩单pdf电子版制作修改
 
Call Girls in Saket 99530🔝 56974 Escort Service
Call Girls in Saket 99530🔝 56974 Escort ServiceCall Girls in Saket 99530🔝 56974 Escort Service
Call Girls in Saket 99530🔝 56974 Escort Service
 
Industrialised data - the key to AI success.pdf
Industrialised data - the key to AI success.pdfIndustrialised data - the key to AI success.pdf
Industrialised data - the key to AI success.pdf
 
RadioAdProWritingCinderellabyButleri.pdf
RadioAdProWritingCinderellabyButleri.pdfRadioAdProWritingCinderellabyButleri.pdf
RadioAdProWritingCinderellabyButleri.pdf
 
dokumen.tips_chapter-4-transient-heat-conduction-mehmet-kanoglu.ppt
dokumen.tips_chapter-4-transient-heat-conduction-mehmet-kanoglu.pptdokumen.tips_chapter-4-transient-heat-conduction-mehmet-kanoglu.ppt
dokumen.tips_chapter-4-transient-heat-conduction-mehmet-kanoglu.ppt
 
Indian Call Girls in Abu Dhabi O5286O24O8 Call Girls in Abu Dhabi By Independ...
Indian Call Girls in Abu Dhabi O5286O24O8 Call Girls in Abu Dhabi By Independ...Indian Call Girls in Abu Dhabi O5286O24O8 Call Girls in Abu Dhabi By Independ...
Indian Call Girls in Abu Dhabi O5286O24O8 Call Girls in Abu Dhabi By Independ...
 
毕业文凭制作#回国入职#diploma#degree澳洲中央昆士兰大学毕业证成绩单pdf电子版制作修改#毕业文凭制作#回国入职#diploma#degree
毕业文凭制作#回国入职#diploma#degree澳洲中央昆士兰大学毕业证成绩单pdf电子版制作修改#毕业文凭制作#回国入职#diploma#degree毕业文凭制作#回国入职#diploma#degree澳洲中央昆士兰大学毕业证成绩单pdf电子版制作修改#毕业文凭制作#回国入职#diploma#degree
毕业文凭制作#回国入职#diploma#degree澳洲中央昆士兰大学毕业证成绩单pdf电子版制作修改#毕业文凭制作#回国入职#diploma#degree
 
vip Sarai Rohilla Call Girls 9999965857 Call or WhatsApp Now Book
vip Sarai Rohilla Call Girls 9999965857 Call or WhatsApp Now Bookvip Sarai Rohilla Call Girls 9999965857 Call or WhatsApp Now Book
vip Sarai Rohilla Call Girls 9999965857 Call or WhatsApp Now Book
 
Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...
Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...
Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...
 
Call Girls In Dwarka 9654467111 Escorts Service
Call Girls In Dwarka 9654467111 Escorts ServiceCall Girls In Dwarka 9654467111 Escorts Service
Call Girls In Dwarka 9654467111 Escorts Service
 
20240419 - Measurecamp Amsterdam - SAM.pdf
20240419 - Measurecamp Amsterdam - SAM.pdf20240419 - Measurecamp Amsterdam - SAM.pdf
20240419 - Measurecamp Amsterdam - SAM.pdf
 
办理(Vancouver毕业证书)加拿大温哥华岛大学毕业证成绩单原版一比一
办理(Vancouver毕业证书)加拿大温哥华岛大学毕业证成绩单原版一比一办理(Vancouver毕业证书)加拿大温哥华岛大学毕业证成绩单原版一比一
办理(Vancouver毕业证书)加拿大温哥华岛大学毕业证成绩单原版一比一
 
科罗拉多大学波尔得分校毕业证学位证成绩单-可办理
科罗拉多大学波尔得分校毕业证学位证成绩单-可办理科罗拉多大学波尔得分校毕业证学位证成绩单-可办理
科罗拉多大学波尔得分校毕业证学位证成绩单-可办理
 
INTERNSHIP ON PURBASHA COMPOSITE TEX LTD
INTERNSHIP ON PURBASHA COMPOSITE TEX LTDINTERNSHIP ON PURBASHA COMPOSITE TEX LTD
INTERNSHIP ON PURBASHA COMPOSITE TEX LTD
 
Call Us ➥97111√47426🤳Call Girls in Aerocity (Delhi NCR)
Call Us ➥97111√47426🤳Call Girls in Aerocity (Delhi NCR)Call Us ➥97111√47426🤳Call Girls in Aerocity (Delhi NCR)
Call Us ➥97111√47426🤳Call Girls in Aerocity (Delhi NCR)
 
Saket, (-DELHI )+91-9654467111-(=)CHEAP Call Girls in Escorts Service Saket C...
Saket, (-DELHI )+91-9654467111-(=)CHEAP Call Girls in Escorts Service Saket C...Saket, (-DELHI )+91-9654467111-(=)CHEAP Call Girls in Escorts Service Saket C...
Saket, (-DELHI )+91-9654467111-(=)CHEAP Call Girls in Escorts Service Saket C...
 
VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130
VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130
VIP Call Girls Service Miyapur Hyderabad Call +91-8250192130
 
RS 9000 Call In girls Dwarka Mor (DELHI)⇛9711147426🔝Delhi
RS 9000 Call In girls Dwarka Mor (DELHI)⇛9711147426🔝DelhiRS 9000 Call In girls Dwarka Mor (DELHI)⇛9711147426🔝Delhi
RS 9000 Call In girls Dwarka Mor (DELHI)⇛9711147426🔝Delhi
 
04242024_CCC TUG_Joins and Relationships
04242024_CCC TUG_Joins and Relationships04242024_CCC TUG_Joins and Relationships
04242024_CCC TUG_Joins and Relationships
 
RA-11058_IRR-COMPRESS Do 198 series of 1998
RA-11058_IRR-COMPRESS Do 198 series of 1998RA-11058_IRR-COMPRESS Do 198 series of 1998
RA-11058_IRR-COMPRESS Do 198 series of 1998
 

UV-Vis Spectroscopy: Instrumentation and Solvent Effects

  • 2. ULTRAVIOLET-VISIBLE SPECTROSCOPY TOPIC: Instrumentation, choice of solvents and solvent effect. Submitted by BINUJA.S.S First year M.Pharm Department of Pharmaceutical chemistry Ezhuthachan college of pharmaceutical sciences
  • 3.  INTRODUCTION  INSTRUMENTATION  INSTRUMENTS  Uv- visible Photometers  Uv – visible Spectrophotometers  CHOICE OF SOLVENTS AND SOLVENT EFFECT  REFERENCE CONTENTS 3
  • 4. INTRODUCTION  Oldest physical method for quantitative analysis and structural elucidation.  qualitative and quantitative analysis of substances.  The spectrophotometer technique is to measures light intensity as a function of wavelength. 4
  • 5.  It does this by:  diffracting the light beam into a spectrum of wavelengths  direct it to an object  receiving the light reflected or returned from the object  detecting the intensities with a charge-coupled device  displaying the results as a graph on the detector. 5
  • 6. 6
  • 7. Instrumentation Instruments for measuring the absorption of UV-Visible are made up of following components. 1. Sources 2. Wavelength selectors 3. Sample containers 4. Detectors 5. Readout device 7
  • 8. Sources REQUIREMENTS OF AN IDEAL SOURCE  Stable and should not allow fluctuations.  Emit light of continuous radiation.  Provide sufficient intensity. 8
  • 9. UV Radiation sources Visible sources  Deuterium lamp  Hydrogen lamp  Tungsten halogen lamp  Xenon arc lamp  Mercury arc lamp  Tungsten halogen lamp  Mercury vapour lamp  Carbonone lamp 9
  • 10. Tungsten Halogen lamp  filament of Tungsten fixed in evacuated condition and then filled with inert gas.  The filament can be heated up to 3000 k, beyond this tungsten starts sublimating.  It is used when polychromatic light is required. To prevent this along with inert gas some amount of halogen is introduced 10
  • 11.  Sublimated form of tungsten reacts with Iodine to form Tungsten –Iodine complex.  Which migrates back to the hot filament where it decomposes and Tungsten get deposited.  Wavelength region:350~2500nm DEMERIT:  It emits the major portion of its radiant energy in near IR region of the spectrum. 11
  • 12. Deuterium and Hydrogen lamp  By electrical excitation of hydrogen or deuterium at low pressure.  A pair of electrodes is enclosed in a glass tube filled with hydrogen gas.  Current is passed → discharge of electrons occurs → excitation of hydrogen molecules → emission of UV radiations in near UV region.  They are stable. 12
  • 13.  Wavelength range: 160-800nm.  Quartz windows must be employed. 13
  • 14. XENON ARC LAMP:  Two tungsten electrodes separated by some distance.  enclosed in a glass tube (for visible) with quartz or fused silica and xenon gas is filled under pressure.  An intense arc is formed between electrods.  Good source of continuous plus additional intense radiation. Its intensity is higher than the hydrogen discharge lamp. DEMERIT:  The lamp since operates at high voltage becomes very hot during operation and hence needs thermal insulation. 14
  • 15.  Wavelength range: 200-1000nm  Peak intensity at about 500nm. 15
  • 16. MERCURY ARC LAMP  Mercury vapour is stored under high pressure and excitation of mercury atoms is done by electric discharge.  Not suitable for continuous spectral studies, because of the presence of sharp lines or bands. 16
  • 17. LIGHT EMITTING DIODE  p-n junction device.  When forward biased, produces radiant energy.  Diodes are made from,(mixtures of)  Gallium aluminium arsenide  Gallium arsenic phosphide  Gallium phosphide  Gallium nitride  Indium gallium nitride. 17
  • 18.  Wavelength region: 375-1000nm. MERITS  Long life time  smaller environmental impact than tungsten filament lamp. 18
  • 19. 19
  • 20. WAVELENGTHSELECTORS Filters and Monochromators Filters  A device that allows light of the required wavelength to pass but absorbs light of other wavelength wholly or partially.  Suitable filter can select a desired wavelength band.  A particular filter may be used for a specific analysis. Absorption filters Interference filters 20
  • 21. Absorption filters - works by selective absorption of unwanted radiation and transmits the radiation which is required.  Examples- Glass and Gelatin filters.  Selection of absorption filter is done according to the following procedure: 21
  • 22.  Draw a filter wheel.  Write the colour VIBGYOR in clockwise or anticlockwise manner, omitting Indigo. 22
  • 23.  If solution to be analyzed is BLUE in colour a filter having a complimentary colour ORANGE is used in the analysis.  Similarly, we can select the required filter in colorimeter, based upon the colour of the solution.  An Absorption glass filter is made of solid sheet of glass that has been coloured by pigments which is dissolved or dispersed in the glass.  The colour in the glass filters are produced by incorporating metal oxides like V, Cr, Mn, Fe, Ni, Co, Cu etc. 23
  • 24. INTERFERENCE FILTERS  Works on the interference phenomenon, causes rejection of unwanted wavelength by selective reflection.  It is constructed by using two parallel glass plates, which are silvered internally and separated by thin film of dielectric material(CaF2, SiO, MgF2).  These filters have a band pass of 10-15nm with peak transmittance of 40-60%. 24
  • 25. Merits -  Provide greater transmittance and narrower band pass (10-15nm) as compare to absorption filter.  Inexpensive  Additional filters can be used to cut off undesired wavelength. 25
  • 26. MONOCHROMATORS  A device used to isolate the radiation of the desired wavelength from wavelength of the continuous spectra. PARTS 1. An entrance slit 2. A collimating lens 3. A dispersing device (usually a prism or grating) 4. A focusing lens 5. An exit slit 26
  • 27. The dispersing element in monochromator is 2 types; • Prism • Grating 27
  • 28. PRISM MONOCHROMATOR  Prism is made from glass, Quartz or fused silica.  Quartz or fused silica is the choice of material of UV spectrum.  Light is passes → dispersion of polychromatic light→ rotation of the prism different wavelengths of the spectrum can be made to pass through in exit slit on the 28
  • 29. 29
  • 32. GRATING MONOCHROMTOR  Are most effective one in converting a polychromatic light to monochromatic light. As a resolution of ±0.1nm could be achieved by using gratings, they are commonly used in spectrophotometers. 2 types. 1. Diffraction grating. 2. Transmission gratings. 32
  • 33. DIFFRACTION GRATING  More refined dispersion of light is obtained by means of diffraction gratings.  consist of large number of parallel lines ( grooves) about 15000-30000/ inch is ruled on highly polished surface of aluminium. 33
  • 34. 34
  • 35. TRANSMISSION GRATING  Similar to diffraction grating  Refraction takes place instead of reflection.  Refraction produces reinforcement.  Occurs when radiation transmitted through grating reinforces with the partially refracted radiation. 35
  • 36. 36
  • 37. SAMPLE CONTAINERS  The cells or cuvettes are used for handling liquid samples.  The cell may either be rectangular or cylindrical in nature. UV region- quartz or fused silica Visible region- glass 37
  • 38. 38
  • 39. DETECTORS  Device which converts light energy into electrical signals.  The transmitted radiation falls on the detector which determines the intensity of radiation absorbed by sample. 3 TYPES.  Barrier layer cell/Photovoltaic cell  Phototubes/ Photo emissive tube  Photomultiplier tube 39
  • 40. Barrier layer cell/Photovoltaic cell Consists of thin film metallic layer coated with silver or gold and acts as an electrode.  metal base plate which acts as another electrode.  These two layers are separated by a semiconductor layer of selenium. 40
  • 41.  Light radiation falls on selenium layer → electrons become mobile → taken up by transparent metal layer.  Creates a potential difference between two electrodes & causes the flow of current.  A flow of current observed in galvanometer which is proportional to the intensity and wavelength of light falling on it. 41
  • 42. 42
  • 43. PHOTOTUBES/ PHOTOEMISSIVE TUBES  Consists of a evacuated glass tube with a photocathode and a collector anode.  surface of photocathode is coated with a layer of elements like caesium, silver oxide or mixture of them.  radiant energy falls on photosensitive cathode, electrons are emitted which are attracted to anode causing current to flow.  More sensitive compared to barrier layer cell. 43
  • 44. 44
  • 45. PHOTOMULTIPLIER TUBE  Multiplication of photoelectrons by secondary emission of electrons.  In a vacuum tube, a primary photo-cathode is fixed which receives radiation from the sample.  Eight to ten dynodes are fixed each with increasing potential of 75-100V higher than preceding one.  last dynode is fixed an anode or electron collector electrode.  extremely sensitive to light and is best suited where weaker or low radiation is received. 45
  • 46. 46
  • 47. 47
  • 48. READOUT DEVICE  The data from a detector are displayed by a readout device, such as an analogue meter, a light beam reflected on a scale, or a digital display , or LCD .  The output can also be transmitted to a computer or printer. 48
  • 50. PHOTOMETER An instrument for measuring the of light or the relative intensity of a pair of lights. Also called an illuminometer. It utilizes filter to isolate a narrow wavelength region.  Single beam  Double beam 50
  • 51. SPECTOPHOTOMETER An instrument measures the ratio, or a function of the two, of the radiant power of two EM beams over a large wavelength region. It utilizes dispersing element (Prisms/Gratings) instead of filters, to scan large wavelength region. Single beam Double beam 51
  • 52. SINGLE BEAM VISIBLE PHOTOMETERS  A direct reading instrument consisting of  tungsten filament/LED as a source  lens to provide a parallel beam  a filter and a photodiode transducer  The current produced by the photodiode is processed with electronics or a computer to give a readout in absorbance.  Sample is inserted into the light path.  The absorbance is calculated as the logarithm of the ratio of these signals. 52
  • 53. 53
  • 54. Double beam visible photometer  Used to measure the absorbance of sample in a flowing stream.  The light beam is split by a 2 branched fibre optic.  One beam is passes – sample  Other beam- reference.  Filters are placed after the cells before the photodiode with nearly identical response.  Electrical outputs→ voltages and signals are processed by a log ratio amplifier or a computer to give a readout proportional to absorbance. 54
  • 55. 55
  • 56. UV absorption photometers  Serve as detectors in HPLC.  Mercury vapour lamp- source.  Wavelength range 254nm. 56
  • 57. Spectrophotometers Visible spectrophotometers  Wavelength range:380-800nm.  Simple and inexpensive.  Eg; Spectronic20  Tungsten filament – source  Diffraction by a simple reflection grating→ radiation passes→ sample/reference 57
  • 58. 58
  • 59. 59
  • 60. UV spectrophotometers Single beam computerised spectrophotometers  Light from the source is carried through lens and/or through aperture to pass through a suitable filter.  The type of filter to be used is governed by the colour of the solution.  The sample solution to be analysed is placed in cuvettes. 60
  • 61. 61
  • 62.  After passing through the solution, the light strikes the surface of detector (barrier-layer cell or phototube) and produces electrical current.  The output of current is measured by the deflection of needle of light-spot galvanometer or micro ammeter. 62
  • 63. 63
  • 64. Double beam spectrophotometers  two beams are formed in the space by a U shaped mirror called as beam splitter or beam chopper.  Chopper is a device consisting of a circular disc. One third of the disc is opaque and one third is transparent, remaining one third is mirrored. It splits the monochromatic beam of light into two beams of equal intensities. 64
  • 65. 65
  • 66.  Eg; Varian Cary100 66
  • 67. CHOICE OF SOLVENTS  It should not itself absorb radiations in the region under investigations.  It should be less polar so that it has minimum interaction with the solute molecule.  Most commonly: 95%ethanol  Cheap, good dissolving power, does not absorb radiation above 210nm.  Typical examples 67
  • 68. Solvent Wavelength(nm) Water 205 Methanol 210 Ethanol 210 Ether 210 Chloroform 245 Carbon tetra chloride 265 Cyclohexane 210 68
  • 69. SOLVENT EFFECT  The position and intensity of an absorption band may shift when the spectrum is recorded in different solvents.  A dilute sample solution is prefered for analysis.  Most commonly used solvent is 95% ethanol. It is best solvent as it is cheap,transparent down to 210nm.  Position as well as intensity of absorption maxima get shifted for a particular chromophore by changing the polarity of solvent.  By increasing polarity of solvent→dienes, conjugated hydrocarbons→ no shift 69
  • 70.  α,β unsaturated carbonyl compound show 2 different shifts. n →Π* transition.  Absorption band moves to shorter wavelength(blue shift) by increasing the polarity of the solvent.  Ground state is more polar as compared to the excited state. Eg: absorption maximum of acetone is at 279nm in hexane as compared to 264nm in water. 70
  • 71. Π→Π* transition.  Absorption band moves to longer wavelength(red shift) by increasing the polarity of the solvent.  The dipole-dipole interaction with the solvent molecules lower the energy of excited state more than that of the ground state. 71
  • 72. Effect of solvent polarity on the various types of bands. K-band  Due to conjugated enes & enones are affected differently by changing the polarity of the solvent.  K bands due to conjugated dienes are not affected by changing the polarity of the solvent, while these bands due to enones shows a red shift by increasing the  polarity of solvent. R band  The absorption shifts to shorter wavelength (blue shift) with increasing polarity of solvent. 72
  • 73. B band  The position as well as the intensity of the band is not shifted by increasing the polarity of the solvent.  But the heterocyclic aromatic compound, a marked hyperchromic shift is observed by increasing the polarity of the solvent. 73
  • 74. REFERENCE  Principles of instrumental analysis – Doglas A Skoog, James holler, Timothy a nieman, 5thedition.  Instrumental methods of analysis by Willards, 7thedition.  Instrumental methods of chemical analysis by H.Kaur  Instrumental methods of chemical analysis by B.K Sharma.  Instrumental methods of chemical analysis by Gurdeep R.Chatwal, Sham K.Anand. 74
  • 76. 76