1. Introduction to Nuclear
Magnetic Resonance
• Topics
– Nuclear spin and magnetism
– Resonance behavior and the Larmor
Frequency
• Larmor frequency
• flip angle
– Energy Absorption and Emission
• NMR spectroscopy
• Energy absorption in tissue (safety issues)
• Relaxometry
– T1,T2,T2* relaxation
2. Nuclear Magnetism
• Nucleons (protons, neutrons) have a quantum
property known as spin.
• Nucleons have been shown to obey Fermi
statistics, and thus have a maximum spin
magnitude of 1/2 Bohr magneton. (spin=1/2)
• In the absence of a magnetic field, nuclear spin
is not an observable
• In the presence of a homogeneous magnetic
field, the energy of the nucleus depends on the
relative orientation of the magnetic field and
the nuclear spin vector
4. Nuclear Magnetic Resonance:
Properties in Matter
• Relaxation
– After we have delivered energy to the nuclei
in our sample at the Larmor frequency,
there are two possible ways for the sample to
lose this energy (back to lowest energy state):
• spontaneous emission
• induced emission
5. • Spontaneous emission:
– negligible effect at RF frequencies (dominant
at visible frequencies)
• Induced emission
– Energy emission requires interaction of the
nucleus with its external environment
The nature of energy emission depends
strongly on the environment of the excited
nucleus (Relaxation)
3ω∝
6. • NMR Spectroscopy is the study of the
chemistry of matter using the NMR
absorption spectrum
• Relaxometry is the study of the chemistry of
matter using the NMR relaxation properties.
MRI generates tissue contrast based (mostly)
on NMR relaxation differences.
8. Alignment of Spins in a
Magnetic Field
spin
magnetic moment
B0 field
M
M=0
9. Energy in a Magnetic Field
(Zeeman Splitting, Spin ½)
E+1/2= −γB0/2 E-1/2= +γB0/2
P+1/2= 0. 5000049 P-1/2= 0.4999951
1.5T, T=310K, P(E)∝exp(−E/kT)
mI = +½ mI = −½
13. • The relative populations of the spin states
can be altered in a well defined way by the
application of a resonant B1 field in the xy-
plane.
• Any fluctuating magnetic field that has a
component in the xy-plane that oscillates at
the resonant frequency can induce
transitions between the spin states.
The T1 Relaxation Process
16. What effect does T1 have on
Images?
t = 0 t = 3s t = 6s t = 9s t = 12s
17. • Assume the steady state has been reached.
• Use a flip angle of θdegrees.
• Find a condition where the transverse
magnetization following the flip is
maximized.
The Ernst Angle
−=
1
expcos
T
TR
θ
22. Transverse Relaxation
• Longitudinal relaxation is driven by field
oscillations in the transverse plane.
• Transverse relaxation is driven by field
oscillations in the longitudinal plane.
• Random fluctuations in B0 experienced by a
nucleus cause the resonant frequency of that
spin to change.
23. Transverse Relaxation
• If the field experienced by the molecule is
purely random then the effect would time
average to zero.
• Correlations in the motion cause a range of
frequencies.
• In solids where there is no molecular
tumbling the range of resonances is very
broad.
28. What is T2
*?
• Spin-spin relaxation represents a loss of
coherence in the transverse magnetization
due to local effects on spin.
• Loss of the coherence of the transverse
magnetization also occurs as a result of bulk
magnetic effects
30. What effect does T2
* have on
Images?
• T2 and T2
* have the same effect on images.
• T2
* effects dominate when there is no spin
echo.
• From now on, we will assume that T2* is
more important, since in imaging it often is.
35. What effect does T2
* have on
Images?
× =
⊗ =
⇓ ⇓ ⇓FT FTFT
Perfect FID T2
*
Decay Actual FID
Perfect Image Point Spread Function Actual Image
36. What effect does T2
* have on
Images?
× =
⊗ =
⇓ ⇓ ⇓FT FTFT
Perfect FID T2
*
Decay Actual FID
Perfect Image Point Spread Function Actual Image
37.
38. Contrast in MRI
• Contrast based on Relaxation Time
• Contrast Agents
• BOLD Contrast
• Perfusion and Diffusion Contrast
39. To start with….
• No Spins, no contrast.
• Spin density is the most fundamental of
contrasts in MRI.
• In our case spin density is related to the
water content of the tissue.
40. Water Content of Tissue
Tissue % Water Content
Grey Matter 70.6
White Matter 84.3
Heart 80.0
Blood 93.0
Bone 12.2
41. Enhancing the Signal
from Spin Density
• Gastrointestinal imaging
– Drinking water
• Lung imaging
– Inhaling hyperpolarized inert gas. A laser
polarises the helium gas increasing its
magnetisation.
42. T1 Contrast
• The shorter the repetition time (TR) the
greater the T1 contrast.
• Prepare the magnetisation with a saturation
or inversion pulse followed by a suitable
delay.
43. T2 or T2
* Contrast
• Sequences without a spin echo will be T2
*-
weighted rather than T2-weighted.
• The longer the echo time (TE) the greater
the T2 contrast.
44. Magnetism
• units of magnetism
– 1T = 10,000 Gauss
– Earth’s field = 0.5 Gauss
• types of magnetism
– ferromagnetism
– paramagnetism
– superparamagnetism
– diamagnetism
45. Magnetic Susceptibility
• the degree to which a material
becomes magnetized when placed
in a magnetic field
• magnetizability
46. Ferromagnetism
• Property of certain metals
– iron
– nickel
– cobalt
• Positive susceptibility
• consist of magnet domains
• able to remain magnetized without
the influence of an external field
47. Paramagnetism
• Oxygen
• metal ions with unpaired electrons
resulting in positive susceptibility
– Fe
– Gd
– Mg
• less than 1/1000 th effect of
ferromagnetism
• shortens T1 and T2 times
48. Superparamagnetism
• halfway between ferro and
paramagnetic
• property of individual domains of
elements that have ferromagnetic
properties in bulk
• iron containing contrast agents
for bowel and lymph nodes are
superparamagnetic
49. Diamagnetism
• no intrinsic atomic magnetic moment
• In a magnetic field weakly repel the field
– negative magnetic susceptibility
• water, copper, nitrogen, barium sulfate,
and most tissues are diamagnetic.
• contributes to the loss of signal seen in
bowel on MRI after administration of
barium sulfate suspensions.
50. Contrast Agents
• Sometimes it is desirable to enhance the
natural contrast in the images.
– Highlighting a tumour.
– Measuring perfusion rates.
• MR contrast agents act to change the
relaxation times of the substrate they are in.
51. Contrast Agents
• Contain an ion with a large number of
unpaired electrons.
• This gives it a larger magnetic moment than
the proton.
• This enhances the local magnetic fields that
fluctuate, due to thermal motion, in the
vicinity of a proton.
52. Contrast Agents
• These are usually paramagnetic ions.
e.g. Gadolinium
• By themselves these ions are highly toxic so
they are bound up in large molecules.
e.g. DTPA
• They tend not to permeate the blood brain
barrier and so enhance tumours that lack
this barrier.
53. Gadolinium
• shortens T1 relaxation of water
– high SI on T1 weighted images in lower
concentrations
• shortens T2 relaxation of water
– lowers SI on T1 weighted images in high
concentrations
• seven unpaired electrons
– high magnetic moment
– unpaired electrons react with protons in
adjacent water molecules shortening their
relaxation time
58. Perfusion Imaging
• Perfusion is the rate constant determining
the delivery of metabolic substrates to the
local tissue and clearance of products of
metabolism.
• If a volume of tissue V is supplied with
arterial blood at a rate of F then perfusion
f = F / V (ml/100g/min)
59. Contrast Agent Methods
• Inject contrast agent.
• Causes a bolus of blood to have a different
signal intensity to the blood in the brain.
• This tagged bolus perfuses the brain.
• Level of signal change gives an indication
of rate of perfusion.
63. What is k-space?
• a mathematical concept
• not a real “space” in the patient nor in
the MR scanner
• key to understanding spatial encoding
of MR images
64. k-space and the MR Image
x
y
f(x,y)
kx
ky
KKK---spacespacespace
F(kx,ky)
ImageImageImage---spacespacespace
ℑ
65. k-space and the MR Image
• each individual point in the MR image
is reconstructed from every point in
the k-space representation of the
image
– like a card shuffling trick: you must have all of
the cards (k-space) to pick the single correct
card from the deck
• all points of k-space must be collected
for a faithful reconstruction of the
image
66. Discrete Fourier Transform
F(kx,ky) is the 2D discrete Fourier transform of the
image f(x,y)
x
y
f(x,y)
kx
ky
ℑ
K-space
F(kx,ky)
f x y
N
F k k e
xk yk
kk
x y
j
N
x j
N
yNN
yx
( , ) ( , )=
+
=
−
=
−
∑∑
1
2
2 2
0
1
0
1 π π
image-space
67. k-space and the MR Image
• If the image is a 256 x 256 matrix size,
then k-space is also 256 x 256 points.
• The individual points in k-space
represent spatial frequencies in the
image.
• Image contrast is represented by low
spatial frequencies; detail is
represented by high spatial frequencies.
69. Spatial Frequencies
• low frequency = contrast
• high frequency = detail
• The most abrupt change occurs at an
edge. Images of edges contain the
highest spatial frequencies.
70. Properties of k-space
• k-space is symmetrical
• all of the points in k-space must be known
to reconstruct the signal faithfully
• truncation of k-space results in loss of
detail, particularly at edges
• most important information centered
around the middle of k-space
• k-space is the Fourier representation of the
waveform
71. MRI and k-space
• The nuclei in an MR experiment
produce a radio signal (wave) that
depends on the strength of the main
magnet and the specific nucleus being
studied (usually H+).
• To reconstruct an MR image we need
to determine the k-space values from
the MR signal.
73. MRI
• Spatial encoding is accomplished by
superimposing gradient fields.
• There are three gradient fields in the
x, y, and z directions.
• Gradients alter the magnetic field
resulting in a change in resonance
frequency or a change in phase.
74. MRI
• For most clinical MR imagers using
superconducting main magnets, the main
magnetic field is oriented in the z direction.
• Gradient fields are located in the x, y, and
z directions.
75. MRI
• The three magnetic gradients work together
to encode the NMR signal with spatial
information.
• Remember: the resonance frequency
depends on the magnetic field strength.
Small alterations in the magnetic field by the
gradient coils will change the resonance
frequency.
76. Gradients
• Consider the example of MR imaging in the
transverse (axial) plane.
Z gradient: slice select
X gradient: frequency encode (readout)
Y gradient: phase encode
77. Slice Selection
• For axial imaging, slice selection occurs
along the long axis of the magnet.
• Superposition of the slice selection gradient
causes non-resonance of tissues that are
located above and below the plane of
interest.
80. Slice Selection
Selection of an axial
slice is accomplished
by the z gradient.
zz gradient directiongradient direction
ω = ω0
graph of the z magnetic gradient
z-axis
β β = β0
β > β0
β < β0
81. Slice Selection
slice location is determined by the null point of the z gradient
β
β0
ω
ω0
RF bandwidth
slice 1
ΖΖ
slice 2 slice 3
Ζ1 Ζ2 Ζ3
82. Frequency Encoding
• Within the imaging plane, a small gradient is
applied left to right to allow for spatial
encoding in the x direction.
• Tissues on the left will have a slightly higher
resonance frequency than tissues on the right.
• The superposition of an x gradient on the
patient is called frequency encoding.
• Frequency encoding enables spatial localization
in the L-R direction only.
84. Frequency Encoding
RF signalRF signal
fromfrom entireentire sliceslice
A/D conversion, 256 pointsA/D conversion, 256 points 1 line of
k-space
85. Phase Encoding
• An additional gradient is applied in the
y direction to encode the image in the
remaining direction.
• Because the x gradient alters the
frequencies in the received signal according
to spatial location, the y gradient must alter
the phase of the signal.
• Thus, the points of k-space are revealed by
recording the digitized RF signal after a
phase encoding gradient application.
86. Phase Encoding
• The technique of phase encoding the second
dimension in the imaging plane is
sometimes referred to as spin warping.
• The phase encoding gradient is “stepped”
during the acquisition of image data for a
single slice. Each step provides a unique
phase encoding.
• For a 256 x 256 square image matrix, 256
unique phase encodings must be performed
for each image slice. The second 256 points
in the x direction are obtained by A to D
conversion of the received signal.
91. • Acquisition of spatially encoded data as
described allows for reconstruction of the
MR image.
• The frequency and phase data are acquired
and form points in a 2D array .
• Reconstruction of the image is provided by
2D inverse Fourier transform of the
2D array.
• This method of spatially encoding the MR
image is called 2D FT imaging.
MR Image Reconstruction
94. Relaxation and Imaging
• FID (free induction decay) is
the relaxation behavior
following a single RF pulse
• most imaging done with
repetitive RF energy deposition
• the interval between the RF
energy pulses is called the TR
interval (time to repetition)
96. Equilibrium
• after 5 or so
repetitions, the
system reaches
equilibrium
• similar to water
flowing into a
leaky bucket
relaxation
RF in
equilibriumequilibrium
97. Differential Relaxation
• short TR
• lower absolute ML
• marked difference
in relative signal
• long TR
• higher absolute ML
• minimal difference
in relative signal
fat protons
water protons
98. Image Contrast and T1
Relaxation
• shorter TRs maximize
differences in T1 relaxation,
generating tissue contrast
• longer TRs minimize
differences in T1 relaxation,
reducing T1 tissue contrast
99. Goals of Imaging
Sequences
• generate an RF signal
perpendicular to β0
• generate tissue contrast
• minimize artifacts
100. Measuring the MR Signal
z
y x
RF signal from
precessing protons
RF signal from
precessing protons
RF antennaRF antenna
β0
107. T2* decay
• occurs between the dephasing and
the rephasing gradients
• rephasing incompletely recovers
the signal
• signal loss is greater with longer
TEs
• decay generates image contrast
108. T2* decay
• T2* decay is always faster than
T2 decay
• gradient echo imaging cannot
recover signal losses from
–magnetic field inhomogeneity
–magnetic susceptibility
–water-fat incoherence
109. T2 and T2* Relaxation
• T2* relaxation influences
contrast in gradient echo
imaging
• T2 relaxation influences
contrast in spin echo imaging
111. Gradient Echo
advantages
• faster imaging
–can use shorter TR and shorter
TEs than SE
• low flip angle deposits less
energy
–more slices per TR than SE
–decreases SAR
• compatible with 3D acquisitions
112. Gradient Echo
disadvantages
• difficult to generate good T2
weighting
• magnetic field inhomogeneities
cause signal loss
–worse with increasing TE times
–susceptibility effects
–dephasing of water and fat
protons
113. Gradient Echo
changing TE
TE 9
FA 30
TE 9
FA 30
TE 30
FA 30
TE 30
FA 30
susceptibility effectsusceptibility effect T2* weightingT2* weighting
115. Gradient Echo
• image contrast depends on
sequence
• conventional GR scan
– aka GRASS, FAST
– decreased FA causes less T1
weighting
– increased TE causes more T2*
weighting
124. 900 Flip
z
y x
z
y x
900 RF
t=t0 t=t0+
900
After
ML=0
MXY=M
Before
ML=M
MXY=0
125. Dephasing in the xy-plane
view from the top
y
x
z Mxy
y
x
z
Mxy≈0
phase coherency phase dispersion
Dephasing begins
immediately after
the 900 RF pulse.
t=0 t=TE/2
900 RF
126. y
x
z Mxy
phase coherency
minus t2 decay
Rephasing in the xy-plane
view from the top
y
x
z
Mxy≈0
phase dispersion
t=TE/2 t=TE
1800 RF
127. z
y x
z
y x
z
y x
z
y x
t=TE/2 t=TE
1800 RF
t=0
900 RF
dephased
rephased
1800 Flip
133. Effects of the 1800 Pulse
• eliminates signal loss due to
field inhomogeneities
• eliminates signal loss due to
susceptibility effects
• eliminates signal loss due to
water/fat dephasing
• all signal decay is caused by
T2 relaxation only
134. Spin Echo
advantages
• high signal to noise
• least artifact prone sequence
• contrast mechanisms easier to
understand
135. Spin Echo
disadvantages
• high SAR than gradient echo
because of 900 and 1800 RF
pulses
• long TR times are incompatible
with 3D acquisitions
136. Spin Echo Contrast
• T1 weighted
–short TR (450-850)
–short TE (10-30)
• T2 weighted
–long TR (2000 +)
–long TE (> 60)
• PD weighted
–long TR, short TE
137. Spin Echo Contrast
T2 Relaxation
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
0 100 200 300 400 500
msec
Mxy
long T2
short T2
T1 Relaxation
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
0 1000 2000 3000 4000 5000
msec
ML
long T1
short T1
T1 weighted - T1 relaxation predominates
•Short TE minimizes differences in T2 relaxation
•Short TR maximizes differences in T1 relaxation
T2 weighted - T2 relaxation predominates
•Long TE maximizes differences in T2 relaxation
•Long TR minimizes differences in T1 relaxation
140. Contrast
• the ability to discriminate different
tissues based on their relative
brightness
141.
142.
143.
144.
145.
146.
147. Caveat
• windowing affects the relative
contrast of tissues
– intensity values of pixels are relative
to one another, unlike CT
• windowing can make a solid tumor
look like a “cyst”
151. Summary
• need visible differences in intensity
to discriminate tissues
• surrounding tissues can make an
intermediate signal tissue appear
dark or bright
• windowing affects image and tissue
contrast
152. Noise
• constant at a given machine setup
• reduces the ability to visualize low
contrast structures
• adds to or subtracts from the
average signal intensity of a given
pixel
153. Noise
• increasing the available signal will
reduce the relative effects of noise
• machine parameters must be
chosen to maximize signal without
significantly extending exam times
• S/N is a relative measure allowing
for comparison in a variety of
circumstances
169. Stronger Main Magnet
S/N effect Downside
• linear increase • less T1 weighting
at high fields
• increased
chemical shift
effects in RO
direction
170. Coil Selection
S/N effect Downside
• increase in signal
with surface coils
• quadrature
provides 40%
increase S/N over
linear
• phased array
increased over
quadrature
• limited coverage
with surface coils
• more complex
coils are more
expensive
171. Larger Voxel Size
S/N effect Downside
• linear increase in
either RO or PE
direction
• linear increase
with increased
slice thickness
• decreased
resolution
172. Decreased Phase Encodings
S/N effect Downside
• square root
increase in signal
to noise
• linear decrease in
scan time
• decreased
resolution in PE
direction
• Gibb’s
phenomenon in
PE direction
173. Increased Signal Averages
S/N effect Downside
• square root
increase in signal
to noise
• linear increase in
scan time
174. Decreased Receiver BW
S/N effect Downside
• square root
increase in signal
to noise
• increase in
chemical shift
artifact in RO
direction
175. Pulse Sequence Parameters
• SE imaging
– increased TR provides nonlinear
increase in SNR with linear increase
in scan time
– decreased TE provides nonlinear
increase in SNR with no effect on
scan time and less T2 weighting
176. Pulse Sequence Parameters
• GE imaging
– complex effects
– maximum SNR typically between 30
and 60 degrees
– long TR sequences (2D)
• increase SNR with increased flip angle
– short TR sequences (TOF & 3D)
• decreased SNR with increased flip angle
177. SNR Application
• pituitary imaging
– baseline:
• 16 cm FOV, 3 mm slice thickness, 192
phase encodes, 4 NEX
– new goal:
• reduced scan time, same SNR
179. Fat Suppression and SNR
• non fat-suppressed image
– each image pixel comprised of signal
from water and fat in the imaging
voxel
• fat-suppression
– reduces total signal by suppression of
fat from the voxel
– reduces SNR