Another Device to Commission?
Begin
Commissioning AlignRT® with
Minimal Disruption
DISCLOSURES
2
I have nothing to disclose
INTRODUCTION
3
• Technology and complexity in radiation oncology continues to grow at a fast pace
• The need for accurate implementation of SRS/SBRT QA programs is crucial
• Increased pressure to keep cost down and be more efficient
How can physicists provide increased value to clinics with increasing demands,
but maintain a high level of safety?
INTRODUCTION
4
• Combine tests to accomplish multiple QA procedures at one time
• Focus first on mission critical processes (think TG-100 and FMEA analysis)
• Use AAPM Task Groups as guidance when developing your program
Now, what about a technology that I don’t know? Like surface imaging?
COMMISSIONING SURFACE IMAGING
5
The approach is the same as every other modality in
radiation therapy
• Start out with the Task Group (AAPM TG-147)1 for
recommendations
• Discuss with your clinicians the goals of the system
(motion management for SRS, positioning for breast
patients, etc)
• Prioritize the QA in terms of the clinical need
– Do you need to perform all QA possible?
1 Willoughby, T., et al. “Quality assurance for nonradiographic radiotherapy localization and positioning systems: Report of Task Group 147.” 8 March 2012
6
THE TECHNOLOGY
• Stereoscopic camera system that is used for positioning and tracking
• Isocenter is determined from a novel calibration procedure2
• Triangulation provides a 3D reference frame using a speckle pattern on the
patient’s surface
ALIGNRT® INTRODUCTION
2 http://www.visionrt.com/content/core-technology
7
COMMON APPLICATION
• Intrafraction motion-management for SRS, SBRT, and respiratory cases
• Positioning for sites where the surface is a good surrogate for the target
– Breast, intracranial, extremities, etc
• General use for all patients
– Gross movements
• The beam can be held (automatically on most systems) when the registered
surfaces are out of the specified tolerance
ALIGNRT® INTRODUCTION
8
Start with the clinical use cases to establish QA procedures in conjunction with the
AAPM TG-147
One size may not fit all, and a good QA program will reflect the clinical need
• What site will be treated most often?
• Will it be used for motion-management, positioning, or both?
• Will the beam be held during treatment delivery (dynamic gating)?
• What tolerances will be needed during position and treatment delivery?
CLINICAL USE CASES
9
Start by mapping out the workflow and establish an End-to-end test, which
incorporates the complete clinical use case
COMMISSIONING TESTS
10
Try to create a test that mimics the clinical use case (E2E)
END-TO-END (E2E) TEST
11
EQUIPMENT NEEDED
• Phantom (preferably anthropomorphic) that has a neutral surface for imaging
– Shouldn’t be a ball or cylinder, due to tracking difficulties
– MAX-HD® (Integrated Medical Technologies) a good choice
– STEEV SRS Phantom (CIRS)
– SRS Head Phantom from IROC (clear, so may need taped to image well)
• Detectors such as Gafchromic® film, microchambers/diodes, OSL/TLD
• Electrometer
• 3DOF Head Adjuster (provided by VisionRT) or 6DOF couch for accurate
positioning
END-TO-END (E2E) TEST
12
END-TO-END (E2E) TEST
SAMPLE PHANTOMS
MAX-HD STEEV SRS (IROC) SRS Phantom
IMT CIRS IROC CIRS
13
SOME NOTES ABOUT EQUIPMENT
• Detectors should be small enough to avoid partial volume effects, especially for
SRS
• Film scanning should use the right calibration procedure
• Make sure film is cut in such a way to ensure reproducibility and localization
– Laser cut films for the phantom are available
• Epson Perfection XL11000 (or the older XL10000) are good choices for film
scanning
END-TO-END (E2E) TEST
14
CT SCANNING
• Use the highest resolution protocol possible (small slice thickness ~1mm, high
mAs, etc)
• Try and use the same immobilization techniques that would be used during
treatment
• Set your DICOM origin (CT origin) at a reproducible localization to minimize setup
errors
• If the phantom supports it, try multiple CT scans with different inserts and cubes
– Again, it’s important to know the origin with certainty for film scanning
END-TO-END (E2E) TEST
15
CONTOURING
• Many treatment planning systems automatically create the BODY contour, so
review for accuracy
• The Target Volume can be the sensitive volume of the detector or the center of
the film
END-TO-END (E2E) TEST
16
TREATMENT PLANNING
• Use a beam configuration that will test the system
– Couch kicks
– VMAT, if applicable, to block cameras
– Dose levels that mimic treatment time (don’t scale MU)
– Use a real plan if available
• Save for a baseline for routine QA
END-TO-END (E2E) TEST
17
LOCALIZATION
• Use the AlignRT system to be the primary localization method
• Verify the positioning with CBCT and lasers
– There may be a small deviation with the DICOM because of the BODY
contour
– If so, position initially with CBCT, capture the reference, then reposition
with the AlignRT system
• Use a 6DOF localization technique to position the phantom
END-TO-END (E2E) TEST
18
DELIVERY
• Make note of gantry occlusion and deltas
– Some variation is acceptable during blockage, but extreme shifts may
indicate improper lighting, ROI, or calibration conditions
• Watch the deltas for drift over the course of the treatment at the couch base
• When rotating the couch, use the AlignRT readout, not the digital indicators, to
reposition
– Stereoscopic systems trump the internal readouts of the LINAC system,
hence the “STEREO” in stereotactic
– This is especially important in older systems, where couch tolerances are
looser
END-TO-END (E2E) TEST
19
ANALYSIS
• Temperature/pressure correct the chamber reading, and use a cross calibration
factor to calculate the ratio of charge/dose (traced by to your ASCL calibrated
detector)
• FilmQA® from Ashland is a good choice for film scanning
– FilmQA analyzes in all 3 RGB channels
– ImageJ is also available for free, but takes some effort to write code
• Always scan the film in the same orientation, and if the calibration films are not
done on the same day, wait a while for the film to develop
END-TO-END (E2E) TEST
20
WHAT AM I LOOKING FOR?
Depends on the clinical goals
• Breast patients could be passing 95% with a gamma index at 3%, 3mm with the
film analysis
• SRS programs treating trigeminal neuralgia cases may need 2%, 1mm
• Output measurements should be within 2% of the expected from the TPS, but
can be tough for small cones because of the output factors
If you pass the E2E, the commissioning process got a whole lot better. But…
END-TO-END (E2E) TEST
21
WHAT IF IT FAILS?
Start looking at each component of the system closer. You will need it for the
overall QA program anyways. Keep in mind:
1. TG-147 is just part of this. Since an E2E test checks the whole system, it could
be anything outlined in TG-40, TG-142, TG-66, etc
2. Focus on the components that are specific to TG-147, as to not get
overwhelmed
END-TO-END (E2E) TEST
22
• Do a “chart check” style review to make sure items transferred properly
– Patient ID
– Isocenter
– Contour (especially BODY)
– Scan orientation
• The coordinate system of the LINAC matches the TPS and DICOM transfer
– For example, the couch coordinates could be opposite if the coordinate
systems don’t match
COMMICATION BETWEEN SYSTEMS
23
• Do a “chart check” style review to make sure items transferred properly
– Patient ID
– Isocenter
– Contour (especially BODY)
– Scan orientation
• The coordinate system of the LINAC matches the TPS and DICOM transfer
– For example, the couch coordinates could be opposite if the coordinate
systems don’t match
• If using an interface for shifts, ensure that is correct by using lasers or another
surrogate to confirm AlignRT shifts performed by the LINAC console
COMMICATION BETWEEN SYSTEMS
24
• An accurate ROI gives stable real-time deltas for any phantom when there are no
occlusions. If the delta jump at couch base, and no blockage
– Ensure the ROI give a unique view
• I.e., not flat, symmetric, nor broken
– The underlying image is intact
• Check lighting condition or skin tone
• It is equally important to not make the ROI so big it doesn’t detect small
movements
– Like contouring the entire head for SRS – just use the small area in the
open mask
REGION OF INTEREST (ROI)
25
• In order to make sure the calibration (both the calibration plate and fine
isocenter) is correct, perform a series of shifts to known positions, and record
the deltas
– Shift 2mm, 1cm, 2cm, 5cm, 10cm in S/I, L/R, and A/P directions
– Kick the couch in increments of 45 degree, for example
– Move the phantom arbitrarily, and see if the 6DOF positioning can get
back to a zero baseline
• If greater than 1mm (for SRS) disagreement is seen, perform isocenter and
monthly calibrations again, and repeat
• Refer to the acceptance testing procedures for vendor baselines as well
SHIFTS
26
• With the camera system on for a significant period of time (like an hour or so),
turn the monitoring on and watch the deltas
• Record the values, and if large deviations are seen, repeat after a longer warm
up is achieved
• Compared to infrared based systems, the drift should be quite minimal
SHIFTS
27
• With the camera system on for a significant period of time (like an hour or so),
turn the monitoring on and watch the deltas
• Record the values, and if large deviations are seen, repeat after a longer warm
up is achieved
• Compared to infrared based systems, the drift should be quite minimal
SHIFTS
28
• With a simple motion phantom, try and move the phantom enough to hold the
beam
• Tighten tolerance to only beam on when the chamber is in position
• Use a simple test plan to isolate it just the motion of the phantom
• Measure with an ion chamber and record the results
– Should be very close to baseline, stationary reading, assuming the
phantom is not shifting during motion, and the gating is working
• This would also test spatial accuracy at the same time
DYNAMIC GATING
29
• As with all medical devices, it is up to the medical physicists to stay current with
the latest customer service bulletins and release notes
• The tests outlined are from a clinical medical physicist’s perspective, but are by
no means meant to be prescriptive
• Training and continuing education are equally, and maybe more, important than
a one-time commissioning – as is the continual QA program
VENDOR RECOMMENDATIONS
30
• VisionRT has already provided a nice workflow for QA
– DailyQA
• Uses the calibration plate, aligned by the therapists, and analyzes it
for constancy
– MonthlyQA
• Again, using the plate, but overrides the triangulated 3D position of
the system. To be done if the system appears to have changed
– Isocenter Calibration
• Using a phantom with fiducials that can be seen on CT/MV, with the
surface imaged, allows the isocenter of the imaging system to be
correlated to the MV isocenter
ESTABLISHING A QA PROGRAM
31
• In addition to the vendor provided phantoms and software, use the results from
the commissioning tests to create baselines for routine QA
ESTABLISHING A QA PROGRAM
32
"If it's not written down, it didn't happen“
• Ensure all the tests you do during commissioning, QA, or anything concerning
the system are well documented
• Not only is it good to establish baselines and trends, it is very useful in helping
other centers when they go live with AlignRT
DOCUMENTATION
33
• Commissioning any system doesn’t have to be difficult, or even that time-
consuming, any long as the problem is well-defined
• By using clinical use cases as guidance, tests will solidify, and the project can be
focused
• By incorporated E2E tests, and grouping other QA procedures in an effective way,
commissioning and routine QA time can be reduced
CONCLUSIONS
34
THANK YOU!
www.alyzenmed.com
1801 South 54th Street
Paragould, AR 72450
870.926.0894
jrogers@alyzenmed.com
Jonathan Rogers, MS, DABR

More Related Content

What's hot

SBRT: The Importance of Continuous Motion Management​
SBRT: The Importance of Continuous Motion Management​SBRT: The Importance of Continuous Motion Management​
SBRT: The Importance of Continuous Motion Management​SGRT Community
 
Total Body Irradiation (TBI) Planning
Total Body Irradiation (TBI) PlanningTotal Body Irradiation (TBI) Planning
Total Body Irradiation (TBI) PlanningSubhash Thakur
 
Image guided adaptive radiotherapy
Image guided adaptive radiotherapyImage guided adaptive radiotherapy
Image guided adaptive radiotherapyapollo seminar group
 
Icru 29,50 &62
Icru 29,50 &62Icru 29,50 &62
Icru 29,50 &62Dhiman Das
 
Managing Motion from End to End
Managing Motion from End to EndManaging Motion from End to End
Managing Motion from End to EndSGRT Community
 
Icru – 83 dr. upasna
Icru – 83  dr. upasnaIcru – 83  dr. upasna
Icru – 83 dr. upasnaUpasna Saxena
 
Respiration motion management
Respiration motion managementRespiration motion management
Respiration motion managementKiran Ramakrishna
 
Assessment of Intrafraction Motion during real-time tracking in the treatment...
Assessment of Intrafraction Motion during real-time tracking in the treatment...Assessment of Intrafraction Motion during real-time tracking in the treatment...
Assessment of Intrafraction Motion during real-time tracking in the treatment...Subrata Roy
 
Immobilisation in Radiotherapy
Immobilisation in RadiotherapyImmobilisation in Radiotherapy
Immobilisation in RadiotherapyYamini Baviskar
 
Volumetric Modulated Arc Therapy
Volumetric Modulated Arc TherapyVolumetric Modulated Arc Therapy
Volumetric Modulated Arc Therapyfondas vakalis
 
Intensity modulated radiation therapy and Image guided radiation therapy
Intensity modulated radiation therapy and Image guided radiation therapy Intensity modulated radiation therapy and Image guided radiation therapy
Intensity modulated radiation therapy and Image guided radiation therapy Ravindra Shende
 
ABC system, Free Breath 4DCT & Symmetry Radiotherapy
ABC system, Free Breath 4DCT & Symmetry RadiotherapyABC system, Free Breath 4DCT & Symmetry Radiotherapy
ABC system, Free Breath 4DCT & Symmetry RadiotherapySanjeet Mandal
 
Stereotactic Body Radiation Therapy
Stereotactic Body Radiation TherapyStereotactic Body Radiation Therapy
Stereotactic Body Radiation Therapyfondas vakalis
 
Icru 50,62,83 volume deliniation
Icru 50,62,83 volume deliniationIcru 50,62,83 volume deliniation
Icru 50,62,83 volume deliniationalthaf jouhar
 
Contouring guidelines Cervix IMRT
Contouring guidelines Cervix IMRTContouring guidelines Cervix IMRT
Contouring guidelines Cervix IMRTDebarshi Lahiri
 

What's hot (20)

SBRT: The Importance of Continuous Motion Management​
SBRT: The Importance of Continuous Motion Management​SBRT: The Importance of Continuous Motion Management​
SBRT: The Importance of Continuous Motion Management​
 
Total Body Irradiation (TBI) Planning
Total Body Irradiation (TBI) PlanningTotal Body Irradiation (TBI) Planning
Total Body Irradiation (TBI) Planning
 
Image guided adaptive radiotherapy
Image guided adaptive radiotherapyImage guided adaptive radiotherapy
Image guided adaptive radiotherapy
 
Icru 29,50 &62
Icru 29,50 &62Icru 29,50 &62
Icru 29,50 &62
 
Modern Radiotherapy
Modern RadiotherapyModern Radiotherapy
Modern Radiotherapy
 
Motion Management in Radiation Therapy
Motion Management in Radiation TherapyMotion Management in Radiation Therapy
Motion Management in Radiation Therapy
 
Managing Motion from End to End
Managing Motion from End to EndManaging Motion from End to End
Managing Motion from End to End
 
Icru – 83 dr. upasna
Icru – 83  dr. upasnaIcru – 83  dr. upasna
Icru – 83 dr. upasna
 
Respiration motion management
Respiration motion managementRespiration motion management
Respiration motion management
 
Assessment of Intrafraction Motion during real-time tracking in the treatment...
Assessment of Intrafraction Motion during real-time tracking in the treatment...Assessment of Intrafraction Motion during real-time tracking in the treatment...
Assessment of Intrafraction Motion during real-time tracking in the treatment...
 
Beam Direction
Beam DirectionBeam Direction
Beam Direction
 
Immobilisation in Radiotherapy
Immobilisation in RadiotherapyImmobilisation in Radiotherapy
Immobilisation in Radiotherapy
 
TSET
TSETTSET
TSET
 
Volumetric Modulated Arc Therapy
Volumetric Modulated Arc TherapyVolumetric Modulated Arc Therapy
Volumetric Modulated Arc Therapy
 
Intensity modulated radiation therapy and Image guided radiation therapy
Intensity modulated radiation therapy and Image guided radiation therapy Intensity modulated radiation therapy and Image guided radiation therapy
Intensity modulated radiation therapy and Image guided radiation therapy
 
ABC system, Free Breath 4DCT & Symmetry Radiotherapy
ABC system, Free Breath 4DCT & Symmetry RadiotherapyABC system, Free Breath 4DCT & Symmetry Radiotherapy
ABC system, Free Breath 4DCT & Symmetry Radiotherapy
 
Stereotactic Body Radiation Therapy
Stereotactic Body Radiation TherapyStereotactic Body Radiation Therapy
Stereotactic Body Radiation Therapy
 
Icru 50,62,83 volume deliniation
Icru 50,62,83 volume deliniationIcru 50,62,83 volume deliniation
Icru 50,62,83 volume deliniation
 
4 D Igrt
4 D Igrt4 D Igrt
4 D Igrt
 
Contouring guidelines Cervix IMRT
Contouring guidelines Cervix IMRTContouring guidelines Cervix IMRT
Contouring guidelines Cervix IMRT
 

Viewers also liked

Prospective Assessment of Deep Inspiration Breath Hold to Prevent Radiation-I...
Prospective Assessment of Deep Inspiration Breath Hold to Prevent Radiation-I...Prospective Assessment of Deep Inspiration Breath Hold to Prevent Radiation-I...
Prospective Assessment of Deep Inspiration Breath Hold to Prevent Radiation-I...SGRT Community
 
Image guided radiation therapy (IGRT)
Image guided radiation therapy (IGRT)Image guided radiation therapy (IGRT)
Image guided radiation therapy (IGRT)Anees Muhammed
 
Qualifying exam presentation 05 20_audio
Qualifying exam presentation 05 20_audioQualifying exam presentation 05 20_audio
Qualifying exam presentation 05 20_audioebassell
 
Preliminary Examination Proposal Slides
Preliminary Examination Proposal SlidesPreliminary Examination Proposal Slides
Preliminary Examination Proposal SlidesManas Tungare
 
TPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation OncologyTPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation OncologyMiami Cancer Institute
 
Fusion of Skeletal and Silhouette-based Features for Human Action Recognition...
Fusion of Skeletal and Silhouette-based Features for Human Action Recognition...Fusion of Skeletal and Silhouette-based Features for Human Action Recognition...
Fusion of Skeletal and Silhouette-based Features for Human Action Recognition...Francisco (Paco) Florez-Revuelta
 
IGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing ProcedureIGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing ProcedureMiami Cancer Institute
 
CVPR2016 Fitting Surface Models to Data 抜粋
CVPR2016 Fitting Surface Models to Data 抜粋CVPR2016 Fitting Surface Models to Data 抜粋
CVPR2016 Fitting Surface Models to Data 抜粋sumisumith
 
Igrt And Resp Gating Final Version
Igrt And Resp Gating Final VersionIgrt And Resp Gating Final Version
Igrt And Resp Gating Final VersionSpectrum Health
 
Jason Morton - SOP Linac Commissioning
Jason Morton - SOP Linac CommissioningJason Morton - SOP Linac Commissioning
Jason Morton - SOP Linac CommissioningGenesisCareUK
 
PhD Research Proposal - Qualifying Exam
PhD Research Proposal - Qualifying ExamPhD Research Proposal - Qualifying Exam
PhD Research Proposal - Qualifying ExamParamita Mirza
 
Graphic design pdf
Graphic design pdfGraphic design pdf
Graphic design pdfJack Lineker
 
Silhouette Rimko
Silhouette RimkoSilhouette Rimko
Silhouette RimkoRimkoc
 
Preliminary Exam Presentation
Preliminary Exam PresentationPreliminary Exam Presentation
Preliminary Exam Presentationggarner
 
Ph.D. Qualifying Exam Presentation (McGill University, Department of Biology))
Ph.D. Qualifying Exam Presentation (McGill University, Department of Biology))Ph.D. Qualifying Exam Presentation (McGill University, Department of Biology))
Ph.D. Qualifying Exam Presentation (McGill University, Department of Biology))nouji87
 
Management dell’organ motion nei trattamenti stereo-RT e radiochirurgici:
Management dell’organ motion nei trattamenti stereo-RT e radiochirurgici: Management dell’organ motion nei trattamenti stereo-RT e radiochirurgici:
Management dell’organ motion nei trattamenti stereo-RT e radiochirurgici: Gemelli Advanced Radiation Therapy
 
Motion capture in the media
Motion capture in the mediaMotion capture in the media
Motion capture in the medialorenmazzon
 
Liu Ren at AI Frontiers: Sensor-aware Augmented Reality
Liu Ren at AI Frontiers: Sensor-aware Augmented RealityLiu Ren at AI Frontiers: Sensor-aware Augmented Reality
Liu Ren at AI Frontiers: Sensor-aware Augmented RealityAI Frontiers
 

Viewers also liked (20)

Prospective Assessment of Deep Inspiration Breath Hold to Prevent Radiation-I...
Prospective Assessment of Deep Inspiration Breath Hold to Prevent Radiation-I...Prospective Assessment of Deep Inspiration Breath Hold to Prevent Radiation-I...
Prospective Assessment of Deep Inspiration Breath Hold to Prevent Radiation-I...
 
PhD Qualifying Exam Slides
PhD Qualifying Exam SlidesPhD Qualifying Exam Slides
PhD Qualifying Exam Slides
 
Image guided radiation therapy (IGRT)
Image guided radiation therapy (IGRT)Image guided radiation therapy (IGRT)
Image guided radiation therapy (IGRT)
 
Qualifying exam presentation 05 20_audio
Qualifying exam presentation 05 20_audioQualifying exam presentation 05 20_audio
Qualifying exam presentation 05 20_audio
 
Preliminary Examination Proposal Slides
Preliminary Examination Proposal SlidesPreliminary Examination Proposal Slides
Preliminary Examination Proposal Slides
 
TPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation OncologyTPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation Oncology
 
Fusion of Skeletal and Silhouette-based Features for Human Action Recognition...
Fusion of Skeletal and Silhouette-based Features for Human Action Recognition...Fusion of Skeletal and Silhouette-based Features for Human Action Recognition...
Fusion of Skeletal and Silhouette-based Features for Human Action Recognition...
 
IGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing ProcedureIGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing Procedure
 
CVPR2016 Fitting Surface Models to Data 抜粋
CVPR2016 Fitting Surface Models to Data 抜粋CVPR2016 Fitting Surface Models to Data 抜粋
CVPR2016 Fitting Surface Models to Data 抜粋
 
Igrt And Resp Gating Final Version
Igrt And Resp Gating Final VersionIgrt And Resp Gating Final Version
Igrt And Resp Gating Final Version
 
Jason Morton - SOP Linac Commissioning
Jason Morton - SOP Linac CommissioningJason Morton - SOP Linac Commissioning
Jason Morton - SOP Linac Commissioning
 
PhD Research Proposal - Qualifying Exam
PhD Research Proposal - Qualifying ExamPhD Research Proposal - Qualifying Exam
PhD Research Proposal - Qualifying Exam
 
Graphic design pdf
Graphic design pdfGraphic design pdf
Graphic design pdf
 
Silhouette Rimko
Silhouette RimkoSilhouette Rimko
Silhouette Rimko
 
Preliminary Exam Presentation
Preliminary Exam PresentationPreliminary Exam Presentation
Preliminary Exam Presentation
 
Ph.D. Qualifying Exam Presentation (McGill University, Department of Biology))
Ph.D. Qualifying Exam Presentation (McGill University, Department of Biology))Ph.D. Qualifying Exam Presentation (McGill University, Department of Biology))
Ph.D. Qualifying Exam Presentation (McGill University, Department of Biology))
 
Management dell’organ motion nei trattamenti stereo-RT e radiochirurgici:
Management dell’organ motion nei trattamenti stereo-RT e radiochirurgici: Management dell’organ motion nei trattamenti stereo-RT e radiochirurgici:
Management dell’organ motion nei trattamenti stereo-RT e radiochirurgici:
 
Motion capture in the media
Motion capture in the mediaMotion capture in the media
Motion capture in the media
 
Motion picture - for art app class sy 2011-2012
Motion picture - for art app class sy 2011-2012Motion picture - for art app class sy 2011-2012
Motion picture - for art app class sy 2011-2012
 
Liu Ren at AI Frontiers: Sensor-aware Augmented Reality
Liu Ren at AI Frontiers: Sensor-aware Augmented RealityLiu Ren at AI Frontiers: Sensor-aware Augmented Reality
Liu Ren at AI Frontiers: Sensor-aware Augmented Reality
 

Similar to Commissioning AlignRT with Minimal Disruption

CT Dose Notifications and Alerts AAPM 2014
CT Dose Notifications and Alerts AAPM 2014CT Dose Notifications and Alerts AAPM 2014
CT Dose Notifications and Alerts AAPM 2014KeLu25
 
Texture analyzer and its accessories
Texture analyzer and its accessoriesTexture analyzer and its accessories
Texture analyzer and its accessoriesRuchikaZalpouri
 
ARIMA Model for analysis of time series data.ppt
ARIMA Model for analysis of time series data.pptARIMA Model for analysis of time series data.ppt
ARIMA Model for analysis of time series data.pptREFOTDEBuea
 
Internal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labInternal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labAnkit Raiyani
 
Best-Practices to Achieve Quality PV Loop Data
Best-Practices to Achieve Quality PV Loop DataBest-Practices to Achieve Quality PV Loop Data
Best-Practices to Achieve Quality PV Loop DataInsideScientific
 
Gr&r studies
Gr&r studiesGr&r studies
Gr&r studiesshilpi020
 
LINAC COMMSSN.ppt
LINAC COMMSSN.pptLINAC COMMSSN.ppt
LINAC COMMSSN.pptBBAdhikari
 
qualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationqualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationSohailPattan
 
Quality of Analytical Procedures
Quality of Analytical ProceduresQuality of Analytical Procedures
Quality of Analytical ProceduresBrendon Naicker
 
Nox t3 complete review final
Nox t3 complete review finalNox t3 complete review final
Nox t3 complete review finalRandy Clare
 
Motion Management in Lung Cancer Radiotherapy
Motion Management in Lung Cancer RadiotherapyMotion Management in Lung Cancer Radiotherapy
Motion Management in Lung Cancer RadiotherapyJyotirup Goswami
 
02trainingmaterialformsa (1) 111030062223-phpapp02
02trainingmaterialformsa (1) 111030062223-phpapp0202trainingmaterialformsa (1) 111030062223-phpapp02
02trainingmaterialformsa (1) 111030062223-phpapp02Junelly Grace Catalan-Tecson
 
Intensity-modulated Radiotherapy
Intensity-modulated RadiotherapyIntensity-modulated Radiotherapy
Intensity-modulated RadiotherapyDr Vijay Raturi
 
Physical and clinical aspects of brachytherapy
Physical and clinical aspects of  brachytherapyPhysical and clinical aspects of  brachytherapy
Physical and clinical aspects of brachytherapyRam Pukar Bharat
 
Good Weighing Practices in QC Lab
Good Weighing Practices in QC LabGood Weighing Practices in QC Lab
Good Weighing Practices in QC LabDr. Amsavel A
 

Similar to Commissioning AlignRT with Minimal Disruption (20)

CT Dose Notifications and Alerts AAPM 2014
CT Dose Notifications and Alerts AAPM 2014CT Dose Notifications and Alerts AAPM 2014
CT Dose Notifications and Alerts AAPM 2014
 
Texture analyzer and its accessories
Texture analyzer and its accessoriesTexture analyzer and its accessories
Texture analyzer and its accessories
 
ARIMA Model.ppt
ARIMA Model.pptARIMA Model.ppt
ARIMA Model.ppt
 
ARIMA Model for analysis of time series data.ppt
ARIMA Model for analysis of time series data.pptARIMA Model for analysis of time series data.ppt
ARIMA Model for analysis of time series data.ppt
 
ARIMA Model.ppt
ARIMA Model.pptARIMA Model.ppt
ARIMA Model.ppt
 
14184864.ppt
14184864.ppt14184864.ppt
14184864.ppt
 
X ray imaging intensifier
X ray imaging intensifierX ray imaging intensifier
X ray imaging intensifier
 
Internal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labInternal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation lab
 
Best-Practices to Achieve Quality PV Loop Data
Best-Practices to Achieve Quality PV Loop DataBest-Practices to Achieve Quality PV Loop Data
Best-Practices to Achieve Quality PV Loop Data
 
Gr&r studies
Gr&r studiesGr&r studies
Gr&r studies
 
LINAC COMMSSN.ppt
LINAC COMMSSN.pptLINAC COMMSSN.ppt
LINAC COMMSSN.ppt
 
qualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationqualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validation
 
Quality assurance in pet
Quality assurance in petQuality assurance in pet
Quality assurance in pet
 
Quality of Analytical Procedures
Quality of Analytical ProceduresQuality of Analytical Procedures
Quality of Analytical Procedures
 
Nox t3 complete review final
Nox t3 complete review finalNox t3 complete review final
Nox t3 complete review final
 
Motion Management in Lung Cancer Radiotherapy
Motion Management in Lung Cancer RadiotherapyMotion Management in Lung Cancer Radiotherapy
Motion Management in Lung Cancer Radiotherapy
 
02trainingmaterialformsa (1) 111030062223-phpapp02
02trainingmaterialformsa (1) 111030062223-phpapp0202trainingmaterialformsa (1) 111030062223-phpapp02
02trainingmaterialformsa (1) 111030062223-phpapp02
 
Intensity-modulated Radiotherapy
Intensity-modulated RadiotherapyIntensity-modulated Radiotherapy
Intensity-modulated Radiotherapy
 
Physical and clinical aspects of brachytherapy
Physical and clinical aspects of  brachytherapyPhysical and clinical aspects of  brachytherapy
Physical and clinical aspects of brachytherapy
 
Good Weighing Practices in QC Lab
Good Weighing Practices in QC LabGood Weighing Practices in QC Lab
Good Weighing Practices in QC Lab
 

More from SGRT Community

Validating the Accuracy, Value, and Investment in SGRT
Validating the Accuracy, Value, and Investment in SGRTValidating the Accuracy, Value, and Investment in SGRT
Validating the Accuracy, Value, and Investment in SGRTSGRT Community
 
Impact of Using SGRT on Treatment Accuracy Based on Transit in Vivo Dosimetry...
Impact of Using SGRT on Treatment Accuracy Based on Transit in Vivo Dosimetry...Impact of Using SGRT on Treatment Accuracy Based on Transit in Vivo Dosimetry...
Impact of Using SGRT on Treatment Accuracy Based on Transit in Vivo Dosimetry...SGRT Community
 
Implementing an End-to-End SGRT Workflow for Breath-Hold SABR
Implementing an End-to-End SGRT Workflow for Breath-Hold SABRImplementing an End-to-End SGRT Workflow for Breath-Hold SABR
Implementing an End-to-End SGRT Workflow for Breath-Hold SABRSGRT Community
 
Feasibility and Tolerability of Surface Guided Radiotherapy in Breath-Hold Li...
Feasibility and Tolerability of Surface Guided Radiotherapy in Breath-Hold Li...Feasibility and Tolerability of Surface Guided Radiotherapy in Breath-Hold Li...
Feasibility and Tolerability of Surface Guided Radiotherapy in Breath-Hold Li...SGRT Community
 
Feasibility Study of Deep Inspiration Breath-Hold Technique with AlignRT InBo...
Feasibility Study of Deep Inspiration Breath-Hold Technique with AlignRT InBo...Feasibility Study of Deep Inspiration Breath-Hold Technique with AlignRT InBo...
Feasibility Study of Deep Inspiration Breath-Hold Technique with AlignRT InBo...SGRT Community
 
How To Go Live with Tattoo and Mark-Free Treatment
How To Go Live with Tattoo and Mark-Free Treatment How To Go Live with Tattoo and Mark-Free Treatment
How To Go Live with Tattoo and Mark-Free Treatment SGRT Community
 
One Clinic With Two Different Systems, and Our Journey Forward with One!
One Clinic With Two Different Systems, and Our Journey Forward with One! One Clinic With Two Different Systems, and Our Journey Forward with One!
One Clinic With Two Different Systems, and Our Journey Forward with One! SGRT Community
 
Measuring and Improving Radiotherapy Delivery Efficiency with SGRT Implementa...
Measuring and Improving Radiotherapy Delivery Efficiency with SGRT Implementa...Measuring and Improving Radiotherapy Delivery Efficiency with SGRT Implementa...
Measuring and Improving Radiotherapy Delivery Efficiency with SGRT Implementa...SGRT Community
 
Validation of Efficiency/ROI Advantages of Postural Video
Validation of Efficiency/ROI Advantages of Postural Video Validation of Efficiency/ROI Advantages of Postural Video
Validation of Efficiency/ROI Advantages of Postural Video SGRT Community
 
Bringing SGRT to Every NHS patient, Every Fraction in Rural Dorset
Bringing SGRT to Every NHS patient, Every Fraction in Rural Dorset Bringing SGRT to Every NHS patient, Every Fraction in Rural Dorset
Bringing SGRT to Every NHS patient, Every Fraction in Rural Dorset SGRT Community
 
Safe Implementation of AlignRT for Breast Patients Across a Multi-Site Center...
Safe Implementation of AlignRT for Breast Patients Across a Multi-Site Center...Safe Implementation of AlignRT for Breast Patients Across a Multi-Site Center...
Safe Implementation of AlignRT for Breast Patients Across a Multi-Site Center...SGRT Community
 
Dealing with SGRT Workload in a Large-Scale University Radiotherapy Department
Dealing with SGRT Workload in a Large-Scale University Radiotherapy Department Dealing with SGRT Workload in a Large-Scale University Radiotherapy Department
Dealing with SGRT Workload in a Large-Scale University Radiotherapy Department SGRT Community
 
Go-Live with Tattoo and Mark Free Treatments
Go-Live with Tattoo and Mark Free Treatments Go-Live with Tattoo and Mark Free Treatments
Go-Live with Tattoo and Mark Free Treatments SGRT Community
 
SGRT for DIBH: from zero to hero 
SGRT for DIBH: from zero to hero  SGRT for DIBH: from zero to hero 
SGRT for DIBH: from zero to hero  SGRT Community
 
SimRT: Workflows and optimizing DIBH planning
SimRT: Workflows and optimizing DIBH planningSimRT: Workflows and optimizing DIBH planning
SimRT: Workflows and optimizing DIBH planningSGRT Community
 
SGRT; Going live with tattoo and mark free treatments 
SGRT; Going live with tattoo and mark free treatments SGRT; Going live with tattoo and mark free treatments 
SGRT; Going live with tattoo and mark free treatments SGRT Community
 
Installation of a Surface Guidance System at a Heavy Ion Gantry Treatment Room​
Installation of a Surface Guidance System at a Heavy Ion Gantry Treatment Room​Installation of a Surface Guidance System at a Heavy Ion Gantry Treatment Room​
Installation of a Surface Guidance System at a Heavy Ion Gantry Treatment Room​SGRT Community
 
IGRT + SGRT for ​ Confident and Efficient SRS​
IGRT + SGRT for ​ Confident and Efficient SRS​IGRT + SGRT for ​ Confident and Efficient SRS​
IGRT + SGRT for ​ Confident and Efficient SRS​SGRT Community
 
From DIBH to Every Patient Every Fraction
From DIBH to Every Patient Every FractionFrom DIBH to Every Patient Every Fraction
From DIBH to Every Patient Every FractionSGRT Community
 
The Next Chapter: SGRT Supporting the Evolving RT Space
The Next Chapter: SGRT Supporting the Evolving RT SpaceThe Next Chapter: SGRT Supporting the Evolving RT Space
The Next Chapter: SGRT Supporting the Evolving RT SpaceSGRT Community
 

More from SGRT Community (20)

Validating the Accuracy, Value, and Investment in SGRT
Validating the Accuracy, Value, and Investment in SGRTValidating the Accuracy, Value, and Investment in SGRT
Validating the Accuracy, Value, and Investment in SGRT
 
Impact of Using SGRT on Treatment Accuracy Based on Transit in Vivo Dosimetry...
Impact of Using SGRT on Treatment Accuracy Based on Transit in Vivo Dosimetry...Impact of Using SGRT on Treatment Accuracy Based on Transit in Vivo Dosimetry...
Impact of Using SGRT on Treatment Accuracy Based on Transit in Vivo Dosimetry...
 
Implementing an End-to-End SGRT Workflow for Breath-Hold SABR
Implementing an End-to-End SGRT Workflow for Breath-Hold SABRImplementing an End-to-End SGRT Workflow for Breath-Hold SABR
Implementing an End-to-End SGRT Workflow for Breath-Hold SABR
 
Feasibility and Tolerability of Surface Guided Radiotherapy in Breath-Hold Li...
Feasibility and Tolerability of Surface Guided Radiotherapy in Breath-Hold Li...Feasibility and Tolerability of Surface Guided Radiotherapy in Breath-Hold Li...
Feasibility and Tolerability of Surface Guided Radiotherapy in Breath-Hold Li...
 
Feasibility Study of Deep Inspiration Breath-Hold Technique with AlignRT InBo...
Feasibility Study of Deep Inspiration Breath-Hold Technique with AlignRT InBo...Feasibility Study of Deep Inspiration Breath-Hold Technique with AlignRT InBo...
Feasibility Study of Deep Inspiration Breath-Hold Technique with AlignRT InBo...
 
How To Go Live with Tattoo and Mark-Free Treatment
How To Go Live with Tattoo and Mark-Free Treatment How To Go Live with Tattoo and Mark-Free Treatment
How To Go Live with Tattoo and Mark-Free Treatment
 
One Clinic With Two Different Systems, and Our Journey Forward with One!
One Clinic With Two Different Systems, and Our Journey Forward with One! One Clinic With Two Different Systems, and Our Journey Forward with One!
One Clinic With Two Different Systems, and Our Journey Forward with One!
 
Measuring and Improving Radiotherapy Delivery Efficiency with SGRT Implementa...
Measuring and Improving Radiotherapy Delivery Efficiency with SGRT Implementa...Measuring and Improving Radiotherapy Delivery Efficiency with SGRT Implementa...
Measuring and Improving Radiotherapy Delivery Efficiency with SGRT Implementa...
 
Validation of Efficiency/ROI Advantages of Postural Video
Validation of Efficiency/ROI Advantages of Postural Video Validation of Efficiency/ROI Advantages of Postural Video
Validation of Efficiency/ROI Advantages of Postural Video
 
Bringing SGRT to Every NHS patient, Every Fraction in Rural Dorset
Bringing SGRT to Every NHS patient, Every Fraction in Rural Dorset Bringing SGRT to Every NHS patient, Every Fraction in Rural Dorset
Bringing SGRT to Every NHS patient, Every Fraction in Rural Dorset
 
Safe Implementation of AlignRT for Breast Patients Across a Multi-Site Center...
Safe Implementation of AlignRT for Breast Patients Across a Multi-Site Center...Safe Implementation of AlignRT for Breast Patients Across a Multi-Site Center...
Safe Implementation of AlignRT for Breast Patients Across a Multi-Site Center...
 
Dealing with SGRT Workload in a Large-Scale University Radiotherapy Department
Dealing with SGRT Workload in a Large-Scale University Radiotherapy Department Dealing with SGRT Workload in a Large-Scale University Radiotherapy Department
Dealing with SGRT Workload in a Large-Scale University Radiotherapy Department
 
Go-Live with Tattoo and Mark Free Treatments
Go-Live with Tattoo and Mark Free Treatments Go-Live with Tattoo and Mark Free Treatments
Go-Live with Tattoo and Mark Free Treatments
 
SGRT for DIBH: from zero to hero 
SGRT for DIBH: from zero to hero  SGRT for DIBH: from zero to hero 
SGRT for DIBH: from zero to hero 
 
SimRT: Workflows and optimizing DIBH planning
SimRT: Workflows and optimizing DIBH planningSimRT: Workflows and optimizing DIBH planning
SimRT: Workflows and optimizing DIBH planning
 
SGRT; Going live with tattoo and mark free treatments 
SGRT; Going live with tattoo and mark free treatments SGRT; Going live with tattoo and mark free treatments 
SGRT; Going live with tattoo and mark free treatments 
 
Installation of a Surface Guidance System at a Heavy Ion Gantry Treatment Room​
Installation of a Surface Guidance System at a Heavy Ion Gantry Treatment Room​Installation of a Surface Guidance System at a Heavy Ion Gantry Treatment Room​
Installation of a Surface Guidance System at a Heavy Ion Gantry Treatment Room​
 
IGRT + SGRT for ​ Confident and Efficient SRS​
IGRT + SGRT for ​ Confident and Efficient SRS​IGRT + SGRT for ​ Confident and Efficient SRS​
IGRT + SGRT for ​ Confident and Efficient SRS​
 
From DIBH to Every Patient Every Fraction
From DIBH to Every Patient Every FractionFrom DIBH to Every Patient Every Fraction
From DIBH to Every Patient Every Fraction
 
The Next Chapter: SGRT Supporting the Evolving RT Space
The Next Chapter: SGRT Supporting the Evolving RT SpaceThe Next Chapter: SGRT Supporting the Evolving RT Space
The Next Chapter: SGRT Supporting the Evolving RT Space
 

Recently uploaded

(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...Taniya Sharma
 
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoybabeytanya
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...CALL GIRLS
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Deliverynehamumbai
 
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...Miss joya
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableNehru place Escorts
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service CoimbatoreCall Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatorenarwatsonia7
 
Bangalore Call Girls Hebbal Kempapura Number 7001035870 Meetin With Bangalor...
Bangalore Call Girls Hebbal Kempapura Number 7001035870  Meetin With Bangalor...Bangalore Call Girls Hebbal Kempapura Number 7001035870  Meetin With Bangalor...
Bangalore Call Girls Hebbal Kempapura Number 7001035870 Meetin With Bangalor...narwatsonia7
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Servicevidya singh
 
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...indiancallgirl4rent
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 

Recently uploaded (20)

(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
 
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
 
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service CoimbatoreCall Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
 
Bangalore Call Girls Hebbal Kempapura Number 7001035870 Meetin With Bangalor...
Bangalore Call Girls Hebbal Kempapura Number 7001035870  Meetin With Bangalor...Bangalore Call Girls Hebbal Kempapura Number 7001035870  Meetin With Bangalor...
Bangalore Call Girls Hebbal Kempapura Number 7001035870 Meetin With Bangalor...
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
 
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 

Commissioning AlignRT with Minimal Disruption

  • 1. Another Device to Commission? Begin Commissioning AlignRT® with Minimal Disruption
  • 3. INTRODUCTION 3 • Technology and complexity in radiation oncology continues to grow at a fast pace • The need for accurate implementation of SRS/SBRT QA programs is crucial • Increased pressure to keep cost down and be more efficient How can physicists provide increased value to clinics with increasing demands, but maintain a high level of safety?
  • 4. INTRODUCTION 4 • Combine tests to accomplish multiple QA procedures at one time • Focus first on mission critical processes (think TG-100 and FMEA analysis) • Use AAPM Task Groups as guidance when developing your program Now, what about a technology that I don’t know? Like surface imaging?
  • 5. COMMISSIONING SURFACE IMAGING 5 The approach is the same as every other modality in radiation therapy • Start out with the Task Group (AAPM TG-147)1 for recommendations • Discuss with your clinicians the goals of the system (motion management for SRS, positioning for breast patients, etc) • Prioritize the QA in terms of the clinical need – Do you need to perform all QA possible? 1 Willoughby, T., et al. “Quality assurance for nonradiographic radiotherapy localization and positioning systems: Report of Task Group 147.” 8 March 2012
  • 6. 6 THE TECHNOLOGY • Stereoscopic camera system that is used for positioning and tracking • Isocenter is determined from a novel calibration procedure2 • Triangulation provides a 3D reference frame using a speckle pattern on the patient’s surface ALIGNRT® INTRODUCTION 2 http://www.visionrt.com/content/core-technology
  • 7. 7 COMMON APPLICATION • Intrafraction motion-management for SRS, SBRT, and respiratory cases • Positioning for sites where the surface is a good surrogate for the target – Breast, intracranial, extremities, etc • General use for all patients – Gross movements • The beam can be held (automatically on most systems) when the registered surfaces are out of the specified tolerance ALIGNRT® INTRODUCTION
  • 8. 8 Start with the clinical use cases to establish QA procedures in conjunction with the AAPM TG-147 One size may not fit all, and a good QA program will reflect the clinical need • What site will be treated most often? • Will it be used for motion-management, positioning, or both? • Will the beam be held during treatment delivery (dynamic gating)? • What tolerances will be needed during position and treatment delivery? CLINICAL USE CASES
  • 9. 9 Start by mapping out the workflow and establish an End-to-end test, which incorporates the complete clinical use case COMMISSIONING TESTS
  • 10. 10 Try to create a test that mimics the clinical use case (E2E) END-TO-END (E2E) TEST
  • 11. 11 EQUIPMENT NEEDED • Phantom (preferably anthropomorphic) that has a neutral surface for imaging – Shouldn’t be a ball or cylinder, due to tracking difficulties – MAX-HD® (Integrated Medical Technologies) a good choice – STEEV SRS Phantom (CIRS) – SRS Head Phantom from IROC (clear, so may need taped to image well) • Detectors such as Gafchromic® film, microchambers/diodes, OSL/TLD • Electrometer • 3DOF Head Adjuster (provided by VisionRT) or 6DOF couch for accurate positioning END-TO-END (E2E) TEST
  • 12. 12 END-TO-END (E2E) TEST SAMPLE PHANTOMS MAX-HD STEEV SRS (IROC) SRS Phantom IMT CIRS IROC CIRS
  • 13. 13 SOME NOTES ABOUT EQUIPMENT • Detectors should be small enough to avoid partial volume effects, especially for SRS • Film scanning should use the right calibration procedure • Make sure film is cut in such a way to ensure reproducibility and localization – Laser cut films for the phantom are available • Epson Perfection XL11000 (or the older XL10000) are good choices for film scanning END-TO-END (E2E) TEST
  • 14. 14 CT SCANNING • Use the highest resolution protocol possible (small slice thickness ~1mm, high mAs, etc) • Try and use the same immobilization techniques that would be used during treatment • Set your DICOM origin (CT origin) at a reproducible localization to minimize setup errors • If the phantom supports it, try multiple CT scans with different inserts and cubes – Again, it’s important to know the origin with certainty for film scanning END-TO-END (E2E) TEST
  • 15. 15 CONTOURING • Many treatment planning systems automatically create the BODY contour, so review for accuracy • The Target Volume can be the sensitive volume of the detector or the center of the film END-TO-END (E2E) TEST
  • 16. 16 TREATMENT PLANNING • Use a beam configuration that will test the system – Couch kicks – VMAT, if applicable, to block cameras – Dose levels that mimic treatment time (don’t scale MU) – Use a real plan if available • Save for a baseline for routine QA END-TO-END (E2E) TEST
  • 17. 17 LOCALIZATION • Use the AlignRT system to be the primary localization method • Verify the positioning with CBCT and lasers – There may be a small deviation with the DICOM because of the BODY contour – If so, position initially with CBCT, capture the reference, then reposition with the AlignRT system • Use a 6DOF localization technique to position the phantom END-TO-END (E2E) TEST
  • 18. 18 DELIVERY • Make note of gantry occlusion and deltas – Some variation is acceptable during blockage, but extreme shifts may indicate improper lighting, ROI, or calibration conditions • Watch the deltas for drift over the course of the treatment at the couch base • When rotating the couch, use the AlignRT readout, not the digital indicators, to reposition – Stereoscopic systems trump the internal readouts of the LINAC system, hence the “STEREO” in stereotactic – This is especially important in older systems, where couch tolerances are looser END-TO-END (E2E) TEST
  • 19. 19 ANALYSIS • Temperature/pressure correct the chamber reading, and use a cross calibration factor to calculate the ratio of charge/dose (traced by to your ASCL calibrated detector) • FilmQA® from Ashland is a good choice for film scanning – FilmQA analyzes in all 3 RGB channels – ImageJ is also available for free, but takes some effort to write code • Always scan the film in the same orientation, and if the calibration films are not done on the same day, wait a while for the film to develop END-TO-END (E2E) TEST
  • 20. 20 WHAT AM I LOOKING FOR? Depends on the clinical goals • Breast patients could be passing 95% with a gamma index at 3%, 3mm with the film analysis • SRS programs treating trigeminal neuralgia cases may need 2%, 1mm • Output measurements should be within 2% of the expected from the TPS, but can be tough for small cones because of the output factors If you pass the E2E, the commissioning process got a whole lot better. But… END-TO-END (E2E) TEST
  • 21. 21 WHAT IF IT FAILS? Start looking at each component of the system closer. You will need it for the overall QA program anyways. Keep in mind: 1. TG-147 is just part of this. Since an E2E test checks the whole system, it could be anything outlined in TG-40, TG-142, TG-66, etc 2. Focus on the components that are specific to TG-147, as to not get overwhelmed END-TO-END (E2E) TEST
  • 22. 22 • Do a “chart check” style review to make sure items transferred properly – Patient ID – Isocenter – Contour (especially BODY) – Scan orientation • The coordinate system of the LINAC matches the TPS and DICOM transfer – For example, the couch coordinates could be opposite if the coordinate systems don’t match COMMICATION BETWEEN SYSTEMS
  • 23. 23 • Do a “chart check” style review to make sure items transferred properly – Patient ID – Isocenter – Contour (especially BODY) – Scan orientation • The coordinate system of the LINAC matches the TPS and DICOM transfer – For example, the couch coordinates could be opposite if the coordinate systems don’t match • If using an interface for shifts, ensure that is correct by using lasers or another surrogate to confirm AlignRT shifts performed by the LINAC console COMMICATION BETWEEN SYSTEMS
  • 24. 24 • An accurate ROI gives stable real-time deltas for any phantom when there are no occlusions. If the delta jump at couch base, and no blockage – Ensure the ROI give a unique view • I.e., not flat, symmetric, nor broken – The underlying image is intact • Check lighting condition or skin tone • It is equally important to not make the ROI so big it doesn’t detect small movements – Like contouring the entire head for SRS – just use the small area in the open mask REGION OF INTEREST (ROI)
  • 25. 25 • In order to make sure the calibration (both the calibration plate and fine isocenter) is correct, perform a series of shifts to known positions, and record the deltas – Shift 2mm, 1cm, 2cm, 5cm, 10cm in S/I, L/R, and A/P directions – Kick the couch in increments of 45 degree, for example – Move the phantom arbitrarily, and see if the 6DOF positioning can get back to a zero baseline • If greater than 1mm (for SRS) disagreement is seen, perform isocenter and monthly calibrations again, and repeat • Refer to the acceptance testing procedures for vendor baselines as well SHIFTS
  • 26. 26 • With the camera system on for a significant period of time (like an hour or so), turn the monitoring on and watch the deltas • Record the values, and if large deviations are seen, repeat after a longer warm up is achieved • Compared to infrared based systems, the drift should be quite minimal SHIFTS
  • 27. 27 • With the camera system on for a significant period of time (like an hour or so), turn the monitoring on and watch the deltas • Record the values, and if large deviations are seen, repeat after a longer warm up is achieved • Compared to infrared based systems, the drift should be quite minimal SHIFTS
  • 28. 28 • With a simple motion phantom, try and move the phantom enough to hold the beam • Tighten tolerance to only beam on when the chamber is in position • Use a simple test plan to isolate it just the motion of the phantom • Measure with an ion chamber and record the results – Should be very close to baseline, stationary reading, assuming the phantom is not shifting during motion, and the gating is working • This would also test spatial accuracy at the same time DYNAMIC GATING
  • 29. 29 • As with all medical devices, it is up to the medical physicists to stay current with the latest customer service bulletins and release notes • The tests outlined are from a clinical medical physicist’s perspective, but are by no means meant to be prescriptive • Training and continuing education are equally, and maybe more, important than a one-time commissioning – as is the continual QA program VENDOR RECOMMENDATIONS
  • 30. 30 • VisionRT has already provided a nice workflow for QA – DailyQA • Uses the calibration plate, aligned by the therapists, and analyzes it for constancy – MonthlyQA • Again, using the plate, but overrides the triangulated 3D position of the system. To be done if the system appears to have changed – Isocenter Calibration • Using a phantom with fiducials that can be seen on CT/MV, with the surface imaged, allows the isocenter of the imaging system to be correlated to the MV isocenter ESTABLISHING A QA PROGRAM
  • 31. 31 • In addition to the vendor provided phantoms and software, use the results from the commissioning tests to create baselines for routine QA ESTABLISHING A QA PROGRAM
  • 32. 32 "If it's not written down, it didn't happen“ • Ensure all the tests you do during commissioning, QA, or anything concerning the system are well documented • Not only is it good to establish baselines and trends, it is very useful in helping other centers when they go live with AlignRT DOCUMENTATION
  • 33. 33 • Commissioning any system doesn’t have to be difficult, or even that time- consuming, any long as the problem is well-defined • By using clinical use cases as guidance, tests will solidify, and the project can be focused • By incorporated E2E tests, and grouping other QA procedures in an effective way, commissioning and routine QA time can be reduced CONCLUSIONS
  • 34. 34 THANK YOU! www.alyzenmed.com 1801 South 54th Street Paragould, AR 72450 870.926.0894 jrogers@alyzenmed.com Jonathan Rogers, MS, DABR