Optical Coherence
Tomography
Rajeswari K


M.Optom
R K o p t o A s O C T
OCT
OCT is a non- contact, noninvasive imaging technique
used to obtain high resolution 10 cross sectional images
of the retina and anterior segment.


Re
fl
ected light is used instead of sound waves.


Infrared ray of 830 nm with 78D internal lens.
2
R K o p t o A s O C T
Basics
Principle of OCT is based on Michaelson’s
interferometer.
3
R K o p t o A s O C T
Theory
Scattering is a fundamental property of a heterogeneous
medium, and occurs because of variations in the
refractive index within tissue
5
R K o p t o A s O C T
Principle
Interferometry is the technique of superimposing
(interfering) two or more waves, to detect differences
between them.
6
Principle
Interferometry works because


Two waves with the same
frequency that have the
same phase will add each
other (constructive)


Two waves that have
opposite phase will
subtract.(destructive)
R K o p t o A s O C T
Instrumentation
Light from a source is directed onto a partially re
fl
ecting
mirror and is split into a reference and a measurement
beam.


The measurement beam re
fl
ected from the specimen
with different time delays according to its internal
microstructure.
8
R K o p t o A s O C T
Working
The light in the reference beam is re
fl
ected from a
reference mirror at a variable distance which produces a
variable time delay.


The light from the specimen, consisting of multiple
echoes, and the light from the reference mirror,
consisting of a single echo at a known delay are
combined and detected.
9
R K o p t o A s O C T
10
R K o p t o A s O C T
Types
Anterior segment OCT


Posterior segment OCT
11
R K o p t o A s O C T
12
ANTERIOR SEGMENT
OCT
R K o p t o A s O C T
Overview
Because anatomy plays a large role in chronic angle closure, evaluating the
angle anatomy is a critical
fi
rst step in treat ing this disease.


Using a combination of various AS imaging technique examiner will get a
clear picture of the angle anatomy and help to make appropriate treatment
decisions.


Gonioscopy is the current gold standard, but imaging techniques such as
anterior segment optical coherence tomography (AS-OCT) provide
additional valuable information.
13
R K o p t o A s O C T
14
History
First described using the same wavelength of
light as retina l OCT
, namely 830nm.


OCT imaging of t he anterior segment with a
longer wavelength of 1310nm was developed
later on and had t he advantages of better
penetration through sclera


Rapid, non contact method that maybe
performed by a technician with patient sitting
Upright.
R K o p t o A s O C T
15
Obtains a high-resolution cross-section of the anterior
segment without touching the eye.


Enables visualisation of the form, site & position of AC
components.


De
fi
nes & measures the distances between them.


Machine automatically eliminates distortion induced by
optical transmission factors.
R K o p t o A s O C T
16
Re
fl
ex Saturation Beam indicates
perpendicularity of the eye to
minimal tilting of the image.
The scan dimensions are 16 x 6 mm and 4 image frames are
averaged.


Resolution is 2-20microns which makes it suita ble for
measurement of
fi
ner eye structures.
R K o p t o A s O C T
17
Qualitative Assessment
An important landmark to identify in ASOCT images is the scleraI spur.


This is visible as an inward projection of the sclera at the junction
between the inner sclera and cornea


Apposition between the iris & inner corneo-scleral wall has been used
as a qualitative method of detecting angle closure
R K o p t o A s O C T
18
Quantitative Assessment
Quantitative measurement of the AC angle is possible with
in-built software & requires identi
fi
cation of the sclera l spur.


The commonly used quantitative parameters


Angle opening distance(AODSOO/AOD750}


Angle recess area(ARASOO/ARA750}


Trabecular space area(TISASOO/T ISA750)
R K o p t o A s O C T
19
Angle opening distance(mm): Perpendicular distance between a
point anterior to the scleral spur and the opposing iris. (500µm)


Angle recess area(mm2): The triangular area bounded by the
anterior iris surface and the inner corneo-scleral wall


Trabecular Iris space area(mm2): the perpendicular distance
between the scleral spur and the opposing iris.
R K o p t o A s O C T
20
R K o p t o A s O C T
21
Newer Modalities
A new swept source FD-OCT device can rapidly obtain
high-resolution scans with 3D reconstruction of the
anterior segment.
Casia SS-1000

Tomey Corporation
R K o p t o A s O C T
Clinical Uses
An adjunct to gonioscopy as well as a substitute when gonioscopy is not feasible.


Useful as a patient education tool, especially when laser peripheral iridotomy is being
recommended.


Photo-therapeutic keratectomy to assess the depth of corneal opacity.


LASIK for CCT & to measure
fl
ap thickness.


Keratoconus & INTACS


Cysts of anterior segment to diagnose etiology.


Biometry for phakic IOL.
22
R K o p t o A s O C T
Glaucoma Imaging &


Newer Developments
R K o p t o A s O C T
24
BLEB
R K o p t o A s O C T
25
R K o p t o A s O C T
Useful tool for evaluating
fi
ltering bleb or GDD in post-
op period.


Clinical description of bleb & Visualising intra-bleb
morphology with AS-OCT may enhance our
understanding of different surgical outcomes.
26
R K o p t o A s O C T
Newer Uses of OCT
Measuring the tear
fi
lm :Although this work is still
mostly showing value in research situations


Measuring contact Lens
fi
t: Scleral CL
27
R K o p t o A s O C T
ADVANTAGES
Non contact.


More physiological.


Shorter imaging time.


Less learning curve for operator.


Target may be used to induce accommodation in eye imaged.


Less inter operator variability.
28
R K o p t o A s O C T
LIMITATION
No penetration of pigmented iris epit helium.
29
R K o p t o A s O C T
30
Completely reduced anterior chamber (A).

The same eye after treatment, now with an open angle (B).
Malignant 

Glaucoma
R K o p t o A s O C T
31
Decedent’s Membrane Detachment
R K o p t o A s O C T
32
R K o p t o A s O C T
33
Granular Dystrophy
R K o p t o A s O C T
34
POSTERIOR SEGMENT
OCT
R K o p t o A s O C T
Types
1. Time domain


2. Fortier domin


3. Spectral domin
35
R K o p t o A s O C T
Time Domin
The Michelson interferometer splits the light from the broadband
source into two paths, the reference and sample arms.


The interference signal between the re
fl
ected reference wave
and the backscattered sample wave is then recorded.


In order to measure the time delays of light echoes coming from
different structures within the eye, the position of the reference
mirror is changed so that the time delay of the reference light
pulse is adjusted accordingly
36
R K o p t o A s O C T
Fortier Domin
In FD-OCT ,the detector arm of the Michelson
interferometer uses a spectrometer instead a single
detector.


The spectrometer measures spectral modulations
produced by interference between the sample and
reference re
fl
ections.
37
R K o p t o A s O C T
TD vs FD
No physical scanning of the reference mirror is required;
thus, FD-OCT can be much faster than TDOCT.


The simultaneous detection of re
fl
ections from a broad
range of depths is much more ef
fi
cient than TD-OCT, in
which signals from various depths are scanned sequentially.


FD-OCT is also fast enough for sequential image frames to
track the pulsation of blood vessels during the cardiac cycle.
38
R K o p t o A s O C T
39
R K o p t o A s O C T
40
R K o p t o A s O C T
Limitation - SLO
Limitation of OCT technology was dif
fi
culty in accurately
localising the cross-sectional images and correlating
them with a conventional en face view of the fundus.


To localise and visually interpret the images, a scanning
laser ophthalmoscopy (SLO) was needed.


This was initiated by OTI technologies (Toronto, Canada)
to develop the Spectral OCT/SLO.
41
R K o p t o A s O C T
Spectral Domin
The Spectral OCT/SLO is a computerized optical scanner device
providing high-resolution, high-de
fi
nition images of the fundus
anatomy.


It integrats SLO’s confocal imaging principles with OCT.


The system simultaneously produces SLO and OCT images that are
created through the same optical path,


It produces a new image format with high resolution tomographic
images.
42
R K o p t o A s O C T
Clinical Uses
Oct is used for diagnosis, monitoring, quantitative and
qualitative assessment of macular pathology


It is used in most of the Vitreo-retinal conditions
43
R K o p t o A s O C T
7. Epiretinal Membrane in
Diabetic Retinopathy.


8. Central Serous Retinopathy.


9. Foveal thining in Retinitis
Pigmentosa and in Myopic
Degeneration.


10.Age Related Macular
Degeneration.


11.Post Retinal Detachment
Surgery.
44
1. PosteriorVitreous Detachment
(PVD)


2. Vitreo-MacularTraction Syndrome
(VMT)


3. Asteriod Hyalosis


4. Intra-retinal Hemorrhage.


5. Macular Edema ( CME, CSME)


6. Macular Holes and cysts.
U
S
E
S
R K o p t o A s O C T
Interpretation
Following things need to observed in OCT:


A. Vitro-retinal interface.


B. RetinalThickness:


(I) Increases: Edema,Traction.


(II) Decreases: Foveal atrophy.
45
R K o p t o A s O C T
C. Re
fl
ectivity:


(I) Hyper-re
fl
ective lesions: Hard Exudates, Blood, Scars.


(II) Hypo-re
fl
ective lesions: Serous Fluid, Hypo-pigmented
lesions of Retinal Pigment Epithelium, Media Hazy.
46
R K o p t o A s O C T
Normal
Highly re
fl
ective structures are shown in bright colures (white and red).


Those with low re
fl
ectivity are represented by dark colors (black and blue).


Intermediate re
fl
ectivity is shown Green.
47
R K o p t o A s O C T
Diabetic Macular Edema
Sponge like retinal thickness.


Macular edema and Cystoid Macular Edema.


Serous Retinal Detachment.


Epi-retinal Membrane.


Vitro-macular traction.
48
R K o p t o A s O C T
49
R K o p t o A s O C T
50
R K o p t o A s O C T
51
R K o p t o A s O C T
52
RD
R K o p t o A s O C T
53
ERM
R K o p t o A s O C T
ARMD
Focal elevation of Retinal Pigment Epithelium.


Large drusen shows irregular elevation of Retinal
Pigment Epithelium with shadow from underlying
choroid.


Large drusen carry high risk of Choroidal Neo-Vascular
Membrane.
54
R K o p t o A s O C T
55
R K o p t o A s O C T
MACULAR HOLE
Three are three type of macular hole we can see:


1. Impending Macular Hole.


2. Lamellar Macular Hole.


3. FullThickness Macular Hole.
56
R K o p t o A s O C T
57
R K o p t o A s O C T
58
R K o p t o A s O C T
59
R K o p t o A s O C T
Disadvantage
Clear ocular media is needed for better performance


Minimum pupillary size of 4mm is required.


It is a costly apparatus.
60
R K o p t o A s O C T
61

More Related Content

What's hot

What's hot (20)

Corneal topography by suraj
Corneal topography by surajCorneal topography by suraj
Corneal topography by suraj
 
AS-OCT
AS-OCTAS-OCT
AS-OCT
 
Contrast sensitivity
Contrast sensitivityContrast sensitivity
Contrast sensitivity
 
Corneal topography
Corneal topographyCorneal topography
Corneal topography
 
Pentacam
Pentacam Pentacam
Pentacam
 
Corneal pachymetry.ppt
Corneal pachymetry.pptCorneal pachymetry.ppt
Corneal pachymetry.ppt
 
Simple & Toric Transposition
Simple & Toric TranspositionSimple & Toric Transposition
Simple & Toric Transposition
 
Autorefractometry: principle and procedure.
Autorefractometry: principle and procedure.Autorefractometry: principle and procedure.
Autorefractometry: principle and procedure.
 
Optical coherence tomography
Optical coherence tomographyOptical coherence tomography
Optical coherence tomography
 
A scan ultrasonography
A scan ultrasonographyA scan ultrasonography
A scan ultrasonography
 
Jackson Cross Cylinder
Jackson Cross CylinderJackson Cross Cylinder
Jackson Cross Cylinder
 
pentacam
pentacampentacam
pentacam
 
Binocular balancing
Binocular balancing Binocular balancing
Binocular balancing
 
Preoperative evaluation for LASIK & PRK
Preoperative evaluation for LASIK & PRKPreoperative evaluation for LASIK & PRK
Preoperative evaluation for LASIK & PRK
 
Examination protocol for binocular vision
Examination protocol for binocular visionExamination protocol for binocular vision
Examination protocol for binocular vision
 
Measuring interpupillary distance
Measuring interpupillary distanceMeasuring interpupillary distance
Measuring interpupillary distance
 
Anterior Segment OCT in Ophthalmology
Anterior Segment OCT in OphthalmologyAnterior Segment OCT in Ophthalmology
Anterior Segment OCT in Ophthalmology
 
Entoptic phenomena
Entoptic phenomenaEntoptic phenomena
Entoptic phenomena
 
Non contact tonometer ppt
Non contact tonometer pptNon contact tonometer ppt
Non contact tonometer ppt
 
Near point of convergence
Near point of convergenceNear point of convergence
Near point of convergence
 

Similar to Optical Coherence Tomography (OCT) Anterior & Posterior

COMPUTED TOMOGRAPHY SCAN
COMPUTED TOMOGRAPHY SCANCOMPUTED TOMOGRAPHY SCAN
COMPUTED TOMOGRAPHY SCAN
Shounak Nandi
 

Similar to Optical Coherence Tomography (OCT) Anterior & Posterior (20)

ASOCT NEW SLIDES.pptx
ASOCT NEW SLIDES.pptxASOCT NEW SLIDES.pptx
ASOCT NEW SLIDES.pptx
 
Role of oct in ophthalmology
Role of oct in ophthalmologyRole of oct in ophthalmology
Role of oct in ophthalmology
 
OCT , Laser therapy for DR , Vitrectomy
OCT , Laser therapy for DR , VitrectomyOCT , Laser therapy for DR , Vitrectomy
OCT , Laser therapy for DR , Vitrectomy
 
COMPUTED TOMOGRAPHY SCAN
COMPUTED TOMOGRAPHY SCANCOMPUTED TOMOGRAPHY SCAN
COMPUTED TOMOGRAPHY SCAN
 
oct-ujjval solanki
oct-ujjval solankioct-ujjval solanki
oct-ujjval solanki
 
OCT MACULA INTERPRETATION.
OCT MACULA  INTERPRETATION.OCT MACULA  INTERPRETATION.
OCT MACULA INTERPRETATION.
 
Oct demystified
Oct demystified  Oct demystified
Oct demystified
 
Final oct
Final octFinal oct
Final oct
 
Oct updateS -AJAY DUDANI
Oct updateS -AJAY DUDANIOct updateS -AJAY DUDANI
Oct updateS -AJAY DUDANI
 
OCT Angiography
OCT AngiographyOCT Angiography
OCT Angiography
 
OCT in glaucoma ppt;1
OCT in glaucoma ppt;1OCT in glaucoma ppt;1
OCT in glaucoma ppt;1
 
AS-OCT.pptx
AS-OCT.pptxAS-OCT.pptx
AS-OCT.pptx
 
Pentacam demystified 2016
Pentacam demystified 2016Pentacam demystified 2016
Pentacam demystified 2016
 
Role of oct in glaucoma
Role of oct in glaucomaRole of oct in glaucoma
Role of oct in glaucoma
 
Optical Coherence Tomography(OCT) in posterior segment diseases
Optical Coherence Tomography(OCT) in posterior segment diseasesOptical Coherence Tomography(OCT) in posterior segment diseases
Optical Coherence Tomography(OCT) in posterior segment diseases
 
spectral domain OCT by Hala Fathi Hannot
spectral domain OCT by Hala Fathi Hannotspectral domain OCT by Hala Fathi Hannot
spectral domain OCT by Hala Fathi Hannot
 
OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY
OPTICAL COHERENCE TOMOGRAPHY  ANGIOGRAPHY  OPTICAL COHERENCE TOMOGRAPHY  ANGIOGRAPHY
OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY
 
Anterior segment and Corneal OCT review
Anterior segment and Corneal OCT reviewAnterior segment and Corneal OCT review
Anterior segment and Corneal OCT review
 
OCT For Everyone..
OCT For Everyone..OCT For Everyone..
OCT For Everyone..
 
Oct introduction
Oct  introductionOct  introduction
Oct introduction
 

Recently uploaded

❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
Rashmi Entertainment
 
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
amritaverma53
 
Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...
Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...
Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...
Call Girls in Nagpur High Profile Call Girls
 

Recently uploaded (20)

❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
 
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
 
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableCall Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
 
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
 
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
 
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their Regulation
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
 
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
 
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
 
Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...
Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...
Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...
 
Bhopal❤CALL GIRL 9352988975 ❤CALL GIRLS IN Bhopal ESCORT SERVICE
Bhopal❤CALL GIRL 9352988975 ❤CALL GIRLS IN Bhopal ESCORT SERVICEBhopal❤CALL GIRL 9352988975 ❤CALL GIRLS IN Bhopal ESCORT SERVICE
Bhopal❤CALL GIRL 9352988975 ❤CALL GIRLS IN Bhopal ESCORT SERVICE
 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
 
Chennai Call Girls Service {7857862533 } ❤️VVIP ROCKY Call Girl in Chennai
Chennai Call Girls Service {7857862533 } ❤️VVIP ROCKY Call Girl in ChennaiChennai Call Girls Service {7857862533 } ❤️VVIP ROCKY Call Girl in Chennai
Chennai Call Girls Service {7857862533 } ❤️VVIP ROCKY Call Girl in Chennai
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanisms
 
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 

Optical Coherence Tomography (OCT) Anterior & Posterior

  • 2. R K o p t o A s O C T OCT OCT is a non- contact, noninvasive imaging technique used to obtain high resolution 10 cross sectional images of the retina and anterior segment. Re fl ected light is used instead of sound waves. Infrared ray of 830 nm with 78D internal lens. 2
  • 3. R K o p t o A s O C T Basics Principle of OCT is based on Michaelson’s interferometer. 3
  • 4.
  • 5. R K o p t o A s O C T Theory Scattering is a fundamental property of a heterogeneous medium, and occurs because of variations in the refractive index within tissue 5
  • 6. R K o p t o A s O C T Principle Interferometry is the technique of superimposing (interfering) two or more waves, to detect differences between them. 6
  • 7. Principle Interferometry works because Two waves with the same frequency that have the same phase will add each other (constructive) Two waves that have opposite phase will subtract.(destructive)
  • 8. R K o p t o A s O C T Instrumentation Light from a source is directed onto a partially re fl ecting mirror and is split into a reference and a measurement beam. The measurement beam re fl ected from the specimen with different time delays according to its internal microstructure. 8
  • 9. R K o p t o A s O C T Working The light in the reference beam is re fl ected from a reference mirror at a variable distance which produces a variable time delay. The light from the specimen, consisting of multiple echoes, and the light from the reference mirror, consisting of a single echo at a known delay are combined and detected. 9
  • 10. R K o p t o A s O C T 10
  • 11. R K o p t o A s O C T Types Anterior segment OCT Posterior segment OCT 11
  • 12. R K o p t o A s O C T 12 ANTERIOR SEGMENT OCT
  • 13. R K o p t o A s O C T Overview Because anatomy plays a large role in chronic angle closure, evaluating the angle anatomy is a critical fi rst step in treat ing this disease. Using a combination of various AS imaging technique examiner will get a clear picture of the angle anatomy and help to make appropriate treatment decisions. Gonioscopy is the current gold standard, but imaging techniques such as anterior segment optical coherence tomography (AS-OCT) provide additional valuable information. 13
  • 14. R K o p t o A s O C T 14 History First described using the same wavelength of light as retina l OCT , namely 830nm. OCT imaging of t he anterior segment with a longer wavelength of 1310nm was developed later on and had t he advantages of better penetration through sclera Rapid, non contact method that maybe performed by a technician with patient sitting Upright.
  • 15. R K o p t o A s O C T 15 Obtains a high-resolution cross-section of the anterior segment without touching the eye. Enables visualisation of the form, site & position of AC components. De fi nes & measures the distances between them. Machine automatically eliminates distortion induced by optical transmission factors.
  • 16. R K o p t o A s O C T 16 Re fl ex Saturation Beam indicates perpendicularity of the eye to minimal tilting of the image. The scan dimensions are 16 x 6 mm and 4 image frames are averaged. Resolution is 2-20microns which makes it suita ble for measurement of fi ner eye structures.
  • 17. R K o p t o A s O C T 17 Qualitative Assessment An important landmark to identify in ASOCT images is the scleraI spur. This is visible as an inward projection of the sclera at the junction between the inner sclera and cornea Apposition between the iris & inner corneo-scleral wall has been used as a qualitative method of detecting angle closure
  • 18. R K o p t o A s O C T 18 Quantitative Assessment Quantitative measurement of the AC angle is possible with in-built software & requires identi fi cation of the sclera l spur. The commonly used quantitative parameters Angle opening distance(AODSOO/AOD750} Angle recess area(ARASOO/ARA750} Trabecular space area(TISASOO/T ISA750)
  • 19. R K o p t o A s O C T 19 Angle opening distance(mm): Perpendicular distance between a point anterior to the scleral spur and the opposing iris. (500µm) Angle recess area(mm2): The triangular area bounded by the anterior iris surface and the inner corneo-scleral wall Trabecular Iris space area(mm2): the perpendicular distance between the scleral spur and the opposing iris.
  • 20. R K o p t o A s O C T 20
  • 21. R K o p t o A s O C T 21 Newer Modalities A new swept source FD-OCT device can rapidly obtain high-resolution scans with 3D reconstruction of the anterior segment. Casia SS-1000 Tomey Corporation
  • 22. R K o p t o A s O C T Clinical Uses An adjunct to gonioscopy as well as a substitute when gonioscopy is not feasible. Useful as a patient education tool, especially when laser peripheral iridotomy is being recommended. Photo-therapeutic keratectomy to assess the depth of corneal opacity. LASIK for CCT & to measure fl ap thickness. Keratoconus & INTACS Cysts of anterior segment to diagnose etiology. Biometry for phakic IOL. 22
  • 23. R K o p t o A s O C T Glaucoma Imaging & Newer Developments
  • 24. R K o p t o A s O C T 24 BLEB
  • 25. R K o p t o A s O C T 25
  • 26. R K o p t o A s O C T Useful tool for evaluating fi ltering bleb or GDD in post- op period. Clinical description of bleb & Visualising intra-bleb morphology with AS-OCT may enhance our understanding of different surgical outcomes. 26
  • 27. R K o p t o A s O C T Newer Uses of OCT Measuring the tear fi lm :Although this work is still mostly showing value in research situations Measuring contact Lens fi t: Scleral CL 27
  • 28. R K o p t o A s O C T ADVANTAGES Non contact. More physiological. Shorter imaging time. Less learning curve for operator. Target may be used to induce accommodation in eye imaged. Less inter operator variability. 28
  • 29. R K o p t o A s O C T LIMITATION No penetration of pigmented iris epit helium. 29
  • 30. R K o p t o A s O C T 30 Completely reduced anterior chamber (A). The same eye after treatment, now with an open angle (B). Malignant Glaucoma
  • 31. R K o p t o A s O C T 31 Decedent’s Membrane Detachment
  • 32. R K o p t o A s O C T 32
  • 33. R K o p t o A s O C T 33 Granular Dystrophy
  • 34. R K o p t o A s O C T 34 POSTERIOR SEGMENT OCT
  • 35. R K o p t o A s O C T Types 1. Time domain 2. Fortier domin 3. Spectral domin 35
  • 36. R K o p t o A s O C T Time Domin The Michelson interferometer splits the light from the broadband source into two paths, the reference and sample arms. The interference signal between the re fl ected reference wave and the backscattered sample wave is then recorded. In order to measure the time delays of light echoes coming from different structures within the eye, the position of the reference mirror is changed so that the time delay of the reference light pulse is adjusted accordingly 36
  • 37. R K o p t o A s O C T Fortier Domin In FD-OCT ,the detector arm of the Michelson interferometer uses a spectrometer instead a single detector. The spectrometer measures spectral modulations produced by interference between the sample and reference re fl ections. 37
  • 38. R K o p t o A s O C T TD vs FD No physical scanning of the reference mirror is required; thus, FD-OCT can be much faster than TDOCT. The simultaneous detection of re fl ections from a broad range of depths is much more ef fi cient than TD-OCT, in which signals from various depths are scanned sequentially. FD-OCT is also fast enough for sequential image frames to track the pulsation of blood vessels during the cardiac cycle. 38
  • 39. R K o p t o A s O C T 39
  • 40. R K o p t o A s O C T 40
  • 41. R K o p t o A s O C T Limitation - SLO Limitation of OCT technology was dif fi culty in accurately localising the cross-sectional images and correlating them with a conventional en face view of the fundus. To localise and visually interpret the images, a scanning laser ophthalmoscopy (SLO) was needed. This was initiated by OTI technologies (Toronto, Canada) to develop the Spectral OCT/SLO. 41
  • 42. R K o p t o A s O C T Spectral Domin The Spectral OCT/SLO is a computerized optical scanner device providing high-resolution, high-de fi nition images of the fundus anatomy. It integrats SLO’s confocal imaging principles with OCT. The system simultaneously produces SLO and OCT images that are created through the same optical path, It produces a new image format with high resolution tomographic images. 42
  • 43. R K o p t o A s O C T Clinical Uses Oct is used for diagnosis, monitoring, quantitative and qualitative assessment of macular pathology It is used in most of the Vitreo-retinal conditions 43
  • 44. R K o p t o A s O C T 7. Epiretinal Membrane in Diabetic Retinopathy. 8. Central Serous Retinopathy. 9. Foveal thining in Retinitis Pigmentosa and in Myopic Degeneration. 10.Age Related Macular Degeneration. 11.Post Retinal Detachment Surgery. 44 1. PosteriorVitreous Detachment (PVD) 2. Vitreo-MacularTraction Syndrome (VMT) 3. Asteriod Hyalosis 4. Intra-retinal Hemorrhage. 5. Macular Edema ( CME, CSME) 6. Macular Holes and cysts. U S E S
  • 45. R K o p t o A s O C T Interpretation Following things need to observed in OCT: A. Vitro-retinal interface. B. RetinalThickness: (I) Increases: Edema,Traction. (II) Decreases: Foveal atrophy. 45
  • 46. R K o p t o A s O C T C. Re fl ectivity: (I) Hyper-re fl ective lesions: Hard Exudates, Blood, Scars. (II) Hypo-re fl ective lesions: Serous Fluid, Hypo-pigmented lesions of Retinal Pigment Epithelium, Media Hazy. 46
  • 47. R K o p t o A s O C T Normal Highly re fl ective structures are shown in bright colures (white and red). Those with low re fl ectivity are represented by dark colors (black and blue). Intermediate re fl ectivity is shown Green. 47
  • 48. R K o p t o A s O C T Diabetic Macular Edema Sponge like retinal thickness. Macular edema and Cystoid Macular Edema. Serous Retinal Detachment. Epi-retinal Membrane. Vitro-macular traction. 48
  • 49. R K o p t o A s O C T 49
  • 50. R K o p t o A s O C T 50
  • 51. R K o p t o A s O C T 51
  • 52. R K o p t o A s O C T 52 RD
  • 53. R K o p t o A s O C T 53 ERM
  • 54. R K o p t o A s O C T ARMD Focal elevation of Retinal Pigment Epithelium. Large drusen shows irregular elevation of Retinal Pigment Epithelium with shadow from underlying choroid. Large drusen carry high risk of Choroidal Neo-Vascular Membrane. 54
  • 55. R K o p t o A s O C T 55
  • 56. R K o p t o A s O C T MACULAR HOLE Three are three type of macular hole we can see: 1. Impending Macular Hole. 2. Lamellar Macular Hole. 3. FullThickness Macular Hole. 56
  • 57. R K o p t o A s O C T 57
  • 58. R K o p t o A s O C T 58
  • 59. R K o p t o A s O C T 59
  • 60. R K o p t o A s O C T Disadvantage Clear ocular media is needed for better performance Minimum pupillary size of 4mm is required. It is a costly apparatus. 60
  • 61. R K o p t o A s O C T 61