SlideShare a Scribd company logo
1 of 8
Download to read offline
702 • CID 2004:39 (1 September) • HEALTHCARE EPIDEMIOLOGY
H E A L T H C A R E E P I D E M I O L O G Y I N V I T E D A R T I C L E
Robert A. Weinstein, Section Editor
Disinfection and Sterilization in Health Care Facilities:
What Clinicians Need to Know
William A. Rutala and David J. Weber
Hospital Epidemiology, University of North Carolina Health Care System, and Division of Infectious Diseases, University of North Carolina School of Medicine,
Chapel Hill
All invasive procedures involve contact between a medical device or surgical instrument and a patient’s sterile tissue or
mucous membranes. A major risk of all such procedures is the introduction of pathogenic microbes that could lead to
infection. Failure to properly disinfect or sterilize reusable medical equipment carries a risk associated with breach of the
host barriers. The level of disinfection or sterilization is dependent on the intended use of the object: critical items (such as
surgical instruments, which contact sterile tissue), semicritical items (such as endoscopes, which contact mucous membranes),
and noncritical items (such as stethoscopes, which contact only intact skin) require sterilization, high-level disinfection, and
low-level disinfection, respectively. Cleaning must always precede high-level disinfection and sterilization. Users must consider
the advantages and disadvantages of specific methods when choosing a disinfection or sterilization process. Adherence to
these recommendations should improve disinfection and sterilization practices in health care facilities, thereby reducing
infections associated with contaminated patient-care items.
In 1996 in the United States, ∼46,500,000 surgical procedures
and an even larger number of invasive medical procedures were
performed [1]. For example, ∼5 million gastrointestinal en-
doscopies are performed per year [1]. Each of these procedures
involves contact by a medical device or surgical instrument
with a patient’s sterile tissue or mucous membranes. A major
risk of all such procedures is the introduction of pathogenic
microbes, which can lead to infection. For example, failure to
properly disinfect or sterilize equipment may lead to person-
to-person transmission via contaminated devices (e.g., Myco-
bacterium tuberculosis–contaminated bronchoscopes).
Achieving disinfection and sterilization through the use of
disinfectants and sterilization practices is essential for ensuring
that medical and surgical instruments do not transmit infec-
tious pathogens to patients. Because it is not necessary to ster-
ilize all patient-care items, health care policies must identify
whether cleaning, disinfection, or sterilization is indicated, pri-
marily on the basis of each item’s intended use.
Multiple studies in many countries have documented lack
Received 15 March 2004; accepted 5 May 2004; electronically published 12 August 2004.
Reprints or correspondence: Dr. William A. Rutala, Div. of Infectious Diseases, 130 Mason
Farm Rd., Bioinformatics, University of North Carolina at Chapel Hill, Chapel Hill, NC 27599-
7030 (brutala@unch.unc.edu).
Clinical Infectious Diseases 2004;39:702–9
ᮊ 2004 by the Infectious Diseases Society of America. All rights reserved.
1058-4838/2004/3905-0015$15.00
of compliance with established guidelines for disinfection and
sterilization [2, 3]. Failure to comply with scientifically based
guidelines has led to numerous outbreaks of infection [3–7].
In this article, a pragmatic approach to the judicious selection
and proper use of disinfection and sterilization processes is
presented that is based on the results of well-designed studies
assessing the efficacy (via laboratory investigations) and effec-
tiveness (via clinical studies) of disinfection and sterilization
procedures.
A RATIONAL APPROACH TO DISINFECTION
AND STERILIZATION
More than 35 years ago, Spaulding [8] devised a rational ap-
proach to disinfection and sterilization of patient-care items or
equipment. This classification scheme is so clear and logical
that it has been retained, refined, and successfully used by in-
fection-control professionals and others when planning meth-
ods for disinfection or sterilization [9–15]. Spaulding believed
that the nature of disinfection could be understood more read-
ily if instruments and items for patient care were divided into
3 categories—namely, critical, semicritical, and noncritical—
on the basis of the degree of risk of infection involved in the
use of the items. This terminology is employed by the Centers
for Disease Control and Prevention (CDC) in the documents
“Guidelines for Environmental Infection Control in Health-
Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
HEALTHCARE EPIDEMIOLOGY • CID 2004:39 (1 September) • 703
Table 1. Methods for disinfection and sterilization of patient-care items and environmental surfaces.
Process, method Level of microbial inactivation Example(s) (processing time) Health care application (example)
Sterilization
High temperature Destroys all microorganisms,
including bacterial spores
Steam (∼40 min) and dry heat (1–6 h,
depending on temperature)
Heat-tolerant critical (surgical instru-
ments) and semicritical patient-care
items
Low temperature Destroys all microorganisms,
including bacterial spores
ETO gas (∼15 h) and hydrogen perox-
ide gas plasma (∼50 min)
Heat-sensitive critical and semicritical
patient-care items
Liquid immersion Destroys all microorganisms,
including bacterial spores
Chemical sterilants:
a
у2.4% glut (∼10
h), 1.12% glut and 1.93% phenol
(12 h), 7.35% HP and 0.23% PA (3
h), 7.5% HP (6 h), 1.0% HP and
0.08% PA (8 h), and у0.2% PA
(∼50 min at 50؇C–56؇C)
Heat-sensitive critical and semicritical
patient-care items that can be
immersed
High-level disinfection
Heat automated Destroys all microorganisms except
high numbers of bacterial spores
Pasteurization (∼50 min) Heat-sensitive semicritical patient-
care items (respiratory-therapy
equipment)
Liquid immersion Destroys all microorganisms except
high numbers of bacterial spores
Chemical sterilants or high-level disin-
fectants:
a
12% glut (20–45 min),
0.55% OPA (12 min), 1.12% glut
and 1.93% phenol (20 min), 7.35%
HP and 0.23% PA (15 min), 7.5%
HP (30 min), 1.0% HP and 0.08%
PA (25 min), and 650–675 ppm
chlorine (10 min)
Heat-sensitive semicritical patient-
care items (GI endoscopes and
bronchoscopes)
Intermediate-level disinfection,
liquid contact
Destroys vegetative bacteria, myco-
bacteria, most viruses, and most
fungi but not bacterial spores
EPA-registered hospital disinfectants
with label claiming tuberculocidal
activity, such as chlorine-based
products and phenolics (at least 60
s)
Noncritical patient-care items (blood-
pressure cuff) or surfaces (bedside
table), with visible blood
Low-level disinfection,
liquid contact
Destroys vegetative bacteria and
some fungi and viruses but not
mycobacteria or spores
EPA-registered hospital disinfectants
with no tuberculocidal claim, such
as chlorine-based products, phenol-
ics, and quaternary ammonium
compounds (at least 60 s), or 70%–
90% alcohol
Noncritical patient-care items (blood-
pressure cuff) or surfaces (bedside
table), with no visible blood
NOTE. Modified from [13], [14], and [17]. AER, automated endoscope reprocessing; EPA, Environmental Protection Agency; ETO, ethylene oxide; FDA,
US Food and Drug Administration; GI, gastrointestinal; glut, glutaraldehyde; HP, hydrogen peroxide; PA, peracetic acid; OPA, ortho-phthalaldehyde.
a
Consult FDA-cleared package inserts for information about FDA-cleared contact time and temperature; see text for discussion of why one product (2%
glut) is used at reduced exposure (20 min at 20؇C). Increasing the temperature by using AER will reduce the contact time (e.g., for OPA, 12 min at 20؇C, but
5 min at 25؇C in AER). Tubing must be completely filled for high-level disinfection and liquid chemical sterilization. Compatibility of material should be investigated
when appropriate (e.g., HP and HP with PA will cause functional damage to endoscopes).
Care Facilities” [16] and “Guideline for Disinfection and Ster-
ilization in Healthcare Facilities” [14].
Critical items. Critical items are those associated with a
high risk of infection if the item is contaminated with any
microorganism, including bacterial spores. Thus, sterilization
of objects that enter sterile tissue or the vascular system is
critical, because any microbial contamination could result in
disease transmission. This category includes surgical instru-
ments, cardiac and urinary catheters, implants, and ultrasound
probes used in sterile body cavities. The items in this category
should be purchased as sterile or should be sterilized by steam
sterilization, if possible. If the item is heat sensitive, it may be
treated with ethylene oxide (ETO) or hydrogen peroxide gas
plasma or with liquid chemical sterilants if other methods are
unsuitable. Tables 1 and 2 list several germicides that are cat-
egorized as chemical sterilants. These include у2.4% glutar-
aldehyde–based formulations, 1.12% glutaraldehyde with
1.93% phenol/phenate, 7.5% stabilized hydrogen peroxide,
7.35% hydrogen peroxide with 0.23% peracetic acid, у0.2%
peracetic acid, and 1.0% hydrogen peroxide with 0.08% per-
acetic acid. The indicated exposure times are within the range
3–12 h, with the exception of у0.2% peracetic acid (sporicidal
time of 12 min at 50ЊC–56ЊC) [19]. Use of liquid chemical
sterilants is a reliable method of sterilization only if cleaning
precedes treatment, which eliminates organic and inorganic
material, and if the proper guidelines for concentration, contact
time, temperature, and pH are followed. Another limitation to
sterilization of devices with liquid chemical sterilants is that the
devices cannot be wrapped during processing in the liquid
chemical sterilant; thus, maintaining sterility after processing
and during storage is impossible. Furthermore, after exposure
to the liquid chemical sterilant, devices may require rinsing
with water that, in general, is not sterile. Therefore, because of
the inherent limitations of the use of liquid chemical sterilants
Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
704 • CID 2004:39 (1 September) • HEALTHCARE EPIDEMIOLOGY
Table 2. Summary of advantages and disadvantages of chemical agents used as chemical sterilants or as high-level disinfectants.
Sterilization method Advantages Disadvantages
Peracetic acid and
hydrogen peroxide
No activation required
Odor or irritation not significant
Concerns regarding compatibility with materials (lead,
brass, copper, and zinc) and both cosmetic and
functional damage
Limited clinical use
Potential for eye and skin damage
Glutaraldehyde Numerous published studies of use
Relatively inexpensive
Excellent compatibility with materials
Respiratory irritation from glutaraldehyde vapor
Pungent and irritating odor
Relatively slow mycobactericidal activity
Coagulates blood and fixes tissue to surfaces
Allergic contact dermatitis
Hydrogen peroxide No activation required
May enhance removal of organic material and organisms
No disposal issues
No odor or irritation issues
Does not coagulate blood or fix tissues to surfaces
Inactivates Cryptosporidium
Published studies of use
Concerns regarding compatibility with materials (brass,
zinc, copper, and nickel/silver plating) and both
cosmetic and functional damage
Serious eye damage with contact
Ortho-phthalaldehyde Fast-acting high-level disinfectant
No activation required
Odor not significant
Claim of excellent compatibility with materials
Claim of not coagulating blood or fixing tissues to
surfaces
Stains protein gray (e.g., skin, mucous membranes,
clothing, and environmental surfaces)
Limited clinical use
More expensive than glutaraldehyde
Eye irritation with contact
Slow sporicidal activity
Repeated exposure may result in hypersensitivity in
some patients with bladder cancer
Peracetic acid Rapid sterilization cycle time (30–45 min)
Low-temperature (50؇C–55؇C) liquid-immersion
sterilization
Environmentally friendly by-products (acetic acid, O2, and
H2O)
Fully automated
Single-use system eliminates need for concentration
testing
Standardized cycle
May enhance removal of organic material and endotoxin
No adverse health effects to operators, under normal
operating conditions
Compatible with many materials and instruments
Does not coagulate blood or fix tissues to surfaces
Sterilant flows through scope, facilitating salt, protein,
and microbe removal
Rapidly sporicidal
Provides procedure standardization (constant dilution,
perfusion of channel, temperatures, and exposure)
Potential incompatibility with materials (e.g., aluminum
anodized coating becomes dull)
Used for immersible instruments only
Biological indicator may not be suitable for routine
monitoring
Only one scope or a small number of instruments can
be processed in a cycle
More expensive (endoscope repairs, operating costs, and
purchase costs) than high-level disinfection
Serious eye and skin damage (concentrated solution)
with contact
Point-of-use system; no sterile storage
NOTE. Modified from [18]. All products are effective in the presence of organic soil, are relatively easy to use, and have a broad spectrum of antimicrobial
activity (bacteria, fungi, viruses, bacterial spores, and mycobacteria). The above characteristics are documented in the literature; contact the manufacturer of the
instrument and sterilant for additional information. All products listed have been cleared by the US Food and Drug Administration (FDA) as chemical sterilants,
except for ortho-phthalaldehyde, which is an FDA-cleared high-level disinfectant.
in a nonautomated reprocessor, their use should be restricted
to reprocessing critical devices that are heat sensitive and in-
compatible with other sterilization methods.
Semicritical items. Semicritical items are those that come
in contact with mucous membranes or nonintact skin. Respi-
ratory-therapy and anesthesia equipment, some endoscopes,
laryngoscope blades, esophageal manometry probes, anorectal
manometry catheters, and diaphragm-fitting rings are included
in this category. These medical devices should be free of all
microorganisms (i.e., mycobacteria, fungi, viruses, and bacte-
ria), although small numbers of bacterial spores may be present.
In general, intact mucous membranes, such as those of the
lungs or the gastrointestinal tract, are resistant to infection by
common bacterial spores but are susceptible to other organ-
isms, such as bacteria, mycobacteria, and viruses. The mini-
mum requirement for semicritical items is high-level disinfec-
tion using chemical disinfectants. Glutaraldehyde, hydrogen
peroxide, ortho-phthalaldehyde (OPA), peracetic acid with hy-
drogen peroxide, and chlorine have been cleared by the US
Food and Drug Administration (FDA) [19] and are dependable
high-level disinfectants when guidelines for effective germicidal
procedures are followed (tables 1 and 2). The exposure time
for most high-level disinfectants varies from 10 to 45 min, at
20ЊC–25ЊC. Outbreaks of infection continue to occur when
ineffective disinfectants, including iodophor, alcohol, and over-
diluted glutaraldehyde [5], are used for so-called high-level
Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
HEALTHCARE EPIDEMIOLOGY • CID 2004:39 (1 September) • 705
disinfection. When a disinfectant is selected for use with certain
patient-care items, the chemical compatibility after extended
use with the items to be disinfected must also be considered.
For example, compatibility testing by Olympus America of
7.5% hydrogen peroxide showed cosmetic and functional
changes in the tested endoscopes (Olympus America, personal
communication). Similarly, Olympus America does not endorse
the use of products containing hydrogen peroxide with pera-
cetic acid, because of cosmetic and functional damage (Olym-
pus America, personal communication).
Semicritical items that will have contact with the mucous
membranes of the respiratory or gastrointestinal tract should
be rinsed with sterile water, filtered water, or tap water, followed
by an alcohol rinse [14, 20, 21]. An alcohol rinse and forced-
air drying markedly reduces the likelihood of contamination
of the instrument (e.g., endoscopes), most likely by eliminating
the wet environment favorable to bacterial growth [21]. After
rinsing, items should be dried and then stored in a manner
that protects them from damage or contamination. There is
no recommendation to use sterile or filtered water, rather than
tap water, for rinsing semicritical equipment that will have
contact with the mucous membranes of the rectum (e.g., rectal
probes or anoscopes) or vagina (e.g., vaginal probes) [14].
Noncritical items. Noncritical items are those that come
in contact with intact skin but not mucous membranes. Intact
skin acts as an effective barrier to most microorganisms; there-
fore, the sterility of items coming in contact with intact skin
is “not critical.” Examples of noncritical items are bedpans,
blood-pressure cuffs, crutches, bed rails, linens, bedside tables,
patient furniture, and floors. In contrast to critical and some
semicritical items, most noncritical reusable items may be de-
contaminated where they are used and do not need to be trans-
ported to a central processing area. There is virtually no doc-
umented risk of transmitting infectious agents to patients via
noncritical items [22] when they are used as noncritical items
and do not contact nonintact skin and/or mucous membranes.
However, these items (e.g., bedside tables or bed rails) could
potentially contribute to secondary transmission, by contam-
inating the hands of health care workers or by contact with
medical equipment that will subsequently come in contact with
patients [23]. Table 1 lists several low-level disinfectants that
may be used for noncritical items. The exposure times for these
disinfectants are 60 s or longer.
CURRENT ISSUES IN DISINFECTION
AND STERILIZATION
Reprocessing of endoscopes. Physicians use endoscopes to
diagnose and treat numerous medical disorders. Although en-
doscopes are a valuable diagnostic and therapeutic tool in mod-
ern medicine and although the incidence of infection associated
with their use has been reported to be very low (∼1 in 1.8
million procedures) [24], more health care–associated out-
breaks of infection have been linked to contaminated endo-
scopes than to any other medical device [3–5]. To prevent the
spread of health care–associated infection, all heat-sensitive en-
doscopes (e.g., gastrointestinal endoscopes, bronchoscopes, and
nasopharyngoscopes) must be properly cleaned and, at a min-
imum, subjected to high-level disinfection after each use. High-
level disinfection can be expected to destroy all microorganisms,
although a few bacterial spores may survive when high numbers
of spores are present.
Recommendations for the cleaning and disinfection of en-
doscopic equipment have been published and should be strictly
followed [14, 20]. Unfortunately, audits have shown that per-
sonnel do not adhere to guidelines on reprocessing [25–27]
and that outbreaks of infection continue to occur [28, 29]. To
ensure that the personnel responsible for reprocessing are prop-
erly trained, initial and annual competency testing should be
required for each individual who is involved in reprocessing
endoscopic instruments [14, 20, 21, 30].
In general, endoscope disinfection or sterilization with a liq-
uid chemical sterilant or high-level disinfectant involves the
following 5 steps, which should be performed after leak testing:
(1) clean: mechanically clean internal and external surfaces,
including brushing internal channels and flushing each internal
channel with water and an enzymatic cleaner; (2) disinfect:
immerse endoscope in high-level disinfectant (or chemical ster-
ilant), perfuse disinfectant (which eliminates air pockets and
ensures contact of the germicide with the internal channels)
into all accessible channels, such as the suction/biopsy channel
and the air/water channel, and expose endoscope for the time
recommended for specific products; (3) rinse: rinse the en-
doscope and all channels with sterile water, filtered water (com-
monly used with automated endoscope reprocessors), or tap
water; (4) dry: rinse the insertion tube and inner channels with
alcohol and dry with forced air, after disinfection and before
storage; and (5) store: store the endoscope in a way that
prevents recontamination and promotes drying (e.g., hung
vertically).
Unfortunately, there is poor compliance with the recom-
mendations for reprocessing endoscopes. In addition, in rare
instances, the scientific literature and recommendations from
professional organizations regarding the use of disinfectants
and sterilants may differ from claims on the manufacturer’s
label. One example is the contact time used to achieve high-
level disinfection with 2% glutaraldehyde. On the basis of FDA
requirements (the FDA regulates liquid sterilants and high-level
disinfectants used on critical and semicritical medical devices),
manufacturers test the efficacy of their germicide formulations
under worst-case conditions (i.e., minimum recommended
concentration of the active ingredient) and in the presence of
organic soil (typically, 5% serum). The soil represents the or-
Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
706 • CID 2004:39 (1 September) • HEALTHCARE EPIDEMIOLOGY
ganic loading to which the device is exposed during actual use
and that would remain on the device in the absence of cleaning.
These stringent test conditions are designed to provide a margin
of safety, by assuring that the contact conditions for the ger-
micide provide complete elimination of the test bacteria (e.g.,
105
–106
cfu M. tuberculosis in organic soil and dried on a scope)
if inoculated into the most difficult areas for the disinfectant
to penetrate and in the absence of cleaning. However, the sci-
entific data demonstrate that M. tuberculosis levels can be re-
duced by at least 8 log10 cfu with cleaning (reduction of 4 log10
cfu) followed by chemical disinfection for 20 min at 20ЊC (re-
duction of 4–6 log10 cfu) [14, 15, 19, 20, 31]. Because of these
data, professional organizations (at least 14 worldwide) that
have endorsed an endoscope-reprocessing guideline recom-
mend contact with 2% glutaraldehyde for 20 min (or !20 min
outside the United States) at 20ЊC to achieve high-level dis-
infection, which differs from the recommendation given on the
manufacturer’s label [20, 32–34].
It is important to emphasize that the FDA tests do not in-
clude cleaning, a critical component of the disinfection process.
When cleaning has been included in the test methodology,
contact with 2% glutaraldehyde for 20 min has been demon-
strated to be effective in eliminating all vegetative bacteria.
Inactivation of Creutzfeldt-Jakob disease (CJD) agent.
CJD is a degenerative neurologic disorder in humans, with an
incidence in the United States of ∼1 case/million population/
year [35]. CJD is thought to be caused by a proteinaceous
infectious agent, or prion. CJD is related to other human trans-
missible spongiform encephalopathies (TSEs), such as kuru
(now eradicated), Gertsmann-Straussler-Sheinker syndrome (1
case/40 million population/year), and fatal insomnia syndrome
(!1 case/40 million population/year). The agents of CJD and
other TSEs exhibit an unusual resistance to conventional chem-
ical and physical decontamination methods. Because the CJD
agent is not readily inactivated by conventional disinfection
and sterilization procedures and because of the invariably fatal
outcome of CJD, the procedures for disinfection and sterili-
zation of the CJD prion have been both conservative and con-
troversial for many years.
The current recommendations consider inactivation data but
also use epidemiological studies of prion transmission, infec-
tivity of human tissues, and efficacy of removing proteins by
cleaning. On the basis of scientific data, only critical devices
(e.g., surgical instruments) and semicritical devices contami-
nated with high-risk tissue (i.e., brain, spinal cord, or eye tissue)
from high-risk patients (e.g., known or suspected infection with
CJD or other prion disease) require special prion reprocessing.
When high-risk tissues, high-risk patients, and critical or sem-
icritical medical devices are involved, one of the following
methods should be used: cleaning of the device and sterilization
using a combination of sodium hydroxide and autoclaving [36]
(e.g., immerse in 1N NaOH for 1 h, remove and rinse in water,
and then transfer to an open pan for autoclaving for 1 h [at
121ЊC in a gravity displacement sterilizer or at 134ЊC in a
porous or prevacuum sterilizer]); autoclaving for 18 min at
134ЊC in a prevacuum sterilizer; or autoclaving for 1 h at 132ЊC
in a gravity displacement sterilizer) [14, 37]. The temperature
should not exceed 134ЊC, because the effectiveness of auto-
claving may decline as the temperature is increased (e.g., to
136ЊC or 138ЊC) [38]. Prion-contaminated medical devices that
are impossible or difficult to clean should be discarded. Flash
sterilization (i.e., steam sterilization of an unwrapped item for
3 min at 132ЊC) should not be used for reprocessing. To min-
imize environmental contamination, noncritical environmental
surfaces should be covered with plastic-backed paper; when
contaminated with high-risk tissues, the paper should be prop-
erly discarded. Noncritical environmental surfaces (e.g., labo-
ratory surfaces) contaminated with high-risk tissues should be
cleaned and then spot decontaminated with a 1:10 dilution of
hypochlorite solution [37].
Emerging pathogens, antibiotic-resistant bacteria, and bio-
terrorism agents. Emerging pathogens are of growing con-
cern to the general public and infection-control professionals.
Relevant pathogens include Cryptosporidium parvum, Helico-
bacter pylori, Escherichia coli O157:H7, HIV, hepatitis C virus,
rotavirus, multidrug-resistant M. tuberculosis, human papillo-
mavirus, and nontuberculosis mycobacteria (e.g., Mycobacte-
rium chelonae). Similarly, recent publications have highlighted
concern about the potential for biological terrorism [39]. The
CDC has categorized several agents as “high priority” because
they can be easily disseminated or transmitted by person-to-
person contact, can cause high mortality, and are likely to cause
public panic and social disruption [40]. These agents include
Bacillus anthracis (anthrax), Yersinia pestis (plague), variola
major (smallpox), Francisella tularensis (tularemia), filoviruses
(Ebola and Marburg [hemorrhagic fever]), and arenaviruses
(Lassa [Lassa fever] and Junin [Argentine hemorrhagic fever])
and related viruses [40].
With rare exceptions (e.g., human papillomavirus), the sus-
ceptibility of each of these pathogens to chemical disinfectants
or sterilants has been studied, and all of these pathogens (or
surrogate microbes, such as feline calicivirus for Norwalk virus,
vaccinia for variola [41], and Bacillus atrophaeus [formerly Ba-
cillus subtilis] for B. anthracis) have been found to be susceptible
to currently available chemical disinfectants or sterilants [42].
Standard sterilization and disinfection procedures for patient-
care equipment (as recommended in this article) are adequate
for sterilization or disinfection of instruments or devices con-
taminated with blood or other body fluids from persons in-
fected with bloodborne pathogens, emerging pathogens, or
bioterrorism agents, with the exception of prions (see previous
section). No changes in procedures for cleaning, disinfecting,
Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
HEALTHCARE EPIDEMIOLOGY • CID 2004:39 (1 September) • 707
Table 3. Summary of advantages and disadvantages of commonly used sterilization technologies.
Sterilization method Advantages Disadvantages
Steam Nontoxic to patient, staff, and environment
Cycle is easy to control and monitor
Rapidly microbicidal
Least affected by organic/inorganic soils, among steriliza-
tion processes listed
Rapid cycle time
Penetrates medical packing and device lumens
Deleterious for heat-sensitive instruments
Microsurgical instruments damaged by repeated
exposure
May leave instruments wet, causing them to rust
Potential for burns
Hydrogen peroxide gas plasma Safe for the environment
Leaves no toxic residuals
Cycle time is 45–73 min, and no aeration is necessary
Used for heat- and moisture-sensitive items, because
process temperature is !50؇C
Equipment is simple to operate, install (208 V outlet),
and monitor
Compatible with most medical devices
Equipment requires electrical outlet only
Cellulose (paper), linens, and liquids cannot be processed
Sterilization chamber is small (∼3.5–7.3 ft3
)
Endoscope or medical-device restrictions based on lu-
men internal diameter and length (see manufacturer’s
recommendations)
Requires synthetic packaging (polypropylene wraps or
polyolefin pouches) or special container tray
Hydrogen peroxide may be toxic at levels 11 ppm TWA
100% ETO Penetrates packaging materials and device lumens
Single-dose cartridge and negative-pressure chamber
minimizes the potential for gas leak and ETO exposure
Equipment is simple to operate and monitor
Compatible with most medical materials
Requires aeration time to remove ETO residue
Sterilization chamber is small (∼4–8.8 ft3
)
ETO is toxic, a carcinogen, and flammable
ETO emission regulated by states, but catalytic cell re-
moves 99.9% of ETO and converts it to CO2 and H2O
ETO cartridges should be stored in flammable liquid–
storage cabinet
Lengthy cycle and aeration time
ETO mixture
a
Penetrates medical packaging and many plastics
Compatible with most medical materials
Cycle is easy to control and monitor
Some states (e.g., CA, NY, and MI) require ETO-emission
reduction of 90%–99.9%
CFC (inert gas that eliminates explosion hazard) banned
in 1995
Potential hazards to staff and patients
Lengthy cycle and aeration time
ETO is toxic, a carcinogen, and flammable
Peracetic acid Rapid cycle time (30–45 min)
Low-temperature (50؇C–55؇C) liquid-immersion
sterilization
Environmentally friendly by-products
Sterilant flows through endoscope, which facilitates salt,
protein, and microbe removal
Point-of-use system; no sterile storage
Biological indicator may not be suitable for routine
monitoring
Used for immersible instruments only
Some incompatibility with materials (e.g., aluminum
anodized coating becomes dull)
Only 1 scope or a small number of instruments
processed in a cycle
Potential for serious eye and skin damage (concentrated
solution) with contact
Must use connector between system and scope to
ensure infusion of sterilant to all channels
NOTE. Modified from [46]. CFC, chlorofluorocarbon; ETO, ethylene oxide; HCFC, hydrochlorofluorocarbon; TWA, time-weighted average.
a
8.6% ETO and 91.4% HCFC; 10% ETO and 90% HCFC; or 8.5% ETO and 91.5% CO2.
or sterilizing need to be made [14, 15]. In addition, there are
no data to show that antibiotic-resistant bacteria (e.g., meth-
icillin-resistant Staphylococcus aureus, vancomycin-resistantEn-
terococcus faecium, and multidrug-resistant M. tuberculosis) are
less sensitive to liquid chemical germicides than are antibiotic-
sensitive bacteria at currently used germicide contact conditions
and concentrations [15, 43, 44].
Advances in disinfection and sterilization methods. In
the past several years, new methods of disinfection and ster-
ilization have been introduced in health care settings. OPA is
a chemical sterilant that received FDA clearance in October
1999. It contains 0.55% 1,2-benzenedicarboxaldehyde. In vitro
studies have demonstrated excellent microbicidal activity [14,
15]. For example, Gregory et al. [45] demonstrated that OPA
has shown superior mycobactericidal activity (reduction of 5
log10 in 5 min), when compared with glutaraldehyde. The ad-
vantages, disadvantages, and characteristics of OPA are listed
in table 2 [15].
The FDA recently cleared a liquid high-level disinfectant (su-
peroxidized water) that contains 650–675 ppm free chlorine
and a new sterilization system using ozone. Because there are
limited data in the scientific literature for assessing the anti-
microbial activity or material compatibility of these processes,
they have not yet been integrated into clinical practice in the
United States [14].
Several methods are used to sterilize patient-care items in
health care, including steam sterilization, ETO, hydrogen per-
oxide gas plasma, and a peracetic acid–immersion system. The
advantages and disadvantages of these systems are listed in table
3 [14].
Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
708 • CID 2004:39 (1 September) • HEALTHCARE EPIDEMIOLOGY
New sterilization technology based on plasma was patented
in 1987 and has been marketed in the United States since 1993.
Gas plasmas have been referred to as the fourth state of matter
(i.e., liquid, solid, gas, and gas plasma). Gas plasmas are gen-
erated in an enclosed chamber in a deep vacuum, by using
radio frequency or microwave energy to excite the gas molecules
and produce charged particles, many of which are in the form
of free radicals. This process has the ability to inactivate a broad
spectrum of microorganisms, including resistant bacterial
spores. Studies have been conducted against vegetative bacteria
(including mycobacteria), yeasts, fungi, viruses, and bacterial
spores [14]. The effectiveness of all sterilization processes can
be altered by lumen length, lumen diameter, inorganic salts,
and organic materials [14].
CONCLUSION
When properly used, disinfection and sterilization can ensure
the safe use of invasive and noninvasive medical devices. The
method of disinfection and sterilization depends on the in-
tended use of the medical device: critical items (those that
contact sterile tissue) must be sterilized prior to use; semicritical
items (those that contact mucous membranes or nonintact
skin) must undergo high-level disinfection; and noncritical
items (those that contact intact skin) should undergo low-level
disinfection. Cleaning should always precede high-level disin-
fection and sterilization. Current disinfection and sterilization
guidelines must be strictly followed.
Acknowledgment
Conflict of interest. W.A.R.: Honoraria from Advanced Sterilization
Products; consultant for Advanced Sterilization Products, Clorox, 3M, and
Metrex; and research funding from Clorox and Reckitt-Benckiser. D.J.B.:
Honoraria from Clorox.
References
1. Centers for Disease Control and Prevention (CDC). Ambulatory and
inpatient procedures in the United States, 1996. Atlanta: CDC, 1998:
1–39.
2. McCarthy GM, Koval JJ, John MA, MacDonald JK. Infection control
practices across Canada: do dentists follow the recommendations? J
Can Dent Assoc 1999;65:506–11.
3. Spach DH, Silverstein FE, Stamm WE. Transmission of infection by
gastrointestinal endoscopy and bronchoscopy. Ann Intern Med
1993;118:117–28.
4. Weber DJ, Rutala WA. Lessons from outbreaks associated with bron-
choscopy. Infect Control Hosp Epidemiol 2001;22:403–8.
5. Weber DJ, Rutala WA, DiMarino AJ Jr. The prevention of infection
following gastrointestinal endoscopy: the importance of prophylaxis
and reprocessing. In: DiMarino AJ Jr, Benjamin SB, eds. Gastrointes-
tinal diseases: an endoscopic approach. Thorofare, NJ: Slack, 2002:
87–106.
6. Meyers H, Brown-Elliott BA, Moore D, et al. An outbreak of Myco-
bacterium chelonae infection following liposuction. Clin Infect Dis
2002;34:1500–7.
7. Lowry PW, Jarvis WR, Oberle AD, et al. Mycobacterium chelonae caus-
ing otitis media in an ear-nose-and-throat practice. N Engl J Med
1988;319:978–82.
8. Spaulding EH. Chemical disinfection of medical and surgical materials.
In: Lawrence C, Block SS, eds. Disinfection, sterilization, and preser-
vation. Philadelphia: Lea & Febiger, 1968:517–31.
9. Favero MS, Bond WW. Chemical disinfection of medical and surgical
materials. In: Block SS, ed. Disinfection, sterilization, and preservation.
Philadelphia: Lippincott, Williams & Wilkins, 2001:881–917.
10. Simmons BP. CDC guidelines for the prevention and control of nos-
ocomial infections: guideline for hospital environmental control. Am
J Infect Control 1983;11:97–120.
11. Garner JS, Favero MS. CDC guideline for handwashing and hospital
environmental control, 1985. Infect Control 1986;7:231–43.
12. Rutala WA. APIC guideline for selection and use of disinfectants. Am
J Infect Control 1990;18:99–117.
13. Rutala WA. APIC guideline for selection and use of disinfectants. As-
sociation for Professionals in Infection Control and Epidemiology
Guidelines Committee of 1994, 1995, and 1996. Am J Infect Control
1996;24:313–42.
14. Rutala WA, Weber DJ. Guideline for disinfection and sterilization in
healthcare facilities: recommendations of the CDC. Healthcare Infec-
tion Control Practices Advisory Committee. MMWR Morb Mortal
Wkly Rep (in press).
15. Rutala WA, Weber DJ. Selection and use of disinfectants in healthcare.
In: Mayhall CG, ed. Hospital epidemiology and infection control. Phil-
adelphia: Lippincott, Williams & Wilkins, 2004:1473–522.
16. Sehulster L, Chinn RYW. Guidelines for environmental infection con-
trol in health-care facilities. Healthcare Infection Control Practices Ad-
visory Committee. MMWR Recomm Rep 2003;52:1–44.
17. Kohn WG, Collins AS, Cleveland JL, Harte JA, Eklund KJ, Malvitz
DM. Guidelines for infection control in dental health-care settings—
2003. MMWR Recomm Rep 2003;52:1–67.
18. Rutala WA, Weber DJ. Disinfection of endoscopes: review of new chem-
ical sterilants used for high-level disinfection. Infect Control Hosp
Epidemiol 1999;20:69–76.
19. Food and Drug Administration (FDA). FDA-cleared sterilants and high
level disinfectants with general claims for processing reusable medical
and dental devices, November 2003. Available at: http://www.fda.gov/
cdrh/ode/germlab.html. Accessed 8 August 2004.
20. Nelson DB, Jarvis WR, Rutala WA, et al. Multi-society guideline for
reprocessing flexible gastrointestinal endoscopes. Infect Control Hosp
Epidemiol 2003;24:532–7.
21. Gerding DN, Peterson LR, Vennes JA. Cleaning and disinfection of
fiberoptic endoscopes: evaluation of glutaraldehyde exposure time and
forced-air drying. Gastroenterology 1982;83:613–8.
22. Weber DJ, Rutala WA. Environmental issues and nosocomialinfections.
In: Wenzel RP, ed. Prevention and control of nosocomial infections.
Baltimore: Williams & Wilkins, 1997:491–514.
23. Weber DJ, Rutala WA. Role of environmental contamination in the
transmission of vancomycin-resistant enterococci. Infect Control Hosp
Epidemiol 1997;18:306–9.
24. Schembre DB. Infectious complications associated with gastrointestinal
endoscopy. Gastrointest Endosc Clin N Am 2000;10:215–32.
25. Jackson FW, Ball MD. Correction of deficiencies in flexible fiberoptic
sigmoidoscope cleaning and disinfection technique in family practice
and internal medicine offices. Arch Fam Med 1997;6:578–82.
26. Orsi GB, Filocamo A, Di Stefano L, Tittobello A. Italian national survey
of digestive endoscopy disinfection procedures. Endoscopy 1997;29:
732–8; quiz:739–40.
27. Honeybourne D, Neumann CS. An audit of bronchoscopy practice in
the United Kingdom: a survey of adherence to national guidelines.
Thorax 1997;52:709–13.
28. Srinivasan A, Wolfenden LL, Song X, et al. An outbreak of Pseudomonas
aeruginosa infections associated with flexible bronchoscopes. N Engl J
Med 2003;348:221–7.
Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
HEALTHCARE EPIDEMIOLOGY • CID 2004:39 (1 September) • 709
29. Cetse JC, Vanhems P. Outbreak of infection associated with broncho-
scopes. N Engl J Med 2003;348:2039–40.
30. Food and Drug Administration (FDA), Centers for Disease Control
and Prevention (CDC). FDA and CDC public health advisory: infec-
tions from endoscopes inadequately reprocessed by an automated en-
doscope reprocessing system. Rockville, MD: FDA, 1999.
31. Rutala WA, Weber DJ. FDA labeling requirements for disinfection of
endoscopes: a counterpoint. Infect Control Hosp Epidemiol 1995;16:
231–5.
32. Kruse A, Rey JF. Guidelines on cleaning and disinfection in GI en-
doscopy: update 1999. The European Society of Gastrointestinal En-
doscopy. Endoscopy 2000;32:77–80.
33. British Society of Gastroenterology. Cleaning and disinfection of equip-
ment for gastrointestinal endoscopy: report of a working party of the
British Society of Gastroenterology Endoscope Committee. Gut
1998;42:585–93.
34. British Thoracic Society. British Thoracic Society guidelines on diag-
nostic flexible bronchoscopy. Thorax 2001;56:1–21.
35. Centers for Disease Control and Prevention. Surveillance for Creutz-
feldt-Jakob disease—United States. MMWR Morb Mortal Wkly Rep
1996;45:665–8.
36. World Health Organization (WHO). WHO infection control guidelines
for transmissible spongiform encephalopathies. Available at: http:
//www.who.int/emc-documents/tse/docs/whocdscsraph2003.pdf. Ac-
cessed 9 August 2004.
37. Rutala WA, Weber DJ. Creutzfeldt-Jakob disease: recommendationsfor
disinfection and sterilization. Clin Infect Dis 2001;32:1348–56.
38. Taylor DM. Inactivation of prions by physical and chemical means. J
Hosp Infect 1999;43(Suppl):S69–76.
39. Henderson DA. The looming threat of bioterrorism. Science 1999;283:
1279–82.
40. Centers for Disease Control and Prevention. Biological and chemical
terrorism: strategic plan for preparedness and response. MMWR Re-
comm Rep 2000;49:1–14.
41. Klein M, DeForest A. The inactivation of viruses by germicides. Chem
Specialists Manuf Assoc Proc 1963;49:116–8.
42. Rutala WA, Weber DJ. Infection control: the role of disinfection and
sterilization. J Hosp Infect 1999;43(Suppl):S43–55.
43. Rutala WA, Stiegel MM, Sarubbi FA, Weber DJ. Susceptibility of an-
tibiotic-susceptible and antibiotic-resistant hospital bacteria to disin-
fectants. Infect Control Hosp Epidemiol 1997;18:417–21.
44. Weber DJ, Rutala WA. Use of germicides in the home and health care
setting: is there a relationship between germicide use and antimicrobial
resistance? Infect Control Hosp Epidemiol (in press).
45. Gregory AW, Schaalje GB, Smart JD, Robison RA. The mycobactericidal
efficacy of ortho-phthalaldehyde and the comparative resistances of
Mycobacterium bovis, Mycobacterium terrae, and Mycobacterium che-
lonae. Infect Control Hosp Epidemiol 1999;20:324–30.
46. Rutala WA, Weber DJ. Clinical effectiveness of low-temperature
sterilization technologies. Infect Control Hosp Epidemiol 1998;19:
798–804.
Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020

More Related Content

What's hot

Infection control protocol
Infection control protocolInfection control protocol
Infection control protocolAbhay Rajpoot
 
Infection control and safety measurs
Infection control and safety measursInfection control and safety measurs
Infection control and safety measursdeepak patel
 
Pk pd analysis and mic interpretation in microbiological reports
Pk pd analysis and mic interpretation in microbiological reportsPk pd analysis and mic interpretation in microbiological reports
Pk pd analysis and mic interpretation in microbiological reportsCentral Govt, India
 
Hospital Infection Control Programme
Hospital Infection Control ProgrammeHospital Infection Control Programme
Hospital Infection Control ProgrammeNursing Path
 
Health care associated infection mitesh
Health care associated infection  miteshHealth care associated infection  mitesh
Health care associated infection miteshmitesh panchal
 
Infection Control
Infection ControlInfection Control
Infection ControlDiaa Swahly
 
Principles and practices in hospital infection control
Principles and practices in hospital infection controlPrinciples and practices in hospital infection control
Principles and practices in hospital infection controlCentral Govt, India
 
Hospital Aquired Infections and infection control in a healthcare setup
Hospital Aquired Infections and infection control in a healthcare setupHospital Aquired Infections and infection control in a healthcare setup
Hospital Aquired Infections and infection control in a healthcare setupSumi Nandwani
 
Infection control practice and standard safety measures
Infection control practice and standard safety measuresInfection control practice and standard safety measures
Infection control practice and standard safety measuresVANITASharma19
 
The history and development of medical laboratory science profession in nigeria
The history and development of medical laboratory science profession in nigeriaThe history and development of medical laboratory science profession in nigeria
The history and development of medical laboratory science profession in nigeriaJennifer Giovanna
 
Infection control protocol and its importance
Infection control protocol and its importanceInfection control protocol and its importance
Infection control protocol and its importanceShehzad Hussain Raja
 
Ppt hospital infection control for small scale hospitals
Ppt hospital infection control for small scale hospitalsPpt hospital infection control for small scale hospitals
Ppt hospital infection control for small scale hospitalsDrNeha Sharma
 
The importance of infection control in patient care
The importance of infection control in patient careThe importance of infection control in patient care
The importance of infection control in patient careMEEQAT HOSPITAL
 
Acs0102 Infection Control In Surgical Practice
Acs0102 Infection Control In Surgical PracticeAcs0102 Infection Control In Surgical Practice
Acs0102 Infection Control In Surgical Practicemedbookonline
 
Infection control protocol in icu
Infection control protocol in icuInfection control protocol in icu
Infection control protocol in icuANJANI WALIA
 
Infection prevention and safety measures
Infection prevention and safety measuresInfection prevention and safety measures
Infection prevention and safety measuresfrank jc
 
Infection control program Concept
Infection control program ConceptInfection control program Concept
Infection control program ConceptTauseef Jawaid
 
INFECTION CONTROL PROGRAMME IN HEALTH CARE FACCILITY
 INFECTION CONTROL PROGRAMME IN HEALTH CARE FACCILITY INFECTION CONTROL PROGRAMME IN HEALTH CARE FACCILITY
INFECTION CONTROL PROGRAMME IN HEALTH CARE FACCILITYTauseef Jawaid
 

What's hot (19)

Infection control protocol
Infection control protocolInfection control protocol
Infection control protocol
 
Infection control and safety measurs
Infection control and safety measursInfection control and safety measurs
Infection control and safety measurs
 
Pk pd analysis and mic interpretation in microbiological reports
Pk pd analysis and mic interpretation in microbiological reportsPk pd analysis and mic interpretation in microbiological reports
Pk pd analysis and mic interpretation in microbiological reports
 
Hospital Infection Control Programme
Hospital Infection Control ProgrammeHospital Infection Control Programme
Hospital Infection Control Programme
 
Health care associated infection mitesh
Health care associated infection  miteshHealth care associated infection  mitesh
Health care associated infection mitesh
 
Infection Control
Infection ControlInfection Control
Infection Control
 
Principles and practices in hospital infection control
Principles and practices in hospital infection controlPrinciples and practices in hospital infection control
Principles and practices in hospital infection control
 
Hospital Aquired Infections and infection control in a healthcare setup
Hospital Aquired Infections and infection control in a healthcare setupHospital Aquired Infections and infection control in a healthcare setup
Hospital Aquired Infections and infection control in a healthcare setup
 
Infection control practice and standard safety measures
Infection control practice and standard safety measuresInfection control practice and standard safety measures
Infection control practice and standard safety measures
 
Surveillance of HAI
Surveillance of HAISurveillance of HAI
Surveillance of HAI
 
The history and development of medical laboratory science profession in nigeria
The history and development of medical laboratory science profession in nigeriaThe history and development of medical laboratory science profession in nigeria
The history and development of medical laboratory science profession in nigeria
 
Infection control protocol and its importance
Infection control protocol and its importanceInfection control protocol and its importance
Infection control protocol and its importance
 
Ppt hospital infection control for small scale hospitals
Ppt hospital infection control for small scale hospitalsPpt hospital infection control for small scale hospitals
Ppt hospital infection control for small scale hospitals
 
The importance of infection control in patient care
The importance of infection control in patient careThe importance of infection control in patient care
The importance of infection control in patient care
 
Acs0102 Infection Control In Surgical Practice
Acs0102 Infection Control In Surgical PracticeAcs0102 Infection Control In Surgical Practice
Acs0102 Infection Control In Surgical Practice
 
Infection control protocol in icu
Infection control protocol in icuInfection control protocol in icu
Infection control protocol in icu
 
Infection prevention and safety measures
Infection prevention and safety measuresInfection prevention and safety measures
Infection prevention and safety measures
 
Infection control program Concept
Infection control program ConceptInfection control program Concept
Infection control program Concept
 
INFECTION CONTROL PROGRAMME IN HEALTH CARE FACCILITY
 INFECTION CONTROL PROGRAMME IN HEALTH CARE FACCILITY INFECTION CONTROL PROGRAMME IN HEALTH CARE FACCILITY
INFECTION CONTROL PROGRAMME IN HEALTH CARE FACCILITY
 

Similar to Disinfection & sterilization journal

Cdc guideline for disinfection and sterilization in healthcare facilities, 2008
Cdc guideline for disinfection and sterilization in healthcare facilities, 2008Cdc guideline for disinfection and sterilization in healthcare facilities, 2008
Cdc guideline for disinfection and sterilization in healthcare facilities, 2008Enrique Guillen
 
Guideline for disinfection and sterilization in healthcare facilities, 2008[1]
Guideline for disinfection and sterilization in healthcare facilities, 2008[1]Guideline for disinfection and sterilization in healthcare facilities, 2008[1]
Guideline for disinfection and sterilization in healthcare facilities, 2008[1]Manel Ferreira
 
stakeholders in IPC.pptx
stakeholders in IPC.pptxstakeholders in IPC.pptx
stakeholders in IPC.pptxNehaPandey199
 
Infection control in community setting
Infection control in community settingInfection control in community setting
Infection control in community settingKaushal Goti
 
Inf control for hcw 2012
Inf control for hcw 2012Inf control for hcw 2012
Inf control for hcw 2012Lee Oi Wah
 
HAIs std precaution , safe injection.pptx
HAIs std precaution , safe injection.pptxHAIs std precaution , safe injection.pptx
HAIs std precaution , safe injection.pptxabdalla ibrahim
 
Decontamination of Medical Devices.docx
Decontamination of Medical Devices.docxDecontamination of Medical Devices.docx
Decontamination of Medical Devices.docxsdfghj21
 
Biosafety and waste management in histopathology
Biosafety and waste management in histopathologyBiosafety and waste management in histopathology
Biosafety and waste management in histopathologyJennifer Giovanna
 
Basic principles of endodontics
Basic principles of endodonticsBasic principles of endodontics
Basic principles of endodonticsTess Boto
 
17 421 pds infection control
17 421 pds infection control17 421 pds infection control
17 421 pds infection controlKalpana Gokul
 
Epcm l15 control of nosocomial (Hospital-Acquired) infection
Epcm l15 control of nosocomial (Hospital-Acquired) infectionEpcm l15 control of nosocomial (Hospital-Acquired) infection
Epcm l15 control of nosocomial (Hospital-Acquired) infectionDr Ghaiath Hussein
 
Hospital acquired infection and its prevention
Hospital acquired infection and its preventionHospital acquired infection and its prevention
Hospital acquired infection and its preventionChaithanya Malalur
 
Hospital acquired infection and its prevention
Hospital acquired infection  and its preventionHospital acquired infection  and its prevention
Hospital acquired infection and its preventionChaithanya Malalur
 
2 Principles of Asepsis & Sterile Technique.ppt
2 Principles of Asepsis & Sterile Technique.ppt2 Principles of Asepsis & Sterile Technique.ppt
2 Principles of Asepsis & Sterile Technique.pptGuezilJoyDelfin2
 
Infection control in healthcare facilities
Infection control in healthcare facilitiesInfection control in healthcare facilities
Infection control in healthcare facilitiesUday Kumar
 
Sanitary Yash ep-WPS Office.pdf
Sanitary Yash ep-WPS Office.pdfSanitary Yash ep-WPS Office.pdf
Sanitary Yash ep-WPS Office.pdfishikachauhangdps
 
Nosocomial infections
Nosocomial infectionsNosocomial infections
Nosocomial infectionsdrsadhana86
 
Assignment on Control and Prevention of Diseases
Assignment on Control and Prevention of DiseasesAssignment on Control and Prevention of Diseases
Assignment on Control and Prevention of DiseasesMuniruzzaman
 
Infection control comitte
Infection control comitte   Infection control comitte
Infection control comitte Heena Parveen
 

Similar to Disinfection & sterilization journal (20)

Cdc guideline for disinfection and sterilization in healthcare facilities, 2008
Cdc guideline for disinfection and sterilization in healthcare facilities, 2008Cdc guideline for disinfection and sterilization in healthcare facilities, 2008
Cdc guideline for disinfection and sterilization in healthcare facilities, 2008
 
Guideline for disinfection and sterilization in healthcare facilities, 2008[1]
Guideline for disinfection and sterilization in healthcare facilities, 2008[1]Guideline for disinfection and sterilization in healthcare facilities, 2008[1]
Guideline for disinfection and sterilization in healthcare facilities, 2008[1]
 
stakeholders in IPC.pptx
stakeholders in IPC.pptxstakeholders in IPC.pptx
stakeholders in IPC.pptx
 
Infection control in community setting
Infection control in community settingInfection control in community setting
Infection control in community setting
 
Inf control for hcw 2012
Inf control for hcw 2012Inf control for hcw 2012
Inf control for hcw 2012
 
HAIs std precaution , safe injection.pptx
HAIs std precaution , safe injection.pptxHAIs std precaution , safe injection.pptx
HAIs std precaution , safe injection.pptx
 
Decontamination of Medical Devices.docx
Decontamination of Medical Devices.docxDecontamination of Medical Devices.docx
Decontamination of Medical Devices.docx
 
Biosafety and waste management in histopathology
Biosafety and waste management in histopathologyBiosafety and waste management in histopathology
Biosafety and waste management in histopathology
 
Basic principles of endodontics
Basic principles of endodonticsBasic principles of endodontics
Basic principles of endodontics
 
17 421 pds infection control
17 421 pds infection control17 421 pds infection control
17 421 pds infection control
 
Epcm l15 control of nosocomial (Hospital-Acquired) infection
Epcm l15 control of nosocomial (Hospital-Acquired) infectionEpcm l15 control of nosocomial (Hospital-Acquired) infection
Epcm l15 control of nosocomial (Hospital-Acquired) infection
 
Hospital acquired infection and its prevention
Hospital acquired infection and its preventionHospital acquired infection and its prevention
Hospital acquired infection and its prevention
 
Hospital acquired infection and its prevention
Hospital acquired infection  and its preventionHospital acquired infection  and its prevention
Hospital acquired infection and its prevention
 
2 Principles of Asepsis & Sterile Technique.ppt
2 Principles of Asepsis & Sterile Technique.ppt2 Principles of Asepsis & Sterile Technique.ppt
2 Principles of Asepsis & Sterile Technique.ppt
 
Infection control in healthcare facilities
Infection control in healthcare facilitiesInfection control in healthcare facilities
Infection control in healthcare facilities
 
Sanitary Yash ep-WPS Office.pdf
Sanitary Yash ep-WPS Office.pdfSanitary Yash ep-WPS Office.pdf
Sanitary Yash ep-WPS Office.pdf
 
Nosocomial infections
Nosocomial infectionsNosocomial infections
Nosocomial infections
 
Assignment on Control and Prevention of Diseases
Assignment on Control and Prevention of DiseasesAssignment on Control and Prevention of Diseases
Assignment on Control and Prevention of Diseases
 
Infection
InfectionInfection
Infection
 
Infection control comitte
Infection control comitte   Infection control comitte
Infection control comitte
 

Recently uploaded

Roles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in PharmacovigilanceRoles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in PharmacovigilanceSamikshaHamane
 
Painted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaPainted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaVirag Sontakke
 
Earth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice greatEarth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice greatYousafMalik24
 
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdfEnzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdfSumit Tiwari
 
Incoming and Outgoing Shipments in 1 STEP Using Odoo 17
Incoming and Outgoing Shipments in 1 STEP Using Odoo 17Incoming and Outgoing Shipments in 1 STEP Using Odoo 17
Incoming and Outgoing Shipments in 1 STEP Using Odoo 17Celine George
 
Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...jaredbarbolino94
 
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxPOINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxSayali Powar
 
Proudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptxProudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptxthorishapillay1
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxGaneshChakor2
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)eniolaolutunde
 
Solving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxSolving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxOH TEIK BIN
 
CELL CYCLE Division Science 8 quarter IV.pptx
CELL CYCLE Division Science 8 quarter IV.pptxCELL CYCLE Division Science 8 quarter IV.pptx
CELL CYCLE Division Science 8 quarter IV.pptxJiesonDelaCerna
 
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdfFraming an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdfUjwalaBharambe
 
Final demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptxFinal demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptxAvyJaneVismanos
 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxRaymartEstabillo3
 
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdfssuser54595a
 
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptxECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptxiammrhaywood
 

Recently uploaded (20)

TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdfTataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
 
Roles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in PharmacovigilanceRoles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in Pharmacovigilance
 
Painted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaPainted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of India
 
Earth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice greatEarth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice great
 
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdfEnzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
 
Incoming and Outgoing Shipments in 1 STEP Using Odoo 17
Incoming and Outgoing Shipments in 1 STEP Using Odoo 17Incoming and Outgoing Shipments in 1 STEP Using Odoo 17
Incoming and Outgoing Shipments in 1 STEP Using Odoo 17
 
Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...
 
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxPOINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
 
Proudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptxProudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptx
 
9953330565 Low Rate Call Girls In Rohini Delhi NCR
9953330565 Low Rate Call Girls In Rohini  Delhi NCR9953330565 Low Rate Call Girls In Rohini  Delhi NCR
9953330565 Low Rate Call Girls In Rohini Delhi NCR
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptx
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)
 
Solving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxSolving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptx
 
CELL CYCLE Division Science 8 quarter IV.pptx
CELL CYCLE Division Science 8 quarter IV.pptxCELL CYCLE Division Science 8 quarter IV.pptx
CELL CYCLE Division Science 8 quarter IV.pptx
 
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdfFraming an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
 
ESSENTIAL of (CS/IT/IS) class 06 (database)
ESSENTIAL of (CS/IT/IS) class 06 (database)ESSENTIAL of (CS/IT/IS) class 06 (database)
ESSENTIAL of (CS/IT/IS) class 06 (database)
 
Final demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptxFinal demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptx
 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
 
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
 
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptxECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
 

Disinfection & sterilization journal

  • 1. 702 • CID 2004:39 (1 September) • HEALTHCARE EPIDEMIOLOGY H E A L T H C A R E E P I D E M I O L O G Y I N V I T E D A R T I C L E Robert A. Weinstein, Section Editor Disinfection and Sterilization in Health Care Facilities: What Clinicians Need to Know William A. Rutala and David J. Weber Hospital Epidemiology, University of North Carolina Health Care System, and Division of Infectious Diseases, University of North Carolina School of Medicine, Chapel Hill All invasive procedures involve contact between a medical device or surgical instrument and a patient’s sterile tissue or mucous membranes. A major risk of all such procedures is the introduction of pathogenic microbes that could lead to infection. Failure to properly disinfect or sterilize reusable medical equipment carries a risk associated with breach of the host barriers. The level of disinfection or sterilization is dependent on the intended use of the object: critical items (such as surgical instruments, which contact sterile tissue), semicritical items (such as endoscopes, which contact mucous membranes), and noncritical items (such as stethoscopes, which contact only intact skin) require sterilization, high-level disinfection, and low-level disinfection, respectively. Cleaning must always precede high-level disinfection and sterilization. Users must consider the advantages and disadvantages of specific methods when choosing a disinfection or sterilization process. Adherence to these recommendations should improve disinfection and sterilization practices in health care facilities, thereby reducing infections associated with contaminated patient-care items. In 1996 in the United States, ∼46,500,000 surgical procedures and an even larger number of invasive medical procedures were performed [1]. For example, ∼5 million gastrointestinal en- doscopies are performed per year [1]. Each of these procedures involves contact by a medical device or surgical instrument with a patient’s sterile tissue or mucous membranes. A major risk of all such procedures is the introduction of pathogenic microbes, which can lead to infection. For example, failure to properly disinfect or sterilize equipment may lead to person- to-person transmission via contaminated devices (e.g., Myco- bacterium tuberculosis–contaminated bronchoscopes). Achieving disinfection and sterilization through the use of disinfectants and sterilization practices is essential for ensuring that medical and surgical instruments do not transmit infec- tious pathogens to patients. Because it is not necessary to ster- ilize all patient-care items, health care policies must identify whether cleaning, disinfection, or sterilization is indicated, pri- marily on the basis of each item’s intended use. Multiple studies in many countries have documented lack Received 15 March 2004; accepted 5 May 2004; electronically published 12 August 2004. Reprints or correspondence: Dr. William A. Rutala, Div. of Infectious Diseases, 130 Mason Farm Rd., Bioinformatics, University of North Carolina at Chapel Hill, Chapel Hill, NC 27599- 7030 (brutala@unch.unc.edu). Clinical Infectious Diseases 2004;39:702–9 ᮊ 2004 by the Infectious Diseases Society of America. All rights reserved. 1058-4838/2004/3905-0015$15.00 of compliance with established guidelines for disinfection and sterilization [2, 3]. Failure to comply with scientifically based guidelines has led to numerous outbreaks of infection [3–7]. In this article, a pragmatic approach to the judicious selection and proper use of disinfection and sterilization processes is presented that is based on the results of well-designed studies assessing the efficacy (via laboratory investigations) and effec- tiveness (via clinical studies) of disinfection and sterilization procedures. A RATIONAL APPROACH TO DISINFECTION AND STERILIZATION More than 35 years ago, Spaulding [8] devised a rational ap- proach to disinfection and sterilization of patient-care items or equipment. This classification scheme is so clear and logical that it has been retained, refined, and successfully used by in- fection-control professionals and others when planning meth- ods for disinfection or sterilization [9–15]. Spaulding believed that the nature of disinfection could be understood more read- ily if instruments and items for patient care were divided into 3 categories—namely, critical, semicritical, and noncritical— on the basis of the degree of risk of infection involved in the use of the items. This terminology is employed by the Centers for Disease Control and Prevention (CDC) in the documents “Guidelines for Environmental Infection Control in Health- Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
  • 2. HEALTHCARE EPIDEMIOLOGY • CID 2004:39 (1 September) • 703 Table 1. Methods for disinfection and sterilization of patient-care items and environmental surfaces. Process, method Level of microbial inactivation Example(s) (processing time) Health care application (example) Sterilization High temperature Destroys all microorganisms, including bacterial spores Steam (∼40 min) and dry heat (1–6 h, depending on temperature) Heat-tolerant critical (surgical instru- ments) and semicritical patient-care items Low temperature Destroys all microorganisms, including bacterial spores ETO gas (∼15 h) and hydrogen perox- ide gas plasma (∼50 min) Heat-sensitive critical and semicritical patient-care items Liquid immersion Destroys all microorganisms, including bacterial spores Chemical sterilants: a у2.4% glut (∼10 h), 1.12% glut and 1.93% phenol (12 h), 7.35% HP and 0.23% PA (3 h), 7.5% HP (6 h), 1.0% HP and 0.08% PA (8 h), and у0.2% PA (∼50 min at 50؇C–56؇C) Heat-sensitive critical and semicritical patient-care items that can be immersed High-level disinfection Heat automated Destroys all microorganisms except high numbers of bacterial spores Pasteurization (∼50 min) Heat-sensitive semicritical patient- care items (respiratory-therapy equipment) Liquid immersion Destroys all microorganisms except high numbers of bacterial spores Chemical sterilants or high-level disin- fectants: a 12% glut (20–45 min), 0.55% OPA (12 min), 1.12% glut and 1.93% phenol (20 min), 7.35% HP and 0.23% PA (15 min), 7.5% HP (30 min), 1.0% HP and 0.08% PA (25 min), and 650–675 ppm chlorine (10 min) Heat-sensitive semicritical patient- care items (GI endoscopes and bronchoscopes) Intermediate-level disinfection, liquid contact Destroys vegetative bacteria, myco- bacteria, most viruses, and most fungi but not bacterial spores EPA-registered hospital disinfectants with label claiming tuberculocidal activity, such as chlorine-based products and phenolics (at least 60 s) Noncritical patient-care items (blood- pressure cuff) or surfaces (bedside table), with visible blood Low-level disinfection, liquid contact Destroys vegetative bacteria and some fungi and viruses but not mycobacteria or spores EPA-registered hospital disinfectants with no tuberculocidal claim, such as chlorine-based products, phenol- ics, and quaternary ammonium compounds (at least 60 s), or 70%– 90% alcohol Noncritical patient-care items (blood- pressure cuff) or surfaces (bedside table), with no visible blood NOTE. Modified from [13], [14], and [17]. AER, automated endoscope reprocessing; EPA, Environmental Protection Agency; ETO, ethylene oxide; FDA, US Food and Drug Administration; GI, gastrointestinal; glut, glutaraldehyde; HP, hydrogen peroxide; PA, peracetic acid; OPA, ortho-phthalaldehyde. a Consult FDA-cleared package inserts for information about FDA-cleared contact time and temperature; see text for discussion of why one product (2% glut) is used at reduced exposure (20 min at 20؇C). Increasing the temperature by using AER will reduce the contact time (e.g., for OPA, 12 min at 20؇C, but 5 min at 25؇C in AER). Tubing must be completely filled for high-level disinfection and liquid chemical sterilization. Compatibility of material should be investigated when appropriate (e.g., HP and HP with PA will cause functional damage to endoscopes). Care Facilities” [16] and “Guideline for Disinfection and Ster- ilization in Healthcare Facilities” [14]. Critical items. Critical items are those associated with a high risk of infection if the item is contaminated with any microorganism, including bacterial spores. Thus, sterilization of objects that enter sterile tissue or the vascular system is critical, because any microbial contamination could result in disease transmission. This category includes surgical instru- ments, cardiac and urinary catheters, implants, and ultrasound probes used in sterile body cavities. The items in this category should be purchased as sterile or should be sterilized by steam sterilization, if possible. If the item is heat sensitive, it may be treated with ethylene oxide (ETO) or hydrogen peroxide gas plasma or with liquid chemical sterilants if other methods are unsuitable. Tables 1 and 2 list several germicides that are cat- egorized as chemical sterilants. These include у2.4% glutar- aldehyde–based formulations, 1.12% glutaraldehyde with 1.93% phenol/phenate, 7.5% stabilized hydrogen peroxide, 7.35% hydrogen peroxide with 0.23% peracetic acid, у0.2% peracetic acid, and 1.0% hydrogen peroxide with 0.08% per- acetic acid. The indicated exposure times are within the range 3–12 h, with the exception of у0.2% peracetic acid (sporicidal time of 12 min at 50ЊC–56ЊC) [19]. Use of liquid chemical sterilants is a reliable method of sterilization only if cleaning precedes treatment, which eliminates organic and inorganic material, and if the proper guidelines for concentration, contact time, temperature, and pH are followed. Another limitation to sterilization of devices with liquid chemical sterilants is that the devices cannot be wrapped during processing in the liquid chemical sterilant; thus, maintaining sterility after processing and during storage is impossible. Furthermore, after exposure to the liquid chemical sterilant, devices may require rinsing with water that, in general, is not sterile. Therefore, because of the inherent limitations of the use of liquid chemical sterilants Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
  • 3. 704 • CID 2004:39 (1 September) • HEALTHCARE EPIDEMIOLOGY Table 2. Summary of advantages and disadvantages of chemical agents used as chemical sterilants or as high-level disinfectants. Sterilization method Advantages Disadvantages Peracetic acid and hydrogen peroxide No activation required Odor or irritation not significant Concerns regarding compatibility with materials (lead, brass, copper, and zinc) and both cosmetic and functional damage Limited clinical use Potential for eye and skin damage Glutaraldehyde Numerous published studies of use Relatively inexpensive Excellent compatibility with materials Respiratory irritation from glutaraldehyde vapor Pungent and irritating odor Relatively slow mycobactericidal activity Coagulates blood and fixes tissue to surfaces Allergic contact dermatitis Hydrogen peroxide No activation required May enhance removal of organic material and organisms No disposal issues No odor or irritation issues Does not coagulate blood or fix tissues to surfaces Inactivates Cryptosporidium Published studies of use Concerns regarding compatibility with materials (brass, zinc, copper, and nickel/silver plating) and both cosmetic and functional damage Serious eye damage with contact Ortho-phthalaldehyde Fast-acting high-level disinfectant No activation required Odor not significant Claim of excellent compatibility with materials Claim of not coagulating blood or fixing tissues to surfaces Stains protein gray (e.g., skin, mucous membranes, clothing, and environmental surfaces) Limited clinical use More expensive than glutaraldehyde Eye irritation with contact Slow sporicidal activity Repeated exposure may result in hypersensitivity in some patients with bladder cancer Peracetic acid Rapid sterilization cycle time (30–45 min) Low-temperature (50؇C–55؇C) liquid-immersion sterilization Environmentally friendly by-products (acetic acid, O2, and H2O) Fully automated Single-use system eliminates need for concentration testing Standardized cycle May enhance removal of organic material and endotoxin No adverse health effects to operators, under normal operating conditions Compatible with many materials and instruments Does not coagulate blood or fix tissues to surfaces Sterilant flows through scope, facilitating salt, protein, and microbe removal Rapidly sporicidal Provides procedure standardization (constant dilution, perfusion of channel, temperatures, and exposure) Potential incompatibility with materials (e.g., aluminum anodized coating becomes dull) Used for immersible instruments only Biological indicator may not be suitable for routine monitoring Only one scope or a small number of instruments can be processed in a cycle More expensive (endoscope repairs, operating costs, and purchase costs) than high-level disinfection Serious eye and skin damage (concentrated solution) with contact Point-of-use system; no sterile storage NOTE. Modified from [18]. All products are effective in the presence of organic soil, are relatively easy to use, and have a broad spectrum of antimicrobial activity (bacteria, fungi, viruses, bacterial spores, and mycobacteria). The above characteristics are documented in the literature; contact the manufacturer of the instrument and sterilant for additional information. All products listed have been cleared by the US Food and Drug Administration (FDA) as chemical sterilants, except for ortho-phthalaldehyde, which is an FDA-cleared high-level disinfectant. in a nonautomated reprocessor, their use should be restricted to reprocessing critical devices that are heat sensitive and in- compatible with other sterilization methods. Semicritical items. Semicritical items are those that come in contact with mucous membranes or nonintact skin. Respi- ratory-therapy and anesthesia equipment, some endoscopes, laryngoscope blades, esophageal manometry probes, anorectal manometry catheters, and diaphragm-fitting rings are included in this category. These medical devices should be free of all microorganisms (i.e., mycobacteria, fungi, viruses, and bacte- ria), although small numbers of bacterial spores may be present. In general, intact mucous membranes, such as those of the lungs or the gastrointestinal tract, are resistant to infection by common bacterial spores but are susceptible to other organ- isms, such as bacteria, mycobacteria, and viruses. The mini- mum requirement for semicritical items is high-level disinfec- tion using chemical disinfectants. Glutaraldehyde, hydrogen peroxide, ortho-phthalaldehyde (OPA), peracetic acid with hy- drogen peroxide, and chlorine have been cleared by the US Food and Drug Administration (FDA) [19] and are dependable high-level disinfectants when guidelines for effective germicidal procedures are followed (tables 1 and 2). The exposure time for most high-level disinfectants varies from 10 to 45 min, at 20ЊC–25ЊC. Outbreaks of infection continue to occur when ineffective disinfectants, including iodophor, alcohol, and over- diluted glutaraldehyde [5], are used for so-called high-level Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
  • 4. HEALTHCARE EPIDEMIOLOGY • CID 2004:39 (1 September) • 705 disinfection. When a disinfectant is selected for use with certain patient-care items, the chemical compatibility after extended use with the items to be disinfected must also be considered. For example, compatibility testing by Olympus America of 7.5% hydrogen peroxide showed cosmetic and functional changes in the tested endoscopes (Olympus America, personal communication). Similarly, Olympus America does not endorse the use of products containing hydrogen peroxide with pera- cetic acid, because of cosmetic and functional damage (Olym- pus America, personal communication). Semicritical items that will have contact with the mucous membranes of the respiratory or gastrointestinal tract should be rinsed with sterile water, filtered water, or tap water, followed by an alcohol rinse [14, 20, 21]. An alcohol rinse and forced- air drying markedly reduces the likelihood of contamination of the instrument (e.g., endoscopes), most likely by eliminating the wet environment favorable to bacterial growth [21]. After rinsing, items should be dried and then stored in a manner that protects them from damage or contamination. There is no recommendation to use sterile or filtered water, rather than tap water, for rinsing semicritical equipment that will have contact with the mucous membranes of the rectum (e.g., rectal probes or anoscopes) or vagina (e.g., vaginal probes) [14]. Noncritical items. Noncritical items are those that come in contact with intact skin but not mucous membranes. Intact skin acts as an effective barrier to most microorganisms; there- fore, the sterility of items coming in contact with intact skin is “not critical.” Examples of noncritical items are bedpans, blood-pressure cuffs, crutches, bed rails, linens, bedside tables, patient furniture, and floors. In contrast to critical and some semicritical items, most noncritical reusable items may be de- contaminated where they are used and do not need to be trans- ported to a central processing area. There is virtually no doc- umented risk of transmitting infectious agents to patients via noncritical items [22] when they are used as noncritical items and do not contact nonintact skin and/or mucous membranes. However, these items (e.g., bedside tables or bed rails) could potentially contribute to secondary transmission, by contam- inating the hands of health care workers or by contact with medical equipment that will subsequently come in contact with patients [23]. Table 1 lists several low-level disinfectants that may be used for noncritical items. The exposure times for these disinfectants are 60 s or longer. CURRENT ISSUES IN DISINFECTION AND STERILIZATION Reprocessing of endoscopes. Physicians use endoscopes to diagnose and treat numerous medical disorders. Although en- doscopes are a valuable diagnostic and therapeutic tool in mod- ern medicine and although the incidence of infection associated with their use has been reported to be very low (∼1 in 1.8 million procedures) [24], more health care–associated out- breaks of infection have been linked to contaminated endo- scopes than to any other medical device [3–5]. To prevent the spread of health care–associated infection, all heat-sensitive en- doscopes (e.g., gastrointestinal endoscopes, bronchoscopes, and nasopharyngoscopes) must be properly cleaned and, at a min- imum, subjected to high-level disinfection after each use. High- level disinfection can be expected to destroy all microorganisms, although a few bacterial spores may survive when high numbers of spores are present. Recommendations for the cleaning and disinfection of en- doscopic equipment have been published and should be strictly followed [14, 20]. Unfortunately, audits have shown that per- sonnel do not adhere to guidelines on reprocessing [25–27] and that outbreaks of infection continue to occur [28, 29]. To ensure that the personnel responsible for reprocessing are prop- erly trained, initial and annual competency testing should be required for each individual who is involved in reprocessing endoscopic instruments [14, 20, 21, 30]. In general, endoscope disinfection or sterilization with a liq- uid chemical sterilant or high-level disinfectant involves the following 5 steps, which should be performed after leak testing: (1) clean: mechanically clean internal and external surfaces, including brushing internal channels and flushing each internal channel with water and an enzymatic cleaner; (2) disinfect: immerse endoscope in high-level disinfectant (or chemical ster- ilant), perfuse disinfectant (which eliminates air pockets and ensures contact of the germicide with the internal channels) into all accessible channels, such as the suction/biopsy channel and the air/water channel, and expose endoscope for the time recommended for specific products; (3) rinse: rinse the en- doscope and all channels with sterile water, filtered water (com- monly used with automated endoscope reprocessors), or tap water; (4) dry: rinse the insertion tube and inner channels with alcohol and dry with forced air, after disinfection and before storage; and (5) store: store the endoscope in a way that prevents recontamination and promotes drying (e.g., hung vertically). Unfortunately, there is poor compliance with the recom- mendations for reprocessing endoscopes. In addition, in rare instances, the scientific literature and recommendations from professional organizations regarding the use of disinfectants and sterilants may differ from claims on the manufacturer’s label. One example is the contact time used to achieve high- level disinfection with 2% glutaraldehyde. On the basis of FDA requirements (the FDA regulates liquid sterilants and high-level disinfectants used on critical and semicritical medical devices), manufacturers test the efficacy of their germicide formulations under worst-case conditions (i.e., minimum recommended concentration of the active ingredient) and in the presence of organic soil (typically, 5% serum). The soil represents the or- Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
  • 5. 706 • CID 2004:39 (1 September) • HEALTHCARE EPIDEMIOLOGY ganic loading to which the device is exposed during actual use and that would remain on the device in the absence of cleaning. These stringent test conditions are designed to provide a margin of safety, by assuring that the contact conditions for the ger- micide provide complete elimination of the test bacteria (e.g., 105 –106 cfu M. tuberculosis in organic soil and dried on a scope) if inoculated into the most difficult areas for the disinfectant to penetrate and in the absence of cleaning. However, the sci- entific data demonstrate that M. tuberculosis levels can be re- duced by at least 8 log10 cfu with cleaning (reduction of 4 log10 cfu) followed by chemical disinfection for 20 min at 20ЊC (re- duction of 4–6 log10 cfu) [14, 15, 19, 20, 31]. Because of these data, professional organizations (at least 14 worldwide) that have endorsed an endoscope-reprocessing guideline recom- mend contact with 2% glutaraldehyde for 20 min (or !20 min outside the United States) at 20ЊC to achieve high-level dis- infection, which differs from the recommendation given on the manufacturer’s label [20, 32–34]. It is important to emphasize that the FDA tests do not in- clude cleaning, a critical component of the disinfection process. When cleaning has been included in the test methodology, contact with 2% glutaraldehyde for 20 min has been demon- strated to be effective in eliminating all vegetative bacteria. Inactivation of Creutzfeldt-Jakob disease (CJD) agent. CJD is a degenerative neurologic disorder in humans, with an incidence in the United States of ∼1 case/million population/ year [35]. CJD is thought to be caused by a proteinaceous infectious agent, or prion. CJD is related to other human trans- missible spongiform encephalopathies (TSEs), such as kuru (now eradicated), Gertsmann-Straussler-Sheinker syndrome (1 case/40 million population/year), and fatal insomnia syndrome (!1 case/40 million population/year). The agents of CJD and other TSEs exhibit an unusual resistance to conventional chem- ical and physical decontamination methods. Because the CJD agent is not readily inactivated by conventional disinfection and sterilization procedures and because of the invariably fatal outcome of CJD, the procedures for disinfection and sterili- zation of the CJD prion have been both conservative and con- troversial for many years. The current recommendations consider inactivation data but also use epidemiological studies of prion transmission, infec- tivity of human tissues, and efficacy of removing proteins by cleaning. On the basis of scientific data, only critical devices (e.g., surgical instruments) and semicritical devices contami- nated with high-risk tissue (i.e., brain, spinal cord, or eye tissue) from high-risk patients (e.g., known or suspected infection with CJD or other prion disease) require special prion reprocessing. When high-risk tissues, high-risk patients, and critical or sem- icritical medical devices are involved, one of the following methods should be used: cleaning of the device and sterilization using a combination of sodium hydroxide and autoclaving [36] (e.g., immerse in 1N NaOH for 1 h, remove and rinse in water, and then transfer to an open pan for autoclaving for 1 h [at 121ЊC in a gravity displacement sterilizer or at 134ЊC in a porous or prevacuum sterilizer]); autoclaving for 18 min at 134ЊC in a prevacuum sterilizer; or autoclaving for 1 h at 132ЊC in a gravity displacement sterilizer) [14, 37]. The temperature should not exceed 134ЊC, because the effectiveness of auto- claving may decline as the temperature is increased (e.g., to 136ЊC or 138ЊC) [38]. Prion-contaminated medical devices that are impossible or difficult to clean should be discarded. Flash sterilization (i.e., steam sterilization of an unwrapped item for 3 min at 132ЊC) should not be used for reprocessing. To min- imize environmental contamination, noncritical environmental surfaces should be covered with plastic-backed paper; when contaminated with high-risk tissues, the paper should be prop- erly discarded. Noncritical environmental surfaces (e.g., labo- ratory surfaces) contaminated with high-risk tissues should be cleaned and then spot decontaminated with a 1:10 dilution of hypochlorite solution [37]. Emerging pathogens, antibiotic-resistant bacteria, and bio- terrorism agents. Emerging pathogens are of growing con- cern to the general public and infection-control professionals. Relevant pathogens include Cryptosporidium parvum, Helico- bacter pylori, Escherichia coli O157:H7, HIV, hepatitis C virus, rotavirus, multidrug-resistant M. tuberculosis, human papillo- mavirus, and nontuberculosis mycobacteria (e.g., Mycobacte- rium chelonae). Similarly, recent publications have highlighted concern about the potential for biological terrorism [39]. The CDC has categorized several agents as “high priority” because they can be easily disseminated or transmitted by person-to- person contact, can cause high mortality, and are likely to cause public panic and social disruption [40]. These agents include Bacillus anthracis (anthrax), Yersinia pestis (plague), variola major (smallpox), Francisella tularensis (tularemia), filoviruses (Ebola and Marburg [hemorrhagic fever]), and arenaviruses (Lassa [Lassa fever] and Junin [Argentine hemorrhagic fever]) and related viruses [40]. With rare exceptions (e.g., human papillomavirus), the sus- ceptibility of each of these pathogens to chemical disinfectants or sterilants has been studied, and all of these pathogens (or surrogate microbes, such as feline calicivirus for Norwalk virus, vaccinia for variola [41], and Bacillus atrophaeus [formerly Ba- cillus subtilis] for B. anthracis) have been found to be susceptible to currently available chemical disinfectants or sterilants [42]. Standard sterilization and disinfection procedures for patient- care equipment (as recommended in this article) are adequate for sterilization or disinfection of instruments or devices con- taminated with blood or other body fluids from persons in- fected with bloodborne pathogens, emerging pathogens, or bioterrorism agents, with the exception of prions (see previous section). No changes in procedures for cleaning, disinfecting, Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
  • 6. HEALTHCARE EPIDEMIOLOGY • CID 2004:39 (1 September) • 707 Table 3. Summary of advantages and disadvantages of commonly used sterilization technologies. Sterilization method Advantages Disadvantages Steam Nontoxic to patient, staff, and environment Cycle is easy to control and monitor Rapidly microbicidal Least affected by organic/inorganic soils, among steriliza- tion processes listed Rapid cycle time Penetrates medical packing and device lumens Deleterious for heat-sensitive instruments Microsurgical instruments damaged by repeated exposure May leave instruments wet, causing them to rust Potential for burns Hydrogen peroxide gas plasma Safe for the environment Leaves no toxic residuals Cycle time is 45–73 min, and no aeration is necessary Used for heat- and moisture-sensitive items, because process temperature is !50؇C Equipment is simple to operate, install (208 V outlet), and monitor Compatible with most medical devices Equipment requires electrical outlet only Cellulose (paper), linens, and liquids cannot be processed Sterilization chamber is small (∼3.5–7.3 ft3 ) Endoscope or medical-device restrictions based on lu- men internal diameter and length (see manufacturer’s recommendations) Requires synthetic packaging (polypropylene wraps or polyolefin pouches) or special container tray Hydrogen peroxide may be toxic at levels 11 ppm TWA 100% ETO Penetrates packaging materials and device lumens Single-dose cartridge and negative-pressure chamber minimizes the potential for gas leak and ETO exposure Equipment is simple to operate and monitor Compatible with most medical materials Requires aeration time to remove ETO residue Sterilization chamber is small (∼4–8.8 ft3 ) ETO is toxic, a carcinogen, and flammable ETO emission regulated by states, but catalytic cell re- moves 99.9% of ETO and converts it to CO2 and H2O ETO cartridges should be stored in flammable liquid– storage cabinet Lengthy cycle and aeration time ETO mixture a Penetrates medical packaging and many plastics Compatible with most medical materials Cycle is easy to control and monitor Some states (e.g., CA, NY, and MI) require ETO-emission reduction of 90%–99.9% CFC (inert gas that eliminates explosion hazard) banned in 1995 Potential hazards to staff and patients Lengthy cycle and aeration time ETO is toxic, a carcinogen, and flammable Peracetic acid Rapid cycle time (30–45 min) Low-temperature (50؇C–55؇C) liquid-immersion sterilization Environmentally friendly by-products Sterilant flows through endoscope, which facilitates salt, protein, and microbe removal Point-of-use system; no sterile storage Biological indicator may not be suitable for routine monitoring Used for immersible instruments only Some incompatibility with materials (e.g., aluminum anodized coating becomes dull) Only 1 scope or a small number of instruments processed in a cycle Potential for serious eye and skin damage (concentrated solution) with contact Must use connector between system and scope to ensure infusion of sterilant to all channels NOTE. Modified from [46]. CFC, chlorofluorocarbon; ETO, ethylene oxide; HCFC, hydrochlorofluorocarbon; TWA, time-weighted average. a 8.6% ETO and 91.4% HCFC; 10% ETO and 90% HCFC; or 8.5% ETO and 91.5% CO2. or sterilizing need to be made [14, 15]. In addition, there are no data to show that antibiotic-resistant bacteria (e.g., meth- icillin-resistant Staphylococcus aureus, vancomycin-resistantEn- terococcus faecium, and multidrug-resistant M. tuberculosis) are less sensitive to liquid chemical germicides than are antibiotic- sensitive bacteria at currently used germicide contact conditions and concentrations [15, 43, 44]. Advances in disinfection and sterilization methods. In the past several years, new methods of disinfection and ster- ilization have been introduced in health care settings. OPA is a chemical sterilant that received FDA clearance in October 1999. It contains 0.55% 1,2-benzenedicarboxaldehyde. In vitro studies have demonstrated excellent microbicidal activity [14, 15]. For example, Gregory et al. [45] demonstrated that OPA has shown superior mycobactericidal activity (reduction of 5 log10 in 5 min), when compared with glutaraldehyde. The ad- vantages, disadvantages, and characteristics of OPA are listed in table 2 [15]. The FDA recently cleared a liquid high-level disinfectant (su- peroxidized water) that contains 650–675 ppm free chlorine and a new sterilization system using ozone. Because there are limited data in the scientific literature for assessing the anti- microbial activity or material compatibility of these processes, they have not yet been integrated into clinical practice in the United States [14]. Several methods are used to sterilize patient-care items in health care, including steam sterilization, ETO, hydrogen per- oxide gas plasma, and a peracetic acid–immersion system. The advantages and disadvantages of these systems are listed in table 3 [14]. Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
  • 7. 708 • CID 2004:39 (1 September) • HEALTHCARE EPIDEMIOLOGY New sterilization technology based on plasma was patented in 1987 and has been marketed in the United States since 1993. Gas plasmas have been referred to as the fourth state of matter (i.e., liquid, solid, gas, and gas plasma). Gas plasmas are gen- erated in an enclosed chamber in a deep vacuum, by using radio frequency or microwave energy to excite the gas molecules and produce charged particles, many of which are in the form of free radicals. This process has the ability to inactivate a broad spectrum of microorganisms, including resistant bacterial spores. Studies have been conducted against vegetative bacteria (including mycobacteria), yeasts, fungi, viruses, and bacterial spores [14]. The effectiveness of all sterilization processes can be altered by lumen length, lumen diameter, inorganic salts, and organic materials [14]. CONCLUSION When properly used, disinfection and sterilization can ensure the safe use of invasive and noninvasive medical devices. The method of disinfection and sterilization depends on the in- tended use of the medical device: critical items (those that contact sterile tissue) must be sterilized prior to use; semicritical items (those that contact mucous membranes or nonintact skin) must undergo high-level disinfection; and noncritical items (those that contact intact skin) should undergo low-level disinfection. Cleaning should always precede high-level disin- fection and sterilization. Current disinfection and sterilization guidelines must be strictly followed. Acknowledgment Conflict of interest. W.A.R.: Honoraria from Advanced Sterilization Products; consultant for Advanced Sterilization Products, Clorox, 3M, and Metrex; and research funding from Clorox and Reckitt-Benckiser. D.J.B.: Honoraria from Clorox. References 1. Centers for Disease Control and Prevention (CDC). Ambulatory and inpatient procedures in the United States, 1996. Atlanta: CDC, 1998: 1–39. 2. McCarthy GM, Koval JJ, John MA, MacDonald JK. Infection control practices across Canada: do dentists follow the recommendations? J Can Dent Assoc 1999;65:506–11. 3. Spach DH, Silverstein FE, Stamm WE. Transmission of infection by gastrointestinal endoscopy and bronchoscopy. Ann Intern Med 1993;118:117–28. 4. Weber DJ, Rutala WA. Lessons from outbreaks associated with bron- choscopy. Infect Control Hosp Epidemiol 2001;22:403–8. 5. Weber DJ, Rutala WA, DiMarino AJ Jr. The prevention of infection following gastrointestinal endoscopy: the importance of prophylaxis and reprocessing. In: DiMarino AJ Jr, Benjamin SB, eds. Gastrointes- tinal diseases: an endoscopic approach. Thorofare, NJ: Slack, 2002: 87–106. 6. Meyers H, Brown-Elliott BA, Moore D, et al. An outbreak of Myco- bacterium chelonae infection following liposuction. Clin Infect Dis 2002;34:1500–7. 7. Lowry PW, Jarvis WR, Oberle AD, et al. Mycobacterium chelonae caus- ing otitis media in an ear-nose-and-throat practice. N Engl J Med 1988;319:978–82. 8. Spaulding EH. Chemical disinfection of medical and surgical materials. In: Lawrence C, Block SS, eds. Disinfection, sterilization, and preser- vation. Philadelphia: Lea & Febiger, 1968:517–31. 9. Favero MS, Bond WW. Chemical disinfection of medical and surgical materials. In: Block SS, ed. Disinfection, sterilization, and preservation. Philadelphia: Lippincott, Williams & Wilkins, 2001:881–917. 10. Simmons BP. CDC guidelines for the prevention and control of nos- ocomial infections: guideline for hospital environmental control. Am J Infect Control 1983;11:97–120. 11. Garner JS, Favero MS. CDC guideline for handwashing and hospital environmental control, 1985. Infect Control 1986;7:231–43. 12. Rutala WA. APIC guideline for selection and use of disinfectants. Am J Infect Control 1990;18:99–117. 13. Rutala WA. APIC guideline for selection and use of disinfectants. As- sociation for Professionals in Infection Control and Epidemiology Guidelines Committee of 1994, 1995, and 1996. Am J Infect Control 1996;24:313–42. 14. Rutala WA, Weber DJ. Guideline for disinfection and sterilization in healthcare facilities: recommendations of the CDC. Healthcare Infec- tion Control Practices Advisory Committee. MMWR Morb Mortal Wkly Rep (in press). 15. Rutala WA, Weber DJ. Selection and use of disinfectants in healthcare. In: Mayhall CG, ed. Hospital epidemiology and infection control. Phil- adelphia: Lippincott, Williams & Wilkins, 2004:1473–522. 16. Sehulster L, Chinn RYW. Guidelines for environmental infection con- trol in health-care facilities. Healthcare Infection Control Practices Ad- visory Committee. MMWR Recomm Rep 2003;52:1–44. 17. Kohn WG, Collins AS, Cleveland JL, Harte JA, Eklund KJ, Malvitz DM. Guidelines for infection control in dental health-care settings— 2003. MMWR Recomm Rep 2003;52:1–67. 18. Rutala WA, Weber DJ. Disinfection of endoscopes: review of new chem- ical sterilants used for high-level disinfection. Infect Control Hosp Epidemiol 1999;20:69–76. 19. Food and Drug Administration (FDA). FDA-cleared sterilants and high level disinfectants with general claims for processing reusable medical and dental devices, November 2003. Available at: http://www.fda.gov/ cdrh/ode/germlab.html. Accessed 8 August 2004. 20. Nelson DB, Jarvis WR, Rutala WA, et al. Multi-society guideline for reprocessing flexible gastrointestinal endoscopes. Infect Control Hosp Epidemiol 2003;24:532–7. 21. Gerding DN, Peterson LR, Vennes JA. Cleaning and disinfection of fiberoptic endoscopes: evaluation of glutaraldehyde exposure time and forced-air drying. Gastroenterology 1982;83:613–8. 22. Weber DJ, Rutala WA. Environmental issues and nosocomialinfections. In: Wenzel RP, ed. Prevention and control of nosocomial infections. Baltimore: Williams & Wilkins, 1997:491–514. 23. Weber DJ, Rutala WA. Role of environmental contamination in the transmission of vancomycin-resistant enterococci. Infect Control Hosp Epidemiol 1997;18:306–9. 24. Schembre DB. Infectious complications associated with gastrointestinal endoscopy. Gastrointest Endosc Clin N Am 2000;10:215–32. 25. Jackson FW, Ball MD. Correction of deficiencies in flexible fiberoptic sigmoidoscope cleaning and disinfection technique in family practice and internal medicine offices. Arch Fam Med 1997;6:578–82. 26. Orsi GB, Filocamo A, Di Stefano L, Tittobello A. Italian national survey of digestive endoscopy disinfection procedures. Endoscopy 1997;29: 732–8; quiz:739–40. 27. Honeybourne D, Neumann CS. An audit of bronchoscopy practice in the United Kingdom: a survey of adherence to national guidelines. Thorax 1997;52:709–13. 28. Srinivasan A, Wolfenden LL, Song X, et al. An outbreak of Pseudomonas aeruginosa infections associated with flexible bronchoscopes. N Engl J Med 2003;348:221–7. Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020
  • 8. HEALTHCARE EPIDEMIOLOGY • CID 2004:39 (1 September) • 709 29. Cetse JC, Vanhems P. Outbreak of infection associated with broncho- scopes. N Engl J Med 2003;348:2039–40. 30. Food and Drug Administration (FDA), Centers for Disease Control and Prevention (CDC). FDA and CDC public health advisory: infec- tions from endoscopes inadequately reprocessed by an automated en- doscope reprocessing system. Rockville, MD: FDA, 1999. 31. Rutala WA, Weber DJ. FDA labeling requirements for disinfection of endoscopes: a counterpoint. Infect Control Hosp Epidemiol 1995;16: 231–5. 32. Kruse A, Rey JF. Guidelines on cleaning and disinfection in GI en- doscopy: update 1999. The European Society of Gastrointestinal En- doscopy. Endoscopy 2000;32:77–80. 33. British Society of Gastroenterology. Cleaning and disinfection of equip- ment for gastrointestinal endoscopy: report of a working party of the British Society of Gastroenterology Endoscope Committee. Gut 1998;42:585–93. 34. British Thoracic Society. British Thoracic Society guidelines on diag- nostic flexible bronchoscopy. Thorax 2001;56:1–21. 35. Centers for Disease Control and Prevention. Surveillance for Creutz- feldt-Jakob disease—United States. MMWR Morb Mortal Wkly Rep 1996;45:665–8. 36. World Health Organization (WHO). WHO infection control guidelines for transmissible spongiform encephalopathies. Available at: http: //www.who.int/emc-documents/tse/docs/whocdscsraph2003.pdf. Ac- cessed 9 August 2004. 37. Rutala WA, Weber DJ. Creutzfeldt-Jakob disease: recommendationsfor disinfection and sterilization. Clin Infect Dis 2001;32:1348–56. 38. Taylor DM. Inactivation of prions by physical and chemical means. J Hosp Infect 1999;43(Suppl):S69–76. 39. Henderson DA. The looming threat of bioterrorism. Science 1999;283: 1279–82. 40. Centers for Disease Control and Prevention. Biological and chemical terrorism: strategic plan for preparedness and response. MMWR Re- comm Rep 2000;49:1–14. 41. Klein M, DeForest A. The inactivation of viruses by germicides. Chem Specialists Manuf Assoc Proc 1963;49:116–8. 42. Rutala WA, Weber DJ. Infection control: the role of disinfection and sterilization. J Hosp Infect 1999;43(Suppl):S43–55. 43. Rutala WA, Stiegel MM, Sarubbi FA, Weber DJ. Susceptibility of an- tibiotic-susceptible and antibiotic-resistant hospital bacteria to disin- fectants. Infect Control Hosp Epidemiol 1997;18:417–21. 44. Weber DJ, Rutala WA. Use of germicides in the home and health care setting: is there a relationship between germicide use and antimicrobial resistance? Infect Control Hosp Epidemiol (in press). 45. Gregory AW, Schaalje GB, Smart JD, Robison RA. The mycobactericidal efficacy of ortho-phthalaldehyde and the comparative resistances of Mycobacterium bovis, Mycobacterium terrae, and Mycobacterium che- lonae. Infect Control Hosp Epidemiol 1999;20:324–30. 46. Rutala WA, Weber DJ. Clinical effectiveness of low-temperature sterilization technologies. Infect Control Hosp Epidemiol 1998;19: 798–804. Downloadedfromhttps://academic.oup.com/cid/article/39/5/702/2022846bygueston24August2020