SlideShare a Scribd company logo
1 of 7
Download to read offline
IOSR Journal of Electronics and Communication Engineering (IOSR-JECE)
e-ISSN: 2278-2834,p- ISSN: 2278-8735.Volume 10, Issue 6, Ver. I (Nov - Dec .2015), PP 48-54
www.iosrjournals.org
DOI: 10.9790/2834-10614854 www.iosrjournals.org 48 | Page
A versatile low cost Electronic Stethoscope design
Olumuyiwa Oludare FAGBOHUN
Department of Electrical Engineering, Faculty of Engineering, Ekiti State University, Ado-Ekiti,Nigeria
Abstract: To develop an electronic stethoscope to amplify the sound of the heart and reduce any background
ambient noise thereby producing a perfect heart beat sound for hearing, and to verify the ability of the
electronic stethoscope in performing a better function than the medical acoustic type without any side effect on
the people, is the aim of this work. The device is developed and consists of two microphones whose signals are
sent through a quad amp which implements an initial gain of three and filters the input via a five pole Sallen-
Key low pass filter. The two filtered signals are sent to a mixer, which allows the user to select which input
microphone they would like to use and adjust the bass and treble characteristics of the signal. Following the
mixer, the signal is sent to another switch, which allows the user to choose between headphone or speaker
output.
I. Introduction
Stethoscope is one of the most important diagnostic tools in the medical world. Over almost two
centuries, the stethoscope has been changed and refined often, but it has never strayed too far from the original
design. Rene Theophile, a French physician, is attributed with the invention of the first stethoscope in 1816
when he was examining an obese patient. His first model of the stethoscope was simply a wooden tube
resembling a candlestick. Over the years, this model has evolved into the device that we now recognize as the
stethoscope through many small changes, such as adding earpieces for each ear and developing the combined
bell and diaphragm chest piece [1]. A stethoscope is a very straightforward device. The chest piece consists of a
shallow, bell-shaped piece and a clear, stiff diaphragm, which is then connected to the metal earpieces by a
flexible tube [1,2]. The bell is used to pick up lower frequency sounds, and the diaphragm is used for higher
frequency sounds. When the chest piece is placed on the skin, vibrations within in the body are amplified by
either the bell or diaphragm. These acoustic pressure waves then travel up through the tubing, resonating to the
earpieces and into the listener’s ears.
Very few changes have been made to the overall design of the stethoscope over the years because it
does its job so well. However, there are a few problems with current models. For example, with the standard
acoustic stethoscope, the listener is not able to amplify the sounds. This is sometimes a problem if he or she is
only getting very quiet sounds through the stethoscope. Also, the earpieces of stethoscopes can be quite
uncomfortable. Although these are not enormous problems, current technology has introduced alternatives to the
acoustic stethoscope.Stethoscopes are not only useful for doctors, but for home mechanics, exterminators,
spying etc [2,3]. Stethoscopes can also be used to check scientific vacuum chambers for leaks, and for various
other small-scale acoustic monitoring tasks. Standard stethoscope provides no amplification which reduces its
use [2,3]. A Stethoscope uses op-amps to amplify the signal and includes a low pass filter to remove background
noise.
A typical electronic stethoscope performs the following functions [3,4];
For listening to the heart beats for any irregular rate of beating.
To verify the operation of the lungs to know if there is fluid build up.
To check the blood pressure running through the veins.
To determine the process of digestive system in the stomach.
For prenatal care during pregnancy.
The field of applied science related to the art of healing by diagnosis, treatment, and prevention of
disease. It encompasses a variety of health care practices evolved to maintain and restore health by the
prevention and treatment of illness in human beings.Contemporary medicine applies biomedical sciences,
biomedical research, genetics and medical technology to diagnose, treat, and prevent injury and disease,
typically through medication or surgery, but also through therapies as diverse as psychotherapy, external splints
& traction, prostheses, biologics, pharmaceuticals, ionizing radiation among others [4].
A versatile low cost Electronic Stethoscope design
DOI: 10.9790/2834-10614854 www.iosrjournals.org 49 | Page
II. Types of stethoscope
2.1. Acousticstethoscope
The stethoscope is an instrument used for auscultation or listing to sounds produced by the body. It is
used primarily to listen to the lungs, heart and intestinal tract. It is also used to listen to blood flow in peripheral
vessels and the heart sounds of development.
Acoustic stethoscopes are familiar to most people, and operate on the transmission of sound from the
chest piece, via air-filled hollow tubes, to the listener's ears. The chest piece usually consists of two sides that
can be placed against the patient for sensing sound; a diaphragm (plastic disc) or bell (hollow cup). If the
diaphragm is placed on the patient, body sounds vibrate the diaphragm, creating acoustic pressure waves which
travel up the tubing to the listener's ears. If the bell is placed on the patient, the vibrations of the skin directly
produce acoustic pressure waves traveling up to the listener's ears [3,4]. The bell transmits low frequency
sounds, while the diaphragm transmits higher frequency sounds. This two-sided stethoscope was invented by
Rappaport and Sprague in the early part of the 20th century. One problem with acoustic stethoscopes was that
the sound level is extremely low. This problem was surmounted in 1999 with the invention of the stratified
continuous (inner) lumen, and the kinetic acoustic mechanism in 2002. Acoustic stethoscopes are the most
commonly used. A recent independent review evaluated twelve common acoustic stethoscopes on the basis of
loudness, clarity, and ergonomics. They did acoustic laboratory testing and recorded heart sounds on volunteers.
The results are listed by brand and model.
2.2 Electronicstethoscope
An electronic stethoscope (stetho-phone) overcomes the low sound levels by electronically amplifying
body sounds. However, amplification of stethoscope contact artefacts, and component cut-offs (frequency
response thresholds of electronic stethoscope microphones, pre-amps, amps, and speakers) limit electronically
amplified stethoscopes' overall utility by amplifying mid-range sounds, while simultaneously attenuating high-
and low- frequency range sounds. Currently, a number of companies offer electronic stethoscopes. Electronic
stethoscopes require conversion of acoustic sound waves to electrical signals which can then be amplified and
processed for optimal listening. Unlike acoustic stethoscopes, which are all based on the same physics,
transducers in electronic stethoscopes vary widely. The simplest and least effective method of sound detection is
achieved by placing a microphone in the chest piece. This method suffers from ambient noise interference and
has fallen out of favour. Another method, used in Welch-Allyn'sMeditron stethoscope, comprises placement of a
piezoelectric crystal at the head of a metal shaft, the bottom of the shaft making contact with a diaphragm [5,6].
3M also uses a piezo-electric crystal placed within foam behind a thick rubber-like diaphragm. Think labs'
Rhythm 32 inventor, Clive Smith uses an Electromagnetic Diaphragm with a conductive inner surface
to form a capacitive sensor. This diaphragm responds to sound waves identically to a conventional acoustic
stethoscope, with changes in an electric field replacing changes in air pressure. This preserves the sound of an
acoustic stethoscope with the benefits of amplification.Because the sounds are transmitted electronically, an
electronic stethoscope can be a wirelessdevice, can be a recording device, and can provide noise reduction,
signal enhancement, and both visual and audio output [6]. Around 2001, Stetho-graphics introduced PC-based
software which enabled a phonocardiograph, graphic representation of cardiologic and pulmonology sounds to
be generated, and interpreted according to related algorithms. All of these features are helpful for purposes of
telemedicine (remote diagnosis) and teaching.Electronic stethoscopes are also used with Computer-aided
Auscultation programs to analyse the recorded heart sounds pathological or innocent heart murmurs.
2.3 Recording stethoscopes
Some electronic stethoscopes feature direct audio output that can be used with an external recording
device, such as a laptop or MP3 recorder. The same connection can be used to listen to the previously-recorded
auscultation through the stethoscope headphones, allowing for more detailed study for general research as well
as evaluation and consultation regarding a particular patient's condition and telemedicine, or remote diagnosis
[7].
III. Methodology of research
The components used for the design of an electronic stethoscope are as listed below:-
a). Chest piece
b). A pre amplifier circuit
c). Operational amplifier circuits
d). a low pass filter circuit
e). Electret microphone
f). Regulators consisting of resistor circuits and capacitors
g). Earpiece
A versatile low cost Electronic Stethoscope design
DOI: 10.9790/2834-10614854 www.iosrjournals.org 50 | Page
h). 12 Volts power source
f). Output (ear piece/ loud speaker)
When the chest piece is placed on the patient, the earpiece allows the internal sound energy from the
patient through the conductor travels through the ear of the clinician or medical personnel. The physical
connection is made between the ear tips and the human ear. The pressure, ear tip seal, ear tip insertion angle are
all variables that can affect the connection.The chest piece usually consist of two parts that can be placed on the
patient for sensing sound i.e. low frequency sound when the diaphragm is placed against the body of the patient
skin, body sound vibrates the diaphragm creating acoustic pressure wave which can travel to the listeners ear.
The large side can be used for adult patient while the small side is especially useful for paediatric or thin patient,
around bandages and carotid assessment. The human body translates the pressure wave into sound and allows
the listener to hear what they are examined.Hence, the chest piece is part of the stethoscope that is placed on a
location where the users wants to hear sound
Figure 1: Chest piece separated from the stethoscope. (The American Journal of Cardiology)
Amplifiers are referred to active circuits since there operation is based on active devices. They are
distinguished from passive circuits by their ability to provide power gain, which is essential to most electronic
systems. This need for power gain is most evident in wireless communication systems, where the signal
received may be in the microvolt range, which is too weak to process without amplification. Small signal
amplifiers are used to increase the voltage amplitude and/or the current amplitude of a signal, to provide some
increase in the power of the signal [11]. Transistors can be used in virtually all small signal amplifiers, either
built into an integrated circuitor used in the discrete form. The choice of the either NPN or PNP bipolar or FET
depends on the design consideration. In this work, a FET transistor was selected. FETs are different from the
bipolar junction transistor in that they amplify voltage changes at the input into current changes at the output.
That is, they depend on the voltage change between the gate and the source to produce a current change between
drain and source. Their large resistance looking into the gate is a property very useful to its selection.
Filters are used in circuits which require the separation of signals according to their frequencies. A low
pass filter is provided to remove high frequency noise and also acts as an anti-aliasing filter, since sigma delta
ADC (codec) is chosen, the requirement on the analog anti-aliasing filters are minimal.This is because over
sampling relaxes the requirement on the base band anti-aliasing filter. Therefore, a simple first order active low-
pass filter is sufficient.The simplest way to make a filter is by using passive components(RLC) which works
well for high frequencies. However, at audio frequencies, the inductors become problematic, as it becomes very
large [12]heavy and expensive. For low frequency application as this design demands, a low pass active filter is
selected, with second order Butterworth at an upper cut off frequency of 400Hz.
Basically, op-amp is such an amplifier which as a very high sensitivity and gain. If it gets a small
amount of input voltage, it can amplify it very well. Thus, the amplification depends on the supply voltage, if
the gain is high enough, the output voltage will be crossing the operating voltage, hence it will saturates and its
output voltage will be the same as its operating voltage.The operational amplifier (Op-Amp) used is in the
inverting mode used as a preamplifier circuit with a design gain of -100. The inverting signal is re-inverted for
balance the phase shift with another op-amp of a gain of -100. The power amplifier is developed with the use of
a push pull op-amp to give the desired current range to drive the output transducer.
The electrets microphone used, picks up a negligible sound and converts it to an electrical format,
reason because an electrets mic is a transducer that converts one form of energy to another (basically sound
energy to electrical energy).
A versatile low cost Electronic Stethoscope design
DOI: 10.9790/2834-10614854 www.iosrjournals.org 51 | Page
IV. Operational analysis of an electronic stethoscope
Some stethoscope must be placed directly on the skin, while others can work effetely through clothing.
For the stethoscope, listeners place the chest piece on the area to be tested and a low-pitched or low frequency
sound is heard.A stethoscope is used in conjunction with a device to measure blood pressure
(sphygmomanometer). The stethoscope detects sound of blood passing through artery.
Figure 2: Block Diagram of an Electronic Stethoscope
Figure 3: Circuit design for an electronic stethoscope.
The operation of the circuit may be understood from the assumption that when a small a.c signal is
applied to the gate, it produces variations in the gate-to-source voltage. This produces variations in the drain
current. As the gate-to-source voltage increases, the drain current also increases, and the voltage drop across the
resistor R2 also increases. This causes the drain voltage to decrease. It means that the positive half cycle of the
output voltage (i.e. the output voltage at the drain is 1800
out- of phase with the input voltage at the gate. The a.c
equivalent circuit of a common source amplifier obtained by short-circuiting the capacitors and the d.c supplies,
with the FET replaced by its low frequency model [15,16], shows that,
Av = Vo/ Vin ………1
And the current through R2 = Rd (by current divider rule ) is
id = rd(gm . vgs) / (Rd + rd) ……2
where gm is the FET transconductance in mA/V or mS; and vgsis the gate –to- source voltage.
The output voltage is,
Vo = - id .Rd = - Rd . rd(gm . vgs) / (Rd + rd)
= - gm .Rd  rd . vgs = - gm . rL .vgs …..3
( rL= (Rd  rd))
Thus,
Vo = - gm . rL .vin ……4
Therefore,
Av = Vo/ Vin = - gm . rL
R1
1.5kΩ
5%
1122
R2
100kΩ
5%
1122
C2
100uF
10%
2211
1
C1
10uF
10%
2 211
R4
4.7kΩ
5%
1122
C3
10uF
10%
2 211
V1
0.001 Vpk
100 Hz
0°
12
V2
12 V
21
Q1
J112
S1D3
G2
U1A
TL072ACD
R3
2.2kΩ
5%
R5
2.2kΩ
5%
R6
47kΩ
5%
V3
12 V
21
V4
12 V
21
U2A
TL072ACD
R7
2.2kΩ
5%
R8
2.2kΩ
5%
R9
47kΩ
5%
V5
12 V
21
V6
12 V
21
XSC1
A B C D
G
T
1
2
3
4
6
5
Q2
2N4403
E3C1
B2
LS1
BUZZER
200 Hz
11
22
R12
1.5kΩ
5%
1122
1
C5
10µF
10%
2 211
C6
100µF
10%
2 211
V7
12 V
21
R14
5kΩ
5%
12
R15
100Ω
5%
1122
D1
1N4001GP
K2A1
D2
1N4001GP
K2A1
R16
5kΩ
5%
12
C7
10µF
10%
2 211
Q3
2N3904
E2C1
B3
Q4
2N3906
E2C1
B3
R10
1kΩ
5%
11 2 2
R11
30Ω
5%
1122
1
C4
10nF
5%
2211
L1
1mH
5%
12
R17
1.5kΩ
5%
1122
Electret
Mic
Control
Module
Output
Stage
A versatile low cost Electronic Stethoscope design
DOI: 10.9790/2834-10614854 www.iosrjournals.org 52 | Page
And the input resistance
Ri = Vin/ iin
with the output resistance
R’o = ( rL= Rd  rd)
And the voltage gain ,
Av = - gm .rL
Substituting the values of the parameters for 2N4416 FET : Vgs = -6v, Pd = 300mW, Vdg= 35, Vdss=
30V , Idss= 5mA and, Igss= 100pA and rd = 100k then ;
rL= Rd  rd = Rd xrd = 10 x 100 = 1000/110 = 9.09k
Rd +rd 10 +100
From shockey’s equation [15],
ID =IDss( 1 – Vgs/ Vp)2
…. 5
Where ,
dID = gm = -2 IDss( 1 – Vgs/ Vp) …….. 6
dIDssVp
substituting the values, we have from equation 6,
ID = 5x 10-3
( 1 – (-6)/(-3) )2
= 5mA
Thus ,
Rs = Vgs/ Ip = 6/ 0.005 = 1.2kΩ
And from
VDs = VDD - ID ( RD + Rs) with VDs = VDD /2 , we have
= 9 + 6 - 5 x 10 -3
RD
And
RD = 2.1 kΩ, where we select 2.2 kΩ
From equation 6,
gm = -2 x 5x 10-3
( 1 – (-6)/(-3) )
-3
= 3.333 x 10-3
S = 3.33mS.
The class B push-pull amplifier is designed for the buffer which consist of capacitors C1, C2 and C3 are
coupling capacitors to prevent the transistor dc bias voltages from being affected by the input circuit or the load
circuit. The amplifier voltage gain is measured by dividing the ac peak to peak output voltage Vo by the ac peak
to peak input voltage Vin. Because the push pull is an emitter follower configuration, the voltage gain should be
close to unity.
The design was simulated on a multism -10 version, with an input sinusoidal voltage of 1microvolt,
and the output response is as shown with measured signals of figure 4.The output voltage from the pre amplifier
circuit with transistor T1 gives 1.8uV, while the output response from the op-amp 2 gives a voltage of 3.54mV.
The output from the op-amp serves as an input to a low pass filter circuit with 101uV and the output fed to a
push pull amplifier which gives a voltage of 6.02V to drive the output device, which in this design is a 200Hz
buzzer.Also, a square wave form was used as an input with results of figure 5. The output voltage from the pre
amplifier circuit with transistor T1 gives 0.100mV, while the output response from the op-amp 2 gives a voltage
of 100mV. The output from the op-amp serves as an input to a low pass filter circuit and the output fed to a push
pull amplifier which gives a voltage of 140mV to drive the output device An impulse voltage of 0.1mV was
used as an input with a design parameter of initial value of -1mV, pulse voltage of 1uV, delay time of zero volts,
rise and fall time of 1ns and a pulse width of 0.5ms. The response the design is as shown in figure 6. The output
voltage from the pre amplifier circuit with transistor T1 gives 2.75uV, while the output response from the op-
amp 2 gives a voltage of 3.64mV. The output from the op-amp serves as an input to a low pass filter circuit and
the output fed to a push pull amplifier which gives a voltage of 6.02Vto drive the output device Apart from the
low cost value of the design at less than N15,000(fifteen thousand Naira only) Nigerian currency compared to
the acoustic value type of about N25,000 (Twenty thousand Naira only); the development gives a very clear
sound from the output response device, with better sensitivity.
A versatile low cost Electronic Stethoscope design
DOI: 10.9790/2834-10614854 www.iosrjournals.org 53 | Page
Figure 4: Analysis of design response at various points of measurement with sinusoidal input.
Figure 5: Analysis of design response at various points of measurement with sinusoidal input.
Figure 6: Analysis of design response at various points of measurement with pulse input.
A versatile low cost Electronic Stethoscope design
DOI: 10.9790/2834-10614854 www.iosrjournals.org 54 | Page
A stethoscope functions in a way that it is used by healthcare professional to examine and evaluate
their patients, typically, it is used to listen to the rhythm or sound made by the body since faints beat cannot be
heard by mere ear except with a stethoscope. For the heart, normally it beats at an even rate, when the
stethoscope is placed on it, it is heard by the healthcare professional both normal and also abnormal sound can
be heard such as; murmur, irregular rate and other sound that might not be normal.Hence, the rhythm of one’s
heartbeat can be heard through a stethoscope to determine if there are any abnormal sounds. For the lungs, it is
helpful to determine if there is a fluid build-up in the lungs and if there are any breathing difficulties.If someone
or a patient has a common flu or cold or suffer from abnormal breathing condition, medical personnel can listen
to the breathing sound by placing the stethoscope on the lungs and a sound is heard that alerts them to
inflammation the air passage and fluid and congestion. To monitor the blood pressure, the blood running
through the vein can be heard with the use of stethoscope, this is another way to count the heartbeats and listen
to the blood running through the veins. When the blood pressure cuff is inflated on the arms, medical personnel
are trained to hear the heart beating in the veins. The first beat they hear is the top number of the blood pressure
and the last beat they hear is the bottom number. In the digestive system i.e. stomach of a patient, the sounds
made by the intestines and other parts of the stomach can be listened to by the stethoscope, this can aid in
determining if there are any stomach disorders or internal abnormalities. It also aids food digestion and even
when the digestive system is empty, the stomach and bowels makes popping and gurgling noises. Some noises
maybe heard by the ears but stethoscope is much more sensitive to these noises, the absence of some could
means an obstruction in your bowels and a large number of noise could signal or signifies an infection or illness.
It is equally of note very useful in prenatal care, as an obstetrician uses a stethoscope to listen a baby’s
heartbeat in the uterus or listen to the baby’s movement in the womb.During pregnancy, the movement of the
baby, the heartbeat of the baby in the womb can be monitored by a stethoscope. Using stethoscope, the listener
can hear normal and abnormal respiratory, cardiac, pleural, arterial, venoms, interne, foetal and intestinal sound.
The advantages of an electronic stethoscope over acoustic type include; i. It is more sensitive than the
acoustic type, ii. the output sound can be varied i.e. (distinct sound output), iii. it can be connected to an external
sound amplifier and the sound can be heard by all persons present in the place for training and lecturing
purposes, iv. It produces a clearer output sound compare to the acoustic type; while the disadvantages include: i.
It is expensive and ii. It has a high cost of maintenance, iii. It is useless if the battery is down or drained.
V. Conclusions
Electronic stethoscope is a device used to measure blood pressure by detecting the blood sound
pressure, heart and lungs beat, digestive system of the stomach, prenatal care (pregnancy). The project aim at
designing an electronic stethoscope, a circuit is design to perform the same function as the conventional
stethoscope.
The entire circuit runs on a 12v battery source using a low cost electronic component like resistors,
capacitors, and op-amps and the required design is achieved.
References
[1]. J. Leyden, (2001): The Chance Invention that Changed Medicine. Saturday Evening Post, pp. 46-73.
[2]. Standris Medical Supply, Inc. (2011):History of Stethoscopes. Retrieved October 23, 2011, from Standris Medical Supply, Inc.
[3]. M. Lambe, (2009):How Does the Stethoscope Work. Retrieved November 15, 2011, from Livestrong.com:
http://www.livestrong.com/article/30071
[4]. M.C. Grenier, K. Gagnon, J. Genest, J. Durand, L.G. Durand, (1998): Clinical Comparison of Acoustic and Electronic Stethoscopes
and Design of a New Electronic Stethoscope. The American Journal of Cardiology, pp. 653-656.
[5]. D.J. Griffiths,(2008):Development of Ionic Polymer Metallic Composites as Sensors. Retrieved October 23, 2011, from .doctor:
[6]. D. Birrenkott, B. Wendorff, J. Ness and C, Durante,(2011): Heart & Breath Sounds Amplifier. Madison: University of Wisconsin-
Madison Department of Biomedical Engineering.
[7]. Opt acoustics. (2011):Technology: Core Sensor Platform. Retrieved October 22, 2011, from Optoacoustics:
http://www.optoacoustics.com/technology/core-sensor-platform
[8]. Analog Devices (2011):ADMP401: Omnidirectional Microphone with Bottom Port and Analog Output. Retrieved October 15,
2011, from Analogue Devices:
[9]. Texas Instruments. (2011): Wireless Connectivity Solutions: Bluetooth Technology. Retrieved October 23, 2011, from Texas
Instruments:
[10]. Texas Instruments.(2011):PurePath Wireless Audio. Retrieved October 23, 2011, from Texas Instruments: J. Cerezo,
(n.d.):Application Note AN-1070: Class D Audio Amplifier Performance Relationship to MOSFET Parameters. Retrieved October
22, 2011, from International Rectifier: http://www.irf.com/technical-info/appnotes/an-1070.pdf
[11]. R. Gomez, (2009):PowerWise Class G versus Class AB Headphone Amplifiers.Retrieved 22 October, 2011, from National
Semiconductor:

More Related Content

What's hot

Assistive technology for hearing and speech disorders
Assistive technology for hearing and speech disordersAssistive technology for hearing and speech disorders
Assistive technology for hearing and speech disordersAlexander Decker
 
Application of DSP in Biomedical science
Application of DSP in Biomedical scienceApplication of DSP in Biomedical science
Application of DSP in Biomedical scienceTaslima Yasmin Tarin
 
Hearing aids & implantable hearing devices
Hearing aids & implantable hearing devicesHearing aids & implantable hearing devices
Hearing aids & implantable hearing devicesAnagha Anand
 
Hearing aid anatomy
Hearing aid anatomyHearing aid anatomy
Hearing aid anatomyDerek Stiles
 
Hearing aids
Hearing aidsHearing aids
Hearing aidsjettfoo
 
Ultrasonic sound
Ultrasonic soundUltrasonic sound
Ultrasonic soundPratiksha
 
Hearing aids and psychoacoustics.ppt
Hearing aids and psychoacoustics.pptHearing aids and psychoacoustics.ppt
Hearing aids and psychoacoustics.pptbethfernandezaud
 
Assistive technology for deaf or hard of hearing
Assistive technology for deaf or hard of hearingAssistive technology for deaf or hard of hearing
Assistive technology for deaf or hard of hearingTural Abdullayev
 
Middle ear implants
Middle ear implantsMiddle ear implants
Middle ear implantsSaef Moniem
 
USES OF ULTRASOUND WAVES
USES OF ULTRASOUND WAVESUSES OF ULTRASOUND WAVES
USES OF ULTRASOUND WAVESSlenaCyrus
 
Speech coding strategies in CI
Speech coding strategies in CISpeech coding strategies in CI
Speech coding strategies in CIDr.Ebtessam Nada
 
Respiratory Rate Measurement
Respiratory Rate MeasurementRespiratory Rate Measurement
Respiratory Rate Measurementgoverdhan765
 
Middle ear implants
Middle ear implantsMiddle ear implants
Middle ear implantsVinod M K
 

What's hot (20)

Hearing aids
Hearing aidsHearing aids
Hearing aids
 
Assistive technology for hearing and speech disorders
Assistive technology for hearing and speech disordersAssistive technology for hearing and speech disorders
Assistive technology for hearing and speech disorders
 
Cochlear Implant
Cochlear ImplantCochlear Implant
Cochlear Implant
 
Lecture 5 description of electro acoustic characteristics of hearing instrume...
Lecture 5 description of electro acoustic characteristics of hearing instrume...Lecture 5 description of electro acoustic characteristics of hearing instrume...
Lecture 5 description of electro acoustic characteristics of hearing instrume...
 
Application of DSP in Biomedical science
Application of DSP in Biomedical scienceApplication of DSP in Biomedical science
Application of DSP in Biomedical science
 
Hearing aids & implantable hearing devices
Hearing aids & implantable hearing devicesHearing aids & implantable hearing devices
Hearing aids & implantable hearing devices
 
Hearing Aid
Hearing AidHearing Aid
Hearing Aid
 
Hearing aid anatomy
Hearing aid anatomyHearing aid anatomy
Hearing aid anatomy
 
Hearing aids
Hearing aidsHearing aids
Hearing aids
 
Ultrasonic sound
Ultrasonic soundUltrasonic sound
Ultrasonic sound
 
Fm system
Fm systemFm system
Fm system
 
Hearing aids and psychoacoustics.ppt
Hearing aids and psychoacoustics.pptHearing aids and psychoacoustics.ppt
Hearing aids and psychoacoustics.ppt
 
Assistive technology for deaf or hard of hearing
Assistive technology for deaf or hard of hearingAssistive technology for deaf or hard of hearing
Assistive technology for deaf or hard of hearing
 
Ultrasonography ppt[1]
Ultrasonography ppt[1]Ultrasonography ppt[1]
Ultrasonography ppt[1]
 
Middle ear implants
Middle ear implantsMiddle ear implants
Middle ear implants
 
USES OF ULTRASOUND WAVES
USES OF ULTRASOUND WAVESUSES OF ULTRASOUND WAVES
USES OF ULTRASOUND WAVES
 
Telemonitoring of Four Characteristic Parameters of Acoustic Vocal Signal in...
Telemonitoring of Four Characteristic Parameters of Acoustic  Vocal Signal in...Telemonitoring of Four Characteristic Parameters of Acoustic  Vocal Signal in...
Telemonitoring of Four Characteristic Parameters of Acoustic Vocal Signal in...
 
Speech coding strategies in CI
Speech coding strategies in CISpeech coding strategies in CI
Speech coding strategies in CI
 
Respiratory Rate Measurement
Respiratory Rate MeasurementRespiratory Rate Measurement
Respiratory Rate Measurement
 
Middle ear implants
Middle ear implantsMiddle ear implants
Middle ear implants
 

Viewers also liked

Jessica Russell Current Resume
Jessica Russell Current ResumeJessica Russell Current Resume
Jessica Russell Current ResumeJessica Russell
 
Hannah_Blumer_Resume and Artist CV
Hannah_Blumer_Resume and Artist CVHannah_Blumer_Resume and Artist CV
Hannah_Blumer_Resume and Artist CVHannah Blumer
 
Resume for Steve White 042115
Resume for Steve White 042115Resume for Steve White 042115
Resume for Steve White 042115Steve White
 
Siu_Resume
Siu_ResumeSiu_Resume
Siu_ResumeNam Siu
 
Resume for employment 2016
Resume for employment 2016Resume for employment 2016
Resume for employment 2016Sebastian Kent
 
Resume - Scott P Salyer - Energy Efficiency, Waste and Sustainability Manger
Resume - Scott P Salyer - Energy Efficiency, Waste and Sustainability MangerResume - Scott P Salyer - Energy Efficiency, Waste and Sustainability Manger
Resume - Scott P Salyer - Energy Efficiency, Waste and Sustainability MangerScott Salyer
 
Full-Stack Development
Full-Stack DevelopmentFull-Stack Development
Full-Stack DevelopmentDhilipsiva DS
 
Top 5 electrical engineer cover letter samples
Top 5 electrical engineer cover letter samplesTop 5 electrical engineer cover letter samples
Top 5 electrical engineer cover letter samplesderipoozas
 
Preetam_Resume_Business Analyst
Preetam_Resume_Business AnalystPreetam_Resume_Business Analyst
Preetam_Resume_Business AnalystPreetam Sahu
 
3.KIRAN Resume_Business Analyst_Insurance
3.KIRAN Resume_Business Analyst_Insurance3.KIRAN Resume_Business Analyst_Insurance
3.KIRAN Resume_Business Analyst_InsuranceKIRAN KUMAR
 
Daksha Desai Resume Business Analyst
Daksha Desai Resume   Business AnalystDaksha Desai Resume   Business Analyst
Daksha Desai Resume Business Analystdakshadesai1
 
Business-Analyst-Resume
Business-Analyst-ResumeBusiness-Analyst-Resume
Business-Analyst-ResumeRanjit nikam
 
Resume - Electrical Engineer (MEP) 9+ Years Exp
Resume - Electrical Engineer (MEP) 9+ Years ExpResume - Electrical Engineer (MEP) 9+ Years Exp
Resume - Electrical Engineer (MEP) 9+ Years ExpSubhash +966535838322
 

Viewers also liked (19)

U130402132138
U130402132138U130402132138
U130402132138
 
H017245157
H017245157H017245157
H017245157
 
Malick Shahzad CV-15
Malick Shahzad CV-15Malick Shahzad CV-15
Malick Shahzad CV-15
 
Jessica Russell Current Resume
Jessica Russell Current ResumeJessica Russell Current Resume
Jessica Russell Current Resume
 
Hannah_Blumer_Resume and Artist CV
Hannah_Blumer_Resume and Artist CVHannah_Blumer_Resume and Artist CV
Hannah_Blumer_Resume and Artist CV
 
Resume for Steve White 042115
Resume for Steve White 042115Resume for Steve White 042115
Resume for Steve White 042115
 
Christopher J Dingle Resume
Christopher J Dingle ResumeChristopher J Dingle Resume
Christopher J Dingle Resume
 
Siu_Resume
Siu_ResumeSiu_Resume
Siu_Resume
 
Resume for employment 2016
Resume for employment 2016Resume for employment 2016
Resume for employment 2016
 
Resume - Scott P Salyer - Energy Efficiency, Waste and Sustainability Manger
Resume - Scott P Salyer - Energy Efficiency, Waste and Sustainability MangerResume - Scott P Salyer - Energy Efficiency, Waste and Sustainability Manger
Resume - Scott P Salyer - Energy Efficiency, Waste and Sustainability Manger
 
Full-Stack Development
Full-Stack DevelopmentFull-Stack Development
Full-Stack Development
 
Top 5 electrical engineer cover letter samples
Top 5 electrical engineer cover letter samplesTop 5 electrical engineer cover letter samples
Top 5 electrical engineer cover letter samples
 
Brittany Washington Resume
Brittany Washington ResumeBrittany Washington Resume
Brittany Washington Resume
 
Preetam_Resume_Business Analyst
Preetam_Resume_Business AnalystPreetam_Resume_Business Analyst
Preetam_Resume_Business Analyst
 
3.KIRAN Resume_Business Analyst_Insurance
3.KIRAN Resume_Business Analyst_Insurance3.KIRAN Resume_Business Analyst_Insurance
3.KIRAN Resume_Business Analyst_Insurance
 
Daksha Desai Resume Business Analyst
Daksha Desai Resume   Business AnalystDaksha Desai Resume   Business Analyst
Daksha Desai Resume Business Analyst
 
ELECTRICAL ENGINEER CV
ELECTRICAL ENGINEER CVELECTRICAL ENGINEER CV
ELECTRICAL ENGINEER CV
 
Business-Analyst-Resume
Business-Analyst-ResumeBusiness-Analyst-Resume
Business-Analyst-Resume
 
Resume - Electrical Engineer (MEP) 9+ Years Exp
Resume - Electrical Engineer (MEP) 9+ Years ExpResume - Electrical Engineer (MEP) 9+ Years Exp
Resume - Electrical Engineer (MEP) 9+ Years Exp
 

Similar to J010614854

Nur 642 stethoscope slideshare
Nur 642 stethoscope slideshareNur 642 stethoscope slideshare
Nur 642 stethoscope slidesharejoanfc12
 
AIOU Code 680 General Introduction to the Hearing Impairment Semester Spring ...
AIOU Code 680 General Introduction to the Hearing Impairment Semester Spring ...AIOU Code 680 General Introduction to the Hearing Impairment Semester Spring ...
AIOU Code 680 General Introduction to the Hearing Impairment Semester Spring ...Zawarali786
 
Phonocardiogram based diagnostic system
Phonocardiogram based diagnostic systemPhonocardiogram based diagnostic system
Phonocardiogram based diagnostic systemijbesjournal
 
Current trends in audiological practices and implications for developing coun...
Current trends in audiological practices and implications for developing coun...Current trends in audiological practices and implications for developing coun...
Current trends in audiological practices and implications for developing coun...Alexander Decker
 
cochlear implantation
cochlear implantationcochlear implantation
cochlear implantationSINGH P K
 
Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...iaemedu
 
Speech processing strategies for cochlear prostheses
Speech processing strategies for cochlear prosthesesSpeech processing strategies for cochlear prostheses
Speech processing strategies for cochlear prosthesesiaemedu
 
Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...iaemedu
 
Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...iaemedu
 
Speech processing strategies for cochlear prostheses
Speech processing strategies for cochlear prosthesesSpeech processing strategies for cochlear prostheses
Speech processing strategies for cochlear prosthesesIAEME Publication
 
Speech processing strategies for cochlear prostheses
Speech processing strategies for cochlear prosthesesSpeech processing strategies for cochlear prostheses
Speech processing strategies for cochlear prosthesesiaemedu
 
Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...iaemedu
 
Phonocardiography(PCG)
Phonocardiography(PCG)Phonocardiography(PCG)
Phonocardiography(PCG)nishanthi p
 
Case study electronic equipments in hospital ece-131
Case study electronic equipments in hospital  ece-131Case study electronic equipments in hospital  ece-131
Case study electronic equipments in hospital ece-131Pushkar N/A
 

Similar to J010614854 (20)

Nur 642 stethoscope slideshare
Nur 642 stethoscope slideshareNur 642 stethoscope slideshare
Nur 642 stethoscope slideshare
 
AIOU Code 680 General Introduction to the Hearing Impairment Semester Spring ...
AIOU Code 680 General Introduction to the Hearing Impairment Semester Spring ...AIOU Code 680 General Introduction to the Hearing Impairment Semester Spring ...
AIOU Code 680 General Introduction to the Hearing Impairment Semester Spring ...
 
Phonocardiogram based diagnostic system
Phonocardiogram based diagnostic systemPhonocardiogram based diagnostic system
Phonocardiogram based diagnostic system
 
Cochlear implant
Cochlear implantCochlear implant
Cochlear implant
 
Current trends in audiological practices and implications for developing coun...
Current trends in audiological practices and implications for developing coun...Current trends in audiological practices and implications for developing coun...
Current trends in audiological practices and implications for developing coun...
 
cochlear implantation
cochlear implantationcochlear implantation
cochlear implantation
 
Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...
 
Speech processing strategies for cochlear prostheses
Speech processing strategies for cochlear prosthesesSpeech processing strategies for cochlear prostheses
Speech processing strategies for cochlear prostheses
 
Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...
 
Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...
 
Speech processing strategies for cochlear prostheses
Speech processing strategies for cochlear prosthesesSpeech processing strategies for cochlear prostheses
Speech processing strategies for cochlear prostheses
 
Speech processing strategies for cochlear prostheses
Speech processing strategies for cochlear prosthesesSpeech processing strategies for cochlear prostheses
Speech processing strategies for cochlear prostheses
 
Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...Speech processing strategies for cochlear prostheses the past, present and fu...
Speech processing strategies for cochlear prostheses the past, present and fu...
 
Phonocardiography(PCG)
Phonocardiography(PCG)Phonocardiography(PCG)
Phonocardiography(PCG)
 
Case study electronic equipments in hospital ece-131
Case study electronic equipments in hospital  ece-131Case study electronic equipments in hospital  ece-131
Case study electronic equipments in hospital ece-131
 
Eas n hp zee
Eas n hp zeeEas n hp zee
Eas n hp zee
 
Bchas zeeshan
Bchas zeeshanBchas zeeshan
Bchas zeeshan
 
Medical dissection lab audiometer
Medical dissection lab  audiometerMedical dissection lab  audiometer
Medical dissection lab audiometer
 
Physics Colloquium
Physics ColloquiumPhysics Colloquium
Physics Colloquium
 
ultrasonography.pptx
ultrasonography.pptxultrasonography.pptx
ultrasonography.pptx
 

More from IOSR Journals (20)

A011140104
A011140104A011140104
A011140104
 
M0111397100
M0111397100M0111397100
M0111397100
 
L011138596
L011138596L011138596
L011138596
 
K011138084
K011138084K011138084
K011138084
 
J011137479
J011137479J011137479
J011137479
 
I011136673
I011136673I011136673
I011136673
 
G011134454
G011134454G011134454
G011134454
 
H011135565
H011135565H011135565
H011135565
 
F011134043
F011134043F011134043
F011134043
 
E011133639
E011133639E011133639
E011133639
 
D011132635
D011132635D011132635
D011132635
 
C011131925
C011131925C011131925
C011131925
 
B011130918
B011130918B011130918
B011130918
 
A011130108
A011130108A011130108
A011130108
 
I011125160
I011125160I011125160
I011125160
 
H011124050
H011124050H011124050
H011124050
 
G011123539
G011123539G011123539
G011123539
 
F011123134
F011123134F011123134
F011123134
 
E011122530
E011122530E011122530
E011122530
 
D011121524
D011121524D011121524
D011121524
 

Recently uploaded

Artificial intelligence in cctv survelliance.pptx
Artificial intelligence in cctv survelliance.pptxArtificial intelligence in cctv survelliance.pptx
Artificial intelligence in cctv survelliance.pptxhariprasad279825
 
Dev Dives: Streamline document processing with UiPath Studio Web
Dev Dives: Streamline document processing with UiPath Studio WebDev Dives: Streamline document processing with UiPath Studio Web
Dev Dives: Streamline document processing with UiPath Studio WebUiPathCommunity
 
Are Multi-Cloud and Serverless Good or Bad?
Are Multi-Cloud and Serverless Good or Bad?Are Multi-Cloud and Serverless Good or Bad?
Are Multi-Cloud and Serverless Good or Bad?Mattias Andersson
 
Pigging Solutions in Pet Food Manufacturing
Pigging Solutions in Pet Food ManufacturingPigging Solutions in Pet Food Manufacturing
Pigging Solutions in Pet Food ManufacturingPigging Solutions
 
Bun (KitWorks Team Study 노별마루 발표 2024.4.22)
Bun (KitWorks Team Study 노별마루 발표 2024.4.22)Bun (KitWorks Team Study 노별마루 발표 2024.4.22)
Bun (KitWorks Team Study 노별마루 발표 2024.4.22)Wonjun Hwang
 
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek SchlawackFwdays
 
WordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your BrandWordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your Brandgvaughan
 
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmaticsKotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmaticscarlostorres15106
 
"Debugging python applications inside k8s environment", Andrii Soldatenko
"Debugging python applications inside k8s environment", Andrii Soldatenko"Debugging python applications inside k8s environment", Andrii Soldatenko
"Debugging python applications inside k8s environment", Andrii SoldatenkoFwdays
 
Designing IA for AI - Information Architecture Conference 2024
Designing IA for AI - Information Architecture Conference 2024Designing IA for AI - Information Architecture Conference 2024
Designing IA for AI - Information Architecture Conference 2024Enterprise Knowledge
 
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...Patryk Bandurski
 
Install Stable Diffusion in windows machine
Install Stable Diffusion in windows machineInstall Stable Diffusion in windows machine
Install Stable Diffusion in windows machinePadma Pradeep
 
Unleash Your Potential - Namagunga Girls Coding Club
Unleash Your Potential - Namagunga Girls Coding ClubUnleash Your Potential - Namagunga Girls Coding Club
Unleash Your Potential - Namagunga Girls Coding ClubKalema Edgar
 
Human Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsHuman Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsMark Billinghurst
 
Streamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project SetupStreamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project SetupFlorian Wilhelm
 
Advanced Test Driven-Development @ php[tek] 2024
Advanced Test Driven-Development @ php[tek] 2024Advanced Test Driven-Development @ php[tek] 2024
Advanced Test Driven-Development @ php[tek] 2024Scott Keck-Warren
 
Unraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdfUnraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdfAlex Barbosa Coqueiro
 
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024BookNet Canada
 
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...shyamraj55
 

Recently uploaded (20)

Artificial intelligence in cctv survelliance.pptx
Artificial intelligence in cctv survelliance.pptxArtificial intelligence in cctv survelliance.pptx
Artificial intelligence in cctv survelliance.pptx
 
Dev Dives: Streamline document processing with UiPath Studio Web
Dev Dives: Streamline document processing with UiPath Studio WebDev Dives: Streamline document processing with UiPath Studio Web
Dev Dives: Streamline document processing with UiPath Studio Web
 
Are Multi-Cloud and Serverless Good or Bad?
Are Multi-Cloud and Serverless Good or Bad?Are Multi-Cloud and Serverless Good or Bad?
Are Multi-Cloud and Serverless Good or Bad?
 
Pigging Solutions in Pet Food Manufacturing
Pigging Solutions in Pet Food ManufacturingPigging Solutions in Pet Food Manufacturing
Pigging Solutions in Pet Food Manufacturing
 
Bun (KitWorks Team Study 노별마루 발표 2024.4.22)
Bun (KitWorks Team Study 노별마루 발표 2024.4.22)Bun (KitWorks Team Study 노별마루 발표 2024.4.22)
Bun (KitWorks Team Study 노별마루 발표 2024.4.22)
 
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
 
WordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your BrandWordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your Brand
 
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmaticsKotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
 
E-Vehicle_Hacking_by_Parul Sharma_null_owasp.pptx
E-Vehicle_Hacking_by_Parul Sharma_null_owasp.pptxE-Vehicle_Hacking_by_Parul Sharma_null_owasp.pptx
E-Vehicle_Hacking_by_Parul Sharma_null_owasp.pptx
 
"Debugging python applications inside k8s environment", Andrii Soldatenko
"Debugging python applications inside k8s environment", Andrii Soldatenko"Debugging python applications inside k8s environment", Andrii Soldatenko
"Debugging python applications inside k8s environment", Andrii Soldatenko
 
Designing IA for AI - Information Architecture Conference 2024
Designing IA for AI - Information Architecture Conference 2024Designing IA for AI - Information Architecture Conference 2024
Designing IA for AI - Information Architecture Conference 2024
 
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
 
Install Stable Diffusion in windows machine
Install Stable Diffusion in windows machineInstall Stable Diffusion in windows machine
Install Stable Diffusion in windows machine
 
Unleash Your Potential - Namagunga Girls Coding Club
Unleash Your Potential - Namagunga Girls Coding ClubUnleash Your Potential - Namagunga Girls Coding Club
Unleash Your Potential - Namagunga Girls Coding Club
 
Human Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsHuman Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR Systems
 
Streamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project SetupStreamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project Setup
 
Advanced Test Driven-Development @ php[tek] 2024
Advanced Test Driven-Development @ php[tek] 2024Advanced Test Driven-Development @ php[tek] 2024
Advanced Test Driven-Development @ php[tek] 2024
 
Unraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdfUnraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdf
 
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
 
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
 

J010614854

  • 1. IOSR Journal of Electronics and Communication Engineering (IOSR-JECE) e-ISSN: 2278-2834,p- ISSN: 2278-8735.Volume 10, Issue 6, Ver. I (Nov - Dec .2015), PP 48-54 www.iosrjournals.org DOI: 10.9790/2834-10614854 www.iosrjournals.org 48 | Page A versatile low cost Electronic Stethoscope design Olumuyiwa Oludare FAGBOHUN Department of Electrical Engineering, Faculty of Engineering, Ekiti State University, Ado-Ekiti,Nigeria Abstract: To develop an electronic stethoscope to amplify the sound of the heart and reduce any background ambient noise thereby producing a perfect heart beat sound for hearing, and to verify the ability of the electronic stethoscope in performing a better function than the medical acoustic type without any side effect on the people, is the aim of this work. The device is developed and consists of two microphones whose signals are sent through a quad amp which implements an initial gain of three and filters the input via a five pole Sallen- Key low pass filter. The two filtered signals are sent to a mixer, which allows the user to select which input microphone they would like to use and adjust the bass and treble characteristics of the signal. Following the mixer, the signal is sent to another switch, which allows the user to choose between headphone or speaker output. I. Introduction Stethoscope is one of the most important diagnostic tools in the medical world. Over almost two centuries, the stethoscope has been changed and refined often, but it has never strayed too far from the original design. Rene Theophile, a French physician, is attributed with the invention of the first stethoscope in 1816 when he was examining an obese patient. His first model of the stethoscope was simply a wooden tube resembling a candlestick. Over the years, this model has evolved into the device that we now recognize as the stethoscope through many small changes, such as adding earpieces for each ear and developing the combined bell and diaphragm chest piece [1]. A stethoscope is a very straightforward device. The chest piece consists of a shallow, bell-shaped piece and a clear, stiff diaphragm, which is then connected to the metal earpieces by a flexible tube [1,2]. The bell is used to pick up lower frequency sounds, and the diaphragm is used for higher frequency sounds. When the chest piece is placed on the skin, vibrations within in the body are amplified by either the bell or diaphragm. These acoustic pressure waves then travel up through the tubing, resonating to the earpieces and into the listener’s ears. Very few changes have been made to the overall design of the stethoscope over the years because it does its job so well. However, there are a few problems with current models. For example, with the standard acoustic stethoscope, the listener is not able to amplify the sounds. This is sometimes a problem if he or she is only getting very quiet sounds through the stethoscope. Also, the earpieces of stethoscopes can be quite uncomfortable. Although these are not enormous problems, current technology has introduced alternatives to the acoustic stethoscope.Stethoscopes are not only useful for doctors, but for home mechanics, exterminators, spying etc [2,3]. Stethoscopes can also be used to check scientific vacuum chambers for leaks, and for various other small-scale acoustic monitoring tasks. Standard stethoscope provides no amplification which reduces its use [2,3]. A Stethoscope uses op-amps to amplify the signal and includes a low pass filter to remove background noise. A typical electronic stethoscope performs the following functions [3,4]; For listening to the heart beats for any irregular rate of beating. To verify the operation of the lungs to know if there is fluid build up. To check the blood pressure running through the veins. To determine the process of digestive system in the stomach. For prenatal care during pregnancy. The field of applied science related to the art of healing by diagnosis, treatment, and prevention of disease. It encompasses a variety of health care practices evolved to maintain and restore health by the prevention and treatment of illness in human beings.Contemporary medicine applies biomedical sciences, biomedical research, genetics and medical technology to diagnose, treat, and prevent injury and disease, typically through medication or surgery, but also through therapies as diverse as psychotherapy, external splints & traction, prostheses, biologics, pharmaceuticals, ionizing radiation among others [4].
  • 2. A versatile low cost Electronic Stethoscope design DOI: 10.9790/2834-10614854 www.iosrjournals.org 49 | Page II. Types of stethoscope 2.1. Acousticstethoscope The stethoscope is an instrument used for auscultation or listing to sounds produced by the body. It is used primarily to listen to the lungs, heart and intestinal tract. It is also used to listen to blood flow in peripheral vessels and the heart sounds of development. Acoustic stethoscopes are familiar to most people, and operate on the transmission of sound from the chest piece, via air-filled hollow tubes, to the listener's ears. The chest piece usually consists of two sides that can be placed against the patient for sensing sound; a diaphragm (plastic disc) or bell (hollow cup). If the diaphragm is placed on the patient, body sounds vibrate the diaphragm, creating acoustic pressure waves which travel up the tubing to the listener's ears. If the bell is placed on the patient, the vibrations of the skin directly produce acoustic pressure waves traveling up to the listener's ears [3,4]. The bell transmits low frequency sounds, while the diaphragm transmits higher frequency sounds. This two-sided stethoscope was invented by Rappaport and Sprague in the early part of the 20th century. One problem with acoustic stethoscopes was that the sound level is extremely low. This problem was surmounted in 1999 with the invention of the stratified continuous (inner) lumen, and the kinetic acoustic mechanism in 2002. Acoustic stethoscopes are the most commonly used. A recent independent review evaluated twelve common acoustic stethoscopes on the basis of loudness, clarity, and ergonomics. They did acoustic laboratory testing and recorded heart sounds on volunteers. The results are listed by brand and model. 2.2 Electronicstethoscope An electronic stethoscope (stetho-phone) overcomes the low sound levels by electronically amplifying body sounds. However, amplification of stethoscope contact artefacts, and component cut-offs (frequency response thresholds of electronic stethoscope microphones, pre-amps, amps, and speakers) limit electronically amplified stethoscopes' overall utility by amplifying mid-range sounds, while simultaneously attenuating high- and low- frequency range sounds. Currently, a number of companies offer electronic stethoscopes. Electronic stethoscopes require conversion of acoustic sound waves to electrical signals which can then be amplified and processed for optimal listening. Unlike acoustic stethoscopes, which are all based on the same physics, transducers in electronic stethoscopes vary widely. The simplest and least effective method of sound detection is achieved by placing a microphone in the chest piece. This method suffers from ambient noise interference and has fallen out of favour. Another method, used in Welch-Allyn'sMeditron stethoscope, comprises placement of a piezoelectric crystal at the head of a metal shaft, the bottom of the shaft making contact with a diaphragm [5,6]. 3M also uses a piezo-electric crystal placed within foam behind a thick rubber-like diaphragm. Think labs' Rhythm 32 inventor, Clive Smith uses an Electromagnetic Diaphragm with a conductive inner surface to form a capacitive sensor. This diaphragm responds to sound waves identically to a conventional acoustic stethoscope, with changes in an electric field replacing changes in air pressure. This preserves the sound of an acoustic stethoscope with the benefits of amplification.Because the sounds are transmitted electronically, an electronic stethoscope can be a wirelessdevice, can be a recording device, and can provide noise reduction, signal enhancement, and both visual and audio output [6]. Around 2001, Stetho-graphics introduced PC-based software which enabled a phonocardiograph, graphic representation of cardiologic and pulmonology sounds to be generated, and interpreted according to related algorithms. All of these features are helpful for purposes of telemedicine (remote diagnosis) and teaching.Electronic stethoscopes are also used with Computer-aided Auscultation programs to analyse the recorded heart sounds pathological or innocent heart murmurs. 2.3 Recording stethoscopes Some electronic stethoscopes feature direct audio output that can be used with an external recording device, such as a laptop or MP3 recorder. The same connection can be used to listen to the previously-recorded auscultation through the stethoscope headphones, allowing for more detailed study for general research as well as evaluation and consultation regarding a particular patient's condition and telemedicine, or remote diagnosis [7]. III. Methodology of research The components used for the design of an electronic stethoscope are as listed below:- a). Chest piece b). A pre amplifier circuit c). Operational amplifier circuits d). a low pass filter circuit e). Electret microphone f). Regulators consisting of resistor circuits and capacitors g). Earpiece
  • 3. A versatile low cost Electronic Stethoscope design DOI: 10.9790/2834-10614854 www.iosrjournals.org 50 | Page h). 12 Volts power source f). Output (ear piece/ loud speaker) When the chest piece is placed on the patient, the earpiece allows the internal sound energy from the patient through the conductor travels through the ear of the clinician or medical personnel. The physical connection is made between the ear tips and the human ear. The pressure, ear tip seal, ear tip insertion angle are all variables that can affect the connection.The chest piece usually consist of two parts that can be placed on the patient for sensing sound i.e. low frequency sound when the diaphragm is placed against the body of the patient skin, body sound vibrates the diaphragm creating acoustic pressure wave which can travel to the listeners ear. The large side can be used for adult patient while the small side is especially useful for paediatric or thin patient, around bandages and carotid assessment. The human body translates the pressure wave into sound and allows the listener to hear what they are examined.Hence, the chest piece is part of the stethoscope that is placed on a location where the users wants to hear sound Figure 1: Chest piece separated from the stethoscope. (The American Journal of Cardiology) Amplifiers are referred to active circuits since there operation is based on active devices. They are distinguished from passive circuits by their ability to provide power gain, which is essential to most electronic systems. This need for power gain is most evident in wireless communication systems, where the signal received may be in the microvolt range, which is too weak to process without amplification. Small signal amplifiers are used to increase the voltage amplitude and/or the current amplitude of a signal, to provide some increase in the power of the signal [11]. Transistors can be used in virtually all small signal amplifiers, either built into an integrated circuitor used in the discrete form. The choice of the either NPN or PNP bipolar or FET depends on the design consideration. In this work, a FET transistor was selected. FETs are different from the bipolar junction transistor in that they amplify voltage changes at the input into current changes at the output. That is, they depend on the voltage change between the gate and the source to produce a current change between drain and source. Their large resistance looking into the gate is a property very useful to its selection. Filters are used in circuits which require the separation of signals according to their frequencies. A low pass filter is provided to remove high frequency noise and also acts as an anti-aliasing filter, since sigma delta ADC (codec) is chosen, the requirement on the analog anti-aliasing filters are minimal.This is because over sampling relaxes the requirement on the base band anti-aliasing filter. Therefore, a simple first order active low- pass filter is sufficient.The simplest way to make a filter is by using passive components(RLC) which works well for high frequencies. However, at audio frequencies, the inductors become problematic, as it becomes very large [12]heavy and expensive. For low frequency application as this design demands, a low pass active filter is selected, with second order Butterworth at an upper cut off frequency of 400Hz. Basically, op-amp is such an amplifier which as a very high sensitivity and gain. If it gets a small amount of input voltage, it can amplify it very well. Thus, the amplification depends on the supply voltage, if the gain is high enough, the output voltage will be crossing the operating voltage, hence it will saturates and its output voltage will be the same as its operating voltage.The operational amplifier (Op-Amp) used is in the inverting mode used as a preamplifier circuit with a design gain of -100. The inverting signal is re-inverted for balance the phase shift with another op-amp of a gain of -100. The power amplifier is developed with the use of a push pull op-amp to give the desired current range to drive the output transducer. The electrets microphone used, picks up a negligible sound and converts it to an electrical format, reason because an electrets mic is a transducer that converts one form of energy to another (basically sound energy to electrical energy).
  • 4. A versatile low cost Electronic Stethoscope design DOI: 10.9790/2834-10614854 www.iosrjournals.org 51 | Page IV. Operational analysis of an electronic stethoscope Some stethoscope must be placed directly on the skin, while others can work effetely through clothing. For the stethoscope, listeners place the chest piece on the area to be tested and a low-pitched or low frequency sound is heard.A stethoscope is used in conjunction with a device to measure blood pressure (sphygmomanometer). The stethoscope detects sound of blood passing through artery. Figure 2: Block Diagram of an Electronic Stethoscope Figure 3: Circuit design for an electronic stethoscope. The operation of the circuit may be understood from the assumption that when a small a.c signal is applied to the gate, it produces variations in the gate-to-source voltage. This produces variations in the drain current. As the gate-to-source voltage increases, the drain current also increases, and the voltage drop across the resistor R2 also increases. This causes the drain voltage to decrease. It means that the positive half cycle of the output voltage (i.e. the output voltage at the drain is 1800 out- of phase with the input voltage at the gate. The a.c equivalent circuit of a common source amplifier obtained by short-circuiting the capacitors and the d.c supplies, with the FET replaced by its low frequency model [15,16], shows that, Av = Vo/ Vin ………1 And the current through R2 = Rd (by current divider rule ) is id = rd(gm . vgs) / (Rd + rd) ……2 where gm is the FET transconductance in mA/V or mS; and vgsis the gate –to- source voltage. The output voltage is, Vo = - id .Rd = - Rd . rd(gm . vgs) / (Rd + rd) = - gm .Rd rd . vgs = - gm . rL .vgs …..3 ( rL= (Rd rd)) Thus, Vo = - gm . rL .vin ……4 Therefore, Av = Vo/ Vin = - gm . rL R1 1.5kΩ 5% 1122 R2 100kΩ 5% 1122 C2 100uF 10% 2211 1 C1 10uF 10% 2 211 R4 4.7kΩ 5% 1122 C3 10uF 10% 2 211 V1 0.001 Vpk 100 Hz 0° 12 V2 12 V 21 Q1 J112 S1D3 G2 U1A TL072ACD R3 2.2kΩ 5% R5 2.2kΩ 5% R6 47kΩ 5% V3 12 V 21 V4 12 V 21 U2A TL072ACD R7 2.2kΩ 5% R8 2.2kΩ 5% R9 47kΩ 5% V5 12 V 21 V6 12 V 21 XSC1 A B C D G T 1 2 3 4 6 5 Q2 2N4403 E3C1 B2 LS1 BUZZER 200 Hz 11 22 R12 1.5kΩ 5% 1122 1 C5 10µF 10% 2 211 C6 100µF 10% 2 211 V7 12 V 21 R14 5kΩ 5% 12 R15 100Ω 5% 1122 D1 1N4001GP K2A1 D2 1N4001GP K2A1 R16 5kΩ 5% 12 C7 10µF 10% 2 211 Q3 2N3904 E2C1 B3 Q4 2N3906 E2C1 B3 R10 1kΩ 5% 11 2 2 R11 30Ω 5% 1122 1 C4 10nF 5% 2211 L1 1mH 5% 12 R17 1.5kΩ 5% 1122 Electret Mic Control Module Output Stage
  • 5. A versatile low cost Electronic Stethoscope design DOI: 10.9790/2834-10614854 www.iosrjournals.org 52 | Page And the input resistance Ri = Vin/ iin with the output resistance R’o = ( rL= Rd rd) And the voltage gain , Av = - gm .rL Substituting the values of the parameters for 2N4416 FET : Vgs = -6v, Pd = 300mW, Vdg= 35, Vdss= 30V , Idss= 5mA and, Igss= 100pA and rd = 100k then ; rL= Rd rd = Rd xrd = 10 x 100 = 1000/110 = 9.09k Rd +rd 10 +100 From shockey’s equation [15], ID =IDss( 1 – Vgs/ Vp)2 …. 5 Where , dID = gm = -2 IDss( 1 – Vgs/ Vp) …….. 6 dIDssVp substituting the values, we have from equation 6, ID = 5x 10-3 ( 1 – (-6)/(-3) )2 = 5mA Thus , Rs = Vgs/ Ip = 6/ 0.005 = 1.2kΩ And from VDs = VDD - ID ( RD + Rs) with VDs = VDD /2 , we have = 9 + 6 - 5 x 10 -3 RD And RD = 2.1 kΩ, where we select 2.2 kΩ From equation 6, gm = -2 x 5x 10-3 ( 1 – (-6)/(-3) ) -3 = 3.333 x 10-3 S = 3.33mS. The class B push-pull amplifier is designed for the buffer which consist of capacitors C1, C2 and C3 are coupling capacitors to prevent the transistor dc bias voltages from being affected by the input circuit or the load circuit. The amplifier voltage gain is measured by dividing the ac peak to peak output voltage Vo by the ac peak to peak input voltage Vin. Because the push pull is an emitter follower configuration, the voltage gain should be close to unity. The design was simulated on a multism -10 version, with an input sinusoidal voltage of 1microvolt, and the output response is as shown with measured signals of figure 4.The output voltage from the pre amplifier circuit with transistor T1 gives 1.8uV, while the output response from the op-amp 2 gives a voltage of 3.54mV. The output from the op-amp serves as an input to a low pass filter circuit with 101uV and the output fed to a push pull amplifier which gives a voltage of 6.02V to drive the output device, which in this design is a 200Hz buzzer.Also, a square wave form was used as an input with results of figure 5. The output voltage from the pre amplifier circuit with transistor T1 gives 0.100mV, while the output response from the op-amp 2 gives a voltage of 100mV. The output from the op-amp serves as an input to a low pass filter circuit and the output fed to a push pull amplifier which gives a voltage of 140mV to drive the output device An impulse voltage of 0.1mV was used as an input with a design parameter of initial value of -1mV, pulse voltage of 1uV, delay time of zero volts, rise and fall time of 1ns and a pulse width of 0.5ms. The response the design is as shown in figure 6. The output voltage from the pre amplifier circuit with transistor T1 gives 2.75uV, while the output response from the op- amp 2 gives a voltage of 3.64mV. The output from the op-amp serves as an input to a low pass filter circuit and the output fed to a push pull amplifier which gives a voltage of 6.02Vto drive the output device Apart from the low cost value of the design at less than N15,000(fifteen thousand Naira only) Nigerian currency compared to the acoustic value type of about N25,000 (Twenty thousand Naira only); the development gives a very clear sound from the output response device, with better sensitivity.
  • 6. A versatile low cost Electronic Stethoscope design DOI: 10.9790/2834-10614854 www.iosrjournals.org 53 | Page Figure 4: Analysis of design response at various points of measurement with sinusoidal input. Figure 5: Analysis of design response at various points of measurement with sinusoidal input. Figure 6: Analysis of design response at various points of measurement with pulse input.
  • 7. A versatile low cost Electronic Stethoscope design DOI: 10.9790/2834-10614854 www.iosrjournals.org 54 | Page A stethoscope functions in a way that it is used by healthcare professional to examine and evaluate their patients, typically, it is used to listen to the rhythm or sound made by the body since faints beat cannot be heard by mere ear except with a stethoscope. For the heart, normally it beats at an even rate, when the stethoscope is placed on it, it is heard by the healthcare professional both normal and also abnormal sound can be heard such as; murmur, irregular rate and other sound that might not be normal.Hence, the rhythm of one’s heartbeat can be heard through a stethoscope to determine if there are any abnormal sounds. For the lungs, it is helpful to determine if there is a fluid build-up in the lungs and if there are any breathing difficulties.If someone or a patient has a common flu or cold or suffer from abnormal breathing condition, medical personnel can listen to the breathing sound by placing the stethoscope on the lungs and a sound is heard that alerts them to inflammation the air passage and fluid and congestion. To monitor the blood pressure, the blood running through the vein can be heard with the use of stethoscope, this is another way to count the heartbeats and listen to the blood running through the veins. When the blood pressure cuff is inflated on the arms, medical personnel are trained to hear the heart beating in the veins. The first beat they hear is the top number of the blood pressure and the last beat they hear is the bottom number. In the digestive system i.e. stomach of a patient, the sounds made by the intestines and other parts of the stomach can be listened to by the stethoscope, this can aid in determining if there are any stomach disorders or internal abnormalities. It also aids food digestion and even when the digestive system is empty, the stomach and bowels makes popping and gurgling noises. Some noises maybe heard by the ears but stethoscope is much more sensitive to these noises, the absence of some could means an obstruction in your bowels and a large number of noise could signal or signifies an infection or illness. It is equally of note very useful in prenatal care, as an obstetrician uses a stethoscope to listen a baby’s heartbeat in the uterus or listen to the baby’s movement in the womb.During pregnancy, the movement of the baby, the heartbeat of the baby in the womb can be monitored by a stethoscope. Using stethoscope, the listener can hear normal and abnormal respiratory, cardiac, pleural, arterial, venoms, interne, foetal and intestinal sound. The advantages of an electronic stethoscope over acoustic type include; i. It is more sensitive than the acoustic type, ii. the output sound can be varied i.e. (distinct sound output), iii. it can be connected to an external sound amplifier and the sound can be heard by all persons present in the place for training and lecturing purposes, iv. It produces a clearer output sound compare to the acoustic type; while the disadvantages include: i. It is expensive and ii. It has a high cost of maintenance, iii. It is useless if the battery is down or drained. V. Conclusions Electronic stethoscope is a device used to measure blood pressure by detecting the blood sound pressure, heart and lungs beat, digestive system of the stomach, prenatal care (pregnancy). The project aim at designing an electronic stethoscope, a circuit is design to perform the same function as the conventional stethoscope. The entire circuit runs on a 12v battery source using a low cost electronic component like resistors, capacitors, and op-amps and the required design is achieved. References [1]. J. Leyden, (2001): The Chance Invention that Changed Medicine. Saturday Evening Post, pp. 46-73. [2]. Standris Medical Supply, Inc. (2011):History of Stethoscopes. Retrieved October 23, 2011, from Standris Medical Supply, Inc. [3]. M. Lambe, (2009):How Does the Stethoscope Work. Retrieved November 15, 2011, from Livestrong.com: http://www.livestrong.com/article/30071 [4]. M.C. Grenier, K. Gagnon, J. Genest, J. Durand, L.G. Durand, (1998): Clinical Comparison of Acoustic and Electronic Stethoscopes and Design of a New Electronic Stethoscope. The American Journal of Cardiology, pp. 653-656. [5]. D.J. Griffiths,(2008):Development of Ionic Polymer Metallic Composites as Sensors. Retrieved October 23, 2011, from .doctor: [6]. D. Birrenkott, B. Wendorff, J. Ness and C, Durante,(2011): Heart & Breath Sounds Amplifier. Madison: University of Wisconsin- Madison Department of Biomedical Engineering. [7]. Opt acoustics. (2011):Technology: Core Sensor Platform. Retrieved October 22, 2011, from Optoacoustics: http://www.optoacoustics.com/technology/core-sensor-platform [8]. Analog Devices (2011):ADMP401: Omnidirectional Microphone with Bottom Port and Analog Output. Retrieved October 15, 2011, from Analogue Devices: [9]. Texas Instruments. (2011): Wireless Connectivity Solutions: Bluetooth Technology. Retrieved October 23, 2011, from Texas Instruments: [10]. Texas Instruments.(2011):PurePath Wireless Audio. Retrieved October 23, 2011, from Texas Instruments: J. Cerezo, (n.d.):Application Note AN-1070: Class D Audio Amplifier Performance Relationship to MOSFET Parameters. Retrieved October 22, 2011, from International Rectifier: http://www.irf.com/technical-info/appnotes/an-1070.pdf [11]. R. Gomez, (2009):PowerWise Class G versus Class AB Headphone Amplifiers.Retrieved 22 October, 2011, from National Semiconductor: