SlideShare a Scribd company logo
1 of 9
Download to read offline
Core Clinic -
Structured Illumination using the
Zeiss Apotome 2
February 20th, 2015
Caterina Di Ciano-Oliveira, Erin Pardy
Research Core Facilities
Introduction: Structured Illumination
What is structured illumination?
• a method of optical sectioning
• i.e. “out-of-focus” light is removed from your image
Principle
• A grid is inserted into the light path: this generates a strong
signal when specimen details are in focus and a weak signal
when specimen details are out of focus
• The grid is translated in 1/3 steps (perpendicular to grid axis)
• Each site on the specimen experiences a series of
illumination intensities, which corresponds to the translation
phase of the grid
• Algorithm analyzes changes in intensity at a given pixel
location in the 3 images and creates an image of the focal
plane, where pixel locations showing the greatest rates of
change have the highest amplitude
• In-focus information is retained in the optical section,
resulting in a “confocal-like” image.
a) Example of illumination mask
Introduction: Structured Illumination
What is it good for?
• moderately bright specimens
• med to low mag
• e.g. tissue sections, Drosophila embryo,
nematodes, zebrafish embryo…
Advantages
• laser free
• affordable
• faster then LSCM
Disadvantages
• requires minimum of 3 images for each z-slice
• algorithm cause dynamic range to decrease by half
• slower then spinning disc
a) Example of illumination mask
Before you Start Acquisition: Calibrations
There are two calibrations that need to be done before:
1) Transmitted light phase calibration: (Caterina)
2) Focus Calibration: (User)
• must be done every time you launch Zen
• should be done for each wavelength used and for each objective
• e.g. for GFP, Rhod and Cy3 with a 20X and a 40X would be 6 total
calibrations.
Requirements for focus calibration:
• a uniformly fluorescent test sample
• e.g. autofluorescent tissue
Before you Start Acquisition: Calibrations
There are two calibrations that need to be done before:
1) Transmitted light phase calibration: (Caterina)
2) Focus Calibration: (User)
• must be done every time you launch Zen
• should be done for each wavelength used and for each objective
• e.g. for GFP, Rhod and Cy3 with a 20X and a 40X would be 6 total
calibrations.
Requirements for focus calibration:
• a uniformly fluorescent test sample
• e.g. autofluorescent tissue
Focus Calibration Workflow
1) Turn on system and launch Zen.
2) Make sure Apotome is pushed all the way in.
3) Put evenly distributed fluorescent calibration sample in stand.
4) Start with the first objective and wavelength you want to calibrate.
5) Look through eye piece and make sure your sample is in focus. (You
should see the grid.)
6) While in Locate tab, click ,“Acquisition” and then “Apotome Focus
Calibration Wizard”.
7) Follow the instructions
• 3 steps: Device Settings, Optical Focusing, Grid Focus Calibration
8) Repeat for each wavelength and for each objective.
Focus Calibration Work Flow
Notes:
• Be sure to select correct light source, filter cube, objective and
camera
- note: Other lamp = X-cite
• Use the recommended grid.
• You can adjust exposure time if needed.
• Often you must use the TFT to redirect light to camera, put in
correct filter cube and turn on RL
Focus Calibration Work Flow
Notes:
• You will see the sample on the screen with the grid.
• Highlight an area of sample that is evenly distributed like as shown in
figure b)
• You can start “Full Scan” and then “Local Scan
• Click finish once done scanning
• Either keep going with more calibrations or click “No” to start acquiring
Acquisition
• Once you have finished calibrations,
go to Acquisition tab.
• Make sure “Apotome Enabled” mode
is checked.
• Make sure correct grid is selected
• Perform acquisitions
• View image
Conventional Sectioning Optical Sectioning

More Related Content

What's hot

Phase contrast microscopy Likhith K
Phase contrast microscopy Likhith KPhase contrast microscopy Likhith K
Phase contrast microscopy Likhith K
LIKHITHK1
 
Spectrophotometry.ppt 18.07.08
Spectrophotometry.ppt 18.07.08Spectrophotometry.ppt 18.07.08
Spectrophotometry.ppt 18.07.08
Amna Azam
 
Chaprter 23 spectrophotometer
Chaprter 23   spectrophotometerChaprter 23   spectrophotometer
Chaprter 23 spectrophotometer
anatanat
 

What's hot (19)

Spectrophotometer-MiniScan EZ by HunterLab
Spectrophotometer-MiniScan EZ by HunterLabSpectrophotometer-MiniScan EZ by HunterLab
Spectrophotometer-MiniScan EZ by HunterLab
 
performance and specifications of spectrophotometer
performance and specifications of spectrophotometerperformance and specifications of spectrophotometer
performance and specifications of spectrophotometer
 
Spectrophotometer
SpectrophotometerSpectrophotometer
Spectrophotometer
 
The Sun Photometer_IEEE Article
The Sun Photometer_IEEE ArticleThe Sun Photometer_IEEE Article
The Sun Photometer_IEEE Article
 
30_4-Jacobs
30_4-Jacobs30_4-Jacobs
30_4-Jacobs
 
Spectrophotometer
SpectrophotometerSpectrophotometer
Spectrophotometer
 
Spectrophotometry
SpectrophotometrySpectrophotometry
Spectrophotometry
 
Phase contrast microscopy Likhith K
Phase contrast microscopy Likhith KPhase contrast microscopy Likhith K
Phase contrast microscopy Likhith K
 
spectrophotometer
spectrophotometerspectrophotometer
spectrophotometer
 
Spectrophotometry.ppt 18.07.08
Spectrophotometry.ppt 18.07.08Spectrophotometry.ppt 18.07.08
Spectrophotometry.ppt 18.07.08
 
Design of spectrophotometry
Design of spectrophotometryDesign of spectrophotometry
Design of spectrophotometry
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Three dimensional particle image velocimetry
Three dimensional particle image velocimetryThree dimensional particle image velocimetry
Three dimensional particle image velocimetry
 
Spectrophotometry
Spectrophotometry Spectrophotometry
Spectrophotometry
 
Prelim
PrelimPrelim
Prelim
 
Spectrometry by kk sahu sir
Spectrometry by kk sahu sirSpectrometry by kk sahu sir
Spectrometry by kk sahu sir
 
Types of spectrometer IMA
Types of spectrometer IMATypes of spectrometer IMA
Types of spectrometer IMA
 
Chaprter 23 spectrophotometer
Chaprter 23   spectrophotometerChaprter 23   spectrophotometer
Chaprter 23 spectrophotometer
 
Sem and tem
Sem and temSem and tem
Sem and tem
 

Similar to StructuredIlluminationKeenan

BASIC PRINCIPLE OF SPECT AND PE.....T.pptx
BASIC PRINCIPLE OF SPECT AND PE.....T.pptxBASIC PRINCIPLE OF SPECT AND PE.....T.pptx
BASIC PRINCIPLE OF SPECT AND PE.....T.pptx
justinfan550
 
Portal Imaging used to clear setup uncertainty
Portal Imaging used to clear setup uncertaintyPortal Imaging used to clear setup uncertainty
Portal Imaging used to clear setup uncertainty
MajoVJJose
 

Similar to StructuredIlluminationKeenan (20)

Lecture (3)external beam therapy (ebt)
Lecture (3)external beam therapy (ebt)Lecture (3)external beam therapy (ebt)
Lecture (3)external beam therapy (ebt)
 
BASIC PRINCIPLE OF SPECT AND PE.....T.pptx
BASIC PRINCIPLE OF SPECT AND PE.....T.pptxBASIC PRINCIPLE OF SPECT AND PE.....T.pptx
BASIC PRINCIPLE OF SPECT AND PE.....T.pptx
 
Confocal microscope presentation pt
Confocal microscope presentation  ptConfocal microscope presentation  pt
Confocal microscope presentation pt
 
Creating 3D neuron reconstructions from image stacks and virtual slides
Creating 3D neuron reconstructions from image stacks and virtual slidesCreating 3D neuron reconstructions from image stacks and virtual slides
Creating 3D neuron reconstructions from image stacks and virtual slides
 
Commissioning of a Surface Guided Radiotherapy System
Commissioning of a Surface Guided Radiotherapy System Commissioning of a Surface Guided Radiotherapy System
Commissioning of a Surface Guided Radiotherapy System
 
Object detection at night
Object detection at nightObject detection at night
Object detection at night
 
Analytical Methods for Characterization of Solid Forms
Analytical Methods for Characterization of Solid FormsAnalytical Methods for Characterization of Solid Forms
Analytical Methods for Characterization of Solid Forms
 
14184864.ppt
14184864.ppt14184864.ppt
14184864.ppt
 
Summer Internship University of IIllinois at Urbana Champaign
Summer Internship University of IIllinois at Urbana ChampaignSummer Internship University of IIllinois at Urbana Champaign
Summer Internship University of IIllinois at Urbana Champaign
 
Cone beam computed tomography
Cone beam computed tomographyCone beam computed tomography
Cone beam computed tomography
 
How WormLab Works
How WormLab WorksHow WormLab Works
How WormLab Works
 
Cbct
CbctCbct
Cbct
 
photometry.pptx
photometry.pptxphotometry.pptx
photometry.pptx
 
Instrumentation of uv visible spectrophotometer
Instrumentation of uv visible spectrophotometerInstrumentation of uv visible spectrophotometer
Instrumentation of uv visible spectrophotometer
 
Portal Imaging used to clear setup uncertainty
Portal Imaging used to clear setup uncertaintyPortal Imaging used to clear setup uncertainty
Portal Imaging used to clear setup uncertainty
 
WRAIR
WRAIRWRAIR
WRAIR
 
Digital image processing 2
Digital image processing   2Digital image processing   2
Digital image processing 2
 
C.elegans Tracking and Analysis
C.elegans Tracking and AnalysisC.elegans Tracking and Analysis
C.elegans Tracking and Analysis
 
Artifacts
ArtifactsArtifacts
Artifacts
 
16 how to set up a gait lab (nov 2014)
16 how to set up a gait lab (nov 2014)16 how to set up a gait lab (nov 2014)
16 how to set up a gait lab (nov 2014)
 

StructuredIlluminationKeenan

  • 1. Core Clinic - Structured Illumination using the Zeiss Apotome 2 February 20th, 2015 Caterina Di Ciano-Oliveira, Erin Pardy Research Core Facilities
  • 2. Introduction: Structured Illumination What is structured illumination? • a method of optical sectioning • i.e. “out-of-focus” light is removed from your image Principle • A grid is inserted into the light path: this generates a strong signal when specimen details are in focus and a weak signal when specimen details are out of focus • The grid is translated in 1/3 steps (perpendicular to grid axis) • Each site on the specimen experiences a series of illumination intensities, which corresponds to the translation phase of the grid • Algorithm analyzes changes in intensity at a given pixel location in the 3 images and creates an image of the focal plane, where pixel locations showing the greatest rates of change have the highest amplitude • In-focus information is retained in the optical section, resulting in a “confocal-like” image. a) Example of illumination mask
  • 3. Introduction: Structured Illumination What is it good for? • moderately bright specimens • med to low mag • e.g. tissue sections, Drosophila embryo, nematodes, zebrafish embryo… Advantages • laser free • affordable • faster then LSCM Disadvantages • requires minimum of 3 images for each z-slice • algorithm cause dynamic range to decrease by half • slower then spinning disc a) Example of illumination mask
  • 4. Before you Start Acquisition: Calibrations There are two calibrations that need to be done before: 1) Transmitted light phase calibration: (Caterina) 2) Focus Calibration: (User) • must be done every time you launch Zen • should be done for each wavelength used and for each objective • e.g. for GFP, Rhod and Cy3 with a 20X and a 40X would be 6 total calibrations. Requirements for focus calibration: • a uniformly fluorescent test sample • e.g. autofluorescent tissue
  • 5. Before you Start Acquisition: Calibrations There are two calibrations that need to be done before: 1) Transmitted light phase calibration: (Caterina) 2) Focus Calibration: (User) • must be done every time you launch Zen • should be done for each wavelength used and for each objective • e.g. for GFP, Rhod and Cy3 with a 20X and a 40X would be 6 total calibrations. Requirements for focus calibration: • a uniformly fluorescent test sample • e.g. autofluorescent tissue
  • 6. Focus Calibration Workflow 1) Turn on system and launch Zen. 2) Make sure Apotome is pushed all the way in. 3) Put evenly distributed fluorescent calibration sample in stand. 4) Start with the first objective and wavelength you want to calibrate. 5) Look through eye piece and make sure your sample is in focus. (You should see the grid.) 6) While in Locate tab, click ,“Acquisition” and then “Apotome Focus Calibration Wizard”. 7) Follow the instructions • 3 steps: Device Settings, Optical Focusing, Grid Focus Calibration 8) Repeat for each wavelength and for each objective.
  • 7. Focus Calibration Work Flow Notes: • Be sure to select correct light source, filter cube, objective and camera - note: Other lamp = X-cite • Use the recommended grid. • You can adjust exposure time if needed. • Often you must use the TFT to redirect light to camera, put in correct filter cube and turn on RL
  • 8. Focus Calibration Work Flow Notes: • You will see the sample on the screen with the grid. • Highlight an area of sample that is evenly distributed like as shown in figure b) • You can start “Full Scan” and then “Local Scan • Click finish once done scanning • Either keep going with more calibrations or click “No” to start acquiring
  • 9. Acquisition • Once you have finished calibrations, go to Acquisition tab. • Make sure “Apotome Enabled” mode is checked. • Make sure correct grid is selected • Perform acquisitions • View image Conventional Sectioning Optical Sectioning