1. Introduction
Absorption Spectroscopic methods of analysis rank among the most widespread and
powerful tools for quantitative analysis. The use of a spectrophotometer to determine the
extent of absorption of various wavelengths of visible light by a given solution is commonly
known as colorimetry. This method is used to determine concentrations of various
chemicals which can give colours either directly or after addition of some other chemicals.
As an example, in the analysis of phosphate, a reaction with orthophosphate is made, to
form the highly coloured molybdenum blue compound. The light absorption of this
compound can then be measured in a spectrophotometer.
Some compounds absorb light in other than the visible range of the spectrum. For
example, nitrates absorb radiation of 220 nm wave length in the UV region.
2. Theory
Absorption Spectroscopic methods of analysis are based upon the fact that compounds
ABSORB light radiation of a specific wavelength. In the analysis, the amount of light
radiation absorbed by a sample is measured. The light absorption is directly related to the
concentration of the coloured compound in the sample.
The wavelength (X) of Maximum Absorption is known for different compounds. For
example, the coloured compound formed for analysis of Phosphate (molybdenum blue)
has maximum light absorption at a= 640 nm. Conversely, a minimum amount of light is
transmitted through the compound at X= 640 nm. This is shown schematically in Figure .1
100
/ s
/ 

840
Wavefengift {nm}
400 640
Wavelength (nm)
Figure 1: Light Absorption and Transmission by Phosphate-molybdenum blue compound.
Schematic diagram showing maximum light absorption (and minimum light transmission) at
A= 640 nm.
The Beer-Lambert Law
The Beer-Lambert Law is illustrated in Figure 2. The Absorbance (or optical density)
and Transmission (or Transmittance) of light through a sample can be calculated by
measuring light intensity entering and exiting the sample,
Figure 2: Light energy of Intensity V passes through a sample with
concentration 'C'. Some light energy is absorbed by the sample.
The amount of light energy exiting the sample has Intensity T
The following terms are defined:
• Light Intensity entering a sample is "l0"
• Light Intensity exiting a sample is "I"
• The concentration of analyte in sample is "C"
• The length of the light path in glass sample cuvette is "L".
• "K" is a constant for a particular solution and wave length
The Beer-Lambert Law is given by the following equations:
Light Absorbance (A) = log (l0/1)= KCL
Light Transmission (T) = l/l0 = 10"KCL
Plots of absorbance and transmission versus concentration of the analyte in sample
according to the above equations is shown in Figure 3,
Concentration Concentration
Figure 3: Beer-Lambert Law relates the amount of light Absorbance (A) by a
solution to the Concentration (C) of a compound in solution and the
length of light path:
• As Concentration (C) increases, light Absorption (A) increases, linearly .
• As Concentration (C) increases, light Transmission (T) decreases,
exponentially
3. The Spectrophotometer Instrument
All spectrophotometer instruments designed to measure the absorption of radiant
energy
have the basic components as follows (Figure 4):
1. a stable source of radiant energy (Light);
2. a wavelength selector to isolate a desired wavelength from the source (filter or
monochromator);
3. transparent container (cuvette) for the sample and the blank;
4. a radiation detector (phototube) to convert the radiant energy received to a
measurable signal; and a readout device that displays the signal from the
detector.
I inht Source
Mnnor.hromator
Samnle
Phototube Measuring
Absorbed / Transmitted
Spectrophotometer
Figure 4: Components of a spectrophotometer
The energy source is to provide a stable source of light radiation, whereas the
wavelength selector permits separation of radiation of the desired wavelength from
other radiation. Light radiation passes through a glass container with sample. The
detector measures the energy after it has passed through the sample. The readout
device calculates the amount of light absorbed by the sample displays the signal from
the detector as absorbance or transmission.
The spectrophotometers which are used for such measurements may vary from simple
and relatively inexpensive colorimeters to highly sophisticated and expensive
instruments that automatically scan the ability of a solution to absorb radiation over a
wide range of wavelengths and record the results of these measurements.
One instrument cannot be used to measure absorbance at all wavelengths because a
given energy source and energy detector is suitable for use over only a limited range of
wavelengths.
True linearity between absorbance and concentration according to Beer-Lambert Law
requires the use of monochromatic light. In addition, a narrow band of light ensures a
greater selectivity since substance with absorption peaks in other close by wavelengths
are less likely to interfere. Further, it increases sensitivity as there is a greatest change
in absorbance per increment of change in concentration.
Both filters and monochromators are used to restrict the radiation wavelength.
Photometers make use of filters, which function by absorbing large protions of the
spectrum while transmitting relatively limited wavelength regions. Spectrophotometers
are instruments equipped with monochromators that permit the continuous variation and
selection of waveiength. The effective bandwidth of a monochromator that is
satisfactory for most applications is about from 1 to 5 nm.
The sample containers, cells or cuvettes, must be fabricated from material that is
transparent to radiation in the spectral region of interest. The commonly used materials
for different wave length regions are:
Quartz or fused silica: UV to 2 ^m in I R
Silicate glass: Above 350 nm to 2 jim in I R
Plastic: visible region
Polished NaCI or AgCI: Wave lengths longer than 2jLim
Cuvettes or cells are provided in pairs that have been carefully matched to make
possible the transmission through the solvent and the sample. Accurate
spectrophotometric analysis requires the use of good quality, matched cells. These
should be regularly checked against one another to detect differences that can arise
from scratches, etching and wear. The most common cell path for UV-visible region is 1
cm. For reasons of economy, cylindrical cells are frequently used. Care must be taken
to duplicate the position of such cells with respect to the light path; otherwise, variations
in path length and in reflection losses will introduce errors.
4. General Measurement Procedures
As explained above, the Beer-Lambert Law forms the basis of the measurement
procedure. The amount of light radiation absorbed by a compound is directly related to
the concentration of the compound.
The general measurement procedure consists of 5 steps:
1. Prepare samples to make coloured compound
2. Make series of standard solutions of known concentrations and treat them in the
same manner as the sample for making coloured compounds
3. Set spectrophotometer to X of maximum light absorption
4. Measure light absorbance of standards
5. Plot standard curve: Absorbance vs. Concentration, as shown in Figure 5
Once the standard plot is made, it is simple to find the concentration of an unknown
sample: Measure the absorption of the unknown, and from the standard plot, read the
related concentration (Figure 6).
Cnnnpntration
Figure 5: Plot of the standard curve: showing the linear relation between light
absorption and concentration of the standards
Figure 6: Finding the concentration of an unknown sample from the standard
curve.
Due to the fact that the overall composition of the sample is seldom the same as that of
the calibration standard, in some cases, the absorption characteristics of the two may
differ. Where such discrepancy is suspected, the standard addition approach may be
used. Here, a known amount of analyte is added to a second aliquot of the sample. The
difference in absorbance is used to calculate the analyte concentration of the sample as
illustrated in Example 1.
Example 1
A 25 ml_ sample after treatment with reagents to generate colour for measurement of
phosphate yielded absorbance of 0.428. Addition of 1.00 ml_ of a solution containing 5.0|ag
phosphorus to a second 25 mL aliquot and development of colour resulted in an absorbance of
0.517. Calculate jag phosphorus in each mL of sample.
Solution:
Correct absorbance for dilution:
Corrected absorbance = 0.517 (26.0/25.0) = 0.538
Absorbance caused by 5|jg phosphorus = 0.538 - 0.428 = 0.110
Therefore, phosphorus in the sample = (5.0/0.11) 0.428
= 19.5[jg, or 19.5/25 = 0.7|jg/mL
OPERATING INSTRUCTIONS FOR SPECTROPHOTOMETER
1. Ensure 230V, 50Hz, single-phase power supply is available in the 3 pin contact 5A
power point,
2. Insert the power cord of the instrument in to a power point.
3. Turn the switch at 'ON" position.
4. Warm up the instrument for 15 minutes.
5. Press %T selector switch.
6. Adjust wavelength control to read the wavelength at which the test is desired.
7. Open the sliding lid of the sample compartment.
8. Turn the filter sliding control at filter position-1 to set zero adjustment.
9. Close the sliding lid of the sample compartment.
10. Turn the coarse and fine controls at their maximum and adjust %T control to get
0.00 displayed on the read out with the filter position-1.
11. Now select the filter position according to the desired wavelength:
Filter Position Wave Length
2 below 395nm
3 395 to 600nm
4 above 600nm
12. Keep the blank cuvette at position-1.
13. Keep the sample cuvette at position -2.
14. Adjust 100%T (%T selector switch should be in pressed position) or 0.0 absorbance
(O.D selector switch should be in pressed position) using coarse and fine controls
knob with blank cuvette.
15. Pull the cuvette position control at position-2.
16. Read out the transmittance/absorbance of the sample.
CHECKING THE PERFORMANCE OF THE SPECTROPHOTOMETER.
1. Preparation of Blank: -
Measure accurately 10 ml of "Anal R" quality HCI of 36-46% concentration. Add it slowly
to 400 ml distilled water filled in one litre volumetric flask. Make it up to mark with
distilled water and mix it up well.
2. Preparation of sample
Weigh exactly 22.2 gm of "Anal R" quality Cobalt Chloride(CoCI2 ).Transfer into one
litre volumetric flask. Dissolve it in 1% Hydrochloric Acid (HCI) and make it up to the
mark.
3. Fill the blank (1%HCI) & sample (CoCI2 Solution) solution in cuvettes.
4. Measure the transmittance/absorbance at wavelength 480 to 540 nm at 5nm
interval.
5. Plot the graph of absorbance versus wavelength. The absorbance should increase
up to 510-515nm wavelength and then decrease at higher wavelengths.
Precautions
1. Always clean the cells thoroughly and rinse at least once with a portion of the
sample, before filling the sample for measurement.
2. Always wipe the exposed surface of the cells dry and free from finger prints, using
tissue paper.
3. Clean cells thoroughly both after and prior to use.
4. Do not leave solutions, particularly strong alkali, in the cells for periods more than an
hour.
5. Ensure that no air bubbles stick to the inner surfaces of the cells, when taking
measurements.
6. Never use a brush or any tool for cleaning which may scratch the optical surfaces.
Cleaning Method
Take 1gm of potassium dichromate, add a little distilled water and very slowly add
approximately 100ml of conc.H2S04. Keep the cells in this acidic solution for a maximum
period of 12hrs. Wash the cells thoroughly with distilled water.

More Related Content

What's hot

Spectrophotometry in clinical chemistry
Spectrophotometry in clinical chemistrySpectrophotometry in clinical chemistry
Spectrophotometry in clinical chemistryOfonmbuk Umoh
 
Spectrophotometry lecture
Spectrophotometry lectureSpectrophotometry lecture
Spectrophotometry lectureCody Taylor
 
How to use a colorimeter
How to use a colorimeterHow to use a colorimeter
How to use a colorimeterDana Porter
 
APPLICATION OF COLORIMETRY IN TRACE MINERAL ANALYSIS
APPLICATION OF COLORIMETRY IN TRACE MINERAL ANALYSISAPPLICATION OF COLORIMETRY IN TRACE MINERAL ANALYSIS
APPLICATION OF COLORIMETRY IN TRACE MINERAL ANALYSISDr. IRSHAD A
 
beer lamberts law, colorimetery, nephlometry and turbidimetry
beer lamberts law, colorimetery, nephlometry and turbidimetrybeer lamberts law, colorimetery, nephlometry and turbidimetry
beer lamberts law, colorimetery, nephlometry and turbidimetryakansha singh
 
Colorimeter and spectrophotometer, Mass Spectrometer
Colorimeter and spectrophotometer, Mass SpectrometerColorimeter and spectrophotometer, Mass Spectrometer
Colorimeter and spectrophotometer, Mass Spectrometerprachann
 
Spectrophotometer
SpectrophotometerSpectrophotometer
SpectrophotometerAMASM
 
YSI Colorimeter Products Intro Webinar
YSI Colorimeter Products Intro WebinarYSI Colorimeter Products Intro Webinar
YSI Colorimeter Products Intro WebinarXylem Inc.
 
Analytical instruments introduction
Analytical instruments introductionAnalytical instruments introduction
Analytical instruments introductionSadiq Rahim
 
performance and specifications of spectrophotometer
performance and specifications of spectrophotometerperformance and specifications of spectrophotometer
performance and specifications of spectrophotometerPulak Das
 

What's hot (20)

Spectrophotometry in clinical chemistry
Spectrophotometry in clinical chemistrySpectrophotometry in clinical chemistry
Spectrophotometry in clinical chemistry
 
Spectrophotometry
SpectrophotometrySpectrophotometry
Spectrophotometry
 
Spectrophotometry lecture
Spectrophotometry lectureSpectrophotometry lecture
Spectrophotometry lecture
 
Colorimeter pdf
Colorimeter pdfColorimeter pdf
Colorimeter pdf
 
How to use a colorimeter
How to use a colorimeterHow to use a colorimeter
How to use a colorimeter
 
Spectrophotometer
SpectrophotometerSpectrophotometer
Spectrophotometer
 
Colorimetry
ColorimetryColorimetry
Colorimetry
 
Turbidity usp
Turbidity uspTurbidity usp
Turbidity usp
 
APPLICATION OF COLORIMETRY IN TRACE MINERAL ANALYSIS
APPLICATION OF COLORIMETRY IN TRACE MINERAL ANALYSISAPPLICATION OF COLORIMETRY IN TRACE MINERAL ANALYSIS
APPLICATION OF COLORIMETRY IN TRACE MINERAL ANALYSIS
 
beer lamberts law, colorimetery, nephlometry and turbidimetry
beer lamberts law, colorimetery, nephlometry and turbidimetrybeer lamberts law, colorimetery, nephlometry and turbidimetry
beer lamberts law, colorimetery, nephlometry and turbidimetry
 
Colorimeter and spectrophotometer, Mass Spectrometer
Colorimeter and spectrophotometer, Mass SpectrometerColorimeter and spectrophotometer, Mass Spectrometer
Colorimeter and spectrophotometer, Mass Spectrometer
 
colorimeter
colorimetercolorimeter
colorimeter
 
Spectrophotometer
SpectrophotometerSpectrophotometer
Spectrophotometer
 
Colorimetry kk
Colorimetry kkColorimetry kk
Colorimetry kk
 
YSI Colorimeter Products Intro Webinar
YSI Colorimeter Products Intro WebinarYSI Colorimeter Products Intro Webinar
YSI Colorimeter Products Intro Webinar
 
Spectrophotometry
SpectrophotometrySpectrophotometry
Spectrophotometry
 
Analytical instruments introduction
Analytical instruments introductionAnalytical instruments introduction
Analytical instruments introduction
 
Spectrophotometry
SpectrophotometrySpectrophotometry
Spectrophotometry
 
Colorimetry
ColorimetryColorimetry
Colorimetry
 
performance and specifications of spectrophotometer
performance and specifications of spectrophotometerperformance and specifications of spectrophotometer
performance and specifications of spectrophotometer
 

Similar to Chapter 8

Spectrophotometry & Calorimetry.pptx
Spectrophotometry & Calorimetry.pptxSpectrophotometry & Calorimetry.pptx
Spectrophotometry & Calorimetry.pptxdrarunsingh4
 
Spectrophotometer.pptx
Spectrophotometer.pptxSpectrophotometer.pptx
Spectrophotometer.pptxDeepa B
 
biophysics 1 spectroscopy
biophysics 1 spectroscopy biophysics 1 spectroscopy
biophysics 1 spectroscopy MSCW Mysore
 
Colorimetry and atomic absorption spectroscopy
Colorimetry and atomic absorption spectroscopy Colorimetry and atomic absorption spectroscopy
Colorimetry and atomic absorption spectroscopy O.P PARIHAR
 
Ismail Tahasildar
Ismail Tahasildar Ismail Tahasildar
Ismail Tahasildar dharwad
 
UV ray spectrophotometer
UV ray spectrophotometerUV ray spectrophotometer
UV ray spectrophotometerGoa App
 
Spectrophoto meter
Spectrophoto meterSpectrophoto meter
Spectrophoto meterNasir Nazeer
 
instrumenation all.pptx
instrumenation all.pptxinstrumenation all.pptx
instrumenation all.pptxreshadnuredin1
 
UV visible Specrosopy and applications
UV visible Specrosopy and applicationsUV visible Specrosopy and applications
UV visible Specrosopy and applicationsNAFIHHASSANMEHABANI
 
UV rays
UV rays UV rays
UV rays Goa App
 
INTRODUCTION TO COLORIMETRY and its basics.ppt
INTRODUCTION TO COLORIMETRY and its basics.pptINTRODUCTION TO COLORIMETRY and its basics.ppt
INTRODUCTION TO COLORIMETRY and its basics.pptSheelaS18
 
photometry.pptx
photometry.pptxphotometry.pptx
photometry.pptxRAJNKIT
 
COLORIMETER & LAMBERTS – BEER’S LAW.pptx
COLORIMETER & LAMBERTS – BEER’S LAW.pptxCOLORIMETER & LAMBERTS – BEER’S LAW.pptx
COLORIMETER & LAMBERTS – BEER’S LAW.pptxRAJNKIT
 

Similar to Chapter 8 (20)

Molecular Spectroscopy
Molecular SpectroscopyMolecular Spectroscopy
Molecular Spectroscopy
 
Spectrophotometry & Calorimetry.pptx
Spectrophotometry & Calorimetry.pptxSpectrophotometry & Calorimetry.pptx
Spectrophotometry & Calorimetry.pptx
 
Spectrophotometer.pptx
Spectrophotometer.pptxSpectrophotometer.pptx
Spectrophotometer.pptx
 
biophysics 1 spectroscopy
biophysics 1 spectroscopy biophysics 1 spectroscopy
biophysics 1 spectroscopy
 
Colorimetry and atomic absorption spectroscopy
Colorimetry and atomic absorption spectroscopy Colorimetry and atomic absorption spectroscopy
Colorimetry and atomic absorption spectroscopy
 
Ismail Tahasildar
Ismail Tahasildar Ismail Tahasildar
Ismail Tahasildar
 
UV ray spectrophotometer
UV ray spectrophotometerUV ray spectrophotometer
UV ray spectrophotometer
 
Spectrophotometry
SpectrophotometrySpectrophotometry
Spectrophotometry
 
Spectrophoto meter
Spectrophoto meterSpectrophoto meter
Spectrophoto meter
 
instrumenation all.pptx
instrumenation all.pptxinstrumenation all.pptx
instrumenation all.pptx
 
UV visible Specrosopy and applications
UV visible Specrosopy and applicationsUV visible Specrosopy and applications
UV visible Specrosopy and applications
 
UV Visible Spectrometry
UV Visible SpectrometryUV Visible Spectrometry
UV Visible Spectrometry
 
P pt priyanka
P pt priyankaP pt priyanka
P pt priyanka
 
UV rays
UV rays UV rays
UV rays
 
INTRODUCTION TO COLORIMETRY and its basics.ppt
INTRODUCTION TO COLORIMETRY and its basics.pptINTRODUCTION TO COLORIMETRY and its basics.ppt
INTRODUCTION TO COLORIMETRY and its basics.ppt
 
photometry.pptx
photometry.pptxphotometry.pptx
photometry.pptx
 
UV MV & MD.pptx
UV MV & MD.pptxUV MV & MD.pptx
UV MV & MD.pptx
 
COLORIMETER & LAMBERTS – BEER’S LAW.pptx
COLORIMETER & LAMBERTS – BEER’S LAW.pptxCOLORIMETER & LAMBERTS – BEER’S LAW.pptx
COLORIMETER & LAMBERTS – BEER’S LAW.pptx
 
Spectroscopy.pptx
Spectroscopy.pptxSpectroscopy.pptx
Spectroscopy.pptx
 
Prabhakar singh ii sem-paper v-colorimeter & spectrophotometer
Prabhakar singh  ii sem-paper v-colorimeter & spectrophotometerPrabhakar singh  ii sem-paper v-colorimeter & spectrophotometer
Prabhakar singh ii sem-paper v-colorimeter & spectrophotometer
 

More from ECRD IN

Spectrophotometer Meter
Spectrophotometer MeterSpectrophotometer Meter
Spectrophotometer MeterECRD IN
 
Quality Assurance and Quality Control
Quality Assurance and Quality ControlQuality Assurance and Quality Control
Quality Assurance and Quality ControlECRD IN
 
Basic Chemistry Concepts
Basic Chemistry ConceptsBasic Chemistry Concepts
Basic Chemistry ConceptsECRD IN
 
Measurement of Respirable Suspended Pm10 Particles
Measurement of Respirable Suspended Pm10 ParticlesMeasurement of Respirable Suspended Pm10 Particles
Measurement of Respirable Suspended Pm10 ParticlesECRD IN
 
Measurement of Hydrogen Ion Concentration (pH)
Measurement of Hydrogen Ion Concentration (pH)Measurement of Hydrogen Ion Concentration (pH)
Measurement of Hydrogen Ion Concentration (pH)ECRD IN
 
Determination of Suspended PM in Atmosphere
Determination of Suspended PM in AtmosphereDetermination of Suspended PM in Atmosphere
Determination of Suspended PM in AtmosphereECRD IN
 
Basic Statistics Concepts
Basic Statistics ConceptsBasic Statistics Concepts
Basic Statistics ConceptsECRD IN
 
Air Pollution, Sources and Characteristics
Air Pollution, Sources and CharacteristicsAir Pollution, Sources and Characteristics
Air Pollution, Sources and CharacteristicsECRD IN
 
Air Pollution Its Origin and Effects
Air Pollution Its Origin and EffectsAir Pollution Its Origin and Effects
Air Pollution Its Origin and EffectsECRD IN
 
Air Pollution and Meteorology
Air Pollution and MeteorologyAir Pollution and Meteorology
Air Pollution and MeteorologyECRD IN
 
Dust-Caused Respiratory Occupational Diseases
Dust-Caused Respiratory Occupational DiseasesDust-Caused Respiratory Occupational Diseases
Dust-Caused Respiratory Occupational DiseasesECRD IN
 
Air Quality Monitoring
Air Quality MonitoringAir Quality Monitoring
Air Quality MonitoringECRD IN
 
Approach & Strategy to Meet new Ambient Air Quality Standards
Approach & Strategy to Meet new Ambient Air Quality StandardsApproach & Strategy to Meet new Ambient Air Quality Standards
Approach & Strategy to Meet new Ambient Air Quality StandardsECRD IN
 
Monitoring CO -NDIR Method
Monitoring CO -NDIR MethodMonitoring CO -NDIR Method
Monitoring CO -NDIR MethodECRD IN
 
Appropriate Instruments & techniques for Complying with Air Quality Standards
Appropriate Instruments & techniques for Complying with Air Quality StandardsAppropriate Instruments & techniques for Complying with Air Quality Standards
Appropriate Instruments & techniques for Complying with Air Quality StandardsECRD IN
 
Techniques of Measurement of Organic Pollutants
Techniques of Measurement of Organic PollutantsTechniques of Measurement of Organic Pollutants
Techniques of Measurement of Organic PollutantsECRD IN
 
Envirotech APM 460 BL
Envirotech APM 460 BLEnvirotech APM 460 BL
Envirotech APM 460 BLECRD IN
 
Monitoring of Gaseous Pollutants
Monitoring of Gaseous PollutantsMonitoring of Gaseous Pollutants
Monitoring of Gaseous PollutantsECRD IN
 
Filter Extraction of Heavy Metal & Benzo Pyrene
Filter Extraction of Heavy Metal & Benzo PyreneFilter Extraction of Heavy Metal & Benzo Pyrene
Filter Extraction of Heavy Metal & Benzo PyreneECRD IN
 
Maintenance of High Volume Dust Repairable Sampler
Maintenance of High Volume Dust Repairable SamplerMaintenance of High Volume Dust Repairable Sampler
Maintenance of High Volume Dust Repairable SamplerECRD IN
 

More from ECRD IN (20)

Spectrophotometer Meter
Spectrophotometer MeterSpectrophotometer Meter
Spectrophotometer Meter
 
Quality Assurance and Quality Control
Quality Assurance and Quality ControlQuality Assurance and Quality Control
Quality Assurance and Quality Control
 
Basic Chemistry Concepts
Basic Chemistry ConceptsBasic Chemistry Concepts
Basic Chemistry Concepts
 
Measurement of Respirable Suspended Pm10 Particles
Measurement of Respirable Suspended Pm10 ParticlesMeasurement of Respirable Suspended Pm10 Particles
Measurement of Respirable Suspended Pm10 Particles
 
Measurement of Hydrogen Ion Concentration (pH)
Measurement of Hydrogen Ion Concentration (pH)Measurement of Hydrogen Ion Concentration (pH)
Measurement of Hydrogen Ion Concentration (pH)
 
Determination of Suspended PM in Atmosphere
Determination of Suspended PM in AtmosphereDetermination of Suspended PM in Atmosphere
Determination of Suspended PM in Atmosphere
 
Basic Statistics Concepts
Basic Statistics ConceptsBasic Statistics Concepts
Basic Statistics Concepts
 
Air Pollution, Sources and Characteristics
Air Pollution, Sources and CharacteristicsAir Pollution, Sources and Characteristics
Air Pollution, Sources and Characteristics
 
Air Pollution Its Origin and Effects
Air Pollution Its Origin and EffectsAir Pollution Its Origin and Effects
Air Pollution Its Origin and Effects
 
Air Pollution and Meteorology
Air Pollution and MeteorologyAir Pollution and Meteorology
Air Pollution and Meteorology
 
Dust-Caused Respiratory Occupational Diseases
Dust-Caused Respiratory Occupational DiseasesDust-Caused Respiratory Occupational Diseases
Dust-Caused Respiratory Occupational Diseases
 
Air Quality Monitoring
Air Quality MonitoringAir Quality Monitoring
Air Quality Monitoring
 
Approach & Strategy to Meet new Ambient Air Quality Standards
Approach & Strategy to Meet new Ambient Air Quality StandardsApproach & Strategy to Meet new Ambient Air Quality Standards
Approach & Strategy to Meet new Ambient Air Quality Standards
 
Monitoring CO -NDIR Method
Monitoring CO -NDIR MethodMonitoring CO -NDIR Method
Monitoring CO -NDIR Method
 
Appropriate Instruments & techniques for Complying with Air Quality Standards
Appropriate Instruments & techniques for Complying with Air Quality StandardsAppropriate Instruments & techniques for Complying with Air Quality Standards
Appropriate Instruments & techniques for Complying with Air Quality Standards
 
Techniques of Measurement of Organic Pollutants
Techniques of Measurement of Organic PollutantsTechniques of Measurement of Organic Pollutants
Techniques of Measurement of Organic Pollutants
 
Envirotech APM 460 BL
Envirotech APM 460 BLEnvirotech APM 460 BL
Envirotech APM 460 BL
 
Monitoring of Gaseous Pollutants
Monitoring of Gaseous PollutantsMonitoring of Gaseous Pollutants
Monitoring of Gaseous Pollutants
 
Filter Extraction of Heavy Metal & Benzo Pyrene
Filter Extraction of Heavy Metal & Benzo PyreneFilter Extraction of Heavy Metal & Benzo Pyrene
Filter Extraction of Heavy Metal & Benzo Pyrene
 
Maintenance of High Volume Dust Repairable Sampler
Maintenance of High Volume Dust Repairable SamplerMaintenance of High Volume Dust Repairable Sampler
Maintenance of High Volume Dust Repairable Sampler
 

Recently uploaded

(PARI) Viman Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
(PARI) Viman Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...(PARI) Viman Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
(PARI) Viman Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...ranjana rawat
 
Mumbai Call Girls, 💞 Prity 9892124323, Navi Mumbai Call girls
Mumbai Call Girls, 💞  Prity 9892124323, Navi Mumbai Call girlsMumbai Call Girls, 💞  Prity 9892124323, Navi Mumbai Call girls
Mumbai Call Girls, 💞 Prity 9892124323, Navi Mumbai Call girlsPooja Nehwal
 
webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...
webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...
webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...Cluster TWEED
 
Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000
Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000
Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000Sapana Sha
 
(DIYA) Call Girls Sinhagad Road ( 7001035870 ) HI-Fi Pune Escorts Service
(DIYA) Call Girls Sinhagad Road ( 7001035870 ) HI-Fi Pune Escorts Service(DIYA) Call Girls Sinhagad Road ( 7001035870 ) HI-Fi Pune Escorts Service
(DIYA) Call Girls Sinhagad Road ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
Spiders by Slidesgo - an introduction to arachnids
Spiders by Slidesgo - an introduction to arachnidsSpiders by Slidesgo - an introduction to arachnids
Spiders by Slidesgo - an introduction to arachnidsprasan26
 
Call Girls Mumbai Gayatri 8617697112 Independent Escort Service Mumbai
Call Girls Mumbai Gayatri 8617697112 Independent Escort Service MumbaiCall Girls Mumbai Gayatri 8617697112 Independent Escort Service Mumbai
Call Girls Mumbai Gayatri 8617697112 Independent Escort Service MumbaiCall girls in Ahmedabad High profile
 
Low Rate Call Girls Bikaner Anika 8250192130 Independent Escort Service Bikaner
Low Rate Call Girls Bikaner Anika 8250192130 Independent Escort Service BikanerLow Rate Call Girls Bikaner Anika 8250192130 Independent Escort Service Bikaner
Low Rate Call Girls Bikaner Anika 8250192130 Independent Escort Service BikanerSuhani Kapoor
 
VIP Call Girls Service Bandlaguda Hyderabad Call +91-8250192130
VIP Call Girls Service Bandlaguda Hyderabad Call +91-8250192130VIP Call Girls Service Bandlaguda Hyderabad Call +91-8250192130
VIP Call Girls Service Bandlaguda Hyderabad Call +91-8250192130Suhani Kapoor
 
Call Girls Service Nagpur Aditi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Aditi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Aditi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Aditi Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
VIP Call Girls Ramanthapur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...
VIP Call Girls Ramanthapur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...VIP Call Girls Ramanthapur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...
VIP Call Girls Ramanthapur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...Suhani Kapoor
 
9873940964 High Profile Call Girls Delhi |Defence Colony ( MAYA CHOPRA ) DE...
9873940964 High Profile  Call Girls  Delhi |Defence Colony ( MAYA CHOPRA ) DE...9873940964 High Profile  Call Girls  Delhi |Defence Colony ( MAYA CHOPRA ) DE...
9873940964 High Profile Call Girls Delhi |Defence Colony ( MAYA CHOPRA ) DE...Delhi Escorts
 
(NANDITA) Hadapsar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
(NANDITA) Hadapsar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...(NANDITA) Hadapsar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
(NANDITA) Hadapsar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...ranjana rawat
 
(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service
(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service
(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...
VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...
VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...Call Girls in Nagpur High Profile
 
Call Girls In Faridabad(Ballabgarh) Book ☎ 8168257667, @4999
Call Girls In Faridabad(Ballabgarh) Book ☎ 8168257667, @4999Call Girls In Faridabad(Ballabgarh) Book ☎ 8168257667, @4999
Call Girls In Faridabad(Ballabgarh) Book ☎ 8168257667, @4999Tina Ji
 
VIP Kolkata Call Girl Kalighat 👉 8250192130 Available With Room
VIP Kolkata Call Girl Kalighat 👉 8250192130  Available With RoomVIP Kolkata Call Girl Kalighat 👉 8250192130  Available With Room
VIP Kolkata Call Girl Kalighat 👉 8250192130 Available With Roomdivyansh0kumar0
 
(ANIKA) Call Girls Wagholi ( 7001035870 ) HI-Fi Pune Escorts Service
(ANIKA) Call Girls Wagholi ( 7001035870 ) HI-Fi Pune Escorts Service(ANIKA) Call Girls Wagholi ( 7001035870 ) HI-Fi Pune Escorts Service
(ANIKA) Call Girls Wagholi ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 

Recently uploaded (20)

(PARI) Viman Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
(PARI) Viman Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...(PARI) Viman Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
(PARI) Viman Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
 
Mumbai Call Girls, 💞 Prity 9892124323, Navi Mumbai Call girls
Mumbai Call Girls, 💞  Prity 9892124323, Navi Mumbai Call girlsMumbai Call Girls, 💞  Prity 9892124323, Navi Mumbai Call girls
Mumbai Call Girls, 💞 Prity 9892124323, Navi Mumbai Call girls
 
webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...
webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...
webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...
 
Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000
Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000
Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000
 
(DIYA) Call Girls Sinhagad Road ( 7001035870 ) HI-Fi Pune Escorts Service
(DIYA) Call Girls Sinhagad Road ( 7001035870 ) HI-Fi Pune Escorts Service(DIYA) Call Girls Sinhagad Road ( 7001035870 ) HI-Fi Pune Escorts Service
(DIYA) Call Girls Sinhagad Road ( 7001035870 ) HI-Fi Pune Escorts Service
 
Call Girls In { Delhi } South Extension Whatsup 9873940964 Enjoy Unlimited Pl...
Call Girls In { Delhi } South Extension Whatsup 9873940964 Enjoy Unlimited Pl...Call Girls In { Delhi } South Extension Whatsup 9873940964 Enjoy Unlimited Pl...
Call Girls In { Delhi } South Extension Whatsup 9873940964 Enjoy Unlimited Pl...
 
Spiders by Slidesgo - an introduction to arachnids
Spiders by Slidesgo - an introduction to arachnidsSpiders by Slidesgo - an introduction to arachnids
Spiders by Slidesgo - an introduction to arachnids
 
Call Girls Mumbai Gayatri 8617697112 Independent Escort Service Mumbai
Call Girls Mumbai Gayatri 8617697112 Independent Escort Service MumbaiCall Girls Mumbai Gayatri 8617697112 Independent Escort Service Mumbai
Call Girls Mumbai Gayatri 8617697112 Independent Escort Service Mumbai
 
Model Call Girl in Rajiv Chowk Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Rajiv Chowk Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Rajiv Chowk Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Rajiv Chowk Delhi reach out to us at 🔝9953056974🔝
 
Low Rate Call Girls Bikaner Anika 8250192130 Independent Escort Service Bikaner
Low Rate Call Girls Bikaner Anika 8250192130 Independent Escort Service BikanerLow Rate Call Girls Bikaner Anika 8250192130 Independent Escort Service Bikaner
Low Rate Call Girls Bikaner Anika 8250192130 Independent Escort Service Bikaner
 
VIP Call Girls Service Bandlaguda Hyderabad Call +91-8250192130
VIP Call Girls Service Bandlaguda Hyderabad Call +91-8250192130VIP Call Girls Service Bandlaguda Hyderabad Call +91-8250192130
VIP Call Girls Service Bandlaguda Hyderabad Call +91-8250192130
 
Call Girls Service Nagpur Aditi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Aditi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Aditi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Aditi Call 7001035870 Meet With Nagpur Escorts
 
VIP Call Girls Ramanthapur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...
VIP Call Girls Ramanthapur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...VIP Call Girls Ramanthapur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...
VIP Call Girls Ramanthapur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...
 
9873940964 High Profile Call Girls Delhi |Defence Colony ( MAYA CHOPRA ) DE...
9873940964 High Profile  Call Girls  Delhi |Defence Colony ( MAYA CHOPRA ) DE...9873940964 High Profile  Call Girls  Delhi |Defence Colony ( MAYA CHOPRA ) DE...
9873940964 High Profile Call Girls Delhi |Defence Colony ( MAYA CHOPRA ) DE...
 
(NANDITA) Hadapsar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
(NANDITA) Hadapsar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...(NANDITA) Hadapsar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
(NANDITA) Hadapsar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
 
(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service
(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service
(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service
 
VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...
VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...
VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...
 
Call Girls In Faridabad(Ballabgarh) Book ☎ 8168257667, @4999
Call Girls In Faridabad(Ballabgarh) Book ☎ 8168257667, @4999Call Girls In Faridabad(Ballabgarh) Book ☎ 8168257667, @4999
Call Girls In Faridabad(Ballabgarh) Book ☎ 8168257667, @4999
 
VIP Kolkata Call Girl Kalighat 👉 8250192130 Available With Room
VIP Kolkata Call Girl Kalighat 👉 8250192130  Available With RoomVIP Kolkata Call Girl Kalighat 👉 8250192130  Available With Room
VIP Kolkata Call Girl Kalighat 👉 8250192130 Available With Room
 
(ANIKA) Call Girls Wagholi ( 7001035870 ) HI-Fi Pune Escorts Service
(ANIKA) Call Girls Wagholi ( 7001035870 ) HI-Fi Pune Escorts Service(ANIKA) Call Girls Wagholi ( 7001035870 ) HI-Fi Pune Escorts Service
(ANIKA) Call Girls Wagholi ( 7001035870 ) HI-Fi Pune Escorts Service
 

Chapter 8

  • 1. Chapter 8 Spectrophotometer meter 1. Introduction Absorption Spectroscopic methods of analysis rank among the most widespread and powerful tools for quantitative analysis. The use of a spectrophotometer to determine the extent of absorption of various wavelengths of visible light by a given solution is commonly known as colorimetry. This method is used to determine concentrations of various chemicals which can give colours either directly or after addition of some other chemicals. As an example, in the analysis of phosphate, a reaction with orthophosphate is made, to form the highly coloured molybdenum blue compound. The light absorption of this compound can then be measured in a spectrophotometer. Some compounds absorb light in other than the visible range of the spectrum. For example, nitrates absorb radiation of 220 nm wave length in the UV region. 2. Theory Absorption Spectroscopic methods of analysis are based upon the fact that compounds ABSORB light radiation of a specific wavelength. In the analysis, the amount of light radiation absorbed by a sample is measured. The light absorption is directly related to the concentration of the coloured compound in the sample. The wavelength (X) of Maximum Absorption is known for different compounds. For example, the coloured compound formed for analysis of Phosphate (molybdenum blue) has maximum light absorption at a= 640 nm. Conversely, a minimum amount of light is transmitted through the compound at X= 640 nm. This is shown schematically in Figure .1 100 / s / 840 Wavefengift {nm} 400 640 Wavelength (nm) Figure 1: Light Absorption and Transmission by Phosphate-molybdenum blue compound. Schematic diagram showing maximum light absorption (and minimum light transmission) at A= 640 nm.
  • 2. The Beer-Lambert Law The Beer-Lambert Law is illustrated in Figure 2. The Absorbance (or optical density) and Transmission (or Transmittance) of light through a sample can be calculated by measuring light intensity entering and exiting the sample, Figure 2: Light energy of Intensity V passes through a sample with concentration 'C'. Some light energy is absorbed by the sample. The amount of light energy exiting the sample has Intensity T The following terms are defined: • Light Intensity entering a sample is "l0" • Light Intensity exiting a sample is "I" • The concentration of analyte in sample is "C" • The length of the light path in glass sample cuvette is "L". • "K" is a constant for a particular solution and wave length The Beer-Lambert Law is given by the following equations: Light Absorbance (A) = log (l0/1)= KCL Light Transmission (T) = l/l0 = 10"KCL Plots of absorbance and transmission versus concentration of the analyte in sample according to the above equations is shown in Figure 3, Concentration Concentration Figure 3: Beer-Lambert Law relates the amount of light Absorbance (A) by a solution to the Concentration (C) of a compound in solution and the length of light path: • As Concentration (C) increases, light Absorption (A) increases, linearly . • As Concentration (C) increases, light Transmission (T) decreases, exponentially
  • 3. 3. The Spectrophotometer Instrument All spectrophotometer instruments designed to measure the absorption of radiant energy have the basic components as follows (Figure 4): 1. a stable source of radiant energy (Light); 2. a wavelength selector to isolate a desired wavelength from the source (filter or monochromator); 3. transparent container (cuvette) for the sample and the blank; 4. a radiation detector (phototube) to convert the radiant energy received to a measurable signal; and a readout device that displays the signal from the detector. I inht Source Mnnor.hromator Samnle Phototube Measuring Absorbed / Transmitted Spectrophotometer Figure 4: Components of a spectrophotometer The energy source is to provide a stable source of light radiation, whereas the wavelength selector permits separation of radiation of the desired wavelength from other radiation. Light radiation passes through a glass container with sample. The detector measures the energy after it has passed through the sample. The readout device calculates the amount of light absorbed by the sample displays the signal from the detector as absorbance or transmission. The spectrophotometers which are used for such measurements may vary from simple and relatively inexpensive colorimeters to highly sophisticated and expensive instruments that automatically scan the ability of a solution to absorb radiation over a wide range of wavelengths and record the results of these measurements. One instrument cannot be used to measure absorbance at all wavelengths because a given energy source and energy detector is suitable for use over only a limited range of wavelengths. True linearity between absorbance and concentration according to Beer-Lambert Law requires the use of monochromatic light. In addition, a narrow band of light ensures a greater selectivity since substance with absorption peaks in other close by wavelengths are less likely to interfere. Further, it increases sensitivity as there is a greatest change in absorbance per increment of change in concentration.
  • 4. Both filters and monochromators are used to restrict the radiation wavelength. Photometers make use of filters, which function by absorbing large protions of the spectrum while transmitting relatively limited wavelength regions. Spectrophotometers are instruments equipped with monochromators that permit the continuous variation and selection of waveiength. The effective bandwidth of a monochromator that is satisfactory for most applications is about from 1 to 5 nm. The sample containers, cells or cuvettes, must be fabricated from material that is transparent to radiation in the spectral region of interest. The commonly used materials for different wave length regions are: Quartz or fused silica: UV to 2 ^m in I R Silicate glass: Above 350 nm to 2 jim in I R Plastic: visible region Polished NaCI or AgCI: Wave lengths longer than 2jLim Cuvettes or cells are provided in pairs that have been carefully matched to make possible the transmission through the solvent and the sample. Accurate spectrophotometric analysis requires the use of good quality, matched cells. These should be regularly checked against one another to detect differences that can arise from scratches, etching and wear. The most common cell path for UV-visible region is 1 cm. For reasons of economy, cylindrical cells are frequently used. Care must be taken to duplicate the position of such cells with respect to the light path; otherwise, variations in path length and in reflection losses will introduce errors. 4. General Measurement Procedures As explained above, the Beer-Lambert Law forms the basis of the measurement procedure. The amount of light radiation absorbed by a compound is directly related to the concentration of the compound. The general measurement procedure consists of 5 steps: 1. Prepare samples to make coloured compound 2. Make series of standard solutions of known concentrations and treat them in the same manner as the sample for making coloured compounds 3. Set spectrophotometer to X of maximum light absorption 4. Measure light absorbance of standards 5. Plot standard curve: Absorbance vs. Concentration, as shown in Figure 5 Once the standard plot is made, it is simple to find the concentration of an unknown sample: Measure the absorption of the unknown, and from the standard plot, read the related concentration (Figure 6).
  • 5. Cnnnpntration Figure 5: Plot of the standard curve: showing the linear relation between light absorption and concentration of the standards Figure 6: Finding the concentration of an unknown sample from the standard curve. Due to the fact that the overall composition of the sample is seldom the same as that of the calibration standard, in some cases, the absorption characteristics of the two may differ. Where such discrepancy is suspected, the standard addition approach may be used. Here, a known amount of analyte is added to a second aliquot of the sample. The difference in absorbance is used to calculate the analyte concentration of the sample as illustrated in Example 1. Example 1 A 25 ml_ sample after treatment with reagents to generate colour for measurement of phosphate yielded absorbance of 0.428. Addition of 1.00 ml_ of a solution containing 5.0|ag phosphorus to a second 25 mL aliquot and development of colour resulted in an absorbance of 0.517. Calculate jag phosphorus in each mL of sample.
  • 6. Solution: Correct absorbance for dilution: Corrected absorbance = 0.517 (26.0/25.0) = 0.538 Absorbance caused by 5|jg phosphorus = 0.538 - 0.428 = 0.110 Therefore, phosphorus in the sample = (5.0/0.11) 0.428 = 19.5[jg, or 19.5/25 = 0.7|jg/mL
  • 7. OPERATING INSTRUCTIONS FOR SPECTROPHOTOMETER 1. Ensure 230V, 50Hz, single-phase power supply is available in the 3 pin contact 5A power point, 2. Insert the power cord of the instrument in to a power point. 3. Turn the switch at 'ON" position. 4. Warm up the instrument for 15 minutes. 5. Press %T selector switch. 6. Adjust wavelength control to read the wavelength at which the test is desired. 7. Open the sliding lid of the sample compartment. 8. Turn the filter sliding control at filter position-1 to set zero adjustment. 9. Close the sliding lid of the sample compartment. 10. Turn the coarse and fine controls at their maximum and adjust %T control to get 0.00 displayed on the read out with the filter position-1. 11. Now select the filter position according to the desired wavelength: Filter Position Wave Length 2 below 395nm 3 395 to 600nm 4 above 600nm 12. Keep the blank cuvette at position-1. 13. Keep the sample cuvette at position -2. 14. Adjust 100%T (%T selector switch should be in pressed position) or 0.0 absorbance (O.D selector switch should be in pressed position) using coarse and fine controls knob with blank cuvette. 15. Pull the cuvette position control at position-2. 16. Read out the transmittance/absorbance of the sample.
  • 8. CHECKING THE PERFORMANCE OF THE SPECTROPHOTOMETER. 1. Preparation of Blank: - Measure accurately 10 ml of "Anal R" quality HCI of 36-46% concentration. Add it slowly to 400 ml distilled water filled in one litre volumetric flask. Make it up to mark with distilled water and mix it up well. 2. Preparation of sample Weigh exactly 22.2 gm of "Anal R" quality Cobalt Chloride(CoCI2 ).Transfer into one litre volumetric flask. Dissolve it in 1% Hydrochloric Acid (HCI) and make it up to the mark. 3. Fill the blank (1%HCI) & sample (CoCI2 Solution) solution in cuvettes. 4. Measure the transmittance/absorbance at wavelength 480 to 540 nm at 5nm interval. 5. Plot the graph of absorbance versus wavelength. The absorbance should increase up to 510-515nm wavelength and then decrease at higher wavelengths. Precautions 1. Always clean the cells thoroughly and rinse at least once with a portion of the sample, before filling the sample for measurement. 2. Always wipe the exposed surface of the cells dry and free from finger prints, using tissue paper. 3. Clean cells thoroughly both after and prior to use. 4. Do not leave solutions, particularly strong alkali, in the cells for periods more than an hour. 5. Ensure that no air bubbles stick to the inner surfaces of the cells, when taking measurements. 6. Never use a brush or any tool for cleaning which may scratch the optical surfaces. Cleaning Method Take 1gm of potassium dichromate, add a little distilled water and very slowly add approximately 100ml of conc.H2S04. Keep the cells in this acidic solution for a maximum period of 12hrs. Wash the cells thoroughly with distilled water.