Photometry techniques like colorimetry, spectrophotometry, and turbidometry measure the intensity of light absorbed or transmitted by a solution. Colorimeters contain a light source, monochromators/filters to select wavelengths, a sample holder (cuvette), photodetectors, and readout devices. The amount of light absorbed follows Beer's and Lambert's laws - absorption increases exponentially with concentration and path length. A colorimeter is used to quantify compounds in biological samples like blood and urine by measuring absorbance and relating it to a standard curve using the Beer-Lambert law. Colorimeters provide a simple and inexpensive way to perform quantitative analysis of colored compounds.
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Colorimetry class
1.
2. Introduction
• Photometry is the most common analytical technique
used in the biochemical laboratory. It is designed to
measure the intensity of a beam of light.
• Photometric principles are applied to the several kinds
of analytical techniques:
(a) where absorbed or transmitted light is measured:
• Colorimetry
• Spectrophotometry
• Atomic absorption, and
• Turbidometry
(b) where emitted light is measured:
• Flame emission photometry
3. Introduction (cont.)
• The components of most photoelectric colorimeters are
basically the same and the basic method of operation is
also similar for all the instruments.
• In analytical chemistry, Colorimetry is a technique
“used to determine the concentration of colored
compounds (analytes) in sample solution” at visible
spectrum of light (400 – 800 nm).
4. • Colorimeter is instrument which is used in the
measurement of the luminious intensity of light.
• Invented by Louis Jules Duboscq in 1870.
5. Colorimetry
Principle:
Colored solutions have the property of absorbing certain
wavelength of light when a monochromatic light is passed
through them.
• The amount of light absorbed or transmitted by a
colored solution is in accordance with two laws:
– Beer’s law
– Lambert’s law
6. Beer’s law :
• When a monochromatic light passes through a colored
solution, amount of light transmitted decreases
exponentially with increase in concentration of colored
substance.
– i.e. the amount of light absorbed by a colored solution is
directly proportion to the conc. Of substance in the colored
solution.
8. Lambert’s law :
• The amount of light transmitted decreases
exponentially with increase in path length (diameter) of
the cuvette or thickness of colored solution through
which light passes.
– i.e. the amount of light absorbed by a colored solution
depends on path length of cuvette or thickness or depth of
the colored solution.
10. Transmittance of a solution containing light
absorbing substance depends upon
1. The nature of light absorbing substance.
2. Wavelength of light and
3. Amount of light absorbing substance in
the light path, which in turn depends on
the concentration of light absorbing
substance and depth of the solution
through which light passes.
11. Preparation of solution for investigation
• In colorimetric estimation it is necessary to
prepare 3 solutions:
BLANK(B)
STANDARD(S)
TEST(T)
11/10/2016 12:05 PM
12. BLANK
To eliminate the effect of light
absorption by the reagent used
Water BLANK
Reagent BLANK
13. STANDARD
Solution of known concentration of
the substance
Both O.D and
concentration are
known
So concentration of
unknown can be
calculated
14. TEST
Test solution is made by
treating a specific volume of
the test sample with reagents
As per
procedure
15. Combined Beer’s- Lambert’s law
• Combined Beer’s- Lambert’s law is thus expressed as
amount of light transmitted through a colored solution
decreases exponentially with increases in conc. Of
colored solution & increase in conc. of colored solution
& increase in the path length of cuvette or thickness of
the colored solution.
• Combining the two laws:
A α C x L
A = K x C x L
Let AT=absorbance of the test solution
CT=concentration of the test solution
AS=absorbance of the standard solution
CS=concentration of the standard solution
16. AT
AS
K x CT x L
K x CS x L
=
AT
AS
CT
CS
=
CT =
AT
AS
x CS
AS = K x CS x LAT = K x CT x L
17. CT =
AT
AS
xCS
Concentration
of TEST sol.
Absorbance of TEST
Absorbance of STANDARD
Con. of STANDARDx=
Concentration
of TEST/100ml
Absorbance of TEST
Absorbance of STANDARD
Concn of Std X 100
x=
X ml
=
ODT
ODS
x CS
18. Concentration of
TEST /100ml
O.D of ‘T’- O.D of ‘B’
O.D of ‘S’- O.D of ‘B’
x=
Volume of ‘T’
Amount of ‘S’
Concentration of
TEST /100ml
x=
Volume of ‘T’
Amount of ‘S’T - B
S - B
x 100
x 100
19. Standard curve (calibration curve)
• The standard curve is prepared to check whether the
method of assaying a particular substance follows
Beer’s Law, i.e. whether the absorbance of the
substance increases in a linear way with its
concentration.
• The standard curve is constructed by plotting a vertical
axis (y – axis, ordinate) for optical densities
(absorbance) and a horizontal axis (x – axis, abscissa)
the concentration of standard solution.
• The concentration of the test/unknown can be
measured from the graph (standard curve).
20. Verification of Beer’s Law
• Prepare 1% standard solution of glucose, i.e. 1gm/dl
1000mg/100ml.
• Make different dilutions of standard solution using the general
formula given as following for obtaining different concentrations
of a solution by dilution with diluent (DW):
Tube no. Conc. (mg%) Amount of mL needed, (V1)
C1 x V1 = C2 x V2
DW (mL) Total vol. V2 (ml)
1 50 1000 x V1 = 50 x 2, V1= 0.1 1.9 2
2 100 1000 x V1 = 100 x 2, V1= 0.2 1.8 2
3 150 1000 x V1 = 150 x 2, V1= 0.3 1.7 2
4 200 1000 x V1 = 200 x 2, V1= 0.4 1.6 2
5 250 1000 x V1 = 250 x 2, V1= 0.5 1.5 2
6 300 1000 x V1 = 300 x 2, V1= 0.6 1.4 2
7 350 1000 x V1 = 350 x 2, V1= 0.7 1.3 2
8 400 1000 x V1 = 400 x 2, V1= 0.8 1.2 2
25. Complimentary color
• Wavelength between 400nm to 760 nm form the visible
spectrum of light
• Light passed through a solution which selectivity absorbs
radiation at fixed wave lengths,then the color of the
transmitted light is complementary to that of the absored
light.
26. Colors and complimentary colors of visible
spectrum
Color of the
solution/ solution
color transmitted
Filter used/ color
absorbed
Wavelength (nm)
Yellow blue Violet 380 – 430
Yellow Blue 430 – 475
Orange Green blue 475 – 495
Red Blue green 495 – 505
Purple Green 505 – 555
Violet Yellow green 555 – 575
Blue yellow 575 – 600
Green blue Orange 600 – 650
Blue green Red 650 - 750
29. Components of Colorimetry
1. Light source:
The light source is usually a tungten lamp, for
wavelength in the visible range (320 – 700nm) and a
deutarium or hydrogen lamps for ultraviolet light
(below 350nm).
a) Tungsten lamp Visible range
b) Deutarium/hydrogen lamp (preferred) UV Rays
c) Black body radiators (Nerst glower) Infrared radiations
30. Light source
1. Tungsten lamp:
filament mode of tungsten sealed in a glass envelope
Filed with inert gas.
Life time is limited due to gaseous tungsten formed by
sublimation.
31. Light source
Carbon arc lamp
• If sufficient intensity of light is not obtained
from tungsten lamp then carbon arc lamp can be
use as a source for color measurement.
32. Monochromators/Filters
• This is a means of selecting a sufficiently narrow wave
band
• Filter will absorb light of unwanted wavelength and
allow only monochromatic light to pass through.
– E.g.: a green filter absorbs all color, except green light which
is allowed to pass through.
Filter
Absorption filter Interference filter
Ex: Glass filter,Gelatin filter
33. Monochromators
• Early colorimeters used Absorption filters (i.e. glass
filter, Gelatin filter) that transmitted a wide segment of
spectrum (50nm or more).
• Newer instrument use Interface filters that consist of
thin layer of magnesium fluoride crystals with a
semitransparent coating of silver on each side.
• Monochromator consists of:
– Entrance slit
– Absorption/ interface filter and
– Prisms or diffraction grating for wavelength selection
– Exit slit
34. Sample Holder/ Cuvette
• Cuvettes are rectangular cell , square cell or
circular one.
• Made up of optical glass for visible wavelength
(quartz or fused silica for UV).
• Common one is square, rectangular to avoid
refraction artifacts.
• Optical path (length) of cuvette is always1cm.
• Capacity may be 3ml/2ml/1ml depending upon
the thickness of the wall of the cuvette.
• For accurate and precise reading cuvette must be
transparent, clean, devoid of any scratches and
there should be no bubble adhering to the inner
surface of the filled cuvette.
35.
36. Photosensitive detectors
• Detectors are the transducers, which convert light
energy to electrical enagery. A detector should be
possess follwing characteristics:
1.Should be sensitive
2.Should have linear response
3.Its noise level Should be low
4.Should have short response time
5.Should stable.
38. Read out devices
• The detector response can be measured by any
of the following devices:
a) Galvanometer
b) Ammeter
c) Recorder
d) Digital readout.
The signal may be transmitted to computer or print
out devices.
39. Criteria for satisfactory colorimetric
estimations
• Stability of color
Color may be fade of air oxidation,
photochemical decomposition, temperature.
• Intensity of color
The color of the solution should be intense in
order to detect small amount of constituents
and for making accurate result in low
concentration.
40. Criteria for satisfactory colorimetric
estimations
• Clarity of the solution
Substance under investigation should be completely
soluble in the solvent, since turbid solution,
suspension or colloidal solution absorb as well as
scatter light.
• Reproducibility
The intensity of the colored solution must be
reproducible. The effect of order of adding reagent,
pH and other variable should be clearly studied
41. Criteria for satisfactory colorimetric
estimations
• Specificity
Color produced should be specific for the desired
constituent. If other constituents interfere with color
reaction they be removed or prevented from or
prevented from functioning through appropriate
treatment like use of other coloring agent, altering
the oxidation state.
• Validity of Beer’s law
The intensity of color should be proportional to
concentration. It can be easily assessed by plotting
absorbance Vs concentration, where a straight line
passing through origin should be obtained.
42. Glass/gel filter is placed in the filter slot
3/4th of cuvette is filled with distilled water
and placed in the cuvette slot
Instrument is switched ‘on’ and allowed to
warm-up for 4-5 minutes
43. Button is adjusted using ‘coarse’ and ‘fine’ knobs to
give zero optical activity in the galvanometer
Blank solution is placed in an identical cuvette and
the OD is read (‘B’)
Blank solution is transferred to the original test tube
44. Test solution is taken in the same cuvette and O.D.
is read (‘T’)
Test solution is transferred back to the original test
tube
Standard solution is taken in same cuvette and O.D.
is read (‘S’)
Standard solution is transferred back to the test
tube
Cuvette is washed
47. • It is widely used in hospital & laboratory for
estimation of biochemical samples , like plasma,
serum, cerebrospinal fluid ( csf ) , urine.
• It is also used to quantitative estimation of
serum components as well as glucose, proteins
and other various biochemical compound.
• They are used by the food industry and by
manufacturers of paints and textiles
48. • They are used to test for water quality, by
screening for chemicals such as chlorine,
fluoride, cyanide, dissolved oxygen, iron,
molybdenum, zinc and hydrazine.
• They are also used to determine the
concentrations of plant nutrients (such as
phosphorus, nitrate and ammonia) in the soil or
hemoglobin in the blood and to identify
substandard and counterfeit drugs.
49. Advantage
It is inexpensive .
Very well applicable for quantitative analysis of
colored compounds.
Easily carriable and transportable.
50. Disadvantage
Cannot be used for colorless compounds.
It does not work in UV and IR regions.
We cannot set specific wavelength, as we have
to set a range as a parameter.
Similar colors from interfering substances can
produce errors in results .
51. Use, care and preventive
maintenance of a Colorimeter:
• Read the user manual carefully.
• Use the correct type of cuvette in the
colorimeter as recommended by the
manufacturer.
• Make sure that the cuvette is clean and it’s
optical surfaces are dry and free from finger
marks and scratches.
52. • Bring filter in to place before switching on the
colorimeter.
• Before reading the absorbance of a solution,
check that it is clear, there are no air bubbles in
it.
• Remove the cuvettes from the instrument when
not in use
53. • Clean the outside of the cuvette with tissue
paper to remove any marks from the optical
surfaces.
• To prolong the life of the lamp, switch off the
colorimeter after use.
• At the end of the day, disconnect It from the
main switch and cover the colorimeter with its
protective cover.
54. • At regular intervals check the mains power
adapter and cable for wear and tear and replace
if damaged.
• Keep in cool place away from corrosive
chemicals or fumes.
55. References
• Clinical chemistry –Michael bishop
• A book of medical science- J.ochei
• Practical biochemistry- Keith Wilson & john
walker
• Clinical chemistry &molecular diagnostics-Tietz