A large single center comparison
with the Aquilion ONE
Introduction
Coronary CT angiography using the Aquilion ONE
with its 320 detector rows and 16cm z-axis
coverage has been shown to have good diagnostic
accuracy to detect significant coronary stenoses
compared to invasive catheter angiography1–2
. In
patients with a heart rate of less than 65 bpm the
entire heart can be acquired in a single heartbeat
for a low radiation dose3
. The use of prospective
gating, where only a portion of the heart cycle is
imaged, SURE
Exposure and the novel iterative
reconstruction algorithm AIDR 3D all assist in
lowering the radiation dose as much as possible.
In patients with higher heart rates (>65 bpm) the
temporal resolution of 175 ms is not fast enough
to freeze cardiac motion, thus multi-segment
reconstruction is necessary to improve temporal
resolution. As multiple beats need to be scanned,
the radiation dose is increased.
In 2012 Toshiba introduced the Aquilion ONE
ViSION Edition. In this system the gantry speed is
increased from 350 ms to 275 ms and thus the
temporal resolution of a scan acquired in a single
heart beat has improved to 137 ms. As a result,
this second generation 320-row scanner is ex-
pected to allow excellent coronary imaging across
a wider range of heart rates as compared to the
first generation scanner. To date two papers have
been published comparing the two systems4–5
.
In the study by Chen et al.4
coronary CT angio-
graphy was performed in 107 consecutive and
unselected patients, who were enrolled regardless
of heart rate, cardiac rhythm, body size, previous
cardiac history (such as coronary bypass grafting)
or clinical indication. The only contraindications
Gladys Lo, Hong Kong Sanatorium & Hospital, Hong Kong, China
Chloe Steveson, Toshiba Medical Systems Corporation, Otawara, Japan
Joanne Schuijf, Toshiba Medical Systems Europe, Zoetermeer, The Netherlands
Coronary CT Angiography with
reduced radiation dose using the
Aquilion ONE ViSION Edition
ONEn=250 ViSIONn=250 Pvalue
BMI – Mean (SD) 24.6 (3.5) 24.6 (3.5) 24.6 (3.5)
Underweight 10 (4) 9 (3.6) 0.815
Normal 157 (62.8) 153 (61.2) 0.712
Overweight 75 (30) 74 (29.6) 0.922
Obese 8 (3.2) 14 (5.6) 0.191
Heart Rate – mean (SD) 62.86 (0.8) 66.75 (1.1) 0.0046*
<50 bpm 35 (14) 19 (7.6) 0.021
51–60 bpm 93 (37.2) 88 (35.2) 0.642
61–75 bpm 84 (33.6) 96 (38.4) 0.264
75–90 bpm 29 (11.6) 28 (11.2) 0.888
>91 bpm 9 (3.6) 19 (7.6) 0.052
Arrhythmia
Yes 8 (3.2) 16 (6.4)
0.094
No 242 (96.8) 234 (93.6)
Table 1: Selected patient demographics
ONEn=250 ViSIONn=250 Pvalue
Number of Beats
1 155 (62) 195 (78) <0.001*
2 72 (18.8) 36 (14.4) <0.001*
3 20 (8) 15 (6) 0.381
4 3 (1.2) 4 (1.6) 1.000
kV
100 174 (69.6) 224 (89.6) <0.001*
120 71 (28.4) 26 (10.4) <0.001*
135 5 (2) 0 (0) 0.061
Radiation Dose – mean (SD) 2.71 (2.47) 2.07 (2.57) 0.0047*
Mean with arrhythmia patients
excluded (SD)
2.49 (1.81) 1.69 (1.52) <0.001*
<1mSv 34 (13.6) 83 (33.2) <0.001*
1 mSv 86 (34.4) 97 (38.8) 0.307
2 mSv 61 (24.4) 32 (12.8) 0.001*
3 mSv 21 (8.4) 11 (4.4) 0.068
>4 mSv 48 (19.2) 27 (3.2) 0.009*
Table 2: Selected scan parameters
2 Coronary CT Angiography with reduced radiation dose using the Aquilion ONE ViSION Edition
were pregnancy or poor renal function. This cohort
was compared with a similar consecutive cohort
imaged with the Aquilion ONE in 2010. The scans
performed on the Aquilion ONE did not use AIDR 3D
or SURE
Exposure. In this study the mean radiation
dose was reduced from 2.67 mSv with Aquilion
ONE to 0.93 mSv with the ViSION, representing
a 65% dose reduction. This significant reduction
is due to the implementation of many dose saving
features such as AIDR 3D and SURE
Exposure in
addition to the increased gantry rotation speed.
A similar comparison was performed by Tomizawa
et al.5
In this study however, a more restricted
population was investigated where the only variable
between the Aquilion ONE and Aquilion ONE ViSION
Edition patient cohorts (n=48 each) was the rota-
tion time. Also in this study, use of the Aquilion ONE
ViSION Edition resulted in a substantial reduction
in radiation dose. From a practical perspective, it is
important to establish whether these observations,
obtained in single academic expert centers, can
be generalized to the larger imaging community.
The current paper therefore compares the two
systems in a radiology department in a private
hospital setting with a higher patient throughput
and a larger number of patients with arrhythmia
than the two published papers.
Patients
We sought to compare the radiation dose reduction
for coronary CT angiography between the Aquilion
ONE and Aquilion ONE ViSION Edition systems when
the only variable was the rotation speed. 500
patients were retrospectively studied, 250 were
imaged with Aquilion ONE from September–
October 2012 and 250 imaged with the ViSION
Edition from December 2012–January 2013.
Patients presented with atypical chest pain, positive
exercise stress test or multiple risk factors for
coronary artery disease. Other parameters such
as iterative reconstruction and SURE
Exposure
settings were the same for both groups. Selected
patient demographics are shown in Table 1. The
BMI was not significantly different between the two
systems. The mean heart rate however was slightly
higher with the ViSION Edition. The percentage
of patients with a very slow heart rate decreased
and the percentage of patients with heart rates
between 61 and 75 bpm increased. The number
of patients with high heart rates also increased
slightly corresponding to an increase in the number
of patients with arrhythmia, though this was not
statistically significant. Overall the patients in each
group were similar. Generally, the cohort studied
0
50
100
150
200
250
1 2 3 4
62%
18.8%
14.4%
8% 6%
1.2% 1.6%
78%
Aquilion ONE
ViSION Edition
0
20
40
60
80
100
120
<1 1 2 3 >4
13.6%
34.4%
38.8%
24.4%
12.8%
8.4%
4.4%
19.2%
3.2%
33.2%
Aquilion ONE
ViSION Edition
Mean Effective Dose
ONE = 2.71 mSv
ViSION = 2.07 mSv
p= 0.0047
0
50
100
150
200
250
1 2 3 4
62%
18.8%
14.4%
8% 6%
1.2% 1.6%
78%
Aquilion ONE
ViSION Edition
0
20
40
60
80
100
120
<1 1 2 3 >4
13.6%
34.4%
38.8%
24.4%
12.8%
8.4%
4.4%
19.2%
3.2%
33.2%
Aquilion ONE
ViSION Edition
0
50
100
150
200
250
100 120 135
69.6%
28.4%
10.4%
2% 0%
89.6%
Aquilion ONE
ViSION Edition
Mean Effective Dose
ONE = 2.71 mSv
ViSION = 2.07 mSv
p= 0.0047
0
50
100
150
200
250
1 2 3 4
62%
18.8%
14.4%
8% 6%
1.2% 1.6%
78%
Aquilion ONE
ViSION Edition
0
20
40
60
80
100
120
<1 1 2 3 >4
13.6%
34.4%
38.8%
24.4%
12.8%
8.4%
4.4%
19.2%
3.2%
33.2%
Aquilion ONE
ViSION Edition
0
50
100
150
200
250
100 120 135
69.6%
28.4%
10.4%
2% 0%
89.6%
Aquilion ONE
ViSION Edition
Mean Effective Dose
ONE = 2.71 mSv
ViSION = 2.07 mSv
p= 0.0047
Figure 1: Number of Beats Figure 2: Radiation Dose
Figure 3: kV selected Figure 4a: 3D VR
Number of Beats
NumberofPatients
NumberofPatients
Radiation Dose (mSv)
kV selected
NumberofPatients
Coronary CT Angiography with reduced radiation dose using the Aquilion ONE ViSION Edition 3
Clinical Case History
This 68 year old woman with BMI 24.2 presented with a history of smoking,
dyslipidemia and hypertension. The heart rate at the time of scan was
70 bpm. A one beat scan was performed with 100 kV. The DLP was 39.0
and the effective dose was 0.5 mSv (k=0.014). No significant coronary
artery disease was demonstrated in this scan.
with the Aquilion ONE ViSION Edition had a higher
mean heart rate and an increased prevalence of
arrhythmia as compared to standard populations
typically studied with coronary CT angiography4
.
Scan Parameters
On both scanners, coronary CT angiography was
performed according to the recommended protocol.
Following a breath exercise, SURE
Cardio automatically
determined the number of beats and exposure
window; however the radiographer was allowed
to modify these settings if desired. SURE
Exposure
was used to select the mA and kV according to
the patient’s size. AIDR 3D iterative reconstruction
and phaseXact were applied to all reconstructions
to ensure the best motion free images at the
lowest dose. Selected scan parameters are shown
in Table 2.
Single beat imaging
The improved temporal resolution on the ViSION
Edition has important implications for the number
of patients that can be imaged in a single heart
beat. Whereas the previous generation scanner
allowed single-heartbeat scanning with heart
rates of up to approximately 65 bpm, with the
ViSION Edition patients with heart rates up to
75 bpm can be imaged in a single heart beat.
At this site, there was a statistically significant
increase in the number of one beat scans per-
formed on the ViSION Edition (78% compared
with 62%), and a corresponding decrease in the
number of 2 beat scans (Figure 1). As expected,
the number of three and four beat scans did not
change as the number of patients with arrhythmia
was similar between the two groups; generally
only patients experiencing arrhythmia during
the acquisition have more than 2 beats imaged.
Overall 92.4% of patients were imaged with
either a one or two beat scan on the ViSION
Edition compared to 80.8% with the Aquilion ONE.
Dose reduction
A radiation dose of less than 1 mSv was achieved
in significantly more patients, 33.2% on the ViSION
Edition compared with 13.6% on the Aquilion ONE.
On the other hand, a general reduction in the num-
ber of patients receiving over 2 mSv was seen as
well (Figure 2). Patients with heart rates between
65 and 75 bpm benefit most from the increased
gantry rotation speed on the ViSION Edition as they
are now imaged with one beat rather than two
beats. The improved temporal resolution has also
allowed the exposure window to be narrowed to a
single gantry rotation in patients with heart rates
Figure 4b: LAD
Figure 4c: LCx Figure 4d: RCA
ULTRASOUND CT MRI X-RAY SERVICES
www.toshiba-medical.eu
©Toshiba Medical Systems Corporation 2014 all rights reserved.
Design and specifications subject to change without notice.
03/2014 MWPCT0017EUC
Printed in Europe
less than 55 bpm. Accordingly, the radiation dose
has been further reduced for these patients too.
The mean radiation dose decreased with the
ViSION Edition (2.07 mSv compared to 2.71 mSv,
p=0.0047). The overall reduction in dose between
the Aquilion ONE and ViSION Edition was 23%.
When the patients with arrhythmia are excluded
from the analysis, similar to the Tomizawa study5
,
the dose reduction was 32%.
The ViSION Edition system is equipped with a
90 kW generator to provide sufficient exposure
required for the 275 ms gantry rotation speed.
The larger generator enables larger patients to be
scanned at 100 kV. In this study 89.6% of patients
were imaged at 100 kV with an average patient
BMI of 25 (Figure 3). Importantly, all these patients
had diagnostic image quality (Figure 4).
Conclusion
Aquilion ONE has proved to be an exceptionally
robust scanner for performing cardiac CT exami-
nations on a wide range of patients with varying
heart rates and rhythms. The ViSION Edition fur-
ther extends this strength by allowing single beat
coronary imaging in patients with heart rates up
to 75 bpm. This study has shown that significant
radiation dose reduction of 23% can be achieved
in a large patient population, echoing similar findings
in the two published papers4, 5
. Importantly, even
in unselected populations, the majority of patients
can now be scanned within a single heart beat,
providing important advantages both for patients
and imaging providers.
References
1. Dewey M, Zimmermann E, Deissenrieder F,
Laule M, Dübel HP, Schlattmann P, Knebel F,
Rutsch W, Hamm B. Noninvasive coronary
angiography by 320-row computed tomography
with lower radiation exposure and maintained
diagnostic accuracy: comparison of results
with cardiac catheterization in a head-to-head
pilot investigation. Circulation. 2009 Sep
8;120(10):867-75.
2. Gaudio C, Pelliccia F, Evangelista A, Tanzilli G,
Paravati V, Pannarale G, Pannitteri G, Barillà F,
Greco C, Franzoni F, Speziale G, Pasceri V.
320-row computed tomography coronary angi-
ography vs. conventional coronary angiography
in patients with suspected coronary artery dis-
ease: A systematic review and meta-analysis.
Int J Cardiol. 2013 Sep 30;168(2):1562-4
3. Hoe J, Toh KH. First experience with 320-row
multidetector CT coronary angiography scan-
ning with prospective electrocardiogram gating
to reduce radiation dose. J Cardiovasc Comput
Tomogr. 2009 Jul-Aug;3(4):257-61.
4. Chen MY, Shanbhag SM, Arai AE, Submillisievert
Median Radiation Dose for Coronary
Angiography with a Second-Generation
320-Detector Row CT Scanner in 107 Consec-
utive Patients, Radiology, 2013, 267(1):76-85
5. Tomizawa N, Maeda E, Akahane M, Torigoe R,
Kiryu S, Ohtomo K, Coronary CT angiography
using the second-generation 320-detector row
CT: assessment of image quality and radiation
dose in various heart rates compared with the
first-generation scanner, Int J Cardiovasc Imag-
ing. 2013 Oct;29(7):1613-8.

More Related Content

What's hot

Aspirin and plavix
Aspirin and plavixAspirin and plavix
Aspirin and plavix
maaaziz
 
Impact of previous stenting on the outcome of (2)
Impact of previous stenting on the outcome of (2)Impact of previous stenting on the outcome of (2)
Impact of previous stenting on the outcome of (2)
escts2012
 

What's hot (20)

Contemporary management of iliofemoral venous thrombosis
Contemporary management of iliofemoral venous thrombosisContemporary management of iliofemoral venous thrombosis
Contemporary management of iliofemoral venous thrombosis
 
Patel TM 201111
Patel TM 201111Patel TM 201111
Patel TM 201111
 
Aspirin and plavix
Aspirin and plavixAspirin and plavix
Aspirin and plavix
 
VIN 2012 - Pellegrino on ECMO
VIN 2012 - Pellegrino on ECMOVIN 2012 - Pellegrino on ECMO
VIN 2012 - Pellegrino on ECMO
 
Da Silva RL - AIMRADIAL 2015 - Spasmolytic
Da Silva RL - AIMRADIAL 2015 - SpasmolyticDa Silva RL - AIMRADIAL 2015 - Spasmolytic
Da Silva RL - AIMRADIAL 2015 - Spasmolytic
 
Scalone G - AIMRADIAL 2015 - Radial in heart transplant patients
Scalone G - AIMRADIAL 2015 - Radial in heart transplant patientsScalone G - AIMRADIAL 2015 - Radial in heart transplant patients
Scalone G - AIMRADIAL 2015 - Radial in heart transplant patients
 
CT Coronary Angiogram VS Cardiac Stress Test
CT Coronary Angiogram VS Cardiac Stress TestCT Coronary Angiogram VS Cardiac Stress Test
CT Coronary Angiogram VS Cardiac Stress Test
 
CT coronary angiography in ED chest pain patients
CT coronary angiography in ED chest pain patientsCT coronary angiography in ED chest pain patients
CT coronary angiography in ED chest pain patients
 
Patel TM 201110
Patel TM 201110Patel TM 201110
Patel TM 201110
 
Impact of previous stenting on the outcome of (2)
Impact of previous stenting on the outcome of (2)Impact of previous stenting on the outcome of (2)
Impact of previous stenting on the outcome of (2)
 
IWO bijeenkomst - 22 april Prof. Lems
IWO bijeenkomst - 22 april Prof. LemsIWO bijeenkomst - 22 april Prof. Lems
IWO bijeenkomst - 22 april Prof. Lems
 
Bernat I - AIMRADIAL 2013 - STEMI-RADIAL trial
Bernat I - AIMRADIAL 2013 - STEMI-RADIAL trialBernat I - AIMRADIAL 2013 - STEMI-RADIAL trial
Bernat I - AIMRADIAL 2013 - STEMI-RADIAL trial
 
Fu Q - AIMRADIAL 2014 - Left vs right radial approach
Fu Q - AIMRADIAL 2014 - Left vs right radial approachFu Q - AIMRADIAL 2014 - Left vs right radial approach
Fu Q - AIMRADIAL 2014 - Left vs right radial approach
 
The road ahead.
The road ahead.The road ahead.
The road ahead.
 
Detecting acute coronary syndrome
Detecting acute coronary syndromeDetecting acute coronary syndrome
Detecting acute coronary syndrome
 
Doppler Ultrasound of the Renal Arteries in Hypertensive Patients in Senegal:...
Doppler Ultrasound of the Renal Arteries in Hypertensive Patients in Senegal:...Doppler Ultrasound of the Renal Arteries in Hypertensive Patients in Senegal:...
Doppler Ultrasound of the Renal Arteries in Hypertensive Patients in Senegal:...
 
Pacchioni A - AIMRADIAL 2014 - Cerebral microembolism
Pacchioni A - AIMRADIAL 2014 - Cerebral microembolismPacchioni A - AIMRADIAL 2014 - Cerebral microembolism
Pacchioni A - AIMRADIAL 2014 - Cerebral microembolism
 
Intermittent claudication
Intermittent claudicationIntermittent claudication
Intermittent claudication
 
Kwok CS - AIMRADIAL 2014 - Gender, syndrome, clinical outcomes
Kwok CS - AIMRADIAL 2014 - Gender, syndrome, clinical outcomesKwok CS - AIMRADIAL 2014 - Gender, syndrome, clinical outcomes
Kwok CS - AIMRADIAL 2014 - Gender, syndrome, clinical outcomes
 
AACVPR 2002
AACVPR 2002AACVPR 2002
AACVPR 2002
 

Similar to White paper "coronary ct angiography with reduced radiation dose using the aquilion one vi sion edition"

The effects of_rosuvastatin_on_plaque_regression_i
The effects of_rosuvastatin_on_plaque_regression_iThe effects of_rosuvastatin_on_plaque_regression_i
The effects of_rosuvastatin_on_plaque_regression_i
19844
 
uptodate on acute kidney injury
uptodate on acute kidney injuryuptodate on acute kidney injury
uptodate on acute kidney injury
Sherif Mohammed
 
Strain cardíaco na avaliação da função cardíaca fetal
Strain cardíaco na avaliação da função cardíaca fetalStrain cardíaco na avaliação da função cardíaca fetal
Strain cardíaco na avaliação da função cardíaca fetal
gisa_legal
 

Similar to White paper "coronary ct angiography with reduced radiation dose using the aquilion one vi sion edition" (20)

The effects of_rosuvastatin_on_plaque_regression_i
The effects of_rosuvastatin_on_plaque_regression_iThe effects of_rosuvastatin_on_plaque_regression_i
The effects of_rosuvastatin_on_plaque_regression_i
 
Scientific news march 2015 samir rafla
Scientific news march 2015 samir raflaScientific news march 2015 samir rafla
Scientific news march 2015 samir rafla
 
Non Invasive testing of myocardial ischemia AA.pptx
Non Invasive testing of myocardial ischemia AA.pptxNon Invasive testing of myocardial ischemia AA.pptx
Non Invasive testing of myocardial ischemia AA.pptx
 
Poster vienne ecr
Poster vienne ecrPoster vienne ecr
Poster vienne ecr
 
Stable ischemic heart disease how is it different from acs..
Stable ischemic heart disease how is it different from acs..Stable ischemic heart disease how is it different from acs..
Stable ischemic heart disease how is it different from acs..
 
50 002
50 00250 002
50 002
 
caudal level of cervical fusion.ppt
caudal level of cervical fusion.pptcaudal level of cervical fusion.ppt
caudal level of cervical fusion.ppt
 
Romagnoli E 201305
Romagnoli E 201305Romagnoli E 201305
Romagnoli E 201305
 
A Speckle Tracking Echocardiographic Study for Correlation Between Global Lef...
A Speckle Tracking Echocardiographic Study for Correlation Between Global Lef...A Speckle Tracking Echocardiographic Study for Correlation Between Global Lef...
A Speckle Tracking Echocardiographic Study for Correlation Between Global Lef...
 
Percutaneous Valve implantation or Operation in aortic stenosis
Percutaneous Valve implantation or Operation in aortic stenosisPercutaneous Valve implantation or Operation in aortic stenosis
Percutaneous Valve implantation or Operation in aortic stenosis
 
Rol actual del cardiodesfibrilador implantable subcutáneo en la prevención de...
Rol actual del cardiodesfibrilador implantable subcutáneo en la prevención de...Rol actual del cardiodesfibrilador implantable subcutáneo en la prevención de...
Rol actual del cardiodesfibrilador implantable subcutáneo en la prevención de...
 
uptodate on acute kidney injury
uptodate on acute kidney injuryuptodate on acute kidney injury
uptodate on acute kidney injury
 
Samir rafla real life outcome of atrial fibrillation ablation-cardio alex 2014
Samir rafla  real life outcome of atrial fibrillation ablation-cardio alex 2014Samir rafla  real life outcome of atrial fibrillation ablation-cardio alex 2014
Samir rafla real life outcome of atrial fibrillation ablation-cardio alex 2014
 
Olivecrona GK et al.
Olivecrona GK et al.Olivecrona GK et al.
Olivecrona GK et al.
 
Ivus jc ultimate trial
Ivus jc ultimate trialIvus jc ultimate trial
Ivus jc ultimate trial
 
Chronic coronary syndromes
Chronic coronary syndromesChronic coronary syndromes
Chronic coronary syndromes
 
What to choose in stable CAD- Medical therapy only or PCI or CABG?
What to choose in stable CAD- Medical therapy only or PCI or CABG?What to choose in stable CAD- Medical therapy only or PCI or CABG?
What to choose in stable CAD- Medical therapy only or PCI or CABG?
 
Discovery ct750 hd book
Discovery ct750 hd bookDiscovery ct750 hd book
Discovery ct750 hd book
 
Sepsisgrandrounds
SepsisgrandroundsSepsisgrandrounds
Sepsisgrandrounds
 
Strain cardíaco na avaliação da função cardíaca fetal
Strain cardíaco na avaliação da função cardíaca fetalStrain cardíaco na avaliação da função cardíaca fetal
Strain cardíaco na avaliação da função cardíaca fetal
 

More from Canon Medical Systems Europe

Canon Medical Systems VISIONS Magazine - issue 32
Canon Medical Systems VISIONS Magazine - issue 32Canon Medical Systems VISIONS Magazine - issue 32
Canon Medical Systems VISIONS Magazine - issue 32
Canon Medical Systems Europe
 
Canon Medical Systems VISIONS Magazine - issue 30
Canon Medical Systems VISIONS Magazine - issue 30Canon Medical Systems VISIONS Magazine - issue 30
Canon Medical Systems VISIONS Magazine - issue 30
Canon Medical Systems Europe
 
Toshiba Medical's VISIONS Magazine - issue 29
Toshiba Medical's VISIONS Magazine - issue 29Toshiba Medical's VISIONS Magazine - issue 29
Toshiba Medical's VISIONS Magazine - issue 29
Canon Medical Systems Europe
 
Toshiba's VISIONS Magazine - issue 28
Toshiba's VISIONS Magazine - issue 28Toshiba's VISIONS Magazine - issue 28
Toshiba's VISIONS Magazine - issue 28
Canon Medical Systems Europe
 
Toshiba's VISIONS Magazine - issue 27
Toshiba's VISIONS Magazine - issue 27Toshiba's VISIONS Magazine - issue 27
Toshiba's VISIONS Magazine - issue 27
Canon Medical Systems Europe
 

More from Canon Medical Systems Europe (20)

Canon Medical Systems VISIONS Magazine - issue 37
Canon Medical Systems VISIONS Magazine - issue 37Canon Medical Systems VISIONS Magazine - issue 37
Canon Medical Systems VISIONS Magazine - issue 37
 
Canon Medical Systems VISIONS Magazine issue #36
Canon Medical Systems VISIONS Magazine issue #36Canon Medical Systems VISIONS Magazine issue #36
Canon Medical Systems VISIONS Magazine issue #36
 
VISIONS Veterinary Special
VISIONS Veterinary SpecialVISIONS Veterinary Special
VISIONS Veterinary Special
 
Canon Medical Systems VISIONS Magazine - issue 35
Canon Medical Systems VISIONS Magazine - issue 35Canon Medical Systems VISIONS Magazine - issue 35
Canon Medical Systems VISIONS Magazine - issue 35
 
VISIONS MR Special
VISIONS MR SpecialVISIONS MR Special
VISIONS MR Special
 
Canon Medical Systems VISIONS Magazine - issue 34
Canon Medical Systems VISIONS Magazine - issue 34Canon Medical Systems VISIONS Magazine - issue 34
Canon Medical Systems VISIONS Magazine - issue 34
 
Canon Medical Systems VISIONS Magazine - issue 33
Canon Medical Systems VISIONS Magazine - issue 33Canon Medical Systems VISIONS Magazine - issue 33
Canon Medical Systems VISIONS Magazine - issue 33
 
Canon Medical Systems VISIONS Magazine - issue 32
Canon Medical Systems VISIONS Magazine - issue 32Canon Medical Systems VISIONS Magazine - issue 32
Canon Medical Systems VISIONS Magazine - issue 32
 
Canon Medical Systems VISIONS Magazine - issue 31
Canon Medical Systems VISIONS Magazine - issue 31Canon Medical Systems VISIONS Magazine - issue 31
Canon Medical Systems VISIONS Magazine - issue 31
 
Canon Medical Systems VISIONS Magazine - issue 30
Canon Medical Systems VISIONS Magazine - issue 30Canon Medical Systems VISIONS Magazine - issue 30
Canon Medical Systems VISIONS Magazine - issue 30
 
Toshiba Medical's VISIONS Magazine - issue 29
Toshiba Medical's VISIONS Magazine - issue 29Toshiba Medical's VISIONS Magazine - issue 29
Toshiba Medical's VISIONS Magazine - issue 29
 
Toshiba's VISIONS Magazine - issue 28
Toshiba's VISIONS Magazine - issue 28Toshiba's VISIONS Magazine - issue 28
Toshiba's VISIONS Magazine - issue 28
 
Toshiba's VISIONS Magazine - issue 27
Toshiba's VISIONS Magazine - issue 27Toshiba's VISIONS Magazine - issue 27
Toshiba's VISIONS Magazine - issue 27
 
Exercise Myocardial Performance - Dr. Guido E. Pieles
Exercise Myocardial Performance - Dr. Guido E. PielesExercise Myocardial Performance - Dr. Guido E. Pieles
Exercise Myocardial Performance - Dr. Guido E. Pieles
 
Toshiba's VISIONS Magazine - issue 26
Toshiba's VISIONS Magazine - issue 26Toshiba's VISIONS Magazine - issue 26
Toshiba's VISIONS Magazine - issue 26
 
Lung SURESubtraction in Everyday Practice - Dr R. Bull
Lung SURESubtraction in Everyday Practice - Dr R. BullLung SURESubtraction in Everyday Practice - Dr R. Bull
Lung SURESubtraction in Everyday Practice - Dr R. Bull
 
Toshiba's VISIONS Magazine - issue 23
Toshiba's VISIONS Magazine - issue 23Toshiba's VISIONS Magazine - issue 23
Toshiba's VISIONS Magazine - issue 23
 
Toshiba's VISIONS Magazine - issue 25
Toshiba's VISIONS Magazine - issue 25Toshiba's VISIONS Magazine - issue 25
Toshiba's VISIONS Magazine - issue 25
 
Dual energy case study
Dual energy case studyDual energy case study
Dual energy case study
 
First Clinical Results of Coronary CT Subtraction
First Clinical Results of Coronary CT SubtractionFirst Clinical Results of Coronary CT Subtraction
First Clinical Results of Coronary CT Subtraction
 

Recently uploaded

Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Dipal Arora
 

Recently uploaded (20)

Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
 
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
 
Call Girls Haridwar Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Haridwar Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 8250077686 Top Class Call Girl Service Available
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
 
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
 

White paper "coronary ct angiography with reduced radiation dose using the aquilion one vi sion edition"

  • 1. A large single center comparison with the Aquilion ONE Introduction Coronary CT angiography using the Aquilion ONE with its 320 detector rows and 16cm z-axis coverage has been shown to have good diagnostic accuracy to detect significant coronary stenoses compared to invasive catheter angiography1–2 . In patients with a heart rate of less than 65 bpm the entire heart can be acquired in a single heartbeat for a low radiation dose3 . The use of prospective gating, where only a portion of the heart cycle is imaged, SURE Exposure and the novel iterative reconstruction algorithm AIDR 3D all assist in lowering the radiation dose as much as possible. In patients with higher heart rates (>65 bpm) the temporal resolution of 175 ms is not fast enough to freeze cardiac motion, thus multi-segment reconstruction is necessary to improve temporal resolution. As multiple beats need to be scanned, the radiation dose is increased. In 2012 Toshiba introduced the Aquilion ONE ViSION Edition. In this system the gantry speed is increased from 350 ms to 275 ms and thus the temporal resolution of a scan acquired in a single heart beat has improved to 137 ms. As a result, this second generation 320-row scanner is ex- pected to allow excellent coronary imaging across a wider range of heart rates as compared to the first generation scanner. To date two papers have been published comparing the two systems4–5 . In the study by Chen et al.4 coronary CT angio- graphy was performed in 107 consecutive and unselected patients, who were enrolled regardless of heart rate, cardiac rhythm, body size, previous cardiac history (such as coronary bypass grafting) or clinical indication. The only contraindications Gladys Lo, Hong Kong Sanatorium & Hospital, Hong Kong, China Chloe Steveson, Toshiba Medical Systems Corporation, Otawara, Japan Joanne Schuijf, Toshiba Medical Systems Europe, Zoetermeer, The Netherlands Coronary CT Angiography with reduced radiation dose using the Aquilion ONE ViSION Edition ONEn=250 ViSIONn=250 Pvalue BMI – Mean (SD) 24.6 (3.5) 24.6 (3.5) 24.6 (3.5) Underweight 10 (4) 9 (3.6) 0.815 Normal 157 (62.8) 153 (61.2) 0.712 Overweight 75 (30) 74 (29.6) 0.922 Obese 8 (3.2) 14 (5.6) 0.191 Heart Rate – mean (SD) 62.86 (0.8) 66.75 (1.1) 0.0046* <50 bpm 35 (14) 19 (7.6) 0.021 51–60 bpm 93 (37.2) 88 (35.2) 0.642 61–75 bpm 84 (33.6) 96 (38.4) 0.264 75–90 bpm 29 (11.6) 28 (11.2) 0.888 >91 bpm 9 (3.6) 19 (7.6) 0.052 Arrhythmia Yes 8 (3.2) 16 (6.4) 0.094 No 242 (96.8) 234 (93.6) Table 1: Selected patient demographics ONEn=250 ViSIONn=250 Pvalue Number of Beats 1 155 (62) 195 (78) <0.001* 2 72 (18.8) 36 (14.4) <0.001* 3 20 (8) 15 (6) 0.381 4 3 (1.2) 4 (1.6) 1.000 kV 100 174 (69.6) 224 (89.6) <0.001* 120 71 (28.4) 26 (10.4) <0.001* 135 5 (2) 0 (0) 0.061 Radiation Dose – mean (SD) 2.71 (2.47) 2.07 (2.57) 0.0047* Mean with arrhythmia patients excluded (SD) 2.49 (1.81) 1.69 (1.52) <0.001* <1mSv 34 (13.6) 83 (33.2) <0.001* 1 mSv 86 (34.4) 97 (38.8) 0.307 2 mSv 61 (24.4) 32 (12.8) 0.001* 3 mSv 21 (8.4) 11 (4.4) 0.068 >4 mSv 48 (19.2) 27 (3.2) 0.009* Table 2: Selected scan parameters
  • 2. 2 Coronary CT Angiography with reduced radiation dose using the Aquilion ONE ViSION Edition were pregnancy or poor renal function. This cohort was compared with a similar consecutive cohort imaged with the Aquilion ONE in 2010. The scans performed on the Aquilion ONE did not use AIDR 3D or SURE Exposure. In this study the mean radiation dose was reduced from 2.67 mSv with Aquilion ONE to 0.93 mSv with the ViSION, representing a 65% dose reduction. This significant reduction is due to the implementation of many dose saving features such as AIDR 3D and SURE Exposure in addition to the increased gantry rotation speed. A similar comparison was performed by Tomizawa et al.5 In this study however, a more restricted population was investigated where the only variable between the Aquilion ONE and Aquilion ONE ViSION Edition patient cohorts (n=48 each) was the rota- tion time. Also in this study, use of the Aquilion ONE ViSION Edition resulted in a substantial reduction in radiation dose. From a practical perspective, it is important to establish whether these observations, obtained in single academic expert centers, can be generalized to the larger imaging community. The current paper therefore compares the two systems in a radiology department in a private hospital setting with a higher patient throughput and a larger number of patients with arrhythmia than the two published papers. Patients We sought to compare the radiation dose reduction for coronary CT angiography between the Aquilion ONE and Aquilion ONE ViSION Edition systems when the only variable was the rotation speed. 500 patients were retrospectively studied, 250 were imaged with Aquilion ONE from September– October 2012 and 250 imaged with the ViSION Edition from December 2012–January 2013. Patients presented with atypical chest pain, positive exercise stress test or multiple risk factors for coronary artery disease. Other parameters such as iterative reconstruction and SURE Exposure settings were the same for both groups. Selected patient demographics are shown in Table 1. The BMI was not significantly different between the two systems. The mean heart rate however was slightly higher with the ViSION Edition. The percentage of patients with a very slow heart rate decreased and the percentage of patients with heart rates between 61 and 75 bpm increased. The number of patients with high heart rates also increased slightly corresponding to an increase in the number of patients with arrhythmia, though this was not statistically significant. Overall the patients in each group were similar. Generally, the cohort studied 0 50 100 150 200 250 1 2 3 4 62% 18.8% 14.4% 8% 6% 1.2% 1.6% 78% Aquilion ONE ViSION Edition 0 20 40 60 80 100 120 <1 1 2 3 >4 13.6% 34.4% 38.8% 24.4% 12.8% 8.4% 4.4% 19.2% 3.2% 33.2% Aquilion ONE ViSION Edition Mean Effective Dose ONE = 2.71 mSv ViSION = 2.07 mSv p= 0.0047 0 50 100 150 200 250 1 2 3 4 62% 18.8% 14.4% 8% 6% 1.2% 1.6% 78% Aquilion ONE ViSION Edition 0 20 40 60 80 100 120 <1 1 2 3 >4 13.6% 34.4% 38.8% 24.4% 12.8% 8.4% 4.4% 19.2% 3.2% 33.2% Aquilion ONE ViSION Edition 0 50 100 150 200 250 100 120 135 69.6% 28.4% 10.4% 2% 0% 89.6% Aquilion ONE ViSION Edition Mean Effective Dose ONE = 2.71 mSv ViSION = 2.07 mSv p= 0.0047 0 50 100 150 200 250 1 2 3 4 62% 18.8% 14.4% 8% 6% 1.2% 1.6% 78% Aquilion ONE ViSION Edition 0 20 40 60 80 100 120 <1 1 2 3 >4 13.6% 34.4% 38.8% 24.4% 12.8% 8.4% 4.4% 19.2% 3.2% 33.2% Aquilion ONE ViSION Edition 0 50 100 150 200 250 100 120 135 69.6% 28.4% 10.4% 2% 0% 89.6% Aquilion ONE ViSION Edition Mean Effective Dose ONE = 2.71 mSv ViSION = 2.07 mSv p= 0.0047 Figure 1: Number of Beats Figure 2: Radiation Dose Figure 3: kV selected Figure 4a: 3D VR Number of Beats NumberofPatients NumberofPatients Radiation Dose (mSv) kV selected NumberofPatients
  • 3. Coronary CT Angiography with reduced radiation dose using the Aquilion ONE ViSION Edition 3 Clinical Case History This 68 year old woman with BMI 24.2 presented with a history of smoking, dyslipidemia and hypertension. The heart rate at the time of scan was 70 bpm. A one beat scan was performed with 100 kV. The DLP was 39.0 and the effective dose was 0.5 mSv (k=0.014). No significant coronary artery disease was demonstrated in this scan. with the Aquilion ONE ViSION Edition had a higher mean heart rate and an increased prevalence of arrhythmia as compared to standard populations typically studied with coronary CT angiography4 . Scan Parameters On both scanners, coronary CT angiography was performed according to the recommended protocol. Following a breath exercise, SURE Cardio automatically determined the number of beats and exposure window; however the radiographer was allowed to modify these settings if desired. SURE Exposure was used to select the mA and kV according to the patient’s size. AIDR 3D iterative reconstruction and phaseXact were applied to all reconstructions to ensure the best motion free images at the lowest dose. Selected scan parameters are shown in Table 2. Single beat imaging The improved temporal resolution on the ViSION Edition has important implications for the number of patients that can be imaged in a single heart beat. Whereas the previous generation scanner allowed single-heartbeat scanning with heart rates of up to approximately 65 bpm, with the ViSION Edition patients with heart rates up to 75 bpm can be imaged in a single heart beat. At this site, there was a statistically significant increase in the number of one beat scans per- formed on the ViSION Edition (78% compared with 62%), and a corresponding decrease in the number of 2 beat scans (Figure 1). As expected, the number of three and four beat scans did not change as the number of patients with arrhythmia was similar between the two groups; generally only patients experiencing arrhythmia during the acquisition have more than 2 beats imaged. Overall 92.4% of patients were imaged with either a one or two beat scan on the ViSION Edition compared to 80.8% with the Aquilion ONE. Dose reduction A radiation dose of less than 1 mSv was achieved in significantly more patients, 33.2% on the ViSION Edition compared with 13.6% on the Aquilion ONE. On the other hand, a general reduction in the num- ber of patients receiving over 2 mSv was seen as well (Figure 2). Patients with heart rates between 65 and 75 bpm benefit most from the increased gantry rotation speed on the ViSION Edition as they are now imaged with one beat rather than two beats. The improved temporal resolution has also allowed the exposure window to be narrowed to a single gantry rotation in patients with heart rates Figure 4b: LAD Figure 4c: LCx Figure 4d: RCA
  • 4. ULTRASOUND CT MRI X-RAY SERVICES www.toshiba-medical.eu ©Toshiba Medical Systems Corporation 2014 all rights reserved. Design and specifications subject to change without notice. 03/2014 MWPCT0017EUC Printed in Europe less than 55 bpm. Accordingly, the radiation dose has been further reduced for these patients too. The mean radiation dose decreased with the ViSION Edition (2.07 mSv compared to 2.71 mSv, p=0.0047). The overall reduction in dose between the Aquilion ONE and ViSION Edition was 23%. When the patients with arrhythmia are excluded from the analysis, similar to the Tomizawa study5 , the dose reduction was 32%. The ViSION Edition system is equipped with a 90 kW generator to provide sufficient exposure required for the 275 ms gantry rotation speed. The larger generator enables larger patients to be scanned at 100 kV. In this study 89.6% of patients were imaged at 100 kV with an average patient BMI of 25 (Figure 3). Importantly, all these patients had diagnostic image quality (Figure 4). Conclusion Aquilion ONE has proved to be an exceptionally robust scanner for performing cardiac CT exami- nations on a wide range of patients with varying heart rates and rhythms. The ViSION Edition fur- ther extends this strength by allowing single beat coronary imaging in patients with heart rates up to 75 bpm. This study has shown that significant radiation dose reduction of 23% can be achieved in a large patient population, echoing similar findings in the two published papers4, 5 . Importantly, even in unselected populations, the majority of patients can now be scanned within a single heart beat, providing important advantages both for patients and imaging providers. References 1. Dewey M, Zimmermann E, Deissenrieder F, Laule M, Dübel HP, Schlattmann P, Knebel F, Rutsch W, Hamm B. Noninvasive coronary angiography by 320-row computed tomography with lower radiation exposure and maintained diagnostic accuracy: comparison of results with cardiac catheterization in a head-to-head pilot investigation. Circulation. 2009 Sep 8;120(10):867-75. 2. Gaudio C, Pelliccia F, Evangelista A, Tanzilli G, Paravati V, Pannarale G, Pannitteri G, Barillà F, Greco C, Franzoni F, Speziale G, Pasceri V. 320-row computed tomography coronary angi- ography vs. conventional coronary angiography in patients with suspected coronary artery dis- ease: A systematic review and meta-analysis. Int J Cardiol. 2013 Sep 30;168(2):1562-4 3. Hoe J, Toh KH. First experience with 320-row multidetector CT coronary angiography scan- ning with prospective electrocardiogram gating to reduce radiation dose. J Cardiovasc Comput Tomogr. 2009 Jul-Aug;3(4):257-61. 4. Chen MY, Shanbhag SM, Arai AE, Submillisievert Median Radiation Dose for Coronary Angiography with a Second-Generation 320-Detector Row CT Scanner in 107 Consec- utive Patients, Radiology, 2013, 267(1):76-85 5. Tomizawa N, Maeda E, Akahane M, Torigoe R, Kiryu S, Ohtomo K, Coronary CT angiography using the second-generation 320-detector row CT: assessment of image quality and radiation dose in various heart rates compared with the first-generation scanner, Int J Cardiovasc Imag- ing. 2013 Oct;29(7):1613-8.