SlideShare a Scribd company logo
1 of 52
Theory and Operation of a Scanning
Electron Microscope Equipped with an
Energy Dispersive Spectrometer
Scott Chumbley
Materials Science and Engineering
Iowa State University
Lecture Goals
What I hope to achieve today is the following…
- Provide some basics concerning SEM operation
and capabilities.
- Briefly discuss the most common signals you have
available to you using an SEM.
- Show some examples of how an SEM can be used
and discuss things to watch for that may cause
your analysis to be flawed.
Outline
I. Introduction
- Basics of Operation
II. Specimen Electron Interactions
- Backscattered Electrons
- Secondary Electrons
- X-rays
III. Examples of Signal Detection and Analysis
- Imaging (Backscattered and Secondary)
- Chemical Analysis (X-rays)
IV. Summary
Optical vs SEM
SEM has numerous
advantages over opticaloptical
microscopymicroscopy…..
Magnification (1000x1000x vs 100,000x)
Resolution (0.2 µm0.2 µm vs 0.005 µm)
Composition Information (limitedlimited vs extensive)
Sample Preparation (flatflat vs varied)
Depth of Focus (15 µm15 µm vs 4 mm)
Example
Pollen imaged using optics (left) and electrons (right).
Even at relatively low magnifications and resolutions the
Depth of Field (DF) for optical systems is low.
Basics of Operation
In operation, the SEM is really nothing more than a very
expensive Black & White Television. Consider…
Both have an electron gun in a vacuum.
Both scan in a serial manner.
The image depends on the quality /
strength of the incoming signal.
For a TV, the quality of the image depends on the strength of
the signal coming from the station. For SEM, it depends on
the signal collected by detectors, which depends on operation.
Basics of Operation
The SEM is analogous to a
conventional optical microscope in
terms of lenses... Condenser
Light Source
Objectives
Oculars
In the SEM we ...
• Heat a filament to get electrons.
• Use electromagnetic lenses to
produce a beam.
• Using scan coils to direct the beam
across the sample.
• Collect the emitted signals to form
an image.
Hardware
The filament is usually
tungsten or Lanthanum
Hexaboride (LaB6).
Typical Electron GunInitial Beam
“image”
Copper coils
Iron shell / casing
Magnetic Field
Final Beam
“image”
The lenses are electromagnetic, i.e.
coils of Cu wire wrapped around an
Fe shell. Changing the current to
the Cu changes the lens strength.
View of an actual lens.
Cu wires covered with
black insulator
Control Knobs
The Condenser Lens affects the resolution and signal strength.
It is usually labeled Spot Size, Probe Current, or Resolution.
The Objective Lens is always set to focus the beam on the
surface of the sample and is labeled Focus.
Scan Coils move the beam left/right and up/down to create a
square raster or a line scan. They control Magnification.
The Objective Aperture defines the angular spread of the
beam and determines how many electrons hit your sample.
Stigmation Coils enable you to make the beam round and get a
sharper image and are labeled Stigmator.
The Complete SEM Column
When you put
everything together
you have an SEM!
Now we need to ask:
How do I get an
image now that I
have e-
s hitting the
sample?
Image Formation
Image formation is achieved by transferring information
pixel by pixel, from sample to the display.
“Pixels” on
sample
“Pixels” on
computer display
Image on computer display
screen reflects sample.
The number of “pixels” is equal for the sample and the display.
Sample
Magnification =
Scan Length on Display
Scan Length on Sample
l
L
L
l
Image Formation
Display of the
collected signal
Sample
Incoming
Electron
Beam
I increase magnification by decreasing the scan size on the sample.
Example
How this all works is usually easily revealed by the contamination
that appears on the surface due to heating of the beam.
Contamination area
shows size of scan when
at a higher
magnification.
Definitions
How I set up my lenses and how much of the electron beam I
allow to hit my sample by inserting different sized apertures
in the column determine my operating conditions, such as….
Resolution: Ability to resolve two closely spaced points.
Depth of Field: Ability to clearly image features at different
distances from the lens or viewer.
Signal Strength: Number of electrons / x-rays are hitting
your sample and / or your detectors.
Working Distance: The distance the sample is from your
objective lens. Since the objective lens is always focused on
the sample, you can’t change one without the other.
Operation
Best depth of field
requires a narrow beam
(sample far from the lens,
small aperture).
Best resolution requires a
small beam (strong lenses,
weak signal, sample close to
the lens, i.e. short working
distance).
Good resolution, weak signal High signal, poor resolution
High signal requires a large diameter beam (weaker lenses,
poorer resolution).
Good depth of field Poor depth of field
Outline
I. Introduction
- Basics of Operation
II. Specimen Electron Interactions
- Backscattered Electrons
- Secondary Electrons
- X-rays
III. Examples of Signal Detection and Analysis
- Imaging (Backscattered and Secondary)
- Chemical Analysis (X-rays)
IV. Summary
Electron / Specimen Interactions
When an electron strikes a sample a large number of signals is
emitted. The most commonly signals are secondary electrons
(SE), backscattered electrons (BSE), and x-rays.
Sample
Backscattered (BSE)
Electrons
Secondary (SE)Secondary (SE)
ElectronsElectrons
X-rays
BSE electrons - high
energy, come off at high
angle, atomic number
dependence, directional
SE electrons - low energy,SE electrons - low energy,
“sucked” in to the“sucked” in to the
detector, depend ondetector, depend on
beam / specimen anglebeam / specimen angle
X-rays - various energies,X-rays - various energies,
chemical informationchemical information
ee--
BeamBeam
Interaction Volume
When the electron beam strikes a sample, the interactions that
result occur over a volume much larger than the size of the beam.
This is called the Interaction Volume.
Electron beam strikes
sample and starts to
interact with the
atoms. This
interaction can be
modeled.
The shape of the
interaction volume
depends on such
things as accelerating
voltage, composition
of the sample, and
angle the beam
enters.
Interaction volume varies from a hemisphere to a bottle shape.
Signal Generation
The signals come from various
depths within the interaction
volume in the sample.
Secondary Electrons - 100ÅSecondary Electrons - 100Å
Characteristic X-rays (K & L) -Characteristic X-rays (K & L) -
Backscattered Electrons
X-rays and BSE can come from
several microns below surface
depending upon sample and
accelerating voltage.
Electron Scattering
Backscattered
electrons come
from elastic
scattering of the
incident electrons
around the
nucleus.
Secondary
electrons come
from inelastic
scattering of the
incident electrons
with the electrons
that surround the
atom.
Best analogy - spacecraft
using gravity to “slingshot”
around planets.
Best analogy - trying to hit a
golf ball through a tree.
Detection
A thin detector over the
sample detects the BSE
electrons that strike
SE can be collected by
placing a positive voltage
on the detector.
SE images appear very 3-dimensional. BSE signal is very directional.
Backscattered Images
BSE electrons produce an image
related to the size of the nucleus.
The larger the nucleus, the more
BSE I get.
Al-Cu alloy. Copper has a larger nucleus so Cu
rich areas appear white, Al rich areas, dark gray.
Important:
1) The observed contrast is
always relative!
2) Sample must be polished
flat!
Steel sample. The lighter regions rich
in Cr, black areas are carbides. But
the sample is entirely steel!
Topo / Compo Imaging
BSE emission is directional.
Manipulation of multiple signals can
produce a topographical or
compositional image.
Top = Topographical (signals subtracted)
Bottom = Compositional (signals added)
The directional nature can
produce confusing images from
rough samples.
SE image BSE image, compo
BSE image, topo BSE image, combination
All 4 images on the left are of the exact same area!
Secondary Electrons
SE produce an image related to the surface roughness of the
material. More SE are emitted from steep slopes and sharp points
than flat surfaces.
SE image from broken steel.
Ridges, steep sides show up
brighter.
Collection efficiency of SE is higher
from the sample side facing the
detector. Thus, shadows from low
points and regions away from the
detector produce a “3-d” image.
X-rays
X-rays are also generated by inelastic collisions. There are two
basic kinds, those that are characteristic of the atom the e-
collides with and continuum x-rays which really are just
background and can’t be used for identification.
Electron from
beam interacts
with the atom’s
electrons and loses
energy.
Energy from the
interaction is
converted into an
x-ray.
Characteristic X-rays
Characteristic X-rays are much more useful! Their energy is
related to the atom because they involve collisions from actual
electrons in the sample.
Innermost shell gives K x-ray 2nd shell gives L x-ray 3rd shell gives M x-ray
K and L x-rays are most commonly observed and used.
Characteristic X-ray Spectrum
By collecting all of the x-rays and counting and separating
them according to energy we can get a spectra that is
characteristic of any particular material or element.
Number of
x-rays
counted
Energy of the collected x-ray
“Dot” Maps
It is possible to have the computer display a colored dot at the
location where the scanning beam was when a particular energy
x-ray was counted. This is called a “dot map”.
Same image of an alloy cast
iron. Different elements are
selected to show as different
colors.
Outline
I. Introduction
- Basics of Operation
II. Specimen Electron Interactions
- Backscattered Electrons
- Secondary Electrons
- X-rays
III. Examples of Signal Detection and Analysis
- Imaging (Backscattered and Secondary)
- Chemical Analysis (X-rays)
IV. Summary
Secondary vs. Backscattered
Copier “A” Toner Copier “B” Toner
The correct choice of imaging signal can allow you
to differentiate between two suspect possibilities…
Imaging of a “period” on paper, produced by two distinct copiers
The difference is clearly seen using BSE, indistinguishable with SE.
(Data from F. Platek, FDA)
Copier Toner Particles
Imaging of Layers
Measurement of
Tablet Coating
Thickness Using BSE
Since this
coating is an
edge, using SE
imaging it would
all appear
bright and the
thin layer would
be lost.
(Data from F. Platek, FDA)
Microscope Operating Parameters
How I operate the microscope will to a large extent determine
what I see. One of the major considerations is the choice of
accelerating voltage.
A low voltage limits
penetration of the beam and
produces an image that is
more surface sensitive.
A lower voltage can
more clearly reveal
surface films or
contaminants.
Imaging and EDS Analysis
5 kV 20 kV
25 kV 30 kV
While imaging may improve, X-ray analysis may suffer…
Higher voltage
gives a clearer
image of other
phases,
HOWEVER,
It comes with
increased
spreading of the
beam (yellow
circles) which
can compromise
EDS analysis.
Interaction Volume
Remember, X-rays are generated from an area much larger
than your beam. Consider these two Al-Y sample analyses…
Solid and dashed lines show
extent of x-ray generation.
Even though the beam is
outside the Y-rich phase
(right), beam spreading
below the surface causes Y
to be detected.
Effect of Overvoltage
In order to excite an x-ray of a particular energy, the beam
needs to have at least 1.5x the energy (overvoltage) just to
generate the x-ray. For example, if my x-ray has an energy of
10 KV…
If Beam Voltage is too low you may
be unable generate higher energy
(shorter wavelength) x-rays!
Pb
15KV
Pb
30KV
20KV
(Data from D. Ward, FBI)
Detector Operating Parameters
The Energy Dispersive Spectrometer is the most often used
detector since it is the easiest, least expensive, flexible, and
gives good results.
Effective use requires a
knowledge of detection
parameters (e.g. deadtime,
count time, pulse processor
time constant), and how the
software operates.
Typical EDS
with schematic
showing the
main coponents.
Peak Misidentification
Commercial routines commonly make mistakes!
4 companies,
all misidentify
various
samples and
elements
Relying too heavily on the software can lead to mistakes.
(Data from D. Newbury)
Example: Processing Time
Effect of Pulse Processor Time
Constant:
Red = 102us, Blue = 3.2us
(Data from D. Ward, FBI)
Note the peak
width of Al,
and the ability
to detect Mg.
As Time Constant increases,
resolution, sensitivity increase.
Note that increasing the time
constant is different than simply
counting for a longer period of time.
Example: Counting Time
Trace element detection and good statistics requires long
acquisition times.
Ti (0.1%)
Ti or Ba?
Peak?
The smaller the amount
desired to detect, the
longer the processing
and counting times
must be.
Rule of thumb: Make sure your
peaks are gaussian and the
background looks like
This:
Not this:
(Data from D. Ward, FBI)
100s
500s
10s
Example: False Peaks
Increasing the beam current will increase the count statistics.
However, a number of false peaks can appear that may
cause problems…
SUM PEAKS result when 2 x-rays enter at the exact same time,
producing false peaks at energy E = E1 + E2
Sometimes Si x-rays
are generated in the
Si detector,
producing false
peaks at energy E =
E1 - ESi called Si
Escape peaks.
These peaks usually result during high count rates.
Sample Parameters
If sample is not flat, BSE emission can
fluoresce x-rays that are not there,
absorption can eliminate x-rays that are
there. It is important to know the
location of the detector with respect to
the sample surface.
Cu sample held in place by Al foil.
Multiple Site EDS
Taking data from
various sites can
produce varying results.
(Data from F. Platek, FDA)
Examples
Similar sample geometry is VERY important when comparing
spectra from two different samples.
Detector
1
2
3
1
2
3
(Data from D. Ward, FBI)
Composition Averaging
When trying to get an average composition from a large
number of particles it is important to obtain numerous spectra
that encompasses as many particles as possible...
In example, the 3 “small area”
spectra show peak ratio
differences (spectra are scaled
on Ti). Summing the spectra
(bottom spectra) approaches
true sample composition.
(Data from D. Ward, FBI)
Quantitative Analysis
Quantification involves comparing data from
your unknown to data from known standard
compositions.
Calibration standard
containing known elements
and / or compounds.
= x (ZAF)i
Ci
C(i)
Ii
I(i)
Ci = wt % of i in unknown
C(i) = wt % of I in standard
Ii = intensity of I x-rays in
unknown
I(i) = intensity of I x-rays in
standard
ZAF = correction factor for
atomic number, absorption
and fluorescence
Samples must be
polished flat to
obtain the best
quantitative results.
The Cliff-Lorimer expression relates the
known compositions to the unknowns in terms
of x-ray intensity.
Methodology
When conducting a quantitative analysis, best results will
involve…
• Using known standards.
• Using a flat sample to minimize absorption / fluorescence.
• Ensuring proper EDS parameters (time constant,
calibration, counting time)
• Ensuring proper microscope parameters (beam current,
overvoltage)
Precision and Accuracy
The questions of importance for any quantitative analysis
method are always:
1. How accurate are my measurements?
2.2. How precisely can I measure something?How precisely can I measure something?
3.3. What are my detection limits?What are my detection limits?
These are always hard questions since the answer can depend
on the system, the operator, and the sample. But, in general, if
you have good data and know what you are doing….
#1: About 5% relative #2: About 5%#2: About 5% #3: About 0.1 wt%#3: About 0.1 wt%
Outline
I. Introduction
- Basics of Operation
II. Specimen Electron Interactions
- Backscattered Electrons
- Secondary Electrons
- X-rays
III. Examples of Signal Detection and Analysis
- Imaging (Backscattered and Secondary)
- Chemical Analysis (X-rays)
IV. Summary
Summary
Hopefully now we have covered all of these basic objectives...
- You know a little bit about the basics of operation
of an SEM.
- You are aware of the most common signals and
information you can get using an SEM.
- You have seen how different results can be
obtained depending upon which signal is used
and how the SEM is operated.
Summary of Lens Operation
Strong
Condenser Lens
Weak Condenser
Lens
Weak Objective
Lens
Strong Objective
Lens
Remember: Condenser lens
= probe current, spot size
Remember: Objective lens =
focus. Can only be changed in
conjunction with working
distance
(from JEOL operation manuals)
Summary of Signals
Secondary electrons are
good for rough samples.
Backscattered electrons are good for
samples with varying composition.
X-rays tell us the elements in the
sample and how much there is.
Conclusions
The SEM is a very flexible instrument for
numerous applications.
- One size does NOT fit all. Choose the signal,
operating parameters for the desired application.
- The information can change depending upon
many different things.
- Trace analysis requires special care and
knowledge of your system and your sample to get
the best results.

More Related Content

What's hot

What's hot (20)

Scanning electron microscopy
Scanning electron microscopyScanning electron microscopy
Scanning electron microscopy
 
Tem noman
Tem nomanTem noman
Tem noman
 
SEM,TEM & AFM
SEM,TEM & AFMSEM,TEM & AFM
SEM,TEM & AFM
 
Transmission Electron Microscopy
Transmission Electron MicroscopyTransmission Electron Microscopy
Transmission Electron Microscopy
 
Electron microscopy
Electron microscopyElectron microscopy
Electron microscopy
 
Electron microscope - SEM and TEM
Electron microscope - SEM and TEMElectron microscope - SEM and TEM
Electron microscope - SEM and TEM
 
Scanning electron microscopy mubbu
Scanning electron microscopy mubbuScanning electron microscopy mubbu
Scanning electron microscopy mubbu
 
Transmission electron microscope (TEM)
Transmission electron microscope (TEM)Transmission electron microscope (TEM)
Transmission electron microscope (TEM)
 
Electron microscope
Electron microscopeElectron microscope
Electron microscope
 
Electron Microscope Shanthakumar
Electron Microscope ShanthakumarElectron Microscope Shanthakumar
Electron Microscope Shanthakumar
 
Types of electron microscope
Types of electron microscopeTypes of electron microscope
Types of electron microscope
 
Transmission electron microscope
Transmission electron microscopeTransmission electron microscope
Transmission electron microscope
 
TRANSMISSION ELECTRON MICROSCOPY
TRANSMISSION ELECTRON MICROSCOPYTRANSMISSION ELECTRON MICROSCOPY
TRANSMISSION ELECTRON MICROSCOPY
 
Electron microscopy 2
Electron microscopy 2Electron microscopy 2
Electron microscopy 2
 
Tem
TemTem
Tem
 
Tariq presntation
Tariq presntationTariq presntation
Tariq presntation
 
Light Microscope and Electron Microscope [Best one]
Light Microscope and Electron Microscope [Best one]Light Microscope and Electron Microscope [Best one]
Light Microscope and Electron Microscope [Best one]
 
Fluorescent microscope
Fluorescent microscope Fluorescent microscope
Fluorescent microscope
 
Sem and tem
Sem and temSem and tem
Sem and tem
 
Transmission electron microscopy
Transmission electron microscopyTransmission electron microscopy
Transmission electron microscopy
 

Similar to Chumbley sem

Scanning Electron Microscopy
Scanning Electron MicroscopyScanning Electron Microscopy
Scanning Electron MicroscopyDr. Siju N Antony
 
Scanning electron microscopy
Scanning electron microscopyScanning electron microscopy
Scanning electron microscopyGanesh Shinde
 
Advanced Characterization Technique - SEM
Advanced Characterization Technique - SEMAdvanced Characterization Technique - SEM
Advanced Characterization Technique - SEMIlyas Hussain
 
Chapter-3. Electron Microscopy.pptx
Chapter-3. Electron Microscopy.pptxChapter-3. Electron Microscopy.pptx
Chapter-3. Electron Microscopy.pptxShivayogiPatil2
 
2018 HM-scanning electron microscope
2018 HM-scanning electron microscope2018 HM-scanning electron microscope
2018 HM-scanning electron microscopeHarsh Mohan
 
Scanning Electron Microscopy (SEM) lecture
Scanning Electron Microscopy (SEM) lectureScanning Electron Microscopy (SEM) lecture
Scanning Electron Microscopy (SEM) lectureSaurabh Bhargava
 
scanning electron microscope (SEM)
scanning electron microscope (SEM)scanning electron microscope (SEM)
scanning electron microscope (SEM)Himanshu Dixit
 
Scaning electron microscop
Scaning electron microscopScaning electron microscop
Scaning electron microscoptaameb
 
Scanning electron microscope
Scanning electron microscopeScanning electron microscope
Scanning electron microscopeSenthil Arasan
 
Introduction to Electron Microscopyy.pdf
Introduction to Electron Microscopyy.pdfIntroduction to Electron Microscopyy.pdf
Introduction to Electron Microscopyy.pdfLapsik1
 
Background Theory Class Resource
Background Theory Class ResourceBackground Theory Class Resource
Background Theory Class Resourceziamughal
 
Material Characterization Techniques.pptx
Material Characterization Techniques.pptxMaterial Characterization Techniques.pptx
Material Characterization Techniques.pptxssuser0cd0f1
 
Material Characterization Techniques.pptx
Material Characterization Techniques.pptxMaterial Characterization Techniques.pptx
Material Characterization Techniques.pptxssuser0cd0f1
 
The SEM can use either backscattered or secondary electrons to form .pdf
The SEM can use either backscattered or secondary electrons to form .pdfThe SEM can use either backscattered or secondary electrons to form .pdf
The SEM can use either backscattered or secondary electrons to form .pdfrushabhshah600
 

Similar to Chumbley sem (20)

Scanning Electron Microscopy
Scanning Electron MicroscopyScanning Electron Microscopy
Scanning Electron Microscopy
 
Scanning electron microscopy
Scanning electron microscopyScanning electron microscopy
Scanning electron microscopy
 
Scanning electron microscopy
Scanning electron microscopyScanning electron microscopy
Scanning electron microscopy
 
Sem
SemSem
Sem
 
Advanced Characterization Technique - SEM
Advanced Characterization Technique - SEMAdvanced Characterization Technique - SEM
Advanced Characterization Technique - SEM
 
Chapter-3. Electron Microscopy.pptx
Chapter-3. Electron Microscopy.pptxChapter-3. Electron Microscopy.pptx
Chapter-3. Electron Microscopy.pptx
 
2018 HM-scanning electron microscope
2018 HM-scanning electron microscope2018 HM-scanning electron microscope
2018 HM-scanning electron microscope
 
Scanning Electron Microscopy (SEM) lecture
Scanning Electron Microscopy (SEM) lectureScanning Electron Microscopy (SEM) lecture
Scanning Electron Microscopy (SEM) lecture
 
Electron Microscopy
Electron MicroscopyElectron Microscopy
Electron Microscopy
 
scanning electron microscope (SEM)
scanning electron microscope (SEM)scanning electron microscope (SEM)
scanning electron microscope (SEM)
 
Scaning electron microscop
Scaning electron microscopScaning electron microscop
Scaning electron microscop
 
Scanning electron microscope
Scanning electron microscopeScanning electron microscope
Scanning electron microscope
 
Introduction to Electron Microscopyy.pdf
Introduction to Electron Microscopyy.pdfIntroduction to Electron Microscopyy.pdf
Introduction to Electron Microscopyy.pdf
 
Sem md ppt
Sem md pptSem md ppt
Sem md ppt
 
Electron Microscope
Electron MicroscopeElectron Microscope
Electron Microscope
 
Background Theory Class Resource
Background Theory Class ResourceBackground Theory Class Resource
Background Theory Class Resource
 
Material Characterization Techniques.pptx
Material Characterization Techniques.pptxMaterial Characterization Techniques.pptx
Material Characterization Techniques.pptx
 
Material Characterization Techniques.pptx
Material Characterization Techniques.pptxMaterial Characterization Techniques.pptx
Material Characterization Techniques.pptx
 
The SEM can use either backscattered or secondary electrons to form .pdf
The SEM can use either backscattered or secondary electrons to form .pdfThe SEM can use either backscattered or secondary electrons to form .pdf
The SEM can use either backscattered or secondary electrons to form .pdf
 
Microscoy description by praveen
Microscoy description by praveenMicroscoy description by praveen
Microscoy description by praveen
 

Chumbley sem

  • 1. Theory and Operation of a Scanning Electron Microscope Equipped with an Energy Dispersive Spectrometer Scott Chumbley Materials Science and Engineering Iowa State University
  • 2. Lecture Goals What I hope to achieve today is the following… - Provide some basics concerning SEM operation and capabilities. - Briefly discuss the most common signals you have available to you using an SEM. - Show some examples of how an SEM can be used and discuss things to watch for that may cause your analysis to be flawed.
  • 3. Outline I. Introduction - Basics of Operation II. Specimen Electron Interactions - Backscattered Electrons - Secondary Electrons - X-rays III. Examples of Signal Detection and Analysis - Imaging (Backscattered and Secondary) - Chemical Analysis (X-rays) IV. Summary
  • 4. Optical vs SEM SEM has numerous advantages over opticaloptical microscopymicroscopy….. Magnification (1000x1000x vs 100,000x) Resolution (0.2 µm0.2 µm vs 0.005 µm) Composition Information (limitedlimited vs extensive) Sample Preparation (flatflat vs varied) Depth of Focus (15 µm15 µm vs 4 mm)
  • 5. Example Pollen imaged using optics (left) and electrons (right). Even at relatively low magnifications and resolutions the Depth of Field (DF) for optical systems is low.
  • 6. Basics of Operation In operation, the SEM is really nothing more than a very expensive Black & White Television. Consider… Both have an electron gun in a vacuum. Both scan in a serial manner. The image depends on the quality / strength of the incoming signal. For a TV, the quality of the image depends on the strength of the signal coming from the station. For SEM, it depends on the signal collected by detectors, which depends on operation.
  • 7. Basics of Operation The SEM is analogous to a conventional optical microscope in terms of lenses... Condenser Light Source Objectives Oculars In the SEM we ... • Heat a filament to get electrons. • Use electromagnetic lenses to produce a beam. • Using scan coils to direct the beam across the sample. • Collect the emitted signals to form an image.
  • 8. Hardware The filament is usually tungsten or Lanthanum Hexaboride (LaB6). Typical Electron GunInitial Beam “image” Copper coils Iron shell / casing Magnetic Field Final Beam “image” The lenses are electromagnetic, i.e. coils of Cu wire wrapped around an Fe shell. Changing the current to the Cu changes the lens strength. View of an actual lens. Cu wires covered with black insulator
  • 9. Control Knobs The Condenser Lens affects the resolution and signal strength. It is usually labeled Spot Size, Probe Current, or Resolution. The Objective Lens is always set to focus the beam on the surface of the sample and is labeled Focus. Scan Coils move the beam left/right and up/down to create a square raster or a line scan. They control Magnification. The Objective Aperture defines the angular spread of the beam and determines how many electrons hit your sample. Stigmation Coils enable you to make the beam round and get a sharper image and are labeled Stigmator.
  • 10. The Complete SEM Column When you put everything together you have an SEM! Now we need to ask: How do I get an image now that I have e- s hitting the sample?
  • 11. Image Formation Image formation is achieved by transferring information pixel by pixel, from sample to the display. “Pixels” on sample “Pixels” on computer display Image on computer display screen reflects sample. The number of “pixels” is equal for the sample and the display. Sample Magnification = Scan Length on Display Scan Length on Sample l L L l
  • 12. Image Formation Display of the collected signal Sample Incoming Electron Beam I increase magnification by decreasing the scan size on the sample.
  • 13. Example How this all works is usually easily revealed by the contamination that appears on the surface due to heating of the beam. Contamination area shows size of scan when at a higher magnification.
  • 14. Definitions How I set up my lenses and how much of the electron beam I allow to hit my sample by inserting different sized apertures in the column determine my operating conditions, such as…. Resolution: Ability to resolve two closely spaced points. Depth of Field: Ability to clearly image features at different distances from the lens or viewer. Signal Strength: Number of electrons / x-rays are hitting your sample and / or your detectors. Working Distance: The distance the sample is from your objective lens. Since the objective lens is always focused on the sample, you can’t change one without the other.
  • 15. Operation Best depth of field requires a narrow beam (sample far from the lens, small aperture). Best resolution requires a small beam (strong lenses, weak signal, sample close to the lens, i.e. short working distance). Good resolution, weak signal High signal, poor resolution High signal requires a large diameter beam (weaker lenses, poorer resolution). Good depth of field Poor depth of field
  • 16. Outline I. Introduction - Basics of Operation II. Specimen Electron Interactions - Backscattered Electrons - Secondary Electrons - X-rays III. Examples of Signal Detection and Analysis - Imaging (Backscattered and Secondary) - Chemical Analysis (X-rays) IV. Summary
  • 17. Electron / Specimen Interactions When an electron strikes a sample a large number of signals is emitted. The most commonly signals are secondary electrons (SE), backscattered electrons (BSE), and x-rays. Sample Backscattered (BSE) Electrons Secondary (SE)Secondary (SE) ElectronsElectrons X-rays BSE electrons - high energy, come off at high angle, atomic number dependence, directional SE electrons - low energy,SE electrons - low energy, “sucked” in to the“sucked” in to the detector, depend ondetector, depend on beam / specimen anglebeam / specimen angle X-rays - various energies,X-rays - various energies, chemical informationchemical information ee-- BeamBeam
  • 18. Interaction Volume When the electron beam strikes a sample, the interactions that result occur over a volume much larger than the size of the beam. This is called the Interaction Volume. Electron beam strikes sample and starts to interact with the atoms. This interaction can be modeled. The shape of the interaction volume depends on such things as accelerating voltage, composition of the sample, and angle the beam enters. Interaction volume varies from a hemisphere to a bottle shape.
  • 19. Signal Generation The signals come from various depths within the interaction volume in the sample. Secondary Electrons - 100ÅSecondary Electrons - 100Å Characteristic X-rays (K & L) -Characteristic X-rays (K & L) - Backscattered Electrons X-rays and BSE can come from several microns below surface depending upon sample and accelerating voltage.
  • 20. Electron Scattering Backscattered electrons come from elastic scattering of the incident electrons around the nucleus. Secondary electrons come from inelastic scattering of the incident electrons with the electrons that surround the atom. Best analogy - spacecraft using gravity to “slingshot” around planets. Best analogy - trying to hit a golf ball through a tree.
  • 21. Detection A thin detector over the sample detects the BSE electrons that strike SE can be collected by placing a positive voltage on the detector. SE images appear very 3-dimensional. BSE signal is very directional.
  • 22. Backscattered Images BSE electrons produce an image related to the size of the nucleus. The larger the nucleus, the more BSE I get. Al-Cu alloy. Copper has a larger nucleus so Cu rich areas appear white, Al rich areas, dark gray. Important: 1) The observed contrast is always relative! 2) Sample must be polished flat! Steel sample. The lighter regions rich in Cr, black areas are carbides. But the sample is entirely steel!
  • 23. Topo / Compo Imaging BSE emission is directional. Manipulation of multiple signals can produce a topographical or compositional image. Top = Topographical (signals subtracted) Bottom = Compositional (signals added) The directional nature can produce confusing images from rough samples. SE image BSE image, compo BSE image, topo BSE image, combination All 4 images on the left are of the exact same area!
  • 24. Secondary Electrons SE produce an image related to the surface roughness of the material. More SE are emitted from steep slopes and sharp points than flat surfaces. SE image from broken steel. Ridges, steep sides show up brighter. Collection efficiency of SE is higher from the sample side facing the detector. Thus, shadows from low points and regions away from the detector produce a “3-d” image.
  • 25. X-rays X-rays are also generated by inelastic collisions. There are two basic kinds, those that are characteristic of the atom the e- collides with and continuum x-rays which really are just background and can’t be used for identification. Electron from beam interacts with the atom’s electrons and loses energy. Energy from the interaction is converted into an x-ray.
  • 26. Characteristic X-rays Characteristic X-rays are much more useful! Their energy is related to the atom because they involve collisions from actual electrons in the sample. Innermost shell gives K x-ray 2nd shell gives L x-ray 3rd shell gives M x-ray K and L x-rays are most commonly observed and used.
  • 27. Characteristic X-ray Spectrum By collecting all of the x-rays and counting and separating them according to energy we can get a spectra that is characteristic of any particular material or element. Number of x-rays counted Energy of the collected x-ray
  • 28. “Dot” Maps It is possible to have the computer display a colored dot at the location where the scanning beam was when a particular energy x-ray was counted. This is called a “dot map”. Same image of an alloy cast iron. Different elements are selected to show as different colors.
  • 29. Outline I. Introduction - Basics of Operation II. Specimen Electron Interactions - Backscattered Electrons - Secondary Electrons - X-rays III. Examples of Signal Detection and Analysis - Imaging (Backscattered and Secondary) - Chemical Analysis (X-rays) IV. Summary
  • 30. Secondary vs. Backscattered Copier “A” Toner Copier “B” Toner The correct choice of imaging signal can allow you to differentiate between two suspect possibilities… Imaging of a “period” on paper, produced by two distinct copiers The difference is clearly seen using BSE, indistinguishable with SE. (Data from F. Platek, FDA) Copier Toner Particles
  • 31. Imaging of Layers Measurement of Tablet Coating Thickness Using BSE Since this coating is an edge, using SE imaging it would all appear bright and the thin layer would be lost. (Data from F. Platek, FDA)
  • 32. Microscope Operating Parameters How I operate the microscope will to a large extent determine what I see. One of the major considerations is the choice of accelerating voltage. A low voltage limits penetration of the beam and produces an image that is more surface sensitive. A lower voltage can more clearly reveal surface films or contaminants.
  • 33. Imaging and EDS Analysis 5 kV 20 kV 25 kV 30 kV While imaging may improve, X-ray analysis may suffer… Higher voltage gives a clearer image of other phases, HOWEVER, It comes with increased spreading of the beam (yellow circles) which can compromise EDS analysis.
  • 34. Interaction Volume Remember, X-rays are generated from an area much larger than your beam. Consider these two Al-Y sample analyses… Solid and dashed lines show extent of x-ray generation. Even though the beam is outside the Y-rich phase (right), beam spreading below the surface causes Y to be detected.
  • 35. Effect of Overvoltage In order to excite an x-ray of a particular energy, the beam needs to have at least 1.5x the energy (overvoltage) just to generate the x-ray. For example, if my x-ray has an energy of 10 KV… If Beam Voltage is too low you may be unable generate higher energy (shorter wavelength) x-rays! Pb 15KV Pb 30KV 20KV (Data from D. Ward, FBI)
  • 36. Detector Operating Parameters The Energy Dispersive Spectrometer is the most often used detector since it is the easiest, least expensive, flexible, and gives good results. Effective use requires a knowledge of detection parameters (e.g. deadtime, count time, pulse processor time constant), and how the software operates. Typical EDS with schematic showing the main coponents.
  • 37. Peak Misidentification Commercial routines commonly make mistakes! 4 companies, all misidentify various samples and elements Relying too heavily on the software can lead to mistakes. (Data from D. Newbury)
  • 38. Example: Processing Time Effect of Pulse Processor Time Constant: Red = 102us, Blue = 3.2us (Data from D. Ward, FBI) Note the peak width of Al, and the ability to detect Mg. As Time Constant increases, resolution, sensitivity increase. Note that increasing the time constant is different than simply counting for a longer period of time.
  • 39. Example: Counting Time Trace element detection and good statistics requires long acquisition times. Ti (0.1%) Ti or Ba? Peak? The smaller the amount desired to detect, the longer the processing and counting times must be. Rule of thumb: Make sure your peaks are gaussian and the background looks like This: Not this: (Data from D. Ward, FBI) 100s 500s 10s
  • 40. Example: False Peaks Increasing the beam current will increase the count statistics. However, a number of false peaks can appear that may cause problems… SUM PEAKS result when 2 x-rays enter at the exact same time, producing false peaks at energy E = E1 + E2 Sometimes Si x-rays are generated in the Si detector, producing false peaks at energy E = E1 - ESi called Si Escape peaks. These peaks usually result during high count rates.
  • 41. Sample Parameters If sample is not flat, BSE emission can fluoresce x-rays that are not there, absorption can eliminate x-rays that are there. It is important to know the location of the detector with respect to the sample surface. Cu sample held in place by Al foil.
  • 42. Multiple Site EDS Taking data from various sites can produce varying results. (Data from F. Platek, FDA)
  • 43. Examples Similar sample geometry is VERY important when comparing spectra from two different samples. Detector 1 2 3 1 2 3 (Data from D. Ward, FBI)
  • 44. Composition Averaging When trying to get an average composition from a large number of particles it is important to obtain numerous spectra that encompasses as many particles as possible... In example, the 3 “small area” spectra show peak ratio differences (spectra are scaled on Ti). Summing the spectra (bottom spectra) approaches true sample composition. (Data from D. Ward, FBI)
  • 45. Quantitative Analysis Quantification involves comparing data from your unknown to data from known standard compositions. Calibration standard containing known elements and / or compounds. = x (ZAF)i Ci C(i) Ii I(i) Ci = wt % of i in unknown C(i) = wt % of I in standard Ii = intensity of I x-rays in unknown I(i) = intensity of I x-rays in standard ZAF = correction factor for atomic number, absorption and fluorescence Samples must be polished flat to obtain the best quantitative results. The Cliff-Lorimer expression relates the known compositions to the unknowns in terms of x-ray intensity.
  • 46. Methodology When conducting a quantitative analysis, best results will involve… • Using known standards. • Using a flat sample to minimize absorption / fluorescence. • Ensuring proper EDS parameters (time constant, calibration, counting time) • Ensuring proper microscope parameters (beam current, overvoltage)
  • 47. Precision and Accuracy The questions of importance for any quantitative analysis method are always: 1. How accurate are my measurements? 2.2. How precisely can I measure something?How precisely can I measure something? 3.3. What are my detection limits?What are my detection limits? These are always hard questions since the answer can depend on the system, the operator, and the sample. But, in general, if you have good data and know what you are doing…. #1: About 5% relative #2: About 5%#2: About 5% #3: About 0.1 wt%#3: About 0.1 wt%
  • 48. Outline I. Introduction - Basics of Operation II. Specimen Electron Interactions - Backscattered Electrons - Secondary Electrons - X-rays III. Examples of Signal Detection and Analysis - Imaging (Backscattered and Secondary) - Chemical Analysis (X-rays) IV. Summary
  • 49. Summary Hopefully now we have covered all of these basic objectives... - You know a little bit about the basics of operation of an SEM. - You are aware of the most common signals and information you can get using an SEM. - You have seen how different results can be obtained depending upon which signal is used and how the SEM is operated.
  • 50. Summary of Lens Operation Strong Condenser Lens Weak Condenser Lens Weak Objective Lens Strong Objective Lens Remember: Condenser lens = probe current, spot size Remember: Objective lens = focus. Can only be changed in conjunction with working distance (from JEOL operation manuals)
  • 51. Summary of Signals Secondary electrons are good for rough samples. Backscattered electrons are good for samples with varying composition. X-rays tell us the elements in the sample and how much there is.
  • 52. Conclusions The SEM is a very flexible instrument for numerous applications. - One size does NOT fit all. Choose the signal, operating parameters for the desired application. - The information can change depending upon many different things. - Trace analysis requires special care and knowledge of your system and your sample to get the best results.

Editor's Notes

  1. The carbides seen between the primary austenite and the secondary austenite are the high temperature M7C3 carbides. These form during the soak at 1100C. The secondary austenite forms during the soak in the salt bath since the austenite/ferrite ration will change. The sigma is a golden brown color. The Austenite appears to be dendritic.
  2. It does not match TTT diagrams for other compositions (e.g.UR50). This indicates that the TTT characteristics of casting and wrought alloys may be comparable for similar chemistries.