Teletherapy Cobalt-60
Machines Vs Linear Accelerator
Prof Amin E AAmin
Dean of the Higher Institute of Optics
Technology
&
Prof of Medical Physics
Radiation Oncology Department
Faculty of Medicine
Ain Shams University
Introduction
ā‘ Both the cobalt-60 teletherapy machine and clinical linac
were introduced in 1950s and used for megavoltage
radiotherapy since then.
ā‘ During the past 50 years:
ā€¢ The basic design of the cobalt machine remained essentially the
same.
ā€¢ Linac progressed to the current design through five generations,
eclipsed the cobalt machine, and became the most widely used
radiation source in modern radiotherapy, with several 1000 units in
clinical practice around the world today.
Cobalt-60 Teletherapy Units Versus Linacs
In comparison with cobalt-60 teletherapy machines linacs have
become very complex in design:
āž¢Because of the multimodality capabilities that have evolved and are
available on most modern linacs (several x-ray energies and
several electron energies).
āž¢Because of an increased use of computer logic and
microprocessors in the control systems of linacs.
āž¢Because of added features, such as high dose rate modes, multileaf
collimation, electron arc therapy, and the dynamic treatment option
on the collimators (dynamic wedge), MLC leaves (IMRT), gantry
or table while the beam is turned on.
COBALT-60 TELETHERAPY UNITS
VERSUS LINACS
Despite clear technological and practical advantages of linacs over
cobalt-60 machines, the latter still occupy an important place in
radiotherapy armamentarium, especially in developing countries,
because of:
ā€¢ Their considerably lower capital and installation cost.
ā€¢ Lower servicing and maintenance cost.
ā€¢ Lesser dependence on reliable electrical power.
ā€¢ Simplicity of design.
ā€¢ Ease of operation.
Cobalt Units Vs Linear Accelerators
ā€¢ Several arguments have been put forward both for and against
Cobalt Units as well as Linear accelerators.
ā€¢ These arguments relate to physics, clinical advantages and
more importantly, the cost consideration.
ā€¢ The following few slides summaries the data that is available.
Cobalt Units Vs Linear Accelerators
Feature Low Energy
Accelerator
Cobalt 60Unit Remarks
Dose rate 600 MU/Min Reducing dose
starts at 200r/min
In Cobalt units dose
will reduce because
of decay, and
depends on source
activity whereas with
Linear Accelerator
one has the guarantee
of constant dose rate
Cobalt Units Vs Linear Accelerators
Feature Low Energy
Accelerator
Cobalt 60Unit Remarks
Energy
Source
DIA
<2mm 20mm ā€«ā€ā€¬Penumbra effect is
higher in cobalt
machines
Cobalt Units Vs Linear Accelerators
Feature Low Energy
Accelerator
Cobalt
60Unit
Remarks
Source
change
None Every 5-10
years (in
average 7
years)
ā€«ā€ā€¬Source cost for 12,000
curie source is very much
on the increasing trend. At
the time of source change,
decayed source has to be
sent back to the
manufacturer, which is
very expensive.
Cobalt Units Vs Linear Accelerators
Feature Low Energy
Accelerator
Cobalt
60Unit
Remarks
D- max 1.5 cm Only 0.5cm ā€«ā€ā€¬Linear accelerator with 6
MV, photon, we are able
to get D-max 1.5 cm. This
is very useful for deep
seated tumours. Skin
sparting is better with
Linac (6 mv).
Cobalt Units Vs Linear Accelerators
Feature Low Energy
Accelerator
Cobalt
60Unit
Remarks
Perchantage
Depth dose
at 10cm
67% 55% Difficult to treat deep
seated tumors in Thorax,
abdomen or pelvis.
Central Axis Depth Dose Distribution
For Different Quality Photon Beams
Cobalt Units Vs Linear Accelerators
Feature Low Energy
Accelerator
Cobalt
60Unit
Remarks
Colimator
trasmission
Not exceed
o.5%
More than 3% ā€”
Minimum
field size
0.5x0.5 cm 5 x 5 cm In cobalt one cannot treat
field size less than 5x5cm.
ā€«ā€ā€¬Radiation
leakage
through the
head
Nil Less than
0.2%
ā€”
Cobalt Units Vs Linear Accelerators
Feature Low Energy
Accelerator
Cobalt
60Unit
Remarks
Penumbra Less than 5
mm
More than 1
cm
With 2 cm. Cobalt source
the Penumbra effect is very
high.
Motorized
wedges
Possible Not possible __
Variable
dose rate
80 to 600
MU/min
Not variable Dose rate if fixed,
decreases with decay of
source
Cobalt Units Vs Linear Accelerators
Feature Low Energy
Accelerator
Cobalt
60Unit
Remarks
Precision Dose delivered
to and arc is
reproducible
within 3% or
1MU which
ever is greater
__ In linear accelerator one
can have reproducibility of
dosimetry system and the
dose delivered is
continuously monitored
and recorded
Cobalt Units Vs Linear Accelerators
Feature Low
Energy
Accelerator
Cobalt
60Unit
Remarks
Multileaf
collimators,
stereotactic
attachments
Possible Not possible Increases accuracy in dose
delivery
IMRT Possible Not possible Increases accuracy in dose
delivery
Cobalt Units Vs Linear Accelerators
Feature Low
Energy
Accelerator
Cobalt
60Unit
Remarks
ā€«ā€ā€¬Asymmetric
collimato jaws
Yes, standard
features
No Electronics are cheaper
Independenty collimator
jaws are very useful for
precise and accurate cancer
treatment
Cobalt Units Vs Linear Accelerators
Feature Low
Energy
Accelerator
Cobalt
60Unit
Remarks
Portal vision
attachment
Yes, possible
(Optinal)
No Allows to view actual
tumour while treatment
Patient
support
system
Superior Inferior, low
accuracy
Cobalt Units Vs Linear Accelerators
Feature Low Energy
Accelerator
Cobalt
60Unit
Remarks
Beam
matching with
second unit
Possible Not
possible
Beam matching is a very good
feature available in Linear
accelerator. This will help
institutions to buy the 2nd
machine with the same beam
characteristics of the existing
units. Thereby patients can be
shifted between different
Linear accelerator. More
flexibility.
Cobalt Units Vs Linear Accelerators
Feature Low
Energy
Accelerator
Cobalt 60Unit Remarks
Source
disposal
Nil Major problem
with cobalt units,
and source cost is
increasing every
year. It is also
becoming
increasingly
difficult to obtain
a source
In the event the source
is imported along with
the cobalt
machine the decayed
source has to be
transported back to the
manufacturers, which
is very expensive.
ā€¢ Cobalt 60 units provide relatively high energy gamma rays for
radiotherapy which are ideally suited for treatment of head and
neck cancers and other superficially located tumors like breast
cancers and soft tissue sarcomas of extremities.
ā€¢ They are not adequate for treatment of deep seated tumors and
have the added disadvantage of decreasing output with decay
of source and the need for source replacement within 5-7
years.
Cobalt Units Vs Linear Accelerators
ā€¢ Disposal of decayed source is another major concern.
ā€¢ The beam characteristics when compared to 6mv Linacs are
inferior and fewer ancillaries are available for cobalt machines
as compared to Linacs.
ā€¢ High energy Linacs in addition to giving two or three x-ray
energies can also generate variable energy electron beam for
treatment.
Cobalt Units Vs Linear Accelerators
ā€¢ While such high energy Linacs are expense, 6 MV Linacs
compare favorably in terms of cost with the Cobalt 60 units.
Even though the initial cost appears to be high, over a ten year
period maintenance costs are less as it does not require change
of source.
ā€¢ It is stressed that service contracts with performance guarantees
for 5-10 years have to be built into the contracts while procuring
the equipment.
Cobalt Units Vs Linear Accelerators
ā€¢ To summarize Linear accelerators have several advantages:
1. Very high energy beams can be created with a machine that
is not very bulky or cumbersome to use.
2. The edges of the beams are much more sharply defined
than those of a cobalt machine, allowing additional
precision in dose delivery.
Advantages Of Linear Accelerators
3. Electron beams can be created (with high energy linacs) that
is of particular value in treating superficial lesions.
4. The dose rate per minute is variable and can be turned up
very high allowing the patient to be located at substantial
distance from the machine in order to create large fields
necessary for total skin or total body irradiation while still
maintaining adequate dose rate. With cobalt, the rate is
determined by the amount of cobalt source in the machine
and cannot be regulated.
Advantages Of Linear Accelerators
5. Advanced techniques (three dimension conformal
radiotherapy, IMRT, VMAT, IGRT) can be applied.
Advantages Of Linear Accelerators
Teletherapy cobalt 60 machines vs linear accelerator

More Related Content

What's hot

importance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapyimportance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapymakhhi
Ā 
Beam modifying devices 1 Radiophysics
Beam modifying devices 1 RadiophysicsBeam modifying devices 1 Radiophysics
Beam modifying devices 1 RadiophysicsDrAyush Garg
Ā 
Beam modification-in-radiotherapy-
Beam modification-in-radiotherapy- Beam modification-in-radiotherapy-
Beam modification-in-radiotherapy- Sathish Kumar
Ā 
Quality assurance of linear accelerator DHX
Quality assurance of linear accelerator DHXQuality assurance of linear accelerator DHX
Quality assurance of linear accelerator DHXSohail Qureshi
Ā 
planning systems in radiotherapy
 planning systems in radiotherapy planning systems in radiotherapy
planning systems in radiotherapyfondas vakalis
Ā 
Multileaf Collimator
Multileaf CollimatorMultileaf Collimator
Multileaf CollimatorVinay Desai
Ā 
New Techniques in Radiotherapy
New Techniques in RadiotherapyNew Techniques in Radiotherapy
New Techniques in RadiotherapySantam Chakraborty
Ā 
TIME DOSE & FRACTIONATION
TIME DOSE & FRACTIONATIONTIME DOSE & FRACTIONATION
TIME DOSE & FRACTIONATIONIsha Jaiswal
Ā 
The vmat vs other recent radiotherapy techniques
The vmat vs other recent radiotherapy techniquesThe vmat vs other recent radiotherapy techniques
The vmat vs other recent radiotherapy techniquesM'dee Phechudi
Ā 
importance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapyimportance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapymakhhi
Ā 
Volumetric Modulated Arc Therapy
Volumetric Modulated Arc TherapyVolumetric Modulated Arc Therapy
Volumetric Modulated Arc Therapyfondas vakalis
Ā 
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEXTISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEXVictor Ekpo
Ā 
Teletherapy Cobalt-60 Machines
Teletherapy Cobalt-60 MachinesTeletherapy Cobalt-60 Machines
Teletherapy Cobalt-60 MachinesAmin Amin
Ā 
telecobalt machine.ppt
telecobalt machine.ppttelecobalt machine.ppt
telecobalt machine.pptAtul Verma
Ā 

What's hot (20)

Dosimetry
DosimetryDosimetry
Dosimetry
Ā 
importance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapyimportance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapy
Ā 
Beam modifying devices 1 Radiophysics
Beam modifying devices 1 RadiophysicsBeam modifying devices 1 Radiophysics
Beam modifying devices 1 Radiophysics
Ā 
Beam modification-in-radiotherapy-
Beam modification-in-radiotherapy- Beam modification-in-radiotherapy-
Beam modification-in-radiotherapy-
Ā 
4dct (2012)
4dct (2012)4dct (2012)
4dct (2012)
Ā 
Quality assurance of linear accelerator DHX
Quality assurance of linear accelerator DHXQuality assurance of linear accelerator DHX
Quality assurance of linear accelerator DHX
Ā 
planning systems in radiotherapy
 planning systems in radiotherapy planning systems in radiotherapy
planning systems in radiotherapy
Ā 
Beam modification
Beam modification Beam modification
Beam modification
Ā 
Session 3 measure beam profile rt
Session 3 measure beam profile rtSession 3 measure beam profile rt
Session 3 measure beam profile rt
Ā 
Multileaf Collimator
Multileaf CollimatorMultileaf Collimator
Multileaf Collimator
Ā 
New Techniques in Radiotherapy
New Techniques in RadiotherapyNew Techniques in Radiotherapy
New Techniques in Radiotherapy
Ā 
TIME DOSE & FRACTIONATION
TIME DOSE & FRACTIONATIONTIME DOSE & FRACTIONATION
TIME DOSE & FRACTIONATION
Ā 
The vmat vs other recent radiotherapy techniques
The vmat vs other recent radiotherapy techniquesThe vmat vs other recent radiotherapy techniques
The vmat vs other recent radiotherapy techniques
Ā 
importance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapyimportance of ct-simulator in radiotherapy
importance of ct-simulator in radiotherapy
Ā 
Electron beam therapy
Electron beam therapyElectron beam therapy
Electron beam therapy
Ā 
Volumetric Modulated Arc Therapy
Volumetric Modulated Arc TherapyVolumetric Modulated Arc Therapy
Volumetric Modulated Arc Therapy
Ā 
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEXTISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
Ā 
Teletherapy Cobalt-60 Machines
Teletherapy Cobalt-60 MachinesTeletherapy Cobalt-60 Machines
Teletherapy Cobalt-60 Machines
Ā 
Isodose lines
Isodose linesIsodose lines
Isodose lines
Ā 
telecobalt machine.ppt
telecobalt machine.ppttelecobalt machine.ppt
telecobalt machine.ppt
Ā 

Similar to Teletherapy cobalt 60 machines vs linear accelerator

Civan over view
Civan over viewCivan over view
Civan over viewyaellasry
Ā 
IMPROVING POWER QUALITY AND FAST ENERGY SAVING METHODS IN POWER DISTRIBUTION ...
IMPROVING POWER QUALITY AND FAST ENERGY SAVING METHODS IN POWER DISTRIBUTION ...IMPROVING POWER QUALITY AND FAST ENERGY SAVING METHODS IN POWER DISTRIBUTION ...
IMPROVING POWER QUALITY AND FAST ENERGY SAVING METHODS IN POWER DISTRIBUTION ...IRJET Journal
Ā 
Portable and mobile radiographic equipments [Autosaved].pptx
Portable and mobile radiographic equipments [Autosaved].pptxPortable and mobile radiographic equipments [Autosaved].pptx
Portable and mobile radiographic equipments [Autosaved].pptxSaritaGaire1
Ā 
Three core extra high voltage cable ā€“ a global innovation
Three core extra high voltage cable ā€“ a global innovationThree core extra high voltage cable ā€“ a global innovation
Three core extra high voltage cable ā€“ a global innovationTata Power Delhi Distribution Limited
Ā 
Active magnetic bearing
Active magnetic bearingActive magnetic bearing
Active magnetic bearingmharish997
Ā 
The inverter core consists of 3 components
The inverter core consists of 3 componentsThe inverter core consists of 3 components
The inverter core consists of 3 componentsrajlaxmi06
Ā 
Ppt 120221105321-phpapp02
Ppt 120221105321-phpapp02Ppt 120221105321-phpapp02
Ppt 120221105321-phpapp02dinagardinesh
Ā 
Lecture (3)external beam therapy (ebt)
Lecture (3)external beam therapy (ebt)Lecture (3)external beam therapy (ebt)
Lecture (3)external beam therapy (ebt)Zyad Ahmed
Ā 
SMAC Food Processing Applications
SMAC Food Processing ApplicationsSMAC Food Processing Applications
SMAC Food Processing ApplicationsJohn Miewald
Ā 
14 hja 12 perpetual travel slide share
14 hja 12 perpetual travel  slide share14 hja 12 perpetual travel  slide share
14 hja 12 perpetual travel slide shareHenning Jacobsen
Ā 
13 hja 01 motive power mst
13 hja 01 motive power mst13 hja 01 motive power mst
13 hja 01 motive power mstHenning Jacobsen
Ā 
High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...
High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...
High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...IJRES Journal
Ā 
High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...
High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...
High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...irjes
Ā 
Linear Accelerator- Commercial Specification & Comparison
Linear Accelerator- Commercial Specification & ComparisonLinear Accelerator- Commercial Specification & Comparison
Linear Accelerator- Commercial Specification & ComparisonArindham Debbarma
Ā 
pipe inspection robot.pptx
pipe inspection robot.pptxpipe inspection robot.pptx
pipe inspection robot.pptxAdityaMishra595951
Ā 
Internship report
Internship report Internship report
Internship report Niteesh Shanbog
Ā 

Similar to Teletherapy cobalt 60 machines vs linear accelerator (20)

Civan over view
Civan over viewCivan over view
Civan over view
Ā 
Ssfcl
SsfclSsfcl
Ssfcl
Ā 
IMPROVING POWER QUALITY AND FAST ENERGY SAVING METHODS IN POWER DISTRIBUTION ...
IMPROVING POWER QUALITY AND FAST ENERGY SAVING METHODS IN POWER DISTRIBUTION ...IMPROVING POWER QUALITY AND FAST ENERGY SAVING METHODS IN POWER DISTRIBUTION ...
IMPROVING POWER QUALITY AND FAST ENERGY SAVING METHODS IN POWER DISTRIBUTION ...
Ā 
Portable and mobile radiographic equipments [Autosaved].pptx
Portable and mobile radiographic equipments [Autosaved].pptxPortable and mobile radiographic equipments [Autosaved].pptx
Portable and mobile radiographic equipments [Autosaved].pptx
Ā 
Three core extra high voltage cable ā€“ a global innovation
Three core extra high voltage cable ā€“ a global innovationThree core extra high voltage cable ā€“ a global innovation
Three core extra high voltage cable ā€“ a global innovation
Ā 
Active magnetic bearing
Active magnetic bearingActive magnetic bearing
Active magnetic bearing
Ā 
The inverter core consists of 3 components
The inverter core consists of 3 componentsThe inverter core consists of 3 components
The inverter core consists of 3 components
Ā 
Ppt 120221105321-phpapp02
Ppt 120221105321-phpapp02Ppt 120221105321-phpapp02
Ppt 120221105321-phpapp02
Ā 
Ppt
PptPpt
Ppt
Ā 
Ppt
PptPpt
Ppt
Ā 
Lecture (3)external beam therapy (ebt)
Lecture (3)external beam therapy (ebt)Lecture (3)external beam therapy (ebt)
Lecture (3)external beam therapy (ebt)
Ā 
SMAC Food Processing Applications
SMAC Food Processing ApplicationsSMAC Food Processing Applications
SMAC Food Processing Applications
Ā 
14 hja 12 perpetual travel slide share
14 hja 12 perpetual travel  slide share14 hja 12 perpetual travel  slide share
14 hja 12 perpetual travel slide share
Ā 
13 hja 01 motive power mst
13 hja 01 motive power mst13 hja 01 motive power mst
13 hja 01 motive power mst
Ā 
High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...
High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...
High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...
Ā 
High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...
High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...
High Step-Up Converter with Voltage Multiplier Module for Renewable Energy Sy...
Ā 
Linear Accelerator- Commercial Specification & Comparison
Linear Accelerator- Commercial Specification & ComparisonLinear Accelerator- Commercial Specification & Comparison
Linear Accelerator- Commercial Specification & Comparison
Ā 
LINAC final.pptx
LINAC final.pptxLINAC final.pptx
LINAC final.pptx
Ā 
pipe inspection robot.pptx
pipe inspection robot.pptxpipe inspection robot.pptx
pipe inspection robot.pptx
Ā 
Internship report
Internship report Internship report
Internship report
Ā 

More from Amin Amin

Laser Light in Cancer Therapy
Laser Light in Cancer TherapyLaser Light in Cancer Therapy
Laser Light in Cancer TherapyAmin Amin
Ā 
Radiation protection course for radiologists L8
Radiation protection course for radiologists L8Radiation protection course for radiologists L8
Radiation protection course for radiologists L8Amin Amin
Ā 
Radiation protection course for radiologists L7
Radiation protection course for radiologists L7Radiation protection course for radiologists L7
Radiation protection course for radiologists L7Amin Amin
Ā 
Radiation protection course for radiologists L6
Radiation protection course for radiologists L6Radiation protection course for radiologists L6
Radiation protection course for radiologists L6Amin Amin
Ā 
Radiation protection course for radiologists L5
Radiation protection course for radiologists L5Radiation protection course for radiologists L5
Radiation protection course for radiologists L5Amin Amin
Ā 
Radiation protection course for radiologists L4
Radiation protection course for radiologists L4Radiation protection course for radiologists L4
Radiation protection course for radiologists L4Amin Amin
Ā 
Radiation protection course for radiologists Lecture 3
Radiation protection course for radiologists Lecture 3Radiation protection course for radiologists Lecture 3
Radiation protection course for radiologists Lecture 3Amin Amin
Ā 
Radiation protection course for radiologists L2
Radiation protection course for radiologists L2Radiation protection course for radiologists L2
Radiation protection course for radiologists L2Amin Amin
Ā 
Radiation protection course for radiologists L1
Radiation protection course for radiologists L1Radiation protection course for radiologists L1
Radiation protection course for radiologists L1Amin Amin
Ā 
Target volume definition
Target volume definitionTarget volume definition
Target volume definitionAmin Amin
Ā 
Radiation safety during therapeutic administration of radioiodine 131
Radiation safety during therapeutic administration of radioiodine 131Radiation safety during therapeutic administration of radioiodine 131
Radiation safety during therapeutic administration of radioiodine 131Amin Amin
Ā 
Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...
Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...
Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...Amin Amin
Ā 
Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...
Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...
Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...Amin Amin
Ā 
Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...
Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...
Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...Amin Amin
Ā 
Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...
Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...
Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...Amin Amin
Ā 
Geometric errors in 3DCRT and IMRT
Geometric errors in 3DCRT and IMRTGeometric errors in 3DCRT and IMRT
Geometric errors in 3DCRT and IMRTAmin Amin
Ā 
Guidelines for patients receiving radioiodine i 131 treatment
Guidelines for patients receiving radioiodine i 131 treatmentGuidelines for patients receiving radioiodine i 131 treatment
Guidelines for patients receiving radioiodine i 131 treatmentAmin Amin
Ā 
Proton beam therapy
Proton beam therapyProton beam therapy
Proton beam therapyAmin Amin
Ā 
Medical linear accelerator
Medical linear acceleratorMedical linear accelerator
Medical linear acceleratorAmin Amin
Ā 
Hypofractionated radiotherapy during covid 19 pandemic
Hypofractionated radiotherapy during covid 19 pandemic Hypofractionated radiotherapy during covid 19 pandemic
Hypofractionated radiotherapy during covid 19 pandemic Amin Amin
Ā 

More from Amin Amin (20)

Laser Light in Cancer Therapy
Laser Light in Cancer TherapyLaser Light in Cancer Therapy
Laser Light in Cancer Therapy
Ā 
Radiation protection course for radiologists L8
Radiation protection course for radiologists L8Radiation protection course for radiologists L8
Radiation protection course for radiologists L8
Ā 
Radiation protection course for radiologists L7
Radiation protection course for radiologists L7Radiation protection course for radiologists L7
Radiation protection course for radiologists L7
Ā 
Radiation protection course for radiologists L6
Radiation protection course for radiologists L6Radiation protection course for radiologists L6
Radiation protection course for radiologists L6
Ā 
Radiation protection course for radiologists L5
Radiation protection course for radiologists L5Radiation protection course for radiologists L5
Radiation protection course for radiologists L5
Ā 
Radiation protection course for radiologists L4
Radiation protection course for radiologists L4Radiation protection course for radiologists L4
Radiation protection course for radiologists L4
Ā 
Radiation protection course for radiologists Lecture 3
Radiation protection course for radiologists Lecture 3Radiation protection course for radiologists Lecture 3
Radiation protection course for radiologists Lecture 3
Ā 
Radiation protection course for radiologists L2
Radiation protection course for radiologists L2Radiation protection course for radiologists L2
Radiation protection course for radiologists L2
Ā 
Radiation protection course for radiologists L1
Radiation protection course for radiologists L1Radiation protection course for radiologists L1
Radiation protection course for radiologists L1
Ā 
Target volume definition
Target volume definitionTarget volume definition
Target volume definition
Ā 
Radiation safety during therapeutic administration of radioiodine 131
Radiation safety during therapeutic administration of radioiodine 131Radiation safety during therapeutic administration of radioiodine 131
Radiation safety during therapeutic administration of radioiodine 131
Ā 
Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...
Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...
Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...
Ā 
Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...
Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...
Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...
Ā 
Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...
Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...
Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...
Ā 
Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...
Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...
Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...
Ā 
Geometric errors in 3DCRT and IMRT
Geometric errors in 3DCRT and IMRTGeometric errors in 3DCRT and IMRT
Geometric errors in 3DCRT and IMRT
Ā 
Guidelines for patients receiving radioiodine i 131 treatment
Guidelines for patients receiving radioiodine i 131 treatmentGuidelines for patients receiving radioiodine i 131 treatment
Guidelines for patients receiving radioiodine i 131 treatment
Ā 
Proton beam therapy
Proton beam therapyProton beam therapy
Proton beam therapy
Ā 
Medical linear accelerator
Medical linear acceleratorMedical linear accelerator
Medical linear accelerator
Ā 
Hypofractionated radiotherapy during covid 19 pandemic
Hypofractionated radiotherapy during covid 19 pandemic Hypofractionated radiotherapy during covid 19 pandemic
Hypofractionated radiotherapy during covid 19 pandemic
Ā 

Recently uploaded

Shazia Iqbal 2024 - Bioorganic Chemistry.pdf
Shazia Iqbal 2024 - Bioorganic Chemistry.pdfShazia Iqbal 2024 - Bioorganic Chemistry.pdf
Shazia Iqbal 2024 - Bioorganic Chemistry.pdfTrustlife
Ā 
Call Girl In Mysore šŸ’ÆNiamh šŸ“²šŸ”7427069034šŸ”Call Girls NošŸ’°Advance Cash On Deliver...
Call Girl In Mysore šŸ’ÆNiamh šŸ“²šŸ”7427069034šŸ”Call Girls NošŸ’°Advance Cash On Deliver...Call Girl In Mysore šŸ’ÆNiamh šŸ“²šŸ”7427069034šŸ”Call Girls NošŸ’°Advance Cash On Deliver...
Call Girl In Mysore šŸ’ÆNiamh šŸ“²šŸ”7427069034šŸ”Call Girls NošŸ’°Advance Cash On Deliver...chaddageeta79
Ā 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxSwetaba Besh
Ā 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationMedicoseAcademics
Ā 
Top 10 Most Beautiful Russian Pornstars List 2024
Top 10 Most Beautiful Russian Pornstars List 2024Top 10 Most Beautiful Russian Pornstars List 2024
Top 10 Most Beautiful Russian Pornstars List 2024locantocallgirl01
Ā 
Female Call Girls Sawai Madhopur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call ...
Female Call Girls Sawai Madhopur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call ...Female Call Girls Sawai Madhopur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call ...
Female Call Girls Sawai Madhopur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call ...Dipal Arora
Ā 
Female Call Girls Tonk Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Serv...
Female Call Girls Tonk  Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Serv...Female Call Girls Tonk  Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Serv...
Female Call Girls Tonk Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Serv...Dipal Arora
Ā 
Part I - Anticipatory Grief: Experiencing grief before the loss has happened
Part I - Anticipatory Grief: Experiencing grief before the loss has happenedPart I - Anticipatory Grief: Experiencing grief before the loss has happened
Part I - Anticipatory Grief: Experiencing grief before the loss has happenedbkling
Ā 
šŸ‘‰ Saharanpur Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl...
šŸ‘‰ Saharanpur Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl...šŸ‘‰ Saharanpur Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl...
šŸ‘‰ Saharanpur Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl...chaddageeta79
Ā 
Intro to disinformation and public health
Intro to disinformation and public healthIntro to disinformation and public health
Intro to disinformation and public healthTina Purnat
Ā 
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan 081901222272 Obat Penggugur Kandu...
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan  081901222272 Obat Penggugur Kandu...Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan  081901222272 Obat Penggugur Kandu...
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan 081901222272 Obat Penggugur Kandu...Halo Docter
Ā 
Lucknow Call Girls Service { 9984666624 } ā¤ļøVVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ā¤ļøVVIP ROCKY Call Girl in Lucknow U...Lucknow Call Girls Service { 9984666624 } ā¤ļøVVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ā¤ļøVVIP ROCKY Call Girl in Lucknow U...Janvi Singh
Ā 
See it and Catch it! Recognizing the Thought Traps that Negatively Impact How...
See it and Catch it! Recognizing the Thought Traps that Negatively Impact How...See it and Catch it! Recognizing the Thought Traps that Negatively Impact How...
See it and Catch it! Recognizing the Thought Traps that Negatively Impact How...bkling
Ā 
MOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATRO
MOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATROMOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATRO
MOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATROKanhu Charan
Ā 
šŸ‘‰ Gulbarga Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl S...
šŸ‘‰ Gulbarga Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl S...šŸ‘‰ Gulbarga Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl S...
šŸ‘‰ Gulbarga Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl S...chaddageeta79
Ā 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan 087776558899
Ā 
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan CytotecJual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotecjualobat34
Ā 
TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...
TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...
TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...rightmanforbloodline
Ā 
Call Now ā˜Ž 9549551166 || Call Girls in Dehradun Escort Service Dehradun
Call Now ā˜Ž 9549551166  || Call Girls in Dehradun Escort Service DehradunCall Now ā˜Ž 9549551166  || Call Girls in Dehradun Escort Service Dehradun
Call Now ā˜Ž 9549551166 || Call Girls in Dehradun Escort Service DehradunJanvi Singh
Ā 
Female Call Girls Jodhpur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Se...
Female Call Girls Jodhpur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Se...Female Call Girls Jodhpur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Se...
Female Call Girls Jodhpur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Se...Dipal Arora
Ā 

Recently uploaded (20)

Shazia Iqbal 2024 - Bioorganic Chemistry.pdf
Shazia Iqbal 2024 - Bioorganic Chemistry.pdfShazia Iqbal 2024 - Bioorganic Chemistry.pdf
Shazia Iqbal 2024 - Bioorganic Chemistry.pdf
Ā 
Call Girl In Mysore šŸ’ÆNiamh šŸ“²šŸ”7427069034šŸ”Call Girls NošŸ’°Advance Cash On Deliver...
Call Girl In Mysore šŸ’ÆNiamh šŸ“²šŸ”7427069034šŸ”Call Girls NošŸ’°Advance Cash On Deliver...Call Girl In Mysore šŸ’ÆNiamh šŸ“²šŸ”7427069034šŸ”Call Girls NošŸ’°Advance Cash On Deliver...
Call Girl In Mysore šŸ’ÆNiamh šŸ“²šŸ”7427069034šŸ”Call Girls NošŸ’°Advance Cash On Deliver...
Ā 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
Ā 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their Regulation
Ā 
Top 10 Most Beautiful Russian Pornstars List 2024
Top 10 Most Beautiful Russian Pornstars List 2024Top 10 Most Beautiful Russian Pornstars List 2024
Top 10 Most Beautiful Russian Pornstars List 2024
Ā 
Female Call Girls Sawai Madhopur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call ...
Female Call Girls Sawai Madhopur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call ...Female Call Girls Sawai Madhopur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call ...
Female Call Girls Sawai Madhopur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call ...
Ā 
Female Call Girls Tonk Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Serv...
Female Call Girls Tonk  Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Serv...Female Call Girls Tonk  Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Serv...
Female Call Girls Tonk Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Serv...
Ā 
Part I - Anticipatory Grief: Experiencing grief before the loss has happened
Part I - Anticipatory Grief: Experiencing grief before the loss has happenedPart I - Anticipatory Grief: Experiencing grief before the loss has happened
Part I - Anticipatory Grief: Experiencing grief before the loss has happened
Ā 
šŸ‘‰ Saharanpur Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl...
šŸ‘‰ Saharanpur Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl...šŸ‘‰ Saharanpur Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl...
šŸ‘‰ Saharanpur Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl...
Ā 
Intro to disinformation and public health
Intro to disinformation and public healthIntro to disinformation and public health
Intro to disinformation and public health
Ā 
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan 081901222272 Obat Penggugur Kandu...
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan  081901222272 Obat Penggugur Kandu...Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan  081901222272 Obat Penggugur Kandu...
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan 081901222272 Obat Penggugur Kandu...
Ā 
Lucknow Call Girls Service { 9984666624 } ā¤ļøVVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ā¤ļøVVIP ROCKY Call Girl in Lucknow U...Lucknow Call Girls Service { 9984666624 } ā¤ļøVVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ā¤ļøVVIP ROCKY Call Girl in Lucknow U...
Ā 
See it and Catch it! Recognizing the Thought Traps that Negatively Impact How...
See it and Catch it! Recognizing the Thought Traps that Negatively Impact How...See it and Catch it! Recognizing the Thought Traps that Negatively Impact How...
See it and Catch it! Recognizing the Thought Traps that Negatively Impact How...
Ā 
MOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATRO
MOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATROMOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATRO
MOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATRO
Ā 
šŸ‘‰ Gulbarga Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl S...
šŸ‘‰ Gulbarga Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl S...šŸ‘‰ Gulbarga Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl S...
šŸ‘‰ Gulbarga Call Girls Service Just Call šŸ‘šŸ‘„7427069034 šŸ‘šŸ‘„ Top Class Call Girl S...
Ā 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Ā 
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan CytotecJual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Ā 
TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...
TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...
TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...
Ā 
Call Now ā˜Ž 9549551166 || Call Girls in Dehradun Escort Service Dehradun
Call Now ā˜Ž 9549551166  || Call Girls in Dehradun Escort Service DehradunCall Now ā˜Ž 9549551166  || Call Girls in Dehradun Escort Service Dehradun
Call Now ā˜Ž 9549551166 || Call Girls in Dehradun Escort Service Dehradun
Ā 
Female Call Girls Jodhpur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Se...
Female Call Girls Jodhpur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Se...Female Call Girls Jodhpur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Se...
Female Call Girls Jodhpur Just Call Dipal šŸ„°8250077686šŸ„° Top Class Call Girl Se...
Ā 

Teletherapy cobalt 60 machines vs linear accelerator

  • 1. Teletherapy Cobalt-60 Machines Vs Linear Accelerator Prof Amin E AAmin Dean of the Higher Institute of Optics Technology & Prof of Medical Physics Radiation Oncology Department Faculty of Medicine Ain Shams University
  • 2. Introduction ā‘ Both the cobalt-60 teletherapy machine and clinical linac were introduced in 1950s and used for megavoltage radiotherapy since then. ā‘ During the past 50 years: ā€¢ The basic design of the cobalt machine remained essentially the same. ā€¢ Linac progressed to the current design through five generations, eclipsed the cobalt machine, and became the most widely used radiation source in modern radiotherapy, with several 1000 units in clinical practice around the world today.
  • 3. Cobalt-60 Teletherapy Units Versus Linacs In comparison with cobalt-60 teletherapy machines linacs have become very complex in design: āž¢Because of the multimodality capabilities that have evolved and are available on most modern linacs (several x-ray energies and several electron energies). āž¢Because of an increased use of computer logic and microprocessors in the control systems of linacs. āž¢Because of added features, such as high dose rate modes, multileaf collimation, electron arc therapy, and the dynamic treatment option on the collimators (dynamic wedge), MLC leaves (IMRT), gantry or table while the beam is turned on.
  • 4. COBALT-60 TELETHERAPY UNITS VERSUS LINACS Despite clear technological and practical advantages of linacs over cobalt-60 machines, the latter still occupy an important place in radiotherapy armamentarium, especially in developing countries, because of: ā€¢ Their considerably lower capital and installation cost. ā€¢ Lower servicing and maintenance cost. ā€¢ Lesser dependence on reliable electrical power. ā€¢ Simplicity of design. ā€¢ Ease of operation.
  • 5. Cobalt Units Vs Linear Accelerators ā€¢ Several arguments have been put forward both for and against Cobalt Units as well as Linear accelerators. ā€¢ These arguments relate to physics, clinical advantages and more importantly, the cost consideration. ā€¢ The following few slides summaries the data that is available.
  • 6. Cobalt Units Vs Linear Accelerators Feature Low Energy Accelerator Cobalt 60Unit Remarks Dose rate 600 MU/Min Reducing dose starts at 200r/min In Cobalt units dose will reduce because of decay, and depends on source activity whereas with Linear Accelerator one has the guarantee of constant dose rate
  • 7. Cobalt Units Vs Linear Accelerators Feature Low Energy Accelerator Cobalt 60Unit Remarks Energy Source DIA <2mm 20mm ā€«ā€ā€¬Penumbra effect is higher in cobalt machines
  • 8. Cobalt Units Vs Linear Accelerators Feature Low Energy Accelerator Cobalt 60Unit Remarks Source change None Every 5-10 years (in average 7 years) ā€«ā€ā€¬Source cost for 12,000 curie source is very much on the increasing trend. At the time of source change, decayed source has to be sent back to the manufacturer, which is very expensive.
  • 9. Cobalt Units Vs Linear Accelerators Feature Low Energy Accelerator Cobalt 60Unit Remarks D- max 1.5 cm Only 0.5cm ā€«ā€ā€¬Linear accelerator with 6 MV, photon, we are able to get D-max 1.5 cm. This is very useful for deep seated tumours. Skin sparting is better with Linac (6 mv).
  • 10. Cobalt Units Vs Linear Accelerators Feature Low Energy Accelerator Cobalt 60Unit Remarks Perchantage Depth dose at 10cm 67% 55% Difficult to treat deep seated tumors in Thorax, abdomen or pelvis.
  • 11. Central Axis Depth Dose Distribution For Different Quality Photon Beams
  • 12. Cobalt Units Vs Linear Accelerators Feature Low Energy Accelerator Cobalt 60Unit Remarks Colimator trasmission Not exceed o.5% More than 3% ā€” Minimum field size 0.5x0.5 cm 5 x 5 cm In cobalt one cannot treat field size less than 5x5cm. ā€«ā€ā€¬Radiation leakage through the head Nil Less than 0.2% ā€”
  • 13. Cobalt Units Vs Linear Accelerators Feature Low Energy Accelerator Cobalt 60Unit Remarks Penumbra Less than 5 mm More than 1 cm With 2 cm. Cobalt source the Penumbra effect is very high. Motorized wedges Possible Not possible __ Variable dose rate 80 to 600 MU/min Not variable Dose rate if fixed, decreases with decay of source
  • 14. Cobalt Units Vs Linear Accelerators Feature Low Energy Accelerator Cobalt 60Unit Remarks Precision Dose delivered to and arc is reproducible within 3% or 1MU which ever is greater __ In linear accelerator one can have reproducibility of dosimetry system and the dose delivered is continuously monitored and recorded
  • 15. Cobalt Units Vs Linear Accelerators Feature Low Energy Accelerator Cobalt 60Unit Remarks Multileaf collimators, stereotactic attachments Possible Not possible Increases accuracy in dose delivery IMRT Possible Not possible Increases accuracy in dose delivery
  • 16. Cobalt Units Vs Linear Accelerators Feature Low Energy Accelerator Cobalt 60Unit Remarks ā€«ā€ā€¬Asymmetric collimato jaws Yes, standard features No Electronics are cheaper Independenty collimator jaws are very useful for precise and accurate cancer treatment
  • 17. Cobalt Units Vs Linear Accelerators Feature Low Energy Accelerator Cobalt 60Unit Remarks Portal vision attachment Yes, possible (Optinal) No Allows to view actual tumour while treatment Patient support system Superior Inferior, low accuracy
  • 18. Cobalt Units Vs Linear Accelerators Feature Low Energy Accelerator Cobalt 60Unit Remarks Beam matching with second unit Possible Not possible Beam matching is a very good feature available in Linear accelerator. This will help institutions to buy the 2nd machine with the same beam characteristics of the existing units. Thereby patients can be shifted between different Linear accelerator. More flexibility.
  • 19. Cobalt Units Vs Linear Accelerators Feature Low Energy Accelerator Cobalt 60Unit Remarks Source disposal Nil Major problem with cobalt units, and source cost is increasing every year. It is also becoming increasingly difficult to obtain a source In the event the source is imported along with the cobalt machine the decayed source has to be transported back to the manufacturers, which is very expensive.
  • 20. ā€¢ Cobalt 60 units provide relatively high energy gamma rays for radiotherapy which are ideally suited for treatment of head and neck cancers and other superficially located tumors like breast cancers and soft tissue sarcomas of extremities. ā€¢ They are not adequate for treatment of deep seated tumors and have the added disadvantage of decreasing output with decay of source and the need for source replacement within 5-7 years. Cobalt Units Vs Linear Accelerators
  • 21. ā€¢ Disposal of decayed source is another major concern. ā€¢ The beam characteristics when compared to 6mv Linacs are inferior and fewer ancillaries are available for cobalt machines as compared to Linacs. ā€¢ High energy Linacs in addition to giving two or three x-ray energies can also generate variable energy electron beam for treatment. Cobalt Units Vs Linear Accelerators
  • 22. ā€¢ While such high energy Linacs are expense, 6 MV Linacs compare favorably in terms of cost with the Cobalt 60 units. Even though the initial cost appears to be high, over a ten year period maintenance costs are less as it does not require change of source. ā€¢ It is stressed that service contracts with performance guarantees for 5-10 years have to be built into the contracts while procuring the equipment. Cobalt Units Vs Linear Accelerators
  • 23. ā€¢ To summarize Linear accelerators have several advantages: 1. Very high energy beams can be created with a machine that is not very bulky or cumbersome to use. 2. The edges of the beams are much more sharply defined than those of a cobalt machine, allowing additional precision in dose delivery. Advantages Of Linear Accelerators
  • 24. 3. Electron beams can be created (with high energy linacs) that is of particular value in treating superficial lesions. 4. The dose rate per minute is variable and can be turned up very high allowing the patient to be located at substantial distance from the machine in order to create large fields necessary for total skin or total body irradiation while still maintaining adequate dose rate. With cobalt, the rate is determined by the amount of cobalt source in the machine and cannot be regulated. Advantages Of Linear Accelerators
  • 25. 5. Advanced techniques (three dimension conformal radiotherapy, IMRT, VMAT, IGRT) can be applied. Advantages Of Linear Accelerators