SlideShare a Scribd company logo
1 of 139
Download to read offline
Radiation Protection Course For Radiologists
Lecture 8 of 8
Procedures for Minimizing Dose to Patients
And X-ray Personnel
Prof Amin E AAmin
Dean of the Higher Institute of Optics Technology
&
Prof of Medical Physics
Radiation Oncology Department
Faculty of Medicine, Ain Shams University
Equipment Specifications
New X-Ray Equipment
• All new medical x-ray equipment and accessories for such
equipment, must conform to the requirements of the Egyptian
law and regulations of installing new radiology equipment.
• These regulations specify standards of design, construction
and performance, with respect to radiation safety, and are
mandatory requirements for new equipment only.
• It is the responsibility of the manufacturer or distributor to
ensure that the equipment conforms to the requirements of
the regulations.
Existing X-ray Equipment
Warning Signs
• The x-ray control panel must bear a permanent and conspicuous
sign warning that hazardous x-radiation is emitted when the
equipment is in operation and prohibiting unauthorized use.
Markings
• All controls, meters, lights and other indicators relevant to the
operation of the equipment must be readily discernible and
clearly labeled or marked as to function.
Indicator Lights
• There must be readily discernible, separate indicators on the
control panel that respectively indicate:
(i) when the control panel is energized and the machine is ready to
produce x-rays, and
(ii) when x-rays are produced.
(iii)When more than one x-ray tube is controlled by one control panel,
there must be clear and visible indication, at or near the tube housing
and on the control panel, of which tube is connected and ready to be
energized.
Focal Spot
• The location of the focal spot must be clearly and accurately
marked on the tube housing.
Filtration
• The external surface of the x-ray tube housing must bear a
permanent mark or label which sets out the minimum
permanent inherent filtration in the useful beam, expressed as
millimetres of aluminum equivalent at a specified peak tube
potential.
Filtration
• The total permanent filtration in the useful beam must be
equivalent to at least the following thicknesses of aluminum:
(i) 0.5 millimetre of aluminum, for machines designed to operate with
x-ray tube potentials below 50 kilovolts peak;
(ii) 1.5 millimetres of aluminum, for machines designed to operate
with x-ray tube potentials from 50 kilovolts peak to 70 kilovolts
peak, and
(iii) 2.5 millimetres of aluminum, for machines designed to operate
with x-ray tube potentials above 70 kilovolts peak;
Mechanical Stability
• The x-ray tube must be securely fixed and correctly aligned
within the tube housing.
• Also, the tube housing must maintain its required exposure
position or movement without excessive drift or vibration
during operation.
• The x-ray tube housing must be supported by mechanical
means. It must not be hand-held during operation.
Exposure Control
• There must be an exposure switch, timer, or other device to
initiate and terminate x-ray production.
• This control must automatically terminate the exposure after a
preset time, product of current and time, or exposure has
elapsed.
Exposure Control
• The exposure switch must:
(i) require continuous pressure by the operator to produce x-rays (i.e.,
“dead man” type);
(ii) if in the form of a footswitch, be so constructed that if overturned
inadvertent exposures do not result;
(iii) be so located that convenient operation outside of a shielded area is
not possible, (except for exposure switches used in conjunction with
mobile x-ray equipment, with spot film devices, in fluoroscopy and with
certain special procedures), and
(iv) for mobile x-ray equipment, be equipped with a cable at least three
meters long.
Indication of Technique Factors
• The actual peak kilovoltage should correspond to the selected
or indicated value to within 5% of the selected or indicated
value.
• The millammeter should be accurate to within 5% of the full
scale reading and be temperature compensated for normal
operating conditions.
X-Ray Tube Shielding
• The shielding of the housing must be such that, at each rating
specified by the manufacturer for that tube, the leakage
radiation, measured at a distance of one metre in any direction
from the focal spot of the x-ray tube, does not exceed 0.1% of
the exposure rate at the same distance along the central axis of
the useful beam.
Stationary General Purpose
Radiographic Equipment
Collimation
• The x-ray tube housing must be equipped with an adjustable
collimator which provides stepless adjustment of the size of
the x-ray field and which must provide the same degree of
shielding towards leakage radiation as is required of the tube
housing.
• The collimator should incorporate means to indicate the size
of the x-ray field at the image receptor. if in the form of a
light localizer, it should give an average illumination of at
least 100 lux at 100 cm or the maximum target-to-image
receptor distance, whichever is less.
Filtration
• The filters must be permanently and securely mounted to the x-
ray port of the tube housing or beam limiting device, or both.
Timer Accuracy
• The timer or exposure control must be such that:
(i) it can be set to control exposures as short as 1/60 second or 5
milliampere-seconds, whichever is greater, and
(ii) at each setting it is accurate to 1/60 second or 7% of that setting,
whichever is greater.
Reproducibility
• The coefficient of variation of any 10 consecutive radiation
measurements, taken at the same source-to-detector distance
within a time period of one hour and with the line voltage for
each measurement within 1% of the mean value for all of the
measurements, must not be greater than 0.05.
• Furthermore, each of the 10 radiation exposures measured
must be within 15% of the mean value of the 10
measurements.
Linearity
• For any fixed value of x-ray tube potential (in kVp), within
the range of values specified for the equipment, the average
ratios of exposure (in milliroentgen or coulomb/kg) to tube
current-exposure time product (in milliampere-seconds) at
any two consecutive tube current settings must not differ by
more than 0.10 times their sum.
Line Voltage Compensator
• The equipment must be equipped with means to compensate for
variations in x-ray tube potential caused by line voltage
fluctuations.
• By these means it must be possible to set and maintain the peak
tube potential (kVp), prior to initiation of an exposure, to within
5% of the selected value, for a line voltage variation of ± 7% of
its nominal value.
Current Stabilization
• The equipment should be provided with a tube current stabilizer
to maintain preset current to within ± 5% of the required value
for exposures of 1/5 second or more.
• Changes in tube potential should not result in a variation of tube
current greater than ± 5%.
Mobile Radiographic Equipment
Mobile Radiographic Equipment
• The requirements given above should be, applied.
• Additional requirements for mobile radiographic equipment are
as follows:
Target-To-Skin Distance
• The equipment must be equipped with means to prevent
operation at target-to-skin distances of less than 30 centimetres.
Exposure Control
• The exposure switch must be of the deadman type and must be
provided with a cable at least 3 metres long.
Capacitor Discharge Equipment
Capacitor discharge type mobile equipment must be equipped with:
– a safety shutter or other means to prevent emission of xradiation from
the exit port of the x-ray tube housing assembly when the exposure
switch or timer is not activated;
– a warning light, or other visual indicator, to indicate that the high
potential capacitor is charged;
– an electrical meter, or other indicator, to indicate the state of charge of
the high potential capacitor, and
– means to enable the high potential capacitor to be discharged to a
residual potential difference of less than 30 volts, without the production
of x-radiation.
Fluoroscopic Equipment
Collimation
• The x-ray tube housing must be equipped with an adjustable
collimator and which must provide the same degree of
shielding towards leakage radiation as is required of the
tube housing.
• The maximum size of x-ray field permitted by the
collimator should not be larger than the useful area of the
input phosphor on the image intensifier. The useful beam
should be centred on the input phosphor and the beam
margins should be tangential to the area being viewed.
Direct Bean Absorbers
• A primary protective barrier must be permanently incorporated
into the equipment to intercept the entire cross section of the
direct beam, for all possible target-to image receptor distances.
The barrier must be so installed that its removal from the
useful beam causes automatic termination of the exposure
• The lead equivalent thickness of the primary protective barrier
must be at least 2 mm for machines capable of operating up to
100 kVp. For each additional kilovolt of operating tube
potential an additional 0.01 mm lead equivalent is required.
Direct Bean Absorbers
• A primary protective barrier must be permanently incorporated
into the equipment to intercept the entire cross section of the
direct beam, for all possible target-to image receptor distances.
• The barrier must be so installed that its removal from the useful
beam causes automatic termination of the exposure
Image Intensification
• All fluoroscopic x-ray machines must be equipped with an
image intensification system.
• The image intensification system must be adequately shielded
so that neither the useful beam nor the scattered radiation from
the intensifier will produce significant exposure to the operator
or other personnel.
Image Intensification
• The image intensification system must be adequately shielded so
that neither the useful beam nor the scattered radiation from the
intensifier will produce significant exposure to the operator or
other personnel.
• The shielding must be such that the exposure rate to an object
having a cross-sectional area of 100 cm2, with no linear dimension
greater than 20 cm, of x-radiation due to transmission of the useful
beam through or scattering from the window of the fluoroscopic
imaging assembly, does not exceed 2 milliroentgen per hour (516
nC/kg/h) at 10 centimetres behind the plane of the image receptor,
for each roentgen per minute of entrance exposure rate.
Target-To-Skin Distance
• Fixed fluoroscopic equipment must incorporate means to
limit the target-to-skin distance to not less than 38 cm.
• In the case of a fluoroscopic machine designed for
special applications that would be impossible at the
above minimum distance, provision may be made for
operation at shorter target-to-skin distances; but in no
case must this distance be less than 20 cm.
Material Between Patient And
Image Receptor
• The aluminum equivalent of the table top must not be greater
than 1 mm when measured at 100 kVp.
Exposure Switch
• The fluoroscopic exposure switch must be of a dead-man type.
Timer
• There must be a cumulative timing device, activated by the
fluoroscopic exposure switch, which requires manual resetting.
• This timer must have a maximum setting not exceeding 5
minutes. Whenever a preset time has been reached, it must give
a clearly audible signal.
Indication of Tube Potential and Current
• Electrical meters or other visual indicators must be provided to
enable continuous monitoring by the operator of tube potential
and current during fluoroscopy.
Entrance Exposure Rate
• At the shortest target-to-skin distance specified for the
equipment, the entrance exposure rate must not exceed 10
roentgen (2.58 mC/kg) per minute for any combination of tube
current and potential. With modern equipment most
fluoroscopy can and should be carried out with exposure rates
of less than 5 roentgen (1.29 mC/kg) per minute.
Bucky Slot Shielding
• Shielding must be provided with the equipment to cover the
Bucky Slot. This should provide the equivalent protection of at
least 0.5 mm of lead at 100 kVp.
Table Shielding
• The side of the fluoroscopic table should have an equivalent
thickness of at least 0.5 mm of lead at 100 kVp to restrict
scattered radiation passing through it.
Operator Shielding
• To protect the operator from radiation scattered above the
table-top, there should be a lead rubber apron of
dimensions not less than 45.7 cm × 45.7 cm (18" × 18").
• The material must have a lead equivalent of not less than
0.5 mm at 100 kVp and the apron must be attached to the
lower edge of the fluoroscopic screen when the latter is
vertical and to the side when the screen is horizontal.
• The apron may consist of several overlapping parts.
Operator Shielding
• In the absence of such a protective apron a retractable lead
shield must be Incorporated in the equipment to project at
least 15 cm above the table when in use.
• It must be mounted on the front side of the table and have a
lead-equivalent thickness of at least 0.5 mm at 100 kVp.
Photofluorographic Equipment
Collimation
• The equipment must be permanently fitted with a
collimator which must confine the beam to the area of the
fluorescent screen.
• The beam must not be greater than 35 cm × 43 cm (14" ×
17") at the screen surface, regardless of the target-to-skin
distance.
Beam Alignment
• The photofluorographic camera and x-ray tube must be coupled
together so that the direct beam always centres on the mid-point
of the screen.
Gonad Shield
• The lower edge of the beam-defining device should be curved
to prevent irradiation of the female gonads.
Maximum Entrance Exposure
• The entrance exposure to a patient must not exceed 200
milliroentgen and should not exceed 100 milliroentgen (25.8
microcoulomb/kg) per film.
Protective Clothing
1. Protective body aprons
2. Gonad shields
3. Protective gloves
Protective Body Aprons
• Protective body aprons used for radiographic or fluoroscopic
examinations with peak x-ray tube potentials of up to 150 kVp
must provide attenuation equivalent to at least 0.5 mm of lead.
Gonad Shields
• Contact-type gonad shields used for routine diagnostic
radiology must have a lead equivalent thickness of at least 0.25
mm and should have a lead equivalent thickness of 0.5 mm at
150 kVp.
• Contact-type gonad shields must be of sufficient size and shape
to exclude the gonads completely from primary beam
irradiation.
Protective Gloves
• Protective gloves used in fluoroscopy must provide attenuation
equivalent to at least 0.25 mm of lead at 150 kVp.
Film Cassettes
• The front panel of the film cassette must not exceed an
aluminum equivalence thickness of 1.0 mm and should not
exceed an aluminum equivalence thickness of 0.5 mm.
• Film cassettes must be completely light-tight. As soon as
the material used for achieving light-tightness becomes
worn and light leakage occurs, either the material or the
cassette should be replaced.
• It is also important to have good film-screen contact;
therefore screens should be cleaned regularly and their
contact with the film checked.
Procedures To Reduce Dose To
X-ray Personnel
Basic Principles of Dose
Reduction
•Time
•Distance
•Shielding
Reduction of External Dose
❖Minimize the time spent near the radiation source
❖Maximize the distance away from the source
❖Make use of available shielding
General Recommendations
1. Only One Investigation
• An x-ray room must not be used for more than one
radiological investigation simultaneously.
2. Only Essential Persons
• Except for those persons whose presence is
essential, no person must be in the x-ray room
when the exposure is carried out.
General Recommendations
3. Be Away from the useful beam
– Personnel must at all times keep as far away from the useful
beam as is practicable. Exposure of personnel to the useful
beam must never be allowed unless the beam is adequately
attenuated by the patient and by protective clothing or screens.
General Recommendations
4. Use Protective Devices
– All personnel must take full advantage of the protective devices
available.
5. Stay Behind Shielding
4. Operators should remain inside the control booth or behind
protective screens when making an x-ray exposure. In cases
where there are reasons that make this impractical, protective
clothing must be worn.
General Recommendations
6. Rules for patient support
– When there is a need to support children or weak patients,
holding devices should be used.
– If parents, escorts or other personnel are called to assist, they
must be provided with protective aprons and gloves, and be
positioned so as to avoid the useful beam.
– No one person should regularly perform these duties.
General Recommendations
7. Personal Dosimetry
– When a lead equivalent protective apron is worn, the personnel
dosimeter must be worn under the apron.
– If extremities are likely to be exposed to significantly higher
doses, additional extremity monitors should be worn.
General Recommendations
8. Personal Dosimetry
– All operators of x-ray equipment, together with personnel (e.g.
nurses) who routinely participate in radiological procedures,
and others likely to receive a radiation dose in excess of 1/20th
of the Dose limit specified in Appendix I, must wear personnel
dosimeters.
General Recommendations
9. Close All Doors
– All entrance doors to an x-ray room, including patient cubicle
and preparation room doors, should be kept closed while a
patient is in the room.
10.Attend The energized machine
– X-ray machines which are energized and ready to produce
radiation must not be left unattended.
General Recommendations
11.In the case of high exposure
– Where radiation doses in excess of 5% of the dose limit specified
in Appendix I are being received regularly by any one person,
appropriate remedial steps must be taken to improve techniques
and protective measures.
General Recommendations
12.Machine Operator
– X-ray equipment must only be operated by, or under the
direct supervision of, qualified individuals.
13.Don’t hold the x-ray housing
– An x-ray housing must not be held by hand during
operation.
Recommendations For Operation
Of Radiographic Units
1. Controlling the machines
– The x-ray exposure should, as a general rule, be controlled
from the control panel located inside the control booth or
behind a shielded wall. In the case of special techniques
where the operator is required to control the exposure while
at the side of the patient, appropriate protective clothing
must be worn.
Recommendations For Operation
Of Radiographic Units
2. Communication with patient
– The operator must have a clear view of the patient during every
exposure and be able to communicate with the patient and/or
attendants without leaving the control booth.
3. Don’t hold the Cassettes
2. Cassettes must never be held by hand during an exposure.
Recommendations For Operation Of
Fluoroscopic Units
1. Protective Clothing
– All persons, with the possible exception of the patient, required to be
in the room during a fluoroscopic procedure should wear protective
aprons. Lead shields or curtains mounted on the fluoroscopic unit
must not be considered a sufficient substitute for the wearing of
protective clothing.
Recommendations For Operation Of
Fluoroscopic Units
2. Protective Gauntlets
– Protective gauntlets should be worn
by the radiologist during palpation in
every fluoroscopic examination.
During fluoroscopy, palpation with
the hand should be kept to a
minimum.
Recommendations For Operation
Of Fluoroscopic Units
3. Protective Clothing
– During fluoroscopy and spotfilm
operation associated with
fluoroscopic operation, where
personnel are required to be at
the side of the patient,
appropriate protective clothing
must be worn by these personnel.
Recommendations For Operation Of
Fluoroscopic Units
4. Shorten The Time
– All fluoroscopic examinations should be carried out as
rapidly as possible and with minimum dose-rates and x-ray
field sizes.
5. Image Intensified Fluoroscopy
– Direct-viewing fluoroscopy must not be carried out; image
intensified fluoroscopy must be used.
Recommendations For Operation
Of Mobile Units
1. Don’t Use Mobile If You Can Use Stationary
– Mobile units must be used only if the condition of the
patient is such as to make it inadvisable for the examination
to be carried out with a stationary unit in the main x-ray
department.
Recommendations For Operation
Of Mobile Units
2. Don’t Irradiate Other Persons
– During operation, the primary beam should be directed from
occupied areas if at all possible, and every effort must be made
to ensure that this beam does not irradiate any other persons in
the vicinity of the patient.
Recommendations For Operation
Of Mobile Units
3. Keep The Distance
– The operator must stand at least 3 metres from the x-ray tube and out of the
direct beam.
4. Use The Shielding
– The operator must be shielded when exposures are made.
Recommendations For Operation
Of Mobile Units
5. Discharge The Capacitor
– In a capacitor discharge unit after an x-ray exposure has been
made there is a residual charge left in the capacitors. This
residual charge can give rise to a “dark current” and result in
x-ray emission even though the exposure switch is not
activated. Therefore, the residual charge must be fully
discharged before the unit is left unattended.
Recommendations For Operation Of
Photofluorographic Units
• Photofluorographic x-ray machines in general can give
significantly higher patient doses than conventional diagnostic
machines.
Recommendations For Operation
Of Photofluorographic Units
1. Only the operator and the patient should be in the room or
vehicle in which the unit is installed when an exposure is
made.
2. When a photofluorographic unit is mounted in a vehicle
appropriate steps must be taken to ensure that persons awaiting
examination and personnel are protected from exposure to x-
radiation. Such persons should be kept well clear of the vehicle
when exposures are being made.
Recommendations For Special
Radiological Procedures
• Cystoscopic radiography
• Hystero-salpingography
• Needle Biopsies of the Lung
• Transhepatic Needle Cholangiography
• Pancreatography, and other similar procedures.
Recommendations For Special
Radiological Procedures
• In these procedures the radiologist and other personnel in the
vicinity of the patient can be subjected to appreciable scattered
radiation from the patient when the x-ray beam is on.
Therefore, the radiologist and other personnel should wear
protective glasses and clothing and should remain as far away
from the patient as practicable, unless adequate scatter shields
can be incorporated into the x-ray equipment.
Recommendations For Special
Radiological Procedures
• The protective devices (e.g., shielded panels, leaded drapes,
extended collimator cones, etc.) provided with the x-ray
equipment should be used whenever they do not interfere
unduly with the diagnostic procedures.
• The smallest x-ray field consistent with the procedure should
be used.
Recommendations For Special
Radiological Procedures
• Angiography. Angiography is potentially one of the sources of
greatest exposure to personnel in diagnostic radiology, since it
requires the presence of a considerable number of personnel
close to the patient and involves fluoroscopy for extended
periods of time and multiple radiographic exposures. For such
procedures all personnel must be aware of the radiation hazards
involved and make every effort to adhere to the
recommendations below:
Recommendations For Special
Radiological Procedures
1. Use Fully protected Device
– Full use must be made of the protective devices provided with x-
ray equipment (e.g., shielded panels, leaded drapes, bucky slot
covers, etc.).
2. Wear Protective Clothing
– All personnel must wear protective clothing and personnel
dosimeters. Protective glasses should also be worn.
Recommendations For Special
Radiological Procedures
3. Go Away or behind shield
– All personnel who are not required to be immediately adjacent to the
patient during the procedure must stand back as far as possible from
the patient and, if at all possible, should stand behind a protective
shield.
4. Use shields if possible
– Where indicated and feasible, special shields in addition to the
protective devices provided with the machine should be used.
Procedures For Minimizing Dose To
Patients
Guidelines For The Prescription Of
Diagnostic X-ray Examinations
• The medical practitioner is in a unique position to reduce
unnecessary radiation exposure to the patient by ensuring that
all examinations are clinically justified. The practitioner can
achieve this by adhering, as much as possible, to certain basic
recommendations. These are as follows:
Guidelines For The Prescription
Of Diagnostic X-ray Examinations
1. The prescription of an x-ray examination of a patient should
only be based on a clinical evaluation of the patient and
should be for the purpose of obtaining diagnostic information.
2. Routine or screening examinations, such as for
preemployment physical examinations, tuberculosis
screening, mass mammographic screening, etc., in which
there is no prior clinical evaluation of the patient, should not
be prescribed.
Guidelines For The Prescription
Of Diagnostic X-ray Examinations
3. It should be determined whether there have been any previous
x-ray examinations which would make further examination
unnecessary, or allow for the ordering of an abbreviated
examination. Relevant previous radiographs or reports should
be examined along with a clinical evaluation of the patient.
4. When a patient is transferred from one physician or hospital
to another any relevant radiographs or reports should
accompany the patient and should be reviewed by the
consulting physician.
Guidelines For The Prescription Of
Diagnostic X-ray Examinations
5. When prescribing a radiological examination, the physician
should specify precisely the clinical indications and
information required.
6. The number of radiographic views, required in an
examination, should be kept to the minimum practicable,
consistent with the clinical objectives of the examination.
7. In prescribing x-ray examinations of pregnant or possibly
pregnant women, full consideration should be taken of the
consequences of foetal exposure.
Guidelines For The Prescription
Of Diagnostic X-ray Examinations
8. If a radiograph contains the required information, repeat
exposures should not be prescribed simply because the
radiograph may not be of the “best” diagnostic quality.
9. Specialized studies should be undertaken only by, or in
close collaboration with, a qualified radiologist. Medical
practitioners should not operate x-ray equipment, or be
responsible for the use of such equipment.
Guidelines For The Prescription
Of Diagnostic X-ray Examinations
10.A patient’s clinical records should include details of x-ray
examinations carried out.
11.The prescription of an x-ray examination of a patient should
only be based on a clinical evaluation of the patient and should
be for the purpose of obtaining diagnostic information.
12.Routine or screening examinations, such as for
preemployment physical examinations, tuberculosis screening,
mass mammographic screening, etc., in which there is no prior
clinical evaluation of the patient, should not be prescribed.
Guidelines For The Prescription Of
Diagnostic X-ray Examinations
13.It should be determined whether there have been any previous
x-ray examinations which would make further examination
unnecessary, or allow for the ordering of an abbreviated
examination. Relevant previous radiographs or reports should
be examined along with a clinical evaluation of the patient.
14.When a patient is transferred from one physician or hospital to
another any relevant radiographs or reports should accompany
the patient and should be reviewed by the consulting physician.
Guidelines For The Prescription
Of Diagnostic X-ray Examinations
15.When prescribing a radiological examination, the physician
should specify precisely the clinical indications and
information required.
16.The number of radiographic views, required in an examination,
should be kept to the minimum practicable, consistent with the
clinical objectives of the examination.
17.In prescribing x-ray examinations of pregnant or possibly
pregnant women, full consideration should be taken of the
consequences of foetal exposure.
Guidelines For The Prescription Of
Diagnostic X-ray Examinations
18.If a radiograph contains the required information, repeat
exposures should not be prescribed simply because the
radiograph may not be of the “best” diagnostic quality.
19.Specialized studies should be undertaken only by, or in close
collaboration with, a qualified radiologist. Medical
practitioners should not operate x-ray equipment, or be
responsible for the use of such equipment.
20.A patient’s clinical records should include details of x-ray
examinations carried out.
Guidelines For Radiography Of
Pregnant Women
• Every effort should be made to avoid unnecessary irradiation of
any woman known to be, or who might be pregnant. This is
particularly important during the earliest stages of pregnancy
when the potential for radiation damage of the rapidly dividing
tissues is the greatest. Clearly, however, in spite of the
possibility of radiation damage, if a radiological examination is
required for the diagnosis or management of an urgent medical
problem it must be done, irrespective of whether the patient
may or may not be pregnant.
Guidelines For Radiography Of
Pregnant Women
• The following recommendations apply to x-ray
examinations involving pregnant or potentially
pregnant women:
Guidelines For Radiography Of
Pregnant Women
1. Ten day Role
– Radiography of the pelvic area in women of child-bearing age
should be undertaken in the ten-day period following the
onset of menstruation, since the risk of pregnancy is very
small during this period.
Guidelines For Radiography Of
Pregnant Women
2. No Elective Abdominal nor Pelvic X-ray
– Only essential investigations should be taken in the case of
pregnant or suspected pregnant women. Elective radiography
of the abdominal and pelvic area in pregnant women must be
avoided. (“Elective” is taken to mean an examination of the
abdomen and pelvis which does not contribute to the diagnosis
or treatment of a women in relation to her immediate illness.)
Guidelines For Radiography Of
Pregnant Women
3. No Photofluorographic
– Pregnant women or women who may be pregnant must
not be accepted for chest photofluorographic (mass
radiographic) examinations.
Guidelines For Radiography Of
Pregnant Women
4. Use protective Shields
– When radiography of the pelvic area
or abdomen is required, the exposure
must be kept to the absolute
minimum necessary and full use
must be made of gonadal shielding
and other protective shielding if the
clinical objectives of the
exmaination will not be
compromised.
Guidelines For Radiography Of
Pregnant Women
5. Foetal dose is lower in prone Position
– If a radiographic examination of the foetus is required the prone
position should be used. This has the effect of shielding the
foetus from the softer x-rays and hence reducing the foetal dose.
Guidelines For Radiography Of
Pregnant Women
6. Radiography should not be used for the determination of
abnormal presentations of the foetus, or for placenta
localizations. Other techniques such as ultrasonography
are better suited for this purpose.
Guidelines For Radiography Of
Pregnant Women
7. Use Well-Collimated X-Ray Beam
– Radiography of the chest, extremities, etc., of a pregnant woman,
for valid clinical reasons, should only be carried out using a well-
collimated x-ray beam and with proper regard for shielding of the
abdominal area.
Guidelines For Radiographic Examination
Of The Breast (Mammography)
The recommendations that follow are intended to provide
guidance to the technologist, the medical physicist and
radiologist in exercising their responsibility towards reduction
of patient dose.
1. The mammographic X-ray system must be designed
specifically for mammography and the image receptor must
be compatible with the system.
2. A film-screen combination that provides good quality
diagnostic results must be used. Direct exposure film must
never be used in mammography.
Guidelines For Radiographic Examination
Of The Breast (Mammography)
3. Except in the case of mammography performed within a
screening program, the technologist must not perform any
examination which has not been prescribed by a physician
responsible for the patient.
4. The dose to the patient must be kept to the lowest practicable
value consistent with clinical objectives, and without loss of
essential diagnostic information. To achieve this, techniques
appropriate to the equipment available should be used and
evaluated from time to time in terms of effectiveness.
Guidelines For Radiographic Examination
Of The Breast (Mammography)
5. Particular care in patient X-ray protection must be taken when
mammographic X-ray examinations of pregnant or possibly
pregnant women are carried out, even though the radiation dose to
the abdomen and fetus are negligible during normal mammographic
X-ray examinations.
6. There may be a need to use special procedures when performing
mammographic X-ray examinations of patients with breast implants
such as extra views, modified positioning, compression and loading
techniques. Particular care of compression techniques must be taken
since excessive compression of the implants during mammographic
X-ray examination may cause rupture of the implant.
Guidelines For Radiographic Examination
Of The Breast (Mammography)
7. In mammography, it is recommended that the field normally be
the full size of the image receptor, but not larger than the image
receptor support, except at the chest wall. The amount of
unexposed areas on the films should be minimal so to avoid the
need for masking. Collimated-down views are useful in some
purposes, but should be considered as a special procedure.
8. Appropriate compression must be used in all mammographic
procedures.
9. Adequate X-ray beam filtration must be used in all
mammographic procedures.
Guidelines For Radiographic Examination
Of The Breast (Mammography)
10.The technologist should use the maximum focal spot to skin
distance consistent with good radiographic loading. For
mammographic procedures, distances of less than 50 cm
should not be used. For mammographic procedures including
magnification, distances of less than 25 cm to the entrance
surface of the breast shall not be used.
11.Full details of the mammographic procedures, including
retakes, carried out should be noted on the patient’s clinical
records.
Guidelines For Radiographic Examination
Of The Breast (Mammography)
12.Irradiation times should be minimized to avoid unnecessary dose
increase from reciprocity law failure and to avoid motion artifacts.
This can be accomplished by the use of sufficiently high tube
current values.
13.Cassettes should be loaded at least 30 minutes in advance to allow
air to escape, and thus improving film/screen contact.
14.The technologist should examine the images after processing in
order to verify that the techniques being used are producing
diagnostic quality images and that the X-ray equipment is
functioning correctly.
Guidelines For Radiographic Examination
Of The Breast (Mammography)
15.An appropriate Quality Assurance program must be
implemented on mammographic X-ray equipment and on
film and Xerographic processing systems.
Guidelines For Radiographic Examination
Of The Breast (Mammography)
16.While recommended dose limits have been defined for radiation
workers and the general population, there is no specific limits
recommended for patients undergoing diagnostic X-rays
procedures. However, it is possible to provide limits on the
amount of radiation breast tissues receive by setting dose limits
on the level of radiation a representative breast phantom would
receives for a given irradiation. These dose limits are presented
in the following table;
Guidelines For Radiographic Examination
Of The Breast (Mammography)
Procedure Mean Glandular Dose Limits
Film-Screen Mammography (no
grid)
1.0 mGy
Film-Screen Mammography (grid
technique)
2.0 mGy
Xeromammography 4.0 mGy
Recommended dose limits per irradiation
Radiological Chest Screening
1. The use of mass radiological screening should only be
considered in areas of high incidence of tuberculosis or in
special population groups who may be particularly susceptible
to lung disease. Selection of population groups for x-ray
screening should be based upon the probability of discovering a
significant number of cases of cardiopulmonary disease and on
the availability of full follow-up facilities for individuals
requiring treatment.
2. Mass x-ray screening of the general population for tuberculosis
and other chest diseases must not be carried out.
Radiological Chest Screening
3. Routine hospital admission chest x-rays should only be taken
in identified high risk groups.
4. Chest x-rays should not be required as a routine part of a pre-
employment medical check-up or for admission to educational
institutions. Since photofluorographic x-ray machines require a
higher level of x-ray exposure than do convential radiographic
machines to produce a film and since the yield of significant
information is low, photofluorographic x-ray machines must
not be used for chest examinations in place of conventional
machines, and should not be used for mass screening when
conventional machines are available.
Guidelines For Carrying Out X-ray
Examinations
General Recommendations
1. The operator must not perform any examination which has not
been prescribed by a physician responsible for the patient.
2. The exposure of the patient must be kept to the lowest
practicable value, consistent with clinical objectives and
without loss of essential diagnostic information. To achieve
this, techniques appropriate to the equipment available should
be used.
3. Particular care, consistent with the recommendations of Section
13.2, must be taken when radiological examinations of
pregnant or potentially pregnant women are carried out.
General Recommendations
4. The x-ray beam must be well-collimated to restrict it as much
as is practicable to the area of diagnostic interest.
5. The x-ray beam size must be limited to the size of the image
receptor or smaller.
6. The x-ray beam should not be directed towards the gonads
unless it is absolutely essential, in which case gonad
shielding must be used whenever the value of the
examination is not impaired by such use. Guidelines on the
use of gonad shielding are given in section 14 of this
guideline.
General Recommendations
7. Shielding should be used where appropriate and practicable to
limit the exposure of body tissues. It is particularly important
that special effort to be made to protect the blood-forming
organs, gonads and thyroids of children.
8. The target-to-skin distance should be as great as possible,
consistent with good radiographic technique.
9. For very young children, special devices should be employed
to restrict movement.
10.Full details of the radiological procedures carried out should be
noted on the patient’s clinical records.
Recommendations For
Radiographic Procedures
1. The edges of the x-ray beam should be seen on all x-ray
films to ensure that no more than the desired area has been
irradiated. The film size used should be as small as possible,
consistent with the diagnostic objectives of the examination.
2. Screen-type film should not be used for non-screen
techniques because it is less sensitive to direct x-radiation
than non-screen film.
Recommendations For
Radiographic Procedures
3. The fastest film or intensifying screen-film combination,
consistent with diagnostically-acceptable results, should
be used. When highest definition is not required a high-
speed film-screen combination should be used.
Recommendations For
Radiographic Procedures
4. X-ray intensifying screens made from rare earth phosphors
should be used where appropriate. Exposure reductions in
excess of 50%, compared to conventional calcium tungstate
systems, are possible using the rare earth-type intersifying
screens and the films developed for use with such screens.
These reductions result from greater x-ray absorption and
increased conversion efficiencies of the new phosphors —
the rare earth systems have a conversion efficiency of x-rays
to light of 15-20%, compared to approximately 5% for
calcium tungstate systems.
Recommendations For
Radiographic Procedures
5. To ensure that patient exposure is kept to a minimum,
consistent with image quality, full advantage should be taken of
a combination of techniques, such as:
a. use of an antiscatter grid or air gap between the patient and the image
receptor;
b. use of the optimum focus-to-film distance appropriate to the
examination;
c. use of the highest kilovoltage which produces films of good quality;
d. use of automatic exposure control devices designed to keep all
exposures and repeat exposures to a minimum.
Recommendations For
Radiographic Procedures
6. The radiographer should see the films after processing in
order to verify that the techniques being used are producing
diagnostic quality films and that the x-ray equipment is
functioning correctly.
7. To avoid the necessity of retakes, it is particularly important
before taking a long series of films that a single preliminary
film of the series should be taken and processed to verify
correctness of settings.
Recommendations For
Radiographic Procedures
8. While dose limits have been defined for radiation workers and the
general population, no specific limits have been recommended, to
date, for patients undergoing diagnostic x-ray procedures. For
patients the risk involved in the exposure must always be
weighed against the medical requirement for accurate diagnosis.
However, it is now feasible to recommend upper limits on patient
skin entrance exposure, for routine, non-specialty, radiographic
examinations. For a reference patient, having the
anthropometrical characteristics shown in Table 1, the skin
entrance exposure (without backscatter) should not exceed the
values indicated in Table 2.
Recommendations For
Radiographic Procedures
Body Part Thickness (cm)
Head (lateral)
Neck (A/P)
Chest (P/A)
Abdomen (A/P)
Foot (D/P)
15
13
23
23
8
Table 1
Anthropometrical Characteristics of the Reference Patient
Recommendations For
Radiographic Procedures
Examination (Projection) ESE (mR)
Chest (P/A)
Skull (Lateral)
Abdomen (A/P)
Cervical Spine (A/P)
Thoracic Spine (A/P)
Full Spine (A/P)
Lumbo-Sacral Spine (A/P)
Retrograde Pyelogram (A/P)
Feet (D/P)
20
170
450
120
400
250
500
500
200
Table 2
Recommended Upper Limits on Skin Entrance Exposure
Recommendations For
Fluoroscopic Procedures
1. In view of the relatively high exposure that results from
fluoroscopy, such procedures should only be carried out
when an equivalent result cannot be obtained from
radiography. Fluoroscopy must not be used as a substitute for
radiography.
2. Flurorscopy must only be carried out by, or under immediate
supervision of, a radiologist or physician properly trained in
fluoroscopic procedures.
3. All fluoroscopic procedures should be carried out as rapidly
as possible with the smallest practical x-ray field sizes.
Recommendations For
Fluoroscopic Procedures
4. The exposure rate used in fluoroscopy should be as low
as possible and must not exceed 10 roentgens (2.58
mC/kg) per minute at the position where the central axis
of the x-ray beam enters the patient. With modern
equipment, most fluoroscopic procedures can be readily
carried out with exposure rates of less than 5 roentgens
(1.29 mC/kg) per minute.
Recommendations For
Fluoroscopic Procedures
5. Image intensification must be used in order to reduce patient
exposure. Image intensifiers can significantly reduce both
exposure rate and exposure time. However, the operator
must monitor the x-ray tube current and voltage on
equipment with automatic brightness control, since both can
rise to high values without the knowledge of the operator,
particularly if the gain of the intensifier is decreased.
6. Television monitoring should be used in conjunction with
the image intensifier.
Recommendations For
Fluoroscopic Procedures
7. Mobile fluoroscopic equipment should only be used for
examinations where it is impractical to transfer patients to
a permanent fluoroscopic installation.
8. Cinefluorography produces the highest patient doses in
diagnostic radiology because the x-ray tube currents and
potentials used are generally higher than those used in
fluoroscopy. Therefore, this technique should not be used
unless significant medical benefit is expected.
Recommendations For
Photofluorographic Procedures
1. The x-ray beam must be restricted to the area of the
fluorescent screen and should be limited to the minimum
size consistent with clinical requirements.
2. Appropriate steps must be taken to exclude or shield the
gonads from the direct beam.
3. The performance of the equipment must be monitored
routinely. Special emphasis should be placed on the speed
of the optical systems used. These should be as fast as
possible and should be replaced when noticeable
deterioration in performance occurs.
Recommendations For Special
Procedures
1. Significant exposure to the patient’s eyes can result during
neurological examinations, such as carotid angiography. In
projections where it does not interfere with the diagnostic
information sought, eye shielding should be used.
2. During cardiac catheterization and angiography significant
exposure of the patient’s thyroid gland can occur. Therefore,
appropriate shielding should be used whenever possible
Guidelines For Reduction Of Gonad Dose
Recommendations For Reducing
Gonad Dose To The Patient
Radiologists and radiographers must pay special attention to
four factors that are important for reducing gonad dose to the
patient:
1. Correct collimation of the x-ray beam. — It is not sufficient
merely to limit the beam to the size of the image receptor.
Care should be taken to further restrict the beam to the
region of the patient’s body that is of diagnostic interest.
Irradiation of any part of the body outside that region
contributes nothing to the objective of the examination and
only increases dose both to the body and the gonads.
Recommendations For Reducing
Gonad Dose To The Patient
2. Gonad shielding. — Appropriate use of specific area
gonad shielding is strongly advised when:
(i) the gonads, of necessity, lie within, or are in close
proximity to, the primary x-ray beam;
(ii) the patient has reasonable reproductive potential; and
(iii) clinical objectives will not be compromised.
Recommendations For Reducing
Gonad Dose To The Patient
3. Appropriate selection of technique factors. — An
appropriate selection of tube voltage, current and filtration
is particularly important for diagnostic procedures in
which the gonads lie within or near the primary x-ray
beam. For example, in fluoroscopy, use of higher tube
voltage and filtration and lower tube current will almost
always reduce the gonad dose.
Recommendations For Reducing
Gonad Dose To The Patient
4. Sensitivity of imaging systems. — The gonad dose is
inversely proportional to the sensitivity of the imaging
system. Thus, doubling the sensitivity halves the gonad
dose; conversely, halving the sensitivity doubles the gonad
dose. It is, therefore, very important to maintain the
sensitivity of the imaging system at its optimum value and
to be alert for any significant deterioration.
Recommendations For Reducing Gonad
Dose To Operators
• Adherence to the safe use procedures specified in Section 11 of
this Safety Code will ensure that gonad dose to radiologists and
radiographers is kept to the minimum practicable.
Radiation Protection Course Equipment Specs

More Related Content

What's hot

Linear accelerator vinay
Linear accelerator vinayLinear accelerator vinay
Linear accelerator vinayVinay Desai
 
Measurement of absorbed dose
Measurement of absorbed doseMeasurement of absorbed dose
Measurement of absorbed dosePurvi Rathod
 
Basic Radiation Protection
Basic Radiation ProtectionBasic Radiation Protection
Basic Radiation ProtectionMae Minguillo
 
Radiation protection in nm
Radiation protection in nmRadiation protection in nm
Radiation protection in nmBlessyPhilip4
 
radiation protection...Koustav Majumder....
radiation protection...Koustav Majumder....radiation protection...Koustav Majumder....
radiation protection...Koustav Majumder....Koustav Majumder
 
CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexVivek Elangovan
 
Viewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic imageViewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic imageSHASHI BHUSHAN
 
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxAREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxZubairUlIslam5
 
Mammographic equipment
Mammographic equipmentMammographic equipment
Mammographic equipmentSharmaRajan4
 
Radiation protection
Radiation protection   Radiation protection
Radiation protection Anil Gupta
 
Radiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionRadiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionNitish Virmani
 
Radiation protection PowerPoint slide PPT pk
Radiation protection PowerPoint slide PPT pkRadiation protection PowerPoint slide PPT pk
Radiation protection PowerPoint slide PPT pkDr pradeep Kumar
 

What's hot (20)

Linear accelerator vinay
Linear accelerator vinayLinear accelerator vinay
Linear accelerator vinay
 
Measurement of absorbed dose
Measurement of absorbed doseMeasurement of absorbed dose
Measurement of absorbed dose
 
X ray generators
X ray generatorsX ray generators
X ray generators
 
Basic Radiation Protection
Basic Radiation ProtectionBasic Radiation Protection
Basic Radiation Protection
 
Radiation protection in nm
Radiation protection in nmRadiation protection in nm
Radiation protection in nm
 
Crso Simpkin Radiation Shielding Design
Crso Simpkin Radiation Shielding DesignCrso Simpkin Radiation Shielding Design
Crso Simpkin Radiation Shielding Design
 
radiation protection...Koustav Majumder....
radiation protection...Koustav Majumder....radiation protection...Koustav Majumder....
radiation protection...Koustav Majumder....
 
Chapter4 radiation monitoring instruments
Chapter4 radiation monitoring instrumentsChapter4 radiation monitoring instruments
Chapter4 radiation monitoring instruments
 
CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose Index
 
Quality assurance
Quality  assuranceQuality  assurance
Quality assurance
 
Evolution of radiation 2012
Evolution of radiation 2012Evolution of radiation 2012
Evolution of radiation 2012
 
Viewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic imageViewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic image
 
Basic of medical linac
Basic of medical linacBasic of medical linac
Basic of medical linac
 
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxAREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
 
Mammographic equipment
Mammographic equipmentMammographic equipment
Mammographic equipment
 
Radiation protection
Radiation protection   Radiation protection
Radiation protection
 
Linear accelerator
Linear acceleratorLinear accelerator
Linear accelerator
 
Radiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionRadiographic Image contrast & image resolution
Radiographic Image contrast & image resolution
 
Radiation protection PowerPoint slide PPT pk
Radiation protection PowerPoint slide PPT pkRadiation protection PowerPoint slide PPT pk
Radiation protection PowerPoint slide PPT pk
 
Nuclear imaging
Nuclear imagingNuclear imaging
Nuclear imaging
 

Similar to Radiation Protection Course Equipment Specs

ATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptxATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptxAALIA ABDULLAH
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfAngelina Protik
 
Radiation safety and protection for dental radiography
Radiation safety and protection for dental radiographyRadiation safety and protection for dental radiography
Radiation safety and protection for dental radiographyNitin Sharma
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCQUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCAnupriyaNR
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptJgdishrathi
 
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working Procedures
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working ProceduresNon-Destructive Testing (NDT) - Industrial Radiography Normal Working Procedures
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working Proceduresshahar_sayuti
 
Portable and mobile radiographic equipments [Autosaved].pptx
Portable and mobile radiographic equipments [Autosaved].pptxPortable and mobile radiographic equipments [Autosaved].pptx
Portable and mobile radiographic equipments [Autosaved].pptxSaritaGaire1
 
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxAkashSharma640477
 
5.radiation protection
5.radiation protection5.radiation protection
5.radiation protectionsamad shaik
 
Radiology department planning
Radiology department planning Radiology department planning
Radiology department planning Prajwith Rai
 
Radiologydepartment_INTRODUCTION AND PLANING.pptx
Radiologydepartment_INTRODUCTION AND PLANING.pptxRadiologydepartment_INTRODUCTION AND PLANING.pptx
Radiologydepartment_INTRODUCTION AND PLANING.pptxDRTRUPTISONTHALIA
 
Validation of uv visible spectrophotometer.pptx
Validation of uv visible spectrophotometer.pptxValidation of uv visible spectrophotometer.pptx
Validation of uv visible spectrophotometer.pptxAnupamaCp2
 
Honeywell is25100tc-spec-sheet
Honeywell is25100tc-spec-sheetHoneywell is25100tc-spec-sheet
Honeywell is25100tc-spec-sheetAlarm Grid
 
mobile image intensifier units.ppt
mobile image intensifier units.pptmobile image intensifier units.ppt
mobile image intensifier units.pptSAMEER AHMAD GANAIE
 

Similar to Radiation Protection Course Equipment Specs (20)

ATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptxATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptx
 
Pixel cp operators manual
Pixel cp operators manualPixel cp operators manual
Pixel cp operators manual
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
 
Radiation safety and protection for dental radiography
Radiation safety and protection for dental radiographyRadiation safety and protection for dental radiography
Radiation safety and protection for dental radiography
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCQUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
 
Fluoroscopy
FluoroscopyFluoroscopy
Fluoroscopy
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.ppt
 
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working Procedures
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working ProceduresNon-Destructive Testing (NDT) - Industrial Radiography Normal Working Procedures
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working Procedures
 
Portable and mobile radiographic equipments [Autosaved].pptx
Portable and mobile radiographic equipments [Autosaved].pptxPortable and mobile radiographic equipments [Autosaved].pptx
Portable and mobile radiographic equipments [Autosaved].pptx
 
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
 
5.radiation protection
5.radiation protection5.radiation protection
5.radiation protection
 
QUALITY CONTROL PDF.pptx
QUALITY CONTROL PDF.pptxQUALITY CONTROL PDF.pptx
QUALITY CONTROL PDF.pptx
 
LINAC final.pptx
LINAC final.pptxLINAC final.pptx
LINAC final.pptx
 
Radiology department planning
Radiology department planning Radiology department planning
Radiology department planning
 
Radiologydepartment_INTRODUCTION AND PLANING.pptx
Radiologydepartment_INTRODUCTION AND PLANING.pptxRadiologydepartment_INTRODUCTION AND PLANING.pptx
Radiologydepartment_INTRODUCTION AND PLANING.pptx
 
Validation of uv visible spectrophotometer.pptx
Validation of uv visible spectrophotometer.pptxValidation of uv visible spectrophotometer.pptx
Validation of uv visible spectrophotometer.pptx
 
Honeywell is25100tc-spec-sheet
Honeywell is25100tc-spec-sheetHoneywell is25100tc-spec-sheet
Honeywell is25100tc-spec-sheet
 
mobile image intensifier units.ppt
mobile image intensifier units.pptmobile image intensifier units.ppt
mobile image intensifier units.ppt
 
Quality control
Quality  control Quality  control
Quality control
 
linac QA.pptx
linac QA.pptxlinac QA.pptx
linac QA.pptx
 

More from Amin Amin

Laser Light in Cancer Therapy
Laser Light in Cancer TherapyLaser Light in Cancer Therapy
Laser Light in Cancer TherapyAmin Amin
 
Radiation protection course for radiologists L6
Radiation protection course for radiologists L6Radiation protection course for radiologists L6
Radiation protection course for radiologists L6Amin Amin
 
Radiation protection course for radiologists L4
Radiation protection course for radiologists L4Radiation protection course for radiologists L4
Radiation protection course for radiologists L4Amin Amin
 
Radiation protection course for radiologists Lecture 3
Radiation protection course for radiologists Lecture 3Radiation protection course for radiologists Lecture 3
Radiation protection course for radiologists Lecture 3Amin Amin
 
Radiation protection course for radiologists L2
Radiation protection course for radiologists L2Radiation protection course for radiologists L2
Radiation protection course for radiologists L2Amin Amin
 
Radiation protection course for radiologists L1
Radiation protection course for radiologists L1Radiation protection course for radiologists L1
Radiation protection course for radiologists L1Amin Amin
 
Target volume definition
Target volume definitionTarget volume definition
Target volume definitionAmin Amin
 
Radiation safety during therapeutic administration of radioiodine 131
Radiation safety during therapeutic administration of radioiodine 131Radiation safety during therapeutic administration of radioiodine 131
Radiation safety during therapeutic administration of radioiodine 131Amin Amin
 
Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...
Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...
Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...Amin Amin
 
Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...
Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...
Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...Amin Amin
 
Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...
Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...
Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...Amin Amin
 
Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...
Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...
Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...Amin Amin
 
Geometric errors in 3DCRT and IMRT
Geometric errors in 3DCRT and IMRTGeometric errors in 3DCRT and IMRT
Geometric errors in 3DCRT and IMRTAmin Amin
 
Guidelines for patients receiving radioiodine i 131 treatment
Guidelines for patients receiving radioiodine i 131 treatmentGuidelines for patients receiving radioiodine i 131 treatment
Guidelines for patients receiving radioiodine i 131 treatmentAmin Amin
 
Proton beam therapy
Proton beam therapyProton beam therapy
Proton beam therapyAmin Amin
 
Teletherapy cobalt 60 machines vs linear accelerator
Teletherapy cobalt 60 machines vs linear acceleratorTeletherapy cobalt 60 machines vs linear accelerator
Teletherapy cobalt 60 machines vs linear acceleratorAmin Amin
 
Medical linear accelerator
Medical linear acceleratorMedical linear accelerator
Medical linear acceleratorAmin Amin
 
Hypofractionated radiotherapy during covid 19 pandemic
Hypofractionated radiotherapy during covid 19 pandemic Hypofractionated radiotherapy during covid 19 pandemic
Hypofractionated radiotherapy during covid 19 pandemic Amin Amin
 
Levels of radiation dose in vascular surgical procedures
Levels of radiation dose in vascular surgical procedures Levels of radiation dose in vascular surgical procedures
Levels of radiation dose in vascular surgical procedures Amin Amin
 
Teletherapy Cobalt-60 Machines
Teletherapy Cobalt-60 MachinesTeletherapy Cobalt-60 Machines
Teletherapy Cobalt-60 MachinesAmin Amin
 

More from Amin Amin (20)

Laser Light in Cancer Therapy
Laser Light in Cancer TherapyLaser Light in Cancer Therapy
Laser Light in Cancer Therapy
 
Radiation protection course for radiologists L6
Radiation protection course for radiologists L6Radiation protection course for radiologists L6
Radiation protection course for radiologists L6
 
Radiation protection course for radiologists L4
Radiation protection course for radiologists L4Radiation protection course for radiologists L4
Radiation protection course for radiologists L4
 
Radiation protection course for radiologists Lecture 3
Radiation protection course for radiologists Lecture 3Radiation protection course for radiologists Lecture 3
Radiation protection course for radiologists Lecture 3
 
Radiation protection course for radiologists L2
Radiation protection course for radiologists L2Radiation protection course for radiologists L2
Radiation protection course for radiologists L2
 
Radiation protection course for radiologists L1
Radiation protection course for radiologists L1Radiation protection course for radiologists L1
Radiation protection course for radiologists L1
 
Target volume definition
Target volume definitionTarget volume definition
Target volume definition
 
Radiation safety during therapeutic administration of radioiodine 131
Radiation safety during therapeutic administration of radioiodine 131Radiation safety during therapeutic administration of radioiodine 131
Radiation safety during therapeutic administration of radioiodine 131
 
Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...
Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...
Radiation Protection Course For Orthopedic Specialists: Lecture 4 of 4: Recom...
 
Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...
Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...
Radiation Protection Course For Orthopedic Specialists: Lecture 3 of 4: Basic...
 
Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...
Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...
Radiation Protection Course For Orthopedic Specialists Lecture 2 of 4 Radiati...
 
Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...
Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...
Radiation Protection Course For Orthopedic Specialists: Lecture 1 of 4: Intro...
 
Geometric errors in 3DCRT and IMRT
Geometric errors in 3DCRT and IMRTGeometric errors in 3DCRT and IMRT
Geometric errors in 3DCRT and IMRT
 
Guidelines for patients receiving radioiodine i 131 treatment
Guidelines for patients receiving radioiodine i 131 treatmentGuidelines for patients receiving radioiodine i 131 treatment
Guidelines for patients receiving radioiodine i 131 treatment
 
Proton beam therapy
Proton beam therapyProton beam therapy
Proton beam therapy
 
Teletherapy cobalt 60 machines vs linear accelerator
Teletherapy cobalt 60 machines vs linear acceleratorTeletherapy cobalt 60 machines vs linear accelerator
Teletherapy cobalt 60 machines vs linear accelerator
 
Medical linear accelerator
Medical linear acceleratorMedical linear accelerator
Medical linear accelerator
 
Hypofractionated radiotherapy during covid 19 pandemic
Hypofractionated radiotherapy during covid 19 pandemic Hypofractionated radiotherapy during covid 19 pandemic
Hypofractionated radiotherapy during covid 19 pandemic
 
Levels of radiation dose in vascular surgical procedures
Levels of radiation dose in vascular surgical procedures Levels of radiation dose in vascular surgical procedures
Levels of radiation dose in vascular surgical procedures
 
Teletherapy Cobalt-60 Machines
Teletherapy Cobalt-60 MachinesTeletherapy Cobalt-60 Machines
Teletherapy Cobalt-60 Machines
 

Recently uploaded

VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...Miss joya
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Servicemakika9823
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableNehru place Escorts
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Miss joya
 
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...Miss joya
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls ServiceMiss joya
 
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...narwatsonia7
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...narwatsonia7
 
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls ServiceMiss joya
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...narwatsonia7
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call girls in Ahmedabad High profile
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...CALL GIRLS
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...Taniya Sharma
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 

Recently uploaded (20)

VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
 
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
 
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 

Radiation Protection Course Equipment Specs

  • 1. Radiation Protection Course For Radiologists Lecture 8 of 8 Procedures for Minimizing Dose to Patients And X-ray Personnel Prof Amin E AAmin Dean of the Higher Institute of Optics Technology & Prof of Medical Physics Radiation Oncology Department Faculty of Medicine, Ain Shams University
  • 3. New X-Ray Equipment • All new medical x-ray equipment and accessories for such equipment, must conform to the requirements of the Egyptian law and regulations of installing new radiology equipment. • These regulations specify standards of design, construction and performance, with respect to radiation safety, and are mandatory requirements for new equipment only. • It is the responsibility of the manufacturer or distributor to ensure that the equipment conforms to the requirements of the regulations.
  • 5. Warning Signs • The x-ray control panel must bear a permanent and conspicuous sign warning that hazardous x-radiation is emitted when the equipment is in operation and prohibiting unauthorized use.
  • 6. Markings • All controls, meters, lights and other indicators relevant to the operation of the equipment must be readily discernible and clearly labeled or marked as to function.
  • 7. Indicator Lights • There must be readily discernible, separate indicators on the control panel that respectively indicate: (i) when the control panel is energized and the machine is ready to produce x-rays, and (ii) when x-rays are produced. (iii)When more than one x-ray tube is controlled by one control panel, there must be clear and visible indication, at or near the tube housing and on the control panel, of which tube is connected and ready to be energized.
  • 8. Focal Spot • The location of the focal spot must be clearly and accurately marked on the tube housing.
  • 9. Filtration • The external surface of the x-ray tube housing must bear a permanent mark or label which sets out the minimum permanent inherent filtration in the useful beam, expressed as millimetres of aluminum equivalent at a specified peak tube potential.
  • 10. Filtration • The total permanent filtration in the useful beam must be equivalent to at least the following thicknesses of aluminum: (i) 0.5 millimetre of aluminum, for machines designed to operate with x-ray tube potentials below 50 kilovolts peak; (ii) 1.5 millimetres of aluminum, for machines designed to operate with x-ray tube potentials from 50 kilovolts peak to 70 kilovolts peak, and (iii) 2.5 millimetres of aluminum, for machines designed to operate with x-ray tube potentials above 70 kilovolts peak;
  • 11. Mechanical Stability • The x-ray tube must be securely fixed and correctly aligned within the tube housing. • Also, the tube housing must maintain its required exposure position or movement without excessive drift or vibration during operation. • The x-ray tube housing must be supported by mechanical means. It must not be hand-held during operation.
  • 12. Exposure Control • There must be an exposure switch, timer, or other device to initiate and terminate x-ray production. • This control must automatically terminate the exposure after a preset time, product of current and time, or exposure has elapsed.
  • 13. Exposure Control • The exposure switch must: (i) require continuous pressure by the operator to produce x-rays (i.e., “dead man” type); (ii) if in the form of a footswitch, be so constructed that if overturned inadvertent exposures do not result; (iii) be so located that convenient operation outside of a shielded area is not possible, (except for exposure switches used in conjunction with mobile x-ray equipment, with spot film devices, in fluoroscopy and with certain special procedures), and (iv) for mobile x-ray equipment, be equipped with a cable at least three meters long.
  • 14. Indication of Technique Factors • The actual peak kilovoltage should correspond to the selected or indicated value to within 5% of the selected or indicated value. • The millammeter should be accurate to within 5% of the full scale reading and be temperature compensated for normal operating conditions.
  • 15. X-Ray Tube Shielding • The shielding of the housing must be such that, at each rating specified by the manufacturer for that tube, the leakage radiation, measured at a distance of one metre in any direction from the focal spot of the x-ray tube, does not exceed 0.1% of the exposure rate at the same distance along the central axis of the useful beam.
  • 17. Collimation • The x-ray tube housing must be equipped with an adjustable collimator which provides stepless adjustment of the size of the x-ray field and which must provide the same degree of shielding towards leakage radiation as is required of the tube housing. • The collimator should incorporate means to indicate the size of the x-ray field at the image receptor. if in the form of a light localizer, it should give an average illumination of at least 100 lux at 100 cm or the maximum target-to-image receptor distance, whichever is less.
  • 18. Filtration • The filters must be permanently and securely mounted to the x- ray port of the tube housing or beam limiting device, or both.
  • 19. Timer Accuracy • The timer or exposure control must be such that: (i) it can be set to control exposures as short as 1/60 second or 5 milliampere-seconds, whichever is greater, and (ii) at each setting it is accurate to 1/60 second or 7% of that setting, whichever is greater.
  • 20. Reproducibility • The coefficient of variation of any 10 consecutive radiation measurements, taken at the same source-to-detector distance within a time period of one hour and with the line voltage for each measurement within 1% of the mean value for all of the measurements, must not be greater than 0.05. • Furthermore, each of the 10 radiation exposures measured must be within 15% of the mean value of the 10 measurements.
  • 21. Linearity • For any fixed value of x-ray tube potential (in kVp), within the range of values specified for the equipment, the average ratios of exposure (in milliroentgen or coulomb/kg) to tube current-exposure time product (in milliampere-seconds) at any two consecutive tube current settings must not differ by more than 0.10 times their sum.
  • 22. Line Voltage Compensator • The equipment must be equipped with means to compensate for variations in x-ray tube potential caused by line voltage fluctuations. • By these means it must be possible to set and maintain the peak tube potential (kVp), prior to initiation of an exposure, to within 5% of the selected value, for a line voltage variation of ± 7% of its nominal value.
  • 23. Current Stabilization • The equipment should be provided with a tube current stabilizer to maintain preset current to within ± 5% of the required value for exposures of 1/5 second or more. • Changes in tube potential should not result in a variation of tube current greater than ± 5%.
  • 25. Mobile Radiographic Equipment • The requirements given above should be, applied. • Additional requirements for mobile radiographic equipment are as follows:
  • 26. Target-To-Skin Distance • The equipment must be equipped with means to prevent operation at target-to-skin distances of less than 30 centimetres.
  • 27. Exposure Control • The exposure switch must be of the deadman type and must be provided with a cable at least 3 metres long.
  • 28. Capacitor Discharge Equipment Capacitor discharge type mobile equipment must be equipped with: – a safety shutter or other means to prevent emission of xradiation from the exit port of the x-ray tube housing assembly when the exposure switch or timer is not activated; – a warning light, or other visual indicator, to indicate that the high potential capacitor is charged; – an electrical meter, or other indicator, to indicate the state of charge of the high potential capacitor, and – means to enable the high potential capacitor to be discharged to a residual potential difference of less than 30 volts, without the production of x-radiation.
  • 30. Collimation • The x-ray tube housing must be equipped with an adjustable collimator and which must provide the same degree of shielding towards leakage radiation as is required of the tube housing. • The maximum size of x-ray field permitted by the collimator should not be larger than the useful area of the input phosphor on the image intensifier. The useful beam should be centred on the input phosphor and the beam margins should be tangential to the area being viewed.
  • 31. Direct Bean Absorbers • A primary protective barrier must be permanently incorporated into the equipment to intercept the entire cross section of the direct beam, for all possible target-to image receptor distances. The barrier must be so installed that its removal from the useful beam causes automatic termination of the exposure • The lead equivalent thickness of the primary protective barrier must be at least 2 mm for machines capable of operating up to 100 kVp. For each additional kilovolt of operating tube potential an additional 0.01 mm lead equivalent is required.
  • 32. Direct Bean Absorbers • A primary protective barrier must be permanently incorporated into the equipment to intercept the entire cross section of the direct beam, for all possible target-to image receptor distances. • The barrier must be so installed that its removal from the useful beam causes automatic termination of the exposure
  • 33. Image Intensification • All fluoroscopic x-ray machines must be equipped with an image intensification system. • The image intensification system must be adequately shielded so that neither the useful beam nor the scattered radiation from the intensifier will produce significant exposure to the operator or other personnel.
  • 34. Image Intensification • The image intensification system must be adequately shielded so that neither the useful beam nor the scattered radiation from the intensifier will produce significant exposure to the operator or other personnel. • The shielding must be such that the exposure rate to an object having a cross-sectional area of 100 cm2, with no linear dimension greater than 20 cm, of x-radiation due to transmission of the useful beam through or scattering from the window of the fluoroscopic imaging assembly, does not exceed 2 milliroentgen per hour (516 nC/kg/h) at 10 centimetres behind the plane of the image receptor, for each roentgen per minute of entrance exposure rate.
  • 35. Target-To-Skin Distance • Fixed fluoroscopic equipment must incorporate means to limit the target-to-skin distance to not less than 38 cm. • In the case of a fluoroscopic machine designed for special applications that would be impossible at the above minimum distance, provision may be made for operation at shorter target-to-skin distances; but in no case must this distance be less than 20 cm.
  • 36. Material Between Patient And Image Receptor • The aluminum equivalent of the table top must not be greater than 1 mm when measured at 100 kVp.
  • 37. Exposure Switch • The fluoroscopic exposure switch must be of a dead-man type.
  • 38. Timer • There must be a cumulative timing device, activated by the fluoroscopic exposure switch, which requires manual resetting. • This timer must have a maximum setting not exceeding 5 minutes. Whenever a preset time has been reached, it must give a clearly audible signal.
  • 39. Indication of Tube Potential and Current • Electrical meters or other visual indicators must be provided to enable continuous monitoring by the operator of tube potential and current during fluoroscopy.
  • 40. Entrance Exposure Rate • At the shortest target-to-skin distance specified for the equipment, the entrance exposure rate must not exceed 10 roentgen (2.58 mC/kg) per minute for any combination of tube current and potential. With modern equipment most fluoroscopy can and should be carried out with exposure rates of less than 5 roentgen (1.29 mC/kg) per minute.
  • 41. Bucky Slot Shielding • Shielding must be provided with the equipment to cover the Bucky Slot. This should provide the equivalent protection of at least 0.5 mm of lead at 100 kVp.
  • 42. Table Shielding • The side of the fluoroscopic table should have an equivalent thickness of at least 0.5 mm of lead at 100 kVp to restrict scattered radiation passing through it.
  • 43. Operator Shielding • To protect the operator from radiation scattered above the table-top, there should be a lead rubber apron of dimensions not less than 45.7 cm × 45.7 cm (18" × 18"). • The material must have a lead equivalent of not less than 0.5 mm at 100 kVp and the apron must be attached to the lower edge of the fluoroscopic screen when the latter is vertical and to the side when the screen is horizontal. • The apron may consist of several overlapping parts.
  • 44. Operator Shielding • In the absence of such a protective apron a retractable lead shield must be Incorporated in the equipment to project at least 15 cm above the table when in use. • It must be mounted on the front side of the table and have a lead-equivalent thickness of at least 0.5 mm at 100 kVp.
  • 46. Collimation • The equipment must be permanently fitted with a collimator which must confine the beam to the area of the fluorescent screen. • The beam must not be greater than 35 cm × 43 cm (14" × 17") at the screen surface, regardless of the target-to-skin distance.
  • 47. Beam Alignment • The photofluorographic camera and x-ray tube must be coupled together so that the direct beam always centres on the mid-point of the screen.
  • 48. Gonad Shield • The lower edge of the beam-defining device should be curved to prevent irradiation of the female gonads.
  • 49. Maximum Entrance Exposure • The entrance exposure to a patient must not exceed 200 milliroentgen and should not exceed 100 milliroentgen (25.8 microcoulomb/kg) per film.
  • 50. Protective Clothing 1. Protective body aprons 2. Gonad shields 3. Protective gloves
  • 51. Protective Body Aprons • Protective body aprons used for radiographic or fluoroscopic examinations with peak x-ray tube potentials of up to 150 kVp must provide attenuation equivalent to at least 0.5 mm of lead.
  • 52. Gonad Shields • Contact-type gonad shields used for routine diagnostic radiology must have a lead equivalent thickness of at least 0.25 mm and should have a lead equivalent thickness of 0.5 mm at 150 kVp. • Contact-type gonad shields must be of sufficient size and shape to exclude the gonads completely from primary beam irradiation.
  • 53. Protective Gloves • Protective gloves used in fluoroscopy must provide attenuation equivalent to at least 0.25 mm of lead at 150 kVp.
  • 54. Film Cassettes • The front panel of the film cassette must not exceed an aluminum equivalence thickness of 1.0 mm and should not exceed an aluminum equivalence thickness of 0.5 mm. • Film cassettes must be completely light-tight. As soon as the material used for achieving light-tightness becomes worn and light leakage occurs, either the material or the cassette should be replaced. • It is also important to have good film-screen contact; therefore screens should be cleaned regularly and their contact with the film checked.
  • 55. Procedures To Reduce Dose To X-ray Personnel
  • 56. Basic Principles of Dose Reduction •Time •Distance •Shielding
  • 57. Reduction of External Dose ❖Minimize the time spent near the radiation source ❖Maximize the distance away from the source ❖Make use of available shielding
  • 58. General Recommendations 1. Only One Investigation • An x-ray room must not be used for more than one radiological investigation simultaneously. 2. Only Essential Persons • Except for those persons whose presence is essential, no person must be in the x-ray room when the exposure is carried out.
  • 59. General Recommendations 3. Be Away from the useful beam – Personnel must at all times keep as far away from the useful beam as is practicable. Exposure of personnel to the useful beam must never be allowed unless the beam is adequately attenuated by the patient and by protective clothing or screens.
  • 60. General Recommendations 4. Use Protective Devices – All personnel must take full advantage of the protective devices available. 5. Stay Behind Shielding 4. Operators should remain inside the control booth or behind protective screens when making an x-ray exposure. In cases where there are reasons that make this impractical, protective clothing must be worn.
  • 61. General Recommendations 6. Rules for patient support – When there is a need to support children or weak patients, holding devices should be used. – If parents, escorts or other personnel are called to assist, they must be provided with protective aprons and gloves, and be positioned so as to avoid the useful beam. – No one person should regularly perform these duties.
  • 62. General Recommendations 7. Personal Dosimetry – When a lead equivalent protective apron is worn, the personnel dosimeter must be worn under the apron. – If extremities are likely to be exposed to significantly higher doses, additional extremity monitors should be worn.
  • 63. General Recommendations 8. Personal Dosimetry – All operators of x-ray equipment, together with personnel (e.g. nurses) who routinely participate in radiological procedures, and others likely to receive a radiation dose in excess of 1/20th of the Dose limit specified in Appendix I, must wear personnel dosimeters.
  • 64. General Recommendations 9. Close All Doors – All entrance doors to an x-ray room, including patient cubicle and preparation room doors, should be kept closed while a patient is in the room. 10.Attend The energized machine – X-ray machines which are energized and ready to produce radiation must not be left unattended.
  • 65. General Recommendations 11.In the case of high exposure – Where radiation doses in excess of 5% of the dose limit specified in Appendix I are being received regularly by any one person, appropriate remedial steps must be taken to improve techniques and protective measures.
  • 66. General Recommendations 12.Machine Operator – X-ray equipment must only be operated by, or under the direct supervision of, qualified individuals. 13.Don’t hold the x-ray housing – An x-ray housing must not be held by hand during operation.
  • 67. Recommendations For Operation Of Radiographic Units 1. Controlling the machines – The x-ray exposure should, as a general rule, be controlled from the control panel located inside the control booth or behind a shielded wall. In the case of special techniques where the operator is required to control the exposure while at the side of the patient, appropriate protective clothing must be worn.
  • 68. Recommendations For Operation Of Radiographic Units 2. Communication with patient – The operator must have a clear view of the patient during every exposure and be able to communicate with the patient and/or attendants without leaving the control booth. 3. Don’t hold the Cassettes 2. Cassettes must never be held by hand during an exposure.
  • 69. Recommendations For Operation Of Fluoroscopic Units 1. Protective Clothing – All persons, with the possible exception of the patient, required to be in the room during a fluoroscopic procedure should wear protective aprons. Lead shields or curtains mounted on the fluoroscopic unit must not be considered a sufficient substitute for the wearing of protective clothing.
  • 70. Recommendations For Operation Of Fluoroscopic Units 2. Protective Gauntlets – Protective gauntlets should be worn by the radiologist during palpation in every fluoroscopic examination. During fluoroscopy, palpation with the hand should be kept to a minimum.
  • 71. Recommendations For Operation Of Fluoroscopic Units 3. Protective Clothing – During fluoroscopy and spotfilm operation associated with fluoroscopic operation, where personnel are required to be at the side of the patient, appropriate protective clothing must be worn by these personnel.
  • 72. Recommendations For Operation Of Fluoroscopic Units 4. Shorten The Time – All fluoroscopic examinations should be carried out as rapidly as possible and with minimum dose-rates and x-ray field sizes. 5. Image Intensified Fluoroscopy – Direct-viewing fluoroscopy must not be carried out; image intensified fluoroscopy must be used.
  • 73. Recommendations For Operation Of Mobile Units 1. Don’t Use Mobile If You Can Use Stationary – Mobile units must be used only if the condition of the patient is such as to make it inadvisable for the examination to be carried out with a stationary unit in the main x-ray department.
  • 74. Recommendations For Operation Of Mobile Units 2. Don’t Irradiate Other Persons – During operation, the primary beam should be directed from occupied areas if at all possible, and every effort must be made to ensure that this beam does not irradiate any other persons in the vicinity of the patient.
  • 75. Recommendations For Operation Of Mobile Units 3. Keep The Distance – The operator must stand at least 3 metres from the x-ray tube and out of the direct beam. 4. Use The Shielding – The operator must be shielded when exposures are made.
  • 76. Recommendations For Operation Of Mobile Units 5. Discharge The Capacitor – In a capacitor discharge unit after an x-ray exposure has been made there is a residual charge left in the capacitors. This residual charge can give rise to a “dark current” and result in x-ray emission even though the exposure switch is not activated. Therefore, the residual charge must be fully discharged before the unit is left unattended.
  • 77. Recommendations For Operation Of Photofluorographic Units • Photofluorographic x-ray machines in general can give significantly higher patient doses than conventional diagnostic machines.
  • 78. Recommendations For Operation Of Photofluorographic Units 1. Only the operator and the patient should be in the room or vehicle in which the unit is installed when an exposure is made. 2. When a photofluorographic unit is mounted in a vehicle appropriate steps must be taken to ensure that persons awaiting examination and personnel are protected from exposure to x- radiation. Such persons should be kept well clear of the vehicle when exposures are being made.
  • 79. Recommendations For Special Radiological Procedures • Cystoscopic radiography • Hystero-salpingography • Needle Biopsies of the Lung • Transhepatic Needle Cholangiography • Pancreatography, and other similar procedures.
  • 80. Recommendations For Special Radiological Procedures • In these procedures the radiologist and other personnel in the vicinity of the patient can be subjected to appreciable scattered radiation from the patient when the x-ray beam is on. Therefore, the radiologist and other personnel should wear protective glasses and clothing and should remain as far away from the patient as practicable, unless adequate scatter shields can be incorporated into the x-ray equipment.
  • 81. Recommendations For Special Radiological Procedures • The protective devices (e.g., shielded panels, leaded drapes, extended collimator cones, etc.) provided with the x-ray equipment should be used whenever they do not interfere unduly with the diagnostic procedures. • The smallest x-ray field consistent with the procedure should be used.
  • 82. Recommendations For Special Radiological Procedures • Angiography. Angiography is potentially one of the sources of greatest exposure to personnel in diagnostic radiology, since it requires the presence of a considerable number of personnel close to the patient and involves fluoroscopy for extended periods of time and multiple radiographic exposures. For such procedures all personnel must be aware of the radiation hazards involved and make every effort to adhere to the recommendations below:
  • 83. Recommendations For Special Radiological Procedures 1. Use Fully protected Device – Full use must be made of the protective devices provided with x- ray equipment (e.g., shielded panels, leaded drapes, bucky slot covers, etc.). 2. Wear Protective Clothing – All personnel must wear protective clothing and personnel dosimeters. Protective glasses should also be worn.
  • 84. Recommendations For Special Radiological Procedures 3. Go Away or behind shield – All personnel who are not required to be immediately adjacent to the patient during the procedure must stand back as far as possible from the patient and, if at all possible, should stand behind a protective shield. 4. Use shields if possible – Where indicated and feasible, special shields in addition to the protective devices provided with the machine should be used.
  • 85. Procedures For Minimizing Dose To Patients
  • 86. Guidelines For The Prescription Of Diagnostic X-ray Examinations • The medical practitioner is in a unique position to reduce unnecessary radiation exposure to the patient by ensuring that all examinations are clinically justified. The practitioner can achieve this by adhering, as much as possible, to certain basic recommendations. These are as follows:
  • 87. Guidelines For The Prescription Of Diagnostic X-ray Examinations 1. The prescription of an x-ray examination of a patient should only be based on a clinical evaluation of the patient and should be for the purpose of obtaining diagnostic information. 2. Routine or screening examinations, such as for preemployment physical examinations, tuberculosis screening, mass mammographic screening, etc., in which there is no prior clinical evaluation of the patient, should not be prescribed.
  • 88. Guidelines For The Prescription Of Diagnostic X-ray Examinations 3. It should be determined whether there have been any previous x-ray examinations which would make further examination unnecessary, or allow for the ordering of an abbreviated examination. Relevant previous radiographs or reports should be examined along with a clinical evaluation of the patient. 4. When a patient is transferred from one physician or hospital to another any relevant radiographs or reports should accompany the patient and should be reviewed by the consulting physician.
  • 89. Guidelines For The Prescription Of Diagnostic X-ray Examinations 5. When prescribing a radiological examination, the physician should specify precisely the clinical indications and information required. 6. The number of radiographic views, required in an examination, should be kept to the minimum practicable, consistent with the clinical objectives of the examination. 7. In prescribing x-ray examinations of pregnant or possibly pregnant women, full consideration should be taken of the consequences of foetal exposure.
  • 90. Guidelines For The Prescription Of Diagnostic X-ray Examinations 8. If a radiograph contains the required information, repeat exposures should not be prescribed simply because the radiograph may not be of the “best” diagnostic quality. 9. Specialized studies should be undertaken only by, or in close collaboration with, a qualified radiologist. Medical practitioners should not operate x-ray equipment, or be responsible for the use of such equipment.
  • 91. Guidelines For The Prescription Of Diagnostic X-ray Examinations 10.A patient’s clinical records should include details of x-ray examinations carried out. 11.The prescription of an x-ray examination of a patient should only be based on a clinical evaluation of the patient and should be for the purpose of obtaining diagnostic information. 12.Routine or screening examinations, such as for preemployment physical examinations, tuberculosis screening, mass mammographic screening, etc., in which there is no prior clinical evaluation of the patient, should not be prescribed.
  • 92. Guidelines For The Prescription Of Diagnostic X-ray Examinations 13.It should be determined whether there have been any previous x-ray examinations which would make further examination unnecessary, or allow for the ordering of an abbreviated examination. Relevant previous radiographs or reports should be examined along with a clinical evaluation of the patient. 14.When a patient is transferred from one physician or hospital to another any relevant radiographs or reports should accompany the patient and should be reviewed by the consulting physician.
  • 93. Guidelines For The Prescription Of Diagnostic X-ray Examinations 15.When prescribing a radiological examination, the physician should specify precisely the clinical indications and information required. 16.The number of radiographic views, required in an examination, should be kept to the minimum practicable, consistent with the clinical objectives of the examination. 17.In prescribing x-ray examinations of pregnant or possibly pregnant women, full consideration should be taken of the consequences of foetal exposure.
  • 94. Guidelines For The Prescription Of Diagnostic X-ray Examinations 18.If a radiograph contains the required information, repeat exposures should not be prescribed simply because the radiograph may not be of the “best” diagnostic quality. 19.Specialized studies should be undertaken only by, or in close collaboration with, a qualified radiologist. Medical practitioners should not operate x-ray equipment, or be responsible for the use of such equipment. 20.A patient’s clinical records should include details of x-ray examinations carried out.
  • 95. Guidelines For Radiography Of Pregnant Women • Every effort should be made to avoid unnecessary irradiation of any woman known to be, or who might be pregnant. This is particularly important during the earliest stages of pregnancy when the potential for radiation damage of the rapidly dividing tissues is the greatest. Clearly, however, in spite of the possibility of radiation damage, if a radiological examination is required for the diagnosis or management of an urgent medical problem it must be done, irrespective of whether the patient may or may not be pregnant.
  • 96. Guidelines For Radiography Of Pregnant Women • The following recommendations apply to x-ray examinations involving pregnant or potentially pregnant women:
  • 97. Guidelines For Radiography Of Pregnant Women 1. Ten day Role – Radiography of the pelvic area in women of child-bearing age should be undertaken in the ten-day period following the onset of menstruation, since the risk of pregnancy is very small during this period.
  • 98. Guidelines For Radiography Of Pregnant Women 2. No Elective Abdominal nor Pelvic X-ray – Only essential investigations should be taken in the case of pregnant or suspected pregnant women. Elective radiography of the abdominal and pelvic area in pregnant women must be avoided. (“Elective” is taken to mean an examination of the abdomen and pelvis which does not contribute to the diagnosis or treatment of a women in relation to her immediate illness.)
  • 99. Guidelines For Radiography Of Pregnant Women 3. No Photofluorographic – Pregnant women or women who may be pregnant must not be accepted for chest photofluorographic (mass radiographic) examinations.
  • 100. Guidelines For Radiography Of Pregnant Women 4. Use protective Shields – When radiography of the pelvic area or abdomen is required, the exposure must be kept to the absolute minimum necessary and full use must be made of gonadal shielding and other protective shielding if the clinical objectives of the exmaination will not be compromised.
  • 101. Guidelines For Radiography Of Pregnant Women 5. Foetal dose is lower in prone Position – If a radiographic examination of the foetus is required the prone position should be used. This has the effect of shielding the foetus from the softer x-rays and hence reducing the foetal dose.
  • 102. Guidelines For Radiography Of Pregnant Women 6. Radiography should not be used for the determination of abnormal presentations of the foetus, or for placenta localizations. Other techniques such as ultrasonography are better suited for this purpose.
  • 103. Guidelines For Radiography Of Pregnant Women 7. Use Well-Collimated X-Ray Beam – Radiography of the chest, extremities, etc., of a pregnant woman, for valid clinical reasons, should only be carried out using a well- collimated x-ray beam and with proper regard for shielding of the abdominal area.
  • 104. Guidelines For Radiographic Examination Of The Breast (Mammography) The recommendations that follow are intended to provide guidance to the technologist, the medical physicist and radiologist in exercising their responsibility towards reduction of patient dose. 1. The mammographic X-ray system must be designed specifically for mammography and the image receptor must be compatible with the system. 2. A film-screen combination that provides good quality diagnostic results must be used. Direct exposure film must never be used in mammography.
  • 105. Guidelines For Radiographic Examination Of The Breast (Mammography) 3. Except in the case of mammography performed within a screening program, the technologist must not perform any examination which has not been prescribed by a physician responsible for the patient. 4. The dose to the patient must be kept to the lowest practicable value consistent with clinical objectives, and without loss of essential diagnostic information. To achieve this, techniques appropriate to the equipment available should be used and evaluated from time to time in terms of effectiveness.
  • 106. Guidelines For Radiographic Examination Of The Breast (Mammography) 5. Particular care in patient X-ray protection must be taken when mammographic X-ray examinations of pregnant or possibly pregnant women are carried out, even though the radiation dose to the abdomen and fetus are negligible during normal mammographic X-ray examinations. 6. There may be a need to use special procedures when performing mammographic X-ray examinations of patients with breast implants such as extra views, modified positioning, compression and loading techniques. Particular care of compression techniques must be taken since excessive compression of the implants during mammographic X-ray examination may cause rupture of the implant.
  • 107. Guidelines For Radiographic Examination Of The Breast (Mammography) 7. In mammography, it is recommended that the field normally be the full size of the image receptor, but not larger than the image receptor support, except at the chest wall. The amount of unexposed areas on the films should be minimal so to avoid the need for masking. Collimated-down views are useful in some purposes, but should be considered as a special procedure. 8. Appropriate compression must be used in all mammographic procedures. 9. Adequate X-ray beam filtration must be used in all mammographic procedures.
  • 108. Guidelines For Radiographic Examination Of The Breast (Mammography) 10.The technologist should use the maximum focal spot to skin distance consistent with good radiographic loading. For mammographic procedures, distances of less than 50 cm should not be used. For mammographic procedures including magnification, distances of less than 25 cm to the entrance surface of the breast shall not be used. 11.Full details of the mammographic procedures, including retakes, carried out should be noted on the patient’s clinical records.
  • 109. Guidelines For Radiographic Examination Of The Breast (Mammography) 12.Irradiation times should be minimized to avoid unnecessary dose increase from reciprocity law failure and to avoid motion artifacts. This can be accomplished by the use of sufficiently high tube current values. 13.Cassettes should be loaded at least 30 minutes in advance to allow air to escape, and thus improving film/screen contact. 14.The technologist should examine the images after processing in order to verify that the techniques being used are producing diagnostic quality images and that the X-ray equipment is functioning correctly.
  • 110. Guidelines For Radiographic Examination Of The Breast (Mammography) 15.An appropriate Quality Assurance program must be implemented on mammographic X-ray equipment and on film and Xerographic processing systems.
  • 111. Guidelines For Radiographic Examination Of The Breast (Mammography) 16.While recommended dose limits have been defined for radiation workers and the general population, there is no specific limits recommended for patients undergoing diagnostic X-rays procedures. However, it is possible to provide limits on the amount of radiation breast tissues receive by setting dose limits on the level of radiation a representative breast phantom would receives for a given irradiation. These dose limits are presented in the following table;
  • 112. Guidelines For Radiographic Examination Of The Breast (Mammography) Procedure Mean Glandular Dose Limits Film-Screen Mammography (no grid) 1.0 mGy Film-Screen Mammography (grid technique) 2.0 mGy Xeromammography 4.0 mGy Recommended dose limits per irradiation
  • 113. Radiological Chest Screening 1. The use of mass radiological screening should only be considered in areas of high incidence of tuberculosis or in special population groups who may be particularly susceptible to lung disease. Selection of population groups for x-ray screening should be based upon the probability of discovering a significant number of cases of cardiopulmonary disease and on the availability of full follow-up facilities for individuals requiring treatment. 2. Mass x-ray screening of the general population for tuberculosis and other chest diseases must not be carried out.
  • 114. Radiological Chest Screening 3. Routine hospital admission chest x-rays should only be taken in identified high risk groups. 4. Chest x-rays should not be required as a routine part of a pre- employment medical check-up or for admission to educational institutions. Since photofluorographic x-ray machines require a higher level of x-ray exposure than do convential radiographic machines to produce a film and since the yield of significant information is low, photofluorographic x-ray machines must not be used for chest examinations in place of conventional machines, and should not be used for mass screening when conventional machines are available.
  • 115. Guidelines For Carrying Out X-ray Examinations
  • 116. General Recommendations 1. The operator must not perform any examination which has not been prescribed by a physician responsible for the patient. 2. The exposure of the patient must be kept to the lowest practicable value, consistent with clinical objectives and without loss of essential diagnostic information. To achieve this, techniques appropriate to the equipment available should be used. 3. Particular care, consistent with the recommendations of Section 13.2, must be taken when radiological examinations of pregnant or potentially pregnant women are carried out.
  • 117. General Recommendations 4. The x-ray beam must be well-collimated to restrict it as much as is practicable to the area of diagnostic interest. 5. The x-ray beam size must be limited to the size of the image receptor or smaller. 6. The x-ray beam should not be directed towards the gonads unless it is absolutely essential, in which case gonad shielding must be used whenever the value of the examination is not impaired by such use. Guidelines on the use of gonad shielding are given in section 14 of this guideline.
  • 118. General Recommendations 7. Shielding should be used where appropriate and practicable to limit the exposure of body tissues. It is particularly important that special effort to be made to protect the blood-forming organs, gonads and thyroids of children. 8. The target-to-skin distance should be as great as possible, consistent with good radiographic technique. 9. For very young children, special devices should be employed to restrict movement. 10.Full details of the radiological procedures carried out should be noted on the patient’s clinical records.
  • 119. Recommendations For Radiographic Procedures 1. The edges of the x-ray beam should be seen on all x-ray films to ensure that no more than the desired area has been irradiated. The film size used should be as small as possible, consistent with the diagnostic objectives of the examination. 2. Screen-type film should not be used for non-screen techniques because it is less sensitive to direct x-radiation than non-screen film.
  • 120. Recommendations For Radiographic Procedures 3. The fastest film or intensifying screen-film combination, consistent with diagnostically-acceptable results, should be used. When highest definition is not required a high- speed film-screen combination should be used.
  • 121. Recommendations For Radiographic Procedures 4. X-ray intensifying screens made from rare earth phosphors should be used where appropriate. Exposure reductions in excess of 50%, compared to conventional calcium tungstate systems, are possible using the rare earth-type intersifying screens and the films developed for use with such screens. These reductions result from greater x-ray absorption and increased conversion efficiencies of the new phosphors — the rare earth systems have a conversion efficiency of x-rays to light of 15-20%, compared to approximately 5% for calcium tungstate systems.
  • 122. Recommendations For Radiographic Procedures 5. To ensure that patient exposure is kept to a minimum, consistent with image quality, full advantage should be taken of a combination of techniques, such as: a. use of an antiscatter grid or air gap between the patient and the image receptor; b. use of the optimum focus-to-film distance appropriate to the examination; c. use of the highest kilovoltage which produces films of good quality; d. use of automatic exposure control devices designed to keep all exposures and repeat exposures to a minimum.
  • 123. Recommendations For Radiographic Procedures 6. The radiographer should see the films after processing in order to verify that the techniques being used are producing diagnostic quality films and that the x-ray equipment is functioning correctly. 7. To avoid the necessity of retakes, it is particularly important before taking a long series of films that a single preliminary film of the series should be taken and processed to verify correctness of settings.
  • 124. Recommendations For Radiographic Procedures 8. While dose limits have been defined for radiation workers and the general population, no specific limits have been recommended, to date, for patients undergoing diagnostic x-ray procedures. For patients the risk involved in the exposure must always be weighed against the medical requirement for accurate diagnosis. However, it is now feasible to recommend upper limits on patient skin entrance exposure, for routine, non-specialty, radiographic examinations. For a reference patient, having the anthropometrical characteristics shown in Table 1, the skin entrance exposure (without backscatter) should not exceed the values indicated in Table 2.
  • 125. Recommendations For Radiographic Procedures Body Part Thickness (cm) Head (lateral) Neck (A/P) Chest (P/A) Abdomen (A/P) Foot (D/P) 15 13 23 23 8 Table 1 Anthropometrical Characteristics of the Reference Patient
  • 126. Recommendations For Radiographic Procedures Examination (Projection) ESE (mR) Chest (P/A) Skull (Lateral) Abdomen (A/P) Cervical Spine (A/P) Thoracic Spine (A/P) Full Spine (A/P) Lumbo-Sacral Spine (A/P) Retrograde Pyelogram (A/P) Feet (D/P) 20 170 450 120 400 250 500 500 200 Table 2 Recommended Upper Limits on Skin Entrance Exposure
  • 127. Recommendations For Fluoroscopic Procedures 1. In view of the relatively high exposure that results from fluoroscopy, such procedures should only be carried out when an equivalent result cannot be obtained from radiography. Fluoroscopy must not be used as a substitute for radiography. 2. Flurorscopy must only be carried out by, or under immediate supervision of, a radiologist or physician properly trained in fluoroscopic procedures. 3. All fluoroscopic procedures should be carried out as rapidly as possible with the smallest practical x-ray field sizes.
  • 128. Recommendations For Fluoroscopic Procedures 4. The exposure rate used in fluoroscopy should be as low as possible and must not exceed 10 roentgens (2.58 mC/kg) per minute at the position where the central axis of the x-ray beam enters the patient. With modern equipment, most fluoroscopic procedures can be readily carried out with exposure rates of less than 5 roentgens (1.29 mC/kg) per minute.
  • 129. Recommendations For Fluoroscopic Procedures 5. Image intensification must be used in order to reduce patient exposure. Image intensifiers can significantly reduce both exposure rate and exposure time. However, the operator must monitor the x-ray tube current and voltage on equipment with automatic brightness control, since both can rise to high values without the knowledge of the operator, particularly if the gain of the intensifier is decreased. 6. Television monitoring should be used in conjunction with the image intensifier.
  • 130. Recommendations For Fluoroscopic Procedures 7. Mobile fluoroscopic equipment should only be used for examinations where it is impractical to transfer patients to a permanent fluoroscopic installation. 8. Cinefluorography produces the highest patient doses in diagnostic radiology because the x-ray tube currents and potentials used are generally higher than those used in fluoroscopy. Therefore, this technique should not be used unless significant medical benefit is expected.
  • 131. Recommendations For Photofluorographic Procedures 1. The x-ray beam must be restricted to the area of the fluorescent screen and should be limited to the minimum size consistent with clinical requirements. 2. Appropriate steps must be taken to exclude or shield the gonads from the direct beam. 3. The performance of the equipment must be monitored routinely. Special emphasis should be placed on the speed of the optical systems used. These should be as fast as possible and should be replaced when noticeable deterioration in performance occurs.
  • 132. Recommendations For Special Procedures 1. Significant exposure to the patient’s eyes can result during neurological examinations, such as carotid angiography. In projections where it does not interfere with the diagnostic information sought, eye shielding should be used. 2. During cardiac catheterization and angiography significant exposure of the patient’s thyroid gland can occur. Therefore, appropriate shielding should be used whenever possible
  • 133. Guidelines For Reduction Of Gonad Dose
  • 134. Recommendations For Reducing Gonad Dose To The Patient Radiologists and radiographers must pay special attention to four factors that are important for reducing gonad dose to the patient: 1. Correct collimation of the x-ray beam. — It is not sufficient merely to limit the beam to the size of the image receptor. Care should be taken to further restrict the beam to the region of the patient’s body that is of diagnostic interest. Irradiation of any part of the body outside that region contributes nothing to the objective of the examination and only increases dose both to the body and the gonads.
  • 135. Recommendations For Reducing Gonad Dose To The Patient 2. Gonad shielding. — Appropriate use of specific area gonad shielding is strongly advised when: (i) the gonads, of necessity, lie within, or are in close proximity to, the primary x-ray beam; (ii) the patient has reasonable reproductive potential; and (iii) clinical objectives will not be compromised.
  • 136. Recommendations For Reducing Gonad Dose To The Patient 3. Appropriate selection of technique factors. — An appropriate selection of tube voltage, current and filtration is particularly important for diagnostic procedures in which the gonads lie within or near the primary x-ray beam. For example, in fluoroscopy, use of higher tube voltage and filtration and lower tube current will almost always reduce the gonad dose.
  • 137. Recommendations For Reducing Gonad Dose To The Patient 4. Sensitivity of imaging systems. — The gonad dose is inversely proportional to the sensitivity of the imaging system. Thus, doubling the sensitivity halves the gonad dose; conversely, halving the sensitivity doubles the gonad dose. It is, therefore, very important to maintain the sensitivity of the imaging system at its optimum value and to be alert for any significant deterioration.
  • 138. Recommendations For Reducing Gonad Dose To Operators • Adherence to the safe use procedures specified in Section 11 of this Safety Code will ensure that gonad dose to radiologists and radiographers is kept to the minimum practicable.