In cell culture, dissolved oxygen (DO) signal noise
due to gas bubble interference is common in small
volume bioreactors which are typically used in
process development and scale-down labs. The
reason DO signal noise is most prevalent in these
settings is due to dynamics, high oxygen demands,
and vertical sensor mounting. This issue will con-
tinue to become more pronounced as processes
yield higher cell densities and higher specific
productivities, resulting in even greater oxygen de-
mands.
Dissolved oxygen, a measure of oxygen content in
the liquid phase, is a key process parameter (KPP)
because of its importance in aerobic metabolism.
DO signal noise is caused by the sensor measuring
oxygen content in gas bubbles when they collide
with the sensors membrane; the content of these
gas bubbles is not a KPP because this oxygen
content is not accessible to the cells. By having
this noise in your DO measurement it does not only
Gas bubbles are a common culprit for causing dissolved oxygen (DO) signal
noise. Gas bubbles, which have a different oxygen content than the media, inter-
act with the sensor's membrane resulting in DO signal noise. Traditional sensor-
cap designs actually amplify this issue because their surface finishes have favor-
able interactions with gas bubbles. METTLER TOLEDO has developed a
one-of-a-kind dissolved oxygen sensor-cap to combat this issue; the new Angled-
OptoCap has an increased hydrophilic surface finish which repels gas bubbles
and a new angled design which minimizes gas bubble exposure.
Eliminating Bubble Noise
In Your Dissolved Oxygen Signal
ApplicationNote
negatively impact process control but it also negatively
impacts experiments which support critical decisions
for investigations, process optimization, tech transfer,
trouble shooting, and quality by design. Building this
signal noise into your design space and set-points can
have unintended consequences during scale-up, and
building this signal noise into your design of experi-
ments adds unnecessary variability. METTLER TOLEDO
has designed a new DO sensor-cap for the InPro6860i
which reduces dissolved oxygen signal noise from
gas bubbles. Click on the link to view a video trial of
the Anti-Bubble OptoCap: Anti-Bubble OptoCap Video
Trial.
Dissolved Oxygen Signal Noise from Gas Bubbles
Although the purpose of a dissolved oxygen sensor is
to measure oxygen in the liquid phase, these sensors
indiscriminately measure oxygen independent of it
being in the gas or liquid phase. When the sensor
membrane is exposed to gas bubbles the oxygen con-
tent in these bubbles impact the DO reading. For oxy-
gen or air bubbles, this results in an increased signal
because the gas bubbles have a higher oxygen con-
tent than that of the process. For nitrogen or carbon
dioxide bubbles, which contain no oxygen, this results
in a reduced DO signal.
The signal noise in the oxygen measurement due to
gas bubbles is not an inaccuracy of the sensor but a
function of the sensor-cap design. There are two pri-
mary sensor-cap characteristics in traditional designs
which promote gas-bubble interference. The largest
contributor is the hydrophobicity of untreated electrop-
olished stainless steel which has favorable interactions
with gas-bubbles. A secondary contributing character-
istic to gas bubble interference is the angle of sensor-
cap. A flat tip serves as a counteracting force to rising
gas bubbles, this increases the exposure to gas-bub-
bles.
The Anti-Bubble OptoCap Designed to Reduce Signal
Noise from Gas Bubbles
The new Anti-Bubble OptoCap reduces DO signal noise
by repelling gas-bubbles and reducing gas bubbles
exposure. This is achieved by engineering two design
characteristics: a more hydrophilic surface treatment
and a 30° cap geometry.
1. More Hydrophillic Surface Treatment
By reducing the contact angle of the sur-
face finish with a proprietary mechanical
surface treatment we were able to design
a more hydrophilic cap. This reduces the
favorable interactions between the cap
and the gas phase resulting in less bub-
ble accumulation and interference. The
new surface finish maintains the N5/Ra
0.4μm surface roughness with the same
stainless steel chemical composition. The
appearance is greyish but it still fulfills
ISO requirements and is suitable for
hygienic applications.
2. 30° Cap Geometry
An angled sensor reduces the probability
that gas bubbles will collide and interfere
with the sensors membrane. The 30°
angle is optimal for reducing gas bubble
interference without impacting the ability
of the sensors to measure dissolved oxy-
gen.
Mettler-Toledo Process Analytics
900 Middlesex Turnpike
Building 8
Billerica, MA 01821
(781) 301-8800
Subject to technical changes
www.mt.com
For more information
OptoCap Characteristics Standard MT OptoCap
(OptoCap BT02T)
New Anti-Bubble OptoCap
(OptoCap BT02THD)
Appearance
Angle Flat 180° Angled 30°
Surface Finish Electropolished Electropolished with proprietary
mechanical surface treated
Surface Roughness N5/Ra 0.4µm N5/Ra 0.4µm
Material Stainless Steel Stainless Steel
Membrane PTFE (FDA USP Class VI) PTFE (FDA USP Class VI)
O-Ring EPDM EPDM
Length (cm) 4.2 4.7
SIP/CIP/Autoclave Compatible? Yes Yes
Meet ISO Standards? Yes Yes
Suitable for Hygienic Applications? Yes Yes
Technical Comparison: Traditional OptoCap vs. Anti-Bubble OptoCap
Performance Comparison:
To validate the improvement of the new Anti-Bubble Opto-
Cap, trials were conducted with traditional sensor-caps
from METTLER TOLEDO and Hamilton, and with the Anti-
Bubble OptoCap. These trials were performed in a 10L
bioreactor with vertical sensor mounting (refer to the
video to see the experimental set-up). As shown in the
first graph ("Traditional Sensor-Caps"), traditional sensor-
caps show a significant amount of signal noise when try-
ing to maintain a 30% DO setpoint. The Anti-Bubble
OptoCap, shown in the second graph, shows no bubble
noise when maintain a 30% DO setpoint under the same
conditions.
%DO
%DO
Video Links:
www.mt.com/inpro6860i
https://www.youtube.com/watch?v=RkU07phMOfY

More Related Content

Viewers also liked

Viewers also liked (9)

Laser based level transmitter for industrial process measurement
Laser based level transmitter for industrial process measurementLaser based level transmitter for industrial process measurement
Laser based level transmitter for industrial process measurement
 
Optimizing gas turbine performance with precise humidity measurements
Optimizing gas turbine performance with precise humidity measurementsOptimizing gas turbine performance with precise humidity measurements
Optimizing gas turbine performance with precise humidity measurements
 
PID Controllers from Precision Digital
PID Controllers from Precision DigitalPID Controllers from Precision Digital
PID Controllers from Precision Digital
 
Continuous On-Line Microbial Monitoring of Pharmaceutical Waters
Continuous On-Line Microbial Monitoring of Pharmaceutical WatersContinuous On-Line Microbial Monitoring of Pharmaceutical Waters
Continuous On-Line Microbial Monitoring of Pharmaceutical Waters
 
Intelligent Sensor Management From Ingold
Intelligent Sensor Management From IngoldIntelligent Sensor Management From Ingold
Intelligent Sensor Management From Ingold
 
Magnetostrictive Liquid Level Transmitter for Process Measurement and Control
Magnetostrictive Liquid Level Transmitter for Process Measurement and ControlMagnetostrictive Liquid Level Transmitter for Process Measurement and Control
Magnetostrictive Liquid Level Transmitter for Process Measurement and Control
 
Optimizing chemcial industry processes
Optimizing chemcial industry processesOptimizing chemcial industry processes
Optimizing chemcial industry processes
 
Gas phase oxygen measurement systems
Gas phase oxygen measurement systemsGas phase oxygen measurement systems
Gas phase oxygen measurement systems
 
Process Analytics Product Catalog
Process Analytics Product CatalogProcess Analytics Product Catalog
Process Analytics Product Catalog
 

More from Alliance Technical Sales, Inc.

More from Alliance Technical Sales, Inc. (20)

Jogler Magnetic Level Gauge, Magnetostrictive Level Transmitter, and Laser Le...
Jogler Magnetic Level Gauge, Magnetostrictive Level Transmitter, and Laser Le...Jogler Magnetic Level Gauge, Magnetostrictive Level Transmitter, and Laser Le...
Jogler Magnetic Level Gauge, Magnetostrictive Level Transmitter, and Laser Le...
 
Electric Heating and Freeze Protection Solutions
Electric Heating and Freeze Protection SolutionsElectric Heating and Freeze Protection Solutions
Electric Heating and Freeze Protection Solutions
 
Mettler Toledo Process Analytics Catalog 2018-2019
Mettler Toledo Process Analytics Catalog 2018-2019Mettler Toledo Process Analytics Catalog 2018-2019
Mettler Toledo Process Analytics Catalog 2018-2019
 
Jogler Industrial Level Measurement Products Catalog
Jogler Industrial Level Measurement Products CatalogJogler Industrial Level Measurement Products Catalog
Jogler Industrial Level Measurement Products Catalog
 
Pharmaceutical Waters Guide for Regulatory Compliance, Analysis and Real-Time...
Pharmaceutical Waters Guide for Regulatory Compliance, Analysis and Real-Time...Pharmaceutical Waters Guide for Regulatory Compliance, Analysis and Real-Time...
Pharmaceutical Waters Guide for Regulatory Compliance, Analysis and Real-Time...
 
Analytical Measurement Solutions for Optimization of Your Brewing Process
Analytical Measurement Solutions for Optimization of Your Brewing ProcessAnalytical Measurement Solutions for Optimization of Your Brewing Process
Analytical Measurement Solutions for Optimization of Your Brewing Process
 
Pre-Assembled Self-Regulating Heating Cable
Pre-Assembled Self-Regulating Heating CablePre-Assembled Self-Regulating Heating Cable
Pre-Assembled Self-Regulating Heating Cable
 
Jogler Liquid Level Gauges, Indicators and Transmitters
Jogler Liquid Level Gauges, Indicators and TransmittersJogler Liquid Level Gauges, Indicators and Transmitters
Jogler Liquid Level Gauges, Indicators and Transmitters
 
Pharmaceutical Water Instrumentation Guide
Pharmaceutical Water Instrumentation GuidePharmaceutical Water Instrumentation Guide
Pharmaceutical Water Instrumentation Guide
 
Pharmaceutical Water Guide - Instrumentation
Pharmaceutical Water Guide - InstrumentationPharmaceutical Water Guide - Instrumentation
Pharmaceutical Water Guide - Instrumentation
 
Indigo 200 Series Transmitters
Indigo 200 Series TransmittersIndigo 200 Series Transmitters
Indigo 200 Series Transmitters
 
Laser Level Transmitter for Process Measurement and Control
Laser Level Transmitter for Process Measurement and ControlLaser Level Transmitter for Process Measurement and Control
Laser Level Transmitter for Process Measurement and Control
 
Portable pH Meter for Process Measurement
Portable pH Meter for Process MeasurementPortable pH Meter for Process Measurement
Portable pH Meter for Process Measurement
 
Dissolved Solids Analyzer Refractometer
Dissolved Solids Analyzer RefractometerDissolved Solids Analyzer Refractometer
Dissolved Solids Analyzer Refractometer
 
Temperature and Relative Humidity Data Logger
Temperature and Relative Humidity Data LoggerTemperature and Relative Humidity Data Logger
Temperature and Relative Humidity Data Logger
 
TOC Total Organic Carbon Analyzer
TOC Total Organic Carbon AnalyzerTOC Total Organic Carbon Analyzer
TOC Total Organic Carbon Analyzer
 
Surface Temperature Sensors and Transmitters for Industrial Applications
Surface Temperature Sensors and Transmitters for Industrial ApplicationsSurface Temperature Sensors and Transmitters for Industrial Applications
Surface Temperature Sensors and Transmitters for Industrial Applications
 
Guide to Oxygen Measurement in Industrial Process Applications
Guide to Oxygen Measurement in Industrial Process ApplicationsGuide to Oxygen Measurement in Industrial Process Applications
Guide to Oxygen Measurement in Industrial Process Applications
 
Turbidity Standards for Turbidity Measurement
Turbidity Standards for Turbidity MeasurementTurbidity Standards for Turbidity Measurement
Turbidity Standards for Turbidity Measurement
 
Automated pH Sensor Cleaning and Calibration
Automated pH Sensor Cleaning and CalibrationAutomated pH Sensor Cleaning and Calibration
Automated pH Sensor Cleaning and Calibration
 

Recently uploaded

Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills KuwaitKuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
jaanualu31
 
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
AldoGarca30
 
Call Girls in South Ex (delhi) call me [🔝9953056974🔝] escort service 24X7
Call Girls in South Ex (delhi) call me [🔝9953056974🔝] escort service 24X7Call Girls in South Ex (delhi) call me [🔝9953056974🔝] escort service 24X7
Call Girls in South Ex (delhi) call me [🔝9953056974🔝] escort service 24X7
9953056974 Low Rate Call Girls In Saket, Delhi NCR
 
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak HamilCara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Kandungan 087776558899
 
"Lesotho Leaps Forward: A Chronicle of Transformative Developments"
"Lesotho Leaps Forward: A Chronicle of Transformative Developments""Lesotho Leaps Forward: A Chronicle of Transformative Developments"
"Lesotho Leaps Forward: A Chronicle of Transformative Developments"
mphochane1998
 
Verification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptxVerification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptx
chumtiyababu
 

Recently uploaded (20)

Orlando’s Arnold Palmer Hospital Layout Strategy-1.pptx
Orlando’s Arnold Palmer Hospital Layout Strategy-1.pptxOrlando’s Arnold Palmer Hospital Layout Strategy-1.pptx
Orlando’s Arnold Palmer Hospital Layout Strategy-1.pptx
 
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills KuwaitKuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
 
Engineering Drawing focus on projection of planes
Engineering Drawing focus on projection of planesEngineering Drawing focus on projection of planes
Engineering Drawing focus on projection of planes
 
Design For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startDesign For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the start
 
Tamil Call Girls Bhayandar WhatsApp +91-9930687706, Best Service
Tamil Call Girls Bhayandar WhatsApp +91-9930687706, Best ServiceTamil Call Girls Bhayandar WhatsApp +91-9930687706, Best Service
Tamil Call Girls Bhayandar WhatsApp +91-9930687706, Best Service
 
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
 
NO1 Top No1 Amil Baba In Azad Kashmir, Kashmir Black Magic Specialist Expert ...
NO1 Top No1 Amil Baba In Azad Kashmir, Kashmir Black Magic Specialist Expert ...NO1 Top No1 Amil Baba In Azad Kashmir, Kashmir Black Magic Specialist Expert ...
NO1 Top No1 Amil Baba In Azad Kashmir, Kashmir Black Magic Specialist Expert ...
 
PE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and propertiesPE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and properties
 
Call Girls in South Ex (delhi) call me [🔝9953056974🔝] escort service 24X7
Call Girls in South Ex (delhi) call me [🔝9953056974🔝] escort service 24X7Call Girls in South Ex (delhi) call me [🔝9953056974🔝] escort service 24X7
Call Girls in South Ex (delhi) call me [🔝9953056974🔝] escort service 24X7
 
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak HamilCara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
 
COST-EFFETIVE and Energy Efficient BUILDINGS ptx
COST-EFFETIVE  and Energy Efficient BUILDINGS ptxCOST-EFFETIVE  and Energy Efficient BUILDINGS ptx
COST-EFFETIVE and Energy Efficient BUILDINGS ptx
 
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptxS1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
 
"Lesotho Leaps Forward: A Chronicle of Transformative Developments"
"Lesotho Leaps Forward: A Chronicle of Transformative Developments""Lesotho Leaps Forward: A Chronicle of Transformative Developments"
"Lesotho Leaps Forward: A Chronicle of Transformative Developments"
 
DC MACHINE-Motoring and generation, Armature circuit equation
DC MACHINE-Motoring and generation, Armature circuit equationDC MACHINE-Motoring and generation, Armature circuit equation
DC MACHINE-Motoring and generation, Armature circuit equation
 
kiln thermal load.pptx kiln tgermal load
kiln thermal load.pptx kiln tgermal loadkiln thermal load.pptx kiln tgermal load
kiln thermal load.pptx kiln tgermal load
 
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
 
Generative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPTGenerative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPT
 
Verification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptxVerification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptx
 
Online electricity billing project report..pdf
Online electricity billing project report..pdfOnline electricity billing project report..pdf
Online electricity billing project report..pdf
 
Work-Permit-Receiver-in-Saudi-Aramco.pptx
Work-Permit-Receiver-in-Saudi-Aramco.pptxWork-Permit-Receiver-in-Saudi-Aramco.pptx
Work-Permit-Receiver-in-Saudi-Aramco.pptx
 

Eliminating Bubble Noise in Dissolved Oxygen Measurements

  • 1. In cell culture, dissolved oxygen (DO) signal noise due to gas bubble interference is common in small volume bioreactors which are typically used in process development and scale-down labs. The reason DO signal noise is most prevalent in these settings is due to dynamics, high oxygen demands, and vertical sensor mounting. This issue will con- tinue to become more pronounced as processes yield higher cell densities and higher specific productivities, resulting in even greater oxygen de- mands. Dissolved oxygen, a measure of oxygen content in the liquid phase, is a key process parameter (KPP) because of its importance in aerobic metabolism. DO signal noise is caused by the sensor measuring oxygen content in gas bubbles when they collide with the sensors membrane; the content of these gas bubbles is not a KPP because this oxygen content is not accessible to the cells. By having this noise in your DO measurement it does not only Gas bubbles are a common culprit for causing dissolved oxygen (DO) signal noise. Gas bubbles, which have a different oxygen content than the media, inter- act with the sensor's membrane resulting in DO signal noise. Traditional sensor- cap designs actually amplify this issue because their surface finishes have favor- able interactions with gas bubbles. METTLER TOLEDO has developed a one-of-a-kind dissolved oxygen sensor-cap to combat this issue; the new Angled- OptoCap has an increased hydrophilic surface finish which repels gas bubbles and a new angled design which minimizes gas bubble exposure. Eliminating Bubble Noise In Your Dissolved Oxygen Signal ApplicationNote
  • 2. negatively impact process control but it also negatively impacts experiments which support critical decisions for investigations, process optimization, tech transfer, trouble shooting, and quality by design. Building this signal noise into your design space and set-points can have unintended consequences during scale-up, and building this signal noise into your design of experi- ments adds unnecessary variability. METTLER TOLEDO has designed a new DO sensor-cap for the InPro6860i which reduces dissolved oxygen signal noise from gas bubbles. Click on the link to view a video trial of the Anti-Bubble OptoCap: Anti-Bubble OptoCap Video Trial. Dissolved Oxygen Signal Noise from Gas Bubbles Although the purpose of a dissolved oxygen sensor is to measure oxygen in the liquid phase, these sensors indiscriminately measure oxygen independent of it being in the gas or liquid phase. When the sensor membrane is exposed to gas bubbles the oxygen con- tent in these bubbles impact the DO reading. For oxy- gen or air bubbles, this results in an increased signal because the gas bubbles have a higher oxygen con- tent than that of the process. For nitrogen or carbon dioxide bubbles, which contain no oxygen, this results in a reduced DO signal. The signal noise in the oxygen measurement due to gas bubbles is not an inaccuracy of the sensor but a function of the sensor-cap design. There are two pri- mary sensor-cap characteristics in traditional designs which promote gas-bubble interference. The largest contributor is the hydrophobicity of untreated electrop- olished stainless steel which has favorable interactions with gas-bubbles. A secondary contributing character- istic to gas bubble interference is the angle of sensor- cap. A flat tip serves as a counteracting force to rising gas bubbles, this increases the exposure to gas-bub- bles. The Anti-Bubble OptoCap Designed to Reduce Signal Noise from Gas Bubbles The new Anti-Bubble OptoCap reduces DO signal noise by repelling gas-bubbles and reducing gas bubbles exposure. This is achieved by engineering two design characteristics: a more hydrophilic surface treatment and a 30° cap geometry. 1. More Hydrophillic Surface Treatment By reducing the contact angle of the sur- face finish with a proprietary mechanical surface treatment we were able to design a more hydrophilic cap. This reduces the favorable interactions between the cap and the gas phase resulting in less bub- ble accumulation and interference. The new surface finish maintains the N5/Ra 0.4μm surface roughness with the same stainless steel chemical composition. The appearance is greyish but it still fulfills ISO requirements and is suitable for hygienic applications. 2. 30° Cap Geometry An angled sensor reduces the probability that gas bubbles will collide and interfere with the sensors membrane. The 30° angle is optimal for reducing gas bubble interference without impacting the ability of the sensors to measure dissolved oxy- gen.
  • 3. Mettler-Toledo Process Analytics 900 Middlesex Turnpike Building 8 Billerica, MA 01821 (781) 301-8800 Subject to technical changes www.mt.com For more information OptoCap Characteristics Standard MT OptoCap (OptoCap BT02T) New Anti-Bubble OptoCap (OptoCap BT02THD) Appearance Angle Flat 180° Angled 30° Surface Finish Electropolished Electropolished with proprietary mechanical surface treated Surface Roughness N5/Ra 0.4µm N5/Ra 0.4µm Material Stainless Steel Stainless Steel Membrane PTFE (FDA USP Class VI) PTFE (FDA USP Class VI) O-Ring EPDM EPDM Length (cm) 4.2 4.7 SIP/CIP/Autoclave Compatible? Yes Yes Meet ISO Standards? Yes Yes Suitable for Hygienic Applications? Yes Yes Technical Comparison: Traditional OptoCap vs. Anti-Bubble OptoCap Performance Comparison: To validate the improvement of the new Anti-Bubble Opto- Cap, trials were conducted with traditional sensor-caps from METTLER TOLEDO and Hamilton, and with the Anti- Bubble OptoCap. These trials were performed in a 10L bioreactor with vertical sensor mounting (refer to the video to see the experimental set-up). As shown in the first graph ("Traditional Sensor-Caps"), traditional sensor- caps show a significant amount of signal noise when try- ing to maintain a 30% DO setpoint. The Anti-Bubble OptoCap, shown in the second graph, shows no bubble noise when maintain a 30% DO setpoint under the same conditions. %DO %DO Video Links: www.mt.com/inpro6860i https://www.youtube.com/watch?v=RkU07phMOfY