SAFETY FEATURES IN
 ANAESTHETIC MACHINES


      -Dr.Prem kumar,
Department of anesthesiology,
  Madras medical college
Why safety features?
 To safeguard human life from unknown
  human errors.
 Safety devices are to prevent delivery of
  hypoxic mixture
 Regulation to prevent excessive pressure
  which is traumatic to patient.
Statistics




Misuse: 70% direct action primary anesthesia provider
         30% contributory actions of ancillary personnel
Classification
 ELECTRICAL   COMPONENTS



 PNEUMATIC   COMPONENTS.
ELECTRICAL
          COMPONENTS

 Master switch
 Power failure indicator
 Reserve power
 Automated check out
PNEUMATIC SYSTEM
   High pressure system

   Intermediate system

   Low pressure system

   Alternative oxygen control
High pressure system
   GAS CYLINDER –
             Color coding
             Pin index
             Cylinder labels
             Safety relief device
COLOR CODING

   Oxygen        black body,white shoulders[white
                   for international]
   Nitrous oxide blue
   Air            white and black
   Carbon dioxide gray
   Helium          brown
   Entonox        black with blue/white shoulders
PIN INDEX SAFETY SYSTEM
 Air             -     1,5
 Oxygen          -     2,5
 Nitrous oxide   -     3,5
 CO2>7.5%         -    1,6
 CO2<7.5%         -    2,6
 Cyclopropane     -     3,6
 N2O-O2            -      7
   The safety relief device is composed of atleast one
    of
          − Frangible disc [bursts under extreme
    pressure]

         - Fusible plug [wood’s metal which has a
    low melting Point ]

          - Safety relief valve [ opens at extreme
    pressure ]
 WOOD’S   METAL
       A fusible alloy that contains 50%
 bismuth, 25% lead, 12.5% tin, and 12.5%
 cadmium, and melts at 158°F (70-72°C);
 used for automatic sprinkler plugs.
   Cylinder pressure indicator [ bourdon’ s pressure
                                  gauge]

   Pressure regulator - PRIMARY

   Washer [ bodok seal ] – rubber made of neoprene

   Hanger yoke assembly – has pin index safety
    system. one way check valve for unidirectional
    gas flow
Bourdon’s pressure gauge
 Indication of incoming gas supply
 In O2 cylinder – indicates amount of gas
 N2O cylinder – pressure is not indicative of
  amount b’cos N2O is stored in liquified
  form.
 Curved tube can rupture with high pressure
  and the gauge has a vent on its back which
  can release the gas in the event of rupture.
Pressure regulator
Bodok seal
Intermediate system
   Pipeline –
              - Gas hoses are named and color coded
              - Non interchangeable quick
                couplers[shrader’s valve-gas specific]
              - NIST
              - Diameter index safety system[DISS]
              - Pipeline pressure indicators
NIST
- In line filter
             - Check valve
             - Gas power outlet [auxiliary oxygen
                  as a source of driving
                  gas for anesthesia ventilator or jet
                  ventilation]
   Oxygen pressure failure device
   Oxygen supply failure alarm [ richter alarm]
OPFD
   Pressure regulators – second stage

   Pressure relief valve

   Oxygen flush - O2+

   Flow control valve – color coding,fluted[O2], size,
    distance, placement of knobs,guard
PRESSURE REGULATOR
Pressure Relief Valve
   Pressure relief valves are installed
    downstream of all pressure regulators
   Valves set at no more than 50% above
    the pressure regulator setting.
   Function: Fully relieving the pressure at
    the set point in case failure of regulator.
   All pressure relief valves have piping
    connections to allow release of to outside
    facility.
Low pressure system
   Flowmeters –
        - Bobbin rotates on flow which prevents it
            from sticking.
        - Bobbin is antistatic to prevent
           sticking to the wall of the flowmeter
         - Fluorescent coating on the back of
            flowmeter panel and on the dot of bobbins
         - Float stop
         - Arrangement
         - Auxiliary oxygen flowmeter
An oxygen leak from the flow tube
 can produce a hypoxic mixture,
regardless of the arrangement of
         the flow tubes
Auxiliary oxygen flowmeter
Hypoxia prevention safety
             devices:
   Proportionating devices –
             - link 25 in datex
               ohmeda[mechanical,pneumatic
               and electronic linkage]
             - S-ORC in draeger,
             - Mandatory minimum oxygen
                 flow
   Vaporizers - Selectatec mechanism,
        interlocking,low filling port, indicator
        of agent,increased wick capacity,
        Fraser sweatman pin safety system
   Unidirectional [check ] valve
   Back pressure relief valve [opens when the
    pressure exceeds 200 cm H2O]
   Common gas outlet
Breathing system
   APL valve open when pressure exceeds 60 cm
    H2O
   Pressure in reservoir bag shouldn’t exceed 60 cm
    H2O.
   Oxygen analyser
   Canister – To prevent co2,carbon monoxide.
    Dessicated absorbents without KOH or ba(OH)2
    and with lesser amounts of NaOH produce less
    heat and no fires.
   VENTILATORS – PRESSURE SENSORS TO
    DETECT EXCESSIVE AIRWAY PRESSURE
    DUE TO VENTILATOR MALFUNCTION

   Alternative oxygen control

   Scavenging system

   Wheels in anesthetic workstation is made of
    antistatic rubber.
Safety features of newer
              machines
   Ø More accurate and/or corrected tidal
    volume through compliance and fresh gas
    compensation
 
   Ø Potential return of sampled gas[SGR]
    to facilitate low-flow

   Ø Fresh gas decoupling prevent
    hyperinflation of the lung
   Electronic PEEP

   Electronic selection of ventilation
    parameters

   Reduced external connections
Thank you

More Related Content

What's hot

Invasive blood pressure_monitoring
Invasive blood pressure_monitoringInvasive blood pressure_monitoring
Invasive blood pressure_monitoring
Ubaidur Rahaman
 
Anaesthesia machine 2014 spmc
Anaesthesia machine 2014 spmcAnaesthesia machine 2014 spmc
Anaesthesia machine 2014 spmc
Deepak Choudhary
 
Non return valves 18 1-11
Non return valves 18 1-11Non return valves 18 1-11
Non return valves 18 1-11
narasimha reddy
 

What's hot (20)

Safety features in anesthesia machine
Safety features in anesthesia machineSafety features in anesthesia machine
Safety features in anesthesia machine
 
Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction
 
Ayre t piece
Ayre t pieceAyre t piece
Ayre t piece
 
Desflurane
DesfluraneDesflurane
Desflurane
 
Vaporizers
Vaporizers Vaporizers
Vaporizers
 
Safety feature of BOYLE'S anesthesia machine
Safety feature of BOYLE'S anesthesia machineSafety feature of BOYLE'S anesthesia machine
Safety feature of BOYLE'S anesthesia machine
 
LAYNGEAL MASK AIRWAY
LAYNGEAL MASK AIRWAYLAYNGEAL MASK AIRWAY
LAYNGEAL MASK AIRWAY
 
Capnography
CapnographyCapnography
Capnography
 
Anaesthesia Workstation checklist and safety features
Anaesthesia Workstation checklist and safety features Anaesthesia Workstation checklist and safety features
Anaesthesia Workstation checklist and safety features
 
Mapleson circuits
Mapleson circuitsMapleson circuits
Mapleson circuits
 
Invasive blood pressure_monitoring
Invasive blood pressure_monitoringInvasive blood pressure_monitoring
Invasive blood pressure_monitoring
 
Anaesthesia machine 2014 spmc
Anaesthesia machine 2014 spmcAnaesthesia machine 2014 spmc
Anaesthesia machine 2014 spmc
 
Capnography
Capnography Capnography
Capnography
 
Supraglottic airway device
Supraglottic airway deviceSupraglottic airway device
Supraglottic airway device
 
Medical gas cylinders
Medical gas cylindersMedical gas cylinders
Medical gas cylinders
 
Anaesthesia machine 1
Anaesthesia machine 1Anaesthesia machine 1
Anaesthesia machine 1
 
Non return valves 18 1-11
Non return valves 18 1-11Non return valves 18 1-11
Non return valves 18 1-11
 
Safety features of anaesthesia machine
Safety features of anaesthesia machineSafety features of anaesthesia machine
Safety features of anaesthesia machine
 
Airway local blocks
Airway local blocksAirway local blocks
Airway local blocks
 
INTERMEDIATE PRESSURE SYSTEM.pptx
INTERMEDIATE PRESSURE SYSTEM.pptxINTERMEDIATE PRESSURE SYSTEM.pptx
INTERMEDIATE PRESSURE SYSTEM.pptx
 

Similar to Safety features in anesthesia machines-madras medical college

Anaesthesia machine
Anaesthesia machineAnaesthesia machine
Anaesthesia machine
imran80
 
Anesthetic machine for student
Anesthetic machine for studentAnesthetic machine for student
Anesthetic machine for student
Doxy Phobia
 
Pneumatic1day
Pneumatic1dayPneumatic1day
Pneumatic1day
Legaspi66
 

Similar to Safety features in anesthesia machines-madras medical college (20)

anaesthesia machine-140810030802-pptx.pptx
anaesthesia machine-140810030802-pptx.pptxanaesthesia machine-140810030802-pptx.pptx
anaesthesia machine-140810030802-pptx.pptx
 
anesthetic machine for student new.ppt
anesthetic machine for student new.pptanesthetic machine for student new.ppt
anesthetic machine for student new.ppt
 
Anaesthesia machine
Anaesthesia machineAnaesthesia machine
Anaesthesia machine
 
Anaesthesia machine manoj
Anaesthesia machine manojAnaesthesia machine manoj
Anaesthesia machine manoj
 
Safety Anesthesia Work Station
Safety Anesthesia Work StationSafety Anesthesia Work Station
Safety Anesthesia Work Station
 
Anaesthesia workstation
Anaesthesia workstationAnaesthesia workstation
Anaesthesia workstation
 
Anesthetic machine for student
Anesthetic machine for studentAnesthetic machine for student
Anesthetic machine for student
 
Anesthesia Machine
Anesthesia MachineAnesthesia Machine
Anesthesia Machine
 
SAFETY FEATURES OF ANAESTHESIA MACHINE
SAFETY FEATURES OF ANAESTHESIA MACHINESAFETY FEATURES OF ANAESTHESIA MACHINE
SAFETY FEATURES OF ANAESTHESIA MACHINE
 
Machine boyles
Machine boylesMachine boyles
Machine boyles
 
computerized-control-universal-brake-test-bench.pptx
computerized-control-universal-brake-test-bench.pptxcomputerized-control-universal-brake-test-bench.pptx
computerized-control-universal-brake-test-bench.pptx
 
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptx
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptxANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptx
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptx
 
Oxygen MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT
Oxygen  MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECTOxygen  MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT
Oxygen MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT
 
Anesthesia machine basics_V1.0_EN讲 - Copy.pdf
Anesthesia machine basics_V1.0_EN讲 - Copy.pdfAnesthesia machine basics_V1.0_EN讲 - Copy.pdf
Anesthesia machine basics_V1.0_EN讲 - Copy.pdf
 
Pneumatic1day
Pneumatic1dayPneumatic1day
Pneumatic1day
 
Anesthetic machine for student
Anesthetic machine for studentAnesthetic machine for student
Anesthetic machine for student
 
Anesthesia machine
Anesthesia machineAnesthesia machine
Anesthesia machine
 
Anaesthesia machine
Anaesthesia machineAnaesthesia machine
Anaesthesia machine
 
gas mgmnt systems.pptx
gas mgmnt systems.pptxgas mgmnt systems.pptx
gas mgmnt systems.pptx
 
anaesthesia machine(1).pptx
anaesthesia machine(1).pptxanaesthesia machine(1).pptx
anaesthesia machine(1).pptx
 

Recently uploaded

Salient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functionsSalient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functions
KarakKing
 
The basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptxThe basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptx
heathfieldcps1
 
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
ZurliaSoop
 

Recently uploaded (20)

FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024
 
Mixin Classes in Odoo 17 How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17  How to Extend Models Using Mixin ClassesMixin Classes in Odoo 17  How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17 How to Extend Models Using Mixin Classes
 
Salient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functionsSalient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functions
 
Kodo Millet PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
Kodo Millet  PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...Kodo Millet  PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
Kodo Millet PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
 
ICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptx
 
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptxSKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
 
Google Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptxGoogle Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptx
 
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdfUGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
 
The basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptxThe basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptx
 
Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...
 
ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701
 
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
 
Unit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxUnit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptx
 
Single or Multiple melodic lines structure
Single or Multiple melodic lines structureSingle or Multiple melodic lines structure
Single or Multiple melodic lines structure
 
Fostering Friendships - Enhancing Social Bonds in the Classroom
Fostering Friendships - Enhancing Social Bonds  in the ClassroomFostering Friendships - Enhancing Social Bonds  in the Classroom
Fostering Friendships - Enhancing Social Bonds in the Classroom
 
Unit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxUnit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptx
 
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxBasic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
 
Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdf
 
How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17
 
Understanding Accommodations and Modifications
Understanding  Accommodations and ModificationsUnderstanding  Accommodations and Modifications
Understanding Accommodations and Modifications
 

Safety features in anesthesia machines-madras medical college

  • 1. SAFETY FEATURES IN ANAESTHETIC MACHINES -Dr.Prem kumar, Department of anesthesiology, Madras medical college
  • 2. Why safety features?  To safeguard human life from unknown human errors.  Safety devices are to prevent delivery of hypoxic mixture  Regulation to prevent excessive pressure which is traumatic to patient.
  • 3. Statistics Misuse: 70% direct action primary anesthesia provider 30% contributory actions of ancillary personnel
  • 4. Classification  ELECTRICAL COMPONENTS  PNEUMATIC COMPONENTS.
  • 5. ELECTRICAL COMPONENTS  Master switch  Power failure indicator  Reserve power  Automated check out
  • 6. PNEUMATIC SYSTEM  High pressure system  Intermediate system  Low pressure system  Alternative oxygen control
  • 7. High pressure system  GAS CYLINDER – Color coding Pin index Cylinder labels Safety relief device
  • 8.
  • 9. COLOR CODING  Oxygen black body,white shoulders[white for international]  Nitrous oxide blue  Air white and black  Carbon dioxide gray  Helium brown  Entonox black with blue/white shoulders
  • 10.
  • 11. PIN INDEX SAFETY SYSTEM  Air - 1,5  Oxygen - 2,5  Nitrous oxide - 3,5  CO2>7.5% - 1,6  CO2<7.5% - 2,6  Cyclopropane - 3,6  N2O-O2 - 7
  • 12.
  • 13.
  • 14. The safety relief device is composed of atleast one of − Frangible disc [bursts under extreme pressure] - Fusible plug [wood’s metal which has a low melting Point ] - Safety relief valve [ opens at extreme pressure ]
  • 15.  WOOD’S METAL A fusible alloy that contains 50% bismuth, 25% lead, 12.5% tin, and 12.5% cadmium, and melts at 158°F (70-72°C); used for automatic sprinkler plugs.
  • 16. Cylinder pressure indicator [ bourdon’ s pressure gauge]  Pressure regulator - PRIMARY  Washer [ bodok seal ] – rubber made of neoprene  Hanger yoke assembly – has pin index safety system. one way check valve for unidirectional gas flow
  • 17. Bourdon’s pressure gauge  Indication of incoming gas supply  In O2 cylinder – indicates amount of gas  N2O cylinder – pressure is not indicative of amount b’cos N2O is stored in liquified form.  Curved tube can rupture with high pressure and the gauge has a vent on its back which can release the gas in the event of rupture.
  • 18.
  • 21.
  • 22. Intermediate system  Pipeline – - Gas hoses are named and color coded - Non interchangeable quick couplers[shrader’s valve-gas specific] - NIST - Diameter index safety system[DISS] - Pipeline pressure indicators
  • 23.
  • 24. NIST
  • 25. - In line filter - Check valve - Gas power outlet [auxiliary oxygen as a source of driving gas for anesthesia ventilator or jet ventilation]  Oxygen pressure failure device  Oxygen supply failure alarm [ richter alarm]
  • 26. OPFD
  • 27. Pressure regulators – second stage  Pressure relief valve  Oxygen flush - O2+  Flow control valve – color coding,fluted[O2], size, distance, placement of knobs,guard
  • 29. Pressure Relief Valve  Pressure relief valves are installed downstream of all pressure regulators  Valves set at no more than 50% above the pressure regulator setting.  Function: Fully relieving the pressure at the set point in case failure of regulator.  All pressure relief valves have piping connections to allow release of to outside facility.
  • 30.
  • 31. Low pressure system  Flowmeters – - Bobbin rotates on flow which prevents it from sticking. - Bobbin is antistatic to prevent sticking to the wall of the flowmeter - Fluorescent coating on the back of flowmeter panel and on the dot of bobbins - Float stop - Arrangement - Auxiliary oxygen flowmeter
  • 32.
  • 33.
  • 34.
  • 35. An oxygen leak from the flow tube can produce a hypoxic mixture, regardless of the arrangement of the flow tubes
  • 37. Hypoxia prevention safety devices:  Proportionating devices – - link 25 in datex ohmeda[mechanical,pneumatic and electronic linkage] - S-ORC in draeger, - Mandatory minimum oxygen flow
  • 38.
  • 39. Vaporizers - Selectatec mechanism, interlocking,low filling port, indicator of agent,increased wick capacity, Fraser sweatman pin safety system  Unidirectional [check ] valve  Back pressure relief valve [opens when the pressure exceeds 200 cm H2O]  Common gas outlet
  • 40.
  • 41. Breathing system  APL valve open when pressure exceeds 60 cm H2O  Pressure in reservoir bag shouldn’t exceed 60 cm H2O.  Oxygen analyser  Canister – To prevent co2,carbon monoxide. Dessicated absorbents without KOH or ba(OH)2 and with lesser amounts of NaOH produce less heat and no fires.
  • 42. VENTILATORS – PRESSURE SENSORS TO DETECT EXCESSIVE AIRWAY PRESSURE DUE TO VENTILATOR MALFUNCTION  Alternative oxygen control  Scavenging system  Wheels in anesthetic workstation is made of antistatic rubber.
  • 43.
  • 44.
  • 45.
  • 46. Safety features of newer machines  Ø More accurate and/or corrected tidal volume through compliance and fresh gas compensation    Ø Potential return of sampled gas[SGR] to facilitate low-flow  Ø Fresh gas decoupling prevent hyperinflation of the lung
  • 47. Electronic PEEP  Electronic selection of ventilation parameters  Reduced external connections