Stiffness Characteristics of Joshi’s External Stabilization System under Axial Compression: a Finite Element Method Based Study Rajeev Kumar*, R C Gupta*, Sanjay Mishra** *Dept of Mechanical Engineering, Institute of Engineering & Technology, Lucknow, UP, India ** Scientist G, Department of Science & Technology (DST), New Delhi, India ABSTRACT A finite element model of fractured tibia with Joshi’s External Stabilizing System (JESS) mounted on it was developed using 3D beam elements in the ANSYS software. The model was loaded in axial compression and the average axial stiffness of the model was calculated. The analytical value of axial stiffness was compared with reported experimental value to validate the finite element model. The validated model was used to carry out parametric studies on the model to determine the axial properties of JESS. It was observed that axial stiffness of JESS increased by 58% when k-wire diameter was varied from 2 mm to 4 mm while keeping other geometric configurations of the device constant; however, the axial stiffness of the device does not show any significant improvement when the diameter of medio-lateral pins in diaphyseal hold were increased. The findings should help in understanding the axial properties of JESS so that it can be used judiciously in clinical applications. 
Keywords – JESS, external fixators, tibial fracture, axial stiffness, k-wires, finite element analysis 
I. INTRODUCTION 
External fixation is a method of stabilization of bone fractures in which a number of percutenous metal pins pass through the fractured bone segments with their ends connected to a rigid frame. Joshi External Stabilizing System (JESS) is an external bone stabilizing device used in the Indian subcontinent. It was designed and fabricated by Dr BB Joshi in late seventies primarily for the treatment of hand trauma [1]. However, due to its low cost and highly versatile nature it evolved with time to be used in the treatment of variety of musculoskeletal disorders. At present, JESS is prominently used in intra-articular distal radial fractures [2], management of idiopathic clubfoot [3], hand trauma and its sequels [4]. Recent applications have been reported about JESS being used in treatment of injuries of tibial plateau and tibial plafond [5]. Many experimental procedures have been employed to study the behaviour of external fixation devices. Yilmaz et al. [6] conducted an experimental study to determine the stiffness characteristics of standard and hybrid Ilizarov circular fixators. Stein et al. [7] performed a biomechanical study on hybrid ring tubular external fixator to measure and compare the mechanical properties of different hybrid fixators. Schrøder et al. [8] performed experimental investigations of four different configurations of the Hoffmann external fixation system to assess its mechanical properties. Experimental procedures are important in establishing basic characteristics of fixation devices. However, they have some inherent shortcomings. 
Such procedures are time consuming, costly and need a large number of experimental data. Also, a minor change in the device requires another set of experiments to collect new data. As a result analytical methods such as finite element method have gained popularity for evaluating the mechanical properties of fixation devices analytically. Many researchers have used finite element analysis to study the mechanical properties of external fixation devices. Rybicki [9] presented the role and approach of finite element analysis in orthopedic studies and the application of FEM in the analysis of stresses in intact bones, analysis of fixation devices and prosthetic devices with a review of works carried out by researchers in these areas. Chao and An [10] classified the commonly used external fixation devices according to their geometrical configurations. They proposed a two dimensional and three dimensional beam elements to develop the finite element model of external fixators. Koo et al. [11] developed a finite element model using three dimensional beam elements in ABAQUS software and analyzed it for various loading conditions. It was found that under constrained axial compression, the pin diameter is the most critical parameter that could affect the system stiffness followed by pin offset. It was also reported that once the pin diameter exceeded certain value, it could not improve the system stiffness significantly. 
Bartel et al. [12, 13] developed a model using symmetric and unsymmetrical beam theory and demonstrated that beam theory, if used within its limitations, could provide an excellent model for understanding the overall behaviour of bone-implant 
RESEARCH ARTICLE OPEN ACCESS
Rajeev Kumar et al Int. Journal of Engineering Research and Applications www.ijera.com 
ISSN : 2248-9622, Vol. 4, Issue 7( Version 5), July 2014, pp.43-47 
www.ijera.com 44 | P a g e 
systems. Prendergast et al. [14] conducted finite element analysis and mechanical testing of unilateral and bilateral external fixators. Watson et al. [15] developed a modular FE model of the components of Illizarov external fixation system to predict mechanical properties of any configuration of the device. The first reported study on characterization of JESS was carried out by Kumar et al. [16] who conducted an experimental and finite element based investigation to determine and compare the axial stiffness of JESS and proposed a validated FE model of JESS under axial compression. Present study aims to analyze of the effect of variation in k-wire diameter on overall axial stiffness of JESS while keeping all other geometrical parameters of the fixator constant using finite element method. It will also study the change in axial stiffness of device when the size of medio-lateral pins in diaphyseal hold is varied. 
II. MATERIALS AND METHODS 
2.1 Geometrical features of JESS A JESS frame used for treatment of proximal tibial fractures can be configured in two parts, a proximal hold or helmet and a diaphyseal hold. The proximal hold consists of two circular rods of 4 mm diameter curved into a three quarter circular rings mounted on the proximal tibia with help of three percutenous pins (k-wires) inserted at about 22.5º with each other. The second three quarter circular ring has a lesser diameter than the first ring and it is added to reinforce the strength of proximal hold. For a JESS used in treatment of metaphyseal tibial fracture, the k-wire diameters may vary from 2.0 mm to 4.0 mm depending upon the clinical requirements. Universal link joints are used to join k-wires to the outer and inner circular rings with proximal tibia. The diaphyseal hold is composed of three parallel pins inserted in the tibia diaphysis in medio- lateral plane (m-l pins) below the fracture fragment and are attached to two Z shaped connecting rods on the either side of tibia using universal link joints. The proximal hold and diaphyseal hold are in turn connected to each other with help of two anterior and two posterior connecting rods joining diaphysis hold to the outer ring of the proximal hold. In addition, one half pin is inserted from anterior in anterior- posterior plane to provide further fragment stability. Fig.1 shows the laboratory specimen of JESS configured on steel tubes simulated as tibial bone which was used for developing a three dimensional finite element model in our study. 2.2 Finite element analysis of JESS 
The aim of this study was to evaluate the effect of geometric variations on the overall axial stiffness of JESS using a validated three dimensional finite element model of JESS as proposed by Kumar et. al. [16]. Therefore, the finite element model of JESS was developed using the identical geometric specifications and material properties as prescribed in the study. 
Fig.1 JESS configured on steel tubes To create the FE model, the structural geometry of the JESS configured on steel tubes was created using ANSYS finite element software. A 15 mm gap was provided between the ends of the hollow steel tubes to represent the metaphyseal tibial fracture. The gap was maintained at 15 mm to ensure the complete load transfer through fixator rather than through the steel tubes. The wireframe model was discretized using 3D beam element (Beam 188). As the different components of the model had different sizes, appropriate section properties were allocated to each component. The discretized model of JESS consisted of 146 nodes and 164 elements. The elastic modulus and Poisson’s ratio for the parts of JESS, k-wires, m-l pins & idealized tibial bone segments were taken as 200 GPa & 0.28 respectively. The idealized pin-bone and pin fixator interfaces and various universal link joints of the JESS were modeled as rigid joints. To measure the axial compression of JESS, the distal end of the model was fixed by setting all degrees of freedom to zero. A vertical compression load was applied to the proximal end. Fig.2 shows the FE model of JESS with applied load and boundary conditions. The model was analyzed for static, linear analysis under axial compression.
Rajeev Kumar et al Int. Journal of Engineering Research and Applications www.ijera.com 
ISSN : 2248-9622, Vol. 4, Issue 7( Version 5), July 2014, pp.43-47 
www.ijera.com 45 | P a g e 
Fig.2 FE model with load and boundary conditions applied The axial stiffness and corresponding inter fragmentary displacements of the JESS frame during the axial loading were calculated from the FE model. Fig.3 shows the deformed FE model of JESS configured on steel tubes under axial compressive load. The validated FE model was used to carry out two parametric studies to assess the effect of geometric variations on the axial properties of JESS. 
Fig.3 Deformed model of JESS under axial compression 
In the first study, the diameter of the k-wires in proximal hold was varied from 2 mm to 4 mm in the step of 0.5 mm and the axial stiffness of the JESS frame was evaluated. All other geometrical parameters were kept identical including the m-l pin diameter. These steps were repeated for another value of m-l pins diameter and variation in axial stiffness was evaluated. This way total four set of data were collected for four different diameters of m-l pins. In the second study, for a selected diameter of k-wire, the diameters of m-l pins in the diaphysis hold were increased from 2.5 mm to 4 mm in a step of 0.5 mm and axial stiffness in each case was calculated. Then k-wire diameter was set to another value and m-l pins diameters were varied to evaluate another set of stiffness data for JESS. 
III. RESULTS 
The axial stiffness of JESS was calculated by finite element model under individual loading condition. The average axial stiffness obtained by FE analysis having k-wire size of 2 mm was compared with the reported experimental value [16] and was found to be comparable. Thus, the FE model can be used as valid model to simulate the axial mechanical properties of JESS. Effect of geometric variations were studied and it was observed that the average axial stiffness of JESS improved by about 58% on varying the diameter of the k-wires from 2 mm to 4 mm in proximal hold with m-l pin diameter kept at 4 mm and keeping all other geometrical parameters constant. Further, by changing the medio-lateral pins diameter in diaphyseal hold from 2.5 mm to 4 mm the average axial stiffness was increased by only about 6.2% while k-wire diameter was kept constant at 4 mm. 
IV. DISCUSSION 
Goodship and Kenwright [18] has reported that the mechanical properties of an external fixation device influence the biological environment at the bone fracture and controlled micro-movement of fracture could facilitate secondary healing in the fractured bone. Mechanical properties of fixation devices also affect the outcome of any fixation process. A very rigid fixator can delay healing; on the other hand an over flexible fixation device may lead to increase chances of pin-bone tract infection, mal- union and even non-union in some cases. Therefore, it is extremely important for a surgeon to have a good knowledge of comparative mechanical properties of the fixation device to use it in clinical applications. The axial stiffness of the JESS was calculated by developing a three dimensional finite element model using 3D beam elements in ANSYS software. It was observed that the average axial stiffness of the JESS increases nonlinearly with increase in the diameter of k-wires and the overall increase in axial stiffness by changing the diameter of k-wires from 2 mm to 4 mm was about 58%. The change in axial stiffness for each size of k-wire with m-l pins diameter maintained at 2.5 mm is shown in figure 4.
Rajeev Kumar et al Int. Journal of Engineering Research and Applications www.ijera.com 
ISSN : 2248-9622, Vol. 4, Issue 7( Version 5), July 2014, pp.43-47 
www.ijera.com 46 | P a g e 
Fig.4 Variation in axial stiffness of JESS with change in k-wire diameter (m-l pins diameter 2.5 mm) It was also observed that incremental gain in axial stiffness from diameter 2 mm to 4 mm drops as the k-wire diameter increases. The gain in axial stiffness was about 19% when the diameter of k-wire was changed from 2 mm to 2.5 mm however, the increase was only 5.5% when the diameter was changed from 3.5 mm to 4 mm. As 4 mm k-wire size is the maximum pin diameter used in a standard JESS under clinical application therefore, axial stiffness was not calculated for higher values of k-wire size beyond 4 mm. The overall improvement in axial stiffness was about 54% on variation of k-wire size when m-l pin diameter was fixed at 2.5 mm and it was about 58% when m-l pin diameter was fixed at 4 mm. Fig.5 shows the effect of variation in k-wire diameters on axial stiffness of JESS. 
Fig.5 Variation in axial stiffness of JESS for different set of m-l pin sizes 
In the second part, k-wire diameters were fixed and m-l pins diameter in diaphysis hold was varied from 2.5 mm to 4 mm in the step of 0.5 mm and for each case, axial stiffness of the device was calculated. These five sets of observations are plotted fig.6. It can be seen that by changing the m-l pins diameter from 2.5 mm to 4 mm while keeping the k- wire diameter constant increases the axial stiffness by a mere 3.5% to 6.2%. It can therefore be suggested that the size of medio-lateral pin does not play a significant role in determining the axial stiffness of the fixator. On the other hand, the k-wire size used in proximal hold plays a major role in controlling the overall axial stiffness of the JESS fixation device and can be manipulated to improve the stiffness of JESS in clinical applications. 
V. CONCLUSION 
Finite element model of JESS configured on steel tubes was developed with 3D beam element in ANSYS software and two parametric studies were conducted. Results show that the axial stiffness of the JESS can be improved by changing the diameter of k- wires in proximal hold. Thus, k-wire sizes play a significant role in determining the axial properties of JESS. 
Fig.6 Variation in axial stiffness for varying medio- lateral pin diameter The results of the study are based on a standard configuration of JESS device commonly used in stabilization of tibial fractures. However, the JESS is available in slightly different modules by different manufacturers in India. The exact configurations of JESS may vary in clinics depending upon surgeon’s own judgment and practice. Therefore, the above results may not be generalized for all types of the JESS configurations. Nevertheless, the present study suggests a way by which axial stiffness of JESS frame can be improved as per the clinical requirements. Conflict of interest statement All authors hereby declare that this study was not funded by any agency and there are no personal relationships with anyone who could influence the results of present study. 
REFERENCES 
[1] R. Prabhu, A legend forever-Dr Brij Bhusan Joshi, Indian Journal Orthopedics, 43(3), 2009, 312. 
[2] S.Thomas, C. John, T. P. Johnny, Intra- articular distal radial fractures – external
Rajeev Kumar et al Int. Journal of Engineering Research and Applications www.ijera.com 
ISSN : 2248-9622, Vol. 4, Issue 7( Version 5), July 2014, pp.43-47 
www.ijera.com 47 | P a g e 
fixation or conventional closed reduction, Journal of Orthopedics, 4(2), 2007, e39. 
[3] S. Suresh, A. Ahmed, V. K. Sharma, Role of Joshi's External Stabilisation System fixator in the management of idiopathic clubfoot, Journal of Orthopedic Surgery, 11(2), 2003, pp. 194–201. 
[4] B. B. Joshi, Joshi’s External Stabilization System (JESS): a simple mini external fixator for the management of hand trauma and its sequels, Injury, 28, 1997, 244-244. [5] B. B. Joshi, Ram Prabhu, N. A. Antao, R. Rohira, M. Vaishnav, L. Maratha, Trauma management by Joshi’s External Stabilization System (JESS), (JESS Research & Development Society, 2007). [6] E. Yilmaz, O. Bllhan, L. Karakurt, N. Arslan, E. Serin, Mechanical performance of hybrid Ilizarov external fixator in comparison with Ilizarov circular external fixator, Clinical Biomech., 18, 2003, 518- 522. [7] H. Stein, R. Mosheiff, F. baumgart, R. Frigg, S. M. Parren, J. Cordey, The hybrid ring tubular extrenal fixator: A biomechanical study, Clinical Biomech., 12 (4), 1997, 259-266. [8] H. A. Schrøder, R. E. Weeth, T. Madsen, Experimental analysis of Hoffmann external fixation in various mountings, Clinical Biomech., 16, 2001, 151-159. [9] Edmund F. Rybicki, The role of finite element models in orthopedics, in R.H. Gallaher, B. R. Simon, P. C. Johnson, J. F. Cross (Ed.), Finite Element in Biomechanics (John Wiley & Sons, USA, 1982) 181-194. [10] E. Y. S. Chao, K. N. An, Biomechanical analysis of external fixation devices for the treatment of open bone fractures, in R.H. Gallaher, B. R. Simon, P. C. Johnson, J. F. Cross (Ed.), Finite Elements in Biomechanics (John Wiley & Sons, USA, 1982) 195-222. [11] T. K. K. Koo, Y. H. Kim, D. B. Choi, K. G. Hua, J. Lin, N. Inove, E. Y. S. Chao, Stiffness analysis of Dynafix external fixator system, Summer Bioengineering Conference, Florida, USA, 2003, 1227-1228. [12] D. L. Bartel, D. T. Davy, T. M. Keaveny, Structural analysis of musculoskeletal system: Beam Theory, Orthopedic Biomechanics: Mechanics & Design in Musculoskeletal Systems, (New Jeresy: Pearson Prentice Hall, 2006) 168-199. 
[13] D. L. Bartel, D. T. Davy, T. M. Keaveny, Fracture fixation devices, Orthopedic Biomechanics: Mechanics & Design in Musculoskeletal Systems, (New Jeresy: Pearson Prentice Hall, 2006) 261-287. 
[14] P. J. Prendergast, S. J. Toland, J. P. Corrigan, Finite element analysis and mechanical testing of external fixator designs, Proc. Instn. Mech. Engrs., 208, 1994, 103-110. [15] M. Watson, K. J. Mathias, N. Maffulli, D. W. Hukins, D. E. T. Shepherd, Finite element modelling of the Illizarov external fixation system, Proc. IMechE, Part H, J Engg. in Medicine, 221(8), 2007, 863-871. [16] Rajeev Kumar, Anupam Gupta, V P Sharma, Sanjay Mishra, Strength of Joshi External Stabilizing System, Journal of Orthopedic Surgery, 19(1), 2011, 72-75. [17] Leonid N. Solomin, The basic principles of external fixation using Illizarov device, (Milan: Springer-Verlog Italia, 2008) 341- 349. [18] A. E. Goodship, J. Kenwright, The influence of induced micromovement upon the healing of experimental tibial fractures, The Journal of Bone & Joint Surgery, 67-B(4), 1885, 650-655.

More Related Content

What's hot

International Journal of Engineering Research and Development
International Journal of Engineering Research and DevelopmentInternational Journal of Engineering Research and Development
International Journal of Engineering Research and DevelopmentIJERD Editor
 
Elise Walker ASB 2012
Elise Walker ASB 2012Elise Walker ASB 2012
Elise Walker ASB 2012Elise Walker
 
Finite Element Investigation of Hybrid and Conventional Knee Implants
Finite Element Investigation of Hybrid and Conventional Knee ImplantsFinite Element Investigation of Hybrid and Conventional Knee Implants
Finite Element Investigation of Hybrid and Conventional Knee ImplantsCSCJournals
 
IJSRED-V2I3P44
IJSRED-V2I3P44IJSRED-V2I3P44
IJSRED-V2I3P44IJSRED
 
Dr jehad al sukhun gives modelling of orbital deformation
Dr jehad al sukhun gives modelling of orbital deformationDr jehad al sukhun gives modelling of orbital deformation
Dr jehad al sukhun gives modelling of orbital deformationjehadsukhun
 
01907-bonejointscience-suturefix
01907-bonejointscience-suturefix01907-bonejointscience-suturefix
01907-bonejointscience-suturefixBrett McKenzie
 
Biomechanical properties of Anterolateral ligament of the knee compared with...
Biomechanical properties of  Anterolateral ligament of the knee compared with...Biomechanical properties of  Anterolateral ligament of the knee compared with...
Biomechanical properties of Anterolateral ligament of the knee compared with...KHALIFA ELMAJRI
 
Fragility Fractures - Pathogenesis-Biomechanical View
Fragility Fractures - Pathogenesis-Biomechanical ViewFragility Fractures - Pathogenesis-Biomechanical View
Fragility Fractures - Pathogenesis-Biomechanical ViewARPUTHA SELVARAJ A
 
Study of Influence of Different Parameters of Head- Restraints on Energy Abso...
Study of Influence of Different Parameters of Head- Restraints on Energy Abso...Study of Influence of Different Parameters of Head- Restraints on Energy Abso...
Study of Influence of Different Parameters of Head- Restraints on Energy Abso...AM Publications
 
EMG Investigations Regarding Handle Size, Grip Force and Stroke Rotation In S...
EMG Investigations Regarding Handle Size, Grip Force and Stroke Rotation In S...EMG Investigations Regarding Handle Size, Grip Force and Stroke Rotation In S...
EMG Investigations Regarding Handle Size, Grip Force and Stroke Rotation In S...Waqas Tariq
 
Vibration analysis of a torpedo battery tray using fea
Vibration analysis of a torpedo battery tray using feaVibration analysis of a torpedo battery tray using fea
Vibration analysis of a torpedo battery tray using feaeSAT Publishing House
 
Upgradation of non ductile reinforced concrete beamcolumn connections using f...
Upgradation of non ductile reinforced concrete beamcolumn connections using f...Upgradation of non ductile reinforced concrete beamcolumn connections using f...
Upgradation of non ductile reinforced concrete beamcolumn connections using f...IAEME Publication
 
Comparison of regression models for estimation of isometric wrist joint torqu...
Comparison of regression models for estimation of isometric wrist joint torqu...Comparison of regression models for estimation of isometric wrist joint torqu...
Comparison of regression models for estimation of isometric wrist joint torqu...Amir Ziai
 
New study on friction in a mom total hip prosthesis with balls in self direct...
New study on friction in a mom total hip prosthesis with balls in self direct...New study on friction in a mom total hip prosthesis with balls in self direct...
New study on friction in a mom total hip prosthesis with balls in self direct...eSAT Publishing House
 
Effect of thread pattern upon osseointegration
Effect of thread pattern upon osseointegrationEffect of thread pattern upon osseointegration
Effect of thread pattern upon osseointegrationNavneet Randhawa
 
Towards A Safer Design of Helmets: Finite Element & Experimental Assessment
Towards A Safer Design of Helmets: Finite Element & Experimental AssessmentTowards A Safer Design of Helmets: Finite Element & Experimental Assessment
Towards A Safer Design of Helmets: Finite Element & Experimental AssessmentSari Kassar
 

What's hot (19)

International Journal of Engineering Research and Development
International Journal of Engineering Research and DevelopmentInternational Journal of Engineering Research and Development
International Journal of Engineering Research and Development
 
H1302034856
H1302034856H1302034856
H1302034856
 
Elise Walker ASB 2012
Elise Walker ASB 2012Elise Walker ASB 2012
Elise Walker ASB 2012
 
Finite Element Investigation of Hybrid and Conventional Knee Implants
Finite Element Investigation of Hybrid and Conventional Knee ImplantsFinite Element Investigation of Hybrid and Conventional Knee Implants
Finite Element Investigation of Hybrid and Conventional Knee Implants
 
IJSRED-V2I3P44
IJSRED-V2I3P44IJSRED-V2I3P44
IJSRED-V2I3P44
 
Dr jehad al sukhun gives modelling of orbital deformation
Dr jehad al sukhun gives modelling of orbital deformationDr jehad al sukhun gives modelling of orbital deformation
Dr jehad al sukhun gives modelling of orbital deformation
 
01907-bonejointscience-suturefix
01907-bonejointscience-suturefix01907-bonejointscience-suturefix
01907-bonejointscience-suturefix
 
Biomechanical properties of Anterolateral ligament of the knee compared with...
Biomechanical properties of  Anterolateral ligament of the knee compared with...Biomechanical properties of  Anterolateral ligament of the knee compared with...
Biomechanical properties of Anterolateral ligament of the knee compared with...
 
4
44
4
 
Fragility Fractures - Pathogenesis-Biomechanical View
Fragility Fractures - Pathogenesis-Biomechanical ViewFragility Fractures - Pathogenesis-Biomechanical View
Fragility Fractures - Pathogenesis-Biomechanical View
 
Study of Influence of Different Parameters of Head- Restraints on Energy Abso...
Study of Influence of Different Parameters of Head- Restraints on Energy Abso...Study of Influence of Different Parameters of Head- Restraints on Energy Abso...
Study of Influence of Different Parameters of Head- Restraints on Energy Abso...
 
Ijciet 06 10_005
Ijciet 06 10_005Ijciet 06 10_005
Ijciet 06 10_005
 
EMG Investigations Regarding Handle Size, Grip Force and Stroke Rotation In S...
EMG Investigations Regarding Handle Size, Grip Force and Stroke Rotation In S...EMG Investigations Regarding Handle Size, Grip Force and Stroke Rotation In S...
EMG Investigations Regarding Handle Size, Grip Force and Stroke Rotation In S...
 
Vibration analysis of a torpedo battery tray using fea
Vibration analysis of a torpedo battery tray using feaVibration analysis of a torpedo battery tray using fea
Vibration analysis of a torpedo battery tray using fea
 
Upgradation of non ductile reinforced concrete beamcolumn connections using f...
Upgradation of non ductile reinforced concrete beamcolumn connections using f...Upgradation of non ductile reinforced concrete beamcolumn connections using f...
Upgradation of non ductile reinforced concrete beamcolumn connections using f...
 
Comparison of regression models for estimation of isometric wrist joint torqu...
Comparison of regression models for estimation of isometric wrist joint torqu...Comparison of regression models for estimation of isometric wrist joint torqu...
Comparison of regression models for estimation of isometric wrist joint torqu...
 
New study on friction in a mom total hip prosthesis with balls in self direct...
New study on friction in a mom total hip prosthesis with balls in self direct...New study on friction in a mom total hip prosthesis with balls in self direct...
New study on friction in a mom total hip prosthesis with balls in self direct...
 
Effect of thread pattern upon osseointegration
Effect of thread pattern upon osseointegrationEffect of thread pattern upon osseointegration
Effect of thread pattern upon osseointegration
 
Towards A Safer Design of Helmets: Finite Element & Experimental Assessment
Towards A Safer Design of Helmets: Finite Element & Experimental AssessmentTowards A Safer Design of Helmets: Finite Element & Experimental Assessment
Towards A Safer Design of Helmets: Finite Element & Experimental Assessment
 

Viewers also liked

A Survey Analysis on CMOS Integrated Circuits with Clock-Gated Logic Structure
A Survey Analysis on CMOS Integrated Circuits with Clock-Gated Logic StructureA Survey Analysis on CMOS Integrated Circuits with Clock-Gated Logic Structure
A Survey Analysis on CMOS Integrated Circuits with Clock-Gated Logic StructureIJERA Editor
 
Amelioration of safety management in infrastructure projects
Amelioration of safety management in infrastructure projectsAmelioration of safety management in infrastructure projects
Amelioration of safety management in infrastructure projectsIJERA Editor
 
Data Collection via Synthetic Aperture Radiometry towards Global System
Data Collection via Synthetic Aperture Radiometry towards Global SystemData Collection via Synthetic Aperture Radiometry towards Global System
Data Collection via Synthetic Aperture Radiometry towards Global SystemIJERA Editor
 
Study of Noise Levels at Commercial and Industrial Areas in an Urban Environment
Study of Noise Levels at Commercial and Industrial Areas in an Urban EnvironmentStudy of Noise Levels at Commercial and Industrial Areas in an Urban Environment
Study of Noise Levels at Commercial and Industrial Areas in an Urban EnvironmentIJERA Editor
 
Performance of water and diluted ethylene glycol as coolants for electronic c...
Performance of water and diluted ethylene glycol as coolants for electronic c...Performance of water and diluted ethylene glycol as coolants for electronic c...
Performance of water and diluted ethylene glycol as coolants for electronic c...IJERA Editor
 
Socially Shared Images with Automated Annotation Process by Using Improved Us...
Socially Shared Images with Automated Annotation Process by Using Improved Us...Socially Shared Images with Automated Annotation Process by Using Improved Us...
Socially Shared Images with Automated Annotation Process by Using Improved Us...IJERA Editor
 
Financing Small Scale Contractors through Mobilization Advance Payments for I...
Financing Small Scale Contractors through Mobilization Advance Payments for I...Financing Small Scale Contractors through Mobilization Advance Payments for I...
Financing Small Scale Contractors through Mobilization Advance Payments for I...IJERA Editor
 
Securing virtualization in real world environments
Securing virtualization in real world environmentsSecuring virtualization in real world environments
Securing virtualization in real world environmentsArun Gopinath
 
Catchment classification: multivariate statistical analyses for physiographic...
Catchment classification: multivariate statistical analyses for physiographic...Catchment classification: multivariate statistical analyses for physiographic...
Catchment classification: multivariate statistical analyses for physiographic...IJERA Editor
 
Future of Supply Chain Management in Various food Production.
Future of Supply Chain Management in Various food Production.Future of Supply Chain Management in Various food Production.
Future of Supply Chain Management in Various food Production.IJERA Editor
 

Viewers also liked (20)

Jess ctev
Jess ctevJess ctev
Jess ctev
 
A42020106
A42020106A42020106
A42020106
 
B045070510
B045070510B045070510
B045070510
 
B42040526
B42040526B42040526
B42040526
 
A Survey Analysis on CMOS Integrated Circuits with Clock-Gated Logic Structure
A Survey Analysis on CMOS Integrated Circuits with Clock-Gated Logic StructureA Survey Analysis on CMOS Integrated Circuits with Clock-Gated Logic Structure
A Survey Analysis on CMOS Integrated Circuits with Clock-Gated Logic Structure
 
Amelioration of safety management in infrastructure projects
Amelioration of safety management in infrastructure projectsAmelioration of safety management in infrastructure projects
Amelioration of safety management in infrastructure projects
 
Bk4201413415
Bk4201413415Bk4201413415
Bk4201413415
 
Data Collection via Synthetic Aperture Radiometry towards Global System
Data Collection via Synthetic Aperture Radiometry towards Global SystemData Collection via Synthetic Aperture Radiometry towards Global System
Data Collection via Synthetic Aperture Radiometry towards Global System
 
O44097379
O44097379O44097379
O44097379
 
Study of Noise Levels at Commercial and Industrial Areas in an Urban Environment
Study of Noise Levels at Commercial and Industrial Areas in an Urban EnvironmentStudy of Noise Levels at Commercial and Industrial Areas in an Urban Environment
Study of Noise Levels at Commercial and Industrial Areas in an Urban Environment
 
Performance of water and diluted ethylene glycol as coolants for electronic c...
Performance of water and diluted ethylene glycol as coolants for electronic c...Performance of water and diluted ethylene glycol as coolants for electronic c...
Performance of water and diluted ethylene glycol as coolants for electronic c...
 
Socially Shared Images with Automated Annotation Process by Using Improved Us...
Socially Shared Images with Automated Annotation Process by Using Improved Us...Socially Shared Images with Automated Annotation Process by Using Improved Us...
Socially Shared Images with Automated Annotation Process by Using Improved Us...
 
Financing Small Scale Contractors through Mobilization Advance Payments for I...
Financing Small Scale Contractors through Mobilization Advance Payments for I...Financing Small Scale Contractors through Mobilization Advance Payments for I...
Financing Small Scale Contractors through Mobilization Advance Payments for I...
 
U04405117121
U04405117121U04405117121
U04405117121
 
D045031724
D045031724D045031724
D045031724
 
C046061219
C046061219C046061219
C046061219
 
Securing virtualization in real world environments
Securing virtualization in real world environmentsSecuring virtualization in real world environments
Securing virtualization in real world environments
 
C045051318
C045051318C045051318
C045051318
 
Catchment classification: multivariate statistical analyses for physiographic...
Catchment classification: multivariate statistical analyses for physiographic...Catchment classification: multivariate statistical analyses for physiographic...
Catchment classification: multivariate statistical analyses for physiographic...
 
Future of Supply Chain Management in Various food Production.
Future of Supply Chain Management in Various food Production.Future of Supply Chain Management in Various food Production.
Future of Supply Chain Management in Various food Production.
 

Similar to Stiffness Characteristics of Joshi’s External Stabilization System under Axial Compression: a Finite Element Method Based Study

Experimental stress analysis and fea of dental implants
Experimental stress analysis and fea of dental implantsExperimental stress analysis and fea of dental implants
Experimental stress analysis and fea of dental implantseSAT Journals
 
ANSYS USERMAT for Prediction of Bone Failure
ANSYS USERMAT for Prediction of Bone FailureANSYS USERMAT for Prediction of Bone Failure
ANSYS USERMAT for Prediction of Bone FailureAnsys
 
IRJET- Design and Analysis of Lumbar Spine using Finite Element Method
IRJET- Design and Analysis of Lumbar Spine using Finite Element MethodIRJET- Design and Analysis of Lumbar Spine using Finite Element Method
IRJET- Design and Analysis of Lumbar Spine using Finite Element MethodIRJET Journal
 
Probabilistic Design of Hollow Circular Composite Structure by using Finite E...
Probabilistic Design of Hollow Circular Composite Structure by using Finite E...Probabilistic Design of Hollow Circular Composite Structure by using Finite E...
Probabilistic Design of Hollow Circular Composite Structure by using Finite E...IJERA Editor
 
AXIAL BEHAVIOR OF STAINLESS STEEL SQUARE THIN-WALLED TUBES STIFFENED INTERNA...
 AXIAL BEHAVIOR OF STAINLESS STEEL SQUARE THIN-WALLED TUBES STIFFENED INTERNA... AXIAL BEHAVIOR OF STAINLESS STEEL SQUARE THIN-WALLED TUBES STIFFENED INTERNA...
AXIAL BEHAVIOR OF STAINLESS STEEL SQUARE THIN-WALLED TUBES STIFFENED INTERNA...IAEME Publication
 
ANALYSIS OF STEEL FRAMES WITH BRACINGS FOR SEISMIC LOADS
ANALYSIS OF STEEL FRAMES WITH BRACINGS FOR SEISMIC LOADSANALYSIS OF STEEL FRAMES WITH BRACINGS FOR SEISMIC LOADS
ANALYSIS OF STEEL FRAMES WITH BRACINGS FOR SEISMIC LOADSIAEME Publication
 
WRASASgDGGS DDGBLFAfbJKCfaiGFBKsIgsdJBIGDSKJ
WRASASgDGGS DDGBLFAfbJKCfaiGFBKsIgsdJBIGDSKJWRASASgDGGS DDGBLFAfbJKCfaiGFBKsIgsdJBIGDSKJ
WRASASgDGGS DDGBLFAfbJKCfaiGFBKsIgsdJBIGDSKJJohn Ajish
 
IRJET- A Review Paper on Mechanical Behavior and Engineering Applications of ...
IRJET- A Review Paper on Mechanical Behavior and Engineering Applications of ...IRJET- A Review Paper on Mechanical Behavior and Engineering Applications of ...
IRJET- A Review Paper on Mechanical Behavior and Engineering Applications of ...IRJET Journal
 
Design And Analysis of Different Artificial Human Hip Stems Based on Fenestra...
Design And Analysis of Different Artificial Human Hip Stems Based on Fenestra...Design And Analysis of Different Artificial Human Hip Stems Based on Fenestra...
Design And Analysis of Different Artificial Human Hip Stems Based on Fenestra...IRJET Journal
 
ANALYSIS AND EXPERIMENTAL STUDY ON STRENGTH AND BEHAVIOUR OF EXTERIOR BEAM-CO...
ANALYSIS AND EXPERIMENTAL STUDY ON STRENGTH AND BEHAVIOUR OF EXTERIOR BEAM-CO...ANALYSIS AND EXPERIMENTAL STUDY ON STRENGTH AND BEHAVIOUR OF EXTERIOR BEAM-CO...
ANALYSIS AND EXPERIMENTAL STUDY ON STRENGTH AND BEHAVIOUR OF EXTERIOR BEAM-CO...IAEME Publication
 
Finite Element Analysis of Osteosynthesis Miniplate for the Reconstruction of...
Finite Element Analysis of Osteosynthesis Miniplate for the Reconstruction of...Finite Element Analysis of Osteosynthesis Miniplate for the Reconstruction of...
Finite Element Analysis of Osteosynthesis Miniplate for the Reconstruction of...UniversitasGadjahMada
 
ANALYTICAL INVESTIGATION ON DYNAMIC BEHAVIOUR OF COMPOSITE BEAM
ANALYTICAL INVESTIGATION ON DYNAMIC BEHAVIOUR OF COMPOSITE BEAMANALYTICAL INVESTIGATION ON DYNAMIC BEHAVIOUR OF COMPOSITE BEAM
ANALYTICAL INVESTIGATION ON DYNAMIC BEHAVIOUR OF COMPOSITE BEAMIAEME Publication
 
Parametric study on reinforced concrete beam
Parametric study on reinforced concrete beamParametric study on reinforced concrete beam
Parametric study on reinforced concrete beamAlexander Decker
 
DESIGN AND ANALYSIS OF BRIDGE WITH TWO ENDS FIXED ON VERTICAL WALL USING FIN...
 DESIGN AND ANALYSIS OF BRIDGE WITH TWO ENDS FIXED ON VERTICAL WALL USING FIN... DESIGN AND ANALYSIS OF BRIDGE WITH TWO ENDS FIXED ON VERTICAL WALL USING FIN...
DESIGN AND ANALYSIS OF BRIDGE WITH TWO ENDS FIXED ON VERTICAL WALL USING FIN...IAEME Publication
 
Analytical Investigation on External Beam-Column Joint Using ANSYS By Varying...
Analytical Investigation on External Beam-Column Joint Using ANSYS By Varying...Analytical Investigation on External Beam-Column Joint Using ANSYS By Varying...
Analytical Investigation on External Beam-Column Joint Using ANSYS By Varying...IJERA Editor
 

Similar to Stiffness Characteristics of Joshi’s External Stabilization System under Axial Compression: a Finite Element Method Based Study (20)

Experimental stress analysis and fea of dental implants
Experimental stress analysis and fea of dental implantsExperimental stress analysis and fea of dental implants
Experimental stress analysis and fea of dental implants
 
ANSYS USERMAT for Prediction of Bone Failure
ANSYS USERMAT for Prediction of Bone FailureANSYS USERMAT for Prediction of Bone Failure
ANSYS USERMAT for Prediction of Bone Failure
 
Ahmed. I .Razooqi
Ahmed. I .RazooqiAhmed. I .Razooqi
Ahmed. I .Razooqi
 
Aj03302190225
Aj03302190225Aj03302190225
Aj03302190225
 
IRJET- Design and Analysis of Lumbar Spine using Finite Element Method
IRJET- Design and Analysis of Lumbar Spine using Finite Element MethodIRJET- Design and Analysis of Lumbar Spine using Finite Element Method
IRJET- Design and Analysis of Lumbar Spine using Finite Element Method
 
Probabilistic Design of Hollow Circular Composite Structure by using Finite E...
Probabilistic Design of Hollow Circular Composite Structure by using Finite E...Probabilistic Design of Hollow Circular Composite Structure by using Finite E...
Probabilistic Design of Hollow Circular Composite Structure by using Finite E...
 
AXIAL BEHAVIOR OF STAINLESS STEEL SQUARE THIN-WALLED TUBES STIFFENED INTERNA...
 AXIAL BEHAVIOR OF STAINLESS STEEL SQUARE THIN-WALLED TUBES STIFFENED INTERNA... AXIAL BEHAVIOR OF STAINLESS STEEL SQUARE THIN-WALLED TUBES STIFFENED INTERNA...
AXIAL BEHAVIOR OF STAINLESS STEEL SQUARE THIN-WALLED TUBES STIFFENED INTERNA...
 
ANALYSIS OF STEEL FRAMES WITH BRACINGS FOR SEISMIC LOADS
ANALYSIS OF STEEL FRAMES WITH BRACINGS FOR SEISMIC LOADSANALYSIS OF STEEL FRAMES WITH BRACINGS FOR SEISMIC LOADS
ANALYSIS OF STEEL FRAMES WITH BRACINGS FOR SEISMIC LOADS
 
WRASASgDGGS DDGBLFAfbJKCfaiGFBKsIgsdJBIGDSKJ
WRASASgDGGS DDGBLFAfbJKCfaiGFBKsIgsdJBIGDSKJWRASASgDGGS DDGBLFAfbJKCfaiGFBKsIgsdJBIGDSKJ
WRASASgDGGS DDGBLFAfbJKCfaiGFBKsIgsdJBIGDSKJ
 
IRJET- A Review Paper on Mechanical Behavior and Engineering Applications of ...
IRJET- A Review Paper on Mechanical Behavior and Engineering Applications of ...IRJET- A Review Paper on Mechanical Behavior and Engineering Applications of ...
IRJET- A Review Paper on Mechanical Behavior and Engineering Applications of ...
 
Design And Analysis of Different Artificial Human Hip Stems Based on Fenestra...
Design And Analysis of Different Artificial Human Hip Stems Based on Fenestra...Design And Analysis of Different Artificial Human Hip Stems Based on Fenestra...
Design And Analysis of Different Artificial Human Hip Stems Based on Fenestra...
 
Gn3611721176
Gn3611721176Gn3611721176
Gn3611721176
 
W4501132137
W4501132137W4501132137
W4501132137
 
ANALYSIS AND EXPERIMENTAL STUDY ON STRENGTH AND BEHAVIOUR OF EXTERIOR BEAM-CO...
ANALYSIS AND EXPERIMENTAL STUDY ON STRENGTH AND BEHAVIOUR OF EXTERIOR BEAM-CO...ANALYSIS AND EXPERIMENTAL STUDY ON STRENGTH AND BEHAVIOUR OF EXTERIOR BEAM-CO...
ANALYSIS AND EXPERIMENTAL STUDY ON STRENGTH AND BEHAVIOUR OF EXTERIOR BEAM-CO...
 
Finite Element Analysis of Osteosynthesis Miniplate for the Reconstruction of...
Finite Element Analysis of Osteosynthesis Miniplate for the Reconstruction of...Finite Element Analysis of Osteosynthesis Miniplate for the Reconstruction of...
Finite Element Analysis of Osteosynthesis Miniplate for the Reconstruction of...
 
ANALYTICAL INVESTIGATION ON DYNAMIC BEHAVIOUR OF COMPOSITE BEAM
ANALYTICAL INVESTIGATION ON DYNAMIC BEHAVIOUR OF COMPOSITE BEAMANALYTICAL INVESTIGATION ON DYNAMIC BEHAVIOUR OF COMPOSITE BEAM
ANALYTICAL INVESTIGATION ON DYNAMIC BEHAVIOUR OF COMPOSITE BEAM
 
IE447FinalPaper
IE447FinalPaperIE447FinalPaper
IE447FinalPaper
 
Parametric study on reinforced concrete beam
Parametric study on reinforced concrete beamParametric study on reinforced concrete beam
Parametric study on reinforced concrete beam
 
DESIGN AND ANALYSIS OF BRIDGE WITH TWO ENDS FIXED ON VERTICAL WALL USING FIN...
 DESIGN AND ANALYSIS OF BRIDGE WITH TWO ENDS FIXED ON VERTICAL WALL USING FIN... DESIGN AND ANALYSIS OF BRIDGE WITH TWO ENDS FIXED ON VERTICAL WALL USING FIN...
DESIGN AND ANALYSIS OF BRIDGE WITH TWO ENDS FIXED ON VERTICAL WALL USING FIN...
 
Analytical Investigation on External Beam-Column Joint Using ANSYS By Varying...
Analytical Investigation on External Beam-Column Joint Using ANSYS By Varying...Analytical Investigation on External Beam-Column Joint Using ANSYS By Varying...
Analytical Investigation on External Beam-Column Joint Using ANSYS By Varying...
 

Recently uploaded

HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICSHARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICSRajkumarAkumalla
 
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVHARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVRajaP95
 
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...RajaP95
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Dr.Costas Sachpazis
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...
High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...
High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...Call Girls in Nagpur High Profile
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).pptssuser5c9d4b1
 
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)Suman Mia
 
Internship report on mechanical engineering
Internship report on mechanical engineeringInternship report on mechanical engineering
Internship report on mechanical engineeringmalavadedarshan25
 
Porous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writingPorous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writingrakeshbaidya232001
 
Current Transformer Drawing and GTP for MSETCL
Current Transformer Drawing and GTP for MSETCLCurrent Transformer Drawing and GTP for MSETCL
Current Transformer Drawing and GTP for MSETCLDeelipZope
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130Suhani Kapoor
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escortsranjana rawat
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxAsutosh Ranjan
 

Recently uploaded (20)

HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICSHARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
 
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVHARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
 
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptxExploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
 
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
 
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
 
High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...
High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...
High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
 
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
 
Internship report on mechanical engineering
Internship report on mechanical engineeringInternship report on mechanical engineering
Internship report on mechanical engineering
 
Porous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writingPorous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writing
 
Current Transformer Drawing and GTP for MSETCL
Current Transformer Drawing and GTP for MSETCLCurrent Transformer Drawing and GTP for MSETCL
Current Transformer Drawing and GTP for MSETCL
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
 
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINEDJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
 
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptx
 

Stiffness Characteristics of Joshi’s External Stabilization System under Axial Compression: a Finite Element Method Based Study

  • 1. Rajeev Kumar et al Int. Journal of Engineering Research and Applications www.ijera.com ISSN : 2248-9622, Vol. 4, Issue 7( Version 5), July 2014, pp.43-47 www.ijera.com 43 | P a g e Stiffness Characteristics of Joshi’s External Stabilization System under Axial Compression: a Finite Element Method Based Study Rajeev Kumar*, R C Gupta*, Sanjay Mishra** *Dept of Mechanical Engineering, Institute of Engineering & Technology, Lucknow, UP, India ** Scientist G, Department of Science & Technology (DST), New Delhi, India ABSTRACT A finite element model of fractured tibia with Joshi’s External Stabilizing System (JESS) mounted on it was developed using 3D beam elements in the ANSYS software. The model was loaded in axial compression and the average axial stiffness of the model was calculated. The analytical value of axial stiffness was compared with reported experimental value to validate the finite element model. The validated model was used to carry out parametric studies on the model to determine the axial properties of JESS. It was observed that axial stiffness of JESS increased by 58% when k-wire diameter was varied from 2 mm to 4 mm while keeping other geometric configurations of the device constant; however, the axial stiffness of the device does not show any significant improvement when the diameter of medio-lateral pins in diaphyseal hold were increased. The findings should help in understanding the axial properties of JESS so that it can be used judiciously in clinical applications. Keywords – JESS, external fixators, tibial fracture, axial stiffness, k-wires, finite element analysis I. INTRODUCTION External fixation is a method of stabilization of bone fractures in which a number of percutenous metal pins pass through the fractured bone segments with their ends connected to a rigid frame. Joshi External Stabilizing System (JESS) is an external bone stabilizing device used in the Indian subcontinent. It was designed and fabricated by Dr BB Joshi in late seventies primarily for the treatment of hand trauma [1]. However, due to its low cost and highly versatile nature it evolved with time to be used in the treatment of variety of musculoskeletal disorders. At present, JESS is prominently used in intra-articular distal radial fractures [2], management of idiopathic clubfoot [3], hand trauma and its sequels [4]. Recent applications have been reported about JESS being used in treatment of injuries of tibial plateau and tibial plafond [5]. Many experimental procedures have been employed to study the behaviour of external fixation devices. Yilmaz et al. [6] conducted an experimental study to determine the stiffness characteristics of standard and hybrid Ilizarov circular fixators. Stein et al. [7] performed a biomechanical study on hybrid ring tubular external fixator to measure and compare the mechanical properties of different hybrid fixators. Schrøder et al. [8] performed experimental investigations of four different configurations of the Hoffmann external fixation system to assess its mechanical properties. Experimental procedures are important in establishing basic characteristics of fixation devices. However, they have some inherent shortcomings. Such procedures are time consuming, costly and need a large number of experimental data. Also, a minor change in the device requires another set of experiments to collect new data. As a result analytical methods such as finite element method have gained popularity for evaluating the mechanical properties of fixation devices analytically. Many researchers have used finite element analysis to study the mechanical properties of external fixation devices. Rybicki [9] presented the role and approach of finite element analysis in orthopedic studies and the application of FEM in the analysis of stresses in intact bones, analysis of fixation devices and prosthetic devices with a review of works carried out by researchers in these areas. Chao and An [10] classified the commonly used external fixation devices according to their geometrical configurations. They proposed a two dimensional and three dimensional beam elements to develop the finite element model of external fixators. Koo et al. [11] developed a finite element model using three dimensional beam elements in ABAQUS software and analyzed it for various loading conditions. It was found that under constrained axial compression, the pin diameter is the most critical parameter that could affect the system stiffness followed by pin offset. It was also reported that once the pin diameter exceeded certain value, it could not improve the system stiffness significantly. Bartel et al. [12, 13] developed a model using symmetric and unsymmetrical beam theory and demonstrated that beam theory, if used within its limitations, could provide an excellent model for understanding the overall behaviour of bone-implant RESEARCH ARTICLE OPEN ACCESS
  • 2. Rajeev Kumar et al Int. Journal of Engineering Research and Applications www.ijera.com ISSN : 2248-9622, Vol. 4, Issue 7( Version 5), July 2014, pp.43-47 www.ijera.com 44 | P a g e systems. Prendergast et al. [14] conducted finite element analysis and mechanical testing of unilateral and bilateral external fixators. Watson et al. [15] developed a modular FE model of the components of Illizarov external fixation system to predict mechanical properties of any configuration of the device. The first reported study on characterization of JESS was carried out by Kumar et al. [16] who conducted an experimental and finite element based investigation to determine and compare the axial stiffness of JESS and proposed a validated FE model of JESS under axial compression. Present study aims to analyze of the effect of variation in k-wire diameter on overall axial stiffness of JESS while keeping all other geometrical parameters of the fixator constant using finite element method. It will also study the change in axial stiffness of device when the size of medio-lateral pins in diaphyseal hold is varied. II. MATERIALS AND METHODS 2.1 Geometrical features of JESS A JESS frame used for treatment of proximal tibial fractures can be configured in two parts, a proximal hold or helmet and a diaphyseal hold. The proximal hold consists of two circular rods of 4 mm diameter curved into a three quarter circular rings mounted on the proximal tibia with help of three percutenous pins (k-wires) inserted at about 22.5º with each other. The second three quarter circular ring has a lesser diameter than the first ring and it is added to reinforce the strength of proximal hold. For a JESS used in treatment of metaphyseal tibial fracture, the k-wire diameters may vary from 2.0 mm to 4.0 mm depending upon the clinical requirements. Universal link joints are used to join k-wires to the outer and inner circular rings with proximal tibia. The diaphyseal hold is composed of three parallel pins inserted in the tibia diaphysis in medio- lateral plane (m-l pins) below the fracture fragment and are attached to two Z shaped connecting rods on the either side of tibia using universal link joints. The proximal hold and diaphyseal hold are in turn connected to each other with help of two anterior and two posterior connecting rods joining diaphysis hold to the outer ring of the proximal hold. In addition, one half pin is inserted from anterior in anterior- posterior plane to provide further fragment stability. Fig.1 shows the laboratory specimen of JESS configured on steel tubes simulated as tibial bone which was used for developing a three dimensional finite element model in our study. 2.2 Finite element analysis of JESS The aim of this study was to evaluate the effect of geometric variations on the overall axial stiffness of JESS using a validated three dimensional finite element model of JESS as proposed by Kumar et. al. [16]. Therefore, the finite element model of JESS was developed using the identical geometric specifications and material properties as prescribed in the study. Fig.1 JESS configured on steel tubes To create the FE model, the structural geometry of the JESS configured on steel tubes was created using ANSYS finite element software. A 15 mm gap was provided between the ends of the hollow steel tubes to represent the metaphyseal tibial fracture. The gap was maintained at 15 mm to ensure the complete load transfer through fixator rather than through the steel tubes. The wireframe model was discretized using 3D beam element (Beam 188). As the different components of the model had different sizes, appropriate section properties were allocated to each component. The discretized model of JESS consisted of 146 nodes and 164 elements. The elastic modulus and Poisson’s ratio for the parts of JESS, k-wires, m-l pins & idealized tibial bone segments were taken as 200 GPa & 0.28 respectively. The idealized pin-bone and pin fixator interfaces and various universal link joints of the JESS were modeled as rigid joints. To measure the axial compression of JESS, the distal end of the model was fixed by setting all degrees of freedom to zero. A vertical compression load was applied to the proximal end. Fig.2 shows the FE model of JESS with applied load and boundary conditions. The model was analyzed for static, linear analysis under axial compression.
  • 3. Rajeev Kumar et al Int. Journal of Engineering Research and Applications www.ijera.com ISSN : 2248-9622, Vol. 4, Issue 7( Version 5), July 2014, pp.43-47 www.ijera.com 45 | P a g e Fig.2 FE model with load and boundary conditions applied The axial stiffness and corresponding inter fragmentary displacements of the JESS frame during the axial loading were calculated from the FE model. Fig.3 shows the deformed FE model of JESS configured on steel tubes under axial compressive load. The validated FE model was used to carry out two parametric studies to assess the effect of geometric variations on the axial properties of JESS. Fig.3 Deformed model of JESS under axial compression In the first study, the diameter of the k-wires in proximal hold was varied from 2 mm to 4 mm in the step of 0.5 mm and the axial stiffness of the JESS frame was evaluated. All other geometrical parameters were kept identical including the m-l pin diameter. These steps were repeated for another value of m-l pins diameter and variation in axial stiffness was evaluated. This way total four set of data were collected for four different diameters of m-l pins. In the second study, for a selected diameter of k-wire, the diameters of m-l pins in the diaphysis hold were increased from 2.5 mm to 4 mm in a step of 0.5 mm and axial stiffness in each case was calculated. Then k-wire diameter was set to another value and m-l pins diameters were varied to evaluate another set of stiffness data for JESS. III. RESULTS The axial stiffness of JESS was calculated by finite element model under individual loading condition. The average axial stiffness obtained by FE analysis having k-wire size of 2 mm was compared with the reported experimental value [16] and was found to be comparable. Thus, the FE model can be used as valid model to simulate the axial mechanical properties of JESS. Effect of geometric variations were studied and it was observed that the average axial stiffness of JESS improved by about 58% on varying the diameter of the k-wires from 2 mm to 4 mm in proximal hold with m-l pin diameter kept at 4 mm and keeping all other geometrical parameters constant. Further, by changing the medio-lateral pins diameter in diaphyseal hold from 2.5 mm to 4 mm the average axial stiffness was increased by only about 6.2% while k-wire diameter was kept constant at 4 mm. IV. DISCUSSION Goodship and Kenwright [18] has reported that the mechanical properties of an external fixation device influence the biological environment at the bone fracture and controlled micro-movement of fracture could facilitate secondary healing in the fractured bone. Mechanical properties of fixation devices also affect the outcome of any fixation process. A very rigid fixator can delay healing; on the other hand an over flexible fixation device may lead to increase chances of pin-bone tract infection, mal- union and even non-union in some cases. Therefore, it is extremely important for a surgeon to have a good knowledge of comparative mechanical properties of the fixation device to use it in clinical applications. The axial stiffness of the JESS was calculated by developing a three dimensional finite element model using 3D beam elements in ANSYS software. It was observed that the average axial stiffness of the JESS increases nonlinearly with increase in the diameter of k-wires and the overall increase in axial stiffness by changing the diameter of k-wires from 2 mm to 4 mm was about 58%. The change in axial stiffness for each size of k-wire with m-l pins diameter maintained at 2.5 mm is shown in figure 4.
  • 4. Rajeev Kumar et al Int. Journal of Engineering Research and Applications www.ijera.com ISSN : 2248-9622, Vol. 4, Issue 7( Version 5), July 2014, pp.43-47 www.ijera.com 46 | P a g e Fig.4 Variation in axial stiffness of JESS with change in k-wire diameter (m-l pins diameter 2.5 mm) It was also observed that incremental gain in axial stiffness from diameter 2 mm to 4 mm drops as the k-wire diameter increases. The gain in axial stiffness was about 19% when the diameter of k-wire was changed from 2 mm to 2.5 mm however, the increase was only 5.5% when the diameter was changed from 3.5 mm to 4 mm. As 4 mm k-wire size is the maximum pin diameter used in a standard JESS under clinical application therefore, axial stiffness was not calculated for higher values of k-wire size beyond 4 mm. The overall improvement in axial stiffness was about 54% on variation of k-wire size when m-l pin diameter was fixed at 2.5 mm and it was about 58% when m-l pin diameter was fixed at 4 mm. Fig.5 shows the effect of variation in k-wire diameters on axial stiffness of JESS. Fig.5 Variation in axial stiffness of JESS for different set of m-l pin sizes In the second part, k-wire diameters were fixed and m-l pins diameter in diaphysis hold was varied from 2.5 mm to 4 mm in the step of 0.5 mm and for each case, axial stiffness of the device was calculated. These five sets of observations are plotted fig.6. It can be seen that by changing the m-l pins diameter from 2.5 mm to 4 mm while keeping the k- wire diameter constant increases the axial stiffness by a mere 3.5% to 6.2%. It can therefore be suggested that the size of medio-lateral pin does not play a significant role in determining the axial stiffness of the fixator. On the other hand, the k-wire size used in proximal hold plays a major role in controlling the overall axial stiffness of the JESS fixation device and can be manipulated to improve the stiffness of JESS in clinical applications. V. CONCLUSION Finite element model of JESS configured on steel tubes was developed with 3D beam element in ANSYS software and two parametric studies were conducted. Results show that the axial stiffness of the JESS can be improved by changing the diameter of k- wires in proximal hold. Thus, k-wire sizes play a significant role in determining the axial properties of JESS. Fig.6 Variation in axial stiffness for varying medio- lateral pin diameter The results of the study are based on a standard configuration of JESS device commonly used in stabilization of tibial fractures. However, the JESS is available in slightly different modules by different manufacturers in India. The exact configurations of JESS may vary in clinics depending upon surgeon’s own judgment and practice. Therefore, the above results may not be generalized for all types of the JESS configurations. Nevertheless, the present study suggests a way by which axial stiffness of JESS frame can be improved as per the clinical requirements. Conflict of interest statement All authors hereby declare that this study was not funded by any agency and there are no personal relationships with anyone who could influence the results of present study. REFERENCES [1] R. Prabhu, A legend forever-Dr Brij Bhusan Joshi, Indian Journal Orthopedics, 43(3), 2009, 312. [2] S.Thomas, C. John, T. P. Johnny, Intra- articular distal radial fractures – external
  • 5. Rajeev Kumar et al Int. Journal of Engineering Research and Applications www.ijera.com ISSN : 2248-9622, Vol. 4, Issue 7( Version 5), July 2014, pp.43-47 www.ijera.com 47 | P a g e fixation or conventional closed reduction, Journal of Orthopedics, 4(2), 2007, e39. [3] S. Suresh, A. Ahmed, V. K. Sharma, Role of Joshi's External Stabilisation System fixator in the management of idiopathic clubfoot, Journal of Orthopedic Surgery, 11(2), 2003, pp. 194–201. [4] B. B. Joshi, Joshi’s External Stabilization System (JESS): a simple mini external fixator for the management of hand trauma and its sequels, Injury, 28, 1997, 244-244. [5] B. B. Joshi, Ram Prabhu, N. A. Antao, R. Rohira, M. Vaishnav, L. Maratha, Trauma management by Joshi’s External Stabilization System (JESS), (JESS Research & Development Society, 2007). [6] E. Yilmaz, O. Bllhan, L. Karakurt, N. Arslan, E. Serin, Mechanical performance of hybrid Ilizarov external fixator in comparison with Ilizarov circular external fixator, Clinical Biomech., 18, 2003, 518- 522. [7] H. Stein, R. Mosheiff, F. baumgart, R. Frigg, S. M. Parren, J. Cordey, The hybrid ring tubular extrenal fixator: A biomechanical study, Clinical Biomech., 12 (4), 1997, 259-266. [8] H. A. Schrøder, R. E. Weeth, T. Madsen, Experimental analysis of Hoffmann external fixation in various mountings, Clinical Biomech., 16, 2001, 151-159. [9] Edmund F. Rybicki, The role of finite element models in orthopedics, in R.H. Gallaher, B. R. Simon, P. C. Johnson, J. F. Cross (Ed.), Finite Element in Biomechanics (John Wiley & Sons, USA, 1982) 181-194. [10] E. Y. S. Chao, K. N. An, Biomechanical analysis of external fixation devices for the treatment of open bone fractures, in R.H. Gallaher, B. R. Simon, P. C. Johnson, J. F. Cross (Ed.), Finite Elements in Biomechanics (John Wiley & Sons, USA, 1982) 195-222. [11] T. K. K. Koo, Y. H. Kim, D. B. Choi, K. G. Hua, J. Lin, N. Inove, E. Y. S. Chao, Stiffness analysis of Dynafix external fixator system, Summer Bioengineering Conference, Florida, USA, 2003, 1227-1228. [12] D. L. Bartel, D. T. Davy, T. M. Keaveny, Structural analysis of musculoskeletal system: Beam Theory, Orthopedic Biomechanics: Mechanics & Design in Musculoskeletal Systems, (New Jeresy: Pearson Prentice Hall, 2006) 168-199. [13] D. L. Bartel, D. T. Davy, T. M. Keaveny, Fracture fixation devices, Orthopedic Biomechanics: Mechanics & Design in Musculoskeletal Systems, (New Jeresy: Pearson Prentice Hall, 2006) 261-287. [14] P. J. Prendergast, S. J. Toland, J. P. Corrigan, Finite element analysis and mechanical testing of external fixator designs, Proc. Instn. Mech. Engrs., 208, 1994, 103-110. [15] M. Watson, K. J. Mathias, N. Maffulli, D. W. Hukins, D. E. T. Shepherd, Finite element modelling of the Illizarov external fixation system, Proc. IMechE, Part H, J Engg. in Medicine, 221(8), 2007, 863-871. [16] Rajeev Kumar, Anupam Gupta, V P Sharma, Sanjay Mishra, Strength of Joshi External Stabilizing System, Journal of Orthopedic Surgery, 19(1), 2011, 72-75. [17] Leonid N. Solomin, The basic principles of external fixation using Illizarov device, (Milan: Springer-Verlog Italia, 2008) 341- 349. [18] A. E. Goodship, J. Kenwright, The influence of induced micromovement upon the healing of experimental tibial fractures, The Journal of Bone & Joint Surgery, 67-B(4), 1885, 650-655.