How to Calibrate an RTD
Using a Dry-Block Calibrator
Kurt Zabriskie
Fluke Calibration
Fluke – American Fork, UT
• Kurt Zabriskie
• 17 Years with Fluke/Hart Scientific
– Western Regional Product Manager - Temperature
• Kurt.Zabriskie@flukecal.com
Agenda
How to Calibrate an RTD Using a Dry-Block Calibrator
• Quick Dry-block introduction
• Dry-block sources of error
• Finding solutions for common errors and problems
• Calculating uncertainties
• How to calibrate probes with odd shapes and sizes
• Temperature range concerns
• Liquid-in-glass thermometers
• RTD calibration example
• Dry-block maintenance
• Summary
Dry-Block Introduction
Dry-Blocks/Dry-Wells
 Moderate accuracy
 Fixed hole diameter
 Fixed immersion depth
 Dry and clean
 Portable
 Faster temperature changes
 Internal reference probe
9100S
Handheld Dry-well
9190A
Ultra-Cool
Field Metrology Well
914X Series
Field Metrology Wells
9103 9140 9141
Dry-Block Calibrators
Dry-Block Sources of Errors
• Immersion effects
• Stem conduction
• Well contact errors
• Display accuracy
• Stability
• Uniformity (radial & axial)
• Loading
Dry-Block – Immersion Effects
• Immersion depth of
the UUT is critical
• Dry-blocks are
generally calibrated
by fully inserting a
reference probe
Dry-Block – Immersion Effects
• By adding a reference
probe, we can be less
concerned with the
control probe location
• The UUT and Reference
can be moved to different
depths
• Immersion depth should
still be considered
• (15 X Probe Diameter)
• Many Fluke dry-blocks
include a reference Input
Dry-Block – Stem Conduction
• May impact the ability to make a good measurement
• The probe sheath acts as heat sink
• Probe diameter matters, length is less critical
• (15 x probe diameter) a good rule to follow
Heat
Dry-Block – Well Contact Errors
• Fit is important
• Loose fitting probes exhibit
erratic behavior
• Wells should be 0.005”-
0.010” larger than the
diameter of the probe for
0.5” diameter probes and
smaller.
• Too snug, and the probe
may become stuck
Dry-Block – Display Accuracy
• Generally a one year
specification on a
spec sheet
• Internal control
sensors are designed
to be robust
• Display accuracy can
be one of the largest
contributors to the
overall uncertainty
Dry-Block – Stability
• Check dry-block
specifications
• Units generally come
with optimized
proportional bands
• Fine tuning the
proportional band
may help at specific
temperatures
Dry-Block – Axial Uniformity
• Axial Uniformity – Variation in the
temperature along the axial length of
the insert (block) within the
measurement zone
• Generally inherent to the dry-block
• The closer the sensing element is to
ambient air, the larger the
uncertainties
• Higher temperatures present larger
errors
• Look for dry-blocks with a calibrated
zone for optimal results
Dry-Block – Radial Uniformity
• Radial Uniformity –
Variation in the
temperature between
different wells of the
insert (block) within the
measurement zone.
• Mostly inherent to the
dry-block design
• Heater placement and
profiling is critical during
engineering of the dry-
block
Dry-Block – Loading
• Loading can impact
uncertainties.
• Some models have a
specification
• Units without
specifications can be
evaluated in the
field
Calculating Uncertainties
• RSS method is generally used to calculate
uncertainties when using a reference
2
3
2
2
2
1 )()()( bbbbtotal 
222
)()()( uniformitystabilityrefbtotal 
Calculating Uncertainties
• GUM compliant example
-38 °C 0 °C 157 °C 232 °C 420 °C 660 °C
Uncertainty Sources: Type mK mK mK mK mK mK
Process Var. (check std) Norm 3.0 3.0 5.0 5.0 5.0 6.0
UUT Precision (noise) Norm 2.8 2.8 2.8 3.3 4.4 5.6
Ref. Precision (noise) Norm 2.8 2.8 2.8 2.8 2.8 5.6
Ref. Calibration Norm 0.2 0.1 0.6 0.5 0.6 1.1
Ref. Drift Rect 1.7 2.0 3.2 3.8 5.1 6.7
Radial Uniformity Rect 10.0 10.0 30.0 30.0 30.0 30.0
Axial Uniformity Rect 6.3 6.3 6.3 6.3 6.3 6.3
Readout (SPRT) Rect 0.2 0.3 0.4 0.5 0.7 0.9
Readout (UUT) Rect 1.3 1.5 2.5 3.1 4.4 6.3
Insulation Leakage Rect 10. 0 10.0 10.0 10.0 10.0 10.0
UUT Repeatability (TPW) Norm 2.8 3.3 5.3 6.3 8.6 11.2
Total (k=2): 20.6 21.0 40.7 41.5 43.9 47.9
Dry-Block – Odd Shaped Probes
• Custom inserts are an
option
• Micro-Baths might be a
good option to explore
Temperature Range Concerns
• –100 to 1200 ºC
• May need to use multiple
units
• It’s ok to switch from one
dry-block to the next Model 9190A:
–95 to 140 ºC
Model 9143:
33 to 350 ºC
Model 9150:
150 to 1200 ºC
Dry-Block – LIG’s
• Liquid and glass thermometers are not
recommended for use in a dry-block calibrator
• Mercury thermometers are on their way out
due to environmental concerns
Dry-Block – Maintenance
• Keep those wells clean, Scotch-Brite pads
and a gun cleaning kit work nicely
• Avoid dropping inserts or other heavy
objects into the well
• Avoid the use of thermo grease
• Recalibrate regularly
• Verify stability
RTD Calibration Example
• Three point RTD Calibration
• –95 ºC, 0 ºC, 140 ºC
• Utilize the 9190A with the “Process
Option” to measure the UUT
• Set the 9190A to each of the three
set points, generally starting with
the lowest point
• Allow for plenty of soak time at
each temperature (15 minutes)
• Record a resistance at each set point
• Utilize at tool such as TableWare to
calculate coefficients
TableWare Software (Model 9933)
When the Super-DAQ is connected to a Fluke
Calibration dry-well, fluid bath, or furnace, it can
control the temperature source to calibrate up to
40 sensors automatically.
You simply program the setpoint temperatures
and their values, select a scan sequence, assign
a reference channel, and set the required stability
band.
The Super-DAQ monitors the temperature
source’s stability through the reference channel,
collects the data from the reference probe and
the “unit under test” (UUT) once stabilized, and
then advances to the next set-point temperature.
After you configure and start the test, you can
walk away to work on other things. The Super-
DAQ just made your day a whole lot easier!
Automate temperature sensor calibration
with the 1586A Super-DAQ
9190A Ultra-Cool
Field Metrology Well
1586A Super-DAQ
with DAQ-STAQ
Application Note and Video
Application Note:
Automating Temperature Sensor
Calibration with the 1586A Super-DAQ
Companion Video:
Automating Temperature Sensor
Calibration with the 1586A Super-DAQ
The application note and video demonstrate the Auto Test function of
the 1586A using a 9142 Field Metrology which can be substituted with
other Fluke Calibration dry-wells, fluid baths, and furnaces.
Dry-Block – Summary
• Dry-blocks are a great option in
many situations
• Highly portable and quick to
change temperature
• Sources of error should be
considered
• On board references are a great
way to combine several
instruments into one
• Be sure to contact us any time for
help with your specific
application
Where Can I Find Out More?
• Comparison Calibration Application Notes
– 914X Series Field Metrology Well
– 9190A Ultra-Cool Field Metrology Well
• Product information at: www.Transcat.com/Fluke
Limited Time Offer
• Free Temperature Probe with Metrology Well Purchase
• Order from Transcat
• Details at: www.transcat.com/Fluke
• 800-800-5001
Questions or Comments?
Email the moderator, Nicole VanWert:
nvanwert@Transcat.com
Transcat: 800-800-5001
www.Transcat.com
For related product information, go to:
www.Transcat.com/Fluke

More Related Content

What's hot

Temperature Transducer
Temperature TransducerTemperature Transducer
Temperature Transducer
AIT
 

What's hot (20)

Industrial Temperature Calibration Selection Guide by Fluke Calibration
Industrial Temperature Calibration Selection Guide by Fluke CalibrationIndustrial Temperature Calibration Selection Guide by Fluke Calibration
Industrial Temperature Calibration Selection Guide by Fluke Calibration
 
Ime unit 5 temperature measurement
Ime unit 5 temperature measurementIme unit 5 temperature measurement
Ime unit 5 temperature measurement
 
Contact temperature sensor calibration
Contact temperature sensor calibrationContact temperature sensor calibration
Contact temperature sensor calibration
 
Temperature measurement part II
Temperature measurement   part IITemperature measurement   part II
Temperature measurement part II
 
Industrial Temperature Measurement
Industrial Temperature MeasurementIndustrial Temperature Measurement
Industrial Temperature Measurement
 
Instrumentation measurement principles
Instrumentation  measurement principlesInstrumentation  measurement principles
Instrumentation measurement principles
 
Thermocouple
ThermocoupleThermocouple
Thermocouple
 
CTM PROVING
CTM PROVINGCTM PROVING
CTM PROVING
 
Measurements lecture 1
Measurements lecture 1Measurements lecture 1
Measurements lecture 1
 
Static characteristics of Instruments
Static characteristics of InstrumentsStatic characteristics of Instruments
Static characteristics of Instruments
 
Phased Array Ultrasonic Testing in lieu of Radiography
Phased Array Ultrasonic Testing in lieu of RadiographyPhased Array Ultrasonic Testing in lieu of Radiography
Phased Array Ultrasonic Testing in lieu of Radiography
 
Pressure gauge calibration process
Pressure gauge calibration processPressure gauge calibration process
Pressure gauge calibration process
 
Ch. no. 5 vibration and strain measurement
Ch. no. 5 vibration and strain measurementCh. no. 5 vibration and strain measurement
Ch. no. 5 vibration and strain measurement
 
Electromagnetic flow measurement
Electromagnetic flow measurementElectromagnetic flow measurement
Electromagnetic flow measurement
 
Level measurement
Level measurementLevel measurement
Level measurement
 
Level measurement theories of operation rev1
Level measurement theories of operation  rev1Level measurement theories of operation  rev1
Level measurement theories of operation rev1
 
Level Measurement
Level MeasurementLevel Measurement
Level Measurement
 
Sensors and its types
Sensors and its typesSensors and its types
Sensors and its types
 
Temperature Measurement
Temperature MeasurementTemperature Measurement
Temperature Measurement
 
Temperature Transducer
Temperature TransducerTemperature Transducer
Temperature Transducer
 

Viewers also liked

Yokogawa Presantesation(1)
Yokogawa Presantesation(1)Yokogawa Presantesation(1)
Yokogawa Presantesation(1)
Maulik kunjadiya
 

Viewers also liked (10)

Temperature calibration
Temperature calibrationTemperature calibration
Temperature calibration
 
Rtd and thermocouples
Rtd and thermocouplesRtd and thermocouples
Rtd and thermocouples
 
Best Seller - Dry Block Temperature Calibrator
Best Seller - Dry Block Temperature CalibratorBest Seller - Dry Block Temperature Calibrator
Best Seller - Dry Block Temperature Calibrator
 
Electronic Calibrator
Electronic CalibratorElectronic Calibrator
Electronic Calibrator
 
Furnace calibration - temperature uniformity survey
Furnace calibration - temperature uniformity surveyFurnace calibration - temperature uniformity survey
Furnace calibration - temperature uniformity survey
 
Understanding AMS2750E (intermediate) v2
Understanding AMS2750E (intermediate) v2Understanding AMS2750E (intermediate) v2
Understanding AMS2750E (intermediate) v2
 
Improving Safety in the Industrial Workplace with Infrared Technology
Improving Safety in the Industrial Workplace with Infrared TechnologyImproving Safety in the Industrial Workplace with Infrared Technology
Improving Safety in the Industrial Workplace with Infrared Technology
 
Calibration
CalibrationCalibration
Calibration
 
Yokogawa Presantesation(1)
Yokogawa Presantesation(1)Yokogawa Presantesation(1)
Yokogawa Presantesation(1)
 
Flowmeter course
Flowmeter courseFlowmeter course
Flowmeter course
 

Similar to How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015

How to calibrate an ir thermometer, frank liebmann, 2013 05-07
How to calibrate an ir thermometer, frank liebmann, 2013 05-07How to calibrate an ir thermometer, frank liebmann, 2013 05-07
How to calibrate an ir thermometer, frank liebmann, 2013 05-07
Jesus Enrique Oliva Salazar
 

Similar to How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015 (20)

Megger regidez dielectrica oil test- OTS seminar
Megger   regidez dielectrica oil test- OTS seminarMegger   regidez dielectrica oil test- OTS seminar
Megger regidez dielectrica oil test- OTS seminar
 
How to calibrate an ir thermometer, frank liebmann, 2013 05-07
How to calibrate an ir thermometer, frank liebmann, 2013 05-07How to calibrate an ir thermometer, frank liebmann, 2013 05-07
How to calibrate an ir thermometer, frank liebmann, 2013 05-07
 
Calibration times march 2011
Calibration times march 2011Calibration times march 2011
Calibration times march 2011
 
Calibration times march 2011
Calibration times march 2011Calibration times march 2011
Calibration times march 2011
 
910 System Overview
910 System Overview910 System Overview
910 System Overview
 
Datasheet Fluke 9140 Series. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 9140 Series. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 9140 Series. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 9140 Series. Hubungi PT. Siwali Swantika 021-45850618
 
project presentation 1 (2).pptx
project presentation 1 (2).pptxproject presentation 1 (2).pptx
project presentation 1 (2).pptx
 
910 System Overview1
910 System Overview1910 System Overview1
910 System Overview1
 
flash-point
flash-pointflash-point
flash-point
 
Fluke Data Acquisition System Selection Guide. Hubungi PT. Siwali Swantika 02...
Fluke Data Acquisition System Selection Guide. Hubungi PT. Siwali Swantika 02...Fluke Data Acquisition System Selection Guide. Hubungi PT. Siwali Swantika 02...
Fluke Data Acquisition System Selection Guide. Hubungi PT. Siwali Swantika 02...
 
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensorsCondumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
 
Temperature Applications for Process Calibration Tools - App Note
Temperature Applications for Process Calibration Tools - App NoteTemperature Applications for Process Calibration Tools - App Note
Temperature Applications for Process Calibration Tools - App Note
 
Reformer Tube Metallurgy: Design Considerations; Failure Mechanisms; Inspecti...
Reformer Tube Metallurgy: Design Considerations; Failure Mechanisms; Inspecti...Reformer Tube Metallurgy: Design Considerations; Failure Mechanisms; Inspecti...
Reformer Tube Metallurgy: Design Considerations; Failure Mechanisms; Inspecti...
 
In-Situ Calibration: Methods & Pitfalls of Thermal Mass Flow Meter Sensor Fie...
In-Situ Calibration: Methods & Pitfalls of Thermal Mass Flow Meter Sensor Fie...In-Situ Calibration: Methods & Pitfalls of Thermal Mass Flow Meter Sensor Fie...
In-Situ Calibration: Methods & Pitfalls of Thermal Mass Flow Meter Sensor Fie...
 
Qualification of high performance liquid chromatography
Qualification of high performance liquid chromatographyQualification of high performance liquid chromatography
Qualification of high performance liquid chromatography
 
pH electrodes-pH sensors CPS41
pH electrodes-pH sensors CPS41pH electrodes-pH sensors CPS41
pH electrodes-pH sensors CPS41
 
How to Select the Right Valve for your Application
How to Select the Right Valve for your ApplicationHow to Select the Right Valve for your Application
How to Select the Right Valve for your Application
 
Combination Guided Wave Radar Transmitter and Magnetic Level Gauge
Combination Guided Wave Radar Transmitter and Magnetic Level GaugeCombination Guided Wave Radar Transmitter and Magnetic Level Gauge
Combination Guided Wave Radar Transmitter and Magnetic Level Gauge
 
2014 PV Performance Modeling Workshop: Toward Reliable Module Temperature Mea...
2014 PV Performance Modeling Workshop: Toward Reliable Module Temperature Mea...2014 PV Performance Modeling Workshop: Toward Reliable Module Temperature Mea...
2014 PV Performance Modeling Workshop: Toward Reliable Module Temperature Mea...
 
Fire foe
Fire foeFire foe
Fire foe
 

More from Transcat

Managing Instrument Calibration and Maintenance Presented by Fluke.pptx
Managing Instrument Calibration and Maintenance Presented by Fluke.pptxManaging Instrument Calibration and Maintenance Presented by Fluke.pptx
Managing Instrument Calibration and Maintenance Presented by Fluke.pptx
Transcat
 

More from Transcat (20)

Predictive Maintenance Best Practices Presented by Fluke.pptx
Predictive Maintenance Best Practices Presented by Fluke.pptxPredictive Maintenance Best Practices Presented by Fluke.pptx
Predictive Maintenance Best Practices Presented by Fluke.pptx
 
Managing Instrument Calibration and Maintenance Presented by Fluke.pptx
Managing Instrument Calibration and Maintenance Presented by Fluke.pptxManaging Instrument Calibration and Maintenance Presented by Fluke.pptx
Managing Instrument Calibration and Maintenance Presented by Fluke.pptx
 
Tools and Techniques for Commissioning and Maintaining PV Systems
Tools and Techniques for Commissioning and Maintaining PV SystemsTools and Techniques for Commissioning and Maintaining PV Systems
Tools and Techniques for Commissioning and Maintaining PV Systems
 
Use of a Druck Pressure Calibrator as a Trouble Shooting Tool Presented by Druck
Use of a Druck Pressure Calibrator as a Trouble Shooting Tool Presented by DruckUse of a Druck Pressure Calibrator as a Trouble Shooting Tool Presented by Druck
Use of a Druck Pressure Calibrator as a Trouble Shooting Tool Presented by Druck
 
Electrical Measurement Safety and Best Practices Webinar Presented by Fluke
Electrical Measurement Safety and Best Practices Webinar Presented by FlukeElectrical Measurement Safety and Best Practices Webinar Presented by Fluke
Electrical Measurement Safety and Best Practices Webinar Presented by Fluke
 
Solar Market Opportunities & Solutions - Presented by Fluke
Solar Market Opportunities & Solutions - Presented by FlukeSolar Market Opportunities & Solutions - Presented by Fluke
Solar Market Opportunities & Solutions - Presented by Fluke
 
Pinpoint the leading causes of downtime with LinkIQ Cable and Network Tester
Pinpoint the leading causes of downtime with LinkIQ Cable and Network TesterPinpoint the leading causes of downtime with LinkIQ Cable and Network Tester
Pinpoint the leading causes of downtime with LinkIQ Cable and Network Tester
 
How to Maintain Your Temperature Calibration Equipment Webinar
How to Maintain Your Temperature Calibration Equipment WebinarHow to Maintain Your Temperature Calibration Equipment Webinar
How to Maintain Your Temperature Calibration Equipment Webinar
 
Identifying Energy Waste in your Plant/Facility Webinar with Fluke
Identifying Energy Waste in your Plant/Facility Webinar with FlukeIdentifying Energy Waste in your Plant/Facility Webinar with Fluke
Identifying Energy Waste in your Plant/Facility Webinar with Fluke
 
Principals of Low Resistance Testing Webinar with Megger
Principals of Low Resistance Testing Webinar with MeggerPrincipals of Low Resistance Testing Webinar with Megger
Principals of Low Resistance Testing Webinar with Megger
 
Leak Detection & Predictive/Preventative Maintenance Solutions Presented by FLIR
Leak Detection & Predictive/Preventative Maintenance Solutions Presented by FLIRLeak Detection & Predictive/Preventative Maintenance Solutions Presented by FLIR
Leak Detection & Predictive/Preventative Maintenance Solutions Presented by FLIR
 
Elevated Body Temperature Measurement with IR Thermometers and TI Cameras Web...
Elevated Body Temperature Measurement with IR Thermometers and TI Cameras Web...Elevated Body Temperature Measurement with IR Thermometers and TI Cameras Web...
Elevated Body Temperature Measurement with IR Thermometers and TI Cameras Web...
 
Field Pressure Calibration Tips Webinar with Fluke Calibration
Field Pressure Calibration Tips Webinar with Fluke CalibrationField Pressure Calibration Tips Webinar with Fluke Calibration
Field Pressure Calibration Tips Webinar with Fluke Calibration
 
Temperature Screening Solutions and Returning to Work Safely Webinar with Ame...
Temperature Screening Solutions and Returning to Work Safely Webinar with Ame...Temperature Screening Solutions and Returning to Work Safely Webinar with Ame...
Temperature Screening Solutions and Returning to Work Safely Webinar with Ame...
 
Using a Portable Bath to Improve Calibration of Sanitary RTD's and Transmitte...
Using a Portable Bath to Improve Calibration of Sanitary RTD's and Transmitte...Using a Portable Bath to Improve Calibration of Sanitary RTD's and Transmitte...
Using a Portable Bath to Improve Calibration of Sanitary RTD's and Transmitte...
 
Elevated Skin Temperature Screening Solutions – How to Get Back to Work Safel...
Elevated Skin Temperature Screening Solutions – How to Get Back to Work Safel...Elevated Skin Temperature Screening Solutions – How to Get Back to Work Safel...
Elevated Skin Temperature Screening Solutions – How to Get Back to Work Safel...
 
Elevated Skin Temperature Screening - How to Get Back to Work Safely with FLIR
Elevated Skin Temperature Screening - How to Get Back to Work Safely with FLIRElevated Skin Temperature Screening - How to Get Back to Work Safely with FLIR
Elevated Skin Temperature Screening - How to Get Back to Work Safely with FLIR
 
Overhead & Underground Electric Safety Webinar Presented by Amprobe
Overhead & Underground Electric Safety Webinar Presented by AmprobeOverhead & Underground Electric Safety Webinar Presented by Amprobe
Overhead & Underground Electric Safety Webinar Presented by Amprobe
 
Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...
Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...
Tips for Calibrating Analog Pressure Gauges Webinar Presented by Fluke Calibr...
 
Blind Spots of Low Voltage Testing Webinar Presented by Megger Baker
Blind Spots of Low Voltage Testing Webinar Presented by Megger BakerBlind Spots of Low Voltage Testing Webinar Presented by Megger Baker
Blind Spots of Low Voltage Testing Webinar Presented by Megger Baker
 

Recently uploaded

Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Safe Software
 
Why Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire businessWhy Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire business
panagenda
 
Architecting Cloud Native Applications
Architecting Cloud Native ApplicationsArchitecting Cloud Native Applications
Architecting Cloud Native Applications
WSO2
 

Recently uploaded (20)

Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
 
How to Troubleshoot Apps for the Modern Connected Worker
How to Troubleshoot Apps for the Modern Connected WorkerHow to Troubleshoot Apps for the Modern Connected Worker
How to Troubleshoot Apps for the Modern Connected Worker
 
Why Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire businessWhy Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire business
 
Architecting Cloud Native Applications
Architecting Cloud Native ApplicationsArchitecting Cloud Native Applications
Architecting Cloud Native Applications
 
AWS Community Day CPH - Three problems of Terraform
AWS Community Day CPH - Three problems of TerraformAWS Community Day CPH - Three problems of Terraform
AWS Community Day CPH - Three problems of Terraform
 
Repurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost Saving
Repurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost SavingRepurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost Saving
Repurposing LNG terminals for Hydrogen Ammonia: Feasibility and Cost Saving
 
WSO2's API Vision: Unifying Control, Empowering Developers
WSO2's API Vision: Unifying Control, Empowering DevelopersWSO2's API Vision: Unifying Control, Empowering Developers
WSO2's API Vision: Unifying Control, Empowering Developers
 
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWEREMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
 
Vector Search -An Introduction in Oracle Database 23ai.pptx
Vector Search -An Introduction in Oracle Database 23ai.pptxVector Search -An Introduction in Oracle Database 23ai.pptx
Vector Search -An Introduction in Oracle Database 23ai.pptx
 
FWD Group - Insurer Innovation Award 2024
FWD Group - Insurer Innovation Award 2024FWD Group - Insurer Innovation Award 2024
FWD Group - Insurer Innovation Award 2024
 
Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...
Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...
Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...
 
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
 
Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...
Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...
Apidays New York 2024 - The Good, the Bad and the Governed by David O'Neill, ...
 
MINDCTI Revenue Release Quarter One 2024
MINDCTI Revenue Release Quarter One 2024MINDCTI Revenue Release Quarter One 2024
MINDCTI Revenue Release Quarter One 2024
 
Corporate and higher education May webinar.pptx
Corporate and higher education May webinar.pptxCorporate and higher education May webinar.pptx
Corporate and higher education May webinar.pptx
 
Rising Above_ Dubai Floods and the Fortitude of Dubai International Airport.pdf
Rising Above_ Dubai Floods and the Fortitude of Dubai International Airport.pdfRising Above_ Dubai Floods and the Fortitude of Dubai International Airport.pdf
Rising Above_ Dubai Floods and the Fortitude of Dubai International Airport.pdf
 
Exploring Multimodal Embeddings with Milvus
Exploring Multimodal Embeddings with MilvusExploring Multimodal Embeddings with Milvus
Exploring Multimodal Embeddings with Milvus
 
Apidays New York 2024 - Passkeys: Developing APIs to enable passwordless auth...
Apidays New York 2024 - Passkeys: Developing APIs to enable passwordless auth...Apidays New York 2024 - Passkeys: Developing APIs to enable passwordless auth...
Apidays New York 2024 - Passkeys: Developing APIs to enable passwordless auth...
 
Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...
Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...
Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...
 
Understanding the FAA Part 107 License ..
Understanding the FAA Part 107 License ..Understanding the FAA Part 107 License ..
Understanding the FAA Part 107 License ..
 

How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015

  • 1. How to Calibrate an RTD Using a Dry-Block Calibrator Kurt Zabriskie Fluke Calibration
  • 2. Fluke – American Fork, UT • Kurt Zabriskie • 17 Years with Fluke/Hart Scientific – Western Regional Product Manager - Temperature • Kurt.Zabriskie@flukecal.com
  • 3. Agenda How to Calibrate an RTD Using a Dry-Block Calibrator • Quick Dry-block introduction • Dry-block sources of error • Finding solutions for common errors and problems • Calculating uncertainties • How to calibrate probes with odd shapes and sizes • Temperature range concerns • Liquid-in-glass thermometers • RTD calibration example • Dry-block maintenance • Summary
  • 4. Dry-Block Introduction Dry-Blocks/Dry-Wells  Moderate accuracy  Fixed hole diameter  Fixed immersion depth  Dry and clean  Portable  Faster temperature changes  Internal reference probe 9100S Handheld Dry-well 9190A Ultra-Cool Field Metrology Well 914X Series Field Metrology Wells 9103 9140 9141 Dry-Block Calibrators
  • 5. Dry-Block Sources of Errors • Immersion effects • Stem conduction • Well contact errors • Display accuracy • Stability • Uniformity (radial & axial) • Loading
  • 6. Dry-Block – Immersion Effects • Immersion depth of the UUT is critical • Dry-blocks are generally calibrated by fully inserting a reference probe
  • 7. Dry-Block – Immersion Effects • By adding a reference probe, we can be less concerned with the control probe location • The UUT and Reference can be moved to different depths • Immersion depth should still be considered • (15 X Probe Diameter) • Many Fluke dry-blocks include a reference Input
  • 8. Dry-Block – Stem Conduction • May impact the ability to make a good measurement • The probe sheath acts as heat sink • Probe diameter matters, length is less critical • (15 x probe diameter) a good rule to follow Heat
  • 9. Dry-Block – Well Contact Errors • Fit is important • Loose fitting probes exhibit erratic behavior • Wells should be 0.005”- 0.010” larger than the diameter of the probe for 0.5” diameter probes and smaller. • Too snug, and the probe may become stuck
  • 10. Dry-Block – Display Accuracy • Generally a one year specification on a spec sheet • Internal control sensors are designed to be robust • Display accuracy can be one of the largest contributors to the overall uncertainty
  • 11. Dry-Block – Stability • Check dry-block specifications • Units generally come with optimized proportional bands • Fine tuning the proportional band may help at specific temperatures
  • 12. Dry-Block – Axial Uniformity • Axial Uniformity – Variation in the temperature along the axial length of the insert (block) within the measurement zone • Generally inherent to the dry-block • The closer the sensing element is to ambient air, the larger the uncertainties • Higher temperatures present larger errors • Look for dry-blocks with a calibrated zone for optimal results
  • 13. Dry-Block – Radial Uniformity • Radial Uniformity – Variation in the temperature between different wells of the insert (block) within the measurement zone. • Mostly inherent to the dry-block design • Heater placement and profiling is critical during engineering of the dry- block
  • 14. Dry-Block – Loading • Loading can impact uncertainties. • Some models have a specification • Units without specifications can be evaluated in the field
  • 15. Calculating Uncertainties • RSS method is generally used to calculate uncertainties when using a reference 2 3 2 2 2 1 )()()( bbbbtotal  222 )()()( uniformitystabilityrefbtotal 
  • 16. Calculating Uncertainties • GUM compliant example -38 °C 0 °C 157 °C 232 °C 420 °C 660 °C Uncertainty Sources: Type mK mK mK mK mK mK Process Var. (check std) Norm 3.0 3.0 5.0 5.0 5.0 6.0 UUT Precision (noise) Norm 2.8 2.8 2.8 3.3 4.4 5.6 Ref. Precision (noise) Norm 2.8 2.8 2.8 2.8 2.8 5.6 Ref. Calibration Norm 0.2 0.1 0.6 0.5 0.6 1.1 Ref. Drift Rect 1.7 2.0 3.2 3.8 5.1 6.7 Radial Uniformity Rect 10.0 10.0 30.0 30.0 30.0 30.0 Axial Uniformity Rect 6.3 6.3 6.3 6.3 6.3 6.3 Readout (SPRT) Rect 0.2 0.3 0.4 0.5 0.7 0.9 Readout (UUT) Rect 1.3 1.5 2.5 3.1 4.4 6.3 Insulation Leakage Rect 10. 0 10.0 10.0 10.0 10.0 10.0 UUT Repeatability (TPW) Norm 2.8 3.3 5.3 6.3 8.6 11.2 Total (k=2): 20.6 21.0 40.7 41.5 43.9 47.9
  • 17. Dry-Block – Odd Shaped Probes • Custom inserts are an option • Micro-Baths might be a good option to explore
  • 18. Temperature Range Concerns • –100 to 1200 ºC • May need to use multiple units • It’s ok to switch from one dry-block to the next Model 9190A: –95 to 140 ºC Model 9143: 33 to 350 ºC Model 9150: 150 to 1200 ºC
  • 19. Dry-Block – LIG’s • Liquid and glass thermometers are not recommended for use in a dry-block calibrator • Mercury thermometers are on their way out due to environmental concerns
  • 20. Dry-Block – Maintenance • Keep those wells clean, Scotch-Brite pads and a gun cleaning kit work nicely • Avoid dropping inserts or other heavy objects into the well • Avoid the use of thermo grease • Recalibrate regularly • Verify stability
  • 21. RTD Calibration Example • Three point RTD Calibration • –95 ºC, 0 ºC, 140 ºC • Utilize the 9190A with the “Process Option” to measure the UUT • Set the 9190A to each of the three set points, generally starting with the lowest point • Allow for plenty of soak time at each temperature (15 minutes) • Record a resistance at each set point • Utilize at tool such as TableWare to calculate coefficients
  • 23. When the Super-DAQ is connected to a Fluke Calibration dry-well, fluid bath, or furnace, it can control the temperature source to calibrate up to 40 sensors automatically. You simply program the setpoint temperatures and their values, select a scan sequence, assign a reference channel, and set the required stability band. The Super-DAQ monitors the temperature source’s stability through the reference channel, collects the data from the reference probe and the “unit under test” (UUT) once stabilized, and then advances to the next set-point temperature. After you configure and start the test, you can walk away to work on other things. The Super- DAQ just made your day a whole lot easier! Automate temperature sensor calibration with the 1586A Super-DAQ 9190A Ultra-Cool Field Metrology Well 1586A Super-DAQ with DAQ-STAQ
  • 24. Application Note and Video Application Note: Automating Temperature Sensor Calibration with the 1586A Super-DAQ Companion Video: Automating Temperature Sensor Calibration with the 1586A Super-DAQ The application note and video demonstrate the Auto Test function of the 1586A using a 9142 Field Metrology which can be substituted with other Fluke Calibration dry-wells, fluid baths, and furnaces.
  • 25. Dry-Block – Summary • Dry-blocks are a great option in many situations • Highly portable and quick to change temperature • Sources of error should be considered • On board references are a great way to combine several instruments into one • Be sure to contact us any time for help with your specific application
  • 26. Where Can I Find Out More? • Comparison Calibration Application Notes – 914X Series Field Metrology Well – 9190A Ultra-Cool Field Metrology Well • Product information at: www.Transcat.com/Fluke
  • 27. Limited Time Offer • Free Temperature Probe with Metrology Well Purchase • Order from Transcat • Details at: www.transcat.com/Fluke • 800-800-5001
  • 28. Questions or Comments? Email the moderator, Nicole VanWert: nvanwert@Transcat.com Transcat: 800-800-5001 www.Transcat.com For related product information, go to: www.Transcat.com/Fluke