Twin ions
1. Measuring the Mass of
an Electron by LC/TOF-
MS
A Study of “Twin Ions”Imma Ferrer & E. Michael Thurman
Department of Analytical Chemistry, University of Almerıa,
Anal. Chem. 2005, 77, 3394-3400
1
2. This Paper Discusses………
The use of LC/MS ion trap to show the formation of fragment
ions of the same nominal mass in positive and negative ion and
their pathway
The use of LC/TOF-MS to measure the accurate mass of
fragment ions and the importance of including the mass of the
electron in ion fragmentation studies
Recent advances in continuous mass calibration allow the
measurement of the electron to one significant figure using
LC/TOF-MS, when combined with the analysis of the NSAID
solution
2
3. How small is an electron?
• Mass : 9.1093829×10−31 kg / 0.00055 u
• Charge : −1.602176565×10−19 C
3
4. Twin Ions
Same fragment ion (nominal mass) formed in either positive or
negative ion electrospray for a suite of analyte
4
5. What is NSAID?
Non steroidal anti inflammatory drugs
Analgesic and antipyretic effects
In higher doses anti inflammatory
Non narcotic
E.g : Aspirin , Ibuprofen ,Naproxen, Diclofenac
5
6. LC / MS Ion Trap Method
• Column : Reverse phase (C8)
• Column temperature : 25 °C
• Mobile phase
A - acetonitrile
B - 0.1% formic acid.
• Linear gradient flow :50 - 100% A in 10 min further maintained for 2
min
• The flow rate :0.6 mL /min
• Injection volume :50 uL
• Maximum ion accumulation
time : 200 ms
• Capillary voltage :3200 V
6
7. LC/TOF-MS Method
• Same LC parameters
• Electrospray interfaced
• Positive ion mode
• Capillary Voltage : 4000 V
• Gas temperature : 300 °C
• A second orthogonal sprayer with a reference solution was
used as a continuous calibration in positive ion (121.0509 &
922.0098 m/z)
7
9. Inductive (i) cleavage of ibuprofen [M + H]+ or [M - H]-
to form the twin ions in either positive or negative ion electrospray
Results & Discussion
9
14. Calculation of Electron Mass from the Difference of the
Twin Ions Based on Each Chromatographic Run
(Negative - Positive) Containing Ibuprofen, Diclofenac,
and Naproxen Using Hand
14
16. Finally…….
It is now possible to reach accuracy values of 1 ppm regularly for small
molecules (which is 0.2 mmu for a mass of m/z 200) using LC/TOF-MS.
The maximum calculated mass precision for the Agilent TOF used in this
study was 0.4 ppm based on the m/z 161 ion using a resolution of 5 000
and an ion intensity of 105 (20 000 ions)
Similar calculations for the m/z 185 ion and m/z 250 ion gave maximum
precision values of 0.2 and 0.3 ppm, respectively
The combination of twin information and routine accuracy of <1 ppm
makes it possible to measure electron mass to one significant figure.
16