SlideShare a Scribd company logo
1 of 1
Download to read offline
Background
Tuning Electrochemical Impedance Spectroscopy
Chi Lin1, David Probst1, Aldin Malkoc1, Jeffrey T. La Belle¹,²,5
1Tempe, Arizona, Arizona State University, School of Biological and Health Systems Engineering; 2 Tempe, Arizona, Arizona State University, School for Engineering Matter, Transport and Energy; 3 Tempe, Arizona, Arizona
State University, School of Electrical, Computer and Energy Engineering;4 Raleigh, North Carolina, Advanced Tear Diagnostics, LLC;5Scottsdale, Arizona, Mayo Clinic Arizona, School of Medicine
- Aliasing is the overlap of two output signals that cause noise to one
another
- Once a signal is aliased it can not be undone or rectified
- Electrochemical impedance spectroscopy (EIS) has become a more
prominent technique for analyzing biomarkers over the last few years
- The largest advantage with EIS is its possibility to measure two or more
biomarkers simultaneously and accurately
- One road block to this achievement is the aliasing of similar biomarker
signals
Functionalization and Techniques
Figure 1: This figure shows the aliasing of two
biomarkers, 1,5 AG with glucose. This is slope
(Ohm/(mg/dl)) versus frequency. The overlap of
signal occurs from 10 Hz up to 1000Hz.
Immobilization onto GDE’s using Il-12 and Nanoparticles
Bare Electrode
MHDA treatment
EDC/NHS application
Biotinylated Il-12 Antibody
Streptavidin nanoparticle
Ethanolamine
Il-12 Antigen
Figure 2: Schematic of immobilization process, the first step is a self assembling
monolayer, MHDA. Then MHDA is activated by EDC/NHS, which in turn antibody binds
to the MHDA. Using streptavidin nanoparticles the nanoparticles were conjugated to the
antibody using biotin-strepaviden interactions
Quantitative Results
Figure 6: The conjugation of quantum dots
to Il-12 antibody shifted the characteristic
peak a magnitude larger, but had low
sensitivity, as well has a very large full width
at half maximum. As compared to untuned,
the quantum dots loss nearly an entire
magnitude of sensitivity.
Figure 3: Overlays of all 4 Nanoparticles
and untuned Il-12. The untuned IL-12
seems to peak at 17 Hz, where all 4
Nanoparticles shift the peak either up or
down. This shift seems to have a direct
correlation to the size of nanoparticle. own.
Discussion
- By analyzing the imaginary impedance rather then complex the
shape of the Slope vs Frequency graph is no longer a low/high
pass filter, but a band pass exhibiting a particular peak
characteristic to that measured marker
- By conjugating nanoparticles to Il-12 the optimal frequency has
shifted
- There was a consistent inverse correlation between the diameter
of the nanoparticle and its effect on the protein properties
- This may be attributed to the surface area of the particle, with
5nm there is a much lower surface area, giving the IL-12 the
ability and room to bond with more particles
- With the bonding to more particles, there is a greater magnitude
of full width at half-maximum
- Quantum dots showed a large shift in frequency, but a much
lower response and sensitivity
Future Work
- Further exploring the electrochemical effects of conjugating
nanoparticles to different types of biomarkers (enzymes)
- Conjugating different types of nanoparticle's to several markers and
measuring them simultaneously by applying several frequencies
unique to a particular analyte to measure
Figure 8: Schematic
of three biomarkers
conjugated to three
different
nanoparticles
References
Acknowledgements
Many thanks to the entire La Belle Group Lab for support, especially the TEIS teams.
[1] J.T. La Belle, et al Methods 61 (2013), 31-59
[2] J.T. La Belle, et al. Analyst, 2011, 136, 1496.
[3] T.L. Adamson, et al. Journal of Diabetes Science and Technology. 8 (2014), 350-355.
Figure 5: This graph shows the relationship
between the diameter of the nanoparticle
and its affect on frequency as well as full
width at half maximum (FWHM). As seen
above the smaller the diameter in
nanoparticle, the greater the shift in
frequency the equation for the relationship
is: y = 259.85e-0.247x. This also holds true
for the full band width, as the diameter
decreases, the full band width increases
the relationship between these two can be
best fit by the following line: y = 805.55e-
0.218x. Both R^2 correlations were about .98
showing a strong relationship between the
size and electrochemical attributes. N=3 for
diameter 5 nm and 20 nm, and N=5 for
diameter of 10 nm.
Figure 4: Overlays of the complex islope
versus frequency. Unlike imaginary
impedance, there is no obvious distinct
peak. Also every frequency below 75 Hz
has aliasing, covering other wanted
signals.
Impedance of AC Circuit
- Z(w) = Z’(w)+Z’’(w)
𝑍 𝑤 = 𝑍°(cos ∅ + 𝑗𝑠𝑖𝑛(∅)
- Z(w) is the Complex
Impedance
- Z’’(w) is the Imaginary
Impedance or is equal to
 𝑍° 𝑗𝑠𝑖𝑛 ∅
- Looking at imaginary
versus total complex
impedance gives us
unique peak previously
not observed
Nano Particle Frequency Shift (Hz) Full Width at Half Maximum
5 Nm Gold 64.2 295.06
10 Nm Gold 4.05 79.946
20 Nm Gold -14.09 10.726
20 nm QD 441.82 2001.5
Table 1: (below) Shows the frequency shift
caused by the addition of each
nanoparticle, as well as the full band
width. As the diameter of the nanoparticle
increases, the shift decreases, to the point
where there is a negative shift, relative to
untuned IL-12
Figure 7: Applying
several frequencies
to measure
simultaneous
markers
0
100
200
300
400
500
600
700
1 10 100 1000 10000 100000
Slope(Ohm/(pg/ml))
Frequency (Hz)
Slope Verses Frequency Il-12 unconjugated
Quantum Dot Slope Versus Frequency
10 Nm GolD Nanoparticle Slope Versus
Frequency
5 Nm Gold Nanoparticle Slope Versus
Frequency
20 Nm Gold Nanoparticle Slope Versus
Frequency
Unconjugated Il-12 Slope Versus
Frequency
-140
-120
-100
-80
-60
-40
-20
0
1 10 100 1000 10000 100000
Slope(Ohm/(pg/ml))
Frequency (Hz)
Imaginary Slope Versus Frequency
10 nm Gold Nano Particle n=4
20 nm Quantum Dot n=3
5 nm Gold Nano Particle n=3
20 nm Gold Nano Particle n=2
IL-12 Untuned Imaginary Slope
-20
-10
0
10
20
30
40
50
60
70
0
50
100
150
200
250
300
350
5 10 20
FrequencyShift(Hz)
FullWidthatHalfMaximum
Diameter of Nanoparticle (nm)
Comparison of Nanoparticel Daimeter and Frequency
Gold Nanoparticle Diameter Versus
Full Band Width
Gold Nano Particle Diameter Versus
Frequency Shift
-2
-1.8
-1.6
-1.4
-1.2
-1
-0.8
-0.6
-0.4
-0.2
0
-120
-100
-80
-60
-40
-20
0
1 10 100 1000 10000 100000
Slope(Ohm/(pg/ml))
Slope(Ohm/(pg/ml))
Frequency
Imaginary Slope Versus Frequency
IL-12 Untuned Slope
20 Nm Quantum Dot
Slope

More Related Content

What's hot

Effective Dosage Utilizing a Therapeutic Laser
Effective Dosage Utilizing a Therapeutic LaserEffective Dosage Utilizing a Therapeutic Laser
Effective Dosage Utilizing a Therapeutic LaserRob Berman
 
Interaction of x-ray and radiation effects
 Interaction of x-ray and radiation effects Interaction of x-ray and radiation effects
Interaction of x-ray and radiation effectsCobain Nirvan
 
PARAMETRIC STUDY OF MICROWAVE ABSORPTION IN LOSSY DIELECTRIC SLABS
PARAMETRIC STUDY OF MICROWAVE ABSORPTION IN LOSSY DIELECTRIC SLABSPARAMETRIC STUDY OF MICROWAVE ABSORPTION IN LOSSY DIELECTRIC SLABS
PARAMETRIC STUDY OF MICROWAVE ABSORPTION IN LOSSY DIELECTRIC SLABSjmicro
 
Electron paramagnetic resonance(epr) spectroscopy
Electron paramagnetic resonance(epr) spectroscopyElectron paramagnetic resonance(epr) spectroscopy
Electron paramagnetic resonance(epr) spectroscopyHalavath Ramesh
 
Effect of isotopic subsitution on the transition frequencies
Effect of isotopic subsitution on the transition frequenciesEffect of isotopic subsitution on the transition frequencies
Effect of isotopic subsitution on the transition frequenciesApurvaSachdeva
 
Factors are affecting the chemical shift values nmr spectra
Factors are affecting the chemical shift values   nmr spectraFactors are affecting the chemical shift values   nmr spectra
Factors are affecting the chemical shift values nmr spectraganapathimurugan1
 
X ray attenuation
X ray attenuationX ray attenuation
X ray attenuationaslam bs
 

What's hot (8)

Meph 569 second major 032
Meph 569 second major 032Meph 569 second major 032
Meph 569 second major 032
 
Effective Dosage Utilizing a Therapeutic Laser
Effective Dosage Utilizing a Therapeutic LaserEffective Dosage Utilizing a Therapeutic Laser
Effective Dosage Utilizing a Therapeutic Laser
 
Interaction of x-ray and radiation effects
 Interaction of x-ray and radiation effects Interaction of x-ray and radiation effects
Interaction of x-ray and radiation effects
 
PARAMETRIC STUDY OF MICROWAVE ABSORPTION IN LOSSY DIELECTRIC SLABS
PARAMETRIC STUDY OF MICROWAVE ABSORPTION IN LOSSY DIELECTRIC SLABSPARAMETRIC STUDY OF MICROWAVE ABSORPTION IN LOSSY DIELECTRIC SLABS
PARAMETRIC STUDY OF MICROWAVE ABSORPTION IN LOSSY DIELECTRIC SLABS
 
Electron paramagnetic resonance(epr) spectroscopy
Electron paramagnetic resonance(epr) spectroscopyElectron paramagnetic resonance(epr) spectroscopy
Electron paramagnetic resonance(epr) spectroscopy
 
Effect of isotopic subsitution on the transition frequencies
Effect of isotopic subsitution on the transition frequenciesEffect of isotopic subsitution on the transition frequencies
Effect of isotopic subsitution on the transition frequencies
 
Factors are affecting the chemical shift values nmr spectra
Factors are affecting the chemical shift values   nmr spectraFactors are affecting the chemical shift values   nmr spectra
Factors are affecting the chemical shift values nmr spectra
 
X ray attenuation
X ray attenuationX ray attenuation
X ray attenuation
 

Viewers also liked

Basics of Electrochemical Impedance Spectroscopy
Basics of Electrochemical Impedance SpectroscopyBasics of Electrochemical Impedance Spectroscopy
Basics of Electrochemical Impedance SpectroscopyGamryInstruments
 
electrochemical impedence spectroscopy
electrochemical impedence spectroscopyelectrochemical impedence spectroscopy
electrochemical impedence spectroscopyAmol Srivastava
 
Impedance parameters
Impedance parametersImpedance parameters
Impedance parametersAhmad Nauman
 
Impedance Spectroscopy
Impedance Spectroscopy Impedance Spectroscopy
Impedance Spectroscopy bisquertGroup
 
Fundamentals of Photovoltaics: Lecture 1
Fundamentals of Photovoltaics: Lecture 1Fundamentals of Photovoltaics: Lecture 1
Fundamentals of Photovoltaics: Lecture 1University of Liverpool
 
Electrochemical impedance spectroscopy
Electrochemical impedance spectroscopyElectrochemical impedance spectroscopy
Electrochemical impedance spectroscopysonmezz
 

Viewers also liked (7)

Basics of Electrochemical Impedance Spectroscopy
Basics of Electrochemical Impedance SpectroscopyBasics of Electrochemical Impedance Spectroscopy
Basics of Electrochemical Impedance Spectroscopy
 
electrochemical impedence spectroscopy
electrochemical impedence spectroscopyelectrochemical impedence spectroscopy
electrochemical impedence spectroscopy
 
S parameters
S parametersS parameters
S parameters
 
Impedance parameters
Impedance parametersImpedance parameters
Impedance parameters
 
Impedance Spectroscopy
Impedance Spectroscopy Impedance Spectroscopy
Impedance Spectroscopy
 
Fundamentals of Photovoltaics: Lecture 1
Fundamentals of Photovoltaics: Lecture 1Fundamentals of Photovoltaics: Lecture 1
Fundamentals of Photovoltaics: Lecture 1
 
Electrochemical impedance spectroscopy
Electrochemical impedance spectroscopyElectrochemical impedance spectroscopy
Electrochemical impedance spectroscopy
 

Similar to BMES2015 TEIS rev 4 (1)

Variation of dose distribution with depth and incident energy using EGSnrc Mo...
Variation of dose distribution with depth and incident energy using EGSnrc Mo...Variation of dose distribution with depth and incident energy using EGSnrc Mo...
Variation of dose distribution with depth and incident energy using EGSnrc Mo...iosrjce
 
Development of fpga based system for neutron flux monitoring in fast breeder ...
Development of fpga based system for neutron flux monitoring in fast breeder ...Development of fpga based system for neutron flux monitoring in fast breeder ...
Development of fpga based system for neutron flux monitoring in fast breeder ...Alexander Decker
 
Basic Concepts of UV & IR Spectroscopy
Basic Concepts of UV & IR SpectroscopyBasic Concepts of UV & IR Spectroscopy
Basic Concepts of UV & IR SpectroscopyDrBasavarajaiahSm
 
BMES2015_ATD_Poster
BMES2015_ATD_PosterBMES2015_ATD_Poster
BMES2015_ATD_PosterGarrett Repp
 
Voltage Transients and Health - Is There a Connection?
Voltage Transients and Health - Is There a Connection?Voltage Transients and Health - Is There a Connection?
Voltage Transients and Health - Is There a Connection?Mikko Ahonen
 
15.30 o4 c aguergaray
15.30 o4 c aguergaray15.30 o4 c aguergaray
15.30 o4 c aguergarayNZIP
 
Development of A 3d Model Eddy Current Testing
Development of A 3d Model Eddy Current TestingDevelopment of A 3d Model Eddy Current Testing
Development of A 3d Model Eddy Current Testinginventionjournals
 
Performance analysis on the basis of a comparative study between multipath ra...
Performance analysis on the basis of a comparative study between multipath ra...Performance analysis on the basis of a comparative study between multipath ra...
Performance analysis on the basis of a comparative study between multipath ra...ijmnct
 
Bio-heat Transfer in Various Transcutaneous Stimulation Models
Bio-heat Transfer in Various Transcutaneous Stimulation ModelsBio-heat Transfer in Various Transcutaneous Stimulation Models
Bio-heat Transfer in Various Transcutaneous Stimulation ModelsTrevor Davis
 
Visualizing Radiation Physics Concepts with photon electron particle tracks
Visualizing Radiation Physics Concepts with photon electron particle tracksVisualizing Radiation Physics Concepts with photon electron particle tracks
Visualizing Radiation Physics Concepts with photon electron particle tracksShahid Naqvi
 
Internal multiple attenuation using inverse scattering: Results from prestack...
Internal multiple attenuation using inverse scattering: Results from prestack...Internal multiple attenuation using inverse scattering: Results from prestack...
Internal multiple attenuation using inverse scattering: Results from prestack...Arthur Weglein
 
My Presentation @Ryerson University
My Presentation @Ryerson UniversityMy Presentation @Ryerson University
My Presentation @Ryerson UniversityJagdish Bhatt
 

Similar to BMES2015 TEIS rev 4 (1) (20)

1a characterization
1a characterization1a characterization
1a characterization
 
Variation of dose distribution with depth and incident energy using EGSnrc Mo...
Variation of dose distribution with depth and incident energy using EGSnrc Mo...Variation of dose distribution with depth and incident energy using EGSnrc Mo...
Variation of dose distribution with depth and incident energy using EGSnrc Mo...
 
Development of fpga based system for neutron flux monitoring in fast breeder ...
Development of fpga based system for neutron flux monitoring in fast breeder ...Development of fpga based system for neutron flux monitoring in fast breeder ...
Development of fpga based system for neutron flux monitoring in fast breeder ...
 
Basic Concepts of UV & IR Spectroscopy
Basic Concepts of UV & IR SpectroscopyBasic Concepts of UV & IR Spectroscopy
Basic Concepts of UV & IR Spectroscopy
 
BMES2015_ATD_Poster
BMES2015_ATD_PosterBMES2015_ATD_Poster
BMES2015_ATD_Poster
 
THz Plasmonics
THz PlasmonicsTHz Plasmonics
THz Plasmonics
 
Voltage Transients and Health - Is There a Connection?
Voltage Transients and Health - Is There a Connection?Voltage Transients and Health - Is There a Connection?
Voltage Transients and Health - Is There a Connection?
 
Publication
PublicationPublication
Publication
 
15.30 o4 c aguergaray
15.30 o4 c aguergaray15.30 o4 c aguergaray
15.30 o4 c aguergaray
 
ac400907q.pdf
ac400907q.pdfac400907q.pdf
ac400907q.pdf
 
B1104010612
B1104010612B1104010612
B1104010612
 
wep153
wep153wep153
wep153
 
Development of A 3d Model Eddy Current Testing
Development of A 3d Model Eddy Current TestingDevelopment of A 3d Model Eddy Current Testing
Development of A 3d Model Eddy Current Testing
 
EM Term Paper
EM Term PaperEM Term Paper
EM Term Paper
 
Performance analysis on the basis of a comparative study between multipath ra...
Performance analysis on the basis of a comparative study between multipath ra...Performance analysis on the basis of a comparative study between multipath ra...
Performance analysis on the basis of a comparative study between multipath ra...
 
Bio-heat Transfer in Various Transcutaneous Stimulation Models
Bio-heat Transfer in Various Transcutaneous Stimulation ModelsBio-heat Transfer in Various Transcutaneous Stimulation Models
Bio-heat Transfer in Various Transcutaneous Stimulation Models
 
Visualizing Radiation Physics Concepts with photon electron particle tracks
Visualizing Radiation Physics Concepts with photon electron particle tracksVisualizing Radiation Physics Concepts with photon electron particle tracks
Visualizing Radiation Physics Concepts with photon electron particle tracks
 
Internal multiple attenuation using inverse scattering: Results from prestack...
Internal multiple attenuation using inverse scattering: Results from prestack...Internal multiple attenuation using inverse scattering: Results from prestack...
Internal multiple attenuation using inverse scattering: Results from prestack...
 
Coates weglein-1996
Coates weglein-1996Coates weglein-1996
Coates weglein-1996
 
My Presentation @Ryerson University
My Presentation @Ryerson UniversityMy Presentation @Ryerson University
My Presentation @Ryerson University
 

BMES2015 TEIS rev 4 (1)

  • 1. Background Tuning Electrochemical Impedance Spectroscopy Chi Lin1, David Probst1, Aldin Malkoc1, Jeffrey T. La Belle¹,²,5 1Tempe, Arizona, Arizona State University, School of Biological and Health Systems Engineering; 2 Tempe, Arizona, Arizona State University, School for Engineering Matter, Transport and Energy; 3 Tempe, Arizona, Arizona State University, School of Electrical, Computer and Energy Engineering;4 Raleigh, North Carolina, Advanced Tear Diagnostics, LLC;5Scottsdale, Arizona, Mayo Clinic Arizona, School of Medicine - Aliasing is the overlap of two output signals that cause noise to one another - Once a signal is aliased it can not be undone or rectified - Electrochemical impedance spectroscopy (EIS) has become a more prominent technique for analyzing biomarkers over the last few years - The largest advantage with EIS is its possibility to measure two or more biomarkers simultaneously and accurately - One road block to this achievement is the aliasing of similar biomarker signals Functionalization and Techniques Figure 1: This figure shows the aliasing of two biomarkers, 1,5 AG with glucose. This is slope (Ohm/(mg/dl)) versus frequency. The overlap of signal occurs from 10 Hz up to 1000Hz. Immobilization onto GDE’s using Il-12 and Nanoparticles Bare Electrode MHDA treatment EDC/NHS application Biotinylated Il-12 Antibody Streptavidin nanoparticle Ethanolamine Il-12 Antigen Figure 2: Schematic of immobilization process, the first step is a self assembling monolayer, MHDA. Then MHDA is activated by EDC/NHS, which in turn antibody binds to the MHDA. Using streptavidin nanoparticles the nanoparticles were conjugated to the antibody using biotin-strepaviden interactions Quantitative Results Figure 6: The conjugation of quantum dots to Il-12 antibody shifted the characteristic peak a magnitude larger, but had low sensitivity, as well has a very large full width at half maximum. As compared to untuned, the quantum dots loss nearly an entire magnitude of sensitivity. Figure 3: Overlays of all 4 Nanoparticles and untuned Il-12. The untuned IL-12 seems to peak at 17 Hz, where all 4 Nanoparticles shift the peak either up or down. This shift seems to have a direct correlation to the size of nanoparticle. own. Discussion - By analyzing the imaginary impedance rather then complex the shape of the Slope vs Frequency graph is no longer a low/high pass filter, but a band pass exhibiting a particular peak characteristic to that measured marker - By conjugating nanoparticles to Il-12 the optimal frequency has shifted - There was a consistent inverse correlation between the diameter of the nanoparticle and its effect on the protein properties - This may be attributed to the surface area of the particle, with 5nm there is a much lower surface area, giving the IL-12 the ability and room to bond with more particles - With the bonding to more particles, there is a greater magnitude of full width at half-maximum - Quantum dots showed a large shift in frequency, but a much lower response and sensitivity Future Work - Further exploring the electrochemical effects of conjugating nanoparticles to different types of biomarkers (enzymes) - Conjugating different types of nanoparticle's to several markers and measuring them simultaneously by applying several frequencies unique to a particular analyte to measure Figure 8: Schematic of three biomarkers conjugated to three different nanoparticles References Acknowledgements Many thanks to the entire La Belle Group Lab for support, especially the TEIS teams. [1] J.T. La Belle, et al Methods 61 (2013), 31-59 [2] J.T. La Belle, et al. Analyst, 2011, 136, 1496. [3] T.L. Adamson, et al. Journal of Diabetes Science and Technology. 8 (2014), 350-355. Figure 5: This graph shows the relationship between the diameter of the nanoparticle and its affect on frequency as well as full width at half maximum (FWHM). As seen above the smaller the diameter in nanoparticle, the greater the shift in frequency the equation for the relationship is: y = 259.85e-0.247x. This also holds true for the full band width, as the diameter decreases, the full band width increases the relationship between these two can be best fit by the following line: y = 805.55e- 0.218x. Both R^2 correlations were about .98 showing a strong relationship between the size and electrochemical attributes. N=3 for diameter 5 nm and 20 nm, and N=5 for diameter of 10 nm. Figure 4: Overlays of the complex islope versus frequency. Unlike imaginary impedance, there is no obvious distinct peak. Also every frequency below 75 Hz has aliasing, covering other wanted signals. Impedance of AC Circuit - Z(w) = Z’(w)+Z’’(w) 𝑍 𝑤 = 𝑍°(cos ∅ + 𝑗𝑠𝑖𝑛(∅) - Z(w) is the Complex Impedance - Z’’(w) is the Imaginary Impedance or is equal to  𝑍° 𝑗𝑠𝑖𝑛 ∅ - Looking at imaginary versus total complex impedance gives us unique peak previously not observed Nano Particle Frequency Shift (Hz) Full Width at Half Maximum 5 Nm Gold 64.2 295.06 10 Nm Gold 4.05 79.946 20 Nm Gold -14.09 10.726 20 nm QD 441.82 2001.5 Table 1: (below) Shows the frequency shift caused by the addition of each nanoparticle, as well as the full band width. As the diameter of the nanoparticle increases, the shift decreases, to the point where there is a negative shift, relative to untuned IL-12 Figure 7: Applying several frequencies to measure simultaneous markers 0 100 200 300 400 500 600 700 1 10 100 1000 10000 100000 Slope(Ohm/(pg/ml)) Frequency (Hz) Slope Verses Frequency Il-12 unconjugated Quantum Dot Slope Versus Frequency 10 Nm GolD Nanoparticle Slope Versus Frequency 5 Nm Gold Nanoparticle Slope Versus Frequency 20 Nm Gold Nanoparticle Slope Versus Frequency Unconjugated Il-12 Slope Versus Frequency -140 -120 -100 -80 -60 -40 -20 0 1 10 100 1000 10000 100000 Slope(Ohm/(pg/ml)) Frequency (Hz) Imaginary Slope Versus Frequency 10 nm Gold Nano Particle n=4 20 nm Quantum Dot n=3 5 nm Gold Nano Particle n=3 20 nm Gold Nano Particle n=2 IL-12 Untuned Imaginary Slope -20 -10 0 10 20 30 40 50 60 70 0 50 100 150 200 250 300 350 5 10 20 FrequencyShift(Hz) FullWidthatHalfMaximum Diameter of Nanoparticle (nm) Comparison of Nanoparticel Daimeter and Frequency Gold Nanoparticle Diameter Versus Full Band Width Gold Nano Particle Diameter Versus Frequency Shift -2 -1.8 -1.6 -1.4 -1.2 -1 -0.8 -0.6 -0.4 -0.2 0 -120 -100 -80 -60 -40 -20 0 1 10 100 1000 10000 100000 Slope(Ohm/(pg/ml)) Slope(Ohm/(pg/ml)) Frequency Imaginary Slope Versus Frequency IL-12 Untuned Slope 20 Nm Quantum Dot Slope