A customer needed to measure temperatures in an immersion tank with an accuracy of ±0.3 °C. The standard probe had an accuracy of ±0.31 °C, exceeding the requirement. Testo AG customized the probe by replacing its sensor with a more precise one, achieving an accuracy of ±0.07 °C. With the instrument's accuracy of ±0.2 °C, the customized probe provided a system accuracy of ±0.27 °C, meeting the customer's needs through a tailored probe solution.