QUASAR programmable respiratory motion platform has
three modes of operation in the platform: oscillation mode,
rotation mode and position mode.
●System is controlled by software
●Motions are programmable
●Provides superior-inferior direction
●Not accurate representation of respiratory motion
●Not precise motion tracking of lung cancer tumors
● Create a completely wireless system
● 2nd round of prototyping
● Improve machining process and materials
● Complete tumor tracking studies
● Create system to allow user to input patient data
There will be three designs for this platform, including: cylindrical insert,
wall system, and bed system.
Bed-system provides superior - inferior direction (y-direction),
including:
○ Thomson rods and bearings
○ Wood bed
○ Nema-17-stepper motor
○ Belt
○ Provide a motion in AP direction for the tumor
○ Maximum moving distance of 8.0 cm. (6.0 cm is required)
Tumor holder
○ Threaded rod will connect with a motor through a coupling
and provide the SI motion
○ The smooth rod will secure a system and prevent any tremor
○ Both rods will place through two holes designed in the tumor
holder
○ Hollow space in one end with a distance of 2.0 cm depth to
Wall-system provides anterior - posterior direction (z-direction),
including:
○ Thomson rod
○ Wood wall
○ Nema-17-stepper motor
○ Tumor holder
○ Be secure perpendicular with a wood bed in a bed system
Abstract
Existing Approach
Project Purpose
Design
Background
Testing and Results
QUASAR Programmable Respiratory Motion Tumor Insert
Daniel Shin; An Tran; Jessica Traub; Kunyu Wu
Faculty Advisor: Dr. Iris Wang, PhD and Zheng Zheng Xu
Department of Biomedical Engineering, State University of New York at Buffalo
Roswell Park Cancer Institute
Acknowledgements
References
• https://www.roswellpark.org/newsroom/roswell-park-logos
• Modulus Medical Devices Inc., Programmable Respiratory Motion Platform - User's Guide
• Cella, David F., Amy E. Bonomi, Stephen R. Lloyd, David S. Tulsky, Edward Kaplan, and Philip Bonomi. "Reliability
and Validity of the Functional Assessment of Cancer Therapy—lung (FACT-L) Quality of Life Instrument."
ScienceDirect 12, no. 3 (June 1995): 199-220. Accessed March 22, 2016. http://www.sciencedirect.com/science/
article/pii/016950029500450F.
• "Lung Cancer." NIHSeniorHealth:. N.p., n.d. Web. 04 Nov. 2015. <http://nihseniorhealth.gov/lungcancer/
lungcancerdefined/01.html>.
• Howlader N, Noone AM, Krapcho M, Garshell J, Miller D, Altekruse SF, Kosary CL, Yu M, Ruhl J, Tatalovich
Z,Mariotto A, Lewis DR, Chen HS, Feuer EJ, Cronin KA (eds). SEER Cancer Statistics Review, 1975-2012, National
Cancer Institute. Bethesda, MD, http://seer.cancer.gov/csr/1975_2012/, based on November 2014 SEER data
submission, posted to the SEER web site, April 2015.
• "CT Quality Assurance Protocol and Its Validation in Various Ministry of Health Hospitals in Oman." World Congress
on Medical Physics and Biomedical Engineering. Ed. Dossel Olaf and Schlegel Wolfgang. 1st ed. Vol. 25/3. Springer,
2009. 76. Print.
• "Staging of Lung Cancer." Patient Information Series. American Thoracic Society Patient Education Series, 9 Sept.
2013. Web. 4 Nov. 2015.
Future Work and Other Applications
Figure 5: Insert 3.5. The hollow space
allow to motion in range of 0 - 3.5
cm. The tumor holder will be able to
move in SI direction with a maximum
distance of 3.5 cm.
Lung and bronchus cancer is a serious and prominent disease
compared to other cancers. Lung cancer is difficult to be treated
without any side effects. Currently, the QUASAR Programmable
Respiratory Motion Phantom is designed to simulate the breathing
motion of a patient, intended to increase the overall testing capabilities
for radiation therapy of lung cancer research. This existing approach
improves the precision and accuracy of such radiation therapies but
only provides one direction, Superior-Inferior (SI). The goal of this
project is to design a tumor insert in order to simulate breathing motion
in the Anterior-Posterior (AP) direction and in the Superior and Inferior
(SI) direction. This design is beneficial to achieve greater precision
and accuracy for tracking lung cancer tumors for radiation therapy
using the unidirectional QUASAR respiratory motion phantom as well
as a way to improve patient treatment.
Non-small cell lung cancer(NSCLC)
●Common but grow and spread faster
●Difficult to detect in early stage
Small cell lung cancer (SCLC)
●Limited Stage - Occur in one lung
●Extensive Stage - Spread to the other lung
Four different stages of NSCLC based on TNM
system
●Stage I - Cancer may be present in the underlying
lung tissues, but the lymph nodes remain unaffected
●Stage II - Cancer may have spread to nearby lymph
nodes
●Stage III - Cancer is continuing to spread from the
lungs to the lymph nodes or to nearby structures
and organs
●Stage IV - Cancer has metastasized throughout the
body and may now affect the liver, bones or brain
Figure 2: Four different stages of non-small cell lung cancer
Design a platform
! Can support around 3 pounds
! Be able to move in anterior - posterior direction (z-direction)
! Be able to move in superior - inferior direction (y-direction)
! Made of material that does not cause noise and scattering of the
X- Ray photons
! Material should have similar attenuation coefficient number to
lung
! Be able to adjust the amplitude and frequency of the motion
! Be able to transport
Project Goals
! Have similar breathe rate to humans for tumor tracking research
(2D)
! Produce anterior - posterior direction (z-direction) motion
! Produce superior - inferior direction (y-direction) motion
! Can be tested under CT machine
! Can be programmed with various amplitudes and frequencies
! Can easily be transported
Software Programs
Figure 7: User Interface - user can adjust the amplitude (A) and period (T)
● Technologies: Image-guided radiation therapy (IGRT) and
respiratory gating
● Use body imaging techniques and high beam radiation to track
and treat the tumor in the body
● To improve the precision and accuracy of radiation therapies:
QUASAR Programmable Respiratory Motion Phantom
○ Simulates the breathing motion of a patient which increases the
overall testing capabilities for radiation therapy
○ Contains a body oval, a drive unit, a respiratory phantom mass,
cylindrical cedar inserts, a chest wall platform and power supply
○ Designed for dosimetric and non-dosimetric measurement while
testing a moving object
○ Contains motion of a cylindrical cedar insert moving in the
Superior and Inferior (SI) direction [1]
○ Chest wall platform works as a third-party motion tracking
system and is linked to the QUASAR phantom
○ Only unidirectional motion. Breathing motion of the lungs are in
3D
• Dimensions of the bed system is satisfied the requirement.
• Tumor holder - Dowel round wood(1-⅛’’) - Reduced machining
cost.
• Cylindrical insert - Laser CNC slices of cork adhered by E6000
glue to achieve complete insert. It satisfied the diameter of the
cylindrical insert space QUASAR. It satisfied the space
requirement for the tumor holder for motion in the Y-direction
and Z-direction.
• Material Testing: Both tumor holders satisfied the CT test seen
in figure 8. One was coated with nail lacquer and the other was
not. They were compared to the existing QUASAR insert
material.
• Software Testing: Each motor was tested individually and
together for their accuracy. The period and amplitude were
measured manually. The results of each trial are seen in table 1.
Preliminary Sketch
Figure 4:
Programmable Motion
Platform. This image
shows a better view of
a whole system,
including: cylindrical
insert, wall system
and bed system.
Our prototype
Cylindrical insert
○ Be the same size as the QUASAR’s insert
○ Provide a space through all the insert for the tumor to move
○ Hollow space will fit the size of a tumor holder
○ Maximum IS motion: 1.0 - 4.0 cm
○ Average IS motion: 2.0 - 3.0 cm
○ Period - 6 seconds
Figure 3: QUASAR
Programmable
Respiratory Platform
Figure 1: QUASAR Programmable
Respiratory Phantom
Table 1: Amplitude and Period Testing and Results
A B
Figure 6: (A) Tumor Insert System. Side View. Finished design with both z-
direction and y-direction, and tumor holder. (B) Tumor Insert System. Corner
view of completed design.
Interface Trial
Failure/
Success
Reason for
Failure
Interface Trial
Failure/
Success
Reason for
Failure
Up
1 F Exceed input
Z -
Amplitude
1 S N/A
2 S N/A 2 S N/A
3 S N/A 3 S N/A
Down
1 F Exceed input
Y -
Amplitude
1 S N/A
2 S N/A 2 S N/A
3 S N/A 3 S N/A
Left
1 F Exceed input
Z - Time
Delay
1 S N/A
2 S N/A 2 S N/A
3 S N/A 3 S N/A
Right
1 F Exceed input
Y – Time
Delay
1 S N/A
2 S N/A 2 S N/A
3 S N/A 3 S N/A
Z – Time
Period
1 F Exceed input
Y – Time
Period
1 F Exceed input
2 F Exceed input 2 F Exceed input
3 F Exceed input 3 F Exceed input
4 S N/A 4 S N/A
Both
Motor
1 S N/A
2 S N/A
1. The Arduino program is uploaded to the Arduino
Board. The function of the code is to control the motor
through the driver.
2. The user interface is written in the Processing
Environment.
3. On the interface, there are three sections to control
the motor in the Y-direction and the Z-direction. Each
motor has two text boxes to let the user input
the amplitude and time period. Running both motors
allows the user input a time delay for either motor. For
each motor, there are two buttons: ‘run’ and ‘stop’. Run
starts the code and the movement of the motor. Stop
causes the motors to terminate movement. There is a
clock and there are controls to adjust the Z-direction up
and down and the Y-direction left and right.
Figure 8: CT Testing
Results. HU Profile
showing both the
tumor holder material
with and with out nail
lacquer compared to
original QUASAR
insert material.

More Related Content

What's hot

Posada_Daniel_2015_memoire
Posada_Daniel_2015_memoirePosada_Daniel_2015_memoire
Posada_Daniel_2015_memoire
Daniel Posada
 
MRI screening procedures
MRI screening proceduresMRI screening procedures
MRI screening procedures
Ahmed Bahnassy
 
Spom Imaging
Spom ImagingSpom Imaging
Spom Imaging
Claraha
 

What's hot (20)

Oncology imaging
Oncology imagingOncology imaging
Oncology imaging
 
Posada_Daniel_2015_memoire
Posada_Daniel_2015_memoirePosada_Daniel_2015_memoire
Posada_Daniel_2015_memoire
 
Radiology Basics
Radiology BasicsRadiology Basics
Radiology Basics
 
Introduction to radiology
Introduction to radiologyIntroduction to radiology
Introduction to radiology
 
Ultasound
UltasoundUltasound
Ultasound
 
YOU CAN NEVER DETECT A DISEASE WITHOUT RADIOLOGY
YOU CAN NEVER DETECT A DISEASE WITHOUT RADIOLOGYYOU CAN NEVER DETECT A DISEASE WITHOUT RADIOLOGY
YOU CAN NEVER DETECT A DISEASE WITHOUT RADIOLOGY
 
Advanced Diagnostic Center (radiology): An Introduction-DIAGNOSTIC RADIOLOGY ...
Advanced Diagnostic Center (radiology): An Introduction-DIAGNOSTIC RADIOLOGY ...Advanced Diagnostic Center (radiology): An Introduction-DIAGNOSTIC RADIOLOGY ...
Advanced Diagnostic Center (radiology): An Introduction-DIAGNOSTIC RADIOLOGY ...
 
Robotic surgery
Robotic surgeryRobotic surgery
Robotic surgery
 
Motion management techniques in RT
Motion management techniques in RT Motion management techniques in RT
Motion management techniques in RT
 
CT Scan: Purpose, Procedure and Risk Factors
CT Scan: Purpose, Procedure and Risk FactorsCT Scan: Purpose, Procedure and Risk Factors
CT Scan: Purpose, Procedure and Risk Factors
 
Respiration motion management
Respiration motion managementRespiration motion management
Respiration motion management
 
MRI screening procedures
MRI screening proceduresMRI screening procedures
MRI screening procedures
 
Radiology instruments
Radiology instrumentsRadiology instruments
Radiology instruments
 
Imaging in cancer therapeutics 2015
Imaging in cancer therapeutics 2015Imaging in cancer therapeutics 2015
Imaging in cancer therapeutics 2015
 
Radiation diagnostics diseases of the brain and spinal cord
Radiation diagnostics diseases of the brain and spinal cord Radiation diagnostics diseases of the brain and spinal cord
Radiation diagnostics diseases of the brain and spinal cord
 
Good mri information
Good mri informationGood mri information
Good mri information
 
Ct brain basics and anatomy
Ct brain basics and anatomyCt brain basics and anatomy
Ct brain basics and anatomy
 
15min teaching demo for consideration as an adjunct professor at a local Univ...
15min teaching demo for consideration as an adjunct professor at a local Univ...15min teaching demo for consideration as an adjunct professor at a local Univ...
15min teaching demo for consideration as an adjunct professor at a local Univ...
 
Spom Imaging
Spom ImagingSpom Imaging
Spom Imaging
 
Presentation1estrada
Presentation1estradaPresentation1estrada
Presentation1estrada
 

Similar to Senior Capstone Project_Final_Poster

med_010_03_030928
med_010_03_030928med_010_03_030928
med_010_03_030928
gapras94
 

Similar to Senior Capstone Project_Final_Poster (20)

Robotic surgery in ENT
Robotic surgery in ENTRobotic surgery in ENT
Robotic surgery in ENT
 
7200 35328
7200 353287200 35328
7200 35328
 
Protable ultrasound
Protable ultrasoundProtable ultrasound
Protable ultrasound
 
Protable ultrasound
Protable ultrasoundProtable ultrasound
Protable ultrasound
 
Radiology_Equipment_Lec-1_Dr. Emad Taleb.pdf
Radiology_Equipment_Lec-1_Dr. Emad Taleb.pdfRadiology_Equipment_Lec-1_Dr. Emad Taleb.pdf
Radiology_Equipment_Lec-1_Dr. Emad Taleb.pdf
 
Ultrasound in the 21st Century: Why Carestream Health's Touch Ultrasounds Mak...
Ultrasound in the 21st Century: Why Carestream Health's Touch Ultrasounds Mak...Ultrasound in the 21st Century: Why Carestream Health's Touch Ultrasounds Mak...
Ultrasound in the 21st Century: Why Carestream Health's Touch Ultrasounds Mak...
 
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundStudy: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
 
Cyberknife
Cyberknife Cyberknife
Cyberknife
 
Elective ii
Elective iiElective ii
Elective ii
 
Ultrasound-1.pptx
Ultrasound-1.pptxUltrasound-1.pptx
Ultrasound-1.pptx
 
Cone Beam Computed Tomography.pptx
Cone Beam Computed Tomography.pptxCone Beam Computed Tomography.pptx
Cone Beam Computed Tomography.pptx
 
Basics of Ultrasound
Basics of Ultrasound Basics of Ultrasound
Basics of Ultrasound
 
med_010_03_030928
med_010_03_030928med_010_03_030928
med_010_03_030928
 
Patient data acquisition and treatment verification.pptx
Patient data acquisition and treatment verification.pptxPatient data acquisition and treatment verification.pptx
Patient data acquisition and treatment verification.pptx
 
Ultrasound Machine-A Revolution In Medical Imaging
Ultrasound Machine-A Revolution In Medical ImagingUltrasound Machine-A Revolution In Medical Imaging
Ultrasound Machine-A Revolution In Medical Imaging
 
Ct scan and its interpretation in omfs
Ct scan and its interpretation in omfsCt scan and its interpretation in omfs
Ct scan and its interpretation in omfs
 
SBRT in lung cancer
SBRT in lung cancerSBRT in lung cancer
SBRT in lung cancer
 
Application of Automation in different sectors
Application of Automation in different sectorsApplication of Automation in different sectors
Application of Automation in different sectors
 
behgal cancer institute
behgal cancer  institute behgal cancer  institute
behgal cancer institute
 
Lung Conditions Prognosis Using CNN Model.pptx
Lung Conditions Prognosis Using CNN Model.pptxLung Conditions Prognosis Using CNN Model.pptx
Lung Conditions Prognosis Using CNN Model.pptx
 

Senior Capstone Project_Final_Poster

  • 1. www.buffalo.edu QUASAR programmable respiratory motion platform has three modes of operation in the platform: oscillation mode, rotation mode and position mode. ●System is controlled by software ●Motions are programmable ●Provides superior-inferior direction ●Not accurate representation of respiratory motion ●Not precise motion tracking of lung cancer tumors ● Create a completely wireless system ● 2nd round of prototyping ● Improve machining process and materials ● Complete tumor tracking studies ● Create system to allow user to input patient data There will be three designs for this platform, including: cylindrical insert, wall system, and bed system. Bed-system provides superior - inferior direction (y-direction), including: ○ Thomson rods and bearings ○ Wood bed ○ Nema-17-stepper motor ○ Belt ○ Provide a motion in AP direction for the tumor ○ Maximum moving distance of 8.0 cm. (6.0 cm is required) Tumor holder ○ Threaded rod will connect with a motor through a coupling and provide the SI motion ○ The smooth rod will secure a system and prevent any tremor ○ Both rods will place through two holes designed in the tumor holder ○ Hollow space in one end with a distance of 2.0 cm depth to Wall-system provides anterior - posterior direction (z-direction), including: ○ Thomson rod ○ Wood wall ○ Nema-17-stepper motor ○ Tumor holder ○ Be secure perpendicular with a wood bed in a bed system Abstract Existing Approach Project Purpose Design Background Testing and Results QUASAR Programmable Respiratory Motion Tumor Insert Daniel Shin; An Tran; Jessica Traub; Kunyu Wu Faculty Advisor: Dr. Iris Wang, PhD and Zheng Zheng Xu Department of Biomedical Engineering, State University of New York at Buffalo Roswell Park Cancer Institute Acknowledgements References • https://www.roswellpark.org/newsroom/roswell-park-logos • Modulus Medical Devices Inc., Programmable Respiratory Motion Platform - User's Guide • Cella, David F., Amy E. Bonomi, Stephen R. Lloyd, David S. Tulsky, Edward Kaplan, and Philip Bonomi. "Reliability and Validity of the Functional Assessment of Cancer Therapy—lung (FACT-L) Quality of Life Instrument." ScienceDirect 12, no. 3 (June 1995): 199-220. Accessed March 22, 2016. http://www.sciencedirect.com/science/ article/pii/016950029500450F. • "Lung Cancer." NIHSeniorHealth:. N.p., n.d. Web. 04 Nov. 2015. <http://nihseniorhealth.gov/lungcancer/ lungcancerdefined/01.html>. • Howlader N, Noone AM, Krapcho M, Garshell J, Miller D, Altekruse SF, Kosary CL, Yu M, Ruhl J, Tatalovich Z,Mariotto A, Lewis DR, Chen HS, Feuer EJ, Cronin KA (eds). SEER Cancer Statistics Review, 1975-2012, National Cancer Institute. Bethesda, MD, http://seer.cancer.gov/csr/1975_2012/, based on November 2014 SEER data submission, posted to the SEER web site, April 2015. • "CT Quality Assurance Protocol and Its Validation in Various Ministry of Health Hospitals in Oman." World Congress on Medical Physics and Biomedical Engineering. Ed. Dossel Olaf and Schlegel Wolfgang. 1st ed. Vol. 25/3. Springer, 2009. 76. Print. • "Staging of Lung Cancer." Patient Information Series. American Thoracic Society Patient Education Series, 9 Sept. 2013. Web. 4 Nov. 2015. Future Work and Other Applications Figure 5: Insert 3.5. The hollow space allow to motion in range of 0 - 3.5 cm. The tumor holder will be able to move in SI direction with a maximum distance of 3.5 cm. Lung and bronchus cancer is a serious and prominent disease compared to other cancers. Lung cancer is difficult to be treated without any side effects. Currently, the QUASAR Programmable Respiratory Motion Phantom is designed to simulate the breathing motion of a patient, intended to increase the overall testing capabilities for radiation therapy of lung cancer research. This existing approach improves the precision and accuracy of such radiation therapies but only provides one direction, Superior-Inferior (SI). The goal of this project is to design a tumor insert in order to simulate breathing motion in the Anterior-Posterior (AP) direction and in the Superior and Inferior (SI) direction. This design is beneficial to achieve greater precision and accuracy for tracking lung cancer tumors for radiation therapy using the unidirectional QUASAR respiratory motion phantom as well as a way to improve patient treatment. Non-small cell lung cancer(NSCLC) ●Common but grow and spread faster ●Difficult to detect in early stage Small cell lung cancer (SCLC) ●Limited Stage - Occur in one lung ●Extensive Stage - Spread to the other lung Four different stages of NSCLC based on TNM system ●Stage I - Cancer may be present in the underlying lung tissues, but the lymph nodes remain unaffected ●Stage II - Cancer may have spread to nearby lymph nodes ●Stage III - Cancer is continuing to spread from the lungs to the lymph nodes or to nearby structures and organs ●Stage IV - Cancer has metastasized throughout the body and may now affect the liver, bones or brain Figure 2: Four different stages of non-small cell lung cancer Design a platform ! Can support around 3 pounds ! Be able to move in anterior - posterior direction (z-direction) ! Be able to move in superior - inferior direction (y-direction) ! Made of material that does not cause noise and scattering of the X- Ray photons ! Material should have similar attenuation coefficient number to lung ! Be able to adjust the amplitude and frequency of the motion ! Be able to transport Project Goals ! Have similar breathe rate to humans for tumor tracking research (2D) ! Produce anterior - posterior direction (z-direction) motion ! Produce superior - inferior direction (y-direction) motion ! Can be tested under CT machine ! Can be programmed with various amplitudes and frequencies ! Can easily be transported Software Programs Figure 7: User Interface - user can adjust the amplitude (A) and period (T) ● Technologies: Image-guided radiation therapy (IGRT) and respiratory gating ● Use body imaging techniques and high beam radiation to track and treat the tumor in the body ● To improve the precision and accuracy of radiation therapies: QUASAR Programmable Respiratory Motion Phantom ○ Simulates the breathing motion of a patient which increases the overall testing capabilities for radiation therapy ○ Contains a body oval, a drive unit, a respiratory phantom mass, cylindrical cedar inserts, a chest wall platform and power supply ○ Designed for dosimetric and non-dosimetric measurement while testing a moving object ○ Contains motion of a cylindrical cedar insert moving in the Superior and Inferior (SI) direction [1] ○ Chest wall platform works as a third-party motion tracking system and is linked to the QUASAR phantom ○ Only unidirectional motion. Breathing motion of the lungs are in 3D • Dimensions of the bed system is satisfied the requirement. • Tumor holder - Dowel round wood(1-⅛’’) - Reduced machining cost. • Cylindrical insert - Laser CNC slices of cork adhered by E6000 glue to achieve complete insert. It satisfied the diameter of the cylindrical insert space QUASAR. It satisfied the space requirement for the tumor holder for motion in the Y-direction and Z-direction. • Material Testing: Both tumor holders satisfied the CT test seen in figure 8. One was coated with nail lacquer and the other was not. They were compared to the existing QUASAR insert material. • Software Testing: Each motor was tested individually and together for their accuracy. The period and amplitude were measured manually. The results of each trial are seen in table 1. Preliminary Sketch Figure 4: Programmable Motion Platform. This image shows a better view of a whole system, including: cylindrical insert, wall system and bed system. Our prototype Cylindrical insert ○ Be the same size as the QUASAR’s insert ○ Provide a space through all the insert for the tumor to move ○ Hollow space will fit the size of a tumor holder ○ Maximum IS motion: 1.0 - 4.0 cm ○ Average IS motion: 2.0 - 3.0 cm ○ Period - 6 seconds Figure 3: QUASAR Programmable Respiratory Platform Figure 1: QUASAR Programmable Respiratory Phantom Table 1: Amplitude and Period Testing and Results A B Figure 6: (A) Tumor Insert System. Side View. Finished design with both z- direction and y-direction, and tumor holder. (B) Tumor Insert System. Corner view of completed design. Interface Trial Failure/ Success Reason for Failure Interface Trial Failure/ Success Reason for Failure Up 1 F Exceed input Z - Amplitude 1 S N/A 2 S N/A 2 S N/A 3 S N/A 3 S N/A Down 1 F Exceed input Y - Amplitude 1 S N/A 2 S N/A 2 S N/A 3 S N/A 3 S N/A Left 1 F Exceed input Z - Time Delay 1 S N/A 2 S N/A 2 S N/A 3 S N/A 3 S N/A Right 1 F Exceed input Y – Time Delay 1 S N/A 2 S N/A 2 S N/A 3 S N/A 3 S N/A Z – Time Period 1 F Exceed input Y – Time Period 1 F Exceed input 2 F Exceed input 2 F Exceed input 3 F Exceed input 3 F Exceed input 4 S N/A 4 S N/A Both Motor 1 S N/A 2 S N/A 1. The Arduino program is uploaded to the Arduino Board. The function of the code is to control the motor through the driver. 2. The user interface is written in the Processing Environment. 3. On the interface, there are three sections to control the motor in the Y-direction and the Z-direction. Each motor has two text boxes to let the user input the amplitude and time period. Running both motors allows the user input a time delay for either motor. For each motor, there are two buttons: ‘run’ and ‘stop’. Run starts the code and the movement of the motor. Stop causes the motors to terminate movement. There is a clock and there are controls to adjust the Z-direction up and down and the Y-direction left and right. Figure 8: CT Testing Results. HU Profile showing both the tumor holder material with and with out nail lacquer compared to original QUASAR insert material.