SlideShare a Scribd company logo
1 of 71
Download to read offline
LINES AND TUBES IN CHEST X RAY
Dr.T.V.S.S.SAI MADHURI
• Many medical devices are used in intensive care unit for long durations.
• Each of them is a double edged sword:intended to save life ,but life
threatening if its in wrong place.
• All catheters have the potential risks of coiling, misplacement, knotting,
and fracture.
• So it is essential for a radiologist to know about the normal position and
malpositions of devices and its associated complications.
• The American College of Radiology (ACR) recommends a CXR immediately
following placement of indwelling tubes, catheters and other devices to
check the position and detect procedure related complications
Commonly used tubes and lines in ICU
Nasogastric tube Pacemaker
Endotracheal tube Pulmonary (swan-ganz)catheter
Tracheostomy tube Intra aortic ballon pump
Central venous line Automated implantable cardioverter
defibrillator
Drainage tube Paediatric lines
Nasogastric tube
Indications :1. Feeding purpose
2. Aspiration of contents
The NG tube has multiple side holes.
There are terminal lead balls to
facilitate identification of the tip.
CORRECT POSITION : NG tube tip should be
10 cm distal to gastro esophageal
junction
MALPOSITIONS OF NG TUBE AND ITS COMPLICATIONS
• In the oesophagus: If the side holes are positioned within the
esophagus there is increased risk of aspiration
• In the trachea or bronchus: Inadvertent insertion into the
trachea and bronchus can cause pneumonia, pulmonary
contusion, or pulmonary laceration.
• Pharyngeal and esophageal perforations can occur but are
rare.
Frontal (A) and lateral (B) radiographs of the neck showigh a NG tube coiled in
the upper esophagus with its tip in
the oropharynx .
Frontal radiograph of the chest shows a NG tube forming a
loop in the left bronchus before the tip reaches the right lower lobe bronchus.
Tube misplacement in right main bronchus
The tube follows the course of the right main bronchus Its tip is projected over the
lower zone of the right lung The NG tube has been inhaled rather than swallowed .
The tube must be removed and repositioned
Endotracheal tube
The endotracheal (ET) tube is inserted for
ventilation of both the lungs and for prevention
of aspiration.
Correct position: ET tube in the neutral position
of the neck is with the tip 5–7 cm above the
carina.
When the carina is not visible, the tip of the ET
tube should be approximately at the level of the
medial ends of the clavicle
It should lie midway between the larynx and
carina so that injury to either structure or
complications like inadvertent extubation or
selective main stem bronchus intubation are
avoided.
Malpositions of ET tube and its complications
• Selective intubation :It can cause collapse of the contralateral lung and
hyperinflation of the ipsilateral lung, or pneumothorax.
• Selective intubation can be clinically occult in about 60% of patients and
only revealed on the CXR,so an immediate CXR after intubation is
warranted as these complications are not uncommon.
• Inadvertent esophageal intubation :It can be detected radiographically by
the presence of an over distended stomach.
• Tracheal stenosis can occur following long-term tube placement.
• Aspiration of tooth can also occur as a complication
Frontal chest radiographs show an endotracheal tube
in the right main bronchus (arrowhead in A), causing hyperinflation of
the ipsilateral lung and partial collapse of the left lung
Frontal radiograph of a neonate shows inadvertent placement
of an endotracheal tube in the esophagus with distension of
the esophagus and stomach with air
TRACHEOSTOMY TUBE
• Tracheostomy tubes, are inserted
through a stoma post-tracheostomy to
help patients unable to breathe
normally.
INDICATIONS OF TRACHEOSTOMY TUBE
• Prolonged ventilator dependence.
• Prophylactic tracheostomy prior to
head and neck cancer treatment.
• Obstructive sleep apnea refractory to
other treatments.
• Chronic aspiration.
• Neuromuscular disease.
• Subglottic stenosis.
CORRECT POSITION:
The tip of the tracheostomy tube should be half way between
the stoma and the carina, at the level of the D3 vertebra.
It should lie parallel to the trachea.
COMPLICATIONS OF TRACHEOSTOMY
• Hemorrhage
• Malposition
• False tract
• Tracheal stenosis
• Surgical emphysema
• Pneumo mediastinum
• Pneumothorax
Frontal chest radiograph shows complications of tracheostomy:
pneumothorax , pneumo mediastinum , and surgical
emphysema
CENTRAL VENOUS LINES
Central venous lines (catheters)
are useful for a variety of
purposes:
• Hemodynamic pressure
monitoring;
• Hemodialysis
• Administration of medications,
nutrition, and fluids.
• Correct positioning of a CVC tip depends on the side of entry
and intended use of the catheter
• For most short-term uses, such as fluid administration or
monitoring of central venous pressure -Positioning the tip of
a central venous catheter within the superior vena cava at
or just above the level of the carina is generally considered
• For the purpose of long term chemotherapy –the tip of cvc
may be placed more inferiorly at the cavo-atrial junction -
the junction of the SVC and right atrium .
• Superior vena cava (SVC) anatomy
• The internal jugular and subclavian veins join to form the
brachiocephalic veins.
• The brachiocephalic veins join to form the SVC(corresponds
to thethe first anterior intercostal space)
• The SVC is located to the right side of the mediastinum above
and below the level of the carina.
• SVC is not visible on a chest X-ray. As the carina is a visible
structure, it can be used as an anatomical landmark to help
determine the level of a CVC tip within the SVC.
• The cavo-atrial junction is located approximately the height of
two vertebral bodies below the level of the carina
VARIOUS APPROACHES FOR CENTRAL
VENOUS LINE
• Right-sided catheters
• CVCs are most commonly inserted via the right
internal jugular vein.
• Right internal jugular catheters are positioned on
the right side of the neck, and pass vertically from
a position above the clavicle.
• This is an ideal position for right-sided catheters
for fluid administration and venous pressure
monitoring, but not for long-term chemotherapy
or dialysis
Right subclavian vein catheter
• Catheters inserted into the subclavian vein pass below the clavicle and
then curve into the SVC.
Left subclavian vein catheter
CV Catheters - Complications
• A chest X-ray taken after central venous catheter placement can
identify immediate complications such as pneumothorax or
pneumomediastinum and to identify incorrect positioning.
• Immediate complications
Complications such as pneumothorax and surgical emphysema may
arise from traumatic placement.
• Incorrect positioning
If catheters are not inserted far enough they may be located in the
jugular, subclavian, or brachiocephalic veins.
Catheters inserted too far may enter the right atrium.
IATROGENIC PNEUMOTHORAX
CVC IN LEFT BRACHIOCEPHALIC
VEIN
The tip of this catheter
is projected over the
left brachiocephalic
vein rather than the
SVC
CATHETER IN RIGHT ATRIUM
This peripherally inserted
central catheter (PICC) was
aimed to be inserted with its tip
at the level of the cavo-atrial
junction (the height of two
vertebral bodies below the
carina)
The PICC has been inserted too
far with its tip in the right atrium
(RA)
HORIZONTAL POSITIONING IN
SVC
Catheters placed via a
left-sided approach are
prone to being positioned
nearly horizontally rather
than vertically within the
SVC.
Catheters which contact
the lateral wall of the SVC
in this way may cause
vessel erosion if positioned
long term, and should
therefore be placed so the
tip is orientated vertically.
INTERNAL JUGULAR
CATHETER
-MISPLACED
This left internal jugular
catheter has entered
the left subclavian vein.
The catheter needs to
be repositioned.
Frontal chest radiograph shows an abnormally medial
course of the catheter in a case of inadvertent carotid
cannulation
INTERCOSTAL DRIANAGE TUBES
• The pleural tube, more commonly known as the intercostal
drainage tube (ICD), is inserted through the 4th intercostal
space in the anterior or mid-axillary line.
• It is then directed posteroinferiorly in cases of effusion and
anterosuperiorly in cases of pneumothorax.
• The ICD tube has a terminal hole as well as side holes; the side
holes can be identified on a CXR by the interruption in the
radio-opaque outline of the tube.
• No side holes should lie outside the chest/pleura
Chest drain-treatment
for pneumothorax
To drain pneumothorax
the tube is positioned
with its tip pointing
superiorly towards the
apex of the pleural
cavity.
CHEST DRAIN - TREATMENT
FOR PLEURAL EFFUSION
To drain pleural effusion
the tube tip is ideally
located towards the
lower part of the
pleural cavity.
MALPOSITIONS OF ICD TUBES
An intercostal drainage
tube is noted in situ. One
hole is situated outside the
rib cage, indicating
Malposition, likely
resulting in insufficient
suction.
Pneumothorax noted on
the left side with a
partially collapsed lung
PULMONARY ARTERY CATHETER(SWAN-GANZ
CATHETER)
• The pulmonary artery catheter, also called a
Swan-Ganz catheter, is used to monitor
circulatory hemodynamics in critically ill
patients by measuring pulmonary capillary
wedge pressure.
• The catheter is inserted via the subclavian vein
or the internal jugular vein.
• The ideal position of the tip of the catheter is in
the right or left main pulmonary arteries, and
the tip should not extend beyond the proximal
interlobar pulmonary artery (within 2 cm of the
hilum)
Swan-Ganz catheter lies in the left pulmonary artery.
COMPLICATIONS OF SWAN GANZ CATHETER
• Occlusion of a pulmonary artery branch may occur if the
catheter is too distal, if there is persistent inflation of the
balloon, or if a clot is formed around the distal tip of the
catheter; such occlusion can result in pulmonary infarction. A
chest radiograph may reveal the infarction as a wedge-shaped
pleural-based pulmonary opacity.
• Other potential complications are intracardiac knotting,
pulmonary artery perforation, arrhythmias, cardiac
perforation, and placement in the inferior vena cava
Anteroposterior chest radiograph shows that tip of catheter is too distal (i.e.,
> 2 cm lateral to hilum). There is wedged-shaped opacity distal to catheter,
consistent with pulmonary infarction.
PACEMAKER
• Pacemakers are used in cases of severe sinus node dysfunction, complete
heart block, and various arrhythmias.
• They have two main elements: a pulse generator and a lead wire with
electrodes.
• The single-lead pacemaker is the most basic type and is positioned with its
tip in the right ventricular apex.
• An atrioventricular two-lead sequential pacemaker has one electrode in
the right atrium and the other at the right ventricular apex .
• Sometimes a third lead is placed in the coronary sinus to pace the left
ventricle.
Frontal chest radiograph (A) shows the optimal position of the electrode of a
single-lead pacemaker. The electrode has been placed in the right ventricular
apex . Frontal chest radiograph (B) shows a two-lead pacemaker that has one
electrode in the right atrium and the other at the right ventricular apex
COMPLICATIONS OF CARDIAC PACEMAKER
• Malposition,
• Intracardiac knotting,
• Fracture,
• Perforation
• Cardiac tamponade,
• Arrhythmias,
• Infection,
• Hemorrhage.
Frontal chest radiograph shows coiling of the lead of a
single-lead pacemaker in the right atrium
Frontal chest radiograph shows abnormal course of the
lead with the electrode tip overlying the liver.This was due
to cardiac perforation by the pacemaker lead,with a fatal outcome
Frontal chest radiograph shows recoil of the pacemaker lead
with its tip in the superior vena cava . This is called Twiddler’s
syndrome
Intra-aortic Balloon Pump
Intra-aortic balloon pump is a long-balloon temporary circulatory assist
device that works on the principle of cardiac counter-pulsation. The IABP is
used to support the circulation.
The balloon, approximately 25-cm long, is mounted on a catheter.
The catheter tip is visible as a 3 x 4-mm rectangular metallic density while
the rest of the catheter is radiolucent.
The catheter is inserted through the femoral artery.
To avoid occlusion of the left subclavian artery and visceral and renal
arteries, its tip should be slightly cephalad to the adjacent carina (2nd–3rd
intercostal space)
• The balloon is inflated with
gas during diastole and
deflates during systole,
resulting in increase in
coronary blood flow and
reduction in left ventricular
after load (and hence,
reduction in myocardial
oxygen consumption).
Normal positioning of intraaortic counter pulsation balloon pump. Magnified
Anteroposterior chest radiograph obtained during systole shows catheter’s
radiopaque tip within upper descending thoracic aorta indicating proper
positioning.
Magnified Anteroposterior chest radiograph obtained during diastole shows
inflated radiolucent balloon as well as radiopaque tip within upper
descending thoracic aorta. Catheter is inflated during diastole to increase
myocardial perfusion and is deflated during systole to decrease left
ventricular after load.
MALPOSITIONED INTRAAORTIC COUNTERPULSATION BALLOON PUMP
Anteroposterior chest radiograph during diastole shows inflated radiolucent
balloon. Radiopaque tip of balloon catheter is abnormally located beyond
aortic arch. It is projected over left common carotid artery. This abnormal
positioning can result in cerebral ischemia.
Malposition intraaortic counter pulsation balloon pump. Anteroposterior
chest radiograph obtained during diastole shows inflated radiolucent balloon
with its radiopaque tip in mid thoracic aorta. Counter pulsation is less
effective when catheter is too low and may result in obstruction of abdominal
aortic branches, such as renal or mesenteric arteries.
Automated Implantable Cardioverter Debrillator
Automated implantable cardioverter
debrillator (AICD) is used in cases of
recurrent refractory ventricular
tachycardia.
It has two electrodes (one electrode in
the right atrium and the other in the
right ventricle).
The lead is wider compared to the
pacemaker lead and has a
‘coiled-spring’ appearance
PACEMAKER VS AUTOMATED IMPLANTABLE CARDIOVERTER
DEBRILLATOR
• Complications are similar to those with
pacemakers and the incidence of
radiographical abnormalities may approach
20%.
Pediatric Lines
• Some catheters are only used in the pediatric
population, for example, the umbilical artery and
venous catheters.
• They are used for vascular access for exchange
transfusion and measurement of blood gases,
pressures, electrolytes, etc.
• The umbilical vein and arteries remain patent for up
to 4–5 days after birth.
Umbilical artery line
The catheter should be passed
through the umbilic artery and enter
the aorta via the internal iliac artery.
In order to avoid placement into
aortic branches, the catheter should
be either in a high position above the
celiac, mesenteric and renal arteries
or in a low position below the inferior
mesenteric artery:
High position: T6-T9
Low position: L3-L5
The high position is advisable since it
leads to less vascular complications
(15).
Umbilical artery line in a
good high position.
Malposition of umbilical
vein line in right portal
vein.
Malposition of umbilical artery line, folded in the
abdominal aorta.
Malposition of umbilical artery line in left iliac artery.
Umbilical vein
catheter
An umbilical vein catheter
should pass through the
umbilical vein into the left
portal vein.
Then through the ductus
venosus into a hepatic vein
and the inferior caval vein
(IVC)
The tip should be
positioned in the IVC at the
level of the diaphragm.
MALPOSITIONS AND ITS COMPLICATIONS
• Low position in the umbilical vein.
• Intrahepatic into the portal venous system, both right and left,
or even into the superior mesenteric or splenic vein.This can
cause thrombosis.
• Perforation of the portal vein can cause haemorrhage or
abscess formation in the liver.
• Position too deep in right atrium or in the left atrium through
a patent foramen ovale or atrial septal defect.
This can lead to cardiac arrhythmias or perforation.
Umbilical vein line positioned in the periphery of the liver through the
right portal vein.
Artery line in the left subclavian artery.
The umbilical vein line is positioned in the umbilical vein and
not deep enough.
The tip of the umbilical vein line is pointed downwards and
probably situated in the mesenteric vein.
Frontal radiograph of the chest and abdomen of a neonate shows the tip of an
umbilical venous catheter in the left atrium; it has passed through a patent
foramen ovale. The tip of the umbilical artery catheter is in the arch of the
aorta (which is undesirable as it is near the origin of the carotid artery)
Thank you

More Related Content

What's hot

Radiograpic views for shoulder joint
Radiograpic views  for shoulder jointRadiograpic views  for shoulder joint
Radiograpic views for shoulder jointGanesan Yogananthem
 
Upper extremity anatomy & positioning
Upper extremity anatomy & positioningUpper extremity anatomy & positioning
Upper extremity anatomy & positioningInfoUtilRT
 
X-RAY PELVIS PROJECTIONS
X-RAY PELVIS PROJECTIONSX-RAY PELVIS PROJECTIONS
X-RAY PELVIS PROJECTIONSJai Kumar
 
Radiological anatomy of mediastinum
Radiological anatomy of  mediastinumRadiological anatomy of  mediastinum
Radiological anatomy of mediastinumAthul Nampoothiri
 
Hepatic arterial anatomy and vascular optimization final
Hepatic arterial anatomy and vascular optimization finalHepatic arterial anatomy and vascular optimization final
Hepatic arterial anatomy and vascular optimization finalpryce27
 
PTBD (Percutaneus trans-hepatic biliary drainage) PPT,PDF PK
PTBD (Percutaneus trans-hepatic biliary drainage) PPT,PDF PKPTBD (Percutaneus trans-hepatic biliary drainage) PPT,PDF PK
PTBD (Percutaneus trans-hepatic biliary drainage) PPT,PDF PKDr pradeep Kumar
 
Radiographic technique of skull
Radiographic technique of skullRadiographic technique of skull
Radiographic technique of skullSaruGosain
 
Medical devices on the chest x ray
Medical  devices on the chest x rayMedical  devices on the chest x ray
Medical devices on the chest x rayd795277
 
XRAY ANKLE POSITIONING.pptx
XRAY ANKLE POSITIONING.pptxXRAY ANKLE POSITIONING.pptx
XRAY ANKLE POSITIONING.pptxdypradio
 
Radiographic views of sacrum and coccyx
Radiographic views of sacrum and coccyxRadiographic views of sacrum and coccyx
Radiographic views of sacrum and coccyxChandan Prasad
 
Ptc )percutaneous transhepatic cholangiography
Ptc )percutaneous transhepatic cholangiographyPtc )percutaneous transhepatic cholangiography
Ptc )percutaneous transhepatic cholangiographyYashawant Yadav
 
Technique 1 Upper limbs 3
Technique 1 Upper limbs 3Technique 1 Upper limbs 3
Technique 1 Upper limbs 3Behzad Ommani
 
Shoulder x-ray
Shoulder x-rayShoulder x-ray
Shoulder x-rayNguyen Ha
 
Ankle joint radiography
Ankle joint radiographyAnkle joint radiography
Ankle joint radiographyNikhil Murkey
 

What's hot (20)

Radiograpic views for shoulder joint
Radiograpic views  for shoulder jointRadiograpic views  for shoulder joint
Radiograpic views for shoulder joint
 
CT Pulmonary Angiography
CT Pulmonary AngiographyCT Pulmonary Angiography
CT Pulmonary Angiography
 
Upper extremity anatomy & positioning
Upper extremity anatomy & positioningUpper extremity anatomy & positioning
Upper extremity anatomy & positioning
 
Virtual colonoscopy
Virtual colonoscopyVirtual colonoscopy
Virtual colonoscopy
 
X-RAY PELVIS PROJECTIONS
X-RAY PELVIS PROJECTIONSX-RAY PELVIS PROJECTIONS
X-RAY PELVIS PROJECTIONS
 
Radiological anatomy of mediastinum
Radiological anatomy of  mediastinumRadiological anatomy of  mediastinum
Radiological anatomy of mediastinum
 
Hepatic arterial anatomy and vascular optimization final
Hepatic arterial anatomy and vascular optimization finalHepatic arterial anatomy and vascular optimization final
Hepatic arterial anatomy and vascular optimization final
 
PTBD (Percutaneus trans-hepatic biliary drainage) PPT,PDF PK
PTBD (Percutaneus trans-hepatic biliary drainage) PPT,PDF PKPTBD (Percutaneus trans-hepatic biliary drainage) PPT,PDF PK
PTBD (Percutaneus trans-hepatic biliary drainage) PPT,PDF PK
 
Radiographic technique of skull
Radiographic technique of skullRadiographic technique of skull
Radiographic technique of skull
 
Dr Harshil radio spine
Dr Harshil radio spineDr Harshil radio spine
Dr Harshil radio spine
 
Medical devices on the chest x ray
Medical  devices on the chest x rayMedical  devices on the chest x ray
Medical devices on the chest x ray
 
the lower limb positioning
the lower limb positioningthe lower limb positioning
the lower limb positioning
 
XRAY ANKLE POSITIONING.pptx
XRAY ANKLE POSITIONING.pptxXRAY ANKLE POSITIONING.pptx
XRAY ANKLE POSITIONING.pptx
 
Radiographic views of sacrum and coccyx
Radiographic views of sacrum and coccyxRadiographic views of sacrum and coccyx
Radiographic views of sacrum and coccyx
 
Ptc )percutaneous transhepatic cholangiography
Ptc )percutaneous transhepatic cholangiographyPtc )percutaneous transhepatic cholangiography
Ptc )percutaneous transhepatic cholangiography
 
CT PA
CT PACT PA
CT PA
 
Technique 1 Upper limbs 3
Technique 1 Upper limbs 3Technique 1 Upper limbs 3
Technique 1 Upper limbs 3
 
Shoulder x-ray
Shoulder x-rayShoulder x-ray
Shoulder x-ray
 
Ankle joint radiography
Ankle joint radiographyAnkle joint radiography
Ankle joint radiography
 
X ray imaging intensifier
X ray imaging intensifierX ray imaging intensifier
X ray imaging intensifier
 

Similar to RADIOLOGY OF LINES AND TUBES IN INTENSIVE CARE SEMINAR.pdf

LINES AND TUBES ON XRAY with pathologies related to them
LINES AND TUBES ON XRAY with pathologies related to themLINES AND TUBES ON XRAY with pathologies related to them
LINES AND TUBES ON XRAY with pathologies related to themmuditsinghal20
 
chest radiology in ICU
   chest radiology in ICU   chest radiology in ICU
chest radiology in ICUEman Mahmoud
 
lines and tubes yash (1).pptx
lines and tubes yash (1).pptxlines and tubes yash (1).pptx
lines and tubes yash (1).pptxdypradio
 
central venous cannulation ug pdf.pdf
central venous cannulation ug pdf.pdfcentral venous cannulation ug pdf.pdf
central venous cannulation ug pdf.pdfsandeepsandeepkundra
 
CENTRAL VENOUS CATHETER
CENTRAL VENOUS CATHETERCENTRAL VENOUS CATHETER
CENTRAL VENOUS CATHETERAvijit Prusty
 
centralvenouscatheter-1.pdf
centralvenouscatheter-1.pdfcentralvenouscatheter-1.pdf
centralvenouscatheter-1.pdfisha sharma
 
Medical devices on the chest x ray
Medical  devices on the chest x rayMedical  devices on the chest x ray
Medical devices on the chest x rayairwave12
 
CHEST X-RAY DEVICES AND ARTIFACTS pptx.pptx
CHEST X-RAY DEVICES AND ARTIFACTS pptx.pptxCHEST X-RAY DEVICES AND ARTIFACTS pptx.pptx
CHEST X-RAY DEVICES AND ARTIFACTS pptx.pptxDR Venkata Ramana
 
Tracheostomy and its care by Dr.Ashwin menon
Tracheostomy and its care by Dr.Ashwin menonTracheostomy and its care by Dr.Ashwin menon
Tracheostomy and its care by Dr.Ashwin menonDr.Ashwin Menon
 
CHEST XRAY TUBES.pptx,,,,,,,,,,,,,,,,,,,
CHEST XRAY TUBES.pptx,,,,,,,,,,,,,,,,,,,CHEST XRAY TUBES.pptx,,,,,,,,,,,,,,,,,,,
CHEST XRAY TUBES.pptx,,,,,,,,,,,,,,,,,,,DR Venkata Ramana
 
Central line placement
Central line placementCentral line placement
Central line placementNeetiJamwal
 
Mitral valve surgical treatment
Mitral valve surgical treatmentMitral valve surgical treatment
Mitral valve surgical treatmentDR NIKUNJ SHEKHADA
 

Similar to RADIOLOGY OF LINES AND TUBES IN INTENSIVE CARE SEMINAR.pdf (20)

LINES AND TUBES ON XRAY with pathologies related to them
LINES AND TUBES ON XRAY with pathologies related to themLINES AND TUBES ON XRAY with pathologies related to them
LINES AND TUBES ON XRAY with pathologies related to them
 
chest radiology in ICU
   chest radiology in ICU   chest radiology in ICU
chest radiology in ICU
 
lines and tubes yash (1).pptx
lines and tubes yash (1).pptxlines and tubes yash (1).pptx
lines and tubes yash (1).pptx
 
Thoracic imaging in icu
Thoracic imaging in icuThoracic imaging in icu
Thoracic imaging in icu
 
Thoracic imaging in ICU
Thoracic imaging in ICUThoracic imaging in ICU
Thoracic imaging in ICU
 
Central venous cannulation
Central venous cannulation Central venous cannulation
Central venous cannulation
 
central venous cannulation ug pdf.pdf
central venous cannulation ug pdf.pdfcentral venous cannulation ug pdf.pdf
central venous cannulation ug pdf.pdf
 
catheters.pptx
catheters.pptxcatheters.pptx
catheters.pptx
 
lines_and_tubes (2).ppt
lines_and_tubes (2).pptlines_and_tubes (2).ppt
lines_and_tubes (2).ppt
 
CENTRAL VENOUS CATHETER
CENTRAL VENOUS CATHETERCENTRAL VENOUS CATHETER
CENTRAL VENOUS CATHETER
 
centralvenouscatheter-1.pdf
centralvenouscatheter-1.pdfcentralvenouscatheter-1.pdf
centralvenouscatheter-1.pdf
 
Chest x rays swati
Chest x rays swatiChest x rays swati
Chest x rays swati
 
central line.pptx
central line.pptx central line.pptx
central line.pptx
 
Medical devices on the chest x ray
Medical  devices on the chest x rayMedical  devices on the chest x ray
Medical devices on the chest x ray
 
CHEST X-RAY DEVICES AND ARTIFACTS pptx.pptx
CHEST X-RAY DEVICES AND ARTIFACTS pptx.pptxCHEST X-RAY DEVICES AND ARTIFACTS pptx.pptx
CHEST X-RAY DEVICES AND ARTIFACTS pptx.pptx
 
Tracheostomy and its care by Dr.Ashwin menon
Tracheostomy and its care by Dr.Ashwin menonTracheostomy and its care by Dr.Ashwin menon
Tracheostomy and its care by Dr.Ashwin menon
 
CHEST XRAY TUBES.pptx,,,,,,,,,,,,,,,,,,,
CHEST XRAY TUBES.pptx,,,,,,,,,,,,,,,,,,,CHEST XRAY TUBES.pptx,,,,,,,,,,,,,,,,,,,
CHEST XRAY TUBES.pptx,,,,,,,,,,,,,,,,,,,
 
Septal puncure ppt
Septal puncure pptSeptal puncure ppt
Septal puncure ppt
 
Central line placement
Central line placementCentral line placement
Central line placement
 
Mitral valve surgical treatment
Mitral valve surgical treatmentMitral valve surgical treatment
Mitral valve surgical treatment
 

Recently uploaded

VIP Call Girl Sector 32 Noida Just Book Me 9711199171
VIP Call Girl Sector 32 Noida Just Book Me 9711199171VIP Call Girl Sector 32 Noida Just Book Me 9711199171
VIP Call Girl Sector 32 Noida Just Book Me 9711199171Call Girls Service Gurgaon
 
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591adityaroy0215
 
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetErnakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh
 
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in ChandigarhChandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in ChandigarhSheetaleventcompany
 
Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...
Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...
Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...Sheetaleventcompany
 
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012adityaroy0215
 
Call Girls Service In Goa 💋 9316020077💋 Goa Call Girls By Russian Call Girl...
Call Girls Service In Goa  💋 9316020077💋 Goa Call Girls  By Russian Call Girl...Call Girls Service In Goa  💋 9316020077💋 Goa Call Girls  By Russian Call Girl...
Call Girls Service In Goa 💋 9316020077💋 Goa Call Girls By Russian Call Girl...russian goa call girl and escorts service
 
Call Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in Anantapur
Call Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in AnantapurCall Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in Anantapur
Call Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in Anantapurgragmanisha42
 
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR Call G...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR   Call G...❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR   Call G...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR Call G...Gfnyt.com
 
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetHubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅gragmanisha42
 
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.ktanvi103
 
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetnagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetSambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetOzhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...indiancallgirl4rent
 
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...indiancallgirl4rent
 

Recently uploaded (20)

VIP Call Girl Sector 32 Noida Just Book Me 9711199171
VIP Call Girl Sector 32 Noida Just Book Me 9711199171VIP Call Girl Sector 32 Noida Just Book Me 9711199171
VIP Call Girl Sector 32 Noida Just Book Me 9711199171
 
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
 
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetErnakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in ChandigarhChandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
 
Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...
Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...
Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...
 
9316020077📞Goa Call Girls Numbers, Call Girls Whatsapp Numbers Goa
9316020077📞Goa  Call Girls  Numbers, Call Girls  Whatsapp Numbers Goa9316020077📞Goa  Call Girls  Numbers, Call Girls  Whatsapp Numbers Goa
9316020077📞Goa Call Girls Numbers, Call Girls Whatsapp Numbers Goa
 
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
 
Call Girls Service In Goa 💋 9316020077💋 Goa Call Girls By Russian Call Girl...
Call Girls Service In Goa  💋 9316020077💋 Goa Call Girls  By Russian Call Girl...Call Girls Service In Goa  💋 9316020077💋 Goa Call Girls  By Russian Call Girl...
Call Girls Service In Goa 💋 9316020077💋 Goa Call Girls By Russian Call Girl...
 
Call Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in Anantapur
Call Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in AnantapurCall Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in Anantapur
Call Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in Anantapur
 
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR Call G...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR   Call G...❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR   Call G...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR Call G...
 
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetHubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
 
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
 
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
 
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetnagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetSambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetOzhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
 
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
 

RADIOLOGY OF LINES AND TUBES IN INTENSIVE CARE SEMINAR.pdf

  • 1. LINES AND TUBES IN CHEST X RAY Dr.T.V.S.S.SAI MADHURI
  • 2. • Many medical devices are used in intensive care unit for long durations. • Each of them is a double edged sword:intended to save life ,but life threatening if its in wrong place. • All catheters have the potential risks of coiling, misplacement, knotting, and fracture. • So it is essential for a radiologist to know about the normal position and malpositions of devices and its associated complications. • The American College of Radiology (ACR) recommends a CXR immediately following placement of indwelling tubes, catheters and other devices to check the position and detect procedure related complications
  • 3. Commonly used tubes and lines in ICU Nasogastric tube Pacemaker Endotracheal tube Pulmonary (swan-ganz)catheter Tracheostomy tube Intra aortic ballon pump Central venous line Automated implantable cardioverter defibrillator Drainage tube Paediatric lines
  • 4. Nasogastric tube Indications :1. Feeding purpose 2. Aspiration of contents The NG tube has multiple side holes. There are terminal lead balls to facilitate identification of the tip. CORRECT POSITION : NG tube tip should be 10 cm distal to gastro esophageal junction
  • 5.
  • 6. MALPOSITIONS OF NG TUBE AND ITS COMPLICATIONS • In the oesophagus: If the side holes are positioned within the esophagus there is increased risk of aspiration • In the trachea or bronchus: Inadvertent insertion into the trachea and bronchus can cause pneumonia, pulmonary contusion, or pulmonary laceration. • Pharyngeal and esophageal perforations can occur but are rare.
  • 7. Frontal (A) and lateral (B) radiographs of the neck showigh a NG tube coiled in the upper esophagus with its tip in the oropharynx .
  • 8. Frontal radiograph of the chest shows a NG tube forming a loop in the left bronchus before the tip reaches the right lower lobe bronchus.
  • 9. Tube misplacement in right main bronchus The tube follows the course of the right main bronchus Its tip is projected over the lower zone of the right lung The NG tube has been inhaled rather than swallowed . The tube must be removed and repositioned
  • 10.
  • 11. Endotracheal tube The endotracheal (ET) tube is inserted for ventilation of both the lungs and for prevention of aspiration. Correct position: ET tube in the neutral position of the neck is with the tip 5–7 cm above the carina. When the carina is not visible, the tip of the ET tube should be approximately at the level of the medial ends of the clavicle It should lie midway between the larynx and carina so that injury to either structure or complications like inadvertent extubation or selective main stem bronchus intubation are avoided.
  • 12.
  • 13. Malpositions of ET tube and its complications • Selective intubation :It can cause collapse of the contralateral lung and hyperinflation of the ipsilateral lung, or pneumothorax. • Selective intubation can be clinically occult in about 60% of patients and only revealed on the CXR,so an immediate CXR after intubation is warranted as these complications are not uncommon. • Inadvertent esophageal intubation :It can be detected radiographically by the presence of an over distended stomach. • Tracheal stenosis can occur following long-term tube placement. • Aspiration of tooth can also occur as a complication
  • 14. Frontal chest radiographs show an endotracheal tube in the right main bronchus (arrowhead in A), causing hyperinflation of the ipsilateral lung and partial collapse of the left lung
  • 15. Frontal radiograph of a neonate shows inadvertent placement of an endotracheal tube in the esophagus with distension of the esophagus and stomach with air
  • 16. TRACHEOSTOMY TUBE • Tracheostomy tubes, are inserted through a stoma post-tracheostomy to help patients unable to breathe normally. INDICATIONS OF TRACHEOSTOMY TUBE • Prolonged ventilator dependence. • Prophylactic tracheostomy prior to head and neck cancer treatment. • Obstructive sleep apnea refractory to other treatments. • Chronic aspiration. • Neuromuscular disease. • Subglottic stenosis.
  • 17. CORRECT POSITION: The tip of the tracheostomy tube should be half way between the stoma and the carina, at the level of the D3 vertebra. It should lie parallel to the trachea.
  • 18.
  • 19. COMPLICATIONS OF TRACHEOSTOMY • Hemorrhage • Malposition • False tract • Tracheal stenosis • Surgical emphysema • Pneumo mediastinum • Pneumothorax
  • 20. Frontal chest radiograph shows complications of tracheostomy: pneumothorax , pneumo mediastinum , and surgical emphysema
  • 21. CENTRAL VENOUS LINES Central venous lines (catheters) are useful for a variety of purposes: • Hemodynamic pressure monitoring; • Hemodialysis • Administration of medications, nutrition, and fluids.
  • 22. • Correct positioning of a CVC tip depends on the side of entry and intended use of the catheter • For most short-term uses, such as fluid administration or monitoring of central venous pressure -Positioning the tip of a central venous catheter within the superior vena cava at or just above the level of the carina is generally considered • For the purpose of long term chemotherapy –the tip of cvc may be placed more inferiorly at the cavo-atrial junction - the junction of the SVC and right atrium .
  • 23. • Superior vena cava (SVC) anatomy • The internal jugular and subclavian veins join to form the brachiocephalic veins. • The brachiocephalic veins join to form the SVC(corresponds to thethe first anterior intercostal space) • The SVC is located to the right side of the mediastinum above and below the level of the carina. • SVC is not visible on a chest X-ray. As the carina is a visible structure, it can be used as an anatomical landmark to help determine the level of a CVC tip within the SVC. • The cavo-atrial junction is located approximately the height of two vertebral bodies below the level of the carina
  • 24.
  • 25. VARIOUS APPROACHES FOR CENTRAL VENOUS LINE • Right-sided catheters • CVCs are most commonly inserted via the right internal jugular vein. • Right internal jugular catheters are positioned on the right side of the neck, and pass vertically from a position above the clavicle. • This is an ideal position for right-sided catheters for fluid administration and venous pressure monitoring, but not for long-term chemotherapy or dialysis
  • 26.
  • 27. Right subclavian vein catheter • Catheters inserted into the subclavian vein pass below the clavicle and then curve into the SVC.
  • 29.
  • 30. CV Catheters - Complications • A chest X-ray taken after central venous catheter placement can identify immediate complications such as pneumothorax or pneumomediastinum and to identify incorrect positioning. • Immediate complications Complications such as pneumothorax and surgical emphysema may arise from traumatic placement. • Incorrect positioning If catheters are not inserted far enough they may be located in the jugular, subclavian, or brachiocephalic veins. Catheters inserted too far may enter the right atrium.
  • 32. CVC IN LEFT BRACHIOCEPHALIC VEIN The tip of this catheter is projected over the left brachiocephalic vein rather than the SVC
  • 33. CATHETER IN RIGHT ATRIUM This peripherally inserted central catheter (PICC) was aimed to be inserted with its tip at the level of the cavo-atrial junction (the height of two vertebral bodies below the carina) The PICC has been inserted too far with its tip in the right atrium (RA)
  • 34. HORIZONTAL POSITIONING IN SVC Catheters placed via a left-sided approach are prone to being positioned nearly horizontally rather than vertically within the SVC. Catheters which contact the lateral wall of the SVC in this way may cause vessel erosion if positioned long term, and should therefore be placed so the tip is orientated vertically.
  • 35. INTERNAL JUGULAR CATHETER -MISPLACED This left internal jugular catheter has entered the left subclavian vein. The catheter needs to be repositioned.
  • 36. Frontal chest radiograph shows an abnormally medial course of the catheter in a case of inadvertent carotid cannulation
  • 37. INTERCOSTAL DRIANAGE TUBES • The pleural tube, more commonly known as the intercostal drainage tube (ICD), is inserted through the 4th intercostal space in the anterior or mid-axillary line. • It is then directed posteroinferiorly in cases of effusion and anterosuperiorly in cases of pneumothorax. • The ICD tube has a terminal hole as well as side holes; the side holes can be identified on a CXR by the interruption in the radio-opaque outline of the tube. • No side holes should lie outside the chest/pleura
  • 38. Chest drain-treatment for pneumothorax To drain pneumothorax the tube is positioned with its tip pointing superiorly towards the apex of the pleural cavity.
  • 39. CHEST DRAIN - TREATMENT FOR PLEURAL EFFUSION To drain pleural effusion the tube tip is ideally located towards the lower part of the pleural cavity.
  • 40. MALPOSITIONS OF ICD TUBES An intercostal drainage tube is noted in situ. One hole is situated outside the rib cage, indicating Malposition, likely resulting in insufficient suction. Pneumothorax noted on the left side with a partially collapsed lung
  • 41. PULMONARY ARTERY CATHETER(SWAN-GANZ CATHETER) • The pulmonary artery catheter, also called a Swan-Ganz catheter, is used to monitor circulatory hemodynamics in critically ill patients by measuring pulmonary capillary wedge pressure. • The catheter is inserted via the subclavian vein or the internal jugular vein. • The ideal position of the tip of the catheter is in the right or left main pulmonary arteries, and the tip should not extend beyond the proximal interlobar pulmonary artery (within 2 cm of the hilum)
  • 42. Swan-Ganz catheter lies in the left pulmonary artery.
  • 43. COMPLICATIONS OF SWAN GANZ CATHETER • Occlusion of a pulmonary artery branch may occur if the catheter is too distal, if there is persistent inflation of the balloon, or if a clot is formed around the distal tip of the catheter; such occlusion can result in pulmonary infarction. A chest radiograph may reveal the infarction as a wedge-shaped pleural-based pulmonary opacity. • Other potential complications are intracardiac knotting, pulmonary artery perforation, arrhythmias, cardiac perforation, and placement in the inferior vena cava
  • 44. Anteroposterior chest radiograph shows that tip of catheter is too distal (i.e., > 2 cm lateral to hilum). There is wedged-shaped opacity distal to catheter, consistent with pulmonary infarction.
  • 45. PACEMAKER • Pacemakers are used in cases of severe sinus node dysfunction, complete heart block, and various arrhythmias. • They have two main elements: a pulse generator and a lead wire with electrodes. • The single-lead pacemaker is the most basic type and is positioned with its tip in the right ventricular apex. • An atrioventricular two-lead sequential pacemaker has one electrode in the right atrium and the other at the right ventricular apex . • Sometimes a third lead is placed in the coronary sinus to pace the left ventricle.
  • 46. Frontal chest radiograph (A) shows the optimal position of the electrode of a single-lead pacemaker. The electrode has been placed in the right ventricular apex . Frontal chest radiograph (B) shows a two-lead pacemaker that has one electrode in the right atrium and the other at the right ventricular apex
  • 47. COMPLICATIONS OF CARDIAC PACEMAKER • Malposition, • Intracardiac knotting, • Fracture, • Perforation • Cardiac tamponade, • Arrhythmias, • Infection, • Hemorrhage.
  • 48. Frontal chest radiograph shows coiling of the lead of a single-lead pacemaker in the right atrium
  • 49. Frontal chest radiograph shows abnormal course of the lead with the electrode tip overlying the liver.This was due to cardiac perforation by the pacemaker lead,with a fatal outcome
  • 50. Frontal chest radiograph shows recoil of the pacemaker lead with its tip in the superior vena cava . This is called Twiddler’s syndrome
  • 51. Intra-aortic Balloon Pump Intra-aortic balloon pump is a long-balloon temporary circulatory assist device that works on the principle of cardiac counter-pulsation. The IABP is used to support the circulation. The balloon, approximately 25-cm long, is mounted on a catheter. The catheter tip is visible as a 3 x 4-mm rectangular metallic density while the rest of the catheter is radiolucent. The catheter is inserted through the femoral artery. To avoid occlusion of the left subclavian artery and visceral and renal arteries, its tip should be slightly cephalad to the adjacent carina (2nd–3rd intercostal space)
  • 52. • The balloon is inflated with gas during diastole and deflates during systole, resulting in increase in coronary blood flow and reduction in left ventricular after load (and hence, reduction in myocardial oxygen consumption).
  • 53. Normal positioning of intraaortic counter pulsation balloon pump. Magnified Anteroposterior chest radiograph obtained during systole shows catheter’s radiopaque tip within upper descending thoracic aorta indicating proper positioning.
  • 54. Magnified Anteroposterior chest radiograph obtained during diastole shows inflated radiolucent balloon as well as radiopaque tip within upper descending thoracic aorta. Catheter is inflated during diastole to increase myocardial perfusion and is deflated during systole to decrease left ventricular after load.
  • 55. MALPOSITIONED INTRAAORTIC COUNTERPULSATION BALLOON PUMP Anteroposterior chest radiograph during diastole shows inflated radiolucent balloon. Radiopaque tip of balloon catheter is abnormally located beyond aortic arch. It is projected over left common carotid artery. This abnormal positioning can result in cerebral ischemia.
  • 56. Malposition intraaortic counter pulsation balloon pump. Anteroposterior chest radiograph obtained during diastole shows inflated radiolucent balloon with its radiopaque tip in mid thoracic aorta. Counter pulsation is less effective when catheter is too low and may result in obstruction of abdominal aortic branches, such as renal or mesenteric arteries.
  • 57. Automated Implantable Cardioverter Debrillator Automated implantable cardioverter debrillator (AICD) is used in cases of recurrent refractory ventricular tachycardia. It has two electrodes (one electrode in the right atrium and the other in the right ventricle). The lead is wider compared to the pacemaker lead and has a ‘coiled-spring’ appearance
  • 58. PACEMAKER VS AUTOMATED IMPLANTABLE CARDIOVERTER DEBRILLATOR
  • 59. • Complications are similar to those with pacemakers and the incidence of radiographical abnormalities may approach 20%.
  • 60. Pediatric Lines • Some catheters are only used in the pediatric population, for example, the umbilical artery and venous catheters. • They are used for vascular access for exchange transfusion and measurement of blood gases, pressures, electrolytes, etc. • The umbilical vein and arteries remain patent for up to 4–5 days after birth.
  • 61. Umbilical artery line The catheter should be passed through the umbilic artery and enter the aorta via the internal iliac artery. In order to avoid placement into aortic branches, the catheter should be either in a high position above the celiac, mesenteric and renal arteries or in a low position below the inferior mesenteric artery: High position: T6-T9 Low position: L3-L5 The high position is advisable since it leads to less vascular complications (15).
  • 62. Umbilical artery line in a good high position. Malposition of umbilical vein line in right portal vein.
  • 63. Malposition of umbilical artery line, folded in the abdominal aorta.
  • 64. Malposition of umbilical artery line in left iliac artery.
  • 65. Umbilical vein catheter An umbilical vein catheter should pass through the umbilical vein into the left portal vein. Then through the ductus venosus into a hepatic vein and the inferior caval vein (IVC) The tip should be positioned in the IVC at the level of the diaphragm.
  • 66. MALPOSITIONS AND ITS COMPLICATIONS • Low position in the umbilical vein. • Intrahepatic into the portal venous system, both right and left, or even into the superior mesenteric or splenic vein.This can cause thrombosis. • Perforation of the portal vein can cause haemorrhage or abscess formation in the liver. • Position too deep in right atrium or in the left atrium through a patent foramen ovale or atrial septal defect. This can lead to cardiac arrhythmias or perforation.
  • 67. Umbilical vein line positioned in the periphery of the liver through the right portal vein. Artery line in the left subclavian artery.
  • 68. The umbilical vein line is positioned in the umbilical vein and not deep enough.
  • 69. The tip of the umbilical vein line is pointed downwards and probably situated in the mesenteric vein.
  • 70. Frontal radiograph of the chest and abdomen of a neonate shows the tip of an umbilical venous catheter in the left atrium; it has passed through a patent foramen ovale. The tip of the umbilical artery catheter is in the arch of the aorta (which is undesirable as it is near the origin of the carotid artery)