SlideShare a Scribd company logo
1 of 78
Submitted By:
Pavana K A
 Introduction
 Theory
 Ion selective electrodes and types
 Applications of potentiometry
 Differential Thermal Analysis(DTA)
 Derivative differential thermal analysis(DDTA)
 Thermo Gravimetric Analysis(TGA)
 Differential scanning calorimetry
 Potentiometry is an electrochemical method of Analysis deals with
the measurement of electric potential or emf of an electrolyte
solution under the condition of constant current.
 Potentiometry is the measurement of electrical potential of an
electrolyte solution to determine its concentration.
 The principle is based on the fact that the potential of the given
sample is directly proportional to the concentration of its electro
active ions or its activity (pH)
 When the pair of electrodes is placed in the sample solution it shows
the potential difference by the addition of the titrant or by the
change in the concentration of the ions.
 The theory of potentiometry is based on the nernst equation.It
gives the basic relationship between the potential generated by an
electrochemical cell and the concentration of the ions.
 The potential E ( Half cell potential) of any electrode is given by
nernst equation.
◦ E : Half cell electrode potential
◦ E0 : Standard electrode potential
◦ R : Universal gas constant( 8.3145 j/k/mole)
◦ T : Temperature (298 K or 250 C)
◦ F : Faraday’s constant (96,500 c/mole)
◦ n : No. of electrons transferred in the half reaction
◦ [Ox] : Concentration of oxidized substances (Reducing agent)
◦ [Red] : Conc. Of reduced species (Oxidizing agent)
By combining the constants and converting base e to Log base 10, the
equation becomes as,
Significance of nernst equation
1. It gives the potential of an electrode.
2. It gives the relationship between the electrode potential and
concentration of electroactive ions.
Limitations of nernst equation
1.At lower concentrations activity of ions(pH) is equal to their
concentration, whereas at higher concentration it is not true.
2.The equation is applicable only when no current flows through
the electrodes, otherwise it changes the ionic activity at the
electrode surface and leads to variation in measured potential.
 It is a device which converts Chemical energy into electrical
energy. It is used to generate potential and electric current
from chemical reactions.
 It consists of two electrodes immersed in an electrolyte
solution. The current is generated by the chemical reactions
which involves releasing and accepting of electrons at the two
electrodes (Redox reactions)
 Each electrode of an electrochemical cell is referred to as Half
cell, one half cell loses electrons(oxidation) other half cell
gains electrons(reduction).
 One half cell is called as reference electrode, which has stable
and constant potential and the other is called as indicator
electrode the potential of which is to be determined.
Electrodes: These are mainly used to measure the voltages.
Types of electrodes
1.Reference Electrode
a. Primary Reference electrode : Standard hydrogen electrode(SHE)
b. Secondary Reference electrode
1.Saturated calomel electrode(SCE)
2.Mercury mercurous sulphate electrode
3.Silver silver chloride electrode
4.Mercury mercuric oxide electrode
2.Indicator Electrodes
a. Hydrogen Electrode
b. Quinhydrone electrode
c. Antimony Antimony Oxide electrode
d. Glass electrode
3.Combination pH Electrode
4.Oxidation reduction electrode
5.Ion Selective Electrodes
A. Glass membrane electrode
B. Liquid membrane electrode
C. Solid state electrode/sensor
D. Biocatalytic Membrane electrode
 An ion-selective electrode (ISE), also known as a specific ion
electrode (SIE), is a transducer (or sensor) that converts the activity
of a specific ion dissolved in a solution into an electrical potential,
which can be measured by a voltmeter or pH meter.
 An ideal I.S.E. consists of a thin membrane across which only the
intended ion can be transported.
 The transport of ions from a high conc. to a low one through a
selective binding with some sites within the membrane creates a
potential difference.
 Ion Selective Electrodes (including the most common pH electrode)
work on the basic principal of the galvanic cell .By measuring the
electric potential generated across a membrane by "selected" ions,
and comparing it to a reference electrode, a net charge is
determined. The strength of this charge is directly proportional to
the concentration of the selected ion. The basic formula is given for
the galvanic cell:
 • Ecell= EISE- ERef
• The glass contain 60 to 75 mole % SiO2,
2 to 20 % Al2O3or LaF3, 0 to 6 % BaO and
CaO, and a variable amount of a group
1A oxide. Inside the glass bulb contains
a dilute HCl solution and silver wire
coated with a layer of silver chloride.
• By altering the composition of the glass,
it is possible to make the electrode
selective for different ions.
• Glass membranes are selective for
monovalent cations because polyvalent
ions cannot easily penetrate the surface
of the membrane.
 The selectivity of glass electrodes is related both to the ability of
the various Monovalent cations to penetrate into the glass
membrane and to the degree of attraction of the cations to the
negative sites within the glass.
 Glass electrodes which are selective for H+(pH electrode),Li+, Na+,
K+, Cs+,Ag+, Ti+and NH4+are commercially available
 Monovalent cations from a solution into which the glass is dipped
can penetrate into the surface of the glass. The concentration of
the analyzed ion in the sample solution changes from that in the
internal reference solution, a potential difference develops across
the membrane.
 The electrode is immersed in the solution and pH is measured.
• A solid state membrane electrode can be a
single crystal, a pellet made from a sparingly
soluble salt, or a sparingly soluble salt
embedded in an inert matrix, e.g., rubber.
• The single crystal and pellet membranes are
homogenous, electrodes containing them are
referred to as homogenous membrane
electrodes.
• The membrane consisting of the sparingly
soluble salt in the inert binding material is a
heterogeneous membrane electrode.
• The lanthanum fluoride (LaF3) membrane is
the only single crystal membrane that is
widely used in ion selective electrodes.
 In the process of “doping”, the resistance of the LaF3 crystal is
decreased by replacing a relatively small number of La3+ions in
the crystals with Eu2+ions(Ionic charge transport)
 Because fluoride can selectively migrate to the crystal, the
lanthanum fluoride membrane is selective for fluoride.
 Vacancies in the crystalline structure have exactly the proper
size, charge, and shape to hold a fluoride ion.
 Fluoride ions migrate from vacancy to vacancy in the defective
LaF3crystal. As a fluoride ion abandons one position in the
crystalline structure, it leaves a hole into which another fluoride
can migrate.
 The result is a crystal which exhibits ionic conductivity.
 The conductance to the membrane, as well as the potential
across the membrane, can be related to the analyte concentration
for many solid state membrane electrodes.
 The membrane of the electrode uses an ion exchanger
permanently embedded in a plastic material that is sealed to
the electrode body.
 The membrane separates the internal filling solution and
reference from the external sample solution.
 The electrode resembles that of solid state electrode.
 The sites are free to move in the active phase(the membrane)
this makes the electrodes selective for multivalent ions over
univalent ions.
 These liquid membrane electrodes are responsive exactly to
those ions ( Ca2+ , ClO4
-, NO3
- and BF4
_ ) That are extremely
difficult to monitor by other techniques.
 A gas permeable membrane is used to isolate the analytes from
possible interferences in the sample.
 A thin buffer layer is used to trap the analyte gas and covert it to
some ionic species that can be detected potentiometrically.
 Gas sensing electrodes are available for the measurement of carbon
dioxide, nitrite and sulphur dioxide.
 These are simple and reliable but they tend to have a relatively slow
response and recovery time(often 30 seconds to 5 minutes)
 They are used to assay the gases dissolved in aqueous solutions.
 It is constructed by enclosing the glass pH membrane in a second,
gas permeable hydrophobic membrane.
 A thin layer of an electrolyte solution is held between the two
membranes.
 They also have a small reference electrode enclosed within the gas
permeable membrane.
 The potential between the internal ISE and the reference
electrode within the outer membrane is monitored.
 The gas permeable membrane holds a constant volume of
solution around the internal ISE into which the gaseous analyte
can diffuse.
 The hydrophobic gas-permeable membrane can be composed
of substance which allows passage of dissolved gas but
prevents the solution within the membrane from escaping
 The materials used are silicon rubber, Teflon polypropylene,
fluorinated ethylene propylene, polyvinylidene fluoride etc…
 Gas from the sample solution passes through the submerged
gas permeable membrane and equilibrates in the electrolyte
solution between the two membranes.
 The gas reacts reversibly with the electrolyte solution to form an ion
to which the ion selective electrode responds.
 Because the activity of the ion that is formed between the two
membranes is proportional to amount of gas dissolved in sample,
the electrode response is directly related to the activity of the gas in
the sample.
 The gases (primarily NH3, SO2 and CO2) which are detected by gas
sensing electrodes based on the pH electrode equilibrate with the
electrolyte solution to alter its pH:
 NH3+ H2O = NH4++ OH-
 SO2+ H2O = HSO3-+ H+
 CO2+ H2O = HCO3-+ H+
 H2S, HCN, HF and chloride can be assayed by using internal
homogenous membrane electrode containing the appropriate silver
salt.
 Disadvantage possesses relatively long response time i.e; require 1-
7minutes after insertion in to a sample solution to reach equilibrium
 They are another form of gas
sensing electrodes invented by
Ruzicka and Hansen.
 A very thin layer of an
appropriate electrolyte solution is
adsorbed on the surface of the
membrane of the glass electrode.
 The electrolyte solution is
adsorbed on glass membrane when
membrane comes in contact with
the sponge containing the
electrolyte solution and a wetting
agent.
 The reference electrode makes contact with the adsorbed
electrolyte layer through a small, porous, ceramic salt bridge.
 The air gap electrode is used to assay ionic species which can be
chemically converted to gases,e.g. HCO3
-
 The HCO3-solution is placed in the sample holder and an acid is
added to convert HCO3
-(aq) to CO2(g).
 The sample holder is placed in position under the electrode and
stirred with a magnetic stirrer and stirrer bar.
 Carbon dioxide which is emitted during the chemical reaction
equilibrates with the electrolyte solution on the glass membrane
and alters the pH of the solution.
 The glass electrode measures the pH of the resulting solution.
 The electrolyte solutions used with air gap electrode are the same as
those used with other gas-sensing electrodes.
 The air-gap electrode has a faster response time due to the thinner
layer of electrolyte solution and a longer lifetime than most of the
other types of sensing electrodes.
 A typical response time for an air-gap electrode is less than a
minute.
 Air-gap electrode is primarily used for analysis of NH4
+, HSO3
-.
 As an example they can be used for the determination of urea in
blood.
 They have the same design as liquid-ion-exchanger membrane
electrodes.
 The liquid-ion-exchanger is replaced in neutral-carrier
membranes with a neutral complexing agent (a neutral carrier)
such as crown ether, which is dissolved in a highly water
insoluble organic solvent.
 The neutral carrier complexes with the analyte at membrane-
sample interface to form a charged complex which is extracted
from the aqueous solution into the organic solvent in the
membrane.
 The selectivity of the membrane for a particular ion depends
upon the ability to extract the ion into the membrane, which in
turn depends upon the ability of the ion to form a complex with
the neutral carrier.
 After complexation and extraction, the species in the neutral-carrier
membrane has the same charge as the extracted ion.
 The solvent in which the neutral carrier is dissolved is usually a high
boiling organic compound such as nitrobenzene (used in Ba2+
selective electrodes), dibutylsebacate (used in K+selective electrode)
and onitrophenyl-n-octylether (used in a Ca2+selective electrode).
 The physical support for the neutral carrier and solvent is usually a
cellulose membrane, more commonly, a PVC membrane. In addition
to K+, Ca2+, Ba2+, neutral-carrier membrane electrodes are also
selective for Li+, H+, Mg2+, NH4+, Sr2+.
 They are considerably smaller than
other forms of ion selective electrodes
because the internal filling solution is
eliminated and the ion selective
membrane is coated directly on the
internal electrode wire.
 The ion selective membranes
utilized in coated wire electrodes
consists of either an ion exchanger or
neutral carrier immobilized in a
polymeric material that is coated on
the electrode.
 They are more sturdy than other
ISEs and can be constructed with
small tips.
 First the metal on the interior of the electrode is sealed into a glass or
some other suitable material so that several mm or less of the wire is
exposed.
 The exposed wire is successively dipped into a solution of the polymeric
material and then into a solution of the ion exchanger or neutral carrier.
 After the electrode has air-dried, the dipping procedure is repeated, if
necessary, until the membrane coating on the wire is the desired
thickness.
 Alternatively, the wire can be dipped into a single solution containing
both the membrane material and the polymerizer.
 The polymeric matrix can be any of materials including PVC, polymethyl
acrylate(PMM) or epoxy.
 The internal electrode can be constructed from metals like platinum,
copper, silver wire and graphite rods.
 The electrode consists of an ion
selective membrane deposited or
coated on the gate of a field effect
transistor (FET). The membrane can
be a sparingly soluble compound such
as silver bromide (solid state
membrane) or some other type of
membrane such as an ion exchanger or
neutral carrier in a PVC matrix.
 Often membranes in a PVC matrix
are used. Membranes in a PVC matrix
can be forced to adhere to the gate of
the FET by placing a polyimide mesh
over the gate prior to coating it with
the membrane.
 The potential at the membrane is partially determined by the
activity of the analyte in solution.
 That potential determines the flow of current through the
drain of the FET.
 The drain current consequently varies with the activity of the
analyte and is the monitored factor.
 It is an ion selective electrode which is coated with an enzyme-
containing acrylamide gel.
 The gel and enzyme are held in place on the surface of the ion
selective electrode by an inert physical support.
 The design is same as gas-sensing electrode.
 The support is a sheet of cellophane or a piece of gauze made
from dacron or nylon.
 The physical support is wrapped around the electrode membrane
and tied in place.
 The acrylamide gel containing the enzyme is coagulated on the
support-electrode combination.
 Enzymes are highly selective biochemical catalysts.
 The selectivity of Biomembrane electrode is due to the selectivity
of the enzymes that are used in electrodes.
 Here the enzyme-catalyzed reaction of the analyte yields an ionic
reaction product which is monitored by the internal ion-selective
electrode.
 The operation of the urea-selective electrode will serve to illustrate
the operation of Biomembrane electrodes.
 The glass membrane of an ammonium-sensitive glass electrode is
coated with an acrylamide gel layer containing the enzyme urease.
 When the electrode is dipped into a solution containing urea, the
following reaction occurs to yield NH4+ :
CO(NH2)2+ H2O 2NH4
++ CO2 .
 The NH4
+formed during the reaction is measured at the ammonium-
selective electrode.
 A working curve is prepared by plotting the potential of the electrode
in standard urea solutions as a function of the logarithm of urea
concentration.
 The urea concentration in the sample is obtained from the
working curve.
 Unfortunately the enzymes used in Biomembrane electrodes
gradually decay and the enzyme containing gel must be
periodically replaced.
 The Biomembrane of urea electrode lasts about 2 weeks.
 Biomembrane electrodes have long response time of 5 or more
minutes.
 Ion-selective electrodes are used in a wide variety of applications for
determining the concentrations of various ions in aqueous solutions.
 The electrodes can be used as an end point detector in a titration.
 Pollution Monitoring: CN, F, S, Cl, NO3 etc., in effluents, and natural
waters.
 Agriculture: NO3, Cl, NH4, K, Ca, I, CN in soils, plant material,
fertilizers and feedstuffs.
 Food Processing: NO3, NO2 in meat preservatives.
 Salt content of meat, fish, dairy products, fruit juices, brewing
solutions.
 F in drinking water and other drinks.
 Detergent Manufacture: Ca, Ba, F for studying effects on water
quality.
 Determination of equilibrium constants.
 Determination of ionic product of water.
 Determination of Dissociation constant of acids.
 Biochemistry: Analysis of Ca, K, Cl in body fluids like blood,
plasma, sweat and serum.
 Paper Manufacture: S and Cl in pulping and recovery cycle
liquors.
 F in skeletal and dental studies.
 Ca in dairy products and beer .
 Explosives: F, Cl, NO3 in explosive materials and combustion
products.
 When a material is heated its structural and chemical composition
can undergo changes such as fusion, melting, crystallization,
oxidation, decomposition, transition, expansion and sintering.
 Using Thermal Analysis such changes can be monitored in every
atmosphere of interest. The obtained information is very useful in
both quality control and problem solving.
 Monitoring a property of the sample as a function of temperature
or Monitors the temperature change associated with a chemical
Reactions.
 Thermogravimetry.
 Differential thermal Analysis.
 Differential scanning colorimetry.
 Dynamic Mechanical Analysis.
 Thermo Mechanical Analysis
 Direct Injection Enthalpimetry.
 Analytical technique In which the difference between the analyte
sample and a non reactive reference material as a function of
furnace temperature or time , while they are subjected to identical
controlled temperature program.
 Any transformation – change in specific heat or an enthaply of
transition can be detected by DTA.
 If zero temperature difference b/w sample & reference material –
sample does not undergo any chemical or physical change.
 The thermal curve is a plot of the temperature difference as a
function of the temperature of one of the two substances.
 In DTA both test sample & an inert reference material (alumina) –
controlled heating or cooling programming.
 The thermal effect may either be endothermic or exothermic and are
caused by physical phenomena such as fusion, crystalline structure
inversion, boiling, vapourisation, sublimation or others. Some
enthalpic effect are also caused by chemical reaction. In this
manner, endo- and exothermal bands and peaks appearing on
thermogram give info, regarding the detection of enthalpic changes.
 Sample holder : Thermocouples, containers,
ceramic or metallic block.
 Furnace
 Temperature Programmer
 Recording System
 The sample and reference are held
in separate containers within the
same furnace. Sample cups/holders
are made of platinum and are fairly
close to each other to ensure
identical heating.
 Temperatures are monitored with
thermocouples placed in the
pedestals that support the sample
cups.(this arrangement varies in
some instruments)
 Reference substance is thermally inert, alumina is often
used.(silicon carbide, glass beads-phase change,not used
throughout the temperature interval)
• This is constructed
with an appropriate
material (wire or
ribbon) wound on a
refractory tube.
• Furnaces possess
the desired
characteristics
for good temperature
regulation and programming.
• These are fairly inexpensive. Generally, the choice of the
resistance material is decided from the intended maximum
temperature of operation and gaseous environments.
 Temperature control unit controls the furnace temperature.
 The combination of the furnace and sensor enables the various
types of the measurement techniques. The computer can be
connected to several instruments which has the other type of
measurement techniques, enables the simultaneous
measurement and analysis.
 In order to control temperature, the three basic elements are
required.-Those are sensor, control element and heater.
 Fluctuations of temperature around the set value can be
minimized if the heat input to the system is progressively
reduced as the temperature approaches the desired value. Such a
controller that anticipates the approach to the set value is known
as a proportional controller.
 In thermo analytical studies, the signal obtained from the
sensors can be recorded in which the signal trace is produced
on paper or film, by ink, heating stylus, electric writing or
optical beam.
 There are two types of recording devices similar to the TG.-
one is deflection type and other is null type.
 Insert a very thin thermocouple into a disposable sample tube 2mm
in diameter and containing 0.1-10 mg of sample.
 Another identical tube is either kept empty or filled with reference
substance, such as quartz, sand, alumina or alundum powder.
 The two tubes are simultaneously inserted into a sample block and
subsequently heated(or cooled) at uniform predetermined
programmed rate.
 DTA- being a dynamic process, it is extremely imp that all aspects of
the technique must be thoroughly standardized so as to obtain
reproducible results. Such aspects are,-pretreatment of specimen, -
particle size & packing of specimen, -dilution of the specimen, -
nature of inert diluent, -crystalline substances must be powdered
and sieved thru 100 mesh sieve, -micelle size is critical for colloidal
particle and – adequate control of atmosphere.
 The furnace temperature is increased at a linear rate. The
temperature between the sample and reference ΔT is
continuously measured and a thermogram is recorded.
 (The plot of ΔT v/s T temperature of the reference)
 A break in the curve indicates that there is either a physical or a
chemical change in the sample.
 If ΔT is (TS-TR) then, the trough in the curve indicates
endothermic process and a crest indicates exothermic process.
 To avoid Oxidation of the sample, the furnace is purged with
inert Gas such as nitrogen to displace air in the furnace.
Sharp Endothermic– changes in crystallanity or fusion
Broad endotherms- dehydration reaction
Physical changes usually result in endothermic curves
Chemical reactions leads to exothermic curves.
Advantages:
 instruments can be used at very high
temperatures
 instruments are highly sensitive
 characteristic transition or reaction
temperatures can be accurately determined
Disadvantages:
 uncertainty of heats of fusion, transition, or
reaction estimations is 20-50%
 Identification of substances – We know that the DTA curve for two
substances is not identical. Therefore, these serve as finger prints for
various substances. Particularly, DTA has become an established
technique for the identification of clays.
 Identification of products – When a substance reacts with another
substance, the products is identified by their specific DTA curves.
Therefore, this technique has been termed ‘reaction DTA’.
 Melting points – As melting points can be easily determined by DTA,
it means that this technique can be used as a direct check of the
purity of the compound.
 Quantitative Analysis – We know that the area of DTA peak is
proportional to the total heat of reaction and hence to the weight of
the sample. Therefore, the quantitative analysis is possible with the
help of standard curves of peak area vs. weight.
 Quality control – DTA technique has been widely used for the quality
control of a large number of substances like cement, glass, soil,
catalysts, textiles, explosives, resins etc
 Thermogravimetry is the measurement of change in weight(mass) of
a sample as a function of temperature (while it is being heated)
 The apparatus used is thermobalance.
 This method is useful for determining sample purity and water,
carbonate, and organic content; and for studying decomposition
reactions.
 Dynamic Thermogravimetry: Sample is subjected to conditions of
continuous increase in temperature, which is linear with time.
 Static/ isothermal TGA: Sample is maintained at constant
temperature for a period of time during which any change in weight
is noted.
 Thermobalance
 Sample Holder
 Furnace
 Precision balance
 Recorder
 Sample Holder: Depending upon the Nature of sample, its weight
and quantity to be handled, different size and shape of sample
holders known as crucibles/shallow dish are employed . They are
made of Glass, quartz, aluminium, Stainless steel, Platinum etc.
 2 types:
1.Shallow Dish/pan: Holding samples which eliminates
Gas, vapours or volatile matter by diffusion during heating.
2.Deep crucibles for general purposes.
 Furnace: Made of high quality metal and varies in shape and size.
Sample can be easily introduces. Temperature Is maintained between
1000o-2000o C By temperature control panel.
 Both linear or fixed temperature heating is maintained by the use of
thermocouples like Chromel-Alumel, Copper-Platinum, Platinum-
Rhodium.
 For Higher temperature Tungsten or rhenium thermocouples are
used.
 Furnace temp. programme : They provide gradual rise of
temperature at a fixed rate . It has a fine control knobs through
which desired temperature with respect to time/rate can be
obtained. The controlling is done by increasing voltage through
the heated element by motor driven variable transformer or by
different thermocouples.
 Recorder: These are of common type having strip and pen
mechanism. Electrical supply duly amplified is fed to recorder
chart. The speed of recorder is variable and adjusted according
to the use.
 Most imp part of the Instrument.
 Accuracy, reproducibility, sensitivity, toughness, toughness etc. are
some important features of thermobalance.
 It should have some ideal properties,
1. it should cover wide range of temperature.
2. it should have high degree of mechanical rigidity and
electronic stability.
3.it should be capable of recording changes in weight
rapidly, accurately and continuously as a function of
temperature and time.
4. The heating rate should be linear in temperature range.
 Recording thermobalances are of two types.
1.Deflection type
2.Null type
In deflection type they are,
1. Beam type: conversion of beam deflection is recorded as a
signal on photographic recorder or measuring device.
2. Helical type: use of spring which contracts or elongates as a
result of weight changes during heating
 Null point balances are more common, A sensor is employed to
detect deviation of beam from a null point. Opposite force of
electrical or mechanical is supplied to the balance beam to
restore null position. The opposing force is calibrated to the
weight change which is recorded directly.
 The sample is continuously weighed while being heated in an
inert atmosphere of nitrogen gas or in vacuum. The plot of
sample weight vs. temperature is called thermogram. The break
in the curve at various temperatures indicates the physical or
chemical change occurring in the sample.
 (1) Instrumental factors
(a) Furnace heating rate
(b) Furnace atmosphere
 (2) Sample characteristics includes
(a) Weight of the sample
(b) Sample particle size
Instrumental factors
 Furnace Heating rate: The temperature at which the compound (or
sample) decompose depends upon the heating rate. When the
heating rate is high, the decomposition temperature is also high. A
heating rate of 3.5°C per minute is usually recommended for reliable
and reproducible TGA.
 Furnace atmosphere: The atmosphere inside the furnace surrounding
the sample has a profound effect on the decomposition temperature
of the sample. A pure N2 gas from a cylinder passed through the
furnace which provides an inert atmosphere.
Sample characteristics
 Weight of the sample: A small weight of the sample is
recommended, using a small weight eliminates the existence of
temperature gradient through the sample.
 Particle size of the sample: The particle size of the sample should
be small and uniform. The use of large particle or crystal may
result in apparent, very rapid weight loss during heating.
 Determination of the purity and thermal stability of both primary and
secondary standard.
 Determination of the composition of complex mixture and
decomposition of complex.
 For studying the sublimation behavior of various substances.
 TGA is used to study the kinetics of the reaction rate constant.
 Used in the study of catalyst: The change in the chemical states of
the catalyst may be studied by TGA techniques. (Zn-ZnCrO4) Zinc-
Zinc chromate is used as the catalyst in the synthesis of methanol.
 Determining the composition of alloys and mixtures.
Thermogravimetry is a valuable technique for assessing the purity of
materials.
 Analytical reagents, especially those used in titrimetric analysis as
primary standards, e.g. Na2CO3, KHP have been examined. Many
primary standards absorb appreciable amounts of water when
exposed to moist atmospheres.TG data can show the extent of this
absorption, hence the most suitable drying temperature for a given
reagent may
 This method can also be extended to the analysis of a three
component mixtures.
 The most important applications of thermogravimetry is in
examining the thermal stability of polymers.be determined.
 This technique is very much similar to the DTA.
 Measurement of difference in heat flow into substance and reference
material as a function of sample temperature when both substances
are exposed to controlled temperature.
 In DSC the sample and the reference compound are gradually heated
in an inert atmosphere if the sample undergoes a physical or
chemical changes by absorption of heat then heat energy is added or
removed from the sample to maintain to maintain both sample and
reference at the same temperature.
 Thermogram in DSC is the plot of energy added to the sample vs
temperature.
 Two basic types of DSC instruments: power compensation DSC
and heat-flux DSC.
Heat flux DSC:
 Both sample and reference pans are heated by a single furnace
through heat sink and heat resistor. Heat flow is proportional to the
heat difference of heat sink and holders. The temperature versus
time profile through a phase transition in a heat flux instrument is
not linear.
 At a phase transition, there is a large change in the heat capacity of
the sample, which leads to a difference in temperatures between the
sample and reference pan.
 A set of mathematical equations convert the signal into heat flow
information. By calibrating the standard material, the unknown
sample quantitative measurement is achievable.
 Sample and reference holders: Al or Pt pans placed on constantan
disc and are connected by a low-resistance heat flow path. A metallic
disc made of constantan alloy is the primary means of heat transfer .
Sample and reference sit on raised constantan discs.
 Sensors
• Chromel® (an alloy made of 90% nickel and 10% chromium)-
constantan area thermocouples (differential heat flow)
• Chromel®-alumel (an alloy consisting of approximately 95% nickel,
2% manganese, 2%
aluminium and 1% silicon) thermocouples (sample temperature)
 Furnace - same block is used for sample and reference.
 Temperature controller:
• The temperature difference between the sample and reference is
converted to differential thermal power, which is supplied to the
heaters to maintain the temperature of the sample and reference at the
program value
 Differential heat flow to sample and reference is measured by
thermocouples which are connected in series, located at the junction
of constantan disc and chromel wafers.
 With this, it is possible to achieve heating or cooling rates of
1000c /min to 00c/min (isothermal).
 It needs mathematical equations to get the heat flow .
 DSC Heat Flow Equation
 Dh/Dt - DSC Heat Flow Signal, Cp - Sample Heat Capacity = Sample
Specific Heat X Sample Weight, Dt/Dt - Heating Rate
 F(t,t) - Heat Flow That Is A Function Of Time At An Absolute
Temperature (Kinetic)
Dh/Dt = Cp Dt/Dt + F(t,t)
Power compensation DSC
 Both sample and reference pans are heated by a different furnaces.
When an event occurs in the sample, sensitive Platinum Resistance
Thermometer (PRT) detects the changes in the sample, and power
(energy) is applied to or removed from the sample furnace to
compensate for the change in heat flow to or from the sample. As a
result, the system is maintained at a “thermal null” state at all times.
The amount of power required to maintain system equilibrium is
directly proportional to the energy changes occurring in the sample.
No complex heat flux equations are necessary with a power
compensation DSC because the system directly measures energy flow
to and from the sample.
 In addition, PC type DSC has enhanced modulated temperature DSC
(StepScan) technique and fast scan DSC (Hyper DSC) for dramatic
improvements in productivity, as well as greater sensitivity.
 Furthermore, the heating and cooling rate of PC types DSC can be as
high as 500°C/min.
 Sample holder
• Aluminum or Platinum pans
 Sensors
• Platinum resistance thermocouples
• Separate sensors and heaters for the sample and reference
 Furnace
• Separate blocks for sample and reference cells
 Temperature controller
• Supply the differential thermal power to the heaters to maintain the
temperature of the sample and reference at the program value
 Accurately weigh samples (approx. 3 to 20 mg). encapsulate in the
metal pans of high thermal conductivity materials (aluminium,
platinum, stainless steel) are used.
 Pan configurations may be open, pinhole or hermetically sealed.
Same pan material and configuration for both sample and reference.
Material should entirely cover the bottom of the pan to ensure
thermal contact.
 Avoid overfilling of the pan to minimize the thermal lag from the
bulk of the material to the sensor . Small sample masses and low
heating rates improve resolution [Do not Decompose The Samples In
DSC Cell]
 Sample Shape: Cut the sample to uniform shape, do not crush the
sample. If the sample to be taken is pellet, cross section is to be
taken. If the sample material is powder then, it is spread uniformly
over the bottom of the sample pan.
 Using Sample Press: When using crimped pans, the pans should not
be over crimped. The bottom of the pans should remain flat, even
after crimping. When using hermetic pans, a little more pressure is
required to crimp the pans. Hermetic pans are sealed by forming a
cold wield on the aluminium pans.
 Sample Size: Smaller samples will increase the resolution but will
decrease the sensitivity.
Larger samples will decrease the resolution but will increase the
sensitivity.
Sample size depends on the type of material being measured.
If the sample is
Extremely reactive in nature - very small samples (<1 mg) are to be
taken.
Pure organics or pharmaceuticals - 1 to 5 mg
Polymers - approximately 10 mg
Composite materials -15 to 20 mg
 Reference Materials - An inert material like α-alumina is generally
used. Empty pan can also be used, if the sample weight is small. With
higher sample weights it is necessary to use a reference material,
because the total weight of the sample and its container should be
approximately the same as the total weight of the reference and its
container . The reference material should be selected so that it
possesses similar thermal characteristics to the sample.
 The most widely used reference material is α-alumina, which must
be of analytical reagent quality . Before use, α-alumina should be
recalcined and stored over magnesium perchlorate in a desiccator .
Kieselguhr is another reference material normally used when the
sample has a fibrous nature.
 If it doesn’t have similar thermal characteristics, dilutions are made.
Dilutions are done by thoroughly mixing suitable proportions of
sample and reference material.
 Purge Gases - Sample may react with air and may oxidize or burn.
The problem is overcome by using inert gases. Inert gases are used
to control moisture in the surrounding atmosphere.
 Commonly used inert gases are nitrogen, helium, argon etc.
 Nitrogen - It is the most commonly used inert gas. It increases the
sensitivity of the experiment. Typical flow rate is 50 ml/min.
 Helium - It has high thermal conductivity . It increases the resolution
of the peaks. The upper temperature limit for this gas is upto 3500c.
Flow rate is 25 ml/min
 Air or oxygen - Sometimes it is deliberately used to view oxidative
effects of the sample. Flow rate is 50 ml/min
 Heating Rate - Faster heating rate will increase the sensitivity but will
decrease the resolution. Slow heating rate will decrease the
sensitivity but will increase the resolution. Good starting point is
100c/min.
 The ultimate goal of calibration and adjustment is to provide a
measuring system that consistently delivers reproducible and
accurate results.
• Evaluation of the thermal resistance of the sample and reference
sensors
• Measurements over the temperature range of interest
 2-Step process
• The temperature difference of two empty crucibles is measured
• The thermal response is then acquired for a standard material,
usually sapphire, on both the sample and reference platforms
 Amplified DSC signal is automatically varied with the
temperature to maintain a constant calorimetric sensitivity with
temperature.
Temperature
 Goal is to match the melting onset temperatures indicated by
the furnace Thermocouple readouts to the known melting points of
standards analyzed by DSC.
 Should be calibrated as close as possible to the desired temperature
range.
 Celebrant's
• High purity
• Accurately known enthalpies
• Thermally stable
• Light stable
• Non-hygroscopic
• Un-reactive (pan, atmosphere)
 The result of a DSC experiment is a curve of heat flux versus
temperature or versus time. There are two different conventions:
exothermic reactions in the sample shown with a positive or
negative peak, depending on the kind of technology used in the
experiment.
 This curve can be used to calculate enthalpies of transitions.
 Area under the peak is directly proportional to heat absorbed or
evolved by the reaction.
 Height of the peak is directly proportional to rate of the
reaction.
Factors affecting DSC curve/ Thermogram
Instrumental factors Sample characteristics
Furnace heating rate Amount of sample
Recording or Chart speed Nature of sample
Furnace atmosphere Sample packing
Geometry of sample holders/
location of sensors
Particle size
Sensitivity of the recording
system
Solubility of evolved gases in
the sample
Composition of sample
containers
Heat of reaction
Thermal conductivity
1) Sample shape:
The shape of the sample has little effect on the quantitative aspect of
DSC but more effect on the qualitative aspects. samples in the form of
a disc film or powder spread on the pan are preferred. In the case of
polymeric sheets, a disc cut with a cork-borer gives good results.
2)Sample size:
About 0.5 to 10mg is usually sufficient. Smaller samples enable faster
scanning, give better shaped peaks with good resolution and provide
better contact with the gaseous environment. With larger samples,
smaller heats of transitions may be measured with greater precision.
3) Atmosphere and geometry of sample holders:
There are a number of variables that affect DSC results includes the
type of pan, heating rate, the nature and mass of the compound, the
particle size distribution, packaging and porosity, pre-treatment and
dilution of the sample. It is used for purity analysis of above 98% pure
compounds.
 Calibration
 Contamination
 Sample preparation – how sample is loaded into a pan
 Residual solvents and moisture.
 Thermal lag
•Heating/Cooling rates
•Sample mass
 Processing errors
 Determination of purity of the sample.
 Detection of polymorphism and quantification of Polymorph,
Detection of metastable polymorphs.
 Detection of isomerism.
 Stability and compatibility studies.
 Lyophilization studies.
 Estimation of amorphous content in excipients.
 Percentage crystallinity Determination.
 Characterization of Membranes, lipids, nucleic acids & micellar
systems.
 Assessment of the effects of structural change on a molecules
stability.
 Measurement of Ultra-light molecular interactions
 Assessment of biocompatibility during manufacturing.
 Protein Stability and Folding
 Liquid Biopharmaceutical Formulations and Process Development
 https://www.hitachi-
hightech.com/global/products/science/tech/ana/thermal/descriptio
ns/ta.html
 Instrumental Methods of Chemical Analysis B. K. Sharma
 Instrumental methods of analysis, seventh edition – Willard, Merritt,
Dean, Settle.
 Instrumental methods of analysis – M. Calvin, S.C. Bhatia .
 Basic Concepts of Analytical Chemistry – Third edition by S.M.
Khopkar .
 Introduction to Instrumental Analysis – Robert D. Braun.
 Instrumental methods in Pharmaceutical Analysis– Dr. U.B.Hadkar
 Pharmaceutical Analysis- volume II – A.V.Kasture, K.R.Mahdik,
S.G.Wadodkar, H.N.More.
 www.slideshare.com
 www.google.co.in
 https://nptel.ac.in/courses/115103030/module4/lec22/1.html
 www.wikipedia.org

More Related Content

What's hot (20)

polarography
polarographypolarography
polarography
 
Polarography[1]
Polarography[1]Polarography[1]
Polarography[1]
 
ELECTROGRAVIMETRY
ELECTROGRAVIMETRYELECTROGRAVIMETRY
ELECTROGRAVIMETRY
 
Potentiometry ppt
Potentiometry pptPotentiometry ppt
Potentiometry ppt
 
LC NMR HYPHENATED TECHNIQUES
LC NMR HYPHENATED TECHNIQUESLC NMR HYPHENATED TECHNIQUES
LC NMR HYPHENATED TECHNIQUES
 
Cosy,nosy
Cosy,nosyCosy,nosy
Cosy,nosy
 
Polarography
PolarographyPolarography
Polarography
 
Mass fragmentation & rules
Mass fragmentation & rulesMass fragmentation & rules
Mass fragmentation & rules
 
CE-MS Hyphenation
CE-MS HyphenationCE-MS Hyphenation
CE-MS Hyphenation
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
Lc nmr
Lc nmrLc nmr
Lc nmr
 
FACTORS INFLUENCING FLUORESCENCE INTENSITY
FACTORS INFLUENCING FLUORESCENCE INTENSITYFACTORS INFLUENCING FLUORESCENCE INTENSITY
FACTORS INFLUENCING FLUORESCENCE INTENSITY
 
Amperometric Titrations
Amperometric TitrationsAmperometric Titrations
Amperometric Titrations
 
2 potentiometry
2 potentiometry2 potentiometry
2 potentiometry
 
Nmr spectroscopy
Nmr spectroscopyNmr spectroscopy
Nmr spectroscopy
 
1 Potentiometry
1  Potentiometry1  Potentiometry
1 Potentiometry
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
Ion pair chromatography for pharmacy students
Ion pair chromatography for pharmacy studentsIon pair chromatography for pharmacy students
Ion pair chromatography for pharmacy students
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
Mass Spectroscopy
Mass Spectroscopy Mass Spectroscopy
Mass Spectroscopy
 

Similar to Potentiometry

Ion sensitive electrode &amp; gas sensitive electrode..
Ion sensitive electrode &amp; gas sensitive electrode..Ion sensitive electrode &amp; gas sensitive electrode..
Ion sensitive electrode &amp; gas sensitive electrode..vishnupriya456
 
Electrochemanal labauxiliary-fin
Electrochemanal labauxiliary-finElectrochemanal labauxiliary-fin
Electrochemanal labauxiliary-finMUBOSScz
 
Electrochemanal labauxiliary-fin
Electrochemanal labauxiliary-finElectrochemanal labauxiliary-fin
Electrochemanal labauxiliary-finMUBOSScz
 
Ion selective electrodes ISE (potentiometery) PH meter
Ion selective electrodes ISE (potentiometery) PH meterIon selective electrodes ISE (potentiometery) PH meter
Ion selective electrodes ISE (potentiometery) PH meterMuhammad Asif Shaheeen
 
Electrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, iseElectrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, iseAngelica Nhoj Gemora
 
Potentiometery BP 1st year.pptx
Potentiometery BP 1st year.pptxPotentiometery BP 1st year.pptx
Potentiometery BP 1st year.pptxshwetasingh31652
 
Pharmaceutical Analysis Potentiometery BP
Pharmaceutical Analysis Potentiometery BPPharmaceutical Analysis Potentiometery BP
Pharmaceutical Analysis Potentiometery BPshwetasingh31652
 
Ion Selective Electrode
Ion Selective ElectrodeIon Selective Electrode
Ion Selective ElectrodeWritu Bashyal
 
Ion Specific Electrodes (ISE)
Ion Specific Electrodes (ISE)Ion Specific Electrodes (ISE)
Ion Specific Electrodes (ISE)Tapeshwar Yadav
 
Electroanalytical chemistry
Electroanalytical chemistryElectroanalytical chemistry
Electroanalytical chemistrywadhava gurumeet
 
Ion selective electrodes(ise)
Ion selective electrodes(ise)Ion selective electrodes(ise)
Ion selective electrodes(ise)Dilshad P.A.
 
[Group 5] electrochemistry, electrophoresis, isoelectric focusing
[Group 5] electrochemistry, electrophoresis, isoelectric focusing[Group 5] electrochemistry, electrophoresis, isoelectric focusing
[Group 5] electrochemistry, electrophoresis, isoelectric focusingRena Faith Baradero
 

Similar to Potentiometry (20)

Ion sensitive electrode &amp; gas sensitive electrode..
Ion sensitive electrode &amp; gas sensitive electrode..Ion sensitive electrode &amp; gas sensitive electrode..
Ion sensitive electrode &amp; gas sensitive electrode..
 
potentiometery
potentiometerypotentiometery
potentiometery
 
Ion selective electrode (hw)85
Ion selective electrode (hw)85Ion selective electrode (hw)85
Ion selective electrode (hw)85
 
Electrochemanal labauxiliary-fin
Electrochemanal labauxiliary-finElectrochemanal labauxiliary-fin
Electrochemanal labauxiliary-fin
 
Electrochemanal labauxiliary-fin
Electrochemanal labauxiliary-finElectrochemanal labauxiliary-fin
Electrochemanal labauxiliary-fin
 
Ion selective electrodes ISE (potentiometery) PH meter
Ion selective electrodes ISE (potentiometery) PH meterIon selective electrodes ISE (potentiometery) PH meter
Ion selective electrodes ISE (potentiometery) PH meter
 
Electrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, iseElectrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, ise
 
pH Measurement
pH MeasurementpH Measurement
pH Measurement
 
Potentiometery BP 1st year.pptx
Potentiometery BP 1st year.pptxPotentiometery BP 1st year.pptx
Potentiometery BP 1st year.pptx
 
Pharmaceutical Analysis Potentiometery BP
Pharmaceutical Analysis Potentiometery BPPharmaceutical Analysis Potentiometery BP
Pharmaceutical Analysis Potentiometery BP
 
Ion Selective Electrode
Ion Selective ElectrodeIon Selective Electrode
Ion Selective Electrode
 
ION SELECTIVE POTENTIOMETRY
ION SELECTIVE POTENTIOMETRYION SELECTIVE POTENTIOMETRY
ION SELECTIVE POTENTIOMETRY
 
Ion Specific Electrodes (ISE)
Ion Specific Electrodes (ISE)Ion Specific Electrodes (ISE)
Ion Specific Electrodes (ISE)
 
Electroanalytical chemistry
Electroanalytical chemistryElectroanalytical chemistry
Electroanalytical chemistry
 
Potentiometry ppt By Chand.pptx
Potentiometry ppt By Chand.pptxPotentiometry ppt By Chand.pptx
Potentiometry ppt By Chand.pptx
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
Reference electrodes
Reference electrodesReference electrodes
Reference electrodes
 
Ion selective electrodes(ise)
Ion selective electrodes(ise)Ion selective electrodes(ise)
Ion selective electrodes(ise)
 
[Group 5] electrochemistry, electrophoresis, isoelectric focusing
[Group 5] electrochemistry, electrophoresis, isoelectric focusing[Group 5] electrochemistry, electrophoresis, isoelectric focusing
[Group 5] electrochemistry, electrophoresis, isoelectric focusing
 
INDICATOR ELECTRODE
INDICATOR ELECTRODEINDICATOR ELECTRODE
INDICATOR ELECTRODE
 

More from Pavana K A

Screening of Dyslipidemic drugs
Screening of Dyslipidemic drugsScreening of Dyslipidemic drugs
Screening of Dyslipidemic drugsPavana K A
 
Safety Pharmacology
Safety PharmacologySafety Pharmacology
Safety PharmacologyPavana K A
 
Role of genomics and proteomics
Role of genomics and proteomicsRole of genomics and proteomics
Role of genomics and proteomicsPavana K A
 
Cell signaling
Cell signalingCell signaling
Cell signalingPavana K A
 
Endocrine pharmacology
Endocrine pharmacologyEndocrine pharmacology
Endocrine pharmacologyPavana K A
 
Anti ulcer screening
Anti ulcer screeningAnti ulcer screening
Anti ulcer screeningPavana K A
 

More from Pavana K A (8)

Screening of Dyslipidemic drugs
Screening of Dyslipidemic drugsScreening of Dyslipidemic drugs
Screening of Dyslipidemic drugs
 
Safety Pharmacology
Safety PharmacologySafety Pharmacology
Safety Pharmacology
 
Role of genomics and proteomics
Role of genomics and proteomicsRole of genomics and proteomics
Role of genomics and proteomics
 
Prokinetics 1
Prokinetics 1Prokinetics 1
Prokinetics 1
 
Gene therapy
Gene therapyGene therapy
Gene therapy
 
Cell signaling
Cell signalingCell signaling
Cell signaling
 
Endocrine pharmacology
Endocrine pharmacologyEndocrine pharmacology
Endocrine pharmacology
 
Anti ulcer screening
Anti ulcer screeningAnti ulcer screening
Anti ulcer screening
 

Recently uploaded

Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsBangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsGfnyt
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
 
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls ServiceMiss joya
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call girls in Ahmedabad High profile
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...Miss joya
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...Taniya Sharma
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomdiscovermytutordmt
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...CALL GIRLS
 
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...Neha Kaur
 
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Miss joya
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Miss joya
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...narwatsonia7
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...Miss joya
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Deliverynehamumbai
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Serviceparulsinha
 

Recently uploaded (20)

Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsBangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
 
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
 
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
 
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
 
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
 

Potentiometry

  • 2.  Introduction  Theory  Ion selective electrodes and types  Applications of potentiometry  Differential Thermal Analysis(DTA)  Derivative differential thermal analysis(DDTA)  Thermo Gravimetric Analysis(TGA)  Differential scanning calorimetry
  • 3.  Potentiometry is an electrochemical method of Analysis deals with the measurement of electric potential or emf of an electrolyte solution under the condition of constant current.  Potentiometry is the measurement of electrical potential of an electrolyte solution to determine its concentration.  The principle is based on the fact that the potential of the given sample is directly proportional to the concentration of its electro active ions or its activity (pH)  When the pair of electrodes is placed in the sample solution it shows the potential difference by the addition of the titrant or by the change in the concentration of the ions.
  • 4.  The theory of potentiometry is based on the nernst equation.It gives the basic relationship between the potential generated by an electrochemical cell and the concentration of the ions.  The potential E ( Half cell potential) of any electrode is given by nernst equation. ◦ E : Half cell electrode potential ◦ E0 : Standard electrode potential ◦ R : Universal gas constant( 8.3145 j/k/mole) ◦ T : Temperature (298 K or 250 C) ◦ F : Faraday’s constant (96,500 c/mole) ◦ n : No. of electrons transferred in the half reaction ◦ [Ox] : Concentration of oxidized substances (Reducing agent) ◦ [Red] : Conc. Of reduced species (Oxidizing agent)
  • 5. By combining the constants and converting base e to Log base 10, the equation becomes as, Significance of nernst equation 1. It gives the potential of an electrode. 2. It gives the relationship between the electrode potential and concentration of electroactive ions. Limitations of nernst equation 1.At lower concentrations activity of ions(pH) is equal to their concentration, whereas at higher concentration it is not true. 2.The equation is applicable only when no current flows through the electrodes, otherwise it changes the ionic activity at the electrode surface and leads to variation in measured potential.
  • 6.  It is a device which converts Chemical energy into electrical energy. It is used to generate potential and electric current from chemical reactions.  It consists of two electrodes immersed in an electrolyte solution. The current is generated by the chemical reactions which involves releasing and accepting of electrons at the two electrodes (Redox reactions)  Each electrode of an electrochemical cell is referred to as Half cell, one half cell loses electrons(oxidation) other half cell gains electrons(reduction).  One half cell is called as reference electrode, which has stable and constant potential and the other is called as indicator electrode the potential of which is to be determined.
  • 7. Electrodes: These are mainly used to measure the voltages. Types of electrodes 1.Reference Electrode a. Primary Reference electrode : Standard hydrogen electrode(SHE) b. Secondary Reference electrode 1.Saturated calomel electrode(SCE) 2.Mercury mercurous sulphate electrode 3.Silver silver chloride electrode 4.Mercury mercuric oxide electrode 2.Indicator Electrodes a. Hydrogen Electrode b. Quinhydrone electrode c. Antimony Antimony Oxide electrode d. Glass electrode 3.Combination pH Electrode 4.Oxidation reduction electrode 5.Ion Selective Electrodes A. Glass membrane electrode B. Liquid membrane electrode C. Solid state electrode/sensor D. Biocatalytic Membrane electrode
  • 8.  An ion-selective electrode (ISE), also known as a specific ion electrode (SIE), is a transducer (or sensor) that converts the activity of a specific ion dissolved in a solution into an electrical potential, which can be measured by a voltmeter or pH meter.  An ideal I.S.E. consists of a thin membrane across which only the intended ion can be transported.  The transport of ions from a high conc. to a low one through a selective binding with some sites within the membrane creates a potential difference.  Ion Selective Electrodes (including the most common pH electrode) work on the basic principal of the galvanic cell .By measuring the electric potential generated across a membrane by "selected" ions, and comparing it to a reference electrode, a net charge is determined. The strength of this charge is directly proportional to the concentration of the selected ion. The basic formula is given for the galvanic cell:  • Ecell= EISE- ERef
  • 9.
  • 10. • The glass contain 60 to 75 mole % SiO2, 2 to 20 % Al2O3or LaF3, 0 to 6 % BaO and CaO, and a variable amount of a group 1A oxide. Inside the glass bulb contains a dilute HCl solution and silver wire coated with a layer of silver chloride. • By altering the composition of the glass, it is possible to make the electrode selective for different ions. • Glass membranes are selective for monovalent cations because polyvalent ions cannot easily penetrate the surface of the membrane.
  • 11.  The selectivity of glass electrodes is related both to the ability of the various Monovalent cations to penetrate into the glass membrane and to the degree of attraction of the cations to the negative sites within the glass.  Glass electrodes which are selective for H+(pH electrode),Li+, Na+, K+, Cs+,Ag+, Ti+and NH4+are commercially available  Monovalent cations from a solution into which the glass is dipped can penetrate into the surface of the glass. The concentration of the analyzed ion in the sample solution changes from that in the internal reference solution, a potential difference develops across the membrane.  The electrode is immersed in the solution and pH is measured.
  • 12. • A solid state membrane electrode can be a single crystal, a pellet made from a sparingly soluble salt, or a sparingly soluble salt embedded in an inert matrix, e.g., rubber. • The single crystal and pellet membranes are homogenous, electrodes containing them are referred to as homogenous membrane electrodes. • The membrane consisting of the sparingly soluble salt in the inert binding material is a heterogeneous membrane electrode. • The lanthanum fluoride (LaF3) membrane is the only single crystal membrane that is widely used in ion selective electrodes.
  • 13.  In the process of “doping”, the resistance of the LaF3 crystal is decreased by replacing a relatively small number of La3+ions in the crystals with Eu2+ions(Ionic charge transport)  Because fluoride can selectively migrate to the crystal, the lanthanum fluoride membrane is selective for fluoride.  Vacancies in the crystalline structure have exactly the proper size, charge, and shape to hold a fluoride ion.  Fluoride ions migrate from vacancy to vacancy in the defective LaF3crystal. As a fluoride ion abandons one position in the crystalline structure, it leaves a hole into which another fluoride can migrate.  The result is a crystal which exhibits ionic conductivity.  The conductance to the membrane, as well as the potential across the membrane, can be related to the analyte concentration for many solid state membrane electrodes.
  • 14.  The membrane of the electrode uses an ion exchanger permanently embedded in a plastic material that is sealed to the electrode body.  The membrane separates the internal filling solution and reference from the external sample solution.  The electrode resembles that of solid state electrode.  The sites are free to move in the active phase(the membrane) this makes the electrodes selective for multivalent ions over univalent ions.  These liquid membrane electrodes are responsive exactly to those ions ( Ca2+ , ClO4 -, NO3 - and BF4 _ ) That are extremely difficult to monitor by other techniques.
  • 15.  A gas permeable membrane is used to isolate the analytes from possible interferences in the sample.  A thin buffer layer is used to trap the analyte gas and covert it to some ionic species that can be detected potentiometrically.  Gas sensing electrodes are available for the measurement of carbon dioxide, nitrite and sulphur dioxide.  These are simple and reliable but they tend to have a relatively slow response and recovery time(often 30 seconds to 5 minutes)  They are used to assay the gases dissolved in aqueous solutions.  It is constructed by enclosing the glass pH membrane in a second, gas permeable hydrophobic membrane.  A thin layer of an electrolyte solution is held between the two membranes.  They also have a small reference electrode enclosed within the gas permeable membrane.
  • 16.
  • 17.  The potential between the internal ISE and the reference electrode within the outer membrane is monitored.  The gas permeable membrane holds a constant volume of solution around the internal ISE into which the gaseous analyte can diffuse.  The hydrophobic gas-permeable membrane can be composed of substance which allows passage of dissolved gas but prevents the solution within the membrane from escaping  The materials used are silicon rubber, Teflon polypropylene, fluorinated ethylene propylene, polyvinylidene fluoride etc…  Gas from the sample solution passes through the submerged gas permeable membrane and equilibrates in the electrolyte solution between the two membranes.
  • 18.  The gas reacts reversibly with the electrolyte solution to form an ion to which the ion selective electrode responds.  Because the activity of the ion that is formed between the two membranes is proportional to amount of gas dissolved in sample, the electrode response is directly related to the activity of the gas in the sample.  The gases (primarily NH3, SO2 and CO2) which are detected by gas sensing electrodes based on the pH electrode equilibrate with the electrolyte solution to alter its pH:  NH3+ H2O = NH4++ OH-  SO2+ H2O = HSO3-+ H+  CO2+ H2O = HCO3-+ H+  H2S, HCN, HF and chloride can be assayed by using internal homogenous membrane electrode containing the appropriate silver salt.  Disadvantage possesses relatively long response time i.e; require 1- 7minutes after insertion in to a sample solution to reach equilibrium
  • 19.  They are another form of gas sensing electrodes invented by Ruzicka and Hansen.  A very thin layer of an appropriate electrolyte solution is adsorbed on the surface of the membrane of the glass electrode.  The electrolyte solution is adsorbed on glass membrane when membrane comes in contact with the sponge containing the electrolyte solution and a wetting agent.  The reference electrode makes contact with the adsorbed electrolyte layer through a small, porous, ceramic salt bridge.
  • 20.  The air gap electrode is used to assay ionic species which can be chemically converted to gases,e.g. HCO3 -  The HCO3-solution is placed in the sample holder and an acid is added to convert HCO3 -(aq) to CO2(g).  The sample holder is placed in position under the electrode and stirred with a magnetic stirrer and stirrer bar.  Carbon dioxide which is emitted during the chemical reaction equilibrates with the electrolyte solution on the glass membrane and alters the pH of the solution.
  • 21.  The glass electrode measures the pH of the resulting solution.  The electrolyte solutions used with air gap electrode are the same as those used with other gas-sensing electrodes.  The air-gap electrode has a faster response time due to the thinner layer of electrolyte solution and a longer lifetime than most of the other types of sensing electrodes.  A typical response time for an air-gap electrode is less than a minute.  Air-gap electrode is primarily used for analysis of NH4 +, HSO3 -.  As an example they can be used for the determination of urea in blood.
  • 22.  They have the same design as liquid-ion-exchanger membrane electrodes.  The liquid-ion-exchanger is replaced in neutral-carrier membranes with a neutral complexing agent (a neutral carrier) such as crown ether, which is dissolved in a highly water insoluble organic solvent.  The neutral carrier complexes with the analyte at membrane- sample interface to form a charged complex which is extracted from the aqueous solution into the organic solvent in the membrane.  The selectivity of the membrane for a particular ion depends upon the ability to extract the ion into the membrane, which in turn depends upon the ability of the ion to form a complex with the neutral carrier.
  • 23.  After complexation and extraction, the species in the neutral-carrier membrane has the same charge as the extracted ion.  The solvent in which the neutral carrier is dissolved is usually a high boiling organic compound such as nitrobenzene (used in Ba2+ selective electrodes), dibutylsebacate (used in K+selective electrode) and onitrophenyl-n-octylether (used in a Ca2+selective electrode).  The physical support for the neutral carrier and solvent is usually a cellulose membrane, more commonly, a PVC membrane. In addition to K+, Ca2+, Ba2+, neutral-carrier membrane electrodes are also selective for Li+, H+, Mg2+, NH4+, Sr2+.
  • 24.  They are considerably smaller than other forms of ion selective electrodes because the internal filling solution is eliminated and the ion selective membrane is coated directly on the internal electrode wire.  The ion selective membranes utilized in coated wire electrodes consists of either an ion exchanger or neutral carrier immobilized in a polymeric material that is coated on the electrode.  They are more sturdy than other ISEs and can be constructed with small tips.
  • 25.  First the metal on the interior of the electrode is sealed into a glass or some other suitable material so that several mm or less of the wire is exposed.  The exposed wire is successively dipped into a solution of the polymeric material and then into a solution of the ion exchanger or neutral carrier.  After the electrode has air-dried, the dipping procedure is repeated, if necessary, until the membrane coating on the wire is the desired thickness.  Alternatively, the wire can be dipped into a single solution containing both the membrane material and the polymerizer.  The polymeric matrix can be any of materials including PVC, polymethyl acrylate(PMM) or epoxy.  The internal electrode can be constructed from metals like platinum, copper, silver wire and graphite rods.
  • 26.  The electrode consists of an ion selective membrane deposited or coated on the gate of a field effect transistor (FET). The membrane can be a sparingly soluble compound such as silver bromide (solid state membrane) or some other type of membrane such as an ion exchanger or neutral carrier in a PVC matrix.  Often membranes in a PVC matrix are used. Membranes in a PVC matrix can be forced to adhere to the gate of the FET by placing a polyimide mesh over the gate prior to coating it with the membrane.
  • 27.  The potential at the membrane is partially determined by the activity of the analyte in solution.  That potential determines the flow of current through the drain of the FET.  The drain current consequently varies with the activity of the analyte and is the monitored factor.
  • 28.  It is an ion selective electrode which is coated with an enzyme- containing acrylamide gel.  The gel and enzyme are held in place on the surface of the ion selective electrode by an inert physical support.  The design is same as gas-sensing electrode.  The support is a sheet of cellophane or a piece of gauze made from dacron or nylon.  The physical support is wrapped around the electrode membrane and tied in place.  The acrylamide gel containing the enzyme is coagulated on the support-electrode combination.  Enzymes are highly selective biochemical catalysts.  The selectivity of Biomembrane electrode is due to the selectivity of the enzymes that are used in electrodes.
  • 29.  Here the enzyme-catalyzed reaction of the analyte yields an ionic reaction product which is monitored by the internal ion-selective electrode.  The operation of the urea-selective electrode will serve to illustrate the operation of Biomembrane electrodes.  The glass membrane of an ammonium-sensitive glass electrode is coated with an acrylamide gel layer containing the enzyme urease.  When the electrode is dipped into a solution containing urea, the following reaction occurs to yield NH4+ : CO(NH2)2+ H2O 2NH4 ++ CO2 .  The NH4 +formed during the reaction is measured at the ammonium- selective electrode.  A working curve is prepared by plotting the potential of the electrode in standard urea solutions as a function of the logarithm of urea concentration.
  • 30.  The urea concentration in the sample is obtained from the working curve.  Unfortunately the enzymes used in Biomembrane electrodes gradually decay and the enzyme containing gel must be periodically replaced.  The Biomembrane of urea electrode lasts about 2 weeks.  Biomembrane electrodes have long response time of 5 or more minutes.
  • 31.  Ion-selective electrodes are used in a wide variety of applications for determining the concentrations of various ions in aqueous solutions.  The electrodes can be used as an end point detector in a titration.  Pollution Monitoring: CN, F, S, Cl, NO3 etc., in effluents, and natural waters.  Agriculture: NO3, Cl, NH4, K, Ca, I, CN in soils, plant material, fertilizers and feedstuffs.  Food Processing: NO3, NO2 in meat preservatives.  Salt content of meat, fish, dairy products, fruit juices, brewing solutions.  F in drinking water and other drinks.  Detergent Manufacture: Ca, Ba, F for studying effects on water quality.  Determination of equilibrium constants.
  • 32.  Determination of ionic product of water.  Determination of Dissociation constant of acids.  Biochemistry: Analysis of Ca, K, Cl in body fluids like blood, plasma, sweat and serum.  Paper Manufacture: S and Cl in pulping and recovery cycle liquors.  F in skeletal and dental studies.  Ca in dairy products and beer .  Explosives: F, Cl, NO3 in explosive materials and combustion products.
  • 33.  When a material is heated its structural and chemical composition can undergo changes such as fusion, melting, crystallization, oxidation, decomposition, transition, expansion and sintering.  Using Thermal Analysis such changes can be monitored in every atmosphere of interest. The obtained information is very useful in both quality control and problem solving.  Monitoring a property of the sample as a function of temperature or Monitors the temperature change associated with a chemical Reactions.  Thermogravimetry.  Differential thermal Analysis.  Differential scanning colorimetry.  Dynamic Mechanical Analysis.  Thermo Mechanical Analysis  Direct Injection Enthalpimetry.
  • 34.  Analytical technique In which the difference between the analyte sample and a non reactive reference material as a function of furnace temperature or time , while they are subjected to identical controlled temperature program.  Any transformation – change in specific heat or an enthaply of transition can be detected by DTA.  If zero temperature difference b/w sample & reference material – sample does not undergo any chemical or physical change.  The thermal curve is a plot of the temperature difference as a function of the temperature of one of the two substances.  In DTA both test sample & an inert reference material (alumina) – controlled heating or cooling programming.  The thermal effect may either be endothermic or exothermic and are caused by physical phenomena such as fusion, crystalline structure inversion, boiling, vapourisation, sublimation or others. Some enthalpic effect are also caused by chemical reaction. In this manner, endo- and exothermal bands and peaks appearing on thermogram give info, regarding the detection of enthalpic changes.
  • 35.
  • 36.  Sample holder : Thermocouples, containers, ceramic or metallic block.  Furnace  Temperature Programmer  Recording System
  • 37.  The sample and reference are held in separate containers within the same furnace. Sample cups/holders are made of platinum and are fairly close to each other to ensure identical heating.  Temperatures are monitored with thermocouples placed in the pedestals that support the sample cups.(this arrangement varies in some instruments)  Reference substance is thermally inert, alumina is often used.(silicon carbide, glass beads-phase change,not used throughout the temperature interval)
  • 38. • This is constructed with an appropriate material (wire or ribbon) wound on a refractory tube. • Furnaces possess the desired characteristics for good temperature regulation and programming. • These are fairly inexpensive. Generally, the choice of the resistance material is decided from the intended maximum temperature of operation and gaseous environments.
  • 39.  Temperature control unit controls the furnace temperature.  The combination of the furnace and sensor enables the various types of the measurement techniques. The computer can be connected to several instruments which has the other type of measurement techniques, enables the simultaneous measurement and analysis.  In order to control temperature, the three basic elements are required.-Those are sensor, control element and heater.  Fluctuations of temperature around the set value can be minimized if the heat input to the system is progressively reduced as the temperature approaches the desired value. Such a controller that anticipates the approach to the set value is known as a proportional controller.
  • 40.
  • 41.  In thermo analytical studies, the signal obtained from the sensors can be recorded in which the signal trace is produced on paper or film, by ink, heating stylus, electric writing or optical beam.  There are two types of recording devices similar to the TG.- one is deflection type and other is null type.
  • 42.  Insert a very thin thermocouple into a disposable sample tube 2mm in diameter and containing 0.1-10 mg of sample.  Another identical tube is either kept empty or filled with reference substance, such as quartz, sand, alumina or alundum powder.  The two tubes are simultaneously inserted into a sample block and subsequently heated(or cooled) at uniform predetermined programmed rate.  DTA- being a dynamic process, it is extremely imp that all aspects of the technique must be thoroughly standardized so as to obtain reproducible results. Such aspects are,-pretreatment of specimen, - particle size & packing of specimen, -dilution of the specimen, - nature of inert diluent, -crystalline substances must be powdered and sieved thru 100 mesh sieve, -micelle size is critical for colloidal particle and – adequate control of atmosphere.
  • 43.  The furnace temperature is increased at a linear rate. The temperature between the sample and reference ΔT is continuously measured and a thermogram is recorded.  (The plot of ΔT v/s T temperature of the reference)  A break in the curve indicates that there is either a physical or a chemical change in the sample.  If ΔT is (TS-TR) then, the trough in the curve indicates endothermic process and a crest indicates exothermic process.  To avoid Oxidation of the sample, the furnace is purged with inert Gas such as nitrogen to displace air in the furnace.
  • 44. Sharp Endothermic– changes in crystallanity or fusion Broad endotherms- dehydration reaction Physical changes usually result in endothermic curves Chemical reactions leads to exothermic curves.
  • 45. Advantages:  instruments can be used at very high temperatures  instruments are highly sensitive  characteristic transition or reaction temperatures can be accurately determined Disadvantages:  uncertainty of heats of fusion, transition, or reaction estimations is 20-50%
  • 46.  Identification of substances – We know that the DTA curve for two substances is not identical. Therefore, these serve as finger prints for various substances. Particularly, DTA has become an established technique for the identification of clays.  Identification of products – When a substance reacts with another substance, the products is identified by their specific DTA curves. Therefore, this technique has been termed ‘reaction DTA’.  Melting points – As melting points can be easily determined by DTA, it means that this technique can be used as a direct check of the purity of the compound.  Quantitative Analysis – We know that the area of DTA peak is proportional to the total heat of reaction and hence to the weight of the sample. Therefore, the quantitative analysis is possible with the help of standard curves of peak area vs. weight.  Quality control – DTA technique has been widely used for the quality control of a large number of substances like cement, glass, soil, catalysts, textiles, explosives, resins etc
  • 47.  Thermogravimetry is the measurement of change in weight(mass) of a sample as a function of temperature (while it is being heated)  The apparatus used is thermobalance.  This method is useful for determining sample purity and water, carbonate, and organic content; and for studying decomposition reactions.  Dynamic Thermogravimetry: Sample is subjected to conditions of continuous increase in temperature, which is linear with time.  Static/ isothermal TGA: Sample is maintained at constant temperature for a period of time during which any change in weight is noted.
  • 48.  Thermobalance  Sample Holder  Furnace  Precision balance  Recorder
  • 49.  Sample Holder: Depending upon the Nature of sample, its weight and quantity to be handled, different size and shape of sample holders known as crucibles/shallow dish are employed . They are made of Glass, quartz, aluminium, Stainless steel, Platinum etc.  2 types: 1.Shallow Dish/pan: Holding samples which eliminates Gas, vapours or volatile matter by diffusion during heating. 2.Deep crucibles for general purposes.  Furnace: Made of high quality metal and varies in shape and size. Sample can be easily introduces. Temperature Is maintained between 1000o-2000o C By temperature control panel.  Both linear or fixed temperature heating is maintained by the use of thermocouples like Chromel-Alumel, Copper-Platinum, Platinum- Rhodium.  For Higher temperature Tungsten or rhenium thermocouples are used.
  • 50.  Furnace temp. programme : They provide gradual rise of temperature at a fixed rate . It has a fine control knobs through which desired temperature with respect to time/rate can be obtained. The controlling is done by increasing voltage through the heated element by motor driven variable transformer or by different thermocouples.  Recorder: These are of common type having strip and pen mechanism. Electrical supply duly amplified is fed to recorder chart. The speed of recorder is variable and adjusted according to the use.
  • 51.  Most imp part of the Instrument.  Accuracy, reproducibility, sensitivity, toughness, toughness etc. are some important features of thermobalance.  It should have some ideal properties, 1. it should cover wide range of temperature. 2. it should have high degree of mechanical rigidity and electronic stability. 3.it should be capable of recording changes in weight rapidly, accurately and continuously as a function of temperature and time. 4. The heating rate should be linear in temperature range.  Recording thermobalances are of two types. 1.Deflection type 2.Null type
  • 52. In deflection type they are, 1. Beam type: conversion of beam deflection is recorded as a signal on photographic recorder or measuring device. 2. Helical type: use of spring which contracts or elongates as a result of weight changes during heating  Null point balances are more common, A sensor is employed to detect deviation of beam from a null point. Opposite force of electrical or mechanical is supplied to the balance beam to restore null position. The opposing force is calibrated to the weight change which is recorded directly.
  • 53.  The sample is continuously weighed while being heated in an inert atmosphere of nitrogen gas or in vacuum. The plot of sample weight vs. temperature is called thermogram. The break in the curve at various temperatures indicates the physical or chemical change occurring in the sample.
  • 54.  (1) Instrumental factors (a) Furnace heating rate (b) Furnace atmosphere  (2) Sample characteristics includes (a) Weight of the sample (b) Sample particle size
  • 55. Instrumental factors  Furnace Heating rate: The temperature at which the compound (or sample) decompose depends upon the heating rate. When the heating rate is high, the decomposition temperature is also high. A heating rate of 3.5°C per minute is usually recommended for reliable and reproducible TGA.  Furnace atmosphere: The atmosphere inside the furnace surrounding the sample has a profound effect on the decomposition temperature of the sample. A pure N2 gas from a cylinder passed through the furnace which provides an inert atmosphere.
  • 56. Sample characteristics  Weight of the sample: A small weight of the sample is recommended, using a small weight eliminates the existence of temperature gradient through the sample.  Particle size of the sample: The particle size of the sample should be small and uniform. The use of large particle or crystal may result in apparent, very rapid weight loss during heating.
  • 57.  Determination of the purity and thermal stability of both primary and secondary standard.  Determination of the composition of complex mixture and decomposition of complex.  For studying the sublimation behavior of various substances.  TGA is used to study the kinetics of the reaction rate constant.  Used in the study of catalyst: The change in the chemical states of the catalyst may be studied by TGA techniques. (Zn-ZnCrO4) Zinc- Zinc chromate is used as the catalyst in the synthesis of methanol.
  • 58.  Determining the composition of alloys and mixtures. Thermogravimetry is a valuable technique for assessing the purity of materials.  Analytical reagents, especially those used in titrimetric analysis as primary standards, e.g. Na2CO3, KHP have been examined. Many primary standards absorb appreciable amounts of water when exposed to moist atmospheres.TG data can show the extent of this absorption, hence the most suitable drying temperature for a given reagent may  This method can also be extended to the analysis of a three component mixtures.  The most important applications of thermogravimetry is in examining the thermal stability of polymers.be determined.
  • 59.  This technique is very much similar to the DTA.  Measurement of difference in heat flow into substance and reference material as a function of sample temperature when both substances are exposed to controlled temperature.  In DSC the sample and the reference compound are gradually heated in an inert atmosphere if the sample undergoes a physical or chemical changes by absorption of heat then heat energy is added or removed from the sample to maintain to maintain both sample and reference at the same temperature.  Thermogram in DSC is the plot of energy added to the sample vs temperature.  Two basic types of DSC instruments: power compensation DSC and heat-flux DSC.
  • 60.
  • 61. Heat flux DSC:  Both sample and reference pans are heated by a single furnace through heat sink and heat resistor. Heat flow is proportional to the heat difference of heat sink and holders. The temperature versus time profile through a phase transition in a heat flux instrument is not linear.  At a phase transition, there is a large change in the heat capacity of the sample, which leads to a difference in temperatures between the sample and reference pan.  A set of mathematical equations convert the signal into heat flow information. By calibrating the standard material, the unknown sample quantitative measurement is achievable.  Sample and reference holders: Al or Pt pans placed on constantan disc and are connected by a low-resistance heat flow path. A metallic disc made of constantan alloy is the primary means of heat transfer . Sample and reference sit on raised constantan discs.
  • 62.  Sensors • Chromel® (an alloy made of 90% nickel and 10% chromium)- constantan area thermocouples (differential heat flow) • Chromel®-alumel (an alloy consisting of approximately 95% nickel, 2% manganese, 2% aluminium and 1% silicon) thermocouples (sample temperature)  Furnace - same block is used for sample and reference.  Temperature controller: • The temperature difference between the sample and reference is converted to differential thermal power, which is supplied to the heaters to maintain the temperature of the sample and reference at the program value
  • 63.  Differential heat flow to sample and reference is measured by thermocouples which are connected in series, located at the junction of constantan disc and chromel wafers.  With this, it is possible to achieve heating or cooling rates of 1000c /min to 00c/min (isothermal).  It needs mathematical equations to get the heat flow .  DSC Heat Flow Equation  Dh/Dt - DSC Heat Flow Signal, Cp - Sample Heat Capacity = Sample Specific Heat X Sample Weight, Dt/Dt - Heating Rate  F(t,t) - Heat Flow That Is A Function Of Time At An Absolute Temperature (Kinetic) Dh/Dt = Cp Dt/Dt + F(t,t)
  • 64. Power compensation DSC  Both sample and reference pans are heated by a different furnaces. When an event occurs in the sample, sensitive Platinum Resistance Thermometer (PRT) detects the changes in the sample, and power (energy) is applied to or removed from the sample furnace to compensate for the change in heat flow to or from the sample. As a result, the system is maintained at a “thermal null” state at all times. The amount of power required to maintain system equilibrium is directly proportional to the energy changes occurring in the sample. No complex heat flux equations are necessary with a power compensation DSC because the system directly measures energy flow to and from the sample.  In addition, PC type DSC has enhanced modulated temperature DSC (StepScan) technique and fast scan DSC (Hyper DSC) for dramatic improvements in productivity, as well as greater sensitivity.  Furthermore, the heating and cooling rate of PC types DSC can be as high as 500°C/min.
  • 65.  Sample holder • Aluminum or Platinum pans  Sensors • Platinum resistance thermocouples • Separate sensors and heaters for the sample and reference  Furnace • Separate blocks for sample and reference cells  Temperature controller • Supply the differential thermal power to the heaters to maintain the temperature of the sample and reference at the program value
  • 66.  Accurately weigh samples (approx. 3 to 20 mg). encapsulate in the metal pans of high thermal conductivity materials (aluminium, platinum, stainless steel) are used.  Pan configurations may be open, pinhole or hermetically sealed. Same pan material and configuration for both sample and reference. Material should entirely cover the bottom of the pan to ensure thermal contact.  Avoid overfilling of the pan to minimize the thermal lag from the bulk of the material to the sensor . Small sample masses and low heating rates improve resolution [Do not Decompose The Samples In DSC Cell]  Sample Shape: Cut the sample to uniform shape, do not crush the sample. If the sample to be taken is pellet, cross section is to be taken. If the sample material is powder then, it is spread uniformly over the bottom of the sample pan.
  • 67.  Using Sample Press: When using crimped pans, the pans should not be over crimped. The bottom of the pans should remain flat, even after crimping. When using hermetic pans, a little more pressure is required to crimp the pans. Hermetic pans are sealed by forming a cold wield on the aluminium pans.  Sample Size: Smaller samples will increase the resolution but will decrease the sensitivity. Larger samples will decrease the resolution but will increase the sensitivity. Sample size depends on the type of material being measured. If the sample is Extremely reactive in nature - very small samples (<1 mg) are to be taken. Pure organics or pharmaceuticals - 1 to 5 mg Polymers - approximately 10 mg Composite materials -15 to 20 mg
  • 68.  Reference Materials - An inert material like α-alumina is generally used. Empty pan can also be used, if the sample weight is small. With higher sample weights it is necessary to use a reference material, because the total weight of the sample and its container should be approximately the same as the total weight of the reference and its container . The reference material should be selected so that it possesses similar thermal characteristics to the sample.  The most widely used reference material is α-alumina, which must be of analytical reagent quality . Before use, α-alumina should be recalcined and stored over magnesium perchlorate in a desiccator . Kieselguhr is another reference material normally used when the sample has a fibrous nature.  If it doesn’t have similar thermal characteristics, dilutions are made. Dilutions are done by thoroughly mixing suitable proportions of sample and reference material.
  • 69.  Purge Gases - Sample may react with air and may oxidize or burn. The problem is overcome by using inert gases. Inert gases are used to control moisture in the surrounding atmosphere.  Commonly used inert gases are nitrogen, helium, argon etc.  Nitrogen - It is the most commonly used inert gas. It increases the sensitivity of the experiment. Typical flow rate is 50 ml/min.  Helium - It has high thermal conductivity . It increases the resolution of the peaks. The upper temperature limit for this gas is upto 3500c. Flow rate is 25 ml/min  Air or oxygen - Sometimes it is deliberately used to view oxidative effects of the sample. Flow rate is 50 ml/min  Heating Rate - Faster heating rate will increase the sensitivity but will decrease the resolution. Slow heating rate will decrease the sensitivity but will increase the resolution. Good starting point is 100c/min.
  • 70.  The ultimate goal of calibration and adjustment is to provide a measuring system that consistently delivers reproducible and accurate results. • Evaluation of the thermal resistance of the sample and reference sensors • Measurements over the temperature range of interest  2-Step process • The temperature difference of two empty crucibles is measured • The thermal response is then acquired for a standard material, usually sapphire, on both the sample and reference platforms  Amplified DSC signal is automatically varied with the temperature to maintain a constant calorimetric sensitivity with temperature.
  • 71. Temperature  Goal is to match the melting onset temperatures indicated by the furnace Thermocouple readouts to the known melting points of standards analyzed by DSC.  Should be calibrated as close as possible to the desired temperature range.  Celebrant's • High purity • Accurately known enthalpies • Thermally stable • Light stable • Non-hygroscopic • Un-reactive (pan, atmosphere)
  • 72.  The result of a DSC experiment is a curve of heat flux versus temperature or versus time. There are two different conventions: exothermic reactions in the sample shown with a positive or negative peak, depending on the kind of technology used in the experiment.  This curve can be used to calculate enthalpies of transitions.  Area under the peak is directly proportional to heat absorbed or evolved by the reaction.  Height of the peak is directly proportional to rate of the reaction.
  • 73. Factors affecting DSC curve/ Thermogram Instrumental factors Sample characteristics Furnace heating rate Amount of sample Recording or Chart speed Nature of sample Furnace atmosphere Sample packing Geometry of sample holders/ location of sensors Particle size Sensitivity of the recording system Solubility of evolved gases in the sample Composition of sample containers Heat of reaction Thermal conductivity
  • 74. 1) Sample shape: The shape of the sample has little effect on the quantitative aspect of DSC but more effect on the qualitative aspects. samples in the form of a disc film or powder spread on the pan are preferred. In the case of polymeric sheets, a disc cut with a cork-borer gives good results. 2)Sample size: About 0.5 to 10mg is usually sufficient. Smaller samples enable faster scanning, give better shaped peaks with good resolution and provide better contact with the gaseous environment. With larger samples, smaller heats of transitions may be measured with greater precision. 3) Atmosphere and geometry of sample holders: There are a number of variables that affect DSC results includes the type of pan, heating rate, the nature and mass of the compound, the particle size distribution, packaging and porosity, pre-treatment and dilution of the sample. It is used for purity analysis of above 98% pure compounds.
  • 75.  Calibration  Contamination  Sample preparation – how sample is loaded into a pan  Residual solvents and moisture.  Thermal lag •Heating/Cooling rates •Sample mass  Processing errors
  • 76.  Determination of purity of the sample.  Detection of polymorphism and quantification of Polymorph, Detection of metastable polymorphs.  Detection of isomerism.  Stability and compatibility studies.  Lyophilization studies.  Estimation of amorphous content in excipients.  Percentage crystallinity Determination.  Characterization of Membranes, lipids, nucleic acids & micellar systems.  Assessment of the effects of structural change on a molecules stability.  Measurement of Ultra-light molecular interactions  Assessment of biocompatibility during manufacturing.  Protein Stability and Folding  Liquid Biopharmaceutical Formulations and Process Development
  • 77.
  • 78.  https://www.hitachi- hightech.com/global/products/science/tech/ana/thermal/descriptio ns/ta.html  Instrumental Methods of Chemical Analysis B. K. Sharma  Instrumental methods of analysis, seventh edition – Willard, Merritt, Dean, Settle.  Instrumental methods of analysis – M. Calvin, S.C. Bhatia .  Basic Concepts of Analytical Chemistry – Third edition by S.M. Khopkar .  Introduction to Instrumental Analysis – Robert D. Braun.  Instrumental methods in Pharmaceutical Analysis– Dr. U.B.Hadkar  Pharmaceutical Analysis- volume II – A.V.Kasture, K.R.Mahdik, S.G.Wadodkar, H.N.More.  www.slideshare.com  www.google.co.in  https://nptel.ac.in/courses/115103030/module4/lec22/1.html  www.wikipedia.org