2. Hyphenated technique
• The hyphenated technique is the combination or
the coupling of the different
analytical techniques. Mainly
chromatographic techniques are combined with
spectroscopic techniques.
• The hyphenated technique is developed from the
coupling of a separation technique and an on-line
spectroscopic detection technology
• e.g., GC-MS, LC-MS, LC-FTIR, LC-NMR, CE-MS, etc
3. Capillary electrophoresis–mass
spectrometry (CE-MS)
• It is an analytical chemistry technique formed by the
combination of the liquid separation process of capillary
electrophoresis with mass spectrometry
• CE-MS combines advantages of both CE and MS to provide
high separation efficiency and molecular mass information in
a single analysis
• It has high resolving power and sensitivity, requires minimal
volume and can analyze at high speed. Ions are typically
formed by electrospray ionization , but they can also be
formed by matrix-assisted laser desorption/ionization or
other ionization techniques.
• It has applications in basic research
in proteomics and quantitative analysis of biomolecules as
well as in clinical medicine
4. • Since its introduction in 1987, new developments and
application has made CE-MS powerful separation and
identification technique
• Use of CE-MS has increased for protein and peptides
analysis and other biomolecules. However, the
development of online CE-MS is not without challenges
• Understanding of CE, the interface setup, ionization
technique and mass detection system is important to
tackle problems while coupling capillary
electrophoresis to mass spectrometry
• It has applications in basic research
in proteomics and quantitative
analysis of biomolecules as well as in clinical medicine
5. • Since its introduction in 1987, new developments and
application has made CE-MS powerful separation and
identification technique. Use of CE-MS has increased for
protein and peptides analysis and other biomolecules.
However, the development of online CE-MS is not without
challenges. Understanding of CE, the interface setup,
ionization technique and mass detection system is important
to tackle problems while coupling capillary electrophoresis to
mass spectrometry
7. History
The original interface between capillary zone
electrophoresis and mass spectrometry was
developed in 1987 by Richard D. Smith and
coworkers at Pacific Northwest National
Laboratory, and who also later were involved in
development of interfaces with other CE
variants, including capillary isotachophoresis
and capillary isoelectric focusing
8. Interfacing CE with MS
• Capillary electrophoresis is a separation technique
which uses high electric field to produce
electroosmotic flow for separation of ions
• Analytes migrate from one end of capillary to other
based on their charge, viscosity and size. Higher the
electric field, greater is the mobility. Mass
spectrometry is an analytical technique that identifies
chemical species depending on their mass-to-charge
ratio. During the process, an ion source will convert
molecules coming from CE to ions that can then be
manipulated using electric and magnetic field. The
separated ions are then measured using a detector
9. • The major problem faced when coupling CE to
MS arises due to insufficient understanding of
fundamental processes when two techniques
are interfaced. The separation and detection
of analytes can be improved with better
interface. CE has been coupled to MS using
various ionization techniques like FAB, ESI,
MALDI, APCI and DESI. The most used
ionization technique is ESI
10. Electrospray ionization interface
The first not CE-MS interface had cathode end of CE
capillary terminated within a stainless steel capillary.
An electrical contact was made at that point completing
the circuit and initiating the electrospray. This interface
system had few drawbacks like mismatch in the flow rates
of two systems. Since then, interface system has been
improved to have continuous flow rate and good
electrical contact. At present, three types of interface
system exist for CE/ESI-MS which are discussed briefly
11. Sheathless interface
• CE capillary is coupled directly to an electrospray ionization source
with a sheathless interface system. The electric contact for ESI is
realized by using capillary coated with conductive metal.Because no
sheath liquid is used, the system has high sensitivity, low flow rates
and minimum background. However, these interface designs, all
have challenges including low mechanical robustness, poor
reproducibility.
• The latest sheathless interface design features porous ESI emitter
through chemical etching. This design effectively provides robust
interfacing with mass spectrometry and addresses the
reproducibility challenges associated with previous designs. This
porous emitter interface has been explored to couple of CITP/CZE
(or transient ITP) which greatly improves sample loading capacity of
CE and enabled ultrasensitive detection of trace analytes
• Another key factor for successful CE-MS interface is the choice of
buffer solution which must be suitable for both CE separation and
ESI operation
12. Another key factor for successful CE-MS interface is the choice of buffer
solution which must be suitable for both CE separation and ESI
operation
13. Sheath flow interface
• With the sheath flow interface, the electrical connection is
established when the CE separation liquid is mixed with
sheath liquid flowing coaxially in a metal capillary tubing.
• Commonly used sheath liquid is 1:1 mixture of water-
methanol with 0.1% acetic acid or formic acid.
• The system is more reliable and has wide selection range of
separation electrolyte.
• There might be some decrease in sensitivity due to sheath
liquid.
14.
15. Liquid junction interface
• This technique uses a stainless steel tee to mix
separation electrolyte from CE capillary with
make up liquid. The CE capillary and ESI needle
are inserted through opposite sides of the tee
and a narrow gap is maintained.
• The electrical contact is established by makeup
liquid surrounding the junction between two
capillaries. This system is easy to operate.
However, the sensitivity is reduced and the
mixing of two liquids could degrade separation
16. Continuous-flow fast atom
bombardment
• CE can be coupled to fast atom bombardment
ionization using a continuous flow interface. The
interface must match the flow rate between the
two systems
• The CF-FAB raquires a relatively high flow rate but
CE need low flow rate for better separation. A
make-up flow can be used using a sheath flow or
liquid junction
17. Applications
• CE-MS ability to separate analytes present in extremely low
concentration with high efficiency at high speed has made
it applicable in all fields of science. CE-MS has been used
for bioanalytical, pharmaceuticals, environmental and
forensic application.
• The major application of CE-MS has been for biological
studies, mostly for protein and peptide analysis. Along with
that, it is used often for routine analysis of pharmaceutical
drugs. There are number of studies reporting
characterization of mixtures of peptides and proteins.
• CE-MS can be used for routine clinical checkup. Body fluids
like blood and urine have been analyzed with CE-MS to
identify biomarkers for renal diseases and cancer.