2. Objectives of this training
Awareness about instrument calibration
Acceptance criteria
Method used
Standards / Master used
Correct way of measurement
Do and Do not for each instrument usage
Correct instrument selection
Knowledge of how to ensure instrument is
acceptable for use
Grooming Inspectors for instrument Calibration
3. Calibration
Calibration- is the process to evaluate
relationship between unknown
values(Gauge) with known values(Master).
1/10th Rule-Master used for
calibration must be 10 times precise than
that instrument used in measurement.
4. Temperature for Calibration.
Internationally accepted
temperature for measurement is
20º and all instruments are
calibrated at this temp.
Temperature needs to be
maintain constantly because of
behavior of materials at
different temperature varies.
5. NABL
National Accreditation Board
For Testing And Calibrating
Laboratories-This is body which
accredits testing and calibration
laboratories for their competence
to perform consistently and
accurately.
6. Least Count
It is least value measured by an instrument.
For Vernier-
L.C= Smallest division on main scale
No of division on vernier scale
=1/50 =0.02 mm
For Micrometer-
L.C=Smallest division on main scale
No of divisions on micrometer scale
=0.5/50=0.01 mm
7. Calibration Sticker
I – Type Of Instrument
II – Sr. No. Of Instrument
III – Location Of Instrument
X
(III)
XXX
(II)
XXX
(I)
8. Masters used for Calibration
Slip Gauges Grade-0.
Dial Calibrator.
Digital Plunger Dial.
Sine Bar.
Electronic Level Meter.
Surface Plate Grade 0.
Wear Gauges.
Lever Dial
Digital Micrometer.
9. Instruments being calibrated In-house
Vernier calipers
Analog & Digital Micrometers
Plunger & Lever Dials (L.C. 0.01mm)
Bore Gauges (L.C. 0.01mm)
Depth Micrometers
Inside Micrometers
Surface Plates
Electronic Level Meters
Height gauges
Thread Plug Gauges up to M12
Bevel Protractors
Measuring Pins
10. Precaution to be taken during Instrument use
Don’t apply additional pressure during measurement to
inside jaws & Outside jaws of verniers. It may Lead to
Out ness of Parallelism of jaws.
Don’t apply extra pressure to Depth blade of vernier
caliper. It may lead to Bending of Blade.
Don’t apply additional pressure to External / Internal
micrometers. It can Lead to have a zero error to
Instruments.
Don’t apply additional pressure to Thread gauges. It
may lead to wear out of gauges where tolerances are
very tight.
Don’t apply additional pressure to Plunger dial. It may
lead to damage of spring or variation in Repeatability of
Plunger dials.
16. External & Depth measurement :
1000
10
± 40
± 30
700
9
500
8
300
7 ± 40
± 30
200
6
150
5
100
4
50
3
30
2
± 20
± 10
10
1
For Depth Measurment
ForExt. Measurement
Ver
Dig
Digital Caliper
Observed Error (mm)
Permissible (m)
Nominal Value
of Slip Size
(mm)
Sr.
No.
Parallelity of External Jaws (Hold 10 mm Slip) :
0.02 upto 300 mm range
0.04 mm above 300 mm range
Parallelism
(mm)
Top of Jaw (mm)
Middle of Jaw (mm)
Bottom of Jaw (mm)
Permissible Error (mm)
Internal Jaws
0.04 mm
0.02 mm
10 mm
or ____ mm
Permissible Error in Measurement
Actual Error
in
Measurement
Permissible
Error in
Parallelism
Actual
Error in
Parallelism
Reading on
Bottom of
the Jaw
Reading
on Top of
the Jaw
Set Slip
Guage
Acceptance Criteria for Vernier Caliper
21. Aceeptance limits for Micrometer
For Ext. Micrometer For Setting Rods
______
______
______
______
______
______
______
______
______
______
±0.002 upto 100
mm
±0.003 upto 300
mm
±0.005 upto 500
mm
±0.020 above
500 mm
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
________
________
________
________
________
________
________
________
________
________
5 microns for
upto 500 mm
range
2.5 _________
5.1 _________
7.7 _________
10.3 ________
12.9 _________
15.0 _________
17.6 _________
20.2 _________
22.8 _________
25.0 _________
ERROR
OBSVD.
(mm)
PERMISSIBLE
ERROR
(mm)
NOMINAL
ROD SIZE
(mm)
ERROR
OBSVD.
(mm)
PERMISSIBLE
ERROR
PARAMETERS
(mm)
22. Plunger And Lever Dials
Standard used-IS-2092 for Plunger Dials
IS- 11498 for Lever Dials
Master used- Dial Calibrator
Features to be calibrated-
Determination of Accuracy.(Rotation)
Clockwise.
Anticlockwise.
Repeatability
Sensitivity
23. Dial Calibration With Dial Calibrator
Dial
Calibrator
Plunger
Dial
Dial
Calibrator
Lever
Dial
24. Acceptance Criteria for Lever Dials
ACLK
For :
0.1mm - 5m
1mm -30m
CLK
TURN 1
Errors in
SPEC. RANGE
0
10
20
30
40
30
20
10
0
DIV OF DIAL
0.002
Sensitivity (mm)
0.003
Repeaibility (mm)
Acceptable
Observed Reading
25. Bore Gauge
Master used- Dial Calibrator
Features to be calibrated
Accuracy
27. Acceptance Criteria fro Bore Gauge
Turn 2
Deviation
( )
1/10th
of Rev -5
1 Rev-10
2 Rev-15
Turn 1
Deviation
( )
Allow. Dev.
as per CAL/
50.70
90
80
70
60
50
40
30
20
10
0
Reading on
Calibrator
Div. of
Dial
28. Electronic Level Meter (Wyler)
Basic Principle
The Measuring principle is based on a
changing of condenser built in two electrodes
and pendulum. The measured change of
capacity is directly influenced by change of
the inclination of the pendulum.This capacity
change is primarily used signal for angle to be
measured.
Master Used – Sine Bar
Feature to be measured
1. Angular Error
32. Acceptance Criteria for Wyler
15 /1000 mm.
Perpendicularity of Vertical Face
2
2
2
2
4
8
Inclination per 200 mm
25
50
75
100
500
1500
For Measuring and Reference Unit
2
2
2
2
2
4
8
Inclination per 200 mm
25
50
75
100
500
1500
For Measuring Unit
1
Observed Error () in 200mm.
Allowable Deviation in per
200 mm.
Height Difference between
slips (m )
Sr. No.
Measuring accuracy for different value of inclination :
33. Surface Plates
Standard Used- IS-7327
Master Used- Electronic Level
Meter (Wyler)
Features to be calibrated
Flatness
35. Inside Micrometer
Standard used- IS-2966
Master used- Slip Gauge
Digital Micrometer
Digital Dial
Features to be checked:
Zero Setting.
Linearity
36. Zero Error & Linearity check
for Inside Micrometer Head
Inside
Micrometer
Head
Digital Micrometer
41. Acceptance Criteria For Depth Micrometer
Set Value(mm) Error (mm)
Size (mm) Permissible
Error
0 – 25 5 Microns
25 – 300 10 Microns
I. Depth Bar 1
i. Smallest Size of Depth Bar +2.5
ii. Smallest Size +10.3
iii. Smallest Size +17.6
iv. Smallest Size +25
_____________
_____________
_____________
_____________
___________
___________
___________
___________
I. Depth Bar 2
II. Depth Bar 3
III. Depth Bar 4
IV. Depth Bar 5
V. Depth Bar 6
VI. Depth Bar 7
VII. Depth Bar 8
VIII. Depth Bar 9
IX. Depth Bar 10
X. Depth Bar 11
XI. Depth Bar 12
25
50
75
100
125
150
175
200
225
250
275
300
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
42. Height Gauge
Standard used- IS 2921
Master used- Slip Gauges Grade 0
Granite Square
Features to be checked:
Perpendicularity
Linearity
48. Bevel Protractor
Standard used- IS-4239
Master used- Sine Bar
Plunger Dial
Slip Gauge
Features to be checked-
Zero Setting
Angle Measurement