2. CONTENTS
Introduction.
Modes of separation.
Instrumentation.
HPLC method development,optimization and
validation.
References.
September 29, 2014 2
3. INTRODUCTION
Most of the drugs in multicomponent dosage forms can be
analyzed by HPLC method because of the several
advantages like:
Improved resolution of the separated substances
Faster separation times and
The improved accuracy, precision, & sensitivity with
which the separated substances may be quantified.
September 29, 2014 3
4. MODES OF SEPARATION IN HPLC
There are different modes of separation in HPLC:
1.Normal phase mode.
2.Reversed phase mode.
3.Reversed phase ion pair chromatography.
4.Affinity chromatography.
5.Size exclusion chromatography
September 29, 2014 4
5. NORMAL PHASE MODE
Stationary phase-polar in nature.
Mobile phase-non polar in nature.
Generally used for separation of non polar compounds.
September 29, 2014 5
6. REvERSED PHASE MODE
Stationary phase-non polar .
Mobile phase-polar in nature.
Generally used for separation of polar compounds.
September 29, 2014 6
7. ION EXCHANGE CHROMATOGRAPHY :
The stationary phase contains ionic groups like NR⁺з, SO⁻з
which interact with the ionic groups of the sample
molecules.
This method is suitable for the separation of charged
molecules only.
ION PAIR CHROMATOGRAPHY :
This may be used for the separation of ionic compounds.
Strong acids & basic compounds may be separated by
reversed phase mode by forming ion pairs with suitable
counter ions.
September 29, 2014 7
8. AFFINITY CHROMATOGRAPHY:
• It uses highly specific biochemical interactions for
separations.
• The stationary phase contains specific groups of
molecules which can absorb the sample if certain steric &
charge related conditions are satisfied.
• This technique can be used to isolate proteins, enzymes,
as wel as antibodies from complex mixture.
SIZE EXCLUSION CHROMATOGRAPHY:
• Separates molecules according to their molecular mass.
• Largest molecules are eluted first and smaller molecules
last.
September 29, 2014 8
9. PRINCIPLE OF SEPARATION
The principle of separation is adsorption.
Separation of components takes place because of the
difference in affinity of compounds towards stationary
phase.
September 29, 2014 9
11. vARIOUS COMPONENTS OF HPLC
1. A solvent delivery system including pump.
2. Sample injection system
3. A chromatographic column
4. Detectors
5. Recorders and integrators
September 29, 2014 11
12. A SOLvENT DELIvERY SYSTEM
A mobile phase is pumped under pressure from one or
several reservoir and flows through the column at a
constant rate.
For normal phase separation eluting power increases
with increasing polarity of the solvent but for reversed
phase separation, eluting power decreases with
increasing polarity.
A degasser is needed to remove dissolved air and other
Septegmbaesr 2e9,s 2 0f1r4om the solvent. 12
14. PUMPS
The pump is one of the most important component of
HPLC, since its performance directly affects retention
time, reproducibility and detector sensitivity.
Three main types of pumps are used in HPLC.
1.Displacement pump
2.Reciprocating pump
3.Pneumatic or constant pressure pump
September 29, 2014 14
15. DISPLACEMENT PUMP: It produce a flow that
tends to independent of viscosity and back pressure and
also output is pulse free but possesses limited capacity
(250ml).
RECIPROCATING PUMP: It has small internal
volume (35-400μl), their high output pressure(up to
10,000psi) and their constant flow rates. But it produces
a pulsed flow.
PNEUMATIC OR CONSTANT PRESSURE
PUMP: They are pulse free . Suffer from limited capacity as well
as a dependence of flow rate on solvent viscosity and column back
pressure. They are limited to pressure less than 2000psi.
September 29, 2014 15
16. SAMPLE INJECTION SYSTEM
There are three important ways of
introducing the sample in to the
injection port.
Loop injection : in which a fixed amount
of volume is introduced by making use
of fixed volume loop injector.
Valve injection: in which, a variable
volume is introduced by making use of
an injection valve.
On column injection: in which, a
variable volume is introduced by
means of a syringe through a septum.
September 29, 2014 16
18. Varian 9010 Solvent Delivery
System
Rheodyne
Injector
%A %B %C Flow Rate Pressure
{H2O} {MeOH} (mL/min) (atmos.)
Ready
A
C
B
Ternary Pump
from solvent
reservoir
Column
to
detector
to column
through
pulse
dampener
to injector
through pump
load
inject
September 29, 2014 18
19. CHROMATOGRAPHIC COLUMN
• The column is usually made up of heavy glass or stainless
steel tubule to withstand high pressure .
• The columns are usually 10-30cm long and 4-10mm inside
diameter containing stationary phase at particle diameter of
25μm or less.
• Column with internal diameter of 5mm give good results
because of compromise between efficiency, sample capacity,
and the amount of packaging and solvent required.
September 29, 2014 19
20. DETECTORS
The function of detector in HPLC is to monitor the mobile
phase as it merges from the column.
Detectors are usually of two types:
1. Bulk property detectors: It compares overall changes
in a physical property of the mobile phase with and
without an eluting solute e.g. refractive index ,dielectric
constant or density.
2.Solute property detectors: It responds to a physical
property of the solute which is not exbited by the pure
mobile phase.e.g.UV absorbance,fluoroscence or
diffusion current.
September 29, 2014 20
21. TYPES OF DETECTORS
There are mainly 4 types of detectors are used in HPLC:
1. Photometric detectors.
Single wavelength detectors.
Multiwavelength detectors.
Variable wavelength detectors.
Programmable detectors.
Diode array detectors .
2. Fluorescence detectors.
3. Refractive index detectors.
4. Electrochemical detectors.
September 29, 2014 21
22. PHOTOMETRIC DETECTORS
• These normally operate in the ultra violet region of the
spectrum .
• Most extensively used in pharmaceutical analysis.
September 29, 2014 22
23. SINGLE WAVELENGTH DETECTORS
Equipped with a low pressure mercury discharge lamp.
The absorbance is measured at the wavelength of
mercury at 254nm.
September 29, 2014 23
24. MULTIWAVELENGTH DETECTORS
• Employ mercury and other discharge sources.
• When used in combination with interference filters
,allow a no of monochromatic wavelengths to be
selected e.g. 206, 226, 280 , 313, 340 or 365 nm.
September 29, 2014 24
26. variable wavelength Detectors
• Use a deuterium light source.
• A grating monochromator to allow selection of any
wavelength in deuterium continuum (190-360nm).
September 29, 2014 26
28. ABS AUFS l RunTime EndTime
0.001 2.000 238 0.00 min 10.0 min
Ready
September 29, 2014 28
29. prograMMable Detectors
Allow the automatic change of wavelength between and
during the chromatographic analysis.
September 29, 2014 29
30. DioDe arraY Detectors
They are microprocessor – controlled photodiode array
spectrophotometers in which light from an UV source passes
through the flow cell into a polychromator which disperses
the beam so that the full spectrum falls on the array of
diodes.
September 29, 2014 30
32. FLUORESCENCE DETECTORS
• These are essentially filter fluorimeter or spectro
-fluorimeters equipped with grating monochromators,
and micro flow cell.
• Their sensitivity depends on the fluorescence
properties of the components in the eluate.
September 29, 2014 32
34. REFRACTIVE INDEX DETECTORS
• Which respond to the change in the bulk property of the
refractive index of the solution of the component in the
mobile solvent system.
• The sensitivity of the refractive index detector is much
less than that of specific solute property detectors, they
are useful for the detection of substances(e.g
,carbohydrates& alcohols) which do not exhibit other
properties that can be used as the basis for specific
detection.
September 29, 2014 34
36. electrocheMical Detectors
• These are based on standard electrochemical principles
involving amperometry,voltametryand polarography.
• These detectors are very sensitive for substances that
are electroactive ,i.e. those that undergo oxidation or
reduction .
• They have found particular application in the assay of
low levels of endogenous catecholamines in biological
tissues,pesticides,tryptophan derivatives and many
drugs.
September 29, 2014 36
39. introDuction:
ANALYTICAL METHOD DEVELOPMENT:
• Method development usually requires selecting the
method requirements and deciding on what type of
instrumentation to utilize and why.
• The wide variety of equipment, columns, eluent and
operational parameters involved makes HPLC method
development .
September 29, 2014 39
40. There are several reasons for developing new methods of
analysis:
1. A suitable method for particular analyte in the specific
matrix is not available.
2.Existing methods may be too error or they may be
unreliable (have poor accuracy or precision)
3.Existing methods may be too expensive, time
consuming.
September 29, 2014 40
41. HPLC method development generally follows the following
steps:
Step 1-selection of the HPLC method and initial system.
Step2-Selection of optimum conditions.
Step3-selectivity optimization.
Step4-system parameter optimization.
Step5-method validation.
September 29, 2014 41
42. Step 3a
Initial HPLC
condition
Step 2
Sample
Step 1 preparation
Method goals and
chemistry
Step 3b
Optimize HPLC
separation
Step 4
Standardization
Step 5
Method validation
Fig 2.2 Pie diagram showing the time that should be spent on different steps of
the method development to meet the commended timeline. The sequence of events
and percentage of time allocated is only suggestive.
September 29, 2014 42
43. step1-selection of hplc MethoD anD
initial conDitions
SELECTION OF HPLC METHOD:
• When selecting an HPLC system it must have a high propability
of actually being able to analyse the sample.
• For example if the sample includes polar analytes then RP-HPLC
would offer both adequate retention and ressolution.
consideration must be given to the following ;
a. sample preparation
b.Types of chromatography
c. Column selection
d.Detector selection
e. Selection of mobile phase composition September 29, 2014 43
44. saMple preparation
Sample preparation is an essential part of HPLC
analysis, to provide a reproducible and homogenous
solution i.e. suitable for injection on to the column.
The aim of sample preparation is a sample that :
1. It is relatively free of interferences
2.Will not damage the column
3.Should compatible with the intended HPLC method ;
September 29, 2014 44
45. TYPES OF CHROMATOGRAPHY
Reversed phase is the choice for the majority of the
samples.
But if acidic or basic analytes are present reversed
phase ion suppression (for weak acids and bases) or
reversed phase ion pairing (for strong acids and bases)
should be used.
For low or medium polarity analytes normal phase
HPLC is used, particularly if the separation of isomers
is required.
For inorganic anion or cation analysis ion exchange
chromatography is best.
September 29, 2014 45
46. Size exclusion chromatography would normally be
considered for analyzing high molecular weight
compounds.
Gradient HPLC only a requirement for complex samples
with a large number of components (20-30) .
Reversed phase HPLC is commonly used in peptide and
small protein analysis using an acetonitrile –water
mobile phase containing 1% trifluoroethanolic acid.
September 29, 2014 46
47. COLUMN SELECTION
A column is chosen based on the
Knowledge of sample
On the expectation of how its components will interact
with the packing material.
the properties of column packing material.
September 29, 2014 47
48. COLUMN SELECTION
1.Knowledge of the Sample: which influences the
choice of Column Bonded Phase characteristics
Knowledge of the sample
• Structure of sample components?
• Number of compounds present?
• Sample matrix?
• pKa values of sample components?
• Concentration range?
• Molecular weight range?
• Solubility?
• Other pertinent data?
Column Chemistry
(bonded phase, bonding
type, endcapping,
carbon load)
}
September 29, 2014 48
49. Pac king material:-
Most HPLC separations are performed on bonded
phase HPLC columns
Octadecyl-derivatized silica gel columns are the most
widely used bonded phase columns in the reverse phase
mode.
Commonly used polar bonded phases certain diol,cyano
or amino functional groups.
Silica based packing materials are used in about 75% of
all HPLC separations performed today.
September 29, 2014 49
50. • due to
the physical stability
the availability of bonded phase
and to the high efficiency of silica based HPLC columns
• Many new resin packings have been introduced in recent
years particularly for biochemical analysis.
• The most well known resin based packings are formed by
the co polymerization of polysterene and divinyl benzene.
• Other packing materials such as alumina, titania and
zirconia have also been employed for bio polymer analysis.
September 29, 2014 50
51. COLUMN DIMENSIONS
Effect on chromatography
Column Dimension
• Short (30-50mm) - short run times, low backpressure
• Long (250-300mm) - higher resolution, long run times
• Narrow (£ 2.1mm) - higher detector sensitivity
• Wide (10-22mm) - high sample loading
September 29, 2014 51
52. DETECTOR SELECTION
Consideration must given to the following:
Do the analytes have chromophores to enable UV detection ?
Is more selective or sensitive detection required?
What detection limits are necessary ?
Will the sample require chemical derivatization to enhance
detect ability and /or improve the chromatography.
September 29, 2014 52
53. DETECTOR SELECTION
A HPLC detector will have a number of performance
characteristics that need to be specified and known before a
particular detector can be chosen for a specific application
and are listed as follows.
1. Dynamic range
2. Response index or linearity
3. Linear dynamic range
4.Detector response
5. Detector noise level
6.Detector sensitivity, or minimum detectable concentration.
7.Total system dispersion
8.Pressure sensitivity
9.Flow arte sensitivity.
10.Operating temperature range. 53
September 29, 2014
54. THE UV DETECTOR
Limited to the detection of those substances that absorb
light in the UV wave length range.
UV detectors detects all sample components that contain
chromophores.
Specifications:
S.NO CHARACTERISTIC FIXED WAVELENGTH
UV DETECTOR
DIODE ARRAY
DETECTOR
1 Sensitivity
(solute benzene)
5×10ˉ⁸gml 1×10ˉ⁷gml
2 Linear dynamic
range
5×10ˉ⁸ to
5×10ˉ⁴gml
10ˉ⁸to
5×10ˉ⁴gml
3 Response index 0.98 - 1.02 0.97- 1.03
September 29, 2014 54
55. THE FLUORESCENCE DETECTOR
It can detect eluted solutes on the basis of fluorescence
,but it can also provide their fluorescence spectra.
Fluorescence and electro chemical detectors should be
used for trace analysis.
Specifications:
S.No CHARACTERISTIC FLUOROSCENCE
DETECTOR
1. Sensitivity
(solute anthracene)
1×10ˉ⁹gml
2 Linear dynamic range 1×10ˉ⁹ to 5×10ˉ⁶gml
3 Response index 0.96 – 1.04
September 29, 2014 55
56. ELECTRICAL CONDUCTIVITY DETECTOR
It is usually employed with an ion suppressor column to
allow salts and buffers to be used in the mobile phase
without affecting the detector output.
Specifications:
S.NO CHARACTERISTIC CONDUCTIVITY
DETECTOR
1. Sensitivity(sodium
chloride)
5×10ˉ⁹gml
2. Linear dynamic range 5×10ˉ⁹to 1×10ˉ6gml
3. Response index 0.97-1.03
September 29, 2014 56
57. REFRACTIVE INDEX DETECTOR
It is one of the least sensitive LC detectors and is used
circumstances where other detector s are inappropriate.
For preparative HPLC it is preferred because it can
handle concentration without overloading the detector
Specifications:
S.NO CHARACTERISTIC RI DETECTOR
1. Sensitivity(solute
benzene)
1×10ˉ⁶g/ml
2. Linear dynamic range 1×10ˉ⁶ -1×10ˉ⁴g/ml
3. Response index 0.97-1.03
September 29, 2014 57
58. MOBILE PHASE SELECTION
The organic phase concentration required for the mobile
phase can be estimated by gradient elution method.
Gradient can be started with 5-10 % of the organic phase in
the mobile phase and the organic phase concentration can
be increased up to 100% within 30-45%.
The elution strength of a mobile phase depends upon its
polarity, the stronger the polarity higher is the elution.
September 29, 2014 58
59. Ionic samples(acidic or basic) can be separated if they
are present in undissociated form. Dissociation of ionic
samples may be suppressed by the selection oh pH.
If the retention times are too long an increase of the
organic phase concentration is needed.
When tailing or fronting is observed, it means that the
mobile phase is not totally compatible with the solutes.
September 29, 2014 59
60. SELECTION OF INITIAL SYSTEM
It could based on :
assessment of the nature of the sample and analytes
together with literature data.
Experience
Expert system software and
Empirical approaches
September 29, 2014 60
61. STEP 2 : SELECTION OF INITIAL CONDITIONS
• This step determines the optimum conditions to adequately
retain all analytes ; i.e.
Ensures no analyte has a capacity factor of less than
0.5(poor retention could result in peak overlapping).
No analyte has a capacity factor greater than 10-15
(excessive retention leads to long analysis time and broad
peaks with poor detectability).
Determination of initial conditions:
• The recommended method involves performing two gradient
runs differ in only in the run time.
• A binary system based on either aceto nitrile/ water or
methanol/water should be used.
September 29, 2014 61
62. STEP 3: SELECTIVITY OPTIMIZATION
The aim of this step is to achieve adequate selectivity.
The mobile phase and stationary phase compositions
need to be taken in to account.
To select these the nature of the analytes must be
considered.
Once the analyte types are identified the relevant
optimization parameters may be selected.
September 29, 2014 62
63. STEP 4: SYSTEM PARAMETER OPTIMIZATION
This is used to find the desired balance between
ressolution and analysis time after satisfactory selectivity
has been achieved.
The parameters involve include column dimensions,
column packing particle size and flow rate.
This parameters may be changed without affecting
capacity factors or selectivity.
September 29, 2014 63
64. TYPES OF OPTIMIZATION
Two types :
Manually and
By using soft wares.
By Manual : separation then can be optimized by change in the
initial mobile phase composition and the slope of the gradient
according to the chromatogram obtained from the preliminary
run.
September 29, 2014 64
65. BY USING SOFT wARES :
Chemo metric protocols available for the development and
optimization of HPLC methods:
Experimental Design (ED)
Factorial design
Plackett-Burman design
D-optimal design
Two-level full factorial design
Central composite design
Box-Behnken design
Doehlert design
Multi-Criteria Decision Making (MCDM)
Overlay plots
Pareto optimality
Utility function
Derringer’s desirability function
September 29, 2014 65
67. The objective of validation of an analytical procedure is
to demonstrate that it is suitable for its intended
purpose.
September 29, 2014 67
68. A brief description of types of tests considered in this
document is provided below:-
Identification tests are intended to ensure the identity
of analyte in a sample. This normally achieved by
comparing the properties of sample with that of the
reference standard.
Testing for impurities can be either a quantitative test
or a limit test for the impurity in a sample.
Assay procedures are intended to measure the analyte
present in a given sample.
September 29, 2014 68
69. • Typical validation characteristics which should be
considered are listed below:-
1.Accuracy
2. Precision
a. repeatability
b. intermediate precision
3. specificity
4. Detection limit
5. Quantitation limit
6. Linearity
7. range.
September 29, 2014 69
70. ACCURACY
The accuracy of an analytical procedure expresses the
closeness of agreement between the value which is
accepted either as a conventional true or an accepted
reference value and the value found.
Accuracy should be assessed using a minimum of 9
determinations over a minimum of 3 concentration
levels covering the specified range
September 29, 2014 70
71. LINEARITY
The linearity of an analytical procedure is its ability
to obtain test results which are directly proportional to
the concentration of analyte in the sample.
Linearity should be evaluated by visual inspection of a
plot of signals as a function of analyte concentration or
content.
For the establishment of linearity, a minimum of 5
concentrations is recommended.
September 29, 2014 71
72. PRECISION
The precision of an analytical procedure expresses the
closeness of agreement between the measurements
obtained from multiple sampling of the same
homogenous sample under the prescribed conditions.
repeatability
Precision intermediate precision
reproducibility
September 29, 2014 72
73. Repeatability:- Expresses the precision under the
same operating conditions over a short interval of time .
Repeatability is also termed intra- assay precision.
• a) a minimum of 9 determinations covering the specified
range for the procedure (e.g., 3 concentrations/3
replicates each);
or
• b) a minimum of 6 determinations at 100% of the test
concentration.
September 29, 2014 73
74. Intermediate precision:- intermediate precision
expresses within – laboratories variations: different days,
different analysts, different equipment e.t.c.
Reproducibility:- reproducibility expresses the precision
between laboratories.
• The standard deviation, relative standard deviation
(coefficient of variation) and confidence interval should
be reported for each type of precision investigated.
September 29, 2014 74
75. DETECTION LIMIT
The detection limit of individual analytical procedure is
the lowest amount of analyte in a sample which can be
detected but not necessarily quantitated as an exact
value.
Several approaches for determining the detection limit
are possible, depending on whether the procedure is a
non-instrumental or instrumental. Approaches other than
those listed below may be acceptable
September 29, 2014 75
76. Based on Visual Evaluation
Visual evaluation may be used for non-instrumental
methods but may also be used with instrumental methods.
The detection limit is determined by the analysis of
samples with known concentrations of analyte and by
establishing the minimum level at which the analyte can
be reliably detected.
Based on Signal-to-Noise
This approach can only be applied to analytical
procedures which exhibit baseline noise.
Determination of the signal-to-noise ratio is performed by
comparing measured signals from samples with known
low concentrations of analyte.
A signal-to-noise ratio between 3 or 2:1 is generally
considered acceptable for estimating the detection limit.
September 29, 2014 76
77. Based on the Standard Deviation of the Response
and the Slope
The detection limit (DL) may be expressed as:
DL = 3.3 σ /s
Where, σ = the standard deviation of the response
S = the slope of the calibration curve
September 29, 2014 77
78. QUANTITATION LIMIT
• The quantitation limit of an individual analytical
procedure is the lowest amount of analyte in a sample
which can be quantitatively determined with suitable
precision and accuracy.
• The quantitation limit is a parameter of quantitative
assays for low levels of compounds in sample matrices
and is used particularly for the determination of
impurities and or degradation products.
September 29, 2014 78
79. Based on Visual Evaluation
Visual evaluation may be used for non-instrumental
methods but may also be used with instrumental methods.
The detection limit is determined by the analysis of
samples with known concentrations of analyte and by
establishing the minimum level at which the analyte can
be reliably detected.
Based on Signal-to-Noise
This approach can only be applied to analytical
procedures which exhibit baseline noise.
Determination of the signal-to-noise ratio is performed by
comparing measured signals from samples with known
low concentrations of analyte.
A signal-to-noise ratio between 10:1 is generally
considered acceptable for estimating the detection limit.
September 29, 2014 79
80. Based on the Standard Deviation of the Response
and the Slope
The quantitation limit (QL ) may be expressed as:
QL = 10 σ /s
Where, σ = the standard deviation of the response
S = the slope of the calibration curve
September 29, 2014 80
81. RANGE
The range of an analytical procedure is the interval
between the upper and lower concentration of the analyte
in the sample for which it has been demon started that the
analytical procedure has a suitable level of precision,
accuracy and linearity.
The following minimum specified ranges should be
considered:
September 29, 2014 81
82. for the assay of a drug substance or a finished (drug)
product: normally from 80 to 120 percent of the test
concentration;
for content uniformity, covering a minimum of 70 to 130
percent of the test concentration, unless a wider more
appropriate range, based on the nature of the dosage
form (e.g., metered dose inhalers), is justified;
for dissolution testing: +/-20 % over the specified range;
September 29, 2014 82
83. ROBUSTNESS
The robustness of an analytical procedure is a
measure of its capacity to remain unaffected by small,
but deliberated variations in method parameters and
provides an indication of its reliability during normal
usage.
September 29, 2014 83
84. REfERENCES
High performance liquid chromatography; principles and
methods in biotechnology by Elena.d katz .
Practical pharmaceutical chemistry part 2 fourth edition by
A.H.Beckett, J.B.stenlake; p.no 157.
Instrumental methods of chemical analysis by
Gurdeep.R.Chatwal, Sham.K Anand, p.no 2.624-2.639.
September 29, 2014 84