SlideShare a Scribd company logo
1 of 28
Laboratory Panel and Radiometer Calibration 28 July 2011, IGARSS, Vancouver Andrew Deadman 1 , Nigel Fox 1 , Irina Behnert 1 D Griffith 2 1  National Physical Laboratory (NPL), United Kingdom 2  Council for Scientific and Industrial Research (CSIR), South Africa
Laboratory Panel and Radiometer Calibration - Outline ,[object Object],[object Object],[object Object],[object Object]
Laboratory Panel and Radiometer Calibration ,[object Object],[object Object],[object Object]
Laboratory Panel and Radiometer Calibration ,[object Object],[object Object],[object Object],[object Object]
Laboratory Panel and Radiometer Calibration ,[object Object]
Laboratory Radiometer Comparison Measured before and after the field measurements. The difference used to check the stability check of the calibration set-up and also the performance of the radiometers. NPL provided a radiance source called TSARS (Transfer Standard Absolute Radiance Source)
Laboratory Radiometer comparison - Results First TSARS measurement – 15 August 2010
Laboratory Radiometer comparison - Results Second TSARS measurement – 26 August 2010
Laboratory Radiometer comparison - Results Difference between two TSARS measurements. Similar trend for most participants.
Laboratory Radiometer comparison Environmental conditions First TSARS calibration 15 Aug 2010 Second TSARS calibration 26 Aug 2010 Environmental conditions outside normal operating conditions for TSARS so absolute calibration was not possible.
Laboratory Radiometer Comparison 2009 TSARS calibration Data from 2009 – difference between before and after field measurements. Shows better TSARS performance
Laboratory Radiometer comparison - 2010 Results Difference between measurement of TSARS over the two days Problem with the performance of one radiometer. No absolute calibration but were able to establish bias between instruments
Laboratory Panel and Radiometer Calibration ,[object Object],[object Object]
Laboratory Panel Comparison 45/0 geometry Stable illumination conditions Same source and detector
Laboratory Panel Comparison Each panel was measured by comparison with an NPL calibrated reference panel. Measured before and after field measurements. Test stability of panels and robustness of the set-up
Laboratory Panel Comparison Set up tested at NPL before measurements in Turkey. Type A uncertainty  < ± 0.2% for most wavelengths of interest. Confidence in set-up.
Laboratory Panel Comparison - Results Results of laboratory panel comparisons. Calibrated reflectance factor values.
Laboratory Panel Comparison - Results Good agreement. No offsets in data or panel ageing affects. Differences less than the estimated calibration uncertainty of ± 1%. Difference in panel reflectance factor values before and after the field campaign.
Laboratory Panel Comparison - Results Comparison of laboratory calibration results and reported reflectance
Reasons for difference in panel reflectance factor values Laboratory  calibration Reported calibration Bidirectional illumination 45 ° ,  nadir view -  45/0 geometry Diffuse illumination, 8 ° view – 8°/hemispherical geometry Reflectance panels are not perfectly lambertian – reflectance factor greater than unity at low sun zenith angles
Laboratory Panel Comparison - Results Some participants had a goniometric calibration of their panel, so a direct comparison is possible. Very good agreement considering three independent methods used traceable to three national standards laboratories
Laboratory Panel and Radiometer Calibration ,[object Object],[object Object]
In-field Panel Comparison Repeat of laboratory exercise, using the sun as the source. Sun zenith angle between 30° and 35°. Each panel compared to NPL reference panel.
Laboratory & In-field panel calibration – expected result NPL laboratory calibration In situ calibrations Reported calibration value
Laboratory & In-field panel calibration – expected result NPL laboratory calibration In situ calibrations Reported calibration value
Reasons for difference in panel reflectance factor values Laboratory  calibration In field calibration Reported calibration Bidirectional illumination 45 º ,  nadir view -  45/0 geometry Diffuse and Direct illumination Sun zenith & irradiance variability Diffuse illumination, 8 º view – 8°/hemispherical geometry
In-field panel comparison Shows deviation of solar irradiance from mean over the period of in-field panel comparison. Corresponds well with AERONET, although there are short term variations not captured by AERONET. Data courtesy of CSIR.
Laboratory Panel and Radiometer Calibration ,[object Object],[object Object],[object Object],[object Object],[object Object]

More Related Content

Viewers also liked

Calibracion de Sondas de Humedad del Suelo ECH2O
Calibracion de Sondas de Humedad del Suelo ECH2OCalibracion de Sondas de Humedad del Suelo ECH2O
Calibracion de Sondas de Humedad del Suelo ECH2OLabFerrer LabFerrer
 
Tecnologia de informacion del senamhi
Tecnologia de informacion del senamhiTecnologia de informacion del senamhi
Tecnologia de informacion del senamhiJorge Chira
 
Jchira expo colombia
Jchira expo colombiaJchira expo colombia
Jchira expo colombiaJorge Chira
 
Tipos de sensores
Tipos de sensoresTipos de sensores
Tipos de sensoresford81
 
10959890 sensores-de-temperatura
10959890 sensores-de-temperatura10959890 sensores-de-temperatura
10959890 sensores-de-temperaturaenriquejanampa
 
SOLAR RADIATION MEASUREMENT
SOLAR RADIATION MEASUREMENTSOLAR RADIATION MEASUREMENT
SOLAR RADIATION MEASUREMENTVanita Thakkar
 
Radiacion ultravioleta y su aplicacion en Fisioterapia
Radiacion ultravioleta y su aplicacion en FisioterapiaRadiacion ultravioleta y su aplicacion en Fisioterapia
Radiacion ultravioleta y su aplicacion en FisioterapiaFernando Castillo
 
Instrument Calibration
Instrument Calibration Instrument Calibration
Instrument Calibration Ramesh Dham
 
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRYINSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRYAshwini Somayaji
 
Principle and instrumentation of UV-visible spectrophotometer.
Principle and instrumentation of UV-visible spectrophotometer.Principle and instrumentation of UV-visible spectrophotometer.
Principle and instrumentation of UV-visible spectrophotometer.Protik Biswas
 
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...mariomS7
 
3 Things Every Sales Team Needs to Be Thinking About in 2017
3 Things Every Sales Team Needs to Be Thinking About in 20173 Things Every Sales Team Needs to Be Thinking About in 2017
3 Things Every Sales Team Needs to Be Thinking About in 2017Drift
 

Viewers also liked (14)

Calibracion de Sondas de Humedad del Suelo ECH2O
Calibracion de Sondas de Humedad del Suelo ECH2OCalibracion de Sondas de Humedad del Suelo ECH2O
Calibracion de Sondas de Humedad del Suelo ECH2O
 
Tecnologia de informacion del senamhi
Tecnologia de informacion del senamhiTecnologia de informacion del senamhi
Tecnologia de informacion del senamhi
 
El sol
El solEl sol
El sol
 
Jchira expo colombia
Jchira expo colombiaJchira expo colombia
Jchira expo colombia
 
Tipos de sensores
Tipos de sensoresTipos de sensores
Tipos de sensores
 
10959890 sensores-de-temperatura
10959890 sensores-de-temperatura10959890 sensores-de-temperatura
10959890 sensores-de-temperatura
 
SOLAR RADIATION MEASUREMENT
SOLAR RADIATION MEASUREMENTSOLAR RADIATION MEASUREMENT
SOLAR RADIATION MEASUREMENT
 
Radiacion ultravioleta y su aplicacion en Fisioterapia
Radiacion ultravioleta y su aplicacion en FisioterapiaRadiacion ultravioleta y su aplicacion en Fisioterapia
Radiacion ultravioleta y su aplicacion en Fisioterapia
 
Instrument Calibration
Instrument Calibration Instrument Calibration
Instrument Calibration
 
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRYINSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
 
Principle and instrumentation of UV-visible spectrophotometer.
Principle and instrumentation of UV-visible spectrophotometer.Principle and instrumentation of UV-visible spectrophotometer.
Principle and instrumentation of UV-visible spectrophotometer.
 
Solar Energy
Solar EnergySolar Energy
Solar Energy
 
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
 
3 Things Every Sales Team Needs to Be Thinking About in 2017
3 Things Every Sales Team Needs to Be Thinking About in 20173 Things Every Sales Team Needs to Be Thinking About in 2017
3 Things Every Sales Team Needs to Be Thinking About in 2017
 

Similar to Deadman IGARSS Laboratory Calibrations.ppt

Photovoltaic Module Energy Yield Measurements: Existing Approaches and Best P...
Photovoltaic Module Energy Yield Measurements: Existing Approaches and Best P...Photovoltaic Module Energy Yield Measurements: Existing Approaches and Best P...
Photovoltaic Module Energy Yield Measurements: Existing Approaches and Best P...Leonardo ENERGY
 
Fda Irvine Usp Eas2009 Final Ger[1] Final
Fda Irvine Usp Eas2009 Final Ger[1] FinalFda Irvine Usp Eas2009 Final Ger[1] Final
Fda Irvine Usp Eas2009 Final Ger[1] FinalGary Ritchie
 
qualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationqualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationSohailPattan
 
TH4_T0_04_thome.pptx
TH4_T0_04_thome.pptxTH4_T0_04_thome.pptx
TH4_T0_04_thome.pptxgrssieee
 
PV Module Characterization Methods at CFV Solar Test Laboratory
PV Module Characterization Methods at CFV Solar Test LaboratoryPV Module Characterization Methods at CFV Solar Test Laboratory
PV Module Characterization Methods at CFV Solar Test LaboratoryCFVSolar
 
Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validationMohiniTawade
 
Csavina: Uncertainty in an uncertain world: Using scientific judgment for ev...
Csavina:  Uncertainty in an uncertain world: Using scientific judgment for ev...Csavina:  Uncertainty in an uncertain world: Using scientific judgment for ev...
Csavina: Uncertainty in an uncertain world: Using scientific judgment for ev...questRCN
 
(H.Parks_Y.Tang_P.Reese_J.Gust_J.Novak)_The_North_American_Josephson_Voltage_...
(H.Parks_Y.Tang_P.Reese_J.Gust_J.Novak)_The_North_American_Josephson_Voltage_...(H.Parks_Y.Tang_P.Reese_J.Gust_J.Novak)_The_North_American_Josephson_Voltage_...
(H.Parks_Y.Tang_P.Reese_J.Gust_J.Novak)_The_North_American_Josephson_Voltage_...Paul Reese
 
Wavelength Accuracy.pptx
Wavelength Accuracy.pptxWavelength Accuracy.pptx
Wavelength Accuracy.pptxsumitdekate1
 
Calibration and validation of analytical instruments
Calibration and validation of analytical instrumentsCalibration and validation of analytical instruments
Calibration and validation of analytical instrumentsSolairajan A
 
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxAkashSharma640477
 
Calibration of IR spectrophotometer ppt
Calibration of IR spectrophotometer pptCalibration of IR spectrophotometer ppt
Calibration of IR spectrophotometer pptSIHAS
 
Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Mostafa Mahmoud
 

Similar to Deadman IGARSS Laboratory Calibrations.ppt (20)

Exploring Sources of Uncertainties in Solar Resource Measurements
Exploring Sources of Uncertainties in Solar Resource MeasurementsExploring Sources of Uncertainties in Solar Resource Measurements
Exploring Sources of Uncertainties in Solar Resource Measurements
 
62 friesen field_data_requirements_for_the_validation_of_pv_module_performanc...
62 friesen field_data_requirements_for_the_validation_of_pv_module_performanc...62 friesen field_data_requirements_for_the_validation_of_pv_module_performanc...
62 friesen field_data_requirements_for_the_validation_of_pv_module_performanc...
 
Photovoltaic Module Energy Yield Measurements: Existing Approaches and Best P...
Photovoltaic Module Energy Yield Measurements: Existing Approaches and Best P...Photovoltaic Module Energy Yield Measurements: Existing Approaches and Best P...
Photovoltaic Module Energy Yield Measurements: Existing Approaches and Best P...
 
Fda Irvine Usp Eas2009 Final Ger[1] Final
Fda Irvine Usp Eas2009 Final Ger[1] FinalFda Irvine Usp Eas2009 Final Ger[1] Final
Fda Irvine Usp Eas2009 Final Ger[1] Final
 
Rte0 mjetotk
Rte0 mjetotk Rte0 mjetotk
Rte0 mjetotk
 
qualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationqualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validation
 
TH4_T0_04_thome.pptx
TH4_T0_04_thome.pptxTH4_T0_04_thome.pptx
TH4_T0_04_thome.pptx
 
2014 PV Performance Modeling Workshop: Results from Flash Testing at Multiple...
2014 PV Performance Modeling Workshop: Results from Flash Testing at Multiple...2014 PV Performance Modeling Workshop: Results from Flash Testing at Multiple...
2014 PV Performance Modeling Workshop: Results from Flash Testing at Multiple...
 
PV Module Characterization Methods at CFV Solar Test Laboratory
PV Module Characterization Methods at CFV Solar Test LaboratoryPV Module Characterization Methods at CFV Solar Test Laboratory
PV Module Characterization Methods at CFV Solar Test Laboratory
 
2014 PV Performance Modeling Workshop: PV Module Characterization Methods at ...
2014 PV Performance Modeling Workshop: PV Module Characterization Methods at ...2014 PV Performance Modeling Workshop: PV Module Characterization Methods at ...
2014 PV Performance Modeling Workshop: PV Module Characterization Methods at ...
 
21 supsi workshop_indoor_para_v3_1
21 supsi workshop_indoor_para_v3_121 supsi workshop_indoor_para_v3_1
21 supsi workshop_indoor_para_v3_1
 
Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validation
 
Csavina: Uncertainty in an uncertain world: Using scientific judgment for ev...
Csavina:  Uncertainty in an uncertain world: Using scientific judgment for ev...Csavina:  Uncertainty in an uncertain world: Using scientific judgment for ev...
Csavina: Uncertainty in an uncertain world: Using scientific judgment for ev...
 
(H.Parks_Y.Tang_P.Reese_J.Gust_J.Novak)_The_North_American_Josephson_Voltage_...
(H.Parks_Y.Tang_P.Reese_J.Gust_J.Novak)_The_North_American_Josephson_Voltage_...(H.Parks_Y.Tang_P.Reese_J.Gust_J.Novak)_The_North_American_Josephson_Voltage_...
(H.Parks_Y.Tang_P.Reese_J.Gust_J.Novak)_The_North_American_Josephson_Voltage_...
 
Wavelength Accuracy.pptx
Wavelength Accuracy.pptxWavelength Accuracy.pptx
Wavelength Accuracy.pptx
 
Calibration and validation of analytical instruments
Calibration and validation of analytical instrumentsCalibration and validation of analytical instruments
Calibration and validation of analytical instruments
 
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
 
2014 PV Performance Modeling Workshop: Outdoor Module Characterization Method...
2014 PV Performance Modeling Workshop: Outdoor Module Characterization Method...2014 PV Performance Modeling Workshop: Outdoor Module Characterization Method...
2014 PV Performance Modeling Workshop: Outdoor Module Characterization Method...
 
Calibration of IR spectrophotometer ppt
Calibration of IR spectrophotometer pptCalibration of IR spectrophotometer ppt
Calibration of IR spectrophotometer ppt
 
Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods
 

More from grssieee

Tangent height accuracy of Superconducting Submillimeter-Wave Limb-Emission S...
Tangent height accuracy of Superconducting Submillimeter-Wave Limb-Emission S...Tangent height accuracy of Superconducting Submillimeter-Wave Limb-Emission S...
Tangent height accuracy of Superconducting Submillimeter-Wave Limb-Emission S...grssieee
 
SEGMENTATION OF POLARIMETRIC SAR DATA WITH A MULTI-TEXTURE PRODUCT MODEL
SEGMENTATION OF POLARIMETRIC SAR DATA WITH A MULTI-TEXTURE PRODUCT MODELSEGMENTATION OF POLARIMETRIC SAR DATA WITH A MULTI-TEXTURE PRODUCT MODEL
SEGMENTATION OF POLARIMETRIC SAR DATA WITH A MULTI-TEXTURE PRODUCT MODELgrssieee
 
TWO-POINT STATISTIC OF POLARIMETRIC SAR DATA TWO-POINT STATISTIC OF POLARIMET...
TWO-POINT STATISTIC OF POLARIMETRIC SAR DATA TWO-POINT STATISTIC OF POLARIMET...TWO-POINT STATISTIC OF POLARIMETRIC SAR DATA TWO-POINT STATISTIC OF POLARIMET...
TWO-POINT STATISTIC OF POLARIMETRIC SAR DATA TWO-POINT STATISTIC OF POLARIMET...grssieee
 
THE SENTINEL-1 MISSION AND ITS APPLICATION CAPABILITIES
THE SENTINEL-1 MISSION AND ITS APPLICATION CAPABILITIESTHE SENTINEL-1 MISSION AND ITS APPLICATION CAPABILITIES
THE SENTINEL-1 MISSION AND ITS APPLICATION CAPABILITIESgrssieee
 
GMES SPACE COMPONENT:PROGRAMMATIC STATUS
GMES SPACE COMPONENT:PROGRAMMATIC STATUSGMES SPACE COMPONENT:PROGRAMMATIC STATUS
GMES SPACE COMPONENT:PROGRAMMATIC STATUSgrssieee
 
PROGRESSES OF DEVELOPMENT OF CFOSAT SCATTEROMETER
PROGRESSES OF DEVELOPMENT OF CFOSAT SCATTEROMETERPROGRESSES OF DEVELOPMENT OF CFOSAT SCATTEROMETER
PROGRESSES OF DEVELOPMENT OF CFOSAT SCATTEROMETERgrssieee
 
DEVELOPMENT OF ALGORITHMS AND PRODUCTS FOR SUPPORTING THE ITALIAN HYPERSPECTR...
DEVELOPMENT OF ALGORITHMS AND PRODUCTS FOR SUPPORTING THE ITALIAN HYPERSPECTR...DEVELOPMENT OF ALGORITHMS AND PRODUCTS FOR SUPPORTING THE ITALIAN HYPERSPECTR...
DEVELOPMENT OF ALGORITHMS AND PRODUCTS FOR SUPPORTING THE ITALIAN HYPERSPECTR...grssieee
 
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...grssieee
 
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...grssieee
 
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...grssieee
 
test 34mb wo animations
test  34mb wo animationstest  34mb wo animations
test 34mb wo animationsgrssieee
 
2011_Fox_Tax_Worksheets.pdf
2011_Fox_Tax_Worksheets.pdf2011_Fox_Tax_Worksheets.pdf
2011_Fox_Tax_Worksheets.pdfgrssieee
 
DLR open house
DLR open houseDLR open house
DLR open housegrssieee
 
DLR open house
DLR open houseDLR open house
DLR open housegrssieee
 
DLR open house
DLR open houseDLR open house
DLR open housegrssieee
 
Tana_IGARSS2011.ppt
Tana_IGARSS2011.pptTana_IGARSS2011.ppt
Tana_IGARSS2011.pptgrssieee
 
Solaro_IGARSS_2011.ppt
Solaro_IGARSS_2011.pptSolaro_IGARSS_2011.ppt
Solaro_IGARSS_2011.pptgrssieee
 

More from grssieee (20)

Tangent height accuracy of Superconducting Submillimeter-Wave Limb-Emission S...
Tangent height accuracy of Superconducting Submillimeter-Wave Limb-Emission S...Tangent height accuracy of Superconducting Submillimeter-Wave Limb-Emission S...
Tangent height accuracy of Superconducting Submillimeter-Wave Limb-Emission S...
 
SEGMENTATION OF POLARIMETRIC SAR DATA WITH A MULTI-TEXTURE PRODUCT MODEL
SEGMENTATION OF POLARIMETRIC SAR DATA WITH A MULTI-TEXTURE PRODUCT MODELSEGMENTATION OF POLARIMETRIC SAR DATA WITH A MULTI-TEXTURE PRODUCT MODEL
SEGMENTATION OF POLARIMETRIC SAR DATA WITH A MULTI-TEXTURE PRODUCT MODEL
 
TWO-POINT STATISTIC OF POLARIMETRIC SAR DATA TWO-POINT STATISTIC OF POLARIMET...
TWO-POINT STATISTIC OF POLARIMETRIC SAR DATA TWO-POINT STATISTIC OF POLARIMET...TWO-POINT STATISTIC OF POLARIMETRIC SAR DATA TWO-POINT STATISTIC OF POLARIMET...
TWO-POINT STATISTIC OF POLARIMETRIC SAR DATA TWO-POINT STATISTIC OF POLARIMET...
 
THE SENTINEL-1 MISSION AND ITS APPLICATION CAPABILITIES
THE SENTINEL-1 MISSION AND ITS APPLICATION CAPABILITIESTHE SENTINEL-1 MISSION AND ITS APPLICATION CAPABILITIES
THE SENTINEL-1 MISSION AND ITS APPLICATION CAPABILITIES
 
GMES SPACE COMPONENT:PROGRAMMATIC STATUS
GMES SPACE COMPONENT:PROGRAMMATIC STATUSGMES SPACE COMPONENT:PROGRAMMATIC STATUS
GMES SPACE COMPONENT:PROGRAMMATIC STATUS
 
PROGRESSES OF DEVELOPMENT OF CFOSAT SCATTEROMETER
PROGRESSES OF DEVELOPMENT OF CFOSAT SCATTEROMETERPROGRESSES OF DEVELOPMENT OF CFOSAT SCATTEROMETER
PROGRESSES OF DEVELOPMENT OF CFOSAT SCATTEROMETER
 
DEVELOPMENT OF ALGORITHMS AND PRODUCTS FOR SUPPORTING THE ITALIAN HYPERSPECTR...
DEVELOPMENT OF ALGORITHMS AND PRODUCTS FOR SUPPORTING THE ITALIAN HYPERSPECTR...DEVELOPMENT OF ALGORITHMS AND PRODUCTS FOR SUPPORTING THE ITALIAN HYPERSPECTR...
DEVELOPMENT OF ALGORITHMS AND PRODUCTS FOR SUPPORTING THE ITALIAN HYPERSPECTR...
 
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
 
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
 
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
EO-1/HYPERION: NEARING TWELVE YEARS OF SUCCESSFUL MISSION SCIENCE OPERATION A...
 
Test
TestTest
Test
 
test 34mb wo animations
test  34mb wo animationstest  34mb wo animations
test 34mb wo animations
 
Test 70MB
Test 70MBTest 70MB
Test 70MB
 
Test 70MB
Test 70MBTest 70MB
Test 70MB
 
2011_Fox_Tax_Worksheets.pdf
2011_Fox_Tax_Worksheets.pdf2011_Fox_Tax_Worksheets.pdf
2011_Fox_Tax_Worksheets.pdf
 
DLR open house
DLR open houseDLR open house
DLR open house
 
DLR open house
DLR open houseDLR open house
DLR open house
 
DLR open house
DLR open houseDLR open house
DLR open house
 
Tana_IGARSS2011.ppt
Tana_IGARSS2011.pptTana_IGARSS2011.ppt
Tana_IGARSS2011.ppt
 
Solaro_IGARSS_2011.ppt
Solaro_IGARSS_2011.pptSolaro_IGARSS_2011.ppt
Solaro_IGARSS_2011.ppt
 

Deadman IGARSS Laboratory Calibrations.ppt

  • 1. Laboratory Panel and Radiometer Calibration 28 July 2011, IGARSS, Vancouver Andrew Deadman 1 , Nigel Fox 1 , Irina Behnert 1 D Griffith 2 1 National Physical Laboratory (NPL), United Kingdom 2 Council for Scientific and Industrial Research (CSIR), South Africa
  • 2.
  • 3.
  • 4.
  • 5.
  • 6. Laboratory Radiometer Comparison Measured before and after the field measurements. The difference used to check the stability check of the calibration set-up and also the performance of the radiometers. NPL provided a radiance source called TSARS (Transfer Standard Absolute Radiance Source)
  • 7. Laboratory Radiometer comparison - Results First TSARS measurement – 15 August 2010
  • 8. Laboratory Radiometer comparison - Results Second TSARS measurement – 26 August 2010
  • 9. Laboratory Radiometer comparison - Results Difference between two TSARS measurements. Similar trend for most participants.
  • 10. Laboratory Radiometer comparison Environmental conditions First TSARS calibration 15 Aug 2010 Second TSARS calibration 26 Aug 2010 Environmental conditions outside normal operating conditions for TSARS so absolute calibration was not possible.
  • 11. Laboratory Radiometer Comparison 2009 TSARS calibration Data from 2009 – difference between before and after field measurements. Shows better TSARS performance
  • 12. Laboratory Radiometer comparison - 2010 Results Difference between measurement of TSARS over the two days Problem with the performance of one radiometer. No absolute calibration but were able to establish bias between instruments
  • 13.
  • 14. Laboratory Panel Comparison 45/0 geometry Stable illumination conditions Same source and detector
  • 15. Laboratory Panel Comparison Each panel was measured by comparison with an NPL calibrated reference panel. Measured before and after field measurements. Test stability of panels and robustness of the set-up
  • 16. Laboratory Panel Comparison Set up tested at NPL before measurements in Turkey. Type A uncertainty < ± 0.2% for most wavelengths of interest. Confidence in set-up.
  • 17. Laboratory Panel Comparison - Results Results of laboratory panel comparisons. Calibrated reflectance factor values.
  • 18. Laboratory Panel Comparison - Results Good agreement. No offsets in data or panel ageing affects. Differences less than the estimated calibration uncertainty of ± 1%. Difference in panel reflectance factor values before and after the field campaign.
  • 19. Laboratory Panel Comparison - Results Comparison of laboratory calibration results and reported reflectance
  • 20. Reasons for difference in panel reflectance factor values Laboratory calibration Reported calibration Bidirectional illumination 45 ° , nadir view - 45/0 geometry Diffuse illumination, 8 ° view – 8°/hemispherical geometry Reflectance panels are not perfectly lambertian – reflectance factor greater than unity at low sun zenith angles
  • 21. Laboratory Panel Comparison - Results Some participants had a goniometric calibration of their panel, so a direct comparison is possible. Very good agreement considering three independent methods used traceable to three national standards laboratories
  • 22.
  • 23. In-field Panel Comparison Repeat of laboratory exercise, using the sun as the source. Sun zenith angle between 30° and 35°. Each panel compared to NPL reference panel.
  • 24. Laboratory & In-field panel calibration – expected result NPL laboratory calibration In situ calibrations Reported calibration value
  • 25. Laboratory & In-field panel calibration – expected result NPL laboratory calibration In situ calibrations Reported calibration value
  • 26. Reasons for difference in panel reflectance factor values Laboratory calibration In field calibration Reported calibration Bidirectional illumination 45 º , nadir view - 45/0 geometry Diffuse and Direct illumination Sun zenith & irradiance variability Diffuse illumination, 8 º view – 8°/hemispherical geometry
  • 27. In-field panel comparison Shows deviation of solar irradiance from mean over the period of in-field panel comparison. Corresponds well with AERONET, although there are short term variations not captured by AERONET. Data courtesy of CSIR.
  • 28.