MEASURING
INSTRUMENTS
Course outcome
C403.2 : Compare AC and DC meters along with
its internal construction
Introduction
• Measurement:
It is the act or result of quantitative
comparison between a predefined standard and
an unknown quantity.
• Instrument:
It is a device or mechanism used to
determine the present value of a quantity under
observation.
Classification of Instruments
• Instruments can be classified as:
1. Absolute instruments
2. Secondary instruments
1. Absolute instruments:
 Absolute instruments indicate the value of the quantity
being measured in terms of constant of instruments and its
deflection.
 No comparison with standard instrument is necessary.
 Example: tangent galvanometer, Rayleigh current balance.
Continued…
2. Secondary instruments:
The secondary instruments need calibration with
respect to the absolute instruments.
The secondary instruments determine the value of the
quantity being measured from the deflection of the
instruments.
Calibration is a must for secondary instrument, without
calibration the deflection obtained is meaningless.
Example: Ammeter, voltmeter, wattmeter etc.
Continued…
• Classification based on the nature of operation:
Secondary instruments are further classified
according to the nature of operation as:
1. Indicating
2. Recording
3. Integrating instruments.
 Indicating instruments indicate the
instantaneous value of quantity under
measurement.
Continued…
Recording instruments give a continuous
record of variation of quantity being measured
(such as voltage, frequency, power etc.).
Recorders are commonly used in power plants,
process industries.
An integrating instrument is one which takes
into consideration the period or the time over
which the quantity is supplied. e.g. ampere-
hour meter, energy meter.
Continued…
• Electrical instruments can also classified as:
1. A.C. instruments
2. D.C. instruments
3. A.C./D.C. instruments
 Some electrical instruments can measure only A.C.
quantity, e.g. induction type instruments, some can
measure only D.C. quantity, e.g. P.M.M.C.
instruments. Some can work on both a.c. and d.c. e.g.
moving iron instruments, dynamometer instruments
etc.
Continued…
• Analog or Digital instruments:
One more way of classifying instrument is:
1. Analog instruments
2. Digital instruments.
 Analog information is continuous and stepless function
of time. Analog instruments are easy to understand,
calibrate and maintain.
 Digital information is in form of discrete pulses or
steps. Digital instruments have higher resolution, high
readability.
Continued…
• Various instruments used in practice:
Following instruments used in day to day life
in order to measure different quantity.
Sr.
No.
Name of the instruments Quantity measured
1. Voltmeter AC or DC voltage
2. Ammeter AC or DC current
3. Wattmeter AC power (Watt)
4. Energy meter Energy (Watt hour)
Operation of Indicating Instruments
• For satisfactory operation of any indicating
instrument, following three torques must act
together appropriately:
1. Deflecting torque
2. Controlling torque
3. Damping torque
Continued…
1. Deflecting Torque:
• It causes the moving system of the instrument to
move from its position of rest.
• Deflecting torque is produced by using any one of
the following effects of electric current:
i. Magnetic effect
ii.Electromagnetic induction effect
iii.Heating effect
iv.Electrostatic effect
Continued…
2. Controlling Torque:
• It limits the movement of moving systems. It also
ensures that magnitude of deflection is always the
same for the given value of input quantity under
measurement.
• Controlling torque acts in the opposite direction
to that of the deflecting torque.
• At steady state,
Deflecting torque = Controlling torque
Continued…
3. Damping Torque:
• Due to deflecting torque, pointer moves in one direction while
due to controlling torque pointer moves in opposite direction.
• Due to these opposite torques, the pointer may oscillate in the
forward and backward direction if the damping torque is not
present.
• Damping torque brings the moving system to rest quickly in its
final position.
• Damping torque acts only when the moving system is actually
moving. If moving system is at rest, damping torque is zero.
Continued…
• Critical Damping:
 Depending on the magnitude of torque, damping may be
classified as underdamped, overdamped or critically damped.
Effect of damping on the deflection of the instrument is
shown in fig.(1).
 If the instrument is underdamped, the pointer will come to
rest after some oscillations.
 If the instrument is overdamped, pointer takes considerable
time to obtain its final deflected position.
 If the damping is critical, without oscillation and in short
time the pointer reach its final steady position.
Continued…
Fig.(1): Effect of damping on the deflection of instrument
Connection Diagrams
• Connection diagram of an Ammeter:
Ammeter is used for the measurement of current.
An ammeter is always connected in series with
the load, the current through which is to be
measured as shown in fig.(1).
Since the resistance offered by an ammeter is
very small, its introduction in series with load
does not alter the circuit conditions.
Continued…
Fig.(1): connection of an ammeter
Continued…
• Connection diagram of a voltmeter:
A voltmeter is used for the measurement of
voltage (potential difference).
So it is connected across the points between
which the potential difference is to be measured.
A voltmeter has a high resistance, so it draws
very small current. The connection of a
voltmeter is as shown in fig.(2).
Continued…
Fig.(2): connection diagram of voltmeter
Continued…
• Comparison of Ammeter and Voltmeter:
Sr.No. Parameter Ammeter Voltmeter
1. Quantity
measured
Current Voltage
2. Connected in Series with load Parallel with
load
3. Resistance Low High
4. Connection
Diagram
Refer fig.(1) Refer fig.(2)
Continued…
• Types of instruments used for Voltmeter and
Ammeters:
The practically used ammeters and voltmeters
can be of the following two types:
1. Permanent ,magnet moving coil (PMMC) type
2. Moving iron (MI) type
3. Hot wire type
4. Induction type
Moving Coil Instruments
• Moving coil instruments (ammeter and
voltmeter) are of two types:
1. Permanent magnet moving coil type (PMMC)
used only for D.C.
2. Dynamometer type can be used for AC as
well DC.
Continued…
1. PMMC Instruments:
 They are also known as d’Arsonval
instruments.
 These instruments works on the
electromagnetic effect of current.
 A permanent magnet used to produce
magnetic flux and coil that carries the current
to be measures moves in this field.
Continued…
• Working principle:
 When a current carrying conductor is placed in a magnetic field, it
experiences a force. It is given by expression,
F = BIL
Where F = Force in Newton,
B = Flux density is tesla,
I = Current is ampere,
L = Length of conductor in meter.
 The current I which is to be measured is passed through the moving coil and
experiences a force which is directly proportional to this current.
 Due to this force the coil moves and the pointer attached to it will also
move.
 The angle through which the pointer moves is proportional to current I.
Continued…
• Construction of PMMC instrument:
 A coil of thin wire is mounted on an aluminum frame (spindle)
positioned between the poles of a U shaped permanent magnet
which is made up of magnetic alloys like alnico.
 The coil is pivoted on the jewelled bearing and thus the coil is
free to rotate. The current is fed to the coil through spiral
springs which are two in numbers.
 The coil which carries a current, which is to be measured,
moves in a strong magnetic field produced by a permanent
magnet and a pointer is attached to the spindle which shows the
measured value.
Continued…
Fig.(1): construction of PMMC instrument
Continued…
• Deflecting Torque:
 It can be proved that the expression for the deflecting torque is
given by,
Td = G x I
where G = constant
I = Current through the moving coil
• Controlling Torque:
 The controlling torque is given by,
Tc = C. θ
where C = Control spring constant in N-m/rad
θ = Deflection of coil from zero position
Continued…
For steady state, the controlling torque is equal
to the deflection torque
∴ Tc = Td
i.e. Cθ = GI
∴ θ I
∝
Thus deflection of the pointer is proportional
to current passed through the coil.
Continued…
• Advantages of PMMC Instruments:
1. The PMMC consumes less power and has great accuracy.
2. It has uniformly divided scale and can cover arc of 270
degree.
3. The PMMC has a high torque to weight ratio.
4. It can be modified as ammeter or voltmeter with suitable
resistance.
5. It has efficient damping characteristics and is not
affected by stray magnetic field.
6. It produces no losses due to hysteresis.
Continued…
• Disadvantages of PMMC Instruments:
1. The moving coil instrument can only be used on D.C
supply as the reversal of current produces reversal of
torque on the coil.
2. It’s very delicate and sometimes uses ac circuit with a
rectifier.
3. It’s costly as compared to moving coil iron
instruments.
4. It may show error due to loss of magnetism of
permanent magnet.
Continued…
• Connection of PMMC instruments as Ammeter or
Voltmeter:
1. Ammeter:
 When the instrument is used as an ammeter, an external
low resistance called shunt is connected in parallel with
the movement so that net resistance of the instrument is
very low and the instrument can be connected in series
with any equipment whose current is to be measured.
 Due to low resistance, voltage drop across instrument is
negligible.
Continued…
Fig.(2a): Connection as an ammeter Fig.(2b): connection as a
voltmeter
Continued…
2. Voltmeter:
When the instrument is used as a voltmeter, an
external resistance is connected in series with
the movement and this series combination is
connected across supply terminals or load
terminals or in general cross any two terminals
where terminal voltage is to be measured.
Due to high series resistance, current flowing
through the instrument is negligible.
Dynamometer type moving Coil
Instruments
 It is modification of PMMC instrument.
 Permanent magnet in PMMC is replaced by two fixed coils.
• Principle of working:
 It works on the principle that whenever a current carrying
conductor is placed in a magnetic field, force is exerted on
the conductor.
 In this case, fixed coils produced magnetic field. Moving
coil carrying the current to be measured is suspended in this
magnetic field.
Continued…
Hence it experiences a torque. The moving
coil then rotates through an angle proportional
to the current flowing through the moving coil.
As the pointer is attached to the moving coil,
we get the pointer deflection proportional to
the current to be measured.
Continued…
• Construction:
 As shown in fig.(1), F1, F2 are two identical circular, air
cored coils. They are connected in series and these coils are
placed in parallel to each other.
 They produced a uniform magnetic field. M is a light moving
coils. It lies in the magnetic field produced by F1, F2.
moving coil is supported by a spindle and jeweled bearing.
 Two control springs wound in opposite directions are used as
leads to pass current in moving coil. These springs produce
the controlling torque. Damping torque is obtained using air-
friction damping.
Continued…
Fig.(1): Dynamometer type moving coil instrument
Continued…
• Advantages:
1. Used of iron is avoided. Hence it is free from
hysteresis and eddy current losses.
2. It is useful for both a.c. and d.c. measurements.
3. High degree of accuracy.
4. It can be used as a transfer instrument. It is
often used as a standard instrument for
calibrating ammeter and volt meters.
Continued…
• Limitations:
1. Scale is nonlinear.
2. Torque to weight ratio is small.
3. It is an expensive instrument.
4. The instrument has low sensitivity.
5. Shielding should be provided to avoid effect
of stray magnetic field.
Moving Iron (MI) instruments
Where ruggedness is more important than high
degree of accuracy, moving iron instruments
are used.
• Types of moving iron instruments:
1. Attraction type
2. Repulsion type
Continued…
1. Attraction type moving iron instrument:
• Construction of the attraction type moving iron instrument is as
shown in the given figure (1).
• The moving iron, i.e. the disc of soft iron, is eccentrically
mounted. Coil is situated around the disc. When the coil is excited
it produces magnetic field.
• Due to magnetic field the moving iron moves from the weaker
field outside the coil to the stronger field inside the coil. Thus
moving iron gets attracted inwards and thus the name attraction
type.
Continued…
• The controlling torque is provided by the
balance weights attached to the moving iron.
Spring also can be used to provide controlling
torque.
• Damping is provided by air friction in which
aluminum piston is attached to the moving
system and moves in a closed air damping
chamber.
Continued…
Fig.(1): attraction type moving instrument
Continued…
2. Repulsion type moving iron instrument:
• In these type of instruments there two vanes present inside the coil in which
one is fixed while other is movable. These both vanes gets similarly
magnetized when coil is excited. Thus there is a force of repulsion between
both vanes causing movement in movable vane. There are further two
different designs of repulsion type moving iron instruments.
i. Radial vane type:
• In radial vane type radial strips of iron are used as moving vanes. The fixed
vane is attached to the coil and moving is attached to the moving spindle.
Continued…
ii. Coaxial vane type:
• In coaxial vane type the fixed as well as moving
vanes are the sections of coaxial cylinders as shown
in above figure.
The controlling torque is provided by the springs or
balancing weights (in vertically mounted
instrument). The damping is provided by air friction
damping same as in attraction type moving iron
instruments.
Continued…
Fig.(2):Repulsion type Moving Iron Instrument
Continued…
• Advantages of Moving iron Instruments:
1. Suitable of a.c. as well as d.c. measurements.
2. Good accuracy.
3. Cheaper in cost as compared to permanents
magnet moving coil instruments.
4. The instrument has high torque to weight ratio.
5. The instrument can be used for low frequency
measurement also.
Continued…
• Disadvantages of Moving iron instruments:
1. Power consumed by the instrument is high as
compared to that of the permanent magnet
moving coil instrument.
2. The scale is non-uniform.
3. Due to the presence of iron part in the operating
system, error due to the hysteresis effect is
introduced. To reduce this effect nickel iron
alloys are used.
Electro-dynamic Wattmeter
• The basic action of this instrument depends
upon the electromagnetic force exerted between
fixed and moving coils carrying current.
• Fig.(1) shows fixed coil FF and moving coil M
carried by spindle.
• Controlling torque is provided by two spiral
springs mounted on the spindle which act as the
leads of coil M.
Continued…
• The deflecting torque is proportional to the
product of the current in the current coil and
voltage across pressure coil. So Td is
proportional to power.
• The scale of this instrument is uniform.
• It is used for power measurement in ac and dc
circuits.
Continued…
Fig.(1): electrodynamic wattmeter
Continued…
• Advantages:
1. It can be used for a.c. as well as d.c.
measurement.
2. It is easy in construction.
3. The instrument is free from hysteresis errors.
4. Uniform scale.
5. Light weight.
6. Consume less power.
Continued…
• Disadvantages:
1. High cost.
2. Errors can occur due to friction, changes in
temperature etc.
3. Large errors at low power factor.
4. It gets affected by the external stray magnetic
fields. So shielding is essential.
Energy Meter
• Fig.(1) shows the construction of energy meter.
• The two exciting coils act as current coil and voltage
coil and the disc acts as a time counting device.
• The disc is kept free to rotate continuously. Speed of
the disc depends on the power supplied to the load.
More the load, higher is the disc speed.
• In this instrument a gear train is provided to count the
revolution of the disc. Number of revolution of the
disc are directly recorded in terms of the energy
consumed.
Continued…
Fig.(1): construction of energy meter
Digital Multimeter
• Advantages:
1. They are having high input impedance. So
there is no loading effect.
2. They are having higher accuracy.
3. They are available in smaller size.
4. Digital display so easy and accurate reading.
5. Automatic range adjustment.
6. Easy to carry.
Continued…
• Disadvantages:
1. Needs battery for its operation.
2. Internal circuit is complex.
3. It is costlier than the analog multimeter.
• Applications:
1. To measure DC voltage
2. To measure the resistance
Clip On Meter
• Working principle:
The principle of operation is shown in above figure,
where it can be seen that the clamp on jaws of the
instrument act as a transformer core and the current
carrying conductor acts as a primary winding.
 The resulting secondary current is then measured by
the instrument (taking the turns ratio of the current
transformer into account).
Current induced in the secondary winding is rectified
and applied to a moving coil meter.
Continued…
Fig.(1): Clip-on meter
Continued…
• Applications:
the clip on ammeter is used for high ac
current measurement without breaking the
circuit.

chapter 2-measuring instrument.pptx moving

  • 1.
  • 2.
    Course outcome C403.2 :Compare AC and DC meters along with its internal construction
  • 3.
    Introduction • Measurement: It isthe act or result of quantitative comparison between a predefined standard and an unknown quantity. • Instrument: It is a device or mechanism used to determine the present value of a quantity under observation.
  • 4.
    Classification of Instruments •Instruments can be classified as: 1. Absolute instruments 2. Secondary instruments 1. Absolute instruments:  Absolute instruments indicate the value of the quantity being measured in terms of constant of instruments and its deflection.  No comparison with standard instrument is necessary.  Example: tangent galvanometer, Rayleigh current balance.
  • 5.
    Continued… 2. Secondary instruments: Thesecondary instruments need calibration with respect to the absolute instruments. The secondary instruments determine the value of the quantity being measured from the deflection of the instruments. Calibration is a must for secondary instrument, without calibration the deflection obtained is meaningless. Example: Ammeter, voltmeter, wattmeter etc.
  • 6.
    Continued… • Classification basedon the nature of operation: Secondary instruments are further classified according to the nature of operation as: 1. Indicating 2. Recording 3. Integrating instruments.  Indicating instruments indicate the instantaneous value of quantity under measurement.
  • 7.
    Continued… Recording instruments givea continuous record of variation of quantity being measured (such as voltage, frequency, power etc.). Recorders are commonly used in power plants, process industries. An integrating instrument is one which takes into consideration the period or the time over which the quantity is supplied. e.g. ampere- hour meter, energy meter.
  • 8.
    Continued… • Electrical instrumentscan also classified as: 1. A.C. instruments 2. D.C. instruments 3. A.C./D.C. instruments  Some electrical instruments can measure only A.C. quantity, e.g. induction type instruments, some can measure only D.C. quantity, e.g. P.M.M.C. instruments. Some can work on both a.c. and d.c. e.g. moving iron instruments, dynamometer instruments etc.
  • 9.
    Continued… • Analog orDigital instruments: One more way of classifying instrument is: 1. Analog instruments 2. Digital instruments.  Analog information is continuous and stepless function of time. Analog instruments are easy to understand, calibrate and maintain.  Digital information is in form of discrete pulses or steps. Digital instruments have higher resolution, high readability.
  • 10.
    Continued… • Various instrumentsused in practice: Following instruments used in day to day life in order to measure different quantity. Sr. No. Name of the instruments Quantity measured 1. Voltmeter AC or DC voltage 2. Ammeter AC or DC current 3. Wattmeter AC power (Watt) 4. Energy meter Energy (Watt hour)
  • 11.
    Operation of IndicatingInstruments • For satisfactory operation of any indicating instrument, following three torques must act together appropriately: 1. Deflecting torque 2. Controlling torque 3. Damping torque
  • 12.
    Continued… 1. Deflecting Torque: •It causes the moving system of the instrument to move from its position of rest. • Deflecting torque is produced by using any one of the following effects of electric current: i. Magnetic effect ii.Electromagnetic induction effect iii.Heating effect iv.Electrostatic effect
  • 13.
    Continued… 2. Controlling Torque: •It limits the movement of moving systems. It also ensures that magnitude of deflection is always the same for the given value of input quantity under measurement. • Controlling torque acts in the opposite direction to that of the deflecting torque. • At steady state, Deflecting torque = Controlling torque
  • 14.
    Continued… 3. Damping Torque: •Due to deflecting torque, pointer moves in one direction while due to controlling torque pointer moves in opposite direction. • Due to these opposite torques, the pointer may oscillate in the forward and backward direction if the damping torque is not present. • Damping torque brings the moving system to rest quickly in its final position. • Damping torque acts only when the moving system is actually moving. If moving system is at rest, damping torque is zero.
  • 15.
    Continued… • Critical Damping: Depending on the magnitude of torque, damping may be classified as underdamped, overdamped or critically damped. Effect of damping on the deflection of the instrument is shown in fig.(1).  If the instrument is underdamped, the pointer will come to rest after some oscillations.  If the instrument is overdamped, pointer takes considerable time to obtain its final deflected position.  If the damping is critical, without oscillation and in short time the pointer reach its final steady position.
  • 16.
    Continued… Fig.(1): Effect ofdamping on the deflection of instrument
  • 17.
    Connection Diagrams • Connectiondiagram of an Ammeter: Ammeter is used for the measurement of current. An ammeter is always connected in series with the load, the current through which is to be measured as shown in fig.(1). Since the resistance offered by an ammeter is very small, its introduction in series with load does not alter the circuit conditions.
  • 18.
  • 19.
    Continued… • Connection diagramof a voltmeter: A voltmeter is used for the measurement of voltage (potential difference). So it is connected across the points between which the potential difference is to be measured. A voltmeter has a high resistance, so it draws very small current. The connection of a voltmeter is as shown in fig.(2).
  • 20.
  • 21.
    Continued… • Comparison ofAmmeter and Voltmeter: Sr.No. Parameter Ammeter Voltmeter 1. Quantity measured Current Voltage 2. Connected in Series with load Parallel with load 3. Resistance Low High 4. Connection Diagram Refer fig.(1) Refer fig.(2)
  • 22.
    Continued… • Types ofinstruments used for Voltmeter and Ammeters: The practically used ammeters and voltmeters can be of the following two types: 1. Permanent ,magnet moving coil (PMMC) type 2. Moving iron (MI) type 3. Hot wire type 4. Induction type
  • 23.
    Moving Coil Instruments •Moving coil instruments (ammeter and voltmeter) are of two types: 1. Permanent magnet moving coil type (PMMC) used only for D.C. 2. Dynamometer type can be used for AC as well DC.
  • 24.
    Continued… 1. PMMC Instruments: They are also known as d’Arsonval instruments.  These instruments works on the electromagnetic effect of current.  A permanent magnet used to produce magnetic flux and coil that carries the current to be measures moves in this field.
  • 25.
    Continued… • Working principle: When a current carrying conductor is placed in a magnetic field, it experiences a force. It is given by expression, F = BIL Where F = Force in Newton, B = Flux density is tesla, I = Current is ampere, L = Length of conductor in meter.  The current I which is to be measured is passed through the moving coil and experiences a force which is directly proportional to this current.  Due to this force the coil moves and the pointer attached to it will also move.  The angle through which the pointer moves is proportional to current I.
  • 26.
    Continued… • Construction ofPMMC instrument:  A coil of thin wire is mounted on an aluminum frame (spindle) positioned between the poles of a U shaped permanent magnet which is made up of magnetic alloys like alnico.  The coil is pivoted on the jewelled bearing and thus the coil is free to rotate. The current is fed to the coil through spiral springs which are two in numbers.  The coil which carries a current, which is to be measured, moves in a strong magnetic field produced by a permanent magnet and a pointer is attached to the spindle which shows the measured value.
  • 27.
  • 28.
    Continued… • Deflecting Torque: It can be proved that the expression for the deflecting torque is given by, Td = G x I where G = constant I = Current through the moving coil • Controlling Torque:  The controlling torque is given by, Tc = C. θ where C = Control spring constant in N-m/rad θ = Deflection of coil from zero position
  • 29.
    Continued… For steady state,the controlling torque is equal to the deflection torque ∴ Tc = Td i.e. Cθ = GI ∴ θ I ∝ Thus deflection of the pointer is proportional to current passed through the coil.
  • 30.
    Continued… • Advantages ofPMMC Instruments: 1. The PMMC consumes less power and has great accuracy. 2. It has uniformly divided scale and can cover arc of 270 degree. 3. The PMMC has a high torque to weight ratio. 4. It can be modified as ammeter or voltmeter with suitable resistance. 5. It has efficient damping characteristics and is not affected by stray magnetic field. 6. It produces no losses due to hysteresis.
  • 31.
    Continued… • Disadvantages ofPMMC Instruments: 1. The moving coil instrument can only be used on D.C supply as the reversal of current produces reversal of torque on the coil. 2. It’s very delicate and sometimes uses ac circuit with a rectifier. 3. It’s costly as compared to moving coil iron instruments. 4. It may show error due to loss of magnetism of permanent magnet.
  • 32.
    Continued… • Connection ofPMMC instruments as Ammeter or Voltmeter: 1. Ammeter:  When the instrument is used as an ammeter, an external low resistance called shunt is connected in parallel with the movement so that net resistance of the instrument is very low and the instrument can be connected in series with any equipment whose current is to be measured.  Due to low resistance, voltage drop across instrument is negligible.
  • 33.
    Continued… Fig.(2a): Connection asan ammeter Fig.(2b): connection as a voltmeter
  • 34.
    Continued… 2. Voltmeter: When theinstrument is used as a voltmeter, an external resistance is connected in series with the movement and this series combination is connected across supply terminals or load terminals or in general cross any two terminals where terminal voltage is to be measured. Due to high series resistance, current flowing through the instrument is negligible.
  • 35.
    Dynamometer type movingCoil Instruments  It is modification of PMMC instrument.  Permanent magnet in PMMC is replaced by two fixed coils. • Principle of working:  It works on the principle that whenever a current carrying conductor is placed in a magnetic field, force is exerted on the conductor.  In this case, fixed coils produced magnetic field. Moving coil carrying the current to be measured is suspended in this magnetic field.
  • 36.
    Continued… Hence it experiencesa torque. The moving coil then rotates through an angle proportional to the current flowing through the moving coil. As the pointer is attached to the moving coil, we get the pointer deflection proportional to the current to be measured.
  • 37.
    Continued… • Construction:  Asshown in fig.(1), F1, F2 are two identical circular, air cored coils. They are connected in series and these coils are placed in parallel to each other.  They produced a uniform magnetic field. M is a light moving coils. It lies in the magnetic field produced by F1, F2. moving coil is supported by a spindle and jeweled bearing.  Two control springs wound in opposite directions are used as leads to pass current in moving coil. These springs produce the controlling torque. Damping torque is obtained using air- friction damping.
  • 38.
  • 39.
    Continued… • Advantages: 1. Usedof iron is avoided. Hence it is free from hysteresis and eddy current losses. 2. It is useful for both a.c. and d.c. measurements. 3. High degree of accuracy. 4. It can be used as a transfer instrument. It is often used as a standard instrument for calibrating ammeter and volt meters.
  • 40.
    Continued… • Limitations: 1. Scaleis nonlinear. 2. Torque to weight ratio is small. 3. It is an expensive instrument. 4. The instrument has low sensitivity. 5. Shielding should be provided to avoid effect of stray magnetic field.
  • 41.
    Moving Iron (MI)instruments Where ruggedness is more important than high degree of accuracy, moving iron instruments are used. • Types of moving iron instruments: 1. Attraction type 2. Repulsion type
  • 42.
    Continued… 1. Attraction typemoving iron instrument: • Construction of the attraction type moving iron instrument is as shown in the given figure (1). • The moving iron, i.e. the disc of soft iron, is eccentrically mounted. Coil is situated around the disc. When the coil is excited it produces magnetic field. • Due to magnetic field the moving iron moves from the weaker field outside the coil to the stronger field inside the coil. Thus moving iron gets attracted inwards and thus the name attraction type.
  • 43.
    Continued… • The controllingtorque is provided by the balance weights attached to the moving iron. Spring also can be used to provide controlling torque. • Damping is provided by air friction in which aluminum piston is attached to the moving system and moves in a closed air damping chamber.
  • 44.
  • 45.
    Continued… 2. Repulsion typemoving iron instrument: • In these type of instruments there two vanes present inside the coil in which one is fixed while other is movable. These both vanes gets similarly magnetized when coil is excited. Thus there is a force of repulsion between both vanes causing movement in movable vane. There are further two different designs of repulsion type moving iron instruments. i. Radial vane type: • In radial vane type radial strips of iron are used as moving vanes. The fixed vane is attached to the coil and moving is attached to the moving spindle.
  • 46.
    Continued… ii. Coaxial vanetype: • In coaxial vane type the fixed as well as moving vanes are the sections of coaxial cylinders as shown in above figure. The controlling torque is provided by the springs or balancing weights (in vertically mounted instrument). The damping is provided by air friction damping same as in attraction type moving iron instruments.
  • 47.
  • 48.
    Continued… • Advantages ofMoving iron Instruments: 1. Suitable of a.c. as well as d.c. measurements. 2. Good accuracy. 3. Cheaper in cost as compared to permanents magnet moving coil instruments. 4. The instrument has high torque to weight ratio. 5. The instrument can be used for low frequency measurement also.
  • 49.
    Continued… • Disadvantages ofMoving iron instruments: 1. Power consumed by the instrument is high as compared to that of the permanent magnet moving coil instrument. 2. The scale is non-uniform. 3. Due to the presence of iron part in the operating system, error due to the hysteresis effect is introduced. To reduce this effect nickel iron alloys are used.
  • 50.
    Electro-dynamic Wattmeter • Thebasic action of this instrument depends upon the electromagnetic force exerted between fixed and moving coils carrying current. • Fig.(1) shows fixed coil FF and moving coil M carried by spindle. • Controlling torque is provided by two spiral springs mounted on the spindle which act as the leads of coil M.
  • 51.
    Continued… • The deflectingtorque is proportional to the product of the current in the current coil and voltage across pressure coil. So Td is proportional to power. • The scale of this instrument is uniform. • It is used for power measurement in ac and dc circuits.
  • 52.
  • 53.
    Continued… • Advantages: 1. Itcan be used for a.c. as well as d.c. measurement. 2. It is easy in construction. 3. The instrument is free from hysteresis errors. 4. Uniform scale. 5. Light weight. 6. Consume less power.
  • 54.
    Continued… • Disadvantages: 1. Highcost. 2. Errors can occur due to friction, changes in temperature etc. 3. Large errors at low power factor. 4. It gets affected by the external stray magnetic fields. So shielding is essential.
  • 55.
    Energy Meter • Fig.(1)shows the construction of energy meter. • The two exciting coils act as current coil and voltage coil and the disc acts as a time counting device. • The disc is kept free to rotate continuously. Speed of the disc depends on the power supplied to the load. More the load, higher is the disc speed. • In this instrument a gear train is provided to count the revolution of the disc. Number of revolution of the disc are directly recorded in terms of the energy consumed.
  • 56.
  • 57.
    Digital Multimeter • Advantages: 1.They are having high input impedance. So there is no loading effect. 2. They are having higher accuracy. 3. They are available in smaller size. 4. Digital display so easy and accurate reading. 5. Automatic range adjustment. 6. Easy to carry.
  • 58.
    Continued… • Disadvantages: 1. Needsbattery for its operation. 2. Internal circuit is complex. 3. It is costlier than the analog multimeter. • Applications: 1. To measure DC voltage 2. To measure the resistance
  • 59.
    Clip On Meter •Working principle: The principle of operation is shown in above figure, where it can be seen that the clamp on jaws of the instrument act as a transformer core and the current carrying conductor acts as a primary winding.  The resulting secondary current is then measured by the instrument (taking the turns ratio of the current transformer into account). Current induced in the secondary winding is rectified and applied to a moving coil meter.
  • 60.
  • 61.
    Continued… • Applications: the clipon ammeter is used for high ac current measurement without breaking the circuit.