AJA
Research Article
A Comparison of Personal Sound
Amplification Products and Hearing
Aids in Ecologically Relevant
Test Environments
Lisa Brody,a Yu-Hsiang Wu,a and Elizabeth Stangla
Purpose: The aim of this study was to compare the
benefit of self-adjusted personal sound amplification
products (PSAPs) to audiologist-fitted hearing aids
based on speech recognition, listening effort, and
sound quality in ecologically relevant test conditions to
estimate real-world effectiveness.
Method: Twenty-five older adults with bilateral mild-to-
moderate hearing loss completed the single-blinded,
crossover study. Participants underwent aided testing
using 3 PSAPs and a traditional hearing aid, as well
as unaided testing. PSAPs were adjusted based on
participant preference, whereas the hearing aid was
configured using best-practice verification protocols.
Audibility provided by the devices was quantified using the
Speech Intelligibility Index (American National Standards
Institute, 2012). Outcome measures assessing speech
recognition, listening effort, and sound quality were
administered in ecologically relevant laboratory conditions
designed to represent real-world speech listening situations.
Results: All devices significantly improved Speech
Intelligibility Index compared to unaided listening, with
the hearing aid providing more audibility than all PSAPs.
Results further revealed that, in general, the hearing aid
improved speech recognition performance and reduced
listening effort significantly more than all PSAPs. Few
differences in sound quality were observed between
devices. All PSAPs improved speech recognition and
listening effort compared to unaided testing.
Conclusions: Hearing aids fitted using best-practice
verification protocols were capable of providing more aided
audibility, better speech recognition performance, and lower
listening effort compared to the PSAPs tested in the current
study. Differences in sound quality between the devices were
minimal. However, because all PSAPs tested in the study
significantly improved participants’ speech recognition
performance and reduced listening effort compared to
unaided listening, PSAPs could serve as a budget-friendly
option for those who cannot afford traditional amplification.
H
earing loss affects more than 37 million adults
in America (National Institute on Deafness and
Other Communication Disorders, 2016). Accord-
ing to a survey of American adults, only about 30% of
those with hearing difficulties own hearing aids (Abrams
& Kihm, 2015), though treatments such as amplification
have been shown to reduce negative consequences of living
with untreated hearing loss (Hougaard, Ruf, Egger, &
Abrams, 2016; Mulrow et al., 1990). One primary reason
for the low uptake of hearing aids is their high cost (Abrams
& Kihm, 2015; Kochkin, 2007). The average pair of hear-
ing aids costs approximately $4,700 (President’s Council of
Advisors on Science and Technology, 2015), which can im-
pose a fina.
Capitol Tech U Doctoral Presentation - April 2024.pptx
AJAResearch ArticleA Comparison of Personal SoundAmp.docx
1. AJA
Research Article
A Comparison of Personal Sound
Amplification Products and Hearing
Aids in Ecologically Relevant
Test Environments
Lisa Brody,a Yu-Hsiang Wu,a and Elizabeth Stangla
Purpose: The aim of this study was to compare the
benefit of self-adjusted personal sound amplification
products (PSAPs) to audiologist-fitted hearing aids
based on speech recognition, listening effort, and
sound quality in ecologically relevant test conditions to
estimate real-world effectiveness.
Method: Twenty-five older adults with bilateral mild-to-
moderate hearing loss completed the single-blinded,
crossover study. Participants underwent aided testing
using 3 PSAPs and a traditional hearing aid, as well
as unaided testing. PSAPs were adjusted based on
participant preference, whereas the hearing aid was
configured using best-practice verification protocols.
Audibility provided by the devices was quantified using the
Speech Intelligibility Index (American National Standards
Institute, 2012). Outcome measures assessing speech
recognition, listening effort, and sound quality were
administered in ecologically relevant laboratory conditions
designed to represent real-world speech listening situations.
2. Results: All devices significantly improved Speech
Intelligibility Index compared to unaided listening, with
the hearing aid providing more audibility than all PSAPs.
Results further revealed that, in general, the hearing aid
improved speech recognition performance and reduced
listening effort significantly more than all PSAPs. Few
differences in sound quality were observed between
devices. All PSAPs improved speech recognition and
listening effort compared to unaided testing.
Conclusions: Hearing aids fitted using best-practice
verification protocols were capable of providing more aided
audibility, better speech recognition performance, and lower
listening effort compared to the PSAPs tested in the current
study. Differences in sound quality between the devices were
minimal. However, because all PSAPs tested in the study
significantly improved participants’ speech recognition
performance and reduced listening effort compared to
unaided listening, PSAPs could serve as a budget-friendly
option for those who cannot afford traditional amplification.
H
earing loss affects more than 37 million adults
in America (National Institute on Deafness and
Other Communication Disorders, 2016). Accord-
ing to a survey of American adults, only about 30% of
those with hearing difficulties own hearing aids (Abrams
& Kihm, 2015), though treatments such as amplification
have been shown to reduce negative consequences of living
with untreated hearing loss (Hougaard, Ruf, Egger, &
Abrams, 2016; Mulrow et al., 1990). One primary reason
for the low uptake of hearing aids is their high cost (Abrams
& Kihm, 2015; Kochkin, 2007). The average pair of hear-
ing aids costs approximately $4,700 (President’s Council of
Advisors on Science and Technology, 2015), which can im-
4. https://doi.org/10.1044/2018_AJA-18-0027
loss and are therefore not regulated as a medical device by
the Food and Drug Administration. Because PSAPs are
intended to be sold directly to consumers, they can be con-
sidered a type of over-the-counter (OTC) amplification
device. Because of their significantly lower cost, PSAPs
find their place among consumers who report some diffi-
culty in hearing but are not yet willing to spend thousands
of dollars on audiologist-fitted hearing aids. PSAPs typi-
cally look like traditional hearing aids, both behind-the-
ear and in-the-ear styles, whereas some resemble ear-level
Bluetooth devices. PSAPs cost between $25 and $500
(President’s Council of Advisors on Science and Technol-
ogy, 2015).
Although PSAPs are not meant for those with hear-
ing loss, it was found that 1.5 million people with hearing
impairments use either a PSAP or OTC device to compen-
sate for their communication difficulties (Kochkin, 2010).
Among a group of surveyed Americans who reported hear-
ing difficulties, 9.4% reported owning a PSAP (Abrams &
Kihm, 2015). Kochkin (2010) found that approximately
72% of those who reportedly used PSAPs had hearing
loss configurations similar to those of patients who used
audiologist-fitted hearing aids. Overall, the demography
of those who purchased PSAPs was similar to those who
purchased audiologist-fitted hearing aids in terms of age,
employment, and education (Kochkin, 2010). A few sig-
nificant differences between the groups were observed by
Kochkin (2010). Specifically, PSAP users earned, on aver-
age, $10,000 less per year than those using audiologist-fitted
hearing aids and were less likely to pursue bilateral amplifi-
cation. In addition, PSAP users, on average, wore their
devices for only 3 hr a day compared to the average 10 hr
5. of use per day reported by hearing aid owners. Lastly, male
individuals were reportedly more likely to purchase PSAPs
compared to female individuals, whereas in the audiologist-
fitted hearing aid market, the gender breakdown was more
equal.
Earlier studies have reported that many OTC devices
provided unsuitable levels of low-frequency gain with in-
sufficient high-frequency gain (Callaway & Punch, 2008;
Chan & McPherson, 2015; Cheng & McPherson, 2000;
Reed, Betz, Lin, & Mamo, 2017; Smith, Wilber, & Cavitt,
2016). For example, Chan and McPherson (2015) reported
that the majority of the evaluated OTC devices in their
study demonstrated linear input–output characteristics,
peak clipping, high levels of equivalent input noise, and lit-
tle usable high-frequency gain. Research further indicated
that many PSAPs or OTC devices had gain frequency re-
sponses that were most appropriate for rising hearing losses
due to a greater low-frequency emphasis (Chan & McPherson,
2015; Cheng & McPherson, 2000). However, recent research
suggested that modern PSAPs could be an appropriate
solution for those with mild-to-moderate hearing loss. For
example, Smith et al. (2016) examined the ability of PSAPs
to match prescriptive targets for 10 hypothetical audio-
grams of varying severity and found that certain higher-end
PSAPs could be appropriately fit to an individual with up
to a moderate degree of hearing loss. Reed, Betz, Lin, et al.
(2017) found that several high-end PSAPs tested had elec-
troacoustic characteristics within the tolerances used to
assess hearing aid function for frequency range, equivalent
input noise, maximum output, and total harmonic distor-
tion measures.
Previous research has also investigated the effect of
PSAPs on patient outcomes in the laboratory. For exam-
6. ple, Xu, Johnson, Cox, and Breitbart (2015) compared the
perceived sound quality of PSAPs relative to hearing aids.
Results showed a significant preference for hearing aids
over PSAPs only when listening to quiet conversation. No
differences in sound quality were observed between any de-
vices for everyday noises or music listening. More recently,
Reed, Betz, Kendig, Korczak, and Lin (2017) compared
five PSAPs with a traditional hearing aid. Results in speech
recognition performance in noise indicated that three of
the five PSAPs performed within 5 percentage points of the
hearing aid.
Research investigating the real-world benefit of PSAPs
and OTC devices is limited. Humes et al. (2017) compared
a direct-to-consumer service delivery model, which could
be used to dispense OTC devices to an audiologist-based
hearing aid service delivery model. The direct-to-consumer
delivery model was shown to be efficacious, yielding out-
comes similar to the traditional audiologist-based model.
However, high-end hearing aids, rather than PSAPs or
OTC devices, were used by Humes et al. Recently, Acosta,
Hines, and Johnson (2018) conducted a double-blinded ran-
domized control field trial to evaluate the effect of self-fitted
OTC devices relative to audiologist-fitted hearing aids.
Outcomes in speech communication and sound aversiveness
were measured 1 week posttrial using the Abbreviated
Profile of Hearing Aid Benefit (Cox & Alexander, 1995).
Results indicated that, although the mean benefit score in
speech communication of OTC devices was higher (better)
than that of hearing aids (Cohen’s d = 0.54), hearing aids
had a higher (better) benefit score in aversiveness than OTC
devices (Cohen’s d = 0.78). However, likely due to the small
sample size of the trial (total n = 17), none of the effects
were statistically significant.
Not only is the effect of PSAPs and OTC devices in
7. the real world unclear (for the reasons mentioned above),
there are also gaps in the literature regarding the benefit
of these devices relative to traditional hearing aids in well-
controlled laboratory environments. For example, in the
study by Xu et al. (2015) that compared the sound quality
of PSAPs and hearing aids, stimuli were prerecorded by
placing the devices on a manikin’s ear. Furthermore, the
hearing aids used to record stimuli were not configured to
compensate for each participant’s hearing loss. Therefore,
it is unclear if the results of Xu et al. (2015) could general-
ize to PSAPs and hearing aids used in the real world. In
the study by Reed, Betz, Kendig, et al. (2017), speech recog-
nition testing was conducted with research participants
wearing the PSAPs adjusted using best-practice verification
protocols by an audiologist, rather than by the participants
themselves. The unrealistic PSAP fitting scheme used by Reed,
Betz, Kendig, et al. (2017) could limit the generalizability of
582 American Journal of Audiology • Vol. 27 • 581–593 •
December 2018
the study. Specifically, two PSAPs used by Reed, Betz,
Kendig, et al. (2017) had smartphone application soft-
ware (app) that allowed users to fine-tune the gain frequency
response of the device. Previous research has shown that
audiologist-driven hearing aid fine-tuning yielded better
outcomes than patient-driven fine-tuning (Boymans &
Dreschler, 2012). Thus, it is unclear if the PSAPs would
still yield similar outcomes compared to audiologist-fitted
hearing aids if the PSAPs were configured by the users.
Furthermore, Reed, Betz, Kendig, et al. (2017) did not spec-
ify if the device’s volume control was available to partici-
pants during testing. Hence, it is also unclear how users
would select the volume levels for PSAPs and how this
8. would affect differences in outcomes between PSAPs and
traditional hearing aids.
To fill gaps in the literature, the purpose of the cur-
rent study was to examine the effect of PSAPs on speech
recognition, listening effort, and sound quality compared
to audiologist-fitted hearing aids using relatively realistic
settings in the laboratory. Specifically, participants were
responsible for selecting their preferred settings in an effort
to simulate the real-world self-fitting process of many
PSAPs on the market. In addition, outcome measures in
the current study were administered using test conditions
that represent real-world listening situations (i.e., ecologi-
cally relevant). It is well established that the effect of
certain hearing aid features, such as directional microphones,
highly depends on the characteristics of listening situations
(e.g., the locations of speech and noise) (Ricketts, 2000).
Because directional microphones are available in many
modern PSAPs, outcomes measured in ecologically rele-
vant conditions could be more generalizable to the real
world.
Method
Participants
Twenty-five adults (12 men and 13 women) were
recruited from the community and completed the study.
Participants were eligible for inclusion in the study based
on the following criteria: (a) bilateral mild-to-moderate sen-
sorineural hearing loss, (b) a maximum threshold of 75 dB
HL up to 4000 Hz, (c) hearing threshold symmetry within
15 dB up to 4000 Hz, and (d) the ability to understand the di-
rections of the experiment and perform experiment-related
tasks. The mean pure-tone thresholds are shown in Figure 1.
Participants ranged in age from 56 to 81 years, with a mean
of 69.6 years (SD = 8.2). All but one participant were ex-
9. perienced hearing aid users, with the mean years of use be-
ing 8.4 years (SD = 9.7).
Amplification Devices and Fitting
A traditional hearing aid and three PSAPs were used
in the current study. The traditional hearing aid was a
ReSound LiNX2 5, denoted as the HA. The HA was
a midlevel, behind-the-ear, receiver-in-the-canal device
equipped with wide dynamic range compression (nine
channels), adaptive directional microphones, digital noise
reduction algorithms, and a smartphone app for volume
adjustment. The three PSAPs used were a Sound World
Solution
s CS50+, a FocusEar RS2, and a Tweak Focus,
denoted as PSAP1, PSAP2, and PSAP3, respectively. The
PSAPs were all behind-the-ear slim-tube style. The prices
of PSAP1, PSAP2, and PSAP3 were $349, $399, and $299
(per device), respectively. The PSAPs were chosen to rep-
resent midlevel to high-end PSAPs. All PSAPs had direc-
tional microphones and noise reduction features. PSAP1
had a corresponding smartphone app that allows users to
adjust the device’s volume and frequency response. Users
could also use the three-channel equalizer feature (bass,
mid, and treble) of the app to adjust the shape of the fre-
10. quency response of PSAP1.
All devices were fitted bilaterally. The HA was fitted
by an audiologist using real-ear measurements matched
to National Acoustics Laboratory (NAL-NL2; Keidser,
Dillon, Flax, Ching, & Brewer, 2011) nonlinear prescrip-
tive targets for an average-level speech input (65 dB SPL)
using a clinically appropriate dome. Two user programs
were configured: the first program (denoted as P1) was
for quiet situations (omnidirectional microphone with noise
reduction turned off), and the second program (P2) was
for noisy situations (adaptive directional microphone with
digital noise reduction enabled). The frequency responses of
the two programs were equalized. The features in P2 were
set to the default as suggested by the manufacturer. The
device’s frequency response was not further fine-tuned based
on user’s feedback.
Figure 1. Mean hearing thresholds for the participants. Right
and left ears are offset from one another. Error bars indicate
1 SD.
Brody et al.: Comparing PSAPs and Hearing Aids 583
11. The PSAPs were fitted using the default earpiece rec-
ommended by the manufacturers (PSAP1 and PSAP2:
closed dome; PSAP3: power dome). PSAP1 had three pre-
determined frequency responses (described as presets),
so participants were given the opportunity to select and
fine-tune the preset (see below for the procedure). PSAP2
and PSAP3 had only one predetermined frequency response
that could not be altered by the end user. All PSAPs had
multiple user programs preconfigured by the manufacturers,
which could not be disabled. For all PSAPs, the default
program (P1) was for listening in quiet environments and
the second program (P2) was for noisy listening situations.
In P2, directional microphones and noise reduction fea-
tures were enabled. PSAP1 and PSAP2 had additional pro-
grams for “entertainment” and telephone use. These
additional programs were not tested in the current study;
only P1 and P2 were used in the experiment.
To better simulate user adjustments of hearing aids
and PSAPs, participants individually selected their pre-
ferred volume levels in each program for each device. Be-
cause PSAP1 had three presets and its smartphone app
had a three-channel equalizer feature, participants were
allowed to select their preferred preset for each program
12. and use the equalizer to fine-tune the frequency response
in addition to adjusting the volume. To determine the
volume level for all devices (and adjust the frequency re-
sponse of PSAP1), participants were positioned in a sound
field created using an eight-loudspeaker array in a sound-
treated booth. Eight Tannoy Di5t loudspeakers (Tannoy)
were located at 0°, 45°, 90°, 135°, 180°, 225°, 270°, and
315° azimuth relative to the participant. The distance be-
tween the loudspeaker and the participant was 1 m. Con-
tinuous sentences from the Connected Speech Test (CST;
Cox, Alexander, & Gilmore, 1987) were presented from 0°
azimuth, and the CST babble noise was presented from
all eight loudspeakers. Speech and babble noise were pre-
sented at 60 and 40 dBA, respectively, for adjustments
made in P1 and at 68 and 61 dBA, respectively, for adjust-
ments in P2. These levels were selected to represent real-
world listening situations (Wu et al., 2018). A Larson
Davis 2560 0.5-in. random incidence microphone and a
Larson Davis System 824 sound-level meter were used to
calibrate the sound field (as well as the sound fields used in
the outcome measures described below). The microphone
was placed at the position of the listener’s head. To cali-
brate the noise, the CST babble presented from each loud-
speaker was first measured. The level and spectrum of
each babble noise were adjusted to be equal across the eight
13. loudspeakers. The overall level of the babble noise (all loud-
speakers presented at once) was then set to 40 dBA (for P1
adjustment) or 61 dBA (for P2).
Participants were instructed to adjust each device
using the following instructions, “Imagine you are having
a conversation and adjust the device so that you can best
understand the speech while maintaining comfort.” Par-
ticipants adjusted left and right devices individually and
could take as much time as they needed. Adjustments were
made by pushing buttons on the devices (PSAP2 and PSAP3)
or using a smartphone app (HA and PSAP1). For PSAP1,
preset selection and fine-tuning were conducted before
adjusting volume levels. A researcher was available to assist
participants (e.g., changing the preset of PSAP1) during
device adjustments. Pairing the smartphone (a Samsung
Galaxy S6) to the devices (HA and PSAP1) was completed
by the researcher before the adjustment process. For all
devices, individual participant-selected settings in each pro-
gram were recorded and were used for all testing in the study.
To illustrate the electroacoustic characteristics of the
devices, a series of Verifit 2 (Audioscan) test box mea-
sures were conducted. The HA was configured to fit the
14. average hearing loss of the participants. All devices were
set using the most commonly selected user settings (e.g.,
volume level and preset) across all participants in both
programs. Figure 2 shows the input–output functions at
2000 Hz of P1 (A) and P2 (B) of each device. The figure
indicates that the HA and PSAP1 utilized compression at
various input levels whereas PSAP2 and PSAP3 demon-
strated less compression. To quantify the device’s directiv-
ity, a Verifit directivity measure (70 dB SPL with a +3 dB
SNR) was conducted with the devices set in P2. The direc-
tivity averaged from 500 to 5000 Hz (American National
Standards Institute, 2010) for the HA, PSAP1, PSAP2,
and PSAP3 was 8.4, 14.8, 9.7, and 9.5 dB, respectively.
Figure 2. Input–output functions at 2000 Hz measured for each
device in P1 (A) and P2 (B). Devices were set to the most
common
selected settings, averaged across all 25 participants. HA =
hearing
aid; PSAP = personal sound amplification product.
584 American Journal of Audiology • Vol. 27 • 581–593 •
December 2018
15. Finally, the Verifit test box noise reduction measure (multi-
talker babble noise stimulus at 70 dB SPL) showed that
the average noise reduction from 500 to 5000 Hz for the
HA, PSAP1, PSAP2, and PSAP3 in P2 was 2.8, 2.9, 3.5,
and 0.4 dB, respectively.
Outcome Measures
Speech Intelligibility Index
The Speech Intelligibility Index (SII; American National
Standards Institute, 2012) was used to quantify speech audi-
bility. Following device fittings and adjustments, an SII
was measured on-ear with a 65-dB SPL speech input using
a probe microphone and the Verifit hearing aid analyzer in
both P1 and P2 for all devices using the participant-selected
settings. An unaided SII was also recorded for comparison
purposes. In addition, a real-ear aided response (REAR)
with an input of 65 dB SPL was recorded for each device in
each program using the subject-selected settings.
Hearing in Noise Test
The Hearing in Noise Test (HINT; Nilsson, Soli, &
Sullivan, 1994) was used to measure participants’ speech
16. recognition thresholds (SRTs) both in quiet and in noise.
The HINT was administered in the same sound field used
for device adjustment. To measure the SRT in quiet, the
HINT sentences were presented from 0° azimuth without
noise and the devices were set to P1. To measure the SRT
in noise, additional uncorrelated HINT noise was pre-
sented using the eight-loudspeaker array. The overall level
of the noise (all loudspeakers presented at once) was fixed
at 65 dBA, and the devices were set to P2. Concatenated
HINT sentences and the HINT speech-shaped noise were
used to calibrate the sound field. In both test conditions
(in quiet and in noise), the listener was asked to repeat a
block of 20 HINT sentences (i.e., two lists). The speech
level was adjusted adaptively, depending on the partici-
pant’s responses, using the one-up-one-down procedure
(4-dB steps for the first four sentences and 2-dB steps for
the remaining sentences). The correct response regarding
each sentence was based on the repetition of all the words
in the sentence, with minor exceptions such as “a” and
“the.” The presentation level (quiet condition) or signal-to-
noise ratio (SNR; noise condition) of the final 17 presenta-
tions was averaged to derive the SRT. Both test conditions
of the HINT were administered for each device as well as
in an unaided condition.
17. Connected Speech Test
Estimated real-world speech recognition perfor-
mance was assessed using the CST. The test is composed
of 48 passages of conversational connected speech bro-
ken up into specific topics. Each passage contains nine
to 10 sentences, and 25 target words are used for scoring.
The CST was selected because passages could be presented
with or without a visual component and using connected
speech offered more ecological validity compared to single
words or unrelated sentences. The CST was conducted in
conditions designed to represent real-world listening situa-
tions. Specifically, Wu et al. (2018) examined the real-world
listening environments of older adults with hearing loss
and developed 12 “prototype listening situations” (PLSs).
These PLSs describe the speech level, noise level, avail-
ability of visual cues, and locations of speech and noise
sources of typical speech listening situations experienced by
older adults with hearing loss. The PLSs could enable more
ecologically valid assessment protocols in the laboratory
to evaluate real-world outcomes (Walden, 1997; Wu et al.,
2018).
In the current study, the six most frequent PLSs de-
18. scribed in Wu et al. (2018) were created for the CST testing
using the eight-loudspeaker sound field described above.
Wu et al. (2018) indicated that these six test environments
would represent 71% of daily speech listening situations.
The six PLSs (denoted as PLS1 to PLS6) were broken down
into two subgroups: quiet and noise. For the three quiet
PLSs, speech and noise were presented at 60 and 40 dBA,
respectively (20 dB SNR). For the noisy PLSs, speech and
noise were presented at 68 and 61 dBA, respectively (7 dB
SNR). Note that noise was used in the quiet PLSs because,
in the real world, listening environments that are completely
quiet are rare (Wu et al., 2018). Within each subgroup, there
were three configurations: speech presented from 0° azimuth
with visual cues present, speech presented from 90° azimuth
with visual cues absent, and speech presented from 0° azi-
muth with visual cues absent. See Figure 3 for the charac-
teristics of the PLSs. Visual cues (i.e., the talker’s face) were
presented on a 17-in. computer monitor, which was placed
right below the loudspeaker at 0° azimuth. For all PLSs,
Figure 3. Characteristics of the six prototype listening
situations
(PLS). Circles indicate the eight-loudspeaker array surrounding
the
participant (center). Black circles indicate location of the
19. speech.
Eye graphic indicates presence of visual cues. For PLS1 to
PLS3,
speech and noise were presented at 60 and 40 dBA,
respectively.
For PLS4 to PLS6, speech and noise were presented at 68 and
61 dBA, respectively.
Brody et al.: Comparing PSAPs and Hearing Aids 585
uncorrelated CST babble was presented from each of the
eight loudspeakers surrounding the listener, and the overall
level (all loudspeakers presented at once) was set to 40 or
61 dBA, depending on the PLS. For each test condition,
two CST passages (50 target words) were presented and
scored. Performance was scored based on the percentage
of target words repeated correctly in each condition. All
devices were set to P1 for testing in quiet environments
(PLS1, PLS2, PLS3) and P2 for testing in noisy environ-
ments (PLS4, PLS5, PLS6). The CST was also adminis-
tered in an unaided condition.
Listening Effort Measure
20. Measures of speech recognition (e.g., the CST) are a
useful way to quantify communication. However, research
suggests that even when speech understanding scores under
two conditions are similar, the level of listening effort
might be different (Sarampalis, Kalluri, Edwards, & Hafter,
2009). To measure listening effort, the participants were
asked to rate their perceived listening effort after listening
and repeating CST sentences in each of the six PLSs. The
participants answered the question “How hard were you
working to achieve your level of speech understanding?”
using a 21-point scale ranging from 0, representing not at
all, to 100, representing very, very hard. This subjective
listening effort measure, rather than objective measures
such as dual-task paradigms, was selected because subjec-
tive measures of listening effort can be more sensitive than
objective measures (Johnson, Xu, Cox, & Pendergraft, 2015;
Seeman & Sims, 2015).
Sound Quality Rating
Subjective sound quality judgments were recorded
as a separate metric to differentiate between devices. Reports
of poor sound quality may be associated with hearing
aid nonuse and dissatisfaction (Solheim, Gay, & Hickson,
21. 2017). Perceived sound quality was measured in a manner
similar to listening effort. Specifically, after listening and
repeating CST sentences in a given condition, the partici-
pants answered the question “How would you judge the
overall sound quality?” using a 21-point scale ranging
from 0, representing very poor, to 100, representing excel-
lent. Sound quality ratings were obtained in only the aided
conditions.
Procedure
The study was approved by the institutional review
board of the University of Iowa. After consenting to the
study protocol, participants’ hearing thresholds were mea-
sured using pure-tone audiometry. If the participant met
all required inclusion criteria, the devices were fitted and
adjusted. Next, outcome measures were administered. For
all testing, a practice condition was administered to con-
firm participants’ understanding of the task. The order of
device condition, PLS, HINT lists, and CST passage pairs
were randomized across participants. The devices were
inserted into the participant’s ears by a researcher so that
participants were blinded to the device. However, researchers
were not blinded to the devices when scoring test materials
22. (single-blinded design). Testing was completed in a series
of two 2-hr sessions. Monetary compensation was pro-
vided to the participants upon completion of the study.
Results
REAR
Figure 4 shows the mean REAR (averaged across
all 50 ears), measured using a 65-dB SPL speech input,
of each device in P1 (A) and P2 (B) with the participant-
selected settings. Mean REAR targets prescribed by
NAL-NL2 averaged across participants are also shown
in the figure. The REARs of the HA and PSAP1 are
closest to the prescribed targets. In contrast, PSAP2
and PSAP3 underamplified speech sounds at frequencies
above 2000 Hz. Root-mean-square deviations between the
NAL-NL2 targets and each device at 500, 1000, 2000, and
4000 Hz were calculated and averaged across all 50 ears.
Root-mean-square deviations for the HA, PSAP1, PSAP2,
and PSAP3 …
University of South Carolina
COMD 500
Article Review Summary
Due date: April 25, 2019 (11:59 pm)
23. Carefully read one of the three research articles posted on
Blackboard. Once you’ve read the article, answer each of the
following questions based on information from the article. Save
the assignment as “Last name_First
name_ArticleReviewAssignment” (example:
“Smith_Jane_ArticleReviewAssignment”) and submit your
assignment through Blackboard.
1. Provide a complete, accurate citation of the article in APA
style below. Use a hanging indent!
Citation:
2. What is the independent variable (or variables)?
3. What is the dependent variable (or variables)?
4. How was the dependent variable measured? List the
instruments, tools, or technology that were used to measure the
dependent variable.
5. Introduction: Summarize the background of the study in a
few sentences. Highlight only the main points—those that are
most relevant to the study you read.
24. 6. Introduction: What was the purpose of the study?
7. Introduction: What were the research questions or
hypotheses? (Usually found in the last paragraph before the
Methods section.)
8. Methods: Who were the participants in the study? List the
number of participants and summarize their demographic
characteristics briefly.
9. Methods: What did the participants do in the study? In other
words, what procedures were used for data collection?
Summarize these procedures briefly.
10. Results and Discussion: Briefly summarize the results of the
study as they relate to the research questions or hypotheses
(from question #7 above). Use normal, everyday language;
statistical terminology is not needed.
11. Discussion: What are the conclusions and implications of
the results that the authors describe in the Discussion section?
12. Reflection: What did you think about this article? What is
the “take away” message, in your opinion?