SlideShare a Scribd company logo
1 of 6
Download to read offline
The Leicester Cough Monitor: preliminary
validation of an automated cough detection
system in chronic cough
S.S. Birring*, T. Fleming*, S. Matos#
, A.A. Raj"
, D.H. Evans#
and I.D. Pavord"
ABSTRACT: Chronic cough is a common condition that presents to both primary and secondary
care. Assessment and management are hampered by the absence of well-validated outcome
measures. The present study comprises the validation of the Leicester Cough Monitor (LCM), an
automated sound-based ambulatory cough monitor.
Cough frequency was measured with the LCM and compared with coughs and other sounds
counted manually over 2 h of a 6-h recording by two observers in nine patients with chronic cough
in order to determine the sensitivity and specificity of the LCM. Automated cough frequency was
also compared with manual counts from one observer in 15 patients with chronic cough and eight
healthy subjects. All subjects underwent 6-h recordings. A subgroup consisting of six control and
five patients with stable chronic cough underwent repeat automated measurements o3 months
apart. A further 50 patients with chronic cough underwent 24-h automated cough monitoring.
The LCM had a sensitivity and specificity of 91 and 99%, respectively, for detecting cough and a
false-positive rate of 2.5 events?h-1
. Mean¡SEM automated cough counts?patient?h-1
was 48¡9 in
patients with chronic cough and 2¡1 in the control group (mean difference 46 counts?patient?h-1
;
95% confidence interval (CI) 20–71). The automated cough counts were repeatable (intra-subject
SD 11.4 coughs?patient?h-1
; intra-class correlation coefficient 0.9). The cough frequency in
patients undergoing 24-h automated monitoring was 19 coughs?patient?h-1
; daytime (08:00–
22:00 h) cough frequency was significantly greater than overnight cough frequency (25 versus
10 coughs?patient?h-1
; mean difference 15 coughs?patient?h-1
, 95% CI 8–22).
The Leicester Cough Monitor is a valid and reliable tool that can be used to assess 24-h cough
frequency in patients with cough. It should be a useful tool to assess patients with cough in
clinical trials and longitudinal studies.
KEYWORDS: Chronic cough, cough counts, cough frequency, cough monitor, Leicester Cough
Monitor
C
hronic cough is a common reason for
referral to respiratory physicians. The
assessment of patients with chronic
cough is often based on the anatomical diagnostic
protocol, which is a systematic evaluation based
on the understanding that most cases are due to
disease of the upper respiratory tract where
cough receptors are most plentiful [1]. The most
common conditions implicated in causing
chronic cough in nonsmokers are asthma, gastro-
oesophageal reflux and rhinitis, or a combination
of these [2].
The identification of an important contribution
by the aforementioned conditions is largely based
on the evaluation of treatment trials [3].
However, there are few well-validated outcome
measures to assess cough severity and treatment
efficacy. Cough visual analogue scores, diary
score cards, quality-of-life questionnaires, cough-
reflex sensitivity measurement and cough moni-
tors have been proposed as potential tools to
assess cough severity [2]. The subjective nature of
symptom scores and quality-of-life question-
naires [4, 5] and the poor specificity of cough-
reflex sensitivity measurement [6] to identify
patients with chronic cough have led to a
renewed interest in the development of auto-
mated ambulatory cough monitors [7–15].
BIRRING et al. [7] have previously shown that
there are marked differences in cough frequency
between patients with chronic cough and healthy
subjects and that these measurements are repea-
table, and have suggested that cough frequency
AFFILIATIONS
*Dept of Respiratory Medicine,
King’s College Hospital, London,
#
Dept of Medical Physics, Leicester
Royal Infirmary, and
"
Institute for Lung Health, Dept of
Respiratory Medicine, Glenfield
Hospital, Leicester, UK.
CORRESPONDENCE
I.D. Pavord
Dept of Respiratory Medicine
Glenfield Hospital Leicester
LE3 9QP
UK
Fax: 44 1162367768
E-mail: ian.pavord@uhl-tr.nhs.uk
Received:
May 10 2007
Accepted after revision:
December 16 2007
STATEMENT OF INTEREST
None declared
European Respiratory Journal
Print ISSN 0903-1936
Online ISSN 1399-3003
EUROPEAN RESPIRATORY JOURNAL VOLUME 31 NUMBER 5 1013
Eur Respir J 2008; 31: 1013–1018
DOI: 10.1183/09031936.00057407
CopyrightßERS Journals Ltd 2008
c
measurement is potentially useful in the assessment of patients
with chronic cough. Currently available cough monitors are
limited by difficulty in achieving unrestricted ambulatory
measurement in the patients’ own environment, an inability to
perform 24-h recording and a lack of automated cough
detection systems. The aim of the present study was to
develop an automated cough monitor (the Leicester Cough
Monitor (LCM)) capable of recording cough sounds for 24 h.
The present study shows the validation of the LCM and
preliminary findings of 6- and 24-h recordings in patients with
chronic cough.
METHODS
Subjects
A total of 15 consecutive patients with an isolated chronic
cough (.3 weeks duration) were recruited from a specialised
cough clinic. The clinic receives referrals from primary and
secondary care largely confined to a population of 970,000
within Leicestershire, UK. The causes of cough in patients with
chronic cough were: cough variant asthma (n54); gastro-
oesophageal reflux (n53); eosinophilic bronchitis (n51);
idiopathic (n53); post-viral (n51); bronchiectasis (n51);
chronic obstructive pulmonary disease (n51); and chronic
bronchitis (n51). Nine out of these patients were randomly
selected for the first stage of validation (cough variant asthma
(n54), eosinophilic bronchitis (n51), idiopathic (n52), post-
viral (n51), bronchiectasis (n51)) and all patients were
included in the second validation stage. Investigations were
carried out according to a standardised algorithm [16]. The
protocol for investigation and treatment and the criteria for
accepting diagnosis were as previously described [16]. Eight
controls were recruited from healthy volunteers responding to
local advertising. Control subjects were asymptomatic, non-
smokers and had normal spirometry and a concentration of
methacholine required to cause a 20% decrease in forced
expiratory volume in one second (FEV1) of .16mg?mL-1
. No
patients had received corticosteroids or other specific treat-
ment for the condition causing cough for o6 weeks prior to
the study. A randomly selected subgroup of six control
subjects and five patients with a stable chronic cough and
stable treatment requirements (three with cough variant
asthma, one with gastro-oesophageal reflux-associated cough
and one with idiopathic chronic cough) participated in cough
frequency repeatability studies 3–6 months after the first, at the
same time of day in order to avoid possible bias from diurnal
variations. A total of 50 further consecutive patients with
chronic cough were recruited in order to evaluate 24-h
recordings with the LCM (idiopathic cough (n526), asthma
(n58), eosinophilic bronchitis (n52), rhinitis (n52), sarcoidosis
(n52), gastro-oesophageal reflux (n53), bronchiectasis (n52),
chronic obstructive pulmonary disease (n52), enlarged tonsils
(n52) and obstructive sleep apnoea (n51)). All subjects gave
full informed written consent to participate. The protocol for
the present study was approved by the Leicestershire Research
Ethics Committee.
Cough monitor
The LCM (fig. 1) is a digital ambulatory cough monitor that
records sound continuously from a free-field microphone
necklace (Sennheiser MKE 2-5; Sennheiser electronic GmbH &
Co. KG, Woedemark, Germany) onto a digital sound recorder
(dimensions 26.7687632 mm; iRiver iFP-799; iRiver Europe
GmbH, Eschborn, Germany) at a sampling frequency of
16 kHz and with an encoding bit rate of 64 kbit?s-1
. The cough
monitor was attached at 09:00 h in all subjects and returned 6–
24 h later. Subjects were told that the LCM was a new
investigative tool being developed to assess the nature of the
cough and were encouraged to resume their normal activity in
their usual environment. When the recording was complete,
data stored on the recorder was downloaded onto a computer,
where it was analysed by an automated cough detection
algorithm (the Leicester Cough Algorithm).
A general outline of the Leicester Cough Algorithm has been
described previously [8, 17]. Briefly, the detection algorithm is
based on Hidden Markov Models, a statistical method that can
be used to characterise the spectral properties of a time-
varying pattern. The cough detection algorithm was imple-
mented based on the keyword-spotting approach, as defined in
speech recognition, in which the objective is to detect the
occurrence of a particular set of keywords in a sequence of
continuous speech. Continuous ambulatory recordings in
patients with chronic cough were used to train statistical
models of the characteristics of cough sounds and of the audio
background. During the detection process, the recorded audio
signal was divided into contiguous 10-s segments to be
analysed by the Hidden Markov Models-based algorithm.
Each 10-s audio segment was recognised by the detection
algorithm as a sequence of variable-length audio sections, each
in turn classified either as background audio or as a possible
cough sound, depending on its statistics [8]. A second
algorithm phase then used brief operator input to facilitate
the automated algorithm in order to eliminate sounds that
Microphone
Recorder
FIGURE 1. The Leicester Cough Monitor.
LCM: A NOVEL COUGH MONITOR S.S. BIRRING ET AL.
1014 VOLUME 31 NUMBER 5 EUROPEAN RESPIRATORY JOURNAL
might have been wrongly classified as cough events in the first
phase. For this, the operator is asked to classify, as cough or
otherwise, a small fraction of the sounds detected in the first
phase as possible cough sounds (the second phase takes 5 min
to carry out for a 24-h recording and ,50 sounds are
classified). The information is then used to create statistical
models that are adapted to the characteristics of the cough
sounds for that particular recording and the remaining sounds
that were not shown to the operator are classified using these
models.
Cough was defined as a characteristic explosive sound. The
algorithm identifies coughs as single events whether they
occur as isolated events or in a cluster (i.e. attempts were made
to determine how many coughs occurred in paroxysms).
Validation
Stage 1
The first stage of validation compared automated cough counts
against those identified by manual sound analysis of the first
and fourth recorded hours of nine randomly selected patients
with chronic cough. Manual analysis of sound recordings
consisted of three blinded observer counts (observer one twice
and observer two once) and cough or noncough sounds were
positively identified when all three counts were in agreement
based on sound and visual inspection of the acoustic trace.
Each cough sound was identified separately, whether it
occured singularly or in a cluster or epoch of coughs. Intra-
and interobserver variability in cough counts was established
from the two blinded analyses performed by observer one and
by comparing the mean of observer one’s counts for these
periods with counts obtained by observer two. The recordings
were analysed twice using the Leicester Cough Algorithm to
assess intra-recording repeatability. In order to classify non-
cough sounds and determine whether particular sounds were
wrongly classified by the automated system, observer one
listened to all 6 h of the nine patients’ recordings and classified
all recognisable sounds. The results of this analysis were
compared with the automated classification.
Stage 2
The second stage of validation was extended to all recordings
and compared automated cough counts against coughs
identified manually by observer one, who analysed the entire
6-h recording. A further 50 patients with chronic cough
underwent 24-h automated cough frequency measurement.
Automated cough frequency was compared for repeatability
studies.
Analysis
Subject characteristics were described using descriptive statis-
tics and expressed as mean¡SEM for parametric data and
median for non-parametric data. Cough frequency was
expressed as individual coughs?patient?h-1
for the duration of
the recording. The validity of the LCM was presented as
sensitivity, specificity and false-positive rate of the automated
algorithm for detecting coughs as measured by observer
manual analysis. Intra- and interobserver variability of manual
cough counts and repeatability data was assessed as intra-class
correlation coefficients and intra-subject SD.
RESULTS
The subject characteristics are shown in table 1.
Validation stage 1 (First and fourth recorded hour)
Mean cough counts were 39 coughs?patient-1
?h-1
by automated
analysis compared with 43 coughs?patient-1
?h-1
by manual
analysis (mean difference -4 coughs?patient-1
?h-1
, 95% confi-
dence interval (CI) -6–13; p50.4). The intra- and interobserver
intra-class correlation coefficients for manual analysis of sound
recordings (between observers one and two) were 0.99 and
0.98, respectively (both p,0.001). The intra- and interobserver
(i.e. between mean of observer one and observer two) intra-
class correlation coefficients for manual analysis of sound
recordings were 0.99 and 0.98, respectively (both p,0.001). The
intra-class correlation coefficient between automated and
manual observer counts was 0.9 (p,0.001; fig. 2a). The
accuracy of manual and automated cough counts appeared
similar in recordings containing paroxysms and those with
isolated coughs. The Leicester Cough Algorithm had a
sensitivity and a specificity of 91% and 99%, respectively, for
detecting cough sounds and a median false positive rate of
2.5 events?patient-1
?h-1
against the gold standard of coughs
detected manually by observer one twice and observer two
once. There was no evidence that any particular sound was
more likely to be classified as a false positive (fig. 3).
Validation stage 2 (6-h recordings)
Mean¡SEM automated cough counts were 48¡9 coughs?
patient-1
?h-1
in patients with chronic cough and 2¡1 coughs?
patient-1
?h-1
in control subjects (mean difference 46 coughs?
patient-1
?h-1
, 95% CI 20–71 coughs?patient-1
?h-1
; p,0.001;
fig. 4). There were no significant differences in cough
frequency between diagnostic groups. The cough analysis
took 2 h to complete, comprising 5 min for data download,
105 min for computer automated analysis (an operator was not
required to be present during phase 1) and 10 min for operator
input (phase 2) and printing results.
The intra-class correlation coefficient between automated and
observer counts was 0.93 (p,0.001; fig. 2b). The LCM had
sensitivity and specificity of 86 and 99%, respectively, for
detecting cough sounds and a median false-positive rate of
1.0 events?patient-1
?h-1
.
The automated cough counts were repeatable in the 11 subjects
who underwent repeatability testing (intra-subject SD 11.4 coughs?
patient-1
?h-1
, intra-class correlation coefficient 0.9; fig. 5).
TABLE 1 Subject characteristics
Control Chronic cough
Subjects (male) n 8 (0) 15 (5)
Age yrs 48¡3 55¡4
Cough duration yrs 5¡2
FEV1 % pred 91¡5 89¡7
FEV1/FVC % 79¡1 75¡4
Data are expressed as mean¡SEM, unless otherwise stated; FEV1: forced
expiratory volume in one second; pred: predicted; FVC: forced vital capacity.
S.S. BIRRING ET AL. LCM: A NOVEL COUGH MONITOR
c
EUROPEAN RESPIRATORY JOURNAL VOLUME 31 NUMBER 5 1015
The cough frequency in patients undergoing 24-h monitoring was
19 coughs?patient-1
?h-1
; daytime (08:00–22:00 h) cough frequency
was significantly greater than overnight cough frequency (25
versus 10 coughs?patient-1
?h-1
; mean difference 15 coughs?pa-
tient-1
?h-1
, 95% CI 8–22 coughs?patient-1
?h-1
; p,0.001; fig. 6).
DISCUSSION
The LCM is a lightweight 24-h automated ambulatory cough
monitor that is easy to use and measures cough in the subjects’
own environment. The present study has shown that it is a
valid and reliable tool for objectively measuring cough
frequency. The high sensitivity and specificity for the detection
of cough sounds is comparable to other routine diagnostic
clinical tools and superior to that reported for other more
cumbersome cough detection systems. Preliminary data is
presented in the current study showing that the cough
frequency measured with the LCM is repeatable over
o3 months, a period relevant to the duration of treatment
trials that form an important part of the assessment of patients
with chronic cough. Repeatability was marginally better than
that of recordings analysed manually [7]. The present data
suggest that the LCM might be a particularly useful outcome
measure in assessing patients with cough and measuring the
response to therapy in the clinic and in clinical trials.
A limitation of the present study is that evaluation of cough
frequency was based on 6-h daytime cough recordings owing
u
u
u
u
u
u
u
uuu
uuu
u u
u
u
u
80
40
0
-40
-80
u u
u
u
u
uuu
uu
u
u
u
u
80
40
0
-40
-80
u
u
uuuuuu
a)
b)
0 50 100 150
Differencebetweenautomatedand
observercoughcounts·patient-1·h-1
Mean of automated and observer coughs·patient-1·h-1
Differencebetweenautomatedand
observercoughcounts·patient-1·h-1
FIGURE 2. Bland-Altman plot of automated versus manual observer cough
counts?patient-1
?h-1
. a) Validation stage 1 at which first and fourth recorded hour
(n59) were analysed. Each hour is depicted individually. b) Validation stage 2 at
which 6-h recordings (n523) were analysed in their entirety. The complete cough
detection algorithm (phases 1 and 2) was tested in each validation stage. –––:
mean difference; ??????: 95% limits of agreement (26 intra-subject SD); e: control
subjects; ¤: chronic cough patients.
31726 sounds:
1834 cough
29892 noncough
Phase 1
2543 accepted sounds:
1573 cough
970 noncough#
Phase 2
1680 accepted sounds:
1528 cough
152 noncough+
29183 rejected sounds:
261 cough
28922 noncough
863 rejected sounds:
45 cough
818 noncough¶
FIGURE 3. False positives characterised manually that were detected by the
automated Leicester Cough Monitor in nine 6-h recordings of patients with chronic
cough. Automated analysis of sound recording is performed during phase 1 and
automated analysis following operator input is performed during phase 2. The
sensitivity for cough detection is slightly lower than that from the gold standard
validation stage 1, since the comparator was a 6-h manual counting by observer
one only. #
: noncough sounds were: speech (n5351), impulsive noise (n5222)
throat clearing (n5119), environmental noise (n5111), laugh (n580), other person
coughing (n531), incomplete coughs (n522), child talking/shouting (n521),
sneeze (n56), telephone ringing (n53), burp (n52) and dog barking (n52).
"
: noncough sounds were: speech (n5293), impulsive noise (n5214), throat
clearing (n596), environmental noise (n587), laugh (n567), other person coughing
(n515), incomplete coughs (n518), child talking/shouting (n517), sneeze (n55),
telephone ringing (n53), burp (n52) and dog barking (n51). +
: noncough sounds
were: speech (n558), impulsive noise (n58), throat clearing (n523), environmental
noise (n524), laugh (n513), other person coughing (n516), incomplete coughs
(n54), child talking/shouting (n54), sneeze (n51) and dog barking (n51).
Chronic cough patients
150
125
100
75
50
25
0
Control subjects
Coughs·patient-1·h-1
uu
uuu
u
u
u
u
u
u
uu
u
u
u u uu uuuu
***
FIGURE 4. Mean¡SEM automated cough counts?patient-1
?h-1
in control
subjects and chronic cough patients (6-h recordings). ***: p,0.001.
LCM: A NOVEL COUGH MONITOR S.S. BIRRING ET AL.
1016 VOLUME 31 NUMBER 5 EUROPEAN RESPIRATORY JOURNAL
to limited battery life (6–8 h) at the inception of the study.
Advances in battery technology, since then, have allowed the
extension of recordings to o24 h. The automated system has
allowed these recordings to be analysed relatively quickly and
accurately and should facilitate the investigation of potential
diurnal variations in cough frequency and the effects of
aggravators of potential cough, such as environmental pollu-
tion, cigarette smoking and gastro-oesophageal reflux. In the
present study, a range of sounds including speech, throat
clearing and environmental noises caused false-positive
detected coughs. There was little evidence that any of these
sounds caused particular difficulties with detection, nor did
cough paroxysms appear to present problems for accurate
manual and automated cough counts. However, greater
experience with the monitor may identify sounds or cough
paroxysms that present particular problems for the algorithm
to classify and allow further refinement of the algorithm. The
present study involved small numbers of subjects and it will be
important to study a larger population of control subjects and
patients with well-defined respiratory disease, before and after
treatment, and in different environments, to fully validate the
cough monitor. The present preliminary work suggests that
such a study will be feasible.
Cough frequency was stable through the day and was
significantly reduced overnight compared with daytimes in
accord with previous data, suggesting a diurnal variation in
cough frequency [5, 9, 13]. Further work is required to
determine the validity and the short- and longer-term
repeatability of 24-h cough recordings.
A limitation of the present study is that only 2 h per recording
were used to compare automated cough counts with those
obtained from manual counting for the validation study.
Manual cough counting is very time consuming and laborious
so only the first and fourth hours of each recording were
manually analysed for consistency. Each recording was
manually counted three times in order to obtain a more robust
measure of the true cough frequency. The LCM had a high
sensitivity and specificity for detecting cough against this gold
standard. This was confirmed in the second part of the
validation study where cough counts derived from automated
analysis of 6-h recordings were compared with cough counts
derived from a single manual observer. The sensitivity of the
cough algorithm was slightly lower with 6-h recordings
compared with 2-h recordings. This is most probably because
of the more robust measure of true cough frequency used for
the 2-h recordings, compared with single observer manual
counts used for the 6-h recordings.
A potential criticism of cough counts derived from audio
recordings is that they might not accurately reflect the true
cough rate since it is not possible to visualise the act of
coughing. However, a recent study [18] compared manual
cough counts from audio with video recordings and found
them to be very similar. That study concluded that manual
cough counts from audio recordings should be regarded as the
gold standard to validate cough monitors since audio record-
ings had superior sound quality to that from video recordings.
The LCM quantifies cough frequency as single episodes of
cough rather than epochs or clusters of coughs and cough
seconds (seconds containing cough), as used by others [12].
The present authors believe that single cough episodes are a
more meaningful measure and are easier to interpret by
physicians and patients. Cough frequency, rather than inten-
sity, was measured since cough events are less influenced by
microphone position and muffling of sounds by covering the
mouth during the act of coughing. Furthermore, cough
intensity determined with sound analysis lacks responsiveness
compared with cough frequency in clinical trials of antitussive
drugs [10]. Cough intensity determined by other parameters
such as airflow or chest wall movement is less practical for
routine clinical measurement. The LCM was validated in
subjects with chronic cough due to a wide range of conditions
so it reliably detects coughs with differing characteristics.
One of the challenges of developing cough monitors in the past
has been differentiating cough sounds from throat clearing,
sneezing, speech and other cough-like sounds. The Leicester
Cough Monitor differentiates cough from other sounds reliably
80
60
40
20
0
-20
-40
-60
-80
0 10 20 30 40 50
u
u
u
u
uuu
u
u
u
Differencebetweentests
coughs·patient-1·h-1
Mean of two tests coughs·patient-1·h-1
FIGURE 5. Bland-Altman plot of automated cough counts?patient-1
?h-1
repeated over 3–6 months in chronic cough patients and control subjects. ––––:
mean difference; ???????: 95% limits of agreement (26 intra-subject SD). e: control
subjects; ¤: chronic cough patients.
09:00
07:00
05:00
03:00
01:00
23:00
19:00
17:00
15:00
13:00
11:00
09:00
21:00
35
30
25
20
15
10
5
0
Coughs·patient-1
Time h
FIGURE 6. 24-h ambulatory, automated cough frequency recordings in 50
chronic cough patients. Data are presented as mean+SEM.
S.S. BIRRING ET AL. LCM: A NOVEL COUGH MONITOR
c
EUROPEAN RESPIRATORY JOURNAL VOLUME 31 NUMBER 5 1017
as indicated by the high sensitivity and particularly high
specificity for detecting cough. The Leicester Cough Monitor
represents a potential advance over existing cough monitors in
that it is portable outside a controlled environment, it does not
require measurement of abdominal electromyography and it
can be set to record for 24 h. The 24-h sound recording is
discarded after the automated analysis process, thereby
ensuring patient privacy. The Leicester Cough Monitor could
be used to validate the presence of excessive cough, to assess
cough severity and to objectively evaluate response to therapy.
Further studies are required in order to assess the use of the
Leicester Cough Monitor in clinical practice.
ACKNOWLEDGEMENTS
The present authors would like to thank the subjects who
participated in the study, C. Long (Dept of Respiratory
Physiology, Glenfield Hospital, Leicester, UK) for assistance
in data analysis, D.D. Vara (Dept of Respiratory Physiology,
Glenfield Hospital) for assistance in the clinical characterisa-
tion of some of the patients and the lung function unit staff at
King’s College Hospital, London, UK (C. Wood, L. Morgan, D.
Cox and D. Land) and Glenfield Hospital.
REFERENCES
1 Irwin RS, Madison JM. The diagnosis and treatment of
cough. N Engl J Med 2000; 343: 1715–1721.
2 Irwin RS. Assessing cough severity and efficacy of therapy
in clinical research: ACCP evidence-based clinical practice
guidelines. Chest 2006; 129: Suppl. 1, 232S–237S.
3 Morice AH, Fontana GA, Sovijarvi AR, et al. The diagnosis
and management of chronic cough. Eur Respir J 2004; 24:
481–492.
4 Birring SS, Prudon B, Carr AJ, Singh SJ, Morgan MD,
Pavord ID. Development of a symptom specific health
status measure for patients with chronic cough: Leicester
Cough Questionnaire (LCQ). Thorax 2003; 58: 339–343.
5 Raj AA, Birring SS. Clinical assessment of chronic cough
severity. Pulm Pharmacol Ther 2007; 20: 334–337.
6 Prudon B, Birring SS, Vara DD, Hall AP, Thompson JP,
Pavord ID. Cough and glottic-stop reflex sensitivity in
health and disease. Chest 2005; 127: 550–557.
7 Birring SS, Matos S, Patel RB, Prudon B, Evans DH,
Pavord ID. Cough frequency, cough sensitivity and health
status in patients with chronic cough. Respir Med 2006; 100:
1105–1109.
8 Matos S, Birring SS, Pavord ID, Evans DH. Detection of
cough signals in continuous audio recordings using
hidden Markov models. IEEE Trans Biomed Eng 2006; 53:
1078–1083.
9 Coyle MA, Keenan DB, Henderson LS, et al. Evaluation of
an ambulatory system for the quantification of cough
frequency in patients with chronic obstructive pulmonary
disease. Cough 2005; 1: 3.
10 Pavesi L, Subburaj S, Porter-Shaw K. Application and
validation of a computerized cough acquisition system for
objective monitoring of acute cough: a meta-analysis. Chest
2001; 120: 1121–1128.
11 Chang AB, Newman RG, Phelan PD, Robertson CF. A new
use for an old Holter monitor: an ambulatory cough meter.
Eur Respir J 1997; 10: 1637–1639.
12 Smith J, Owen E, Earis J, Woodcock A. Effect of codeine on
objective measurement of cough in chronic obstructive
pulmonary disease. J Allergy Clin Immunol 2006; 117: 831–835.
13 Hsu JY, Stone RA, Logan-Sinclair RB, Worsdell M,
Busst CM, Chung KF. Coughing frequency in patients
with persistent cough: assessment using a 24 hour
ambulatory recorder. Eur Respir J 1994; 7: 1246–1253.
14 Barry SJ, Dane AD, Morice AH, Walmsley AD. The
automatic recognition and counting of cough. Cough
2006; 2: 8.
15 Paul IM, Wai K, Jewell SJ, Shaffer ML, Varadan VV.
Evaluation of a new self-contained, ambulatory, objective
cough monitor. Cough 2006; 2: 7.
16 Brightling CE, Ward R, Goh KL, Wardlaw AJ, Pavord ID.
Eosinophilic bronchitis is an important cause of chronic
cough. Am J Respir Crit Care Med 1999; 160: 406–410.
17 Matos S, Birring SS, Pavord ID, Evans DH. An automated
system for 24-hour monitoring of cough frequency: the
Leicester cough monitor. IEEE Trans Biomed Eng 2007; 54:
1472–1479.
18 Smith JA, Earis JE, Woodcock AA. Establishing a gold
standard for manual cough counting: video versus digital
audio recordings. Cough 2006; 2: 6.
LCM: A NOVEL COUGH MONITOR S.S. BIRRING ET AL.
1018 VOLUME 31 NUMBER 5 EUROPEAN RESPIRATORY JOURNAL

More Related Content

What's hot

Ventilator associated pneumonia
Ventilator associated pneumonia Ventilator associated pneumonia
Ventilator associated pneumonia Maher AlQuaimi
 
2021 clinchestmedpneumothoraxclassificationetiology convertido
2021 clinchestmedpneumothoraxclassificationetiology convertido2021 clinchestmedpneumothoraxclassificationetiology convertido
2021 clinchestmedpneumothoraxclassificationetiology convertidoIsaacMoreano1
 
SARI CRITICAL CARE TRAINING CLINICAL SYNDROMES
SARI CRITICAL CARE TRAINING CLINICAL SYNDROMESSARI CRITICAL CARE TRAINING CLINICAL SYNDROMES
SARI CRITICAL CARE TRAINING CLINICAL SYNDROMESSandro Zorzi
 
Antibiotic Strategy in Lower Respiratory Tract Infections (part 1)
Antibiotic Strategy in Lower Respiratory Tract Infections (part 1)Antibiotic Strategy in Lower Respiratory Tract Infections (part 1)
Antibiotic Strategy in Lower Respiratory Tract Infections (part 1)Gamal Agmy
 
Mers co v - journal reading
Mers co v - journal readingMers co v - journal reading
Mers co v - journal readingSoroy Lardo
 
Bundles to prevent ventilator associated pneumonia
Bundles to prevent ventilator associated pneumoniaBundles to prevent ventilator associated pneumonia
Bundles to prevent ventilator associated pneumoniapravin2k2
 
Antibiotic Strategy in Lower Respiratory Tract Infections (part 2)
Antibiotic Strategy in Lower Respiratory Tract Infections (part 2)Antibiotic Strategy in Lower Respiratory Tract Infections (part 2)
Antibiotic Strategy in Lower Respiratory Tract Infections (part 2)Gamal Agmy
 
Ats guidelines for cap 2019
Ats guidelines for cap 2019Ats guidelines for cap 2019
Ats guidelines for cap 2019santoshbhskr
 
Radiological Presentation of COVID 19
Radiological Presentation of COVID 19Radiological Presentation of COVID 19
Radiological Presentation of COVID 19Gamal Agmy
 
Renin - Angiotensin - Aldosterone System Inhibitors in Patients with Covid-19
Renin - Angiotensin - Aldosterone System Inhibitors in Patients with Covid-19Renin - Angiotensin - Aldosterone System Inhibitors in Patients with Covid-19
Renin - Angiotensin - Aldosterone System Inhibitors in Patients with Covid-19Valentina Corona
 
Pulmonary involvement in peoples living with HIV
Pulmonary involvement in peoples living with HIVPulmonary involvement in peoples living with HIV
Pulmonary involvement in peoples living with HIVChetan Ganteppanavar
 
Ventilator-Associated Event (VAE2)
Ventilator-Associated Event (VAE2)Ventilator-Associated Event (VAE2)
Ventilator-Associated Event (VAE2)Ahmed AlGahtani, RRT
 
Interstitial pulmonary fibrosis
Interstitial pulmonary fibrosisInterstitial pulmonary fibrosis
Interstitial pulmonary fibrosisashish ranjan
 

What's hot (20)

Prone ventilation
Prone ventilationProne ventilation
Prone ventilation
 
Ventilator associated pneumonia
Ventilator associated pneumonia Ventilator associated pneumonia
Ventilator associated pneumonia
 
2021 clinchestmedpneumothoraxclassificationetiology convertido
2021 clinchestmedpneumothoraxclassificationetiology convertido2021 clinchestmedpneumothoraxclassificationetiology convertido
2021 clinchestmedpneumothoraxclassificationetiology convertido
 
SARI CRITICAL CARE TRAINING CLINICAL SYNDROMES
SARI CRITICAL CARE TRAINING CLINICAL SYNDROMESSARI CRITICAL CARE TRAINING CLINICAL SYNDROMES
SARI CRITICAL CARE TRAINING CLINICAL SYNDROMES
 
Prone Positioning in Severe Acute Distress Syndrome - Posição Prona em Indiv...
Prone Positioning in Severe Acute  Distress Syndrome - Posição Prona em Indiv...Prone Positioning in Severe Acute  Distress Syndrome - Posição Prona em Indiv...
Prone Positioning in Severe Acute Distress Syndrome - Posição Prona em Indiv...
 
Antibiotic Strategy in Lower Respiratory Tract Infections (part 1)
Antibiotic Strategy in Lower Respiratory Tract Infections (part 1)Antibiotic Strategy in Lower Respiratory Tract Infections (part 1)
Antibiotic Strategy in Lower Respiratory Tract Infections (part 1)
 
H1N1
H1N1 H1N1
H1N1
 
Mers co v - journal reading
Mers co v - journal readingMers co v - journal reading
Mers co v - journal reading
 
Hospital pneumonia
Hospital pneumoniaHospital pneumonia
Hospital pneumonia
 
Bundles to prevent ventilator associated pneumonia
Bundles to prevent ventilator associated pneumoniaBundles to prevent ventilator associated pneumonia
Bundles to prevent ventilator associated pneumonia
 
Antibiotic Strategy in Lower Respiratory Tract Infections (part 2)
Antibiotic Strategy in Lower Respiratory Tract Infections (part 2)Antibiotic Strategy in Lower Respiratory Tract Infections (part 2)
Antibiotic Strategy in Lower Respiratory Tract Infections (part 2)
 
NEUMONIA ASOCIADA AL VENTILADOR GUIA IDSA 2016
NEUMONIA ASOCIADA AL VENTILADOR GUIA IDSA 2016NEUMONIA ASOCIADA AL VENTILADOR GUIA IDSA 2016
NEUMONIA ASOCIADA AL VENTILADOR GUIA IDSA 2016
 
Ats guidelines for cap 2019
Ats guidelines for cap 2019Ats guidelines for cap 2019
Ats guidelines for cap 2019
 
Radiological Presentation of COVID 19
Radiological Presentation of COVID 19Radiological Presentation of COVID 19
Radiological Presentation of COVID 19
 
Renin - Angiotensin - Aldosterone System Inhibitors in Patients with Covid-19
Renin - Angiotensin - Aldosterone System Inhibitors in Patients with Covid-19Renin - Angiotensin - Aldosterone System Inhibitors in Patients with Covid-19
Renin - Angiotensin - Aldosterone System Inhibitors in Patients with Covid-19
 
Pulmonary involvement in peoples living with HIV
Pulmonary involvement in peoples living with HIVPulmonary involvement in peoples living with HIV
Pulmonary involvement in peoples living with HIV
 
Ventilator-Associated Event (VAE2)
Ventilator-Associated Event (VAE2)Ventilator-Associated Event (VAE2)
Ventilator-Associated Event (VAE2)
 
EOlife X
EOlife XEOlife X
EOlife X
 
Vap prevention 2014 ppt
Vap prevention 2014 pptVap prevention 2014 ppt
Vap prevention 2014 ppt
 
Interstitial pulmonary fibrosis
Interstitial pulmonary fibrosisInterstitial pulmonary fibrosis
Interstitial pulmonary fibrosis
 

Similar to The leicester cough monitor preliminary

Evaluation of lung ultrasound for the diagnosis of pneumonia in the ED
Evaluation of lung ultrasound for the diagnosis of pneumonia in the EDEvaluation of lung ultrasound for the diagnosis of pneumonia in the ED
Evaluation of lung ultrasound for the diagnosis of pneumonia in the EDMario Robusti
 
Management Of Community Acquired Pneumonia
Management  Of Community Acquired PneumoniaManagement  Of Community Acquired Pneumonia
Management Of Community Acquired PneumoniaAshraf ElAdawy
 
Frequency and management of respiratory incidents in invasive home ventilation
Frequency and management of respiratory incidents in invasive home ventilationFrequency and management of respiratory incidents in invasive home ventilation
Frequency and management of respiratory incidents in invasive home ventilationMissing Man
 
PATTERN OF HOSPITAL-ACQUIRED PNEUMONIA IN INTENSIVE CARE UNIT OF SUEZ CANAL U...
PATTERN OF HOSPITAL-ACQUIRED PNEUMONIA IN INTENSIVE CARE UNIT OF SUEZ CANAL U...PATTERN OF HOSPITAL-ACQUIRED PNEUMONIA IN INTENSIVE CARE UNIT OF SUEZ CANAL U...
PATTERN OF HOSPITAL-ACQUIRED PNEUMONIA IN INTENSIVE CARE UNIT OF SUEZ CANAL U...Khaled Mohamed
 
Plan presentation on copd ………………………………….
Plan presentation on copd ………………………………….Plan presentation on copd ………………………………….
Plan presentation on copd ………………………………….Roop
 
Study of clinical and etiological profile of community acquired pneumonia in ...
Study of clinical and etiological profile of community acquired pneumonia in ...Study of clinical and etiological profile of community acquired pneumonia in ...
Study of clinical and etiological profile of community acquired pneumonia in ...iosrjce
 
Community Acquired Pneumonia
Community Acquired PneumoniaCommunity Acquired Pneumonia
Community Acquired PneumoniaAshraf ElAdawy
 
Patient ventilator interactions during mechanical ventilation
Patient ventilator interactions during mechanical ventilationPatient ventilator interactions during mechanical ventilation
Patient ventilator interactions during mechanical ventilationDr.Mahmoud Abbas
 
Noninvasive Tests for Asthma Diagnosis
Noninvasive Tests for Asthma DiagnosisNoninvasive Tests for Asthma Diagnosis
Noninvasive Tests for Asthma DiagnosisAli Taki
 
Arterial CO2 Tension on Admission as a marker of in-hospital mortality
Arterial CO2 Tension on Admission as a marker of in-hospital mortality Arterial CO2 Tension on Admission as a marker of in-hospital mortality
Arterial CO2 Tension on Admission as a marker of in-hospital mortality GerardJamero1
 
Diagnosis of infection pleural effussion by reagen strips
Diagnosis of infection pleural effussion by reagen stripsDiagnosis of infection pleural effussion by reagen strips
Diagnosis of infection pleural effussion by reagen stripsMarcelo Alvarez
 
Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...
Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...
Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...Hauke Sann
 
Bacterial flora in sputum and antibiotic sensitivity in exacerbations of bron...
Bacterial flora in sputum and antibiotic sensitivity in exacerbations of bron...Bacterial flora in sputum and antibiotic sensitivity in exacerbations of bron...
Bacterial flora in sputum and antibiotic sensitivity in exacerbations of bron...Dr.Aslam calicut
 
Uniportal video assisted thoracoscopic bronchial sleeve lobectomy in five pat...
Uniportal video assisted thoracoscopic bronchial sleeve lobectomy in five pat...Uniportal video assisted thoracoscopic bronchial sleeve lobectomy in five pat...
Uniportal video assisted thoracoscopic bronchial sleeve lobectomy in five pat...Clinical Surgery Research Communications
 
A study on “impact of artificial intelligence in covid19 diagnosis”
A study on “impact of artificial intelligence in covid19 diagnosis”A study on “impact of artificial intelligence in covid19 diagnosis”
A study on “impact of artificial intelligence in covid19 diagnosis”Dr. C.V. Suresh Babu
 
Use of Capnograph in Breathlessness Patients
Use of Capnograph in Breathlessness PatientsUse of Capnograph in Breathlessness Patients
Use of Capnograph in Breathlessness Patientsnhliza
 

Similar to The leicester cough monitor preliminary (20)

Evaluation of lung ultrasound for the diagnosis of pneumonia in the ED
Evaluation of lung ultrasound for the diagnosis of pneumonia in the EDEvaluation of lung ultrasound for the diagnosis of pneumonia in the ED
Evaluation of lung ultrasound for the diagnosis of pneumonia in the ED
 
Management Of Community Acquired Pneumonia
Management  Of Community Acquired PneumoniaManagement  Of Community Acquired Pneumonia
Management Of Community Acquired Pneumonia
 
Frequency and management of respiratory incidents in invasive home ventilation
Frequency and management of respiratory incidents in invasive home ventilationFrequency and management of respiratory incidents in invasive home ventilation
Frequency and management of respiratory incidents in invasive home ventilation
 
XpertTBCExtrapulm.pdf
XpertTBCExtrapulm.pdfXpertTBCExtrapulm.pdf
XpertTBCExtrapulm.pdf
 
PATTERN OF HOSPITAL-ACQUIRED PNEUMONIA IN INTENSIVE CARE UNIT OF SUEZ CANAL U...
PATTERN OF HOSPITAL-ACQUIRED PNEUMONIA IN INTENSIVE CARE UNIT OF SUEZ CANAL U...PATTERN OF HOSPITAL-ACQUIRED PNEUMONIA IN INTENSIVE CARE UNIT OF SUEZ CANAL U...
PATTERN OF HOSPITAL-ACQUIRED PNEUMONIA IN INTENSIVE CARE UNIT OF SUEZ CANAL U...
 
Plan presentation on copd ………………………………….
Plan presentation on copd ………………………………….Plan presentation on copd ………………………………….
Plan presentation on copd ………………………………….
 
Study of clinical and etiological profile of community acquired pneumonia in ...
Study of clinical and etiological profile of community acquired pneumonia in ...Study of clinical and etiological profile of community acquired pneumonia in ...
Study of clinical and etiological profile of community acquired pneumonia in ...
 
Community Acquired Pneumonia
Community Acquired PneumoniaCommunity Acquired Pneumonia
Community Acquired Pneumonia
 
Patient ventilator interactions during mechanical ventilation
Patient ventilator interactions during mechanical ventilationPatient ventilator interactions during mechanical ventilation
Patient ventilator interactions during mechanical ventilation
 
Noninvasive Tests for Asthma Diagnosis
Noninvasive Tests for Asthma DiagnosisNoninvasive Tests for Asthma Diagnosis
Noninvasive Tests for Asthma Diagnosis
 
Arterial CO2 Tension on Admission as a marker of in-hospital mortality
Arterial CO2 Tension on Admission as a marker of in-hospital mortality Arterial CO2 Tension on Admission as a marker of in-hospital mortality
Arterial CO2 Tension on Admission as a marker of in-hospital mortality
 
Weaning Tips & Tricks
Weaning Tips & TricksWeaning Tips & Tricks
Weaning Tips & Tricks
 
Diagnosis of infection pleural effussion by reagen strips
Diagnosis of infection pleural effussion by reagen stripsDiagnosis of infection pleural effussion by reagen strips
Diagnosis of infection pleural effussion by reagen strips
 
Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...
Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...
Roka P.L. et al.: Breath-by-breath regional expiratory time constants by elec...
 
Inda glyco
Inda glycoInda glyco
Inda glyco
 
Bacterial flora in sputum and antibiotic sensitivity in exacerbations of bron...
Bacterial flora in sputum and antibiotic sensitivity in exacerbations of bron...Bacterial flora in sputum and antibiotic sensitivity in exacerbations of bron...
Bacterial flora in sputum and antibiotic sensitivity in exacerbations of bron...
 
Tuberculosis diagnostics
Tuberculosis diagnosticsTuberculosis diagnostics
Tuberculosis diagnostics
 
Uniportal video assisted thoracoscopic bronchial sleeve lobectomy in five pat...
Uniportal video assisted thoracoscopic bronchial sleeve lobectomy in five pat...Uniportal video assisted thoracoscopic bronchial sleeve lobectomy in five pat...
Uniportal video assisted thoracoscopic bronchial sleeve lobectomy in five pat...
 
A study on “impact of artificial intelligence in covid19 diagnosis”
A study on “impact of artificial intelligence in covid19 diagnosis”A study on “impact of artificial intelligence in covid19 diagnosis”
A study on “impact of artificial intelligence in covid19 diagnosis”
 
Use of Capnograph in Breathlessness Patients
Use of Capnograph in Breathlessness PatientsUse of Capnograph in Breathlessness Patients
Use of Capnograph in Breathlessness Patients
 

Recently uploaded

Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...EduSkills OECD
 
Privatization and Disinvestment - Meaning, Objectives, Advantages and Disadva...
Privatization and Disinvestment - Meaning, Objectives, Advantages and Disadva...Privatization and Disinvestment - Meaning, Objectives, Advantages and Disadva...
Privatization and Disinvestment - Meaning, Objectives, Advantages and Disadva...RKavithamani
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeThiyagu K
 
The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13Steve Thomason
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptxVS Mahajan Coaching Centre
 
Separation of Lanthanides/ Lanthanides and Actinides
Separation of Lanthanides/ Lanthanides and ActinidesSeparation of Lanthanides/ Lanthanides and Actinides
Separation of Lanthanides/ Lanthanides and ActinidesFatimaKhan178732
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactdawncurless
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Sapana Sha
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104misteraugie
 
Hybridoma Technology ( Production , Purification , and Application )
Hybridoma Technology  ( Production , Purification , and Application  ) Hybridoma Technology  ( Production , Purification , and Application  )
Hybridoma Technology ( Production , Purification , and Application ) Sakshi Ghasle
 
Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationnomboosow
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)eniolaolutunde
 
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdfssuser54595a
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxGaneshChakor2
 
mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docxPoojaSen20
 
URLs and Routing in the Odoo 17 Website App
URLs and Routing in the Odoo 17 Website AppURLs and Routing in the Odoo 17 Website App
URLs and Routing in the Odoo 17 Website AppCeline George
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxheathfieldcps1
 
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptxContemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptxRoyAbrique
 

Recently uploaded (20)

Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
 
Privatization and Disinvestment - Meaning, Objectives, Advantages and Disadva...
Privatization and Disinvestment - Meaning, Objectives, Advantages and Disadva...Privatization and Disinvestment - Meaning, Objectives, Advantages and Disadva...
Privatization and Disinvestment - Meaning, Objectives, Advantages and Disadva...
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and Mode
 
The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
 
Separation of Lanthanides/ Lanthanides and Actinides
Separation of Lanthanides/ Lanthanides and ActinidesSeparation of Lanthanides/ Lanthanides and Actinides
Separation of Lanthanides/ Lanthanides and Actinides
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impact
 
TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdfTataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104
 
Hybridoma Technology ( Production , Purification , and Application )
Hybridoma Technology  ( Production , Purification , and Application  ) Hybridoma Technology  ( Production , Purification , and Application  )
Hybridoma Technology ( Production , Purification , and Application )
 
Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communication
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)
 
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptx
 
mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docx
 
Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1
 
URLs and Routing in the Odoo 17 Website App
URLs and Routing in the Odoo 17 Website AppURLs and Routing in the Odoo 17 Website App
URLs and Routing in the Odoo 17 Website App
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
 
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptxContemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
 

The leicester cough monitor preliminary

  • 1. The Leicester Cough Monitor: preliminary validation of an automated cough detection system in chronic cough S.S. Birring*, T. Fleming*, S. Matos# , A.A. Raj" , D.H. Evans# and I.D. Pavord" ABSTRACT: Chronic cough is a common condition that presents to both primary and secondary care. Assessment and management are hampered by the absence of well-validated outcome measures. The present study comprises the validation of the Leicester Cough Monitor (LCM), an automated sound-based ambulatory cough monitor. Cough frequency was measured with the LCM and compared with coughs and other sounds counted manually over 2 h of a 6-h recording by two observers in nine patients with chronic cough in order to determine the sensitivity and specificity of the LCM. Automated cough frequency was also compared with manual counts from one observer in 15 patients with chronic cough and eight healthy subjects. All subjects underwent 6-h recordings. A subgroup consisting of six control and five patients with stable chronic cough underwent repeat automated measurements o3 months apart. A further 50 patients with chronic cough underwent 24-h automated cough monitoring. The LCM had a sensitivity and specificity of 91 and 99%, respectively, for detecting cough and a false-positive rate of 2.5 events?h-1 . Mean¡SEM automated cough counts?patient?h-1 was 48¡9 in patients with chronic cough and 2¡1 in the control group (mean difference 46 counts?patient?h-1 ; 95% confidence interval (CI) 20–71). The automated cough counts were repeatable (intra-subject SD 11.4 coughs?patient?h-1 ; intra-class correlation coefficient 0.9). The cough frequency in patients undergoing 24-h automated monitoring was 19 coughs?patient?h-1 ; daytime (08:00– 22:00 h) cough frequency was significantly greater than overnight cough frequency (25 versus 10 coughs?patient?h-1 ; mean difference 15 coughs?patient?h-1 , 95% CI 8–22). The Leicester Cough Monitor is a valid and reliable tool that can be used to assess 24-h cough frequency in patients with cough. It should be a useful tool to assess patients with cough in clinical trials and longitudinal studies. KEYWORDS: Chronic cough, cough counts, cough frequency, cough monitor, Leicester Cough Monitor C hronic cough is a common reason for referral to respiratory physicians. The assessment of patients with chronic cough is often based on the anatomical diagnostic protocol, which is a systematic evaluation based on the understanding that most cases are due to disease of the upper respiratory tract where cough receptors are most plentiful [1]. The most common conditions implicated in causing chronic cough in nonsmokers are asthma, gastro- oesophageal reflux and rhinitis, or a combination of these [2]. The identification of an important contribution by the aforementioned conditions is largely based on the evaluation of treatment trials [3]. However, there are few well-validated outcome measures to assess cough severity and treatment efficacy. Cough visual analogue scores, diary score cards, quality-of-life questionnaires, cough- reflex sensitivity measurement and cough moni- tors have been proposed as potential tools to assess cough severity [2]. The subjective nature of symptom scores and quality-of-life question- naires [4, 5] and the poor specificity of cough- reflex sensitivity measurement [6] to identify patients with chronic cough have led to a renewed interest in the development of auto- mated ambulatory cough monitors [7–15]. BIRRING et al. [7] have previously shown that there are marked differences in cough frequency between patients with chronic cough and healthy subjects and that these measurements are repea- table, and have suggested that cough frequency AFFILIATIONS *Dept of Respiratory Medicine, King’s College Hospital, London, # Dept of Medical Physics, Leicester Royal Infirmary, and " Institute for Lung Health, Dept of Respiratory Medicine, Glenfield Hospital, Leicester, UK. CORRESPONDENCE I.D. Pavord Dept of Respiratory Medicine Glenfield Hospital Leicester LE3 9QP UK Fax: 44 1162367768 E-mail: ian.pavord@uhl-tr.nhs.uk Received: May 10 2007 Accepted after revision: December 16 2007 STATEMENT OF INTEREST None declared European Respiratory Journal Print ISSN 0903-1936 Online ISSN 1399-3003 EUROPEAN RESPIRATORY JOURNAL VOLUME 31 NUMBER 5 1013 Eur Respir J 2008; 31: 1013–1018 DOI: 10.1183/09031936.00057407 CopyrightßERS Journals Ltd 2008 c
  • 2. measurement is potentially useful in the assessment of patients with chronic cough. Currently available cough monitors are limited by difficulty in achieving unrestricted ambulatory measurement in the patients’ own environment, an inability to perform 24-h recording and a lack of automated cough detection systems. The aim of the present study was to develop an automated cough monitor (the Leicester Cough Monitor (LCM)) capable of recording cough sounds for 24 h. The present study shows the validation of the LCM and preliminary findings of 6- and 24-h recordings in patients with chronic cough. METHODS Subjects A total of 15 consecutive patients with an isolated chronic cough (.3 weeks duration) were recruited from a specialised cough clinic. The clinic receives referrals from primary and secondary care largely confined to a population of 970,000 within Leicestershire, UK. The causes of cough in patients with chronic cough were: cough variant asthma (n54); gastro- oesophageal reflux (n53); eosinophilic bronchitis (n51); idiopathic (n53); post-viral (n51); bronchiectasis (n51); chronic obstructive pulmonary disease (n51); and chronic bronchitis (n51). Nine out of these patients were randomly selected for the first stage of validation (cough variant asthma (n54), eosinophilic bronchitis (n51), idiopathic (n52), post- viral (n51), bronchiectasis (n51)) and all patients were included in the second validation stage. Investigations were carried out according to a standardised algorithm [16]. The protocol for investigation and treatment and the criteria for accepting diagnosis were as previously described [16]. Eight controls were recruited from healthy volunteers responding to local advertising. Control subjects were asymptomatic, non- smokers and had normal spirometry and a concentration of methacholine required to cause a 20% decrease in forced expiratory volume in one second (FEV1) of .16mg?mL-1 . No patients had received corticosteroids or other specific treat- ment for the condition causing cough for o6 weeks prior to the study. A randomly selected subgroup of six control subjects and five patients with a stable chronic cough and stable treatment requirements (three with cough variant asthma, one with gastro-oesophageal reflux-associated cough and one with idiopathic chronic cough) participated in cough frequency repeatability studies 3–6 months after the first, at the same time of day in order to avoid possible bias from diurnal variations. A total of 50 further consecutive patients with chronic cough were recruited in order to evaluate 24-h recordings with the LCM (idiopathic cough (n526), asthma (n58), eosinophilic bronchitis (n52), rhinitis (n52), sarcoidosis (n52), gastro-oesophageal reflux (n53), bronchiectasis (n52), chronic obstructive pulmonary disease (n52), enlarged tonsils (n52) and obstructive sleep apnoea (n51)). All subjects gave full informed written consent to participate. The protocol for the present study was approved by the Leicestershire Research Ethics Committee. Cough monitor The LCM (fig. 1) is a digital ambulatory cough monitor that records sound continuously from a free-field microphone necklace (Sennheiser MKE 2-5; Sennheiser electronic GmbH & Co. KG, Woedemark, Germany) onto a digital sound recorder (dimensions 26.7687632 mm; iRiver iFP-799; iRiver Europe GmbH, Eschborn, Germany) at a sampling frequency of 16 kHz and with an encoding bit rate of 64 kbit?s-1 . The cough monitor was attached at 09:00 h in all subjects and returned 6– 24 h later. Subjects were told that the LCM was a new investigative tool being developed to assess the nature of the cough and were encouraged to resume their normal activity in their usual environment. When the recording was complete, data stored on the recorder was downloaded onto a computer, where it was analysed by an automated cough detection algorithm (the Leicester Cough Algorithm). A general outline of the Leicester Cough Algorithm has been described previously [8, 17]. Briefly, the detection algorithm is based on Hidden Markov Models, a statistical method that can be used to characterise the spectral properties of a time- varying pattern. The cough detection algorithm was imple- mented based on the keyword-spotting approach, as defined in speech recognition, in which the objective is to detect the occurrence of a particular set of keywords in a sequence of continuous speech. Continuous ambulatory recordings in patients with chronic cough were used to train statistical models of the characteristics of cough sounds and of the audio background. During the detection process, the recorded audio signal was divided into contiguous 10-s segments to be analysed by the Hidden Markov Models-based algorithm. Each 10-s audio segment was recognised by the detection algorithm as a sequence of variable-length audio sections, each in turn classified either as background audio or as a possible cough sound, depending on its statistics [8]. A second algorithm phase then used brief operator input to facilitate the automated algorithm in order to eliminate sounds that Microphone Recorder FIGURE 1. The Leicester Cough Monitor. LCM: A NOVEL COUGH MONITOR S.S. BIRRING ET AL. 1014 VOLUME 31 NUMBER 5 EUROPEAN RESPIRATORY JOURNAL
  • 3. might have been wrongly classified as cough events in the first phase. For this, the operator is asked to classify, as cough or otherwise, a small fraction of the sounds detected in the first phase as possible cough sounds (the second phase takes 5 min to carry out for a 24-h recording and ,50 sounds are classified). The information is then used to create statistical models that are adapted to the characteristics of the cough sounds for that particular recording and the remaining sounds that were not shown to the operator are classified using these models. Cough was defined as a characteristic explosive sound. The algorithm identifies coughs as single events whether they occur as isolated events or in a cluster (i.e. attempts were made to determine how many coughs occurred in paroxysms). Validation Stage 1 The first stage of validation compared automated cough counts against those identified by manual sound analysis of the first and fourth recorded hours of nine randomly selected patients with chronic cough. Manual analysis of sound recordings consisted of three blinded observer counts (observer one twice and observer two once) and cough or noncough sounds were positively identified when all three counts were in agreement based on sound and visual inspection of the acoustic trace. Each cough sound was identified separately, whether it occured singularly or in a cluster or epoch of coughs. Intra- and interobserver variability in cough counts was established from the two blinded analyses performed by observer one and by comparing the mean of observer one’s counts for these periods with counts obtained by observer two. The recordings were analysed twice using the Leicester Cough Algorithm to assess intra-recording repeatability. In order to classify non- cough sounds and determine whether particular sounds were wrongly classified by the automated system, observer one listened to all 6 h of the nine patients’ recordings and classified all recognisable sounds. The results of this analysis were compared with the automated classification. Stage 2 The second stage of validation was extended to all recordings and compared automated cough counts against coughs identified manually by observer one, who analysed the entire 6-h recording. A further 50 patients with chronic cough underwent 24-h automated cough frequency measurement. Automated cough frequency was compared for repeatability studies. Analysis Subject characteristics were described using descriptive statis- tics and expressed as mean¡SEM for parametric data and median for non-parametric data. Cough frequency was expressed as individual coughs?patient?h-1 for the duration of the recording. The validity of the LCM was presented as sensitivity, specificity and false-positive rate of the automated algorithm for detecting coughs as measured by observer manual analysis. Intra- and interobserver variability of manual cough counts and repeatability data was assessed as intra-class correlation coefficients and intra-subject SD. RESULTS The subject characteristics are shown in table 1. Validation stage 1 (First and fourth recorded hour) Mean cough counts were 39 coughs?patient-1 ?h-1 by automated analysis compared with 43 coughs?patient-1 ?h-1 by manual analysis (mean difference -4 coughs?patient-1 ?h-1 , 95% confi- dence interval (CI) -6–13; p50.4). The intra- and interobserver intra-class correlation coefficients for manual analysis of sound recordings (between observers one and two) were 0.99 and 0.98, respectively (both p,0.001). The intra- and interobserver (i.e. between mean of observer one and observer two) intra- class correlation coefficients for manual analysis of sound recordings were 0.99 and 0.98, respectively (both p,0.001). The intra-class correlation coefficient between automated and manual observer counts was 0.9 (p,0.001; fig. 2a). The accuracy of manual and automated cough counts appeared similar in recordings containing paroxysms and those with isolated coughs. The Leicester Cough Algorithm had a sensitivity and a specificity of 91% and 99%, respectively, for detecting cough sounds and a median false positive rate of 2.5 events?patient-1 ?h-1 against the gold standard of coughs detected manually by observer one twice and observer two once. There was no evidence that any particular sound was more likely to be classified as a false positive (fig. 3). Validation stage 2 (6-h recordings) Mean¡SEM automated cough counts were 48¡9 coughs? patient-1 ?h-1 in patients with chronic cough and 2¡1 coughs? patient-1 ?h-1 in control subjects (mean difference 46 coughs? patient-1 ?h-1 , 95% CI 20–71 coughs?patient-1 ?h-1 ; p,0.001; fig. 4). There were no significant differences in cough frequency between diagnostic groups. The cough analysis took 2 h to complete, comprising 5 min for data download, 105 min for computer automated analysis (an operator was not required to be present during phase 1) and 10 min for operator input (phase 2) and printing results. The intra-class correlation coefficient between automated and observer counts was 0.93 (p,0.001; fig. 2b). The LCM had sensitivity and specificity of 86 and 99%, respectively, for detecting cough sounds and a median false-positive rate of 1.0 events?patient-1 ?h-1 . The automated cough counts were repeatable in the 11 subjects who underwent repeatability testing (intra-subject SD 11.4 coughs? patient-1 ?h-1 , intra-class correlation coefficient 0.9; fig. 5). TABLE 1 Subject characteristics Control Chronic cough Subjects (male) n 8 (0) 15 (5) Age yrs 48¡3 55¡4 Cough duration yrs 5¡2 FEV1 % pred 91¡5 89¡7 FEV1/FVC % 79¡1 75¡4 Data are expressed as mean¡SEM, unless otherwise stated; FEV1: forced expiratory volume in one second; pred: predicted; FVC: forced vital capacity. S.S. BIRRING ET AL. LCM: A NOVEL COUGH MONITOR c EUROPEAN RESPIRATORY JOURNAL VOLUME 31 NUMBER 5 1015
  • 4. The cough frequency in patients undergoing 24-h monitoring was 19 coughs?patient-1 ?h-1 ; daytime (08:00–22:00 h) cough frequency was significantly greater than overnight cough frequency (25 versus 10 coughs?patient-1 ?h-1 ; mean difference 15 coughs?pa- tient-1 ?h-1 , 95% CI 8–22 coughs?patient-1 ?h-1 ; p,0.001; fig. 6). DISCUSSION The LCM is a lightweight 24-h automated ambulatory cough monitor that is easy to use and measures cough in the subjects’ own environment. The present study has shown that it is a valid and reliable tool for objectively measuring cough frequency. The high sensitivity and specificity for the detection of cough sounds is comparable to other routine diagnostic clinical tools and superior to that reported for other more cumbersome cough detection systems. Preliminary data is presented in the current study showing that the cough frequency measured with the LCM is repeatable over o3 months, a period relevant to the duration of treatment trials that form an important part of the assessment of patients with chronic cough. Repeatability was marginally better than that of recordings analysed manually [7]. The present data suggest that the LCM might be a particularly useful outcome measure in assessing patients with cough and measuring the response to therapy in the clinic and in clinical trials. A limitation of the present study is that evaluation of cough frequency was based on 6-h daytime cough recordings owing u u u u u u u uuu uuu u u u u u 80 40 0 -40 -80 u u u u u uuu uu u u u u 80 40 0 -40 -80 u u uuuuuu a) b) 0 50 100 150 Differencebetweenautomatedand observercoughcounts·patient-1·h-1 Mean of automated and observer coughs·patient-1·h-1 Differencebetweenautomatedand observercoughcounts·patient-1·h-1 FIGURE 2. Bland-Altman plot of automated versus manual observer cough counts?patient-1 ?h-1 . a) Validation stage 1 at which first and fourth recorded hour (n59) were analysed. Each hour is depicted individually. b) Validation stage 2 at which 6-h recordings (n523) were analysed in their entirety. The complete cough detection algorithm (phases 1 and 2) was tested in each validation stage. –––: mean difference; ??????: 95% limits of agreement (26 intra-subject SD); e: control subjects; ¤: chronic cough patients. 31726 sounds: 1834 cough 29892 noncough Phase 1 2543 accepted sounds: 1573 cough 970 noncough# Phase 2 1680 accepted sounds: 1528 cough 152 noncough+ 29183 rejected sounds: 261 cough 28922 noncough 863 rejected sounds: 45 cough 818 noncough¶ FIGURE 3. False positives characterised manually that were detected by the automated Leicester Cough Monitor in nine 6-h recordings of patients with chronic cough. Automated analysis of sound recording is performed during phase 1 and automated analysis following operator input is performed during phase 2. The sensitivity for cough detection is slightly lower than that from the gold standard validation stage 1, since the comparator was a 6-h manual counting by observer one only. # : noncough sounds were: speech (n5351), impulsive noise (n5222) throat clearing (n5119), environmental noise (n5111), laugh (n580), other person coughing (n531), incomplete coughs (n522), child talking/shouting (n521), sneeze (n56), telephone ringing (n53), burp (n52) and dog barking (n52). " : noncough sounds were: speech (n5293), impulsive noise (n5214), throat clearing (n596), environmental noise (n587), laugh (n567), other person coughing (n515), incomplete coughs (n518), child talking/shouting (n517), sneeze (n55), telephone ringing (n53), burp (n52) and dog barking (n51). + : noncough sounds were: speech (n558), impulsive noise (n58), throat clearing (n523), environmental noise (n524), laugh (n513), other person coughing (n516), incomplete coughs (n54), child talking/shouting (n54), sneeze (n51) and dog barking (n51). Chronic cough patients 150 125 100 75 50 25 0 Control subjects Coughs·patient-1·h-1 uu uuu u u u u u u uu u u u u uu uuuu *** FIGURE 4. Mean¡SEM automated cough counts?patient-1 ?h-1 in control subjects and chronic cough patients (6-h recordings). ***: p,0.001. LCM: A NOVEL COUGH MONITOR S.S. BIRRING ET AL. 1016 VOLUME 31 NUMBER 5 EUROPEAN RESPIRATORY JOURNAL
  • 5. to limited battery life (6–8 h) at the inception of the study. Advances in battery technology, since then, have allowed the extension of recordings to o24 h. The automated system has allowed these recordings to be analysed relatively quickly and accurately and should facilitate the investigation of potential diurnal variations in cough frequency and the effects of aggravators of potential cough, such as environmental pollu- tion, cigarette smoking and gastro-oesophageal reflux. In the present study, a range of sounds including speech, throat clearing and environmental noises caused false-positive detected coughs. There was little evidence that any of these sounds caused particular difficulties with detection, nor did cough paroxysms appear to present problems for accurate manual and automated cough counts. However, greater experience with the monitor may identify sounds or cough paroxysms that present particular problems for the algorithm to classify and allow further refinement of the algorithm. The present study involved small numbers of subjects and it will be important to study a larger population of control subjects and patients with well-defined respiratory disease, before and after treatment, and in different environments, to fully validate the cough monitor. The present preliminary work suggests that such a study will be feasible. Cough frequency was stable through the day and was significantly reduced overnight compared with daytimes in accord with previous data, suggesting a diurnal variation in cough frequency [5, 9, 13]. Further work is required to determine the validity and the short- and longer-term repeatability of 24-h cough recordings. A limitation of the present study is that only 2 h per recording were used to compare automated cough counts with those obtained from manual counting for the validation study. Manual cough counting is very time consuming and laborious so only the first and fourth hours of each recording were manually analysed for consistency. Each recording was manually counted three times in order to obtain a more robust measure of the true cough frequency. The LCM had a high sensitivity and specificity for detecting cough against this gold standard. This was confirmed in the second part of the validation study where cough counts derived from automated analysis of 6-h recordings were compared with cough counts derived from a single manual observer. The sensitivity of the cough algorithm was slightly lower with 6-h recordings compared with 2-h recordings. This is most probably because of the more robust measure of true cough frequency used for the 2-h recordings, compared with single observer manual counts used for the 6-h recordings. A potential criticism of cough counts derived from audio recordings is that they might not accurately reflect the true cough rate since it is not possible to visualise the act of coughing. However, a recent study [18] compared manual cough counts from audio with video recordings and found them to be very similar. That study concluded that manual cough counts from audio recordings should be regarded as the gold standard to validate cough monitors since audio record- ings had superior sound quality to that from video recordings. The LCM quantifies cough frequency as single episodes of cough rather than epochs or clusters of coughs and cough seconds (seconds containing cough), as used by others [12]. The present authors believe that single cough episodes are a more meaningful measure and are easier to interpret by physicians and patients. Cough frequency, rather than inten- sity, was measured since cough events are less influenced by microphone position and muffling of sounds by covering the mouth during the act of coughing. Furthermore, cough intensity determined with sound analysis lacks responsiveness compared with cough frequency in clinical trials of antitussive drugs [10]. Cough intensity determined by other parameters such as airflow or chest wall movement is less practical for routine clinical measurement. The LCM was validated in subjects with chronic cough due to a wide range of conditions so it reliably detects coughs with differing characteristics. One of the challenges of developing cough monitors in the past has been differentiating cough sounds from throat clearing, sneezing, speech and other cough-like sounds. The Leicester Cough Monitor differentiates cough from other sounds reliably 80 60 40 20 0 -20 -40 -60 -80 0 10 20 30 40 50 u u u u uuu u u u Differencebetweentests coughs·patient-1·h-1 Mean of two tests coughs·patient-1·h-1 FIGURE 5. Bland-Altman plot of automated cough counts?patient-1 ?h-1 repeated over 3–6 months in chronic cough patients and control subjects. ––––: mean difference; ???????: 95% limits of agreement (26 intra-subject SD). e: control subjects; ¤: chronic cough patients. 09:00 07:00 05:00 03:00 01:00 23:00 19:00 17:00 15:00 13:00 11:00 09:00 21:00 35 30 25 20 15 10 5 0 Coughs·patient-1 Time h FIGURE 6. 24-h ambulatory, automated cough frequency recordings in 50 chronic cough patients. Data are presented as mean+SEM. S.S. BIRRING ET AL. LCM: A NOVEL COUGH MONITOR c EUROPEAN RESPIRATORY JOURNAL VOLUME 31 NUMBER 5 1017
  • 6. as indicated by the high sensitivity and particularly high specificity for detecting cough. The Leicester Cough Monitor represents a potential advance over existing cough monitors in that it is portable outside a controlled environment, it does not require measurement of abdominal electromyography and it can be set to record for 24 h. The 24-h sound recording is discarded after the automated analysis process, thereby ensuring patient privacy. The Leicester Cough Monitor could be used to validate the presence of excessive cough, to assess cough severity and to objectively evaluate response to therapy. Further studies are required in order to assess the use of the Leicester Cough Monitor in clinical practice. ACKNOWLEDGEMENTS The present authors would like to thank the subjects who participated in the study, C. Long (Dept of Respiratory Physiology, Glenfield Hospital, Leicester, UK) for assistance in data analysis, D.D. Vara (Dept of Respiratory Physiology, Glenfield Hospital) for assistance in the clinical characterisa- tion of some of the patients and the lung function unit staff at King’s College Hospital, London, UK (C. Wood, L. Morgan, D. Cox and D. Land) and Glenfield Hospital. REFERENCES 1 Irwin RS, Madison JM. The diagnosis and treatment of cough. N Engl J Med 2000; 343: 1715–1721. 2 Irwin RS. Assessing cough severity and efficacy of therapy in clinical research: ACCP evidence-based clinical practice guidelines. Chest 2006; 129: Suppl. 1, 232S–237S. 3 Morice AH, Fontana GA, Sovijarvi AR, et al. The diagnosis and management of chronic cough. Eur Respir J 2004; 24: 481–492. 4 Birring SS, Prudon B, Carr AJ, Singh SJ, Morgan MD, Pavord ID. Development of a symptom specific health status measure for patients with chronic cough: Leicester Cough Questionnaire (LCQ). Thorax 2003; 58: 339–343. 5 Raj AA, Birring SS. Clinical assessment of chronic cough severity. Pulm Pharmacol Ther 2007; 20: 334–337. 6 Prudon B, Birring SS, Vara DD, Hall AP, Thompson JP, Pavord ID. Cough and glottic-stop reflex sensitivity in health and disease. Chest 2005; 127: 550–557. 7 Birring SS, Matos S, Patel RB, Prudon B, Evans DH, Pavord ID. Cough frequency, cough sensitivity and health status in patients with chronic cough. Respir Med 2006; 100: 1105–1109. 8 Matos S, Birring SS, Pavord ID, Evans DH. Detection of cough signals in continuous audio recordings using hidden Markov models. IEEE Trans Biomed Eng 2006; 53: 1078–1083. 9 Coyle MA, Keenan DB, Henderson LS, et al. Evaluation of an ambulatory system for the quantification of cough frequency in patients with chronic obstructive pulmonary disease. Cough 2005; 1: 3. 10 Pavesi L, Subburaj S, Porter-Shaw K. Application and validation of a computerized cough acquisition system for objective monitoring of acute cough: a meta-analysis. Chest 2001; 120: 1121–1128. 11 Chang AB, Newman RG, Phelan PD, Robertson CF. A new use for an old Holter monitor: an ambulatory cough meter. Eur Respir J 1997; 10: 1637–1639. 12 Smith J, Owen E, Earis J, Woodcock A. Effect of codeine on objective measurement of cough in chronic obstructive pulmonary disease. J Allergy Clin Immunol 2006; 117: 831–835. 13 Hsu JY, Stone RA, Logan-Sinclair RB, Worsdell M, Busst CM, Chung KF. Coughing frequency in patients with persistent cough: assessment using a 24 hour ambulatory recorder. Eur Respir J 1994; 7: 1246–1253. 14 Barry SJ, Dane AD, Morice AH, Walmsley AD. The automatic recognition and counting of cough. Cough 2006; 2: 8. 15 Paul IM, Wai K, Jewell SJ, Shaffer ML, Varadan VV. Evaluation of a new self-contained, ambulatory, objective cough monitor. Cough 2006; 2: 7. 16 Brightling CE, Ward R, Goh KL, Wardlaw AJ, Pavord ID. Eosinophilic bronchitis is an important cause of chronic cough. Am J Respir Crit Care Med 1999; 160: 406–410. 17 Matos S, Birring SS, Pavord ID, Evans DH. An automated system for 24-hour monitoring of cough frequency: the Leicester cough monitor. IEEE Trans Biomed Eng 2007; 54: 1472–1479. 18 Smith JA, Earis JE, Woodcock AA. Establishing a gold standard for manual cough counting: video versus digital audio recordings. Cough 2006; 2: 6. LCM: A NOVEL COUGH MONITOR S.S. BIRRING ET AL. 1018 VOLUME 31 NUMBER 5 EUROPEAN RESPIRATORY JOURNAL