Definition
An anaesthetic gas scavenging system is a system or device that
collects excess anaesthetic gas from the breathing system and
discharges it outside the working environment.
• It is used to collect gases or aerosolized medications from the patient exhalations
or found near the patient area because of some treatment related activities.
Often associated with delivery of anaesthesia but may also include other patient
related activities .
Introduction
When patients are administered anaesthetic gas, its molecules escape into the room since the use
of gases and agents exceeds the amount necessary for the patient and can impact the
performance of medical teams.
Inhaled anesthetic agents include two different classes of chemicals: nitrous oxide and
halogenated agents. Halogenated agents currently in use include halothane , enflurane,
isoflurane, desflurane , and sevoflurane .
Since the late 1960s there has been speculation that trace anaesthetic gases/vapours may have a
harmful effect on operating theatre personnel, but it has been conclusively proved from currently
available studies that there is no association between occupational exposure to trace levels of
waste anaesthetic vapours in scavenged operating theatres and adverse health effects. Nitrous
Oxide does some in high concentration.
However, it is desirable to vent out the exhaled anaesthetic vapours and maintain a vapour-free
theatre environment. At present, the Occupational Safety and Health Administration (OSHA) has
no permissible exposure limits regulating these agents.
In the Operating Room the Anaesthetic Gas Scavenging System collects and removes waste gases
from the patient breathing circuit and the patient ventilation circuit. It can also collect any type of
gases or aerosolized medications that is intended only for the patient should not be breath by
others. AGSS systems to ensure the safety of your patients and staff.
Waste anesthetic gases (WAGs).
• Healthcare workers in a variety of settings( operating rooms, recovery
area, labour rooms, even radiology suites) can be exposed to the
anesthetic gases that are released or leak out during medical
procedures.
• Gases most commonly used include nitrous oxide, isoflurane,
desflurane, and sevoflurane.
• These gases and vapours are known as waste anesthetic gases
(WAGs).
Reasons for scavenging
• On the basis of epidemiological studies carried out since the mid 1960s, it has
been suggested that operating theatre staff and their wives and children are
subject to an increased risk of various pathologies, including  spontaneous
abortion, stillbirth, congenital abnormalities, low birth weight, altered sex ratio,
infertility, cancer, liver disease and kidney disease.
• Vessey and Nunn (1980) have critically reviewed both the epidemiological and
the animal studies, and conclude that, while the evidence for most of the above
hazards is unconvincing, the increased risk of abortion in female theatre staff is
reasonably well substantiated.
• Nitrous oxide inhibits the enzyme methionine synthase by inactivating its bound
co-factor methyl cobalamin (a form of vitamin B12), probably causing impairment
of deoxyribonucleic acid synthesis, which could well produce fetotoxic effects
(Shareet al., 1983).
Evidences of Adverse effects on Prolonged
Exposed.
• Nitrous Oxide: Several summaries of the epidemiologic studies of exposure
to N2O and reviews of the topic generally including animal and
retrospective studies (Purdham 1986; Kestenberg 1988; and NIOSH 1994)
have been published.
• They report a consistent excess of spontaneous abortion in exposed
women. Evidence for congenital abnormalities is less strongly associated
with exposure.
• Halogenated Hydrocarbons : Moreover, there is biological plausibility that
adds to the concern that high levels of unscavenged waste anesthetic
gases present a potential for adverse neurological effects or reproductive
risk to exposed workers or developmental anomalies in their offspring
(Cohen et al. 1980; Rowland et al. 1992).
Level of different agents which showed no
significant adverse effects on animal studies.
• 100 particles per million (ppm) for nitrous oxide
• 50 ppm for enflurane
• 50 ppm for isoflurane
• 10 ppm for halothane
• 20 ppm for sevoflurane (recommended by Abbot Laboratories)
• No limit set for desflurane although a 50 ppm target is advisable due to its
similarity to enflurane.
These levels were chosen because they are well below the levels at which
any significant adverse effects occurred in animals and represent levels at
which there is no evidence to suggest human health would be affected.
Recommended levels for different agents by
different agencies.
• In the United States, the maximum accepted concentrations of any
Halogenated agent should be less than 2 ppm.
• When such agents are used in combination with nitrous oxide, levels of
less than 0.5 ppm should be achieved.
• Nitrous oxide, (when used as the sole anaesthetic agent), at 8-hour time-
weighted average concentrations should be less than 25 ppm during the
administration of an anaesthetic.
• Holland has a limit of 25 ppm for nitrous oxide, whereas Italy, Sweden,
Norway and Denmark set 100 ppm as their limit for exposure to nitrous
oxide.
 It is not possible to set uniform levels without sufficient data.
In 2000, OSHA revised its recommendations on
waste anesthetic gases
In 2000, OSHA revised its recommendations on waste anesthetic gases in the light of current knowledge.
• The revised recommendations are published on the Internet for informational purposes only and are
regularly updated as information becomes available. The document is not published in the standard OSHA
manual on occupational hazards, however.
• The recommendations are advisory and have not been promulgated as a standard; rather, they are to be
seen as guidelines.
1. OSHA recommends scavenging of waste anesthetic gases in all anesthetizing locations and advocates work
practices to reduce trace levels of anesthetic gases in the ambient air.
2. A documented maintenance program should be in place for all anesthetic delivery machines, and an
ongoing education program for all personnel to inform them of these recommendations must exist.
3.OSHA recommends a program for monitoring trace anesthetic gases and also recommends a pre
employment medical examination for all employees.
4. Each institution also should have a mechanism in place for employees to report any work-related health
problems.
Association of Anaesthetists of Great Britain and
Ireland (1975) recommendation.
• As a result of the increasing concern about the possible hazards of
exposure to waste anaesthetic gases, the Association of Anaesthetists
of Great Britain and Ireland (1975) has recommended that all
anaesthetizing locations should be equipped with scavenging
systems, and this is now official DHSS policy (Department of Health
and Social Security, 1976).
• NIOSH recommendation to OSHA: Workers should not be exposed to
an eight hour time-weighted average of > 2 ppm halogenated agents
(not > 0.5 ppm if nitrous oxide is in use) or > 25 ppm nitrous oxide.
Target level for operating theatre pollution
In most jurisdictions, there is a legal requirement to scavenge waste gases to
maintain the level of waste gases in the Operating Room below the legally
acceptable limit.
 In the UK: the limits are typically  100ppm for nitrous oxide and 50ppm for
halogenated volatile anaesthetic agents (except halothane which is 10ppm).
In U.S.A NIOSH Recommendations published; 1977 by the (National Institute for
Occupational Safety and Health), states that:-
o Nitrous oxide maximum 25 ppm and
o Halogenated volatile gases maximum 2 ppm.
(Sharer and colleagues (1983) have criticized the NIOSH limit for nitrous oxide as
being too restrictive, and have suggested a limit of 200 p.p.m., based on a
consideration of the data on methionine synthase inactivation and fetotoxicity).
Anaesthetic Gases &
Agents in use.
Name of Agents Year of Introduction Present Use
Diethyl ether 1842 No
Nitrous oxide 1844 yes
Chloroform 1847 NO
Cyclopropane 1933 NO
Trichloroethylene 1934 No
Fluroxene 1954 NO
Halothane 1956 yes
Methoxyflurane 1960 Infrequently
Enflurane 1974 yes
Isoflurane 1980 yes
Desflurane 1992 yes
Sevoflurane 1995 yes
Nitrous Oxide and only few of the
anaesthetic agents are in use. Most
commonly used drugs are Sevoflurane,
Isoflurane , Desflurane & Halothane
Adverse effects
 National Institute for Occupational Safety and Health ,2007( NIOSH) states that “exposure to high
concentrations of waste anesthetic gases - even for a short time - may cause the following health
effects:
• Headache
• Irritability
• Fatigue
• Nausea
• Drowsiness
• Difficulties with judgment and coordination
• Liver and kidney disease”
These health effects were mainly noted for older anesthetics (e.g., trichloroethylene,
methoxyflurane) that are no longer commonly in use.
 Studies are inconclusive on the potential health effects from occupational exposure to some of
the newer anesthetics, such as isoflurane
NIOSH studies
• NIOSH (2007) continues: “Although some studies report no adverse health effects
from long-term exposure to low concentrations of waste anesthetic gases, several
studies have linked such exposure to miscarriages, genetic damage, and cancer
among operating-room workers.
• Studies have also reported miscarriages in the spouses of exposed workers and
birth defects in their offspring.”
• NIOSH (2015) later reports that “Some studies have documented adverse health
effects (e.g., headaches, fatigue, irritability, birth defects, miscarriages, liver and
kidney disease, cancer) from excessive exposure to anesthetic gas
 Scavenger systems were forced on anesthesia machines by the
‘recommendations’ of the National Institute for Occupational Safety and Health
(NIOSH).
The recommendations for room air concentrations of waste anesthetic gases is
25 ppm for nitrous oxide and 2 ppm for halogenated agents.
Who is exposed to waste anesthetic gases?
The following hospital workers may be exposed to waste anesthetic
gases:
• Anesthesiologists
• Dentists
• Nurse anesthetists
• Operating-room nurses
• Operating-room technicians
• Other operating-room personnel
• Recovery-room nurses
• Other recovery-room personnel
• Surgeons
Objective
 The objective of Anaesthetic gas scavenging systems(AGSS) to
prevent medical staff from inhaling the anaesthetic nitrous oxide and
halogenated hydrocarbons administered to patients during surgery.
In most jurisdictions, there is a legal requirement to scavenge waste
gases to maintain the level of waste gases in the Operating Room
below the legally acceptable limit.
In addition to the legal requirement there is an Occupational Health
requirement to maintain a safe workplace and limit exposure to
potentially harmful gases and agents.
Guidelines for controlling pollution in OR
• There is no association between occupational exposure to
anaesthetic agents trace levels and adverse health effects.
• There are no agreed international standards of the maximum
accepted concentrations of agents in the theatre environment.
• Routine monitoring and testing (PPM) are mandatory.
Factors responsible for theatre pollution
Anaesthesia Techniques Anaesthesia Machines
• Leaks from various connections
like tubing connections if not
fitting properly, soda lime
canister etc.
• Others: Cryosurgery Unit,
Cardiopulmonary bypass circuit
if vapor is used.
Poorly fitting facemask
Pediatric breathing system( T-piece)
Un-cuffed tracheal Tubes
Gases coming out through APL valve
Gases coming through ventilator exhaust
Exhalation gases during recovery
Spillage during filling of vaporizers
How to reduce pollution in OR?
 1. Adequate theatre ventilation and air conditioning, with frequent and rapid changing of the circulating air (15–20 times per hour).
 Theatres that are unventilated are four times as contaminated with anaesthetic gases and vapours compared to those with proper
ventilation.
 A non-recirculating ventilation system is usually used. A recirculating ventilation system is not recommended. In labour wards, where
anaesthetic agents including Entonox are used, rooms should be well ventilated with a minimum of five air changes per hour.
 2. Use of the circle breathing system.
 This system recycles the exhaled anaesthetic vapours, absorbing CO2.
 It requires a very low fresh gas flow, so reducing the amount of inhalational agents used.
 3. Total intravenous anaesthesia.
 4. Regional anaesthesia.
 5. Avoiding spillage and using fume cupboards during vaporizer filling.. Modern vaporizers use special agent-specific filling devices as a
safety feature and to reduce spillage and pollution.
 Good technique will also help lessen exposure.
 6. Scavenging of Anaesthetic gases.
Good technique will also help lessen
exposure.
• Good mask fit
• Avoid unscavengeable techniques if possible (insufflation)
• Prevent flow from breathing system into room air (only turn on agent and nitrous
oxide after mask is on face, turn them off before suctioning)
• Washout anesthetics (into the breathing circuit) at the end of the anesthetic
• Don’t spill liquid agent
• Use low flows
• Use cuffed tracheal tubes when possible
• Check the machine regularly for leaks
• Disconnect nitrous oxide pipeline connection at wall at the day’s end (beginning?)
• Total intravenous anesthesia
Scavenging of waste Gases
• In any location in which inhalation anaesthetics are administered,
there should be an adequate and reliable system for scavenging the
waste anaesthetic gases.
• A scavenging system need to be capable of collecting the waste
anaesthetic gases from the breathing system and disposing them
safely.
Unscavenged operating theatres can show a very high levels of N2O.
(400–3000 ppm).
Scavenging of waste gases may be a passive or active system.
A passive scavenging system.
• A passive scavenging system operates without the use of suction,
since the positive pressure of gas in the breathing circuit pushes
waste anesthetic gases into the scavenging system.
• One-way valves in the interface help to move waste anesthetic gases
outdoors or into a non-recirculating air ventilation system.
• A passive scavenging system does not involve the use of a vacuum
pump or suction.
Active scavenging systems
• Active scavenging systems use suction, or a vacuum pump, to actively
remove waste anesthetic gases from the breathing circuit and draw
these gases into a scavenging system.
• The interface of an active system requires a negative pressure relief
valve in order to prevent negative pressure from reaching the
breathing circuit and affecting the patient’s lungs.
Comparison of both the systems
 Active System
• Uses a device like suction to draw
gases from breathing system.
• Mostly uses a compressor to draw
the gases and agents.
• Expensive
• Requires maintenance.
• Requires a means to protect the
patient's airway from the
application of suction, or buildup of
positive pressure.
Passive System
• This system uses no suction, pressure
in the gas line helps to drive out the
gases out of the machine.
• It must be located adjacent to an
outside place, the pipe simply passes
through a whole in the wall.
• It is inexpensive
• Dose not require maintenance.
• Requires the patient to be protected
from a positive pressure buildup only
if there is obstruction in the exit limb.
The basic functional components of an Anaesthetic Gas
Scavenging System are as follows:
• 1.A collecting assembly / shroud with a relief valve by which the
waste gas leaves the breathing or ventilation circuit.
• 2.A transfer system of tubing to conduct waste gases to the
Scavenging Interface.
• 3.The Scavenging Interface, and
• 4.A Disposal line to conduct the waste gas to a passive evacuation
system, or a Waste Anaesthetic Gas Disposal/Medical Vacuum system
via a station outlet.
Schematic diagram showing the
different components of scavenging
system
1.A collecting assembly / shroud with a relief valve by which the
waste gas leaves the breathing or ventilation circuit.
2.A transfer system of tubing to conduct waste gases to the
Scavenging Interface.
3.The Scavenging Interface, and
4.A Disposal line to conduct the waste gas to a passive evacuation
system,
A collecting Assembly( Receiving system)
Gas collection assembly, (tubes connected to APL and vent relief valve)
 A reservoir bag can be used receiving or collecting system .
 Two spring-loaded valves in the system guard against excessive positive
(1000 Pa) in case of a distal obstruction or negative (–50 Pa) pressures in
case of increased demand in the scavenging system.
Without these valves, excessive positive pressure increases the risk of
barotrauma, should there be an obstruction beyond the receiving system.
Excessive negative pressure could lead to the collapse of the reservoir bag
of the breathing system and the risk of rebreathing.
Collecting System
Transfer System
It is a simple wide bore pipe without any leak which
can transfer from the collecting system to Scavenging
interface. Can be copper pipe.
Transfer tubing (19 or 30 mm, sometimes yellow color-
coded)
The transfer hose should always be fitted with a
pressure relief valve (10cmH2O) such that if the
transfer hose were to become occluded for any reason
the patient would still have an expiratory path.
Scavenging Interface
• The scavenger interface is the most important component.
• The extraction flow rate ensures that waste anaesthetic gases are adequately removed from the
system. It protects the breathing circuit from excess positive or negative pressure.
• Positive-pressure relief valve is mandatory to vent excess gas in case of occlusion distal to a
closed interface. If active disposal system in use, must have negative pressure relief valve as well.
• Reservoir highly desirable with active systems.
• The induced flow rate should be as low as possible with the system ideally being passive between
the patient and the receiving unit and active from the receiving unit to the exhaust point.
• The importance of this flow rate is that if it falls too low it may not be sufficient for waste gas
removal.
• If too high, this would lead to the waste gases spilling out from the base of the receiving reservoir
unit into the immediate working environment; it may lead to an increase of the induced flow at
the patient connection port.
Closed interface
Closed interface communicates with
atmosphere only through valves.
Should adjust vacuum pressor so that
reservoir bag neither flattened not over-
distended.
Active Scavenging Interface closed
system.
The Ohio scavenging interface has
connections for the outlets from the
breathing system and ventilator (A),
one or two reservoir bags (B), and
the vacuum line (V).
The suction is controlled by the
needle valve (N).
There are both positive (P1) and
negative (P2) pressure relief valves
in case the reservoir bag becomes
empty or too full.
Function of the Positive and Negative Pressure
Relief Valves
The scavenging system has negative and positive pressure relief valves to
prevent excessively high or low pressures from developing in the
scavenging manifold and being transmitted to the patient via the breathing
circuit.
The positive pressure relief valve: will lift up and open at approximately 5 cm
H2O. This happens when the vacuum is too weak or if the vacuum adjustment valve is
completely closed.
A slight hissing or rattling sound can be heard (indicating the venting of exhaust gases
into the room) during the expiratory phase of mechanical ventilation or when the manual
bag is squeezed. In addition, the scavenging bag will be fully inflated.
Negative pressure relief valve: to prevent negative pressure from being
transmitted to the breathing circuit, will actuate at approximately -0.25 cm H2O and
allow room air to enter the scavenging system to "break" the vacuum. The scavenging
bag will be collapsed flat.
Open interface
Open interface has no valves, and is open to
atmosphere (allows both negative and
positive pressure relief).
Should be used only with active systems.
Remember that slight hissing from an open
interface is normal- there is no audible
indication of waste gas leaks.
Safer for the patient (no hazard of positive or
negative pressure being applied to the
airway as a result of scavenger failure).
The risk of occupational exposure for
providers ignorant of their proper use is
higher with the open interface (Anesth
Analg 1992;75:1073).
Home made Simple
active scavenging
interface
A cylindrical plastic jar.
Suction through valve attached at the
bottom.
At the top there are two inlets, one for waste
gas pipe insertion the other one for air to
enter if creates excessive negative pressure.
The disposal system
 The disposal system is a wide-bore copper pipe leading to the
atmosphere directly or via the theatre ventilation system.
Gas disposal assembly (active or passive, active system most
common uses the hospital suction system)
Passive disposal system
• In the past, the standard of collecting waste gases has been through
passive methods, which include (depending on the flow rate of the
delivery gas) pushing the waste gas into an activated charcoal
canister.
Activated charcoal cartridges.
• Activated charcoal cartridges can also be used for passive
scavenging.
• It is important to note the limitations of these cartridges: they are
less effective at high anesthesia flow rates, they must be changed
every 12 hours, and care must be taken to avoid occluding the
cartridge’s air exit holes.
• When used correctly, however, activated charcoal cartridges can
effectively decrease waste anesthetic gas exposure.
F/air Canister
A sensible answer to anesthesia gas
problems in the operating room, the F/AIR
anesthesia gas filter was specifically designed
to remove waste anesthesia gases such as
ISOFLURANE, HALOTHANE, ENFLURANE,
SEVOFLURANE, etc. from the operating room
environment.
Economical to use, with an average useful life
of 12 to 15 hours, the F/AIR is easily
adaptable to anesthesia machines.
The F/AIR can remove no less than 50 grams
of pure halogenated anesthetic gases and
can then be conveniently discarded.
The F-AirÂŽ canister must be changed after 8
hrs of normal use or a weight gain of 50
grams.
Disposable
Ready to use
WAG Activated Charcoal
Canisters
Activated Charcoal Filtration System
Contains 2 pounds of the highest quality
activated charcoal available to make the WAG
the most effective filter on the market.
Will adsorb a minimum of 200 grams
Halogenated Agent. (4 times the amount of F-
Air Canisters).
Unique design traps Isoflurane/Sevoflurane
Agent molecules in the canister.
Relief holes on top ensure that the canister can
sit upright on a flat surface.
Universal tubing connections: 19mm (standard
in industry), 22mm, and fitting for 15mm male
to use ¼” ID Tubing.
Passed independent testing according to OSHA
regulations of 2ppm.
No breakthrough exhibited at +/- 60 LPM flow
rate (Verified via infra red spectrophotometer).
Drawing showing the
flow system of gases.
Diagram of a passive scavenging system.
Modern scavenging system has four components.
a). Collecting System:
b).Transferring System:.
c). Receiving System:
d). Gas disposal Assembly: May be a passive or active
disposal unit.
Passive System
The passive system is simple to construct with zero running cost.
Components are :-
1. The collecting and transfer system which consists of a shroud connected to
the adjustable pressure limiting (APL) valve (or expiratory valve of the
ventilator).
2. A 30-mm connector attached to transfer tubing leads to a receiving system.
3. Receiving System.
4. Gas Disposal Unit or assembly.
The 30-mm wide-bore connector is designed as a safety measure in order to
prevent accidental misconnection to other ports of the breathing system .
Draw backs of passive scavenging system
• Scavenging systems are either passive or active.
• Passive systems have a large tube or canister with an inlet, an outlet,
and one open end.
• The waste anesthetic gases from the ventilator and ‘pop-off’ valve
enter the inlet.
• Suction is applied at the outlet.
• If the suction rate exceeds the rate of entry of anesthetic gases, room
air is drawn into the open end.
• If entry of anesthetic gases exceeds the rate of suction, excess gases
exit via the open end into the room.
Active system
• The purpose of the system from the plant to the terminal outlet is to provide the extraction flow
rates to remove waste anaesthetic gases from the receiving unit reservoir.
• The role of the receiving unit is to provide a safe interface between the patient and the extraction
flow rates.
Components
• The collecting and transfer system is similar to that of the passive
system .
• A fan or Vacuum unit to establish flow.
Adverse effects of active system.
• Active systems contain negative and positive relief valves to release excess
pressures.
• Active systems have been associated with higher exposure of gases in the
environment than passive systems.
• When the reservoir bag overinflates due to an excess inflow of anesthetic gases,
pressure increases until the positive pressure relief valve opens to vent excessive
gases into the room.
• A valve malfunction will allow the pressure increase in the patient circuit to
increase to the limit of the elastic properties of the particular reservoir bag.
• High pressures can be detected in the patient circuit by the circuit pressure gauge
and/or by the high pressure gauge of the ventilator.
• If high pressure persists the patient may develop decreased cardiac output
secondary to high intrathoracic pressure, and/or pneumothorax.
Adverse effects of active system(Conti-)
• When the suction rate exceeds the anesthetic gas inflow rate, the
reservoir bag deflates.
• Usually any excessive negative pressure is offset by room air entering
the negative pressure relief valve.
• Some scavenger systems have a second or backup relief valve.
• The pressure gauge of the ventilator may detect an excessive negative
pressure state.
• If the negative pressure relief valve system malfunctions, it could be
possible for the suction to remove anesthetic gases from the patient
circuit.
Mechanism of action
• 1. The exhaled gases are driven by either the patient’s respiratory efforts or the
ventilator.
• 2. The receiving system should be mounted on the anaesthetic machine to minimize the
length of transfer tubing, therefore minimizing resistance to flow.
• Problems in practice and safety features
• 1. Connecting the scavenging system to the exit grille of the theatre ventilation is
possible. Recirculation or reversing of the flow is a problem in this situation.
• 2. Excess positive or negative pressures caused by the wind at the outlet might affect the
performance and even reverse the flow.
• 3. The outlet should be fitted with a wire mesh to protect against insects.
• 4. Compressing or occluding the passive hose may lead to the escape of gases/vapours
into the operating theatre and thereby polluting it.
• The disposal hose should be made of non-compressible materials and not placed on the
floor.
Requirements of standard Scavenging System
The draft BS( British Standard) on scavenging equipment includes the
following provisions for patient safety:
(1) At a continuous flow of air 30 litre min"1, the scavenging system
should not impose a greater resistance than 50 Pa.
(2) In the event of total obstruction of the system, or failure of the
power supply, the resistance should not be greater than 1 kPa at a
flow of 30 litremin"1. If the transfer tubing is less than 1 m in length,
the necessary pressure relief may be afforded by the receiving
system.
(3) The scavenging system shall not create a sub-atmospheric
pressure in the collecting system of more than 50 Pa.
Monitoring the functioning of Scavenging system.
 Sampling procedures for evaluating waste anaesthetic vapour concentrations
in air should be conducted for nitrous oxide and halogenated agents on a yearly
basis in the UK and on a quarterly basis in the USA in each location where
anaesthesia is administered.
 Monitoring should include:
• Leak testing of equipment
• Sampling air in the theatre personnel breathing zone.
Planned preventative maintenance (PPM) programme. Anaesthetic equipment,
gas scavenging, gas supply, flowmeters and ventilation systems must be subject
to a maintenance programme.
, The general ventilation system and the scavenging equipment should be
examined and tested by a responsible person, at least once annually .
WHAT IS AN IDEAL SCAVENGING SYSTEM?
.
1.Should not affect the ventilation and oxygenation of the patient.
2.Should not affect the dynamics of the breathing system.
3.A well-designed scavenging system should consist of
i. a collecting device for gases from the breathing system or ventilator at the site of
overflow,
ii. a system to carry waste anaesthetic gases from the collecting device and a method for
limiting both positive and negative pressure variations in the breathing system.
4.The performance of the scavenging system should be part of the
anaesthetic machine check.
5.Scavenging systems can be of a passive or active type.
WHAT IS AN IDEAL SCAVENGING SYSTEM (CONT-)?
6.The scavenger system must be independent of the main hospital
ventilation system.
7. The scavenging system should be kept in good repair to prevent
leaks using a maintenance and inspection program as listed below.
8.Gases are not discharged near the air intake of the building or
surrounding buildings.
9. In the event of a “Code RED”, hospital ventilation is shut down to
reduce the potential spread of a fire but the scavenger system must
continue to work.
Consider using a WAG collection canister that can be attached to an
anesthetic gas machine before the scavenger. It will capture the
agents (except for nitrous oxide).This technology reduces emissions to
the atmosphere by capturing the agents and recycling them.
An effective waste anesthetic gas management program
includes:
• Engineering Controls,
• Work Practices,
• Air Monitoring and,
• Hazard Communication and Training.
Air Monitoring
• One of the tools used to measure exposure to waste anesthetic gases is air
monitoring.
• Monitoring may be continuous or periodic but should adequately measure
exposure in the work areas and surrounding areas.
• Monitoring can aid in identifying the presence and location of leaked gases
and the effectiveness of corrective measures.
• As most halogenated anesthetic gases cannot be detected by smell (unless
they are in high concentration) proper monitoring becomes all the more
critical.
• Nitrous oxide is an odourless and colourless gas and can only be detected
by WAG monitor.
Hazard Communication and Training
• Employers should develop and implement a written hazard communication program regarding WAGs
that includes description of the physical and health hazards of anesthetic agents in use, the compiling
and availability of up to date material safety data sheets on all anesthetic gases used; proper labelling
of canisters, tanks, and containers; and a comprehensive employee training and information program.
• The training program should list steps, the workers can take to protect themselves from the hazards of
WAGs.
• The program should include information on steps taken by the employer;
- such as engineering controls,
- clearly outline emergency procedures to contain spills,
- describe safe work practices and the use of any personal protective equipment, and
- detail the use of continuous monitoring devices.
• The training program should clearly outline all methods and observable indicators that can detect the
presence and release of anesthetic gases.
• Spills should be treated as emergencies. Spills of anesthetic agents must only be cleaned up and
controlled by properly trained and equipped personnel
Good technique that lessens expose to waste
anaesthetic gases.
• Good mask fit
• Avoid unscavengeable techniques if possible (insufflation)
• Prevent flow from breathing system into room air (only turn on agent and nitrous
oxide after mask is on face, turn them off before suctioning)
• Washout anesthetics (into the breathing circuit) at the end of the anesthetic
• Don’t spill liquid agent
• Use low flows
• Use cuffed tracheal tubes when possible
• Check the machine regularly for leaks
• Disconnect nitrous oxide pipeline connection at wall at the day’s end (beginning?)
• Total intravenous anesthesia
Certain devices & systems for an ideal
scavenging system.
Passive Outlet for
AGSS
Separate Suction System for Anaesthesia gas scavenging
Exhaust Fan Use to
Scavenge AGSS
Lack Circuit(J. A. LACK)
Connection to be fitted to APL valve
Devices using N2O
Thanks.

More Related Content

What's hot

Medical gas cylinders
Medical gas cylindersMedical gas cylinders
Medical gas cylindersAshish Dhandare
 
Face masks, laryngeal tube, airways yuvaraj
Face masks, laryngeal tube, airways  yuvarajFace masks, laryngeal tube, airways  yuvaraj
Face masks, laryngeal tube, airways yuvarajhavalprit
 
Anaesthesia workstation
Anaesthesia workstationAnaesthesia workstation
Anaesthesia workstationZIKRULLAH MALLICK
 
Anaesthesia Vaporizers
Anaesthesia VaporizersAnaesthesia Vaporizers
Anaesthesia VaporizersRahul Varshney
 
Management of Bronchospasm during General Anaesthesia
Management of Bronchospasm during General Anaesthesia Management of Bronchospasm during General Anaesthesia
Management of Bronchospasm during General Anaesthesia Ashwin Haridas
 
Safety features in anesthesia machine
Safety features in anesthesia machineSafety features in anesthesia machine
Safety features in anesthesia machineomar143
 
Low pressure system in anaesthesia machine
Low pressure system in anaesthesia machineLow pressure system in anaesthesia machine
Low pressure system in anaesthesia machineSwadheen Rout
 
Humidifiers in anaesthesia and critical care
Humidifiers in anaesthesia and critical careHumidifiers in anaesthesia and critical care
Humidifiers in anaesthesia and critical careTuhin Mistry
 
SAFETY FEATURES OF ANAESTHESIA MACHINE
SAFETY FEATURES OF ANAESTHESIA MACHINESAFETY FEATURES OF ANAESTHESIA MACHINE
SAFETY FEATURES OF ANAESTHESIA MACHINEMAHESWARI JAIKUMAR
 
Anaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supplyAnaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supplyUnnikrishnan Prathapadas
 
Mapleson breathing systems
Mapleson breathing systemsMapleson breathing systems
Mapleson breathing systemsdrdeepak016
 
Facemask , oral and nasal airways
Facemask , oral and nasal airwaysFacemask , oral and nasal airways
Facemask , oral and nasal airwaysDR SHADAB KAMAL
 
Breathing circuit's
Breathing circuit'sBreathing circuit's
Breathing circuit'sImran Sheikh
 
Physiology of inhalational anaesthetic agents
Physiology of inhalational anaesthetic  agentsPhysiology of inhalational anaesthetic  agents
Physiology of inhalational anaesthetic agentsDr Ravi Shankar Sharma
 
Neuromuscular Monitoring
Neuromuscular MonitoringNeuromuscular Monitoring
Neuromuscular MonitoringMohtasib Madaoo
 
Anesthesia workstation
Anesthesia workstationAnesthesia workstation
Anesthesia workstationRashmit Shrestha
 
Minimum alveolar concentration (mac)
Minimum alveolar concentration (mac)Minimum alveolar concentration (mac)
Minimum alveolar concentration (mac)Torrentz Tiku
 
Circle system low flow anesthesia
Circle system low flow anesthesiaCircle system low flow anesthesia
Circle system low flow anesthesiaDrgeeta Choudhary
 

What's hot (20)

Medical gas cylinders
Medical gas cylindersMedical gas cylinders
Medical gas cylinders
 
Face masks, laryngeal tube, airways yuvaraj
Face masks, laryngeal tube, airways  yuvarajFace masks, laryngeal tube, airways  yuvaraj
Face masks, laryngeal tube, airways yuvaraj
 
Anaesthesia workstation
Anaesthesia workstationAnaesthesia workstation
Anaesthesia workstation
 
Anaesthesia Vaporizers
Anaesthesia VaporizersAnaesthesia Vaporizers
Anaesthesia Vaporizers
 
Management of Bronchospasm during General Anaesthesia
Management of Bronchospasm during General Anaesthesia Management of Bronchospasm during General Anaesthesia
Management of Bronchospasm during General Anaesthesia
 
Safety features in anesthesia machine
Safety features in anesthesia machineSafety features in anesthesia machine
Safety features in anesthesia machine
 
Low pressure system in anaesthesia machine
Low pressure system in anaesthesia machineLow pressure system in anaesthesia machine
Low pressure system in anaesthesia machine
 
Humidifiers in anaesthesia and critical care
Humidifiers in anaesthesia and critical careHumidifiers in anaesthesia and critical care
Humidifiers in anaesthesia and critical care
 
SAFETY FEATURES OF ANAESTHESIA MACHINE
SAFETY FEATURES OF ANAESTHESIA MACHINESAFETY FEATURES OF ANAESTHESIA MACHINE
SAFETY FEATURES OF ANAESTHESIA MACHINE
 
Breathing circuits
Breathing circuitsBreathing circuits
Breathing circuits
 
Anaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supplyAnaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supply
 
Mapleson breathing systems
Mapleson breathing systemsMapleson breathing systems
Mapleson breathing systems
 
Facemask , oral and nasal airways
Facemask , oral and nasal airwaysFacemask , oral and nasal airways
Facemask , oral and nasal airways
 
Breathing circuit's
Breathing circuit'sBreathing circuit's
Breathing circuit's
 
Physiology of inhalational anaesthetic agents
Physiology of inhalational anaesthetic  agentsPhysiology of inhalational anaesthetic  agents
Physiology of inhalational anaesthetic agents
 
Neuromuscular Monitoring
Neuromuscular MonitoringNeuromuscular Monitoring
Neuromuscular Monitoring
 
Anesthesia workstation
Anesthesia workstationAnesthesia workstation
Anesthesia workstation
 
Minimum alveolar concentration (mac)
Minimum alveolar concentration (mac)Minimum alveolar concentration (mac)
Minimum alveolar concentration (mac)
 
Vaporizers
Vaporizers Vaporizers
Vaporizers
 
Circle system low flow anesthesia
Circle system low flow anesthesiaCircle system low flow anesthesia
Circle system low flow anesthesia
 

Similar to Scavenging of waste anaesthetic gases.

Industrial Ototoxicity
Industrial OtotoxicityIndustrial Ototoxicity
Industrial OtotoxicityAmira Badr
 
Scavanging system
Scavanging systemScavanging system
Scavanging systemKIMS
 
Organophosporous Poisoning - Forensic Medicine
Organophosporous Poisoning - Forensic MedicineOrganophosporous Poisoning - Forensic Medicine
Organophosporous Poisoning - Forensic Medicinesuaveguava12
 
Nitrous Oxide Sedation in Pediatric Dentistry
Nitrous Oxide Sedation in Pediatric DentistryNitrous Oxide Sedation in Pediatric Dentistry
Nitrous Oxide Sedation in Pediatric DentistryMedicineAndFamily
 
elements hazard HSE
elements hazard HSEelements hazard HSE
elements hazard HSEReza Taryghat
 
Nitrous oxide sedation
Nitrous oxide sedationNitrous oxide sedation
Nitrous oxide sedationRahaf Sn
 
Hospital Pharmacy Isolator Solutions for USP <797> Compliance from Esco
Hospital Pharmacy Isolator Solutions for USP <797> Compliance from EscoHospital Pharmacy Isolator Solutions for USP <797> Compliance from Esco
Hospital Pharmacy Isolator Solutions for USP <797> Compliance from EscoEsco Group
 
Acute Respiratory Distress Syndrome
Acute Respiratory Distress SyndromeAcute Respiratory Distress Syndrome
Acute Respiratory Distress SyndromeRikin Hasnani
 
organophosphate poisoning1 medicne..pptx
organophosphate poisoning1 medicne..pptxorganophosphate poisoning1 medicne..pptx
organophosphate poisoning1 medicne..pptxAnnaKhurshid
 
Chemical based hazards
Chemical based hazardsChemical based hazards
Chemical based hazardsMukesh Patil
 
Nitrous oxide inhalation sedation
Nitrous oxide inhalation sedationNitrous oxide inhalation sedation
Nitrous oxide inhalation sedationaravindhanarumugam1
 
Nitrous oxide and its current status
Nitrous oxide and its current statusNitrous oxide and its current status
Nitrous oxide and its current statusRicha Kumar
 

Similar to Scavenging of waste anaesthetic gases. (20)

1447.full
1447.full1447.full
1447.full
 
Industrial Ototoxicity
Industrial OtotoxicityIndustrial Ototoxicity
Industrial Ototoxicity
 
DESFLURANE AND XENON
DESFLURANE AND XENONDESFLURANE AND XENON
DESFLURANE AND XENON
 
Scavanging system
Scavanging systemScavanging system
Scavanging system
 
44-313-1-PB
44-313-1-PB44-313-1-PB
44-313-1-PB
 
Organophosporous Poisoning - Forensic Medicine
Organophosporous Poisoning - Forensic MedicineOrganophosporous Poisoning - Forensic Medicine
Organophosporous Poisoning - Forensic Medicine
 
General anesthetics
General anestheticsGeneral anesthetics
General anesthetics
 
Nitrous Oxide Sedation in Pediatric Dentistry
Nitrous Oxide Sedation in Pediatric DentistryNitrous Oxide Sedation in Pediatric Dentistry
Nitrous Oxide Sedation in Pediatric Dentistry
 
elements hazard HSE
elements hazard HSEelements hazard HSE
elements hazard HSE
 
Nitrous oxide sedation
Nitrous oxide sedationNitrous oxide sedation
Nitrous oxide sedation
 
OP P
OP POP P
OP P
 
oxygen therapy.pptx
oxygen therapy.pptxoxygen therapy.pptx
oxygen therapy.pptx
 
Toxicity of Pesticides
Toxicity of PesticidesToxicity of Pesticides
Toxicity of Pesticides
 
Hospital Pharmacy Isolator Solutions for USP <797> Compliance from Esco
Hospital Pharmacy Isolator Solutions for USP <797> Compliance from EscoHospital Pharmacy Isolator Solutions for USP <797> Compliance from Esco
Hospital Pharmacy Isolator Solutions for USP <797> Compliance from Esco
 
Acute Respiratory Distress Syndrome
Acute Respiratory Distress SyndromeAcute Respiratory Distress Syndrome
Acute Respiratory Distress Syndrome
 
organophosphate poisoning1 medicne..pptx
organophosphate poisoning1 medicne..pptxorganophosphate poisoning1 medicne..pptx
organophosphate poisoning1 medicne..pptx
 
Chemical based hazards
Chemical based hazardsChemical based hazards
Chemical based hazards
 
Nitrous oxide inhalation sedation
Nitrous oxide inhalation sedationNitrous oxide inhalation sedation
Nitrous oxide inhalation sedation
 
Medical gases.pptx
Medical gases.pptxMedical gases.pptx
Medical gases.pptx
 
Nitrous oxide and its current status
Nitrous oxide and its current statusNitrous oxide and its current status
Nitrous oxide and its current status
 

More from KIMS

Vaccinations along all age groups in India
Vaccinations along all age groups in IndiaVaccinations along all age groups in India
Vaccinations along all age groups in IndiaKIMS
 
PFA ( Presentation).pptx
PFA ( Presentation).pptxPFA ( Presentation).pptx
PFA ( Presentation).pptxKIMS
 
Recovery Position ( New).pptx
Recovery Position ( New).pptxRecovery Position ( New).pptx
Recovery Position ( New).pptxKIMS
 
INSULI THERAPY.pptx
INSULI THERAPY.pptxINSULI THERAPY.pptx
INSULI THERAPY.pptxKIMS
 
Circulation of Blood.pptx
Circulation of Blood.pptxCirculation of Blood.pptx
Circulation of Blood.pptxKIMS
 
Dengue.pptx
Dengue.pptxDengue.pptx
Dengue.pptxKIMS
 
PPH.pptx
PPH.pptxPPH.pptx
PPH.pptxKIMS
 
OHA.pptx
OHA.pptxOHA.pptx
OHA.pptxKIMS
 
Surgical hand wash
Surgical hand washSurgical hand wash
Surgical hand washKIMS
 
Hyperlipidaemia.
Hyperlipidaemia.Hyperlipidaemia.
Hyperlipidaemia.KIMS
 
Vaporizers
VaporizersVaporizers
VaporizersKIMS
 
Physics in anesthesia
Physics in anesthesiaPhysics in anesthesia
Physics in anesthesiaKIMS
 
Regional anaesthesia in Pediatrics
Regional anaesthesia in Pediatrics Regional anaesthesia in Pediatrics
Regional anaesthesia in Pediatrics KIMS
 
Massive transfusion
Massive transfusionMassive transfusion
Massive transfusionKIMS
 
Fluid resuscitation in trauma
Fluid resuscitation in traumaFluid resuscitation in trauma
Fluid resuscitation in traumaKIMS
 
Propofol infusion syndrome.
Propofol infusion syndrome.Propofol infusion syndrome.
Propofol infusion syndrome.KIMS
 
Low flow Anesthesia system
Low flow  Anesthesia systemLow flow  Anesthesia system
Low flow Anesthesia systemKIMS
 
Non invasive ventilations
Non invasive ventilationsNon invasive ventilations
Non invasive ventilationsKIMS
 
Non invasive ventilations
Non invasive ventilationsNon invasive ventilations
Non invasive ventilationsKIMS
 
Neonatal resuscitation
Neonatal resuscitationNeonatal resuscitation
Neonatal resuscitationKIMS
 

More from KIMS (20)

Vaccinations along all age groups in India
Vaccinations along all age groups in IndiaVaccinations along all age groups in India
Vaccinations along all age groups in India
 
PFA ( Presentation).pptx
PFA ( Presentation).pptxPFA ( Presentation).pptx
PFA ( Presentation).pptx
 
Recovery Position ( New).pptx
Recovery Position ( New).pptxRecovery Position ( New).pptx
Recovery Position ( New).pptx
 
INSULI THERAPY.pptx
INSULI THERAPY.pptxINSULI THERAPY.pptx
INSULI THERAPY.pptx
 
Circulation of Blood.pptx
Circulation of Blood.pptxCirculation of Blood.pptx
Circulation of Blood.pptx
 
Dengue.pptx
Dengue.pptxDengue.pptx
Dengue.pptx
 
PPH.pptx
PPH.pptxPPH.pptx
PPH.pptx
 
OHA.pptx
OHA.pptxOHA.pptx
OHA.pptx
 
Surgical hand wash
Surgical hand washSurgical hand wash
Surgical hand wash
 
Hyperlipidaemia.
Hyperlipidaemia.Hyperlipidaemia.
Hyperlipidaemia.
 
Vaporizers
VaporizersVaporizers
Vaporizers
 
Physics in anesthesia
Physics in anesthesiaPhysics in anesthesia
Physics in anesthesia
 
Regional anaesthesia in Pediatrics
Regional anaesthesia in Pediatrics Regional anaesthesia in Pediatrics
Regional anaesthesia in Pediatrics
 
Massive transfusion
Massive transfusionMassive transfusion
Massive transfusion
 
Fluid resuscitation in trauma
Fluid resuscitation in traumaFluid resuscitation in trauma
Fluid resuscitation in trauma
 
Propofol infusion syndrome.
Propofol infusion syndrome.Propofol infusion syndrome.
Propofol infusion syndrome.
 
Low flow Anesthesia system
Low flow  Anesthesia systemLow flow  Anesthesia system
Low flow Anesthesia system
 
Non invasive ventilations
Non invasive ventilationsNon invasive ventilations
Non invasive ventilations
 
Non invasive ventilations
Non invasive ventilationsNon invasive ventilations
Non invasive ventilations
 
Neonatal resuscitation
Neonatal resuscitationNeonatal resuscitation
Neonatal resuscitation
 

Recently uploaded

pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...Call Girls Noida
 
Bangalore call girl 👯‍♀️@ Simran Independent Call Girls in Bangalore GIUXUZ...
Bangalore call girl  👯‍♀️@ Simran Independent Call Girls in Bangalore  GIUXUZ...Bangalore call girl  👯‍♀️@ Simran Independent Call Girls in Bangalore  GIUXUZ...
Bangalore call girl 👯‍♀️@ Simran Independent Call Girls in Bangalore GIUXUZ...Gfnyt
 
Jalandhar Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
Jalandhar  Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...Jalandhar  Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
Jalandhar Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...Call Girls Service Chandigarh Ayushi
 
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR Call G...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR   Call G...❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR   Call G...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR Call G...Gfnyt.com
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxAyush Gupta
 
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...Gfnyt.com
 
Call Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In Raipur
Call Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In RaipurCall Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In Raipur
Call Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In Raipurgragmanisha42
 
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar SumanCall Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar SumanCall Girls Service Chandigarh Ayushi
 
Call Girl Gurgaon Saloni 9711199012 Independent Escort Service Gurgaon
Call Girl Gurgaon Saloni 9711199012 Independent Escort Service GurgaonCall Girl Gurgaon Saloni 9711199012 Independent Escort Service Gurgaon
Call Girl Gurgaon Saloni 9711199012 Independent Escort Service GurgaonCall Girls Service Gurgaon
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Memriyagarg453
 
Dehradun Call Girls Service 8854095900 Real Russian Girls Looking Models
Dehradun Call Girls Service 8854095900 Real Russian Girls Looking ModelsDehradun Call Girls Service 8854095900 Real Russian Girls Looking Models
Dehradun Call Girls Service 8854095900 Real Russian Girls Looking Modelsindiancallgirl4rent
 
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...Niamh verma
 
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.ktanvi103
 
Hot Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
Hot  Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In ChandigarhHot  Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
Hot Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In ChandigarhVip call girls In Chandigarh
 
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetChandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meetpriyashah722354
 
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋Sheetaleventcompany
 
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 

Recently uploaded (20)

Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service LucknowCall Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
 
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
 
Bangalore call girl 👯‍♀️@ Simran Independent Call Girls in Bangalore GIUXUZ...
Bangalore call girl  👯‍♀️@ Simran Independent Call Girls in Bangalore  GIUXUZ...Bangalore call girl  👯‍♀️@ Simran Independent Call Girls in Bangalore  GIUXUZ...
Bangalore call girl 👯‍♀️@ Simran Independent Call Girls in Bangalore GIUXUZ...
 
Jalandhar Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
Jalandhar  Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...Jalandhar  Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
Jalandhar Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
 
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR Call G...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR   Call G...❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR   Call G...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Meghna Jaipur Call Girls Number CRTHNR Call G...
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptx
 
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
 
Call Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In Raipur
Call Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In RaipurCall Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In Raipur
Call Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In Raipur
 
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar SumanCall Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
 
Call Girl Gurgaon Saloni 9711199012 Independent Escort Service Gurgaon
Call Girl Gurgaon Saloni 9711199012 Independent Escort Service GurgaonCall Girl Gurgaon Saloni 9711199012 Independent Escort Service Gurgaon
Call Girl Gurgaon Saloni 9711199012 Independent Escort Service Gurgaon
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
 
Dehradun Call Girls Service 8854095900 Real Russian Girls Looking Models
Dehradun Call Girls Service 8854095900 Real Russian Girls Looking ModelsDehradun Call Girls Service 8854095900 Real Russian Girls Looking Models
Dehradun Call Girls Service 8854095900 Real Russian Girls Looking Models
 
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
 
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
 
Hot Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
Hot  Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In ChandigarhHot  Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
Hot Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
 
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service DehradunCall Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
 
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetChandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
 
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
 
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
 
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
 

Scavenging of waste anaesthetic gases.

  • 1. Theatre Pollution & Scavenger system in OR Dr. P K Maharana. KIMS, BHUBANESWAR, India.
  • 2. Definition An anaesthetic gas scavenging system is a system or device that collects excess anaesthetic gas from the breathing system and discharges it outside the working environment. • It is used to collect gases or aerosolized medications from the patient exhalations or found near the patient area because of some treatment related activities. Often associated with delivery of anaesthesia but may also include other patient related activities .
  • 3. Introduction When patients are administered anaesthetic gas, its molecules escape into the room since the use of gases and agents exceeds the amount necessary for the patient and can impact the performance of medical teams. Inhaled anesthetic agents include two different classes of chemicals: nitrous oxide and halogenated agents. Halogenated agents currently in use include halothane , enflurane, isoflurane, desflurane , and sevoflurane . Since the late 1960s there has been speculation that trace anaesthetic gases/vapours may have a harmful effect on operating theatre personnel, but it has been conclusively proved from currently available studies that there is no association between occupational exposure to trace levels of waste anaesthetic vapours in scavenged operating theatres and adverse health effects. Nitrous Oxide does some in high concentration. However, it is desirable to vent out the exhaled anaesthetic vapours and maintain a vapour-free theatre environment. At present, the Occupational Safety and Health Administration (OSHA) has no permissible exposure limits regulating these agents. In the Operating Room the Anaesthetic Gas Scavenging System collects and removes waste gases from the patient breathing circuit and the patient ventilation circuit. It can also collect any type of gases or aerosolized medications that is intended only for the patient should not be breath by others. AGSS systems to ensure the safety of your patients and staff.
  • 4. Waste anesthetic gases (WAGs). • Healthcare workers in a variety of settings( operating rooms, recovery area, labour rooms, even radiology suites) can be exposed to the anesthetic gases that are released or leak out during medical procedures. • Gases most commonly used include nitrous oxide, isoflurane, desflurane, and sevoflurane. • These gases and vapours are known as waste anesthetic gases (WAGs).
  • 5. Reasons for scavenging • On the basis of epidemiological studies carried out since the mid 1960s, it has been suggested that operating theatre staff and their wives and children are subject to an increased risk of various pathologies, including  spontaneous abortion, stillbirth, congenital abnormalities, low birth weight, altered sex ratio, infertility, cancer, liver disease and kidney disease. • Vessey and Nunn (1980) have critically reviewed both the epidemiological and the animal studies, and conclude that, while the evidence for most of the above hazards is unconvincing, the increased risk of abortion in female theatre staff is reasonably well substantiated. • Nitrous oxide inhibits the enzyme methionine synthase by inactivating its bound co-factor methyl cobalamin (a form of vitamin B12), probably causing impairment of deoxyribonucleic acid synthesis, which could well produce fetotoxic effects (Shareet al., 1983).
  • 6. Evidences of Adverse effects on Prolonged Exposed. • Nitrous Oxide: Several summaries of the epidemiologic studies of exposure to N2O and reviews of the topic generally including animal and retrospective studies (Purdham 1986; Kestenberg 1988; and NIOSH 1994) have been published. • They report a consistent excess of spontaneous abortion in exposed women. Evidence for congenital abnormalities is less strongly associated with exposure. • Halogenated Hydrocarbons : Moreover, there is biological plausibility that adds to the concern that high levels of unscavenged waste anesthetic gases present a potential for adverse neurological effects or reproductive risk to exposed workers or developmental anomalies in their offspring (Cohen et al. 1980; Rowland et al. 1992).
  • 7. Level of different agents which showed no significant adverse effects on animal studies. • 100 particles per million (ppm) for nitrous oxide • 50 ppm for enflurane • 50 ppm for isoflurane • 10 ppm for halothane • 20 ppm for sevoflurane (recommended by Abbot Laboratories) • No limit set for desflurane although a 50 ppm target is advisable due to its similarity to enflurane. These levels were chosen because they are well below the levels at which any significant adverse effects occurred in animals and represent levels at which there is no evidence to suggest human health would be affected.
  • 8. Recommended levels for different agents by different agencies. • In the United States, the maximum accepted concentrations of any Halogenated agent should be less than 2 ppm. • When such agents are used in combination with nitrous oxide, levels of less than 0.5 ppm should be achieved. • Nitrous oxide, (when used as the sole anaesthetic agent), at 8-hour time- weighted average concentrations should be less than 25 ppm during the administration of an anaesthetic. • Holland has a limit of 25 ppm for nitrous oxide, whereas Italy, Sweden, Norway and Denmark set 100 ppm as their limit for exposure to nitrous oxide.  It is not possible to set uniform levels without sufficient data.
  • 9. In 2000, OSHA revised its recommendations on waste anesthetic gases In 2000, OSHA revised its recommendations on waste anesthetic gases in the light of current knowledge. • The revised recommendations are published on the Internet for informational purposes only and are regularly updated as information becomes available. The document is not published in the standard OSHA manual on occupational hazards, however. • The recommendations are advisory and have not been promulgated as a standard; rather, they are to be seen as guidelines. 1. OSHA recommends scavenging of waste anesthetic gases in all anesthetizing locations and advocates work practices to reduce trace levels of anesthetic gases in the ambient air. 2. A documented maintenance program should be in place for all anesthetic delivery machines, and an ongoing education program for all personnel to inform them of these recommendations must exist. 3.OSHA recommends a program for monitoring trace anesthetic gases and also recommends a pre employment medical examination for all employees. 4. Each institution also should have a mechanism in place for employees to report any work-related health problems.
  • 10. Association of Anaesthetists of Great Britain and Ireland (1975) recommendation. • As a result of the increasing concern about the possible hazards of exposure to waste anaesthetic gases, the Association of Anaesthetists of Great Britain and Ireland (1975) has recommended that all anaesthetizing locations should be equipped with scavenging systems, and this is now official DHSS policy (Department of Health and Social Security, 1976). • NIOSH recommendation to OSHA: Workers should not be exposed to an eight hour time-weighted average of > 2 ppm halogenated agents (not > 0.5 ppm if nitrous oxide is in use) or > 25 ppm nitrous oxide.
  • 11. Target level for operating theatre pollution In most jurisdictions, there is a legal requirement to scavenge waste gases to maintain the level of waste gases in the Operating Room below the legally acceptable limit.  In the UK: the limits are typically  100ppm for nitrous oxide and 50ppm for halogenated volatile anaesthetic agents (except halothane which is 10ppm). In U.S.A NIOSH Recommendations published; 1977 by the (National Institute for Occupational Safety and Health), states that:- o Nitrous oxide maximum 25 ppm and o Halogenated volatile gases maximum 2 ppm. (Sharer and colleagues (1983) have criticized the NIOSH limit for nitrous oxide as being too restrictive, and have suggested a limit of 200 p.p.m., based on a consideration of the data on methionine synthase inactivation and fetotoxicity).
  • 12. Anaesthetic Gases & Agents in use. Name of Agents Year of Introduction Present Use Diethyl ether 1842 No Nitrous oxide 1844 yes Chloroform 1847 NO Cyclopropane 1933 NO Trichloroethylene 1934 No Fluroxene 1954 NO Halothane 1956 yes Methoxyflurane 1960 Infrequently Enflurane 1974 yes Isoflurane 1980 yes Desflurane 1992 yes Sevoflurane 1995 yes Nitrous Oxide and only few of the anaesthetic agents are in use. Most commonly used drugs are Sevoflurane, Isoflurane , Desflurane & Halothane
  • 13. Adverse effects  National Institute for Occupational Safety and Health ,2007( NIOSH) states that “exposure to high concentrations of waste anesthetic gases - even for a short time - may cause the following health effects: • Headache • Irritability • Fatigue • Nausea • Drowsiness • Difficulties with judgment and coordination • Liver and kidney disease” These health effects were mainly noted for older anesthetics (e.g., trichloroethylene, methoxyflurane) that are no longer commonly in use.  Studies are inconclusive on the potential health effects from occupational exposure to some of the newer anesthetics, such as isoflurane
  • 14. NIOSH studies • NIOSH (2007) continues: “Although some studies report no adverse health effects from long-term exposure to low concentrations of waste anesthetic gases, several studies have linked such exposure to miscarriages, genetic damage, and cancer among operating-room workers. • Studies have also reported miscarriages in the spouses of exposed workers and birth defects in their offspring.” • NIOSH (2015) later reports that “Some studies have documented adverse health effects (e.g., headaches, fatigue, irritability, birth defects, miscarriages, liver and kidney disease, cancer) from excessive exposure to anesthetic gas  Scavenger systems were forced on anesthesia machines by the ‘recommendations’ of the National Institute for Occupational Safety and Health (NIOSH). The recommendations for room air concentrations of waste anesthetic gases is 25 ppm for nitrous oxide and 2 ppm for halogenated agents.
  • 15. Who is exposed to waste anesthetic gases? The following hospital workers may be exposed to waste anesthetic gases: • Anesthesiologists • Dentists • Nurse anesthetists • Operating-room nurses • Operating-room technicians • Other operating-room personnel • Recovery-room nurses • Other recovery-room personnel • Surgeons
  • 16. Objective  The objective of Anaesthetic gas scavenging systems(AGSS) to prevent medical staff from inhaling the anaesthetic nitrous oxide and halogenated hydrocarbons administered to patients during surgery. In most jurisdictions, there is a legal requirement to scavenge waste gases to maintain the level of waste gases in the Operating Room below the legally acceptable limit. In addition to the legal requirement there is an Occupational Health requirement to maintain a safe workplace and limit exposure to potentially harmful gases and agents.
  • 17. Guidelines for controlling pollution in OR • There is no association between occupational exposure to anaesthetic agents trace levels and adverse health effects. • There are no agreed international standards of the maximum accepted concentrations of agents in the theatre environment. • Routine monitoring and testing (PPM) are mandatory.
  • 18. Factors responsible for theatre pollution Anaesthesia Techniques Anaesthesia Machines • Leaks from various connections like tubing connections if not fitting properly, soda lime canister etc. • Others: Cryosurgery Unit, Cardiopulmonary bypass circuit if vapor is used. Poorly fitting facemask Pediatric breathing system( T-piece) Un-cuffed tracheal Tubes Gases coming out through APL valve Gases coming through ventilator exhaust Exhalation gases during recovery Spillage during filling of vaporizers
  • 19. How to reduce pollution in OR?  1. Adequate theatre ventilation and air conditioning, with frequent and rapid changing of the circulating air (15–20 times per hour).  Theatres that are unventilated are four times as contaminated with anaesthetic gases and vapours compared to those with proper ventilation.  A non-recirculating ventilation system is usually used. A recirculating ventilation system is not recommended. In labour wards, where anaesthetic agents including Entonox are used, rooms should be well ventilated with a minimum of five air changes per hour.  2. Use of the circle breathing system.  This system recycles the exhaled anaesthetic vapours, absorbing CO2.  It requires a very low fresh gas flow, so reducing the amount of inhalational agents used.  3. Total intravenous anaesthesia.  4. Regional anaesthesia.  5. Avoiding spillage and using fume cupboards during vaporizer filling.. Modern vaporizers use special agent-specific filling devices as a safety feature and to reduce spillage and pollution.  Good technique will also help lessen exposure.  6. Scavenging of Anaesthetic gases.
  • 20. Good technique will also help lessen exposure. • Good mask fit • Avoid unscavengeable techniques if possible (insufflation) • Prevent flow from breathing system into room air (only turn on agent and nitrous oxide after mask is on face, turn them off before suctioning) • Washout anesthetics (into the breathing circuit) at the end of the anesthetic • Don’t spill liquid agent • Use low flows • Use cuffed tracheal tubes when possible • Check the machine regularly for leaks • Disconnect nitrous oxide pipeline connection at wall at the day’s end (beginning?) • Total intravenous anesthesia
  • 21. Scavenging of waste Gases • In any location in which inhalation anaesthetics are administered, there should be an adequate and reliable system for scavenging the waste anaesthetic gases. • A scavenging system need to be capable of collecting the waste anaesthetic gases from the breathing system and disposing them safely. Unscavenged operating theatres can show a very high levels of N2O. (400–3000 ppm). Scavenging of waste gases may be a passive or active system.
  • 22. A passive scavenging system. • A passive scavenging system operates without the use of suction, since the positive pressure of gas in the breathing circuit pushes waste anesthetic gases into the scavenging system. • One-way valves in the interface help to move waste anesthetic gases outdoors or into a non-recirculating air ventilation system. • A passive scavenging system does not involve the use of a vacuum pump or suction.
  • 23. Active scavenging systems • Active scavenging systems use suction, or a vacuum pump, to actively remove waste anesthetic gases from the breathing circuit and draw these gases into a scavenging system. • The interface of an active system requires a negative pressure relief valve in order to prevent negative pressure from reaching the breathing circuit and affecting the patient’s lungs.
  • 24. Comparison of both the systems  Active System • Uses a device like suction to draw gases from breathing system. • Mostly uses a compressor to draw the gases and agents. • Expensive • Requires maintenance. • Requires a means to protect the patient's airway from the application of suction, or buildup of positive pressure. Passive System • This system uses no suction, pressure in the gas line helps to drive out the gases out of the machine. • It must be located adjacent to an outside place, the pipe simply passes through a whole in the wall. • It is inexpensive • Dose not require maintenance. • Requires the patient to be protected from a positive pressure buildup only if there is obstruction in the exit limb.
  • 25. The basic functional components of an Anaesthetic Gas Scavenging System are as follows: • 1.A collecting assembly / shroud with a relief valve by which the waste gas leaves the breathing or ventilation circuit. • 2.A transfer system of tubing to conduct waste gases to the Scavenging Interface. • 3.The Scavenging Interface, and • 4.A Disposal line to conduct the waste gas to a passive evacuation system, or a Waste Anaesthetic Gas Disposal/Medical Vacuum system via a station outlet.
  • 26. Schematic diagram showing the different components of scavenging system 1.A collecting assembly / shroud with a relief valve by which the waste gas leaves the breathing or ventilation circuit. 2.A transfer system of tubing to conduct waste gases to the Scavenging Interface. 3.The Scavenging Interface, and 4.A Disposal line to conduct the waste gas to a passive evacuation system,
  • 27. A collecting Assembly( Receiving system) Gas collection assembly, (tubes connected to APL and vent relief valve)  A reservoir bag can be used receiving or collecting system .  Two spring-loaded valves in the system guard against excessive positive (1000 Pa) in case of a distal obstruction or negative (–50 Pa) pressures in case of increased demand in the scavenging system. Without these valves, excessive positive pressure increases the risk of barotrauma, should there be an obstruction beyond the receiving system. Excessive negative pressure could lead to the collapse of the reservoir bag of the breathing system and the risk of rebreathing.
  • 29. Transfer System It is a simple wide bore pipe without any leak which can transfer from the collecting system to Scavenging interface. Can be copper pipe. Transfer tubing (19 or 30 mm, sometimes yellow color- coded) The transfer hose should always be fitted with a pressure relief valve (10cmH2O) such that if the transfer hose were to become occluded for any reason the patient would still have an expiratory path.
  • 30. Scavenging Interface • The scavenger interface is the most important component. • The extraction flow rate ensures that waste anaesthetic gases are adequately removed from the system. It protects the breathing circuit from excess positive or negative pressure. • Positive-pressure relief valve is mandatory to vent excess gas in case of occlusion distal to a closed interface. If active disposal system in use, must have negative pressure relief valve as well. • Reservoir highly desirable with active systems. • The induced flow rate should be as low as possible with the system ideally being passive between the patient and the receiving unit and active from the receiving unit to the exhaust point. • The importance of this flow rate is that if it falls too low it may not be sufficient for waste gas removal. • If too high, this would lead to the waste gases spilling out from the base of the receiving reservoir unit into the immediate working environment; it may lead to an increase of the induced flow at the patient connection port.
  • 31. Closed interface Closed interface communicates with atmosphere only through valves. Should adjust vacuum pressor so that reservoir bag neither flattened not over- distended.
  • 32. Active Scavenging Interface closed system. The Ohio scavenging interface has connections for the outlets from the breathing system and ventilator (A), one or two reservoir bags (B), and the vacuum line (V). The suction is controlled by the needle valve (N). There are both positive (P1) and negative (P2) pressure relief valves in case the reservoir bag becomes empty or too full.
  • 33. Function of the Positive and Negative Pressure Relief Valves The scavenging system has negative and positive pressure relief valves to prevent excessively high or low pressures from developing in the scavenging manifold and being transmitted to the patient via the breathing circuit. The positive pressure relief valve: will lift up and open at approximately 5 cm H2O. This happens when the vacuum is too weak or if the vacuum adjustment valve is completely closed. A slight hissing or rattling sound can be heard (indicating the venting of exhaust gases into the room) during the expiratory phase of mechanical ventilation or when the manual bag is squeezed. In addition, the scavenging bag will be fully inflated. Negative pressure relief valve: to prevent negative pressure from being transmitted to the breathing circuit, will actuate at approximately -0.25 cm H2O and allow room air to enter the scavenging system to "break" the vacuum. The scavenging bag will be collapsed flat.
  • 34. Open interface Open interface has no valves, and is open to atmosphere (allows both negative and positive pressure relief). Should be used only with active systems. Remember that slight hissing from an open interface is normal- there is no audible indication of waste gas leaks. Safer for the patient (no hazard of positive or negative pressure being applied to the airway as a result of scavenger failure). The risk of occupational exposure for providers ignorant of their proper use is higher with the open interface (Anesth Analg 1992;75:1073).
  • 35. Home made Simple active scavenging interface A cylindrical plastic jar. Suction through valve attached at the bottom. At the top there are two inlets, one for waste gas pipe insertion the other one for air to enter if creates excessive negative pressure.
  • 36. The disposal system  The disposal system is a wide-bore copper pipe leading to the atmosphere directly or via the theatre ventilation system. Gas disposal assembly (active or passive, active system most common uses the hospital suction system)
  • 37. Passive disposal system • In the past, the standard of collecting waste gases has been through passive methods, which include (depending on the flow rate of the delivery gas) pushing the waste gas into an activated charcoal canister.
  • 38. Activated charcoal cartridges. • Activated charcoal cartridges can also be used for passive scavenging. • It is important to note the limitations of these cartridges: they are less effective at high anesthesia flow rates, they must be changed every 12 hours, and care must be taken to avoid occluding the cartridge’s air exit holes. • When used correctly, however, activated charcoal cartridges can effectively decrease waste anesthetic gas exposure.
  • 39. F/air Canister A sensible answer to anesthesia gas problems in the operating room, the F/AIR anesthesia gas filter was specifically designed to remove waste anesthesia gases such as ISOFLURANE, HALOTHANE, ENFLURANE, SEVOFLURANE, etc. from the operating room environment. Economical to use, with an average useful life of 12 to 15 hours, the F/AIR is easily adaptable to anesthesia machines. The F/AIR can remove no less than 50 grams of pure halogenated anesthetic gases and can then be conveniently discarded. The F-AirÂŽ canister must be changed after 8 hrs of normal use or a weight gain of 50 grams. Disposable Ready to use
  • 40. WAG Activated Charcoal Canisters Activated Charcoal Filtration System Contains 2 pounds of the highest quality activated charcoal available to make the WAG the most effective filter on the market. Will adsorb a minimum of 200 grams Halogenated Agent. (4 times the amount of F- Air Canisters). Unique design traps Isoflurane/Sevoflurane Agent molecules in the canister. Relief holes on top ensure that the canister can sit upright on a flat surface. Universal tubing connections: 19mm (standard in industry), 22mm, and fitting for 15mm male to use ¼” ID Tubing. Passed independent testing according to OSHA regulations of 2ppm. No breakthrough exhibited at +/- 60 LPM flow rate (Verified via infra red spectrophotometer).
  • 41. Drawing showing the flow system of gases.
  • 42. Diagram of a passive scavenging system. Modern scavenging system has four components. a). Collecting System: b).Transferring System:. c). Receiving System: d). Gas disposal Assembly: May be a passive or active disposal unit.
  • 43. Passive System The passive system is simple to construct with zero running cost. Components are :- 1. The collecting and transfer system which consists of a shroud connected to the adjustable pressure limiting (APL) valve (or expiratory valve of the ventilator). 2. A 30-mm connector attached to transfer tubing leads to a receiving system. 3. Receiving System. 4. Gas Disposal Unit or assembly. The 30-mm wide-bore connector is designed as a safety measure in order to prevent accidental misconnection to other ports of the breathing system .
  • 44. Draw backs of passive scavenging system • Scavenging systems are either passive or active. • Passive systems have a large tube or canister with an inlet, an outlet, and one open end. • The waste anesthetic gases from the ventilator and ‘pop-off’ valve enter the inlet. • Suction is applied at the outlet. • If the suction rate exceeds the rate of entry of anesthetic gases, room air is drawn into the open end. • If entry of anesthetic gases exceeds the rate of suction, excess gases exit via the open end into the room.
  • 45. Active system • The purpose of the system from the plant to the terminal outlet is to provide the extraction flow rates to remove waste anaesthetic gases from the receiving unit reservoir. • The role of the receiving unit is to provide a safe interface between the patient and the extraction flow rates. Components • The collecting and transfer system is similar to that of the passive system . • A fan or Vacuum unit to establish flow.
  • 46. Adverse effects of active system. • Active systems contain negative and positive relief valves to release excess pressures. • Active systems have been associated with higher exposure of gases in the environment than passive systems. • When the reservoir bag overinflates due to an excess inflow of anesthetic gases, pressure increases until the positive pressure relief valve opens to vent excessive gases into the room. • A valve malfunction will allow the pressure increase in the patient circuit to increase to the limit of the elastic properties of the particular reservoir bag. • High pressures can be detected in the patient circuit by the circuit pressure gauge and/or by the high pressure gauge of the ventilator. • If high pressure persists the patient may develop decreased cardiac output secondary to high intrathoracic pressure, and/or pneumothorax.
  • 47. Adverse effects of active system(Conti-) • When the suction rate exceeds the anesthetic gas inflow rate, the reservoir bag deflates. • Usually any excessive negative pressure is offset by room air entering the negative pressure relief valve. • Some scavenger systems have a second or backup relief valve. • The pressure gauge of the ventilator may detect an excessive negative pressure state. • If the negative pressure relief valve system malfunctions, it could be possible for the suction to remove anesthetic gases from the patient circuit.
  • 48. Mechanism of action • 1. The exhaled gases are driven by either the patient’s respiratory efforts or the ventilator. • 2. The receiving system should be mounted on the anaesthetic machine to minimize the length of transfer tubing, therefore minimizing resistance to flow. • Problems in practice and safety features • 1. Connecting the scavenging system to the exit grille of the theatre ventilation is possible. Recirculation or reversing of the flow is a problem in this situation. • 2. Excess positive or negative pressures caused by the wind at the outlet might affect the performance and even reverse the flow. • 3. The outlet should be fitted with a wire mesh to protect against insects. • 4. Compressing or occluding the passive hose may lead to the escape of gases/vapours into the operating theatre and thereby polluting it. • The disposal hose should be made of non-compressible materials and not placed on the floor.
  • 49. Requirements of standard Scavenging System The draft BS( British Standard) on scavenging equipment includes the following provisions for patient safety: (1) At a continuous flow of air 30 litre min"1, the scavenging system should not impose a greater resistance than 50 Pa. (2) In the event of total obstruction of the system, or failure of the power supply, the resistance should not be greater than 1 kPa at a flow of 30 litremin"1. If the transfer tubing is less than 1 m in length, the necessary pressure relief may be afforded by the receiving system. (3) The scavenging system shall not create a sub-atmospheric pressure in the collecting system of more than 50 Pa.
  • 50. Monitoring the functioning of Scavenging system.  Sampling procedures for evaluating waste anaesthetic vapour concentrations in air should be conducted for nitrous oxide and halogenated agents on a yearly basis in the UK and on a quarterly basis in the USA in each location where anaesthesia is administered.  Monitoring should include: • Leak testing of equipment • Sampling air in the theatre personnel breathing zone. Planned preventative maintenance (PPM) programme. Anaesthetic equipment, gas scavenging, gas supply, flowmeters and ventilation systems must be subject to a maintenance programme. , The general ventilation system and the scavenging equipment should be examined and tested by a responsible person, at least once annually .
  • 51. WHAT IS AN IDEAL SCAVENGING SYSTEM? . 1.Should not affect the ventilation and oxygenation of the patient. 2.Should not affect the dynamics of the breathing system. 3.A well-designed scavenging system should consist of i. a collecting device for gases from the breathing system or ventilator at the site of overflow, ii. a system to carry waste anaesthetic gases from the collecting device and a method for limiting both positive and negative pressure variations in the breathing system. 4.The performance of the scavenging system should be part of the anaesthetic machine check. 5.Scavenging systems can be of a passive or active type.
  • 52. WHAT IS AN IDEAL SCAVENGING SYSTEM (CONT-)? 6.The scavenger system must be independent of the main hospital ventilation system. 7. The scavenging system should be kept in good repair to prevent leaks using a maintenance and inspection program as listed below. 8.Gases are not discharged near the air intake of the building or surrounding buildings. 9. In the event of a “Code RED”, hospital ventilation is shut down to reduce the potential spread of a fire but the scavenger system must continue to work. Consider using a WAG collection canister that can be attached to an anesthetic gas machine before the scavenger. It will capture the agents (except for nitrous oxide).This technology reduces emissions to the atmosphere by capturing the agents and recycling them.
  • 53. An effective waste anesthetic gas management program includes: • Engineering Controls, • Work Practices, • Air Monitoring and, • Hazard Communication and Training.
  • 54. Air Monitoring • One of the tools used to measure exposure to waste anesthetic gases is air monitoring. • Monitoring may be continuous or periodic but should adequately measure exposure in the work areas and surrounding areas. • Monitoring can aid in identifying the presence and location of leaked gases and the effectiveness of corrective measures. • As most halogenated anesthetic gases cannot be detected by smell (unless they are in high concentration) proper monitoring becomes all the more critical. • Nitrous oxide is an odourless and colourless gas and can only be detected by WAG monitor.
  • 55. Hazard Communication and Training • Employers should develop and implement a written hazard communication program regarding WAGs that includes description of the physical and health hazards of anesthetic agents in use, the compiling and availability of up to date material safety data sheets on all anesthetic gases used; proper labelling of canisters, tanks, and containers; and a comprehensive employee training and information program. • The training program should list steps, the workers can take to protect themselves from the hazards of WAGs. • The program should include information on steps taken by the employer; - such as engineering controls, - clearly outline emergency procedures to contain spills, - describe safe work practices and the use of any personal protective equipment, and - detail the use of continuous monitoring devices. • The training program should clearly outline all methods and observable indicators that can detect the presence and release of anesthetic gases. • Spills should be treated as emergencies. Spills of anesthetic agents must only be cleaned up and controlled by properly trained and equipped personnel
  • 56. Good technique that lessens expose to waste anaesthetic gases. • Good mask fit • Avoid unscavengeable techniques if possible (insufflation) • Prevent flow from breathing system into room air (only turn on agent and nitrous oxide after mask is on face, turn them off before suctioning) • Washout anesthetics (into the breathing circuit) at the end of the anesthetic • Don’t spill liquid agent • Use low flows • Use cuffed tracheal tubes when possible • Check the machine regularly for leaks • Disconnect nitrous oxide pipeline connection at wall at the day’s end (beginning?) • Total intravenous anesthesia
  • 57. Certain devices & systems for an ideal scavenging system.
  • 58.
  • 60. Separate Suction System for Anaesthesia gas scavenging
  • 61. Exhaust Fan Use to Scavenge AGSS
  • 63. Connection to be fitted to APL valve