Image Guided Radiation
Therapy
Moderator: Prof. Bhavana Rai
Dr. Namrata Das
1st year Junior Resident
12-10-2018
OUTLINE
• Introduction
- Evolution of radiotherapy techniques
- ICRU definitions
• Types of Error
• Timeline when errors are addressed by image guidance
- Positioning and Immobilization
- Image Acquisition and Registration: 4DCT
- Target Delineation: advanced imaging modalities
- Patient Setup confirmation: Types of in room imaging
• Offline, online correction strategies
- No Action Level (NAL), Shrinking Action Level (SAL)
• Clinical application of IGRT:
- Head and neck cancer
- Lung cancer
- Prostate cancer
- Upper GI malignancy: Liver
• Summary
12-10-2018
12-10-2018
Introduction
The purpose of radiotherapy is to maximize tumour kill and minimize normal
tissue damage by delivering a precisely measured dose uniformly to the target
tissue.
• Radiotherapy treatment planning has always used imaging.
• History of development of radiotherapy has closely followed the development of new
imaging techniques.
12-10-2018
Evolution of Radiotherapy Treatment Techniques
2D Planning
3D Planning/
Conformal
Radiotherapy
X Ray Based:
Discovery of X Ray and
subsequent boom during the
World War
Volumetric Imaging:
1970s: Computed
Tomography
1980s: Magnetic
Resonance Imaging
3D CRT: geometrical field shaping
IMRT: geometrical field shaping+
intensity modulation
IGRT: geometrical field shaping +
intensity modulation + image
guidance
12-10-2018
2D/ Conventional
Planning
3D/
Conformal
Planning
Drawbacks
• Inability to visualize tumor or normal
tissue
• Only square/rectangle fields
• Large safety margins resulting in high dose
to normal tissue
ICRU 29 (1978): Dose specification for the reporting of external beam
radiotherapy with photons and electrons
12-10-2018
Evolution of the concept of treatment volumes
Target
volume
Treated Volume
Irradiated Volume
Irradiated Volume: volume which receives an absorbed
dose considered to be significant in relation to tissue
tolerance
Target Volume: Tissues that are to be irradiated to a
specific dose
Treatment volume: Volume enclosed by a relevant
isodose surface that is selected and specified by the
radiation oncologist as being appropriate to achieve the
purpose of treatment
12-10-2018
ICRU 50 (1993): Prescribing, Recording, Reporting Photon Beam
Therapy
Treated Volume
Irradiated Volume
PTV
CTV
GTV
Planning Treatment Volume: Geometrical concept defined to
select appropriate beam sizes, beam arrangements taking
into consideration the net effect of all possible geometrical
variations, inaccuracies in order to ensure prescription done
is actually absorbed in the CTV.
Clinical Target Volume: Tissue volume that contains GTV and/or
microscopic malignant disease which is to be eliminated – has
to be treated respecting anatomical boundaries.
Gross Tumour Volume: Gross palpable or visible or
demonstratable extent and localization of tumour
12-10-2018
ICRU 62 (1999): Supplement to Rep. 50
Planning Treatment Volume (PTV) = Internal Target Volume (ITV) + Setup Margin
(SM)
ITV = CTV+ Internal Margin (IM)
Treated Volume
Irradiated Volume
PTV
CTV
GTV
ITV
Internal Margin (IM): A margin that must be added to the
CTV to compensate for the expected physiological
movements and the variations in size, shape and position of
the CTV during therapy in relation to the internal reference
point.
Internal Target Volume (ITV): CTV +IM
Setup Margin (SM): It is the margin that must be added to
account specifically for uncertainties (inaccuracies and lack of
reproducibility) in patient positioning and alignment of the
therapeutic beams during treatment planning and through all
treatment sessions.
Conformal Radiotherapy: 3D CRT
3 Dimensional Conformal Radiotherapy
(3D CRT)
It is the use of 3 dimensional
anatomical information to plan and
deliver treatment so that the
resultant dose distribution conforms
as closely as possible to the target
volume in 3 dimensions with
minimum dose to the surrounding
normal tissue.
Advantage over 2D:
• increased tumour dose, reduced
normal tissue dose
• Reduced safety margin
Drawback:
• Forward planning: trial and error
method
• Increased dose spillage
• Ability to sculpt tumour limited
12-10-2018
Transition from 2-D Radiotherapy to 3-D Conformal and Intensity Modulated Radiotherapy. IAEA-TECDOC-1588. May 2008.
Intensity Modulated Radiotherapy is
a special form of 3D conformal dose
delivery, enhanced by generating a
non-uniform photon fluence within
each beam, calculated by an inverse
treatment planning process
designed to meet specified
dosimetric objectives.
Advantage over 3D CRT:
• Able to achieve concave dose
distribution
• Allows dose painting:
inhomogeneous dose prescription
to different targets
• Reduced margins
• Inverse planning
Drawback: Chance of missing target
higher due to reduced margins at the
time of treatment
12-10-2018
Conformal Radiotherapy: IMRT
• Uses conformal uniform fields
leading to convex distribution
• By varying the fluence of beams,
more fluence can be delivered to
rays reaching target as compared
to normal tissue. Allows concave
dose distribution
12-10-2018
Need for Image Guided Radiotherapy
The uncertainty in target localization while delivering highly conformal
radiotherapy is specially important for:
• Tumours prone to intrafraction motion: lung cancer
• Tumours prone to interfraction motion: prostate cancer
• Tumours prone to deformation: head and neck cancers
• Treatment delivery by IMRT, SBRT (high dose per fraction with narrow
margins)
Thus there is a need for image guidance at various stages of the treatment
process.
12-10-2018
Definition
• Multiple definitions - some too extensive, others too narrow
• ACR, ASTRO practice guidelines,
“a procedure that refines the delivery of therapeutic radiation by
applying image based targeted re-localization to allow proper patient
repositioning for the purpose of ensuring accurate treatment and
minimizing the volume of normal tissue exposed to radiation.”
• IGRT Committee, RTOG, 2001,
“IGRT refers broadly to treatment delivery using modern imaging
methods such as CT, MRI, PET and USG in target and non target
structures for RT definition, design and delivery”.
12-10-2018
IGRT can be defined as the use of:
• Advanced imaging modalities, especially those incorporating
functional and/or biological information, to augment target and
normal tissue delineation
• In-room imaging to adjust for target motion or positional
uncertainty (interfractional and intrafractional), and, potentially,
to adapt treatment to tumour response
Image Guided Radiation Therapy. A clinical perspective. Arno Mundt, John Roeske.
12-10-2018
Use of imaging for:
• Target and normal tissue delineation
• Margin design and treatment planning
• Setup adjustments and delivery
12-10-2018
Use of advanced imaging
modality for accurate
delineation of target
Use of imaging for setup
verification and adjustment
Errors
• Error is any mismatch between the true anatomy and pathology
during treatment and planning
12-10-2018
Types of Errors
• Deviation that varies in direction
and magnitude for each fraction
• Cumulative dose distribution gets
blurred with random errors
• Direction and magnitude
remains same for each fraction
RANDOM ERRORS SYSTEMATIC ERRORS
12-10-2018
Planned
Isocentre
Y
X
Systematic
error
Random
errors
12-10-2018
Systematic Error
External
Wrong isocentre shift;
discordance between planning
CT and LINAC laser systems
Solution:
On Board Imaging
Internal
Physical changes since planning
CT: drainage of pyometra
Shrinkage of tumour
Solution:
Adaptive replanning
Systematic errors
A. CHOICE OF IMAGING
All tumours and organs cannot be
identified using a planning CT
12-10-2018
B. IMAGE FUSION
Fusing two or more image sets for
delineation can lead to some mismatch –
leading to suboptimal delineation
12-10-2018
C. IMAGE DISTORTION
Planning images can have artefacts due to
motion, foreign objects or matching
related imaging artefacts
1. Movement induced artefact: Fast free
breathing CT Scans produce these errors
C. IMAGE DISTORTION
Planning images can have artefacts due to motion, foreign objects
or matching related imaging artefacts
Prosthetics, foreign metallic/plastic objects can
distort images
Systematic errors
12-10-2018
Systematic errors
Anatomical changes due to:
• Tumour regression: the shrink in tumour will lead
to a change in the dose distribution of OARs (lung,
head and neck)
• Weight loss: changes the anatomic contour of
patients
• Resolving atelectasis: Opening up of blocked
airways
12-10-2018
Random Error
External
Daily setup variation
On board imaging
Internal
Daily position change due to
bowel, bladder filling
Respiratory Motion
OBI
Generation of individualized
ITV margin
Respiratory gating/tracking
Random Error: organ motion and deformation
Seconds:
respiration,
heartbeat
Minutes: Peristaltic
motion, rectal
motion
Days: Radiation
induced changes in
mucosa, bladder,
bowel distension
Weeks: changes in organ
shape and position
12-10-2018
Random Error: organ motion in the pelvis
• Rectal filling and bowel filling changes: inter-fraction variability of
position and rotation
12-10-2018
• Sites affected by respiratory motion:
Lung and mediastinum, breast and
chest wall, upper abdomen (liver,
biliary tract, stomach, pancreas)
• Motion of lung tumours:
- Supero-inferior (longitudinal)
direction: maximum movement
- Lower lobe tumours: greater
movement in SI direction
- Anterior tumours: greater AP motion
- Central tumours have less motion
than peripheral
12-10-2018
Random Error: respiratory motion
Highly variable motion range:
• Lung: SI = 3.9 cm, AP = 2.4cm, LR = 2.4cm
• Liver: Normal breathing: 10-25 mm, deep
breathing = 37-55 mm
Timeline when errors can be addressed
• Positioning and Immobilization
• Image Acquisition and Registration
• Target Delineation
• Treatment Planning
• Patient specific dosimetric verification of treatment plan
• Patient Setup confirmation
• Treatment Delivery
• Quality assurance
12-10-2018
Immobilization in IGRT
Rationale:
• Reduces interfraction and intrafraction variation in position
• Reduces intrafraction motion of anatomic region or specific targets
• Reduces dependence on frequent imaging
• Reduce setup and imaging time
• External or internal
12-10-2018
Immobilization aids
12-10-2018
BRAIN HEAD AND NECK
- 3D CRT – thermoplastic mask
- Stereotactic treatment: Rigid frames
(<=1mm)/non-invasive and frameless systems with image
guidance (<=2mm)
Thermoplastic mask – 3-5mm
setup errors in the region of the
head and face
12-10-2018
Immobilization aids
PROSTATE THORAX AND
UPPER ABDOMEN
Timeline when errors can be addressed
•Positioning and Immobilization
•Image Acquisition and Registration
•Target Delineation
•Treatment Planning
•Setup confirmation
•Treatment Delivery
•Quality assurance
12-10-2018
4D CT Simulation
In 4D CT, a series of CT scans is reconstructed at different phases of the
breathing cycle.
The three components of a normal CT simulator are:
(1) A large bore (75-85cm) to accommodate various treatment positions along
with treatment accessories with a flat couch insert to simulate treatment
machine couch.
(2) An integrated laser marking system
(3) Virtual simulation and visualization
12-10-2018
12-10-2018
INFRA RED CAMERA PLASTIC MARKER
WITH INFRA RED
REFLECTORS
4D CT: Real Time Position Management (RPM)
Steps:
• Patient trained to breath regularly
• Setup/immobilize the patient in the
treatment position
• Infrared red reflector placed over
diaphragm
• Images acquired for multiple phases
of respiration, about 8-25 datasets at
each couch position.
• System sorts the images into different
3D image sets (binned) with the help
of respiratory signal
12-10-2018
CT Scan
Axial scan trigger,
1st couch position
Axial scan trigger,
2nd couch position
Exhalation
Inhalation
Scan Scan Scan
Axial scan trigger,
3rd couch position
Prospective CT Image Acquisition
X-ray on
Exhalation
Inhalation
1st couch
position
2nd couch
position
3rd couch
position
“Image acquired”
signal to RPM
system
(Ford 2003, Vedam 2003)
Retrospective 4D CT Image Acquisition
Limitations of 4D CT
• Depends upon uniform patient respiratory pattern
• The range of motion may not be similar during treatment so image
guidance during treatment delivery is important
12-10-2018
Derivation of ITV
• Population Based: may over or underestimate motion.
Not ideal.
• Individual Based:
1. Combining Breath hold scans
2. Combining 4DCT images
3. MIP: Maximum intensity projection
12-10-2018
12-10-2018
PTV
CTV
GTV
ITV
ITV = CTV + IM
PTV = ITV + SM
1. ITV: Combining 4DCT images
Target is contoured in every phase
of respiration and then fused to
generate the ITV.
Advantage: Captures the entire
trajectory of motion
Disadvantage: Time consuming
12-10-2018
2. ITV: MIP
Maximum Intensity Projection (MIP):
MIP is an image set generated which
includes all the possible locations of a
tumour.
Maximum range of motion of the
hyperdense tumour is considered and
images are stacked together to yield
the MIP.
Advantage: accurate and automatic
method of generating the ITV
Disadvantage: beneficial only for lung
tumours surrounded by normal
parenchyma (not those in contact with
mediastinum, chest wall, diaphragm)
12-10-2018
Timeline
•Positioning and Immobilization
•Image Acquisition and Registration
•Target Delineation
•Treatment Planning
•Setup confirmation
•Treatment Delivery
•Quality assurance
12-10-2018
Tissue delineation using advanced imaging
• Image fusion for radiotherapy is very important to correctly define
tumour and normal tissue
• Necessary to wrap high doses around tumour
• CT image essential for dose calculation
• However, target delineation on CT not accurate for all sites
• To avoid systematic errors from inaccurate delineation, fusion of
other modalities important
12-10-2018
Different imaging modalities
12-10-2018
3 steps of image fusion
• Registration: Computes the difference between images
• Fusion: Fuses the images with the calculated differences
• Validation: Check if images are matching well
12-10-2018
PET-CT fusion
In lung cancers, PET-CT fusion helps
in:
• Primary tumour delineation
• Nodal volume
Incorporation of FDG PET in CT based
planning:
• Increase or decrease in target
volume (usually decrease)
• Reduces interobserver variability
• Distinguish between collapse and
consolidation
12-10-2018
Timeline
•Positioning and Immobilization
•Image Acquisition and Registration
•Target Delineation
•Treatment Planning
•Setup confirmation
•Treatment Delivery
•Quality assurance
12-10-2018
Setup Error
It is the difference in position of the target (or surrogate bony
anatomy) between planning and treatment after a treatment setup
based on external setup markers.
• Skin marks/tattoos are the set of coordinates based on which
patients are initially et up.
• Treatment verification imaging using internal boy anatomy
(2D verification) or soft tissue target (3D verification) does not
exactly match planning images (DRRs or CT).
• This difference in position is called setup error
12-10-2018
Causes of setup error
• Technical/QA
- Laser misalignment, isocentre alignment not coinciding
• Initial setup
-isocentre shifts
• Positioning and Immobilization
- Imperfect arms and legs can change position of external skin marks
- Suboptimal immobilization may lead to changes in between setup and
verification/treatment
• Daily Setup
-positioning
-suboptimal marking of external markers
12-10-2018
6 axes of motion
Motion and errors are described in 6 axes, 3 translational and 3
rotational
THREE TRANSLATIONAL:
lateral, longitudinal,
vertical
THREE ROTATIONAL:
pitch, roll, yaw
12-10-2018
Imaging Techniques
USG Based
BAT
- SonArray
- I-Beam
- Restitu
Video Based
- AlignRT
Planar Imaging
Based
KV Xray OBI
(On Board
Imaging)
Gantry
Mounted
- Varian OBI
- Elekta
Synergy
-IRIS
Stereoscopic/
Room
Mounted
- Cyberknife
- RTRT
-Extractrac
System
(BrainLab)
MV X Ray
Imaging
EPID
(Electronic
Portal Imaging
Device)
Volumetric
imaging based
Fan Beam
- Helical
Tomotherapy
- In room CT
Cone Beam
KVCBCT
- Varian OBI
- Elekta
MVCBCT
Siemens
12-10-2018
Imaging Techniques
USG Based
BAT
- SonArray
- I-Beam
- Restitu
Video Based
- AlignRT
Planar Imaging
Based
KV Xray OBI
(On Board
Imaging)
Gantry
Mounted
- Varian OBI
- Elekta
Synergy
-IRIS
Stereoscopic/
Room
Mounted
- Cyberknife
- RTRT
-Extractrac
System
(BrainLab)
MV X Ray
Imaging
EPID
(Electronic
Portal Imaging
Device)
Volumetric
imaging based
Fan Beam
- Helical
Tomotherapy
- In room CT
Cone Beam
KVCBCT
- Varian OBI
- Elekta
MVCBCT
Siemens
12-10-2018
2D imaging: Gantry mounted
Portal Imaging
• Portal imaging is a method to verify that the area planned by a
radiation field (or portal) matches with the area treated on a patient.
• Aim: To match bony anatomy covered in plan with that of the covered
by portal curing treatment
• Earliest form of image guidance
12-10-2018
Steps of obtaining a portal radiograph
Film in a cassette holder (with phosphor screen) placed
across patient
MV treatment field (portal) exposed
Film manually taken to a developing room and
developed image compared with planning radiograph
12-10-2018
Portal imaging
Advantages
• Inexpensive
• First method of verification
Limitations
• Time consuming, cumbersome
• Poor quality (megavoltage)
• Setup data could not be stored
Digital (electronic) imaging solved most of these drawbacks.
12-10-2018
Electronic portal Imaging (EPI)
EPI uses a detector, usually attached to the gantry opposite the
treatment beam, to produce high quality digital images rapidly.
- MV treatment beam used
- Allows fast identification of setup errors and online correction
12-10-2018
EPIDs on Linear Accelerators
• IGRT EPID is a flat panel detector
mounted on retractable arms
opposite the treatment head
• Arms often capable of
movement to allow them to
place detector at variable
distance from source
EPID
12-10-2018
Amorphous silicon based EPIDs
• Tiles of photodiodes arranged on
glass substrate on a flat panel
array
• Incident photons pass through a
phosphor and emit light
• Photodiode captures light, emits
electronic signal
Three types of commercial electronic portal imaging devices:
1. Camera based
2. Liquid Ionization Chamber Matrix Based
3. Amorphous Silicon Based
12-10-2018
Advantages
• Simple addition to linear
accelerators
• By identifying bony anatomy and
implanted fiducials, maybe
sufficient for several sites
• Less mechanical calibration since
treatment and imaging beam
same
• Non IGRT use: Portal dosimetry
Disadvantage
• Relatively poor contrast in 2D
images
• Lack of soft tissue detail
• Relatively high dose
• Images need to be taken in two
lanes for translational errors
12-10-2018
Gantry Mounted KV 2D Imaging
• There is additional kilovoltage 2D imaging system (source and detector) mounted
on the gantry
12-10-2018
12-10-2018
12-10-2018
Advantages
• Allows KV imaging of superior
quality, less imaging dose
• Allows orthogonal image pair
(KV and MV) without turning
gantry
Disadvantages
• Separate attachment
• Calibration of treatment and kV
imaging isocentres required
• Cannot be used when the couch
is rotated (to avoid collisions)
12-10-2018
Summary: 2D imaging - Gantry mounted
• 2D imaging using MV EPIDs and kV imagers have applicability in sites
where matching of bony surrogates or implanted fiducials suffice
- brain tumours
- gynaecological tumours
-prostate cancer (implanted fiducials)
• Good reproducibility, ease of use
• Limited when soft tissue details need to be addressed
• Unable to account for intrafraction variability
12-10-2018
2D imaging: Floor ceiling mounted
stereoscopic imaging
Setup of a floor-ceiling mounted
system:
• Pair of KV source of imager –
one mounted on ceiling, another
on floor
• Two such orthogonal pairs with
imaging planes intersecting at
treatment isocenter
12-10-2018
Advantages of floor-ceiling mounted systems
• Fast imaging
• Intrafraction motion imaging possible since imaging while treatment beam on is
possible
• Imaging dose is low as two KV images are used instead of a KV/MV pair
• Therefore more frequent imaging can be done
Frameless stereotaxy
For stereotactic treatment requiring highly
conformal treatment either small or no
margins
Respiratory Motion Gating/ Tracking
External surrogates (chest wall motion) can be used
to determine phase and amplitude of respiration
Lung/GI lesion motion can be predicted/tracked
INDICATIONS
12-10-2018
Platforms for floor ceiling imaging
1. Extractac System (BrainLab)
(1) Two KV X Ray generators
placed in recesses in the floor
(2) Two amorphous silicon flat
panel imagers on ceiling
(3) Robotic couch capable of 6
dimensional correction
(4) Optical tracking camera
monitoring surface markers
12-10-2018
12-10-2018
Platforms for floor ceiling imaging
2. Cyberknife System
Parts:
1. LINAC: X-band linac on a
robotic arm
2. KV 2D Imager Pairs
3. Robotic couch
4. Optical tracking system
12-10-2018
Cyberknife – robotic workspace
• Robotic manipulator designed to
move the LINAC in a fixed,
predetermined workspace.
• Nodes: Preassigned points in
workspace where LINAC can
stop and deliver RT
• At each node, beams can be
delivered form different angles
12-10-2018
Difference in how Cyberknife responds to
shifts
Traditional Linear accelerators
Setup errors/ intrafraction change
determined by image guidance
Corrections made by shifting
patient/couch to bring back target
to planning position
Cyberknife
Setup errors/ intrafraction change
determined by image guidance
Corrections made by continuously
changing the LINAC position/angle
to adjust the treatment beam to
match the planning position
before treatment is delivered
12-10-2018
Platforms for floor ceiling imaging
3. RTRT System
• Works by fluoroscopic tracking of
implanted fiducials
• Gold fiducials, 1.5-2mm diameter
• Placement methods:
bronchoscopic for peripheral lung
tumours, operative for spinal
tumours
• Fluroscopic cameras with moving
object recognition software
capture intrafraction motion
(gating)
• Disadvantage: exposure high
12-10-2018
Summary: 2D imaging - Ceiling mounted
• Floor ceiling mounted imaging fast as not related to gantry motion
• Intrafraction imaging possible – useful in stereotactic
hypofractionated treatments with long treatment times and in sites
with high organ motion uncertainty
• Frameless intracranial and body stereotactic platforms rely on
frequent stereoscopic imaging
• Limitation: no soft tissue visualization
12-10-2018
Imaging Techniques
USG Based
BAT
- SonArray
- I-Beam
- Restitu
Video Based
- AlignRT
Planar Imaging
Based
KV Xray OBI
(On Board
Imaging)
Gantry
Mounted
- Varian OBI
- Elekta
Synergy
-IRIS
Stereoscopic/
Room
Mounted
- Cyberknife
- RTRT
-Extractrac
System
(BrainLab)
MV X Ray
Imaging
EPID
(Electronic
Portal Imaging
Device)
Volumetric
imaging based
Fan Beam
- Helical
Tomotherapy
- In room CT
Cone Beam
KVCBCT
- Varian OBI
- Elekta
MVCBCT
Siemens
12-10-2018
3D imaging: KV CT based IGRT
• Involves use of kilovoltage X Rays from a source separate from the
LINAC beam but within the treatment room to generate a 3D
(volumetric) CT image
EPIDS
Use a pair of MV/KV images
to obtain setup information
in 3 axes (X, Y, Z) based on
bony anatomy
In room CT
Many KV/MV images
acquired as the gantry
rotates and reconstructing
them to obtain volume
images – setup information
in 3 axes (X,Y,Z) and
anatomical changes based in
bony and soft tissue anatomy
12-10-2018
From 2D to 3D
12-10-2018
Types of KV CT
CT on rails
Separate diagnostic CT unit is
places in the treatment room with
rails connecting the CT with the
couch
Cone beam CT (CBCT)
Imaging beam is shaped like a
cone and is captured by a flat
panel detector
12-10-2018
CT on rails
12-10-2018
Cone Beam CBCT
12-10-2018
3D imaging: MV CT based IGRT
MV imaging involves use of megavoltage beams from the same LINAC
that delivers treatment to take verification images for patient setup and
position
Evolution from EPIDs
EPIDs
Use a pair of MV images to
obtain setup information in
3 axes based on bony
anatomy
MV CT
Use many MV images to
obtain setup information in 3
axes and deformation based
on bony and soft-tissue
anatomy
12-10-2018
MVCBCT: Imaging beam
• Imaging beam is different form treatment beam
• Treatment beam: 6 MV, 300 MU/min
• Imaging beam: 6 MV, 50 MU/min
• Aim is to minimize imaging dose while acquiring good quality images
• New Imaging Beam Line (IBL) uses:
- 4 MV beam
- graphite/diamond target (produce low energy photons)
- no flattening filter (prevent beam hardening)
Purpose is to develop a photon beam with optimal energy
12-10-2018
Types of MVCBCT
12-10-2018
MVCBCT
12-10-2018
Tomotherapy: Fan beam based MVCT
12-10-2018
USG Based Image Guidance
• Relatively inexpensive modality
that allows soft tissue visualization
without use of ionizing radiation
• Commonly used for prostate
cancer
• Diminished use now due to fiducial
based 2D X ray or CBCT based
verification
12-10-2018
Electromagnetic tracking
• Method of position verification of target with implanted beacons
before and during treatment using electromagnetic signals instead of
ionizing radiation
12-10-2018
Commercially available systems
12-10-2018
12-10-2018
Extractac (Brainlab)
kV X ray
CBCT
Fluoroscopy
EPID
Timeline
•Positioning and Immobilization
•Image Acquisition and Registration
•Target Delineation
•Treatment Planning
•Setup confirmation
•Treatment Delivery
•Quality assurance
12-10-2018
12-10-2018
Uncertainties in planning and delivery due to respiratory motion
Breath hold techniques
1. Voluntary Breath Hold
- Patient holds breath without mechanical
assistance
- Deep Inspiratory Breath hold:
• Deep inspiration breath hold (DIBH) :- Free-
breathing interval followed by a breath hold
at approximately 100% vital capacity.
• The patient is trained to take a deep breath
and exhale slowly and then to take another
deep breath and hold it for as long as
he/she can.
12-10-2018
2. Assisted Breath hold
Assisted techniques
mechanically block the air
passage of the spirometer for
short periods of time to maintain
a static lung volume (and tumour
position)
12-10-2018
Breath hold techniques
• Gating is delivery of radiation during a
particular part of the breathing cycle
• Breathing cycle is monitored with the help
of external signals or fiducials
• Beam on-off synchronised with motion
surrogate
• Real Time Radiotherapy (RTRT) system:
gating using internal signals
12-10-2018
Respiratory Gating
• Tumour tracking involves the
radiation beam which is designed
to treat the tumour with small
margins by following the tumour
• Achieved by:
- Movement of the linear
accelerator assembly
- Movement of MLCs
• Machines: Cyberknife, Vero
12-10-2018
Tumour tracking
Offline and online correction strategies
• Offline verification: When in room imaging before treatment is
compared to the reference (planning) image for setup errors at a time
after treatment has been delivered, it is called offline verification.
• Setup errors detected for that set of fractions are only used for setup
corrections for future treatment.
• Offline verification protocols are used when daily imaging is not
essential.
12-10-2018
Image for first few
fractions
Match after
treatment
Use setup errors from
first few treatments
to change setup for
future treatment
• Online verification: When in room imaging before treatment is
compared to the reference (planning) image for setup errors/organ
motion before treatment has been delivered with a view to base setup
corrections for the same day, it is called online verification.
• Setup errors detected for that fraction are used for setup corrections
the same day’s treatment.
Image daily
Match before
treatment
Use match for
correction before
each day’s
treatment
12-10-2018
Imaging protocol
12-10-2018
OFFLINE ONLINE
Corrects systematic errors only Corrects systematic and random errors
Less resource intensive More resource intensive
Less imaging dose More imaging dose
Conventionally fractionated treatments
with good immobilization and less target
motion
Stereotactic, hypofractionated treatments
(with small PTV margins). Sites with
significant variations in target motion,
deformation
Offline protocol
Done one of several established correction protocols:
1) Shrinking Action Level (SAL) protocol
2) No Action Level (NAL) protocol
12-10-2018
No Action Level Protocol
• Measure the errors in the first 3-5 fractions
• Calculate the mean
• Implement it as an isocentre shift for the remaining fractions
12-10-2018
Example of NAL implementation
12-10-2018
X Y Z
Fraction 1 0.7 0.3 0.6
Fraction 2 0.3 0.4 0.5
Fraction 3 0.5 0.5 0.4
Mean 0.5 0.4 0.5
The mean value after 3-5 fractions
is implemented as an isocentre
adjustment during setup for all
subsequent fractions no matter
how small its value. There is no
threshold or action level.
Setup errors are weekly
checked to verify if there
are no residual errors. If
yes, they are corrected.
Example of Shrinking
Action Level Protocol
• SAL: for every fraction, the mean
error is compared with the
action level.
• If mean error remains lower
than action level for 5 fractions,
no need to correct setup and
further imaging is stopped after
5 fractions.
• Advantage of SAL: Less need for
setup correction
12-10-2018
X Mean
error
Action
Level, A0
Action
Fraction 1 0.7 0.7 0.7 Continue
Fraction 2 0.3 0.5 0.5 Continue
Fraction 3 0.5 0.5 0.4 Correct by
0.4
An = A0/ 𝑁,
An = Action level at fraction N
A0 = Initial action level
Practically, NAL is widely used for its
simplicity and less imaging requirement.
Derivation of setup PTV margins
• Planning Target Volume: geometrical concept that takes into account the net
effect of all possible geometrical variations, in order to ensure that the
prescribed dose is actually absorbed in the CTV.
• PTV = ITV + setup margin
• Setup margins are derived from the systematic and random errors using
margin recipes or formulas.
Van Herk (2000):
Set up margin
PTV = 2.5ƹ + 0.7σ where ƹ = systematic error, σ = random error
To ensure minimum CTV dose of 95% to 90% of the patients
• Margin recipes give more importance to systematic errors. Random errors
cause dose blurring but random errors cause persistent underdosage.
12-10-2018
12-10-2018
Head and neck
• Immobilization is easy and effective – S clamp orfit cast.
• Target delineation: MRI, PET.
• Role of adaptive planning: Tumour shrinkage/weight loss
12-10-2018
Lung cancer
• Image acquisition: Special strategies like breath hold or 4D CT
• Advanced imaging modality: PET CT for identification of primary and
nodal volume
• Individualized margins
• Motion management strategies for reducing requirement of large PTV
margins: motion restriction, breath hold, gating, tracking
12-10-2018
Prostate cancer
• Treatment uncertainties arise:
rectal distension, rotation of
seminal vesicle position,
intrafraction motion
• Interfraction motion: 2D imaging
with implanted fiducials, X based
volumetric imaging and USG based
IGRT
• Daily online image guidance with
soft tissue matching
12-10-2018
Upper GI malignancy: Liver, pancreas
• Appropriate imaging: Triphasic CT +/- fusion with MRI
• Consider fiducial placement: important SBRT
• Respirator motion management strategy: 4D CT, motion guided
delivery
• Imaging verification for motion and setup: Online verification for SBRT
12-10-2018
Summary
The use of modern imaging modalities incorporating functional or
biological information to augment target delineation and the use of
imaging to adjust for target motion or positional uncertainty and
potentially to adapt treatment to tumor response
12-10-2018
Thank you
12-10-2018

More Related Content

What's hot

Volumetric Modulated Arc Therapy
Volumetric Modulated Arc TherapyVolumetric Modulated Arc Therapy
Volumetric Modulated Arc Therapyfondas vakalis
 
Principles of beam direction and use of simulators
Principles of beam direction and use of simulators Principles of beam direction and use of simulators
Principles of beam direction and use of simulators Anil Gupta
 
GAMMA KNIFE , X-KNIFE AND CYBERKNIFE
GAMMA KNIFE , X-KNIFE AND CYBERKNIFE GAMMA KNIFE , X-KNIFE AND CYBERKNIFE
GAMMA KNIFE , X-KNIFE AND CYBERKNIFE Pooja Gupta
 
Advances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and DeliveryAdvances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and DeliveryMiami Cancer Institute
 
Patient Positioning and Immobilization Devices In Radiotherapy Planning
Patient Positioning and Immobilization Devices In Radiotherapy PlanningPatient Positioning and Immobilization Devices In Radiotherapy Planning
Patient Positioning and Immobilization Devices In Radiotherapy PlanningSubhash Thakur
 
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptxQUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptxSuryaSuganthan2
 
Electron arc therapy
Electron arc therapyElectron arc therapy
Electron arc therapyVinay Desai
 
Commissioning of Truebeam LINAC
Commissioning of Truebeam LINACCommissioning of Truebeam LINAC
Commissioning of Truebeam LINACVIneeth C
 
Photon Vs proton beam therapy
Photon Vs proton beam therapyPhoton Vs proton beam therapy
Photon Vs proton beam therapyDr Vijay Raturi
 
Radioisotopes and dose rates used for brachytherapy
Radioisotopes and dose rates used for brachytherapyRadioisotopes and dose rates used for brachytherapy
Radioisotopes and dose rates used for brachytherapySubhash Thakur
 
The role of a physicist in radiation oncology
The role of a physicist in radiation oncologyThe role of a physicist in radiation oncology
The role of a physicist in radiation oncologyMohamed Ali Morsy
 
Stereotactic Body Radiation Therapy
Stereotactic Body Radiation TherapyStereotactic Body Radiation Therapy
Stereotactic Body Radiation Therapyfondas vakalis
 
Immobilisation in Radiotherapy
Immobilisation in RadiotherapyImmobilisation in Radiotherapy
Immobilisation in RadiotherapyYamini Baviskar
 

What's hot (20)

Volumetric Modulated Arc Therapy
Volumetric Modulated Arc TherapyVolumetric Modulated Arc Therapy
Volumetric Modulated Arc Therapy
 
Principles of beam direction and use of simulators
Principles of beam direction and use of simulators Principles of beam direction and use of simulators
Principles of beam direction and use of simulators
 
GAMMA KNIFE , X-KNIFE AND CYBERKNIFE
GAMMA KNIFE , X-KNIFE AND CYBERKNIFE GAMMA KNIFE , X-KNIFE AND CYBERKNIFE
GAMMA KNIFE , X-KNIFE AND CYBERKNIFE
 
Advances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and DeliveryAdvances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and Delivery
 
Helical Tomotherapy
Helical TomotherapyHelical Tomotherapy
Helical Tomotherapy
 
Qc
QcQc
Qc
 
Patient Positioning and Immobilization Devices In Radiotherapy Planning
Patient Positioning and Immobilization Devices In Radiotherapy PlanningPatient Positioning and Immobilization Devices In Radiotherapy Planning
Patient Positioning and Immobilization Devices In Radiotherapy Planning
 
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptxQUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
 
3 dcrt
3 dcrt3 dcrt
3 dcrt
 
Image Guided Radiation Therapy (IGRT)
Image Guided Radiation Therapy (IGRT)Image Guided Radiation Therapy (IGRT)
Image Guided Radiation Therapy (IGRT)
 
Electron arc therapy
Electron arc therapyElectron arc therapy
Electron arc therapy
 
Commissioning of Truebeam LINAC
Commissioning of Truebeam LINACCommissioning of Truebeam LINAC
Commissioning of Truebeam LINAC
 
Photon Vs proton beam therapy
Photon Vs proton beam therapyPhoton Vs proton beam therapy
Photon Vs proton beam therapy
 
Radioisotopes and dose rates used for brachytherapy
Radioisotopes and dose rates used for brachytherapyRadioisotopes and dose rates used for brachytherapy
Radioisotopes and dose rates used for brachytherapy
 
The role of a physicist in radiation oncology
The role of a physicist in radiation oncologyThe role of a physicist in radiation oncology
The role of a physicist in radiation oncology
 
Modern Radiotherapy
Modern RadiotherapyModern Radiotherapy
Modern Radiotherapy
 
Stereotactic Body Radiation Therapy
Stereotactic Body Radiation TherapyStereotactic Body Radiation Therapy
Stereotactic Body Radiation Therapy
 
Immobilisation in Radiotherapy
Immobilisation in RadiotherapyImmobilisation in Radiotherapy
Immobilisation in Radiotherapy
 
Linear Accelerator
Linear AcceleratorLinear Accelerator
Linear Accelerator
 
Cobalt & linac
Cobalt & linacCobalt & linac
Cobalt & linac
 

Similar to Image Guided Radiotherapy

Radiotherapy planning in carcinoma cervix dr rekha
Radiotherapy planning in carcinoma cervix dr rekhaRadiotherapy planning in carcinoma cervix dr rekha
Radiotherapy planning in carcinoma cervix dr rekhaDr Rekha Arya
 
IMRT and 3D CRT in cervical Cancers
IMRT and 3D CRT in cervical CancersIMRT and 3D CRT in cervical Cancers
IMRT and 3D CRT in cervical CancersSantam Chakraborty
 
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...Abhishek Soni
 
Target volume definition
Target volume definitionTarget volume definition
Target volume definitionAmin Amin
 
Getting Ahead of the Expanding Landscape: Radiopharmaceutical Dosimetry
Getting Ahead of the Expanding Landscape: Radiopharmaceutical DosimetryGetting Ahead of the Expanding Landscape: Radiopharmaceutical Dosimetry
Getting Ahead of the Expanding Landscape: Radiopharmaceutical DosimetryMedpace
 
Image guided adaptive radiotherapy
Image guided adaptive radiotherapyImage guided adaptive radiotherapy
Image guided adaptive radiotherapyapollo seminar group
 
Image registration and data fusion techniques.pptx latest save
Image registration and data fusion techniques.pptx latest saveImage registration and data fusion techniques.pptx latest save
Image registration and data fusion techniques.pptx latest saveM'dee Phechudi
 
EPI - Bringing It In-House Series #4
EPI - Bringing It In-House Series #4EPI - Bringing It In-House Series #4
EPI - Bringing It In-House Series #4CannonDesign
 
PET GUIDED TARGET CONTOURING GUIDELINES.pptx
PET GUIDED TARGET CONTOURING GUIDELINES.pptxPET GUIDED TARGET CONTOURING GUIDELINES.pptx
PET GUIDED TARGET CONTOURING GUIDELINES.pptxGaurav Jaswal
 
Icru 29,50 &62
Icru 29,50 &62Icru 29,50 &62
Icru 29,50 &62Dhiman Das
 
Current and emerging partnerships in imaging and radiotherapy by musila mutala
Current and emerging partnerships in imaging and radiotherapy by musila mutalaCurrent and emerging partnerships in imaging and radiotherapy by musila mutala
Current and emerging partnerships in imaging and radiotherapy by musila mutalaKesho Conference
 
THE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRTTHE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRTMelissa McClement
 
Overview of ART in lung cancer
Overview of ART in lung cancerOverview of ART in lung cancer
Overview of ART in lung cancerNaveen Mummudi
 
State Of The Art Crt Imrt
State Of The Art Crt ImrtState Of The Art Crt Imrt
State Of The Art Crt Imrtfondas vakalis
 
New Techniques in Radiotherapy
New Techniques in RadiotherapyNew Techniques in Radiotherapy
New Techniques in RadiotherapySantam Chakraborty
 
New microsoft office power point presentation
New microsoft office power point presentationNew microsoft office power point presentation
New microsoft office power point presentationSathish Kumar
 

Similar to Image Guided Radiotherapy (20)

Radiotherapy planning in carcinoma cervix dr rekha
Radiotherapy planning in carcinoma cervix dr rekhaRadiotherapy planning in carcinoma cervix dr rekha
Radiotherapy planning in carcinoma cervix dr rekha
 
IMRT and 3D CRT in cervical Cancers
IMRT and 3D CRT in cervical CancersIMRT and 3D CRT in cervical Cancers
IMRT and 3D CRT in cervical Cancers
 
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...
Three dimensional conformal radiotherapy - 3D-CRT and IMRT - Intensity modula...
 
Target volume definition
Target volume definitionTarget volume definition
Target volume definition
 
Getting Ahead of the Expanding Landscape: Radiopharmaceutical Dosimetry
Getting Ahead of the Expanding Landscape: Radiopharmaceutical DosimetryGetting Ahead of the Expanding Landscape: Radiopharmaceutical Dosimetry
Getting Ahead of the Expanding Landscape: Radiopharmaceutical Dosimetry
 
Image guided adaptive radiotherapy
Image guided adaptive radiotherapyImage guided adaptive radiotherapy
Image guided adaptive radiotherapy
 
IMRT and 3DCRT
IMRT and 3DCRT IMRT and 3DCRT
IMRT and 3DCRT
 
Image registration and data fusion techniques.pptx latest save
Image registration and data fusion techniques.pptx latest saveImage registration and data fusion techniques.pptx latest save
Image registration and data fusion techniques.pptx latest save
 
Thastro08 igrt1
Thastro08 igrt1Thastro08 igrt1
Thastro08 igrt1
 
ICRU 83
ICRU 83ICRU 83
ICRU 83
 
EPI - Bringing It In-House Series #4
EPI - Bringing It In-House Series #4EPI - Bringing It In-House Series #4
EPI - Bringing It In-House Series #4
 
ICRU 83
ICRU 83ICRU 83
ICRU 83
 
PET GUIDED TARGET CONTOURING GUIDELINES.pptx
PET GUIDED TARGET CONTOURING GUIDELINES.pptxPET GUIDED TARGET CONTOURING GUIDELINES.pptx
PET GUIDED TARGET CONTOURING GUIDELINES.pptx
 
Icru 29,50 &62
Icru 29,50 &62Icru 29,50 &62
Icru 29,50 &62
 
Current and emerging partnerships in imaging and radiotherapy by musila mutala
Current and emerging partnerships in imaging and radiotherapy by musila mutalaCurrent and emerging partnerships in imaging and radiotherapy by musila mutala
Current and emerging partnerships in imaging and radiotherapy by musila mutala
 
THE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRTTHE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRT
 
Overview of ART in lung cancer
Overview of ART in lung cancerOverview of ART in lung cancer
Overview of ART in lung cancer
 
State Of The Art Crt Imrt
State Of The Art Crt ImrtState Of The Art Crt Imrt
State Of The Art Crt Imrt
 
New Techniques in Radiotherapy
New Techniques in RadiotherapyNew Techniques in Radiotherapy
New Techniques in Radiotherapy
 
New microsoft office power point presentation
New microsoft office power point presentationNew microsoft office power point presentation
New microsoft office power point presentation
 

More from Namrata Das

Landmark trials in breast cancer.pptx
Landmark trials in breast cancer.pptxLandmark trials in breast cancer.pptx
Landmark trials in breast cancer.pptxNamrata Das
 
Radiotherapy and Cetuximab in head and neck cancer.pptx
Radiotherapy and Cetuximab in head and neck cancer.pptxRadiotherapy and Cetuximab in head and neck cancer.pptx
Radiotherapy and Cetuximab in head and neck cancer.pptxNamrata Das
 
Testing, genetic counselling and its implications in Gynaecological Cancers
Testing, genetic counselling and its implications in Gynaecological CancersTesting, genetic counselling and its implications in Gynaecological Cancers
Testing, genetic counselling and its implications in Gynaecological CancersNamrata Das
 
Importance of oxygen in radiotherapy
Importance of oxygen in radiotherapyImportance of oxygen in radiotherapy
Importance of oxygen in radiotherapyNamrata Das
 
Radiation induced cell kill and damage
Radiation induced cell kill and damageRadiation induced cell kill and damage
Radiation induced cell kill and damageNamrata Das
 
Medical Statistics used in Oncology
Medical Statistics used in OncologyMedical Statistics used in Oncology
Medical Statistics used in OncologyNamrata Das
 
SRS, SRT CNS Tumours
SRS, SRT CNS TumoursSRS, SRT CNS Tumours
SRS, SRT CNS TumoursNamrata Das
 

More from Namrata Das (7)

Landmark trials in breast cancer.pptx
Landmark trials in breast cancer.pptxLandmark trials in breast cancer.pptx
Landmark trials in breast cancer.pptx
 
Radiotherapy and Cetuximab in head and neck cancer.pptx
Radiotherapy and Cetuximab in head and neck cancer.pptxRadiotherapy and Cetuximab in head and neck cancer.pptx
Radiotherapy and Cetuximab in head and neck cancer.pptx
 
Testing, genetic counselling and its implications in Gynaecological Cancers
Testing, genetic counselling and its implications in Gynaecological CancersTesting, genetic counselling and its implications in Gynaecological Cancers
Testing, genetic counselling and its implications in Gynaecological Cancers
 
Importance of oxygen in radiotherapy
Importance of oxygen in radiotherapyImportance of oxygen in radiotherapy
Importance of oxygen in radiotherapy
 
Radiation induced cell kill and damage
Radiation induced cell kill and damageRadiation induced cell kill and damage
Radiation induced cell kill and damage
 
Medical Statistics used in Oncology
Medical Statistics used in OncologyMedical Statistics used in Oncology
Medical Statistics used in Oncology
 
SRS, SRT CNS Tumours
SRS, SRT CNS TumoursSRS, SRT CNS Tumours
SRS, SRT CNS Tumours
 

Recently uploaded

Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryJyoti singh
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...rajnisinghkjn
 
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...amritaverma53
 
Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...
Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...
Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...minkseocompany
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsMedicoseAcademics
 
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...gragneelam30
 
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...gragneelam30
 
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...chanderprakash5506
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana GuptaLifecare Centre
 
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowtanudubay92
 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationMedicoseAcademics
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...soniyagrag336
 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxSwetaba Besh
 
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...Rashmi Entertainment
 
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...call girls hydrabad
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableSteve Davis
 
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...Janvi Singh
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...TanyaAhuja34
 
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...dishamehta3332
 

Recently uploaded (20)

Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
 
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
 
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
 
Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...
Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...
Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanisms
 
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
 
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
 
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their Regulation
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
 
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
 
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
 
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
 

Image Guided Radiotherapy

  • 1. Image Guided Radiation Therapy Moderator: Prof. Bhavana Rai Dr. Namrata Das 1st year Junior Resident 12-10-2018
  • 2. OUTLINE • Introduction - Evolution of radiotherapy techniques - ICRU definitions • Types of Error • Timeline when errors are addressed by image guidance - Positioning and Immobilization - Image Acquisition and Registration: 4DCT - Target Delineation: advanced imaging modalities - Patient Setup confirmation: Types of in room imaging • Offline, online correction strategies - No Action Level (NAL), Shrinking Action Level (SAL) • Clinical application of IGRT: - Head and neck cancer - Lung cancer - Prostate cancer - Upper GI malignancy: Liver • Summary 12-10-2018
  • 3. 12-10-2018 Introduction The purpose of radiotherapy is to maximize tumour kill and minimize normal tissue damage by delivering a precisely measured dose uniformly to the target tissue.
  • 4. • Radiotherapy treatment planning has always used imaging. • History of development of radiotherapy has closely followed the development of new imaging techniques. 12-10-2018 Evolution of Radiotherapy Treatment Techniques 2D Planning 3D Planning/ Conformal Radiotherapy X Ray Based: Discovery of X Ray and subsequent boom during the World War Volumetric Imaging: 1970s: Computed Tomography 1980s: Magnetic Resonance Imaging 3D CRT: geometrical field shaping IMRT: geometrical field shaping+ intensity modulation IGRT: geometrical field shaping + intensity modulation + image guidance
  • 5. 12-10-2018 2D/ Conventional Planning 3D/ Conformal Planning Drawbacks • Inability to visualize tumor or normal tissue • Only square/rectangle fields • Large safety margins resulting in high dose to normal tissue
  • 6. ICRU 29 (1978): Dose specification for the reporting of external beam radiotherapy with photons and electrons 12-10-2018 Evolution of the concept of treatment volumes Target volume Treated Volume Irradiated Volume Irradiated Volume: volume which receives an absorbed dose considered to be significant in relation to tissue tolerance Target Volume: Tissues that are to be irradiated to a specific dose Treatment volume: Volume enclosed by a relevant isodose surface that is selected and specified by the radiation oncologist as being appropriate to achieve the purpose of treatment
  • 7. 12-10-2018 ICRU 50 (1993): Prescribing, Recording, Reporting Photon Beam Therapy Treated Volume Irradiated Volume PTV CTV GTV Planning Treatment Volume: Geometrical concept defined to select appropriate beam sizes, beam arrangements taking into consideration the net effect of all possible geometrical variations, inaccuracies in order to ensure prescription done is actually absorbed in the CTV. Clinical Target Volume: Tissue volume that contains GTV and/or microscopic malignant disease which is to be eliminated – has to be treated respecting anatomical boundaries. Gross Tumour Volume: Gross palpable or visible or demonstratable extent and localization of tumour
  • 8. 12-10-2018 ICRU 62 (1999): Supplement to Rep. 50 Planning Treatment Volume (PTV) = Internal Target Volume (ITV) + Setup Margin (SM) ITV = CTV+ Internal Margin (IM) Treated Volume Irradiated Volume PTV CTV GTV ITV Internal Margin (IM): A margin that must be added to the CTV to compensate for the expected physiological movements and the variations in size, shape and position of the CTV during therapy in relation to the internal reference point. Internal Target Volume (ITV): CTV +IM Setup Margin (SM): It is the margin that must be added to account specifically for uncertainties (inaccuracies and lack of reproducibility) in patient positioning and alignment of the therapeutic beams during treatment planning and through all treatment sessions.
  • 9. Conformal Radiotherapy: 3D CRT 3 Dimensional Conformal Radiotherapy (3D CRT) It is the use of 3 dimensional anatomical information to plan and deliver treatment so that the resultant dose distribution conforms as closely as possible to the target volume in 3 dimensions with minimum dose to the surrounding normal tissue. Advantage over 2D: • increased tumour dose, reduced normal tissue dose • Reduced safety margin Drawback: • Forward planning: trial and error method • Increased dose spillage • Ability to sculpt tumour limited 12-10-2018 Transition from 2-D Radiotherapy to 3-D Conformal and Intensity Modulated Radiotherapy. IAEA-TECDOC-1588. May 2008.
  • 10. Intensity Modulated Radiotherapy is a special form of 3D conformal dose delivery, enhanced by generating a non-uniform photon fluence within each beam, calculated by an inverse treatment planning process designed to meet specified dosimetric objectives. Advantage over 3D CRT: • Able to achieve concave dose distribution • Allows dose painting: inhomogeneous dose prescription to different targets • Reduced margins • Inverse planning Drawback: Chance of missing target higher due to reduced margins at the time of treatment 12-10-2018 Conformal Radiotherapy: IMRT
  • 11. • Uses conformal uniform fields leading to convex distribution • By varying the fluence of beams, more fluence can be delivered to rays reaching target as compared to normal tissue. Allows concave dose distribution 12-10-2018
  • 12. Need for Image Guided Radiotherapy The uncertainty in target localization while delivering highly conformal radiotherapy is specially important for: • Tumours prone to intrafraction motion: lung cancer • Tumours prone to interfraction motion: prostate cancer • Tumours prone to deformation: head and neck cancers • Treatment delivery by IMRT, SBRT (high dose per fraction with narrow margins) Thus there is a need for image guidance at various stages of the treatment process. 12-10-2018
  • 13. Definition • Multiple definitions - some too extensive, others too narrow • ACR, ASTRO practice guidelines, “a procedure that refines the delivery of therapeutic radiation by applying image based targeted re-localization to allow proper patient repositioning for the purpose of ensuring accurate treatment and minimizing the volume of normal tissue exposed to radiation.” • IGRT Committee, RTOG, 2001, “IGRT refers broadly to treatment delivery using modern imaging methods such as CT, MRI, PET and USG in target and non target structures for RT definition, design and delivery”. 12-10-2018
  • 14. IGRT can be defined as the use of: • Advanced imaging modalities, especially those incorporating functional and/or biological information, to augment target and normal tissue delineation • In-room imaging to adjust for target motion or positional uncertainty (interfractional and intrafractional), and, potentially, to adapt treatment to tumour response Image Guided Radiation Therapy. A clinical perspective. Arno Mundt, John Roeske. 12-10-2018
  • 15. Use of imaging for: • Target and normal tissue delineation • Margin design and treatment planning • Setup adjustments and delivery 12-10-2018 Use of advanced imaging modality for accurate delineation of target Use of imaging for setup verification and adjustment
  • 16. Errors • Error is any mismatch between the true anatomy and pathology during treatment and planning 12-10-2018 Types of Errors • Deviation that varies in direction and magnitude for each fraction • Cumulative dose distribution gets blurred with random errors • Direction and magnitude remains same for each fraction RANDOM ERRORS SYSTEMATIC ERRORS
  • 18. 12-10-2018 Systematic Error External Wrong isocentre shift; discordance between planning CT and LINAC laser systems Solution: On Board Imaging Internal Physical changes since planning CT: drainage of pyometra Shrinkage of tumour Solution: Adaptive replanning
  • 19. Systematic errors A. CHOICE OF IMAGING All tumours and organs cannot be identified using a planning CT 12-10-2018 B. IMAGE FUSION Fusing two or more image sets for delineation can lead to some mismatch – leading to suboptimal delineation
  • 20. 12-10-2018 C. IMAGE DISTORTION Planning images can have artefacts due to motion, foreign objects or matching related imaging artefacts 1. Movement induced artefact: Fast free breathing CT Scans produce these errors C. IMAGE DISTORTION Planning images can have artefacts due to motion, foreign objects or matching related imaging artefacts Prosthetics, foreign metallic/plastic objects can distort images Systematic errors
  • 21. 12-10-2018 Systematic errors Anatomical changes due to: • Tumour regression: the shrink in tumour will lead to a change in the dose distribution of OARs (lung, head and neck) • Weight loss: changes the anatomic contour of patients • Resolving atelectasis: Opening up of blocked airways
  • 22. 12-10-2018 Random Error External Daily setup variation On board imaging Internal Daily position change due to bowel, bladder filling Respiratory Motion OBI Generation of individualized ITV margin Respiratory gating/tracking
  • 23. Random Error: organ motion and deformation Seconds: respiration, heartbeat Minutes: Peristaltic motion, rectal motion Days: Radiation induced changes in mucosa, bladder, bowel distension Weeks: changes in organ shape and position 12-10-2018
  • 24. Random Error: organ motion in the pelvis • Rectal filling and bowel filling changes: inter-fraction variability of position and rotation 12-10-2018
  • 25. • Sites affected by respiratory motion: Lung and mediastinum, breast and chest wall, upper abdomen (liver, biliary tract, stomach, pancreas) • Motion of lung tumours: - Supero-inferior (longitudinal) direction: maximum movement - Lower lobe tumours: greater movement in SI direction - Anterior tumours: greater AP motion - Central tumours have less motion than peripheral 12-10-2018 Random Error: respiratory motion Highly variable motion range: • Lung: SI = 3.9 cm, AP = 2.4cm, LR = 2.4cm • Liver: Normal breathing: 10-25 mm, deep breathing = 37-55 mm
  • 26. Timeline when errors can be addressed • Positioning and Immobilization • Image Acquisition and Registration • Target Delineation • Treatment Planning • Patient specific dosimetric verification of treatment plan • Patient Setup confirmation • Treatment Delivery • Quality assurance 12-10-2018
  • 27. Immobilization in IGRT Rationale: • Reduces interfraction and intrafraction variation in position • Reduces intrafraction motion of anatomic region or specific targets • Reduces dependence on frequent imaging • Reduce setup and imaging time • External or internal 12-10-2018
  • 28. Immobilization aids 12-10-2018 BRAIN HEAD AND NECK - 3D CRT – thermoplastic mask - Stereotactic treatment: Rigid frames (<=1mm)/non-invasive and frameless systems with image guidance (<=2mm) Thermoplastic mask – 3-5mm setup errors in the region of the head and face
  • 30. Timeline when errors can be addressed •Positioning and Immobilization •Image Acquisition and Registration •Target Delineation •Treatment Planning •Setup confirmation •Treatment Delivery •Quality assurance 12-10-2018
  • 31. 4D CT Simulation In 4D CT, a series of CT scans is reconstructed at different phases of the breathing cycle. The three components of a normal CT simulator are: (1) A large bore (75-85cm) to accommodate various treatment positions along with treatment accessories with a flat couch insert to simulate treatment machine couch. (2) An integrated laser marking system (3) Virtual simulation and visualization 12-10-2018
  • 32. 12-10-2018 INFRA RED CAMERA PLASTIC MARKER WITH INFRA RED REFLECTORS
  • 33. 4D CT: Real Time Position Management (RPM) Steps: • Patient trained to breath regularly • Setup/immobilize the patient in the treatment position • Infrared red reflector placed over diaphragm • Images acquired for multiple phases of respiration, about 8-25 datasets at each couch position. • System sorts the images into different 3D image sets (binned) with the help of respiratory signal 12-10-2018
  • 34. CT Scan Axial scan trigger, 1st couch position Axial scan trigger, 2nd couch position Exhalation Inhalation Scan Scan Scan Axial scan trigger, 3rd couch position Prospective CT Image Acquisition
  • 35. X-ray on Exhalation Inhalation 1st couch position 2nd couch position 3rd couch position “Image acquired” signal to RPM system (Ford 2003, Vedam 2003) Retrospective 4D CT Image Acquisition
  • 36. Limitations of 4D CT • Depends upon uniform patient respiratory pattern • The range of motion may not be similar during treatment so image guidance during treatment delivery is important 12-10-2018
  • 37. Derivation of ITV • Population Based: may over or underestimate motion. Not ideal. • Individual Based: 1. Combining Breath hold scans 2. Combining 4DCT images 3. MIP: Maximum intensity projection 12-10-2018 12-10-2018 PTV CTV GTV ITV ITV = CTV + IM PTV = ITV + SM
  • 38. 1. ITV: Combining 4DCT images Target is contoured in every phase of respiration and then fused to generate the ITV. Advantage: Captures the entire trajectory of motion Disadvantage: Time consuming 12-10-2018
  • 39. 2. ITV: MIP Maximum Intensity Projection (MIP): MIP is an image set generated which includes all the possible locations of a tumour. Maximum range of motion of the hyperdense tumour is considered and images are stacked together to yield the MIP. Advantage: accurate and automatic method of generating the ITV Disadvantage: beneficial only for lung tumours surrounded by normal parenchyma (not those in contact with mediastinum, chest wall, diaphragm) 12-10-2018
  • 40. Timeline •Positioning and Immobilization •Image Acquisition and Registration •Target Delineation •Treatment Planning •Setup confirmation •Treatment Delivery •Quality assurance 12-10-2018
  • 41. Tissue delineation using advanced imaging • Image fusion for radiotherapy is very important to correctly define tumour and normal tissue • Necessary to wrap high doses around tumour • CT image essential for dose calculation • However, target delineation on CT not accurate for all sites • To avoid systematic errors from inaccurate delineation, fusion of other modalities important 12-10-2018
  • 43. 3 steps of image fusion • Registration: Computes the difference between images • Fusion: Fuses the images with the calculated differences • Validation: Check if images are matching well 12-10-2018
  • 44. PET-CT fusion In lung cancers, PET-CT fusion helps in: • Primary tumour delineation • Nodal volume Incorporation of FDG PET in CT based planning: • Increase or decrease in target volume (usually decrease) • Reduces interobserver variability • Distinguish between collapse and consolidation 12-10-2018
  • 45. Timeline •Positioning and Immobilization •Image Acquisition and Registration •Target Delineation •Treatment Planning •Setup confirmation •Treatment Delivery •Quality assurance 12-10-2018
  • 46. Setup Error It is the difference in position of the target (or surrogate bony anatomy) between planning and treatment after a treatment setup based on external setup markers. • Skin marks/tattoos are the set of coordinates based on which patients are initially et up. • Treatment verification imaging using internal boy anatomy (2D verification) or soft tissue target (3D verification) does not exactly match planning images (DRRs or CT). • This difference in position is called setup error 12-10-2018
  • 47. Causes of setup error • Technical/QA - Laser misalignment, isocentre alignment not coinciding • Initial setup -isocentre shifts • Positioning and Immobilization - Imperfect arms and legs can change position of external skin marks - Suboptimal immobilization may lead to changes in between setup and verification/treatment • Daily Setup -positioning -suboptimal marking of external markers 12-10-2018
  • 48. 6 axes of motion Motion and errors are described in 6 axes, 3 translational and 3 rotational THREE TRANSLATIONAL: lateral, longitudinal, vertical THREE ROTATIONAL: pitch, roll, yaw 12-10-2018
  • 49. Imaging Techniques USG Based BAT - SonArray - I-Beam - Restitu Video Based - AlignRT Planar Imaging Based KV Xray OBI (On Board Imaging) Gantry Mounted - Varian OBI - Elekta Synergy -IRIS Stereoscopic/ Room Mounted - Cyberknife - RTRT -Extractrac System (BrainLab) MV X Ray Imaging EPID (Electronic Portal Imaging Device) Volumetric imaging based Fan Beam - Helical Tomotherapy - In room CT Cone Beam KVCBCT - Varian OBI - Elekta MVCBCT Siemens 12-10-2018
  • 50. Imaging Techniques USG Based BAT - SonArray - I-Beam - Restitu Video Based - AlignRT Planar Imaging Based KV Xray OBI (On Board Imaging) Gantry Mounted - Varian OBI - Elekta Synergy -IRIS Stereoscopic/ Room Mounted - Cyberknife - RTRT -Extractrac System (BrainLab) MV X Ray Imaging EPID (Electronic Portal Imaging Device) Volumetric imaging based Fan Beam - Helical Tomotherapy - In room CT Cone Beam KVCBCT - Varian OBI - Elekta MVCBCT Siemens 12-10-2018
  • 51. 2D imaging: Gantry mounted Portal Imaging • Portal imaging is a method to verify that the area planned by a radiation field (or portal) matches with the area treated on a patient. • Aim: To match bony anatomy covered in plan with that of the covered by portal curing treatment • Earliest form of image guidance 12-10-2018
  • 52. Steps of obtaining a portal radiograph Film in a cassette holder (with phosphor screen) placed across patient MV treatment field (portal) exposed Film manually taken to a developing room and developed image compared with planning radiograph 12-10-2018
  • 53. Portal imaging Advantages • Inexpensive • First method of verification Limitations • Time consuming, cumbersome • Poor quality (megavoltage) • Setup data could not be stored Digital (electronic) imaging solved most of these drawbacks. 12-10-2018
  • 54. Electronic portal Imaging (EPI) EPI uses a detector, usually attached to the gantry opposite the treatment beam, to produce high quality digital images rapidly. - MV treatment beam used - Allows fast identification of setup errors and online correction 12-10-2018
  • 55. EPIDs on Linear Accelerators • IGRT EPID is a flat panel detector mounted on retractable arms opposite the treatment head • Arms often capable of movement to allow them to place detector at variable distance from source EPID 12-10-2018
  • 56. Amorphous silicon based EPIDs • Tiles of photodiodes arranged on glass substrate on a flat panel array • Incident photons pass through a phosphor and emit light • Photodiode captures light, emits electronic signal Three types of commercial electronic portal imaging devices: 1. Camera based 2. Liquid Ionization Chamber Matrix Based 3. Amorphous Silicon Based 12-10-2018
  • 57. Advantages • Simple addition to linear accelerators • By identifying bony anatomy and implanted fiducials, maybe sufficient for several sites • Less mechanical calibration since treatment and imaging beam same • Non IGRT use: Portal dosimetry Disadvantage • Relatively poor contrast in 2D images • Lack of soft tissue detail • Relatively high dose • Images need to be taken in two lanes for translational errors 12-10-2018
  • 58. Gantry Mounted KV 2D Imaging • There is additional kilovoltage 2D imaging system (source and detector) mounted on the gantry 12-10-2018
  • 61. Advantages • Allows KV imaging of superior quality, less imaging dose • Allows orthogonal image pair (KV and MV) without turning gantry Disadvantages • Separate attachment • Calibration of treatment and kV imaging isocentres required • Cannot be used when the couch is rotated (to avoid collisions) 12-10-2018
  • 62. Summary: 2D imaging - Gantry mounted • 2D imaging using MV EPIDs and kV imagers have applicability in sites where matching of bony surrogates or implanted fiducials suffice - brain tumours - gynaecological tumours -prostate cancer (implanted fiducials) • Good reproducibility, ease of use • Limited when soft tissue details need to be addressed • Unable to account for intrafraction variability 12-10-2018
  • 63. 2D imaging: Floor ceiling mounted stereoscopic imaging Setup of a floor-ceiling mounted system: • Pair of KV source of imager – one mounted on ceiling, another on floor • Two such orthogonal pairs with imaging planes intersecting at treatment isocenter 12-10-2018
  • 64. Advantages of floor-ceiling mounted systems • Fast imaging • Intrafraction motion imaging possible since imaging while treatment beam on is possible • Imaging dose is low as two KV images are used instead of a KV/MV pair • Therefore more frequent imaging can be done Frameless stereotaxy For stereotactic treatment requiring highly conformal treatment either small or no margins Respiratory Motion Gating/ Tracking External surrogates (chest wall motion) can be used to determine phase and amplitude of respiration Lung/GI lesion motion can be predicted/tracked INDICATIONS 12-10-2018
  • 65. Platforms for floor ceiling imaging 1. Extractac System (BrainLab) (1) Two KV X Ray generators placed in recesses in the floor (2) Two amorphous silicon flat panel imagers on ceiling (3) Robotic couch capable of 6 dimensional correction (4) Optical tracking camera monitoring surface markers 12-10-2018
  • 67. Platforms for floor ceiling imaging 2. Cyberknife System Parts: 1. LINAC: X-band linac on a robotic arm 2. KV 2D Imager Pairs 3. Robotic couch 4. Optical tracking system 12-10-2018
  • 68. Cyberknife – robotic workspace • Robotic manipulator designed to move the LINAC in a fixed, predetermined workspace. • Nodes: Preassigned points in workspace where LINAC can stop and deliver RT • At each node, beams can be delivered form different angles 12-10-2018
  • 69. Difference in how Cyberknife responds to shifts Traditional Linear accelerators Setup errors/ intrafraction change determined by image guidance Corrections made by shifting patient/couch to bring back target to planning position Cyberknife Setup errors/ intrafraction change determined by image guidance Corrections made by continuously changing the LINAC position/angle to adjust the treatment beam to match the planning position before treatment is delivered 12-10-2018
  • 70. Platforms for floor ceiling imaging 3. RTRT System • Works by fluoroscopic tracking of implanted fiducials • Gold fiducials, 1.5-2mm diameter • Placement methods: bronchoscopic for peripheral lung tumours, operative for spinal tumours • Fluroscopic cameras with moving object recognition software capture intrafraction motion (gating) • Disadvantage: exposure high 12-10-2018
  • 71. Summary: 2D imaging - Ceiling mounted • Floor ceiling mounted imaging fast as not related to gantry motion • Intrafraction imaging possible – useful in stereotactic hypofractionated treatments with long treatment times and in sites with high organ motion uncertainty • Frameless intracranial and body stereotactic platforms rely on frequent stereoscopic imaging • Limitation: no soft tissue visualization 12-10-2018
  • 72. Imaging Techniques USG Based BAT - SonArray - I-Beam - Restitu Video Based - AlignRT Planar Imaging Based KV Xray OBI (On Board Imaging) Gantry Mounted - Varian OBI - Elekta Synergy -IRIS Stereoscopic/ Room Mounted - Cyberknife - RTRT -Extractrac System (BrainLab) MV X Ray Imaging EPID (Electronic Portal Imaging Device) Volumetric imaging based Fan Beam - Helical Tomotherapy - In room CT Cone Beam KVCBCT - Varian OBI - Elekta MVCBCT Siemens 12-10-2018
  • 73. 3D imaging: KV CT based IGRT • Involves use of kilovoltage X Rays from a source separate from the LINAC beam but within the treatment room to generate a 3D (volumetric) CT image EPIDS Use a pair of MV/KV images to obtain setup information in 3 axes (X, Y, Z) based on bony anatomy In room CT Many KV/MV images acquired as the gantry rotates and reconstructing them to obtain volume images – setup information in 3 axes (X,Y,Z) and anatomical changes based in bony and soft tissue anatomy 12-10-2018
  • 74. From 2D to 3D 12-10-2018
  • 75. Types of KV CT CT on rails Separate diagnostic CT unit is places in the treatment room with rails connecting the CT with the couch Cone beam CT (CBCT) Imaging beam is shaped like a cone and is captured by a flat panel detector 12-10-2018
  • 78. 3D imaging: MV CT based IGRT MV imaging involves use of megavoltage beams from the same LINAC that delivers treatment to take verification images for patient setup and position Evolution from EPIDs EPIDs Use a pair of MV images to obtain setup information in 3 axes based on bony anatomy MV CT Use many MV images to obtain setup information in 3 axes and deformation based on bony and soft-tissue anatomy 12-10-2018
  • 79. MVCBCT: Imaging beam • Imaging beam is different form treatment beam • Treatment beam: 6 MV, 300 MU/min • Imaging beam: 6 MV, 50 MU/min • Aim is to minimize imaging dose while acquiring good quality images • New Imaging Beam Line (IBL) uses: - 4 MV beam - graphite/diamond target (produce low energy photons) - no flattening filter (prevent beam hardening) Purpose is to develop a photon beam with optimal energy 12-10-2018
  • 82. Tomotherapy: Fan beam based MVCT 12-10-2018
  • 83. USG Based Image Guidance • Relatively inexpensive modality that allows soft tissue visualization without use of ionizing radiation • Commonly used for prostate cancer • Diminished use now due to fiducial based 2D X ray or CBCT based verification 12-10-2018
  • 84. Electromagnetic tracking • Method of position verification of target with implanted beacons before and during treatment using electromagnetic signals instead of ionizing radiation 12-10-2018
  • 86. 12-10-2018 Extractac (Brainlab) kV X ray CBCT Fluoroscopy EPID
  • 87. Timeline •Positioning and Immobilization •Image Acquisition and Registration •Target Delineation •Treatment Planning •Setup confirmation •Treatment Delivery •Quality assurance 12-10-2018
  • 88. 12-10-2018 Uncertainties in planning and delivery due to respiratory motion
  • 89. Breath hold techniques 1. Voluntary Breath Hold - Patient holds breath without mechanical assistance - Deep Inspiratory Breath hold: • Deep inspiration breath hold (DIBH) :- Free- breathing interval followed by a breath hold at approximately 100% vital capacity. • The patient is trained to take a deep breath and exhale slowly and then to take another deep breath and hold it for as long as he/she can. 12-10-2018
  • 90. 2. Assisted Breath hold Assisted techniques mechanically block the air passage of the spirometer for short periods of time to maintain a static lung volume (and tumour position) 12-10-2018 Breath hold techniques
  • 91. • Gating is delivery of radiation during a particular part of the breathing cycle • Breathing cycle is monitored with the help of external signals or fiducials • Beam on-off synchronised with motion surrogate • Real Time Radiotherapy (RTRT) system: gating using internal signals 12-10-2018 Respiratory Gating
  • 92. • Tumour tracking involves the radiation beam which is designed to treat the tumour with small margins by following the tumour • Achieved by: - Movement of the linear accelerator assembly - Movement of MLCs • Machines: Cyberknife, Vero 12-10-2018 Tumour tracking
  • 93. Offline and online correction strategies • Offline verification: When in room imaging before treatment is compared to the reference (planning) image for setup errors at a time after treatment has been delivered, it is called offline verification. • Setup errors detected for that set of fractions are only used for setup corrections for future treatment. • Offline verification protocols are used when daily imaging is not essential. 12-10-2018 Image for first few fractions Match after treatment Use setup errors from first few treatments to change setup for future treatment
  • 94. • Online verification: When in room imaging before treatment is compared to the reference (planning) image for setup errors/organ motion before treatment has been delivered with a view to base setup corrections for the same day, it is called online verification. • Setup errors detected for that fraction are used for setup corrections the same day’s treatment. Image daily Match before treatment Use match for correction before each day’s treatment 12-10-2018
  • 95. Imaging protocol 12-10-2018 OFFLINE ONLINE Corrects systematic errors only Corrects systematic and random errors Less resource intensive More resource intensive Less imaging dose More imaging dose Conventionally fractionated treatments with good immobilization and less target motion Stereotactic, hypofractionated treatments (with small PTV margins). Sites with significant variations in target motion, deformation
  • 96. Offline protocol Done one of several established correction protocols: 1) Shrinking Action Level (SAL) protocol 2) No Action Level (NAL) protocol 12-10-2018
  • 97. No Action Level Protocol • Measure the errors in the first 3-5 fractions • Calculate the mean • Implement it as an isocentre shift for the remaining fractions 12-10-2018
  • 98. Example of NAL implementation 12-10-2018 X Y Z Fraction 1 0.7 0.3 0.6 Fraction 2 0.3 0.4 0.5 Fraction 3 0.5 0.5 0.4 Mean 0.5 0.4 0.5 The mean value after 3-5 fractions is implemented as an isocentre adjustment during setup for all subsequent fractions no matter how small its value. There is no threshold or action level. Setup errors are weekly checked to verify if there are no residual errors. If yes, they are corrected.
  • 99. Example of Shrinking Action Level Protocol • SAL: for every fraction, the mean error is compared with the action level. • If mean error remains lower than action level for 5 fractions, no need to correct setup and further imaging is stopped after 5 fractions. • Advantage of SAL: Less need for setup correction 12-10-2018 X Mean error Action Level, A0 Action Fraction 1 0.7 0.7 0.7 Continue Fraction 2 0.3 0.5 0.5 Continue Fraction 3 0.5 0.5 0.4 Correct by 0.4 An = A0/ 𝑁, An = Action level at fraction N A0 = Initial action level Practically, NAL is widely used for its simplicity and less imaging requirement.
  • 100. Derivation of setup PTV margins • Planning Target Volume: geometrical concept that takes into account the net effect of all possible geometrical variations, in order to ensure that the prescribed dose is actually absorbed in the CTV. • PTV = ITV + setup margin • Setup margins are derived from the systematic and random errors using margin recipes or formulas. Van Herk (2000): Set up margin PTV = 2.5ƹ + 0.7σ where ƹ = systematic error, σ = random error To ensure minimum CTV dose of 95% to 90% of the patients • Margin recipes give more importance to systematic errors. Random errors cause dose blurring but random errors cause persistent underdosage. 12-10-2018
  • 102. Head and neck • Immobilization is easy and effective – S clamp orfit cast. • Target delineation: MRI, PET. • Role of adaptive planning: Tumour shrinkage/weight loss 12-10-2018
  • 103. Lung cancer • Image acquisition: Special strategies like breath hold or 4D CT • Advanced imaging modality: PET CT for identification of primary and nodal volume • Individualized margins • Motion management strategies for reducing requirement of large PTV margins: motion restriction, breath hold, gating, tracking 12-10-2018
  • 104. Prostate cancer • Treatment uncertainties arise: rectal distension, rotation of seminal vesicle position, intrafraction motion • Interfraction motion: 2D imaging with implanted fiducials, X based volumetric imaging and USG based IGRT • Daily online image guidance with soft tissue matching 12-10-2018
  • 105. Upper GI malignancy: Liver, pancreas • Appropriate imaging: Triphasic CT +/- fusion with MRI • Consider fiducial placement: important SBRT • Respirator motion management strategy: 4D CT, motion guided delivery • Imaging verification for motion and setup: Online verification for SBRT 12-10-2018
  • 106. Summary The use of modern imaging modalities incorporating functional or biological information to augment target delineation and the use of imaging to adjust for target motion or positional uncertainty and potentially to adapt treatment to tumor response 12-10-2018