SlideShare a Scribd company logo
1 of 64
Ultrasound Physics
Md Serajus Salekin Chowdhury
Senior Scientific Officer
Bangladesh Atomic Energy Commission
Waves
A wave can be described as a disturbance that travels through a medium from one location to another
location. For an example, when a pebble is thrown in a pond, it creates a disturbance and forms
ripples which travel together as a front in a straight-line direction, or the waves may be circular
waves that originate from the point where the disturbances occurs in the pond.
In physics, the wave is defined as the energy transferred through medium with regular vibration
or oscillating motion. A few examples of waves are: water wave, light wave, electromagnetic wave,
sound wave, seismic wave (earthquakes).
One way to categorize waves is on the basis of the direction of movement of the distinct particles of
the medium relative to the direction that the waves travel. Waves can be classified into two types
Longitudinal Wave and Transverse Wave.
Figure : Waves
Longitudinal Wave
Longitudinal waves occur when the oscillations are parallel to the direction of propagation. It is a
wave in which particles of the medium move in a direction parallel to the direction that the wave
moves. A sound wave traveling through air is a classic example of a longitudinal wave.
Figure: Longitudinal Wave
Transverse Wave
Transverse waves occur when a disturbance creates oscillations perpendicular at right angles to the
direction of energy transfer. It is a wave in which particles of the medium move in a direction
perpendicular to the direction that the wave moves. The electromagnetic radiations which include
light waves are a classic example of transverse waves.
Figure: Transverse Wave
Amplitude
Amplitude of a wave is defined as the distance from the rest position of the vibrating particle of the
medium to the highest point of the crest or the lowest point of the trough. It is denoted with a or A
and measured in meter.
Figure: Amplitude
Wavelength
Waves make characteristic patterns as they travel through space. The wavelength, λ, of a wave is the
distance from any point on one wave to the same point on the next wave along. (The symbol is a
Greek letter, 'lambda'.) To avoid confusion, it is best to measure wavelength from the top of a crest to
the top of the next crest, or from the bottom of a trough to the bottom of the next trough. Wavelength
is measured in meter.
Figure: Wavelength
Wavelength
Since a longitudinal wave does not contain crests and troughs, its wavelength must be measured
differently. A longitudinal wave consists of a repeating pattern of compressions and rarefactions.
Thus, the wavelength is commonly measured as the distance from one compression to the next
adjacent compression or the distance from one rarefaction to the next adjacent rarefaction.
Figure: Wavelength
Frequency
The frequency, f, of a wave is the number of waves passing a point in a certain time. We normally use
a time of one second, so we can say, “the number of waves passing a point in one second is called
frequency of the wave”. The unit of frequency is hertz (Hz), i.e., one hertz is equal to one wave per
second.
For water waves and sound waves the unit hertz is usually good enough but radio and TV waves have
such a high frequency that the kilohertz (kHz) or even the megahertz (MHz) are better units.
1 kHz = 1,000 Hz
1 MHz = 1,000,000 Hz
Time Period
Time taken for one complete cycle or vibration by the wave generating source is called time period.
In other word it can be defined as the time taken by a wave to travel the distance of a complete
wavelength. It is simply reciprocal of the frequency (1/f). Unit of time period is second.
𝑇𝑖𝑚𝑒 𝑃𝑒𝑟𝑖𝑜𝑑 𝑇 =
1
𝐹𝑟𝑒𝑞𝑢𝑒𝑛𝑐𝑦 (𝑓)
Wave Speed/ Wave Velocity
The wave speed/velocity is measured as the distance travelled by the crests or trough in a given time
interval. The relationship is similar to the relationship of speed, distance and time in the classical
physics.
Therefore, wave speed/wave velocity can be represented mathematically as,
SW = DW /T
Where,
SW = speed of the wave
DW = distance travelled by crest or trough of the wave
T = time taken to travel the distance DW
The wave speed varies for different types of waves. The Speed of light wave is 3 ×× 108108 m/s.
Frequency, Wavelength & Wave velocity Relationship
The diagrams at the right show several "snapshots" of the production of a wave within a
rope. The motion of the disturbance along the medium after every one-fourth of a period
is depicted. Observe that in the time it takes from the first to the last snapshot, the hand
has made one complete back-and-forth motion. A period has elapsed. Observe that
during this same amount of time, the leading edge of the disturbance has moved a
distance equal to one complete wavelength. So in a time of one period, the wave has
moved a distance of one wavelength. Combining this information with the equation for
speed (speed = distance/time), it can be said that the speed of a wave is also the
wavelength/period.
Since the period is the reciprocal of the frequency, the expression 1/f can be substituted
into the above equation for period. Rearranging the equation yields a new equation of
the form:
Speed = Wavelength • Frequency
The above equation is known as the wave equation. It states the mathematical
relationship between the speed (v) of a wave and its wavelength (λ) and frequency (f).
Using the symbols v, λ, and f, the equation can be rewritten as
v = f • λ
Sound
Sound is a mechanical wave that results from the back and forth vibration of the particles of the
medium through which the sound wave is moving. If a sound wave is moving from left to right
through air, then particles of air will be displaced both rightward and leftward as the energy of the
sound wave passes through it. The motion of the particles is parallel (and anti-parallel) to the
direction of the energy transport. This is what characterizes sound waves in air as longitudinal
waves.
Because of the longitudinal motion of the air particles, there are regions in the air where the air
particles are compressed together and other regions where the air particles are spread apart. These
regions are known as compressions and rarefactions respectively. The compressions are regions of
high air pressure while the rarefactions are regions of low air pressure. The diagram below depicts a
sound wave created by a tuning fork and propagated through the air in an open tube. The
compressions and rarefactions are labeled.
Figure: Propagation of sound
Types of Sound Wave
According to audibility, more precisely frequency limit, sound waves are grouped in three classes:
1. Infrasound,
2. Audible sound, and
3.Ultrasound.
Infrasound
Infrasound are the sound waves with frequencies below the lower limit of human audibility. Usually
hearing becomes gradually less sensitive as frequency decreases and below 20 Hz or cycle per
second our ear become insensitive to sound. The study of such sound waves is referred to sometimes
as infrasonics, covering sounds beneath 20 Hz down to 0.1Hz and rarely to 0.001 Hz. This frequency
range is utilized for monitoring earthquakes, charting rock and petroleum formations below the earth,
and also in ballistocardiography and seismocardiography to study the mechanics of the heart.
Audible Sound
Vibrations transmitted through an elastic solid or a liquid or gas, with frequencies in the approximate
range of 20 to 20,000 hertz, capable of being detected by human organs of hearing are known as
audible sound.
Ultrasound
Ultrasounds are sound waves with frequencies higher than the upper audible (<20,000 Hz) limit of
human hearing. Ultrasound has the same physical property as audible sound, the only difference is
we cannot hear it.
Basic Ultrasound
Medical Ultrasound, also known as diagnostic sonography or ultrasonography, is a diagnostic
imaging technique based on the application of ultrasound. It is used to see internal body structures
such as tendons, muscles, joints, vessels and internal organs. Its aim is often to find a source of a
disease or to exclude any pathology. The practice of examining pregnant women using ultrasound is
called obstetric ultrasound, and is widely used.
Ultrasonic images are made by sending pulses of ultrasound into tissue using a probe. The
sound echoes off the tissue; with different tissues reflecting varying degrees of sound. These echoes
are recorded and displayed as an image to the operator.
Types of Ultrasound Images
Many different types of images can be formed using sonographic instruments. The most well-known
type is a B-mode image, which displays the acoustic impedance of a two-dimensional cross-section
of tissue. Other types of image can display blood flow, motion of tissue over time, the location of
blood, the presence of specific molecules, the stiffness of tissue, or the anatomy of a three-
dimensional region.
Generation of Ultrasound
An ultrasound wave is usually generated and detected by a piezoelectric crystal that mounted on
transducer, popularly known as ultrasound probe. (We will discuss about it later)
Piezoelectric crystals have an unique vice versa phenomena. If you put it in any electric field it will
deforms with the variation of the field. On the other hand it produces electric field in response to
mechanical stress.
An ultrasound wave is generated when an electric field is applied to an array of piezoelectric
crystals located on the transducer surface. Electrical stimulation causes mechanical distortion of the
crystals resulting in vibration and production of sound waves (i.e. mechanical energy).
Each piezoelectric crystal produces an ultrasound wave. The summation of all waves generated by
the piezoelectric crystals forms the ultrasound beam. Ultrasound waves are generated in pulses
(intermittent trains of pressure waves) and each pulse commonly consists of 2 or 3 sound cycles of
the same frequency.
Generation of Ultrasound
The pulse length (PL) is the distance traveled per pulse. Waves of short pulse lengths improve axial
resolution for ultrasound imaging. The PL cannot be reduced to less than 2 or 3 sound cycles by the
damping materials within the transducer.
Pulse Repetition Frequency (PRF) is the rate of pulses emitted by the transducer (number of pulses
per unit time). Ultrasound pulses must be spaced with enough time between pulses to permit the
sound to reach the target of interest and return to the transducer before the next pulse is generated.
The PRF for medical imaging ranges from 1-10 kHz. For example, if the PRF = 5 kHz and the time
between pulses is 0.2 msec, it will take 0.1 msec to reach the target and 0.1 msec to return to the
transducer. This means the pulse will travel 15.4 cm before the next pulse is emitted (1,540 m/sec x
0.1 msec = 0.154 m in 0.1 msec = 15.4 cm).
Modes of Ultrasound
The principal modes of ultrasound in echocardiography are:
1. 2-D or 2 dimensional mode
2. M-mode or motion mode
3. Colour flow Doppler imaging
4. Pulse wave Doppler
5. Continuous wave Doppler
6. Tissue Doppler
Effect of Ultrasound on Foetus
Having an ultrasound won't affect your baby. Ultrasound sends sound waves through your womb
(uterus), which bounce off your baby's body. The echoes are turned into an image on a screen, so
your sonographer can see your baby’s position and movements. The frequency or length of the sound
waves depends on how far along your pregnancy is and the type of scan being carried out.
Studies have found no link between ultrasound and birth weight, childhood cancers, dyslexia
or hearing.
Almost all women want a scan during pregnancy, so it's hard to find women who haven't had one
make a comparison. That in itself may reassure you.
Effect of Ultrasound on Fetus
During an ultrasound scan, the equipment generates a small amount of heat which is absorbed by the
part of the body that's being scanned. Antenatal scans produce less than one degree C. This means
they're fine for you and your baby. It's only if the temperature of the scanned body tissue rises by four
degrees C that harm may be caused.
The most routine type of scanning is used to get 2D pictures of your baby. This uses a low intensity
of ultrasound spread over a large area, which causes very minimal heating. What's more, the fluid
around your baby and any movements she may make will help to spread any heat. This may help to
reassure you further.
3D and 4D scans are just as safe as 2D scans, because the image is made up of sections of two-
dimensional images converted into a picture. The power intensity is the same as it is for 2D scanning.
Types of Ultrasound Probe
Probes are generally described by the size and shape of their face ("footprint"). Selecting the right
probe for the situation is essential to get good images, although there may be times where more than
one probe may be appropriate for a given exam. The essential element of each ultrasound transducer
is a piezoelectric crystal, serving both to generate and to receive ultrasound waves.
The ultrasound transducers differ in construction according to:
- piezoelectric crystal arrangement,
- aperture (footprint),
- operating frequency (which is directly related to the penetration depth)
The following types of transducers/probes are most often used in the critical ultrasound imaging:
1. sector or phased array,
2. linear, and
3. convex (standard or micro-convex).
Sector or Phased Array Probe
A phased array probe generates an image from an electronically steered beam in a close array,
generating an image that comes from a point and is good for getting between ribs such as in cardiac
ultrasound.
Figure: Sector or Phased Array Probe
Sector or Phased Array Probe
Features:
 Piezoelectric crystal arrangement: phased-array (most commonly used)
 Footprint size: small and field of view will be spreaded widely at
 Operating frequency (bandwidth): 1-5 MHz (usually 3.5-5 MHz)
 Element radius: 32 ch – 128 ch
 Ultrasound beam shape: sector, almost triangular
Uses:
 Abdominal application.
 Transesophageal application.
 Brain diagnosis,
 Echocardiography,
 Gynecological ultrasound, etc.
Linear or Vascular Probe
A linear (also sometimes called vascular) probes are generally high frequency, better for imaging
superficial structures and vessels.
Figure: Linear or Vascular Probe
Linear or Vascular Probe
Features:
 Piezoelectric crystal arrangement: linear
 Footprint size: usually big (small for the hockey transducers)
 Operating frequency (bandwidth): 3-12 mhz (usually 5-7.5 mhz)
 Ultrasound beam shape: rectangular
 Element radius: 64 ch – 256 ch
Uses:
 Vascular application:
 Arteria carotis
 MT,
FMD measurement for early detection of arterial sclerosis
Linear or Vascular Probe
Uses:
 venipuncture, blood vessel visualization
 Breast, Thyroid
 Tendon, arthrogenous
 Intraoperative, laparoscopy
 The thickness measurement of Body fat and Muscles for daily healthcare check and
locomotive syndrome check
 Photoacoustic imaging, ultrasonic velocity change imaging.
Convex or curvilinear Probe
A convex or curvilinear probes may have a wider footprint and lower frequency for transabdominal
imaging, or in a tighter array (wider field of view) and higher frequency for endocavitary imaging.
Figure: Convex or Curvilinear Probe
Linear or Vascular Probe
Features:
 piezoelectric crystal arrangement: curvilinear, along the aperture
 footprint size: big (small for the micro-convex transducers)
 operating frequency (bandwidth): 1-5 MHz (usually 3.5-5 MHz)
 ultrasound beam shape: sector; the ultrasound beam shape and size vary with distance from the
transducer, that causes the lack of lateral resolution at greater depths
 Element radius: 64 ch – 192 ch
Uses:
 Typical abdominal application.
 Transvaginal and transrectal application.
 Pelvic and lung (micro-convex transducer) ultrasound
 Diagnoses organs.
Ultrasound Probe Care
An ultrasound transducer is the most important and usually the most expensive element of the
ultrasound machine, so it should be used carefully, which means the following:
• Do not throw, drop or knock the transducer,
• Do not allow to spoil the transducer`s duct,
• Wipe the gel from the transducer after each use,
• Do not sluice with alcohol-based confections.
Image Orientation
Ultrasound orientation is essential to understand what is being seen in point-of-care ultrasound. There
are two key aspects: how the indicator is oriented relative to the screen and how the probe and
indicator are placed and oriented relative to the patient. The conventional ultrasound image is a two-
dimensional plane composed of frames made up of scan lines that are updated many times a second
to create a moving image.
Ultrasound orientation can be challenging because it involves understanding how a two-dimensional
plane may cut through a three-dimensional object in any of the three standard planes (sagittal,
transverse, or coronal), as well as any oblique orientation between those planes. However, one of the
great advantages of ultrasound is the ability to obtain images from a variety of different approaches,
and a thorough understanding of orientation will allow you to take advantage of this.
Indicator‐to-Screen Orientation
There are two rules for indicator-to-screen orientation in standard emergency ultrasound imaging,
which uses the same convention as general radiology imaging:
1. The top of the screen is closer to the probe. Bottom of the screen shows structures farther
away from the probe.
2. The left side of the screen, as it is viewed, corresponds to the side of the probe marked with
an indicator.
Note: Imaging performed by cardiology specialists uses a different and opposite convention for rule
2.
The "indicator" may differ widely between manufacturers, and is typically a bump or a groove. It is
important to verify your orientation prior to beginning any exam. This should be done by placing a
small amount of gel on the side of the probe where you believe the indicator to be, and confirming
that the side with the gel corresponds to the left of the screen as it is viewed (Figs. a and b).
Indicator‐to-Screen Orientation
There are two rules for indicator-to-screen orientation in standard emergency ultrasound imaging,
which uses the same convention as general radiology imaging:
1. The top of the screen is closer to the probe. Bottom of the screen shows structures farther
away from the probe.
2. The left side of the screen, as it is viewed, corresponds to the side of the probe marked with
an indicator.
Note: Imaging performed by cardiology specialists uses a different and opposite convention for rule
2.
The "indicator" may differ widely between manufacturers, and is typically a bump or a groove. It is
important to verify your orientation prior to beginning any exam. This should be done by placing a
small amount of gel on the side of the probe where you believe the indicator to be, and confirming
that the side with the gel corresponds to the left of the screen as it is viewed (Figs. a and b).
Indicator‐to-Screen Orientation
(a) Shows a thumb placed on the indicator, with gel placed on the face of the probe on that side.
(b) (b) Shows what should be seen on the screen in general imaging orientation.
Indicator‐to-Patient Orientation
Once the indicator-to-screen orientation is understood and verified, the probe is placed on the patient
and images are viewed on the screen.
The top of the screen will show structures closer to where the probe is placed, with the bottom of the
screen showing structures farther away from the probe face. The left of the screen will show
structures toward where the indicator is directed.
Image Planes
Longitudinal (Sagittal) - transducer is placed along the patient`s
long axis with the index directed cranial (the cranial structures
are displayed on the left side of the screen, i.e. the marked side of
the screen).
Transverse (Axial) - transducer is placed perpendicular to the
patient`s long axis with the index directed (usually) to the
patient`s right (the right structures are displayed on the left side
of the screen, i.e. the marked side of the screen).
Frontal (Coronal) - transducer is placed lateral with the index
directed cranial (the cranial structures are displayed on the left
side of the screen, i.e. the marked side of the screen; the
structures proximal to the probe are located at the top of the
screen).
Manipulation of Transducer
An image acquisition and optimization of the views is achieved the proper manipulation of the
transducer:
 tilting allows optimization of the image and imaging different structures in the same
tomographic plane,
 angulation of the transducer slides the ultrasound image from side to side,
 rotation the transducer at a single position changes the image plane.
An unexperienced sonographer should manipulate the probe carefully (limited movements) in one
image plane. The chaotic movements do not allow obtaining a reliable image views and proper data
acquisition.
Modes of Ultrasound
1. A-mode (A=amplitude)
The amplitude of reflected ultrasound is displayed on
an oscilloscope screen. It is just of historical
importance. The A-mode is now used only in
ophthalmology.
 Each pulse produces a new a one-dimensional
display or image
 Line of information on the display
 An uncommon display, except in
ophthalmologic sonography used
for precise intraocular length measurements Figure: A-mode Ultrasound
Modes of Ultrasound
2. B-mode (B=brightness) (2D in echocardiography)
This is now the essential imaging modality in the
diagnostic ultrasound. An amplitude of the reflected
ultrasound signals is converted into a gray scale image.
Owing to the wide gray scale (most of the ultrasound
machines use 256 shades of gray) even very small
differences in echogenicity are possible to visualize.
 Basis for gray scale,
two-dimensional (2D) imaging
 US unit tracks the position of the transducer to place a
dot on the screen corresponding to the transducer
position (X, Y locations), creating a 2D image
Figure: B-mode Ultrasound
Modes of Ultrasound
3. M-mode (M=motion)
It reflects a motion of the heart structures over
time. Nowadays, the integration of 2D and M-
mode images is possible. Due to its excellent
temporal resolution (high sampling rate), M-
mode is extremely valuable for accurate
evaluation of rapid movements.
• One-dimension image used to investigate
moving structures with respect to time
• Evaluates motion pattern of moving
structures such as in the heart
Figure: M-mode Ultrasound
Modes of Ultrasound
4. D-mode (D=Doppler)
This imaging mode is based on the Doppler
effect, i.e. change in frequency (Doppler shift)
caused by the reciprocal movement of the
sound generator and the observer. Diagnostic
ultrasound uses the change in frequency of
ultrasound signal backscattered from red blood
cells. The frequency of the reflected ultrasound
wave increases or decreases according to the
direction of blood flow in relation to the
transducer.
Figure: M-mode Ultrasound
Types of Doppler
Doppler Duplex
Doppler Duplex technique is based on the
simultaneous B-mode real-time and Doppler
imaging. A gray scale display serves for the
localization of flow measurement site. Both
spectral Doppler techniques (Continuous-Wave
and Pulsed) can be used in Doppler Duplex
imaging
Figure: Doppler Duplex Ultrasound
Types of Doppler
CW Doppler (Continuous-Wave Doppler)
(CWD)
This imaging mode requires two piezoelectric
crystals. One continuously transmits and the
other one receives the Doppler signals along the
scan-line. CWD is very useful in high velocity
signals recording. It defines blood flow
direction but has no value in Doppler signal
source identification.
Figure: CW Doppler Ultrasound
Types of Doppler
PW Doppler (Pulsed-Wave Doppler) (PWD)
The same piezoelectric crystal transmits and
receives the ultrasound signal. The reflected
signal returns to the transducer after a definite
delay that is known as pulse repetition
frequency (PRF). PWD allows measurement
from a small, specific blood volume (depth of
interest), which is defined by a sample volume.
It provides both blood flow direction and
precise determination of of Doppler signal
source. The main limitation is failure to display
the high velocity signals. The maximum
detectable frequency shift (the Nyquist limit) is
determined by the value of half the PRF. After
the velocity exceeds the limit, aliasing occurs Figure: PW Doppler Ultrasound
Types of Doppler
Color Doppler
Color Doppler imaging is based on the PWD
assumption. However, in Color Doppler
imaging, multiple sample volume are evaluated
along each scanning line. Velocities are
displayed using a color scale. Velocities toward
the transducer are red, and velocities away from
the transducer are blue.
Figure: Color Doppler Ultrasound
Types of Doppler
Power Doppler
Power Doppler is a variant of Color Doppler
technique which displays the magnitude of the
Doppler signal rather than the Doppler
frequency. It is often used to increase sensitivity
to low flows and velocities. Power Doppler
does not display flow direction or different
velocities.
Figure: Power Doppler Ultrasound
Artifacts
An ultrasound artifact is a structure in an image which does not directly similar with actual tissue
being scanned.
Artifact assumes different forms including :
 Structures in the image that are not actually present
 Objects that should be represented but are missing from the image.
 Structures which are misregistered on the image.
Common Artifacts
1. Reverberation
2. Acoustic Shadowing
3. Acoustic Enhancement
4. Edge Shadowing
5. Beam Width Artifact
6. Slice Thickness Artifact
7. Side Lobe Artifact
8. Mirror Image
9. Double Image
10. Equipment-generated Artifact
11. Refraction Artifact
REVERBERATION
 This is the production of false echoes due to repeated reflections between two interfaces with a
high acoustic impedance mismatch.
 The echo from the interface is received by the transducer and displayed on the image.
 Some of the energy in the returned echo is reflected at the transducer face, and return to the
reflecting interface as if it was a weak transmitted pulse, returning as a second echo.
 As the time taken for the second echo to arrive is twice that taken by the depth.
 This sequence of reflection and transmission can occur many times, with the third echo taking
three times as long to return to the transducer and being displayed at three times the depth, and so
on.
 The reverberation echoes will be equally spaced because the time for each additional echo is
multiple of the time of return of the first echo.
 These reverberation echoes will be strong because of the high acoustic mismatch.
 This artifact will be seen at the skin-transducer interface and behind bowel gas.
REVERBERATION
Rectification:
 Increase the amount of gel used.
 Use a stand-off pad.
 Reduce the gain.
 Move the position of the transducer.
ACOUSTIC SHADOWING
 This appears as an area of low amplitude echoes
behind an area of strongly attenuating tissue.
 It is caused by severe attenuation of the beam at
an interface, resulting in very little sound being
transmitted beyond.
 The attenuation can be due to either absorption
or reflection of the sound waves, or a
combination of the two.
 Acoustic shadowing will occur at interfaces with
large acoustic mismatch such as:
Soft tissue and gas
Soft tissue and bone or calculus
ACOUSTIC SHADOWING
ACOUSTIC ENHANCEMENT
 This artifact appears as a localized area of increased echo amplitude behind an area of low
attenuation.
 On a scan it will appears as an area of increased brightness, and can commonly be seen distal to
fluid-filled structures such as the urinary bladder, GB or a cyst.
 The artifact arises due to the application of the time-gain compensation(TGC) to areas of low
attenuating structures such as fluid.
 It is caused by the low level of attenuation of the beam as it passes through fluid relative to the
greater attenuation of the beam in the adjacent more solid tissue.
 This artifact can often be an useful diagnostic aid, particularly when scanning a soft-tissue mass
or cyst containing low level echoes.
 These echoes may often cause the structure to disappear in the image as it blend into the
surrounding echo pattern.
ACOUSTIC ENHANCEMENT
EDGE SHADOWING
 A combination of refraction and reflection occurring at the edges of rounded structures will result
in edge shadowing artifact.
 It arises due to refraction of the beam caused by both the curvature of the rounded edges and
difference in speed of two materials.
 When the ultrasound beam reaches the rounded edge of a structure, reflection will occur, with an
angle of incidence equal to the angle of reflection.
 The outer part of the beam will be totally reflected, but the reminder of the beam passes through
the rounded structure and is refracted.
 This combination of reflection and refraction of the beam at the edges of a rounded structure
results in a thin strip of tissue behind the edge not being insonated and causes a shadow.
EDGE SHADOWING
BEAM WIDTH ARTIFACT
 This artifact can be
demonstrated by scanning a
point reflector in a phantom,
where the display will clearly
portray this as a line.
 During routine scanning, the
artifact can be seen when
spurious echoes are displayed
in an echo-free area.
 Correct positioning of the focal
zone will help to reduce this
artifact.
 The focal zone is controlled by
electronically narrowing the
beam
SLICE THICKNESS ARTIFACT
 These occurs due to the thickness of the beam, and are similar to beam width artifacts.
 These artifacts will typically be seen in transverse views of the urinary bladder when structures
adjacent to the slice through the bladder being scanned will be incorporated into the image.
 These echoes are then displayed as if they were arising from within the bladder.
 Although the appearance of this artifact is similar to the beam width artifact, the differentiating
factors is that the reflector causing the slice thickness artifact will not be seen on the display.
 This artifact is a result of inherent characteristics of the transducer, and apart from trying a
different transducer, cannot be eliminated.
SIDE LOBE ARTIFACT
 The energy within the ultrasound
beam exists as several side lobes
radiating at a number of angles from
a central lobe.
 Echoes are generated by these side
lobes in addition to the main lobe,
but all the returning echoes are
assumed by the transducer to have
arisen from the central axis of the
main lobe.
 Side lobe echoes will therefore be
misregistered in the display.
 This artifact can often be seen in
area such as the urinary bladder and
may also arise within a cyst.
MIRROR IMAGE ARTIFACT
 These artifacts results in a mirror image of a structure occurring in an ultrasound display.
 They arise due to specular reflection of the beam at a large smooth interface.
 An area close to a specular reflector will be imaged twice, once by the original ultrasound beam
and once by the beam after it has reflected off the specular reflector.
 Mirror image artifacts are most commonly seen where there is a large acoustic mismatch, such as a
fluid-air interface.
 Typically this artifact can occur during the scanning of a full bladder, when air in the rectum
behind the bladder act as specular reflector and mirror image of the bladder is displayed
posteriorly.
 It will then have the appearance of a large cyst behind the bladder.
 It can also be seen when scanning the liver, and the diaphragm act as a specular reflector.
MIRROR IMAGE ARTIFACT
DOUBLE IMAGE ARTIFACT
 This image is caused by refraction of the beam
and may occur in areas such as the rectus
abdominis muscle on the anterior abdominal
wall.
 In the transverse plane the edges of the muscle
act as a lens and the ultrasound beam to be
refracted and this causes the single structure to
be interrogated by two separate refracted
beams.
 Two sets of echoes will therefore be returned
and these will cause display of two structures
in the image. This results in, for example, two
images of the transverse aorta side by side in
the abdomen.
EQUIPMENT GENERATED ARTIFACT
 Incorrect use of the equipment controls can lead to artifact appearing.
 Misuse of controls such as the gain or TGC can result in echoes being recorded as too bright or too
dark.
 Care must be taken when setting these controls, to ensure an even brightness throughout the image.
 If the dynamic range control is incorrectly set, this can lead to an image which has too much
contrast, and result in the loss of subtle echo information.
 Gain must be in medium level.
 Blurring of a moving image can occur if the frame rate is too low or if the persistence is too high.
 It is important to ensure that the frame rate is capable of recording a moving structure at the speed.
 Use of multiple focal zones can give rise to a prominent banding effect within the image.
EQUIPMENT GENERATED ARTIFACT
REFRACTION ARTIFCT
 The refraction is the change of the sound direction on passing from one medium to another.
 In ultrasound, refraction is due to sound velocity mismatches combines with oblique angles of
incidence, most commonly with convex scanheads.
 When the ultrasound wave crosses at an oblique angle the interface of two materials, through
which the waves propagate at different velocities, refraction occurs, caused by bending of the wave
beam.
 Refraction artifact cause spatial distortion and loss of resolution in the image.

More Related Content

What's hot

Ultrasound imaging system
Ultrasound imaging system Ultrasound imaging system
Ultrasound imaging system Riadh Al-Haidari
 
Physic Of Ultrasound
Physic Of UltrasoundPhysic Of Ultrasound
Physic Of UltrasoundKhalis Karim
 
Physical Principles Of Ultrasound
Physical Principles Of UltrasoundPhysical Principles Of Ultrasound
Physical Principles Of Ultrasoundu.surgery
 
Types of Ultrasound Probes
Types of Ultrasound ProbesTypes of Ultrasound Probes
Types of Ultrasound Probeshafsatalat1
 
ultrasound physics
ultrasound physicsultrasound physics
ultrasound physicsRakesh Ca
 
Principles of Ultrasound Physics.pptx
Principles of Ultrasound Physics.pptxPrinciples of Ultrasound Physics.pptx
Principles of Ultrasound Physics.pptxGemechisAsefa1
 
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptxLec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptxAli Ayaz
 
MRI Parameters for Imaging
MRI Parameters for ImagingMRI Parameters for Imaging
MRI Parameters for ImagingAnjan Dangal
 
Physics of ultrasound imaging
Physics of ultrasound imagingPhysics of ultrasound imaging
Physics of ultrasound imagingmtuzer
 
Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)Prajwith Rai
 
Ultrasound Transducer Types
Ultrasound Transducer TypesUltrasound Transducer Types
Ultrasound Transducer TypesMette Lyng
 
ultrasound transducers and resolution
ultrasound transducers and resolutionultrasound transducers and resolution
ultrasound transducers and resolutionVallabhaneni Bhupal
 
Doppler Effect - Ultrasound
Doppler Effect - UltrasoundDoppler Effect - Ultrasound
Doppler Effect - UltrasoundVictor Ekpo
 
Understanding ultrasound
Understanding ultrasoundUnderstanding ultrasound
Understanding ultrasoundMohammad Amir
 
Image reconstrsuction in ct pdf
Image reconstrsuction in ct pdfImage reconstrsuction in ct pdf
Image reconstrsuction in ct pdfmitians
 

What's hot (20)

Ultrasound imaging system
Ultrasound imaging system Ultrasound imaging system
Ultrasound imaging system
 
Physic Of Ultrasound
Physic Of UltrasoundPhysic Of Ultrasound
Physic Of Ultrasound
 
Physical Principles Of Ultrasound
Physical Principles Of UltrasoundPhysical Principles Of Ultrasound
Physical Principles Of Ultrasound
 
Types of Ultrasound Probes
Types of Ultrasound ProbesTypes of Ultrasound Probes
Types of Ultrasound Probes
 
ultrasound physics
ultrasound physicsultrasound physics
ultrasound physics
 
Ultrasound physics
Ultrasound physicsUltrasound physics
Ultrasound physics
 
Principles of Ultrasound Physics.pptx
Principles of Ultrasound Physics.pptxPrinciples of Ultrasound Physics.pptx
Principles of Ultrasound Physics.pptx
 
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptxLec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
 
MRI Parameters for Imaging
MRI Parameters for ImagingMRI Parameters for Imaging
MRI Parameters for Imaging
 
Physics of ultrasound imaging
Physics of ultrasound imagingPhysics of ultrasound imaging
Physics of ultrasound imaging
 
Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)
 
Ultrasound Transducer Types
Ultrasound Transducer TypesUltrasound Transducer Types
Ultrasound Transducer Types
 
ultrasound transducers and resolution
ultrasound transducers and resolutionultrasound transducers and resolution
ultrasound transducers and resolution
 
Basics Physics of ultrasound
Basics Physics of ultrasoundBasics Physics of ultrasound
Basics Physics of ultrasound
 
Doppler Effect - Ultrasound
Doppler Effect - UltrasoundDoppler Effect - Ultrasound
Doppler Effect - Ultrasound
 
Understanding ultrasound
Understanding ultrasoundUnderstanding ultrasound
Understanding ultrasound
 
Image reconstrsuction in ct pdf
Image reconstrsuction in ct pdfImage reconstrsuction in ct pdf
Image reconstrsuction in ct pdf
 
Mri safety
Mri safetyMri safety
Mri safety
 
Mri artifacts
Mri artifactsMri artifacts
Mri artifacts
 
Usg artifacts
Usg artifactsUsg artifacts
Usg artifacts
 

Viewers also liked

Poly Cystic Ovary Syndrome (PCOS)- Live a Stress Free, Active life with Homeo...
Poly Cystic Ovary Syndrome (PCOS)- Live a Stress Free, Active life with Homeo...Poly Cystic Ovary Syndrome (PCOS)- Live a Stress Free, Active life with Homeo...
Poly Cystic Ovary Syndrome (PCOS)- Live a Stress Free, Active life with Homeo...Welcome Cure LLP
 
Clinical Applications of Chest Sonography
Clinical Applications of Chest SonographyClinical Applications of Chest Sonography
Clinical Applications of Chest SonographyGamal Agmy
 
Ultrasound Physics & Knobology
Ultrasound Physics & KnobologyUltrasound Physics & Knobology
Ultrasound Physics & Knobologyu.surgery
 
Ultrasound diagnostics fin
Ultrasound diagnostics finUltrasound diagnostics fin
Ultrasound diagnostics finMUBOSScz
 
Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifactsansaripv
 
GEMC- Right Upper Quadrant Ultrasound- Resident Training
GEMC- Right Upper Quadrant Ultrasound- Resident TrainingGEMC- Right Upper Quadrant Ultrasound- Resident Training
GEMC- Right Upper Quadrant Ultrasound- Resident TrainingOpen.Michigan
 

Viewers also liked (6)

Poly Cystic Ovary Syndrome (PCOS)- Live a Stress Free, Active life with Homeo...
Poly Cystic Ovary Syndrome (PCOS)- Live a Stress Free, Active life with Homeo...Poly Cystic Ovary Syndrome (PCOS)- Live a Stress Free, Active life with Homeo...
Poly Cystic Ovary Syndrome (PCOS)- Live a Stress Free, Active life with Homeo...
 
Clinical Applications of Chest Sonography
Clinical Applications of Chest SonographyClinical Applications of Chest Sonography
Clinical Applications of Chest Sonography
 
Ultrasound Physics & Knobology
Ultrasound Physics & KnobologyUltrasound Physics & Knobology
Ultrasound Physics & Knobology
 
Ultrasound diagnostics fin
Ultrasound diagnostics finUltrasound diagnostics fin
Ultrasound diagnostics fin
 
Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifacts
 
GEMC- Right Upper Quadrant Ultrasound- Resident Training
GEMC- Right Upper Quadrant Ultrasound- Resident TrainingGEMC- Right Upper Quadrant Ultrasound- Resident Training
GEMC- Right Upper Quadrant Ultrasound- Resident Training
 

Similar to Ultrasound physics

Vibrational Biophysics Iqqm Morgan
Vibrational Biophysics Iqqm MorganVibrational Biophysics Iqqm Morgan
Vibrational Biophysics Iqqm Morgansolarsonic
 
Programmed learning material
Programmed learning materialProgrammed learning material
Programmed learning material14162083
 
Programmed learning material
Programmed learning materialProgrammed learning material
Programmed learning material9924066834
 
Q-3 Transverse vs Longitudinal waves Science 7 - Week 3-4.pptx
Q-3 Transverse vs Longitudinal waves Science 7 - Week 3-4.pptxQ-3 Transverse vs Longitudinal waves Science 7 - Week 3-4.pptx
Q-3 Transverse vs Longitudinal waves Science 7 - Week 3-4.pptxRoseAnnFabialaLeanil
 
PHYS 103Report 3 InstructionsSound and ResonanceDownload thi.docx
PHYS 103Report 3 InstructionsSound and ResonanceDownload thi.docxPHYS 103Report 3 InstructionsSound and ResonanceDownload thi.docx
PHYS 103Report 3 InstructionsSound and ResonanceDownload thi.docxkarlhennesey
 
Module-3-The-Waves - the different types of waves
Module-3-The-Waves - the different types of wavesModule-3-The-Waves - the different types of waves
Module-3-The-Waves - the different types of wavesJunBryanAcob
 
3.1 form 4 general wave properties
3.1 form 4 general wave properties3.1 form 4 general wave properties
3.1 form 4 general wave propertieschris lembalemba
 
Y11 phy270115wavemot
Y11 phy270115wavemotY11 phy270115wavemot
Y11 phy270115wavemot3SNEducation
 
Physics ii djy 2013 ppt wave characteristics
Physics ii djy 2013 ppt    wave characteristicsPhysics ii djy 2013 ppt    wave characteristics
Physics ii djy 2013 ppt wave characteristicsDavid Young
 
03-03-08 - Waves
03-03-08 - Waves03-03-08 - Waves
03-03-08 - Waveswjerlinger
 
Unit 4 2014 ppt wave characteristics
Unit 4 2014  ppt    wave characteristicsUnit 4 2014  ppt    wave characteristics
Unit 4 2014 ppt wave characteristicsDavid Young
 
General properties of wave
General properties of waveGeneral properties of wave
General properties of wavesunnyiek
 
3.1 form 4 general wave properties
3.1 form 4 general wave properties3.1 form 4 general wave properties
3.1 form 4 general wave propertieschris lembalemba
 

Similar to Ultrasound physics (20)

Vibrational Biophysics Iqqm Morgan
Vibrational Biophysics Iqqm MorganVibrational Biophysics Iqqm Morgan
Vibrational Biophysics Iqqm Morgan
 
Programmed learning material
Programmed learning materialProgrammed learning material
Programmed learning material
 
Programmed learning material
Programmed learning materialProgrammed learning material
Programmed learning material
 
Q-3 Transverse vs Longitudinal waves Science 7 - Week 3-4.pptx
Q-3 Transverse vs Longitudinal waves Science 7 - Week 3-4.pptxQ-3 Transverse vs Longitudinal waves Science 7 - Week 3-4.pptx
Q-3 Transverse vs Longitudinal waves Science 7 - Week 3-4.pptx
 
PHYS 103Report 3 InstructionsSound and ResonanceDownload thi.docx
PHYS 103Report 3 InstructionsSound and ResonanceDownload thi.docxPHYS 103Report 3 InstructionsSound and ResonanceDownload thi.docx
PHYS 103Report 3 InstructionsSound and ResonanceDownload thi.docx
 
Module-3-The-Waves - the different types of waves
Module-3-The-Waves - the different types of wavesModule-3-The-Waves - the different types of waves
Module-3-The-Waves - the different types of waves
 
Wave Motion
Wave MotionWave Motion
Wave Motion
 
Waves
WavesWaves
Waves
 
Physics P1.5
Physics  P1.5Physics  P1.5
Physics P1.5
 
3.1 form 4 general wave properties
3.1 form 4 general wave properties3.1 form 4 general wave properties
3.1 form 4 general wave properties
 
Y11 phy270115wavemot
Y11 phy270115wavemotY11 phy270115wavemot
Y11 phy270115wavemot
 
Physics ii djy 2013 ppt wave characteristics
Physics ii djy 2013 ppt    wave characteristicsPhysics ii djy 2013 ppt    wave characteristics
Physics ii djy 2013 ppt wave characteristics
 
03-03-08 - Waves
03-03-08 - Waves03-03-08 - Waves
03-03-08 - Waves
 
Waves
WavesWaves
Waves
 
Waves ii
Waves iiWaves ii
Waves ii
 
Chapter 1 waves
Chapter 1 wavesChapter 1 waves
Chapter 1 waves
 
Unit 4 2014 ppt wave characteristics
Unit 4 2014  ppt    wave characteristicsUnit 4 2014  ppt    wave characteristics
Unit 4 2014 ppt wave characteristics
 
General properties of wave
General properties of waveGeneral properties of wave
General properties of wave
 
Waves ch. 8
Waves ch. 8Waves ch. 8
Waves ch. 8
 
3.1 form 4 general wave properties
3.1 form 4 general wave properties3.1 form 4 general wave properties
3.1 form 4 general wave properties
 

Recently uploaded

Solving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxSolving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxOH TEIK BIN
 
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptxECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptxiammrhaywood
 
Employee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxEmployee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxNirmalaLoungPoorunde1
 
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdfssuser54595a
 
Final demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptxFinal demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptxAvyJaneVismanos
 
Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...jaredbarbolino94
 
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...Marc Dusseiller Dusjagr
 
Painted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaPainted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaVirag Sontakke
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptxVS Mahajan Coaching Centre
 
CELL CYCLE Division Science 8 quarter IV.pptx
CELL CYCLE Division Science 8 quarter IV.pptxCELL CYCLE Division Science 8 quarter IV.pptx
CELL CYCLE Division Science 8 quarter IV.pptxJiesonDelaCerna
 
Capitol Tech U Doctoral Presentation - April 2024.pptx
Capitol Tech U Doctoral Presentation - April 2024.pptxCapitol Tech U Doctoral Presentation - April 2024.pptx
Capitol Tech U Doctoral Presentation - April 2024.pptxCapitolTechU
 
MICROBIOLOGY biochemical test detailed.pptx
MICROBIOLOGY biochemical test detailed.pptxMICROBIOLOGY biochemical test detailed.pptx
MICROBIOLOGY biochemical test detailed.pptxabhijeetpadhi001
 
DATA STRUCTURE AND ALGORITHM for beginners
DATA STRUCTURE AND ALGORITHM for beginnersDATA STRUCTURE AND ALGORITHM for beginners
DATA STRUCTURE AND ALGORITHM for beginnersSabitha Banu
 
Procuring digital preservation CAN be quick and painless with our new dynamic...
Procuring digital preservation CAN be quick and painless with our new dynamic...Procuring digital preservation CAN be quick and painless with our new dynamic...
Procuring digital preservation CAN be quick and painless with our new dynamic...Jisc
 
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPTECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPTiammrhaywood
 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxRaymartEstabillo3
 
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdfEnzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdfSumit Tiwari
 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon AUnboundStockton
 

Recently uploaded (20)

Solving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxSolving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptx
 
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptxECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
 
Employee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxEmployee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptx
 
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
 
Final demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptxFinal demo Grade 9 for demo Plan dessert.pptx
Final demo Grade 9 for demo Plan dessert.pptx
 
Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...
 
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
 
Painted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaPainted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of India
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
 
CELL CYCLE Division Science 8 quarter IV.pptx
CELL CYCLE Division Science 8 quarter IV.pptxCELL CYCLE Division Science 8 quarter IV.pptx
CELL CYCLE Division Science 8 quarter IV.pptx
 
ESSENTIAL of (CS/IT/IS) class 06 (database)
ESSENTIAL of (CS/IT/IS) class 06 (database)ESSENTIAL of (CS/IT/IS) class 06 (database)
ESSENTIAL of (CS/IT/IS) class 06 (database)
 
Model Call Girl in Bikash Puri Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Bikash Puri  Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Bikash Puri  Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Bikash Puri Delhi reach out to us at 🔝9953056974🔝
 
Capitol Tech U Doctoral Presentation - April 2024.pptx
Capitol Tech U Doctoral Presentation - April 2024.pptxCapitol Tech U Doctoral Presentation - April 2024.pptx
Capitol Tech U Doctoral Presentation - April 2024.pptx
 
MICROBIOLOGY biochemical test detailed.pptx
MICROBIOLOGY biochemical test detailed.pptxMICROBIOLOGY biochemical test detailed.pptx
MICROBIOLOGY biochemical test detailed.pptx
 
DATA STRUCTURE AND ALGORITHM for beginners
DATA STRUCTURE AND ALGORITHM for beginnersDATA STRUCTURE AND ALGORITHM for beginners
DATA STRUCTURE AND ALGORITHM for beginners
 
Procuring digital preservation CAN be quick and painless with our new dynamic...
Procuring digital preservation CAN be quick and painless with our new dynamic...Procuring digital preservation CAN be quick and painless with our new dynamic...
Procuring digital preservation CAN be quick and painless with our new dynamic...
 
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPTECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
 
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdfEnzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon A
 

Ultrasound physics

  • 1. Ultrasound Physics Md Serajus Salekin Chowdhury Senior Scientific Officer Bangladesh Atomic Energy Commission
  • 2. Waves A wave can be described as a disturbance that travels through a medium from one location to another location. For an example, when a pebble is thrown in a pond, it creates a disturbance and forms ripples which travel together as a front in a straight-line direction, or the waves may be circular waves that originate from the point where the disturbances occurs in the pond. In physics, the wave is defined as the energy transferred through medium with regular vibration or oscillating motion. A few examples of waves are: water wave, light wave, electromagnetic wave, sound wave, seismic wave (earthquakes). One way to categorize waves is on the basis of the direction of movement of the distinct particles of the medium relative to the direction that the waves travel. Waves can be classified into two types Longitudinal Wave and Transverse Wave. Figure : Waves
  • 3. Longitudinal Wave Longitudinal waves occur when the oscillations are parallel to the direction of propagation. It is a wave in which particles of the medium move in a direction parallel to the direction that the wave moves. A sound wave traveling through air is a classic example of a longitudinal wave. Figure: Longitudinal Wave
  • 4. Transverse Wave Transverse waves occur when a disturbance creates oscillations perpendicular at right angles to the direction of energy transfer. It is a wave in which particles of the medium move in a direction perpendicular to the direction that the wave moves. The electromagnetic radiations which include light waves are a classic example of transverse waves. Figure: Transverse Wave
  • 5. Amplitude Amplitude of a wave is defined as the distance from the rest position of the vibrating particle of the medium to the highest point of the crest or the lowest point of the trough. It is denoted with a or A and measured in meter. Figure: Amplitude
  • 6. Wavelength Waves make characteristic patterns as they travel through space. The wavelength, λ, of a wave is the distance from any point on one wave to the same point on the next wave along. (The symbol is a Greek letter, 'lambda'.) To avoid confusion, it is best to measure wavelength from the top of a crest to the top of the next crest, or from the bottom of a trough to the bottom of the next trough. Wavelength is measured in meter. Figure: Wavelength
  • 7. Wavelength Since a longitudinal wave does not contain crests and troughs, its wavelength must be measured differently. A longitudinal wave consists of a repeating pattern of compressions and rarefactions. Thus, the wavelength is commonly measured as the distance from one compression to the next adjacent compression or the distance from one rarefaction to the next adjacent rarefaction. Figure: Wavelength
  • 8. Frequency The frequency, f, of a wave is the number of waves passing a point in a certain time. We normally use a time of one second, so we can say, “the number of waves passing a point in one second is called frequency of the wave”. The unit of frequency is hertz (Hz), i.e., one hertz is equal to one wave per second. For water waves and sound waves the unit hertz is usually good enough but radio and TV waves have such a high frequency that the kilohertz (kHz) or even the megahertz (MHz) are better units. 1 kHz = 1,000 Hz 1 MHz = 1,000,000 Hz
  • 9. Time Period Time taken for one complete cycle or vibration by the wave generating source is called time period. In other word it can be defined as the time taken by a wave to travel the distance of a complete wavelength. It is simply reciprocal of the frequency (1/f). Unit of time period is second. 𝑇𝑖𝑚𝑒 𝑃𝑒𝑟𝑖𝑜𝑑 𝑇 = 1 𝐹𝑟𝑒𝑞𝑢𝑒𝑛𝑐𝑦 (𝑓)
  • 10. Wave Speed/ Wave Velocity The wave speed/velocity is measured as the distance travelled by the crests or trough in a given time interval. The relationship is similar to the relationship of speed, distance and time in the classical physics. Therefore, wave speed/wave velocity can be represented mathematically as, SW = DW /T Where, SW = speed of the wave DW = distance travelled by crest or trough of the wave T = time taken to travel the distance DW The wave speed varies for different types of waves. The Speed of light wave is 3 ×× 108108 m/s.
  • 11. Frequency, Wavelength & Wave velocity Relationship The diagrams at the right show several "snapshots" of the production of a wave within a rope. The motion of the disturbance along the medium after every one-fourth of a period is depicted. Observe that in the time it takes from the first to the last snapshot, the hand has made one complete back-and-forth motion. A period has elapsed. Observe that during this same amount of time, the leading edge of the disturbance has moved a distance equal to one complete wavelength. So in a time of one period, the wave has moved a distance of one wavelength. Combining this information with the equation for speed (speed = distance/time), it can be said that the speed of a wave is also the wavelength/period. Since the period is the reciprocal of the frequency, the expression 1/f can be substituted into the above equation for period. Rearranging the equation yields a new equation of the form: Speed = Wavelength • Frequency The above equation is known as the wave equation. It states the mathematical relationship between the speed (v) of a wave and its wavelength (λ) and frequency (f). Using the symbols v, λ, and f, the equation can be rewritten as v = f • λ
  • 12. Sound Sound is a mechanical wave that results from the back and forth vibration of the particles of the medium through which the sound wave is moving. If a sound wave is moving from left to right through air, then particles of air will be displaced both rightward and leftward as the energy of the sound wave passes through it. The motion of the particles is parallel (and anti-parallel) to the direction of the energy transport. This is what characterizes sound waves in air as longitudinal waves. Because of the longitudinal motion of the air particles, there are regions in the air where the air particles are compressed together and other regions where the air particles are spread apart. These regions are known as compressions and rarefactions respectively. The compressions are regions of high air pressure while the rarefactions are regions of low air pressure. The diagram below depicts a sound wave created by a tuning fork and propagated through the air in an open tube. The compressions and rarefactions are labeled. Figure: Propagation of sound
  • 13. Types of Sound Wave According to audibility, more precisely frequency limit, sound waves are grouped in three classes: 1. Infrasound, 2. Audible sound, and 3.Ultrasound. Infrasound Infrasound are the sound waves with frequencies below the lower limit of human audibility. Usually hearing becomes gradually less sensitive as frequency decreases and below 20 Hz or cycle per second our ear become insensitive to sound. The study of such sound waves is referred to sometimes as infrasonics, covering sounds beneath 20 Hz down to 0.1Hz and rarely to 0.001 Hz. This frequency range is utilized for monitoring earthquakes, charting rock and petroleum formations below the earth, and also in ballistocardiography and seismocardiography to study the mechanics of the heart.
  • 14. Audible Sound Vibrations transmitted through an elastic solid or a liquid or gas, with frequencies in the approximate range of 20 to 20,000 hertz, capable of being detected by human organs of hearing are known as audible sound. Ultrasound Ultrasounds are sound waves with frequencies higher than the upper audible (<20,000 Hz) limit of human hearing. Ultrasound has the same physical property as audible sound, the only difference is we cannot hear it.
  • 15. Basic Ultrasound Medical Ultrasound, also known as diagnostic sonography or ultrasonography, is a diagnostic imaging technique based on the application of ultrasound. It is used to see internal body structures such as tendons, muscles, joints, vessels and internal organs. Its aim is often to find a source of a disease or to exclude any pathology. The practice of examining pregnant women using ultrasound is called obstetric ultrasound, and is widely used. Ultrasonic images are made by sending pulses of ultrasound into tissue using a probe. The sound echoes off the tissue; with different tissues reflecting varying degrees of sound. These echoes are recorded and displayed as an image to the operator. Types of Ultrasound Images Many different types of images can be formed using sonographic instruments. The most well-known type is a B-mode image, which displays the acoustic impedance of a two-dimensional cross-section of tissue. Other types of image can display blood flow, motion of tissue over time, the location of blood, the presence of specific molecules, the stiffness of tissue, or the anatomy of a three- dimensional region.
  • 16. Generation of Ultrasound An ultrasound wave is usually generated and detected by a piezoelectric crystal that mounted on transducer, popularly known as ultrasound probe. (We will discuss about it later) Piezoelectric crystals have an unique vice versa phenomena. If you put it in any electric field it will deforms with the variation of the field. On the other hand it produces electric field in response to mechanical stress. An ultrasound wave is generated when an electric field is applied to an array of piezoelectric crystals located on the transducer surface. Electrical stimulation causes mechanical distortion of the crystals resulting in vibration and production of sound waves (i.e. mechanical energy). Each piezoelectric crystal produces an ultrasound wave. The summation of all waves generated by the piezoelectric crystals forms the ultrasound beam. Ultrasound waves are generated in pulses (intermittent trains of pressure waves) and each pulse commonly consists of 2 or 3 sound cycles of the same frequency.
  • 17. Generation of Ultrasound The pulse length (PL) is the distance traveled per pulse. Waves of short pulse lengths improve axial resolution for ultrasound imaging. The PL cannot be reduced to less than 2 or 3 sound cycles by the damping materials within the transducer. Pulse Repetition Frequency (PRF) is the rate of pulses emitted by the transducer (number of pulses per unit time). Ultrasound pulses must be spaced with enough time between pulses to permit the sound to reach the target of interest and return to the transducer before the next pulse is generated. The PRF for medical imaging ranges from 1-10 kHz. For example, if the PRF = 5 kHz and the time between pulses is 0.2 msec, it will take 0.1 msec to reach the target and 0.1 msec to return to the transducer. This means the pulse will travel 15.4 cm before the next pulse is emitted (1,540 m/sec x 0.1 msec = 0.154 m in 0.1 msec = 15.4 cm).
  • 18. Modes of Ultrasound The principal modes of ultrasound in echocardiography are: 1. 2-D or 2 dimensional mode 2. M-mode or motion mode 3. Colour flow Doppler imaging 4. Pulse wave Doppler 5. Continuous wave Doppler 6. Tissue Doppler
  • 19. Effect of Ultrasound on Foetus Having an ultrasound won't affect your baby. Ultrasound sends sound waves through your womb (uterus), which bounce off your baby's body. The echoes are turned into an image on a screen, so your sonographer can see your baby’s position and movements. The frequency or length of the sound waves depends on how far along your pregnancy is and the type of scan being carried out. Studies have found no link between ultrasound and birth weight, childhood cancers, dyslexia or hearing. Almost all women want a scan during pregnancy, so it's hard to find women who haven't had one make a comparison. That in itself may reassure you.
  • 20. Effect of Ultrasound on Fetus During an ultrasound scan, the equipment generates a small amount of heat which is absorbed by the part of the body that's being scanned. Antenatal scans produce less than one degree C. This means they're fine for you and your baby. It's only if the temperature of the scanned body tissue rises by four degrees C that harm may be caused. The most routine type of scanning is used to get 2D pictures of your baby. This uses a low intensity of ultrasound spread over a large area, which causes very minimal heating. What's more, the fluid around your baby and any movements she may make will help to spread any heat. This may help to reassure you further. 3D and 4D scans are just as safe as 2D scans, because the image is made up of sections of two- dimensional images converted into a picture. The power intensity is the same as it is for 2D scanning.
  • 21. Types of Ultrasound Probe Probes are generally described by the size and shape of their face ("footprint"). Selecting the right probe for the situation is essential to get good images, although there may be times where more than one probe may be appropriate for a given exam. The essential element of each ultrasound transducer is a piezoelectric crystal, serving both to generate and to receive ultrasound waves. The ultrasound transducers differ in construction according to: - piezoelectric crystal arrangement, - aperture (footprint), - operating frequency (which is directly related to the penetration depth) The following types of transducers/probes are most often used in the critical ultrasound imaging: 1. sector or phased array, 2. linear, and 3. convex (standard or micro-convex).
  • 22. Sector or Phased Array Probe A phased array probe generates an image from an electronically steered beam in a close array, generating an image that comes from a point and is good for getting between ribs such as in cardiac ultrasound. Figure: Sector or Phased Array Probe
  • 23. Sector or Phased Array Probe Features:  Piezoelectric crystal arrangement: phased-array (most commonly used)  Footprint size: small and field of view will be spreaded widely at  Operating frequency (bandwidth): 1-5 MHz (usually 3.5-5 MHz)  Element radius: 32 ch – 128 ch  Ultrasound beam shape: sector, almost triangular Uses:  Abdominal application.  Transesophageal application.  Brain diagnosis,  Echocardiography,  Gynecological ultrasound, etc.
  • 24. Linear or Vascular Probe A linear (also sometimes called vascular) probes are generally high frequency, better for imaging superficial structures and vessels. Figure: Linear or Vascular Probe
  • 25. Linear or Vascular Probe Features:  Piezoelectric crystal arrangement: linear  Footprint size: usually big (small for the hockey transducers)  Operating frequency (bandwidth): 3-12 mhz (usually 5-7.5 mhz)  Ultrasound beam shape: rectangular  Element radius: 64 ch – 256 ch Uses:  Vascular application:  Arteria carotis  MT, FMD measurement for early detection of arterial sclerosis
  • 26. Linear or Vascular Probe Uses:  venipuncture, blood vessel visualization  Breast, Thyroid  Tendon, arthrogenous  Intraoperative, laparoscopy  The thickness measurement of Body fat and Muscles for daily healthcare check and locomotive syndrome check  Photoacoustic imaging, ultrasonic velocity change imaging.
  • 27. Convex or curvilinear Probe A convex or curvilinear probes may have a wider footprint and lower frequency for transabdominal imaging, or in a tighter array (wider field of view) and higher frequency for endocavitary imaging. Figure: Convex or Curvilinear Probe
  • 28. Linear or Vascular Probe Features:  piezoelectric crystal arrangement: curvilinear, along the aperture  footprint size: big (small for the micro-convex transducers)  operating frequency (bandwidth): 1-5 MHz (usually 3.5-5 MHz)  ultrasound beam shape: sector; the ultrasound beam shape and size vary with distance from the transducer, that causes the lack of lateral resolution at greater depths  Element radius: 64 ch – 192 ch Uses:  Typical abdominal application.  Transvaginal and transrectal application.  Pelvic and lung (micro-convex transducer) ultrasound  Diagnoses organs.
  • 29. Ultrasound Probe Care An ultrasound transducer is the most important and usually the most expensive element of the ultrasound machine, so it should be used carefully, which means the following: • Do not throw, drop or knock the transducer, • Do not allow to spoil the transducer`s duct, • Wipe the gel from the transducer after each use, • Do not sluice with alcohol-based confections.
  • 30. Image Orientation Ultrasound orientation is essential to understand what is being seen in point-of-care ultrasound. There are two key aspects: how the indicator is oriented relative to the screen and how the probe and indicator are placed and oriented relative to the patient. The conventional ultrasound image is a two- dimensional plane composed of frames made up of scan lines that are updated many times a second to create a moving image. Ultrasound orientation can be challenging because it involves understanding how a two-dimensional plane may cut through a three-dimensional object in any of the three standard planes (sagittal, transverse, or coronal), as well as any oblique orientation between those planes. However, one of the great advantages of ultrasound is the ability to obtain images from a variety of different approaches, and a thorough understanding of orientation will allow you to take advantage of this.
  • 31. Indicator‐to-Screen Orientation There are two rules for indicator-to-screen orientation in standard emergency ultrasound imaging, which uses the same convention as general radiology imaging: 1. The top of the screen is closer to the probe. Bottom of the screen shows structures farther away from the probe. 2. The left side of the screen, as it is viewed, corresponds to the side of the probe marked with an indicator. Note: Imaging performed by cardiology specialists uses a different and opposite convention for rule 2. The "indicator" may differ widely between manufacturers, and is typically a bump or a groove. It is important to verify your orientation prior to beginning any exam. This should be done by placing a small amount of gel on the side of the probe where you believe the indicator to be, and confirming that the side with the gel corresponds to the left of the screen as it is viewed (Figs. a and b).
  • 32. Indicator‐to-Screen Orientation There are two rules for indicator-to-screen orientation in standard emergency ultrasound imaging, which uses the same convention as general radiology imaging: 1. The top of the screen is closer to the probe. Bottom of the screen shows structures farther away from the probe. 2. The left side of the screen, as it is viewed, corresponds to the side of the probe marked with an indicator. Note: Imaging performed by cardiology specialists uses a different and opposite convention for rule 2. The "indicator" may differ widely between manufacturers, and is typically a bump or a groove. It is important to verify your orientation prior to beginning any exam. This should be done by placing a small amount of gel on the side of the probe where you believe the indicator to be, and confirming that the side with the gel corresponds to the left of the screen as it is viewed (Figs. a and b).
  • 33. Indicator‐to-Screen Orientation (a) Shows a thumb placed on the indicator, with gel placed on the face of the probe on that side. (b) (b) Shows what should be seen on the screen in general imaging orientation.
  • 34. Indicator‐to-Patient Orientation Once the indicator-to-screen orientation is understood and verified, the probe is placed on the patient and images are viewed on the screen. The top of the screen will show structures closer to where the probe is placed, with the bottom of the screen showing structures farther away from the probe face. The left of the screen will show structures toward where the indicator is directed.
  • 35. Image Planes Longitudinal (Sagittal) - transducer is placed along the patient`s long axis with the index directed cranial (the cranial structures are displayed on the left side of the screen, i.e. the marked side of the screen). Transverse (Axial) - transducer is placed perpendicular to the patient`s long axis with the index directed (usually) to the patient`s right (the right structures are displayed on the left side of the screen, i.e. the marked side of the screen). Frontal (Coronal) - transducer is placed lateral with the index directed cranial (the cranial structures are displayed on the left side of the screen, i.e. the marked side of the screen; the structures proximal to the probe are located at the top of the screen).
  • 36. Manipulation of Transducer An image acquisition and optimization of the views is achieved the proper manipulation of the transducer:  tilting allows optimization of the image and imaging different structures in the same tomographic plane,  angulation of the transducer slides the ultrasound image from side to side,  rotation the transducer at a single position changes the image plane. An unexperienced sonographer should manipulate the probe carefully (limited movements) in one image plane. The chaotic movements do not allow obtaining a reliable image views and proper data acquisition.
  • 37. Modes of Ultrasound 1. A-mode (A=amplitude) The amplitude of reflected ultrasound is displayed on an oscilloscope screen. It is just of historical importance. The A-mode is now used only in ophthalmology.  Each pulse produces a new a one-dimensional display or image  Line of information on the display  An uncommon display, except in ophthalmologic sonography used for precise intraocular length measurements Figure: A-mode Ultrasound
  • 38. Modes of Ultrasound 2. B-mode (B=brightness) (2D in echocardiography) This is now the essential imaging modality in the diagnostic ultrasound. An amplitude of the reflected ultrasound signals is converted into a gray scale image. Owing to the wide gray scale (most of the ultrasound machines use 256 shades of gray) even very small differences in echogenicity are possible to visualize.  Basis for gray scale, two-dimensional (2D) imaging  US unit tracks the position of the transducer to place a dot on the screen corresponding to the transducer position (X, Y locations), creating a 2D image Figure: B-mode Ultrasound
  • 39. Modes of Ultrasound 3. M-mode (M=motion) It reflects a motion of the heart structures over time. Nowadays, the integration of 2D and M- mode images is possible. Due to its excellent temporal resolution (high sampling rate), M- mode is extremely valuable for accurate evaluation of rapid movements. • One-dimension image used to investigate moving structures with respect to time • Evaluates motion pattern of moving structures such as in the heart Figure: M-mode Ultrasound
  • 40. Modes of Ultrasound 4. D-mode (D=Doppler) This imaging mode is based on the Doppler effect, i.e. change in frequency (Doppler shift) caused by the reciprocal movement of the sound generator and the observer. Diagnostic ultrasound uses the change in frequency of ultrasound signal backscattered from red blood cells. The frequency of the reflected ultrasound wave increases or decreases according to the direction of blood flow in relation to the transducer. Figure: M-mode Ultrasound
  • 41. Types of Doppler Doppler Duplex Doppler Duplex technique is based on the simultaneous B-mode real-time and Doppler imaging. A gray scale display serves for the localization of flow measurement site. Both spectral Doppler techniques (Continuous-Wave and Pulsed) can be used in Doppler Duplex imaging Figure: Doppler Duplex Ultrasound
  • 42. Types of Doppler CW Doppler (Continuous-Wave Doppler) (CWD) This imaging mode requires two piezoelectric crystals. One continuously transmits and the other one receives the Doppler signals along the scan-line. CWD is very useful in high velocity signals recording. It defines blood flow direction but has no value in Doppler signal source identification. Figure: CW Doppler Ultrasound
  • 43. Types of Doppler PW Doppler (Pulsed-Wave Doppler) (PWD) The same piezoelectric crystal transmits and receives the ultrasound signal. The reflected signal returns to the transducer after a definite delay that is known as pulse repetition frequency (PRF). PWD allows measurement from a small, specific blood volume (depth of interest), which is defined by a sample volume. It provides both blood flow direction and precise determination of of Doppler signal source. The main limitation is failure to display the high velocity signals. The maximum detectable frequency shift (the Nyquist limit) is determined by the value of half the PRF. After the velocity exceeds the limit, aliasing occurs Figure: PW Doppler Ultrasound
  • 44. Types of Doppler Color Doppler Color Doppler imaging is based on the PWD assumption. However, in Color Doppler imaging, multiple sample volume are evaluated along each scanning line. Velocities are displayed using a color scale. Velocities toward the transducer are red, and velocities away from the transducer are blue. Figure: Color Doppler Ultrasound
  • 45. Types of Doppler Power Doppler Power Doppler is a variant of Color Doppler technique which displays the magnitude of the Doppler signal rather than the Doppler frequency. It is often used to increase sensitivity to low flows and velocities. Power Doppler does not display flow direction or different velocities. Figure: Power Doppler Ultrasound
  • 46. Artifacts An ultrasound artifact is a structure in an image which does not directly similar with actual tissue being scanned. Artifact assumes different forms including :  Structures in the image that are not actually present  Objects that should be represented but are missing from the image.  Structures which are misregistered on the image.
  • 47. Common Artifacts 1. Reverberation 2. Acoustic Shadowing 3. Acoustic Enhancement 4. Edge Shadowing 5. Beam Width Artifact 6. Slice Thickness Artifact 7. Side Lobe Artifact 8. Mirror Image 9. Double Image 10. Equipment-generated Artifact 11. Refraction Artifact
  • 48. REVERBERATION  This is the production of false echoes due to repeated reflections between two interfaces with a high acoustic impedance mismatch.  The echo from the interface is received by the transducer and displayed on the image.  Some of the energy in the returned echo is reflected at the transducer face, and return to the reflecting interface as if it was a weak transmitted pulse, returning as a second echo.  As the time taken for the second echo to arrive is twice that taken by the depth.  This sequence of reflection and transmission can occur many times, with the third echo taking three times as long to return to the transducer and being displayed at three times the depth, and so on.  The reverberation echoes will be equally spaced because the time for each additional echo is multiple of the time of return of the first echo.  These reverberation echoes will be strong because of the high acoustic mismatch.  This artifact will be seen at the skin-transducer interface and behind bowel gas.
  • 49. REVERBERATION Rectification:  Increase the amount of gel used.  Use a stand-off pad.  Reduce the gain.  Move the position of the transducer.
  • 50. ACOUSTIC SHADOWING  This appears as an area of low amplitude echoes behind an area of strongly attenuating tissue.  It is caused by severe attenuation of the beam at an interface, resulting in very little sound being transmitted beyond.  The attenuation can be due to either absorption or reflection of the sound waves, or a combination of the two.  Acoustic shadowing will occur at interfaces with large acoustic mismatch such as: Soft tissue and gas Soft tissue and bone or calculus
  • 52. ACOUSTIC ENHANCEMENT  This artifact appears as a localized area of increased echo amplitude behind an area of low attenuation.  On a scan it will appears as an area of increased brightness, and can commonly be seen distal to fluid-filled structures such as the urinary bladder, GB or a cyst.  The artifact arises due to the application of the time-gain compensation(TGC) to areas of low attenuating structures such as fluid.  It is caused by the low level of attenuation of the beam as it passes through fluid relative to the greater attenuation of the beam in the adjacent more solid tissue.  This artifact can often be an useful diagnostic aid, particularly when scanning a soft-tissue mass or cyst containing low level echoes.  These echoes may often cause the structure to disappear in the image as it blend into the surrounding echo pattern.
  • 54. EDGE SHADOWING  A combination of refraction and reflection occurring at the edges of rounded structures will result in edge shadowing artifact.  It arises due to refraction of the beam caused by both the curvature of the rounded edges and difference in speed of two materials.  When the ultrasound beam reaches the rounded edge of a structure, reflection will occur, with an angle of incidence equal to the angle of reflection.  The outer part of the beam will be totally reflected, but the reminder of the beam passes through the rounded structure and is refracted.  This combination of reflection and refraction of the beam at the edges of a rounded structure results in a thin strip of tissue behind the edge not being insonated and causes a shadow.
  • 56. BEAM WIDTH ARTIFACT  This artifact can be demonstrated by scanning a point reflector in a phantom, where the display will clearly portray this as a line.  During routine scanning, the artifact can be seen when spurious echoes are displayed in an echo-free area.  Correct positioning of the focal zone will help to reduce this artifact.  The focal zone is controlled by electronically narrowing the beam
  • 57. SLICE THICKNESS ARTIFACT  These occurs due to the thickness of the beam, and are similar to beam width artifacts.  These artifacts will typically be seen in transverse views of the urinary bladder when structures adjacent to the slice through the bladder being scanned will be incorporated into the image.  These echoes are then displayed as if they were arising from within the bladder.  Although the appearance of this artifact is similar to the beam width artifact, the differentiating factors is that the reflector causing the slice thickness artifact will not be seen on the display.  This artifact is a result of inherent characteristics of the transducer, and apart from trying a different transducer, cannot be eliminated.
  • 58. SIDE LOBE ARTIFACT  The energy within the ultrasound beam exists as several side lobes radiating at a number of angles from a central lobe.  Echoes are generated by these side lobes in addition to the main lobe, but all the returning echoes are assumed by the transducer to have arisen from the central axis of the main lobe.  Side lobe echoes will therefore be misregistered in the display.  This artifact can often be seen in area such as the urinary bladder and may also arise within a cyst.
  • 59. MIRROR IMAGE ARTIFACT  These artifacts results in a mirror image of a structure occurring in an ultrasound display.  They arise due to specular reflection of the beam at a large smooth interface.  An area close to a specular reflector will be imaged twice, once by the original ultrasound beam and once by the beam after it has reflected off the specular reflector.  Mirror image artifacts are most commonly seen where there is a large acoustic mismatch, such as a fluid-air interface.  Typically this artifact can occur during the scanning of a full bladder, when air in the rectum behind the bladder act as specular reflector and mirror image of the bladder is displayed posteriorly.  It will then have the appearance of a large cyst behind the bladder.  It can also be seen when scanning the liver, and the diaphragm act as a specular reflector.
  • 61. DOUBLE IMAGE ARTIFACT  This image is caused by refraction of the beam and may occur in areas such as the rectus abdominis muscle on the anterior abdominal wall.  In the transverse plane the edges of the muscle act as a lens and the ultrasound beam to be refracted and this causes the single structure to be interrogated by two separate refracted beams.  Two sets of echoes will therefore be returned and these will cause display of two structures in the image. This results in, for example, two images of the transverse aorta side by side in the abdomen.
  • 62. EQUIPMENT GENERATED ARTIFACT  Incorrect use of the equipment controls can lead to artifact appearing.  Misuse of controls such as the gain or TGC can result in echoes being recorded as too bright or too dark.  Care must be taken when setting these controls, to ensure an even brightness throughout the image.  If the dynamic range control is incorrectly set, this can lead to an image which has too much contrast, and result in the loss of subtle echo information.  Gain must be in medium level.  Blurring of a moving image can occur if the frame rate is too low or if the persistence is too high.  It is important to ensure that the frame rate is capable of recording a moving structure at the speed.  Use of multiple focal zones can give rise to a prominent banding effect within the image.
  • 64. REFRACTION ARTIFCT  The refraction is the change of the sound direction on passing from one medium to another.  In ultrasound, refraction is due to sound velocity mismatches combines with oblique angles of incidence, most commonly with convex scanheads.  When the ultrasound wave crosses at an oblique angle the interface of two materials, through which the waves propagate at different velocities, refraction occurs, caused by bending of the wave beam.  Refraction artifact cause spatial distortion and loss of resolution in the image.