Comparison of reticulocyte count by VCS technology; Beckman Coulter
Hematology LH785 & fluorescent flow cytometry; XT4000i automate
hematology analyzers
Kridsada Sirisabhabhorn1 Anurak Choeymang2 Supaporn Pumpa1 Naiyana poottasane1 and Palakorn Puttaruk1
1Department of Medical Technology Laboratory, Thammasat University hospital, Pathumthani, Thailand
2Pharmacology and Toxicology Unit, Faculty of Allied Health Sciences, Thammasat University, Pathumthani, Thailand
Reticulocyte count is significant for indication a new erythrocyte synthesis. Especially, usually applied for diagnosis and monitor in new born blood and anemia
diseases. The traditional method for investigation of reticulocyte count with 1% brilliant cresyl blue stain have been long time practice, it not suitable for work load unit
and rapid diagnosis requirement today. The new principle of reticulocyte detection was applied with 2 automate hematology analyzers which Coulter®LH785; VCS
(volume, conductivity and light scatter) technology and Sysmex®XT-4000i; fluorescent flow cytometry for improve reticulocyte count diagnosis. This study propose for
evaluate VCS technology by two automate hematology analyzers, compared with traditional method. All 105 cases were investigated by comparison among
Coulter®LH785, Sysmex® XT-4000i and traditional method approved percentage reticulocyte count by 3 professional medical technologists. Pearson correlation was used
for consider correlation. The results demonstrated that Coulter®LH785 and Sysmex® XT-4000i correlation coefficients (r) as 0.889, p-values < 0.05 and 0.936, p-values
< 0.05 respectively. Moreover, the comparisons between Coulter®LH785 and Sysmex® XT-4000i illustrated correlation coefficients (r) as 0.902. The traditional method of
human reticulocyte countable was not different significantly statistic analysis by ANOVAs test. Conclusion, 3 methods of Coulter®LH785, Sysmex® XT-4000i automate
hematology analyzers and traditional reticulocyte counts are good correlation within p-values < 0.05. Moreover, it is helpful for work flow management than traditional
method and reduce turn around time (TAT) of reticulocyte investigation.
Abstract
Key words: reticulocyte, fluorescent flow cytometry, VCS technology
Manual LH785 XT4000i
Manual Pearson Correlation 1 .889(**) .936(**)
Sig. (2-tailed) .000 .000
N 105 105 105
LH785 Pearson Correlation .889(**) 1 .902(**)
Sig. (2-tailed) .000 .000
N 105 105 105
XT4000i Pearson Correlation .936(**) .902(**) 1
Sig. (2-tailed) .000 .000
N 105 105 105
Methods & Correlations
** Correlation is significant at the 0.01 level (2-tailed).
0
1
2
3
4
5
6
7
8
9
0 2 4 6 8 10
XT400i
Manual
0
1
2
3
4
5
6
7
8
0 2 4 6 8 10
LH785
Manual
0
1
2
3
4
5
6
7
8
0 2 4 6 8 10
XT400i
LH785
Introduction
Materials and Methods
Complication of conventional method used for investigation % reticulocyte increase job workload and leads to long TAT (Turn Around Time) and costumer
dissatisfaction. Moreover manual method needs microscope specialist for % reticulocyte counting. Today the principle of fluorescent flow cytometry, Sysmex®XT-4000i and
VCS technology, Coulter®LH785 can reduce work flow of conventional method for % reticulocyte count analysis. However measuring accuracy of each method is
significant for reliable report. The objective in this study was to evaluated comparative both two principle of automate analyzer above with conventional or manual method
, 1% brilliant cresyl blue stain using Pearson Correlation. The study was shown benefit that confidence reliable % reticulocyte count values and provided short procedure
of analysis which suitable each analytical department.
Conventional method with 1% brilliant cresyl blue stain
Fluorescent flow cytometry, Sysmex®XT-4000i
VCS technology, Coulter®LH785
Results
Conclusion and discussion
In this study conclude that two automate analyzers which composed principle of fluorescent flow cytometry, Sysmex®XT-4000i and VCS technology, Coulter®LH785
were presented good correlation for investigated % reticulocyte counting by correlation coefficients (r) values as 0.936 and 0.889 within p-values < 0.05 when compared
with conventional method. Additionally two automate analyzers were demonstrated correlation coefficients (r) as 0.902 within p-values < 0.05 thus also good correlation
each other. The performance ensuring in conventional method by three special medical technologists were shown no different significantly statistic analysis by ANOVAs
test. The automated analyzers can provide helpful in % reticulocyte investigation and improve diagnosis approach to standard, reduced individual human errors and
eliminated process of crysel blue staining and slide preparation. Lastly, quality assurance necessary should monitoring update always by medical technologist. This study
declared no conflict of interest.
References
R=0.902R=0.889R=0.936
Conventional method
Sysmex®XT-4000i
Coulter®LH785
Graph 1. Illustrated correlation curve of % reticulocyte count between
Sysmex®XT-4000i and conventional method (manual method)
Graph 2. Illustrated correlation curve of % reticulocyte count between
Coulter®LH785 and conventional method (manual method)
Graph 3. Illustrated correlation curve of % reticulocyte count between
Coulter®LH785 and Sysmex®XT-4000i
Pearson Correlations
Table 1. Illustrated
correlation results 3 method
for investigated percent of
reticulocyte count which
compared using Pearson
correlation, p-values < 0.05
All 105 cases were investigated % reticulocyte count with 3 principles mention above. All methods were investigated their relationships using Pearson correlation,
which explored correlation coefficients (r) p-values < 0.05. Using 3 medical technologists who were ensued not different significantly statistic analysis by ANOVAs test in
conventional method evaluated mean values of % reticulocyte count
Sysmex. Reticulocytes and their significance. Sysmex Xtra Online, 2010: 1-12.
PCL Holding Co., Ltd. Standard Operating Procedure ( SOP ) For Coulter ReticPakTM

More Related Content

Similar to Comparison of reticulocyte count by VCS technology; Beckman Coulter Hematology LH785 & fluorescent flow cytometry; XT4000i automate hematology analyzers/Kridsada31

EXPERIMENTAL IMPLEMENTATION OF EMBARRASINGLY PARALLEL PROCESS IN ANALYSIS OF ...
EXPERIMENTAL IMPLEMENTATION OF EMBARRASINGLY PARALLEL PROCESS IN ANALYSIS OF ...EXPERIMENTAL IMPLEMENTATION OF EMBARRASINGLY PARALLEL PROCESS IN ANALYSIS OF ...
EXPERIMENTAL IMPLEMENTATION OF EMBARRASINGLY PARALLEL PROCESS IN ANALYSIS OF ...ijesajournal
 
Automated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling PosterAutomated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling PosterRick Youngblood
 
LC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsLC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsSCIEX
 
Presentation Shahbaz
Presentation ShahbazPresentation Shahbaz
Presentation ShahbazZahoor Ahmed
 
20180805-Hematology-Calibration-1.pptx
20180805-Hematology-Calibration-1.pptx20180805-Hematology-Calibration-1.pptx
20180805-Hematology-Calibration-1.pptxPUPUTPUJIANTI
 
Development and Validation of RP HPLC Method for Estimation of Vortioxetine i...
Development and Validation of RP HPLC Method for Estimation of Vortioxetine i...Development and Validation of RP HPLC Method for Estimation of Vortioxetine i...
Development and Validation of RP HPLC Method for Estimation of Vortioxetine i...ijtsrd
 
6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final
6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final
6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_finalLawrence Hwang
 
Towards the Automated Calculation of Clinical Quality Indicators
Towards the Automated Calculation of Clinical Quality IndicatorsTowards the Automated Calculation of Clinical Quality Indicators
Towards the Automated Calculation of Clinical Quality IndicatorsKathrin Dentler
 
A novel approach in identification of blood group using laser technology
A novel approach in identification of blood group using laser technologyA novel approach in identification of blood group using laser technology
A novel approach in identification of blood group using laser technologyeSAT Publishing House
 
L25 b met_complete_blood_count
L25 b met_complete_blood_countL25 b met_complete_blood_count
L25 b met_complete_blood_countShakeel Mujahid
 
Internal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labInternal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labAnkit Raiyani
 
Quality assurance part_2
Quality assurance part_2Quality assurance part_2
Quality assurance part_2ThorikulHuda2
 
Anaesth int care 40 (1007 15) 2012
Anaesth int care 40 (1007 15) 2012Anaesth int care 40 (1007 15) 2012
Anaesth int care 40 (1007 15) 2012AnestesioRecife
 
Protromibina equipo.pdf
Protromibina equipo.pdfProtromibina equipo.pdf
Protromibina equipo.pdfleroleroero1
 
Real-time and Non-Invasive Detection of Haemoglobin level using CNN
Real-time and Non-Invasive Detection of Haemoglobin level using CNNReal-time and Non-Invasive Detection of Haemoglobin level using CNN
Real-time and Non-Invasive Detection of Haemoglobin level using CNNIRJET Journal
 
Pp 90505-lc-ms-serum-profiling-biomarker-msacleu2019-pp90505-en
Pp 90505-lc-ms-serum-profiling-biomarker-msacleu2019-pp90505-enPp 90505-lc-ms-serum-profiling-biomarker-msacleu2019-pp90505-en
Pp 90505-lc-ms-serum-profiling-biomarker-msacleu2019-pp90505-enAlexander Boichenko
 

Similar to Comparison of reticulocyte count by VCS technology; Beckman Coulter Hematology LH785 & fluorescent flow cytometry; XT4000i automate hematology analyzers/Kridsada31 (20)

EXPERIMENTAL IMPLEMENTATION OF EMBARRASINGLY PARALLEL PROCESS IN ANALYSIS OF ...
EXPERIMENTAL IMPLEMENTATION OF EMBARRASINGLY PARALLEL PROCESS IN ANALYSIS OF ...EXPERIMENTAL IMPLEMENTATION OF EMBARRASINGLY PARALLEL PROCESS IN ANALYSIS OF ...
EXPERIMENTAL IMPLEMENTATION OF EMBARRASINGLY PARALLEL PROCESS IN ANALYSIS OF ...
 
Automated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling PosterAutomated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling Poster
 
LC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsLC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminants
 
Presentation Shahbaz
Presentation ShahbazPresentation Shahbaz
Presentation Shahbaz
 
20180805-Hematology-Calibration-1.pptx
20180805-Hematology-Calibration-1.pptx20180805-Hematology-Calibration-1.pptx
20180805-Hematology-Calibration-1.pptx
 
SPECTROPHOTOMETRIC MEASUREMENT OF SOLUBILITY TEST TURBIDITY AS AN IMPROVED DI...
SPECTROPHOTOMETRIC MEASUREMENT OF SOLUBILITY TEST TURBIDITY AS AN IMPROVED DI...SPECTROPHOTOMETRIC MEASUREMENT OF SOLUBILITY TEST TURBIDITY AS AN IMPROVED DI...
SPECTROPHOTOMETRIC MEASUREMENT OF SOLUBILITY TEST TURBIDITY AS AN IMPROVED DI...
 
Development and Validation of RP HPLC Method for Estimation of Vortioxetine i...
Development and Validation of RP HPLC Method for Estimation of Vortioxetine i...Development and Validation of RP HPLC Method for Estimation of Vortioxetine i...
Development and Validation of RP HPLC Method for Estimation of Vortioxetine i...
 
6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final
6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final
6-8-2015 AACC Poster HIV p24 S-PLEX - Stengelin_final
 
Towards the Automated Calculation of Clinical Quality Indicators
Towards the Automated Calculation of Clinical Quality IndicatorsTowards the Automated Calculation of Clinical Quality Indicators
Towards the Automated Calculation of Clinical Quality Indicators
 
JOURNAL CLUB PRESENTATION (20L81S0707-PA)
JOURNAL CLUB PRESENTATION (20L81S0707-PA)JOURNAL CLUB PRESENTATION (20L81S0707-PA)
JOURNAL CLUB PRESENTATION (20L81S0707-PA)
 
A novel approach in identification of blood group using laser technology
A novel approach in identification of blood group using laser technologyA novel approach in identification of blood group using laser technology
A novel approach in identification of blood group using laser technology
 
L25 b met_complete_blood_count
L25 b met_complete_blood_countL25 b met_complete_blood_count
L25 b met_complete_blood_count
 
Internal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labInternal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation lab
 
Quality assurance part_2
Quality assurance part_2Quality assurance part_2
Quality assurance part_2
 
Anaesth int care 40 (1007 15) 2012
Anaesth int care 40 (1007 15) 2012Anaesth int care 40 (1007 15) 2012
Anaesth int care 40 (1007 15) 2012
 
Protromibina equipo.pdf
Protromibina equipo.pdfProtromibina equipo.pdf
Protromibina equipo.pdf
 
Dharma S - AIMRADIAL 2014 - Nitroglycerin and radial artery occlusion
 Dharma S - AIMRADIAL 2014 - Nitroglycerin and radial artery occlusion Dharma S - AIMRADIAL 2014 - Nitroglycerin and radial artery occlusion
Dharma S - AIMRADIAL 2014 - Nitroglycerin and radial artery occlusion
 
Chap3 1
Chap3 1Chap3 1
Chap3 1
 
Real-time and Non-Invasive Detection of Haemoglobin level using CNN
Real-time and Non-Invasive Detection of Haemoglobin level using CNNReal-time and Non-Invasive Detection of Haemoglobin level using CNN
Real-time and Non-Invasive Detection of Haemoglobin level using CNN
 
Pp 90505-lc-ms-serum-profiling-biomarker-msacleu2019-pp90505-en
Pp 90505-lc-ms-serum-profiling-biomarker-msacleu2019-pp90505-enPp 90505-lc-ms-serum-profiling-biomarker-msacleu2019-pp90505-en
Pp 90505-lc-ms-serum-profiling-biomarker-msacleu2019-pp90505-en
 

More from kridsada31

Waraporn Bubpha, Autcharaporn Srisuwan, Palakorn Puttaruk and Kridsada Sirisa...
Waraporn Bubpha, Autcharaporn Srisuwan, Palakorn Puttaruk and Kridsada Sirisa...Waraporn Bubpha, Autcharaporn Srisuwan, Palakorn Puttaruk and Kridsada Sirisa...
Waraporn Bubpha, Autcharaporn Srisuwan, Palakorn Puttaruk and Kridsada Sirisa...kridsada31
 
Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk, Medical Techno...
Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk, Medical Techno...Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk, Medical Techno...
Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk, Medical Techno...kridsada31
 
Kridsada Sirisabhabhorn, Supaporn Pumpa, Surapong Pornprasitseang and Palakor...
Kridsada Sirisabhabhorn, Supaporn Pumpa, Surapong Pornprasitseang and Palakor...Kridsada Sirisabhabhorn, Supaporn Pumpa, Surapong Pornprasitseang and Palakor...
Kridsada Sirisabhabhorn, Supaporn Pumpa, Surapong Pornprasitseang and Palakor...kridsada31
 
Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk Medical Technol...
Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk Medical Technol...Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk Medical Technol...
Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk Medical Technol...kridsada31
 
Increase in the prevalence of Clostridium difficileproducing toxin B in fecal...
Increase in the prevalence of Clostridium difficileproducing toxin B in fecal...Increase in the prevalence of Clostridium difficileproducing toxin B in fecal...
Increase in the prevalence of Clostridium difficileproducing toxin B in fecal...kridsada31
 
Thecomparison methods of amphetamines detection in urine samples in Thammasat...
Thecomparison methods of amphetamines detection in urine samples in Thammasat...Thecomparison methods of amphetamines detection in urine samples in Thammasat...
Thecomparison methods of amphetamines detection in urine samples in Thammasat...kridsada31
 
Comparison and Evaluation of FUS-2000 Urinalysis Hybrid and Conventional Micr...
Comparison and Evaluation of FUS-2000 Urinalysis Hybrid and Conventional Micr...Comparison and Evaluation of FUS-2000 Urinalysis Hybrid and Conventional Micr...
Comparison and Evaluation of FUS-2000 Urinalysis Hybrid and Conventional Micr...kridsada31
 
Commonly renal diseases associated with urine elements/Kridsada31
Commonly renal diseases associated with urine elements/Kridsada31Commonly renal diseases associated with urine elements/Kridsada31
Commonly renal diseases associated with urine elements/Kridsada31kridsada31
 
Urine sediment 55/Kridsada31
Urine sediment 55/Kridsada31Urine sediment 55/Kridsada31
Urine sediment 55/Kridsada31kridsada31
 
เกณฑ์การรายงานผลการตรวจวิเคราะห์ตะกอนปัสสาวะ หน่วยจุลทรรศน์และปรสิตวิทยา/Krid...
เกณฑ์การรายงานผลการตรวจวิเคราะห์ตะกอนปัสสาวะ หน่วยจุลทรรศน์และปรสิตวิทยา/Krid...เกณฑ์การรายงานผลการตรวจวิเคราะห์ตะกอนปัสสาวะ หน่วยจุลทรรศน์และปรสิตวิทยา/Krid...
เกณฑ์การรายงานผลการตรวจวิเคราะห์ตะกอนปัสสาวะ หน่วยจุลทรรศน์และปรสิตวิทยา/Krid...kridsada31
 
Malaria situation in Thammasat Charlerm Prakiet hospital Pathumthani/Kridsada31
Malaria situation in Thammasat Charlerm Prakiet hospital Pathumthani/Kridsada31Malaria situation in Thammasat Charlerm Prakiet hospital Pathumthani/Kridsada31
Malaria situation in Thammasat Charlerm Prakiet hospital Pathumthani/Kridsada31kridsada31
 
Trend of parasitic infection rates of Thammasat University Hospital’s patient...
Trend of parasitic infection rates of Thammasat University Hospital’s patient...Trend of parasitic infection rates of Thammasat University Hospital’s patient...
Trend of parasitic infection rates of Thammasat University Hospital’s patient...kridsada31
 
Improvement of bacterial estimation in urinalysis by iQ®200 ELITE®: capture i...
Improvement of bacterial estimation in urinalysis by iQ®200 ELITE®: capture i...Improvement of bacterial estimation in urinalysis by iQ®200 ELITE®: capture i...
Improvement of bacterial estimation in urinalysis by iQ®200 ELITE®: capture i...kridsada31
 
The External Quality Assessment [EQA] analysis achievement for excellent perf...
The External Quality Assessment [EQA] analysis achievement for excellent perf...The External Quality Assessment [EQA] analysis achievement for excellent perf...
The External Quality Assessment [EQA] analysis achievement for excellent perf...kridsada31
 
The introduction of well trained experience with iQ®200i microscopic analyzer...
The introduction of well trained experience with iQ®200i microscopic analyzer...The introduction of well trained experience with iQ®200i microscopic analyzer...
The introduction of well trained experience with iQ®200i microscopic analyzer...kridsada31
 

More from kridsada31 (15)

Waraporn Bubpha, Autcharaporn Srisuwan, Palakorn Puttaruk and Kridsada Sirisa...
Waraporn Bubpha, Autcharaporn Srisuwan, Palakorn Puttaruk and Kridsada Sirisa...Waraporn Bubpha, Autcharaporn Srisuwan, Palakorn Puttaruk and Kridsada Sirisa...
Waraporn Bubpha, Autcharaporn Srisuwan, Palakorn Puttaruk and Kridsada Sirisa...
 
Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk, Medical Techno...
Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk, Medical Techno...Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk, Medical Techno...
Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk, Medical Techno...
 
Kridsada Sirisabhabhorn, Supaporn Pumpa, Surapong Pornprasitseang and Palakor...
Kridsada Sirisabhabhorn, Supaporn Pumpa, Surapong Pornprasitseang and Palakor...Kridsada Sirisabhabhorn, Supaporn Pumpa, Surapong Pornprasitseang and Palakor...
Kridsada Sirisabhabhorn, Supaporn Pumpa, Surapong Pornprasitseang and Palakor...
 
Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk Medical Technol...
Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk Medical Technol...Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk Medical Technol...
Kridsada Sirisabhabhorn, Supaporn Pumpa and Palakorn Puttaruk Medical Technol...
 
Increase in the prevalence of Clostridium difficileproducing toxin B in fecal...
Increase in the prevalence of Clostridium difficileproducing toxin B in fecal...Increase in the prevalence of Clostridium difficileproducing toxin B in fecal...
Increase in the prevalence of Clostridium difficileproducing toxin B in fecal...
 
Thecomparison methods of amphetamines detection in urine samples in Thammasat...
Thecomparison methods of amphetamines detection in urine samples in Thammasat...Thecomparison methods of amphetamines detection in urine samples in Thammasat...
Thecomparison methods of amphetamines detection in urine samples in Thammasat...
 
Comparison and Evaluation of FUS-2000 Urinalysis Hybrid and Conventional Micr...
Comparison and Evaluation of FUS-2000 Urinalysis Hybrid and Conventional Micr...Comparison and Evaluation of FUS-2000 Urinalysis Hybrid and Conventional Micr...
Comparison and Evaluation of FUS-2000 Urinalysis Hybrid and Conventional Micr...
 
Commonly renal diseases associated with urine elements/Kridsada31
Commonly renal diseases associated with urine elements/Kridsada31Commonly renal diseases associated with urine elements/Kridsada31
Commonly renal diseases associated with urine elements/Kridsada31
 
Urine sediment 55/Kridsada31
Urine sediment 55/Kridsada31Urine sediment 55/Kridsada31
Urine sediment 55/Kridsada31
 
เกณฑ์การรายงานผลการตรวจวิเคราะห์ตะกอนปัสสาวะ หน่วยจุลทรรศน์และปรสิตวิทยา/Krid...
เกณฑ์การรายงานผลการตรวจวิเคราะห์ตะกอนปัสสาวะ หน่วยจุลทรรศน์และปรสิตวิทยา/Krid...เกณฑ์การรายงานผลการตรวจวิเคราะห์ตะกอนปัสสาวะ หน่วยจุลทรรศน์และปรสิตวิทยา/Krid...
เกณฑ์การรายงานผลการตรวจวิเคราะห์ตะกอนปัสสาวะ หน่วยจุลทรรศน์และปรสิตวิทยา/Krid...
 
Malaria situation in Thammasat Charlerm Prakiet hospital Pathumthani/Kridsada31
Malaria situation in Thammasat Charlerm Prakiet hospital Pathumthani/Kridsada31Malaria situation in Thammasat Charlerm Prakiet hospital Pathumthani/Kridsada31
Malaria situation in Thammasat Charlerm Prakiet hospital Pathumthani/Kridsada31
 
Trend of parasitic infection rates of Thammasat University Hospital’s patient...
Trend of parasitic infection rates of Thammasat University Hospital’s patient...Trend of parasitic infection rates of Thammasat University Hospital’s patient...
Trend of parasitic infection rates of Thammasat University Hospital’s patient...
 
Improvement of bacterial estimation in urinalysis by iQ®200 ELITE®: capture i...
Improvement of bacterial estimation in urinalysis by iQ®200 ELITE®: capture i...Improvement of bacterial estimation in urinalysis by iQ®200 ELITE®: capture i...
Improvement of bacterial estimation in urinalysis by iQ®200 ELITE®: capture i...
 
The External Quality Assessment [EQA] analysis achievement for excellent perf...
The External Quality Assessment [EQA] analysis achievement for excellent perf...The External Quality Assessment [EQA] analysis achievement for excellent perf...
The External Quality Assessment [EQA] analysis achievement for excellent perf...
 
The introduction of well trained experience with iQ®200i microscopic analyzer...
The introduction of well trained experience with iQ®200i microscopic analyzer...The introduction of well trained experience with iQ®200i microscopic analyzer...
The introduction of well trained experience with iQ®200i microscopic analyzer...
 

Comparison of reticulocyte count by VCS technology; Beckman Coulter Hematology LH785 & fluorescent flow cytometry; XT4000i automate hematology analyzers/Kridsada31

  • 1. Comparison of reticulocyte count by VCS technology; Beckman Coulter Hematology LH785 & fluorescent flow cytometry; XT4000i automate hematology analyzers Kridsada Sirisabhabhorn1 Anurak Choeymang2 Supaporn Pumpa1 Naiyana poottasane1 and Palakorn Puttaruk1 1Department of Medical Technology Laboratory, Thammasat University hospital, Pathumthani, Thailand 2Pharmacology and Toxicology Unit, Faculty of Allied Health Sciences, Thammasat University, Pathumthani, Thailand Reticulocyte count is significant for indication a new erythrocyte synthesis. Especially, usually applied for diagnosis and monitor in new born blood and anemia diseases. The traditional method for investigation of reticulocyte count with 1% brilliant cresyl blue stain have been long time practice, it not suitable for work load unit and rapid diagnosis requirement today. The new principle of reticulocyte detection was applied with 2 automate hematology analyzers which Coulter®LH785; VCS (volume, conductivity and light scatter) technology and Sysmex®XT-4000i; fluorescent flow cytometry for improve reticulocyte count diagnosis. This study propose for evaluate VCS technology by two automate hematology analyzers, compared with traditional method. All 105 cases were investigated by comparison among Coulter®LH785, Sysmex® XT-4000i and traditional method approved percentage reticulocyte count by 3 professional medical technologists. Pearson correlation was used for consider correlation. The results demonstrated that Coulter®LH785 and Sysmex® XT-4000i correlation coefficients (r) as 0.889, p-values < 0.05 and 0.936, p-values < 0.05 respectively. Moreover, the comparisons between Coulter®LH785 and Sysmex® XT-4000i illustrated correlation coefficients (r) as 0.902. The traditional method of human reticulocyte countable was not different significantly statistic analysis by ANOVAs test. Conclusion, 3 methods of Coulter®LH785, Sysmex® XT-4000i automate hematology analyzers and traditional reticulocyte counts are good correlation within p-values < 0.05. Moreover, it is helpful for work flow management than traditional method and reduce turn around time (TAT) of reticulocyte investigation. Abstract Key words: reticulocyte, fluorescent flow cytometry, VCS technology Manual LH785 XT4000i Manual Pearson Correlation 1 .889(**) .936(**) Sig. (2-tailed) .000 .000 N 105 105 105 LH785 Pearson Correlation .889(**) 1 .902(**) Sig. (2-tailed) .000 .000 N 105 105 105 XT4000i Pearson Correlation .936(**) .902(**) 1 Sig. (2-tailed) .000 .000 N 105 105 105 Methods & Correlations ** Correlation is significant at the 0.01 level (2-tailed). 0 1 2 3 4 5 6 7 8 9 0 2 4 6 8 10 XT400i Manual 0 1 2 3 4 5 6 7 8 0 2 4 6 8 10 LH785 Manual 0 1 2 3 4 5 6 7 8 0 2 4 6 8 10 XT400i LH785 Introduction Materials and Methods Complication of conventional method used for investigation % reticulocyte increase job workload and leads to long TAT (Turn Around Time) and costumer dissatisfaction. Moreover manual method needs microscope specialist for % reticulocyte counting. Today the principle of fluorescent flow cytometry, Sysmex®XT-4000i and VCS technology, Coulter®LH785 can reduce work flow of conventional method for % reticulocyte count analysis. However measuring accuracy of each method is significant for reliable report. The objective in this study was to evaluated comparative both two principle of automate analyzer above with conventional or manual method , 1% brilliant cresyl blue stain using Pearson Correlation. The study was shown benefit that confidence reliable % reticulocyte count values and provided short procedure of analysis which suitable each analytical department. Conventional method with 1% brilliant cresyl blue stain Fluorescent flow cytometry, Sysmex®XT-4000i VCS technology, Coulter®LH785 Results Conclusion and discussion In this study conclude that two automate analyzers which composed principle of fluorescent flow cytometry, Sysmex®XT-4000i and VCS technology, Coulter®LH785 were presented good correlation for investigated % reticulocyte counting by correlation coefficients (r) values as 0.936 and 0.889 within p-values < 0.05 when compared with conventional method. Additionally two automate analyzers were demonstrated correlation coefficients (r) as 0.902 within p-values < 0.05 thus also good correlation each other. The performance ensuring in conventional method by three special medical technologists were shown no different significantly statistic analysis by ANOVAs test. The automated analyzers can provide helpful in % reticulocyte investigation and improve diagnosis approach to standard, reduced individual human errors and eliminated process of crysel blue staining and slide preparation. Lastly, quality assurance necessary should monitoring update always by medical technologist. This study declared no conflict of interest. References R=0.902R=0.889R=0.936 Conventional method Sysmex®XT-4000i Coulter®LH785 Graph 1. Illustrated correlation curve of % reticulocyte count between Sysmex®XT-4000i and conventional method (manual method) Graph 2. Illustrated correlation curve of % reticulocyte count between Coulter®LH785 and conventional method (manual method) Graph 3. Illustrated correlation curve of % reticulocyte count between Coulter®LH785 and Sysmex®XT-4000i Pearson Correlations Table 1. Illustrated correlation results 3 method for investigated percent of reticulocyte count which compared using Pearson correlation, p-values < 0.05 All 105 cases were investigated % reticulocyte count with 3 principles mention above. All methods were investigated their relationships using Pearson correlation, which explored correlation coefficients (r) p-values < 0.05. Using 3 medical technologists who were ensued not different significantly statistic analysis by ANOVAs test in conventional method evaluated mean values of % reticulocyte count Sysmex. Reticulocytes and their significance. Sysmex Xtra Online, 2010: 1-12. PCL Holding Co., Ltd. Standard Operating Procedure ( SOP ) For Coulter ReticPakTM