4. Basic principles
• Best fit on regular surface
• Irregular surface needs customization
• Source to treatment depth best if 5mm
• Beyond 5 mm ,there will be overdose
• Keep the 150/200 isodose on mould
• 100% isodose on skin
• Follow DNR ratio
11. A) Axial CECT showing the Puffed Cheek technique in
B) That separates the buccal and gingival surfaces with air depicting that epicenter of lesion
(arrows) is in the buccal mucosa .The lesion does not abut mandible as appears in A
21. Reaction
Usually no reaction during discharge
Look at fortnight after discharge
If good reaction
Means
brachy has shown its action
Do not worry
No need to be sorry
Saline wash with steroid ointment
That is my commitment
It will heal
That is my deal
22.
23.
24. How to prevent bad reaction?
• Follow DNR
• Dose non uniformity ratio
103. TARGET DELINEATION
• Target definition of the Gross Tumor Volume (GTV)
and Clinical Target Volume (CTV) are usually
performed by clinical examination aided by imaging
and intraoperative findings.
• In brachytherapy, typically no additional margin is
required to ensure that theCTV receives the planned
dose and therefore.
• Unlike external beam irradiation, CTV and PTV
(Planning Target Volume) are considered equivalent
107. PLANNING
• ACCURATELY MEASURED CATHTER LENGTH BEFORE CT
• CATHTER NUMBERING AND MARKED
• CATHTER RECONSTRUCTION
• PLANNED BY IPSA
• OPTIMIZED GRAPHICALLY
• TREATMENT CALCULATION
• VERIFICATION
• EXECUTION
108. TARGET DOSE EVALUATION
• CTV coverage is adequate (i.e., D90 is at least 90% of
the prescribed dose)
• The prescription dose is usually the minimum dose
received by the CTV or a CTV surrogate (i.e., the D90 >
100, V100 > 90%)
• Dose non-uniformity ratio (DNR= V100/V150)
• DNR should be equal or lower than 0.36
• The optimal spacing between applicators is <15 mm
109. TARGET DOSE EVALUATION
COVERAGE INDEX
CTV VOLUME 0.32CC DESIRABLE
DNR INDEX V150 / V100 = 34.04/96.42 =0.35 0.36 T0 0.5
D90
98.34% >90%-100%
V100
96.42% >90%
V150
34.04% ALARA
V200
8.94% ALARA
V300
2.67% ALARA
110. ISODOSE IN ALL PLANES
BLUE 85%
RED 100%
ORANGE 150%
WHITE 200%
GREEN 300%
124. SUMMARY
• Follow GEC-ESTRO guidelines
• Train the brain of your dentist
• Follow steps
• Wires/source should not touch necked to any surface
• Use lead /IOP
• Reactions usually observed after treatment- do not
get panic , wait
• Counsel