Planning Of Radiological
Facility
By: Tanish khandelwal
Introduction
• In planning a radiology department from radiation safety
view point , the safety of all the concerned including
radiation workers, public at large, and patient ha to be
considered
• The aim of planning is to bring down the levels of
radiation exposure to as low as possible but well within
the permissible limits and to reduce exposures to
patients as well.
• The various sectors which contribute to the reduction of
doses are following:
…..CONT
1. Proper choice of the equipment
2. Proper design of the installation
3. Participation of qualified staff
Planning of Diagnostic Radiation
Department
1. Installation Design
• The installation planning includes the selection of
layout of various facilities , proper size of room,
structural shielding and location of control panel.
A. Location
•The site or room should be located
as far as possible from areas of high
occupancy and general traffic such
as maternity and pediatric wards and
other department of hospital that are
not directly related to radiation and it
use.
B. Layout
• The layout of the various facilities such as
dark room , doctors room, waiting area, etc., in
the x- ray department should aim at providing
integrated facilities so that handling of x-ray
equipment and related operations can be done
with adequate protection.
C. Room Size
• The room size should be such as to permit
installation, use and servicing of equipment with
safety and convenience for operating personnel,
servicing personnel and patient and to keep control
panel at reasonable safe distance which is minimum 2
metres from the machine
D. Shielding
• Appropriate shielding should be provided for the
walls, the ceiling and the floor of the room so that
the doses received by the occupationally exposed are
kept well below 20 msv and to the population well
below 1 msv.
E. Opening and ventilation
• Unshielded openings , if provided in the room for
ventilation or natural light must be located above a
height of 2m from the ground or finished floor level
outside the room.
F. Waiting area
• In order to avoid the crowding of patients and
relatives near the entrance door , waiting area must
be provided outside and adjacent the equipment
room.
G. Warning light and placard
• A suitable warning signal such as the red light must be
provided at a conspicuous place outside the room and
kept ON when the unit is in use, to prevent entry of
persons not connected with the examination or
treatment.
• An appropriate warning placard must also be posted
outside the room entrance or door.
2. PROPER CHOICE OF
EQUIPMENT
• Radiological examination should be carried out only
with those machines which are intended to be used
for such examination. For example a mobile or
portable x-ray machine with less protection should
not be used for routine and regular radiography.
• Units with ordinary patient table cannot be used for
special examination involving more personnel and
complicated movements of patient.
3. PARTICIPATION OF QUALIFIED
AND
TRAINED STAFF
• The participation of qualified staff is very important in
reducing the exposures. An unqualified person may take
more time for screening or may wrongly present the
radiograph compared to a qualified doctor.
• The radiologist and the technician both should be
trained in radiation safety aspects also so that they may
adopt all the steps to reduce the exposure to the
minimum level.
ESTABLISHING A DIAGNOSTIC
X-RAY FACILITY
1. ROOM SIZE
• The room housing an x-ray unit shall be not
less than 18 meter square for general purpose
radiography and conventional fluoroscopy
equipment.
• X-ray rooms shall not be less than 4 meter.
2. Wall Thickness
• If the X-ray installation is located in a residential complex, if shall be ensured
that
1. Walls of the X-ray rooms on which primary X-ray beams falls is not less than
35 cm or 14 inch thick brick or equivalent.
2. Walls of the X-ray rooms on which scattered X-ray falls is not less than 23 cm
or 9 inch thick brick or equivalent, and
3. There is a shielding equivalent to at least 23 cm or 9 inch thick brick or 1.7
mm lead in front of the doors and windows of the X-ray room to protect the
adjacent areas, either by general public or not under possession of the owner
of the X-ray room.
4. The ceiling must have a thickness of concrete not less than 6 inch or 13.5 cm.
3. CONTROL ROOM
• For equipment operating at 125 kV or above, the control panel must be
installed in a separate control room located outside but contiguous to
the machine room and provide with appropriate shielding, direct
viewing and oral communication facility between the operator and the
patient.
4.DOORS
• Doors to be lined with 1.5 mm thick lead sheet with proper
overlapping at the joint and junction and wall of 9 inch thickness of
brick and ceiling of 6 inch of concrete.
5.Dark Room
• The dark room should be located in such a way that the primary beam
is not directed on it.
• Appropriate shielding must be provided for the dark room.
6.Approval
• Two copies of the x-ray room layout drawn are sent for approval to
head, Radiological Safety Division, Atomic Energy Regulatory Board.
• AERB payable in Mumbai, towards charges for the approval of layout.
7. Placard
• A warning placard must be posted outside the room entrance or door.
Planning Of Radiology Department.pptx
Planning Of Radiology Department.pptx

More Related Content

What's hot

Dark Room Construction
Dark Room ConstructionDark Room Construction
Dark Room ConstructionTarique Ajij
 
Unit ii radio diagnosis and imaging services
Unit ii radio diagnosis and imaging servicesUnit ii radio diagnosis and imaging services
Unit ii radio diagnosis and imaging servicesanjalatchi
 
responsibility of radiographer
responsibility of radiographerresponsibility of radiographer
responsibility of radiographerBISHAL KHANAL
 
Planning of Nuclear Medicine Facility
Planning of Nuclear Medicine FacilityPlanning of Nuclear Medicine Facility
Planning of Nuclear Medicine FacilityMustafeed Uddin
 
Aerb guidelines for Room installation
Aerb guidelines for Room installationAerb guidelines for Room installation
Aerb guidelines for Room installationSelf
 
Mobile & portable radiography
Mobile & portable radiographyMobile & portable radiography
Mobile & portable radiographyPriyanka Parimala
 
Aerb guidelines for x ray and ct installation
Aerb guidelines for x ray and ct installationAerb guidelines for x ray and ct installation
Aerb guidelines for x ray and ct installationAabid Rahiman
 
Responsibility of radiographer
Responsibility of radiographerResponsibility of radiographer
Responsibility of radiographermohammed althaf
 
Radiation safety program
Radiation safety programRadiation safety program
Radiation safety programSatendra Kumar
 
X ray generators and its type
X ray generators and its typeX ray generators and its type
X ray generators and its typeDeepaknegi09
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x raySushilPattar
 
Mammographic equipment and its advancement
Mammographic equipment and its advancementMammographic equipment and its advancement
Mammographic equipment and its advancementYashawant Yadav
 
Primer in quality improvement in radiology department
Primer in quality improvement in radiology departmentPrimer in quality improvement in radiology department
Primer in quality improvement in radiology departmentAhmed Bahnassy
 
Quality Management for Diagnostic Imaging
Quality Management for Diagnostic ImagingQuality Management for Diagnostic Imaging
Quality Management for Diagnostic Imagingwnabxrayd
 
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptxLec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptxAli Ayaz
 

What's hot (20)

Dark Room Construction
Dark Room ConstructionDark Room Construction
Dark Room Construction
 
Unit ii radio diagnosis and imaging services
Unit ii radio diagnosis and imaging servicesUnit ii radio diagnosis and imaging services
Unit ii radio diagnosis and imaging services
 
Qc in xray
Qc in xrayQc in xray
Qc in xray
 
responsibility of radiographer
responsibility of radiographerresponsibility of radiographer
responsibility of radiographer
 
Pacs
PacsPacs
Pacs
 
Planning of Nuclear Medicine Facility
Planning of Nuclear Medicine FacilityPlanning of Nuclear Medicine Facility
Planning of Nuclear Medicine Facility
 
Radiation regulatory bodies
Radiation regulatory bodiesRadiation regulatory bodies
Radiation regulatory bodies
 
Quality assurance
Quality  assuranceQuality  assurance
Quality assurance
 
Aerb guidelines for Room installation
Aerb guidelines for Room installationAerb guidelines for Room installation
Aerb guidelines for Room installation
 
Mobile & portable radiography
Mobile & portable radiographyMobile & portable radiography
Mobile & portable radiography
 
Components of ct
Components of ctComponents of ct
Components of ct
 
Aerb guidelines for x ray and ct installation
Aerb guidelines for x ray and ct installationAerb guidelines for x ray and ct installation
Aerb guidelines for x ray and ct installation
 
Responsibility of radiographer
Responsibility of radiographerResponsibility of radiographer
Responsibility of radiographer
 
Radiation safety program
Radiation safety programRadiation safety program
Radiation safety program
 
X ray generators and its type
X ray generators and its typeX ray generators and its type
X ray generators and its type
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x ray
 
Mammographic equipment and its advancement
Mammographic equipment and its advancementMammographic equipment and its advancement
Mammographic equipment and its advancement
 
Primer in quality improvement in radiology department
Primer in quality improvement in radiology departmentPrimer in quality improvement in radiology department
Primer in quality improvement in radiology department
 
Quality Management for Diagnostic Imaging
Quality Management for Diagnostic ImagingQuality Management for Diagnostic Imaging
Quality Management for Diagnostic Imaging
 
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptxLec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
 

Similar to Planning Of Radiology Department.pptx

Radiologydepartment_INTRODUCTION AND PLANING.pptx
Radiologydepartment_INTRODUCTION AND PLANING.pptxRadiologydepartment_INTRODUCTION AND PLANING.pptx
Radiologydepartment_INTRODUCTION AND PLANING.pptxDRTRUPTISONTHALIA
 
Radioghrapic imaging Room.pptx
Radioghrapic imaging Room.pptxRadioghrapic imaging Room.pptx
Radioghrapic imaging Room.pptxShah Syed Taqqi
 
Planning-Organizing-imaging-services-pdf.pdf
Planning-Organizing-imaging-services-pdf.pdfPlanning-Organizing-imaging-services-pdf.pdf
Planning-Organizing-imaging-services-pdf.pdfyaskumaumaridriss
 
RADIATION SAFETY AND PROTECTION-2.pptx
RADIATION SAFETY AND PROTECTION-2.pptxRADIATION SAFETY AND PROTECTION-2.pptx
RADIATION SAFETY AND PROTECTION-2.pptxPooja461465
 
Ot infection control rkch
Ot infection control rkchOt infection control rkch
Ot infection control rkchsabahjak
 
literature study for EMERGENCY DEPARTMENT in hospital
literature study for EMERGENCY DEPARTMENT in hospitalliterature study for EMERGENCY DEPARTMENT in hospital
literature study for EMERGENCY DEPARTMENT in hospitalBalaji Ar
 
(Hospital presentation)
(Hospital presentation)(Hospital presentation)
(Hospital presentation)Bee Key Verma
 
Operating theatre suite planning and design
Operating theatre suite planning and designOperating theatre suite planning and design
Operating theatre suite planning and designMahmoud Shaqria
 
1.introdaction OR.ppt
1.introdaction OR.ppt1.introdaction OR.ppt
1.introdaction OR.pptTadesseFenta1
 
radiation protection.pptx
radiation protection.pptxradiation protection.pptx
radiation protection.pptxPooja461465
 
5.radiation protection
5.radiation protection5.radiation protection
5.radiation protectionsamad shaik
 
Superficial radiotherapy (sxt)_following_ear_lobe
Superficial radiotherapy (sxt)_following_ear_lobeSuperficial radiotherapy (sxt)_following_ear_lobe
Superficial radiotherapy (sxt)_following_ear_lobemenkantozz
 
Handbook to build_an_hospital_crf
Handbook to build_an_hospital_crfHandbook to build_an_hospital_crf
Handbook to build_an_hospital_crfsubh111
 
Guidelines on Radiation Safety in Dentistry
Guidelines on Radiation Safety in DentistryGuidelines on Radiation Safety in Dentistry
Guidelines on Radiation Safety in DentistryChow Peng Yue
 
Guidelines for layout as per AERB
Guidelines for layout as per AERBGuidelines for layout as per AERB
Guidelines for layout as per AERBDr Jitu Lal Meena
 
Emergency department planning and design
Emergency department planning and designEmergency department planning and design
Emergency department planning and designMahmoud Shaqria
 

Similar to Planning Of Radiology Department.pptx (20)

Radiologydepartment_INTRODUCTION AND PLANING.pptx
Radiologydepartment_INTRODUCTION AND PLANING.pptxRadiologydepartment_INTRODUCTION AND PLANING.pptx
Radiologydepartment_INTRODUCTION AND PLANING.pptx
 
Radioghrapic imaging Room.pptx
Radioghrapic imaging Room.pptxRadioghrapic imaging Room.pptx
Radioghrapic imaging Room.pptx
 
Planning-Organizing-imaging-services-pdf.pdf
Planning-Organizing-imaging-services-pdf.pdfPlanning-Organizing-imaging-services-pdf.pdf
Planning-Organizing-imaging-services-pdf.pdf
 
Introduction to icu.pptx
Introduction to icu.pptxIntroduction to icu.pptx
Introduction to icu.pptx
 
RADIATION SAFETY AND PROTECTION-2.pptx
RADIATION SAFETY AND PROTECTION-2.pptxRADIATION SAFETY AND PROTECTION-2.pptx
RADIATION SAFETY AND PROTECTION-2.pptx
 
Ot infection control rkch
Ot infection control rkchOt infection control rkch
Ot infection control rkch
 
literature study for EMERGENCY DEPARTMENT in hospital
literature study for EMERGENCY DEPARTMENT in hospitalliterature study for EMERGENCY DEPARTMENT in hospital
literature study for EMERGENCY DEPARTMENT in hospital
 
SURGICAL THEATRE DESIGN.pptx
SURGICAL THEATRE DESIGN.pptxSURGICAL THEATRE DESIGN.pptx
SURGICAL THEATRE DESIGN.pptx
 
Icu
IcuIcu
Icu
 
(Hospital presentation)
(Hospital presentation)(Hospital presentation)
(Hospital presentation)
 
Operating theatre suite planning and design
Operating theatre suite planning and designOperating theatre suite planning and design
Operating theatre suite planning and design
 
1.introdaction OR.ppt
1.introdaction OR.ppt1.introdaction OR.ppt
1.introdaction OR.ppt
 
radiation protection.pptx
radiation protection.pptxradiation protection.pptx
radiation protection.pptx
 
5.radiation protection
5.radiation protection5.radiation protection
5.radiation protection
 
Superficial radiotherapy (sxt)_following_ear_lobe
Superficial radiotherapy (sxt)_following_ear_lobeSuperficial radiotherapy (sxt)_following_ear_lobe
Superficial radiotherapy (sxt)_following_ear_lobe
 
Handbook to build_an_hospital_crf
Handbook to build_an_hospital_crfHandbook to build_an_hospital_crf
Handbook to build_an_hospital_crf
 
Guidelines on Radiation Safety in Dentistry
Guidelines on Radiation Safety in DentistryGuidelines on Radiation Safety in Dentistry
Guidelines on Radiation Safety in Dentistry
 
Guidelines for layout as per AERB
Guidelines for layout as per AERBGuidelines for layout as per AERB
Guidelines for layout as per AERB
 
radiation
radiation radiation
radiation
 
Emergency department planning and design
Emergency department planning and designEmergency department planning and design
Emergency department planning and design
 

Recently uploaded

No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...
No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...Vip call girls In Chandigarh
 
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment BookingModels Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Call Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any TimeCall Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any Timedelhimodelshub1
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed RuleShelby Lewis
 
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service HyderabadVIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabaddelhimodelshub1
 
Russian Escorts Delhi | 9711199171 | all area service available
Russian Escorts Delhi | 9711199171 | all area service availableRussian Escorts Delhi | 9711199171 | all area service available
Russian Escorts Delhi | 9711199171 | all area service availablesandeepkumar69420
 
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...delhimodelshub1
 
Call Girls Uppal 7001305949 all area service COD available Any Time
Call Girls Uppal 7001305949 all area service COD available Any TimeCall Girls Uppal 7001305949 all area service COD available Any Time
Call Girls Uppal 7001305949 all area service COD available Any Timedelhimodelshub1
 
Call Girls Gurgaon Parul 9711199012 Independent Escort Service Gurgaon
Call Girls Gurgaon Parul 9711199012 Independent Escort Service GurgaonCall Girls Gurgaon Parul 9711199012 Independent Escort Service Gurgaon
Call Girls Gurgaon Parul 9711199012 Independent Escort Service GurgaonCall Girls Service Gurgaon
 
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...soniya singh
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxAyush Gupta
 
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliCall Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliHigh Profile Call Girls Chandigarh Aarushi
 
Low Rate Call Girls In Bommanahalli Just Call 7001305949
Low Rate Call Girls In Bommanahalli Just Call 7001305949Low Rate Call Girls In Bommanahalli Just Call 7001305949
Low Rate Call Girls In Bommanahalli Just Call 7001305949ps5894268
 
Call Girls Secunderabad 7001305949 all area service COD available Any Time
Call Girls Secunderabad 7001305949 all area service COD available Any TimeCall Girls Secunderabad 7001305949 all area service COD available Any Time
Call Girls Secunderabad 7001305949 all area service COD available Any Timedelhimodelshub1
 
Call Girls Service Bommasandra - Call 7001305949 Rs-3500 with A/C Room Cash o...
Call Girls Service Bommasandra - Call 7001305949 Rs-3500 with A/C Room Cash o...Call Girls Service Bommasandra - Call 7001305949 Rs-3500 with A/C Room Cash o...
Call Girls Service Bommasandra - Call 7001305949 Rs-3500 with A/C Room Cash o...narwatsonia7
 
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...delhimodelshub1
 

Recently uploaded (20)

No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...
No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...
 
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment BookingModels Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
 
Call Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any TimeCall Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any Time
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
 
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service HyderabadVIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
 
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service GuwahatiCall Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
 
VIP Call Girls Lucknow Isha 🔝 9719455033 🔝 🎶 Independent Escort Service Lucknow
VIP Call Girls Lucknow Isha 🔝 9719455033 🔝 🎶 Independent Escort Service LucknowVIP Call Girls Lucknow Isha 🔝 9719455033 🔝 🎶 Independent Escort Service Lucknow
VIP Call Girls Lucknow Isha 🔝 9719455033 🔝 🎶 Independent Escort Service Lucknow
 
College Call Girls Dehradun Kavya 🔝 7001305949 🔝 📍 Independent Escort Service...
College Call Girls Dehradun Kavya 🔝 7001305949 🔝 📍 Independent Escort Service...College Call Girls Dehradun Kavya 🔝 7001305949 🔝 📍 Independent Escort Service...
College Call Girls Dehradun Kavya 🔝 7001305949 🔝 📍 Independent Escort Service...
 
Russian Escorts Delhi | 9711199171 | all area service available
Russian Escorts Delhi | 9711199171 | all area service availableRussian Escorts Delhi | 9711199171 | all area service available
Russian Escorts Delhi | 9711199171 | all area service available
 
Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service LucknowCall Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
 
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
 
Call Girls Uppal 7001305949 all area service COD available Any Time
Call Girls Uppal 7001305949 all area service COD available Any TimeCall Girls Uppal 7001305949 all area service COD available Any Time
Call Girls Uppal 7001305949 all area service COD available Any Time
 
Call Girls Gurgaon Parul 9711199012 Independent Escort Service Gurgaon
Call Girls Gurgaon Parul 9711199012 Independent Escort Service GurgaonCall Girls Gurgaon Parul 9711199012 Independent Escort Service Gurgaon
Call Girls Gurgaon Parul 9711199012 Independent Escort Service Gurgaon
 
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptx
 
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliCall Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
 
Low Rate Call Girls In Bommanahalli Just Call 7001305949
Low Rate Call Girls In Bommanahalli Just Call 7001305949Low Rate Call Girls In Bommanahalli Just Call 7001305949
Low Rate Call Girls In Bommanahalli Just Call 7001305949
 
Call Girls Secunderabad 7001305949 all area service COD available Any Time
Call Girls Secunderabad 7001305949 all area service COD available Any TimeCall Girls Secunderabad 7001305949 all area service COD available Any Time
Call Girls Secunderabad 7001305949 all area service COD available Any Time
 
Call Girls Service Bommasandra - Call 7001305949 Rs-3500 with A/C Room Cash o...
Call Girls Service Bommasandra - Call 7001305949 Rs-3500 with A/C Room Cash o...Call Girls Service Bommasandra - Call 7001305949 Rs-3500 with A/C Room Cash o...
Call Girls Service Bommasandra - Call 7001305949 Rs-3500 with A/C Room Cash o...
 
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
 

Planning Of Radiology Department.pptx

  • 2. Introduction • In planning a radiology department from radiation safety view point , the safety of all the concerned including radiation workers, public at large, and patient ha to be considered • The aim of planning is to bring down the levels of radiation exposure to as low as possible but well within the permissible limits and to reduce exposures to patients as well. • The various sectors which contribute to the reduction of doses are following:
  • 3. …..CONT 1. Proper choice of the equipment 2. Proper design of the installation 3. Participation of qualified staff
  • 4. Planning of Diagnostic Radiation Department 1. Installation Design • The installation planning includes the selection of layout of various facilities , proper size of room, structural shielding and location of control panel.
  • 5. A. Location •The site or room should be located as far as possible from areas of high occupancy and general traffic such as maternity and pediatric wards and other department of hospital that are not directly related to radiation and it use.
  • 6. B. Layout • The layout of the various facilities such as dark room , doctors room, waiting area, etc., in the x- ray department should aim at providing integrated facilities so that handling of x-ray equipment and related operations can be done with adequate protection.
  • 7. C. Room Size • The room size should be such as to permit installation, use and servicing of equipment with safety and convenience for operating personnel, servicing personnel and patient and to keep control panel at reasonable safe distance which is minimum 2 metres from the machine
  • 8. D. Shielding • Appropriate shielding should be provided for the walls, the ceiling and the floor of the room so that the doses received by the occupationally exposed are kept well below 20 msv and to the population well below 1 msv.
  • 9. E. Opening and ventilation • Unshielded openings , if provided in the room for ventilation or natural light must be located above a height of 2m from the ground or finished floor level outside the room.
  • 10. F. Waiting area • In order to avoid the crowding of patients and relatives near the entrance door , waiting area must be provided outside and adjacent the equipment room.
  • 11. G. Warning light and placard • A suitable warning signal such as the red light must be provided at a conspicuous place outside the room and kept ON when the unit is in use, to prevent entry of persons not connected with the examination or treatment. • An appropriate warning placard must also be posted outside the room entrance or door.
  • 12. 2. PROPER CHOICE OF EQUIPMENT • Radiological examination should be carried out only with those machines which are intended to be used for such examination. For example a mobile or portable x-ray machine with less protection should not be used for routine and regular radiography. • Units with ordinary patient table cannot be used for special examination involving more personnel and complicated movements of patient.
  • 13. 3. PARTICIPATION OF QUALIFIED AND TRAINED STAFF • The participation of qualified staff is very important in reducing the exposures. An unqualified person may take more time for screening or may wrongly present the radiograph compared to a qualified doctor. • The radiologist and the technician both should be trained in radiation safety aspects also so that they may adopt all the steps to reduce the exposure to the minimum level.
  • 15. 1. ROOM SIZE • The room housing an x-ray unit shall be not less than 18 meter square for general purpose radiography and conventional fluoroscopy equipment. • X-ray rooms shall not be less than 4 meter.
  • 16. 2. Wall Thickness • If the X-ray installation is located in a residential complex, if shall be ensured that 1. Walls of the X-ray rooms on which primary X-ray beams falls is not less than 35 cm or 14 inch thick brick or equivalent. 2. Walls of the X-ray rooms on which scattered X-ray falls is not less than 23 cm or 9 inch thick brick or equivalent, and 3. There is a shielding equivalent to at least 23 cm or 9 inch thick brick or 1.7 mm lead in front of the doors and windows of the X-ray room to protect the adjacent areas, either by general public or not under possession of the owner of the X-ray room. 4. The ceiling must have a thickness of concrete not less than 6 inch or 13.5 cm.
  • 17. 3. CONTROL ROOM • For equipment operating at 125 kV or above, the control panel must be installed in a separate control room located outside but contiguous to the machine room and provide with appropriate shielding, direct viewing and oral communication facility between the operator and the patient.
  • 18. 4.DOORS • Doors to be lined with 1.5 mm thick lead sheet with proper overlapping at the joint and junction and wall of 9 inch thickness of brick and ceiling of 6 inch of concrete.
  • 19. 5.Dark Room • The dark room should be located in such a way that the primary beam is not directed on it. • Appropriate shielding must be provided for the dark room.
  • 20. 6.Approval • Two copies of the x-ray room layout drawn are sent for approval to head, Radiological Safety Division, Atomic Energy Regulatory Board. • AERB payable in Mumbai, towards charges for the approval of layout.
  • 21. 7. Placard • A warning placard must be posted outside the room entrance or door.