recent advances in anesthesia machine design (2).ppt
1.
2.
3. Anesthesia Machine
History of anesthesia machine
Components of modern anesthesia machine
Safety features of modern anesthesia machine
Anesthesia machine Check list
5. Morton’s ether inhaler… the exact story of
anesthesia machine starts from 1864 Oct. 16th when
Morton demonstrated the first successful ether
anesthesia for dental operation .
6. John Snow apparatus
Snow created the first of his series of ether inhalers
that featured unidirectional valves within a malleable,
well fitting mask of his own design, which closely
resembles the form of a modern face mask. The face
piece was connected to the vaporizer by a breathing
tube.
7. Joseph Clover
He was the first physician to administer chloroform
in known concentrations through the “Clover bag”.
8. Boyle Apperatus
in 1917 developed his continuous-flow anesthesia
machine, the design of which is the forerunner of all
modern anesthetic machines, it was first, to use N2O-O2
system with water sight feed type of flowmeter.
9. Boyle’s machine consist of
A high-pressure supply of gases. It housed two oxygen and
two nitrous oxide cylinders in a wooden box.
Pressure gauges on oxygen cylinders and fine-adjustment
reducing valves. These produce a manageable breathing
system pressure. It had a spirit flame to warm these and
prevent obstruction of gas flow from ice.
Flowmeters to control gas flow rate and adjust
proportions of gas delivered.
A metal and glass vaporizer bottle for ether.
10. Evolution of Carbon Dioxide Absorbent
Franz Kuhn, a German surgeon. His use of
canisters developed for mine rescues by Dragger
was innovative.
11. Early anesthesia vaporizers
The Junker’s inhaler (forerunner of gases under
pressure) was the first device in which an attempt
was made to increase the gas liquid interface.
The Copper Kettle was the first temperature
compensated accurate vaporizer.
The TECOTA(Temperature Compensated
Trichloroethylene Air) vaporizer appeared when
halothane was first marketed in England. It is an
effective temperature-compensated, agent specific
vaporizer.
“Fluotec,” the first of a series of agent-specific
“tec” vaporizers for use in the operating room.
12.
13. Evolution of the anesthesia
ventilator
In 1907, the first intermittent positive-pressure
device, the Dragger “Pulmotor,” was developed
to inflate the lungs during cardiothoracic
surgery.
A further boost to the development of
automatic artificial ventilators occurred in 1952,
when a catastrophic poliomyelitis epidemic
struck Denmark. There was a very high
incidence of bulbar paralysis. By using
tracheostomy and manual positive pressure
ventilation physicians overcome the catastrophe.
19. Medical compressed gases supply
Gases used in the administration of anesthesia
(oxygen,nitrous oxide,air) are most often delivered to
the anesthesia machine via two sources pipline and
cylinder:
Pipeline supply source:
In the majority of modern hospitals, piped medical gases and vaccum (
PMGV ) system have been installed to distribute oxygen, nitrous oxide,
medical air and vacuum. The pipelines’ outlets act as self-closing
sockets, each specifically configured, coloured and labelled for one gas.
Oxygen, nitrous oxide and air are delivered to the anesthetic room at a
pressure (4bar, 60 pounds per square inch (psi)).
20.
The PMGV services comprise five sections
Bulk store.
Distribution pipelines in the hospital.
Terminal outlets, situated usually on the walls or
ceilings of the operating theatre and other sites.
flexible hoses connecting the terminal outlet to the
anesthetic machine. colour coded throughout their
length (oxygen white, nitrous oxide blue, vacuum
yellow ).
Connections between flexible hoses and Anesthesia
machine.
21.
22. Bulk stores:
•Oxygen:
The standard bulk oxygen supply should consist of three
cryogenic oxygen storage tanks: a primary tank, A; a secondary
tank, B; and a reserve tank, C (Tanks A and B are physically
situated next to each other. The reserve tank (C) is located
approximately 1 block (approximately 305 m) away.
In small hospitals, oxygen may be supplied to the PMGV from
a bank of several oxygen cylinders attached to a manifold.
23. Oxygen cylinder manifolds consist of two groups of
large cylinders (size J). The two groups alternate in
supplying oxygen to the pipelines. In both groups, all
cylinder valves are open so that they empty
simultaneously.
All cylinders have non-return valves. The supply
automatically changes from one group to the other
when the first group of cylinders is nearly empty. The
changeover also activates an electrical signalling
system which alerts staff to change the empty
cylinders.
24. Nitrous oxide
Nitrous oxide , also may be supplied from banks of
cylinders connected to manifolds similar to those used
for oxygen.
Medical compressed air
Compressed air is supplied from a bank of cylinders into
the PMGV system . Air of medical quality is required as
industrial compressed air may contain fine particles of oil
.
25.
26. Piped medical vaccum
Piped medical vacuum is provided by large vacuum pumps
which discharge via a filter and silencer to a suitable point ,
usually at roof level , where gases are vented to atmosphere.
Terminal outlets
The terminal outlets are colour-coded with noninterchangeable
connections specific to each gas .
Cylinder supply source:
Oxygen on anesthesia machines, the cylinder serves as a
backup supply its of E size and have an internal volume of
4.8 L. about 650 L. colour coded with PISS prevent wrong
connection.
27. Nitrous oxide: cylinders arecolour coded blue with
PISS.
Entonox(oxygen and nitrous)cylinders colour
coded white and blue.
Heliox(helium and oxygen)
28. Anesthesia gas delivery system
Anesthesia gas delivery systems or anesthesia machines should have
aback bar which is ISO pin type to attach vaporizers easily and
attachments to use the breathing circuits(CGO) and top shelf to keep
monitors and a tabletop to keep anesthetic drugs and equipments.
Gas inlet receives gas from pipline or cylinder
Pressure regulators reducing pressure entering the machine
1. Fail safe: Oxygen pressure failure devices signal low oxygen
pressuere and cut of nitrous oxide supply
2. Flow control valves and flowmeters control flow rate:These reduce
the pressure from 4 bar to just above atmospheric
3. . A fresh gas outlet delivers final gas composition to breathing circuit
29.
30. Design Limitations of Conventional
Anesthesia Machine
Presence of many external connections subjected
to kinking, obstruction and disconnection.
Lack of protection against barotruma.
Lack of automated checkout.
Lack of advanced ventilation features.
Lack of low-flow adaptation.
Vaporizer risks.
32. Hypoxic inspired gas mixture.
Hypoventilation.
Hypercapnia (absent/ineffective CO2 absorbent).
Excessive airway pressure.
Inhalation of foreign material.
Over-dosage of the anesthetic agent.
Inadequate anesthetic concentration.
Fires
Exposure to waste anesthetic gases
33. Safety features in modern anesthesia
machine
Poor equipment design has been identified as a major
factor in many adverse events which directly affect
patient safety. Modern anesthetic machines are
designed with many safety devices which include :
Electrical system :
Electrical power is essential to anesthesia machine as
well as monitors to function normally. Modern
anesthesia machines have a power failure alarm and
back-up battery which designed to provide temporary
power until emergency generator is on
34. safety feature in gas supply systems
design :
Pipeline system :
Pipeline design must provide reserve supplies and
reserve plant to ensure continuous supply, non
interchangeability between different gas services,
monitoring and alarm system and correct installation,
testing, certification of the system and correct marking.
35. Connectors and Connections
Diameter Index Safety System(DISS) ; which prevents
hoses being accidentally plugged into the wrong wall
socket.
Colour coding; flexible colour-coded hoses; connect
the outlets to the anesthetic machine .The anesthetic
machine end should be permanently fixed .
Connectors ; each has a different size and
configuration to prevent misconnection.
Pin Index Safety System (PISS); prevent attachment of
wrong cylinders.
36.
37. Pressure Regulator
providing a constant low pressure suitable for the
machine from the variable high-pressure cylinders.
Some machines use two regulators whereby the second
regulator delivers the gas to flowmeters at just above
atmospheric pressure.
Unidirectional flow valve permits gas to enter the
machine not to get out.
Oxygen Flush this switch directs a high pressure flow
of oxygen direct to the CGO from the source, either
pipeline or cylinder, bypassing all intermediate meters
and vaporizers.
38. Flowmeters :
Flowmeter Knobs: are of the same colours of
corresponding gas and of the same order of
flowmeters. Oxygen flowmeters should have a
unique profile.
Tubes: are made leak-proof and have an antistatic
coating on both surfaces, preventing the bobbin
from sticking.
The bobbin: is visible throughout the length of
the tube.
39.
40. Hypoxia prevention devices:
the main goal of design of modern anesthesia machine is to prevent
delivery of hypoxic gas mixture.
Mandatory minimum oxygen flow: modern machines have a
minimum pre-set oxygen flow, which will automatically start once
the machine is powered on.
Minimum oxygen ratio . In contemporary machines, the flows of
oxygen and nitrous oxide are interlinked so that a fresh gas mixture
containing at least 25%.
oxygen analyzer which equipped with an audible alarm is to alert
to low concentrations of oxygen.
fail-safe valve: a fail-safe valve is present in the gas line supplying
each of the flowmeters except oxygen. Controlled by oxygen supply
pressure, the valve shuts off the supply pressure of all other gases
(nitrous oxide or air) as the oxygen supply pressure decreases.
41.
42. Safety features of Vaporizers
Keyed fillers :
An agent-specific colour- and shape-coded device. Low filling
port and presence of maximum filling mark prevent over filling.
Vaporizer interlock :
Interlock ensures that only one vaporizer is turned on.
Anti-spill mechanism:
The Tec 5 generation of vaporizers and later models have an
anti-spill device, which prevents volatile agents from entering the
bypass chamber.
Release button
Must be activated before dial can be turned on.
43.
44. Monitoring
Oxygen analyser ; General anaesthesia should never be
administered without an oxygen analyser in the
breathing circuit.
Airway pressure monitoring
Gas volume monitoring
Disconnection monitors are an integral component of
newer anaesthesia machines.
45. Criteria to determine obsolescence
of anaesthesia machines
Anaesthesia machines may be considered obsolete
(to be discarded) if they do not offer the following
mandatory safety features:
46. Lack of certain safety features
Minimum oxygen ratio device in a machine, which
can provide nitrous oxide as well.
Oxygen failure safety device.
Oxygen supply pressure failure alarm.
Vaporizer interlock device (may be waived off if
machine is capable of accepting only one vaporizer).
Pin index safety system and Non-interchangeable
gas connectors like DISS.
47. Presence of unacceptable features
Measured flow (flow meter-controlled) vaporizers.
(e.g., Copper Kettle, Verni-trol).
More than one flow control knob for a single gas
delivered to the CGO of the machine.
Vaporizer with rotary concentration dials such that the
anaesthetic vapour concentration
increases when the dial is turned clockwise.
Connection (s) in scavenging system of the same (i.e.,
15-mm or 22-mm) diameter as a breathing system
connection.
48. Lack of certain desirable features to be
considered
Means to isolate the adjustable pressure-limiting valve during
mechanical ventilation.
Oxygen flow control knob that is fluted and larger than the
other flow control knobs.
Oxygen flush control protected from accidental
activation.
Main on/off switch for electrical power to integral monitors
and alarms.
Anti-disconnection device at the fresh gas outlet.
Airway pressure alarm (for detecting sustained positive
pressure, negative pressure and high peak pressure).