APPLICATIONS
TN-1104
Page 1 of 8For additional technical notes, visit www.phenomenex.com
Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analyses
When Exploiting the Properties of 2.6 μm Kinetex®
Core-Shell Columns
James Rudge1, David Neville2
, Richard Houghton2
, Chris Message1
, and Jeremy Bierman3
1
Phenomenex LTD, Queens Avenue, Hurdsfield Industrial Estate, Macclesfield, Cheshire, SK10 2BN, United Kingdom
3
Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 90501 USA
2
Quotient Bioresearch Ltd., Newmarket Road, Fordham, Cambridgeshire, CB7 5WW, United Kingdom
When compared to 1.7 μm fully porous materials, the ultra-high
efficiency and low backpressures provided by Kinetex core-shell
2.6 μm columns, provides users opportunities to go beyond what
is traditionally accepted for UHPLC runs. To match backpressures
exerted by 1.7 µm fully porous columns, large increases in core-
shell column length and/or system flow rate is required. When gra-
dient conditions are maintained, increases in flow rate and length
result in an increase in peak capacity on the order of 26 %. When
MS conditions are optimized to compensate for new flow rates,
sensitivity increases of up to 3.5 times are observed and further-
more, peak capacity figures increase by almost 70 %.
Introduction
The introduction of Kinetex core-shell columns has brought dra-
matic benefits to chromatographers. The high efficiency shown
from the 2.6 μm core-shell particle has allowed users to run ultra-
high performance separations on HPLC and UHPLC systems alike.
Moreover, the larger particle size of the Kinetex 2.6 μm material
compared to 1.7 μm fully porous particles significantly reduces
backpressures by a factor of ~40 %. This decrease in backpressure
provides users an opportunity to exploit flow rates and/or column
lengths to allow for faster UHPLC run times, higher peak capaci-
ties, and improvements in peak sensitivities.
Improvements in peak capacity can be realized if gradient rates are
not changed and flow rates and/or column lengths are increased.
The reason for this is that when the same gradient rate is applied,
faster flow rates act to volumetrically lower gradients. However,
if column length is increased and gradient time is also increased
(with respect to the change in length) then it is known that again
peak capacity values will increase due to the increased efficiency
of the longer column, but this comes at the expense of increased
run times.
Improved peak capacities observed over identical gradient win-
dows are due to a reduction in peak widths. If smaller peak widths
are observed, then this should give an opportunity for larger peak
heights and hence better peak sensitivity. However, attention should
be given to optimizing the detector to realize this benefit fully.
The work discussed in this technical note demonstrates perfor-
mance advantages when operating 2.6 µm Kinetex core-shell par-
ticle columns under conditions that deliver backpressures match-
ing that of a 50 x 2.1 mm fully porous 1.7 µm C18 column that is
operated at traditional UHPLC flow rates (0.8 mL per minute) for a
2.1 mm ID column.
Materials and Methods
All analytes were purchased from Sigma-Aldrich.
Laboratory:
All work was carried out by Quotient Bioresearch Laboratories.
Columns Used:
A fully porous 1.7 μm C18, 50 x 2.1 mm column was compared
with Kinetex 2.6 μm C18, 50 x 2.1 mm, 75 x 2.1 mm, and 100 x
2.1 mm columns from Phenomenex.
Instrumentation:
Waters ACQUITY®
Binary UPLC®
AB SCIEX API 5000™
Mobile Phase Preparation:
Solutions of 0.1 % formic acid in water and acetonitrile
Standard Solution Preparation:
A 0.5 ng/mL solution containing the following analytes was
prepared in water.
1. Haloperidol
2. Diltiazem
3. Terfenadine
4. Cimetidine
5. Acetaminophen
6. Sulfathiazole
7. Pindolol
8. Quinidine
9. Acebutolol
10. Chlorpheniramine
11. Tripolidine
12. Prednisolone
13. Nortriptyline
14. 2-Hydroxy-5-methyl benzaldehyde
15.	Hexanophenone
Experiment 1 – Performance comparison between 2.6 µm
Kinetex core-shell column and 1.7 µm fully porous column at the
same flow rate and then at the same pressure on varying column
lengths:
A 5 µL sample of all 15 components was injected with a gradient
chromatographic separation using an initial flow rate of 0.8 mL/
min over 2.5 minutes on a fully porous C18 1.7 µm, 50 x 2.1 mm
column and then on a Kinetex C18 2.6 µm, 50 x 2.1 mm column.
To match the backpressure of the fully porous column (531 bar,
7700 psi), the flow rate was then increased to 1.4 mL/min on the
Kinetex column. Next a Kinetex C18 2.6 µm, 75 x 2.1 mm column
APPLICATIONS
TN-1104
Page 2 of 8 For additional technical notes, visit www.phenomenex.com
was taken and run at 1.1 mL/min, again to match the backpres-
sure of the fully porous column. Finally, a Kinetex C18 2.6 µm,
100 x 2.1 mm Kinetex column was taken and run at 0.85 mL/
min, again to run at the same backpressure of the fully porous 1.7
µm column at 0.8 mL per minute. All columns were maintained at
40 ºC and data was collected using MS detection (For initial MS
conditions see Table 6, column 2).
Experiment 2 – Observed performance under optimized running
conditions with an optimized MS detector:
Aliquots of 5 µL of the analyte mix were repeatedly injected with
a gradient chromatographic separation using an initial flow rate of
0.8 mL/min on a Kinetex C18 2.6 µm, 50 x 2.1 mm column and
the total run time was 2.5 minutes. Peak height, peak width, and
peak capacity data was collected. Following this, the instrument
flow rate was increased to 1.4 mL/min and the data was collected
again. The columns were maintained at 40 ºC with MS detection
where the source was optimized for each flow rate (see “Opt 0.8
mL/min” and “Opt 1.4 mL/min” columns of Table 2). To increase
the number of data points across each peak, subsets of analytes
were grouped into 4 groups of 3 and MRM data was collected for
three masses per run. Table 1 lists each group and their respec-
tive parent and product ions.
Experiment 3 – Measured effect of optimized MS detector on
observed performance under sub-optimal running conditions:
Aliquots of 5 µL of the analyte mix were repeatedly injected with
a gradient chromatographic separation using an initial flow rate
of 0.8 mL/min on a fully porous C18 1.7 µm, 50 x 2.1 mm column
and on a Kinetex C18 2.6 µm, 50 x 2.1 mm column. The total
run time was 2.5 minutes and peak height, peak width, and peak
capacity data was collected at a flow rate optimized for 0.8 mL
per minute (Program A). Using the Kinetex column, the instrument
flow rate was increased to 1.4 mL/min (Program B) and the data
was collected again. Finally, at 1.4 mL per minute, the gradient
was adjusted with respect to the change in flow rate from 0.8 to
1.4 mL per minute. The columns were maintained at 40 ºC with
MS detection where the source was optimized for each flow rate
(see “Opt 0.8 mL/min” and “Opt 1.4 mL/min” columns of Table 2).
Like Experiment 2, to increase the number of data points across
each peak, subsets of analytes were grouped into 4 groups of 3
and MRM data was collected for three masses per run. Table 1
lists each group and their respective parent and product ions.
Table 1.
List of grouped analytes and their respective parent and product ions.
Group
Number Sample Name
Q1 mass
(m/z)
Q3 mass
(m/z)
Dwell time
(ms)
Group 1 Acetaminophen 152.2 110.1 50
Cimetidine 253.2 159.1 50
Pindolol 249.3 116.1 50
Group 2 Sulfathiazole 256.1 91.9 50
Chlorpheniramine 275.2 230 50
Triprolidine 279.4 208.1 50
Group 3 Diltiazem 415.3 178.1 50
Terfenadine 472.4 436.4 50
Acebutolol 337.3 116.2 50
Group 4 Haloperidol 376.3 123 50
Quinidine 325.3 79.1 50
Nortriptyline 264.3 91.1 50
Conditions are the same for all columns, except where noted.
Column: Kinetex 2.6 μm C18 100 Å
Fully Porous 1.7 μm C18
Dimensions: Kinetex: 50 x 2.1 mm, 75 x 2.1 mm, and 100 x 2.1 mm
Fully Porous: 50 x 2.1 mm
Flow Rate: 0.8, 1.1, and 1.4 and 0.85 mL/min
Mobile Phase: A: 0.1% Formic acid in Water
B: 0.1% Formic acid Acetonitrile
Gradient:
Temperature: 40 ºC
Time (min) % A
0.00 95
0.25 95
1.80 10
1.90 10
1.91
2.50
95
95
Conditions are the same for all columns, except where noted.
Column: Kinetex 2.6 μm C18 100 Å
Fully Porous 1.7 μm C18
Dimensions: 50 x 2.1 mm
Flow Rate: 0.8 and 1.4 mL/min
Mobile Phase: A: 0.1% Formic acid in Water
B: 0.1% Formic acid Acetonitrile
Gradient:
Temperature: 40 ºC
Time (min) % A
0.00 95
0.25 95
1.80 10
1.90 10
1.91
2.50
95
95
Conditions are the same for all columns, except where noted.
Column: Kinetex 2.6 μm C18 100 Å
Fully Porous 1.7 μm C18
Dimensions: 50 x 2.1 mm
Flow Rate: 0.8 and 1.4 mL/min
Mobile Phase: A: 0.1% Formic acid in Water
B: 0.1% Formic acid Acetonitrile
Gradient:
Temperature: 40 ºC
Program A (0.8 mL/min)
Time (min) % A
0.00 95
0.25 95
1.80 10
1.90 10
1.91
2.50
95
95
Program B (1.4 mL/min)
Time (min) % A
0.00 95
0.25 95
1.80 10
1.90 10
1.91
2.50
95
95
APPLICATIONS
TN-1104
Page 3 of 8For additional technical notes, visit www.phenomenex.com
fully porous column. The resulting peak capacity (n = 34.6) was
almost the same as the fully porous material (see Table 3). The
reason for this was that there were almost half the number of col-
umn volumes passed down the longer column. So although the
100 mm length column would have been a lot more efficient than
the 50 mm length 1.7 µm fully porous column, the gradient had
been volumetrically steepened and so any gains in performance
due to efficiency would have been negated by a contraction of
the gradient.
Results and Discussion
Experiment 1
The fully porous and Kinetex®
50 x 2.1 mm columns were initially
run at 0.8 mL per minute. The backpressures for both columns
were recorded and adjusted gradient peak capacity values were
then calculated.
Figures 1 and 2 illustrate that both the fully porous and Kinetex
50 x 2.1 mm columns gave comparable peak capacity figures (see
Table 3). The small change was due to the difference in adjusted
retention window (1.285 min for the fully porous column and 1.187
min for Kinetex). In fact, the average peak widths were slightly
narrower for the Kinetex (0.036 min) compared to the fully porous
column (0.037 min). As expected there was a large disparity in
backpressure, 7700 psi for the fully porous column vs 4600 psi
for Kinetex. This change in backpressure was exploited where the
flow rate on the Kinetex column was increased to 1.4 mL/min to
match the backpressure of the 1.7 µm fully porous column. Figure
3 and Table 3 shows that the resulting change in flow rate caused
the average peak width to reduce by ~30 % and the peak capacity
to increase by 26 %. The increase in performance was attributed
to the fact that the number of column volumes had increased per
percent of organic thus had volumetrically lowered the gradient.
When the column length of the Kinetex was increased to 75 mm
and the flow rate was altered to 1.1 mL per minute such that the
backpressure matched the 50 x 2.1 mm fully porous column at 0.8
mL per minute, a performance gain was again seen (see figure 4,
table 3). In fact, the increase in peak capacity (n = 42.3) almost
matched the peak capacity of the Kinetex 50 x 2.1 mm column
run at 1.4 mL per minute (n = 43.5). However, the reason behind
this gain was not due to a volumetric lowering of the gradient, but
due to the increased efficiency manifested in the longer column
length. The volumetric gradient remained almost identical to the
50 x 2.1 mm column run at 0.8 mL/min.
To see if increasing the length to 100 mm provided any further
benefits, a Kinetex 2.6 µm, 100 x 2.1 mm was then run at 0.85 mL
per minute to match the backpressure of the 50 x 2.1 mm, 1.7 µm
Table 2.
Mass spectrometer (MS) source conditions
Mass Spectrometric Source Conditions:
Parameter Initial Opt
0.8 mL/min
Opt
1.4 mL/min
CAD (PSI) 7 4 7
Curtain gas (PSI) 30 30 30
Gas1 (PSI) 50 40 50
Gas2 (PSI) 40 50 50
IS (eV) 5500 2000 2000
TEM (°C) 650 750 750
Opt = Optimized flow rate
Figure 1.
LC/MS chromatogram of the fully porous 1.7 µm, C18, 50 x 2.1 mm at
0.8 mL/min
14
5
4
6
8
7
9
10
11
12
2
1
13
3
15
1.4e6
19861
1.3e6
1.2e6
1.1e6
1.0e6
9.0e5
8.0e5
7.0e5
6.0e5
5.0e5
4.0e5
3.0e5
2.0e5
1.0e5
0.0
0.2 0.4 0.6 0.8 1.0 1.2 1.4
Intensity,cps
1.6 1.8 2.0 2.2 min
AppID19861
7700 psi (531 bar)
Peak Capacity (n) = 34.3
Figure 2.
LC/MS chromatogram of Kinetex 2.6 µm C18, 50 x 2.1 mm at
0.8 mL/min
14
5
4
6
8
7
9
10 11
12
2
1
13
3
15
7.0e5
19911
6.5e5
6.0e5
5.5e5
5.0e5
4.5e5
4.0e5
3.5e5
3.0e5
2.5e5
2.0e5
1.5e5
1.0e5
5.0e4
0.0
0.2 0.4 0.6 0.8 1.0 1.2 1.4
Intensity,cps
1.6 1.8 2.0 2.2 min
9.9e5
9.5e5
9.0e5
8.5e5
8.0e5
7.5e5
0,0
0,2 0,4 0,6 0,8 1,0 1,2 1,4 1,6 1,8 2,0 2,2 min
AppID19911
4600 psi (317 bar)
Peak Capacity (n) = 32.9
APPLICATIONS
TN-1104
Page 4 of 8 For additional technical notes, visit www.phenomenex.com
From the above results it was concluded that Kinetex 2.6 µm core-
shell columns run at 1.4 mL per minute was optimal in terms of
peak width and peak capacity. Nevertheless, it was noticed that
although column performance had improved, peak sensitivities
had dropped. It was hypothesized that the reason for this drop in
sensitivity was due to the fact that the MS had not been optimized
for the change in flow rate and hence source modifications would
be required to rectify this. Moreover it was also concluded that an
increase in the mass dwell time per MRM may also improve peak
sensitivity. Work was then done to optimize the MS and acquire
masses in small groups to allow for an increase in the mass dwell
time.
Experiment 2
To observe if peak sensitivity is recovered when running under
more optimal MS conditions for the faster flow rate, it was decided
to run just the basic analytes in the mix (9) as these provided the
best peak sensitivity and peak shape in Experiment 1 (Figure 6,
7, and 8). The results (see Table 4) indicated again a big increase
in peak capacity (52 %), however, unexpectedly peak heights in-
creased too, ranging from a 7 % increase to around 3.5 times
increase (Figure 9). It was concluded that the MS optimization
allowed for more effective desolvation thus allowing more ions
into the MS source. Furthermore, due to running small groups, the
MS was able to detect the peak apexes more effectively under the
optimized conditions.
Figure 5.
LC/MS chromatogram of Kinetex 2.6 µm C18, 100 x 2.1 mm at
0.85 mL/min
min
19940
1
2
8
6
5
4
11
10
9
3
12
1315
14
2 4 6 8 10 12 14 16 18 20 22 24min
7
Intensity,cps
1.08e6
1.05e6
1.00e6
9.50e5
9.00e5
8.50e5
8.00e5
7.50e5
7.00e5
6.50e5
6.00e5
5.50e5
5.00e5
4.50e5
4.00e5
3.50e5
3.00e5
2.50e5
2.00e5
1.50e5
1.00e5
5.00e4
0.0
AppID19940
7700 psi (531 bar)
Peak Capacity (n) = 34.6
Table 3.
Performance comparison summary of Kinetex 2.6 µm vs. 1.7 µm fully
porous columns for Experiment 1
Sample Name PWave
(min)
dRT
(min)
Peak
Capacity
(n)
Fully Porous C18 1.7 µm, 50 x 2.1 mm
@ 0.8 mL/min
0.037 1.29 34.4
Kinetex C18 2.6 µm, 50 x 2.1 mm
@ 0.8 mL/min
0.036 1.19 32.9
Kinetex C18 2.6 µm, 50 x 2.1 mm
@ 1.4 mL/min
0.025 1.08 43.6
Kinetex C18 2.6 µm, 75 x 2.1 mm
@ 1.1 mL/min
0.026 1.12 42.4
Kinetex C18 2.6 µm, 100 x 2.1 mm
@ 0.85 mL/min
0.034 1.17 34.6
Figure 4.
LC/MS chromatogram of Kinetex 2.6 µm C18, 75 x 2.1 mm at 1.1 mL/min
1
2
8
6
5
4
11
10
19939
9
3
12
1315
14
2 4 6 8 10 12 14 16 18 20 22 24min
7
Intensity,cps
1.10e6
1.05e6
1.00e6
9.50e5
9.00e5
8.50e5
8.00e5
7.50e5
7.00e5
6.50e5
6.00e5
5.50e5
5.00e5
4.50e5
4.00e5
3.50e5
3.00e5
2.50e5
2.00e5
1.50e5
1.00e5
5.00e4
0.0
AppID19939
7700 psi (531 bar)
Peak Capacity (n) = 42.3
Figure 3.
LC/MS chromatogram of Kinetex®
2.6 µm C18, 50 x 2.1 mm at
1.4 mL/min
14
5
4
6
8
7
910
11
12
2
1
13
3
15
1.4e6
19862
1.3e6
1.2e6
1.1e6
1.0e6
9.0e5
8.0e5
7.0e5
6.0e5
5.0e5
4.0e5
3.0e5
2.0e5
1.0e5
0.0
0.2 0.4 0.6 0.8 1.0 1.2 1.4
Intensity,cps
1.6 1.8 2.0 2.2 min
AppID19862
7700 psi (531 bar)
Peak Capacity (n) = 43.5
APPLICATIONS
TN-1104
Page 5 of 8For additional technical notes, visit www.phenomenex.com
Figure 6.
Group 1 analytes on Kinetex®
C18 2.6 µm, 50 x 2.1 mm at 0.8 and 1.4 mL/min respectively under optimal MS conditions for each flow rate.
19941
1.00e6
5.00e5
0.5
3
Intensity,cps
0.0
1.0 1.5 2.0 2.5 min
1
6.0e5
4.0e5
2.0e5
0.5
Intensity,cps
0.0
1.0 1.5 2.0 2.5 min
0.5
2
Intensity,cps
1.00e5
5.00e4
0.0
1.0 1.5 2.5 min
19945
1.21e5
Intensity,cps
1.00e5
5.00e4
0.0
0.5
2
1.0 1.5 2.0 min
1.4e5
Intensity,cps
1.0e5
5.0e4
0.0
0.5
3
1.0 1.5 2.0 min
3.0e5
Intensity,cps
2.0e5
1.0e5
0.0
0.5
1
1.0 1.5 2.0 min
19946
4.0e5
Intensity,cps
2.0e5
0.0
0.5
1
1.0 1.5 2.0 min
1.9e4
Intensity,cps
1.5e4
1.0e4
5000.00
0.0
0.5
2
1.0 1.5 2.0 min
1.22e6
Intensity,cps
1.00e6
5.00e5
0.0
0.5
3
1.0 1.5 2.0 min
19942
4.0e5
Intensity,cps
2.0e5
0.0
0.5
2
1.0 1.5 2.0 min
5.2e5
Intensity,cps
4.0e5
2.0e5
0.0
0.5
3
1.0 1.5 2.0 min
4.7e5
Intensity,cps
4.0e5
2.0e5
0.0
0.5
1
1.0 1.5 2.0 min
19943
2.8e4
Intensity,cps
2.0e4
1.0e4
0.0
0.5
2
1.0 1.5 2.0 min
1.4e6
Intensity,cps
1.0e6
5.0e6
0.0
0.5
3
1.0 1.5 2.0 min
4.8e5
Intensity,cps
4.0e5
2.0e5
0.0
0.5
1
1.0 1.5 2.0 min
0.5
19944
2
Intensity,cps
9.0e4
5.0e4
0.0
1.0 1.5 2.0 2.5 min
4.0e5
0.5
1
Intensity,cps
2.0e5
0.0
1.0 1.5 2.0 2.5 min
0.5
3
Intensity,cps
9.0e5
5.0e5
0.0
1.0 1.5 2.0 2.5 min
AppID19944
AppID19941
Figure 7.
Group 2 analytes on Kinetex C18 2.6 µm, 50 x 2.1 mm at 0.8 and 1.4 mL/min respectively under optimal MS conditions for each flow rate.
AppID19942
AppID19945
Figure 8.
Group 3 analytes on Kinetex C18 2.6 µm, 50 x 2.1 mm at 0.8 and 1.4 mL/min respectively under optimal MS conditions for each flow rate.
AppID19943
AppID19946
1.4 mL per min 0.8 mL per min
0.8 mL per min 1.4 mL per min
0.8 mL per min
0.8 mL per min 1.4 mL per min
1.4 mL per min
APPLICATIONS
TN-1104
Page 6 of 8 For additional technical notes, visit www.phenomenex.com
It was concluded that the MS optimization allowed for more
effective desolvation and this helped to improve peak sensitivities.
Furthermore, as the peak widths were so narrow the MS was able
to detect the peak apexes more effectively under the optimized
conditions. Thus it was decided to compare the fully porous 1.7µm
column with Kinetex 2.6 µm under optimized MS conditions.
Experiment 3
Kinetex®
C18 2.6 µm, 50 x 2.1 mm columns were compared with
fully porous 1.7 µm columns at 0.8 mL/min under optimal mass
spec conditions for that flow rate. It was found that the Kinetex
columns gave a slightly, but not significantly, higher peak capac-
ity. However, when the Kinetex columns were run at 1.4 mL per
minute significant increases in peak capacity (around 70 %) were
observed (see Figure 11 and Table 5). Finally, when the gradi-
ent rate was increased such that the number of column volumes
matched the run at 0.8 mL/min, the consequence was to reduce
the run time to 1.5 minutes. This resulted in reducing the peak
capacity to the same level as the fully porous column run at 0.8
mL/min over the original 1.5 minutes. Furthermore, as shown in
Figure 10, sensitivity improvements were again observed when
the Kinetex column was run at the faster 1.4 mL per minute flow
rate.
Figure 10.
Sensitivity results for Experiment 3
Figure 11.
Peak Capacity Data for Experiment 3
Comparing average peak capacity (nave
) on a Kinetex C18 2.6 µm 50 x 2.1 mm column vs. a fully
porous C18 1.7 µm 50 x 2.1 mm column under various flow and gradient conditions
1.7 µm FP column
0.8 mL/min over 2.5 min
Kinetex 2.6 µm
0.8 mL/min over 2.5 min
Kinetex 2.6 µm
1.4 mL/min over 2.5 min
Kinetex 2.6 µm
1.4 mL/min over 1.5 min
60
50
40
30
20
10
0
Condition
Table 5.
Summary of peak capacity data for Experiment 3
Conditions
PWave
(min)
dRT
(min)
Peak
Capacity
(n)
Fully Porous C18 1.7 µm, 50 x 2.1 mm
@ 0.8 mL/min (2.5 min)
0.029 0.87 30.1
Kinetex C18 2.6 µm, 50 x 2.1 mm
@ 0.8 mL/min (2.5 min)
0.027 0.88 32.5
Kinetex C18 2.6 µm, 50 x 2.1 mm
@ 1.4 mL/min (2.5 min)
0.017 0.89 51.6
Kinetex C18 2.6 µm, 50 x 2.1 mm
@ 1.4 mL/min (1.5 min)
0.017 0.50 30.0
Figure 9.
Peak height data for all analytes tested under optimized MS conditions for
each flow rate. (The 0.8 mL per minute data was normalized to 1.)
1.00
1.50
2.00
2.50
3.00
3.50
4.00
AcebutololCimetidine Pindolol Sulfathiazole Tripolidine Diltiazem TerfenadineAcetaminophen Chlorpheniramine
Analyte
Changeinsensitivitywherevaluestestedat0.8mL/minis
Change in sensitivity from 0.8 to 1.4 mL/min
Group 1
Group 2
Group 3
normalisedto1
Table 4.
Performance comparison summary of Kinetex run at 0.8 and 1.4 mL per
minute respectively
Conditions
PWave
(min)
dRT
(min)
Peak
Capacity
(n)
Kinetex C18 2.6 µm, 50 x 2.1 mm
@ 0.8 mL/min
0.024 0.88 36.0
Kinetex C18 2.6 µm, 50 x 2.1 mm
@ 1.4 mL/min
0.016 0.89 54.9
0
200000
400000
600000
800000
1000000
1200000
1400000
Comparing peak heights on a Kinetex C18 2.6 µm 50 x 2.1 mm column vs. a Fully Porous
C18 1.7 µm 50 x 2.1 column under various flow and gradient conditions
Counts(m/z)
Analyte
Cimetidine
Pindolol
Sulfathiazole
Tripolidine
Diltiazem
Terfenadine
Acebutolol
Haloperidol
Quinidine
Nortriptyline
1.7 µm FP column 0.8 mL/min
over 2.5 min
Kinetex 2.6 µm 0.8 mL/min
over 2.5 min
Kinetex 2.6 µm 1.4 mL/min
over 2.5 min
Kinetex 2.6 µm 1.4 mL/min
per min over 1.5 min
Chlorpheniramine
Acetaminophen
APPLICATIONS
TN-1104
Page 7 of 8For additional technical notes, visit www.phenomenex.com
Conclusion
Kinetex columns packed with 2.6 µm core-shell particles offer big
advantages over fully porous 1.7 µm columns. The lower back-
pressure can be utilized to run columns at flow rates which would
normally be impossible to reach on UHPLC instruments due to
instrument and material pressure limits. The observed reduced
resistance to flow of Kinetex compared to 1.7 µm columns, but
comparable efficiency can be used to gain much higher peak ca-
pacities when run under fast gradient conditions. Moreover, by
steepening the gradient conditions with respect to the change in
flow rate, then run times can be reduced without loss of perfor-
mance. When MS conditions are optimized to the change in flow
rate, then significant sensitivity improvements are also observed.
*	 SecurityGuard Ultra cartridges require holder, Part No. AJ0-9000
When using Kinetex 1.7 µm, increased performance can be achieved, however higher pressure-capable
instrumentation is required.
**	Columns are pH stable from 1.5-10 under isocratic conditions. Columns are pH stable 1.5-8.5 under
gradient conditions.
	
2.1 mm ID Kinetex columns are pressure stable up to 1000 bar.
2.6 µm Material Characteristics
Packing
Material
Total
Particle
Size
(µm)
Porous
Shell
(µm)
Solid
Core
(µm)
Pore
Size
(Å)
Effective
Surface
Area
(m2
/g)
Effective
Carbon
Load
%
pH
Stability
Pressure
Stability
Kinetex XB-C18 2.6 0.35 1.9 100 200 10 1.5 - 8.5**
1000/600
bar
Kinetex C18 2.6 0.35 1.9 100 200 12 1.5 - 8.5**
Kinetex C8 2.6 0.35 1.9 100 200 8 1.5 - 8.5**
Kinetex PFP 2.6 0.35 1.9 100 200 9 1.5 - 8.5**
Kinetex HILIC 2.6 0.35 1.9 100 200 0 2.0 - 7.5
Kinetex
Phenyl-Hexyl
2.6 0.35 1.9 100 200 11 1.5 - 8.5**
1.7 µm Material Characteristics
Packing
Material
Total
Particle
Size
(µm)
Porous
Shell
(µm)
Sol-
id
Core
(µm)
Pore
Size
(Å)
Effective
Surface
Area
(m2
/g)
Effective
Carbon
Load
%
pH
Stability
Pressure
Stability
Kinetex XB-C18 1.7 0.23 1.25 100 200 10 1.5 - 8.5**
1000 bar
Kinetex C18 1.7 0.23 1.25 100 200 12 1.5 - 8.5**
Kinetex C8 1.7 0.23 1.25 100 200 8 1.5 - 8.5**
Kinetex PFP 1.7 0.23 1.25 100 200 9 1.5 - 8.5**
Kinetex HILIC 1.7 0.23 1.25 100 200 0 2.0 - 7.5
Kinetex
Phenyl-Hexyl
1.7 0.23 1.25 100 200 11 1.5 - 8.5**
2.6 μm MidBore™
Columns (mm)
SecurityGuard
Ultra Cartridges
30 x 3.0 50 x 3.0 75 x 3.0 100 x 3.0 150 x 3.0 3/pk
XB-C18 00A-4496-Y0 00B-4496-Y0 00C-4496-Y0 00D-4496-Y0 00F-4496-Y0 AJ0-8775
C18 00A-4462-Y0 00B-4462-Y0 00C-4462-Y0 00D-4462-Y0 00F-4462-Y0 AJ0-8775
C8 00A-4497-Y0 00B-4497-Y0 00C-4497-Y0 00D-4497-Y0 00F-4497-Y0 AJ0-8777
PFP 00A-4477-Y0 00B-4477-Y0 00C-4477-Y0 00D-4477-Y0 00F-4477-Y0 AJ0-8780
HILIC 00A-4461-Y0 –– –– –– 00F-4461-Y0 AJ0-8779
Phenyl-Hexyl –– –– –– –– –– AJ0-8781
for 3.0 mm ID
Kinetex®
Ordering Information
2.6 μm Analytical Columns (mm)
SecurityGuard
Ultra Cartridges
30 x 4.6 50 x 4.6 75 x 4.6 100 x 4.6 150 x 4.6 3/pk
XB-C18 –– 00B-4496-E0 00C-4496-E0 00D-4496-E0 00F-4496-E0 AJ0-8768
C18 00A-4462-E0 00B-4462-E0 00C-4462-E0 00D-4462-E0 00F-4462-E0 AJ0-8768
C8 –– 00B-4497-E0 00C-4497-E0 00D-4497-E0 00F-4497-E0 AJ0-8770
PFP 00A-4477-E0 00B-4477-E0 00C-4477-E0 00D-4477-E0 00F-4477-E0 AJ0-8773
HILIC 00A-4461-E0 00B-4461-E0 00C-4461-E0 00D-4461-E0 00F-4461-E0 AJ0-8772
Phenyl-Hexyl –– 00B-4495-E0 –– 00D-4495-E0 00F-4495-E0 AJ0-8774
for 4.6 mm ID
1.7 μm Minibore Columns (mm)
SecurityGuard
Ultra Cartridges*
30 x 2.1 50 x 2.1 100 x 2.1 150 x 2.1 3/pk
XB-C18 00A-4498-AN 00B-4498-AN 00D-4498-AN 00F-4498-AN AJ0-8782
C18 00A-4475-AN 00B-4475-AN 00D-4475-AN 00F-4475-AN AJ0-8782
C8 00A-4499-AN 00B-4499-AN 00D-4499-AN 00F-4499-AN AJ0-8784
PFP 00A-4476-AN 00B-4476-AN 00D-4476-AN 00F-4476-AN AJ0-8787
HILIC 00A-4474-AN 00B-4474-AN 00D-4474-AN –– AJ0-8786
Phenyl-Hexyl –– 00B-4500-AN 00D-4500-AN 00F-4500-AN AJ0-8788
for 2.1 mm ID
1.7 μm MidBore Columns (mm)
SecurityGuard
Ultra Cartridges
30 x 3.0 50 x 3.0 100 x 3.0 3/pk
XB-C18 00A-4498-Y0 00B-4498-Y0 00D-4498-Y0 AJ0-8775
C18 –– 00B-4475-Y0 00D-4475-Y0 AJ0-8775
C8 00A-4499-Y0 00B-4499-Y0 00D-4499-Y0 AJ0-8777
PFP –– –– 00D-4476-Y0 AJ0-8780
HILIC –– 00B-4474-Y0 –– AJ0-8779
Phenyl-Hexyl –– –– –– AJ0-8781
for 3.0 mm ID
2.6 μm Minibore Columns (mm)
SecurityGuard
Ultra Cartridges
30 x 2.1 50 x 2.1 75x 2.1 100 x 2.1 150 x 2.1 3/pk
XB-C18 00A-4496-AN 00B-4496-AN 00C-4496-AN 00D-4496-AN 00F-4496-AN AJ0-8782
C18 00A-4462-AN 00B-4462-AN 00C-4462-AN 00D-4462-AN 00F-4462-AN AJ0-8782
C8 00A-4497-AN 00B-4497-AN 00C-4497-AN 00D-4497-AN 00F-4497-AN AJ0-8784
PFP 00A-4477-AN 00B-4477-AN 00C-4477-AN 00D-4477-AN 00F-4477-AN AJ0-8787
HILIC 00A-4461-AN 00B-4461-AN 00C-4461-AN 00D-4461-AN 00F-4461-AN AJ0-8786
Phenyl-Hexyl –– 00B-4495-AN –– 00D-4495-AN –– AJ0-8788
for 2.1 mm ID
APPLICATIONS
TN-1104
Page 8 of 8
Terms and Conditions
Subject to Phenomenex Standard Terms & Conditions, which may be viewed at
www.phenomenex.com/TermsAndConditions
Trademarks
Kinetex is a registered trademark of Phenomenex, Inc. in the United States, European Union,
and other jurisdictions. MidBore is a registered trademark of Phenomenex, Inc. Waters, AC-
QUITY, and UPLC are registered trademarks of Waters Corporation. API 5000 is a trademark
of AB Sciex Pte. Ltd. AB SCIEX™
is being used under license.
Disclaimers
Comparative separations may not be representative of all applications. Phenomenex is not
affiliated with Waters Corporation.
© 2011 Phenomenex, Inc. All rights reserved.
TN89190811_L
t:
f:
Australia
02-9428-6444
02-9428-6445
auinfo@phenomenex.com
t:
f:
Austria
01-319-1301
01-319-1300
anfrage@phenomenex.com
t:
f:
Belgium
+31 (0)30-2418700
+31 (0)30-2383749
beinfo@phenomenex.com
t:
f:
Canada
(800) 543-3681
(310) 328-7768
info@phenomenex.com
t:
f:
Denmark
4824 8048
4810 6265
nordicinfo@phenomenex.com
t:
f:
Finland
+358 (0)9 4789 0063
+45 4810 6265
nordicinfo@phenomenex.com
t:
f:
France
01 30 09 21 10
01 30 09 21 11
franceinfo@phenomenex.com
t:
f:
Germany
06021-58830-0
06021-58830-11
anfrage@phenomenex.com
t:
f:
India
040-3012 2400
040-3012 2411
indiainfo@phenomenex.com
t:
f:
Ireland
01 247 5405
+44 1625-501796
eireinfo@phenomenex.com
t:
f:
Italy
051 6327511
051 6327555
italiainfo@phenomenex.com
t:
f:
Luxembourg
+31 (0)30-2418700
+31 (0)30-2383749
nlinfo@phenomenex.com
t:
f:
Mexico
001-800-844-5226
001-310-328-7768
tecnicomx@phenomenex.com
t:
f:
The Netherlands
030-2418700
030-2383749
nlinfo@phenomenex.com
t:
f:
New Zealand
09-4780951
09-4780952
nzinfo@phenomenex.com
t:
f:
Norway
+47 810 02 005
+45 4810 6265
nordicinfo@phenomenex.com
t:
f:
Puerto Rico
(800) 541-HPLC
(310) 328-7768
info@phenomenex.com
t:
f:
United Kingdom
01625-501367
01625-501796
ukinfo@phenomenex.com
t:
f:
All other countries:
Corporate Office USA
(310) 212-0555
(310) 328-7768
info@phenomenex.com
www.phenomenex.com
Phenomenex products are available worldwide. For the distributor in your country,
contact Phenomenex USA, International Department at international@phenomenex.com
If Kinetex analytical columns do not provide at least an equivalent
separation as compared to a competing column of the same phase
and dimensions, return the column with comparative data within 45
days for a FULL REFUND.

More Related Content

What's hot

Development and validation of RP – HPLC method for the estimation of Tylosin ...
Development and validation of RP – HPLC method for the estimation of Tylosin ...Development and validation of RP – HPLC method for the estimation of Tylosin ...
Development and validation of RP – HPLC method for the estimation of Tylosin ...pharmaindexing
 
A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...
A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...
A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...Alok Singh
 
Cleaning Validation for Pharmaceutical Manufacturing Using Online SFE/SFC
Cleaning Validation for Pharmaceutical Manufacturing Using Online SFE/SFCCleaning Validation for Pharmaceutical Manufacturing Using Online SFE/SFC
Cleaning Validation for Pharmaceutical Manufacturing Using Online SFE/SFCShimadzu Scientific Instruments
 
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...pharmaindexing
 
Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...
Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...
Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...SriramNagarajan15
 
Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...
Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...
Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...pharmaindexing
 
PNAS-2016-Gielen-1606927113
PNAS-2016-Gielen-1606927113PNAS-2016-Gielen-1606927113
PNAS-2016-Gielen-1606927113Raphaelle Hours
 
Multiple Method Development and Validation for Simultaneous Estimation of Chl...
Multiple Method Development and Validation for Simultaneous Estimation of Chl...Multiple Method Development and Validation for Simultaneous Estimation of Chl...
Multiple Method Development and Validation for Simultaneous Estimation of Chl...ijtsrd
 
A Systemic Study of Chitin Digestion Protocols
A Systemic Study of Chitin Digestion ProtocolsA Systemic Study of Chitin Digestion Protocols
A Systemic Study of Chitin Digestion ProtocolsArjuna Karikaran
 
Analytical method development and validation for the estimation of Nimesulide...
Analytical method development and validation for the estimation of Nimesulide...Analytical method development and validation for the estimation of Nimesulide...
Analytical method development and validation for the estimation of Nimesulide...SriramNagarajan18
 
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water MatricesRapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water MatricesShimadzu Scientific Instruments
 
Analysis of Micronutrients in Fortified Breakfast Cereal by Flame Atomic Abso...
Analysis of Micronutrients in Fortified Breakfast Cereal by Flame Atomic Abso...Analysis of Micronutrients in Fortified Breakfast Cereal by Flame Atomic Abso...
Analysis of Micronutrients in Fortified Breakfast Cereal by Flame Atomic Abso...PerkinElmer, Inc.
 
Spectrophotometric Estimation of Rosuvastatin Calcium in Bulk and Pharmaceuti...
Spectrophotometric Estimation of Rosuvastatin Calcium in Bulk and Pharmaceuti...Spectrophotometric Estimation of Rosuvastatin Calcium in Bulk and Pharmaceuti...
Spectrophotometric Estimation of Rosuvastatin Calcium in Bulk and Pharmaceuti...pharmaindexing
 
Traditional Kashmiri Recipe “Shangri-Kahwa” as a Stimulant Drink and Effectiv...
Traditional Kashmiri Recipe “Shangri-Kahwa” as a Stimulant Drink and Effectiv...Traditional Kashmiri Recipe “Shangri-Kahwa” as a Stimulant Drink and Effectiv...
Traditional Kashmiri Recipe “Shangri-Kahwa” as a Stimulant Drink and Effectiv...SriramNagarajan15
 
Development and validation of hplc method for determination of theophylline a...
Development and validation of hplc method for determination of theophylline a...Development and validation of hplc method for determination of theophylline a...
Development and validation of hplc method for determination of theophylline a...IJSIT Editor
 
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...Shimadzu Scientific Instruments
 

What's hot (19)

Development and validation of RP – HPLC method for the estimation of Tylosin ...
Development and validation of RP – HPLC method for the estimation of Tylosin ...Development and validation of RP – HPLC method for the estimation of Tylosin ...
Development and validation of RP – HPLC method for the estimation of Tylosin ...
 
A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...
A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...
A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...
 
Cleaning Validation for Pharmaceutical Manufacturing Using Online SFE/SFC
Cleaning Validation for Pharmaceutical Manufacturing Using Online SFE/SFCCleaning Validation for Pharmaceutical Manufacturing Using Online SFE/SFC
Cleaning Validation for Pharmaceutical Manufacturing Using Online SFE/SFC
 
Ultra-High Speed Analysis of USP Methods
Ultra-High Speed Analysis of USP MethodsUltra-High Speed Analysis of USP Methods
Ultra-High Speed Analysis of USP Methods
 
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
 
Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...
Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...
Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...
 
Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...
Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...
Method Development and Validation of Naftopidil by Reverse Phase-HPLC in Bulk...
 
PNAS-2016-Gielen-1606927113
PNAS-2016-Gielen-1606927113PNAS-2016-Gielen-1606927113
PNAS-2016-Gielen-1606927113
 
Multiple Method Development and Validation for Simultaneous Estimation of Chl...
Multiple Method Development and Validation for Simultaneous Estimation of Chl...Multiple Method Development and Validation for Simultaneous Estimation of Chl...
Multiple Method Development and Validation for Simultaneous Estimation of Chl...
 
A Systemic Study of Chitin Digestion Protocols
A Systemic Study of Chitin Digestion ProtocolsA Systemic Study of Chitin Digestion Protocols
A Systemic Study of Chitin Digestion Protocols
 
Analytical method development and validation for the estimation of Nimesulide...
Analytical method development and validation for the estimation of Nimesulide...Analytical method development and validation for the estimation of Nimesulide...
Analytical method development and validation for the estimation of Nimesulide...
 
High Sensitivity HPLC Profiling of 2-AB Glycans
High Sensitivity HPLC Profiling of 2-AB GlycansHigh Sensitivity HPLC Profiling of 2-AB Glycans
High Sensitivity HPLC Profiling of 2-AB Glycans
 
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water MatricesRapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
 
Analysis of Micronutrients in Fortified Breakfast Cereal by Flame Atomic Abso...
Analysis of Micronutrients in Fortified Breakfast Cereal by Flame Atomic Abso...Analysis of Micronutrients in Fortified Breakfast Cereal by Flame Atomic Abso...
Analysis of Micronutrients in Fortified Breakfast Cereal by Flame Atomic Abso...
 
Spectrophotometric Estimation of Rosuvastatin Calcium in Bulk and Pharmaceuti...
Spectrophotometric Estimation of Rosuvastatin Calcium in Bulk and Pharmaceuti...Spectrophotometric Estimation of Rosuvastatin Calcium in Bulk and Pharmaceuti...
Spectrophotometric Estimation of Rosuvastatin Calcium in Bulk and Pharmaceuti...
 
Traditional Kashmiri Recipe “Shangri-Kahwa” as a Stimulant Drink and Effectiv...
Traditional Kashmiri Recipe “Shangri-Kahwa” as a Stimulant Drink and Effectiv...Traditional Kashmiri Recipe “Shangri-Kahwa” as a Stimulant Drink and Effectiv...
Traditional Kashmiri Recipe “Shangri-Kahwa” as a Stimulant Drink and Effectiv...
 
Development and validation of hplc method for determination of theophylline a...
Development and validation of hplc method for determination of theophylline a...Development and validation of hplc method for determination of theophylline a...
Development and validation of hplc method for determination of theophylline a...
 
Increasing the Throughput of UHPLC
Increasing the Throughput of UHPLCIncreasing the Throughput of UHPLC
Increasing the Throughput of UHPLC
 
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
 

Viewers also liked

Do workers work enough? International comparison of working hours and wage di...
Do workers work enough? International comparison of working hours and wage di...Do workers work enough? International comparison of working hours and wage di...
Do workers work enough? International comparison of working hours and wage di...Malte Nyfos Mathiasen
 
Young courtney presentation
Young courtney presentationYoung courtney presentation
Young courtney presentationCourtney Young
 
How Neighborhoods affect Choice and Well-being
How Neighborhoods affect Choice and Well-beingHow Neighborhoods affect Choice and Well-being
How Neighborhoods affect Choice and Well-beingMalte Nyfos Mathiasen
 
Research Project Prospectus - The 60s Generation’s Life Course of Political P...
Research Project Prospectus - The 60s Generation’s Life Course of Political P...Research Project Prospectus - The 60s Generation’s Life Course of Political P...
Research Project Prospectus - The 60s Generation’s Life Course of Political P...Malte Nyfos Mathiasen
 
Bachelor Thesis M.S. Jesus Restricted Version [Only available in Dutch]
Bachelor Thesis M.S. Jesus Restricted Version [Only available in Dutch]Bachelor Thesis M.S. Jesus Restricted Version [Only available in Dutch]
Bachelor Thesis M.S. Jesus Restricted Version [Only available in Dutch]Marson Jesus
 
Co-opting labor-based Parties and Unions
Co-opting labor-based Parties and UnionsCo-opting labor-based Parties and Unions
Co-opting labor-based Parties and UnionsMalte Nyfos Mathiasen
 
USP 621 Allowable Adjustment to Chromatography HPLC Methods
USP 621 Allowable Adjustment to Chromatography HPLC MethodsUSP 621 Allowable Adjustment to Chromatography HPLC Methods
USP 621 Allowable Adjustment to Chromatography HPLC MethodsSandy Simmons
 
Torres de hanoi
Torres de hanoiTorres de hanoi
Torres de hanoiSMCangry
 
Column care for your Aeris Widepore HPLC
Column care for your Aeris Widepore HPLCColumn care for your Aeris Widepore HPLC
Column care for your Aeris Widepore HPLCSandy Simmons
 
Método de lagrange
Método de lagrangeMétodo de lagrange
Método de lagrangeSMCangry
 
ShannonBoone_Interactive Prototype
ShannonBoone_Interactive PrototypeShannonBoone_Interactive Prototype
ShannonBoone_Interactive PrototypeShannon Boone
 
Angular 2 Essential Training
Angular 2 Essential Training Angular 2 Essential Training
Angular 2 Essential Training Patrick Schroeder
 

Viewers also liked (13)

Do workers work enough? International comparison of working hours and wage di...
Do workers work enough? International comparison of working hours and wage di...Do workers work enough? International comparison of working hours and wage di...
Do workers work enough? International comparison of working hours and wage di...
 
Young courtney presentation
Young courtney presentationYoung courtney presentation
Young courtney presentation
 
How Neighborhoods affect Choice and Well-being
How Neighborhoods affect Choice and Well-beingHow Neighborhoods affect Choice and Well-being
How Neighborhoods affect Choice and Well-being
 
Research Project Prospectus - The 60s Generation’s Life Course of Political P...
Research Project Prospectus - The 60s Generation’s Life Course of Political P...Research Project Prospectus - The 60s Generation’s Life Course of Political P...
Research Project Prospectus - The 60s Generation’s Life Course of Political P...
 
Bachelor Thesis M.S. Jesus Restricted Version [Only available in Dutch]
Bachelor Thesis M.S. Jesus Restricted Version [Only available in Dutch]Bachelor Thesis M.S. Jesus Restricted Version [Only available in Dutch]
Bachelor Thesis M.S. Jesus Restricted Version [Only available in Dutch]
 
Co-opting labor-based Parties and Unions
Co-opting labor-based Parties and UnionsCo-opting labor-based Parties and Unions
Co-opting labor-based Parties and Unions
 
USP 621 Allowable Adjustment to Chromatography HPLC Methods
USP 621 Allowable Adjustment to Chromatography HPLC MethodsUSP 621 Allowable Adjustment to Chromatography HPLC Methods
USP 621 Allowable Adjustment to Chromatography HPLC Methods
 
Torres de hanoi
Torres de hanoiTorres de hanoi
Torres de hanoi
 
Column care for your Aeris Widepore HPLC
Column care for your Aeris Widepore HPLCColumn care for your Aeris Widepore HPLC
Column care for your Aeris Widepore HPLC
 
Método de lagrange
Método de lagrangeMétodo de lagrange
Método de lagrange
 
ShannonBoone_Interactive Prototype
ShannonBoone_Interactive PrototypeShannonBoone_Interactive Prototype
ShannonBoone_Interactive Prototype
 
boone_assignment 6
boone_assignment 6boone_assignment 6
boone_assignment 6
 
Angular 2 Essential Training
Angular 2 Essential Training Angular 2 Essential Training
Angular 2 Essential Training
 

Similar to Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analyses When Exploiting the Properties of 2.6 μm Kinetex® Core-Shell Columns

UPLC INTRODUCTION AND REVIEW
UPLC INTRODUCTION AND REVIEWUPLC INTRODUCTION AND REVIEW
UPLC INTRODUCTION AND REVIEWAnnette Loubriel
 
jaypatelppt2update2018FINAL.pptx
jaypatelppt2update2018FINAL.pptxjaypatelppt2update2018FINAL.pptx
jaypatelppt2update2018FINAL.pptxJayPatel895875
 
Bock Acs Paper Id1119039 08192007 Poster
Bock Acs Paper Id1119039 08192007 PosterBock Acs Paper Id1119039 08192007 Poster
Bock Acs Paper Id1119039 08192007 PosterMarybock
 
108 HPLC. Comparing Narrow Bore column with a standard Bore of 4.6 mm ID for ...
108 HPLC. Comparing Narrow Bore column with a standard Bore of 4.6 mm ID for ...108 HPLC. Comparing Narrow Bore column with a standard Bore of 4.6 mm ID for ...
108 HPLC. Comparing Narrow Bore column with a standard Bore of 4.6 mm ID for ...Hossein Hodjat
 
High performance thin layer chromatography
High performance thin layer chromatography High performance thin layer chromatography
High performance thin layer chromatography RevanHiwale1
 
Application of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
Application of a Dual Injection (U)HPLC for the Analysis of PharmaceuticalsApplication of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
Application of a Dual Injection (U)HPLC for the Analysis of PharmaceuticalsShimadzu Scientific Instruments
 
110 Separation of 5 peptides on Narrow Bore column of 2.1 mm ID. Acidic and ...
110  Separation of 5 peptides on Narrow Bore column of 2.1 mm ID. Acidic and ...110  Separation of 5 peptides on Narrow Bore column of 2.1 mm ID. Acidic and ...
110 Separation of 5 peptides on Narrow Bore column of 2.1 mm ID. Acidic and ...Hossein Hodjat
 
Filtration Strategies for Optimal Development and Purification of a FMD Virus...
Filtration Strategies for Optimal Development and Purification of a FMD Virus...Filtration Strategies for Optimal Development and Purification of a FMD Virus...
Filtration Strategies for Optimal Development and Purification of a FMD Virus...Merck Life Sciences
 
Comparison of Formulation Analysis by UPLC FINAL
Comparison of Formulation Analysis by UPLC FINALComparison of Formulation Analysis by UPLC FINAL
Comparison of Formulation Analysis by UPLC FINALJessica Sitko
 
Ultra perfoemence liquid chromatography bhagyashri ppt 23.pptx
Ultra perfoemence liquid chromatography bhagyashri ppt 23.pptxUltra perfoemence liquid chromatography bhagyashri ppt 23.pptx
Ultra perfoemence liquid chromatography bhagyashri ppt 23.pptxBhagyashriPatil73
 
Ultra performance liquid chromatography
Ultra performance liquid chromatographyUltra performance liquid chromatography
Ultra performance liquid chromatographybiniyapatel
 
A hands on guide to ultrafiltration1
A hands on guide to ultrafiltration1A hands on guide to ultrafiltration1
A hands on guide to ultrafiltration1Arash Abed
 
Automated SPE-LC/MS/MS method development using ITSP
Automated SPE-LC/MS/MS method development using ITSPAutomated SPE-LC/MS/MS method development using ITSP
Automated SPE-LC/MS/MS method development using ITSPMark Hayward
 
Method Development Poster-MicroLiter
Method Development Poster-MicroLiterMethod Development Poster-MicroLiter
Method Development Poster-MicroLiterRick Youngblood
 
Tips and tricks_hplc_troubleshooting
Tips and tricks_hplc_troubleshootingTips and tricks_hplc_troubleshooting
Tips and tricks_hplc_troubleshootingFlavio De Jorge
 
Ultra Performance Liquid Chromatography
Ultra Performance Liquid ChromatographyUltra Performance Liquid Chromatography
Ultra Performance Liquid ChromatographyMahewash Sana Pathan
 

Similar to Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analyses When Exploiting the Properties of 2.6 μm Kinetex® Core-Shell Columns (20)

Pesticide Residue Analysis Webinar Series: Tips and Tricks for the Whole Work...
Pesticide Residue Analysis Webinar Series: Tips and Tricks for the Whole Work...Pesticide Residue Analysis Webinar Series: Tips and Tricks for the Whole Work...
Pesticide Residue Analysis Webinar Series: Tips and Tricks for the Whole Work...
 
UPLC INTRODUCTION AND REVIEW
UPLC INTRODUCTION AND REVIEWUPLC INTRODUCTION AND REVIEW
UPLC INTRODUCTION AND REVIEW
 
jaypatelppt2update2018FINAL.pptx
jaypatelppt2update2018FINAL.pptxjaypatelppt2update2018FINAL.pptx
jaypatelppt2update2018FINAL.pptx
 
Bock Acs Paper Id1119039 08192007 Poster
Bock Acs Paper Id1119039 08192007 PosterBock Acs Paper Id1119039 08192007 Poster
Bock Acs Paper Id1119039 08192007 Poster
 
108 HPLC. Comparing Narrow Bore column with a standard Bore of 4.6 mm ID for ...
108 HPLC. Comparing Narrow Bore column with a standard Bore of 4.6 mm ID for ...108 HPLC. Comparing Narrow Bore column with a standard Bore of 4.6 mm ID for ...
108 HPLC. Comparing Narrow Bore column with a standard Bore of 4.6 mm ID for ...
 
Uplc
UplcUplc
Uplc
 
High performance thin layer chromatography
High performance thin layer chromatography High performance thin layer chromatography
High performance thin layer chromatography
 
Application of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
Application of a Dual Injection (U)HPLC for the Analysis of PharmaceuticalsApplication of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
Application of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
 
110 Separation of 5 peptides on Narrow Bore column of 2.1 mm ID. Acidic and ...
110  Separation of 5 peptides on Narrow Bore column of 2.1 mm ID. Acidic and ...110  Separation of 5 peptides on Narrow Bore column of 2.1 mm ID. Acidic and ...
110 Separation of 5 peptides on Narrow Bore column of 2.1 mm ID. Acidic and ...
 
Filtration Strategies for Optimal Development and Purification of a FMD Virus...
Filtration Strategies for Optimal Development and Purification of a FMD Virus...Filtration Strategies for Optimal Development and Purification of a FMD Virus...
Filtration Strategies for Optimal Development and Purification of a FMD Virus...
 
Comparison of Formulation Analysis by UPLC FINAL
Comparison of Formulation Analysis by UPLC FINALComparison of Formulation Analysis by UPLC FINAL
Comparison of Formulation Analysis by UPLC FINAL
 
Ultra perfoemence liquid chromatography bhagyashri ppt 23.pptx
Ultra perfoemence liquid chromatography bhagyashri ppt 23.pptxUltra perfoemence liquid chromatography bhagyashri ppt 23.pptx
Ultra perfoemence liquid chromatography bhagyashri ppt 23.pptx
 
Ultra performance liquid chromatography
Ultra performance liquid chromatographyUltra performance liquid chromatography
Ultra performance liquid chromatography
 
A hands on guide to ultrafiltration1
A hands on guide to ultrafiltration1A hands on guide to ultrafiltration1
A hands on guide to ultrafiltration1
 
Uplc ppt
Uplc ppt Uplc ppt
Uplc ppt
 
Automated SPE-LC/MS/MS method development using ITSP
Automated SPE-LC/MS/MS method development using ITSPAutomated SPE-LC/MS/MS method development using ITSP
Automated SPE-LC/MS/MS method development using ITSP
 
Method Development Poster-MicroLiter
Method Development Poster-MicroLiterMethod Development Poster-MicroLiter
Method Development Poster-MicroLiter
 
Tips and tricks_hplc_troubleshooting
Tips and tricks_hplc_troubleshootingTips and tricks_hplc_troubleshooting
Tips and tricks_hplc_troubleshooting
 
Uplc
UplcUplc
Uplc
 
Ultra Performance Liquid Chromatography
Ultra Performance Liquid ChromatographyUltra Performance Liquid Chromatography
Ultra Performance Liquid Chromatography
 

Recently uploaded

Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...CALL GIRLS
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowSonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowRiya Pathan
 
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service CoimbatoreCall Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatorenarwatsonia7
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...Miss joya
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknownarwatsonia7
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...narwatsonia7
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service JaipurHigh Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipurparulsinha
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Gabriel Guevara MD
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Servicesonalikaur4
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Menarwatsonia7
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowNehru place Escorts
 

Recently uploaded (20)

Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowSonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
 
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service CoimbatoreCall Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
 
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service JaipurHigh Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
 

Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analyses When Exploiting the Properties of 2.6 μm Kinetex® Core-Shell Columns

  • 1. APPLICATIONS TN-1104 Page 1 of 8For additional technical notes, visit www.phenomenex.com Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analyses When Exploiting the Properties of 2.6 μm Kinetex® Core-Shell Columns James Rudge1, David Neville2 , Richard Houghton2 , Chris Message1 , and Jeremy Bierman3 1 Phenomenex LTD, Queens Avenue, Hurdsfield Industrial Estate, Macclesfield, Cheshire, SK10 2BN, United Kingdom 3 Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 90501 USA 2 Quotient Bioresearch Ltd., Newmarket Road, Fordham, Cambridgeshire, CB7 5WW, United Kingdom When compared to 1.7 μm fully porous materials, the ultra-high efficiency and low backpressures provided by Kinetex core-shell 2.6 μm columns, provides users opportunities to go beyond what is traditionally accepted for UHPLC runs. To match backpressures exerted by 1.7 µm fully porous columns, large increases in core- shell column length and/or system flow rate is required. When gra- dient conditions are maintained, increases in flow rate and length result in an increase in peak capacity on the order of 26 %. When MS conditions are optimized to compensate for new flow rates, sensitivity increases of up to 3.5 times are observed and further- more, peak capacity figures increase by almost 70 %. Introduction The introduction of Kinetex core-shell columns has brought dra- matic benefits to chromatographers. The high efficiency shown from the 2.6 μm core-shell particle has allowed users to run ultra- high performance separations on HPLC and UHPLC systems alike. Moreover, the larger particle size of the Kinetex 2.6 μm material compared to 1.7 μm fully porous particles significantly reduces backpressures by a factor of ~40 %. This decrease in backpressure provides users an opportunity to exploit flow rates and/or column lengths to allow for faster UHPLC run times, higher peak capaci- ties, and improvements in peak sensitivities. Improvements in peak capacity can be realized if gradient rates are not changed and flow rates and/or column lengths are increased. The reason for this is that when the same gradient rate is applied, faster flow rates act to volumetrically lower gradients. However, if column length is increased and gradient time is also increased (with respect to the change in length) then it is known that again peak capacity values will increase due to the increased efficiency of the longer column, but this comes at the expense of increased run times. Improved peak capacities observed over identical gradient win- dows are due to a reduction in peak widths. If smaller peak widths are observed, then this should give an opportunity for larger peak heights and hence better peak sensitivity. However, attention should be given to optimizing the detector to realize this benefit fully. The work discussed in this technical note demonstrates perfor- mance advantages when operating 2.6 µm Kinetex core-shell par- ticle columns under conditions that deliver backpressures match- ing that of a 50 x 2.1 mm fully porous 1.7 µm C18 column that is operated at traditional UHPLC flow rates (0.8 mL per minute) for a 2.1 mm ID column. Materials and Methods All analytes were purchased from Sigma-Aldrich. Laboratory: All work was carried out by Quotient Bioresearch Laboratories. Columns Used: A fully porous 1.7 μm C18, 50 x 2.1 mm column was compared with Kinetex 2.6 μm C18, 50 x 2.1 mm, 75 x 2.1 mm, and 100 x 2.1 mm columns from Phenomenex. Instrumentation: Waters ACQUITY® Binary UPLC® AB SCIEX API 5000™ Mobile Phase Preparation: Solutions of 0.1 % formic acid in water and acetonitrile Standard Solution Preparation: A 0.5 ng/mL solution containing the following analytes was prepared in water. 1. Haloperidol 2. Diltiazem 3. Terfenadine 4. Cimetidine 5. Acetaminophen 6. Sulfathiazole 7. Pindolol 8. Quinidine 9. Acebutolol 10. Chlorpheniramine 11. Tripolidine 12. Prednisolone 13. Nortriptyline 14. 2-Hydroxy-5-methyl benzaldehyde 15. Hexanophenone Experiment 1 – Performance comparison between 2.6 µm Kinetex core-shell column and 1.7 µm fully porous column at the same flow rate and then at the same pressure on varying column lengths: A 5 µL sample of all 15 components was injected with a gradient chromatographic separation using an initial flow rate of 0.8 mL/ min over 2.5 minutes on a fully porous C18 1.7 µm, 50 x 2.1 mm column and then on a Kinetex C18 2.6 µm, 50 x 2.1 mm column. To match the backpressure of the fully porous column (531 bar, 7700 psi), the flow rate was then increased to 1.4 mL/min on the Kinetex column. Next a Kinetex C18 2.6 µm, 75 x 2.1 mm column
  • 2. APPLICATIONS TN-1104 Page 2 of 8 For additional technical notes, visit www.phenomenex.com was taken and run at 1.1 mL/min, again to match the backpres- sure of the fully porous column. Finally, a Kinetex C18 2.6 µm, 100 x 2.1 mm Kinetex column was taken and run at 0.85 mL/ min, again to run at the same backpressure of the fully porous 1.7 µm column at 0.8 mL per minute. All columns were maintained at 40 ºC and data was collected using MS detection (For initial MS conditions see Table 6, column 2). Experiment 2 – Observed performance under optimized running conditions with an optimized MS detector: Aliquots of 5 µL of the analyte mix were repeatedly injected with a gradient chromatographic separation using an initial flow rate of 0.8 mL/min on a Kinetex C18 2.6 µm, 50 x 2.1 mm column and the total run time was 2.5 minutes. Peak height, peak width, and peak capacity data was collected. Following this, the instrument flow rate was increased to 1.4 mL/min and the data was collected again. The columns were maintained at 40 ºC with MS detection where the source was optimized for each flow rate (see “Opt 0.8 mL/min” and “Opt 1.4 mL/min” columns of Table 2). To increase the number of data points across each peak, subsets of analytes were grouped into 4 groups of 3 and MRM data was collected for three masses per run. Table 1 lists each group and their respec- tive parent and product ions. Experiment 3 – Measured effect of optimized MS detector on observed performance under sub-optimal running conditions: Aliquots of 5 µL of the analyte mix were repeatedly injected with a gradient chromatographic separation using an initial flow rate of 0.8 mL/min on a fully porous C18 1.7 µm, 50 x 2.1 mm column and on a Kinetex C18 2.6 µm, 50 x 2.1 mm column. The total run time was 2.5 minutes and peak height, peak width, and peak capacity data was collected at a flow rate optimized for 0.8 mL per minute (Program A). Using the Kinetex column, the instrument flow rate was increased to 1.4 mL/min (Program B) and the data was collected again. Finally, at 1.4 mL per minute, the gradient was adjusted with respect to the change in flow rate from 0.8 to 1.4 mL per minute. The columns were maintained at 40 ºC with MS detection where the source was optimized for each flow rate (see “Opt 0.8 mL/min” and “Opt 1.4 mL/min” columns of Table 2). Like Experiment 2, to increase the number of data points across each peak, subsets of analytes were grouped into 4 groups of 3 and MRM data was collected for three masses per run. Table 1 lists each group and their respective parent and product ions. Table 1. List of grouped analytes and their respective parent and product ions. Group Number Sample Name Q1 mass (m/z) Q3 mass (m/z) Dwell time (ms) Group 1 Acetaminophen 152.2 110.1 50 Cimetidine 253.2 159.1 50 Pindolol 249.3 116.1 50 Group 2 Sulfathiazole 256.1 91.9 50 Chlorpheniramine 275.2 230 50 Triprolidine 279.4 208.1 50 Group 3 Diltiazem 415.3 178.1 50 Terfenadine 472.4 436.4 50 Acebutolol 337.3 116.2 50 Group 4 Haloperidol 376.3 123 50 Quinidine 325.3 79.1 50 Nortriptyline 264.3 91.1 50 Conditions are the same for all columns, except where noted. Column: Kinetex 2.6 μm C18 100 Å Fully Porous 1.7 μm C18 Dimensions: Kinetex: 50 x 2.1 mm, 75 x 2.1 mm, and 100 x 2.1 mm Fully Porous: 50 x 2.1 mm Flow Rate: 0.8, 1.1, and 1.4 and 0.85 mL/min Mobile Phase: A: 0.1% Formic acid in Water B: 0.1% Formic acid Acetonitrile Gradient: Temperature: 40 ºC Time (min) % A 0.00 95 0.25 95 1.80 10 1.90 10 1.91 2.50 95 95 Conditions are the same for all columns, except where noted. Column: Kinetex 2.6 μm C18 100 Å Fully Porous 1.7 μm C18 Dimensions: 50 x 2.1 mm Flow Rate: 0.8 and 1.4 mL/min Mobile Phase: A: 0.1% Formic acid in Water B: 0.1% Formic acid Acetonitrile Gradient: Temperature: 40 ºC Time (min) % A 0.00 95 0.25 95 1.80 10 1.90 10 1.91 2.50 95 95 Conditions are the same for all columns, except where noted. Column: Kinetex 2.6 μm C18 100 Å Fully Porous 1.7 μm C18 Dimensions: 50 x 2.1 mm Flow Rate: 0.8 and 1.4 mL/min Mobile Phase: A: 0.1% Formic acid in Water B: 0.1% Formic acid Acetonitrile Gradient: Temperature: 40 ºC Program A (0.8 mL/min) Time (min) % A 0.00 95 0.25 95 1.80 10 1.90 10 1.91 2.50 95 95 Program B (1.4 mL/min) Time (min) % A 0.00 95 0.25 95 1.80 10 1.90 10 1.91 2.50 95 95
  • 3. APPLICATIONS TN-1104 Page 3 of 8For additional technical notes, visit www.phenomenex.com fully porous column. The resulting peak capacity (n = 34.6) was almost the same as the fully porous material (see Table 3). The reason for this was that there were almost half the number of col- umn volumes passed down the longer column. So although the 100 mm length column would have been a lot more efficient than the 50 mm length 1.7 µm fully porous column, the gradient had been volumetrically steepened and so any gains in performance due to efficiency would have been negated by a contraction of the gradient. Results and Discussion Experiment 1 The fully porous and Kinetex® 50 x 2.1 mm columns were initially run at 0.8 mL per minute. The backpressures for both columns were recorded and adjusted gradient peak capacity values were then calculated. Figures 1 and 2 illustrate that both the fully porous and Kinetex 50 x 2.1 mm columns gave comparable peak capacity figures (see Table 3). The small change was due to the difference in adjusted retention window (1.285 min for the fully porous column and 1.187 min for Kinetex). In fact, the average peak widths were slightly narrower for the Kinetex (0.036 min) compared to the fully porous column (0.037 min). As expected there was a large disparity in backpressure, 7700 psi for the fully porous column vs 4600 psi for Kinetex. This change in backpressure was exploited where the flow rate on the Kinetex column was increased to 1.4 mL/min to match the backpressure of the 1.7 µm fully porous column. Figure 3 and Table 3 shows that the resulting change in flow rate caused the average peak width to reduce by ~30 % and the peak capacity to increase by 26 %. The increase in performance was attributed to the fact that the number of column volumes had increased per percent of organic thus had volumetrically lowered the gradient. When the column length of the Kinetex was increased to 75 mm and the flow rate was altered to 1.1 mL per minute such that the backpressure matched the 50 x 2.1 mm fully porous column at 0.8 mL per minute, a performance gain was again seen (see figure 4, table 3). In fact, the increase in peak capacity (n = 42.3) almost matched the peak capacity of the Kinetex 50 x 2.1 mm column run at 1.4 mL per minute (n = 43.5). However, the reason behind this gain was not due to a volumetric lowering of the gradient, but due to the increased efficiency manifested in the longer column length. The volumetric gradient remained almost identical to the 50 x 2.1 mm column run at 0.8 mL/min. To see if increasing the length to 100 mm provided any further benefits, a Kinetex 2.6 µm, 100 x 2.1 mm was then run at 0.85 mL per minute to match the backpressure of the 50 x 2.1 mm, 1.7 µm Table 2. Mass spectrometer (MS) source conditions Mass Spectrometric Source Conditions: Parameter Initial Opt 0.8 mL/min Opt 1.4 mL/min CAD (PSI) 7 4 7 Curtain gas (PSI) 30 30 30 Gas1 (PSI) 50 40 50 Gas2 (PSI) 40 50 50 IS (eV) 5500 2000 2000 TEM (°C) 650 750 750 Opt = Optimized flow rate Figure 1. LC/MS chromatogram of the fully porous 1.7 µm, C18, 50 x 2.1 mm at 0.8 mL/min 14 5 4 6 8 7 9 10 11 12 2 1 13 3 15 1.4e6 19861 1.3e6 1.2e6 1.1e6 1.0e6 9.0e5 8.0e5 7.0e5 6.0e5 5.0e5 4.0e5 3.0e5 2.0e5 1.0e5 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 Intensity,cps 1.6 1.8 2.0 2.2 min AppID19861 7700 psi (531 bar) Peak Capacity (n) = 34.3 Figure 2. LC/MS chromatogram of Kinetex 2.6 µm C18, 50 x 2.1 mm at 0.8 mL/min 14 5 4 6 8 7 9 10 11 12 2 1 13 3 15 7.0e5 19911 6.5e5 6.0e5 5.5e5 5.0e5 4.5e5 4.0e5 3.5e5 3.0e5 2.5e5 2.0e5 1.5e5 1.0e5 5.0e4 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 Intensity,cps 1.6 1.8 2.0 2.2 min 9.9e5 9.5e5 9.0e5 8.5e5 8.0e5 7.5e5 0,0 0,2 0,4 0,6 0,8 1,0 1,2 1,4 1,6 1,8 2,0 2,2 min AppID19911 4600 psi (317 bar) Peak Capacity (n) = 32.9
  • 4. APPLICATIONS TN-1104 Page 4 of 8 For additional technical notes, visit www.phenomenex.com From the above results it was concluded that Kinetex 2.6 µm core- shell columns run at 1.4 mL per minute was optimal in terms of peak width and peak capacity. Nevertheless, it was noticed that although column performance had improved, peak sensitivities had dropped. It was hypothesized that the reason for this drop in sensitivity was due to the fact that the MS had not been optimized for the change in flow rate and hence source modifications would be required to rectify this. Moreover it was also concluded that an increase in the mass dwell time per MRM may also improve peak sensitivity. Work was then done to optimize the MS and acquire masses in small groups to allow for an increase in the mass dwell time. Experiment 2 To observe if peak sensitivity is recovered when running under more optimal MS conditions for the faster flow rate, it was decided to run just the basic analytes in the mix (9) as these provided the best peak sensitivity and peak shape in Experiment 1 (Figure 6, 7, and 8). The results (see Table 4) indicated again a big increase in peak capacity (52 %), however, unexpectedly peak heights in- creased too, ranging from a 7 % increase to around 3.5 times increase (Figure 9). It was concluded that the MS optimization allowed for more effective desolvation thus allowing more ions into the MS source. Furthermore, due to running small groups, the MS was able to detect the peak apexes more effectively under the optimized conditions. Figure 5. LC/MS chromatogram of Kinetex 2.6 µm C18, 100 x 2.1 mm at 0.85 mL/min min 19940 1 2 8 6 5 4 11 10 9 3 12 1315 14 2 4 6 8 10 12 14 16 18 20 22 24min 7 Intensity,cps 1.08e6 1.05e6 1.00e6 9.50e5 9.00e5 8.50e5 8.00e5 7.50e5 7.00e5 6.50e5 6.00e5 5.50e5 5.00e5 4.50e5 4.00e5 3.50e5 3.00e5 2.50e5 2.00e5 1.50e5 1.00e5 5.00e4 0.0 AppID19940 7700 psi (531 bar) Peak Capacity (n) = 34.6 Table 3. Performance comparison summary of Kinetex 2.6 µm vs. 1.7 µm fully porous columns for Experiment 1 Sample Name PWave (min) dRT (min) Peak Capacity (n) Fully Porous C18 1.7 µm, 50 x 2.1 mm @ 0.8 mL/min 0.037 1.29 34.4 Kinetex C18 2.6 µm, 50 x 2.1 mm @ 0.8 mL/min 0.036 1.19 32.9 Kinetex C18 2.6 µm, 50 x 2.1 mm @ 1.4 mL/min 0.025 1.08 43.6 Kinetex C18 2.6 µm, 75 x 2.1 mm @ 1.1 mL/min 0.026 1.12 42.4 Kinetex C18 2.6 µm, 100 x 2.1 mm @ 0.85 mL/min 0.034 1.17 34.6 Figure 4. LC/MS chromatogram of Kinetex 2.6 µm C18, 75 x 2.1 mm at 1.1 mL/min 1 2 8 6 5 4 11 10 19939 9 3 12 1315 14 2 4 6 8 10 12 14 16 18 20 22 24min 7 Intensity,cps 1.10e6 1.05e6 1.00e6 9.50e5 9.00e5 8.50e5 8.00e5 7.50e5 7.00e5 6.50e5 6.00e5 5.50e5 5.00e5 4.50e5 4.00e5 3.50e5 3.00e5 2.50e5 2.00e5 1.50e5 1.00e5 5.00e4 0.0 AppID19939 7700 psi (531 bar) Peak Capacity (n) = 42.3 Figure 3. LC/MS chromatogram of Kinetex® 2.6 µm C18, 50 x 2.1 mm at 1.4 mL/min 14 5 4 6 8 7 910 11 12 2 1 13 3 15 1.4e6 19862 1.3e6 1.2e6 1.1e6 1.0e6 9.0e5 8.0e5 7.0e5 6.0e5 5.0e5 4.0e5 3.0e5 2.0e5 1.0e5 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 Intensity,cps 1.6 1.8 2.0 2.2 min AppID19862 7700 psi (531 bar) Peak Capacity (n) = 43.5
  • 5. APPLICATIONS TN-1104 Page 5 of 8For additional technical notes, visit www.phenomenex.com Figure 6. Group 1 analytes on Kinetex® C18 2.6 µm, 50 x 2.1 mm at 0.8 and 1.4 mL/min respectively under optimal MS conditions for each flow rate. 19941 1.00e6 5.00e5 0.5 3 Intensity,cps 0.0 1.0 1.5 2.0 2.5 min 1 6.0e5 4.0e5 2.0e5 0.5 Intensity,cps 0.0 1.0 1.5 2.0 2.5 min 0.5 2 Intensity,cps 1.00e5 5.00e4 0.0 1.0 1.5 2.5 min 19945 1.21e5 Intensity,cps 1.00e5 5.00e4 0.0 0.5 2 1.0 1.5 2.0 min 1.4e5 Intensity,cps 1.0e5 5.0e4 0.0 0.5 3 1.0 1.5 2.0 min 3.0e5 Intensity,cps 2.0e5 1.0e5 0.0 0.5 1 1.0 1.5 2.0 min 19946 4.0e5 Intensity,cps 2.0e5 0.0 0.5 1 1.0 1.5 2.0 min 1.9e4 Intensity,cps 1.5e4 1.0e4 5000.00 0.0 0.5 2 1.0 1.5 2.0 min 1.22e6 Intensity,cps 1.00e6 5.00e5 0.0 0.5 3 1.0 1.5 2.0 min 19942 4.0e5 Intensity,cps 2.0e5 0.0 0.5 2 1.0 1.5 2.0 min 5.2e5 Intensity,cps 4.0e5 2.0e5 0.0 0.5 3 1.0 1.5 2.0 min 4.7e5 Intensity,cps 4.0e5 2.0e5 0.0 0.5 1 1.0 1.5 2.0 min 19943 2.8e4 Intensity,cps 2.0e4 1.0e4 0.0 0.5 2 1.0 1.5 2.0 min 1.4e6 Intensity,cps 1.0e6 5.0e6 0.0 0.5 3 1.0 1.5 2.0 min 4.8e5 Intensity,cps 4.0e5 2.0e5 0.0 0.5 1 1.0 1.5 2.0 min 0.5 19944 2 Intensity,cps 9.0e4 5.0e4 0.0 1.0 1.5 2.0 2.5 min 4.0e5 0.5 1 Intensity,cps 2.0e5 0.0 1.0 1.5 2.0 2.5 min 0.5 3 Intensity,cps 9.0e5 5.0e5 0.0 1.0 1.5 2.0 2.5 min AppID19944 AppID19941 Figure 7. Group 2 analytes on Kinetex C18 2.6 µm, 50 x 2.1 mm at 0.8 and 1.4 mL/min respectively under optimal MS conditions for each flow rate. AppID19942 AppID19945 Figure 8. Group 3 analytes on Kinetex C18 2.6 µm, 50 x 2.1 mm at 0.8 and 1.4 mL/min respectively under optimal MS conditions for each flow rate. AppID19943 AppID19946 1.4 mL per min 0.8 mL per min 0.8 mL per min 1.4 mL per min 0.8 mL per min 0.8 mL per min 1.4 mL per min 1.4 mL per min
  • 6. APPLICATIONS TN-1104 Page 6 of 8 For additional technical notes, visit www.phenomenex.com It was concluded that the MS optimization allowed for more effective desolvation and this helped to improve peak sensitivities. Furthermore, as the peak widths were so narrow the MS was able to detect the peak apexes more effectively under the optimized conditions. Thus it was decided to compare the fully porous 1.7µm column with Kinetex 2.6 µm under optimized MS conditions. Experiment 3 Kinetex® C18 2.6 µm, 50 x 2.1 mm columns were compared with fully porous 1.7 µm columns at 0.8 mL/min under optimal mass spec conditions for that flow rate. It was found that the Kinetex columns gave a slightly, but not significantly, higher peak capac- ity. However, when the Kinetex columns were run at 1.4 mL per minute significant increases in peak capacity (around 70 %) were observed (see Figure 11 and Table 5). Finally, when the gradi- ent rate was increased such that the number of column volumes matched the run at 0.8 mL/min, the consequence was to reduce the run time to 1.5 minutes. This resulted in reducing the peak capacity to the same level as the fully porous column run at 0.8 mL/min over the original 1.5 minutes. Furthermore, as shown in Figure 10, sensitivity improvements were again observed when the Kinetex column was run at the faster 1.4 mL per minute flow rate. Figure 10. Sensitivity results for Experiment 3 Figure 11. Peak Capacity Data for Experiment 3 Comparing average peak capacity (nave ) on a Kinetex C18 2.6 µm 50 x 2.1 mm column vs. a fully porous C18 1.7 µm 50 x 2.1 mm column under various flow and gradient conditions 1.7 µm FP column 0.8 mL/min over 2.5 min Kinetex 2.6 µm 0.8 mL/min over 2.5 min Kinetex 2.6 µm 1.4 mL/min over 2.5 min Kinetex 2.6 µm 1.4 mL/min over 1.5 min 60 50 40 30 20 10 0 Condition Table 5. Summary of peak capacity data for Experiment 3 Conditions PWave (min) dRT (min) Peak Capacity (n) Fully Porous C18 1.7 µm, 50 x 2.1 mm @ 0.8 mL/min (2.5 min) 0.029 0.87 30.1 Kinetex C18 2.6 µm, 50 x 2.1 mm @ 0.8 mL/min (2.5 min) 0.027 0.88 32.5 Kinetex C18 2.6 µm, 50 x 2.1 mm @ 1.4 mL/min (2.5 min) 0.017 0.89 51.6 Kinetex C18 2.6 µm, 50 x 2.1 mm @ 1.4 mL/min (1.5 min) 0.017 0.50 30.0 Figure 9. Peak height data for all analytes tested under optimized MS conditions for each flow rate. (The 0.8 mL per minute data was normalized to 1.) 1.00 1.50 2.00 2.50 3.00 3.50 4.00 AcebutololCimetidine Pindolol Sulfathiazole Tripolidine Diltiazem TerfenadineAcetaminophen Chlorpheniramine Analyte Changeinsensitivitywherevaluestestedat0.8mL/minis Change in sensitivity from 0.8 to 1.4 mL/min Group 1 Group 2 Group 3 normalisedto1 Table 4. Performance comparison summary of Kinetex run at 0.8 and 1.4 mL per minute respectively Conditions PWave (min) dRT (min) Peak Capacity (n) Kinetex C18 2.6 µm, 50 x 2.1 mm @ 0.8 mL/min 0.024 0.88 36.0 Kinetex C18 2.6 µm, 50 x 2.1 mm @ 1.4 mL/min 0.016 0.89 54.9 0 200000 400000 600000 800000 1000000 1200000 1400000 Comparing peak heights on a Kinetex C18 2.6 µm 50 x 2.1 mm column vs. a Fully Porous C18 1.7 µm 50 x 2.1 column under various flow and gradient conditions Counts(m/z) Analyte Cimetidine Pindolol Sulfathiazole Tripolidine Diltiazem Terfenadine Acebutolol Haloperidol Quinidine Nortriptyline 1.7 µm FP column 0.8 mL/min over 2.5 min Kinetex 2.6 µm 0.8 mL/min over 2.5 min Kinetex 2.6 µm 1.4 mL/min over 2.5 min Kinetex 2.6 µm 1.4 mL/min per min over 1.5 min Chlorpheniramine Acetaminophen
  • 7. APPLICATIONS TN-1104 Page 7 of 8For additional technical notes, visit www.phenomenex.com Conclusion Kinetex columns packed with 2.6 µm core-shell particles offer big advantages over fully porous 1.7 µm columns. The lower back- pressure can be utilized to run columns at flow rates which would normally be impossible to reach on UHPLC instruments due to instrument and material pressure limits. The observed reduced resistance to flow of Kinetex compared to 1.7 µm columns, but comparable efficiency can be used to gain much higher peak ca- pacities when run under fast gradient conditions. Moreover, by steepening the gradient conditions with respect to the change in flow rate, then run times can be reduced without loss of perfor- mance. When MS conditions are optimized to the change in flow rate, then significant sensitivity improvements are also observed. * SecurityGuard Ultra cartridges require holder, Part No. AJ0-9000 When using Kinetex 1.7 µm, increased performance can be achieved, however higher pressure-capable instrumentation is required. ** Columns are pH stable from 1.5-10 under isocratic conditions. Columns are pH stable 1.5-8.5 under gradient conditions. 2.1 mm ID Kinetex columns are pressure stable up to 1000 bar. 2.6 µm Material Characteristics Packing Material Total Particle Size (µm) Porous Shell (µm) Solid Core (µm) Pore Size (Å) Effective Surface Area (m2 /g) Effective Carbon Load % pH Stability Pressure Stability Kinetex XB-C18 2.6 0.35 1.9 100 200 10 1.5 - 8.5** 1000/600 bar Kinetex C18 2.6 0.35 1.9 100 200 12 1.5 - 8.5** Kinetex C8 2.6 0.35 1.9 100 200 8 1.5 - 8.5** Kinetex PFP 2.6 0.35 1.9 100 200 9 1.5 - 8.5** Kinetex HILIC 2.6 0.35 1.9 100 200 0 2.0 - 7.5 Kinetex Phenyl-Hexyl 2.6 0.35 1.9 100 200 11 1.5 - 8.5** 1.7 µm Material Characteristics Packing Material Total Particle Size (µm) Porous Shell (µm) Sol- id Core (µm) Pore Size (Å) Effective Surface Area (m2 /g) Effective Carbon Load % pH Stability Pressure Stability Kinetex XB-C18 1.7 0.23 1.25 100 200 10 1.5 - 8.5** 1000 bar Kinetex C18 1.7 0.23 1.25 100 200 12 1.5 - 8.5** Kinetex C8 1.7 0.23 1.25 100 200 8 1.5 - 8.5** Kinetex PFP 1.7 0.23 1.25 100 200 9 1.5 - 8.5** Kinetex HILIC 1.7 0.23 1.25 100 200 0 2.0 - 7.5 Kinetex Phenyl-Hexyl 1.7 0.23 1.25 100 200 11 1.5 - 8.5** 2.6 μm MidBore™ Columns (mm) SecurityGuard Ultra Cartridges 30 x 3.0 50 x 3.0 75 x 3.0 100 x 3.0 150 x 3.0 3/pk XB-C18 00A-4496-Y0 00B-4496-Y0 00C-4496-Y0 00D-4496-Y0 00F-4496-Y0 AJ0-8775 C18 00A-4462-Y0 00B-4462-Y0 00C-4462-Y0 00D-4462-Y0 00F-4462-Y0 AJ0-8775 C8 00A-4497-Y0 00B-4497-Y0 00C-4497-Y0 00D-4497-Y0 00F-4497-Y0 AJ0-8777 PFP 00A-4477-Y0 00B-4477-Y0 00C-4477-Y0 00D-4477-Y0 00F-4477-Y0 AJ0-8780 HILIC 00A-4461-Y0 –– –– –– 00F-4461-Y0 AJ0-8779 Phenyl-Hexyl –– –– –– –– –– AJ0-8781 for 3.0 mm ID Kinetex® Ordering Information 2.6 μm Analytical Columns (mm) SecurityGuard Ultra Cartridges 30 x 4.6 50 x 4.6 75 x 4.6 100 x 4.6 150 x 4.6 3/pk XB-C18 –– 00B-4496-E0 00C-4496-E0 00D-4496-E0 00F-4496-E0 AJ0-8768 C18 00A-4462-E0 00B-4462-E0 00C-4462-E0 00D-4462-E0 00F-4462-E0 AJ0-8768 C8 –– 00B-4497-E0 00C-4497-E0 00D-4497-E0 00F-4497-E0 AJ0-8770 PFP 00A-4477-E0 00B-4477-E0 00C-4477-E0 00D-4477-E0 00F-4477-E0 AJ0-8773 HILIC 00A-4461-E0 00B-4461-E0 00C-4461-E0 00D-4461-E0 00F-4461-E0 AJ0-8772 Phenyl-Hexyl –– 00B-4495-E0 –– 00D-4495-E0 00F-4495-E0 AJ0-8774 for 4.6 mm ID 1.7 μm Minibore Columns (mm) SecurityGuard Ultra Cartridges* 30 x 2.1 50 x 2.1 100 x 2.1 150 x 2.1 3/pk XB-C18 00A-4498-AN 00B-4498-AN 00D-4498-AN 00F-4498-AN AJ0-8782 C18 00A-4475-AN 00B-4475-AN 00D-4475-AN 00F-4475-AN AJ0-8782 C8 00A-4499-AN 00B-4499-AN 00D-4499-AN 00F-4499-AN AJ0-8784 PFP 00A-4476-AN 00B-4476-AN 00D-4476-AN 00F-4476-AN AJ0-8787 HILIC 00A-4474-AN 00B-4474-AN 00D-4474-AN –– AJ0-8786 Phenyl-Hexyl –– 00B-4500-AN 00D-4500-AN 00F-4500-AN AJ0-8788 for 2.1 mm ID 1.7 μm MidBore Columns (mm) SecurityGuard Ultra Cartridges 30 x 3.0 50 x 3.0 100 x 3.0 3/pk XB-C18 00A-4498-Y0 00B-4498-Y0 00D-4498-Y0 AJ0-8775 C18 –– 00B-4475-Y0 00D-4475-Y0 AJ0-8775 C8 00A-4499-Y0 00B-4499-Y0 00D-4499-Y0 AJ0-8777 PFP –– –– 00D-4476-Y0 AJ0-8780 HILIC –– 00B-4474-Y0 –– AJ0-8779 Phenyl-Hexyl –– –– –– AJ0-8781 for 3.0 mm ID 2.6 μm Minibore Columns (mm) SecurityGuard Ultra Cartridges 30 x 2.1 50 x 2.1 75x 2.1 100 x 2.1 150 x 2.1 3/pk XB-C18 00A-4496-AN 00B-4496-AN 00C-4496-AN 00D-4496-AN 00F-4496-AN AJ0-8782 C18 00A-4462-AN 00B-4462-AN 00C-4462-AN 00D-4462-AN 00F-4462-AN AJ0-8782 C8 00A-4497-AN 00B-4497-AN 00C-4497-AN 00D-4497-AN 00F-4497-AN AJ0-8784 PFP 00A-4477-AN 00B-4477-AN 00C-4477-AN 00D-4477-AN 00F-4477-AN AJ0-8787 HILIC 00A-4461-AN 00B-4461-AN 00C-4461-AN 00D-4461-AN 00F-4461-AN AJ0-8786 Phenyl-Hexyl –– 00B-4495-AN –– 00D-4495-AN –– AJ0-8788 for 2.1 mm ID
  • 8. APPLICATIONS TN-1104 Page 8 of 8 Terms and Conditions Subject to Phenomenex Standard Terms & Conditions, which may be viewed at www.phenomenex.com/TermsAndConditions Trademarks Kinetex is a registered trademark of Phenomenex, Inc. in the United States, European Union, and other jurisdictions. MidBore is a registered trademark of Phenomenex, Inc. Waters, AC- QUITY, and UPLC are registered trademarks of Waters Corporation. API 5000 is a trademark of AB Sciex Pte. Ltd. AB SCIEX™ is being used under license. Disclaimers Comparative separations may not be representative of all applications. Phenomenex is not affiliated with Waters Corporation. © 2011 Phenomenex, Inc. All rights reserved. TN89190811_L t: f: Australia 02-9428-6444 02-9428-6445 auinfo@phenomenex.com t: f: Austria 01-319-1301 01-319-1300 anfrage@phenomenex.com t: f: Belgium +31 (0)30-2418700 +31 (0)30-2383749 beinfo@phenomenex.com t: f: Canada (800) 543-3681 (310) 328-7768 info@phenomenex.com t: f: Denmark 4824 8048 4810 6265 nordicinfo@phenomenex.com t: f: Finland +358 (0)9 4789 0063 +45 4810 6265 nordicinfo@phenomenex.com t: f: France 01 30 09 21 10 01 30 09 21 11 franceinfo@phenomenex.com t: f: Germany 06021-58830-0 06021-58830-11 anfrage@phenomenex.com t: f: India 040-3012 2400 040-3012 2411 indiainfo@phenomenex.com t: f: Ireland 01 247 5405 +44 1625-501796 eireinfo@phenomenex.com t: f: Italy 051 6327511 051 6327555 italiainfo@phenomenex.com t: f: Luxembourg +31 (0)30-2418700 +31 (0)30-2383749 nlinfo@phenomenex.com t: f: Mexico 001-800-844-5226 001-310-328-7768 tecnicomx@phenomenex.com t: f: The Netherlands 030-2418700 030-2383749 nlinfo@phenomenex.com t: f: New Zealand 09-4780951 09-4780952 nzinfo@phenomenex.com t: f: Norway +47 810 02 005 +45 4810 6265 nordicinfo@phenomenex.com t: f: Puerto Rico (800) 541-HPLC (310) 328-7768 info@phenomenex.com t: f: United Kingdom 01625-501367 01625-501796 ukinfo@phenomenex.com t: f: All other countries: Corporate Office USA (310) 212-0555 (310) 328-7768 info@phenomenex.com www.phenomenex.com Phenomenex products are available worldwide. For the distributor in your country, contact Phenomenex USA, International Department at international@phenomenex.com If Kinetex analytical columns do not provide at least an equivalent separation as compared to a competing column of the same phase and dimensions, return the column with comparative data within 45 days for a FULL REFUND.