3. HARKEN’s -TEN COMMANDMENTS
The IDEAL prosthesis MUST
1. technically practical to insert.
2. inserted in a physiological site (generally the normal anatomic site).
3. capable of permanent fixation
4. chemically inert and not damage blood elements.
5. offer no resistance to physiological flows.
Ann Thorac Surg 1989;48:18-19
4. HARKEN’s -TEN COMMANDMENTS
6. close promptly (less than 0.05 second).
7. remain closed during the appropriate phase of the cardiac cycle.
8. not propagate emboli.
9. have lasting physical and geometric features.
10 It must not annoy the patient.
Ann Thorac Surg
1989;48:18-19
7. EVOLUTION OF PROSTHETIC HEART
VALVES
• 1950’s : Idea of prosthetic
heart valve consisting of cage
with mobile spherical poppet
• Resembled a bottle stopper.
9. HUFNAGEL’S BALL AND CAGE DEVICE
• No anticoagulation was used
• Drawbacks:
– Mortality
– cumbersome insertion during brief cross‐clamp period
– Valve noise : hollow nylon ball coated with silicone
rubber
– Poor hemodynamics
– embolization and thrombosis
10. DR. JUDSON CHESTERMAN-1955
• First ball valve implantation in
mitral position
• Cage and poppet device :
Perspex
• Fixed to outside of heart with
two buttons attached to device
• Patient died 14 hours after
surgery
12. DWIGHT EMARY HARKEN- 1960
• Double cage-ball valve
• Cage : stainless steel
• Lucite ball- changed to silicone-
rubber
• sewing ring Ivalon or Teflon
backing.
• Placed subcoronary location
• Two survivors among first seven
patients
13. STARR-EDWARDS BALL VALVE-1960
• "The need exists and it will eventually be
met. That is the law that all nature
follows.“-Lowell Edward
• Our job is not to design a valve identical
to nature‘s,not to see how close we can
come to duplicating a natural
phenomenon, but to overcome the
clinical problem of the diseased heart
valve. If we can do this with a valve
similar to the natural one-fine. But we
must evaluate on the basis of function
rather than form.—Albert Starr
(Tex Heart Inst J
14. DEVELOPMENT OF STARR-EDWARDS
BALL VALVE
• Bileaflet valve : silicone-
rubber leaflets hinged on a
central crossbar made of
solid Teflon
•Teflon cloth margin for
fixation.
•Thrombus formation
originate at suture line and
grow by direct extension
onto leaflets. (Tex Heart Inst J
15. DEVELOPMENT OF STARR-EDWARDS BALL
VALVE
• Adopted idea of ball and cage
valve from Ellis and Bulbulian
• Cage : Lucite cage
(methacrylate)thick struts
• Ball : compression moulded
Silicone elastomer rubber
• Sewing ring : Knitted teflon
• 1960 in mitral position
16. DEVELOPMENT OF STARR-EDWARDS BALL
VALVE
• First successful human
implantation: Philip
Admunson on September 21,
1960
• He had undergone 2 previous
commissurotomies and was
in NYHA functional class IV
• Admunson survived for 15
years after the implantation
18. SE model 6120/1260
• Heat cured ball – disapperance of ball variance
• Thrombo embolic problem: extending cloth of sewing ring
to edge of inflow orifice(decreasing amount of metal
exposed)
• Lowering of cage
• Barium sulfate impregnated silastic ball to make radio-
opaque.
• Teflon and polypropylene sewing ring.
19. DRAWBACK OF STARR-EDWARDS
BALL VALVE
• Ball variance.
– Hollow stellite-21 balls
– Heat cured silicone rubber balls
• High profile
• Difficult implantation in small ventricles and
small aortic root
• Inherently high gradients
20. DRAWBACK OF STARR-EDWARDS BALL
VALVE
• Less favorable thromboembolic profiles
• Wear and tear of cloth
• Edwards Lifesciences (Irvine, CA) discontinued
production of the Starr-Edwards valve in 2007.
21. MAGOVERN-CROMIE BALL VALVE-1962
• Sutureless valve
• Ball-Silicone rubber
with barium
• cage-titanium
• 25-year experi- ence
– 728 patients between
1962 and 1988
– ball variance occurred in
14 patients (2%)
23. DEBAKEY-SURGITOOL CAGED BALL
VALVE,1967
• hollow pyrolytic carbon
poppet
• first use of new carbon
material developed by Dr
Jack Bokros
• Drawback:
• strut wear and strut
fracture
24. DR JACK BOKROS
• Invented pyrolytic carbon :
Pyrolyte
• Exceptional
biocompatibility (highly
thromboresistant)
• Founded Medical Carbon
Research Institute ,Austin
• silicone-free pyrolytic
carbon: On-X Valve.
25. PYROLYTIC CARBON
• Isotropic form of carbon.
• distorted lattice structure with random unassociated carbon
atoms
• Formed by pyrolysis of hydrocarbon gas creating random
crystallization
• Excellent stability, strength, wear resistance, fatigue resistance
and biocompatibility
• Originally developed for the encapsulation of nuclear fuel rods
26. NINA STARR BRAUNWALD -1960
• Flexible polyurethane-
Dacron fabric mitral
valve prosthesis with
attached Teflon-tape
chordae ten- doneae
• 44 year old female with
mitral regurgitation
Nina Starr
Braunwald (1928–
1992)
27. BRAUNWALD-CUTTER BALL VALVE,1968
• Cloth covered caged ball valve
• Struts : knit Dacron tubing
• Inflow ring : ultrathin polypropylene
mesh fabric
• Drawbacks:
– fabric wear
– silicone poppet abrasion in aortic valves
leading to poppet escape.
28.
29.
30. NONTILTING DISC VALVES
• Closing component was a poppet that was held in a cage
(open position) or obturated the ring (closed position)
• Differed in material of disc, housing, ring and cage
design.
• Advantages:
– low-profile design
– easier implantation
– very little opening resistance
– very short closure delay (and therefore very little
regurgitation)
37. TILTING DISC VALVES
• Designed on principle of tilting disc : differences
in disc housing and angle of tilting
• Tilting angle : eventual diagnostics of valve failure
• Discs are radio-opaque , fluoroscopy imaging
– normal mobility
– restricted range of motion
– complete occluder blockade.
38. LILLEHEI-CRUZ-KASTER TILTING DISC
VALVE, 1963
• Free-floating disc tilting on
edge of orifice ring
• Good hemodynamic qualities
• Disadvantage: Area of stasis
between open disc & aortic
wall
39. BJORK-SHILEY FLAT DISC VALVE-1969
• Flat occluder disc
• Disc : Delrin(POM=polyoxymethylene);
Pyrolyte disc
• Stellite housing
• Disc tilting up to 60°
• Inlet and outlet struts welded to flange
•
• Early failures of inlet strut welds eliminated
with change to welding process.
41. BJORK-SHILEY CONVEXO-CONCAVE
TILTING DISC VALVE-1976
Advantages:
• Decreased thromboembolic complications by 50 %
• superior hemodynamic characteristics
• valve completely open with half flow required with
straight disc
• much more rapid reaction on closure resulting in
reduced regurgitation
42. BJORK-SHILEY CONVEXO-CONCAVE TILTING
DISC VALVE-1976
• 1986: Removed from market due to serious safety
concerns.
• outlet strut fracture causing death in 2/3rd of patients
• After recall, not all Bjork Shiley valves were removed
from patients
• 1991 : class action lawsuit filed against Pfizer
• 1992: lawsuit settled, with Pfizer expecting to pay
between $155 and $205 million total.
44. LILLEHEI-KASTER TILTING DISC
PROSTHESIS, 1970
• Pivot point moved forward to cord
measuring one-third of
circumference of orifice.
• Lateral guides replaced cage of Cruz
valve.
• Seating : titanium
• Disc : Pyrolyte.
• Disc opening = 80° and closing = 18°
45. OMNISCIENCE-1978 AND
OMNICARBON-1984
• Omniscience valve:
– two tabs as catch mechanism.
– titanium housing ,Pyrolyte disc.
• Omnicarbon Valve
– disc and housing -pyrolytic
carbon.
– disc opens up to 80° and closes
at 12°, thus achieving the tilting
range of 68°.
46. MEDTRONIC-HALL-KASTER TILTING
DISC, 1977
• Titanium housing
• Pyrolytic carbon disc with a
small central perforation.
• Disc slides over a guidewire
through its central perforation
to tilt open
• opens up to 75° (aortic valve)
and 70° (mitral valve)
47.
48. TTK CHITRA VALVE
• Distributed by TTK (Tiruvellore Thattai
Krishnamachari) pharma Chennai
• The first human implant was December 6,
1990 at Sree Chitra Tirunal Institute for
Medical Sciences and Technology, Trivandrum
Marthanda Varma Sankaran
Valiathan
51. Kalke-Lillehei Bileafet Valve
• Based on configuration of Indian tidal floodgates
• peripheral hinging leaflets and central opening.
• Implanted on May 20, 1968, woman with advanced
rheumatic mitral disease.
• She developed low cardiac output and died 48
hours postoperatively.
52. St. Jude Medical Heart Valve-1977
• Pyrolytic carbon over graphite substrate for
housing and leaflets
• Opening angle = 85o : central near laminar flow
• SJM Standard series
• SJM Masters series
• SJM HP (hemodynamic plus, since 1992)
• SJM Regent (since 1998)
55. SJM Masters HP(Hemodynamic Plus)
• Sewing cuff reduced-> supra-annular
placement
• Increased EOA
• Minimizes interference with subvalvular
structures in mitral position.
• Aortic and Mitral sizes: 17 to 27 mm.
56. SJM REGENT
• Supra-annular placement only for aortic position(size 19-27
mm)
• Significantly larger EOAs
– Less gradient even in valve sizes as small as 19 mm
– Renders root enlargement practically unnecessary
• Low-implant height
• Flexi cuff and Standard cuff
57. ATS (Advancing The Standard) MEDICAL
MECHANICAL HEART VALVE
• Pyrolytic carbon over graphite
substrate for housing and leaflets
• opening angle is 85o.
• Open pivot :decrease blood stasis
and thrombus formation near
hinge
• Standard series
• Advanced Performance series
59. Medtronic Open Pivot Mechanical
Valve
• Originally developed and owned by ATS Medical,Inc,
(Minneapolis, MN)
• Open pivot design
– Eliminates shallow recessess in hinge area where clots may
form
– Continous gentle flow of blood across valve-> low
hemolysis,low level of clotting
– Continous passive washing of valve
60. ON-X MECHANICAL PROSTHESIS -1996
• Housing: pure, non-silicon
carbide alloyed pyrolytic-carbon
• Leaflets : pyrolytic carbon-coated
over tungsten-loaded graphite
substrate
• curved housing in flow geometry
and an orifice diameter-to-
housing height ratio to minimize
vena contracta phenomenon and
facilitate laminar blood flow.
• opening angle 90o
65. TERMINOLOGY & PARAMETERS
• Valve size :outer diameter of the valve housing
(TAD - tissue annulus diameter)
• Internal orifice diameter (IOD) of the valve is
smaller than labeled valve size
• ESRD (external sewing ring diameter) larger
than TAD
69. PATIENT–PROSTHESIS MISMATCH
(PPM).
• First described in 1978 by Rahimtoola
• “Mismatch can be considered to be present when the effective
prosthetic valve area, after insertion into the patient, is less than
that of a normal human valve”
• Smaller than expected effective orifice area (EOA) in relation to
the patient's body surface area (BSA) will result in higher
transvalvar gradients.
Circulation vol 58 ,No
1,July 1978
70. PATIENT–PROSTHESIS MISMATCH (PPM).
• Aim : Implant a valve large enough to avoid hemodynamically
significant patient–prosthesis mismatch (PPM).
• Aortic position -IEOA of implanted valve should be > 0.85
cm2/m2.
• Mitral position- IEOA of implanted valve should be > 1.2
cm2/m2.
• Severe patient–prosthesis mismatch occurs
– IEOA <0.65 cm2/m2 in aortic position
– IEOA <0.9 cm/m2 in mitral position
71. PERFORMANCE INDEX (PI)
– Better indicator of hydraulic function efficiency of
particular prosthesis
– Ratio of effective orifice area (EOA) to sewing ring area
– Size-independent measure of valve’s resistance
characteristics
– Bileaflet valves typically have higher PI’s than tilted-disc
models, which in turn have higher PI’s than caged-ball
model
72. GOOD HEMODYNAMICS
• minimal resistance to forward blood flow
• only trivial regurgitant backflow as the occluder closes
• minimal turbulence and stasis in vivo during physiologic
flow conditions
• durable enough to last a lifetime
• constructed of biomaterials that are nonantigenic,
nontoxic, nonimmunogenic, nondegradable, and
noncarcinogenic.
• low incidence of thromboembolism.
73. HEMODYNAMICS
Factors determining opening resistance to blood flow
• orifice diameter
• size, shape, and weight of occluder
• opening angle
• orientation of leaflet or disk occluders with respect
to plane of annular orifice for any given annular size
76. HEMODYNAMICS
Thrombosis:
• Areas of perivalvular blood stagnation and
turbulence
• increase platelet aggregation
• activation of the coagulation proteins
• thrombus formation.
77. HEMODYNAMICS
Dynamic regurgitation
• sum of closing volume and leakage volume
Closing volume
• function of EOA and time needed for closure.
• Closure time is influenced by difference between the
opening and closing angles of occluder and valve ring.
78. HEMODYNAMICS
Leakage volume(washing jets)
• inherent to design of valve
• depends on amount of time valve remains in closed
position
• small amount of regurgitant volume can be beneficial by
• minimizing stasis and reducing platelet aggregation
• decreases incidence of valve thrombosis and valve-
related thromboembolism
79. Hemodynamics: Ball and cage valves
• Minimal leakage
• Leakage volumes - indicate pathologic process
• Annular area for flow creates turbulenece
• Cage may contact ventricular wall during contraction
• Partial obstruction by ball in aortic position
• Increase risk of hemolysis and thromboembolic
complications
80. Hemodynamics : Tilting disc valve
• Less obstruction to flow
• Gradient of 6- 7 mm Hg
• Opening angle high- less
gradient, more
regurgitation
• Opening angle low- more
gradient and less
regurgitation
81. Hemodynamics : Bileaflet valve
• more uniform central, and laminar flow
• less turbulence
• decreased transvalvular pressure gradients
• favorable hemodynamics in smaller sizes makes it
especially useful in children.
• large EOA for each valve size at the expense of
greater regurgitant volumes, especially at low heart
rate
85. valve housing Sewing ring leaflets Opening
angle
Implantatio
n
ROTATABLE
SJM PYROLYTIC
CARBON
COATED
GRAPHITE
SUBSTRATE
POLYESTER/
PTFE
PYROLYTIC
CARBON
COATED
GRAPHITE
SUBSTRATE
850 Masters:
intraannular
HP and
Regent:
supra
annular
YES
MEDTRONIC 100%
PYROLYTIC
CARBON
with
titanium
strengthnin
g band
Double-
velour
POLYESTER
Cuff marker
present
PYROLYTIC
CARBON
COATED
GRAPHITE
SUBSTRATE,
20%
tungsten
impregnate
d
850 Standard:
intra
annular
AP: supra
annular
yes
On-X GRAPHITE
SUBSTRATE
coated with
On-X
carbon-pure
unalloyed
PTFE
mounted
using
titanium
retaining
rings and 5-
On-X
CARBON
COATED
GRAPHITE
SUBSTRATE,
10%
900 Mitral:
supra
annular
Aortic(19-
25mm):intra
supra-
yes
86.
87. MORBIDITY AND MORTALITY
GUIDELINES
• The councils of the Society of Thoracic Surgeons
(STS) and the American Association of Thoracic
Surgery (AATS) formulated the Ad Hoc Liaison
Committee for Standardizing Definitions of
Prosthetic Heart Valve Morbidity.
• The initial report of this committee was issued in
1988 with an update in 1996
• The report strictly defines types of morbidity and
mortality that can occur after valvular surgery.
88. Early Mortality
• Early mortality is to be reported as all-cause
mortality at 30, 60, or 90 days and depicted by
actuarial estimates (with number remaining at
risk and confidence intervals [CIs]) or as
simple percentages, regardless of the patient’s
location, be it home or in a health care facility
89. STRUCTURAL VALVE DETERIORATION
• Includes dysfunction or deterioration
involving the operated valve (exclusive of
infection or thrombosis), as determined by
reoperation, autopsy, or clinical
investigation.
90. STRUCTURAL VALVE DETERIORATION
changes intrinsic to the valve:
• such as wear, fracture, poppet escape, calcification,
leaflet tear, stent creep
• suture line disruption of components of a prosthetic
valve
• new chordal rupture, leaflet disruption, or leaflet
retraction of a re- paired valve.
91. NONSTRUCTURAL DYSFUNCTION
• abnormality not intrinsic to the valve itself
that results in stenosis or regurgitation of
the operated valve or hemolysis.
• do not directly involve valve components
yet result in dysfunction
92. NONSTRUCTURAL DYSFUNCTION
• entrapment by pannus, tissue, or suture
• paravalvular leak
• inappropriate sizing or positioning
• residual leak or obstruction after valve
implantation or repair
• clinically important intravascular hemolytic
anemia
93. VALVE THROMBOSIS
• any thrombus not caused by infection attached to or
near an operated valve that occludes part of the blood
flow path, interferes with valve function, or is sufficiently
large to warrant treatment.
• Valve thrombus found at autopsy in a patient whose
cause of death was not valve related or found at
operation for an unrelated indication should also be
counted as valve thrombosis.
94. EMBOLISM
• Embolism is any embolic event that occurs in
the absence of infection after the immediate
perioperative period.
• Embolism may be manifested by a neurologic
event or a noncerebral embolic event.
95. BLEEDING EVENT
• A bleeding event is any episode of major
internal or external bleeding that causes
death, hospitalization, or permanent injury
(eg, vision loss) or necessitates transfusion.
96. OPERATED VALVE ENDOCARDITIS
• Operated valve endocarditis is any infection
involving a valve on which an operation has
been performed.
• Positive blood cultures are not required for the
diagnosis of operated valve endocarditis.
97. OPERATED VALVE ENDOCARDITIS
The diagnosis is based on one of the following criteria:
• (1) reoperation with evidence of abscess, paravalvular leak, pus,
or vegetation confirmed as secondary to infection by histologic
or bacteriologic studies
• (2) autopsy findings of abscess, pus, or vegetation involving a
repaired or replaced valve
• (3) in the absence of reoperation or autopsy, meeting of the
Duke Criteria for endocarditis
98. Morbidity data
• Rate for nonstructural dysfunction of mechanical valves
– 0.2 to 0.8 (events/ patient-years) for the aortic position
– 0.3 to 1.4 (events/patient-years) for the mitral position
• Rate of thrombosis of mechanical valves
– 0 to 0.2 (events/patient-years) in the aortic position
– 0.4 to 0.8 (events/patient-years) for the mitral position
99. Morbidity data
Rate of thromboembolism
• 1.4 to 2.5 (events/patient-years) for the aortic position
• 1.8 to 3.6 (events/patient-years) for the mitral position
Rate of bleeding event
• 0.8 to 2.5 (events/patient-years) for the aortic position
• 1.2 to 2.2 (events/patient-years) for the mitral position
Rates for prosthetic valve endocarditis
• 0.4 to 0.7 (events/patient-years) for both the aortic and mitral
positions.
Editor's Notes
In principle, his device resembled a bottle stopper (002) for which a patent
was obtained by JB Williams2 in 1858.
Harvard Medical School, Boston, Massachusetts USA
During the sixties, a number of innovative designs were pieced together for experimental as well as clinical application.
The development of the original ball-and-cage valve design can be attributed to the bottle stopper for which a patent
was obtained by JB Williams in 1858.
The problem of mitral regurgitation also received zealous attention from several quarters. ,
CITY GENERAL HOSPTIAL, SHEFFIELD, ENGLAND
when poppet got twisted out of position
JOHN H GIBBON
Philadelphia 6 May 1953 successfully performed cardiopulmonary bypass for open-heart closure of ASD
introduction of cardiopulmonary bypass opened the era of implanting valve prostheses in their native positions
Dr Harken working with Mr W. C. Birtwell from Davol Rubber Company, Rhode Island, started the modern era of prosthetic valve replacement
following excision of the diseased cusps.
concerned that the ball could intrude into the aortic wall. Thus, he designed his valve with a second outer concentric cage.
Both the patients required successive valve replacement, one at 3 years for perivalvular leak and another at 22 years for bacterial endocarditis. The ball valve removed after 22 years had no deterioration
Their first design was a silicone rubber bileaflet valve. This approach proved to be unsuccessful.
Ball variance : gross abnormality with cracks and fracture manifesting as loss of substance,loss of spherical shape,yellowish discolouration due to absorption of lipids.
PyC is generally deposited as a structural coating over a substrate pre-form
It is usually formed in a fluidized bed furnace
Americal thoracic surgeon In 1960, at the age of 32, she led the operative team at the U.S. National Institutes of Health (NIH) that implanted the first successful artificial mitral human heart valve replacement
The valve had two leaflets. Teflon chordae attached to the device were brought out through the ventricular muscle and secured outside the heart at the time of implantation
Due to these drawbacks, the non-tilting disc valves fell into dis use and were replaced by modern tilting disc valves.
first commercially available prosthetic valve with sewing ring configurations dedicated for intra-annular, (177) supra-annular (178) and subannular placement. (179) The aortic valve was available only for intra-annular implantation.
The convex-concave shape of the disc is of significant advantage, not only decreasing thromboembolic com- plications by 5 0 % , but with its superior hemodynamic characteristics, it will keep the valve completely open with half the flow required with a straight disc. It also has a much more rapid reaction on closure resulting in reduced regurgitation. The 70-degree opening angle of the monostrut valve also decreases the in vivo gradient by an average 15% across valve sizes 2 1-33 mm as com- pared to the spherical disc valve.
The convex-concave shape of the disc is of significant advantage, not only decreasing thromboembolic com- plications by 5 0 % , but with its superior hemodynamic characteristics, it will keep the valve completely open with half the flow required with a straight disc. It also has a much more rapid reaction on closure resulting in reduced regurgitation. The 70-degree opening angle of the monostrut valve also decreases the in vivo gradient by an average 15% across valve sizes 2 1-33 mm as com- pared to the spherical disc valve.
It was manufactured first by Shiley Laboratories, then later by Pfizer after that company purchased Shiley.
fracturing at the place where it was welded onto the metal valve ring
Eventually, 619 of the 80,000 convexo-concave valves implanted fractured in this way, with the patient dying in two-thirds of those cases.
Because of the risk associated with removal surgery, which included an approximated death rate of about five percent, patients with weak or defective heart valves did not undergo the
removal surgery, leaving the potentially defective heart valves implanted.
on behalf of patients who had received the implant and were concerned it might fracture.
The lawsuit alleged Pfizer Inc hid potential defects from patients.
Pfizer agreed to create a fund to pay patients for a cardiac consultation and further agreed to set aside funds for research to identify which patients had a significant risk of having the
heart valve fracture.
orifice ring and integral struts are constructed from a single piece of cobalt–chromium alloy to give it a weld-free mechanism.
The 70-degree opening angle of the monostrut valve also decreases the in vivo gradient by an average 15% across valve sizes 2 1-33 mm as com- pared to the spherical disc valve.
Leaflet motion is by rotation and translation.
The aortic and mitral valves differ in their sewing ring configuration
To eliminate the area of stasis behind the open Cruz valve ,kaster
e Omniscience valve was introduced in 1978 to replace the Lillehei-Kaster valve.
Karl Victor Hall, MD (1917 to present) was the chairman of the Department of Surgery at the Rikshospitalet in Oslo, Norway. As a cardiac surgeon, Dr Hall used the available valves of the time until the mid-1970s, when he sensed that improvements could be made in the tilting disc concept.
To accomplish his goal, Dr Hall needed financial help and scientific support. Arne Woien, a friend, physicist, and businessman, came to his aid. As prosthetic heart valve expertise was not available in Norway, Mr Woien suggested that they contact Mr Robert Kaster, who had returned to Minneapolis from New York. In 1974, Medtronic, Inc, of Minneapolis distributed the Lillehei- Kaster valve. Mr Woien was the Medtronic European representative. It was through this connection that Mr Woien and Mr Kaster met.
Doctor Hall, Mr Woien, and Mr Kaster worked to- gether to develop a satisfactory design,
Valve washing was improved by a relatively larger minor orifice and a disc that lifted out of the housing and rotated with opening. It had a moderately high profile in the open position and a low transvalvular gradient. Occluder impingement was possible because its position at the equator of the valve housing made it susceptible to obstruction from retained valve elements, sutures cut too long, or pannus. Loss of structural integrity, however, was never reported. The valve could be rotated after implan- tation to achieve the desired orientation. Several studies reported a low incidence of valve-related morbidity and mortality.
Medtronic discontinued manu- facturing the valve in September of 2009, removing the last of the tilting disc prosthesis available for clinical use.
Tilting Disc
– pivoted eccentrically in the metallic frame.
– MADE FROM ULTRA HMW POLY ETHYLENE •
The sewing ring – POLYETHYLENE TEREPTHALATE (PET) – fitted snugly around the frame – used to suture the valve in the intended position in the heart.
FRAME: COBALT CHROMIUM tungsten ALLOY( HAYNES 25) •
The frame and the disc are hydro dynamically designed to reduce drag and inertia and polished to minimize the chances of clotting.
Bhagavant Kalke, MD, began work in Dr Lillehei’s Min- neapolis laboratory in 1964 and continued with Dr Lille- hei after their move to Cornell Medical School in 1967. Doctor Kalke’s
Later he placed the pivot sites for the two rigid leaflets at the equator of the annular ring.
bileaflet design positioned the leaflets to open with their hinging axis toward the periphery of the metal annulus. As this did not work well, the pivot axis was moved to the diameter of the retaining ring. A single wire guard placed 90 degrees from the main hinging axis of the leaflets and extending over the leaflets aided the control of leaflet excursion
reduced sewing ring significant enlargement of EOA(effective orifice area)
avoiding PPM(patient-prosthesis mismatch) in patients with small aortic annulus and large body-surface area.
Its two pyrolytic carbon semilunar lea ets open up to 85° and close at 30°, which yields a tilting range of 55°. e valve mechanism has remained unchanged since its release, but a rotatable sewing cu was added and other ring re nements were developed. e The primary facilitator of the development of the St. Jude Medical valve was Mr Manny Villafana, founder of Car- diac Pacemakers, Inc. In 1976, Xinon (Chris) Posis, an industrial engineer, designed a prototype of a bileaflet valve with the pivots near the periphery and a central opening. Mr Posis obtained additional suggestions for his design from Dr Demetre Nicoloff, a cardiovascular sur- geon at the University of Minnesota. Posis and Nicoloff showed their prototype valve to Manny Villafana who felt the peripheral hinges would not be suitable. Subse- quently, Mr Posis along with another engineer, Donald Hanson, redesigned the valve with the hinge mechanism located near the central axis of the housing. It was Mr Villafana’s idea that the entire valve be fabricated of Jack Bokros’ pyrolytic carbon, which had been widely adapted for most monoleaflet valves in the 1970s. Working with Jack Bokros, the engineers Posis and Hanson modified the initial pivot mechanism and came up with the con- cept of a leaflet-tab rotating in a “butterfly recess” in the inner wall of the housing. Doctor Nicoloff implanted the first St. Jude valve on October 3, 1977.
It was suggested by Mr Villafana that they call their new valve the Nicoloff valve. Doctor Nicoloff declined. At that time, Mr Villafana’s son was recovering from a serious illness. The St Jude valve was proposed as a name by Mr Villafana. Church liturgy teaches that St Jude Thaddeus is the patron saint of difficult cases.
In the original prosthesis, the housing could not be rotated. The prosthesis has been altered in the Masters (Fig. 40b) series to rotate within the sewing ring and provide radio-opacity of the annular-rotating mechanism
In the standard cuff St. Jude Medical mechanical prosthesis, part of the cuff fabric is intra-annular, whereas in the HP Series prosthesis this fabric has shifted to an entirely supra-annular position. The St. Jude Medical Regent™ prosthesis shifts the carbon rim from intra-annular to entirely supra-annular.
introduced and are designed to optimize hemodynamics.
in patients with small aortic annulus and large body-surface area.
standard SJM model size (21 mm) had an e ective ori ce area (EOA) of only 1.51 cm2, the type HP 2.03 cm2, and the latest model, Regent, 2.47 cm2. Nonetheless, even a “small-size” SJM Regent 19-mm valve has an EOA of 1.84 cm2, i.e., su cient enough to prevent a signi cant patient–prosthesis mismatch in a patient with a body-surface area of 2 m2.
allowing for an increase in orifice size for a given tissue annulus diameter..
redesigned sewing cuff to facilitate supra-annular placement, allowing for an increase in orifice size for a given tissue annulus diameter..
Minimizes interference with subvalvular structures in the mitral position.
St. Jude Medical heart valves are MR conditional
Renders root enlargement practically unnecessary
Significant reduction in left ventricular (LV) mass
Low-implant height
FlexCuff has a flanged and more pliable sewing ring that easily conforms to annulus.
even a “small-size” SJM Regent 19-mm valve has an EOA of 1.84 cm2, i.e., sufficient enough to prevent a signi cant patient–prosthesis mismatch in a patient with a body-surface area of 2 m2.
Due to lack of silicon doping in valves carbon construction
Increased EOA ,decreases ability of retained valve tissue to interfere with opening and closing
Allows the valve to wash itself
e whole arti cial valve diameter is then larger for several millimeters (the thickness of the sewing ring is added), i.e., the external sewing ring diameter (ESRD).
valve housing and its sewing ring oc- clude an important portion of the patient’s valve orifice; therefore,
much larger ori ce area and therefore better hemodynamic parameters can be implanted into a given annulus.
valve housing and its sewing ring oc- clude an important portion of the patient’s valve orifice; therefore,
much larger ori ce area and therefore better hemodynamic parameters can be implanted into a given annulus.
larger EOAs, reduced transprosthetic gradients
EOA measure of how much the prosthesis impedes blood flow through the valve. A higher EOA corresponds to a smaller energy loss. The performance index (PI) normalizes the EOA by valve size and is a size-independent measure of the valve’s resistance characteristics. Bileaflet valves typically have higher PI’s than tilted-disc models, which in turn have higher PI’s than caged-ball models
Indeed, it is generally superior in newer compared with older generations of prostheses, in mechanical compared with stented bioprosthetic valves,16 in stentless compared with stented bioprosthetic valves,17,18 and in supraannular compared with intra-annular stented bioprostheses.
Least resistance to transvalvular blood flow is provided by a large ratio of orifice to total annular area.
A wide opening angle improves the effective orifice area and results in decreased pressure gradients
With an increasing orifice diameter- more energy is lost across the valve as more backflow passes through the valve
is a feature of all prosthetic valves
that passes through valve while it is closed.
Ball valve allowed
in the ball-and-cage design, and in contrast to other mechanical valves, the
The central ball occluder causes lateralization of forward flow and results in turbulence and cavitation
three different flow areas through the valve orifice
at any annulus diameter size and cardiac output compared with the caged-ball and single leaflet tilting valves
Low-profile prostheses simplify the surgical implant.
The lowest profile is that of the bileaflet valve,
The increased flow orifice of the bileaflet valve is clearly shown as compared to ball valves and tilting disc valves
SJM: double-velour knitted polyester fibre in polyester cuff allows rapid,controlled endothelial ingrowth over entire sewing cuff
Polyester sewing cuff have suture markers for valve orientation/placement of sutures
he calculation of events per patient-year(s) is the number of incident cases divided by the amount of person-time at risk. The calculation can be accomplished by adding the number of patients in the group and multiplying that number times the years that patients are in a study in order to calculate the patient-years (denominator). Then divide the number of events (numerator) by the denominator.
Example: 100 patients are followed for 2 years. In this case, there are 200 patient-years of follow-up.
If there were 8 myocardial infarctions in the group, the rate would be 8 MIs per 200 patient years or 4 MIs per 100 patient-years.