Enhanced Simultaneous Total ion and Selective Multi-Ion GC/MS Using a
Combination of New Software and Low-Bleed Columns
Martin Okiro and Robert Wohleb
VICI Gig Harbor Group Inc, P.O. Box 2657, Gig Harbor, WA 98335
Key Words: GC/MS; Enhanced MSD (E-MSD); simultaneous single ion monitoring
Summary
A new mass spectrophotometer (MS) data acquisition system comprised of hardware, firmware and
software has been developed to tune, calibrate and acquire data on single or quadrupole mass
spectrometers. The data acquisition routine is so flexible and powerful that several data collection
methods can be utilized during the course of a single run. For instance, total ion chromatography and
single-ion monitoring can be accomplished simultaneously. Consequently, with a single injection, a
total ion chromatogram (TIC) can be generated as well as additional chromatograms each targeting up
to 32 individual ions. A combination of this enhanced MSD and exceptionally low-bleed columns
results in a powerful new analytical tool allowing detection limit improvement by a factor of ten on
EPA method 8270 semivolatiles. Results further show that the TIC’s component responses remains
uncompromised and degree of confidence in compound identification is improved.
1 Introduction
The use of gas chromatography coupled with mass spectrometer (GC/MS) for the evaluation of
complex mixtures, such as those targeted by EPA Methods 8270 and 624 for the identification and
quantitation of semivolatile and volatile organic compounds respectively, has been amply documented
in the literature [1,2]. The procedure typically involves the generation of a total ion chromatogram
(TIC). When the target compound levels are below the normal detection limits, the EPA allows
separate selected ion monitoring (SIM), but the practice is expressly discouraged due to vastly reduced
confidence in compound identification, except in cases where multiple ions are monitored for each
compound. This paper describes a novel system that allows for the simultaneous acquisition of scan
and SIMs. Since the data is acquired simultaneously, the uncertainty in compound identification is
minimized and sample analysis time is reduced.
The E-MSD data acquisition system employs a combination of hardware, firmware and software to
tune, calibrate and acquire data on single or quadrupole mass spectrophotometers. The data acquisition
system allows the setup of several data collection methods or experiments for each sample injection.
Thus, you can for instance acquire both a Scan and multi-ion SIM from the same run thereby
eliminating injection port bias. A further benefit of the system is the availability of an enhanced
electrometer option with an improved real dynamic range.
GC/MS chromatograms of semi-volatile mixtures were obtained using the normal scan mode of
operation and the simultaneous scan/SIM mode. Chromatogram characteristics were examined to
compare the two modes of operation.
2 Materials and Methods
The instrumentation employed in this study includes a Humble Instruments & Services, Inc. (HISI)
Enhanced MSD (5970 Mass Spectrometer with an A.S.k ModCtrl board). Computer hardware and
software consists of a PC with HP ChemStation for data processing and MStation hardware and
software for the control of the 5970 (A product of Acquisition Solutions k). The GC was a standard HP
5890 equipped with a split splitless injection port. The column was a VB-5 (5% phenyl 95% methyl
polysiloxane) 30mx25mmx25µm.
All standards were prepared in accordance with SW846 and EPA method 8270. Concentrates were
purchased from Restek Corporation and Chem Service. Standards were prepared at levels of 0.5, 1.0,
2.0, 5.0 and 10.0 µg/ml in dichloromethane. Samples were manually injected.
The chromatograms were obtained by injecting 2ul of the various semivolatile standards into the GC
injection port. The GC conditions were as follows: Injector temp 280o
C, splitless; oven initial temp
40o
C for 2 minutes, followed by 6o
C/min to 320o
C.
3 Results and Discussion
The total ion chromatogram for a 1ppm semi-volatile standard mixture is shown in Fig.1.
Figure 1. Total Ion Chromatogram of semi-volatile Mixture.
Selected slices of the total ion chromatograms run in the regular and simultaneous scan/SIM modes are
compared in Fig. 2. Since peak heights and shapes are comparable in both modes of operation, method
sensitivity is not compromised by the scan/SIM mode of operation even though fewer scans/per second
are performed. Standard level is 0.5 ppm
Figure 2
A comparison of the TIC obtained from the regular scan mode and a TIC of all targeted ions obtained
in the SIM/scan mode (SIM component) of a 1.0 ppm standard is shown in Fig. 3. Targeted ions from
the TICs shown in Fig. 3 are displayed in Fig. 4. In Fig. 3, the large peak at 23.7 minutes in the TIC
run under scan mode is eliminated when the chromatogram is run under SIM mode, and only the
targeted peak at 23.6 minutes remains. In addition, the signal to noise level is improved in the
SIM/scan mode.
Figure 3
Figure 4
.
Figure 5
Conclusions
Despite the fact that operating the instrument in
the multiple experiment mode lowered the
number of scans/sec from 0.944 to 0.578 when
targeting six ions in the SIM mode, there was
no significant loss of sensitivity in the
simultaneous scan/SIM run. Calibrations for
both experiments were equally linear. In
addition, the method detection limit for targeted
problematic compounds was enhanced by a
factor greater than ten.
The Combination of E-MSD software and
exceptionally low bleed ValcoBond capillary
columns can be employed to significantly
improve limits of detection in various GC/MS
analytical procedures while also improving the
degree of confidence in acquired data.
5 Literature Cited
[1] Eichelberger, J.W., Harris, L.E., and Budde, W.L, Anal. Chem. 1975, 47, 995-1000.
[2 ] Engel, T.M., et al., “Screening of Semivolatile Organic Compounds for Extractability and
Aqueous Stability by SW-846, Method 3510,” Contract 68-03-3224, U.S. Environmental
Protection Agency, Environmental Monitoring and Support Laboratory, Cincinnati, OH. June
5, 1987.
6 Acknowledgements
The authors would like to thank HISI Analytical Technologies (a division of Humble Instruments &
Supplies, Inc. - www.humble-inc.com) for loan of the equipment used in the study.
E-MSD is a product of Acquisition Solutions k (A.S.k) . For information on E-MSD software and hardware
upgrade go on the web at www.flash.net/~acqsol/

More Related Content

What's hot

Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...Mudasir Gulzar
 
PAPER FOR ANALYTICALnew
PAPER FOR ANALYTICALnewPAPER FOR ANALYTICALnew
PAPER FOR ANALYTICALnewPhilip Alabi
 
NEAFS_DSA_FTIR_Image_NBOMes and LSD_Poster
NEAFS_DSA_FTIR_Image_NBOMes and LSD_PosterNEAFS_DSA_FTIR_Image_NBOMes and LSD_Poster
NEAFS_DSA_FTIR_Image_NBOMes and LSD_PosterAmanda Moore
 
Blood Sugar (Glucose) Measurement, Monitoring and Data Analysis: A Review on ...
Blood Sugar (Glucose) Measurement, Monitoring and Data Analysis: A Review on ...Blood Sugar (Glucose) Measurement, Monitoring and Data Analysis: A Review on ...
Blood Sugar (Glucose) Measurement, Monitoring and Data Analysis: A Review on ...Md Kafiul Islam
 
IRJET - Survey on Etched Fiber Bragg Grating based Senor Applications
IRJET - Survey on Etched Fiber Bragg Grating based Senor ApplicationsIRJET - Survey on Etched Fiber Bragg Grating based Senor Applications
IRJET - Survey on Etched Fiber Bragg Grating based Senor ApplicationsIRJET Journal
 
SIFT-MS_Instant_Quant_of_Comps_in_Air
SIFT-MS_Instant_Quant_of_Comps_in_AirSIFT-MS_Instant_Quant_of_Comps_in_Air
SIFT-MS_Instant_Quant_of_Comps_in_AirJason Worden
 
Deep MALDI for Oncology Biomarkers Study
Deep MALDI for Oncology Biomarkers StudyDeep MALDI for Oncology Biomarkers Study
Deep MALDI for Oncology Biomarkers StudyRohan Thakur
 
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDATocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDAShimadzu Scientific Instruments
 
Interlaboratory study animal proteins in feed 2013 v2
Interlaboratory study animal proteins in feed 2013 v2Interlaboratory study animal proteins in feed 2013 v2
Interlaboratory study animal proteins in feed 2013 v2Rikilt (Wageningen UR)
 
Radiological Consequences of the Fukushima Nuclear Accident - 28 March 2011
Radiological Consequences of the Fukushima Nuclear Accident - 28 March 2011Radiological Consequences of the Fukushima Nuclear Accident - 28 March 2011
Radiological Consequences of the Fukushima Nuclear Accident - 28 March 2011International Atomic Energy Agency
 
Motion artifact reduction in photoplethysmography utilizing empirical mode de...
Motion artifact reduction in photoplethysmography utilizing empirical mode de...Motion artifact reduction in photoplethysmography utilizing empirical mode de...
Motion artifact reduction in photoplethysmography utilizing empirical mode de...guest0802300
 
Saci And Ion Exchange Chromatography
Saci And Ion Exchange ChromatographySaci And Ion Exchange Chromatography
Saci And Ion Exchange ChromatographySimone Cristoni
 
2018 sensors and actuators
2018 sensors and actuators2018 sensors and actuators
2018 sensors and actuatorsFilipa Correia
 
Focus on aggregation: types, causes, characterization, and impact
Focus on aggregation: types, causes, characterization, and impactFocus on aggregation: types, causes, characterization, and impact
Focus on aggregation: types, causes, characterization, and impactKBI Biopharma
 

What's hot (17)

Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
 
PAPER FOR ANALYTICALnew
PAPER FOR ANALYTICALnewPAPER FOR ANALYTICALnew
PAPER FOR ANALYTICALnew
 
NEAFS_DSA_FTIR_Image_NBOMes and LSD_Poster
NEAFS_DSA_FTIR_Image_NBOMes and LSD_PosterNEAFS_DSA_FTIR_Image_NBOMes and LSD_Poster
NEAFS_DSA_FTIR_Image_NBOMes and LSD_Poster
 
Blood Sugar (Glucose) Measurement, Monitoring and Data Analysis: A Review on ...
Blood Sugar (Glucose) Measurement, Monitoring and Data Analysis: A Review on ...Blood Sugar (Glucose) Measurement, Monitoring and Data Analysis: A Review on ...
Blood Sugar (Glucose) Measurement, Monitoring and Data Analysis: A Review on ...
 
IRJET - Survey on Etched Fiber Bragg Grating based Senor Applications
IRJET - Survey on Etched Fiber Bragg Grating based Senor ApplicationsIRJET - Survey on Etched Fiber Bragg Grating based Senor Applications
IRJET - Survey on Etched Fiber Bragg Grating based Senor Applications
 
SIFT-MS_Instant_Quant_of_Comps_in_Air
SIFT-MS_Instant_Quant_of_Comps_in_AirSIFT-MS_Instant_Quant_of_Comps_in_Air
SIFT-MS_Instant_Quant_of_Comps_in_Air
 
Deep MALDI for Oncology Biomarkers Study
Deep MALDI for Oncology Biomarkers StudyDeep MALDI for Oncology Biomarkers Study
Deep MALDI for Oncology Biomarkers Study
 
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDATocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
 
Interlaboratory study animal proteins in feed 2013 v2
Interlaboratory study animal proteins in feed 2013 v2Interlaboratory study animal proteins in feed 2013 v2
Interlaboratory study animal proteins in feed 2013 v2
 
1 perkinelmerleadtestingoverview(5 pgs)
1 perkinelmerleadtestingoverview(5 pgs)1 perkinelmerleadtestingoverview(5 pgs)
1 perkinelmerleadtestingoverview(5 pgs)
 
Radiological Consequences of the Fukushima Nuclear Accident - 28 March 2011
Radiological Consequences of the Fukushima Nuclear Accident - 28 March 2011Radiological Consequences of the Fukushima Nuclear Accident - 28 March 2011
Radiological Consequences of the Fukushima Nuclear Accident - 28 March 2011
 
Motion artifact reduction in photoplethysmography utilizing empirical mode de...
Motion artifact reduction in photoplethysmography utilizing empirical mode de...Motion artifact reduction in photoplethysmography utilizing empirical mode de...
Motion artifact reduction in photoplethysmography utilizing empirical mode de...
 
Saci And Ion Exchange Chromatography
Saci And Ion Exchange ChromatographySaci And Ion Exchange Chromatography
Saci And Ion Exchange Chromatography
 
Materials for glucose measurement
Materials for glucose measurementMaterials for glucose measurement
Materials for glucose measurement
 
2018 sensors and actuators
2018 sensors and actuators2018 sensors and actuators
2018 sensors and actuators
 
Graphene
GrapheneGraphene
Graphene
 
Focus on aggregation: types, causes, characterization, and impact
Focus on aggregation: types, causes, characterization, and impactFocus on aggregation: types, causes, characterization, and impact
Focus on aggregation: types, causes, characterization, and impact
 

Similar to EMSDVici

Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Shimadzu Scientific Instruments
 
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...KBI Biopharma
 
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...PerkinElmer, Inc.
 
A Simple and an Innovative Gas Chromatography Method to Quantify Isopentane i...
A Simple and an Innovative Gas Chromatography Method to Quantify Isopentane i...A Simple and an Innovative Gas Chromatography Method to Quantify Isopentane i...
A Simple and an Innovative Gas Chromatography Method to Quantify Isopentane i...IOSR Journals
 
5-MEO-DALT produces short-lived psychedelic effects
5-MEO-DALT produces short-lived psychedelic effects5-MEO-DALT produces short-lived psychedelic effects
5-MEO-DALT produces short-lived psychedelic effectsBRC SHOP
 
Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...Kate?ina Svobodov
 
High throughput, data independent acquisition for qualitative and quantitativ...
High throughput, data independent acquisition for qualitative and quantitativ...High throughput, data independent acquisition for qualitative and quantitativ...
High throughput, data independent acquisition for qualitative and quantitativ...SCIEX
 
Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...
Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...
Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...SCIEX
 
Journal Combinatorial Chemistry 2006 v8 p820
Journal Combinatorial Chemistry 2006 v8 p820Journal Combinatorial Chemistry 2006 v8 p820
Journal Combinatorial Chemistry 2006 v8 p820Peter Tidswell
 
Revisions to EPA Method 624 for Analysis of VOCs by GCMS
Revisions to EPA Method 624 for Analysis of VOCs by GCMSRevisions to EPA Method 624 for Analysis of VOCs by GCMS
Revisions to EPA Method 624 for Analysis of VOCs by GCMSShimadzu Scientific Instruments
 
PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...
PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...
PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...PerkinElmer, Inc.
 
An algorithm for detection of tuberculosis bacilli in Ziehl-Neelsen sputum sm...
An algorithm for detection of tuberculosis bacilli in Ziehl-Neelsen sputum sm...An algorithm for detection of tuberculosis bacilli in Ziehl-Neelsen sputum sm...
An algorithm for detection of tuberculosis bacilli in Ziehl-Neelsen sputum sm...IJECEIAES
 
A novel algorithm for detection of tuberculosis bacilli in sputum smear fluor...
A novel algorithm for detection of tuberculosis bacilli in sputum smear fluor...A novel algorithm for detection of tuberculosis bacilli in sputum smear fluor...
A novel algorithm for detection of tuberculosis bacilli in sputum smear fluor...IJECEIAES
 
Quality assurance part_2
Quality assurance part_2Quality assurance part_2
Quality assurance part_2ThorikulHuda2
 
A Novel Method to Improve Measurement Results of Flame Photometry Using Image...
A Novel Method to Improve Measurement Results of Flame Photometry Using Image...A Novel Method to Improve Measurement Results of Flame Photometry Using Image...
A Novel Method to Improve Measurement Results of Flame Photometry Using Image...CSCJournals
 
FL52308_E 0212M_GCMS_crops_REVISED FINAL
FL52308_E 0212M_GCMS_crops_REVISED FINALFL52308_E 0212M_GCMS_crops_REVISED FINAL
FL52308_E 0212M_GCMS_crops_REVISED FINALJennifer Christianson
 

Similar to EMSDVici (20)

Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
 
Enhancing Volatile Organic Compounds in Water
Enhancing Volatile Organic Compounds in Water Enhancing Volatile Organic Compounds in Water
Enhancing Volatile Organic Compounds in Water
 
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
New Software Methods Enhance Sedimentation Velocity Analysis of Protein Aggre...
 
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
 
A Simple and an Innovative Gas Chromatography Method to Quantify Isopentane i...
A Simple and an Innovative Gas Chromatography Method to Quantify Isopentane i...A Simple and an Innovative Gas Chromatography Method to Quantify Isopentane i...
A Simple and an Innovative Gas Chromatography Method to Quantify Isopentane i...
 
5-MEO-DALT produces short-lived psychedelic effects
5-MEO-DALT produces short-lived psychedelic effects5-MEO-DALT produces short-lived psychedelic effects
5-MEO-DALT produces short-lived psychedelic effects
 
Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...Multi-residue pesticide analysis of food samples using acetonitrile extractio...
Multi-residue pesticide analysis of food samples using acetonitrile extractio...
 
HPLC2005
HPLC2005HPLC2005
HPLC2005
 
High throughput, data independent acquisition for qualitative and quantitativ...
High throughput, data independent acquisition for qualitative and quantitativ...High throughput, data independent acquisition for qualitative and quantitativ...
High throughput, data independent acquisition for qualitative and quantitativ...
 
Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...
Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...
Signature Peptide MRM Optimization Made Easy for Therapeutic Protein and Pept...
 
ISES 2015 Max Mascelloni Cell exposure
ISES 2015 Max Mascelloni Cell exposureISES 2015 Max Mascelloni Cell exposure
ISES 2015 Max Mascelloni Cell exposure
 
Journal Combinatorial Chemistry 2006 v8 p820
Journal Combinatorial Chemistry 2006 v8 p820Journal Combinatorial Chemistry 2006 v8 p820
Journal Combinatorial Chemistry 2006 v8 p820
 
Revisions to EPA Method 624 for Analysis of VOCs by GCMS
Revisions to EPA Method 624 for Analysis of VOCs by GCMSRevisions to EPA Method 624 for Analysis of VOCs by GCMS
Revisions to EPA Method 624 for Analysis of VOCs by GCMS
 
PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...
PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...
PerkinElmer: Whole Tablet Measurements Using the Frontier Tablet Autosampler ...
 
An algorithm for detection of tuberculosis bacilli in Ziehl-Neelsen sputum sm...
An algorithm for detection of tuberculosis bacilli in Ziehl-Neelsen sputum sm...An algorithm for detection of tuberculosis bacilli in Ziehl-Neelsen sputum sm...
An algorithm for detection of tuberculosis bacilli in Ziehl-Neelsen sputum sm...
 
Acamprosate Analytical
Acamprosate AnalyticalAcamprosate Analytical
Acamprosate Analytical
 
A novel algorithm for detection of tuberculosis bacilli in sputum smear fluor...
A novel algorithm for detection of tuberculosis bacilli in sputum smear fluor...A novel algorithm for detection of tuberculosis bacilli in sputum smear fluor...
A novel algorithm for detection of tuberculosis bacilli in sputum smear fluor...
 
Quality assurance part_2
Quality assurance part_2Quality assurance part_2
Quality assurance part_2
 
A Novel Method to Improve Measurement Results of Flame Photometry Using Image...
A Novel Method to Improve Measurement Results of Flame Photometry Using Image...A Novel Method to Improve Measurement Results of Flame Photometry Using Image...
A Novel Method to Improve Measurement Results of Flame Photometry Using Image...
 
FL52308_E 0212M_GCMS_crops_REVISED FINAL
FL52308_E 0212M_GCMS_crops_REVISED FINALFL52308_E 0212M_GCMS_crops_REVISED FINAL
FL52308_E 0212M_GCMS_crops_REVISED FINAL
 

EMSDVici

  • 1. Enhanced Simultaneous Total ion and Selective Multi-Ion GC/MS Using a Combination of New Software and Low-Bleed Columns Martin Okiro and Robert Wohleb VICI Gig Harbor Group Inc, P.O. Box 2657, Gig Harbor, WA 98335 Key Words: GC/MS; Enhanced MSD (E-MSD); simultaneous single ion monitoring Summary A new mass spectrophotometer (MS) data acquisition system comprised of hardware, firmware and software has been developed to tune, calibrate and acquire data on single or quadrupole mass spectrometers. The data acquisition routine is so flexible and powerful that several data collection methods can be utilized during the course of a single run. For instance, total ion chromatography and single-ion monitoring can be accomplished simultaneously. Consequently, with a single injection, a total ion chromatogram (TIC) can be generated as well as additional chromatograms each targeting up to 32 individual ions. A combination of this enhanced MSD and exceptionally low-bleed columns results in a powerful new analytical tool allowing detection limit improvement by a factor of ten on EPA method 8270 semivolatiles. Results further show that the TIC’s component responses remains uncompromised and degree of confidence in compound identification is improved. 1 Introduction The use of gas chromatography coupled with mass spectrometer (GC/MS) for the evaluation of complex mixtures, such as those targeted by EPA Methods 8270 and 624 for the identification and quantitation of semivolatile and volatile organic compounds respectively, has been amply documented in the literature [1,2]. The procedure typically involves the generation of a total ion chromatogram (TIC). When the target compound levels are below the normal detection limits, the EPA allows separate selected ion monitoring (SIM), but the practice is expressly discouraged due to vastly reduced confidence in compound identification, except in cases where multiple ions are monitored for each compound. This paper describes a novel system that allows for the simultaneous acquisition of scan and SIMs. Since the data is acquired simultaneously, the uncertainty in compound identification is minimized and sample analysis time is reduced. The E-MSD data acquisition system employs a combination of hardware, firmware and software to tune, calibrate and acquire data on single or quadrupole mass spectrophotometers. The data acquisition system allows the setup of several data collection methods or experiments for each sample injection. Thus, you can for instance acquire both a Scan and multi-ion SIM from the same run thereby eliminating injection port bias. A further benefit of the system is the availability of an enhanced electrometer option with an improved real dynamic range. GC/MS chromatograms of semi-volatile mixtures were obtained using the normal scan mode of operation and the simultaneous scan/SIM mode. Chromatogram characteristics were examined to compare the two modes of operation.
  • 2. 2 Materials and Methods The instrumentation employed in this study includes a Humble Instruments & Services, Inc. (HISI) Enhanced MSD (5970 Mass Spectrometer with an A.S.k ModCtrl board). Computer hardware and software consists of a PC with HP ChemStation for data processing and MStation hardware and software for the control of the 5970 (A product of Acquisition Solutions k). The GC was a standard HP 5890 equipped with a split splitless injection port. The column was a VB-5 (5% phenyl 95% methyl polysiloxane) 30mx25mmx25µm. All standards were prepared in accordance with SW846 and EPA method 8270. Concentrates were purchased from Restek Corporation and Chem Service. Standards were prepared at levels of 0.5, 1.0, 2.0, 5.0 and 10.0 µg/ml in dichloromethane. Samples were manually injected. The chromatograms were obtained by injecting 2ul of the various semivolatile standards into the GC injection port. The GC conditions were as follows: Injector temp 280o C, splitless; oven initial temp 40o C for 2 minutes, followed by 6o C/min to 320o C. 3 Results and Discussion The total ion chromatogram for a 1ppm semi-volatile standard mixture is shown in Fig.1. Figure 1. Total Ion Chromatogram of semi-volatile Mixture. Selected slices of the total ion chromatograms run in the regular and simultaneous scan/SIM modes are compared in Fig. 2. Since peak heights and shapes are comparable in both modes of operation, method sensitivity is not compromised by the scan/SIM mode of operation even though fewer scans/per second are performed. Standard level is 0.5 ppm
  • 3. Figure 2 A comparison of the TIC obtained from the regular scan mode and a TIC of all targeted ions obtained in the SIM/scan mode (SIM component) of a 1.0 ppm standard is shown in Fig. 3. Targeted ions from the TICs shown in Fig. 3 are displayed in Fig. 4. In Fig. 3, the large peak at 23.7 minutes in the TIC run under scan mode is eliminated when the chromatogram is run under SIM mode, and only the targeted peak at 23.6 minutes remains. In addition, the signal to noise level is improved in the SIM/scan mode.
  • 5. Figure 5 Conclusions Despite the fact that operating the instrument in the multiple experiment mode lowered the number of scans/sec from 0.944 to 0.578 when targeting six ions in the SIM mode, there was no significant loss of sensitivity in the simultaneous scan/SIM run. Calibrations for both experiments were equally linear. In addition, the method detection limit for targeted problematic compounds was enhanced by a factor greater than ten. The Combination of E-MSD software and exceptionally low bleed ValcoBond capillary columns can be employed to significantly improve limits of detection in various GC/MS analytical procedures while also improving the degree of confidence in acquired data. 5 Literature Cited [1] Eichelberger, J.W., Harris, L.E., and Budde, W.L, Anal. Chem. 1975, 47, 995-1000. [2 ] Engel, T.M., et al., “Screening of Semivolatile Organic Compounds for Extractability and Aqueous Stability by SW-846, Method 3510,” Contract 68-03-3224, U.S. Environmental Protection Agency, Environmental Monitoring and Support Laboratory, Cincinnati, OH. June 5, 1987. 6 Acknowledgements The authors would like to thank HISI Analytical Technologies (a division of Humble Instruments & Supplies, Inc. - www.humble-inc.com) for loan of the equipment used in the study. E-MSD is a product of Acquisition Solutions k (A.S.k) . For information on E-MSD software and hardware upgrade go on the web at www.flash.net/~acqsol/