Qualification of analytical
instruments
Prepared by
Bhumi P. Suratiya
M.Pharm (Sem-II)
MQA201T
Guided By
Dr. Prasanna Pradhan
(M.Pharm, PhD)
(Associate professor)
Pharmaceutical Quality Assurance
Pioneer Pharmacy College, Vadodara
Content
2
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
 Introduction
 Validation
 Calibration
 Qualification
 Qualification of analytical instruments:
 UV-Visible spectrophotometer
• DQ, IQ,OQ,PQ
 FTIR
• DQ, IQ,OQ,PQ
 References
Introduction
 Validation: Establishing documented evidence that
provides a high degree of assurance that a process, system,
equipment or assay will consistently produce the desired
results according to predetermined specifications and
quality attributes.
 It is a requirement for Good Manufacturing Practices and other
regulatory requirements.
3
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
 Calibration
Calibration is a necessary component to ensure of the
authenticity of Qualification and Validation.
 Calibration is a process that demonstrates a particular
instrument or device produces results within specified
limits, as compared to those produced by a definite standard
over an appropriate range of measurements.
4
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
 Qualification
 Action of proving and documenting that any equipment,
utilities, and systems actually and consistently leads to the
expected results.
 Qualification should be completed before process
validation is performed.
 Qualification Phases:
 The Qualification phases consist of mainly 5 phases.
 Its start with User Requirement Specification (URS).
 URS is a document which states the laboratory
requirement and technical with operational
requirements that should be met. 5
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
 The other phases consist of
• Design qualification(DQ)
• Installation qualification(IQ)
• Operational qualification(OQ)
• Performance qualification(PQ)
6
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
oDesign Qualification:
Documented evidence which
shows that the plant design agrees
with the design specifications of
the customer.
Operational qualification:
Documented evidence which
shows that all parts of the plant
and equipment work within their
specifications and parameters
Process.
Installation Qualification:
Written evidence is given that all
parts of equipment are installed
according to the equipment
supplier's and purchase
specifications.
o Performance
Qualification: Provides
documented evidence that all
parts of a plant and other
processes produce products of
specified quality under conditions
of normal production for a longer
period of time.
Qualification
7
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Design
Qualification
Installation
Qualification
Operational
Qualification
Performance
Qualification
Timing and Applicability
Prior to purchase of a
new model of
instrument
At installation of each
instrument (new, old, or
existing unqualified)
After installation or
major repair of each
instrument
Periodically at specified intervals for
each instrument
Activities
Assurance of
manufacturer's DQ
Description  Fixed parameters Preventive maintenance and repairs
Assurance of adequate
support availability
from manufacturer
Instrument delivery Establish practices to address
operation, calibration, maintenance,
and change control
Instrument's fitness for
use in laboratory
Utilities/facility  Environment
Assembly and installation
Network and data storage  Secure data storage,
backup, and archive
Installation verification  Instrument function
tests
 Performance checks
Table:- Timing, Applicability, and Activities for Each Phase of Analytical Instrument Qualification: Activities under
each phase are usually performed as indicated in the table. In some cases, it may be more appropriate to combine a given activity or perform
with another phase. Such activities are shown linked (↔). It is more important that the activity is performed, but not so important under which
phase it is performed.
8
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Class Type of class instrument or
equipment
Process of
qualification
A Simple Magnetic stirrer No qualification
B Moderate pH meter/oven
Partial qualification
routine calibration,
performance check
maintenance
C Complex FT-IR/HPLC/GC
Full qualification-all
elements of qualification
including software
validation
 Instruments can generally be classified as belonging to groups A, B, or
C. It should be noted that the same type of instrument can fit into one
or more categories, depending on its intended use.
9
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Qualification of
UV-Visible
spectrophotometer
10
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
 Supplier must provide documented evidence that the product has
been designed, developed and manufactured in a quality
environment e.g. ISO 9001:2000 certification.
 Design Qualification Should be performed before the purchase of
new model of UV Spectrophotometer.
 The documents of design qualification should explain:
 What the user want the instrument to do?
 Examples of functional and operational specifications are:
o Optics: Double beam/ Single beam
o Measurable range: e.g 190-1100 nm
o Wavelength Accuracy: e.g. ±0.5 nm or better
Design Qualification(DQ)
11
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
o Criteria for selection of the vendor
o Warranty and maintenance support
o Discount
o Cost of Annual Maintenance Contract (AMC) after
expiry of standard warranty.
Design Qualification(DQ)
12
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Installation Qualification(IQ)
 IQ is the documented collection of activities necessary to establish that
an instrument is delivered as designed and specified, and is properly
installed in the selected environment, and that this environment is
suitable for the instrument.
 According to USP, the IQ requirements provide evidence that the
hardware and software are properly installed in the desired location.
 The instrument has been checked and verified as undamaged.
 The appropriate documentation has been supplied and it is of correct
issue and uniquely identified by part number, version number and date.
13
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Operational Qualification(OQ)
 OQ is the documented collection of activities necessary to demonstrate
that an instrument will function according to its operational
specification in the selected environment.
 Following are the test performed during Operational . Qualification for
UV Spectrophotometer
Wavelength Accuracy
Control of Absorbance
Resolution
Noise
Stability
Limit of Stray Light
14
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Parameter Test Acceptance criteria
Wavelength
accuracy
• Perform to understand deviation of the wavelength
reading from the known wavelength of the band
• Usually, Holmium oxide solution (4% holmium
oxide in 10% perchloric acid) is used
• Holmium oxide solution gives characteristic peaks
at 241.1. 249.9, 278.1, 287.2. 333.5, 345.4 etc.
• ±1 nm in UV region (200-400
nm). ± 2 nm (400-700 nm)
• Three repeated scans of same
peak standard deviation of the
mean must not exceed 0.5 nm
Stray light • This may be due to poor design or faulty
monochromator or may be because of operator,
usually not problem in new instrument but increases
with age of optics and its degradation.
• Higher absorbance ranges (above 1) are more
susceptible.
• For the range of 300-350 nm, aqueous sodium
nitrite (50 g/L) is scanned
The transmittance of the solution
in a 1cm cell should be less than
0.01 or the absorbance value
should be greater than 2
15
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Parameter Test Acceptance criteria
Resolution
power
• Insufficient resolution in any spectrophotometer may loose some
features of fine spectrum or unable to show difference of
absorbance values between two close wavelengths.
• Measure the ratio of the absorbance of a 0.020% (v/v) solution of
toluene in hexane (UV grade) at the maximum and minimum at
about 269 and 266 nm. respectively, using hexane as the
reference.
• The ratio of the
absorbance at 269 nm and
absorbance at 266 nm
should be greater than 1.5
Noise • Noise originating from light source and electronic systems in UV
spectrophotometer and effects the measurement of absorbance in
both lower and higher wavelengths.
• To estimate the noise of spectrophotometer air is scanned (no
sample) for 10 min in absorption mode and Root Mean Square
(RMS) is calculated and recorded at 500 nm.
• The RMS noise should be
less than 0.001 AU
16
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Parameter Test
Acceptance
criteria
Baseline
flatness
• Baseline is used to nullify the effect of environment in measurement by
removing background noise to get true absorption profile of analyte.
• In double beam UV spectrophotometer baseline correction is usually performed
by keeping solvent in one cuvette and test solution in other. This will
automatically correct the baseline and subtract the background noise by solvent.
• This test is performed by scanning air in the absorbance mode in UV region and
highest and lowest deflections in the absorbance unit are recorded.
• The
measurement is
typically
less than 0.01 A
U
Stability • The changes in the components and age are major factors making instability of
response by UV Spectrophotometer resulting error in readings
• The error caused by this factor may be positive or negative (Positive: more than
actual observed, Negative: less than actual response observed)
• For checking stability air is for 60 mins in absorbance mode particular
wavelength at (generally 340 nm) and deflections in the absorbance are
recorded.
• The deflection
is less than
0.002 AU/ hr
17
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Parameter Test Acceptance criteria
Photometric
accuracy
• Photometric accuracy is determined by comparing the
difference between the measured absorbance of the
reference material and the established value.
• Six replicate measurements of
the 0.006% w/v of the potassium
dichromate solution at 235 257,
313 and 350 nm should be less
than 0.5% RSD.
Linearity • Ideally, the absorbance should increase with increase
in the concentration of analyte in solution
• For this various dilutions of Potassium Dichromate
are prepared ranging from 20, 40, 60, 80, and 100
mg/L in 0.005 M sulfuric acid and scanned by
keeping 0.005 M sulphuric acid in reference cuvette.
• The calibration curve is prepared by measuring
absorbance of each solution and correlation
coefficient is calculated.
• Correlation coefficient r > 0.999
18
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Performance Qualification(PQ)
 Performance of instrument may change gradually over time because of
normal wear of parts, failure or change of its components.
 The purpose of the PQ is to provide evidence that instrument is fit for
its routine use.
 Generally, performance of spectrophotometer measurements are
performed under identical condition for the test specimen and the
reference substance.
19
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Qualification of
FTIR
20
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Instrumentation of FTIR
21
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
 Supplier must provide documented evidence that the product has been
designed, developed and manufactured in a quality environment e.g.
ISO 9001:2000 certification.
 Supplier must provide phone and on-site support in case of defects.
 Supplier must provide Information through the internet on availability
of new firmware upgrades.
Design Qualification(DQ)
22
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Attribute Specifications
Detector range
The optical bench shall include a
DTGS(Deuterated triglycine sulfate)
detector with a 7400-350cm-1.
Compressed air interferometer -
Air cooled standard infrared source -
Spectral resolution
The instrument shall have a spectral
resolution not exceeding than 1.07400-
350cm-1.
The below is a sample of technical attributes that may be considered in
design qualification of an FT-IR spectrometer and may be adapted if
necessary.
23
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Installation Qualification(IQ)
 Equipment is compared as received with purchase order, including
software, accessories, spare parts and consumables.
 Documentation checked for completeness of operating,
maintenance instructions, standard operating procedures for
testing and safety, validation certificates and health and safety
instruments.
 Equipment is checked for any damage.
 The supplier’s instruction for installation is read.
24
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
 The instruments are switched on and ensured that all modules
power up and an electronic self test is performed. Any deviations
are recorded.
 Software is installed on computer by following the manufacturer’s
recommendation.
 Correct software installation is verified.
 A backup copy of software is made.
 Peripherals (e.g. printers and equipment) modules are configured.
 A list with a description of all hardware are identified an made
including drawings where appropriate.
25
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
 A list with a description of all software installed on the
computer is made.
 Equipment's manuals and SOPs are listed.
 An installation report is prepared.
26
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Operational Qualification
List of
parameters
which
should be
checked:-
 Wavelength accuracy
 Wavelength resolution
 Wave number scale
 Detector Energy Ratio
 Signal/noise Ratio
 Zero Test
 Contamination Test
27
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Wavelength accuracy
Test procedure
Acceptance
limits
Test frequency Remarks
Measurement of
polystyrene
spectrum at 1144,
1680, 2167 and
2307 nm. The
results are
compared with
reference values.
±2 nm Daily Standard should
be traceable to
national standard.
28
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Wavelength resolution
Test procedure Acceptance limits Test frequency
Resolution of
polystyrene at 2870/2851
and at 1589/1583 nm.
 T(band 2870 cm-1 - band
2851 cm-1 )18
 T(band 1589 cm-1 - band
1583 cm-1 )12
Daily
29
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Wave number Scale
Test procedure
Transmission
minima(cm-1 )
Acceptable
tolerance(cm-1 )
Recording the spectrum
of a polystyrene film
which has transmission
minima(absorption
maxima) at the wave
numbers given in the
next column.
3060.0 ±1.0
2849.5 ±1.0
1942.9 ±1.0
1601.2 ±1.0
1583.0 ±1.0
1154.5 ±1.0
1028.3 ±1.0 30
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Detector Energy Ratio
Test procedure Acceptance limits
The minimum energy ratio value for at
least one of the following is measured
and it is compared to the vendor’s
specifications:
o Energy at 3990 cm-1 /energy at
2000 cm-1
o Energy at 3400 cm-1 /energy at
1300 cm-1
Energy ratio test specifications vary
for each spectrometer configuration
and must be referred to the
manufacturer’s specifications.
31
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Signal/noise Ratio
Test
procedure
Peak to peak
noise between
RMS(root mean
square) noise
between
Limits(%T)
The maximum
noise level is
recorded for
each of the
regions given
in the next two
columns.
• 4050 cm-1 and
3950 cm-1
• 2050 cm-1 and
1950 cm-1
• 050 cm-1 and
950 cm-1
• 550 cm-1 and
450 cm-1
• 4050 cm-1 and
3950 cm-1
• 2050 cm-1 and
1950 cm-1
• 1050 cm-1 and
950 cm-1
• 550 cm-1 and
450 cm-1
Noise level test
specifications vary
for each
spectrometer
configuration and
must be referred to
the manufacturer’s
specifications.
32
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Zero Test
Test procedure Limits(%T)
 When using a polystyrene film of approximately 35µm in
thickness as standard at the wavelength of 2925 cm-1 and 700
cm-1, almost complete absorption of the irradiated energy can
be observed.
 With this test, the remaining transmission is measured.
 As the maximum absorption can be observed at 700 cm-1
negative values may be observed.
 The objective of the test is to evaluate if despite the fact that
there is almost complete absorption, energy is still detectable.
 Non-valid results are an indication of a non-linear behavior of
the detector and the electronic system.
Results vary
for each
spectrometer
configuration
and must be
referred to the
manufacturer’s
specifications.
33
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Contamination Test
Test procedure Wave number(cm-1 ) Upper limit(A)
 The automated function of the
instrument is used to perform
this test (If automation is not
available then a background
spectrum is recorded).
 This test checks the presence of
peaks that signal a
contamination problem.
3100 -2800 0.1
1800 -1600 0.1
1400 -1100 0.2
34
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Performance Qualification(PQ)
Wavenumber Accuracy
Wavenumber reproducibility
Transmittance reproducibility
35
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Wavenumber accuracy
 The wave number scale is usually calibrated by the use of several
characteristics.
 Wave numbers of a polystyrene film:
• 3060.0 (±1.5)cm-1
• 2849.5 (±1.5)cm-1
• 1942.9 (±1.5)cm-1
• 1601.2 (±1.0)cm-1
• 1583.0 (±1.0)cm-1
• 1154.5 (±1.0)cm-1
• 1028.3(±1.0)cm-1
 The software then
judges whether the
values are within the
allowable range and the
program labels the
results ‘PASS’ if all the
peak numbers are
within the range. 36
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Wavenumber reproducibility
 This program specifies three points to measure the peak wave
numbers.
 Then it obtains the actual peak wave numbers at each point by
measuring the polystyrene film twice.
 It should satisfy 5 cm-1 around 3000 cm-1 of polystyrene
absorption wave number, 1 cm-1 around 1000 cm-1 .
 The software determines whether the difference between each of
two measurements are within the allowable range and it labels
the results ‘PASS’ if they are within the range.
37
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Transmittance reproducibility
 This program specifies peak wave numbers at three points and
the transmittance at each point is measured twice.
 The transmittance reproducibility should satisfy 0.5 %T when
the several points of polystyrene absorption from 3000 cm-1 to
1000 cm-1 are measured twice.
 Then it is determined whether the differences between the two
data are within the range.
 The FT-IR abnormalities or large changes over short term and
long term are assessed by these tests.
38
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
The two parameters checked by this program are:
Energy spectrum test:-power
spectra obtained in the inspection
are compared with reference data
and the spectra are checked for
changes over long periods. 100%T
line spectra are calculated for power
spectra and are measured
continuously in inspection and the
spectra are checked for changes
over short periods.
A.Polystyrene test:- evaluation is
performed using differences
between spectra obtained for
polystyrene film in inspection and
the stored reference data. When the
differences are within the standard,
it passes the first.
39
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
References
1. Gouthami B, Venkateshwarlu G, Venkateshwarlu P. Calibration and
Validation of HPLC, GC and UV-VIS Spectroscopy. Int J Mdm. chem
Appl Sci. 2014,1(4):27-34
2. N.R.Anupriya, “Qualification of UV Visible Spectrophotometry, June
2023, https://www.slideshare.net/AnupriyaNR/qualification-of-
uvvisible-spectrophotometer-ftir-dsc-hplc.
3. Indian Pharmacopoeia Commission , Ghaziabad , Indian Pharmacopoeia
2022, (1): 221-223.
4. Sharma P. Validation in pharmaceutical industry. 2nd ed. Delhi:
Vandana publications; 2013.P:167
40
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
References
5. David Rudd, qualification of analytical equipment, method validation
in pharmaceutical analysis, part 2, (chapter 4), 2005
6. D.Gowrisankar, K. Abbulu, O.Bala souri, K.Sujana, Validation and
Calibration of analytical instruments, 2010; 2(2).
41
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx

More Related Content

What's hot

Validation utility system
Validation utility systemValidation utility system
Validation utility systemArpitSuralkar
 
Pilot plant scale up for parentrals
Pilot plant scale up for parentralsPilot plant scale up for parentrals
Pilot plant scale up for parentralsParag Behura
 
Manufacturing Operations and Controls
Manufacturing Operations and ControlsManufacturing Operations and Controls
Manufacturing Operations and ControlsSayali Powar
 
Quality by design (QbD) and process analytical technology (PAT)
Quality by design (QbD) and process analytical technology (PAT)Quality by design (QbD) and process analytical technology (PAT)
Quality by design (QbD) and process analytical technology (PAT)shreyashChaudhari6
 
QUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTER
QUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTERQUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTER
QUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTERUshaKhanal3
 
IPQC of Sterile Ointment.pptx
IPQC of Sterile Ointment.pptxIPQC of Sterile Ointment.pptx
IPQC of Sterile Ointment.pptxDhruvi50
 
Documentation In Pharmaceutical Industry.pptx
Documentation In Pharmaceutical Industry.pptxDocumentation In Pharmaceutical Industry.pptx
Documentation In Pharmaceutical Industry.pptxsaurabh11102000
 
Validation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steamValidation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steamJp Prakash
 
Quality by Design and Process Analytical Technology
Quality by Design and Process Analytical TechnologyQuality by Design and Process Analytical Technology
Quality by Design and Process Analytical TechnologyMANIKANDAN V
 
Advanced sterile product manufacturing technology
Advanced sterile product manufacturing technologyAdvanced sterile product manufacturing technology
Advanced sterile product manufacturing technologyPriyankaSananse
 
Auditing of vendors and production department
Auditing of vendors and production departmentAuditing of vendors and production department
Auditing of vendors and production departmentArpitSuralkar
 
Qualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemQualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemAkshay Trivedi , Maliba Pharmacy College
 
Auditing of vendors and production department
Auditing of vendors and production departmentAuditing of vendors and production department
Auditing of vendors and production departmentPRANJAY PATIL
 

What's hot (20)

Pharmaceutical validation of water system
Pharmaceutical validation of  water system Pharmaceutical validation of  water system
Pharmaceutical validation of water system
 
Qualification of UV spectrophotometer
Qualification of UV spectrophotometer Qualification of UV spectrophotometer
Qualification of UV spectrophotometer
 
Validation utility system
Validation utility systemValidation utility system
Validation utility system
 
Pilot plant scale up for parentrals
Pilot plant scale up for parentralsPilot plant scale up for parentrals
Pilot plant scale up for parentrals
 
CLEANING METHOD DEVELOPMENT
CLEANING METHOD DEVELOPMENT CLEANING METHOD DEVELOPMENT
CLEANING METHOD DEVELOPMENT
 
Manufacturing Operations and Controls
Manufacturing Operations and ControlsManufacturing Operations and Controls
Manufacturing Operations and Controls
 
Quality by design (QbD) and process analytical technology (PAT)
Quality by design (QbD) and process analytical technology (PAT)Quality by design (QbD) and process analytical technology (PAT)
Quality by design (QbD) and process analytical technology (PAT)
 
Qualification of glassware
Qualification of glasswareQualification of glassware
Qualification of glassware
 
QUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTER
QUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTERQUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTER
QUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTER
 
QUALIFICATION OF FT-IR
QUALIFICATION OF FT-IRQUALIFICATION OF FT-IR
QUALIFICATION OF FT-IR
 
IPQC of Sterile Ointment.pptx
IPQC of Sterile Ointment.pptxIPQC of Sterile Ointment.pptx
IPQC of Sterile Ointment.pptx
 
QUALIFICATION OF HPTLC,HPTLC
QUALIFICATION  OF HPTLC,HPTLCQUALIFICATION  OF HPTLC,HPTLC
QUALIFICATION OF HPTLC,HPTLC
 
Documentation In Pharmaceutical Industry.pptx
Documentation In Pharmaceutical Industry.pptxDocumentation In Pharmaceutical Industry.pptx
Documentation In Pharmaceutical Industry.pptx
 
Validation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steamValidation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steam
 
Quality by Design and Process Analytical Technology
Quality by Design and Process Analytical TechnologyQuality by Design and Process Analytical Technology
Quality by Design and Process Analytical Technology
 
Advanced sterile product manufacturing technology
Advanced sterile product manufacturing technologyAdvanced sterile product manufacturing technology
Advanced sterile product manufacturing technology
 
Auditing of vendors and production department
Auditing of vendors and production departmentAuditing of vendors and production department
Auditing of vendors and production department
 
Qualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemQualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility System
 
Case study on Out of Specification (OOS).
Case study on Out of Specification (OOS).Case study on Out of Specification (OOS).
Case study on Out of Specification (OOS).
 
Auditing of vendors and production department
Auditing of vendors and production departmentAuditing of vendors and production department
Auditing of vendors and production department
 

Similar to qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx

Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments PUNEET NIRMAL
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCQUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCAnupriyaNR
 
Validation of uv visible spectrophotometer.pptx
Validation of uv visible spectrophotometer.pptxValidation of uv visible spectrophotometer.pptx
Validation of uv visible spectrophotometer.pptxAnupamaCp2
 
Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validationMohiniTawade
 
Qualification of GC & FTIR
Qualification of GC & FTIRQualification of GC & FTIR
Qualification of GC & FTIRMehulJain143
 
Qualification of INFARED SPECTROSCOPY BY HIMAJA.D
Qualification of INFARED SPECTROSCOPY BY HIMAJA.DQualification of INFARED SPECTROSCOPY BY HIMAJA.D
Qualification of INFARED SPECTROSCOPY BY HIMAJA.Dhimaja donthula
 
Qualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETERQualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETERDr.K.Venkateswara raju
 
Wavelength Accuracy.pptx
Wavelength Accuracy.pptxWavelength Accuracy.pptx
Wavelength Accuracy.pptxsumitdekate1
 
Calibration and validation of analytical instruments
Calibration and validation of analytical instrumentsCalibration and validation of analytical instruments
Calibration and validation of analytical instrumentsSolairajan A
 
Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)PRAVADA
 
QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT 1.pptx
QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT 1.pptxQUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT 1.pptx
QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT 1.pptxTaushifulHoque
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER Ankit Kumar
 
qualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationqualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationSohailPattan
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instrumentsSanathoiba Singha
 
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working Procedures
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working ProceduresNon-Destructive Testing (NDT) - Industrial Radiography Normal Working Procedures
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working Proceduresshahar_sayuti
 
validation HPLC and HPTLC D-A star.pptx
validation HPLC and HPTLC D-A star.pptxvalidation HPLC and HPTLC D-A star.pptx
validation HPLC and HPTLC D-A star.pptxDevaPundkar
 
Performance qualification of High performance liquid chromatography
Performance qualification of High performance liquid chromatography Performance qualification of High performance liquid chromatography
Performance qualification of High performance liquid chromatography Atchaya Thalapathy
 

Similar to qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx (20)

Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments
 
Qualification of high performance liquid chromatography
Qualification of high performance liquid chromatographyQualification of high performance liquid chromatography
Qualification of high performance liquid chromatography
 
Qualification
QualificationQualification
Qualification
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCQUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
 
Validation of uv visible spectrophotometer.pptx
Validation of uv visible spectrophotometer.pptxValidation of uv visible spectrophotometer.pptx
Validation of uv visible spectrophotometer.pptx
 
Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validation
 
Qualification of GC & FTIR
Qualification of GC & FTIRQualification of GC & FTIR
Qualification of GC & FTIR
 
Qualification of INFARED SPECTROSCOPY BY HIMAJA.D
Qualification of INFARED SPECTROSCOPY BY HIMAJA.DQualification of INFARED SPECTROSCOPY BY HIMAJA.D
Qualification of INFARED SPECTROSCOPY BY HIMAJA.D
 
Qualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETERQualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETER
 
Wavelength Accuracy.pptx
Wavelength Accuracy.pptxWavelength Accuracy.pptx
Wavelength Accuracy.pptx
 
Calibration and validation of analytical instruments
Calibration and validation of analytical instrumentsCalibration and validation of analytical instruments
Calibration and validation of analytical instruments
 
Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)
 
QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT 1.pptx
QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT 1.pptxQUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT 1.pptx
QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT 1.pptx
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER
 
qualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationqualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validation
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instruments
 
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working Procedures
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working ProceduresNon-Destructive Testing (NDT) - Industrial Radiography Normal Working Procedures
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working Procedures
 
Qualification of HPLC
Qualification of HPLCQualification of HPLC
Qualification of HPLC
 
validation HPLC and HPTLC D-A star.pptx
validation HPLC and HPTLC D-A star.pptxvalidation HPLC and HPTLC D-A star.pptx
validation HPLC and HPTLC D-A star.pptx
 
Performance qualification of High performance liquid chromatography
Performance qualification of High performance liquid chromatography Performance qualification of High performance liquid chromatography
Performance qualification of High performance liquid chromatography
 

Recently uploaded

9711147426✨Call In girls Gurgaon Sector 31. SCO 25 escort service
9711147426✨Call In girls Gurgaon Sector 31. SCO 25 escort service9711147426✨Call In girls Gurgaon Sector 31. SCO 25 escort service
9711147426✨Call In girls Gurgaon Sector 31. SCO 25 escort servicejennyeacort
 
VIP High Class Call Girls Jamshedpur Anushka 8250192130 Independent Escort Se...
VIP High Class Call Girls Jamshedpur Anushka 8250192130 Independent Escort Se...VIP High Class Call Girls Jamshedpur Anushka 8250192130 Independent Escort Se...
VIP High Class Call Girls Jamshedpur Anushka 8250192130 Independent Escort Se...Suhani Kapoor
 
办理学位证纽约大学毕业证(NYU毕业证书)原版一比一
办理学位证纽约大学毕业证(NYU毕业证书)原版一比一办理学位证纽约大学毕业证(NYU毕业证书)原版一比一
办理学位证纽约大学毕业证(NYU毕业证书)原版一比一fhwihughh
 
Industrialised data - the key to AI success.pdf
Industrialised data - the key to AI success.pdfIndustrialised data - the key to AI success.pdf
Industrialised data - the key to AI success.pdfLars Albertsson
 
꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call
꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call
꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Callshivangimorya083
 
INTERNSHIP ON PURBASHA COMPOSITE TEX LTD
INTERNSHIP ON PURBASHA COMPOSITE TEX LTDINTERNSHIP ON PURBASHA COMPOSITE TEX LTD
INTERNSHIP ON PURBASHA COMPOSITE TEX LTDRafezzaman
 
办理学位证中佛罗里达大学毕业证,UCF成绩单原版一比一
办理学位证中佛罗里达大学毕业证,UCF成绩单原版一比一办理学位证中佛罗里达大学毕业证,UCF成绩单原版一比一
办理学位证中佛罗里达大学毕业证,UCF成绩单原版一比一F sss
 
Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...
Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...
Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...ThinkInnovation
 
代办国外大学文凭《原版美国UCLA文凭证书》加州大学洛杉矶分校毕业证制作成绩单修改
代办国外大学文凭《原版美国UCLA文凭证书》加州大学洛杉矶分校毕业证制作成绩单修改代办国外大学文凭《原版美国UCLA文凭证书》加州大学洛杉矶分校毕业证制作成绩单修改
代办国外大学文凭《原版美国UCLA文凭证书》加州大学洛杉矶分校毕业证制作成绩单修改atducpo
 
Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝
Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝
Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝soniya singh
 
Call Girls In Dwarka 9654467111 Escorts Service
Call Girls In Dwarka 9654467111 Escorts ServiceCall Girls In Dwarka 9654467111 Escorts Service
Call Girls In Dwarka 9654467111 Escorts ServiceSapana Sha
 
Amazon TQM (2) Amazon TQM (2)Amazon TQM (2).pptx
Amazon TQM (2) Amazon TQM (2)Amazon TQM (2).pptxAmazon TQM (2) Amazon TQM (2)Amazon TQM (2).pptx
Amazon TQM (2) Amazon TQM (2)Amazon TQM (2).pptxAbdelrhman abooda
 
Data Science Jobs and Salaries Analysis.pptx
Data Science Jobs and Salaries Analysis.pptxData Science Jobs and Salaries Analysis.pptx
Data Science Jobs and Salaries Analysis.pptxFurkanTasci3
 
Brighton SEO | April 2024 | Data Storytelling
Brighton SEO | April 2024 | Data StorytellingBrighton SEO | April 2024 | Data Storytelling
Brighton SEO | April 2024 | Data StorytellingNeil Barnes
 
Call Girls In Mahipalpur O9654467111 Escorts Service
Call Girls In Mahipalpur O9654467111  Escorts ServiceCall Girls In Mahipalpur O9654467111  Escorts Service
Call Girls In Mahipalpur O9654467111 Escorts ServiceSapana Sha
 
High Class Call Girls Noida Sector 39 Aarushi 🔝8264348440🔝 Independent Escort...
High Class Call Girls Noida Sector 39 Aarushi 🔝8264348440🔝 Independent Escort...High Class Call Girls Noida Sector 39 Aarushi 🔝8264348440🔝 Independent Escort...
High Class Call Girls Noida Sector 39 Aarushi 🔝8264348440🔝 Independent Escort...soniya singh
 
Customer Service Analytics - Make Sense of All Your Data.pptx
Customer Service Analytics - Make Sense of All Your Data.pptxCustomer Service Analytics - Make Sense of All Your Data.pptx
Customer Service Analytics - Make Sense of All Your Data.pptxEmmanuel Dauda
 

Recently uploaded (20)

VIP Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Boo...
VIP Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Boo...VIP Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Boo...
VIP Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Boo...
 
9711147426✨Call In girls Gurgaon Sector 31. SCO 25 escort service
9711147426✨Call In girls Gurgaon Sector 31. SCO 25 escort service9711147426✨Call In girls Gurgaon Sector 31. SCO 25 escort service
9711147426✨Call In girls Gurgaon Sector 31. SCO 25 escort service
 
VIP High Class Call Girls Jamshedpur Anushka 8250192130 Independent Escort Se...
VIP High Class Call Girls Jamshedpur Anushka 8250192130 Independent Escort Se...VIP High Class Call Girls Jamshedpur Anushka 8250192130 Independent Escort Se...
VIP High Class Call Girls Jamshedpur Anushka 8250192130 Independent Escort Se...
 
办理学位证纽约大学毕业证(NYU毕业证书)原版一比一
办理学位证纽约大学毕业证(NYU毕业证书)原版一比一办理学位证纽约大学毕业证(NYU毕业证书)原版一比一
办理学位证纽约大学毕业证(NYU毕业证书)原版一比一
 
Decoding Loan Approval: Predictive Modeling in Action
Decoding Loan Approval: Predictive Modeling in ActionDecoding Loan Approval: Predictive Modeling in Action
Decoding Loan Approval: Predictive Modeling in Action
 
Industrialised data - the key to AI success.pdf
Industrialised data - the key to AI success.pdfIndustrialised data - the key to AI success.pdf
Industrialised data - the key to AI success.pdf
 
꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call
꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call
꧁❤ Greater Noida Call Girls Delhi ❤꧂ 9711199171 ☎️ Hard And Sexy Vip Call
 
INTERNSHIP ON PURBASHA COMPOSITE TEX LTD
INTERNSHIP ON PURBASHA COMPOSITE TEX LTDINTERNSHIP ON PURBASHA COMPOSITE TEX LTD
INTERNSHIP ON PURBASHA COMPOSITE TEX LTD
 
办理学位证中佛罗里达大学毕业证,UCF成绩单原版一比一
办理学位证中佛罗里达大学毕业证,UCF成绩单原版一比一办理学位证中佛罗里达大学毕业证,UCF成绩单原版一比一
办理学位证中佛罗里达大学毕业证,UCF成绩单原版一比一
 
Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...
Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...
Predictive Analysis - Using Insight-informed Data to Determine Factors Drivin...
 
代办国外大学文凭《原版美国UCLA文凭证书》加州大学洛杉矶分校毕业证制作成绩单修改
代办国外大学文凭《原版美国UCLA文凭证书》加州大学洛杉矶分校毕业证制作成绩单修改代办国外大学文凭《原版美国UCLA文凭证书》加州大学洛杉矶分校毕业证制作成绩单修改
代办国外大学文凭《原版美国UCLA文凭证书》加州大学洛杉矶分校毕业证制作成绩单修改
 
Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝
Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝
Call Girls in Defence Colony Delhi 💯Call Us 🔝8264348440🔝
 
Call Girls In Dwarka 9654467111 Escorts Service
Call Girls In Dwarka 9654467111 Escorts ServiceCall Girls In Dwarka 9654467111 Escorts Service
Call Girls In Dwarka 9654467111 Escorts Service
 
Call Girls in Saket 99530🔝 56974 Escort Service
Call Girls in Saket 99530🔝 56974 Escort ServiceCall Girls in Saket 99530🔝 56974 Escort Service
Call Girls in Saket 99530🔝 56974 Escort Service
 
Amazon TQM (2) Amazon TQM (2)Amazon TQM (2).pptx
Amazon TQM (2) Amazon TQM (2)Amazon TQM (2).pptxAmazon TQM (2) Amazon TQM (2)Amazon TQM (2).pptx
Amazon TQM (2) Amazon TQM (2)Amazon TQM (2).pptx
 
Data Science Jobs and Salaries Analysis.pptx
Data Science Jobs and Salaries Analysis.pptxData Science Jobs and Salaries Analysis.pptx
Data Science Jobs and Salaries Analysis.pptx
 
Brighton SEO | April 2024 | Data Storytelling
Brighton SEO | April 2024 | Data StorytellingBrighton SEO | April 2024 | Data Storytelling
Brighton SEO | April 2024 | Data Storytelling
 
Call Girls In Mahipalpur O9654467111 Escorts Service
Call Girls In Mahipalpur O9654467111  Escorts ServiceCall Girls In Mahipalpur O9654467111  Escorts Service
Call Girls In Mahipalpur O9654467111 Escorts Service
 
High Class Call Girls Noida Sector 39 Aarushi 🔝8264348440🔝 Independent Escort...
High Class Call Girls Noida Sector 39 Aarushi 🔝8264348440🔝 Independent Escort...High Class Call Girls Noida Sector 39 Aarushi 🔝8264348440🔝 Independent Escort...
High Class Call Girls Noida Sector 39 Aarushi 🔝8264348440🔝 Independent Escort...
 
Customer Service Analytics - Make Sense of All Your Data.pptx
Customer Service Analytics - Make Sense of All Your Data.pptxCustomer Service Analytics - Make Sense of All Your Data.pptx
Customer Service Analytics - Make Sense of All Your Data.pptx
 

qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx

  • 1. Qualification of analytical instruments Prepared by Bhumi P. Suratiya M.Pharm (Sem-II) MQA201T Guided By Dr. Prasanna Pradhan (M.Pharm, PhD) (Associate professor) Pharmaceutical Quality Assurance Pioneer Pharmacy College, Vadodara
  • 2. Content 2 B h u m i S u r a t i y a M p h a r m S e m 2  Introduction  Validation  Calibration  Qualification  Qualification of analytical instruments:  UV-Visible spectrophotometer • DQ, IQ,OQ,PQ  FTIR • DQ, IQ,OQ,PQ  References
  • 3. Introduction  Validation: Establishing documented evidence that provides a high degree of assurance that a process, system, equipment or assay will consistently produce the desired results according to predetermined specifications and quality attributes.  It is a requirement for Good Manufacturing Practices and other regulatory requirements. 3 B h u m i S u r a t i y a M p h a r m S e m 2
  • 4.  Calibration Calibration is a necessary component to ensure of the authenticity of Qualification and Validation.  Calibration is a process that demonstrates a particular instrument or device produces results within specified limits, as compared to those produced by a definite standard over an appropriate range of measurements. 4 B h u m i S u r a t i y a M p h a r m S e m 2
  • 5.  Qualification  Action of proving and documenting that any equipment, utilities, and systems actually and consistently leads to the expected results.  Qualification should be completed before process validation is performed.  Qualification Phases:  The Qualification phases consist of mainly 5 phases.  Its start with User Requirement Specification (URS).  URS is a document which states the laboratory requirement and technical with operational requirements that should be met. 5 B h u m i S u r a t i y a M p h a r m S e m 2
  • 6.  The other phases consist of • Design qualification(DQ) • Installation qualification(IQ) • Operational qualification(OQ) • Performance qualification(PQ) 6 B h u m i S u r a t i y a M p h a r m S e m 2
  • 7. oDesign Qualification: Documented evidence which shows that the plant design agrees with the design specifications of the customer. Operational qualification: Documented evidence which shows that all parts of the plant and equipment work within their specifications and parameters Process. Installation Qualification: Written evidence is given that all parts of equipment are installed according to the equipment supplier's and purchase specifications. o Performance Qualification: Provides documented evidence that all parts of a plant and other processes produce products of specified quality under conditions of normal production for a longer period of time. Qualification 7 B h u m i S u r a t i y a M p h a r m S e m 2
  • 8. Design Qualification Installation Qualification Operational Qualification Performance Qualification Timing and Applicability Prior to purchase of a new model of instrument At installation of each instrument (new, old, or existing unqualified) After installation or major repair of each instrument Periodically at specified intervals for each instrument Activities Assurance of manufacturer's DQ Description  Fixed parameters Preventive maintenance and repairs Assurance of adequate support availability from manufacturer Instrument delivery Establish practices to address operation, calibration, maintenance, and change control Instrument's fitness for use in laboratory Utilities/facility  Environment Assembly and installation Network and data storage  Secure data storage, backup, and archive Installation verification  Instrument function tests  Performance checks Table:- Timing, Applicability, and Activities for Each Phase of Analytical Instrument Qualification: Activities under each phase are usually performed as indicated in the table. In some cases, it may be more appropriate to combine a given activity or perform with another phase. Such activities are shown linked (↔). It is more important that the activity is performed, but not so important under which phase it is performed. 8 B h u m i S u r a t i y a M p h a r m S e m 2
  • 9. Class Type of class instrument or equipment Process of qualification A Simple Magnetic stirrer No qualification B Moderate pH meter/oven Partial qualification routine calibration, performance check maintenance C Complex FT-IR/HPLC/GC Full qualification-all elements of qualification including software validation  Instruments can generally be classified as belonging to groups A, B, or C. It should be noted that the same type of instrument can fit into one or more categories, depending on its intended use. 9 B h u m i S u r a t i y a M p h a r m S e m 2
  • 11.  Supplier must provide documented evidence that the product has been designed, developed and manufactured in a quality environment e.g. ISO 9001:2000 certification.  Design Qualification Should be performed before the purchase of new model of UV Spectrophotometer.  The documents of design qualification should explain:  What the user want the instrument to do?  Examples of functional and operational specifications are: o Optics: Double beam/ Single beam o Measurable range: e.g 190-1100 nm o Wavelength Accuracy: e.g. ±0.5 nm or better Design Qualification(DQ) 11 B h u m i S u r a t i y a M p h a r m S e m 2
  • 12. o Criteria for selection of the vendor o Warranty and maintenance support o Discount o Cost of Annual Maintenance Contract (AMC) after expiry of standard warranty. Design Qualification(DQ) 12 B h u m i S u r a t i y a M p h a r m S e m 2
  • 13. Installation Qualification(IQ)  IQ is the documented collection of activities necessary to establish that an instrument is delivered as designed and specified, and is properly installed in the selected environment, and that this environment is suitable for the instrument.  According to USP, the IQ requirements provide evidence that the hardware and software are properly installed in the desired location.  The instrument has been checked and verified as undamaged.  The appropriate documentation has been supplied and it is of correct issue and uniquely identified by part number, version number and date. 13 B h u m i S u r a t i y a M p h a r m S e m 2
  • 14. Operational Qualification(OQ)  OQ is the documented collection of activities necessary to demonstrate that an instrument will function according to its operational specification in the selected environment.  Following are the test performed during Operational . Qualification for UV Spectrophotometer Wavelength Accuracy Control of Absorbance Resolution Noise Stability Limit of Stray Light 14 B h u m i S u r a t i y a M p h a r m S e m 2
  • 15. Parameter Test Acceptance criteria Wavelength accuracy • Perform to understand deviation of the wavelength reading from the known wavelength of the band • Usually, Holmium oxide solution (4% holmium oxide in 10% perchloric acid) is used • Holmium oxide solution gives characteristic peaks at 241.1. 249.9, 278.1, 287.2. 333.5, 345.4 etc. • ±1 nm in UV region (200-400 nm). ± 2 nm (400-700 nm) • Three repeated scans of same peak standard deviation of the mean must not exceed 0.5 nm Stray light • This may be due to poor design or faulty monochromator or may be because of operator, usually not problem in new instrument but increases with age of optics and its degradation. • Higher absorbance ranges (above 1) are more susceptible. • For the range of 300-350 nm, aqueous sodium nitrite (50 g/L) is scanned The transmittance of the solution in a 1cm cell should be less than 0.01 or the absorbance value should be greater than 2 15 B h u m i S u r a t i y a M p h a r m S e m 2
  • 16. Parameter Test Acceptance criteria Resolution power • Insufficient resolution in any spectrophotometer may loose some features of fine spectrum or unable to show difference of absorbance values between two close wavelengths. • Measure the ratio of the absorbance of a 0.020% (v/v) solution of toluene in hexane (UV grade) at the maximum and minimum at about 269 and 266 nm. respectively, using hexane as the reference. • The ratio of the absorbance at 269 nm and absorbance at 266 nm should be greater than 1.5 Noise • Noise originating from light source and electronic systems in UV spectrophotometer and effects the measurement of absorbance in both lower and higher wavelengths. • To estimate the noise of spectrophotometer air is scanned (no sample) for 10 min in absorption mode and Root Mean Square (RMS) is calculated and recorded at 500 nm. • The RMS noise should be less than 0.001 AU 16 B h u m i S u r a t i y a M p h a r m S e m 2
  • 17. Parameter Test Acceptance criteria Baseline flatness • Baseline is used to nullify the effect of environment in measurement by removing background noise to get true absorption profile of analyte. • In double beam UV spectrophotometer baseline correction is usually performed by keeping solvent in one cuvette and test solution in other. This will automatically correct the baseline and subtract the background noise by solvent. • This test is performed by scanning air in the absorbance mode in UV region and highest and lowest deflections in the absorbance unit are recorded. • The measurement is typically less than 0.01 A U Stability • The changes in the components and age are major factors making instability of response by UV Spectrophotometer resulting error in readings • The error caused by this factor may be positive or negative (Positive: more than actual observed, Negative: less than actual response observed) • For checking stability air is for 60 mins in absorbance mode particular wavelength at (generally 340 nm) and deflections in the absorbance are recorded. • The deflection is less than 0.002 AU/ hr 17 B h u m i S u r a t i y a M p h a r m S e m 2
  • 18. Parameter Test Acceptance criteria Photometric accuracy • Photometric accuracy is determined by comparing the difference between the measured absorbance of the reference material and the established value. • Six replicate measurements of the 0.006% w/v of the potassium dichromate solution at 235 257, 313 and 350 nm should be less than 0.5% RSD. Linearity • Ideally, the absorbance should increase with increase in the concentration of analyte in solution • For this various dilutions of Potassium Dichromate are prepared ranging from 20, 40, 60, 80, and 100 mg/L in 0.005 M sulfuric acid and scanned by keeping 0.005 M sulphuric acid in reference cuvette. • The calibration curve is prepared by measuring absorbance of each solution and correlation coefficient is calculated. • Correlation coefficient r > 0.999 18 B h u m i S u r a t i y a M p h a r m S e m 2
  • 19. Performance Qualification(PQ)  Performance of instrument may change gradually over time because of normal wear of parts, failure or change of its components.  The purpose of the PQ is to provide evidence that instrument is fit for its routine use.  Generally, performance of spectrophotometer measurements are performed under identical condition for the test specimen and the reference substance. 19 B h u m i S u r a t i y a M p h a r m S e m 2
  • 22.  Supplier must provide documented evidence that the product has been designed, developed and manufactured in a quality environment e.g. ISO 9001:2000 certification.  Supplier must provide phone and on-site support in case of defects.  Supplier must provide Information through the internet on availability of new firmware upgrades. Design Qualification(DQ) 22 B h u m i S u r a t i y a M p h a r m S e m 2
  • 23. Attribute Specifications Detector range The optical bench shall include a DTGS(Deuterated triglycine sulfate) detector with a 7400-350cm-1. Compressed air interferometer - Air cooled standard infrared source - Spectral resolution The instrument shall have a spectral resolution not exceeding than 1.07400- 350cm-1. The below is a sample of technical attributes that may be considered in design qualification of an FT-IR spectrometer and may be adapted if necessary. 23 B h u m i S u r a t i y a M p h a r m S e m 2
  • 24. Installation Qualification(IQ)  Equipment is compared as received with purchase order, including software, accessories, spare parts and consumables.  Documentation checked for completeness of operating, maintenance instructions, standard operating procedures for testing and safety, validation certificates and health and safety instruments.  Equipment is checked for any damage.  The supplier’s instruction for installation is read. 24 B h u m i S u r a t i y a M p h a r m S e m 2
  • 25.  The instruments are switched on and ensured that all modules power up and an electronic self test is performed. Any deviations are recorded.  Software is installed on computer by following the manufacturer’s recommendation.  Correct software installation is verified.  A backup copy of software is made.  Peripherals (e.g. printers and equipment) modules are configured.  A list with a description of all hardware are identified an made including drawings where appropriate. 25 B h u m i S u r a t i y a M p h a r m S e m 2
  • 26.  A list with a description of all software installed on the computer is made.  Equipment's manuals and SOPs are listed.  An installation report is prepared. 26 B h u m i S u r a t i y a M p h a r m S e m 2
  • 27. Operational Qualification List of parameters which should be checked:-  Wavelength accuracy  Wavelength resolution  Wave number scale  Detector Energy Ratio  Signal/noise Ratio  Zero Test  Contamination Test 27 B h u m i S u r a t i y a M p h a r m S e m 2
  • 28. Wavelength accuracy Test procedure Acceptance limits Test frequency Remarks Measurement of polystyrene spectrum at 1144, 1680, 2167 and 2307 nm. The results are compared with reference values. ±2 nm Daily Standard should be traceable to national standard. 28 B h u m i S u r a t i y a M p h a r m S e m 2
  • 29. Wavelength resolution Test procedure Acceptance limits Test frequency Resolution of polystyrene at 2870/2851 and at 1589/1583 nm.  T(band 2870 cm-1 - band 2851 cm-1 )18  T(band 1589 cm-1 - band 1583 cm-1 )12 Daily 29 B h u m i S u r a t i y a M p h a r m S e m 2
  • 30. Wave number Scale Test procedure Transmission minima(cm-1 ) Acceptable tolerance(cm-1 ) Recording the spectrum of a polystyrene film which has transmission minima(absorption maxima) at the wave numbers given in the next column. 3060.0 ±1.0 2849.5 ±1.0 1942.9 ±1.0 1601.2 ±1.0 1583.0 ±1.0 1154.5 ±1.0 1028.3 ±1.0 30 B h u m i S u r a t i y a M p h a r m S e m 2
  • 31. Detector Energy Ratio Test procedure Acceptance limits The minimum energy ratio value for at least one of the following is measured and it is compared to the vendor’s specifications: o Energy at 3990 cm-1 /energy at 2000 cm-1 o Energy at 3400 cm-1 /energy at 1300 cm-1 Energy ratio test specifications vary for each spectrometer configuration and must be referred to the manufacturer’s specifications. 31 B h u m i S u r a t i y a M p h a r m S e m 2
  • 32. Signal/noise Ratio Test procedure Peak to peak noise between RMS(root mean square) noise between Limits(%T) The maximum noise level is recorded for each of the regions given in the next two columns. • 4050 cm-1 and 3950 cm-1 • 2050 cm-1 and 1950 cm-1 • 050 cm-1 and 950 cm-1 • 550 cm-1 and 450 cm-1 • 4050 cm-1 and 3950 cm-1 • 2050 cm-1 and 1950 cm-1 • 1050 cm-1 and 950 cm-1 • 550 cm-1 and 450 cm-1 Noise level test specifications vary for each spectrometer configuration and must be referred to the manufacturer’s specifications. 32 B h u m i S u r a t i y a M p h a r m S e m 2
  • 33. Zero Test Test procedure Limits(%T)  When using a polystyrene film of approximately 35µm in thickness as standard at the wavelength of 2925 cm-1 and 700 cm-1, almost complete absorption of the irradiated energy can be observed.  With this test, the remaining transmission is measured.  As the maximum absorption can be observed at 700 cm-1 negative values may be observed.  The objective of the test is to evaluate if despite the fact that there is almost complete absorption, energy is still detectable.  Non-valid results are an indication of a non-linear behavior of the detector and the electronic system. Results vary for each spectrometer configuration and must be referred to the manufacturer’s specifications. 33 B h u m i S u r a t i y a M p h a r m S e m 2
  • 34. Contamination Test Test procedure Wave number(cm-1 ) Upper limit(A)  The automated function of the instrument is used to perform this test (If automation is not available then a background spectrum is recorded).  This test checks the presence of peaks that signal a contamination problem. 3100 -2800 0.1 1800 -1600 0.1 1400 -1100 0.2 34 B h u m i S u r a t i y a M p h a r m S e m 2
  • 35. Performance Qualification(PQ) Wavenumber Accuracy Wavenumber reproducibility Transmittance reproducibility 35 B h u m i S u r a t i y a M p h a r m S e m 2
  • 36. Wavenumber accuracy  The wave number scale is usually calibrated by the use of several characteristics.  Wave numbers of a polystyrene film: • 3060.0 (±1.5)cm-1 • 2849.5 (±1.5)cm-1 • 1942.9 (±1.5)cm-1 • 1601.2 (±1.0)cm-1 • 1583.0 (±1.0)cm-1 • 1154.5 (±1.0)cm-1 • 1028.3(±1.0)cm-1  The software then judges whether the values are within the allowable range and the program labels the results ‘PASS’ if all the peak numbers are within the range. 36 B h u m i S u r a t i y a M p h a r m S e m 2
  • 37. Wavenumber reproducibility  This program specifies three points to measure the peak wave numbers.  Then it obtains the actual peak wave numbers at each point by measuring the polystyrene film twice.  It should satisfy 5 cm-1 around 3000 cm-1 of polystyrene absorption wave number, 1 cm-1 around 1000 cm-1 .  The software determines whether the difference between each of two measurements are within the allowable range and it labels the results ‘PASS’ if they are within the range. 37 B h u m i S u r a t i y a M p h a r m S e m 2
  • 38. Transmittance reproducibility  This program specifies peak wave numbers at three points and the transmittance at each point is measured twice.  The transmittance reproducibility should satisfy 0.5 %T when the several points of polystyrene absorption from 3000 cm-1 to 1000 cm-1 are measured twice.  Then it is determined whether the differences between the two data are within the range.  The FT-IR abnormalities or large changes over short term and long term are assessed by these tests. 38 B h u m i S u r a t i y a M p h a r m S e m 2
  • 39. The two parameters checked by this program are: Energy spectrum test:-power spectra obtained in the inspection are compared with reference data and the spectra are checked for changes over long periods. 100%T line spectra are calculated for power spectra and are measured continuously in inspection and the spectra are checked for changes over short periods. A.Polystyrene test:- evaluation is performed using differences between spectra obtained for polystyrene film in inspection and the stored reference data. When the differences are within the standard, it passes the first. 39 B h u m i S u r a t i y a M p h a r m S e m 2
  • 40. References 1. Gouthami B, Venkateshwarlu G, Venkateshwarlu P. Calibration and Validation of HPLC, GC and UV-VIS Spectroscopy. Int J Mdm. chem Appl Sci. 2014,1(4):27-34 2. N.R.Anupriya, “Qualification of UV Visible Spectrophotometry, June 2023, https://www.slideshare.net/AnupriyaNR/qualification-of- uvvisible-spectrophotometer-ftir-dsc-hplc. 3. Indian Pharmacopoeia Commission , Ghaziabad , Indian Pharmacopoeia 2022, (1): 221-223. 4. Sharma P. Validation in pharmaceutical industry. 2nd ed. Delhi: Vandana publications; 2013.P:167 40 B h u m i S u r a t i y a M p h a r m S e m 2
  • 41. References 5. David Rudd, qualification of analytical equipment, method validation in pharmaceutical analysis, part 2, (chapter 4), 2005 6. D.Gowrisankar, K. Abbulu, O.Bala souri, K.Sujana, Validation and Calibration of analytical instruments, 2010; 2(2). 41 B h u m i S u r a t i y a M p h a r m S e m 2

Editor's Notes

  1. Stray light is defined as the detected light of any wavelength that is outside the band width of the wavelength selected.
  2. Resolution power- The resolution of the UV-VIS spectrometer is related to its spectral band width. The smaller the band width the finer the resolution. The SBW depends on the slit width and the dispersive power of the monochromator. Noise - Noise is the measurement affects the accuracy at the both end of the absorbance scale. Photon noise from the light source affects the accuracy of the measurement leads to low absorbance.
  3. Base line flatness - he flat baseline test demonstrates that the ability of the instrument to normalize the light intensity measurement and the spectral out put at different wavelength throughout the spectral range. Stability - The lamp intensity is a function of the lamp age, temperature fluctuation and wavelength of the measurement. These changes can lead to errors in the value of the measurements, over an extended period of time.
  4. Linearity - The linearity dynamic range of measurement is limited by stray light at high absorbance and by noise at low absorbance. The accuracy of the quantification of the sample depends on the precision and linearity of the measurements.
  5. Linearity - The linearity dynamic range of measurement is limited by stray light at high absorbance and by noise at low absorbance. The accuracy of the quantification of the sample depends on the precision and linearity of the measurements.