What is Pulse Oximeter?
Pulse Oximeters are non-Invasive medical
sensors used to continuously measure the
Oxygen saturation of hemoglobin in the blood. It
displays the percentage of blood that is loaded
with oxygen.
Principle of Pulse oximeter
• The principle of pulse oximetry is based on the differential absorption
characteristics of oxygenated and de-oxygenated hemoglobin.
Oxygenated hemoglobin absorbs more infrared light and allows more
red light to pass through. Whereas Deoxygenated hemoglobin absorbs
more red light and allows more infrared light to pass through.
• Each pulse oximeter sensor probe contains two light-emitting diodes
one emitting red light and the other emitting near-infrared light, it also
has a photo-detector. The photo-detector measures the intensity of
transmitted light at each wavelength. And using the differences in the
reading the blood oxygen content is calculated. The probe is placed on
a suitable part of the body, usually a fingertip or ear lobe.
Principle of Pulse Oximeter
Methods for Monitoring Oxygen Saturation in
Blood
• Transmission Method
• Reflectance Method
Transmission Method
• In the transmission method the transmitter i.e. the LED & the receiver
i.e. the photo-detector are placed on opposite sides of the finger. In this
method, this finger will be placed between the LED’s & the
photodetector. When the finger has been placed a part of the light will
be absorbed by the finger and some part will reach the photodetector.
Reflectance Method
• In the Reflective method, the LED and the photo-detector are placed
on the same side i.e. next to each other. In the reflective method, there
will be some fixed light reflection back to the sensor due to the finger.
how is Oxygen Saturation Calculated?
• Sensor consists of two transmitting LEDs a Red led of approximately 650
nm and an infrared led which has a wavelength of 950 nm.
• Oxygenated hemoglobin (RED line) absorbs more infrared light than red
light. Also, the De-Oxygenated hemoglobin (blue line) absorbs more red
light as compared to infrared light. The oxygen content in the blood can be
easily calculated by comparing how much red light (R) is absorbed as
compared to infrared (IR) light.
• Depending on the amount of Oxygenated hemoglobin or De-Oxygenated
hemoglobin the ratio of red light absorbed vs Infrared light (R/IR) absorbed
will change. Typically R/IR ratio of 0.5 equates to approximately 100%
SpO2, a ratio of 1.0 to approximately 82% SpO2, while a ratio of 2.0 equates
to 0% SpO2.
New Microsoft PowerPoint Presentation.pptx
New Microsoft PowerPoint Presentation.pptx
New Microsoft PowerPoint Presentation.pptx

More Related Content

Similar to New Microsoft PowerPoint Presentation.pptx

Pulseoximetry
PulseoximetryPulseoximetry
Pulseoximetry
drriyas03
 

Similar to New Microsoft PowerPoint Presentation.pptx (20)

Pulse oximetry
Pulse oximetryPulse oximetry
Pulse oximetry
 
Pulse oximeter
Pulse  oximeterPulse  oximeter
Pulse oximeter
 
Pulse oximetry ppt
Pulse oximetry pptPulse oximetry ppt
Pulse oximetry ppt
 
Pulse oximetry capnography
Pulse oximetry capnography Pulse oximetry capnography
Pulse oximetry capnography
 
Ppt0000002
Ppt0000002Ppt0000002
Ppt0000002
 
Co32559562
Co32559562Co32559562
Co32559562
 
Pulse Oxymetry
Pulse OxymetryPulse Oxymetry
Pulse Oxymetry
 
Pulseoximetry
PulseoximetryPulseoximetry
Pulseoximetry
 
International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER) International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER)
 
Pulse oximetry
Pulse oximetryPulse oximetry
Pulse oximetry
 
Pulseox[1]
Pulseox[1]Pulseox[1]
Pulseox[1]
 
Pulse oximetry journal
Pulse oximetry journalPulse oximetry journal
Pulse oximetry journal
 
Scientific Concept SpO2 of Pulse Oximeter
Scientific Concept SpO2 of Pulse OximeterScientific Concept SpO2 of Pulse Oximeter
Scientific Concept SpO2 of Pulse Oximeter
 
Pulse oximetry
Pulse oximetryPulse oximetry
Pulse oximetry
 
Ndir gas sensor
Ndir gas sensorNdir gas sensor
Ndir gas sensor
 
photoplethysmogram
photoplethysmogram photoplethysmogram
photoplethysmogram
 
PULSE OXIMETRY
PULSE OXIMETRYPULSE OXIMETRY
PULSE OXIMETRY
 
Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...
Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...
Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...
 
Oximeter
OximeterOximeter
Oximeter
 
Pulse Oxymetry , Inspired & Expired Gas Monitoring
Pulse Oxymetry ,  Inspired & Expired Gas MonitoringPulse Oxymetry ,  Inspired & Expired Gas Monitoring
Pulse Oxymetry , Inspired & Expired Gas Monitoring
 

Recently uploaded

UNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and workingUNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and working
rknatarajan
 
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
dharasingh5698
 
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Christo Ananth
 
AKTU Computer Networks notes --- Unit 3.pdf
AKTU Computer Networks notes ---  Unit 3.pdfAKTU Computer Networks notes ---  Unit 3.pdf
AKTU Computer Networks notes --- Unit 3.pdf
ankushspencer015
 

Recently uploaded (20)

(INDIRA) Call Girl Meerut Call Now 8617697112 Meerut Escorts 24x7
(INDIRA) Call Girl Meerut Call Now 8617697112 Meerut Escorts 24x7(INDIRA) Call Girl Meerut Call Now 8617697112 Meerut Escorts 24x7
(INDIRA) Call Girl Meerut Call Now 8617697112 Meerut Escorts 24x7
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . ppt
 
Online banking management system project.pdf
Online banking management system project.pdfOnline banking management system project.pdf
Online banking management system project.pdf
 
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
 
Vivazz, Mieres Social Housing Design Spain
Vivazz, Mieres Social Housing Design SpainVivazz, Mieres Social Housing Design Spain
Vivazz, Mieres Social Housing Design Spain
 
chapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineeringchapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineering
 
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
 
Glass Ceramics: Processing and Properties
Glass Ceramics: Processing and PropertiesGlass Ceramics: Processing and Properties
Glass Ceramics: Processing and Properties
 
Call for Papers - International Journal of Intelligent Systems and Applicatio...
Call for Papers - International Journal of Intelligent Systems and Applicatio...Call for Papers - International Journal of Intelligent Systems and Applicatio...
Call for Papers - International Journal of Intelligent Systems and Applicatio...
 
UNIT-III FMM. DIMENSIONAL ANALYSIS
UNIT-III FMM.        DIMENSIONAL ANALYSISUNIT-III FMM.        DIMENSIONAL ANALYSIS
UNIT-III FMM. DIMENSIONAL ANALYSIS
 
Unit 1 - Soil Classification and Compaction.pdf
Unit 1 - Soil Classification and Compaction.pdfUnit 1 - Soil Classification and Compaction.pdf
Unit 1 - Soil Classification and Compaction.pdf
 
UNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and workingUNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and working
 
University management System project report..pdf
University management System project report..pdfUniversity management System project report..pdf
University management System project report..pdf
 
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
 
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
 
PVC VS. FIBERGLASS (FRP) GRAVITY SEWER - UNI BELL
PVC VS. FIBERGLASS (FRP) GRAVITY SEWER - UNI BELLPVC VS. FIBERGLASS (FRP) GRAVITY SEWER - UNI BELL
PVC VS. FIBERGLASS (FRP) GRAVITY SEWER - UNI BELL
 
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
 
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordCCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
 
AKTU Computer Networks notes --- Unit 3.pdf
AKTU Computer Networks notes ---  Unit 3.pdfAKTU Computer Networks notes ---  Unit 3.pdf
AKTU Computer Networks notes --- Unit 3.pdf
 
Roadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and RoutesRoadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and Routes
 

New Microsoft PowerPoint Presentation.pptx

  • 1.
  • 2. What is Pulse Oximeter? Pulse Oximeters are non-Invasive medical sensors used to continuously measure the Oxygen saturation of hemoglobin in the blood. It displays the percentage of blood that is loaded with oxygen.
  • 3. Principle of Pulse oximeter • The principle of pulse oximetry is based on the differential absorption characteristics of oxygenated and de-oxygenated hemoglobin. Oxygenated hemoglobin absorbs more infrared light and allows more red light to pass through. Whereas Deoxygenated hemoglobin absorbs more red light and allows more infrared light to pass through.
  • 4. • Each pulse oximeter sensor probe contains two light-emitting diodes one emitting red light and the other emitting near-infrared light, it also has a photo-detector. The photo-detector measures the intensity of transmitted light at each wavelength. And using the differences in the reading the blood oxygen content is calculated. The probe is placed on a suitable part of the body, usually a fingertip or ear lobe. Principle of Pulse Oximeter
  • 5. Methods for Monitoring Oxygen Saturation in Blood • Transmission Method • Reflectance Method
  • 6. Transmission Method • In the transmission method the transmitter i.e. the LED & the receiver i.e. the photo-detector are placed on opposite sides of the finger. In this method, this finger will be placed between the LED’s & the photodetector. When the finger has been placed a part of the light will be absorbed by the finger and some part will reach the photodetector.
  • 7. Reflectance Method • In the Reflective method, the LED and the photo-detector are placed on the same side i.e. next to each other. In the reflective method, there will be some fixed light reflection back to the sensor due to the finger.
  • 8. how is Oxygen Saturation Calculated? • Sensor consists of two transmitting LEDs a Red led of approximately 650 nm and an infrared led which has a wavelength of 950 nm. • Oxygenated hemoglobin (RED line) absorbs more infrared light than red light. Also, the De-Oxygenated hemoglobin (blue line) absorbs more red light as compared to infrared light. The oxygen content in the blood can be easily calculated by comparing how much red light (R) is absorbed as compared to infrared (IR) light. • Depending on the amount of Oxygenated hemoglobin or De-Oxygenated hemoglobin the ratio of red light absorbed vs Infrared light (R/IR) absorbed will change. Typically R/IR ratio of 0.5 equates to approximately 100% SpO2, a ratio of 1.0 to approximately 82% SpO2, while a ratio of 2.0 equates to 0% SpO2.