1. OCT in coronary artery disease
Your light in the dark tunnel
Ahmed ElBorae, MSc, MRCPUK (London), EAPCI
Assistant lecturer of Cardiology, Cairo University
5. OCT concept
• Depends on a rotating near infra-red pulses and interferometer
• Generate picture according to the amplitude and time delay of the reflected
waves
Tissue
Interferometer
Image
Pulses of infrared light
Coherence
Represents “Optical biopsy”
6. Clinical Evidence
CLI-OPCI I and II:
OCT improves outcomes
vs. angiography
OPUS-Class Study
Reliability of OCT measurement
vs. IVUS and angiography
OCT Safety and Efficacy
Non-occlusive OCT study
Past
Present–2015
Future
ILUMIEN IV
OCT vs Angiography Randomized
Outcomes Trial
Other areas under
consideration:
 Non-contrast flush media
 BVS
 Virtual OFR
ILUMIEN I:
Define and evaluate OCT
stent guidance parameters
and determine impact on
physician decision making (84%)
ILUMIEN II
OCT vs. IVUS propensity
matched comparison of stent
expansion.
ILUMIEN III
OCT vs IVUS vs Angiography
prospective randomized trial to
evaluate OCT EEL-guided PCI
Courtesy of Ziad Ali
7. Percutaneous interventional cardiovascular medicine – The PCR-EAPCI Textbook
Francesco Prati1,2,3, Alessandro Sticchi1,2, Evelyn Regar4
OCT Vs IVUS
“Light vs. sound”
X 10
8. Agenda
• Basics
• How to perform an OCT study?
• Algorithmic approach
• Clinical application
• Limitations
• Future direction
9. 1- OCT Catheter preparation
B- Connect “Clockwise”
“DOC” Sterile cover
A- Flush 3 cc 100% contrast
Percutaneous interventional cardiovascular medicine – The PCR-EAPCI Textbook
Francesco Prati1,2,3, Alessandro Sticchi1,2, Evelyn Regar4
10. • Compatible with 6F guide catheter, vessel > 2 mm
• Advance till lens marker distal to area of interest
• Don’t forget nitroglycerin, full heparinization
• Remember blood is the enemy of infrared light
2- Advance the catheter
Percutaneous interventional cardiovascular medicine – The PCR-EAPCI Textbook
Francesco Prati1,2,3, Alessandro Sticchi1,2, Evelyn Regar4
Proximal marker
Lens marker
Pullback length
11. • A- Purge catheter& click on “Calibrate”
• B- Press “enable pullback”
• C- Inject either ( Contrast or Dextran or mix)
rate 3-4 ml/ s
• D- Revise run and remove catheter
3- Start pullback run
•Modes
•The 75 mm Survey Mode – fast but less frame rate
•The 54 mm High Resolution Mode – twice the frame rate
OCTaid,mount Sainai app
52. Thank You
James Fujimoto, MIT, USA
God father of OCT
Gary Tearney, MIT,USA
1st invivo OCT in human
David Huang, MIT, USA
1st OCT imaging invitro 1990