ATOMIC EMISSION SPECTROSCOPY
ATOMIC EMISSION SPECTROSCOPY
 The sample is converted into excited gaseous atoms and ions
which emit light at characteristic wavelengths (wavenumbers).
 The analyte is identified by emission at a known wavelength
(qual. analysis) and its concentration is determined from the
intensity of the emission at that wavelength (quant. analysis).
 Atoms or ions in the gas phase can be excited by high
temperatures or by light absorption.
 A fraction of the atoms or ions lose their excitation energy by
giving off a photon.
 The most intense emission line for a neutral atom is the
resonance line (same line for absorption and emission e.g.
589nm for Na atom).
Atomic Emission Spectrometers.
Emission sources: The source must
a. vaporize the sample,
b. break down all compounds to atoms and ions,
c. and excite the atoms and ions.
• Very high temperatures are required, much higher than in atomic absorption
Methods to atomize and excite the analytes:
a. electrical arcs (low voltage, high current)
b. electrical sparks (high voltage, low current
c. flames have been used
.
Monochromators/ Polychromators
 A polychromator has multiple exit slits with multiple detectors
PMT Because there are many emission lines from the plasma,
often separated by only a few Angstroms, high resolution is
required in monochromators and polychromators.
 Designs are based on an echelle grating operating at high
orders combined with an order-sorting prism
Detectors
• Usually a PMT (multiple PMT’s in a polychromator)
• Both the polychromator + multiple PMTs and the echelle can
monitor concentrations of 10 or more elements from one sample
within a few minutes.
• These devices are used heavily by the metallurgy industry to
monitor composition of alloys.
Internal Standard vs. external standard method
• Because AES can easily perform multi-element
analysis, an internal standard (an element not
naturally found in the samples) is often incorporated
at a constant concentration into each sample.
• To compensate for fluctuation of plasma emission
intensity (caused by fluctuations in plasma
temperature), the emission intensity of each element
line is divided by the emission intensity of an internal
standard.
Internal standard Method
An Internal Standard is a substance that is added in a
constant amount to all samples, blanks and calibration
standards in an analysis.
Calibration involves plotting the ratio of the analyte signal
to the internal standard signal as a function of analyte
concentration of the standards.
This ratio for the samples is then used to obtain their
analyte concentrations from a calibration curve.
Internal standard can compensate for several types of
both random and systematic errors.
Internal Standard Method
 Most convenient when variations in analytical sample size,
position, or matrix limit the precision of a technique.
 Or when the sensitivity of instrument changes with time
Internal Standard Procedure
 Prepare a set of standard solutions for analyte (A) as with the
calibration curve method, but add a constant amount of a
second species (B called internal standard) to each solution.
 Add the same amount of IS to the blank and sample solution.
 Prepare a plot of SA/SB versus [A].
Example: Pb by ICP Emission
Each Pb solution contains 100 ppm Cu as IS
[Pb]
(ppm) Pb Cu Pb/Cu
20 112 1347 0.083
40 243 1527 0.159
60 326 1383 0.236
80 355 1135 0.313
100 558 1440 0.388
Signal
No Internal Standard Correction
y = 5.02x + 17.6
R² = 0.9413
0
100
200
300
400
500
600
0
20
40
60
80
100
120
Internal Standard Correction
y = 0.0038x + 0.0066
R² = 1
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
0.45
0
20
40
60
80
100
120
unknown sample of Pb (containing100 ppm Cu
as IS) gives a ratio of 0.241.
Conc. of Pb in the unknown = 78.3 ppm
Back to sources: flame
Flame photometer:
 flame emission spectrometry is used for the
quantitative determination of small amounts of metal
salts in solution.
 Flam emission is commonly used in clinical labs for
determination the concentration of Na or K ions in
blood.
In this method:
 A. sample solution sprayed or aspirated as fine mist
into flame;
 B. conversion of sample solution into an aerosol by
atomizer (scent spray)
principle.
 Then Heat of the flame vaporizes sample constituents.
Still no chemical change. By heat of the flame + action of
the reducing gas (fuel), molecules & ions of the sample
species are decomposed and reduced to give ATOMS.
Na+ + e- --> Na
 Heat of the flame causes excitation of some atoms into
higher electronic states. Excited atoms revert to ground
state by emission of light energy, h, of characteristic
wavelength; measured by detector.
 The reading that the instrument gives is used to calculate
the concentration of the element to be determined.
The Sherwood Model 41O
Experimental Aspects of Flame Photometry
• Propane-air or natural gas-air gives good flame - strong heat,
minimal background light emission.
• But always need to run a solvent blank for setting zero emission.
• Solutions diluted to fall within linear part of emission curve.
• Anion and cation interference effects can cause errors
(enhancement or suppression). "Radiation buffer" for dilution of
standards and samples to swamp out inconsistencies.
• Internal standard (lithium) useful to counter random flame
instability and random dilution errors.
Plasma not flame
• ICP torch is the most important.
• Like the flame AA experiment, the sample is
introduced into the instrument as a solution fed into a
nebulizer, creating a fine spray of droplets in a flowing
gas stream
Inductively coupled plasma (ICP):
• A plasma (partially ionized gas) of argon is formed and
sustained by intense electric and magnetic fields inside a coiled
radio antenna (27 MHz, 1-2 kW power).
A triple-tube quartz torch is used to:
• (a) generate annular plasma,
• (b) introduce the sample as droplets in an Ar gas stream.
• Temperatures are about 10,000 K at hottest part of plasma.
Observation is made 1-3 cm above hottest part; local
temperatures are still high (ca. 6000 K).
Advantages:
(a) a very wide dynamic range (3 to 6 orders-of-magnitude) on
calibration curves;
(b) stable operation over long periods of time,
(c) almost no chemical interferences due to the very high
temperatures.
Disadvantages:
(a) expensive to buy and operate;
(b) rapid consumption of high purity Ar gas (several L/min).
Atomic Emission Spectroscopy (Chapter10.pptx
Atomic Emission Spectroscopy (Chapter10.pptx

More Related Content

Similar to Atomic Emission Spectroscopy (Chapter10.pptx

1606732404-atomic-absorption-emission-2.ppt
1606732404-atomic-absorption-emission-2.ppt1606732404-atomic-absorption-emission-2.ppt
1606732404-atomic-absorption-emission-2.pptMuhammadJavedIqbal40
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopyMahendra G S
 
1606732404-atomic-absorption-emission-2.ppt
1606732404-atomic-absorption-emission-2.ppt1606732404-atomic-absorption-emission-2.ppt
1606732404-atomic-absorption-emission-2.pptRanaNoman21
 
Mridu-ICP-MS.pptx
Mridu-ICP-MS.pptxMridu-ICP-MS.pptx
Mridu-ICP-MS.pptxSweetuCutie
 
Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.
Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.
Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.Acmas Technologies Pvt. Ltd.
 
Principles of Inductively coupled plasma spectrometry.pptx
Principles of Inductively coupled plasma spectrometry.pptxPrinciples of Inductively coupled plasma spectrometry.pptx
Principles of Inductively coupled plasma spectrometry.pptxShweta Pandey
 
Atomic absorption spectrophotometer
Atomic absorption spectrophotometerAtomic absorption spectrophotometer
Atomic absorption spectrophotometerTapeshwar Yadav
 
A presentation ,analytical methods-for-determination-of-metals-in-environment...
A presentation ,analytical methods-for-determination-of-metals-in-environment...A presentation ,analytical methods-for-determination-of-metals-in-environment...
A presentation ,analytical methods-for-determination-of-metals-in-environment...Adnan Sohail
 
Atomic Fluorescence Spectroscopy (AFS)
Atomic Fluorescence Spectroscopy (AFS)Atomic Fluorescence Spectroscopy (AFS)
Atomic Fluorescence Spectroscopy (AFS)Sajjad Ullah
 
atomic absorption spectroscopy
atomic absorption spectroscopyatomic absorption spectroscopy
atomic absorption spectroscopysharmasuriti
 
flame emission and AAS.pptx
flame emission and AAS.pptxflame emission and AAS.pptx
flame emission and AAS.pptxDeepa B
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopySwatiChoudhary51
 
Atomic Absorption Spectroscopy by Yousef Bahrami
Atomic Absorption Spectroscopy by Yousef BahramiAtomic Absorption Spectroscopy by Yousef Bahrami
Atomic Absorption Spectroscopy by Yousef BahramiYousefBahrami
 
Bahrami presentation- AAS
Bahrami presentation- AASBahrami presentation- AAS
Bahrami presentation- AASYousefBahrami
 
Atomic absorption & flame emission spectrophotometry by Dr. Anurag Yadav
Atomic absorption & flame emission spectrophotometry by Dr. Anurag YadavAtomic absorption & flame emission spectrophotometry by Dr. Anurag Yadav
Atomic absorption & flame emission spectrophotometry by Dr. Anurag YadavDr Anurag Yadav
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopyAjay Dominic
 

Similar to Atomic Emission Spectroscopy (Chapter10.pptx (20)

1606732404-atomic-absorption-emission-2.ppt
1606732404-atomic-absorption-emission-2.ppt1606732404-atomic-absorption-emission-2.ppt
1606732404-atomic-absorption-emission-2.ppt
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopy
 
1606732404-atomic-absorption-emission-2.ppt
1606732404-atomic-absorption-emission-2.ppt1606732404-atomic-absorption-emission-2.ppt
1606732404-atomic-absorption-emission-2.ppt
 
Mridu-ICP-MS.pptx
Mridu-ICP-MS.pptxMridu-ICP-MS.pptx
Mridu-ICP-MS.pptx
 
Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.
Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.
Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.
 
Principles of Inductively coupled plasma spectrometry.pptx
Principles of Inductively coupled plasma spectrometry.pptxPrinciples of Inductively coupled plasma spectrometry.pptx
Principles of Inductively coupled plasma spectrometry.pptx
 
Atomic absorption spectrophotometer
Atomic absorption spectrophotometerAtomic absorption spectrophotometer
Atomic absorption spectrophotometer
 
A presentation ,analytical methods-for-determination-of-metals-in-environment...
A presentation ,analytical methods-for-determination-of-metals-in-environment...A presentation ,analytical methods-for-determination-of-metals-in-environment...
A presentation ,analytical methods-for-determination-of-metals-in-environment...
 
Atomic Fluorescence Spectroscopy (AFS)
Atomic Fluorescence Spectroscopy (AFS)Atomic Fluorescence Spectroscopy (AFS)
Atomic Fluorescence Spectroscopy (AFS)
 
atomic absorption spectroscopy
atomic absorption spectroscopyatomic absorption spectroscopy
atomic absorption spectroscopy
 
AES Plasma
AES PlasmaAES Plasma
AES Plasma
 
Atomic spectroscopy
Atomic spectroscopyAtomic spectroscopy
Atomic spectroscopy
 
ICP-AES.pptx
ICP-AES.pptxICP-AES.pptx
ICP-AES.pptx
 
flame emission and AAS.pptx
flame emission and AAS.pptxflame emission and AAS.pptx
flame emission and AAS.pptx
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Atomic Absorption Spectroscopy by Yousef Bahrami
Atomic Absorption Spectroscopy by Yousef BahramiAtomic Absorption Spectroscopy by Yousef Bahrami
Atomic Absorption Spectroscopy by Yousef Bahrami
 
Bahrami presentation- AAS
Bahrami presentation- AASBahrami presentation- AAS
Bahrami presentation- AAS
 
Atomic absorption & flame emission spectrophotometry by Dr. Anurag Yadav
Atomic absorption & flame emission spectrophotometry by Dr. Anurag YadavAtomic absorption & flame emission spectrophotometry by Dr. Anurag Yadav
Atomic absorption & flame emission spectrophotometry by Dr. Anurag Yadav
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Icp ms
Icp msIcp ms
Icp ms
 

More from AhmadHashlamon

02slidLarge value of face area Large value of face area
02slidLarge value of face area Large value of face area02slidLarge value of face area Large value of face area
02slidLarge value of face area Large value of face areaAhmadHashlamon
 
kinectic_and_drug_stability microemulsions
kinectic_and_drug_stability microemulsionskinectic_and_drug_stability microemulsions
kinectic_and_drug_stability microemulsionsAhmadHashlamon
 
04slidemicroemulsions microemulsions microemulsions microemulsions microemuls...
04slidemicroemulsions microemulsions microemulsions microemulsions microemuls...04slidemicroemulsions microemulsions microemulsions microemulsions microemuls...
04slidemicroemulsions microemulsions microemulsions microemulsions microemuls...AhmadHashlamon
 
skeletal musclemicroemulsions microemulsions microemulsions microemulsions mi...
skeletal musclemicroemulsions microemulsions microemulsions microemulsions mi...skeletal musclemicroemulsions microemulsions microemulsions microemulsions mi...
skeletal musclemicroemulsions microemulsions microemulsions microemulsions mi...AhmadHashlamon
 
microemulsions microemulsions microemulsions microemulsions
microemulsions microemulsions microemulsions microemulsionsmicroemulsions microemulsions microemulsions microemulsions
microemulsions microemulsions microemulsions microemulsionsAhmadHashlamon
 
Herbal_Medicine_Developmggggffgfent.pptx
Herbal_Medicine_Developmggggffgfent.pptxHerbal_Medicine_Developmggggffgfent.pptx
Herbal_Medicine_Developmggggffgfent.pptxAhmadHashlamon
 
performulation studies.pptx
performulation studies.pptxperformulation studies.pptx
performulation studies.pptxAhmadHashlamon
 
Oxidation-Reduction-Reactions-and-Titrations.ppt
Oxidation-Reduction-Reactions-and-Titrations.pptOxidation-Reduction-Reactions-and-Titrations.ppt
Oxidation-Reduction-Reactions-and-Titrations.pptAhmadHashlamon
 
Complexometric-Titrations part 2.ppt
Complexometric-Titrations part 2.pptComplexometric-Titrations part 2.ppt
Complexometric-Titrations part 2.pptAhmadHashlamon
 
Data-Handling part 2.ppt
Data-Handling part 2.pptData-Handling part 2.ppt
Data-Handling part 2.pptAhmadHashlamon
 
1 Precipitation-titrations.ppt
1 Precipitation-titrations.ppt1 Precipitation-titrations.ppt
1 Precipitation-titrations.pptAhmadHashlamon
 
Complexometric-Titrations part 1.ppt
Complexometric-Titrations part 1.pptComplexometric-Titrations part 1.ppt
Complexometric-Titrations part 1.pptAhmadHashlamon
 
Data-Handling part 1 .ppt
Data-Handling part 1 .pptData-Handling part 1 .ppt
Data-Handling part 1 .pptAhmadHashlamon
 
Acid-Base-Equilibria.ppt
Acid-Base-Equilibria.pptAcid-Base-Equilibria.ppt
Acid-Base-Equilibria.pptAhmadHashlamon
 
General-Equilibrium-Concepts.ppt
General-Equilibrium-Concepts.pptGeneral-Equilibrium-Concepts.ppt
General-Equilibrium-Concepts.pptAhmadHashlamon
 
Stoichiometric-calculations part 2.ppt
Stoichiometric-calculations part 2.pptStoichiometric-calculations part 2.ppt
Stoichiometric-calculations part 2.pptAhmadHashlamon
 

More from AhmadHashlamon (20)

02slidLarge value of face area Large value of face area
02slidLarge value of face area Large value of face area02slidLarge value of face area Large value of face area
02slidLarge value of face area Large value of face area
 
kinectic_and_drug_stability microemulsions
kinectic_and_drug_stability microemulsionskinectic_and_drug_stability microemulsions
kinectic_and_drug_stability microemulsions
 
04slidemicroemulsions microemulsions microemulsions microemulsions microemuls...
04slidemicroemulsions microemulsions microemulsions microemulsions microemuls...04slidemicroemulsions microemulsions microemulsions microemulsions microemuls...
04slidemicroemulsions microemulsions microemulsions microemulsions microemuls...
 
skeletal musclemicroemulsions microemulsions microemulsions microemulsions mi...
skeletal musclemicroemulsions microemulsions microemulsions microemulsions mi...skeletal musclemicroemulsions microemulsions microemulsions microemulsions mi...
skeletal musclemicroemulsions microemulsions microemulsions microemulsions mi...
 
microemulsions microemulsions microemulsions microemulsions
microemulsions microemulsions microemulsions microemulsionsmicroemulsions microemulsions microemulsions microemulsions
microemulsions microemulsions microemulsions microemulsions
 
Herbal_Medicine_Developmggggffgfent.pptx
Herbal_Medicine_Developmggggffgfent.pptxHerbal_Medicine_Developmggggffgfent.pptx
Herbal_Medicine_Developmggggffgfent.pptx
 
C13NMR 2.pptx
C13NMR 2.pptxC13NMR 2.pptx
C13NMR 2.pptx
 
Clothes fiber.pptx
Clothes fiber.pptxClothes fiber.pptx
Clothes fiber.pptx
 
performulation studies.pptx
performulation studies.pptxperformulation studies.pptx
performulation studies.pptx
 
introduction.pptx
introduction.pptxintroduction.pptx
introduction.pptx
 
chapter_3_part_1.pptx
chapter_3_part_1.pptxchapter_3_part_1.pptx
chapter_3_part_1.pptx
 
Oxidation-Reduction-Reactions-and-Titrations.ppt
Oxidation-Reduction-Reactions-and-Titrations.pptOxidation-Reduction-Reactions-and-Titrations.ppt
Oxidation-Reduction-Reactions-and-Titrations.ppt
 
Complexometric-Titrations part 2.ppt
Complexometric-Titrations part 2.pptComplexometric-Titrations part 2.ppt
Complexometric-Titrations part 2.ppt
 
Data-Handling part 2.ppt
Data-Handling part 2.pptData-Handling part 2.ppt
Data-Handling part 2.ppt
 
1 Precipitation-titrations.ppt
1 Precipitation-titrations.ppt1 Precipitation-titrations.ppt
1 Precipitation-titrations.ppt
 
Complexometric-Titrations part 1.ppt
Complexometric-Titrations part 1.pptComplexometric-Titrations part 1.ppt
Complexometric-Titrations part 1.ppt
 
Data-Handling part 1 .ppt
Data-Handling part 1 .pptData-Handling part 1 .ppt
Data-Handling part 1 .ppt
 
Acid-Base-Equilibria.ppt
Acid-Base-Equilibria.pptAcid-Base-Equilibria.ppt
Acid-Base-Equilibria.ppt
 
General-Equilibrium-Concepts.ppt
General-Equilibrium-Concepts.pptGeneral-Equilibrium-Concepts.ppt
General-Equilibrium-Concepts.ppt
 
Stoichiometric-calculations part 2.ppt
Stoichiometric-calculations part 2.pptStoichiometric-calculations part 2.ppt
Stoichiometric-calculations part 2.ppt
 

Recently uploaded

OKC Thunder Reveal Game 2 Playoff T Shirts
OKC Thunder Reveal Game 2 Playoff T ShirtsOKC Thunder Reveal Game 2 Playoff T Shirts
OKC Thunder Reveal Game 2 Playoff T Shirtsrahman018755
 
Nicola Mining Inc. Corporate Presentation April 2024
Nicola Mining Inc. Corporate Presentation April 2024Nicola Mining Inc. Corporate Presentation April 2024
Nicola Mining Inc. Corporate Presentation April 2024nicola_mining
 
High Profile Call Girls Kolkata Gayatri 🤌 8250192130 🚀 Vip Call Girls Kolkata
High Profile Call Girls Kolkata Gayatri 🤌  8250192130 🚀 Vip Call Girls KolkataHigh Profile Call Girls Kolkata Gayatri 🤌  8250192130 🚀 Vip Call Girls Kolkata
High Profile Call Girls Kolkata Gayatri 🤌 8250192130 🚀 Vip Call Girls Kolkataanamikaraghav4
 
定制(UWIC毕业证书)英国卡迪夫城市大学毕业证成绩单原版一比一
定制(UWIC毕业证书)英国卡迪夫城市大学毕业证成绩单原版一比一定制(UWIC毕业证书)英国卡迪夫城市大学毕业证成绩单原版一比一
定制(UWIC毕业证书)英国卡迪夫城市大学毕业证成绩单原版一比一Fir La
 
Methanex Investor Presentation (April 2024)
Methanex Investor Presentation (April 2024)Methanex Investor Presentation (April 2024)
Methanex Investor Presentation (April 2024)Methanex Corporation
 
Russian Call Girls Kolkata Amaira 🤌 8250192130 🚀 Vip Call Girls Kolkata
Russian Call Girls Kolkata Amaira 🤌  8250192130 🚀 Vip Call Girls KolkataRussian Call Girls Kolkata Amaira 🤌  8250192130 🚀 Vip Call Girls Kolkata
Russian Call Girls Kolkata Amaira 🤌 8250192130 🚀 Vip Call Girls Kolkataanamikaraghav4
 
《加州大学圣克鲁兹分校学位证书复制》Q微信741003700美国学历疑难问题指南|挂科被加州大学圣克鲁兹分校劝退没有毕业证怎么办?《UCSC毕业证购买|加...
《加州大学圣克鲁兹分校学位证书复制》Q微信741003700美国学历疑难问题指南|挂科被加州大学圣克鲁兹分校劝退没有毕业证怎么办?《UCSC毕业证购买|加...《加州大学圣克鲁兹分校学位证书复制》Q微信741003700美国学历疑难问题指南|挂科被加州大学圣克鲁兹分校劝退没有毕业证怎么办?《UCSC毕业证购买|加...
《加州大学圣克鲁兹分校学位证书复制》Q微信741003700美国学历疑难问题指南|挂科被加州大学圣克鲁兹分校劝退没有毕业证怎么办?《UCSC毕业证购买|加...wyqazy
 
Cyberagent_For New Investors_EN_240424.pdf
Cyberagent_For New Investors_EN_240424.pdfCyberagent_For New Investors_EN_240424.pdf
Cyberagent_For New Investors_EN_240424.pdfCyberAgent, Inc.
 
Collective Mining | Corporate Presentation - April 2024
Collective Mining | Corporate Presentation - April 2024Collective Mining | Corporate Presentation - April 2024
Collective Mining | Corporate Presentation - April 2024CollectiveMining1
 
VIP 7001035870 Find & Meet Hyderabad Call Girls Miyapur high-profile Call Girl
VIP 7001035870 Find & Meet Hyderabad Call Girls Miyapur high-profile Call GirlVIP 7001035870 Find & Meet Hyderabad Call Girls Miyapur high-profile Call Girl
VIP 7001035870 Find & Meet Hyderabad Call Girls Miyapur high-profile Call Girladitipandeya
 
Short-, Mid-, and Long-term gxxoals.pptx
Short-, Mid-, and Long-term gxxoals.pptxShort-, Mid-, and Long-term gxxoals.pptx
Short-, Mid-, and Long-term gxxoals.pptxHenryBriggs2
 
VIP Kolkata Call Girl Entally 👉 8250192130 Available With Room
VIP Kolkata Call Girl Entally 👉 8250192130  Available With RoomVIP Kolkata Call Girl Entally 👉 8250192130  Available With Room
VIP Kolkata Call Girl Entally 👉 8250192130 Available With Roomdivyansh0kumar0
 
No 1 AMil Baba In Islamabad No 1 Amil Baba In Lahore No 1 Amil Baba In Faisl...
No 1 AMil Baba In Islamabad  No 1 Amil Baba In Lahore No 1 Amil Baba In Faisl...No 1 AMil Baba In Islamabad  No 1 Amil Baba In Lahore No 1 Amil Baba In Faisl...
No 1 AMil Baba In Islamabad No 1 Amil Baba In Lahore No 1 Amil Baba In Faisl...First NO1 World Amil baba in Faisalabad
 

Recently uploaded (20)

Falcon Invoice Discounting - Best Platform
Falcon Invoice Discounting - Best PlatformFalcon Invoice Discounting - Best Platform
Falcon Invoice Discounting - Best Platform
 
OKC Thunder Reveal Game 2 Playoff T Shirts
OKC Thunder Reveal Game 2 Playoff T ShirtsOKC Thunder Reveal Game 2 Playoff T Shirts
OKC Thunder Reveal Game 2 Playoff T Shirts
 
Model Call Girl in Udyog Vihar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Udyog Vihar Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Udyog Vihar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Udyog Vihar Delhi reach out to us at 🔝9953056974🔝
 
Nicola Mining Inc. Corporate Presentation April 2024
Nicola Mining Inc. Corporate Presentation April 2024Nicola Mining Inc. Corporate Presentation April 2024
Nicola Mining Inc. Corporate Presentation April 2024
 
High Profile Call Girls Kolkata Gayatri 🤌 8250192130 🚀 Vip Call Girls Kolkata
High Profile Call Girls Kolkata Gayatri 🤌  8250192130 🚀 Vip Call Girls KolkataHigh Profile Call Girls Kolkata Gayatri 🤌  8250192130 🚀 Vip Call Girls Kolkata
High Profile Call Girls Kolkata Gayatri 🤌 8250192130 🚀 Vip Call Girls Kolkata
 
Call Girls in South Ex⎝⎝9953056974⎝⎝ Escort Delhi NCR
Call Girls in South Ex⎝⎝9953056974⎝⎝ Escort Delhi NCRCall Girls in South Ex⎝⎝9953056974⎝⎝ Escort Delhi NCR
Call Girls in South Ex⎝⎝9953056974⎝⎝ Escort Delhi NCR
 
young Call girls in Dwarka sector 1🔝 9953056974 🔝 Delhi escort Service
young Call girls in Dwarka sector 1🔝 9953056974 🔝 Delhi escort Serviceyoung Call girls in Dwarka sector 1🔝 9953056974 🔝 Delhi escort Service
young Call girls in Dwarka sector 1🔝 9953056974 🔝 Delhi escort Service
 
定制(UWIC毕业证书)英国卡迪夫城市大学毕业证成绩单原版一比一
定制(UWIC毕业证书)英国卡迪夫城市大学毕业证成绩单原版一比一定制(UWIC毕业证书)英国卡迪夫城市大学毕业证成绩单原版一比一
定制(UWIC毕业证书)英国卡迪夫城市大学毕业证成绩单原版一比一
 
young call girls in Hauz Khas,🔝 9953056974 🔝 escort Service
young call girls in Hauz Khas,🔝 9953056974 🔝 escort Serviceyoung call girls in Hauz Khas,🔝 9953056974 🔝 escort Service
young call girls in Hauz Khas,🔝 9953056974 🔝 escort Service
 
Methanex Investor Presentation (April 2024)
Methanex Investor Presentation (April 2024)Methanex Investor Presentation (April 2024)
Methanex Investor Presentation (April 2024)
 
Russian Call Girls Kolkata Amaira 🤌 8250192130 🚀 Vip Call Girls Kolkata
Russian Call Girls Kolkata Amaira 🤌  8250192130 🚀 Vip Call Girls KolkataRussian Call Girls Kolkata Amaira 🤌  8250192130 🚀 Vip Call Girls Kolkata
Russian Call Girls Kolkata Amaira 🤌 8250192130 🚀 Vip Call Girls Kolkata
 
《加州大学圣克鲁兹分校学位证书复制》Q微信741003700美国学历疑难问题指南|挂科被加州大学圣克鲁兹分校劝退没有毕业证怎么办?《UCSC毕业证购买|加...
《加州大学圣克鲁兹分校学位证书复制》Q微信741003700美国学历疑难问题指南|挂科被加州大学圣克鲁兹分校劝退没有毕业证怎么办?《UCSC毕业证购买|加...《加州大学圣克鲁兹分校学位证书复制》Q微信741003700美国学历疑难问题指南|挂科被加州大学圣克鲁兹分校劝退没有毕业证怎么办?《UCSC毕业证购买|加...
《加州大学圣克鲁兹分校学位证书复制》Q微信741003700美国学历疑难问题指南|挂科被加州大学圣克鲁兹分校劝退没有毕业证怎么办?《UCSC毕业证购买|加...
 
Cyberagent_For New Investors_EN_240424.pdf
Cyberagent_For New Investors_EN_240424.pdfCyberagent_For New Investors_EN_240424.pdf
Cyberagent_For New Investors_EN_240424.pdf
 
Collective Mining | Corporate Presentation - April 2024
Collective Mining | Corporate Presentation - April 2024Collective Mining | Corporate Presentation - April 2024
Collective Mining | Corporate Presentation - April 2024
 
Model Call Girl in Uttam Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Uttam Nagar Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Uttam Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Uttam Nagar Delhi reach out to us at 🔝9953056974🔝
 
VIP 7001035870 Find & Meet Hyderabad Call Girls Miyapur high-profile Call Girl
VIP 7001035870 Find & Meet Hyderabad Call Girls Miyapur high-profile Call GirlVIP 7001035870 Find & Meet Hyderabad Call Girls Miyapur high-profile Call Girl
VIP 7001035870 Find & Meet Hyderabad Call Girls Miyapur high-profile Call Girl
 
Short-, Mid-, and Long-term gxxoals.pptx
Short-, Mid-, and Long-term gxxoals.pptxShort-, Mid-, and Long-term gxxoals.pptx
Short-, Mid-, and Long-term gxxoals.pptx
 
Escort Service Call Girls In Shalimar Bagh, 99530°56974 Delhi NCR
Escort Service Call Girls In Shalimar Bagh, 99530°56974 Delhi NCREscort Service Call Girls In Shalimar Bagh, 99530°56974 Delhi NCR
Escort Service Call Girls In Shalimar Bagh, 99530°56974 Delhi NCR
 
VIP Kolkata Call Girl Entally 👉 8250192130 Available With Room
VIP Kolkata Call Girl Entally 👉 8250192130  Available With RoomVIP Kolkata Call Girl Entally 👉 8250192130  Available With Room
VIP Kolkata Call Girl Entally 👉 8250192130 Available With Room
 
No 1 AMil Baba In Islamabad No 1 Amil Baba In Lahore No 1 Amil Baba In Faisl...
No 1 AMil Baba In Islamabad  No 1 Amil Baba In Lahore No 1 Amil Baba In Faisl...No 1 AMil Baba In Islamabad  No 1 Amil Baba In Lahore No 1 Amil Baba In Faisl...
No 1 AMil Baba In Islamabad No 1 Amil Baba In Lahore No 1 Amil Baba In Faisl...
 

Atomic Emission Spectroscopy (Chapter10.pptx

  • 2. ATOMIC EMISSION SPECTROSCOPY  The sample is converted into excited gaseous atoms and ions which emit light at characteristic wavelengths (wavenumbers).  The analyte is identified by emission at a known wavelength (qual. analysis) and its concentration is determined from the intensity of the emission at that wavelength (quant. analysis).  Atoms or ions in the gas phase can be excited by high temperatures or by light absorption.  A fraction of the atoms or ions lose their excitation energy by giving off a photon.  The most intense emission line for a neutral atom is the resonance line (same line for absorption and emission e.g. 589nm for Na atom).
  • 3. Atomic Emission Spectrometers. Emission sources: The source must a. vaporize the sample, b. break down all compounds to atoms and ions, c. and excite the atoms and ions. • Very high temperatures are required, much higher than in atomic absorption Methods to atomize and excite the analytes: a. electrical arcs (low voltage, high current) b. electrical sparks (high voltage, low current c. flames have been used .
  • 4. Monochromators/ Polychromators  A polychromator has multiple exit slits with multiple detectors PMT Because there are many emission lines from the plasma, often separated by only a few Angstroms, high resolution is required in monochromators and polychromators.  Designs are based on an echelle grating operating at high orders combined with an order-sorting prism
  • 5. Detectors • Usually a PMT (multiple PMT’s in a polychromator) • Both the polychromator + multiple PMTs and the echelle can monitor concentrations of 10 or more elements from one sample within a few minutes. • These devices are used heavily by the metallurgy industry to monitor composition of alloys.
  • 6. Internal Standard vs. external standard method • Because AES can easily perform multi-element analysis, an internal standard (an element not naturally found in the samples) is often incorporated at a constant concentration into each sample. • To compensate for fluctuation of plasma emission intensity (caused by fluctuations in plasma temperature), the emission intensity of each element line is divided by the emission intensity of an internal standard.
  • 7. Internal standard Method An Internal Standard is a substance that is added in a constant amount to all samples, blanks and calibration standards in an analysis. Calibration involves plotting the ratio of the analyte signal to the internal standard signal as a function of analyte concentration of the standards. This ratio for the samples is then used to obtain their analyte concentrations from a calibration curve. Internal standard can compensate for several types of both random and systematic errors.
  • 8. Internal Standard Method  Most convenient when variations in analytical sample size, position, or matrix limit the precision of a technique.  Or when the sensitivity of instrument changes with time
  • 9. Internal Standard Procedure  Prepare a set of standard solutions for analyte (A) as with the calibration curve method, but add a constant amount of a second species (B called internal standard) to each solution.  Add the same amount of IS to the blank and sample solution.  Prepare a plot of SA/SB versus [A].
  • 10. Example: Pb by ICP Emission Each Pb solution contains 100 ppm Cu as IS [Pb] (ppm) Pb Cu Pb/Cu 20 112 1347 0.083 40 243 1527 0.159 60 326 1383 0.236 80 355 1135 0.313 100 558 1440 0.388 Signal
  • 11.
  • 12. No Internal Standard Correction y = 5.02x + 17.6 R² = 0.9413 0 100 200 300 400 500 600 0 20 40 60 80 100 120
  • 13. Internal Standard Correction y = 0.0038x + 0.0066 R² = 1 0 0.05 0.1 0.15 0.2 0.25 0.3 0.35 0.4 0.45 0 20 40 60 80 100 120
  • 14. unknown sample of Pb (containing100 ppm Cu as IS) gives a ratio of 0.241. Conc. of Pb in the unknown = 78.3 ppm
  • 16. Flame photometer:  flame emission spectrometry is used for the quantitative determination of small amounts of metal salts in solution.  Flam emission is commonly used in clinical labs for determination the concentration of Na or K ions in blood. In this method:  A. sample solution sprayed or aspirated as fine mist into flame;  B. conversion of sample solution into an aerosol by atomizer (scent spray)
  • 17. principle.  Then Heat of the flame vaporizes sample constituents. Still no chemical change. By heat of the flame + action of the reducing gas (fuel), molecules & ions of the sample species are decomposed and reduced to give ATOMS. Na+ + e- --> Na  Heat of the flame causes excitation of some atoms into higher electronic states. Excited atoms revert to ground state by emission of light energy, h, of characteristic wavelength; measured by detector.  The reading that the instrument gives is used to calculate the concentration of the element to be determined.
  • 19. Experimental Aspects of Flame Photometry • Propane-air or natural gas-air gives good flame - strong heat, minimal background light emission. • But always need to run a solvent blank for setting zero emission. • Solutions diluted to fall within linear part of emission curve. • Anion and cation interference effects can cause errors (enhancement or suppression). "Radiation buffer" for dilution of standards and samples to swamp out inconsistencies. • Internal standard (lithium) useful to counter random flame instability and random dilution errors.
  • 20.
  • 21.
  • 22.
  • 23. Plasma not flame • ICP torch is the most important. • Like the flame AA experiment, the sample is introduced into the instrument as a solution fed into a nebulizer, creating a fine spray of droplets in a flowing gas stream
  • 24.
  • 25.
  • 26. Inductively coupled plasma (ICP): • A plasma (partially ionized gas) of argon is formed and sustained by intense electric and magnetic fields inside a coiled radio antenna (27 MHz, 1-2 kW power). A triple-tube quartz torch is used to: • (a) generate annular plasma, • (b) introduce the sample as droplets in an Ar gas stream. • Temperatures are about 10,000 K at hottest part of plasma. Observation is made 1-3 cm above hottest part; local temperatures are still high (ca. 6000 K).
  • 27. Advantages: (a) a very wide dynamic range (3 to 6 orders-of-magnitude) on calibration curves; (b) stable operation over long periods of time, (c) almost no chemical interferences due to the very high temperatures. Disadvantages: (a) expensive to buy and operate; (b) rapid consumption of high purity Ar gas (several L/min).