This document evaluates the performance of the Clart Entherpex kit, which allows for the simultaneous detection and identification of eight human herpesviruses and enteroviruses via multiplex PCR and microarray analysis. The evaluation found the kit had high sensitivity for herpesviruses down to 500 copies/ml or less, and enteroviruses down to 1480 copies/ml. It correctly identified 96% of previously positive herpesvirus samples and 100% of enterovirus samples. However, further studies with more CSF samples are needed to characterize kit performance before clinical use.