Measuring pH Using the Orion
3 Star Plus pH Meter with Printer




        Measuring pH Using the Orion 3 Star Plus pH Meter with Printer   1
Orion 3 Star Plus pH Meter




       Front of Meter & Keypad



     Measuring pH Using the Orion 3 Star Plus pH Meter with Printer   2
Start Up
• Turn Printer ON
• Press POWER Key on
  Meter
    Display will read:
        Star PLUS
        Software Version and
         current method
    Header information will
     print:
        Meter Serial Number
        Meter Software Version
        Method currently set
• Check print quality
                  Measuring pH Using the Orion 3 Star Plus pH Meter with Printer   3
Calibrate Method 1


      Open
     Fill Hole

                                                  Rinse electrode and
                                Select Method (#
                                                  blot carefully
                                decimals) to be used
                                for samples




                                   Press MEASURE
                                   to set method


 Measuring pH Using the Orion 3 Star Plus pH Meter with Printer         4
Calibrate Method 1
                                                            CAL 1 Displayed..




Press CALIBRATE




                        • pH will stop flashing when
                          reading stable
                        • Buffer value will be displayed
                                  Immerse electrode
                          with leading zeroes
                                  in first buffer

                   Measuring pH Using the Orion 3 Star Plus pH Meter with Printer   5
Calibrate Method 1
                                After pressing CALIBRATE,
                                • CAL 2 displayed
                                • pH flashes while reading in
                                  progress




                                                           • Remove electrode from
                                                             first buffer, rinse & blot
                                                           • Immerse electrode in
                                                             second buffer

            • pH will stop flashing when
              reading stable
            • Buffer value will be displayed
              with leading zeroes


 Measuring pH Using the Orion 3 Star Plus pH Meter with Printer                       6
Calibrate Method 1




                       After pressing CALIBRATE,• Remove electrode from
                       • CAL 3 displayed
                         pH will stop flashing when
                                                  second buffer, rinse & blot
                       • pH flashes while reading in
                         reading stable         • Immerse electrode in third
                       • progress
                         Buffer value will be displayed
                                                  buffer
                         with leading zeroes




 Measuring pH Using the Orion 3 Star Plus pH Meter with Printer            7
Completing Calibration
                               Press MEASURE to print
                               & save calibration results




                                                             • Check date and time
                                                               (not visible in photo)
                                                             • Check mV for each
                                                               buffer
                                                             • Check slope 1 and slope
                                                               2
                                                             • Initial & date printout




   Measuring pH Using the Orion 3 Star Plus pH Meter with Printer                   8
Verifying Instrument Performance
                                                                • Remove electrode from
                                                                  buffer, rinse & blot
                                                                • Immerse electrode in QC
Press MEASURE                                                     (6.86) buffer


                                                     pH will stop flashing when
                                                      reading stable


                                 Press MEASURE
                                 to print result


                                 • Initial & date
                                 • Identify sample as
                                   QC Buffer




                Measuring pH Using the Orion 3 Star Plus pH Meter with Printer              9
Documenting Calibration

 Calibration & Initial
 QC Buffer on form

• Check Dates & Times
• Check Method
• Check Calibration #
• Check Limits:
  • 4.01 +177 mV ± 30 mV
  • 7.00 0 mV ± 30 mV
  • 10.01 -177 mV ± 30 mV
  • Slope 1 and 2:
         95% - 102%
  • QC Buffer: 6.86 ± 0.05
      Method 1: 6.8 – 6.9
      Method 2: 6.81 – 6.91




                              Measuring pH Using the Orion 3 Star Plus pH Meter with Printer   10
Measuring Sample pH
                                                                           • Remove electrode from
                                                                             buffer or storage solution,
                                                                             rinse & blot
                                                                           • Immerse electrode in
Press MEASURE                                                                sample


                                                      pH will stop flashing when
                                                       reading stable



                          Press MEASURE
                          again to print result




                                • Initial & date
                                • Identify sample
                Measuring pH Using the Orion 3 Star Plus pH Meter with Printer                     11
Reviewing Results
• Check Dates & Times
  • If > 2 hours since last QC Buffer run,
    instrument performance must be
    verified by running the QC Buffer as a
    Batch control

    • Check Method
      • 1 decimal for any product QC and most
        reagents/solution
      • 2 decimals if reagent procedure
        requires more sensitivity
      • NO ROUNDING

         • Check Calibration #
            • Match to the most recent calibration for
              the method selected

             • Attach product results to Product QC
               Forms
             • Attach reagent/solution pH
               measurements to Reagent pH Log
               [19.4.frm024]
                           Measuring pH Using the Orion 3 Star Plus pH Meter with Printer   12

Measuring pH with orion 3star with printer workshop

  • 1.
    Measuring pH Usingthe Orion 3 Star Plus pH Meter with Printer Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 1
  • 2.
    Orion 3 StarPlus pH Meter Front of Meter & Keypad Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 2
  • 3.
    Start Up • TurnPrinter ON • Press POWER Key on Meter  Display will read:  Star PLUS  Software Version and current method  Header information will print:  Meter Serial Number  Meter Software Version  Method currently set • Check print quality Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 3
  • 4.
    Calibrate Method 1 Open Fill Hole Rinse electrode and Select Method (# blot carefully decimals) to be used for samples Press MEASURE to set method Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 4
  • 5.
    Calibrate Method 1 CAL 1 Displayed.. Press CALIBRATE • pH will stop flashing when reading stable • Buffer value will be displayed Immerse electrode with leading zeroes in first buffer Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 5
  • 6.
    Calibrate Method 1 After pressing CALIBRATE, • CAL 2 displayed • pH flashes while reading in progress • Remove electrode from first buffer, rinse & blot • Immerse electrode in second buffer • pH will stop flashing when reading stable • Buffer value will be displayed with leading zeroes Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 6
  • 7.
    Calibrate Method 1 After pressing CALIBRATE,• Remove electrode from • CAL 3 displayed pH will stop flashing when second buffer, rinse & blot • pH flashes while reading in reading stable • Immerse electrode in third • progress Buffer value will be displayed buffer with leading zeroes Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 7
  • 8.
    Completing Calibration Press MEASURE to print & save calibration results • Check date and time (not visible in photo) • Check mV for each buffer • Check slope 1 and slope 2 • Initial & date printout Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 8
  • 9.
    Verifying Instrument Performance • Remove electrode from buffer, rinse & blot • Immerse electrode in QC Press MEASURE (6.86) buffer pH will stop flashing when reading stable Press MEASURE to print result • Initial & date • Identify sample as QC Buffer Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 9
  • 10.
    Documenting Calibration Calibration& Initial QC Buffer on form • Check Dates & Times • Check Method • Check Calibration # • Check Limits: • 4.01 +177 mV ± 30 mV • 7.00 0 mV ± 30 mV • 10.01 -177 mV ± 30 mV • Slope 1 and 2: 95% - 102% • QC Buffer: 6.86 ± 0.05 Method 1: 6.8 – 6.9 Method 2: 6.81 – 6.91 Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 10
  • 11.
    Measuring Sample pH • Remove electrode from buffer or storage solution, rinse & blot • Immerse electrode in Press MEASURE sample pH will stop flashing when reading stable Press MEASURE again to print result • Initial & date • Identify sample Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 11
  • 12.
    Reviewing Results • CheckDates & Times • If > 2 hours since last QC Buffer run, instrument performance must be verified by running the QC Buffer as a Batch control • Check Method • 1 decimal for any product QC and most reagents/solution • 2 decimals if reagent procedure requires more sensitivity • NO ROUNDING • Check Calibration # • Match to the most recent calibration for the method selected • Attach product results to Product QC Forms • Attach reagent/solution pH measurements to Reagent pH Log [19.4.frm024] Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 12