Biomedical Instrumentation I
Electrical Safety
This Week
 Safety in the clinical environment: Electrical safety
 Physiological effects of electricity
 Susceptibility parameters
 Distribution of electrical power
 Isolated power systems
 Macroshock hazards
 Microshock hazards
 Electrical safety codes and standards
 Protection
 Power distribution
 Ground fault circuit interrupters (GFCI)
 Equipment design
 Electrical safety analyzers / Testing electrical systems
Safety in Clinical Environment
 Electrical hazards
 Electrical shocks (micro and macro) due to equipment failure, failure of
power delivery systems, ground failures, burns, fire, etc.
 Mechanical hazards
 mobility aids, transfer devices, prosthetic devices, mechanical assist
devices, patient support devices
 Environmental hazards
 Solid wastes, noise, utilities (natural gas), building structures, etc.
 Biological hazards
 Infection control, viral outbreak, isolation, decontamination, sterilization,
waste disposal issues
 Radiation hazards
 Use of radioactive materials, radiation devices (MRI, CT, PET), exposure
control
Electrical Safety
 Many sources of energy, potentially hazardous substances,
instruments and procedures
 Use of fire, compressed air, water, chemicals, drugs, microorganisms,
waste, sound, electricity, radiation, natural and unnatural disaster,
negligence, sources of radiation, etc.
 Medical procedures expose patients to increased risks of hazards due
to skin and membranes being penetrated / altered
 10,000 device related injuries in the US every year! Typically due to
 Improper use
 Inadequate training
 Lack of experience
 Improper (lack of) use of manuals
 Device failure
Physiological Effects of
Electricity
 For electricity to have an effect on the human body:
 An electrical potential difference must be present
 The individual must be part of the electrical circuit, that is, a current must
enter the body at one point and leave it at another.
 However, what causes the physiological effect is NOT voltage, but
rather CURRENT.
 A high voltage (KV,(103V)) applied over a large impedance (rough skin)
may not cause much (any) damage
 A low voltage applied over very small impedances (heart tissue) may cause
grave consequences (ventricular fibrillation)
 The magnitude of the current is simply the applied voltage divided
by the total effective impedance the current faces; skin : largest.
 Electricity can have one of three effects:
 Electrical stimulation of excitable tissue (muscles, nerve)
 Resistive heating of tissue
 Electrical burns / tissue damage for direct current and high voltages
Physiological Effects of
Electricity
Physiological effects of electricity. Threshold or estimated mean values are given for each effect in a
70 kg human for a 1 to 3 s exposure to 60 Hz current applied via copper wires grasped by the hands.
Dry skin impedance:93 kΩ / cm2
Electrode gel on skin: 10.8 kΩ / cm2
Penetrated skin: 200 Ω / cm2
The real physiological effect depends on the actual path of the current
 Threshold of perception: The minimal current that an individual can detect. For AC (with
wet hands) can be as small as 0.5 mA at 60 Hz. For DC, 2 ~10 mA
 Let-go current: The maximal current at which the subject can voluntarily withdraw. 6 ~ 100
mA, at which involuntary muscle contractions, reflex withdrawals, secondary physical effects
(falling, hitting head) may also occur
 Respiratory Paralysis / Pain / Fatigue At as low as 20 mA, involuntary contractions of
respiratory muscles can cause asphyxiation / respiratory arrest, if the current is not interrupted.
Strong involuntary contraction of other muscles can cause pain and fatigue
 Ventricular fibrillation 75 ~ 400 mA can cause heart muscles to contract uncontrollably,
altering the normal propagation of the electrical activity of the heart. HR can raise up to 300
bpm, rapid, disorganized and too high to pump any meaningful amount of blood  ventricular
fibrillation. Normal rhythm can only return using a defibrillator
 Sustained myocardial contraction / Burns and physical injury At 1 ~6 A, the
entire heart muscle contracts and heart stops beating. This will not cause irreversible tissue
damage, however, as normal rhythm will return once the current is removed. At or after 10A,
however, burns can occur, particularly at points of entry and exit.
Physiological Effects of
Electricity
Important Susceptibility
Parameters
 Threshold and let-go current variability
Distributions of perception thresholds and let-go currents These data depend on surface area of
contact, moistened hand grasping AWG No. 8 copper wire, 70 kg human, 60Hz, 1~3 s. of exposure
Important Susceptibility
Parameters
 Frequency
 Note that the minimal let-go
current happens at the precise
frequency of commercial
power-line, 50-60Hz.
 Let-go current rises below 10
Hz and above several hundred
Hz.
Let-go current versus frequency
Percentile values indicate variability of
let-go current among individuals. Let-
go currents for women are about two-
thirds the values for men.
Important Susceptibility
Parameters
 Duration
 The longer the duration, the smaller
the current at which ventricular
fibrillation occurs
 Shock must occur long enough to
coincide with the most vulnerable
period occurring during the T wave.
 Weight
 Fibrillation threshold increases with
body weight (from 50mA for 6kg
dogs to 130 mA for 24 kg dogs.
Fibrillation current versus shock duration.
Thresholds for ventricular fibrillation in animals
for 60 Hz AC current. Duration of current (0.2 to
5 s) and weight of animal body were varied.
Important Susceptibility
Parameters
 Points of entry
 The magnitude of the current required to fibrillate the heart is far greater if the
current is applied directly to heart; externally applied current loses much of its
amplitude due to current distributions. Large, externally applied currents cause
macroshock.
 If catheters are used, the natural protection provided by the skin 15 kΩ ~ 2 MΩ) is
bypassed, greatly reducing the amount of current req’d to cause fibrillation. Even
smallest currents (80 ~ 600 μA), causing microshock, may result in fibrillation.
Safety limit for microshocks is 10 μA.
 The precise point of entry, even externally is very important: If both points of
entry and exit are on the same extremity, the risk of fibrillation is greatly reduced
even at high currents (e.g. the current req’d for fibrillation through Lead I (LA-RA)
electrodes is higher than for Leads II (LL-RA) and III (LL-LA).
Important Susceptibility
Parameters
 Points of entry
Effect of entry points on current distribution (a) Macroshock, externally applied current spreads through-
out the body. (b) Microshock, all the current applied through an intracardiac catheter flows through the heart.
Distribution of
Electrical Power
Simplified electric-power distribution for 115 V circuits. Power frequency is 60 Hz
(230 V)
Distribution of Power
 If electrical devices were perfect, only two wires
would be adequate (hot and return), with all power
confined to these two wires. However, there are
two major departures from this ideal case:
• A fault may occur, through miswiring, component
failure, etc., causing an electrical potential between
an exposed surface (metal casing of the device) and a
grounded surface (wet floor, metal case of another
device etc.) Any person who bridges these two
surfaces is subject to macroshock.
• Even if a fault does not occur, imperfect insulation or electromagnetic coupling (capacitive
or inductive) may produce an electrical potential relative to the ground. A susceptible patient
providing a path for this leakage current to flow to the ground is subject to microshock.
• The additional ground line provides a good line of defense! (how / why?)
GROUND!
 The additional line that is connected directly to the earth-ground
provides the following:
 In case of a fault (short circuit between hot conductor and metal casing),
a large current will use the path through the ground wire (instead of the
patient) and not only protect the patient, but also cause the circuit breaker
to open. The ability of the grounding system to conduct high currents to
ground is crucial for this to work!
 If there is no fault, the ground wire serves to conduct the leakage current
safely back to the electrical power source – again, as long as the
grounding system provides a low-resistance pathway to the ground
 Leakage current recommended by ECRI are established to prevent injury
in case the grounding system fails and a patient touches an electrically
active surface (10 ~ 100 μA).
Isolated Power Distribution
Not grounded !
• Normally, when there is a ground-fault from hot wire to ground, a large current is drawn causing a
potential hazard, as the device will stop functioning when the circuit breakers open !
• This can be prevented by using the isolated system, which separates ground from neutral, making
neutral and hot electrically identical. A single ground-fault will not cause large currents, as long as
both hot conductors are initially isolated from ground!
• In fact, in such an isolated system, if a single ground-fault occurs, the system simply reverts back to
the normal ground-referenced system.
•A line isolation monitor is used with such system that continuously monitors for the first ground
fault, during which case it simply informs the operators to fix the problem. The single ground fault
does NOT constitute a hazard!
Macroshock
 Most electrical devices have a metal cabinet, which constitutes a hazard, in
case of an insulation failure or shortened component between the hot power
lead and the chassis. There is then 115 ~ 230 V between the chassis and any
other grounded object.
 The first line of defense available to patients is their skin.
 The outer layer provides 15 kΩ to 1 MΩ depending on the part of the body,
moisture and sweat present, 1% of that of dry skin if skin is broken,
 Internal resistance of the body is 200Ω for each limb, and 100Ω for the trunk, thus
internal body resistance between any two limbs is about 500Ω (somewhat higher
for obese people due to high resistivity of the adipose tissue)!
 Any procedure that reduces or eliminates the skin resistance increases the risk of
electrical shock, including biopotential electrode gel, electronic thermometers
placed in ears, mouth, rectum, intravenous catheters, etc.
 A third wire, grounded to earth, can greatly reduce the effect of macroshock,
as the resistance of that path would be much smaller then even that of
internal body resistance!
Macroshock Hazards
• Direct faults between the hot conductor and the ground is not common,
and technically speaking, ground connection is not necessary during
normal operation.
• In fact, a ground fault will not be detected during normal operation of
the device, only when someone touches it, the hazard becomes known.
Therefore, ground wire in devices and receptacles must be periodically
tested.
Macroshock Hazards
Microshock Hazards
Small currents inevitably flow between adjacent insulated conductors at different
potentials  leakage currents which flow through stray capacitances, insulation,
dust and moisture
Leakage current flowing to the chassis flows safely to the ground, if a low-resistance
ground wire is available.
Microshock Hazards
• If ground wire is broken, the chassis
potential rises above the ground; a
patient who has a grounded connection
to the heart (e.g. through a catheter)
receives a microshock if s/he touches
the chassis.
• If there is a connection from the
chassis to the patient’s heart, and a
connection to the ground anywhere in
the body, this also causes microshock.
• Note that the hazard for microshock
only exists if there is a direct
connection to the heart. Otherwise,
even the internal resistance of the body
is high enough top prevent the
microshocks.
Microshock via
Ground Potentials
Microshocks can also occur if
different devices are not at the
exact same ground potential.
In fact, the microshock can
occur even when a device that
does not connected to the patient
has a ground fault!
A fairly common ground
wire resistance of 0.1Ω can
easily cause a a 500mV
potential difference if
initiated due to a, say
5A of ground fault.
If the patient resistance is less
then 50kΩ, this would cause an
above safe current of 10μA
Microshock Via
Ground Potentials
Safety Codes & Standards
 Limits on leakage current are instituted and regulated by the safety
codes instituted in part by the National Fire Protection Association
(NFPA), American National Standards Institute (ANSI), Association
for the Advancement of Medical Instrumentation (AAMI), and
Emergency Care Research Institute (ECRI).
Basic Approaches to
Shock Protection
 There are two major ways to protect patients from shocks:
 Completely isolate and insulate patient from all sources of electric current
 Keep all conductive surfaces within reach of the patient at the same
voltage
 Neither can be fully achieved  some combination of these two
 Grounding system
 Isolated power-distribution system
 Ground-fault circuit interrupters (GFCI)
Grounding Systems
Low resistance (0.15 Ω) ground that can carry
currents up to the circuit-breaker ratings protects
patients by keeping all conductive surfaces and
receptacle grounds at the same potential.
Protects patients from
• Macroshocks
• Microshocks
• Ground faults elsewhere (!)
The difference between the receptacle grounds
and other surface should be no more then 40 mV)
All the receptacle grounds and conductive surfaces
in the vicinity of the patient are connected to the
patient-equipment grounding point. Each patient-
equipment grounding point is connected to the
reference grounding point that makes a single
connection to the building ground.
Isolated Power Systems
 A good equipotential grounding system cannot eliminate large
current that may result from major ground-faults (which are rather
rare).
 Isolated power systems can protect against such major (single)
ground faults
 Provide considerable protection against macroshocks, particularly around
wet conditions
 However, they are expensive !
 Used only at locations where flammable anesthetics are used. Additional
minor protection against microshocks does not justify the high cost of
these systems to be used everywhere in the clinical environment
Ground – Fault
Circuit Interrupters (GFCI)
 Disconnects source of electric current when a ground fault greater than about 6 mA occurs!
When there is no fault, Ihot=Ineutral. The GFCI detects the difference between these two
currents. If the difference is above a threshold, that means the rest of the current must be
flowing through elsewhere, either the chassis or the patient !!!.
The detection is done through the monitoring the voltage induced by the two coils (hot and
neutral) in the differential transformer!
GFCI
The National Electric Code (NEC - 1996) requires that all
circuits serving bathrooms, garages, outdoor receptacles,
swimming pools and construction sites be fitted with GFCI.
Note that GFCI protect against major ground faults only, not
against microshocks.
Patient care areas are typically not fitted with GFCI, since the
loss of power to life support equipment can also be equally
deadly!
Protection through
Equipment Design
 Strain-relief devices for cords, where cord enters the equipment and
between the cord and plug
 Reduction of leakage current through proper layout and insulation to
minimize the capacitance between all hot conductors and the chassis
 Double insulation to prevent the contact of the patient with the
chassis or any other conducting surface (outer case being insulating
material, plastic knobs, etc.)
 Operation at low voltages; solid state devices operating at <10V are
far less likely to cause macroshocks
 Electrical isolation in circuit design
Electrical Isolation
CM
CM
CMRR
SIG ISO
ISO
Error
Isolation
barrier
Isolation
Capacitance
and resistance
-
+
-
+
Input common
(a)
*IMRR in v/v
Output
common
o =
CM
CMRR
ISO
IMRR
SIG ± Gain
±
RF
ISO
IMRR*
~
~
~
~
~
Error
• Main features of an isolation amplifier:
• High ohmic isolation between input and output (>10MΩ)
• High isolation mode voltage (>1000V)
• High common mode rejection ration (>100 dB)
Transformer Isolation
Amplifiers
-
+
(b)
+ 15 V DC
o
Power
return
25 kHz
- 7.5 V
+ISO
Out
SIG
-ISO
Out
+ 7.5 V
In com
In -
In +
FB
±
5 V
F.S.
Oscillator
25 kHz
Signal
Mod
Rect and
filter
Power
Demod
± 5 V
F.S.
AD202
Hi
Lo
Optical Isolation Amplifier
Isolation barrier
Input
control
Output
control
-V
+V
+o
-
+
o = i
RK
RG
CR3 CR1 CR2
i2
i
i
i
i1
RG
AI AII
i3
i2
o
RK = 1M W
-
+
-
+
+
-
(c)
~
1 2
Electrical Safety Analyzers
Wiring / Receptacle Testing
 Three LED receptacle tester:
 Simple device used to test common wiring problems (can detect only 8 of
possible 64 states)
 Will not detect ground/neutral reversal, or when ground/neutral are hot
and hot is grounded (GFCI would detect the latter)
Electrical Safety Analyzers
Testing Electrical Appliances
 Ground-pin-to-chassis resistance: Should be <0.15Ω during the life
of the appliance
Ground-pin-to-chassis resistance test
Electrical Safety Analyzers
Testing Electrical Appliances
 Chassis leakage current: The leakage current should not exceed 500μA with single fault for
devices not intended for patient contact, and not exceed 300 μA for those that are intended for
patient contact.
To exposed conductive
surface or if none, then 10 by
20 cm metal foil in contact
with the exposed surface
Insulating surface
Current meter
I
Test circuit
Open switch
for appliances
not intended to
contact a patient
Grounding-contact
switch (use in
OPEN position)
Polarity- reversing
switch (use both
positions)
Appliance power switch
(use both OFF and ON positions)
This connection
is at service
entrance or on
supply side of
separately derived
system
Building
ground
H (black)
N (white)
Appliance
G
N
H
H = hot
N = neutral (grounded)
G = grounding conductor
I < 500 μA for facility Ðowned housekeeping and maintenance appliances
I > 300 μA for appliances intended for use in the patient vicinity
120 V
G (green)
Internal
Circuitry
Electrical Safety Analyzers
Testing Electrical Appliances
 Leakage current in patient leads:
 Potentially most damaging leakage is the one with patient leads, since they
typically have low impedance patient contacts
 Current should be restricted to 50μA for non-isolated leads and to 10 μA
for isolated leads (used with catheters / electrodes that make connection
to the heart)
 Leakage current between any pair of leads, or between a single lead and
other patient connections should also be controlled
 Leakage in case of line voltage appearing on the patient should also be
restricted.
Leakage current Testers
Test for leakage current from
patient leads to ground
Leakage Current testers
Test for leakage current
between patient leads
Leakage Current Testers
Test for ac isolation current
Isolation current is the current
that passes through patient
leads to ground if and when
line voltage appears on the
patient. This should also be
limited to 50μA
And this concludes…

More Related Content

What's hot

Biomedical digital signal processing : Digital Hearing Aid
Biomedical digital signal processing : Digital Hearing Aid Biomedical digital signal processing : Digital Hearing Aid
Biomedical digital signal processing : Digital Hearing Aid Pooja Yadav
 
Biomedical Engineering
Biomedical EngineeringBiomedical Engineering
Biomedical Engineeringsubkal
 
Biomedical signal processing
Biomedical signal processingBiomedical signal processing
Biomedical signal processingAbdul Kader
 
HEART LUNG MACHINE
HEART LUNG MACHINEHEART LUNG MACHINE
HEART LUNG MACHINEsathish sak
 
Medical instrumentation- patient monitoring systems
Medical instrumentation- patient monitoring systems Medical instrumentation- patient monitoring systems
Medical instrumentation- patient monitoring systems Poornima D
 
IMPEDANCE MATCHING CIRCUIT AND FILTERS
IMPEDANCE MATCHING CIRCUIT AND FILTERSIMPEDANCE MATCHING CIRCUIT AND FILTERS
IMPEDANCE MATCHING CIRCUIT AND FILTERSBharathasreejaG
 
Wireless telemetry systems
Wireless telemetry systemsWireless telemetry systems
Wireless telemetry systemsSneha Suluru
 
Sensors for Biomedical Devices and systems
Sensors for Biomedical Devices and systemsSensors for Biomedical Devices and systems
Sensors for Biomedical Devices and systemsGunjan Patel
 
Transducers for bio medical
Transducers for bio medicalTransducers for bio medical
Transducers for bio medicalSLIET
 
Amplifiers and biopotential amplifiers new
Amplifiers and biopotential amplifiers newAmplifiers and biopotential amplifiers new
Amplifiers and biopotential amplifiers newM. Raihan
 
biomedical signal processing
biomedical signal processingbiomedical signal processing
biomedical signal processingMeenakshi Sood
 
Unit 3 biomedical
Unit 3 biomedicalUnit 3 biomedical
Unit 3 biomedicalAnu Antony
 
Sensory Substitution
Sensory SubstitutionSensory Substitution
Sensory SubstitutionNatasha Gupta
 
safety from electric shock
safety from electric shocksafety from electric shock
safety from electric shock2461998
 

What's hot (20)

Biomedical digital signal processing : Digital Hearing Aid
Biomedical digital signal processing : Digital Hearing Aid Biomedical digital signal processing : Digital Hearing Aid
Biomedical digital signal processing : Digital Hearing Aid
 
Biomedical Engineering
Biomedical EngineeringBiomedical Engineering
Biomedical Engineering
 
Biomedical signal processing
Biomedical signal processingBiomedical signal processing
Biomedical signal processing
 
HEART LUNG MACHINE
HEART LUNG MACHINEHEART LUNG MACHINE
HEART LUNG MACHINE
 
Medical instrumentation- patient monitoring systems
Medical instrumentation- patient monitoring systems Medical instrumentation- patient monitoring systems
Medical instrumentation- patient monitoring systems
 
Ecg Signal Processing
Ecg Signal ProcessingEcg Signal Processing
Ecg Signal Processing
 
IMPEDANCE MATCHING CIRCUIT AND FILTERS
IMPEDANCE MATCHING CIRCUIT AND FILTERSIMPEDANCE MATCHING CIRCUIT AND FILTERS
IMPEDANCE MATCHING CIRCUIT AND FILTERS
 
Instrumentation amplifier in ECG
Instrumentation amplifier  in ECG Instrumentation amplifier  in ECG
Instrumentation amplifier in ECG
 
PCG
PCGPCG
PCG
 
Electromyography (emg) basics
Electromyography (emg)   basicsElectromyography (emg)   basics
Electromyography (emg) basics
 
Wireless telemetry systems
Wireless telemetry systemsWireless telemetry systems
Wireless telemetry systems
 
Electrical safety
Electrical safetyElectrical safety
Electrical safety
 
Sensors for Biomedical Devices and systems
Sensors for Biomedical Devices and systemsSensors for Biomedical Devices and systems
Sensors for Biomedical Devices and systems
 
Transducers for bio medical
Transducers for bio medicalTransducers for bio medical
Transducers for bio medical
 
Amplifiers and biopotential amplifiers new
Amplifiers and biopotential amplifiers newAmplifiers and biopotential amplifiers new
Amplifiers and biopotential amplifiers new
 
biomedical signal processing
biomedical signal processingbiomedical signal processing
biomedical signal processing
 
Unit 3 biomedical
Unit 3 biomedicalUnit 3 biomedical
Unit 3 biomedical
 
Sensory Substitution
Sensory SubstitutionSensory Substitution
Sensory Substitution
 
safety from electric shock
safety from electric shocksafety from electric shock
safety from electric shock
 
Transducers
TransducersTransducers
Transducers
 

Similar to electrical safety_lecture.pdf

1728511503435_Electrical Safety Analyzer.ppt
1728511503435_Electrical Safety Analyzer.ppt1728511503435_Electrical Safety Analyzer.ppt
1728511503435_Electrical Safety Analyzer.pptRGCE
 
Electrical safety in OR
Electrical safety in ORElectrical safety in OR
Electrical safety in ORYamini Dhengle
 
ELECTRICAL SAFETY pptx
ELECTRICAL SAFETY pptxELECTRICAL SAFETY pptx
ELECTRICAL SAFETY pptxmohamedrifky10
 
Energy devices in surgery
Energy devices in surgeryEnergy devices in surgery
Energy devices in surgeryRam Raksha
 
ELECTRICAL SAFETY-corrected.pptx
ELECTRICAL SAFETY-corrected.pptxELECTRICAL SAFETY-corrected.pptx
ELECTRICAL SAFETY-corrected.pptxmohamedrifky10
 
ELECTRICAL SAFETY IN OPERATION THEATRE .pptx
ELECTRICAL SAFETY IN OPERATION THEATRE .pptxELECTRICAL SAFETY IN OPERATION THEATRE .pptx
ELECTRICAL SAFETY IN OPERATION THEATRE .pptxSujata Walode
 
Electrical safety training
Electrical safety trainingElectrical safety training
Electrical safety trainingTarkeeb
 
Report on Electrical Safety
Report on Electrical SafetyReport on Electrical Safety
Report on Electrical Safetyson_nikhil
 
General Electrical Safety Training by BGSU
General Electrical Safety Training by BGSUGeneral Electrical Safety Training by BGSU
General Electrical Safety Training by BGSUAtlantic Training, LLC.
 
electrical engineering, electrical fire, and safety, electrical safety audits...
electrical engineering, electrical fire, and safety, electrical safety audits...electrical engineering, electrical fire, and safety, electrical safety audits...
electrical engineering, electrical fire, and safety, electrical safety audits...Marcep Inc.
 
Electrical safety for anesthesiologists
Electrical safety for anesthesiologistsElectrical safety for anesthesiologists
Electrical safety for anesthesiologistsnibas_700
 
electricalsafety-150402105714-conversion-gate01 (2).pdf
electricalsafety-150402105714-conversion-gate01 (2).pdfelectricalsafety-150402105714-conversion-gate01 (2).pdf
electricalsafety-150402105714-conversion-gate01 (2).pdfMubshiraTC2
 
She training module electrical safety
She training module electrical safetyShe training module electrical safety
She training module electrical safetysunilkumarkushwaha12
 

Similar to electrical safety_lecture.pdf (20)

1728511503435_Electrical Safety Analyzer.ppt
1728511503435_Electrical Safety Analyzer.ppt1728511503435_Electrical Safety Analyzer.ppt
1728511503435_Electrical Safety Analyzer.ppt
 
ES.pptx
ES.pptxES.pptx
ES.pptx
 
Electrical safety in OR
Electrical safety in ORElectrical safety in OR
Electrical safety in OR
 
ELECTRICAL SAFETY pptx
ELECTRICAL SAFETY pptxELECTRICAL SAFETY pptx
ELECTRICAL SAFETY pptx
 
Energy devices in surgery
Energy devices in surgeryEnergy devices in surgery
Energy devices in surgery
 
Electrical safety
Electrical safetyElectrical safety
Electrical safety
 
ELECTRICAL SAFETY.pptx
ELECTRICAL SAFETY.pptxELECTRICAL SAFETY.pptx
ELECTRICAL SAFETY.pptx
 
ELECTRICAL SAFETY-corrected.pptx
ELECTRICAL SAFETY-corrected.pptxELECTRICAL SAFETY-corrected.pptx
ELECTRICAL SAFETY-corrected.pptx
 
Electric shock
Electric shockElectric shock
Electric shock
 
ELECTRICAL SAFETY IN OPERATION THEATRE .pptx
ELECTRICAL SAFETY IN OPERATION THEATRE .pptxELECTRICAL SAFETY IN OPERATION THEATRE .pptx
ELECTRICAL SAFETY IN OPERATION THEATRE .pptx
 
Electrical safety training
Electrical safety trainingElectrical safety training
Electrical safety training
 
Report on Electrical Safety
Report on Electrical SafetyReport on Electrical Safety
Report on Electrical Safety
 
Elecrosurgery in hysteroscopy
Elecrosurgery in hysteroscopyElecrosurgery in hysteroscopy
Elecrosurgery in hysteroscopy
 
electrical-safety.pptx
electrical-safety.pptxelectrical-safety.pptx
electrical-safety.pptx
 
General Electrical Safety Training by BGSU
General Electrical Safety Training by BGSUGeneral Electrical Safety Training by BGSU
General Electrical Safety Training by BGSU
 
Energy sources in surgery
Energy sources in surgery Energy sources in surgery
Energy sources in surgery
 
electrical engineering, electrical fire, and safety, electrical safety audits...
electrical engineering, electrical fire, and safety, electrical safety audits...electrical engineering, electrical fire, and safety, electrical safety audits...
electrical engineering, electrical fire, and safety, electrical safety audits...
 
Electrical safety for anesthesiologists
Electrical safety for anesthesiologistsElectrical safety for anesthesiologists
Electrical safety for anesthesiologists
 
electricalsafety-150402105714-conversion-gate01 (2).pdf
electricalsafety-150402105714-conversion-gate01 (2).pdfelectricalsafety-150402105714-conversion-gate01 (2).pdf
electricalsafety-150402105714-conversion-gate01 (2).pdf
 
She training module electrical safety
She training module electrical safetyShe training module electrical safety
She training module electrical safety
 

More from Shashikanth Boorla

IPC_PPT_CDT4_PCS_Process response.ppt
IPC_PPT_CDT4_PCS_Process response.pptIPC_PPT_CDT4_PCS_Process response.ppt
IPC_PPT_CDT4_PCS_Process response.pptShashikanth Boorla
 
IPC_PPT_CDT1-1_PCS & importance.ppt
IPC_PPT_CDT1-1_PCS & importance.pptIPC_PPT_CDT1-1_PCS & importance.ppt
IPC_PPT_CDT1-1_PCS & importance.pptShashikanth Boorla
 
IPC_PPT_CDT_PCS_Process model & response of Interacting system.ppt
IPC_PPT_CDT_PCS_Process model & response of Interacting system.pptIPC_PPT_CDT_PCS_Process model & response of Interacting system.ppt
IPC_PPT_CDT_PCS_Process model & response of Interacting system.pptShashikanth Boorla
 
Interacting & non Interacting.pptx
Interacting & non Interacting.pptxInteracting & non Interacting.pptx
Interacting & non Interacting.pptxShashikanth Boorla
 
Geometric Configurations for Robots.pptx
Geometric Configurations for Robots.pptxGeometric Configurations for Robots.pptx
Geometric Configurations for Robots.pptxShashikanth Boorla
 
Economic and Social Aspects of Robotics.pptx
Economic and Social Aspects of Robotics.pptxEconomic and Social Aspects of Robotics.pptx
Economic and Social Aspects of Robotics.pptxShashikanth Boorla
 

More from Shashikanth Boorla (7)

IPC_PPT_CDT4_PCS_Process response.ppt
IPC_PPT_CDT4_PCS_Process response.pptIPC_PPT_CDT4_PCS_Process response.ppt
IPC_PPT_CDT4_PCS_Process response.ppt
 
IPC_PPT_CDT1-1_PCS & importance.ppt
IPC_PPT_CDT1-1_PCS & importance.pptIPC_PPT_CDT1-1_PCS & importance.ppt
IPC_PPT_CDT1-1_PCS & importance.ppt
 
IPC_PPT_CDT_PCS_Process model & response of Interacting system.ppt
IPC_PPT_CDT_PCS_Process model & response of Interacting system.pptIPC_PPT_CDT_PCS_Process model & response of Interacting system.ppt
IPC_PPT_CDT_PCS_Process model & response of Interacting system.ppt
 
Interacting & non Interacting.pptx
Interacting & non Interacting.pptxInteracting & non Interacting.pptx
Interacting & non Interacting.pptx
 
types of robots.pptx
types of robots.pptxtypes of robots.pptx
types of robots.pptx
 
Geometric Configurations for Robots.pptx
Geometric Configurations for Robots.pptxGeometric Configurations for Robots.pptx
Geometric Configurations for Robots.pptx
 
Economic and Social Aspects of Robotics.pptx
Economic and Social Aspects of Robotics.pptxEconomic and Social Aspects of Robotics.pptx
Economic and Social Aspects of Robotics.pptx
 

Recently uploaded

Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknowgragteena
 
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...delhimodelshub1
 
Jalandhar Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
Jalandhar  Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...Jalandhar  Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
Jalandhar Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...Call Girls Service Chandigarh Ayushi
 
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...High Profile Call Girls Chandigarh Aarushi
 
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...High Profile Call Girls Chandigarh Aarushi
 
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliCall Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliHigh Profile Call Girls Chandigarh Aarushi
 
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service HyderabadCall Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabaddelhimodelshub1
 
Call Girls Service Chandigarh Gori WhatsApp ❤9115573837 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤9115573837 VIP Call Girls Chandi...Call Girls Service Chandigarh Gori WhatsApp ❤9115573837 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤9115573837 VIP Call Girls Chandi...Niamh verma
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxAyush Gupta
 
Dehradun Call Girls Service ❤️🍑 9675010100 👄🫦Independent Escort Service Dehradun
Dehradun Call Girls Service ❤️🍑 9675010100 👄🫦Independent Escort Service DehradunDehradun Call Girls Service ❤️🍑 9675010100 👄🫦Independent Escort Service Dehradun
Dehradun Call Girls Service ❤️🍑 9675010100 👄🫦Independent Escort Service DehradunNiamh verma
 
Hot Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
Hot  Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In ChandigarhHot  Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
Hot Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In ChandigarhVip call girls In Chandigarh
 
Leading transformational change: inner and outer skills
Leading transformational change: inner and outer skillsLeading transformational change: inner and outer skills
Leading transformational change: inner and outer skillsHelenBevan4
 
Call Girl Hyderabad Madhuri 9907093804 Independent Escort Service Hyderabad
Call Girl Hyderabad Madhuri 9907093804 Independent Escort Service HyderabadCall Girl Hyderabad Madhuri 9907093804 Independent Escort Service Hyderabad
Call Girl Hyderabad Madhuri 9907093804 Independent Escort Service Hyderabaddelhimodelshub1
 
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsiindian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana TulsiHigh Profile Call Girls Chandigarh Aarushi
 

Recently uploaded (20)

Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
 
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
 
Jalandhar Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
Jalandhar  Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...Jalandhar  Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
Jalandhar Female Call Girls Contact Number 9053900678 💚Jalandhar Female Call...
 
Call Girls in Lucknow Esha 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
Call Girls in Lucknow Esha 🔝 8923113531  🔝 🎶 Independent Escort Service LucknowCall Girls in Lucknow Esha 🔝 8923113531  🔝 🎶 Independent Escort Service Lucknow
Call Girls in Lucknow Esha 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
 
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
 
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
 
Russian Call Girls in Dehradun Komal 🔝 7001305949 🔝 📍 Independent Escort Serv...
Russian Call Girls in Dehradun Komal 🔝 7001305949 🔝 📍 Independent Escort Serv...Russian Call Girls in Dehradun Komal 🔝 7001305949 🔝 📍 Independent Escort Serv...
Russian Call Girls in Dehradun Komal 🔝 7001305949 🔝 📍 Independent Escort Serv...
 
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
 
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliCall Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
 
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service HyderabadCall Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
 
Call Girls Service Chandigarh Gori WhatsApp ❤9115573837 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤9115573837 VIP Call Girls Chandi...Call Girls Service Chandigarh Gori WhatsApp ❤9115573837 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤9115573837 VIP Call Girls Chandi...
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptx
 
Dehradun Call Girls Service ❤️🍑 9675010100 👄🫦Independent Escort Service Dehradun
Dehradun Call Girls Service ❤️🍑 9675010100 👄🫦Independent Escort Service DehradunDehradun Call Girls Service ❤️🍑 9675010100 👄🫦Independent Escort Service Dehradun
Dehradun Call Girls Service ❤️🍑 9675010100 👄🫦Independent Escort Service Dehradun
 
Hot Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
Hot  Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In ChandigarhHot  Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
Hot Call Girl In Chandigarh 👅🥵 9053'900678 Call Girls Service In Chandigarh
 
Leading transformational change: inner and outer skills
Leading transformational change: inner and outer skillsLeading transformational change: inner and outer skills
Leading transformational change: inner and outer skills
 
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service DehradunCall Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
 
Call Girl Hyderabad Madhuri 9907093804 Independent Escort Service Hyderabad
Call Girl Hyderabad Madhuri 9907093804 Independent Escort Service HyderabadCall Girl Hyderabad Madhuri 9907093804 Independent Escort Service Hyderabad
Call Girl Hyderabad Madhuri 9907093804 Independent Escort Service Hyderabad
 
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service GuwahatiCall Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
 
Russian Call Girls South Delhi 9711199171 discount on your booking
Russian Call Girls South Delhi 9711199171 discount on your bookingRussian Call Girls South Delhi 9711199171 discount on your booking
Russian Call Girls South Delhi 9711199171 discount on your booking
 
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsiindian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
 

electrical safety_lecture.pdf

  • 2. This Week  Safety in the clinical environment: Electrical safety  Physiological effects of electricity  Susceptibility parameters  Distribution of electrical power  Isolated power systems  Macroshock hazards  Microshock hazards  Electrical safety codes and standards  Protection  Power distribution  Ground fault circuit interrupters (GFCI)  Equipment design  Electrical safety analyzers / Testing electrical systems
  • 3. Safety in Clinical Environment  Electrical hazards  Electrical shocks (micro and macro) due to equipment failure, failure of power delivery systems, ground failures, burns, fire, etc.  Mechanical hazards  mobility aids, transfer devices, prosthetic devices, mechanical assist devices, patient support devices  Environmental hazards  Solid wastes, noise, utilities (natural gas), building structures, etc.  Biological hazards  Infection control, viral outbreak, isolation, decontamination, sterilization, waste disposal issues  Radiation hazards  Use of radioactive materials, radiation devices (MRI, CT, PET), exposure control
  • 4. Electrical Safety  Many sources of energy, potentially hazardous substances, instruments and procedures  Use of fire, compressed air, water, chemicals, drugs, microorganisms, waste, sound, electricity, radiation, natural and unnatural disaster, negligence, sources of radiation, etc.  Medical procedures expose patients to increased risks of hazards due to skin and membranes being penetrated / altered  10,000 device related injuries in the US every year! Typically due to  Improper use  Inadequate training  Lack of experience  Improper (lack of) use of manuals  Device failure
  • 5. Physiological Effects of Electricity  For electricity to have an effect on the human body:  An electrical potential difference must be present  The individual must be part of the electrical circuit, that is, a current must enter the body at one point and leave it at another.  However, what causes the physiological effect is NOT voltage, but rather CURRENT.  A high voltage (KV,(103V)) applied over a large impedance (rough skin) may not cause much (any) damage  A low voltage applied over very small impedances (heart tissue) may cause grave consequences (ventricular fibrillation)  The magnitude of the current is simply the applied voltage divided by the total effective impedance the current faces; skin : largest.  Electricity can have one of three effects:  Electrical stimulation of excitable tissue (muscles, nerve)  Resistive heating of tissue  Electrical burns / tissue damage for direct current and high voltages
  • 6. Physiological Effects of Electricity Physiological effects of electricity. Threshold or estimated mean values are given for each effect in a 70 kg human for a 1 to 3 s exposure to 60 Hz current applied via copper wires grasped by the hands. Dry skin impedance:93 kΩ / cm2 Electrode gel on skin: 10.8 kΩ / cm2 Penetrated skin: 200 Ω / cm2 The real physiological effect depends on the actual path of the current
  • 7.  Threshold of perception: The minimal current that an individual can detect. For AC (with wet hands) can be as small as 0.5 mA at 60 Hz. For DC, 2 ~10 mA  Let-go current: The maximal current at which the subject can voluntarily withdraw. 6 ~ 100 mA, at which involuntary muscle contractions, reflex withdrawals, secondary physical effects (falling, hitting head) may also occur  Respiratory Paralysis / Pain / Fatigue At as low as 20 mA, involuntary contractions of respiratory muscles can cause asphyxiation / respiratory arrest, if the current is not interrupted. Strong involuntary contraction of other muscles can cause pain and fatigue  Ventricular fibrillation 75 ~ 400 mA can cause heart muscles to contract uncontrollably, altering the normal propagation of the electrical activity of the heart. HR can raise up to 300 bpm, rapid, disorganized and too high to pump any meaningful amount of blood  ventricular fibrillation. Normal rhythm can only return using a defibrillator  Sustained myocardial contraction / Burns and physical injury At 1 ~6 A, the entire heart muscle contracts and heart stops beating. This will not cause irreversible tissue damage, however, as normal rhythm will return once the current is removed. At or after 10A, however, burns can occur, particularly at points of entry and exit. Physiological Effects of Electricity
  • 8. Important Susceptibility Parameters  Threshold and let-go current variability Distributions of perception thresholds and let-go currents These data depend on surface area of contact, moistened hand grasping AWG No. 8 copper wire, 70 kg human, 60Hz, 1~3 s. of exposure
  • 9. Important Susceptibility Parameters  Frequency  Note that the minimal let-go current happens at the precise frequency of commercial power-line, 50-60Hz.  Let-go current rises below 10 Hz and above several hundred Hz. Let-go current versus frequency Percentile values indicate variability of let-go current among individuals. Let- go currents for women are about two- thirds the values for men.
  • 10. Important Susceptibility Parameters  Duration  The longer the duration, the smaller the current at which ventricular fibrillation occurs  Shock must occur long enough to coincide with the most vulnerable period occurring during the T wave.  Weight  Fibrillation threshold increases with body weight (from 50mA for 6kg dogs to 130 mA for 24 kg dogs. Fibrillation current versus shock duration. Thresholds for ventricular fibrillation in animals for 60 Hz AC current. Duration of current (0.2 to 5 s) and weight of animal body were varied.
  • 11. Important Susceptibility Parameters  Points of entry  The magnitude of the current required to fibrillate the heart is far greater if the current is applied directly to heart; externally applied current loses much of its amplitude due to current distributions. Large, externally applied currents cause macroshock.  If catheters are used, the natural protection provided by the skin 15 kΩ ~ 2 MΩ) is bypassed, greatly reducing the amount of current req’d to cause fibrillation. Even smallest currents (80 ~ 600 μA), causing microshock, may result in fibrillation. Safety limit for microshocks is 10 μA.  The precise point of entry, even externally is very important: If both points of entry and exit are on the same extremity, the risk of fibrillation is greatly reduced even at high currents (e.g. the current req’d for fibrillation through Lead I (LA-RA) electrodes is higher than for Leads II (LL-RA) and III (LL-LA).
  • 12. Important Susceptibility Parameters  Points of entry Effect of entry points on current distribution (a) Macroshock, externally applied current spreads through- out the body. (b) Microshock, all the current applied through an intracardiac catheter flows through the heart.
  • 13. Distribution of Electrical Power Simplified electric-power distribution for 115 V circuits. Power frequency is 60 Hz (230 V)
  • 14. Distribution of Power  If electrical devices were perfect, only two wires would be adequate (hot and return), with all power confined to these two wires. However, there are two major departures from this ideal case: • A fault may occur, through miswiring, component failure, etc., causing an electrical potential between an exposed surface (metal casing of the device) and a grounded surface (wet floor, metal case of another device etc.) Any person who bridges these two surfaces is subject to macroshock. • Even if a fault does not occur, imperfect insulation or electromagnetic coupling (capacitive or inductive) may produce an electrical potential relative to the ground. A susceptible patient providing a path for this leakage current to flow to the ground is subject to microshock. • The additional ground line provides a good line of defense! (how / why?)
  • 15. GROUND!  The additional line that is connected directly to the earth-ground provides the following:  In case of a fault (short circuit between hot conductor and metal casing), a large current will use the path through the ground wire (instead of the patient) and not only protect the patient, but also cause the circuit breaker to open. The ability of the grounding system to conduct high currents to ground is crucial for this to work!  If there is no fault, the ground wire serves to conduct the leakage current safely back to the electrical power source – again, as long as the grounding system provides a low-resistance pathway to the ground  Leakage current recommended by ECRI are established to prevent injury in case the grounding system fails and a patient touches an electrically active surface (10 ~ 100 μA).
  • 16. Isolated Power Distribution Not grounded ! • Normally, when there is a ground-fault from hot wire to ground, a large current is drawn causing a potential hazard, as the device will stop functioning when the circuit breakers open ! • This can be prevented by using the isolated system, which separates ground from neutral, making neutral and hot electrically identical. A single ground-fault will not cause large currents, as long as both hot conductors are initially isolated from ground! • In fact, in such an isolated system, if a single ground-fault occurs, the system simply reverts back to the normal ground-referenced system. •A line isolation monitor is used with such system that continuously monitors for the first ground fault, during which case it simply informs the operators to fix the problem. The single ground fault does NOT constitute a hazard!
  • 17. Macroshock  Most electrical devices have a metal cabinet, which constitutes a hazard, in case of an insulation failure or shortened component between the hot power lead and the chassis. There is then 115 ~ 230 V between the chassis and any other grounded object.  The first line of defense available to patients is their skin.  The outer layer provides 15 kΩ to 1 MΩ depending on the part of the body, moisture and sweat present, 1% of that of dry skin if skin is broken,  Internal resistance of the body is 200Ω for each limb, and 100Ω for the trunk, thus internal body resistance between any two limbs is about 500Ω (somewhat higher for obese people due to high resistivity of the adipose tissue)!  Any procedure that reduces or eliminates the skin resistance increases the risk of electrical shock, including biopotential electrode gel, electronic thermometers placed in ears, mouth, rectum, intravenous catheters, etc.  A third wire, grounded to earth, can greatly reduce the effect of macroshock, as the resistance of that path would be much smaller then even that of internal body resistance!
  • 18. Macroshock Hazards • Direct faults between the hot conductor and the ground is not common, and technically speaking, ground connection is not necessary during normal operation. • In fact, a ground fault will not be detected during normal operation of the device, only when someone touches it, the hazard becomes known. Therefore, ground wire in devices and receptacles must be periodically tested.
  • 20. Microshock Hazards Small currents inevitably flow between adjacent insulated conductors at different potentials  leakage currents which flow through stray capacitances, insulation, dust and moisture Leakage current flowing to the chassis flows safely to the ground, if a low-resistance ground wire is available.
  • 21. Microshock Hazards • If ground wire is broken, the chassis potential rises above the ground; a patient who has a grounded connection to the heart (e.g. through a catheter) receives a microshock if s/he touches the chassis. • If there is a connection from the chassis to the patient’s heart, and a connection to the ground anywhere in the body, this also causes microshock. • Note that the hazard for microshock only exists if there is a direct connection to the heart. Otherwise, even the internal resistance of the body is high enough top prevent the microshocks.
  • 22. Microshock via Ground Potentials Microshocks can also occur if different devices are not at the exact same ground potential. In fact, the microshock can occur even when a device that does not connected to the patient has a ground fault! A fairly common ground wire resistance of 0.1Ω can easily cause a a 500mV potential difference if initiated due to a, say 5A of ground fault. If the patient resistance is less then 50kΩ, this would cause an above safe current of 10μA
  • 24. Safety Codes & Standards  Limits on leakage current are instituted and regulated by the safety codes instituted in part by the National Fire Protection Association (NFPA), American National Standards Institute (ANSI), Association for the Advancement of Medical Instrumentation (AAMI), and Emergency Care Research Institute (ECRI).
  • 25. Basic Approaches to Shock Protection  There are two major ways to protect patients from shocks:  Completely isolate and insulate patient from all sources of electric current  Keep all conductive surfaces within reach of the patient at the same voltage  Neither can be fully achieved  some combination of these two  Grounding system  Isolated power-distribution system  Ground-fault circuit interrupters (GFCI)
  • 26. Grounding Systems Low resistance (0.15 Ω) ground that can carry currents up to the circuit-breaker ratings protects patients by keeping all conductive surfaces and receptacle grounds at the same potential. Protects patients from • Macroshocks • Microshocks • Ground faults elsewhere (!) The difference between the receptacle grounds and other surface should be no more then 40 mV) All the receptacle grounds and conductive surfaces in the vicinity of the patient are connected to the patient-equipment grounding point. Each patient- equipment grounding point is connected to the reference grounding point that makes a single connection to the building ground.
  • 27. Isolated Power Systems  A good equipotential grounding system cannot eliminate large current that may result from major ground-faults (which are rather rare).  Isolated power systems can protect against such major (single) ground faults  Provide considerable protection against macroshocks, particularly around wet conditions  However, they are expensive !  Used only at locations where flammable anesthetics are used. Additional minor protection against microshocks does not justify the high cost of these systems to be used everywhere in the clinical environment
  • 28. Ground – Fault Circuit Interrupters (GFCI)  Disconnects source of electric current when a ground fault greater than about 6 mA occurs! When there is no fault, Ihot=Ineutral. The GFCI detects the difference between these two currents. If the difference is above a threshold, that means the rest of the current must be flowing through elsewhere, either the chassis or the patient !!!. The detection is done through the monitoring the voltage induced by the two coils (hot and neutral) in the differential transformer!
  • 29. GFCI The National Electric Code (NEC - 1996) requires that all circuits serving bathrooms, garages, outdoor receptacles, swimming pools and construction sites be fitted with GFCI. Note that GFCI protect against major ground faults only, not against microshocks. Patient care areas are typically not fitted with GFCI, since the loss of power to life support equipment can also be equally deadly!
  • 30. Protection through Equipment Design  Strain-relief devices for cords, where cord enters the equipment and between the cord and plug  Reduction of leakage current through proper layout and insulation to minimize the capacitance between all hot conductors and the chassis  Double insulation to prevent the contact of the patient with the chassis or any other conducting surface (outer case being insulating material, plastic knobs, etc.)  Operation at low voltages; solid state devices operating at <10V are far less likely to cause macroshocks  Electrical isolation in circuit design
  • 31. Electrical Isolation CM CM CMRR SIG ISO ISO Error Isolation barrier Isolation Capacitance and resistance - + - + Input common (a) *IMRR in v/v Output common o = CM CMRR ISO IMRR SIG ± Gain ± RF ISO IMRR* ~ ~ ~ ~ ~ Error • Main features of an isolation amplifier: • High ohmic isolation between input and output (>10MΩ) • High isolation mode voltage (>1000V) • High common mode rejection ration (>100 dB)
  • 32. Transformer Isolation Amplifiers - + (b) + 15 V DC o Power return 25 kHz - 7.5 V +ISO Out SIG -ISO Out + 7.5 V In com In - In + FB ± 5 V F.S. Oscillator 25 kHz Signal Mod Rect and filter Power Demod ± 5 V F.S. AD202 Hi Lo
  • 33. Optical Isolation Amplifier Isolation barrier Input control Output control -V +V +o - + o = i RK RG CR3 CR1 CR2 i2 i i i i1 RG AI AII i3 i2 o RK = 1M W - + - + + - (c) ~ 1 2
  • 34. Electrical Safety Analyzers Wiring / Receptacle Testing  Three LED receptacle tester:  Simple device used to test common wiring problems (can detect only 8 of possible 64 states)  Will not detect ground/neutral reversal, or when ground/neutral are hot and hot is grounded (GFCI would detect the latter)
  • 35. Electrical Safety Analyzers Testing Electrical Appliances  Ground-pin-to-chassis resistance: Should be <0.15Ω during the life of the appliance Ground-pin-to-chassis resistance test
  • 36. Electrical Safety Analyzers Testing Electrical Appliances  Chassis leakage current: The leakage current should not exceed 500μA with single fault for devices not intended for patient contact, and not exceed 300 μA for those that are intended for patient contact. To exposed conductive surface or if none, then 10 by 20 cm metal foil in contact with the exposed surface Insulating surface Current meter I Test circuit Open switch for appliances not intended to contact a patient Grounding-contact switch (use in OPEN position) Polarity- reversing switch (use both positions) Appliance power switch (use both OFF and ON positions) This connection is at service entrance or on supply side of separately derived system Building ground H (black) N (white) Appliance G N H H = hot N = neutral (grounded) G = grounding conductor I < 500 μA for facility Ðowned housekeeping and maintenance appliances I > 300 μA for appliances intended for use in the patient vicinity 120 V G (green) Internal Circuitry
  • 37. Electrical Safety Analyzers Testing Electrical Appliances  Leakage current in patient leads:  Potentially most damaging leakage is the one with patient leads, since they typically have low impedance patient contacts  Current should be restricted to 50μA for non-isolated leads and to 10 μA for isolated leads (used with catheters / electrodes that make connection to the heart)  Leakage current between any pair of leads, or between a single lead and other patient connections should also be controlled  Leakage in case of line voltage appearing on the patient should also be restricted.
  • 38. Leakage current Testers Test for leakage current from patient leads to ground
  • 39. Leakage Current testers Test for leakage current between patient leads
  • 40. Leakage Current Testers Test for ac isolation current Isolation current is the current that passes through patient leads to ground if and when line voltage appears on the patient. This should also be limited to 50μA