SlideShare a Scribd company logo
1 of 4
5.0 RESULT / DATA ANALYSIS
Table 1: Absorbance value at 516 nm of Carmoisine Standard at different concentration
Concentration (ppm) Absorbance (516 nm) Molar absorptivity (L.mol−1
. cm−1
)
0 (Blank) 0 0
5 0.342 3.4366 × 104
15 0.739 2.4753 × 104
25 1.106 2.2228 × 104
35 1.455 2.0887 × 104
45 1.792 2.0008 × 104
Table 2: Absorbance value at 516 nm of Unknown Samples at different concentration
Unknown Concentration (ppm) Absorbance (516 nm) Molar absorptivity
(L. mol−1
. cm−1
)
A 27.8 1.005 1.8163 × 104
B 36.4 1.317 1.8179 × 104
Figure 1: Standard calibration curve of Carmoisine
y = 0.0362x + 0.1828
R² = 0.9989
0
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
1.8
2
0 5 10 15 20 25 30 35 40 45 50
Absorbance(516nm)
Concentration (ppm)
Standart Calibration Curve of Carmoisine
6.0 DISCUSSION
The purpose of this experiment were to determine Lambda Max for Carmoisine sample
(wavelength scan). To determine ɛ the molar absorptivity of Carmoisine and to prepare a serial
dilution and generate a standard calibration graph for sample quantitation (photometric scan).
The Carmoisine standard solution samples were run by using Uv-Vis Spectrophotometer.
Ultraviolet-visible spectroscopy or ultraviolet-visible spectrophotometry (UV-Vis or UV/Vis)
refers to absorption spectroscopy in the ultraviolet-visible spectral region. This means it uses light
in the visible and adjacent (near-UV and near-infrared (NIR)) ranges. UV-Vis spectrophotometer
is used in the quantitative determination of concentrations of the absorber in the solutions of
transition metal ions and highly conjugated organic compounds (IIT Kanpur, 2012). Besides, UV
spectrophotometer principle follows the Beer-Lambert Law. This law states that whenever a beam
of monochromatic light is passed through a solution with an absorbing substance, the decreasing
rate of the radiation intensity along with the thickness of the absorbing solution is actually
proportional to the concentration of the solution and the incident radiation (Choudhary, A. 2017).
The experiment was started with 100 ppm of stock solution of Carmoisine was prepared
to dilute the stock solution into different concentration which were 5 ppm, 15 ppm, 25 ppm, 35
ppm and 45 ppm with distilled water into 50 ml volumetric flask. The volume needed to prepare
the serial dilution from 100 ppm of stock solution were 2.5 ml, 7.5 ml, 12.5 ml, 17.5 ml and
22.5 ml respectively. The volume of serial dilution were calculated by using the
formula M1V1 = M2V2. Moreover, the unknown sample was prepared with the same method
as serial dilution but the volume of stock solution must be between 2.5 ml to 22.5 ml.
After setting up the UV-Vis, the blank solution was placed into instrument and baseline
was click. After the blank solution became zero, the standard solution of 45 ppm was placed to
another cuvette. The maximum wavelength was recorded at 516 nm. The maximum
wavelength was the wavelength with the higher absorbance. For the photometric scan, the
cuvette was filled with standard solution 5 ppm to 45 ppm to determine the absorbance and
prepared for the standard calibration curve.
According to the result obtained, the absorbance reading showed for the standard
solution 5 ppm, 15 ppm, 25 ppm, 35 ppm and 45 ppm were 0.342, 0.739, 1.106, 1.455 and
1.792 respectively. Based on graph plotted, the absorbance (516 nm) was directly proportional
to the concentration (ppm). Because of when the concentration was higher, more radiation is
absorbed and the absorbance also become higher. In addition, for the unknown samples the
absorbance obtained were 1.005 and 1.317. The concentration of unknown samples have found
by using Beer’s Lambert law formula A = εbc with the slope obtained was 0.0362 which
indicates the molar absorptivity of Carmoisine. Thus, the concentration of unknown samples
were 27.8 ppm and 36.4 ppm. Furthermore, the graph showed the R2
value that is a statistical
measure of how close the data are to the fitted regression line (Ogee, A. 2013). The correlation
coefficient varies about +1 to -1. The graph showed the correlation coefficient was 0.9989
which the R2
value was nearest to the +1 which indicates the accurate of the result.
As conclusion, there are several possibilities precaution that should be considered.
Firstly, make sure use the pipette to measure the volume of Carmoisine to get the accuracy.
Secondly, make sure wipe and do not touched the clean sides of cuvette and carrying out the
set-up procedure in the correct order.
REFERENCES
1. Ogee, A. (2013). Regression Analysis: How Do I Interpret R-squared and Assess the
Goodness-of-Fit?. Retrieved from https://blog.minitab.com/blog/adventures- in-
statistics-2/regression-analysis-how-do-i-interpret-r-squared-and-assess-the-goodness-
of-fit.
2. Wenzel, T. (2019). Beer’s Law. Retrieved from
https://chem.libretexts.org/Bookshelves/Analytical_Chemistry/Supplemental_Module
s_(Analytical_Chemistry)/Analytical_Sciences_Digital_Library/Active_Learning/In_
Class_Activities.
3. IIT Kanpur. (2012). UV-Vis Spectrophotometer. Retrieved from
https://www.iitk.ac.in/dordold/index.php?option=com_content&view=category&layo
ut=blog&id=219&Itemid=238.
4.
Result & discussion uv vis (exp 3) - for merge

More Related Content

What's hot

Compaction in Tablet Manufacturing
Compaction in Tablet ManufacturingCompaction in Tablet Manufacturing
Compaction in Tablet Manufacturing
Suraj Choudhary
 

What's hot (20)

Microcapsules and microspheres
Microcapsules and microspheresMicrocapsules and microspheres
Microcapsules and microspheres
 
Microemulsion
MicroemulsionMicroemulsion
Microemulsion
 
Nanoparticles
NanoparticlesNanoparticles
Nanoparticles
 
Liposome
LiposomeLiposome
Liposome
 
Capsules
Capsules Capsules
Capsules
 
Infrared Spectrophotometry.ppt
Infrared Spectrophotometry.pptInfrared Spectrophotometry.ppt
Infrared Spectrophotometry.ppt
 
Ion exchange chromatography - by Shree
Ion   exchange chromatography - by ShreeIon   exchange chromatography - by Shree
Ion exchange chromatography - by Shree
 
Liposomes
Liposomes Liposomes
Liposomes
 
Polymeric Micelle
Polymeric MicellePolymeric Micelle
Polymeric Micelle
 
Niosomes
NiosomesNiosomes
Niosomes
 
High performed liquid chromatography (HPLC) or High pressure liquid chromatog...
High performed liquid chromatography (HPLC) or High pressure liquid chromatog...High performed liquid chromatography (HPLC) or High pressure liquid chromatog...
High performed liquid chromatography (HPLC) or High pressure liquid chromatog...
 
Micro emulsions and multiple emulsions
Micro emulsions and multiple emulsionsMicro emulsions and multiple emulsions
Micro emulsions and multiple emulsions
 
Soaps and syndetbars
Soaps and syndetbarsSoaps and syndetbars
Soaps and syndetbars
 
SUSPENSION.ppt
SUSPENSION.pptSUSPENSION.ppt
SUSPENSION.ppt
 
Compaction in Tablet Manufacturing
Compaction in Tablet ManufacturingCompaction in Tablet Manufacturing
Compaction in Tablet Manufacturing
 
ROLE OF LIPOSOME IN NOVEL DRUG DELIVERY SYSTEM
ROLE OF LIPOSOME IN NOVEL DRUG DELIVERY SYSTEMROLE OF LIPOSOME IN NOVEL DRUG DELIVERY SYSTEM
ROLE OF LIPOSOME IN NOVEL DRUG DELIVERY SYSTEM
 
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
 
Microspheres
MicrospheresMicrospheres
Microspheres
 
Micronization - Solubility Enhancement by Micronization
Micronization - Solubility Enhancement by MicronizationMicronization - Solubility Enhancement by Micronization
Micronization - Solubility Enhancement by Micronization
 
Viscosity measurements
Viscosity measurementsViscosity measurements
Viscosity measurements
 

Similar to Result & discussion uv vis (exp 3) - for merge

a simple multi-utility miniature device used as test tube to 3D cell culture ...
a simple multi-utility miniature device used as test tube to 3D cell culture ...a simple multi-utility miniature device used as test tube to 3D cell culture ...
a simple multi-utility miniature device used as test tube to 3D cell culture ...
guo jun chen
 
exp 2 adsorbtion from solution
exp 2 adsorbtion from solutionexp 2 adsorbtion from solution
exp 2 adsorbtion from solution
Rashid Alsuwaidi
 
Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...
Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...
Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...
ijtsrd
 
Lab1 spectrophotometry
Lab1 spectrophotometryLab1 spectrophotometry
Lab1 spectrophotometry
Cody Taylor
 
AppNote_MFI Manual Viscosity Study
AppNote_MFI Manual Viscosity StudyAppNote_MFI Manual Viscosity Study
AppNote_MFI Manual Viscosity Study
Colin Merrifield
 
To estimate the concentration of KMnO4 and CuSO4 solutions, colorimetrically
To estimate the concentration of KMnO4 and CuSO4 solutions, colorimetricallyTo estimate the concentration of KMnO4 and CuSO4 solutions, colorimetrically
To estimate the concentration of KMnO4 and CuSO4 solutions, colorimetrically
Mithil Fal Desai
 
Scanned with CamScanner1 STANDARIZATION OF A B.docx
Scanned with CamScanner1 STANDARIZATION OF A B.docxScanned with CamScanner1 STANDARIZATION OF A B.docx
Scanned with CamScanner1 STANDARIZATION OF A B.docx
todd331
 
Experiment A- Dye Adsorption Technical Memo
Experiment A- Dye Adsorption Technical MemoExperiment A- Dye Adsorption Technical Memo
Experiment A- Dye Adsorption Technical Memo
Vanessa Ferrero
 
Analysis of analgesics and antipyretics.indu
Analysis of analgesics and  antipyretics.induAnalysis of analgesics and  antipyretics.indu
Analysis of analgesics and antipyretics.indu
hdghcfgfgftf
 
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET By Ninad M...
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET  By Ninad M...SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET  By Ninad M...
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET By Ninad M...
Ninad Phadke
 

Similar to Result & discussion uv vis (exp 3) - for merge (20)

a simple multi-utility miniature device used as test tube to 3D cell culture ...
a simple multi-utility miniature device used as test tube to 3D cell culture ...a simple multi-utility miniature device used as test tube to 3D cell culture ...
a simple multi-utility miniature device used as test tube to 3D cell culture ...
 
Reducing the Effects of Sample Solvent on UHPLC Analyses
Reducing the Effects of Sample Solvent on UHPLC AnalysesReducing the Effects of Sample Solvent on UHPLC Analyses
Reducing the Effects of Sample Solvent on UHPLC Analyses
 
exp 2 adsorbtion from solution
exp 2 adsorbtion from solutionexp 2 adsorbtion from solution
exp 2 adsorbtion from solution
 
Stability indicating method development and validation for the estimation of ...
Stability indicating method development and validation for the estimation of ...Stability indicating method development and validation for the estimation of ...
Stability indicating method development and validation for the estimation of ...
 
Application of UV-Visible spectroscopy
Application of UV-Visible spectroscopyApplication of UV-Visible spectroscopy
Application of UV-Visible spectroscopy
 
Uv and visible appications
Uv and visible appicationsUv and visible appications
Uv and visible appications
 
Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...
Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...
Validated RP-HPLC Method for the Determination of Nelaribine in Bulk and Tabl...
 
Lab1 spectrophotometry
Lab1 spectrophotometryLab1 spectrophotometry
Lab1 spectrophotometry
 
B37
B37B37
B37
 
Titrimetric Methods
Titrimetric Methods Titrimetric Methods
Titrimetric Methods
 
Serial Dilution Vs Accuracy
Serial Dilution Vs AccuracySerial Dilution Vs Accuracy
Serial Dilution Vs Accuracy
 
Spectrophotometry Lecture
Spectrophotometry LectureSpectrophotometry Lecture
Spectrophotometry Lecture
 
AppNote_MFI Manual Viscosity Study
AppNote_MFI Manual Viscosity StudyAppNote_MFI Manual Viscosity Study
AppNote_MFI Manual Viscosity Study
 
To estimate the concentration of KMnO4 and CuSO4 solutions, colorimetrically
To estimate the concentration of KMnO4 and CuSO4 solutions, colorimetricallyTo estimate the concentration of KMnO4 and CuSO4 solutions, colorimetrically
To estimate the concentration of KMnO4 and CuSO4 solutions, colorimetrically
 
Estimation of chromium (vi) by spectrophotometric method
Estimation of chromium (vi) by spectrophotometric methodEstimation of chromium (vi) by spectrophotometric method
Estimation of chromium (vi) by spectrophotometric method
 
Determination of Chloramphenicol in Bulk Drug and Pharmaceutical Dosage Forms...
Determination of Chloramphenicol in Bulk Drug and Pharmaceutical Dosage Forms...Determination of Chloramphenicol in Bulk Drug and Pharmaceutical Dosage Forms...
Determination of Chloramphenicol in Bulk Drug and Pharmaceutical Dosage Forms...
 
Scanned with CamScanner1 STANDARIZATION OF A B.docx
Scanned with CamScanner1 STANDARIZATION OF A B.docxScanned with CamScanner1 STANDARIZATION OF A B.docx
Scanned with CamScanner1 STANDARIZATION OF A B.docx
 
Experiment A- Dye Adsorption Technical Memo
Experiment A- Dye Adsorption Technical MemoExperiment A- Dye Adsorption Technical Memo
Experiment A- Dye Adsorption Technical Memo
 
Analysis of analgesics and antipyretics.indu
Analysis of analgesics and  antipyretics.induAnalysis of analgesics and  antipyretics.indu
Analysis of analgesics and antipyretics.indu
 
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET By Ninad M...
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET  By Ninad M...SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET  By Ninad M...
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET By Ninad M...
 

Recently uploaded

Integrated Test Rig For HTFE-25 - Neometrix
Integrated Test Rig For HTFE-25 - NeometrixIntegrated Test Rig For HTFE-25 - Neometrix
Integrated Test Rig For HTFE-25 - Neometrix
Neometrix_Engineering_Pvt_Ltd
 
Verification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptxVerification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptx
chumtiyababu
 

Recently uploaded (20)

Online electricity billing project report..pdf
Online electricity billing project report..pdfOnline electricity billing project report..pdf
Online electricity billing project report..pdf
 
kiln thermal load.pptx kiln tgermal load
kiln thermal load.pptx kiln tgermal loadkiln thermal load.pptx kiln tgermal load
kiln thermal load.pptx kiln tgermal load
 
HAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKAR
HAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKARHAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKAR
HAND TOOLS USED AT ELECTRONICS WORK PRESENTED BY KOUSTAV SARKAR
 
Online food ordering system project report.pdf
Online food ordering system project report.pdfOnline food ordering system project report.pdf
Online food ordering system project report.pdf
 
Hostel management system project report..pdf
Hostel management system project report..pdfHostel management system project report..pdf
Hostel management system project report..pdf
 
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
 
PE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and propertiesPE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and properties
 
Computer Lecture 01.pptxIntroduction to Computers
Computer Lecture 01.pptxIntroduction to ComputersComputer Lecture 01.pptxIntroduction to Computers
Computer Lecture 01.pptxIntroduction to Computers
 
Integrated Test Rig For HTFE-25 - Neometrix
Integrated Test Rig For HTFE-25 - NeometrixIntegrated Test Rig For HTFE-25 - Neometrix
Integrated Test Rig For HTFE-25 - Neometrix
 
Block diagram reduction techniques in control systems.ppt
Block diagram reduction techniques in control systems.pptBlock diagram reduction techniques in control systems.ppt
Block diagram reduction techniques in control systems.ppt
 
Wadi Rum luxhotel lodge Analysis case study.pptx
Wadi Rum luxhotel lodge Analysis case study.pptxWadi Rum luxhotel lodge Analysis case study.pptx
Wadi Rum luxhotel lodge Analysis case study.pptx
 
Bhubaneswar🌹Call Girls Bhubaneswar ❤Komal 9777949614 💟 Full Trusted CALL GIRL...
Bhubaneswar🌹Call Girls Bhubaneswar ❤Komal 9777949614 💟 Full Trusted CALL GIRL...Bhubaneswar🌹Call Girls Bhubaneswar ❤Komal 9777949614 💟 Full Trusted CALL GIRL...
Bhubaneswar🌹Call Girls Bhubaneswar ❤Komal 9777949614 💟 Full Trusted CALL GIRL...
 
Moment Distribution Method For Btech Civil
Moment Distribution Method For Btech CivilMoment Distribution Method For Btech Civil
Moment Distribution Method For Btech Civil
 
COST-EFFETIVE and Energy Efficient BUILDINGS ptx
COST-EFFETIVE  and Energy Efficient BUILDINGS ptxCOST-EFFETIVE  and Energy Efficient BUILDINGS ptx
COST-EFFETIVE and Energy Efficient BUILDINGS ptx
 
Engineering Drawing focus on projection of planes
Engineering Drawing focus on projection of planesEngineering Drawing focus on projection of planes
Engineering Drawing focus on projection of planes
 
NO1 Top No1 Amil Baba In Azad Kashmir, Kashmir Black Magic Specialist Expert ...
NO1 Top No1 Amil Baba In Azad Kashmir, Kashmir Black Magic Specialist Expert ...NO1 Top No1 Amil Baba In Azad Kashmir, Kashmir Black Magic Specialist Expert ...
NO1 Top No1 Amil Baba In Azad Kashmir, Kashmir Black Magic Specialist Expert ...
 
Generative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPTGenerative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPT
 
Employee leave management system project.
Employee leave management system project.Employee leave management system project.
Employee leave management system project.
 
Verification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptxVerification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptx
 
GEAR TRAIN- BASIC CONCEPTS AND WORKING PRINCIPLE
GEAR TRAIN- BASIC CONCEPTS AND WORKING PRINCIPLEGEAR TRAIN- BASIC CONCEPTS AND WORKING PRINCIPLE
GEAR TRAIN- BASIC CONCEPTS AND WORKING PRINCIPLE
 

Result & discussion uv vis (exp 3) - for merge

  • 1. 5.0 RESULT / DATA ANALYSIS Table 1: Absorbance value at 516 nm of Carmoisine Standard at different concentration Concentration (ppm) Absorbance (516 nm) Molar absorptivity (L.mol−1 . cm−1 ) 0 (Blank) 0 0 5 0.342 3.4366 × 104 15 0.739 2.4753 × 104 25 1.106 2.2228 × 104 35 1.455 2.0887 × 104 45 1.792 2.0008 × 104 Table 2: Absorbance value at 516 nm of Unknown Samples at different concentration Unknown Concentration (ppm) Absorbance (516 nm) Molar absorptivity (L. mol−1 . cm−1 ) A 27.8 1.005 1.8163 × 104 B 36.4 1.317 1.8179 × 104 Figure 1: Standard calibration curve of Carmoisine y = 0.0362x + 0.1828 R² = 0.9989 0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2 0 5 10 15 20 25 30 35 40 45 50 Absorbance(516nm) Concentration (ppm) Standart Calibration Curve of Carmoisine
  • 2. 6.0 DISCUSSION The purpose of this experiment were to determine Lambda Max for Carmoisine sample (wavelength scan). To determine ɛ the molar absorptivity of Carmoisine and to prepare a serial dilution and generate a standard calibration graph for sample quantitation (photometric scan). The Carmoisine standard solution samples were run by using Uv-Vis Spectrophotometer. Ultraviolet-visible spectroscopy or ultraviolet-visible spectrophotometry (UV-Vis or UV/Vis) refers to absorption spectroscopy in the ultraviolet-visible spectral region. This means it uses light in the visible and adjacent (near-UV and near-infrared (NIR)) ranges. UV-Vis spectrophotometer is used in the quantitative determination of concentrations of the absorber in the solutions of transition metal ions and highly conjugated organic compounds (IIT Kanpur, 2012). Besides, UV spectrophotometer principle follows the Beer-Lambert Law. This law states that whenever a beam of monochromatic light is passed through a solution with an absorbing substance, the decreasing rate of the radiation intensity along with the thickness of the absorbing solution is actually proportional to the concentration of the solution and the incident radiation (Choudhary, A. 2017). The experiment was started with 100 ppm of stock solution of Carmoisine was prepared to dilute the stock solution into different concentration which were 5 ppm, 15 ppm, 25 ppm, 35 ppm and 45 ppm with distilled water into 50 ml volumetric flask. The volume needed to prepare the serial dilution from 100 ppm of stock solution were 2.5 ml, 7.5 ml, 12.5 ml, 17.5 ml and 22.5 ml respectively. The volume of serial dilution were calculated by using the formula M1V1 = M2V2. Moreover, the unknown sample was prepared with the same method as serial dilution but the volume of stock solution must be between 2.5 ml to 22.5 ml. After setting up the UV-Vis, the blank solution was placed into instrument and baseline was click. After the blank solution became zero, the standard solution of 45 ppm was placed to another cuvette. The maximum wavelength was recorded at 516 nm. The maximum wavelength was the wavelength with the higher absorbance. For the photometric scan, the cuvette was filled with standard solution 5 ppm to 45 ppm to determine the absorbance and prepared for the standard calibration curve.
  • 3. According to the result obtained, the absorbance reading showed for the standard solution 5 ppm, 15 ppm, 25 ppm, 35 ppm and 45 ppm were 0.342, 0.739, 1.106, 1.455 and 1.792 respectively. Based on graph plotted, the absorbance (516 nm) was directly proportional to the concentration (ppm). Because of when the concentration was higher, more radiation is absorbed and the absorbance also become higher. In addition, for the unknown samples the absorbance obtained were 1.005 and 1.317. The concentration of unknown samples have found by using Beer’s Lambert law formula A = εbc with the slope obtained was 0.0362 which indicates the molar absorptivity of Carmoisine. Thus, the concentration of unknown samples were 27.8 ppm and 36.4 ppm. Furthermore, the graph showed the R2 value that is a statistical measure of how close the data are to the fitted regression line (Ogee, A. 2013). The correlation coefficient varies about +1 to -1. The graph showed the correlation coefficient was 0.9989 which the R2 value was nearest to the +1 which indicates the accurate of the result. As conclusion, there are several possibilities precaution that should be considered. Firstly, make sure use the pipette to measure the volume of Carmoisine to get the accuracy. Secondly, make sure wipe and do not touched the clean sides of cuvette and carrying out the set-up procedure in the correct order. REFERENCES 1. Ogee, A. (2013). Regression Analysis: How Do I Interpret R-squared and Assess the Goodness-of-Fit?. Retrieved from https://blog.minitab.com/blog/adventures- in- statistics-2/regression-analysis-how-do-i-interpret-r-squared-and-assess-the-goodness- of-fit. 2. Wenzel, T. (2019). Beer’s Law. Retrieved from https://chem.libretexts.org/Bookshelves/Analytical_Chemistry/Supplemental_Module s_(Analytical_Chemistry)/Analytical_Sciences_Digital_Library/Active_Learning/In_ Class_Activities. 3. IIT Kanpur. (2012). UV-Vis Spectrophotometer. Retrieved from https://www.iitk.ac.in/dordold/index.php?option=com_content&view=category&layo ut=blog&id=219&Itemid=238. 4.