2. 1st Detect Proprietarywww.1stdetect.com
Air Toxic Monitoring
US EPA Requires Oil Refineries To Monitor Benzene Levels
Other Hazardous Air Pollutants (HAP) include 1,3-butadiene, toluene, ethylbenzene,
and xylenes
Method 325 established by the EPA for fenceline monitoring of benzene at petroleum
refineries
• Method sets standards for sampler placement, sample collection, preparation, and
analysis
• Recommends a passive sampling approach
‒ 12 to 24 passive samplers placed at 15 to 30 degree intervals along the perimeter
of each refinery
‒ Sampling cycle: 14 days
3. 1st Detect Proprietarywww.1stdetect.com
Passive Sampling - Limitations and Challenges
Passive Sampling
Samples analyzed off-line
Inability to identify emission source
Miniature Mass Spectrometer As An Alternative
Approach To Passive Sampling
4. 1st Detect Proprietarywww.1stdetect.com
iONTRAC
Why iONTRAC?
Program and monitor gas streams in real-time
Data from unused channels stored for future analysis
No carrier gases required
Operator controlled or autonomous unattended operation
Continuous monitoring
Out-of-spec product alarms
Cost Effective
Miniature ion trap mass spectrometer
Mass Range: 35 – 500 amu
Dynamic Range: 2 orders of magnitude
Gases Monitored: User programmable (up to 8 ions)
5. 1st Detect Proprietarywww.1stdetect.com
GOAL
To Demonstrate the Feasibility of the iONTRAC to Monitor HAPs
Monitor benzene and ethylbenzene excursions
Evaluate the iONTRAC’s performance under ambient conditions
6. 1st Detect Proprietarywww.1stdetect.com
Ambient Monitoring Technical Approach
iONTRAC equipped with a preconcentrator utilizing a Carbopack X sorbent
• Sampling Time: 70 s
• Desorption Time: 60 s
Instrument was placed in parking lot at 1st Detect’s facility in Webster TX
System challenged with benzene and ethylbenzene at varying distances
Mixtures of benzene and ethylbenzene analyzed
Meteorological data recorded with each simulated emission
Sampling Mode: Continuous
Sample period: 5 days
• Hold Time: 200 s
• Desorption Temp: 300C
12. 1st Detect Proprietarywww.1stdetect.com
Detection Levels
Benzene
Benzene and Ethylbenzene Detected at 200 ft
20 ft
Intensity as a Function of Distance
Intensity(a.u.)
Avg.Intensity(a.u.)
Intensity(a.u.)
200 ft
13. 1st Detect Proprietarywww.1stdetect.com
Summary
iONTRAC offers continuous real-time monitoring of VOCs in
ambient air with:
• Rapid clear down rates (< 2 min)
• Stable and rapid response time
• Detection VOCs levels to a least 200 ft
• Simultaneous monitoring of multiple VOCs
• Faster detection times than passive sampling (<5 min)
14. 1st Detect Proprietarywww.1stdetect.com
Future Directions
Optimize sampling and detection time
Test additional compounds
Increase the number of sampling ports on the iONTRAC
Test units on-site at petroleum refineries
Where Do We Go From Here?